The sensor electronics with the stored sensor data are in the connecting
head of the electrode. The data include, among other things, the senso r type
and series number. In addition, the calibration data are stored in the sensor
with each calibration and the calibration history is recorded (the last 10 calibrations). The data is recalled by the meter when the sensor is connected
and is used for measurement and for measured value documentation.
Storing the calibration data in the sensor ensures that the correct slope and
asymmetry are automatically used if the sensor is operated with different
meters. On the other hand, different calibrated sensors can be used with one
meter without the need to recalibrate.
The digital transmission technique guarantees the failure-free communication with the meter even with long connection cables. The sensor firmware
can be updated via the meter.
ba76137d0410/2017
3
Page 4
General information4110(W)
2C ommissioning, measuring, calibration
2.1Commissioning
Scope of delivery Electrode 4110(W)
Operating manual
CommissioningPrepare the electrode for measuring as follows:
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Please keep the watering cap. It is required for the electrode to
be stored. Always keep the watering cap clean.
Remove any gas bubbles behind the pH membrane by shaking.
Connect the electrode to the meter.
4110– via the sensor cable to a free IDS connec-
tor on the meter
4110W– via a connecting cable (accessory) to a
free IDS connector on the meter
or
– wireless via an IDS WA-S adapter (acces-
sory) to a WA-capable meter
Accessories for the connection of the 4110W
sensor to the meter: See chapter 7 W
PARTSANDACCESSORIES.
EAR
Opening and closing the IDS plug connection,
see section 2.2 O
PLUGCONNECTION.
IDS
PENINGANDCLOSINGTHE
Calibrate the electrode according to the operating manual of the meter
and observe the following rules while doing so.
4
ba76137d0410/2017
Page 5
4110(W)General information
2
4
3
1
Plug connection
closed and locked
Guidance
Groove
2.2Opening and closing the IDS plug connection
This section only applies to IDS plug variants (4110W).
Opening the plug
connection
Closing the plug
connection
If necessary, clean the plug connection.
Open the locking device (step 1).
Use your thumb and index finger to press the clips of the connector
together, and pull the connector out of the plug (step 2).
Make sure that the plug connection is completely dry and clean.
Align the guidance of the connector with the groove in the plug and insert
the connector in the unlocked plug until it catches (step 3).
Close the locking device (step 4).
ba76137d0410/2017
5
Page 6
Storage4110(W)
2.3Calibration and measurement: General rules
Avoid the carryover of any solution (sample or buffer solution) from one
measurement to the next by taking the following measures:
– Shortly rinse the calibration and sample beakers with the solution the
beakers are to be filled with next.
– Between measurements, rinse the electrode with the solution that fol-
lows. Alternatively, you can also rinse the electrode with deionized
water and then carefully dab it dry.
To measure in aqueous solutions, it is recommended to immerse the elec-
trode in a vertical or slightly tilted position.
Observe the correct depth of immersion and make sure the contact
between the junction and test sample is thorough. The junction is in the
area of the bottom end of the shaft (see arrow).
Subsequent
calibrations
During short
measuring breaks
4110(W)
Caution: Only the shaft part of the electrode may be immersed!
For measurements in aqueous solutions, provide approximately the same
stirring conditions for measuring as for calibrating.
The frequency of subsequent calibrations depends on the application. Many
meters provide an option where you can enter a calibration interval. After the
calibration interval has expired, the meter will automatically remind you of the
due calibration.
3Storage
Immerse the electrode in reference electrolyte (KCl 3 mol/l, Ag+ free). Prior
to the next measurement, shortly rinse the electrode with the test sample or
deionized water.
Do not scratch the pH membrane.
6
ba76137d0410/2017
Page 7
4110(W)Aging
Overnight or
longer
Put the clean electrode in the watering cap that is filled with reference electrolyte (KCl 3 mol/l, Ag
NOTE
pH electrodes must not be stored dry or in deionized water. The electrode could be permanently damaged by this. If the liquid in the
watering cap has dried up, condition the electrode in reference electrolyte (KCl 3 mol/l, Ag+ free) for at least 24 hours.
During longer storing periods, salt sediments may develop on the
watering cap. They do not affect the measuring characteristics
and can easily be removed with deionized water when the electrode is put into operation again.
+
free).
4A ging
pH electrodes are consumables. Every pH electrode undergoes a natural
aging process. With aging, the responding behavior becomes slower and the
electrode slope and asymmetry change. Moreover, extreme operating conditions can considerably shorten the lifetime of the electrode. These are:
Reference electrolyte solution 250 ml
to fill the watering cap
(KCl 3 mol/l, Ag
DescriptionModelOrder no.
IDS connection cable, 1.5 mIDS-CABLE-
IDS connection cable, 3 mIDS-CABLE-3903 851Y
DescriptionModelOrder no.
WA capable IDS meter
+ radio module for IDS meter
Radio module for plug head sensorIDS WA-S108 141Y
+
-free)
KCl-250109 705Y
1.5
see Internet
903 850Y
ba76137d0410/2017
8Disposal
At the end of its operational lifetime, the electrode must be returned to the
disposal or return system statutory in your country (electronic waste). If you
have any questions, please contact your supplier.
9
Page 10
Disposal4110(W)
10
ba76137d0410/2017
Page 11
4110Contact Information
Contact Information
Ordering & Technical Support
Telephone
:(937) 767-1058
Fax
Email
Mail:YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available)Name and Phone Number
Model number or brief descriptionBilling and shipping address
QuantityPurchase Order or Credit Card
:(800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
YSI has authorized service centers throughout the United States and
Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form
with cleaning certification. The form must be completely filled out for an
YSI Service Center to accept the instrument for service. The Product
Return form may be downloaded at www.ysi.com
‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76137d0410/2017
11
Page 12
Contact Information4110
12
ba76137d0410/2017
Page 13
Page 14
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions
to meet our world's water needs. Developing new technologies that will improve
the way water is used, conserved, and re-used in the future is central to our work.
We move, treat, analyze, and return water to the environment, and we help people
use water efficiently, in their homes, buildings, factories and farms. In more than
150 countries, we have strong, long-standing relationships with customers who
know us for our powerful combination of leading product brands and applications
expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |'m|zīlə
YSI
1725 Brannum Lane
Yellow Springs, OH 45387
Tel: +1 937-767-7241; 800-765-4974
Fax: +1 937-767-1058
Email: [email protected]
Web: www.ysi.com