The compact, digital precision meter MultiLab 4010-1W enables you to carry
out pH measurements, ORP measurements, conductivity measurements and
dissolved oxygen (D.O.) measurements quickly and reliably.
The MultiLab 4010-1W provides the maximum degree of operating comfort,
reliability and measuring certainty for all applications.
The MultiLab 4010-1W supports you in your work with the following functions:
Automatic sensor recognition
CMC (continuous measurement control)
QSC (sensor quality control)
Electronic access control
Data transmission via the USB interface (USB-B).
1 Keypad
2 Display
3 Connectors
ba76141e0708/20187
Page 8
OverviewMultiLab 4010-1W
1
2
3
4
5
4010 -1WP
1.2MultiLab 4010P-1W meter with integrated printer
The integrated printer of the MultiLab 4010P-1W enables to document measurements according to GLP requirements.
The information concerning the printer of the MultiLab 4010P-1W
is given in a separate chapter (see section 12 P
M
ULTILAB 4010P-1W), page 87).
RINTER (ONLY
1 Keypad
2 Display
3 Connectors
4Printer
5 Control panel of the printer
1.3Sensors
A measuring system ready to measure consists of the MultiLab 4010-1W meter
and a suitable sensor.
Suitable sensors are IDS pH sensors, IDS ORP sensors, IDS conductivity
sensors and IDS D.O. sensors.
1.3.1IDS sensors
IDS sensors
support the automatic sensor recognition
show only the settings relevant to the specific sensor in the setting menu
process signals in the sensor digitally so that precise and interference-free
measurements are enabled even with long cables
facilitate to assign a sensor to a measured parameter with differently colored
couplings
8ba76141e0708/2018
Page 9
MultiLab 4010-1WOverview
have quick-lock couplings with which to fix the sensors to the meter.
Sensor data from
IDS sensors
IDS sensors transmit the following sensor data to the meter:
SENSOR ID
– Sensor name
– Sensor series number
Calibration data
Measurement settings
The calibration data are updated in the IDS sensor after e ach calibration procedure. A message is displayed while the data are being updated in the sensor.
In the measured value display, you can display the sensor name
and series number of the selected sensor with the [Info] softkey.
You can then display further sensor data stored in the sensor with
the [More] softkey (see section 4.1.5 S
ENSORINFO, page 18).
1.3.2Wireless operation of IDS sensors
With the aid of the adapters in the IDS WA Kit, IDS sensors with plug head
connectors (variant P) can be wirelessly connected to your MultiLab 4010-1W.
(MultiLab 4010-1 IDS: from software version V2.00)
Two adapters, one at the IDS meter (IDS WA-M) and one at the sensor (IDS
WA-S), replace the sensor cable with an energy-saving Bluetooth LE radio
connection.
Further information on the wireless operation of IDS sensors:
Web resources
Operating manual of the IDS WA Kit.
1.3.3Automatic sensor recognition
The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibrating
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor serial number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to hide menus automatically that do not concern this sensor
ba76141e0708/20189
Page 10
OverviewMultiLab 4010-1W
To be able to use the automatic sensor recognition, a meter that supports the
automatic sensor recognition (e.g. MultiLab 4010-1W) and a digital IDS sensor
are required.
In digital IDS sensors, sensor data are stored that clearly identify the sensor.
The sensor data are automatically taken over by the meter.
10ba76141e0708/2018
Page 11
MultiLab 4010-1WSafety
2S afety
2.1Safety information
2.1.1Safety information in the operating manual
This operating manual provides important information on the safe operation of
the meter. Read this operating manual thoroughly and make yourself familiar
with the meter before putting it into operation or working with it. The operating
manual must be kept in the vicinity of the meter so you can always find the information you need.
Important safety instructions are highlighted in this operating manual. They are
indicated by the warning symbol (triangle) in the left column. The signal word
(e.g. "Caution") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not
followed.
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not followed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
2.1.2Safety signs on the meter
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
2.1.3Further documents providing safety information
The following documents provide additional information, which you should
observe for your safety when working with the measuring system:
•Operating manuals of sensors and other accessories
•Safety datasheets of calib ration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
ba76141e0708/201811
Page 12
SafetyMultiLab 4010-1W
2.2Safe operation
2.2.1Authorized use
The authorized use of the meter consists exclusively of the measurement of the
pH, ORP, conductivity and dissolved oxygen in a laboratory environment.
Only the operation and running of
the meter according to the instructions and technical specifications given
in this operating manual is authorized (see section 15 T
96).
Any other use is considered unauthorized.
2.2.2Requirements for safe operation
Note the following points for safe operation:
•The meter may only be operated accord ing to the authorized use specified
above.
•The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
•The meter may only be operated under the environmental conditions
mentioned in this operating manual.
•The meter may only be opened if this is explicitly described in this operating
manual (example: Inserting the batteries).
ECHNICALDATA, page
2.2.3Unauthorized use
The meter must not be put into operation if:
•it is visibly damaged (e.g. after being transported)
•it was stored under adverse conditions for a lengthy period of time (storing
conditions, see section 15 T
ECHNICALDATA, page 96).
12ba76141e0708/2018
Page 13
MultiLab 4010-1WCommissioning
3C ommissioning
3.1Scope of delivery
MultiLab 4010-1W / MultiLab 4010P-1W
4 batteries 1.5 V Mignon type AA
Power pack
USB cable (A plug on mini B plug)
Stand
Stand holder
Comprehensive operating manual
Short instructions
CD-ROM with
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1.Open the battery compartment (1) on the underside of the
meter.
1 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned correctly.
±signs on the batteries must correspond to the ± signs in
The
the battery compartment.
2.Place four batteries (type Mignon AA) in the battery compartment.
3.Close the battery compartment (1).
4.Set the date and time
(see section 4.4.5 E
page 24).
XAMPLE 2 ONNAVIGATION: SETTINGTHEDATEANDTIME,
14ba76141e0708/2018
Page 15
MultiLab 4010-1WCommissioning
1
2
3.3.2Connecting the power pack
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see section 15.2
G
ENERALDATA, page 96).
CAUTION
Use original power packs only (see section 15.2 GENERALDATA,
page 96).
1.Connect the plug of the power pack to the socket for the power pack on
the MultiLab 4010-1W.
2.Connect the original power pack to an easily accessible power outlet.
3.3.3Mounting the stand
The stand base can be mounted at the right side of the meter.
ba76141e0708/201815
Page 16
OperationMultiLab 4010-1W
4Operation
4.1General operating principles
4.1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g.
and release) in this operating manual.
A long keystroke (press and keep depressed for approx. 2 sec) is indicated by
the underscore behind the key symbol (e.g.
<ENTER>) generally indicates a short keystroke (press
<M>:Selects the measured parameter / Quits the settings
<CAL>:
<CAL__>:
<STO>:
<STO_ >:
<RCL>:
<RCL__>:
<><>:
<__>< __>:
<ENTER>:
<ENTER__>:
Calls up the calibration procedure
Displays the calibration data
Saves a measured value manually
Opens the menu for the automatic save function
Displays the manually stored measured values
Displays the automatically stored measured values
Menu control, navigation
Increments, decrements values
Increments, decrements values continuously
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
<AR>Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
16ba76141e0708/2018
Page 17
MultiLab 4010-1WOperation
3
2
1
4
5
6
8
HOLD AR
7
AutoCal TEC
i
03.04.2013
08:00
USB output
YSI
4.1.2Display
Example
pH
1 Status information (sensor)
2 Measured value
3 Measured parameter
4 Continuous measurement control (CMC function)
5 Sensor symbol (calibration evaluation, calibration interval)
6 Measured temperature (with unit)
7 Status information (meter)
8 Softkeys and date + time
4.1.3Status information (meter)
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (
<AR> key)
ZeroCalThe zero point is calibrated.
Batteries are almost empty
Data are automatically output to the USB-B interface at
intervals
ba76141e0708/201817
Page 18
OperationMultiLab 4010-1W
1
3
5
4
2
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
4.1.4Socket field
1 IDS sensors: (pH, ORP, conductivity, D.O.)
2 USB B (device) interface
3 Stirrer (interface for BOD sensor)
4 Power pack
5 Service interface
CAUTION
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
YSI IDS sensors and IDS adapters meet these requirements.
4.1.5Sensor info
You can display the current sensor data and sensor settings of a connected
sensor at any time. The sensor data are available in the measured value
display with the
<F1>/[Info] softkey.
1.In the measured value display:
Display the sensor data (sensor name, series number) with [
18ba76141e0708/2018
<F1>Info].
Page 19
MultiLab 4010-1WOperation
25.0
7.007
pH
°C
0
14
More
4110
B09250001
pH resolution0.001
mV resolution0.1
BufferYSI
Calibration interval7d
Unit for slopemV/pH
QSC:off
Software versionV 1.00
4110
B092500013
03.04.2013
08:00
Back
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
2.Display further sensor data (settings) with <F1>/[More].
ba76141e0708/201819
4.2Switching on
1.Switch the meter on with
The meter performs a self-test.
2.Connect the sensor.
The meter is ready to measure.
4.3Switching off
1.Switch the meter off with
<On/Off>.
<On/Off>.
Page 20
OperationMultiLab 4010-1W
4.4Navigation
4.4.1Operating modes
Operating
mode
MeasuringThe measurement data of the connected sensor are shown
CalibrationThe course of a calibration with calibration information, func-
Storing in
memory
Transmitting data
SettingThe system menu or a sensor menu with submenus, set-
4.4.2Measured value display
In the measured value display, you can
open the menu for calibration and measurement settings with
(short
keystroke)
Explanation
in the measured value display
tions and settings is displayed
The meter stores measuring data automatically or manually
The meter transmits measuring data and calibration records
to a USB-B interface automatically or manually.
tings and functions is displayed
<ENTER>
open the system menu with the sensor-independent settings by pressing
<ENTER>Storage & config for a <ENTER__>long keystroke, approx.
2s).
change the display in the selected measuring screen (e. g. pH ><− mV) by
pressing
4.4.3Menus and dialogs
The menus for settings and dialogs in procedures contain further sube lements.
The selection is done with the
played with a frame.
Submenus
The name of the submenu is displayed at the upper edge of the frame. Su bmenus are opened by confirming with
<M>.
<><> keys. The current selection is dis-
<ENTER>. Example:
20ba76141e0708/2018
Page 21
MultiLab 4010-1WOperation
General
Interface
Clock function
Service information
Reset
System
03.04.2013
08:00
Back
Language:Deutsch
Audio signal:off
Illumination:on
Contrast:12
Shutoff time:1 h
Temperature unit:°C
Stability control
:on
General
03.04.2013
08:00
Back
Calibration record
Calibration data storage
Buffer:YSI
Single-point calibration:yes
Calibration interval:7 d
Unit for slope:mV/pH
[
i
] 4.00 7.00 10.00 (25 °C)
pH
03.04.2013
08:00
Back
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with
the setting can be changed with
<>< > and <ENTER>. Example:
<ENTER>. Subsequently,
ba76141e0708/201821
Functions
Functions are designated by the name of the function. They are immediately
carried out by confirming with
<ENTER>. Example: Display the Calibration
record function.
Messages
i
Information is marked by the [
] symbol. It cannot be selected. Example:
Page 22
OperationMultiLab 4010-1W
Calibration record
Calibration data storage
Buffer:YSI
Single-point calibration:yes
Calibration interval:7 d
Unit for slope:mV/pH
[
i
] 4.00 7.00 10.00 (25 °C)
pH
03.04.2013
08:00
Back
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
System
Data storage
Storage & config
03.04.2013
08:00
Back
4.4.4Navigation example 1: Setting the language
1.Press the
<On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
2.Open the Storage & config menu with
The instrument is in the setting mode.
<ENTER__>.
22ba76141e0708/2018
3.Select the System submenu with
<><>.
The current selection is displayed with a frame.
4.Open the System submenu with
<ENTER>.
Page 23
MultiLab 4010-1WOperation
General
Interface
Clock function
Service information
Reset
System
03.04.2013
08:00
Back
Language:Deutsch
Audio signal:off
Illumination:on
Contrast:50 %
Shutoff time:1 h
Temperature unit:°C
Stability control
:on
General
03.04.2013
08:00
Back
Language:Deutsch
Audio signal:off
Illumination:on
Contrast:50 %
Shutoff time:1 h
Temperature unit:°C
Stability control
:on
General
03.04.2013
08:00
Back
5.Select the General submenu with <><>.
The current selection is displayed with a frame.
6.Open the General
submenu with <ENTER>.
7.Open the setting mode for the Language with
<ENTER>.
ba76141e0708/201823
8.Select the required language with
9.Confirm the setting with
<ENTER>.
The meter switches to the measuring mode.
The selected language is active.
<><>.
Page 24
OperationMultiLab 4010-1W
Date format:dd.mm.yy
Date:03.04.2013
Time:14:53:40
Clock function
03.04.2013
08:00
Back
4.4.5Example 2 on navigation: Setting the date and time
The meter has a clock with a date function. The date and time are indicated in
the status line of the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
The date and time are reset to default after a fall of the supply voltage (empty batteries).
Setting the date,
time and date format
The date format can be switched from the display of day, month, year
(dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1.In the measured value display:
Open the Storage & config menu with
<ENTER__>.
The instrument is in the setting mode.
2.Select and confirm the System / Clock function menu with
<><> and
<ENTER>.
The setting menu for the date and time opens up.
24ba76141e0708/2018
3.Select and confirm the Time menu with
The hours are highlighted.
4.Change and confirm the setting with
<><> and <ENTER>.
The minutes are highlighted.
5.Change and confirm the setting with
<><> and <ENTER>.
The seconds are highlighted.
<><> and <ENTER>.
Page 25
MultiLab 4010-1WOperation
6.Change and confirm the setting with <><> and <ENTER>.
The time is set.
7.If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8.To make further settings, switch to the next higher menu level with
or
Switch to the measured value display with
<M>.
The instrument is in the measuring mode.
<F1>.
ba76141e0708/201825
Page 26
pH valueMultiLab 4010-1W
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
5pH value
5.1Measuring
5.1.1Measuring the pH value
NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
1.Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
Stability control
(AutoRead)
& HOLD function
2.If necessary, select the measured parameter with
3.Adjust the temperature of the solutions and measure the current temperature if the measurement is made without a temperature sensor.
4.If necessary, calibrate or check the IDS pH sensor.
5.Immerse the IDS pH sensor in the test sample.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
<M>.
26ba76141e0708/2018
Hold function
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the setting for automatic Stability control (see section 9.5.3 A
TROL, page 74) in the System menu.
To start the Stability control function manually, the HOLD function must be
enabled.
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
UTOMATIC STABILITYCON-
Page 27
MultiLab 4010-1WpH value
You can terminate the HOLD function and the Stability control func-
tion with
<AR> or <M> at any time.
Stability control
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the measured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manually with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
ENSOR-INDEPENDENT
or
Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
parameter
pH value15 seconds
Temperature15 seconds
Time intervalStability in the time inter-
val
∆ : better than 0.01 pH
∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
5.1.2Measuring the temperature
For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
ba76141e0708/201827
Page 28
pH valueMultiLab 4010-1W
5.2pH calibration
5.2.1Why calibrate?
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
5.2.2When do you have to calibrate?
Routinely within the framework of the company quality assurance
When the calibration interval has expired
The buffer solutions used are automatically identified during the calibration
procedure. The relevant nominal values are used automatically.
Prerequisite: The buffer set used has been set in the meter (see section
9.1.2 B
Manual calibration (ConCal)
Any buffer solutions can be used. The nominal values of the buffer solutions
are manually entered by the user during the calibration procedure.
5.2.4Carrying out automatic calibration (AutoCal)
Make sure that, in the menu for measurement and calibration settings (in the
menu pH/
(see section 9.1.1 S
UFFERSETSFORCALIBRATION, page 64).
For most applications, automatic calibration (AutoCal) is the safest
and most convenient way of executing a calibration.
<ENTER>/Calibration / Buffer), the correct buffer set is selected
ETTINGSFORPH MEASUREMENTS, page 63).
Use one to five buffer solutions of the selected buffer set in any order.
Below, calibration with YSI buffers (YSI) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the procedure is identical.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
28ba76141e0708/2018
Page 29
MultiLab 4010-1WpH value
pH1
-180.0
24.8 °C
AutoCal YSI
01.07.2011
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal YSI
10.000
01.07.2011
08:00
AR
Buffer
mV
1.Connect the pH sensor to the meter.
The pH measuring window is displayed.
2.Keep the buffer solutions ready.
3.Start the calibration with
The calibration display for the first buffer appears (voltage display).
4.Thoroughly rinse the sensor with deionized water.
5.Immerse the sensor in the first buffer solution.
6.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
<CAL>.
<ENTER>.
ba76141e0708/201829
7.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
8.If necessary, finish the calibration procedure as a single-point calibration with
The calibration record is displayed.
<M>.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDSpH sensor.
Page 30
pH valueMultiLab 4010-1W
pH
0.0
24.8 °C
AutoCal YSI
7.000
01.07.2011
08:00
AR
Buffer
mV
pH
180.0
24.8 °C
AutoCal YSI
4.000
01.07.2011
08:00
AR
Buffer
mV
Continuing with two-
point calibration
9.Thoroughly rinse the sensor with deionized water.
10.Immerse the pH sensor in buffer solution 2.
11.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
12.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
<ENTER>.
The calibration display for the next buffer appears (voltage display).
Continuing with
three- to five-point
calibration
1 3.If necessary, finish the calibration procedure as a two-point calibration
<M>.
with
The calibration record is displayed.
14.Thoroughly rinse the sensor with deionized water.
15.Immerse the sensor in the next buffer solution.
16.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
<ENTER>.
30ba76141e0708/2018
Page 31
MultiLab 4010-1WpH value
pH
0.0
24.8 °C
ConCal
1
01.07.2011
08:00
Buffer
mV
17.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
Calibration is automatically completed after the last buffer of a buffer set has been measured. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.5Carrying out manual calibration (ConCal)
Make sure that, in the menu for measurement and calibration settings (in the
menu pH/
(see section 9.1.1 S
Use one to five buffer solutions in any order.
The pH values of the buffer solutions have to differ by at least one pH unit.
<ENTER>/Calibration / Buffer) the buffer set ConCal is selected
ETTINGSFORPH MEASUREMENTS, page 63).
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH sensor to the meter.
The pH measuring window is displayed.
2.Keep the buffer solutions ready.
3.Start the calibration with
The calibration display for the first buffer appears (voltage display).
<CAL>.
4.Thoroughly rinse the sensor with deionized water.
5.Immerse the pH sensor in buffer solution 1.
6.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
<ENTER>.
ba76141e0708/201831
Page 32
pH valueMultiLab 4010-1W
pH
0.0
24.8 °C
ConCal
7.000
01.07.2011
08:00
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
01.07.2011
08:00
Buffer
7.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
The pH value of the buffer solution is displayed.
Continuing with two-
point calibration
8.Set the nominal buffer value for the measured temperature with
<><>.
9.Accept the calibration value with
The calibration display for the next buffer appears (voltage display).
1 0 .If necessary, finish the calibration procedure as a single-point calibration
with
<M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDSpH sensor.
11.Thoroughly rinse the sensor with deionized water.
12.Immerse the pH sensor in buffer solution 2.
13.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
<ENTER>.
<ENTER>.
32ba76141e0708/2018
Page 33
MultiLab 4010-1WpH value
pH
4.035
24.8 °C
ConCal
2
01.07.2011
08:00
Buffer
pH
9.958
24.8 °C
ConCal
3
01.07.2011
08:00
Buffer
14.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
The pH value of the buffer solution is displayed.
15.Set the nominal buffer value for the measured temperature with
<><>.
Continuing with
three- to five-point
calibration
16.Accept the calibration value with
The calibration display for the next buffer appears (voltage display).
17.If necessary, finish the calibration procedure as a two-point calibration
<M>.
with
The calibration record is displayed.
18.Thoroughly rinse the sensor with deionized water.
19.Immerse the sensor in the next buffer solution.
20.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
21.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
<ENTER>.
<ENTER>.
The pH value of the buffer solution is displayed.
ba76141e0708/201833
Page 34
pH valueMultiLab 4010-1W
22.Set the nominal buffer value for the measured temperature with
<>< >.
23.Accept the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
24.If necessary, use
<M> to finish the calibration.
The calibration record is displayed.
or
Continue calibrating using the next buffer with
<ENTER>.
After the fifth buffer has been measured the calibration is automatically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.6Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
CalibrationDetermined
Displayed calibration data
values
1-pointAsymmetry Zero point = Asymmetry
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-pointAsymmetry
Slope
3-point to 5point
Asymmetry
Slope
Zero point = Asymmetry
Slope = Slope
Zero point = Asymmetry
Slope = Slope
The calibration line is calculated by
linear regression.
You can display the slope in the units, mV/pH or % (see section
9.1.1 S
ETTINGSFORPH MEASUREMENTS, page 63).
5.2.7Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
34ba76141e0708/2018
Page 35
MultiLab 4010-1WpH value
The calibration record of the last calibration is available in the menu Calibration
/ Calibration record. To open it in the measured value display, press the
<CAL__> key.
The calibration records of the last 10 calibrations are available in the menu Calibration / Calibration data storage. To open the Calibration menu in the
measured value display, press the
<ENTER> key.
Menu itemSetting/
function
Calibration /
Calibration data storage / Display
Calibration /
Calibration data storage / Output to
RS232/USB
-Displays the calibration records.
-Outputs the calibration data storage
Explanation
Further options:
Scroll through the calibration
records with
Output the displayed calibration
record to the interface with
[USB output].
Output all calibration records to
the interface with
output].
Quit the display with
<ENTER>.
or
Switch directly to the measured
value display with
to the interface (see section 12
RANSMITTINGDATA, page 120).
T
<><>.
<F2>/
<F2__>[USB
<F1>/[Back]
<M>.
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
DisplayCalibration
record
+++
++
+
Zero point
[mV]
-15 ... +15-60,5 ... -58,0
-20 ... <-15
or
>+15 ... +20
-25 ... <-20
or
>+20 ... +25
Slope [mV/pH]
>-58.0 ... -57.0
-61.0 ... <-60.5
or
>-57.0 ... -56.0
ba76141e0708/201835
Page 36
pH valueMultiLab 4010-1W
4010-1W
Ser. no. 11292113
CALIBRATIONpH
01.02.2014 15:55
Ser. no. 10501234
YSI
Buffer 14.00
Buffer 27.00
Buffer 310.00
Voltage 1184.0 mV
Voltage 23.0 mV
Voltage 3-177.0 mV
Temperatur 124.0 °C
Temperatur 224.0 °C
Temperatur 324.0 °C
Slope-60.2 mV/pH
Asymmetry4.0 mV
Sensor +++
etc...
Calibration record
(USB output)
DisplayCalibration
record
-
Zero point
[mV]
-30 ... <-25
or
->+25 ... +30
Slope [mV/pH]
-62.0 ... <-61.0
or
>-56.0 ... -50.0
Clean the IDS sensor according to the sensor operating manual
ErrorError
Error elimination (see section 14 W
<-30
or
>+30
HATTODOIF..., page 92)
<-62.0
or
> -50,0
For pH IDS sensors you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 5.2.9 QSC
SENSORQUALITYCONTROL), page 38).
(
FUNCTION
36ba76141e0708/2018
Page 37
MultiLab 4010-1WpH value
1
2
4
3
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
5.2.8Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is displayed in the measured value display. Here you can very clearly see whether
or not the current measured value is in the calibrated part of the measuring
range.
The following information is displayed:
1 Currently measured pH value (needle)
2 Marking lines for all nominal buffer values used with the last valid cali-
bration
3 Measuring range for which a valid calibration is available. Measured
values in this range are suitable for documentation.
4 Measuring range for which no valid calibration is available (shaded).
Measured values in this range are not suitable for documentation.
Calibrate the meter with buffers covering this measuring range.
If the current measured value is outside the calibrated range, this
area is shaded stronger.
If a measured value is outside the measuring range pH 0 - 14, overflow arrows are displayed at the left or right edge of the measuring
range.
The limits of the calibrated range are determined by the buffers used for calibration:
Lower limit: Buffer with lowest pH value - 2 pH units
Upper limit: Buffer with highest pH value + 2 pH units
ba76141e0708/201837
Page 38
pH valueMultiLab 4010-1W
1
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
5.2.9QSC function (sensor quality control)
General information
on the QSC function
The QSC function (Quality Sensor Control) is a new sensor evaluation for
digital IDS sensors. It evaluates the condition of an IDS pH sensor individually
and with a very fine grading.
The QSC scale shows the current sensor evaluation with an indicator on the
display.
1QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
Sensor evaluation
with / without
QSC function
QSC calibrationThe QSC function is enabled by once carrying out an additional three-point cali-
In the USB output the sensor evaluation is given as a percentage (1-100).
The finely graded sensor evaluation of the QSC function promptly calls your
attention to changes of the sensor.
Thus you can do what is necessary to restore the optimum measuring quality
(e.g. clean, calibrate or replace the sensor). clean, calibrate or replace the
sensor).
With QSC functionWithout QSC function (sensor sym-
bol)
Very fine grading of the sensor evaluation (100 grades)
The reference value is individually
determined for each sensor during
the QSC initial calibration.
Low tolerances for zero point and
slope when using QSC buffer solutions
Additional QSC calibration required
(with special QSC buffer set)
bration with special QSC buffer solutions. It covers the measuring range of the
sensor (pH 2 to pH 11). The QSC initial calibration determines the actual condition of the sensor and stores it as a reference in the sensor.
To meet the high requirements of a QSC initial calibration, the QSC initial calibration should optimally be carried out with the initial commissioning of the
sensor.
Rough grading of the sensor evaluation (4 grades)
A theoretical reference value is used
for all sensors
Greater tolerances for zero point and
slope when using commercial buffer
sets
No additional calibration required
38ba76141e0708/2018
Page 39
MultiLab 4010-1WpH value
pH1
-180.0
24.8 °C
AutoCal OSC-Kit
01.07.2011
08:00
Buffer
mV
Carry out the normal calibrations for your special measuring range with your
usual standard solutions as previously done.
As soon as the QSC function was enabled for an IDS sensor, it is
not possible to return to the sensor evaluation with the sensor
symbol for this sensor.
Carrying out a QSC
initial calibration
Carry out the QSC initial calibration very carefully. It determines the
reference value for the sensor. This reference value cannot be
overwritten or reset.
As soon as the QSC function was enabled, it is not possible to
return to the sensor evaluation with the sensor symbol.
1.Prepare for calibration the meter, sensor and the buffer solutions of the
QSC-Kit.
2.Open the menu for measurement settings with
3.In the QSC menu, select First calibration with
The calibration display appears. AutoCal QSC-Kit is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC calibration.
<ENTER>.
<>< >.
ba76141e0708/201839
4.Start the measurement with
Calibration with the buffers of the QSC-Kit is done like a normal threepoint calibration (see section 5.2.4 C
TION (AUTOCAL), page 28).
5.As soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
initial calibration.
The QSC initial calibration is completed. The sensor is calibrated. If you want
to calibrate with special buffers for your measurements, you can subsequently
carry out a normal calibration with your buffers. The reference values determined with the QSC calibration are also used for the evaluation of normal calibrations. In the measured value display, the QSC scale of the QSC function is
always displayed. A double arrow on the QSC scale indicates the current sen-
<ENTER>.
ARRYINGOUTAUTOMATICCALIBRA-
Page 40
pH valueMultiLab 4010-1W
1
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
sor evaluation.
1QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
Carrying out a QSC
control calibration
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).
You can carry out QSC control calibrations at greater intervals than normal calibrations.
1.Prepare for calibration the meter, sensor and the buffer solutions of the
QSC-Kit.
2.Open the menu for measurement settings with
3.In the QSC menu, select Control calibration with
The calibration display appears. AutoCal QSC-Kit is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC control calibration.
4.Start the measurement with
Calibration with the buffers of the QSC-Kit is done like a normal threepoint calibration (see section 5.2.4 C
TION (AUTOCAL), page 28).
As soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
control calibration.
<ENTER>.
ARRYINGOUTAUTOMATICCALIBRA-
<ENTER>.
<><>.
40ba76141e0708/2018
Page 41
MultiLab 4010-1WORP
U
157.0 mV
01.02.2014
08:00
i
USB output
6ORP
6.1Measuring
6.1.1Measuring the ORP
NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.
1.Connect the ORP sensor to the meter.
The ORP measuring window is displayed.
Stability control
(AutoRead)
& HOLD function
2.Temper the test sample or measure the current temperature.
3.Check the meter with the ORP sensor.
4.Immerse the ORP sensor in the test sample.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the setting for automatic Stability control (see section 9.5.3 A
TROL, page 74) in the System menu.
To start the Stability control function manually, the HOLD function must be
enabled.
UTOMATIC STABILITYCON-
ba76141e0708/201841
Page 42
ORPMultiLab 4010-1W
Hold function
Stability control
1.Freeze the measured value with
<AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the HOLD function and the Stability control func-
tion with
<AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the mea sured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manually with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
Criteria for a stable
measured value
ENSOR-INDEPENDENT
3.Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
parameter
ORP15 seconds
Temperature15 seconds
Time intervalStability in the time
interval
∆ : better than 0.3 mV
∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
6.1.2Measuring the temperature
For reproducible ORP measurements, it is essential to measure the tempera-
42ba76141e0708/2018
Page 43
MultiLab 4010-1WORP
ture of the test sample.
When a sensor with integrated temperature sensor is operated, the tempera-
ture value is displayed and recorded together with the measured value.
6.2ORP calibration
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
ba76141e0708/201843
Page 44
Dissolved oxygenMultiLab 4010-1W
O
2
mg/l
7.92
25.0
°C
760 mm Hg
01.02.2014
08:00
i
USB output
7Dissolved oxygen
The general procedures for measuring, calibrating and setting are very similar
with the various types of IDS-D.O. sensors.
Any individual special features of the IDS-D.O. sensors are indicated in the
general procedures.
The individual feature of the FDO 4410 is, for example, the high stability of its
measuring characteristics during its total lifetime. Therefore, this sensor does
not have to be calibrated regularly. It can simply be checked (FDO Check)
instead.
A sensor with stirrer can only be operated while the meter is linepowered.
7.1Measuring
7.1.1Measuring D.O.
1.Connect the IDS-D.O. sensor or a D.O. sensor with IDS adapter to the
meter.
The D.O. measuring screen is displayed.
2.If necessary, select the measured parameter with
3.Check or calibrate the meter with the sensor.
Only in special cases does the FDO 4410-D.O. sensor require ca libration. Regular checking with the aid of the FDO Check is sufficient.
4.Immerse the D.O. sensor in the test sample.
<M>.
measured parameter
44ba76141e0708/2018
Selecting the
displayed
You can switch between the following displays with
D.O. concentration [mg/l]
D.O. saturation [%]
<M>:
Page 45
MultiLab 4010-1WDissolved oxygen
Salinity correctionWhen measuring the D.O. concentration [mg/l] of solutions with a salt content
of more than 1 psu, a salinity correction is required. For this, you have to me asure and input the salinity of the measured medium first.
When the salinity correction is switched on, the [Sal] indicator is displayed in
the measuring screen.
You can switch the salinity correction on or off and enter the salinity
in the menu for calibration and measurement settings (see section
9.3.1 S
CALIBRATIONSETTINGS), page 67).
ETTINGSFOR D.O. SENSORS (MENUFORMEASUREMENTAND
Air pressure correc-
tion
(DO % Saturation
local function)
Stability control
(AutoRead)
& HOLD function
The integrated air pressure sensor of the MultiLab 4010-1W measures the
current air pressure.
During calibration, the air pressure correction function is automatically activated. While the parameter oxygen saturation [%] is displayed, the air pressure
correction is applied if the DO % Saturation local function is active.
The current air pressure is shown on the display when an IDS oxygen sensor
is connected.
The air pressure correction (function DO % Saturation local) is
switched on or off in the menu for calibration and measurement settings (see section 9.3.3 DO % S
ATURATIONLOCAL, page 69).
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
Irrespective of the setting for automatic Stability control (see section 9.5.3
UTOMATIC STABILITYCONTROL, page 74) in the System menu, yo u can start a
A
measurement with Stability control manually at any time.
To start the Stability control function manually, the HOLD function must be
enabled.
Hold function
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the HOLD function and the Stability control func-
tion with
ba76141e0708/201845
<AR> or <M> at any time.
Page 46
Dissolved oxygenMultiLab 4010-1W
Stability control
2.Using
<ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the measured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manually with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
ENSOR-INDEPENDENT
or
Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured parameterTime intervalStability in the time
∆ : better than 0.03 mg/l
∆ : better than 0.4 %
∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
7.1.2Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the temperature of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor integrated in the IDS sensor.
46ba76141e0708/2018
Page 47
MultiLab 4010-1WDissolved oxygen
7.2FDO Check (checking the FDO 4410)
The sensor is checked in water vapor-saturated air in the check and storage
vessel FDO Check.
7.2.1Why should you check the sensor?
With the FDO Check procedure you can find out in a simple manner whether
the FDO 4410 D.O. sensor should be cleaned or calibrated.
7.2.2When should you check the sensor?
Checking can be useful in the following cases:
When the check interval has expired
(the [check] status indicator is displayed).
If the measured values seem to be implausible
If you assume that the sensor cap is contaminated or at the end of
its lifetime
After the sensor cap was exchanged
Routinely within the framework of the company quality assurance
7.2.3Perform a FDO Check
Proceed as follows to carry out the FDO Check:
1.Connect the D.O. sensor to the meter.
2.Place the D.O. sensor in the FDO Check check and storage vessel.
3.Close the check and storage vessel FDO Check.
The sponge in the check and storage beaker must be moist (not
wet). Leave the sensor in the check and storage beaker long
enough so it can
adapt to the ambient temperature.
4.In the menu for measurement and calibration settings (
<ENTER> / FDO
Check / Start FDO Check), start the FDO Check.
The meter switches to the measured parameter, %.
During the FDO Check the DO % Saturation local function is automatically activated.
ba76141e0708/201847
Page 48
Dissolved oxygenMultiLab 4010-1W
O
2
99.8
%
25.0
°C
FDO Check
[HOLD]
01.02.2014
08:00
[L][t90= 30 s]
5.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
6.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>.
<ENTER>.
The measured value is frozen.
7.Switch to the measured value display with
The check measurement is not documented.
After the FDO Check the DO % Saturation local function is automatically
reset to the setting before the FDO Check.
7.2.4Evaluation
The evaluation is based on the accuracy required and de fined by the user (e.g.
± 2 %). Together with the nominal value (100 %) this results in a validity scope
for the check of 98 to 102 % (see example).
If the measured value is within the validity scope, no cleaning or user
calibration is required.
If the measured value is outside the validity scope, the sensor shaft and
membrane should be cleaned, and the check should then be repeated
(see section 5.4.1).
<M>.
Example:
– Required accuracy: ± 2 %.
– In water vapor-saturated air or air-saturated water, the nominal
value for the relative D.O. saturation (abbreviated: saturation) is
100 %.
– Therefore, the validity scope is 98 ... 102 %
– The check resulted in a measured value of 99.3 %
The measurement error is within the specified validity scope.
No cleaning or user calibration is required.
48ba76141e0708/2018
Page 49
MultiLab 4010-1WDissolved oxygen
7.3Calibration
7.3.1Why calibrate?
D.O. sensors age. Aging changes the zero point and slope of the D.O. sensor.
As a result, an inexact measured value is displayed. Calibration determines
and stores the current values of the zero point and slope.
The FDO 4410 D.O. sensor ages so little it does not have to be regularly calibrated.
To detect changes of the sensor as early as possible, the
FDO Check procedure can be useful (see section 7.2 FDO C
CHECKINGTHE FDO 4410), page 47).
(
7.3.2When to calibrate?
If your evaluation of the FDO Check suggests calibrating (only FDO 4410)
When the calibration interval has expired
HECK
When your accuracy requirements are especially high
Routinely within the framework of the company quality assurance
After a Zero calibration.
7.3.3Calibration procedures
The MultiLab 4010-1W provides 2 calibration procedures:
Calibration in water vapor-saturated air.
Calibration via a comparison measurement (e.g. Winkler titration according
to DIN EN 25813 or ISO 5813). At the same time, the relative slope is
adapted to the comparison measurement by a correction multiplier. When
the correction multiplier is active, the [Factor] indicator appears in the mea-
suring window.
For both calibration procedures, an additional Zero calibration is
possible (see section 7.3.6 Z
EROCALIBRATION, page 52).
7.3.4Calibration in water vapor-saturated air
For this calibration procedure, the Comparison meas. setting must be set to off
in the Calibration menu.
As the calibration vessel use a BOD bottle that contains a small amount of
clean water (approx. 40 ml). The sensor must not be immersed in the water.
To calibrate the FDO 4410, use the calibration and storage vessel
FDO Check. The sponge in the check and storage beaker must be
moist (not wet).
ba76141e0708/201849
Page 50
Dissolved oxygenMultiLab 4010-1W
O
2
99.8 %
25.0
°C
Cal
01.02.2014
08:00
DO Sat
Proceed as follows to calibrate the D.O. sensor:
1.Connect the D.O. sensor to the meter.
2.Put the D.O. sensor into the calibration vessel.
3.For FDO 4410 with the check and storage vessel FDO Check:
Close the check and storage vessel FDO Check.
Leave the sensor in the calibration vessel or check and storage
vessel (FDO 4410) long enough (at least 15 minutes) until the air is
saturated with water vapor and the sensor has adapted to the
ambient temperature.
4.Start the calibration with
5.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
6.Wait for the AutoRead measurement to be finished (beep, [HOLD][AR]
status indicator).
The calibration record is displayed and output to the interface.
<CAL>.
<ENTER>.
7.Switch to the measured value display with
7.3.5Calibrating with Comparison meas. (e.g. Winkler titration)
With the calibration procedure Comparison meas., the measured value of the
sensor is adjusted to the nominal value of a comparison solution via a correc-
50ba76141e0708/2018
tion multiplier. The current correction multiplier is documented in the sensor
i
menu (
For this calibration procedure, the Comparison meas. setting must be set to on
in the Calibration menu.
Factor = x.xxx) and in the calibration record.
<F1>/[continue].
Page 51
MultiLab 4010-1WDissolved oxygen
O
2
7.92
mg/l
25.1
°C
Compare
03.04.2013
08:00
DO Adjust
O
2
7.92
mg/l
20.3°C
Compare
03.04.2013
08:00
DO Adjust
Proceed as follows to calibrate the D.O. sensor:
1.Connect the D.O. sensor to the meter.
2.Immerse the D.O. sensor in the reference solution.
3.Start the calibration with
4.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.Wait for the measurement with stability control to be finished (beep,
[HOLD][AR] status indicator).
<CAL>.
<ENTER>.
6.Using
to the nominal value (value of the comparison measurement). Subsequently take over the adjustment with
The meter switches to the measured value display.
The status display [Factor] is active.
ba76141e0708/201851
<><>, set the measured value to adjust the displayed value
<ENTER>.
Page 52
Dissolved oxygenMultiLab 4010-1W
O
2
0.9
%
Cal
03.04.2013
08:00
25.1°C
DO Zero
7.3.6Zero calibration
With a Zero calibration, the zero point of the sensor is redetermined by calibrating in a zero solution.
For most applications, an additional Zero calibration is not required and not
recommended. Only with very rare applications can the accuracy of a calibration be improved if a Zero calibration was carried out before.
The FDO 4410 D.O. sensor ages so little it does not have to be zero
point calibrated.
A zero point calibration is not allowed for this sensor.
The zero point calibration is best carried out prior to calibrating with a calibration procedure (e.g. calibration in water vapor-saturated air or calibration with
comparison measurement).
1.Connect the D.O. sensor to the meter.
2.Place the D.O. sensor in a solution that does not contain any dissolved
oxygen.
A solution not containing any dissolved oxygen can be prepared by
dissolving approx. 8 to 10 g sodium sulfite (Na
water. Carefully mix the solution. It may take up to 60 minutes until
the solution is free of oxygen.
3.In the menu for measurement and calibration settings (
ibration / Zero calibration), start the Zero calibration.
The calibration point for the measured value 0 (DO Zero) is displayed.
) in 500 ml tap-
2SO3
<ENTER> / Cal-
4.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
5.Wait for the AutoRead measurement to be finished (beep, [HOLD][AR]
status indicator).
The current value is set to zero.
The calibration record is displayed.
<ENTER>.
52ba76141e0708/2018
Page 53
MultiLab 4010-1WDissolved oxygen
6.Switch to the measured value display with
The zero point is calibrated.
The [ZeroCal] status indicator is displayed.
7.Carry out a calibration (see section 7.3.3 C
page 49).
7.3.7Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is available in the menu Calibration
/ Calibration record. To open it in the measured value display, press the
<F1>/[continue].
ALIBRATIONPROCEDURES,
<CAL__> key.
The calibration records of the last 10 calibrations are available in the menu Calibration / Calibration data storage / Display. To open the Calibration menu in the
measured value display, press the
<ENTER> key.
Menu itemSetting/
function
Calibration /
Calibration data storage / Display
Calibration /
Calibration data storage / Output to
RS232/USB
-Displays the calibration records.
-Outputs the calibration data stor-
Explanation
Further options:
Scroll through the calibration
records with
Output the displayed calibra-
tion record to the interface with
<><>.
<F2>/[USB output].
Output all calibration records to
the interface with
<F2__>[USB output].
Quit the display with
[Back] or
Switch directly to the mea-
sured value display with
age to the interface (see section
RANSMITTINGDATA, page
12 T
120).
<ENTER>.
<F1>/
<M>.
Calibration evalua-
tion
ba76141e0708/201853
After calibration, the meter automatically evaluates the current status of the ca l-
Page 54
Dissolved oxygenMultiLab 4010-1W
CALIBRATION Ox
03.04.2013 07:43:33
FDO 4410
Ser. no. 12B100016
SC-FDO 12B100015
Sensor+++
ibration. The evaluation appears on the display and in the calibration record.
For evaluation, the characteristic curve of the sensor that was
determined is compared to the characteristic curve of an ideal sensor at similar environmental conditions (relative slope S): S = S
sor
/ S
ideal sensor
. An ideal sensor has a slope of 1.
sen-
Calibration evalua-
tion FDO 4410
Calibration evalua-
tion ProOBOD,
4100 ProBOD,
YSI 5010 with 4011
Adapter
DisplayCalibration recordRelative slope
+++S = 0.94 ... 1.06
++S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
+S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
ErrorErrorS < 0.90
Error elimination (see section 14 W
HATTODOIF...,
or
S > 1.10
page 92)
DisplayCalibration record
+++
Calibration record
(USB output,
example FDO 4410)
Error elimination (see section 14 W
HATTODOIF..., page 92)
54ba76141e0708/2018
Page 55
MultiLab 4010-1WConductivity
μS/cm
[Tr25][nLF][ 0.835 1/cm]
1417
25.0
°C
01.02.2014
08:00
i
USB output
8Conductivity
8.1Measuring
8.1.1Measuring the conductivity
NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
1.Connect the conductivity sensor to the meter.
The conductivity measuring window is displayed.
The measuring cell and cell constant for the connected IDS conductivity
sensor are automatically taken over.
Selecting the
displayed
measured parameter
2.If necessary, press
3.Immerse the conductivity sensor in the test sample.
You can switch between the following displays with
Total dissolved solids TDS [mg/l] / [g/l]
The multiplier to calculate the total dissolved solids is set to 1.00 in the factory.
You can adjust this multiplier to meet your requirements in the range 0.40 ...
1.00. The multiplier is set in the menu for the parameter, TDS.
Determining the TDS multiplier:
See section 19.3 C
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ALCULATETHE TDS MULTIPLIER, page 109.
Page 56
ConductivityMultiLab 4010-1W
ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the setting for automatic Stability control (see section 9.5.3 A
TROL, page 74) in the System menu.
UTOMATIC STABILITYCON-
To start the Stability control function manually, the HOLD function must be
enabled.
Hold function
Stability control
1.Freeze the measured value with
<AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the HOLD function and the Stability control func-
tion with
<AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the measured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manually with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
ENSOR-INDEPENDENT
or
Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
Criteria for a stable
The Stability control function checks whether the measured values are stable
measured value
56ba76141e0708/2018
Page 57
MultiLab 4010-1WConductivity
within the monitored time interval.
Measured
Time intervalStability in the time interval
parameter
Conductivity ϰ10 seconds
∆ ϰ: better than
1.0% of measured value
Temperature15 seconds
∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
8.1.2Measuring the temperature
For reproducible conductivity measurements, it is essential to measure the
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
8.2Temperature compensation
The calculation of the temperature compensation is based on the preset reference temperature, 20 °C or 25 °C. It appears on the display as Tr20 or Tr25.
You can select one of the following temperature compensation methods:
Nonlinear temperature compensation (nLF) according to EN 27 888
Linear temperature compensation (Lin) with adjustable coefficient
No temperature compensation (off)
The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 9.4.1 S
TINGSFOR IDS CONDUCTIVITYSENSORS, page 70).
ET-
Application tipsSelect the following temperature compensations give n in the table according to
the respective test sample:
Test sampleTemperature compensationDisplay
Natural water
(ground water, surface
nLF
according to EN 27 888
nLF
water, drinking water)
Ultrapure waternLF
nLF
according to EN 27 888
Other aqueous solutions
Lin
adjustable temperature coeffi-
Lin
cient 0.000 ... 10.000 %/K
ba76141e0708/201857
Page 58
ConductivityMultiLab 4010-1W
Test sampleTemperature compensationDisplay
Salinity (seawater)Automatic nLF according to
Sal, nLF
IOT (International Oceanographic Tables)
8.3Calibration
8.3.1Why calibrate?
Aging slightly changes the cell constant, e. g. due to coatings. As a result, an
inexact measured value is displayed. The original characteristics of the cell can
often be restored by cleaning the cell. Calibration determines the current value
of the cell constant and stores this value in the meter.
Thus, you should calibrate at regular intervals.
8.3.2When to calibrate?
Routinely within the framework of the company quality assurance
When the calibration interval has expired
8.3.3Calibration procedure
The MultiLab 4010-1W provides 2 calibration procedures:
Determining the cell constant
Calibration in the check- and calibration standard 0.01 mol/l KCl (1413 µS/
cm @25 °C)
Safe and simple calibration procedure for IDS conductivity sensors with a
cell constant in the range 0.450 ... 0.500 cm
-1
.
Determining the cell constant
Calibration with freely selectable check- and calibration standard
Comprehensive and flexible calibration procedure for all IDS conductivity
sensors, irrespective of the cell constant.
Which calibration procedures can be used depends on the conductivity sensor
used. The menu for measurement settings automatically displays only those
settings and calibration procedures that are available for the sensor.
With conductivity sensors for which a cell constant was set, no
sensor symbol appears on the display.
If a calibration interval was specified, it is not active.
8.3.4Determining the cell constant (calibration in the check- and cali-
bration standard)
For this calibration procedure, IDS conductivity sensors with a cell constant in
58ba76141e0708/2018
Page 59
MultiLab 4010-1WConductivity
1/cm
0.475
25.0
°C
03.04.2013
08:00
the range 0.450 ... 0.500 cm-1 are suitable, e.g. 4310.
IDS conductivity sensors with other cell constants are unsuitable for this simple
calibration procedure. As an alternative, you can manually determine the cell
constant and adjust it in the menu (see section 8.3.5 S
You can determine the actual cell constant of the IDS conductivity sensor within
a valid range (e.g. 4310: 0.450 ... 0.500 cm
The cell constant is determined in the check- and calibration standard 0.01 mol/
l KCl (1413 µS/cm @ 25 °C).
In the default condition, the calibrated cell constant is set to the nominal cell
constant of the IDS sensor (e.g. 4310: 0.475 cm
For this calibration procedure, the Type setting must be set to cal. Proceed as
follows to determine the cell constant:
1.Connect the conductivity sensor to the meter.
2.In the measured value display, select the conductivity parameter with
-1
).
-1
).
<M>.
3.Start the calibration with
The cell constant that was calibrated last is displayed.
<CAL>.
4.Immerse the conductivity sensor in the check- and calibration standard
0.01 mol/l KCl (1413 µS/cm @ 25 °C).
5.Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
6.Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>.
<ENTER>.
The calibration record is displayed and output to the interface.
8.3.5Setting the cell constant
(calibration with freely selectable check- and calibration standard)
You can set within a valid range the cell constant of the IDS conductivity sensor
(setting range: see sensor operating manual).
With a freely selectable check- and calibration standard with known conductivity nominal value (within the measuring range of the sensor), you can exactly
adjust the cell constant to the check- and calibration standard by the changing
conductivity value displayed.
In the default condition, the cell constant is set to the nominal cell constant of
the IDS sensor.
For this calibration procedure, the Type setting must be set to man. Proceed as
follows to set the cell constant:
1.Connect the conductivity sensor to the meter.
2.In the measured value display, select the conductivity parameter with
<M>.
3.Open the menu for measurement settings with
<ENTER>.
4.Select the setting of the cell constant
(4310: Menu Type: manandCell const. man
4320: Menu Cell constant)
The current conductivity value is displayed in the status line.
5.Set the suitable temperature compensation and reference temperature
for the check- and calibration standard.
6.Immerse the conductivity sensor in the check- and calibration standard.
Wait for a stable measured value.
60ba76141e0708/2018
7.Using
<><>, adjust the cell constant until the displayed conductivity
value ([i] ϰ = ...) corresponds to the nominal value.
8.Switch to the measured value display with
The setting of the cell constant is taken over.
No calibration record is output.
<M>.
Page 61
MultiLab 4010-1WConductivity
8.3.6Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is available in the menu Calibration
/ Calibration record. To open it in the measured value display, press the
<CAL__> key.
The calibration records of the last 10 calibrations are available in the menu Calibration / Calibration data storage / Display. To open the Calibration menu in the
measured value display, press the
<ENTER> key.
Menu itemSetting/
function
Calibration /
-Displays the calibration records.
Calibration data
storage / Display
Calibration /
-Outputs the calibration data storage
Calibration data
storage / Output to
RS232/USB
Explanation
Further options:
Scroll through the calibration
records with
<><>.
Output the displayed calibration
record to the interface with
<F2>/[USB output].
Output all calibration records to
the interface with
output].
Quit the display with
[Back] or
<ENTER>.
Switch directly to the measured
value display with
to the interface (see section 12
RANSMITTINGDATA, page 120).
T
<F2__>[USB
<F1>/
<M>.
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the ca libration. The evaluation appears on the display and in the calibration record.
DisplayCalibration recordCell constant [cm-1]
+++Within the range
0.450 ... 0.500 cm
-1
ErrorErrorOutside the range
Error elimination (see section 14
HATTODOIF..., page 92)
W
ba76141e0708/201861
0.450 ... 0.500 cm
-1
Page 62
ConductivityMultiLab 4010-1W
CALIBRATION Cond
03.04.2013 07:43:33
4310
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor +++
Calibration record
(USB output)
62ba76141e0708/2018
Page 63
MultiLab 4010-1WSettings
9Settings
9.1pH measurement settings
9.1.1Settings for pH measurements
SettingsThe settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required measured parameter in the measured value display and press the
After completing the settings, switch to the measured value display with
Default settings are printed in bold.
<ENTER> key.
<M>.
Menu itemPossible
setting
Calibration /
Calibration record
Calibration /
Calibration data storage / Display
Calibration /
Calibration data storage / Output to
RS232/USB
Calibration / BufferYSI
Calibration / Singlepoint calibration
Calibration /
Calibration interval
-Displays the calibration record of the last calibration.
-Displays the last calibration records (max. 10)
-Outputs the calibration data storage to the interface
ConCal
NIST/DIN
...
yes
no
1 ... 7 ... 999 dCalibration interval for the IDS pH sensor (in days).
Explanation
Buffer sets to be used for pH calibration.
More buffers and details: see section 9.1.2 B
FORCALIBRATION, page 64 and section 5.2 PH CALIBRATION,
page 28.
Quick calibration with 1 buffer
The meter reminds you to calibrate regularly by the flashing
sensor symbol in the measuring screen.
UFFERSETS
Calibration / Unit for
slope
QSC / First calibration-Starts the initial calibration with QSC buffers.
QSC / Record of first
calibration
QSC / Control calibration
Man. temperature-25 ... +25 ...
ba76141e0708/201863
mV/pH
%
-Displays the calibration record of the QSC initial calibration.
-Starts the control calibration with QSC buffers.
+130 °C
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).
This menu item is only available as long as no initial calibration was carried out with the connected IDS sensor.
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.
Entry of the manually determined temperature.
This menu item is only available if an IDS adapter is connected.
Page 64
SettingsMultiLab 4010-1W
Menu itemPossible
Explanation
setting
pH resolution0.001
Resolution of the pH display
0.01
0.1
mV resolution0.1
Resolution of the mV display
1
Reset-Resets all sensor settings to the delivery condition (see sec-
tion 9.6.1 R
ESETTINGTHEMEASUREMENTSETTINGS, pag e 75).
9.1.2Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
No.Buffer set *pH valuesat
1YSI *4.000
25 °C
7.000
10.000
2ConCalAnyAny
3NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1.679
4.006
6.865
25 °C
9.180
12.454
4TEC
Technical buffers
2.000
4.010
25 °C
7.000
10.011
5Merck 1*4.000
20°C
7.000
9.000
6Merck 2 *1.000
20°C
6.000
8.000
13.000
7Merck 3 *4.660
20°C
6.880
9.220
8Merck 4 *2.000
20°C
4.000
7.000
10.000
64ba76141e0708/2018
Page 65
MultiLab 4010-1WSettings
No.Buffer set *pH valuesat
9Merck 5 *4.010
7.000
10.000
10DIN 192671.090
4.650
6.790
9.230
11Mettler Toledo USA *1.679
4.003
7.002
10.013
12Mettler Toledo EU *1.995
4.005
7.002
9.208
13Fisher *2.007
4.002
7.004
10.002
14Fluka BS *4.006
6.984
8.957
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
15Radiometer *1.678
4.005
7.000
9.180
16Baker *4.006
6.991
10.008
17Metrohm *3.996
7.003
8.999
18Beckman *4.005
7.005
10.013
19Hamilton Duracal *4.005
7.002
10.013
20Precisa *3.996
7.003
8.999
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
21Reagecon TEC *2.000
25 °C
4.010
7.000
10.000
ba76141e0708/201865
Page 66
SettingsMultiLab 4010-1W
No.Buffer set *pH valuesat
22Reagecon 20 *2.000
20°C
4.000
7.000
10.000
13.000
23Reagecon 25 *2.000
25 °C
4.000
7.000
10.000
13.000
24Chemsolute *2.000
20°C
4.000
7.000
10.000
25USABlueBook
*4.000
25 °C
7.000
10.000
* Brand names or trade names are tradema rks of their respective owners protected
by law.
Setting the calibra-
tion interval
The buffers are selected in the menu, pH /
/ Buffer (see section 9.1.1 S
ETTINGSFORPH MEASUREMENTS, page
<ENTER> / Calibration
63).
9.1.3Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.9 QSC
FUNCTION (SENSORQUALITYCONTROL), page
38).
After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
1.Open the menu for measurement settings with <ENTER>.
2.In the Calibration / Calibration interval menu, set the calibration interval
<><>.
with
66ba76141e0708/2018
Page 67
MultiLab 4010-1WSettings
3.Confirm the setting with
4.Quit the menu with
<ENTER>.
<M>.
9.2ORP measurement settings
9.2.1Settings for ORP measurements
The settings are made in the menu for measuring se ttings of the ORP measurement. To open the settings, display the required measured parameter in the
measured value display and press the
settings, switch to the measured value display with
Default settings are printed in bold.
Menu itemPossible settingExplanation
mV resolution0.1
1
Reset-Resets all sensor settings to the
<ENTER> key. After completing the
<M>.
Resolution of the mV display
delivery condition (see section
9.6.1 R
SETTINGS, page 75).
ESETTINGTHEMEASUREMENT
9.3D.O. measurement settings
9.3.1Settings for D.O. sensors
(menu for measurement and calibration settings)
SettingsThe settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the
settings, switch to the measured value display with
The possible settings are individually displayed for each sensor. Below, the setting menu is shown with all settings for all IDS-D.O. sensors.
Default settings are printed in bold.
Menu itemPossible
setting
Calibration /
Calibration record
Calibration /
Calibration data
storage / Display
-Displays the calibration record of
-Displays the last calibration records
<ENTER> key. After completing the
<M>.
Explanation
the last calibration.
(max. 10)
Calibration /
Calibration data
storage / Output to
RS232/USB
ba76141e0708/201867
-Outputs the calibration data storage to the interface
Page 68
SettingsMultiLab 4010-1W
Menu itemPossible
setting
Calibration / Zero
calibration(only for
4100 ProBOD, 5010
with 4011 AdapterProOBOD)
Calibration /
Calibration interval
Calibration /
Comparison meas.
Calibration / Sensor
cap coefficients
(only for ProOBOD)
1 ... 180 ...
999 d
on
off
K1
...
K5
KC
Explanation
Starts the 0-point calibration (see
section 7.3.6 Z
EROCALIBRATION,
page 52)
Calibration interval for the D.O.
sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Enables to adjust the measured
value with the aid of a comparison
measurement, e.g. Winkler titration.
For details, see section 7.3 C
BRATION, page 49.
ALI-
Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2
NTER SENSORCAPCOEFFICIENTS
E
ROOBOD), page 69.
(P
Calibration /
Membrane cap
(only for 4100 ProBOD, 5010 with
4011 Adapter)
FDO Check /
Start FDO Check
(only for FDO 4410)
FDO Check / Check
interval
(only for FDO 4410)
Salinity/Sal correction (only for param-
eter mg/l)
Salinity/Salinity
(only for parameter
mg/l)
Black
Yellow
After exchanging the sensor cap,
select the cap type here.
-Starts the check with the
FDO Check
1 ... 60 ...
999 d
Interval for the FDO Check (in
days).
The meter reminds you to check
the sensor regularly by FDO Check
status indicator in the measuring
screen.
on
off
Manual salt content correction for
concentration measurements.
0.0 ... 70.0Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if
the manual salinity correction is
switched on.
68ba76141e0708/2018
Page 69
MultiLab 4010-1WSettings
Menu itemPossible
setting
Resolution
(only for ProOBOD))
Response time t90
0.1
1
30 ... 300 sResponse time of the signal filter (in
(only for FDO 4410)
DO % Saturation
local
on
off
Explanation
Set a high or low resolution.
The setting of the resolution is separately stored for each measured
parameter.
seconds).
A signal filter in the sensor reduces
the limits of variation of the measured value. The signal filter is
characterized by the response time
t90. This is the time after which 90
% of a signal change is displayed.
DO % Saturation local is a procedure that takes the local air pressure into account for each
saturation measurement.
For details, see section 9.3.3 DO %
ATURATIONLOCAL, page 69
S
Reset-Resets all sensor settings to the
delivery condition (see section
9.6.1 R
SETTINGS, page 75).
ESETTINGTHEMEASUREMENT
9.3.2Enter Sensor cap coefficients (ProOBOD)
The values of the coefficients are provided with the sensor cap.
1.Change the digit of the highlighted position with
2.Go to the next position with
3.Confirm with
<ENTER> when a coefficient is completely entered.
< ><F2>/[ ].
<><>
.
9.3.3DO % Saturation local
Irrespective of the height or air pressure, the calibration value is set to 100 %.
The function DO % Saturation local fulfills the EU regulations for the parameter
oxygen saturation [%].
When the DO % Saturation local is enabled the display shows an [L] for the
parameter oxygen saturation.
D.O. mg/L readings are unaffected by the selection of the DO % Saturation local function.
ba76141e0708/201869
Page 70
SettingsMultiLab 4010-1W
9.4Cond measurement settings
9.4.1Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the measured value display and press the
tings, switch to the measured value display with
The possible settings are individually displayed for each sensor. Below the setting menu is displayed for two IDS sensors (4310, 4320).
Default settings are printed in bold.
<ENTER> key. After completing the set-
<M>.
General setting
menu for
conductivity
Menu itemPossible
Explanation
setting
Calibration /
Calibration record
Calibration /
Calibration data stor-
-Displays the calibration record of
the last calibration.
-Displays the last calibration
records (max. 10)
age / Display
Calibration /
Calibration data stor-
-Outputs the calibration data stor-
age to the interface
age / Output to
RS232/USB
Calibration /
Calibration interval
1 ... 150 ...
999 d
Calibration interval for the IDS
conductivity sensor (in days).
The meter reminds you to calibrate regularly by the flashing
sensor symbol in the measuring
screen.
Reset-Resets all sensor settings to the
delivery condition (see section
9.6.1 R
MENTSETTINGS, page 75).
ESETTINGTHEMEASURE-
70ba76141e0708/2018
Page 71
MultiLab 4010-1WSettings
Setting menu 4310
Menu itemPossible
Explanation
setting
TypeUsed measuring cell
CalMeasuring cells whose cell con-
stants are determined through
calibration in the check- and calibration standard 0.01 mol/l KCl
(1413µS/cm @25°C).
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
manFreely (manually) adjustable cell
constant in the range 0.450 ...
-1
.
Cell const. man0.450 ...
0.475 ...
0.500 cm
0.500 cm
Display and setting options for
the manually adjustable cell con-
-1
stant.
This menu item is only available
when Typeman is set.
The current conductivity value is
displayed in the status line.
Temp. comp. (TC) /
Method
nLF
Lin
off
Procedure for temperature compensation (see section 8.2 T
PERATURECOMPENSATION, page
57).
This setting is only available for
the measured parameters, conductivity (ϰ) and resistivity (ρ).
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear temperature compensation.
This menu item is only available
when the linear temperature
compensation is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for
the measured parameters, conductivity (ϰ) and resistivity (ρ).
Multiplier for TDS0.40 ... 1.00Multiplier for TDS value
EM-
ba76141e0708/201871
Page 72
SettingsMultiLab 4010-1W
Setting menu
4320
Menu itemPossible
setting
Cell constant0.090 ...
0.100 ...
0.110 cm
Temp. comp. (TC) /
Method
nLF
Lin
off
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Explanation
Display and setting options for the
cell constant.
-1
The current conductivity value is
displayed in the status line.
Procedure for temperature compensation (see section 8.2 T
PERATURECOMPENSATION, page
EM-
57).
This setting is only available for the
measured parameters, conductivity
(ϰ) and resistivity (ρ).
Coefficient of the linear temperature compensation.
This menu item is only available
when the linear temperature compensation is set.
Reference temperature
This setting is only available for the
measured parameters, conductivity
(ϰ) and resistivity (ρ).
Multiplier for TDS0.40 ... 1.00Multiplier for TDS value
72ba76141e0708/2018
Page 73
MultiLab 4010-1WSettings
9.5Sensor-independent settings
9.5.1System
To open the Storage & config menu, press the
sured value display. After completing the settings, switch to the measured
value display with
<M>.
Default settings are printed in bold.
Menu itemPossible settingExplanation
System / General / Language
System / General / Audio
signal
System / General / Illumi-
nation
English
(more)
on
off
Auto
on
Selects the menu language
Switching on/off the beep
Switches the display illumination on/off
off
System / General / Con-
0 ... 50 ... 100 Changes the display contrast
trast
System / General / Shutoff
time
System / General / Tem-
perature unit
10 min ... 1h ...
24 h
°C
°F
Adjusts the switch-off time
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed with the
selected unit.
<ENTER__> key in the mea-
System / General / Air
pressure unit
System / General / Stability control
System / Interface / Baud
rate
System / Interface / Output format
Only with: Output format
CSV:
System / Interface /
Decimal separator
System / Interface /
Output header
mbar
mmHg
inHg
on
off
1200, 2400,
4800, 9600,
19200
ASCII
CSV
Dot (xx.x)
Comma (xx,x)
Air pressure unit
Switches on or off the automatic stability control
during measurement (see section 9.5.3 A
TABILITYCONTROL, page 74 )
S
UTOMATIC
Baud rate of the USB Device interface
Output format for data transmission
For details, see section 11 T
RANSMITTINGDATA,
page 84
Decimal separator
Output of a header for Output format: CSV
ba76141e0708/201873
Page 74
SettingsMultiLab 4010-1W
Menu itemPossible settingExplanation
System / Clock functionDate format
Datum
Time
System / Service information
Settings of time and date.
For details, see section 4.4.5 E
TION: SETTINGTHEDATEANDTIME, page 24
XAMPLE 2 ONNAVIGA-
Hardware version and software version of the meter
are displayed.
System / Reset-Resets the system settings to the default values.
For details, see section 9.6.2 R
TEMSETTINGS, page 77
ESETTINGTHESYS-
9.5.2Data storage
This menu contains all functions to display, edit and erase stored measured
values.
Detailed information on the memory functions of the
MultiLab 4010-1W are given in section 10 D
ATASTORAGE, page 78.
9.5.3Automatic Stability control
The automatic Stability control function continuously checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
You can activate or switch off the automatic Stability control function (see
section 9.5 S
ENSOR-INDEPENDENTSETTINGS, page 73).
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
9.5.4Automatic switch-off function
The instrument has an automatic switch-off function in order to save the
batteries (see section 9.5.1 S
YSTEM, page 73). The automatic switchoff function
switches off the meter if no key is pressed for an adjustable period.
The automatic switchoff function is not active
if the power pack is connected
if a USB-B cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
74ba76141e0708/2018
Page 75
MultiLab 4010-1WSettings
9.5.5Display illumination
The meter automatically switches off the display illumination if no key is
pressed for 20 seconds.
The illumination is switched on with the next keystroke again.
You can also generally switch on the display illumination (see section 9.5.1
YSTEM, page 73).
S
9.6Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1 Resetting the measurement settings
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
pHThe following settings for pH measurements are reset to the default settings
with the Reset function:
SettingDefault settings
Buffer YSI
Calibration interval7 d
Unit for slopemV/pH
Measured parameterpH
Unit for slope0.001
mV resolution0.1
Asymmetry0 mV
Slope-59.2 mV
Man. temperature25 °C
Single-point calibrationoff
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
ba76141e0708/201875
Page 76
SettingsMultiLab 4010-1W
ORPThe following settings for ORP measurements are reset to the default settings
with the Reset function:
SettingDefault settings
mV resolution0.1
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
Dissolved oxygenThe following settings for D.O. measurements are reset to th e default settin gs
with the Reset function:
SettingDefault settings
Calibration interval180 d
Measured parameterD.O. concentration (mg/l)
Salinity (value)0.0
Salinity (function)off
Number of calibration points1
Resolution0.1
DO % Saturation localoff
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
ConductivityThe following settings for conductivity measurements are reset to the default
settings with the Reset function:
SettingDefault settings
Calibration interval150 d
Measured parameter
ϰ
Cell constant (c)Depending on the connected measur-
ing cell:
-1
0.475 cm
0.475 cm
0.100 cm
(calibrated)
-1
(set)
-1
Temperature compensationnLF
Reference temperature25 °C
Temperature coefficient (TC) of
2.000 %/K
the linear temperature compensation
TDS multiplier1.00
76ba76141e0708/2018
Page 77
MultiLab 4010-1WSettings
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
9.6.2Resetting the system settings
The following system settings can be reset to the default condition:
SettingDefault settings
LanguageEnglish
Audio signalon
Baud rate4800 Baud
Output formatASCII
Decimal separator.
Contrast50
IlluminationAuto
Shutoff time1 h
Temperature unit°C
Stability controlon
The resetting of the system settings is done in the menu Storage & config /
System / Reset. To open the menu Storage & config in the measured value
display, press the
<ENTER__> key.
ba76141e0708/201877
Page 78
Data storageMultiLab 4010-1W
03.04.2013 07:43:33
pH 7.000 24.8 °C AR +++
ID number:1
continue
Manual data storage 4 von 500
03.04.2013
08:00
Back
10Data storage
You can transmit measured values (datasets) to the data storage:
Manual data storage (see section 10.1 M
Automatic data storage at intervals (see section 10.2 A
AGEATINTERVALS, page 79)
ANUALDATASTORAGE, page 78)
UTOMATICDATASTOR-
Each data storing process transmits the current dataset to the USB interface.
10.1Manual data storage
You can transmit a measurement dataset to the data storage as follows. The
dataset is at the same time output to the interface:
1.Press the
The menu for manual data storage appears.
<STO> key shortly.
When the storage is
full
78ba76141e0708/2018
2.If necessary, change and confirm the ID number (1 ... 10000) with
<><> and <ENTER>.
The dataset is stored. The meter switches to the measured value display.
When all storage locations are occupied, it is not possible to continue storing.
Then you can e.g. transmit the stored data to a PC (see section 10.3.1 M
INGTHEMEASUREMENTDATASTORAGE, page 81) and subsequently erase the
data storage (see section 10.3.2 E
RASINGTHEMEASUREMENTDATASTORAGE,
page 82).
ANAG-
Page 79
MultiLab 4010-1WData storage
ID number 1
Interval 30 s
Duration 180 min
continue
0d03h00min
01d17h33min
Automatic data storage
1
2
3
03.04.2013
08:00
Back
10.2Automatic data storage at intervals
The storage interval (Interval) determines the time interval between automatic
data storage processes. Each data storing process transmits the current dataset to the USB interface.
Configuring the
automatic storing
function
1.Press the
<STO_ > key.
The menu for automatic data storage appears.
1 Specified entire storing duration
2 Max. available storage duration
3 Graphical display of the storage usage
SettingsYou can configure the automatic data storage function with the following set-
tings:
Menu itemPossible settingExplanation
ID number1 ... 10000ID number for the dataset series.
Interval1 s, 5 s, 10 s,
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the storage duration.
Duration1 min ... x minStorage duration.
Specifies after which time the automatic data storage should be terminated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
ba76141e0708/201879
Page 80
Data storageMultiLab 4010-1W
25.0
7.007
pH
°C
0
14
0d03h00min
1
2
03.04.2013
08:00
Stop automatic storage?
yes
no
Warning
03.04.2013
08:00
Starting the auto-
matic storing func-
tion
To start the automatic data storage function, select continue with <><>
and confirm with
play.
1 Remaining storing duration
2 Graphical display of the storage duration
The active automatic data storage function can be recognized by the progress
bar in the status line. The progress bar indicates the remaining storage duration.
<ENTER>. The meter switches to the measured value dis-
Terminating the
automatic storage
function prema-
turely
If the automatic data storage function is activated, only the following keys are active:
and the automatic switch-off function are deactivated.
Proceed as follows to switch off the automatic data storage function before the
adjusted storage duration has expired:
1.Press the <STO_ > key.
The following window appears.
<M><STO_ > and <On/Off>. The other keys
80ba76141e0708/2018
2.Using
The meter switches to the measured value display.
The automatic data storage function is terminated.
<><>, select yes and confirm with <ENTER>.
Page 81
MultiLab 4010-1WData storage
10.3Measurement data storage
10.3.1 Managing the measurement data storage
For each measurement data storage (automatic and manual) the following
functions are available:
Display
Output to RS232/USB
Erase
The storage is edited in the menu Storage & config/ Data storage. To open the
Storage & config menu, press the
display.
<ENTER__> key in the measured value
Settings
Open the manual or automatic data storage directly with the
<RCL__> key.
The settings are explained here using the manual data storage as
an example. The same settings and functions are available for the
automatic data storage.
Menu itemSetting/
function
Data storage /
Manual data storage
/ Display
-Displays all measurement data-
Explanation
sets page by page.
Further options:
Scroll through the datasets with
<>< >.
Output the displayed dataset to
the interface with
output].
<RCL> or
<F2>/[USB
Quit the display with
[Back].
Data storage /
Manual data storage
/ Output to RS232/
USB
Data storage /
Manual data storage
/ Erase
ba76141e0708/201881
-Outputs all stored measurement
data to the interface.
-Erases the entire manual measurement data storage.
Note:
All calibration data remain stored
when this action is performed.
Quitting the displayTo quit the display of stored measurement datasets, you have the following
options:
Switch directly to the measured value display with
Quit the display and move to the next higher menu level with
10.3.2 Erasing the measurement data storage
82ba76141e0708/2018
Erasing the measurement data storage (see section 10.3.1 M
MEASUREMENTDATASTORAGE, page 81).
10.3.3 Measurement dataset
A complete dataset consists of:
Date/time
Meter name, series number
Sensor name, series number
ID number
<M>.
<F1>/[Back].
ANAGINGTHE
Page 83
MultiLab 4010-1WData storage
Measured value of the connected sensor
Measured temperature value of the connected sensor
AutoRead info: AR appears with the measured value if the AutoRead crite-
rion was met while storing (stable measured value). Otherwise, the AR display is missing.
Calibration evaluation:
- 4 levels (+++, ++, +, -, or no evaluation) or
- QSC (percentage)
10.3.4 Storage locations
The MultiLab 4010-1W meter has two measurement data storages. The
measured values recorded either manually or automatic are stored separately
in individual measurement data storages.
Data storageMaximum number of datasets
Manual data storage500
Automatic data storage4500
ba76141e0708/201883
Page 84
Transmitting dataMultiLab 4010-1W
11Transmitting data
The meter has the following interfaces:
USB-B interface (USB Device)
e.g. to connect a PC
Via the USB-B interface (USB Device) you can transmit data to a PC or printer
and update the meter software.
11.1Transmitting data to a PC
Via the USB-B interface (USB Device) you can transmit data to a PC.
PC system require-
ments
Installation of the
USB driver
Connecting a PC
Microsoft Windows
(for details, see enclosed installation CD, Driver directory)
Installed USB driver for the meter (see CD-ROM or Internet)
Settings for the USB/RS232 interface on the PC and meter in agreement
Program to receive the measurement data on the PC
(e.g. MultiLab Importer, see CD-ROM or Internet)
1.Insert the supplied installation CD in the CD drive of your PC.
or
Download the USB driver from the Internet and unpack the files and
folders.
2.Start the driver installation (32 bits or 64 bits) suitable for your operating system.
Follow the Windows installation instructions as necessary.
1.Connect the MultiLab 4010-1W to the PC via the USB-B interface
(USB Device).
The meter is listed as a virtual COM interface among the connections
in the Windows instrument manager.
Adjusting the set-
tings for the data
transmission
84ba76141e0708/2018
2.Set the same transmission data at the meter and PC:
Baud rate: to be selected in the range 1200 ... 19200
Set at the PC only:
The following rule applies: With the exception of the menus, shortly
pressing the
<F2>/[USB output] key generally outputs the display
contents to the interface (displayed measured values, measuring
datasets, calibration records).
11.2MultiLab Importer
With the aid of the MultiLab Importer software, you can record and evaluate
measurement data with a PC.
More detailed information can be found in the MultiLab Importer
operating manual.
ba76141e0708/201885
Page 86
Transmitting dataMultiLab 4010-1W
11.3BOD Analyst Pro
With the aid of the BOD Analyst Pro software you can administrate BOD
measurement on the PC and automatically calculate the measured values.
More detailed information can be found in the BOD Analyst Pro
operating manual.
86ba76141e0708/2018
Page 87
MultiLab 4010-1WPrinter (only MultiLab 4010P-1W)
2
1
3
4
12Printer (only MultiLab 4010P-1W)
1 Operating LED
– Is illuminated: Printer is switched on
Switch on the printer
– Flashes: Printer is switched off
2 Control panel with the keys
3 Lever to open the paper compartment
4 Lid of the paper compartment
<FEED> and <ON/OFF>
12.1Commissioning / switching the printer on or off
1.Connect the power pack to the MultiLab 4010P-1W.
The LED lights up green. The printer is ready for operation.
or
When the printer had been switched off (LED flashing):
Switch on the printer with
The LED lights up green. The printer is ready for operation.
If there is a USB-B connection at the same time (e.g. to a PC), the
data are output to the PC only.
<ON/OFF>.
Switch printer off
ba76141e0708/201887
1.Switch off the printer with <ON/OFF>.
The LED flashes. The printer is switched off.
Page 88
Printer (only MultiLab 4010P-1W)MultiLab 4010-1W
12.2Operation / printing
Data are only output to the printer if the following requirements are met
Data are transmitted manually or automatically (see section 11 T
TINGDATA, page 84)
RANSMIT-
The printer is switched on (LED illuminated)
There is no USB connection.
12.3Printer settings
To open the Storage & config menu, press the <F1__>[Menü] key in the measured value display. After completing the settings, switch to the measured
value display with
Default settings are printed in bold.
Menu itemPossible set-
System / Printer /
Font size
<M>.
ting
12x20
8x16
7x16
Explanation
Select the font size for the printer
A print sample (font of the printer) to
view the available font sizes can be
output with
<ON/OFF_ >.
System / Printer /
Print test page
-The printer prints the meter information from the menu System / Ser-vice information.
The current printer settings are
used for the print.
12.4Maintenance
12.4.1 Changing the roll of paper (thermal paper)
1.Pull the lever (3) until the lid (4) of the paper compartment opens.
2.If necessary, remove any remains of the old roll of paper.
3.Insert the new roll of paper. The start of the paper should poke out of
the paper compartment.
4.Close the lid (4) of the paper compartment so that it clicks into place.
5.If necessary, move the printer paper forward with
<FEED>.
88ba76141e0708/2018
Page 89
MultiLab 4010-1WPrinter (only MultiLab 4010P-1W)
12.5What to do if... / printer
Integrated printer
does not print
CauseRemedy
– Printer switched off
(LED flashing)
– Switch on the printer
(LED illuminated)
– No power pack connected– Connect the power pack
– USB cable connected– Disconnect the USB cable from the
meter
Printer operating -
paper not being
printed
Integrated printer
prints automatically
– Function "store automatically at
intervals" with long intervals is
switched on
– Switch off the function (see section
10.2 A
INTERVALS, page 79)
UTOMATICDATASTORAGEAT
– No paper available– Insert a roll of paper
CauseRemedy
– Paper inserted the wrong way up– Turn the roll of paper around and
insert it with the other side upwards
CauseRemedy
– The function "store automatically
at intervals" or "transmit data
automatically at intervals" is
switched on
– Switch off the functions (see sec-
tion 10.2 A
ATINTERVALS, page 79 or
section 11 T
UTOMATICDATASTORAGE
RANSMITTINGDATA,
page 84)
ba76141e0708/201889
Page 90
Maintenance, cleaning, disposalMultiLab 4010-1W
1
13Maintenance, cleaning, disposal
13.1Maintenance
13.1.1 General maintenance activities
The only maintenance activity required is replacing the batteries.
See the relevant operating manuals of the IDS sensors for instructions on maintenance.
13.1.2 Replacing the batteries
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1.Open the battery compartment (1) on the underside of the
meter.
1 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The
±signs on the batteries must correspond to the ± signs in
the battery compartment.
2.Place four batteries (type Mignon AA) in the battery compart-
ment.
3.Close the battery compartment (1).
90ba76141e0708/2018
Page 91
MultiLab 4010-1WMaintenance, cleaning, disposal
4.Set the date and time
(see section 4.4.5 EXAMPLE 2 ONNAVIGATION: SETTINGTHE
DATEANDTIME, page 24).
Dispose of used batteries according to the local regulations of your
country.
End users within the European Union are obligated to return used
batteries
(even ecologically compatible ones) to a collection point set up for
recycling
purposes.
Batteries are marked with the crossed-out waste container
symbol. Therefore, they may not be disposed with the domestic
waste.
13.2Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lintfree cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain solvents. Remove any splashes immediately.
13.3Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
meter against damage during transport.
13.4Disposal
At the end of its operational lifetime, the meter must be returned to the
disposal or return system statutory in your country. If you have any questions,
please contact your supplier.
ba76141e0708/201891
Page 92
What to do if...MultiLab 4010-1W
14What to do if...
14.1pH
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
The measured value is outside the measuring range.
CauseRemedy
IDS pH sensor:
– The measured value is outside
the measuring range of the
meter
– Air bubble in front of the junction– Remove air bubble
– Cable broken– Replace the sensor
– Gel electrolyte dried out– Replace the sensor
CauseRemedy
IDS pH sensor:
– The values determined for zero
point and slope of the IDS pH
sensor are outside the allowed
limits.
– Use suitable IDS pH sensor
(e.g. sway or stir the solution)
–Recalibrate
– Sensor soiled– Clean the sensor
– Sensor broken– Replace the sensor
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 0.5 A
Output: 9 Vdc, 1100 mA
Connection max. overvoltage category II
ShenZhen RiHuiDa Power Supply Co Ltd
RHD10W090110
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 0.4 A
Output: 9 Vdc, 1100 mA
delivery: Euro, US, UK and Australian.
96ba76141e0708/2018
Page 97
MultiLab 4010-1WTechnical data
USB interface (device)TypeUSB 1.1
USB-B (Device), PC
Baud rateAdjustable:
1200, 2400, 4800, 9600, 19200 Baud
Data bits8
Stop bits2
ParityNone
HandshakeRTS/CTS
Cable length max. 3 m (9.843 feet)
Guidelines
and norms used
EMCEC directive 2014/30/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safetyEC directive 2014/35/EC
EN 61010-1
IP protection classEN 60529
ba76141e0708/201897
Page 98
Firmware updateMultiLab 4010-1W
16Firmware update
16.1Firmware update for the meter MultiLab 4010-1W
You can find available firmware update files for your meter on the Internet. With
the "Firmware Update " program and a PC you can update the firmware of the
MultiLab 4010-1W to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB-B interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the MultiLab 4010-1W).
1.Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type), the new
data are displayed there.
2.In the windows start menu, open the update folder and start the firmware update program for the meter
3.Using the USB interface cable, connect the MultiLab 4010-1W to a
USB interface (virtual COM port) of the PC.
4.Switch on the MultiLab 4010-1W.
5.In the firmware update program, start the update process with OK.
6.Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes up to 15 minutes. A terminatory message is displayed after a successful programming process. The firmware update is completed.
7.Disconnect the MultiLab 4010-1W from the PC.
The MultiLab 4010-1W is ready for operation again.
After switching the meter off and on you can check whether the meter has taken
over the new software version (see Y
VERSIONISINTHEMETEROR IDS SENSOR, PAGE 95).
OUWANTTOKNOWWHICHSOFTWARE
98ba76141e0708/2018
Page 99
MultiLab 4010-1WFirmware update
16.2Firmware-Update for IDS Sensors
With the "Firmware Update" program and a PC you can update the firmware of
an IDS sensor to the newest version.
You can find available firmware update files for your IDS sensor on the Internet.
For updating, use cables to connect the IDS sensor to the MultiLab 4010-1W,
and the MultiLab 4010-1W to a PC.
For the update via the USB-B interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the MultiLab 4010-1W).
1.Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the sensor (or sensor type), the
new data are displayed there.
2.In the windows start menu, open the update folder and start the firmware update program for the IDS sensor
3.Connect the IDS sensor to the MultiLab 4010-1W meter.
4.Using the USB interface cable, connect the MultiLab 4010-1W to a
USB interface (virtual COM port) of the PC.
5.Switch on the MultiLab 4010-1W.
6.In the firmware update program, start the update process with OK.
7.Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes up to 5 minutes. A terminatory message is displayed after a successful programming process. The firmware update is completed.
8.Disconnect the MultiLab 4010-1W from the PC.
Meter and sensor are ready for operation again.
After switching the meter off and on you can check whether the sensor has
taken over the new software version (see Y
VERSIONISINTHEMETEROR IDS SENSOR, PAGE 95).
OUWANTTOKNOWWHICHSOFTWARE
ba76141e0708/201899
Page 100
GlossaryMultiLab 4010-1W
17Glossary
pH/ORP
Asymmetrysee zero point
Electromotive force of
an electrode
JunctionThe junction is a porous body in the housing wall of reference elec-
ORP
(U)
pH valueThe pH value is a measure of the acidic or basic effect of an aqueous
PotentiometryName of a measuring technique. The signal (depending on the mea-
SlopeThe slope of a linear calibration function.
The electromotive force U of the combination electrode is the measurable electromotive force of an electrode in a solution. It equals the sum
of all the galvanic voltages of the combination electrode. Its dependency on the pH results in the electrode function, which is characterized
by the parameters, slope and zero point.
trodes or electrolyte bridges. It arranges the electrical contact between
two solutions and makes the electrolyte exchange more difficult. The
expression, junction, is also used for ground or junction-less transitions.
The ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface (e.
g. a gold or platinum surface).
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
Zero pointThe zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified
temperature is zero. Normally, this is at 25 °C.
Conductivity
Cell constant (c)Characteristic quantity of a conductivity measuring cell, depending on
the geometry.
Conductivity (ϰ)Short form of the expression, specific electrical conductivity.
It corresponds to the reciprocal value of the resistivity.
It is a measured value of the ability of a substance to conduct an electric
current. In water analysis, the electrical conductivity is a dimension for
the ionized substances in a solution.
Reference tempera-
ture
Resistivity (ρ)Short name for the specific electrolytic resistance. It corresponds to the
Fixed temperature value to compare temperature-dependent measured
values. For conductivity measurements, the measured value is converted to a conductivity value at a reference temperature of 20 °C or 25
°C.
reciprocal value of the electrical conductivity.
100ba76141e0708/2018
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.