3B Scientific Test Vessel User Manual

3B SCIENTIFIC® PHYSICS
Ultrasonic CW generator U10006
Laser diode for Debye-Sears effect U10007
Test vessel U10008
Instruction sheet
8/03 ALF
®
3
2
1
Ultrasound generator with accessories for the Debye­Sears experiment and for demonstrating standing ul­trasonic waves.
Safety instructions
The slits in the device are for ventilation and must
be kept clear to avoid overheating the equipment. It is recommended that the device be placed on its own stand (also included).
Before switching on the device make sure that it is
adjusted to the mains voltage you intend to use. Make sure that you keep within the operating speci­fications.
Never try to push objects through the slits in the
device since this could cause short circuits or elec­tric shocks.
Only use 3B’s own ultrasound transducer with the
“PROBE” connection. Be careful, voltages may be as high as 70 V.
Do not use the ultrasound transducer for a lengthy
period without contact to fluids. Otherwise the transducer may overheat and be destroyed.
Do not turn on the U10007 laser diode if there are
1 Laser diode
2 Test vessel
3 Ultrasonic transducer
people standing in the beam. This is a class II laser with power < 1 mW. Do not look into the beam or aim it at other people or animals.
Beware: high-powered ultrasound! Do not use the
probe on people or animals.
Contents
1. Introduction ............................................ 7
2. Components ............................................ 7
2.1 Ultrasound generator - operating elements 7
2.2 Set-up for test container 8
2.3 Lens on glass slide 8
2.4 Laser diode (U10008) 8
3. Experiment procedure .............................. 8
3.1 General instructions for experiments 8
3.2 Debye-Sears effect 9
3.3 Projection of standing ultrasonic waves 9
4. Technical details ..................................... 10
1.1. Ultrasound generator 10
4.2. Ultrasonic transducer 10
4.3. Test vessel 10
6
1. Introduction
In 1932 Debye and Sears demonstrated for the first time how light is refracted when passing through a fluid ex­cited to high-frequency oscillation. The maxima and minima in density act here like the grid elements of an optical diffraction grating. The grating constant corresponds to the ultrasonic wavelength and is thus dependent on the frequency of the ultrasound and the speed of sound in the medium through which the sound is travelling. The CW (continuous wave) generator with its accompa­nying broad-band ultrasound transducer, along with an adjustable test vesseland its integrated laser holder, mean that it is possible for the first time to demon­strate this phenomenon to schoolchildren and college students alike using simple and compact equipment. The equipment can demonstrate both frequency de­pendence (with four different frequencies) and the ul­trasonic wavelength in various fluids, allowing the speed of sound in the medium to be calculated. It is also possible to project standing ultrasonic waves by inserting an optical lens between the laser source and the ultrasonic waves. The waves are then projected by the diverging laser beam.
2. Components
2.1 Ultrasound generator - operating elements
The ultrasound generator generates continuous high­power ultrasonic waves (CW = continuous wave). The voltage bmcan be adjusted between 5 and 65 V. There is an LCD display for the voltage output bl. The trans­mitted voltage can be switched off separately 8. The frequency can be set to one of four frequencies (1, 2, 4 or 8 MHz) 9. An additional monitor output 7 allows the output frequency to be determined precisely with the aid of an oscilloscope or a frequency counter. A suitable output 3is provided to supply power to a laser diode. This can also be switched off separately
2.
Adjustment of mains voltage
The ultrasound generator can operate with 230 V or 115 V mains voltage. A voltage selector switch is situ­ated on the rear of the casing and is hidden beneath the covering to the right of the rear panel. A screw­driver is used to alter the switch setting (see illustra­tion). If the mains voltage is altered, the fuse needs to be changed to suit the mains voltage. A T630 mA fuse should be used with 115 V mains voltage and a T315 mA fuse for 230 V mains.
Warning:
Unplug from the mains before altering the mains set­ting. No voltage may be applied to the device while making the adjustment,
1 2 3 4 5 6 7
1 Mains switch 2 On/off switch for laser 3 Connector socket for laser 4 Control light for laser 5 Connector socket for ultrasonic transducer 6 Control light for ultrasonic transducer 7 Monitor output 8 On/off switch for ultrasound 9 Frequency selection switch bl Voltage display bm Voltage setting
bm
bl
9
8
1
2
115
230
1 Remove the two screws from the rear panelling and take off
the panel. The voltage selector switch is revealed.
2 Fuse
115 V 630 mA 230 V 315 mA
3 An arrow indicates the voltage that is currently set. Adjust using
a screwdriver inserted into the slit.
3
7
2.2 Set-up for U10008 test container
1 Adjustment screw for changing the angle of the transducer 2 Lid 3 Transducer holder 4 Ultrasonic transducer (supplied with U10006) 5 Securing screw for laser diode 6 Slot for lens holder 7 Laser holder 8 Securing screw for ultrasonic transducer 9 Glass vessel
The following illustration shows how the lens should be properly placed in the test container.
1
2
3
4
5
6
7
8
9
2.4 Laser diode (U10007)
Standing ultrasonic waves are generated in a special container. It allows the angle of incidence of the waves to be set precisely to the perpendicular by means of a special holder for the ultrasound transducer. The test vessel consists of a glass vessel with a lid and a holder for the transducer, three adjustment screws to set the alignment for the standing wave and a laser holder perpendicular to the axis of the waves which also has a holder for a lens so that the ultrasonic waves and the Debye-Sears effect can be projected.
2.3 Lens on a glass slide
A plane-convex lens is attached to a rectangular slide. This is inserted into the slot in the laser holder of the test container for experiments involving projection.
1 2 3
1 Plano-convex lens (f=100 cm) 2 Glass slide 3 Grip
1 2 3 4
1 Laser beam window 2 Laser module jacket 3 Plug for connecting to ultrasound generator 4 Connecting lead
Class II laser diode for demonstrating the Debye-Sears effect and projecting ultrasonic waves for use with the ultrasound generator and test container.
3. Experiment procedure
3.1 General instructions for experiments
Pay attention to the following instructions if the ex­periment is to work:
Use water that contains as little air as possible since
air bubbles disturb both the ultrasound field and the refraction of the laser beam.
Any air bubbles around the probe should be re-
moved.
Allow for the maximum distance between the test
container and the projection screen.
When no measurements are being made, the ul-
trasound should be switched off so that the test fluid does not heat up.
8
For precise measurements, also measure the tem-
perature and include this in the comparison.
When using high voltage at any frequency and when
the transducer is properly aligned, at least 3 orders of diffraction should be visible.
The projection experiment is much more sensitive
to the angle of the transducer than the refraction experiment. Thus for this experiment the conditions for generating standing waves need to be adhered to more precisely.
3.2 Debye-Sears effect
The wavelength of the ultra­sonic waves in the Debye-Sears experiment (photo left, 4 MHz in water) can be determined for various different test fluids (water, glycerine, cooking oil). This means measuring the dis­tance s between the ultra­sound transducer and the re­fracted image. Then the num­ber of orders of refraction N and distance between the -nth and +nth bands x can be de-
termined. Since the wavelength of the laser light
λ
L
known then:
λ
2N s
(1)
λ
s
gives the ultrasonic wavelength
L
=
x
λ
. The individual vari-
s
ables can be calculated as in the following diagram.
λ
L
λ
s
s
N = 3
N = 2 N = 1
N = 0 N = -1
N = -2
N = -3
The ultrasonic frequency n is measured at the monitor socket. Then the speed of sound c in the fluid is given by:
is
2. Glycerine
v = 4 MHz, s = 2.90 m, N = 2, x = 1.6 cm,
λ
= 650 nm
L
therefore: λ = 471.2 µm, c = 1885 m/s (Table: 1900 m/s at 25°C)
3.3 Projection of standing ultrasonic waves
Direct projection of ultrasonic waves can be an inter­esting extension to the experiment. The sound wave is pro­jected by inserting a convex lens into the laser beam so that the beam is di­verged. The density variations in the standing wave then appear as light and dark regions on the projection screen (see photograph right). To determine the wavelength from the diffraction im­age and the geometry involved, as well as the focal length f of the lens in air (100 mm in this case), corrections due to the glass walls of the vessel and the test fluid also need to be considered.
N = 6 N = 5
N = 4 N = 3
x
N = 2
g
1
a1a
x
The light refraction method as described in 3.2 is thus
g
2
2
s
N = 1
N = 0
better suited for calculating the wavelength precisely. The precise equation for
λ
in the projection experi-
s
ment is:
gna
1g1
f
−−
(3)
2
==
λ
s
N
x
sf
−−+−
 
n
FL
ggnaa
12g12
+
n
FL
 
 
(2) c =
λ
ν
s
Example results:
1. Water
v = 4 MHz, s = 2.90 m, N = 4, x = 4.1 cm, therefore:
λ
= 367.8 µm, c = 1471 m/s
s
(Table: 1480 m/s at 20°C)
λ
= 650 nm
L
The distance a1 between the glass wall towards the lens and the distance a2 can be approximated as half the internal width of 9.6 cm. The thickness of the glass g and g2 is about 4 or 5 mm. The refractive indices nFL for the test fluid and ng for the glass may be measured or taken from tables.
9
1
N is the number of bright maxima and x the separa­tion between them. The speed of sound in the medium can now be calcu­lated once again from the measured frequency n us­ing equation (2).
Note:
With the U10008 test vessel it is only possible to set up a good projection of the standing ultrasonic wave at a frequency of 4 MHz. At 2 MHz and especially at 1 MHz ultrasonic waves reflected from the bottom of the con­tainer are superimposed destructively on the imping­ing ultrasonic beam. Only for 4 MHz waves is the dif­ference in the path length of the order of several wave­lengths. At 8 MHz the amplitude of the emitted wave is too small and also the absorption is much greater (the absorption coefficient is proportional to the square of the frequency). This means that no clear standing waves are produced.
Example results:
Water:
a1 = a2 = 4.8 cm, f = 10 cm, nFL = 1.33, ng = 1.45, s =3.03 m, n = 4 MHz, x = 8.9 cm, g1 = 5 mm, g2 = 4 mm
therefore: λ = 397 µm, c = 1590 m/s (Table: 1480 m/s at 20°C)
4. Technical details
4.1 Ultrasound generator (U10006)
Frequency: 1, 2, 4, 8 MHz adjustable Output signal: Sine wave, continuously adjustable
between 5 Vpp to approximately 65 Vpp. Can be switched off. Red control light
Display: 3-digit LCD, one decimal place,
height of digits 10 mm
Monitor signal: TTL, Frequency signal
Laser output: 3 V DC, max. 300 mW at socket
5.5 mm outer Ø, 2.5 mm inner Ø, Can be switched off.
Red control light Dimensions: 256 x 86 x 156 mm Mains voltage: 115/230 V, 50/60 Hz Power consumption: max. 60 VA; 5 VA minimum power
(laser and ultrasound switched
off) Fuse: T 315 mA (230 V) / T 630 mA (115 V)
Ultrasonic transducer
Probe diameter: 27 mm Active surface area: 2 cm² Cable length: 1 m with BNC socket
4.2 Test vessel (U10008)
Test vessel: 100 x 100 x 120 mm Volume: 900 ml approx. Laser holder: 17 mm internal diameter Transducer holder: adjustable, designed for the
supplied transducer
Lens on glass slide
Lens slide: 76 mm x 26 mm Glass slide with
roughened gripping surface Lens: Plano-convex, f = 100 mm
(in air), 16 mm Ø
4.3 Laser diode (U10007)
Beam spot: < 6 mm at 3 m Wavelength: 650 nm Power: < 1 mW, laser safety class II Supply voltage: 3 V DC Current consumption: max. 35 mA Connecting plug: 1 m cable with plug of 5.5 mm
outer Ø, 2.5 mm inner Ø Dimensions: 80 mm x 17 mm Ø
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany • www.3bscientific.com • Technical amendments are possible
10
Loading...