YSI 4010-2 User Manual

OPERATIONS MANUAL
ba76144e01 02/2013
4010-2/3 pH Adapter DIN / BNC
PH ADAPTER FOR MULTILAB 4010-2 AND MULTILAB 4010-3
4010-2/3 pH Adapter
For the most recent version of the manual, please visit www.ysi.com.
1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241
800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com
Copyright © 2013 Xylem Inc.
2 ba76144e01 02/2013
4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC General information

General information

The 4010-2/3 pH Adapter enables the connection of an analog pH electrode (with or without temperature sensor) to the digital meters MultiLab 4010-2 and 4010-3.
Two models of the adapter are available:
Connection of the analog electrode Adapter
DIN 4010-2/3 pH Adapter DIN
BNC 4010-2/3 pH Adapter BNC
Automatic adapter
recognition
Calibration data
and calibration
history
The measuring electronics with the stored adapter data is in the 4010-2/3 pH Adapter. The data include, among other things, the adapter type and series number. With each calibration, the calibration data is written in the adapter and the calibration history is recorded. The data are called up by the meter and used for measurement and documentation of measured values.
If the adapter firmware is enhanced by YSI, it can be updated via the meter.
Note
More detailed information on the automatic adapter recognition or sensor rec­ognition is given in the operating manual of the meter.
The meter uses for measurement the calibration data stored in the adapter, regardless of the mounted electrode.
Note
Therefore, you always have to recalibrate after exchanging the pH electrode. Otherwise, the IDS pH sensor will work with the calibration data of the previously calibrated electrode. This could lead to incorrect measured values. When evaluating the calibration history, please note that in between calibrations, the electrode may have been exchanged.
ba76144e01 02/2013
3
General information 4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC
12
3

Commissioning, measuring, calibration

Commissioning
Example
4010-3 with
mounted 4010-2/3
pH Adapter DIN
1 Unscrew the 4 screws of the cover plate (the screw driver is included
in the scope of delivery).
2 Insert the 4010-2/3 pH Adapter in the open mounting slot.
3 Attach the adapter to the meter with the 4 screws.
1 pH electrode
2 Reference electrode
3 Temperature sensor (NTC 30)
Note
The further commissioning steps are given in the operating manual of the pH electrode.
4 Connect the electrode to the socket of the adapter.
5 Calibrate the electrode according to the operating manual of the me-
ter.
All further details are given in the operating manuals of the pH electrode and meter.
4
ba76144e01 02/2013
4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC General information

Technical data

General data Electrodes that can be connected All pH electrodes with DIN or BNC
plug head connector
Application
characteristics
Measuring
characteristics of
the IDS measuring
electronics in the
4010-2/3 pH
Adapter
Material, electrical
connection
Electrical data of
the electrode
connection
pH measuring range Depending on the electrode used
Allowed temperature range 0 ... 80 °C (32 ... 176 °F)
Typical application Applications with special pH elec-
trodes
Measured pa­rameter
Measuring range Resolution Accuracy
(± 1 digit)
pH -2.000 ... 20.000 0.001 ± 0.004
U [mV] -1000.0 ... +1000.0 0.1 ± 0.2
T [°C] 0.1 ± 0.2
Material of the plug enclosure ABS
Electrode connection DIN or BNC
Meter connection Mounting slot with contacts
Input resistance > 5 * 1012 ohm
Input current < 1 * 10
-12
A
ba76144e01 02/2013
Note
The latest version of the present operating manual can be found on the Inter­net under www.YSI.com
.
5
General information 4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC
6
ba76144e01 02/2013
4010-2/3 pH Adapter Contact Information

Contact Information

Ordering & Technical Support

Telephone
Fax
: (937) 767-1058
Email
Mail: YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available) Name and Phone Number Model number or brief description Billing and shipping address Quantity Purchase Order or Credit Card
: (800) 897-4151
(937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET
: environmental@ysi.com
1725 Brannum Lane Yellow Springs, OH 45387 USA
: www.ysi.com

Service Information

YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com ‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76144e01 02/2013
7
Contact Information 4010-2/3 pH Adapter
8
ba76144e01 02/2013
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. We move, treat, analyze, and return water to the environment, and we help people use water efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |' m|zīlə
YSI 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: environmental@ysi.com Web: www.ysi.com
©Xylem Inc
Loading...