3B Scientific Teltron NMR Supplementary Set User Manual

f
-
3B SCIENTIFIC
NMR Supplementary Set 1000642
06/12 ALF
®
PHYSICS
1 Magnet unit 1a Magnet 1b Yoke 2 Discs for assembly 3 NMR probe 4 Polystyrene sample 5 Comparison sample 6 Glycerine sample 7 Teflon sample
1. Safety instructions
The permanent magnets can generate consid­erable forces of attraction and repulsion with the result that there is a risk of squashing or splin­tering.
Be especially cautious when inserting mag­nets into the basic unit.
Never use the magnets except as specified.
Magnetic fields can erase data from magnetic media and affect or destroy electronic or me­chanical components like heart pacemakers.
People with pacemakers should not conduct this experiment.
2. Description
The NMR supplement set is to be used with the ESR/NMR basic set (1000637 resp. 1000638) for investigating nuclear spin resonance in glycerine, polystyrene and Teflon.
The set consists of an NMR probe with a high­frequency coil, a strong and uniform permanent magnet, a glycerine sample, a polystyrene sam­ple, a Teflon sample, an empty comparison sam­ple and two discs for assembling the apparatus.
3. Technical data
Magnetic flux of permanent magnets: 300 mT approx.
Frequency range: 11 15 MHz approx.
4. Maintenance and storage
Remove any debris from the surfaces of the permanent magnet poles and of the assem­bly discs using a cloth with some isopropanol.
Keep permanent magnets in a dry place.
5. Disposal
The packaging should
be disposed of at local recycling points.
If the equipment itsel
needs to be scrapped, it is safe to dispose of all com ponents other than the probe in domestic waste.
The probe should be disposed of in containers dedicated to the disposal of electrical refuse.
1
6. Equipment required in addition
1 ESR/NMR basic set (230 V, 50/60 Hz) 1000638 or 1 ESR/NMR basic set (115 V, 50/60 Hz) 1000637
1 Analogue oscilloscope, 2x30 MHz 1002727 2 High-frequency cables 1002746
alternatively
1 3B NETlog™ unit (230 V, 50/60 Hz) 1000540
or
1 3B NETlog™ unit (115, 50/60 Hz) 1000539
assembly discs (see Fig. 4).
Push back the two discs with your thumbs and pull on the coils to move the magnet supports into their end positions (see Fig. 5).
Tighten both knurled screws by hand at the same time. Make sure that the magnets are accurately aligned on top of the assembly discs. If necessary, slot the discs all the way back into the probe chamber and then push the magnet supports into their end positions.
1 3B NETlab™ 1000544
2 High-frequency cables, BNC/4-mm plug 1002748 1 PC
7. Operation
7.1 Assembly of the basic unit
The discs, the pole surfaces of the magnets and the probe chamber in the basic unit must all be free of grease, dust and debris.
If necessary they should be cleaned using isopropanol.
Insert the discs into both sides of the sample chamber (see Fig. 1).
Fig. 2 Permanent magnet inserted into basic unit
Fig. 1 Sample chamber with assembly discs inserted
Twist the metal rod into the yoke of the magnet unit. Place the magnet unit in the clip on the basic unit as shown in Fig. 2.
Take the magnet unit in both hands and push the yoke downwards with your thumbs (see Fig. 3).
Push the coils over the magnets, making sure the direction if the windings is the same. The embossed arrows on the coils must point in the same direction.
Clean any grit or metal shavings off the two magnets with a cloth.
Move the completed magnetic unit onto the assembly discs, making sure to carry out the steps that follow: (see Fig. 4).
Separate the two coils with your hands, pull­ing them outwards to increase the gap be­tween them. You can rest your hands on the knurled screws while doing this. Push the magnets about a quarter of the way onto the
Fig. 3 Removing the yoke from the permanent mag­net
Fig. 4 Pulling the two permanent magnets apart
2
Loading...
+ 4 hidden pages