YSI 4010-1W Operating Manual

Page 1
OPERATING MANUAL
4010P-1W
4010-1W
4010-1W
IDS WA
IDS WA
ba76141e07 08/2018
MultiLab 4010(P)-1(W)
DIGITAL METER FOR (WIRELESS) IDS SENSORS
Page 2
MultiLab 4010-1W
For the most recent version of the manual, please visit
www.ysi.com
.
Contact YSI
1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241
800-765-4974 Email: info@ysi.com Internet: www.ysi.com
Copyright © 2018 Xylem Inc.
2 ba76141e07 08/2018
Page 3
MultiLab 4010-1W Contents
Contents
1 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 MultiLab 4010-1W . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 MultiLab 4010P-1W meter with integrated printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3.1 IDS sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3.2 Wireless operation of IDS sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.3 Automatic sensor recognition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.1 Safety information in the operating manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.2 Safety signs on the meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.3 Further documents providing safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 Safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.2 Requirements for safe operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.3 Unauthorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3.1 Inserting the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3.2 Connecting the power pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3.3 Mounting the stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1 General operating principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1.3 Status information (meter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1.4 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.1.5 Sensor info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2 Switching on. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3 Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.4 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.4.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.4.2 Measured value display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.4.3 Menus and dialogs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.4.4 Navigation example 1: Setting the language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
ba76141e07 08/2018 3
Page 4
Contents MultiLab 4010-1W
4.4.5 Example 2 on navigation: Setting the date and time . . . . . . . . . . . . . . . . . . . . . . . . . 24
5 pH value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.1.1 Measuring the pH value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.1.2 Measuring the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.2 pH calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2.1 Why calibrate?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2.2 When do you have to calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2.3 Calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2.4 Carrying out automatic calibration (AutoCal) . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2.5 Carrying out manual calibration (ConCal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2.6 Calibration points. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.7 Calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.8 Continuous measurement control (CMC function) . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2.9 QSC function (sensor quality control) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6 ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.1.1 Measuring the ORP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.1.2 Measuring the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2 ORP calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7 Dissolved oxygen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.1.1 Measuring D.O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.1.2 Measuring the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.2 FDO Check (checking the FDO 4410) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2.1 Why should you check the sensor?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2.2 When should you check the sensor?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2.3 Perform a FDO Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2.4 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.3 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3.1 Why calibrate?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3.2 When to calibrate?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3.3 Calibration procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3.4 Calibration in water vapor-saturated air. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3.5 Calibrating with Comparison meas. (e.g. Winkler titration) . . . . . . . . . . . . . . . . . . . . 50
7.3.6 Zero calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.3.7 Calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.1.1 Measuring the conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.1.2 Measuring the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.2 Temperature compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.3 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.3.1 Why calibrate?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.3.2 When to calibrate?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.3.3 Calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.3.4 Determining the cell constant (calibration in the check- and calibration standard) . . 58
4 ba76141e07 08/2018
Page 5
MultiLab 4010-1W Contents
8.3.5 Setting the cell constant
(calibration with freely selectable check- and calibration standard). . . . . . . . . . . . . . 60
8.3.6 Calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
9 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.1 pH measurement settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.1.1 Settings for pH measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.1.2 Buffer sets for calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
9.1.3 Calibration interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
9.2 ORP measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
9.2.1 Settings for ORP measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
9.3 D.O. measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
9.3.1 Settings for D.O. sensors
(menu for measurement and calibration settings) . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
9.3.2 Enter Sensor cap coefficients (ProOBOD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
9.3.3 DO % Saturation local . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
9.4 Cond measurement settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
9.4.1 Settings for IDS conductivity sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
9.5 Sensor-independent settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.5.1 System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.5.2 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.5.3 Automatic Stability control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.5.4 Automatic switch-off function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.5.5 Display illumination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.6 Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.6.1 Resetting the measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.6.2 Resetting the system settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
10 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
10.1 Manual data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
10.2 Automatic data storage at intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
10.3 Measurement data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
10.3.1 Managing the measurement data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
10.3.2 Erasing the measurement data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
10.3.3 Measurement dataset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
10.3.4 Storage locations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
11 Transmitting data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
11.1 Transmitting data to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
11.2 MultiLab Importer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
11.3 BOD Analyst Pro. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
12 Printer (only MultiLab 4010P-1W) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
12.1 Commissioning / switching the printer on or off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
12.2 Operation / printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
12.3 Printer settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
12.4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
12.4.1 Changing the roll of paper (thermal paper) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
12.5 What to do if... / printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
ba76141e07 08/2018 5
Page 6
Contents MultiLab 4010-1W
13 Maintenance, cleaning, disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
13.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
13.1.1 General maintenance activities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
13.1.2 Replacing the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
13.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
13.3 Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
13.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
14 What to do if.... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
14.1 pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
14.2 Dissolved oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
14.3 Conductivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
14.4 General topics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
15 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
15.1 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
15.2 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
16 Firmware update. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
16.1 Firmware update for the meter MultiLab 4010-1W . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
16.2 Firmware-Update for IDS Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
17 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
18 Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
19 Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
19.1 Oxygen solubility table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
19.2 DO% Calibration values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
19.3 Calculate the TDS Multiplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
20 Contact Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
20.1 Ordering & Technical Support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
20.2 Service Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
6 ba76141e07 08/2018
Page 7
MultiLab 4010-1W Overview
1
2
3
4010-1W

1 Overview

1.1 MultiLab 4010-1W

The compact, digital precision meter MultiLab 4010-1W enables you to carry out pH measurements, ORP measurements, conductivity measurements and dissolved oxygen (D.O.) measurements quickly and reliably.
The MultiLab 4010-1W provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The MultiLab 4010-1W supports you in your work with the following functions:
Automatic sensor recognition CMC (continuous measurement control)QSC (sensor quality control)Electronic access controlData transmission via the USB interface (USB-B).
1 Keypad 2 Display 3 Connectors
ba76141e07 08/2018 7
Page 8
Overview MultiLab 4010-1W
1
2
3
4
5
4010 -1WP

1.2 MultiLab 4010P-1W meter with integrated printer

The integrated printer of the MultiLab 4010P-1W enables to document mea­surements according to GLP requirements.
The information concerning the printer of the MultiLab 4010P-1W is given in a separate chapter (see section 12 P M
ULTILAB 4010P-1W), page 87).
RINTER (ONLY
1 Keypad 2 Display 3 Connectors 4Printer 5 Control panel of the printer

1.3 Sensors

A measuring system ready to measure consists of the MultiLab 4010-1W meter and a suitable sensor.
Suitable sensors are IDS pH sensors, IDS ORP sensors, IDS conductivity sensors and IDS D.O. sensors.

1.3.1 IDS sensors

IDS sensors
support the automatic sensor recognitionshow only the settings relevant to the specific sensor in the setting menuprocess signals in the sensor digitally so that precise and interference-free
measurements are enabled even with long cables
facilitate to assign a sensor to a measured parameter with differently colored
couplings
8 ba76141e07 08/2018
Page 9
MultiLab 4010-1W Overview
have quick-lock couplings with which to fix the sensors to the meter.
Sensor data from
IDS sensors
IDS sensors transmit the following sensor data to the meter: SENSOR ID
– Sensor name – Sensor series number
Calibration dataMeasurement settings
The calibration data are updated in the IDS sensor after e ach calibration proce­dure. A message is displayed while the data are being updated in the sensor.
In the measured value display, you can display the sensor name and series number of the selected sensor with the [Info] softkey. You can then display further sensor data stored in the sensor with the [More] softkey (see section 4.1.5 S
ENSOR INFO, page 18).

1.3.2 Wireless operation of IDS sensors

With the aid of the adapters in the IDS WA Kit, IDS sensors with plug head connectors (variant P) can be wirelessly connected to your MultiLab 4010-1W. (MultiLab 4010-1 IDS: from software version V2.00)
Two adapters, one at the IDS meter (IDS WA-M) and one at the sensor (IDS WA-S), replace the sensor cable with an energy-saving Bluetooth LE radio connection.
Further information on the wireless operation of IDS sensors:
Web resources Operating manual of the IDS WA Kit.

1.3.3 Automatic sensor recognition

The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibratingto assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor serial number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to hide menus automatically that do not concern this sensor
ba76141e07 08/2018 9
Page 10
Overview MultiLab 4010-1W
To be able to use the automatic sensor recognition, a meter that supports the automatic sensor recognition (e.g. MultiLab 4010-1W) and a digital IDS sensor are required.
In digital IDS sensors, sensor data are stored that clearly identify the sensor. The sensor data are automatically taken over by the meter.
10 ba76141e07 08/2018
Page 11
MultiLab 4010-1W Safety

2 S afety

2.1 Safety information

2.1.1 Safety information in the operating manual

This operating manual provides important information on the safe operation of the meter. Read this operating manual thoroughly and make yourself familiar with the meter before putting it into operation or working with it. The operating manual must be kept in the vicinity of the meter so you can always find the infor­mation you need.
Important safety instructions are highlighted in this operating manual. They are indicated by the warning symbol (triangle) in the left column. The signal word (e.g. "Caution") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to seri­ous (irreversible) injury or death if the safety instruction is not followed.
CAUTION
indicates a possibly dangerous situation that can lead to slight (reversible) injury if the safety instruction is not fol­lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged if the actions mentioned are not taken.

2.1.2 Safety signs on the meter

Note all labels, information signs and safety symbols on the meter and in the battery compartment. A warning symbol (triangle) without text refers to safety information in this operating manual.

2.1.3 Further documents providing safety information

The following documents provide additional information, which you should observe for your safety when working with the measuring system:
Operating manuals of sensors and other accessories
Safety datasheets of calib ration or maintenance accessories (such as buffer solutions, electrolyte solutions, etc.)
ba76141e07 08/2018 11
Page 12
Safety MultiLab 4010-1W

2.2 Safe operation

2.2.1 Authorized use

The authorized use of the meter consists exclusively of the measurement of the pH, ORP, conductivity and dissolved oxygen in a laboratory environment.
Only the operation and running of the meter according to the instructions and technical specifications given in this operating manual is authorized (see section 15 T
96).
Any other use is considered unauthorized.

2.2.2 Requirements for safe operation

Note the following points for safe operation:
The meter may only be operated accord ing to the authorized use specified above.
The meter may only be supplied with power by the energy sources mentioned in this operating manual.
The meter may only be operated under the environmental conditions mentioned in this operating manual.
The meter may only be opened if this is explicitly described in this operating manual (example: Inserting the batteries).
ECHNICAL DATA, page

2.2.3 Unauthorized use

The meter must not be put into operation if:
it is visibly damaged (e.g. after being transported)
it was stored under adverse conditions for a lengthy period of time (storing conditions, see section 15 T
ECHNICAL DATA, page 96).
12 ba76141e07 08/2018
Page 13
MultiLab 4010-1W Commissioning

3 C ommissioning

3.1 Scope of delivery

MultiLab 4010-1W / MultiLab 4010P-1W4 batteries 1.5 V Mignon type AAPower packUSB cable (A plug on mini B plug)StandStand holderComprehensive operating manualShort instructions CD-ROM with
– USB drivers – detailed operating manual – Software MultiLab Importer

3.2 Power supply

The MultiLab 4010-1W is supplied with power in the following ways: Line power operation via the supplied power pack
A sensor with stirrer can only be operated with line power.
Battery operation (4 batteries, 1.5 V Mignon type AA)USB operation via a connected USB-B cable

3.3 Initial commissioning

Perform the following activities:
Insert the supplied batteriesFor mains operation: Connect the power packIf necessary, mount the standSwitch on the meter (see
section 4.2 S
WITCHING ON, page 19)
Set the date and time (see
section 4.4.5 E
24)
ba76141e07 08/2018 13
XAMPLE 2 ON NAVIGATION: SETTING THE DATE AND TIME, page
Page 14
Commissioning MultiLab 4010-1W
1

3.3.1 Inserting the batteries

You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH). In order to charge the batteries, an external charging device is required.
1. Open the battery compartment (1) on the underside of the
meter.
1 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned cor­rectly.
± signs on the batteries must correspond to the ± signs in
The the battery compartment.
2. Place four batteries (type Mignon AA) in the battery compartment.
3. Close the battery compartment (1).
4. Set the date and time
(see section 4.4.5 E page 24).
XAMPLE 2 ON NAVIGATION: SETTING THE DATE AND TIME,
14 ba76141e07 08/2018
Page 15
MultiLab 4010-1W Commissioning
1
2

3.3.2 Connecting the power pack

CAUTION
The line voltage at the operating site must lie within the input voltage range of the original power pack (see section 15.2
G
ENERAL DATA, page 96).
CAUTION
Use original power packs only (see section 15.2 GENERAL DATA, page 96).
1. Connect the plug of the power pack to the socket for the power pack on
the MultiLab 4010-1W.
2. Connect the original power pack to an easily accessible power outlet.

3.3.3 Mounting the stand

The stand base can be mounted at the right side of the meter.
ba76141e07 08/2018 15
Page 16
Operation MultiLab 4010-1W

4Operation

4.1 General operating principles

4.1.1 Keypad

In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. and release) in this operating manual. A long keystroke (press and keep depressed for approx. 2 sec) is indicated by the underscore behind the key symbol (e.g.
<ENTER>) generally indicates a short keystroke (press
<ENTER__>).
<F1>: <F1__>:
Softkeys providing situation dependent functions, e.g.:
<F1>/[
i
]: View information on a sensor
<F2>: <F2__>:
<On/Off>: Switches the meter on or off
<M>: Selects the measured parameter / Quits the settings
<CAL>: <CAL__>:
<STO>: <STO_ >:
<RCL>: <RCL__>:
<><>: <__><__>:
<ENTER>: <ENTER__>:
Calls up the calibration procedure Displays the calibration data
Saves a measured value manually Opens the menu for the automatic save function
Displays the manually stored measured values Displays the automatically stored measured values
Menu control, navigation Increments, decrements values Increments, decrements values continuously
Opens the menu for measurement settings / confirms entries Opens the menu for system settings
<AR> Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
16 ba76141e07 08/2018
Page 17
MultiLab 4010-1W Operation
3
2
1
4
5
6
8
HOLD AR
7
AutoCal TEC
i
03.04.2013 08:00
USB output
YSI

4.1.2 Display

Example
pH
1 Status information (sensor) 2 Measured value 3 Measured parameter 4 Continuous measurement control (CMC function) 5 Sensor symbol (calibration evaluation, calibration interval) 6 Measured temperature (with unit) 7 Status information (meter) 8 Softkeys and date + time

4.1.3 Status information (meter)

AR Stability control (AutoRead) is active HOLD Measured value is frozen (
<AR> key)
ZeroCal The zero point is calibrated.
Batteries are almost empty
Data are automatically output to the USB-B interface at intervals
ba76141e07 08/2018 17
Page 18
Operation MultiLab 4010-1W
1
3
5
4
2
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output

4.1.4 Socket field

1 IDS sensors: (pH, ORP, conductivity, D.O.) 2 USB B (device) interface 3 Stirrer (interface for BOD sensor) 4 Power pack 5 Service interface
CAUTION
Only connect sensors to the meter that cannot return any volt­ages or currents that are not allowed (> SELV and > current circuit with current limiting). YSI IDS sensors and IDS adapters meet these requirements.

4.1.5 Sensor info

You can display the current sensor data and sensor settings of a connected sensor at any time. The sensor data are available in the measured value display with the
<F1>/[Info] softkey.
1. In the measured value display: Display the sensor data (sensor name, series number) with [
18 ba76141e07 08/2018
<F1>Info].
Page 19
MultiLab 4010-1W Operation
25.0
7.007
pH
°C
0
14
More
4110
B09250001
pH resolution 0.001 mV resolution 0.1 Buffer YSI Calibration interval 7d Unit for slope mV/pH QSC: off Software version V 1.00
4110 B092500013
03.04.2013 08:00
Back
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output
2. Display further sensor data (settings) with <F1>/[More].
ba76141e07 08/2018 19

4.2 Switching on

1. Switch the meter on with The meter performs a self-test.
2. Connect the sensor. The meter is ready to measure.

4.3 Switching off

1. Switch the meter off with
<On/Off>.
<On/Off>.
Page 20
Operation MultiLab 4010-1W

4.4 Navigation

4.4.1 Operating modes

Operating mode
Measuring The measurement data of the connected sensor are shown
Calibration The course of a calibration with calibration information, func-
Storing in memory
Transmit­ting data
Setting The system menu or a sensor menu with submenus, set-

4.4.2 Measured value display

In the measured value display, you can open the menu for calibration and measurement settings with
(short
keystroke)
Explanation
in the measured value display
tions and settings is displayed The meter stores measuring data automatically or manually
The meter transmits measuring data and calibration records to a USB-B interface automatically or manually.
tings and functions is displayed
<ENTER>
open the system menu with the sensor-independent settings by pressing
<ENTER>Storage & config for a <ENTER__>long keystroke, approx.
2s).
change the display in the selected measuring screen (e. g. pH ><− mV) by
pressing

4.4.3 Menus and dialogs

The menus for settings and dialogs in procedures contain further sube lements. The selection is done with the played with a frame.
Submenus
The name of the submenu is displayed at the upper edge of the frame. Su b­menus are opened by confirming with
<M>.
<><> keys. The current selection is dis-
<ENTER>. Example:
20 ba76141e07 08/2018
Page 21
MultiLab 4010-1W Operation
General Interface Clock function Service information Reset
System
03.04.2013 08:00
Back
Language: Deutsch Audio signal: off Illumination: on Contrast: 12 Shutoff time: 1 h Temperature unit: °C Stability control
: on
General
03.04.2013 08:00
Back
Calibration record Calibration data storage Buffer: YSI Single-point calibration: yes Calibration interval: 7 d Unit for slope: mV/pH
[
i
] 4.00 7.00 10.00 (25 °C)
pH
03.04.2013 08:00
Back
Settings
Settings are indicated by a colon. The current setting is displayed on the right-hand side. The setting mode is opened with the setting can be changed with
<><> and <ENTER>. Example:
<ENTER>. Subsequently,
ba76141e07 08/2018 21
Functions
Functions are designated by the name of the function. They are immediately carried out by confirming with
<ENTER>. Example: Display the Calibration
record function.
Messages
i
Information is marked by the [
] symbol. It cannot be selected. Example:
Page 22
Operation MultiLab 4010-1W
Calibration record Calibration data storage Buffer: YSI Single-point calibration: yes Calibration interval: 7 d Unit for slope: mV/pH
[
i
] 4.00 7.00 10.00 (25 °C)
pH
03.04.2013 08:00
Back
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output
System Data storage
Storage & config
03.04.2013 08:00
Back

4.4.4 Navigation example 1: Setting the language

1. Press the
<On/Off> key.
The measured value display appears. The instrument is in the measuring mode.
2. Open the Storage & config menu with The instrument is in the setting mode.
<ENTER__>.
22 ba76141e07 08/2018
3. Select the System submenu with
<><>.
The current selection is displayed with a frame.
4. Open the System submenu with
<ENTER>.
Page 23
MultiLab 4010-1W Operation
General Interface Clock function Service information Reset
System
03.04.2013 08:00
Back
Language: Deutsch Audio signal: off Illumination: on Contrast: 50 % Shutoff time: 1 h Temperature unit: °C Stability control
: on
General
03.04.2013 08:00
Back
Language: Deutsch Audio signal: off Illumination: on Contrast: 50 % Shutoff time: 1 h Temperature unit: °C Stability control
: on
General
03.04.2013 08:00
Back
5. Select the General submenu with <><>.
The current selection is displayed with a frame.
6. Open the General
submenu with <ENTER>.
7. Open the setting mode for the Language with
<ENTER>.
ba76141e07 08/2018 23
8. Select the required language with
9. Confirm the setting with
<ENTER>.
The meter switches to the measuring mode. The selected language is active.
<><>.
Page 24
Operation MultiLab 4010-1W
Date format: dd.mm.yy Date: 03.04.2013 Time: 14:53:40
Clock function
03.04.2013 08:00
Back

4.4.5 Example 2 on navigation: Setting the date and time

The meter has a clock with a date function. The date and time are indicated in the status line of the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well.
The correct setting of the date and time and date format is important for the fol­lowing functions and displays:
Current date and timeCalibration dateIdentification of stored measured values.
Therefore, check the time at regular intervals.
The date and time are reset to default after a fall of the supply volt­age (empty batteries).
Setting the date,
time and date format
The date format can be switched from the display of day, month, year (dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1. In the measured value display: Open the Storage & config menu with
<ENTER__>.
The instrument is in the setting mode.
2. Select and confirm the System / Clock function menu with
<><> and
<ENTER>.
The setting menu for the date and time opens up.
24 ba76141e07 08/2018
3. Select and confirm the Time menu with The hours are highlighted.
4. Change and confirm the setting with
<><> and <ENTER>.
The minutes are highlighted.
5. Change and confirm the setting with
<><> and <ENTER>.
The seconds are highlighted.
<><> and <ENTER>.
Page 25
MultiLab 4010-1W Operation
6. Change and confirm the setting with <><> and <ENTER>. The time is set.
7. If necessary, set the Date and Date format. The setting is made similarly to that of the time.
8. To make further settings, switch to the next higher menu level with or Switch to the measured value display with
<M>.
The instrument is in the measuring mode.
<F1>.
ba76141e07 08/2018 25
Page 26
pH value MultiLab 4010-1W
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output

5 pH value

5.1 Measuring

5.1.1 Measuring the pH value

NOTE
When connecting a grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated.
1. Connect the IDS pH sensor to the meter. The pH measuring window is displayed.
Stability control
(AutoRead)
& HOLD function
2. If necessary, select the measured parameter with
3. Adjust the temperature of the solutions and measure the current tem­perature if the measurement is made without a temperature sensor.
4. If necessary, calibrate or check the IDS pH sensor.
5. Immerse the IDS pH sensor in the test sample.
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproduc­ibility of measured values.
<M>.
26 ba76141e07 08/2018
Hold function
The measured parameter flashes on the display
as soon as the measured value is outside the stability range when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the set­ting for automatic Stability control (see section 9.5.3 A
TROL, page 74) in the System menu.
To start the Stability control function manually, the HOLD function must be enabled.
1. Freeze the measured value with <AR>. The [HOLD] status indicator is displayed. The HOLD function is active.
UTOMATIC STABILITY CON-
Page 27
MultiLab 4010-1W pH value
You can terminate the HOLD function and the Stability control func- tion with
<AR> or <M> at any time.
Stability control
2. Using <ENTER>, activate the Stability control function manually. The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the mea­sured parameter flashes. The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears, the display of the mea­sured parameter stops flashing, and there is a beep. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu­ally with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
3. Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
ENSOR-INDEPENDENT
or Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the previous indication.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable within the monitored time interval.
Measured parameter
pH value 15 seconds Temperature 15 seconds
Time interval Stability in the time inter-
val
: better than 0.01 pH : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.

5.1.2 Measuring the temperature

For reproducible pH measurements, it is essential to measure the temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in the IDS sensor.
ba76141e07 08/2018 27
Page 28
pH value MultiLab 4010-1W

5.2 pH calibration

5.2.1 Why calibrate?

pH electrodes age. This changes the zero point (asymmetry) and slope of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines and stores the current values of the zero point and slope of the electrode. Thus, you should calibrate at regular intervals.

5.2.2 When do you have to calibrate?

Routinely within the framework of the company quality assuranceWhen the calibration interval has expired

5.2.3 Calibration procedure

The MultiLab 4010-1W provides 2 calibration procedures: Automatic calibration (AutoCal)
The buffer solutions used are automatically identified during the calibration procedure. The relevant nominal values are used automatically.
Prerequisite: The buffer set used has been set in the meter (see section
9.1.2 B
Manual calibration (ConCal)
Any buffer solutions can be used. The nominal values of the buffer solutions are manually entered by the user during the calibration procedure.

5.2.4 Carrying out automatic calibration (AutoCal)

Make sure that, in the menu for measurement and calibration settings (in the menu pH/ (see section 9.1.1 S
UFFER SETS FOR CALIBRATION, page 64).
For most applications, automatic calibration (AutoCal) is the safest and most convenient way of executing a calibration.
<ENTER>/Calibration / Buffer), the correct buffer set is selected
ETTINGS FOR PH MEASUREMENTS, page 63).
Use one to five buffer solutions of the selected buffer set in any order. Below, calibration with YSI buffers (YSI) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the proce­dure is identical.
If single-point calibration was set in the menu, the calibration proce­dure is automatically finished with the measurement of buffer solu­tion 1 and the calibration record is displayed.
28 ba76141e07 08/2018
Page 29
MultiLab 4010-1W pH value
pH 1
-180.0
24.8 °C
AutoCal YSI
01.07.2011 08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal YSI
10.000
01.07.2011 08:00
AR
Buffer
mV
1. Connect the pH sensor to the meter. The pH measuring window is displayed.
2. Keep the buffer solutions ready.
3. Start the calibration with The calibration display for the first buffer appears (voltage display).
4. Thoroughly rinse the sensor with deionized water.
5. Immerse the sensor in the first buffer solution.
6. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
<CAL>.
<ENTER>.
ba76141e07 08/2018 29
7. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
8. If necessary, finish the calibration procedure as a single-point calibra­tion with The calibration record is displayed.
<M>.
For single-point calibration, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the zero point of the IDS­pH sensor.
Page 30
pH value MultiLab 4010-1W
pH
0.0
24.8 °C
AutoCal YSI
7.000
01.07.2011 08:00
AR
Buffer
mV
pH
180.0
24.8 °C
AutoCal YSI
4.000
01.07.2011 08:00
AR
Buffer
mV
Continuing with two-
point calibration
9. Thoroughly rinse the sensor with deionized water.
10. Immerse the pH sensor in buffer solution 2.
11. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
12. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
<ENTER>.
The calibration display for the next buffer appears (voltage display).
Continuing with
three- to five-point
calibration
1 3. If necessary, finish the calibration procedure as a two-point calibration
<M>.
with The calibration record is displayed.
14. Thoroughly rinse the sensor with deionized water.
15. Immerse the sensor in the next buffer solution.
16. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
<ENTER>.
30 ba76141e07 08/2018
Page 31
MultiLab 4010-1W pH value
pH
0.0
24.8 °C
ConCal
1
01.07.2011 08:00
Buffer
mV
17. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
Calibration is automatically completed after the last buffer of a buf­fer set has been measured. Then the calibration record is dis­played.
The calibration line is determined by linear regression.

5.2.5 Carrying out manual calibration (ConCal)

Make sure that, in the menu for measurement and calibration settings (in the menu pH/ (see section 9.1.1 S
Use one to five buffer solutions in any order. The pH values of the buffer solutions have to differ by at least one pH unit.
<ENTER>/Calibration / Buffer) the buffer set ConCal is selected
ETTINGS FOR PH MEASUREMENTS, page 63).
If single-point calibration was set in the menu, the calibration proce­dure is automatically finished with the measurement of buffer solu­tion 1 and the calibration record is displayed.
1. Connect the pH sensor to the meter. The pH measuring window is displayed.
2. Keep the buffer solutions ready.
3. Start the calibration with The calibration display for the first buffer appears (voltage display).
<CAL>.
4. Thoroughly rinse the sensor with deionized water.
5. Immerse the pH sensor in buffer solution 1.
6. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
<ENTER>.
ba76141e07 08/2018 31
Page 32
pH value MultiLab 4010-1W
pH
0.0
24.8 °C
ConCal
7.000
01.07.2011 08:00
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
01.07.2011 08:00
Buffer
 
7. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
The pH value of the buffer solution is displayed.
Continuing with two-
point calibration
8. Set the nominal buffer value for the measured temperature with
<><>.
9. Accept the calibration value with The calibration display for the next buffer appears (voltage display).
1 0 . If necessary, finish the calibration procedure as a single-point calibration
with
<M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the zero point of the IDS­pH sensor.
11. Thoroughly rinse the sensor with deionized water.
12. Immerse the pH sensor in buffer solution 2.
13. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
<ENTER>.
<ENTER>.
32 ba76141e07 08/2018
Page 33
MultiLab 4010-1W pH value
pH
4.035
24.8 °C
ConCal
2
01.07.2011 08:00
Buffer
 
pH
9.958
24.8 °C
ConCal
3
01.07.2011 08:00
Buffer
 
14. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
The pH value of the buffer solution is displayed.
15. Set the nominal buffer value for the measured temperature with
<><>.
Continuing with
three- to five-point
calibration
16. Accept the calibration value with The calibration display for the next buffer appears (voltage display).
17. If necessary, finish the calibration procedure as a two-point calibration
<M>.
with The calibration record is displayed.
18. Thoroughly rinse the sensor with deionized water.
19. Immerse the sensor in the next buffer solution.
20. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
21. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>.
<ENTER>.
<ENTER>.
The pH value of the buffer solution is displayed.
ba76141e07 08/2018 33
Page 34
pH value MultiLab 4010-1W
22. Set the nominal buffer value for the measured temperature with
<><>.
23. Accept the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
24. If necessary, use
<M> to finish the calibration.
The calibration record is displayed. or Continue calibrating using the next buffer with
<ENTER>.
After the fifth buffer has been measured the calibration is automat­ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.

5.2.6 Calibration points

Calibration can be performed using one to five buffer solutions in any order (single-point to five-point calibration). The meter determines the following values and calculates the calibration line as follows:
Calibration Determined
Displayed calibration data
values
1-point Asymmetry Zero point = Asymmetry
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point Asymmetry
Slope
3-point to 5­point
Asymmetry Slope
Zero point = AsymmetrySlope = Slope
Zero point = AsymmetrySlope = Slope
The calibration line is calculated by linear regression.
You can display the slope in the units, mV/pH or % (see section
9.1.1 S
ETTINGS FOR PH MEASUREMENTS, page 63).

5.2.7 Calibration data

The calibration record is automatically transmitted to the interface after calibrating.
34 ba76141e07 08/2018
Page 35
MultiLab 4010-1W pH value
The calibration record of the last calibration is available in the menu Calibration / Calibration record. To open it in the measured value display, press the
<CAL__> key.
The calibration records of the last 10 calibrations are available in the menu Cali­bration / Calibration data storage. To open the Calibration menu in the
measured value display, press the
<ENTER> key.
Menu item Setting/
function
Calibration / Calibration data stor­age / Display
Calibration / Calibration data stor­age / Output to RS232/USB
- Displays the calibration records.
- Outputs the calibration data storage
Explanation
Further options: Scroll through the calibration
records with
Output the displayed calibration
record to the interface with [USB output].
Output all calibration records to
the interface with output].
Quit the display with
<ENTER>.
or
Switch directly to the measured
value display with
to the interface (see section 12
RANSMITTING DATA, page 120).
T
<><>.
<F2>/
<F2__>[USB
<F1>/[Back]
<M>.
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the calibration. The zero point and slope are evaluated separately. The worse evaluation of both is taken into account. The evaluation appears on the display and in the calibration record.
Display Calibration
record
+++
++
+
Zero point [mV]
-15 ... +15 -60,5 ... -58,0
-20 ... <-15 or >+15 ... +20
-25 ... <-20 or >+20 ... +25
Slope [mV/pH]
>-58.0 ... -57.0
-61.0 ... <-60.5 or >-57.0 ... -56.0
ba76141e07 08/2018 35
Page 36
pH value MultiLab 4010-1W
4010-1W
Ser. no. 11292113 CALIBRATIONpH
01.02.2014 15:55 Ser. no. 10501234
YSI Buffer 1 4.00 Buffer 2 7.00 Buffer 3 10.00 Voltage 1 184.0 mV Voltage 2 3.0 mV Voltage 3 -177.0 mV Temperatur 1 24.0 °C Temperatur 2 24.0 °C Temperatur 3 24.0 °C Slope -60.2 mV/pH Asymmetry 4.0 mV
Sensor +++
etc...
Calibration record
(USB output)
Display Calibration
record
-
Zero point [mV]
-30 ... <-25 or
->+25 ... +30
Slope [mV/pH]
-62.0 ... <-61.0 or >-56.0 ... -50.0
Clean the IDS sensor according to the sensor operating manual
Error Error
Error elimination (see section 14 W
<-30 or >+30
HAT TO DO IF..., page 92)
<-62.0 or > -50,0
For pH IDS sensors you can optionally enable a more finely graded calibration evaluation (QSC) (see section 5.2.9 QSC
SENSOR QUALITY CONTROL), page 38).
(
FUNCTION
36 ba76141e07 08/2018
Page 37
MultiLab 4010-1W pH value
1
2
4
3
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output

5.2.8 Continuous measurement control (CMC function)

The Continuous Measurement Control (CMC function) facilitates to evaluate the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is dis­played in the measured value display. Here you can very clearly see whether or not the current measured value is in the calibrated part of the measuring range.
The following information is displayed:
1 Currently measured pH value (needle) 2 Marking lines for all nominal buffer values used with the last valid cali-
bration
3 Measuring range for which a valid calibration is available. Measured
values in this range are suitable for documentation.
4 Measuring range for which no valid calibration is available (shaded).
Measured values in this range are not suitable for documentation. Calibrate the meter with buffers covering this measuring range. If the current measured value is outside the calibrated range, this area is shaded stronger. If a measured value is outside the measuring range pH 0 - 14, over­flow arrows are displayed at the left or right edge of the measuring range.
The limits of the calibrated range are determined by the buffers used for cali­bration:
Lower limit: Buffer with lowest pH value - 2 pH units Upper limit: Buffer with highest pH value + 2 pH units
ba76141e07 08/2018 37
Page 38
pH value MultiLab 4010-1W
1
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output

5.2.9 QSC function (sensor quality control)

General information
on the QSC function
The QSC function (Quality Sensor Control) is a new sensor evaluation for digital IDS sensors. It evaluates the condition of an IDS pH sensor individually and with a very fine grading.
The QSC scale shows the current sensor evaluation with an indicator on the display.
1QSC scale
The double arrow on the QSC scale indicates the current sensor evaluation
Sensor evaluation
with / without QSC function
QSC calibration The QSC function is enabled by once carrying out an additional three-point cali-
In the USB output the sensor evaluation is given as a percentage (1-100). The finely graded sensor evaluation of the QSC function promptly calls your
attention to changes of the sensor. Thus you can do what is necessary to restore the optimum measuring quality (e.g. clean, calibrate or replace the sensor). clean, calibrate or replace the sensor).
With QSC function Without QSC function (sensor sym-
bol)
Very fine grading of the sensor eval­uation (100 grades)
The reference value is individually determined for each sensor during the QSC initial calibration.
Low tolerances for zero point and slope when using QSC buffer solu­tions
Additional QSC calibration required (with special QSC buffer set)
bration with special QSC buffer solutions. It covers the measuring range of the sensor (pH 2 to pH 11). The QSC initial calibration determines the actual condi­tion of the sensor and stores it as a reference in the sensor. To meet the high requirements of a QSC initial calibration, the QSC initial cali­bration should optimally be carried out with the initial commissioning of the sensor.
Rough grading of the sensor evalua­tion (4 grades)
A theoretical reference value is used for all sensors
Greater tolerances for zero point and slope when using commercial buffer sets
No additional calibration required
38 ba76141e07 08/2018
Page 39
MultiLab 4010-1W pH value
pH 1
-180.0
24.8 °C
AutoCal OSC-Kit
01.07.2011 08:00
Buffer
mV
Carry out the normal calibrations for your special measuring range with your usual standard solutions as previously done.
As soon as the QSC function was enabled for an IDS sensor, it is not possible to return to the sensor evaluation with the sensor symbol for this sensor.
Carrying out a QSC
initial calibration
Carry out the QSC initial calibration very carefully. It determines the reference value for the sensor. This reference value cannot be overwritten or reset. As soon as the QSC function was enabled, it is not possible to return to the sensor evaluation with the sensor symbol.
1. Prepare for calibration the meter, sensor and the buffer solutions of the QSC-Kit.
2. Open the menu for measurement settings with
3. In the QSC menu, select First calibration with The calibration display appears. AutoCal QSC-Kit is displayed as the buffer. Exclusively use the QSC-Kit for the QSC calibration. If you use other buffers, you will have no valid QSC calibration.
<ENTER>.
<><>.
ba76141e07 08/2018 39
4. Start the measurement with Calibration with the buffers of the QSC-Kit is done like a normal three­point calibration (see section 5.2.4 C
TION (AUTOCAL), page 28).
5. As soon as the three-point calibration has been successfully carried out you can decide whether to accept or discard the calibration as the QSC initial calibration.
The QSC initial calibration is completed. The sensor is calibrated. If you want to calibrate with special buffers for your measurements, you can subsequently carry out a normal calibration with your buffers. The reference values deter­mined with the QSC calibration are also used for the evaluation of normal cali­brations. In the measured value display, the QSC scale of the QSC function is always displayed. A double arrow on the QSC scale indicates the current sen-
<ENTER>.
ARRYING OUT AUTOMATIC CALIBRA-
Page 40
pH value MultiLab 4010-1W
1
25.0
7.007
pH
°C
0
14
01.02.2014 08:00
i
USB output
sor evaluation.
1QSC scale
The double arrow on the QSC scale indicates the current sensor evaluation
Carrying out a QSC
control calibration
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably changed (after some normal calibrations).
You can carry out QSC control calibrations at greater intervals than normal cali­brations.
1. Prepare for calibration the meter, sensor and the buffer solutions of the QSC-Kit.
2. Open the menu for measurement settings with
3. In the QSC menu, select Control calibration with The calibration display appears. AutoCal QSC-Kit is displayed as the buffer. Exclusively use the QSC-Kit for the QSC calibration. If you use other buffers, you will have no valid QSC control calibration.
4. Start the measurement with Calibration with the buffers of the QSC-Kit is done like a normal three­point calibration (see section 5.2.4 C
TION (AUTOCAL), page 28).
As soon as the three-point calibration has been successfully carried out you can decide whether to accept or discard the calibration as the QSC control calibration.
<ENTER>.
ARRYING OUT AUTOMATIC CALIBRA-
<ENTER>.
<><>.
40 ba76141e07 08/2018
Page 41
MultiLab 4010-1W ORP
U
157.0 mV
01.02.2014 08:00
i
USB output

6ORP

6.1 Measuring

6.1.1 Measuring the ORP

NOTE
When connecting a grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated.
IDS ORP sensors are not calibrated. However, you can check IDS ORP sensors using a test solution.
1. Connect the ORP sensor to the meter. The ORP measuring window is displayed.
Stability control
(AutoRead)
& HOLD function
2. Temper the test sample or measure the current temperature.
3. Check the meter with the ORP sensor.
4. Immerse the ORP sensor in the test sample.
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproduc­ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the set­ting for automatic Stability control (see section 9.5.3 A
TROL, page 74) in the System menu.
To start the Stability control function manually, the HOLD function must be enabled.
UTOMATIC STABILITY CON-
ba76141e07 08/2018 41
Page 42
ORP MultiLab 4010-1W
Hold function
Stability control
1. Freeze the measured value with
<AR>.
The [HOLD] status indicator is displayed. The HOLD function is active.
You can terminate the HOLD function and the Stability control func- tion with
<AR> or <M> at any time.
2. Using <ENTER>, activate the Stability control function manually. The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the mea­sured parameter flashes. The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears, the display of the mea ­sured parameter stops flashing, and there is a beep. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu­ally with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
Criteria for a stable
measured value
ENSOR-INDEPENDENT
3. Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
or Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the pre­vious indication.
The Stability control function checks whether the measured values are stable within the monitored time interval.
Measured parameter
ORP 15 seconds Temperature 15 seconds
Time interval Stability in the time
interval
: better than 0.3 mV : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.

6.1.2 Measuring the temperature

For reproducible ORP measurements, it is essential to measure the tempera-
42 ba76141e07 08/2018
Page 43
MultiLab 4010-1W ORP
ture of the test sample. When a sensor with integrated temperature sensor is operated, the tempera-
ture value is displayed and recorded together with the measured value.

6.2 ORP calibration

ORP electrodes are not calibrated. You can, however, check ORP electrodes by measuring the ORP of a test solution and comparing the value with the nominal value.
ba76141e07 08/2018 43
Page 44
Dissolved oxygen MultiLab 4010-1W
O
2
mg/l
7.92
25.0
°C
760 mm Hg
01.02.2014 08:00
i
USB output

7 Dissolved oxygen

The general procedures for measuring, calibrating and setting are very similar with the various types of IDS-D.O. sensors. Any individual special features of the IDS-D.O. sensors are indicated in the general procedures.
The individual feature of the FDO 4410 is, for example, the high stability of its measuring characteristics during its total lifetime. Therefore, this sensor does not have to be calibrated regularly. It can simply be checked (FDO Check) instead.
A sensor with stirrer can only be operated while the meter is line­powered.

7.1 Measuring

7.1.1 Measuring D.O.

1. Connect the IDS-D.O. sensor or a D.O. sensor with IDS adapter to the meter. The D.O. measuring screen is displayed.
2. If necessary, select the measured parameter with
3. Check or calibrate the meter with the sensor.
Only in special cases does the FDO 4410-D.O. sensor require ca l­ibration. Regular checking with the aid of the FDO Check is suffi­cient.
4. Immerse the D.O. sensor in the test sample.
<M>.
measured parameter
44 ba76141e07 08/2018
Selecting the
displayed
You can switch between the following displays with
D.O. concentration [mg/l]D.O. saturation [%]
<M>:
Page 45
MultiLab 4010-1W Dissolved oxygen
Salinity correction When measuring the D.O. concentration [mg/l] of solutions with a salt content
of more than 1 psu, a salinity correction is required. For this, you have to me a­sure and input the salinity of the measured medium first.
When the salinity correction is switched on, the [Sal] indicator is displayed in the measuring screen.
You can switch the salinity correction on or off and enter the salinity in the menu for calibration and measurement settings (see section
9.3.1 S
CALIBRATION SETTINGS), page 67).
ETTINGS FOR D.O. SENSORS (MENU FOR MEASUREMENT AND
Air pressure correc-
tion
(DO % Saturation
local function)
Stability control
(AutoRead)
& HOLD function
The integrated air pressure sensor of the MultiLab 4010-1W measures the current air pressure.
During calibration, the air pressure correction function is automatically acti­vated. While the parameter oxygen saturation [%] is displayed, the air pressure correction is applied if the DO % Saturation local function is active.
The current air pressure is shown on the display when an IDS oxygen sensor is connected.
The air pressure correction (function DO % Saturation local) is switched on or off in the menu for calibration and measurement set­tings (see section 9.3.3 DO % S
ATURATION LOCAL, page 69).
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproduc­ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range when the automatic Stability control is switched off.
Irrespective of the setting for automatic Stability control (see section 9.5.3
UTOMATIC STABILITY CONTROL, page 74) in the System menu, yo u can start a
A measurement with Stability control manually at any time.
To start the Stability control function manually, the HOLD function must be enabled.
Hold function
1. Freeze the measured value with <AR>. The [HOLD] status indicator is displayed. The HOLD function is active.
You can terminate the HOLD function and the Stability control func- tion with
ba76141e07 08/2018 45
<AR> or <M> at any time.
Page 46
Dissolved oxygen MultiLab 4010-1W
Stability control
2. Using
<ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the mea­sured parameter flashes. The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears, the display of the mea­sured parameter stops flashing, and there is a beep. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu­ally with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
3. Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
ENSOR-INDEPENDENT
or Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the previous indication.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable within the monitored time interval.
Measured parameter Time interval Stability in the time
interval
D.O. concentration 20 seconds D.O. saturation 20 seconds Temperature 15 seconds
: better than 0.03 mg/l : better than 0.4 % : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the mon­itored time interval. The actual duration is mostly longer.

7.1.2 Measuring the temperature

For reproducible D.O. measurements, it is essential to measure the tempera­ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte­grated in the IDS sensor.
46 ba76141e07 08/2018
Page 47
MultiLab 4010-1W Dissolved oxygen

7.2 FDO Check (checking the FDO 4410)

The sensor is checked in water vapor-saturated air in the check and storage vessel FDO Check.

7.2.1 Why should you check the sensor?

With the FDO Check procedure you can find out in a simple manner whether the FDO 4410 D.O. sensor should be cleaned or calibrated.

7.2.2 When should you check the sensor?

Checking can be useful in the following cases: When the check interval has expired
(the [check] status indicator is displayed).
If the measured values seem to be implausibleIf you assume that the sensor cap is contaminated or at the end of
its lifetime
After the sensor cap was exchangedRoutinely within the framework of the company quality assurance

7.2.3 Perform a FDO Check

Proceed as follows to carry out the FDO Check:
1. Connect the D.O. sensor to the meter.
2. Place the D.O. sensor in the FDO Check check and storage vessel.
3. Close the check and storage vessel FDO Check.
The sponge in the check and storage beaker must be moist (not wet). Leave the sensor in the check and storage beaker long enough so it can adapt to the ambient temperature.
4. In the menu for measurement and calibration settings (
<ENTER> / FDO
Check / Start FDO Check), start the FDO Check. The meter switches to the measured parameter, %. During the FDO Check the DO % Saturation local function is automati­cally activated.
ba76141e07 08/2018 47
Page 48
Dissolved oxygen MultiLab 4010-1W
O
2
99.8
%
25.0
°C
FDO Check
[HOLD]
01.02.2014 08:00
[L][t90= 30 s]
5. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
6. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>.
<ENTER>.
The measured value is frozen.
7. Switch to the measured value display with The check measurement is not documented. After the FDO Check the DO % Saturation local function is automatically reset to the setting before the FDO Check.

7.2.4 Evaluation

The evaluation is based on the accuracy required and de fined by the user (e.g. ± 2 %). Together with the nominal value (100 %) this results in a validity scope for the check of 98 to 102 % (see example).
If the measured value is within the validity scope, no cleaning or user calibration is required.
If the measured value is outside the validity scope, the sensor shaft and membrane should be cleaned, and the check should then be repeated (see section 5.4.1).
<M>.
Example: – Required accuracy: ± 2 %. – In water vapor-saturated air or air-saturated water, the nominal
value for the relative D.O. saturation (abbreviated: saturation) is
100 %. – Therefore, the validity scope is 98 ... 102 % – The check resulted in a measured value of 99.3 %
The measurement error is within the specified validity scope. No cleaning or user calibration is required.
48 ba76141e07 08/2018
Page 49
MultiLab 4010-1W Dissolved oxygen

7.3 Calibration

7.3.1 Why calibrate?

D.O. sensors age. Aging changes the zero point and slope of the D.O. sensor. As a result, an inexact measured value is displayed. Calibration determines and stores the current values of the zero point and slope.
The FDO 4410 D.O. sensor ages so little it does not have to be reg­ularly calibrated. To detect changes of the sensor as early as possible, the FDO Check procedure can be useful (see section 7.2 FDO C
CHECKING THE FDO 4410), page 47).
(

7.3.2 When to calibrate?

If your evaluation of the FDO Check suggests calibrating (only FDO 4410)When the calibration interval has expired
HECK
When your accuracy requirements are especially highRoutinely within the framework of the company quality assuranceAfter a Zero calibration.

7.3.3 Calibration procedures

The MultiLab 4010-1W provides 2 calibration procedures:
Calibration in water vapor-saturated air. Calibration via a comparison measurement (e.g. Winkler titration according
to DIN EN 25813 or ISO 5813). At the same time, the relative slope is
adapted to the comparison measurement by a correction multiplier. When
the correction multiplier is active, the [Factor] indicator appears in the mea-
suring window.
For both calibration procedures, an additional Zero calibration is possible (see section 7.3.6 Z
ERO CALIBRATION, page 52).

7.3.4 Calibration in water vapor-saturated air

For this calibration procedure, the Comparison meas. setting must be set to off in the Calibration menu.
As the calibration vessel use a BOD bottle that contains a small amount of clean water (approx. 40 ml). The sensor must not be immersed in the water.
To calibrate the FDO 4410, use the calibration and storage vessel FDO Check. The sponge in the check and storage beaker must be moist (not wet).
ba76141e07 08/2018 49
Page 50
Dissolved oxygen MultiLab 4010-1W
O
2
99.8 %
25.0
°C
Cal
01.02.2014 08:00
DO Sat
Proceed as follows to calibrate the D.O. sensor:
1. Connect the D.O. sensor to the meter.
2. Put the D.O. sensor into the calibration vessel.
3. For FDO 4410 with the check and storage vessel FDO Check: Close the check and storage vessel FDO Check.
Leave the sensor in the calibration vessel or check and storage vessel (FDO 4410) long enough (at least 15 minutes) until the air is saturated with water vapor and the sensor has adapted to the ambient temperature.
4. Start the calibration with
5. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
6. Wait for the AutoRead measurement to be finished (beep, [HOLD][AR] status indicator). The calibration record is displayed and output to the interface.
<CAL>.
<ENTER>.
7. Switch to the measured value display with

7.3.5 Calibrating with Comparison meas. (e.g. Winkler titration)

With the calibration procedure Comparison meas., the measured value of the sensor is adjusted to the nominal value of a comparison solution via a correc-
50 ba76141e07 08/2018
tion multiplier. The current correction multiplier is documented in the sensor
i
menu ( For this calibration procedure, the Comparison meas. setting must be set to on
in the Calibration menu.
Factor = x.xxx) and in the calibration record.
<F1>/[continue].
Page 51
MultiLab 4010-1W Dissolved oxygen
O
2
7.92
mg/l
25.1
°C
Compare
03.04.2013 08:00
DO Adjust
O
2
7.92
mg/l
20.3°C
Compare
03.04.2013 08:00
 
DO Adjust
Proceed as follows to calibrate the D.O. sensor:
1. Connect the D.O. sensor to the meter.
2. Immerse the D.O. sensor in the reference solution.
3. Start the calibration with
4. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
5. Wait for the measurement with stability control to be finished (beep, [HOLD][AR] status indicator).
<CAL>.
<ENTER>.
6. Using to the nominal value (value of the comparison measurement). Subse­quently take over the adjustment with The meter switches to the measured value display. The status display [Factor] is active.
ba76141e07 08/2018 51
<><>, set the measured value to adjust the displayed value
<ENTER>.
Page 52
Dissolved oxygen MultiLab 4010-1W
O
2
0.9
%
Cal
03.04.2013 08:00
25.1°C
DO Zero

7.3.6 Zero calibration

With a Zero calibration, the zero point of the sensor is redetermined by cali­brating in a zero solution.
For most applications, an additional Zero calibration is not required and not recommended. Only with very rare applications can the accuracy of a calibra­tion be improved if a Zero calibration was carried out before.
The FDO 4410 D.O. sensor ages so little it does not have to be zero point calibrated. A zero point calibration is not allowed for this sensor.
The zero point calibration is best carried out prior to calibrating with a calibra­tion procedure (e.g. calibration in water vapor-saturated air or calibration with comparison measurement).
1. Connect the D.O. sensor to the meter.
2. Place the D.O. sensor in a solution that does not contain any dissolved oxygen.
A solution not containing any dissolved oxygen can be prepared by dissolving approx. 8 to 10 g sodium sulfite (Na water. Carefully mix the solution. It may take up to 60 minutes until the solution is free of oxygen.
3. In the menu for measurement and calibration settings ( ibration / Zero calibration), start the Zero calibration. The calibration point for the measured value 0 (DO Zero) is displayed.
) in 500 ml tap-
2SO3
<ENTER> / Cal-
4. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
5. Wait for the AutoRead measurement to be finished (beep, [HOLD][AR] status indicator). The current value is set to zero. The calibration record is displayed.
<ENTER>.
52 ba76141e07 08/2018
Page 53
MultiLab 4010-1W Dissolved oxygen
6. Switch to the measured value display with The zero point is calibrated. The [ZeroCal] status indicator is displayed.
7. Carry out a calibration (see section 7.3.3 C page 49).

7.3.7 Calibration data

The calibration record is automatically transmitted to the interface after calibrating.
The calibration record of the last calibration is available in the menu Calibration / Calibration record. To open it in the measured value display, press the
<F1>/[continue].
ALIBRATION PROCEDURES,
<CAL__> key.
The calibration records of the last 10 calibrations are available in the menu Cali­bration / Calibration data storage / Display. To open the Calibration menu in the
measured value display, press the
<ENTER> key.
Menu item Setting/
function
Calibration / Calibration data stor­age / Display
Calibration / Calibration data stor­age / Output to RS232/USB
- Displays the calibration records.
- Outputs the calibration data stor-
Explanation
Further options: Scroll through the calibration
records with
Output the displayed calibra-
tion record to the interface with
<><>.
<F2>/[USB output].
Output all calibration records to
the interface with
<F2__>[USB output].
Quit the display with
[Back] or
Switch directly to the mea-
sured value display with
age to the interface (see section
RANSMITTING DATA, page
12 T
120).
<ENTER>.
<F1>/
<M>.
Calibration evalua-
tion
ba76141e07 08/2018 53
After calibration, the meter automatically evaluates the current status of the ca l-
Page 54
Dissolved oxygen MultiLab 4010-1W
CALIBRATION Ox
03.04.2013 07:43:33 FDO 4410
Ser. no. 12B100016 SC-FDO 12B100015
Sensor +++
ibration. The evaluation appears on the display and in the calibration record.
For evaluation, the characteristic curve of the sensor that was determined is compared to the characteristic curve of an ideal sen­sor at similar environmental conditions (relative slope S): S = S
sor
/ S
ideal sensor
. An ideal sensor has a slope of 1.
sen-
Calibration evalua-
tion FDO 4410
Calibration evalua-
tion ProOBOD,
4100 ProBOD,
YSI 5010 with 4011
Adapter
Display Calibration record Relative slope
+++ S = 0.94 ... 1.06
++ S = 0.92 ... 0.94
or S = 1.06 ... 1.08
+ S = 0.90 ... 0.92
or
S = 1.08 ... 1.10 Error Error S < 0.90 Error elimination (see section 14 W
HAT TO DO IF...,
or
S > 1.10 page 92)
Display Calibration record
+++
Calibration record
(USB output,
example FDO 4410)
Error elimination (see section 14 W
HAT TO DO IF..., page 92)
54 ba76141e07 08/2018
Page 55
MultiLab 4010-1W Conductivity
μS/cm
[Tr25][nLF][ 0.835 1/cm]
1417
25.0
°C
01.02.2014 08:00
i
USB output

8 Conductivity

8.1 Measuring

8.1.1 Measuring the conductivity

NOTE
When connecting a grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated.
1. Connect the conductivity sensor to the meter. The conductivity measuring window is displayed. The measuring cell and cell constant for the connected IDS conductivity sensor are automatically taken over.
Selecting the
displayed
measured parameter
2. If necessary, press
3. Immerse the conductivity sensor in the test sample.
You can switch between the following displays with
Conductivity [μS/cm] / [mS/cm]Resistivity [Ω·cm] / [kΩ·cm] / [MΩ·cm]Salinity Sal [ ] (
psu)
<M> to select the measured parameter ϰ.
<M>:
Stability control
& HOLD function
ba76141e07 08/2018 55
(AutoRead)
Total dissolved solids TDS [mg/l] / [g/l] The multiplier to calculate the total dissolved solids is set to 1.00 in the factory.
You can adjust this multiplier to meet your requirements in the range 0.40 ...
1.00. The multiplier is set in the menu for the parameter, TDS.
Determining the TDS multiplier: See section 19.3 C
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproduc-
ALCULATE THE TDS MULTIPLIER, page 109.
Page 56
Conductivity MultiLab 4010-1W
ibility of measured values. The measured parameter flashes on the display
as soon as the measured value is outside the stability range when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the set­ting for automatic Stability control (see section 9.5.3 A
TROL, page 74) in the System menu.
UTOMATIC STABILITY CON-
To start the Stability control function manually, the HOLD function must be enabled.
Hold function
Stability control
1. Freeze the measured value with
<AR>.
The [HOLD] status indicator is displayed. The HOLD function is active.
You can terminate the HOLD function and the Stability control func- tion with
<AR> or <M> at any time.
2. Using <ENTER>, activate the Stability control function manually. The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the measured parameter flashes. The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears, the display of the mea­sured parameter stops flashing, and there is a beep. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu­ally with
<ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
3. Using
You can switch off the beep (see section 9.5 S
SETTINGS, page 73).
<ENTER>, start a further measurement with stability control.
ENSOR-INDEPENDENT
or Release the frozen measured value again with
<AR> or <M>.
The [AR] status display disappears. The display switches back to the previous indication.
Criteria for a stable
The Stability control function checks whether the measured values are stable
measured value
56 ba76141e07 08/2018
Page 57
MultiLab 4010-1W Conductivity
within the monitored time interval.
Measured
Time interval Stability in the time interval
parameter
Conductivity ϰ 10 seconds
ϰ: better than
1.0% of measured value
Temperature 15 seconds
: better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the mon­itored time interval. The actual duration is mostly longer.

8.1.2 Measuring the temperature

For reproducible conductivity measurements, it is essential to measure the temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in the IDS sensor.

8.2 Temperature compensation

The calculation of the temperature compensation is based on the preset refer­ence temperature, 20 °C or 25 °C. It appears on the display as Tr20 or Tr25.
You can select one of the following temperature compensation methods:
 Nonlinear temperature compensation (nLF) according to EN 27 888  Linear temperature compensation (Lin) with adjustable coefficient  No temperature compensation (off)
The reference temperature and temperature compensation are set in the menu for the parameter, conductivity (see section 9.4.1 S
TINGS FOR IDS CONDUCTIVITY SENSORS, page 70).
ET-
Application tips Select the following temperature compensations give n in the table according to
the respective test sample:
Test sample Temperature compensation Display
Natural water (ground water, surface
nLF according to EN 27 888
nLF
water, drinking water) Ultrapure water nLF
nLF
according to EN 27 888
Other aqueous solu­tions
Lin
adjustable temperature coeffi-
Lin
cient 0.000 ... 10.000 %/K
ba76141e07 08/2018 57
Page 58
Conductivity MultiLab 4010-1W
Test sample Temperature compensation Display
Salinity (seawater) Automatic nLF according to
Sal, nLF
IOT (International Oceano­graphic Tables)

8.3 Calibration

8.3.1 Why calibrate?

Aging slightly changes the cell constant, e. g. due to coatings. As a result, an inexact measured value is displayed. The original characteristics of the cell can often be restored by cleaning the cell. Calibration determines the current value of the cell constant and stores this value in the meter. Thus, you should calibrate at regular intervals.

8.3.2 When to calibrate?

Routinely within the framework of the company quality assuranceWhen the calibration interval has expired

8.3.3 Calibration procedure

The MultiLab 4010-1W provides 2 calibration procedures: Determining the cell constant
Calibration in the check- and calibration standard 0.01 mol/l KCl (1413 µS/ cm @25 °C) Safe and simple calibration procedure for IDS conductivity sensors with a cell constant in the range 0.450 ... 0.500 cm
-1
.
Determining the cell constant
Calibration with freely selectable check- and calibration standard Comprehensive and flexible calibration procedure for all IDS conductivity sensors, irrespective of the cell constant.
Which calibration procedures can be used depends on the conductivity sensor used. The menu for measurement settings automatically displays only those settings and calibration procedures that are available for the sensor.
With conductivity sensors for which a cell constant was set, no sensor symbol appears on the display. If a calibration interval was specified, it is not active.
8.3.4 Determining the cell constant (calibration in the check- and cali-
bration standard)
For this calibration procedure, IDS conductivity sensors with a cell constant in
58 ba76141e07 08/2018
Page 59
MultiLab 4010-1W Conductivity
1/cm
0.475
25.0
°C
03.04.2013 08:00
the range 0.450 ... 0.500 cm-1 are suitable, e.g. 4310. IDS conductivity sensors with other cell constants are unsuitable for this simple
calibration procedure. As an alternative, you can manually determine the cell constant and adjust it in the menu (see section 8.3.5 S
STANT (CALIBRATION WITH FREELY SELECTABLE CHECK- AND CALIBRATION STAN- DARD), page 60).
ETTING THE CELL CON-
You can determine the actual cell constant of the IDS conductivity sensor within a valid range (e.g. 4310: 0.450 ... 0.500 cm
The cell constant is determined in the check- and calibration standard 0.01 mol/ l KCl (1413 µS/cm @ 25 °C).
In the default condition, the calibrated cell constant is set to the nominal cell constant of the IDS sensor (e.g. 4310: 0.475 cm
For this calibration procedure, the Type setting must be set to cal. Proceed as follows to determine the cell constant:
1. Connect the conductivity sensor to the meter.
2. In the measured value display, select the conductivity parameter with
-1
).
-1
).
<M>.
3. Start the calibration with The cell constant that was calibrated last is displayed.
<CAL>.
4. Immerse the conductivity sensor in the check- and calibration standard
0.01 mol/l KCl (1413 µS/cm @ 25 °C).
5. Start the measurement with The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes.
6. Wait for the end of the measurement with stability control (beep, [HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>.
<ENTER>.
The calibration record is displayed and output to the interface.
7. Switch to the measured value display with
ba76141e07 08/2018 59
<F1>/[continue].
Page 60
Conductivity MultiLab 4010-1W
Calibration Type man Cell const. man: 0.475 1/cm Temp. comp. (TC): Multiplier for TDS: 1.00 Reset:
[
i
] ϰ = 1432 µS/cm
ϰ
03.04.2013 08:00

8.3.5 Setting the cell constant (calibration with freely selectable check- and calibration standard)

You can set within a valid range the cell constant of the IDS conductivity sensor (setting range: see sensor operating manual).
With a freely selectable check- and calibration standard with known conduc­tivity nominal value (within the measuring range of the sensor), you can exactly adjust the cell constant to the check- and calibration standard by the changing conductivity value displayed.
In the default condition, the cell constant is set to the nominal cell constant of the IDS sensor.
For this calibration procedure, the Type setting must be set to man. Proceed as follows to set the cell constant:
1. Connect the conductivity sensor to the meter.
2. In the measured value display, select the conductivity parameter with
<M>.
3. Open the menu for measurement settings with
<ENTER>.
4. Select the setting of the cell constant
(4310: Menu Type: man and Cell const. man
4320: Menu Cell constant)
The current conductivity value is displayed in the status line.
5. Set the suitable temperature compensation and reference temperature
for the check- and calibration standard.
6. Immerse the conductivity sensor in the check- and calibration standard.
Wait for a stable measured value.
60 ba76141e07 08/2018
7. Using
<><>, adjust the cell constant until the displayed conductivity
value ([i] ϰ = ...) corresponds to the nominal value.
8. Switch to the measured value display with
The setting of the cell constant is taken over. No calibration record is output.
<M>.
Page 61
MultiLab 4010-1W Conductivity

8.3.6 Calibration data

The calibration record is automatically transmitted to the interface after calibrating.
The calibration record of the last calibration is available in the menu Calibration / Calibration record. To open it in the measured value display, press the
<CAL__> key.
The calibration records of the last 10 calibrations are available in the menu Cali­bration / Calibration data storage / Display. To open the Calibration menu in the
measured value display, press the
<ENTER> key.
Menu item Setting/
function
Calibration /
- Displays the calibration records.
Calibration data storage / Display
Calibration /
- Outputs the calibration data storage
Calibration data storage / Output to RS232/USB
Explanation
Further options: Scroll through the calibration
records with
<><>.
Output the displayed calibration
record to the interface with
<F2>/[USB output].
Output all calibration records to
the interface with output].
Quit the display with
[Back] or
<ENTER>.
Switch directly to the measured
value display with
to the interface (see section 12
RANSMITTING DATA, page 120).
T
<F2__>[USB
<F1>/
<M>.
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the ca l­ibration. The evaluation appears on the display and in the calibration record.
Display Calibration record Cell constant [cm-1]
+++ Within the range
0.450 ... 0.500 cm
-1
Error Error Outside the range Error elimination (see section 14
HAT TO DO IF..., page 92)
W
ba76141e07 08/2018 61
0.450 ... 0.500 cm
-1
Page 62
Conductivity MultiLab 4010-1W
CALIBRATION Cond
03.04.2013 07:43:33 4310
Ser. no. 09250033 Cell constant 0.476 1/cm 25.0 °C
Sensor +++
Calibration record
(USB output)
62 ba76141e07 08/2018
Page 63
MultiLab 4010-1W Settings

9 Settings

9.1 pH measurement settings

9.1.1 Settings for pH measurements

Settings The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea­sured parameter in the measured value display and press the After completing the settings, switch to the measured value display with
Default settings are printed in bold.
<ENTER> key.
<M>.
Menu item Possible
setting
Calibration / Calibration record
Calibration / Calibration data stor­age / Display
Calibration / Calibration data stor­age / Output to RS232/USB
Calibration / Buffer YSI
Calibration / Single­point calibration
Calibration / Calibration interval
- Displays the calibration record of the last calibration.
- Displays the last calibration records (max. 10)
- Outputs the calibration data storage to the interface
ConCal NIST/DIN ...
yes
no 1 ... 7 ... 999 dCalibration interval for the IDS pH sensor (in days).
Explanation
Buffer sets to be used for pH calibration. More buffers and details: see section 9.1.2 B
FOR CALIBRATION, page 64 and section 5.2 PH CALIBRATION,
page 28. Quick calibration with 1 buffer
The meter reminds you to calibrate regularly by the flashing sensor symbol in the measuring screen.
UFFER SETS
Calibration / Unit for slope
QSC / First calibration - Starts the initial calibration with QSC buffers.
QSC / Record of first calibration
QSC / Control calibra­tion
Man. temperature -25 ... +25 ...
ba76141e07 08/2018 63
mV/pH
%
- Displays the calibration record of the QSC initial calibration.
- Starts the control calibration with QSC buffers.
+130 °C
Unit of the slope. The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).
This menu item is only available as long as no initial calibra­tion was carried out with the connected IDS sensor.
This menu item is only available if an initial calibration was carried out with the connected IDS sensor.
Entry of the manually determined temperature. This menu item is only available if an IDS adapter is con­nected.
Page 64
Settings MultiLab 4010-1W
Menu item Possible
Explanation
setting
pH resolution 0.001
Resolution of the pH display
0.01
0.1
mV resolution 0.1
Resolution of the mV display
1
Reset - Resets all sensor settings to the delivery condition (see sec-
tion 9.6.1 R
ESETTING THE MEASUREMENT SETTINGS, pag e 75).

9.1.2 Buffer sets for calibration

You can use the buffer sets quoted in the table for an automatic calibration. The pH values are valid for the specified temperature values. The temperature dependence of the pH values is taken into consideration during the calibration.
No. Buffer set * pH values at
1 YSI * 4.000
25 °C
7.000
10.000
2 ConCal Any Any 3 NIST/DIN
DIN buffers according to DIN 19266 and NIST Traceable Buffers
1.679
4.006
6.865
25 °C
9.180
12.454
4 TEC
Technical buffers
2.000
4.010
25 °C
7.000
10.011
5 Merck 1* 4.000
20°C
7.000
9.000
6 Merck 2 * 1.000
20°C
6.000
8.000
13.000
7 Merck 3 * 4.660
20°C
6.880
9.220
8 Merck 4 * 2.000
20°C
4.000
7.000
10.000
64 ba76141e07 08/2018
Page 65
MultiLab 4010-1W Settings
No. Buffer set * pH values at
9 Merck 5 * 4.010
7.000
10.000
10 DIN 19267 1.090
4.650
6.790
9.230
11 Mettler Toledo USA * 1.679
4.003
7.002
10.013
12 Mettler Toledo EU * 1.995
4.005
7.002
9.208
13 Fisher * 2.007
4.002
7.004
10.002
14 Fluka BS * 4.006
6.984
8.957
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
15 Radiometer * 1.678
4.005
7.000
9.180
16 Baker * 4.006
6.991
10.008
17 Metrohm * 3.996
7.003
8.999
18 Beckman * 4.005
7.005
10.013
19 Hamilton Duracal * 4.005
7.002
10.013
20 Precisa * 3.996
7.003
8.999
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
21 Reagecon TEC * 2.000
25 °C
4.010
7.000
10.000
ba76141e07 08/2018 65
Page 66
Settings MultiLab 4010-1W
No. Buffer set * pH values at
22 Reagecon 20 * 2.000
20°C
4.000
7.000
10.000
13.000
23 Reagecon 25 * 2.000
25 °C
4.000
7.000
10.000
13.000
24 Chemsolute * 2.000
20°C
4.000
7.000
10.000
25 USABlueBook
*4.000
25 °C
7.000
10.000
* Brand names or trade names are tradema rks of their respective owners protected
by law.
Setting the calibra-
tion interval
The buffers are selected in the menu, pH / / Buffer (see section 9.1.1 S
ETTINGS FOR PH MEASUREMENTS, page
<ENTER> / Calibration
63).

9.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol. After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.9 QSC
FUNCTION (SENSOR QUALITY CONTROL), page
38).
After the specified calibration interval has expired the sensor symbol or the QSC scale flashes. It is still possible to measure.
To ensure the high measuring accuracy of the measuring system, calibrate after the calibration interval has expired.
The calibration interval is set to 7 days in the factory. You can change the interval (1 ... 999 days):
1. Open the menu for measurement settings with <ENTER>.
2. In the Calibration / Calibration interval menu, set the calibration interval
<><>.
with
66 ba76141e07 08/2018
Page 67
MultiLab 4010-1W Settings
3. Confirm the setting with
4. Quit the menu with
<ENTER>.
<M>.

9.2 ORP measurement settings

9.2.1 Settings for ORP measurements

The settings are made in the menu for measuring se ttings of the ORP measure­ment. To open the settings, display the required measured parameter in the measured value display and press the settings, switch to the measured value display with Default settings are printed in bold.
Menu item Possible setting Explanation
mV resolution 0.1
1
Reset - Resets all sensor settings to the
<ENTER> key. After completing the
<M>.
Resolution of the mV display
delivery condition (see section
9.6.1 R
SETTINGS, page 75).
ESETTING THE MEASUREMENT

9.3 D.O. measurement settings

9.3.1 Settings for D.O. sensors (menu for measurement and calibration settings)

Settings The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the measured value display and press the settings, switch to the measured value display with
The possible settings are individually displayed for each sensor. Below, the set­ting menu is shown with all settings for all IDS-D.O. sensors.
Default settings are printed in bold.
Menu item Possible
setting
Calibration / Calibration record
Calibration / Calibration data storage / Display
- Displays the calibration record of
- Displays the last calibration records
<ENTER> key. After completing the
<M>.
Explanation
the last calibration.
(max. 10)
Calibration / Calibration data storage / Output to RS232/USB
ba76141e07 08/2018 67
- Outputs the calibration data stor­age to the interface
Page 68
Settings MultiLab 4010-1W
Menu item Possible
setting
Calibration / Zero calibration(only for
4100 ProBOD, 5010 with 4011 Adapter­ProOBOD)
Calibration / Calibration interval
Calibration / Comparison meas.
Calibration / Sensor cap coefficients
(only for ProOBOD)
1 ... 180 ... 999 d
on
off
K1 ... K5 KC
Explanation
Starts the 0-point calibration (see section 7.3.6 Z
ERO CALIBRATION,
page 52)
Calibration interval for the D.O. sensor (in days). The meter reminds you to calibrate regularly by the flashing sensor symbol in the measuring screen.
Enables to adjust the measured value with the aid of a comparison measurement, e.g. Winkler titra­tion. For details, see section 7.3 C
BRATION, page 49.
ALI-
Enter the cap coefficients here after exchanging the sensor cap. For details, see section 9.3.2
NTER SENSOR CAP COEFFICIENTS
E
ROOBOD), page 69.
(P
Calibration / Membrane cap
(only for 4100 Pro­BOD, 5010 with 4011 Adapter)
FDO Check / Start FDO Check
(only for FDO 4410)
FDO Check / Check interval
(only for FDO 4410)
Salinity/Sal correc­tion (only for param-
eter mg/l) Salinity/Salinity
(only for parameter mg/l)
Black Yellow
After exchanging the sensor cap, select the cap type here.
- Starts the check with the
FDO Check
1 ... 60 ... 999 d
Interval for the FDO Check (in days). The meter reminds you to check the sensor regularly by FDO Check status indicator in the measuring screen.
on
off
Manual salt content correction for concentration measurements.
0.0 ... 70.0 Salinity or salinity equivalent for the salt content correction. This menu item is only available if the manual salinity correction is switched on.
68 ba76141e07 08/2018
Page 69
MultiLab 4010-1W Settings
Menu item Possible
setting
Resolution
(only for ProOBOD))
Response time t90
0.1
1
30 ... 300 s Response time of the signal filter (in
(only for FDO 4410)
DO % Saturation local
on
off
Explanation
Set a high or low resolution. The setting of the resolution is sep­arately stored for each measured parameter.
seconds). A signal filter in the sensor reduces the limits of variation of the mea­sured value. The signal filter is characterized by the response time t90. This is the time after which 90 % of a signal change is displayed.
DO % Saturation local is a proce­dure that takes the local air pres­sure into account for each saturation measurement. For details, see section 9.3.3 DO %
ATURATION LOCAL, page 69
S
Reset - Resets all sensor settings to the
delivery condition (see section
9.6.1 R
SETTINGS, page 75).
ESETTING THE MEASUREMENT

9.3.2 Enter Sensor cap coefficients (ProOBOD)

The values of the coefficients are provided with the sensor cap.
1. Change the digit of the highlighted position with
2. Go to the next position with
3. Confirm with
<ENTER> when a coefficient is completely entered.
< ><F2>/[ ].
<><>
.

9.3.3 DO % Saturation local

Irrespective of the height or air pressure, the calibration value is set to 100 %. The function DO % Saturation local fulfills the EU regulations for the parameter oxygen saturation [%].
When the DO % Saturation local is enabled the display shows an [L] for the parameter oxygen saturation.
D.O. mg/L readings are unaffected by the selection of the DO % Saturation local function.
ba76141e07 08/2018 69
Page 70
Settings MultiLab 4010-1W

9.4 Cond measurement settings

9.4.1 Settings for IDS conductivity sensors

The settings are made in the menu for the measured parameter, conductivity. To open the settings, display the required measured parameter in the mea­sured value display and press the tings, switch to the measured value display with The possible settings are individually displayed for each sensor. Below the set­ting menu is displayed for two IDS sensors (4310, 4320).
Default settings are printed in bold.
<ENTER> key. After completing the set-
<M>.
General setting
menu for
conductivity
Menu item Possible
Explanation
setting
Calibration / Calibration record
Calibration / Calibration data stor-
- Displays the calibration record of
the last calibration.
- Displays the last calibration records (max. 10)
age / Display Calibration /
Calibration data stor-
- Outputs the calibration data stor-
age to the interface
age / Output to RS232/USB
Calibration / Calibration interval
1 ... 150 ... 999 d
Calibration interval for the IDS
conductivity sensor (in days). The meter reminds you to cali­brate regularly by the flashing sensor symbol in the measuring screen.
Reset - Resets all sensor settings to the
delivery condition (see section
9.6.1 R
MENT SETTINGS, page 75).
ESETTING THE MEASURE-
70 ba76141e07 08/2018
Page 71
MultiLab 4010-1W Settings
Setting menu 4310
Menu item Possible
Explanation
setting
Type Used measuring cell
Cal Measuring cells whose cell con-
stants are determined through calibration in the check- and cali­bration standard 0.01 mol/l KCl (1413µS/cm @25°C). Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is displayed in the status line.
man Freely (manually) adjustable cell
constant in the range 0.450 ...
-1
.
Cell const. man 0.450 ...
0.475 ...
0.500 cm
0.500 cm Display and setting options for
the manually adjustable cell con-
-1
stant. This menu item is only available when Type man is set. The current conductivity value is displayed in the status line.
Temp. comp. (TC) / Method
nLF
Lin off
Procedure for temperature com­pensation (see section 8.2 T
PERATURE COMPENSATION, page
57). This setting is only available for the measured parameters, con­ductivity (ϰ) and resistivity (ρ).
Temp. comp. (TC) / Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear tempera­ture compensation. This menu item is only available when the linear temperature compensation is set.
Temp. comp. (TC) / Reference temp.
20 °C
25 °C
Reference temperature This setting is only available for the measured parameters, con­ductivity (ϰ) and resistivity (ρ).
Multiplier for TDS 0.40 ... 1.00 Multiplier for TDS value
EM-
ba76141e07 08/2018 71
Page 72
Settings MultiLab 4010-1W
Setting menu
4320
Menu item Possible
setting
Cell constant 0.090 ...
0.100 ...
0.110 cm
Temp. comp. (TC) / Method
nLF
Lin off
Temp. comp. (TC) / Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Temp. comp. (TC) / Reference temp.
20 °C
25 °C
Explanation
Display and setting options for the cell constant.
-1
The current conductivity value is displayed in the status line.
Procedure for temperature com­pensation (see section 8.2 T
PERATURE COMPENSATION, page
EM-
57). This setting is only available for the measured parameters, conductivity (ϰ) and resistivity (ρ).
Coefficient of the linear tempera­ture compensation. This menu item is only available when the linear temperature com­pensation is set.
Reference temperature This setting is only available for the measured parameters, conductivity (ϰ) and resistivity (ρ).
Multiplier for TDS 0.40 ... 1.00 Multiplier for TDS value
72 ba76141e07 08/2018
Page 73
MultiLab 4010-1W Settings

9.5 Sensor-independent settings

9.5.1 System

To open the Storage & config menu, press the sured value display. After completing the settings, switch to the measured value display with
<M>.
Default settings are printed in bold.
Menu item Possible setting Explanation
System / General / Lan­guage
System / General / Audio signal
System / General / Illumi- nation
English
(more)
on
off
Auto
on
Selects the menu language
Switching on/off the beep
Switches the display illumination on/off
off
System / General / Con-
0 ... 50 ... 100 Changes the display contrast
trast System / General / Shutoff
time System / General / Tem-
perature unit
10 min ... 1h ... 24 h
°C
°F
Adjusts the switch-off time
Temperature unit, degrees Celsius or degrees Fahrenheit. All temperature values are displayed with the selected unit.
<ENTER__> key in the mea-
System / General / Air pressure unit
System / General / Stabil­ity control
System / Interface / Baud rate
System / Interface / Out­put format
Only with: Output format CSV:
System / Interface /
Decimal separator
System / Interface /
Output header
mbar
mmHg
inHg
on
off
1200, 2400, 4800, 9600, 19200
ASCII
CSV
Dot (xx.x)
Comma (xx,x)
Air pressure unit
Switches on or off the automatic stability control during measurement (see section 9.5.3 A
TABILITY CONTROL, page 74 )
S
UTOMATIC
Baud rate of the USB Device interface
Output format for data transmission For details, see section 11 T
RANSMITTING DATA,
page 84
Decimal separator
Output of a header for Output format: CSV
ba76141e07 08/2018 73
Page 74
Settings MultiLab 4010-1W
Menu item Possible setting Explanation
System / Clock function Date format
Datum Time
System / Service informa­tion
Settings of time and date. For details, see section 4.4.5 E
TION: SETTING THE DATE AND TIME, page 24
XAMPLE 2 ON NAVIGA-
Hardware version and software version of the meter are displayed.
System / Reset - Resets the system settings to the default values.
For details, see section 9.6.2 R
TEM SETTINGS, page 77
ESETTING THE SYS-

9.5.2 Data storage

This menu contains all functions to display, edit and erase stored measured values.
Detailed information on the memory functions of the MultiLab 4010-1W are given in section 10 D
ATA STORAGE, page 78.

9.5.3 Automatic Stability control

The automatic Stability control function continuously checks the stability of the measurement signal. The stability has a considerable impact on the reproduc­ibility of measured values.
You can activate or switch off the automatic Stability control function (see section 9.5 S
ENSOR-INDEPENDENT SETTINGS, page 73).
The measured parameter flashes on the display
as soon as the measured value is outside the stability range when the automatic Stability control is switched off.

9.5.4 Automatic switch-off function

The instrument has an automatic switch-off function in order to save the batteries (see section 9.5.1 S
YSTEM, page 73). The automatic switchoff function
switches off the meter if no key is pressed for an adjustable period. The automatic switchoff function is not active
if the power pack is connectedif a USB-B cable is connected if the Automatic data storage function is active, or with automatic data trans-
mission
74 ba76141e07 08/2018
Page 75
MultiLab 4010-1W Settings

9.5.5 Display illumination

The meter automatically switches off the display illumination if no key is pressed for 20 seconds.
The illumination is switched on with the next keystroke again. You can also generally switch on the display illumination (see section 9.5.1
YSTEM, page 73).
S

9.6 Reset

You can reset (initialize) all sensor settings and sensor-independent settings separately from each other.

9.6.1 Resetting the measurement settings

The calibration data are reset to the default settings together with the measuring parameters. Recalibrate after performing a reset.
pH The following settings for pH measurements are reset to the default settings
with the Reset function:
Setting Default settings
Buffer YSI Calibration interval 7 d Unit for slope mV/pH Measured parameter pH Unit for slope 0.001 mV resolution 0.1 Asymmetry 0 mV Slope -59.2 mV Man. temperature 25 °C Single-point calibration off
The sensor settings are reset under the Reset menu item in the menu for cali­bration and measurement settings. To open the settings, display the required measured parameter in the measured value display and press the
<ENTER>
key.
ba76141e07 08/2018 75
Page 76
Settings MultiLab 4010-1W
ORP The following settings for ORP measurements are reset to the default settings
with the Reset function:
Setting Default settings
mV resolution 0.1
The sensor settings are reset under the Reset menu item in the menu for cali­bration and measurement settings. To open the settings, display the required measured parameter in the measured value display and press the
<ENTER>
key.
Dissolved oxygen The following settings for D.O. measurements are reset to th e default settin gs
with the Reset function:
Setting Default settings
Calibration interval 180 d Measured parameter D.O. concentration (mg/l) Salinity (value) 0.0 Salinity (function) off Number of calibration points 1 Resolution 0.1 DO % Saturation local off
The sensor settings are reset under the Reset menu item in the menu for cali­bration and measurement settings. To open the settings, display the required measured parameter in the measured value display and press the
<ENTER>
key.
Conductivity The following settings for conductivity measurements are reset to the default
settings with the Reset function:
Setting Default settings
Calibration interval 150 d Measured parameter
ϰ
Cell constant (c) Depending on the connected measur-
ing cell:
-1
0.475 cm
0.475 cm
0.100 cm
(calibrated)
-1
(set)
-1
Temperature compensation nLF Reference temperature 25 °C Temperature coefficient (TC) of
2.000 %/K
the linear temperature compensa­tion
TDS multiplier 1.00
76 ba76141e07 08/2018
Page 77
MultiLab 4010-1W Settings
The sensor settings are reset under the Reset menu item in the menu for cali­bration and measurement settings. To open the settings, display the required measured parameter in the measured value display and press the
<ENTER>
key.

9.6.2 Resetting the system settings

The following system settings can be reset to the default condition:
Setting Default settings
Language English Audio signal on Baud rate 4800 Baud Output format ASCII Decimal separator . Contrast 50 Illumination Auto Shutoff time 1 h Temperature unit °C Stability control on
The resetting of the system settings is done in the menu Storage & config / System / Reset. To open the menu Storage & config in the measured value
display, press the
<ENTER__> key.
ba76141e07 08/2018 77
Page 78
Data storage MultiLab 4010-1W
03.04.2013 07:43:33 pH 7.000 24.8 °C AR +++
ID number: 1 continue
Manual data storage 4 von 500
03.04.2013 08:00
Back

10 Data storage

You can transmit measured values (datasets) to the data storage:
Manual data storage (see section 10.1 MAutomatic data storage at intervals (see section 10.2 A
AGE AT INTERVALS, page 79)
ANUAL DATA STORAGE, page 78)
UTOMATIC DATA STOR-
Each data storing process transmits the current dataset to the USB interface.

10.1 Manual data storage

You can transmit a measurement dataset to the data storage as follows. The dataset is at the same time output to the interface:
1. Press the The menu for manual data storage appears.
<STO> key shortly.
When the storage is
full
78 ba76141e07 08/2018
2. If necessary, change and confirm the ID number (1 ... 10000) with
<><> and <ENTER>.
The dataset is stored. The meter switches to the measured value dis­play.
When all storage locations are occupied, it is not possible to continue storing. Then you can e.g. transmit the stored data to a PC (see section 10.3.1 M
ING THE MEASUREMENT DATA STORAGE, page 81) and subsequently erase the
data storage (see section 10.3.2 E
RASING THE MEASUREMENT DATA STORAGE,
page 82).
ANAG-
Page 79
MultiLab 4010-1W Data storage
ID number 1 Interval 30 s Duration 180 min continue
0d03h00min
0 1d17h33min
Automatic data storage
1
2
3
03.04.2013 08:00
Back

10.2 Automatic data storage at intervals

The storage interval (Interval) determines the time interval between automatic data storage processes. Each data storing process transmits the current data­set to the USB interface.
Configuring the
automatic storing
function
1. Press the
<STO_ > key.
The menu for automatic data storage appears.
1 Specified entire storing duration 2 Max. available storage duration 3 Graphical display of the storage usage
Settings You can configure the automatic data storage function with the following set-
tings:
Menu item Possible setting Explanation
ID number 1 ... 10000 ID number for the dataset series. Interval 1 s, 5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min
Storage interval. The lower limit of the storage interval can be restricted by the number of free storage locations. The upper limit is restricted by the stor­age duration.
Duration 1 min ... x min Storage duration.
Specifies after which time the auto­matic data storage should be termi­nated. The lower limit of the storage duration is restricted by the storage interval. The upper limit is restricted by the number of free storage locations.
ba76141e07 08/2018 79
Page 80
Data storage MultiLab 4010-1W
25.0
7.007
pH
°C
0
14
0d03h00min
1
2
03.04.2013 08:00
Stop automatic storage?
yes no
Warning
03.04.2013 08:00
Starting the auto-
matic storing func-
tion
To start the automatic data storage function, select continue with <><> and confirm with play.
1 Remaining storing duration 2 Graphical display of the storage duration
The active automatic data storage function can be recognized by the progress bar in the status line. The progress bar indicates the remaining storage dura­tion.
<ENTER>. The meter switches to the measured value dis-
Terminating the
automatic storage
function prema-
turely
If the automatic data storage function is activated, only the follow­ing keys are active: and the automatic switch-off function are deactivated.
Proceed as follows to switch off the automatic data storage function before the adjusted storage duration has expired:
1. Press the <STO_ > key.
The following window appears.
<M><STO_ > and <On/Off>. The other keys
80 ba76141e07 08/2018
2. Using The meter switches to the measured value display. The automatic data storage function is terminated.
<><>, select yes and confirm with <ENTER>.
Page 81
MultiLab 4010-1W Data storage

10.3 Measurement data storage

10.3.1 Managing the measurement data storage

For each measurement data storage (automatic and manual) the following functions are available:
DisplayOutput to RS232/USBErase
The storage is edited in the menu Storage & config/ Data storage. To open the Storage & config menu, press the
display.
<ENTER__> key in the measured value
Settings
Open the manual or automatic data storage directly with the
<RCL__> key.
The settings are explained here using the manual data storage as an example. The same settings and functions are available for the automatic data storage.
Menu item Setting/
function
Data storage / Manual data storage
/ Display
- Displays all measurement data-
Explanation
sets page by page. Further options:
Scroll through the datasets with
<><>.
Output the displayed dataset to
the interface with output].
<RCL> or
<F2>/[USB
Quit the display with
[Back].
Data storage / Manual data storage / Output to RS232/ USB
Data storage / Manual data storage
/ Erase
ba76141e07 08/2018 81
- Outputs all stored measurement data to the interface.
- Erases the entire manual mea­surement data storage. Note: All calibration data remain stored when this action is performed.
<F1>/
Page 82
Data storage MultiLab 4010-1W
Manual data storage 3 of 64
03.04.2013 07:43:33 ID numbe r: 1 4110 B092500013
pH 7.000 24.8 °C AR Sensor:
+++
03.04.2013 08:00
Back
USB output
03.04.2013 07:43:33
4010-1W
Ser. no. 09250023 ID number 2 4110
Ser. no. B092500013 pH 6.012 24.8 °C, AR, S: +++ ________________________________________
03.04.2013 07:43:53
4010-1W
Ser. no. 09250013 ID number 2 4110
Ser. no. B092500013 pH 6.012 24.8 °C, AR, S: +++ ________________________________________
Display presentation
of a dataset
Representation of a
dataset
(USB output)
Quitting the display To quit the display of stored measurement datasets, you have the following
options:
Switch directly to the measured value display with Quit the display and move to the next higher menu level with

10.3.2 Erasing the measurement data storage

82 ba76141e07 08/2018
Erasing the measurement data storage (see section 10.3.1 M
MEASUREMENT DATA STORAGE, page 81).

10.3.3 Measurement dataset

A complete dataset consists of:
Date/timeMeter name, series numberSensor name, series numberID number
<M>.
<F1>/[Back].
ANAGING THE
Page 83
MultiLab 4010-1W Data storage
Measured value of the connected sensorMeasured temperature value of the connected sensorAutoRead info: AR appears with the measured value if the AutoRead crite-
rion was met while storing (stable measured value). Otherwise, the AR dis­play is missing.
Calibration evaluation:
- 4 levels (+++, ++, +, -, or no evaluation) or
- QSC (percentage)

10.3.4 Storage locations

The MultiLab 4010-1W meter has two measurement data storages. The measured values recorded either manually or automatic are stored separately in individual measurement data storages.
Data storage Maximum number of datasets
Manual data storage 500 Automatic data storage 4500
ba76141e07 08/2018 83
Page 84
Transmitting data MultiLab 4010-1W

11 Transmitting data

The meter has the following interfaces: USB-B interface (USB Device)
e.g. to connect a PC
Via the USB-B interface (USB Device) you can transmit data to a PC or printer and update the meter software.

11.1 Transmitting data to a PC

Via the USB-B interface (USB Device) you can transmit data to a PC.
PC system require-
ments
Installation of the
USB driver
Connecting a PC
Microsoft Windows
(for details, see enclosed installation CD, Driver directory)
Installed USB driver for the meter (see CD-ROM or Internet)Settings for the USB/RS232 interface on the PC and meter in agreementProgram to receive the measurement data on the PC
(e.g. MultiLab Importer, see CD-ROM or Internet)
1. Insert the supplied installation CD in the CD drive of your PC. or Download the USB driver from the Internet and unpack the files and folders.
2. Start the driver installation (32 bits or 64 bits) suitable for your operat­ing system. Follow the Windows installation instructions as necessary.
1. Connect the MultiLab 4010-1W to the PC via the USB-B interface (USB Device). The meter is listed as a virtual COM interface among the connections in the Windows instrument manager.
Adjusting the set-
tings for the data
transmission
84 ba76141e07 08/2018
2. Set the same transmission data at the meter and PC:
Baud rate: to be selected in the range 1200 ... 19200Set at the PC only:
– Handshake: RTS/CTS – Parity: none – Data bits: 8 – Stop bits: 1
Page 85
MultiLab 4010-1W Transmitting data
Starting the program
for the data
reception
Transmitting data
(options)
3. On the PC, start the program for the data reception, e.g.: MultiLab Importer (see section 11.2 M
ULTILAB IMPORTER, page 85)
Terminal program
Data Control Operation / description
Current measured values of all connected sensors
Manual With
Simultaneously with every manual
automatic, at
With
intervals
<F2>/[USB output].
data storage process (see section 10.1 M
ANUAL DATA STORAGE, page 78).
<F2__>[USB output].
Then you can set the transmission interval.
Simultaneously with every automatic
data storing process (see section 10.2 A
UTOMATIC DATA STORAGE AT INTER-
VALS, page 79).
Stored mea­sured values
Manual Displayed dataset with
put] after calling up from the data stor-
<F2>[USB out-
age.
All datasets with the Output to RS232/
USB function.
(see section 10.3.1 M
MEASUREMENT DATA STORAGE, page
ANAGING THE
81).
calibration records
Manual Calibration record with
output] (see section 5.2.7 C
DATA, page 34; section 7.3.7 CALIBRA- TION DATA, page 53; section 8.3.6 CAL- IBRATION DATA, page 61).
<F2>/[USB
ALIBRATION
automatic At the end of a calibration procedure.
The following rule applies: With the exception of the menus, shortly pressing the
<F2>/[USB output] key generally outputs the display
contents to the interface (displayed measured values, measuring datasets, calibration records).

11.2 MultiLab Importer

With the aid of the MultiLab Importer software, you can record and evaluate measurement data with a PC.
More detailed information can be found in the MultiLab Importer operating manual.
ba76141e07 08/2018 85
Page 86
Transmitting data MultiLab 4010-1W

11.3 BOD Analyst Pro

With the aid of the BOD Analyst Pro software you can administrate BOD measurement on the PC and automatically calculate the measured values.
More detailed information can be found in the BOD Analyst Pro operating manual.
86 ba76141e07 08/2018
Page 87
MultiLab 4010-1W Printer (only MultiLab 4010P-1W)
2
1
3
4

12 Printer (only MultiLab 4010P-1W)

1 Operating LED
– Is illuminated: Printer is switched on
Switch on the printer
– Flashes: Printer is switched off
2 Control panel with the keys 3 Lever to open the paper compartment 4 Lid of the paper compartment
<FEED> and <ON/OFF>

12.1 Commissioning / switching the printer on or off

1. Connect the power pack to the MultiLab 4010P-1W.
The LED lights up green. The printer is ready for operation. or When the printer had been switched off (LED flashing): Switch on the printer with The LED lights up green. The printer is ready for operation.
If there is a USB-B connection at the same time (e.g. to a PC), the data are output to the PC only.
<ON/OFF>.
Switch printer off
ba76141e07 08/2018 87
1. Switch off the printer with <ON/OFF>.
The LED flashes. The printer is switched off.
Page 88
Printer (only MultiLab 4010P-1W) MultiLab 4010-1W

12.2 Operation / printing

Data are only output to the printer if the following requirements are met Data are transmitted manually or automatically (see section 11 T
TING DATA, page 84)
RANSMIT-
The printer is switched on (LED illuminated)There is no USB connection.

12.3 Printer settings

To open the Storage & config menu, press the <F1__>[Menü] key in the mea­sured value display. After completing the settings, switch to the measured value display with
Default settings are printed in bold.
Menu item Possible set-
System / Printer / Font size
<M>.
ting
12x20
8x16 7x16
Explanation
Select the font size for the printer A print sample (font of the printer) to
view the available font sizes can be output with
<ON/OFF_ >.
System / Printer / Print test page
- The printer prints the meter informa­tion from the menu System / Ser- vice information. The current printer settings are used for the print.

12.4 Maintenance

12.4.1 Changing the roll of paper (thermal paper)

1. Pull the lever (3) until the lid (4) of the paper compartment opens.
2. If necessary, remove any remains of the old roll of paper.
3. Insert the new roll of paper. The start of the paper should poke out of the paper compartment.
4. Close the lid (4) of the paper compartment so that it clicks into place.
5. If necessary, move the printer paper forward with
<FEED>.
88 ba76141e07 08/2018
Page 89
MultiLab 4010-1W Printer (only MultiLab 4010P-1W)

12.5 What to do if... / printer

Integrated printer
does not print
Cause Remedy
– Printer switched off
(LED flashing)
– Switch on the printer
(LED illuminated) – No power pack connected – Connect the power pack – USB cable connected – Disconnect the USB cable from the
meter
Printer operating -
paper not being
printed
Integrated printer
prints automatically
– Function "store automatically at
intervals" with long intervals is switched on
– Switch off the function (see section
10.2 A
INTERVALS, page 79)
UTOMATIC DATA STORAGE AT
– No paper available – Insert a roll of paper
Cause Remedy
– Paper inserted the wrong way up – Turn the roll of paper around and
insert it with the other side upwards
Cause Remedy
– The function "store automatically
at intervals" or "transmit data automatically at intervals" is switched on
– Switch off the functions (see sec-
tion 10.2 A
AT INTERVALS, page 79 or
section 11 T
UTOMATIC DATA STORAGE
RANSMITTING DATA,
page 84)
ba76141e07 08/2018 89
Page 90
Maintenance, cleaning, disposal MultiLab 4010-1W
1

13 Maintenance, cleaning, disposal

13.1 Maintenance

13.1.1 General maintenance activities

The only maintenance activity required is replacing the batteries.
See the relevant operating manuals of the IDS sensors for instruc­tions on maintenance.

13.1.2 Replacing the batteries

You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH). In order to charge the batteries, an external charging device is required.
1. Open the battery compartment (1) on the underside of the
meter.
1 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned cor­rectly. The
± signs on the batteries must correspond to the ± signs in
the battery compartment.
2. Place four batteries (type Mignon AA) in the battery compart-
ment.
3. Close the battery compartment (1).
90 ba76141e07 08/2018
Page 91
MultiLab 4010-1W Maintenance, cleaning, disposal
4. Set the date and time
(see section 4.4.5 EXAMPLE 2 ON NAVIGATION: SETTING THE
DATE AND TIME, page 24).
Dispose of used batteries according to the local regulations of your country.
End users within the European Union are obligated to return used batteries (even ecologically compatible ones) to a collection point set up for recycling purposes.
Batteries are marked with the crossed-out waste container symbol. Therefore, they may not be disposed with the domestic waste.

13.2 Cleaning

Occasionally wipe the outside of the measuring instrument with a damp, lint­free cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain sol­vents. Remove any splashes immediately.

13.3 Packing

This meter is sent out in a protective transport packing. We recommend: Keep the packing material. The original packing protects the meter against damage during transport.

13.4 Disposal

At the end of its operational lifetime, the meter must be returned to the disposal or return system statutory in your country. If you have any questions, please contact your supplier.
ba76141e07 08/2018 91
Page 92
What to do if... MultiLab 4010-1W

14 What to do if...

14.1 pH

More information and instructions on cleaning and exchange of sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
The measured value is outside the measuring range.
Cause Remedy
IDS pH sensor: – The measured value is outside
the measuring range of the meter
– Air bubble in front of the junction – Remove air bubble
– Cable broken – Replace the sensor – Gel electrolyte dried out – Replace the sensor
Cause Remedy
IDS pH sensor: – The values determined for zero
point and slope of the IDS pH sensor are outside the allowed limits.
– Use suitable IDS pH sensor
(e.g. sway or stir the solution)
–Recalibrate
– Sensor soiled – Clean the sensor – Sensor broken – Replace the sensor
Buffer solutions:
– The used buffer solutions do not
agree with the set buffer set
– Buffer solutions too old – Use only once.
– Buffer solutions depleted – Change solutions
92 ba76141e07 08/2018
– Set different buffer set or
– Use different buffer solutions
Note the shelf life
Page 93
MultiLab 4010-1W What to do if...
No stable measured
Cause Remedy
value
IDS pH sensor: – pH sensor soiled – Clean the pH sensor
Test sample: – pH value not stable – Measure with air excluded if neces-
sary
– Temperature not stable – Adjust temperature if necessary
IDS pH sensor + test sample: – Conductivity too low – Use suitable IDS pH sensor – Temperature too high – Use suitable IDS pH sensor
Obviously incorrect
measured values
– Organic liquids – Use suitable IDS pH sensor
Cause Remedy
IDS pH sensor: – Sensor unsuitable – Use suitable IDS sensor – Temperature difference
between buffer and test sample
– Adjust temperature of buffer or sam-
ple solutions
too great
– Measurement procedure not
– Follow special procedure
suitable

14.2 Dissolved oxygen

More information and instructions on cleaning and exchange of sensors are given in the documentation of your sensor.
Error message
The measured value is outside the measuring range.
OFL, UFL
Cause Remedy
– Measured value outside the
– Use a suitable IDS D.O. sensor
measuring range
ba76141e07 08/2018 93
Page 94
What to do if... MultiLab 4010-1W
Error message,
Error
Cause Remedy
– Sensor contaminated – Clean the sensor – Measured temperature value
outside the operating conditions
– Keep to the temperature range for
the test sample (display of OFL/UFL instead of a temperature value)
– Defective sensor – Calibration
– Exchange the sensor cap
– Replace the sensor – Calibration failed – Recalibrate – D.O. concentration during zero
point calibration too high.
– Immerse the sensor in an oxygen-
free solution

14.3 Conductivity

More information and instructions on cleaning and exchange of sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
The measured value is outside the measuring range.
Cause Remedy
– Measured value outside the
– Use suitable IDS conductivity sensor
measuring range
Cause Remedy
– Sensor contaminated – Clean the sensor and replace it if
necessary
– Calibration solution not suitable – Check the calibration solutions
More information and instructions on cleaning and exchange of sensors are given in the documentation of your sensor.
94 ba76141e07 08/2018
Page 95
MultiLab 4010-1W What to do if...

14.4 General topics

Sensor symbol
flashes
Display
Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
Cause Remedy
– Calibration interval expired – Recalibrate the measuring system
Cause Remedy
– Batteries almost empty – Replace the batteries (see section
13.1 M
AINTENANCE, page 90)
Cause Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<ENTER> and <On/Off>
key simultaneously
Cause Remedy
– E. g., a question by the service
department
– Switch on the meter.
– Open the menu,
<ENTER__> / Stor-
age & config / System / Service infor­mation. The instrument data are
displayed.
or
– Connect the sensor.
Press softkey [
<F1>Info<F1>] /
[More] The sensor data are displayed (see section 4.1.5 S
ENSOR INFO, page
18)
ba76141e07 08/2018 95
Page 96
Technical data MultiLab 4010-1W

15 Technical data

15.1 Measuring ranges, resolution, accuracy

Measuring ranges, accuracy
* Available only if a D.O. sensor is connected
Variable Measuring range Accuracy
Air pressure (abso­lute)*
Further data are given in the documentation of your sensor.
225 ... 825 mm Hg ± 3 mm Hg

15.2 General data

Dimensions MultiLab 4010-1W approx. 230 x 190 x 80 mm
(9.06 x 7.48 x 3.15 inches)
Weight MultiLab 4010-1W approx. 0.8 kg (1.76 pounds) Mechanical structure Type of protection
(MultiLab 4010-1W)
Electrical safety Protective class III Test certificates CE
IP 43
Ambient conditions Storage -25 °C ... +65 °C
Operation +5 °C ... +55 °C
With the power pack connected: +5 °C ... +40 °C
Admissible relative humidity
Power supply Batteries 4 x 1.5 V alkali-manganese batteries, type
Operational life Approx. 150 h* Power pack Helmsman Industrial Co Ltd SEI0901100P
Primary plugs Primary plugs contained in the scope of
* The operational life is shorter is the display illumination is switched on permanently
Yearly mean: < 75 % 30 days/year: 95 % Other days: 85 %
AA
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 0.5 A Output: 9 Vdc, 1100 mA Connection max. overvoltage category II
ShenZhen RiHuiDa Power Supply Co Ltd RHD10W090110 Input: 100 ... 240 V ~ / 50 ... 60 Hz / 0.4 A Output: 9 Vdc, 1100 mA
delivery: Euro, US, UK and Australian.
96 ba76141e07 08/2018
Page 97
MultiLab 4010-1W Technical data
USB interface (device) Type USB 1.1
USB-B (Device), PC
Baud rate Adjustable:
1200, 2400, 4800, 9600, 19200 Baud Data bits 8 Stop bits 2 Parity None Handshake RTS/CTS Cable length max. 3 m (9.843 feet)
Guidelines and norms used
EMC EC directive 2014/30/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A Meter safety EC directive 2014/35/EC
EN 61010-1 IP protection class EN 60529
ba76141e07 08/2018 97
Page 98
Firmware update MultiLab 4010-1W

16 Firmware update

16.1 Firmware update for the meter MultiLab 4010-1W

You can find available firmware update files for your meter on the Internet. With the "Firmware Update " program and a PC you can update the firmware of the MultiLab 4010-1W to the newest version.
For the update you have to connect the meter to a PC. For the update via the USB-B interface, the following is required:
a free USB interface (virtual COM port) on the PC the driver for the USB interface (on the enclosed CD-ROM)the USB cable (included in the scope of delivery of the MultiLab 4010-1W).
1. Install the downloaded firmware update on a PC. An update folder is created in the Windows start menu. If an update folder already exists for the meter (or meter type), the new data are displayed there.
2. In the windows start menu, open the update folder and start the firm­ware update program for the meter
3. Using the USB interface cable, connect the MultiLab 4010-1W to a USB interface (virtual COM port) of the PC.
4. Switch on the MultiLab 4010-1W.
5. In the firmware update program, start the update process with OK.
6. Follow the instructions of the firmware update program. During the programming process, a corresponding message and a progress bar (in %) are displayed. The programming process takes up to 15 minutes. A terminatory mes­sage is displayed after a successful programming process. The firm­ware update is completed.
7. Disconnect the MultiLab 4010-1W from the PC. The MultiLab 4010-1W is ready for operation again.
After switching the meter off and on you can check whether the meter has taken over the new software version (see Y
VERSION IS IN THE METER OR IDS SENSOR, PAGE 95).
OU WANT TO KNOW WHICH SOFTWARE
98 ba76141e07 08/2018
Page 99
MultiLab 4010-1W Firmware update

16.2 Firmware-Update for IDS Sensors

With the "Firmware Update" program and a PC you can update the firmware of an IDS sensor to the newest version. You can find available firmware update files for your IDS sensor on the Internet.
For updating, use cables to connect the IDS sensor to the MultiLab 4010-1W, and the MultiLab 4010-1W to a PC.
For the update via the USB-B interface, the following is required:
a free USB interface (virtual COM port) on the PC the driver for the USB interface (on the enclosed CD-ROM)the USB cable (included in the scope of delivery of the MultiLab 4010-1W).
1. Install the downloaded firmware update on a PC. An update folder is created in the Windows start menu. If an update folder already exists for the sensor (or sensor type), the new data are displayed there.
2. In the windows start menu, open the update folder and start the firm­ware update program for the IDS sensor
3. Connect the IDS sensor to the MultiLab 4010-1W meter.
4. Using the USB interface cable, connect the MultiLab 4010-1W to a USB interface (virtual COM port) of the PC.
5. Switch on the MultiLab 4010-1W.
6. In the firmware update program, start the update process with OK.
7. Follow the instructions of the firmware update program. During the programming process, a corresponding message and a progress bar (in %) are displayed. The programming process takes up to 5 minutes. A terminatory mes­sage is displayed after a successful programming process. The firm­ware update is completed.
8. Disconnect the MultiLab 4010-1W from the PC. Meter and sensor are ready for operation again.
After switching the meter off and on you can check whether the sensor has taken over the new software version (see Y
VERSION IS IN THE METER OR IDS SENSOR, PAGE 95).
OU WANT TO KNOW WHICH SOFTWARE
ba76141e07 08/2018 99
Page 100
Glossary MultiLab 4010-1W

17 Glossary

pH/ORP
Asymmetry see zero point
Electromotive force of
an electrode
Junction The junction is a porous body in the housing wall of reference elec-
ORP
(U)
pH value The pH value is a measure of the acidic or basic effect of an aqueous
Potentiometry Name of a measuring technique. The signal (depending on the mea-
Slope The slope of a linear calibration function.
The electromotive force U of the combination electrode is the measur­able electromotive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the combination electrode. Its depen­dency on the pH results in the electrode function, which is characterized by the parameters, slope and zero point.
trodes or electrolyte bridges. It arranges the electrical contact between two solutions and makes the electrolyte exchange more difficult. The expression, junction, is also used for ground or junction-less transitions.
The ORP is caused by oxidizing or reducing substances dissolved in water if these substances become effective on an electrode surface (e. g. a gold or platinum surface).
solution. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH value is the value of a pH measurement.
sured parameter) of the electrode is the electrical potential. The electri­cal current remains constant.
Zero point The zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified temperature is zero. Normally, this is at 25 °C.
Conductivity
Cell constant (c) Characteristic quantity of a conductivity measuring cell, depending on
the geometry.
Conductivity (ϰ) Short form of the expression, specific electrical conductivity.
It corresponds to the reciprocal value of the resistivity. It is a measured value of the ability of a substance to conduct an electric current. In water analysis, the electrical conductivity is a dimension for the ionized substances in a solution.
Reference tempera-
ture
Resistivity (ρ) Short name for the specific electrolytic resistance. It corresponds to the
Fixed temperature value to compare temperature-dependent measured values. For conductivity measurements, the measured value is con­verted to a conductivity value at a reference temperature of 20 °C or 25 °C.
reciprocal value of the electrical conductivity.
100 ba76141e07 08/2018
Loading...