4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC General information

General information

The 4010-2/3 pH Adapter enables the connection of an analog pH electrode (with or without temperature sensor) to the digital meters MultiLab 4010-2 and 4010-3.
Two models of the adapter are available:
Connection of the analog electrode Adapter
DIN 4010-2/3 pH Adapter DIN
BNC 4010-2/3 pH Adapter BNC
Automatic adapter
recognition
Calibration data
and calibration
history
The measuring electronics with the stored adapter data is in the 4010-2/3 pH Adapter. The data include, among other things, the adapter type and series number. With each calibration, the calibration data is written in the adapter and the calibration history is recorded. The data are called up by the meter and used for measurement and documentation of measured values.
If the adapter firmware is enhanced by YSI, it can be updated via the meter.
Note
More detailed information on the automatic adapter recognition or sensor rec­ognition is given in the operating manual of the meter.
The meter uses for measurement the calibration data stored in the adapter, regardless of the mounted electrode.
Note
Therefore, you always have to recalibrate after exchanging the pH electrode. Otherwise, the IDS pH sensor will work with the calibration data of the previously calibrated electrode. This could lead to incorrect measured values. When evaluating the calibration history, please note that in between calibrations, the electrode may have been exchanged.