3B Scientific AM50 User Manual

3B SCIENTIFIC3B SCIENTIFIC
3B SCIENTIFIC®
3B SCIENTIFIC3B SCIENTIFIC
PHYSICSPHYSICS
PHYSICS
PHYSICSPHYSICS
U17450 Analog Multimeter with Zero-point Center/Left
Instruction Sheet
10/04 MC/ALF
1 Rotary switch to select the
3
4
2
1
measurement range
6
electrical zero-point in the center of the scale
3 Safety connection sockets
5
4 Slide switch to select the
operating mode
5 Adjustment screw to set the
zero-point mechanically
6 Nub to latch the housing
shut
Handheld meter for current and voltage measurement.
1. Safety instructions
Before using the analog multime-
ter, make sure you read the oper­ating instructions carefully and that you comply with them com­pletely.
The safety of the multimeter and
the person operating it can only be guaranteed if it is used in ac­cordance with the instructions. Do not operate or handle this unit incorrectly or inappropriately.
The device may only be used by
persons, who are aware of the haz­ards of contact (for voltages over 30 V rms) and can undertake the appropriate safety precautions. This also includes the appearance
of unforeseen voltages e.g. in de­fective units or charged capacitors.
In the case of voltage and current
measurements the nominal volt­age between the phase and neu­tral conductor may not exceed 300 V according to CAT II (in circuits that are directly connected to the mains) and CAT III (in building wir­ing installations) 300 V.
The analog multimeter may be
used for measurements in circuits with corona discharge (high volt­age).
In measurements involving RF cir-
cuits special care must be taken due to the existence of dangerous hybrid voltages.
The appropriate permissible mea-
surement range may not be ex­ceeded. Always change from a higher measurement range to a
6
lower measurement range.
Before using the device, check the
housing and the measurement cables for any damage.
Do not conduct measurements in
a damp environment. Workplace, hands, shoes and floor must be dry.
Before opening the housing all
measurement leads are discon­nected from the device.
2. Description, technical data
Active analog multimeter with a slide switch for selection of the op­erating mode and a rotary switch to select the measurement range as well as a scale with the mirrored background for parallax-free read­ings with a zero-point adjustable to the center or the left of the scale. By adjusting the electrical zero-point to the center of the scale bipolar DC voltage and current measurements can be conducted without worrying about polarity. The device is ex­tremely robust in terms of load ca­pacity and is equipped with excel­lent overload protection due to its two anti-parallel diodes as well as moving coil movement which is not sensitive to external electromagnet­ic fields. The safety connection sock­ets offer protection against acciden­tal touch contact. After approx. 45 mins the battery is automatically switched off. The power can be re­established by turning off and on using the slide switch for the oper­ating mode. The robust plastic hous­ing and the spring-connected bear­ing jewels of the movement guar­antee protection against damage
caused by mechanical stress. Measurement ranges: Voltage measurement:
DC
Meas. range Internal resistance
100 mV 10 M 300 mV 10 M
1 V 10 M 3 V 10 M
10 V 10 M 100 V 10 M 300 V 10 M
AC
Meas. range Internal resistance
3 V 1 M 10 V 1 M 30 V 1 M
100 V 1 M 300 V 1 M
Current measurement AC/DC:
Meas. range Voltage drop
0.1 mA 55 mV 1 mA 55 mV
10 mA 55 mV
100 mA 55 mV
1 A 53 mV 3 A 51 mV
Accuracy: DC class 2; AC class 3 Effective variables and nominal op­erating ranges: Temperature 0 – 40° C: ± 2% / K Frequency for all measurement ranges: ±2.5% at 30 Hz up to
1.5 kHz ± 5% at 1.5 kHz
up to 3 kHz Reference conditions: Ambient temperature: +23° C ± 2 K Frequency: 50 to 60 Hz Waveform: Sinusoidal
7
Overload protection:
+
F3 fuse, 15 H/250 V in accordance with DIN VDE 0820 Section 22/EN 60 127-2 as circuit protection in the case of overload; movement protect­ed by 2 antiparallel-connected di­odes Protection class: IEC 1010-1/EN 61010-1/ VDE 0411-1 Overvoltage category: CAT III Nominal voltage: 300 V Degree of pollution: 2 Test voltage: 3.7 kV~ EMV: Electromagnetic compatibility Jamming: EN 50081-1:1992 Interference immunity: EN 50082-1:1992 Power supply: 1 x 9 V flat cell bat­tery, IEC 6F22 Dimensions: 98 x 138 x 35 mm Weight: approx. 0.3 kg
3. Operation
3.1 Readying for use
Insert battery into the battery
compartment. To do this remove the section of housing by pressing in the nub (6), e.g. using a screw­driver. Then insert the battery and connected it to the battery clip. Replace the housing section and snap it into place.
Check the mechanical zero-point.
The measuring instrument may not be connected at this time. Set the sliding switch (4) to the “0” position. The needle must be lo­cated in the “ OFF” position when the multimeter is in a hori­zontal position. If necessary make the corresponding adjustments with the adjustment screw (5).
Check the electrical zero-point. Set
the slide switch (4) into the “ ” position. Select the measurement range, the needle must be set to zero-point in the center of the scale, otherwise correct using the rotary knob (2).
Check the battery. To do this set
the slide switch (4) to “ ” , and then adjust the rotary switch (1) to“ ”. If the needle does not de­flect into the battery test zone des­ignated “ ” the battery must be replaced.
3.2 General instructions
When performing measurements
always set the rotary switch (1) to the highest measurement range. Then turn the switch to lower rang­es until you obtain optimum nee­dle deflection.
When the power supply is inter-
rupted by the automatic battery switch-off (after approx. 45 min.) switch the slide switch (4) on and off again.
When the multimeter is not in use,
disconnect all measurement leads from the meter, reset the rotary switch (1) to the highest range, set the slide switch (4) to the “0” set­ting and, if necessary, remove the battery.
3.3 DC voltage measurement
_
8
3.3.1 Operating mode: electrical
_
+
3.3.2 Operating mode: electrical zero-point center
Set the slide switch (4) to the “ ”
setting.
Using the rotary switch (1) select
the corresponding measurement range “V ...”.
The needle should now be posi-
tioned in the center of the scale.
Connect the multimeter and take
a reading from the lower scale.
3.4 Measuring alternating voltage, directly up to 300 V
~
Set the slide switch (4) to “
setting.
Using the rotary switch (1) select
the corresponding measurement range “V~”.
Connect the multimeter and take
a reading from the upper scale.
To reduce the effects of the fre-
quency, connect the socket “ directly to ground or to the point with the lowest potential with re­spect to ground.
3.5 Measuring AC voltage with superimposed DC voltage
⊥⊥
⊥⊥
~
With the aid of a capacitor (recom-
mended: 4.7 µF/630 V) it is possi-
ble to isolate DC-voltage compo­nents in an amplifier output stage, for example. The resultant mea­surement error is less than 0.2% at a measurement frequency of 50 Hz.
Proceed with the measurement as
stated in point 3.4.
Measurement of the DC voltage
components is performed as de­scribed under 3.3.
To avoid overloading, the set mea-
surement range must be greater than the initially determined DC voltage components.
Caution: before switching to a low-
er measurement range both volt­age components must be checked.
3.6 Current measurement
When performing current mea-
surements the multimeter must be connected in series with the load in the circuit, which has the lowest potential with respect to ground.
3.6.1 DC measurement, directly
3.6.1.1 Operating mode: electrical zero-point left
Switch the slide switch (4) to the
” setting.
Using the rotary switch (1), select
the corresponding measurement range “A ...”.
Connect the multimeter take a
reading from the upper scale.
9
3.6.1.2 Operating mode: electrical
_
+
~
zero-point center
Adjust the slide switch (4) to the
“ ” setting.
Using the rotary switch (1), select
the corresponding measurement range “A ...”.
The needle should be located in
the center of the scale.
Connect the multimeter and take
a reading from the lower scale.
3.6.2 Measuring DC current using
shunts
3.6.2.1 Operating mode: Zero-point
left
Set the slide switch (4) to the
”.
Set the rotary switch (1) to the “V ...
100 mV” setting.
Connect the multimeter and the
take a reading from the upper scale.
3.6.2.2 Operating mode: Zero-point
center
Set the slide switch (4) to the “ ”
setting.
Set the rotar y switch (1) to the “V ...
100 mV” position.
The needle should now point to
the center of the scale.
Connect the multimeter and take
a reading from the lower scale.
3.6.3 Measuring AC current
Set the slide switch (4) to the “
setting.
Use the rotary switch (1) to select
the corresponding measurement range “A~”.
Connect the multimeter and take
a reading from the upper scale.
4 Maintenance
4.1 Cleaning
Only use a paintbrush or soft tow-
el to clean the multimeter. If stat­ic electrical charge builds up on the view window, this can be elim­inated using a damp rag or an an­tistatic agent.
4.2 Replacing the battery
If during the battery test the nee-
dle no longer deflects to the bat­tery test zone depicted with “ ”, the battery must be replaced (see
3.1).
4.3 Replacing the fuse
The multimeter is equipped with
a safety fuse F3,15/250. The fuse holder is located on the printed circuit board. To replace the fuse the device must be opened as de­scribed in section 3.1.
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Ger many • www.3bscientific.com • Technical amendments are possible
10
Loading...