YOKOGAWA XL100, XL112, XL111, XL114, XL121 User Manual

...
User's Manual
XL100 Datum-Y
クイックマニュアル QuickSetupManual
IMXL120P
1stEdition:Apr.2007(YMI)
XL111,XL112,XL114  ポータブルデータロガ−(Datum-Y) XL121,XL122,XL124  ポータブルデータステーション(Datum-Y)
XL111,XL112,XL114  PortableDataLogger(Datum-Y) XL121,XL122,XL124  PortableDataStation(Datum-Y)
1
IM XL120P
Introduction
Thank you for purchasing our XL100 Portable Data Station/Portable Data
Logger.
This Quick Setup Manual briefly describes the key operations as well as setting
examples of the XL100 upon actual measurement, so that you can operate the
XL100 for the first time.
In addition to this manual, the User’s Manual and Communication Function
Manual contained in the CD-ROM are available separately. The User’s Manual
provides detailed information regarding all of the functions and operations of the
XL100 excluding the communication functions. The Communication Function
Manual provides information necessary for using communication functions and
creating communication programs. Use them together with this Quick Setup
Manual. The Communication Function Manual is available only for the Portable
Data Station.
After reading this manual, keep it in an easily accessible place for later
reference. This manual will come in handy when you are unsure of how to
operate the product.
Notes
• The contents of this manual are subject to change without prior notice.
• Figures and illustrations representing display views in this manual may differ
from actual views.
• Every effort has been made to ensure accuracy in the preparation of this
manual. However, should any doubts arise or errors come to your attention,
please contact the vendor from which you purchased the product.
• The contents of this manual may not be transcribed or reproduced, in part or
in their entirety, without prior permission.
Trademarks
The company and product names referred to in this document are either
trademarks or registered trademarks of their respective holders.
Revisions
First Edition: April, 2007
1st Edition: April 2007 (YMI)
All Rights Reserved, Copyright © 2005, Yokogawa Meters & Instruments Corporation
2
IM XL120P
Safety Precautions
When operating the instrument, be sure to observe the cautionary notes given
in “Safety Precautions” on pages 4 and 5 and section 3.1, “Handling
Precautions” in the User’s Manual. If you use the instrument in any way other
than as instructed, the instrument’s protective measures may be impaired.
The following safety symbols are used on the instrument and in this manual.
WARNING
Indicates a hazard that may result in the loss of life or serious injury of the user unless the described instruction is abided by.
CAUTION
Indicates a hazard that may result in an injury to the user and/or physical damage to the product or other equipment unless the described instruction is abided by.
Note
Indicates information that is essential for handling the instrument or should be noted in order to familiarize yourself with the instrument’s operating procedures and/or functions.
TIP
Indicates information that complements the present topic.
3
IM XL120P
Contents
Introduction ............................................................................. 1
Safety Precautions.................................................................. 2
1. Checking the Contents of the Package ........................ 4
2. Flow of Operation ........................................................... 5
3. Names and Functions of Parts ...................................... 6
4. How to View the Display ................................................. 9
5. Introduction of the Main Functions............................. 15
Input Type and Calculation .................................................................. 15
Alarm Function ..................................................................................... 16
Saving Data ........................................................................................... 17
Triggers ................................................................................................. 18
File Operations ..................................................................................... 18
Communication Function .................................................................... 19
6. Operation Mode and Basic Key Operations ............... 21
Operation Modes and Switching the Operation Mode ...................... 21
Switching the Display in Free Running Mode or Logging Mode...... 22
Switching the menu in Setting Mode .................................................. 22
Key Operations for Entering Characters ............................................ 23
Key Operations for Entering Values .................................................... 23
7. Signal Wiring ................................................................. 24
8. Connecting to the Power Supply and Turning the
Power Switch ON/OFF .................................................. 26
Connecting the Power Supply ............................................................. 26
Turning the Power Switch ON/OFF ..................................................... 27
9. Setting the Input Channel ............................................ 29
10. Setting the Data Save Operation ................................. 33
11. Confirming the Settings and Performing the Measure-
ment ............................................................................... 38
12. Inserting an External Storage Medium and Saving Data . 39
Inserting an External Storage Medium ............................................... 39
Starting the Data Save Operation ....................................................... 40
Stopping the Data Save Operation ..................................................... 40
13. Analyzing the Saved Data ............................................ 41
Loading the Saved Data File ................................................................ 41
Loading the Measured Data ................................................................. 43
Displaying Statistical Calculation Values ........................................... 43
14. Troubleshooting ............................................................ 44
Index....................................................................................... 45
4
IM XL120P
1. Checking the Contents of the Package
Unpack the box and check the contents before operating the instrument. Should
the product you have received be the wrong model, lack any items, or show any
problems in its appearance, contact the vendor from whom you purchased the
product.
Instrument Main Unit
Check the model and suffix code printed on the nameplate on the rear panel to
ensure that the XL100 is exactly as specified in your purchase order.
XL100
No.
Model
When inquiring about the product to the vender, please also give the vendor this number.
Accessories
Make sure that the package contains all the accessories listed below and that
they are all free from any damage.
These are attached to the XL100.
Terminal block unit
Rubber boot
AC adapter
Quick setup manual (this manual)
CD-ROM
Contains Standard Software and PDF manuals (User's Manual, Quick Setup Manual, and Communication Function Manual).
1.
b
+
-
c h
1
b
+
-
c h
2
b
+
-
c h
3
b
+
-
c h
4
b
+
-
c h
5
b
+
-
c h
6
b
+
-
c h
7
+
-
c h
8
(Ex.: 95052)
Screwdriver for terminals
Side cover (Supplied with the XL121, XL122, and XL124)
TIP
For details on peripherals and spare parts, see page 3 in the User’s Manual.
5
IM XL120P
2. Flow of Operation
Connect the wires
Connect the power supply and turn the power ON
Configure
• Set the input
• Save the data
Set other items (as necessary)
• Alarm
• Display
• Calculation
• Communications
• Hardware
• System
Confirm the settings and measure
Save the data
Free Running Mode
Logging Mode
Set the language
Initial startup only
Analyze the data
• Review Mode
• Logging & Review Mode
Process the file (as necessary)
• Rename the file
• Delete the files
• Copy the data
Reference chapter/ section in the User’s Manual
3.3-3.5
Page 24
Page 26
Page 28
Page 29
Page 33
Page 38
Page 39
Page 41
Reference page in this manual
3.6, 4.1
Chapter 5
9.1
Chapter 6 Chapter 7 Chapter 8 Chapter 10 Chapter 4, 11, etc.
11.6, 11.7
9.1
9.2
9.7
9.8
9.9
6
IM XL120P
3. Names and Functions of Parts
Front Panel
HOME REVIEW FILE
SETTING
HOLD
TIME / DIV
SAVE
RANGE
DISPLAY
GROUP
MARK
SET
ESC
SELECT
START / STOP
Display
Displays measured data, operation status, setup menu, settings, etc.
Terminal block unit
Terminal block unit where probes are connected.
Keys
Operation status LED
POWER : Illuminates when the power is ON CHARGE : Illuminates when the battery is being charged START : Illuminates while logging
Status display section
Data display, setting menu, and setting display section
For the wiring procedure, see page 23.
For the viewing the displays, see page 9.
Keys
HOME REVIEW FILE
SETTING
HOLD
TIME / DIV
SAVE
RANGE
DISPLAY
GROUP
MARK
SET
ESC
SELECT
START / STOP
1
2
3
4
5
6
7
8
9
11
10
12
13
14
15
16
7
IM XL120P
1. HOME Key
Press this key to enable Free Running Mode for measuring instantaneous values (see page 21).
2. REVIEW Key
Press this key to enable Logging & Review Mode in which past measured data can be viewed while logging (see page 21) or enable Review Mode in which saved data can be analyzed (see page 21).
3. FILE Key
Press this key to enable File Operation Mode in which file names can be changed, measured data can be copied, setting data can be saved or loaded, and so on (see page 21).
4. SETTING Key
Press this key to set measurement conditions, conditions for saving measured data, alarm conditions, etc.
5. HOLD Key
Press this key to hold the display so that the measured values are not updated or to release the display. In addition, hold this key down to enable or disable key lock.
6. TIME/DIV Key
Press this key to switch the time axis (the time per grid (division)).
7. ESC Key
Press this key to cancel a key operation.
8. SET Key
Press this key to set settings entered through the keys.
9. START/STOP Key
Press this key to start/stop logging.
10. RANGE Key
Press this key to change the input range or span (scale).
11. DISPLAY Key
Press this key to switch the display in Free Running Mode or Logging Mode (see page
22). Press this key also to switch between marker display and statistical calculation display in Review Mode.
12. SAVE Key
Press this key to manually save or print the measured data or screen data.
13. GROUP Key
Press this key to switch the displayed group of measurement, calculation, and communication input channels.
14. Fast Forward Key
Press this key to move the marker to the left or right by 1 division on the review display (see page 43).
15. MARK Key
Press this key to select a marker to be activated on the review display (see page 43).
16. Arrow/SELECT Key
Press the arrow keys to select items on the display. Press this key also to move the marker to the left or right on the review display (see page 43). Press the SELECT key to confirm a selection.
3. Names and Functions of Parts
8
IM XL120P
Side Panel
For details on connecting the power supply and turning the power ON/OFF, see pages 26 and 27.
For details on how to insert the CF card or SD card, see page 39.
CF card slot
SD card slot*
Trigger input/output terminal
RS-485 connector*
Ethernet port*
Power switch
AC adapter jack
RS-232 connector*
For a description, see section 3.9.
For a description, see section 3.8.
For a description, see section 3.8.
For a description, see section 4.7.
For a description, see section 4.7.
For a description, see section 3.5.
For a description, see section 3.4.
For a description, see section 3.8.
For a description, see section 3.8.
For a description, see section 4.1.
For a description, see section 3.6.
Digital I/O connector
USB port for USB memory
USB port for connecting a PC*
* Supported only on the XL121, XL122, and XL124
USB
SG
USB
RS485 RS232LAN
CF
SD
TRIGGER
ALARM
LOGIC/PULSE
GND
IN
OUT
DC INPUT
POWER
12V
20W MAX
The reference sections are those of the User’s Manual.
Rear Panel
Screwdriver for terminals (accessory)
Press the screwdriver towards the spring (right in this figure) to detach it.
Battery cover
Holds the lithium ion battery (94009) sold separately .
Name plate
3. Names and Functions of Parts
9
IM XL120P
4. How to View the Display
Status Display Section
1
2
3
4
5
6
8
7
1. Operation Mode
Displays the mode: Free Running, Logging, Logging & Review, Setup, or
File Operation.
2. User name
Displays the login user name when the key login function (see section 11.7
in the User’s Manual) is turned ON.
3. Group Name (For the procedure to set groups, see section 7.2 in the
User’s Manual)
Displays the group name of the displayed measurement channel.
4. Alarm Status (For a description of the alarm function, see page 16)
The status is displayed using different icon colors as follows:
Gray: No alarm setting Yellow-green: Alarm setting enabled Red: Alarm activated
5. Alarm Output Status
The status is displayed using different icon colors for each alarm output
channel (1 to 4) as follows:
Gray: No alarm setting Yellow-green: Alarm setting enabled Red: Alarm outputting
6. Date/Time (For the procedure to set the date/time, see section 4.3 in the
User’s Manual)
Displays the year, month, day, hour, minute, and second.
7. Sampling Interval
Displays the sampling interval (measurement/save interval of measured
data) when in Free Running, Logging, or Logging & Review Mode.
8. Various Icons
The following icons are used to display the operation status, interface usage
status, etc.
An icon shown when the data save destination is set to internal memory. The icon blinks when there is access to the internal memory. The icon is gray when the data save destination is not set to internal memory.
10
IM XL120P
An icon shown when the data save destination is set to internal memory and the save mode is set to DIVISION. The icon blinks when there is access to the internal memory.
An icon shown when the data save destination is set to internal memory and the memory full operation is set to REPEAT. The icon blinks when there is access to the internal memory.
An icon shown when the data save destination is set to internal memory and the memory full operation is set to DELETE. The icon blinks when there is access to the internal memory.
An icon shown when the data save destination is set to internal memory, the save mode is set to DIVISION, and the memory full operation is set to REPEAT. The icon blinks when there is access to the internal memory.
An icon shown when the data save destination is set to internal memory, the save mode is set to DIVISION, and the memory full operation is set to DELETE. The icon blinks when there is access to the internal memory.
An icon shown when the data save destination is set to CF card. The icon blinks when there is access to the CF card. The icon is gray when the data save destination is not set to CF card.
An icon shown when the data save destination is set to CF card and the save mode is set to DIVISION. The icon blinks when there is access to the CF card.
An icon shown when the data save destination is set to CF card and the memory full operation is set to REPEAT. The icon blinks when there is access to the CF card.
An icon shown when the data save destination is set to CF card and the memory full operation is set to DELETE. The icon blinks when there is access to the CF card.
An icon shown when the data save destination is set to CF card, the save mode is set to DIVISION, and the memory full operation is set to REPEAT. The icon blinks when there is access to the CF card.
An icon shown when the data save destination is set to CF card, the save mode is set to DIVISION, and the memory full operation is set to DELETE. The icon blinks when there is access to the CF card.
An icon shown when the data save destination is set to SD card. The icon blinks when there is access to the SD card. The icon is gray when the data save destination is not set to SD card. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the data save destination is set to SD card and the save mode is set to DIVISION. The icon blinks when there is access to the SD card. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the data save destination is set to SD card and the memory full operation is set to REPEAT. The icon blinks when there is access to the SD card. (Supported only on the XL121, XL122, and XL124.)
4. How to View the Display
11
IM XL120P
An icon shown when the data save destination is set to SD card and the memory full operation is set to DELETE. The icon blinks when there is access to the SD card. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the data save destination is set to SD card, the save mode is set to DIVISION, and the memory full operation is set to REPEAT. The icon blinks when there is access to the SD card. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the data save destination is set to SD card, the save mode is set to DIVISION, and the memory full operation is set to DELETE. The icon blinks when there is access to the SD card. (Supported only on the XL121, XL122, and XL124.)
An icon shown when there is data saved in the backup memory. The icon blinks when there is access to the backup memory. The icon is gray when there is no data saved to the backup memory.
An icon shown when the interface is set to LAN, LAN/RS-232, or LAN/RS-485. For other cases, the icon is gray. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the interface is set to USB. For other cases, the icon is gray. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the communication protocol is set to Modbus (slave). The icon is gray when set to Modbus (master). (Supported only on the XL121, XL122, and XL124.)
An icon shown when the communication protocol is set to Modbus (master). The icon is gray when set to Modbus (slave). (Supported only on the XL121, XL122, and XL124.)
An icon shown when the printer output is turned ON and the sampling interval is greater than or equal to 1 minute. The icon is gray when the printer output is OFF. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the printer output is turned ON and the sampling interval is less than or equal to 30 seconds. This indicates that only manual print is valid using the SAVE key. (Supported only on the XL121, XL122, and XL124.)
An icon shown when the display update is held. The icon is gray when the display is not held.
An icon shown when key lock is enabled. The icon is gray when key lock is disabled.
An icon shown when the AC adapter is connected.
An icon shown when the AC adapter is not connected, and the XL100 is
running on a battery. Shows the remaining battery power using four levels
(
).
4. How to View the Display
12
IM XL120P
Data Display Section
Waveform & Digital Display
4
1
6
7
3
2
5
10
12
11
8
9
1. Waveform
Waveforms of measured data, calculated data, and communication input data. Waveforms of logic input are shown at the lower section of the screen as shown in the figure above.
2. Time Axis
Displays the time axis (time per grid (division)) specified by the TIME/DIV key.
3. Alarm Line
Displayed with a dotted line at the position of the alarm value of the selected channel (active channel).
4. Scale Upper Limit
Shows the display upper limit of the active channel.
5. Scale Lower Limit
Shows the display lower limit of the active channel.
6. Pen
Displayed at the current value position of each channel. The active channel is shown highlighted in reverse video.
7. Usage Indication Bar of the Storage Media
Displays using a blue bar the amount of space used with respect to the total space on the storage medium that is specified to be the save destination of the measured data.
Total space on the save destination medium
Used space
Free space
8. Grid
The grid can be turned ON/OFF.
4. How to View the Display
13
IM XL120P
9. Elapsed Time
Displays the elapsed time from the start of the logging operation.
10. Digital Display
Displays the current values of the measured data, calculated data, and
communication input data using numeric values.When an alarm is occurring, the
value is shown in red in reverse video.
11. Unit
Displays preset characters such as °C or an arbitrary specified characters (up to 6
characters).
12. Channel No./Tag
Displays the channel number and the specified tag (up to 8 characters). The
active channel is shown highlighted (reverse video).
Other Data Displays
In addition to the waveform & digital display, other displays are available including the
waveform display that does not show numeric values. For a description of the data
displays below, see section 2.3, “Data Display” in the User’s Manual.
• Digital Display
Displays the numeric values of the instantaneous values and statistical
calculation values.
• Bar Graph Display
Displays a bar graph in place of a waveform.
4. How to View the Display
14
IM XL120P
• Review Display
Displays the waveforms of data saved in the past.
• Logging & Review Display
Displays both the waveforms of data currently being logged and the
waveforms of data saved in the past.
• Alarm Summary Display
Displays the alarm status in a list.
• Log Display
Displays the log data of error messages and communications.
4. How to View the Display
15
IM XL120P
5. Introduction of the Main Functions
Input Type and Calculation
As shown in the table below, the available input types are analog input, which includes DC voltage, thermocouple, and RTD, and other inputs, which consist of pulse signal (1 channel) and logic signals (2 channels). In addition, the arithmetic calculations between two inputs can be performed and assigned to a calculation channel and displayed in the same fashion as measured values. The statistics of measured values can also be displayed. For details on the input settings, see chapter 5, “Setting the Input Channels.” For details on calculation, see chapter 8, “Setting the Calculation of Measured Data.”
Input/Calculation Description
DC voltage Measures a DC voltage in the range of ±100 mV to ±50 V.
Thermocouple Selectable from the following types: R, S, B, K, E, J, T, N, W,
L, and U.
RTD Selectable from Pt100 and JPt100 types.
Pulse signal Displays the pulse input as number of revolutions, integrated
value, or instantaneous value.
Logic signal Displays the logic waveform at the lower section of the display by
taking input voltage less than or equal to 0.9 V to be OFF (0) and input voltage greater than or equal to 2.1 V to be ON (1).
Calculation Performs arithmetic calculations using measured data,
calculated data, communication data, and arbitrary assigned constants and displays the result.
Statistical calculation Calculates and displays the maximum, minimum, average,
peak (P-P), or rms value of the measured value.
Voltage
Thermocouple
RTD
Signal input
Displays the measured/calculated data
Pulse/Logic input
TIP
You must connect a digital I/O cable sold separately to the input terminal (digital I/O connector) to apply pulse or logic signals. (See section 3.4 in the User's Manual.)
16
IM XL120P
Alarm Function
Generates an alarm when the measured/calculated value meets a certain
condition. When an alarm occurs, information notifying the alarm occurrence is
displayed on the screen. In addition, an alarm signal can be delivered from the
output terminal (digital I/O connector) on the rear panel of the XL100 by
connecting a digital I/O cable (option). You can select the alarm conditions from
the following table.
For details on the settings, see chapter 6, “Setting Alarms” in the User’s
Manual.
Input Type Setting Alarm Condition
Level OFF Not set alarm conditions. or Pulse Hi An alarm occurs when the measured/calculated
value is greater than or equal to the alarm value.
Lo An alarm occurs when the measured/calculated
value is less than or equal to the alarm value.
Window IN An alarm occurs when the measured/calculated
value is within the lower limits and upper limits of the alarm range.
Window OUT An alarm occurs when the measured/calculated
value is outside the lower limits and upper limits of the alarm range.
Logic OFF Not set alarm conditions.
Hi An alarm occurs when the logic input changes
from low to high.
Lo An alarm occurs when the logic input changes
from high to low.
Alarm value
Alarm release
Measured value
Alarm occurrence
Hi
Lo
Measured value
Alarm value
Alarm occurrence
Alarm value
Alarm release
Measured value
Alarm occurrence
Window IN
Window OUT
Measured value
Alarm release
Alarm value
Alarm occurrence
Alarm release
5. Introduction of the Main Functions
17
IM XL120P
Saving Data
Measured data, calculated data, setting data, and so forth can be saved to the XL100 internal memory or an external storage medium (CF card or SD card) that is inserted in the XL100.
SD
CF
Save the data
Supported only on the XL121, XL122, and XL124
TIP
The XL100 is equipped with a USB port for USB memories. However, data cannot be saved directly to a USB memory (see section 3.9 in the User’s Manual).
The types of data that can be saved are as follows:
Type Description
Logging data The instantaneous values of the measured/calculated data
can be saved at a specified sampling interval. The data save operation is started or stopped with the START/STOP key. The save operation can also be started or stopped when a specific event (see “Trigger” on the next page) occurs. The logging data contains alarm information.
Manual sampled data The measured/calculated data (instantaneous values) of all
channels can be saved by pressing the SAVE key in Free Running Mode.
Alarm data The same information as the alarm summary display can be
saved by pressing the SAVE key during alarm summary display.
Screen image data The image data of the screen being displayed can be saved
by pressing the SAVE key in Free Running Mode, etc.
Setting data The setting data of the XL100 can be saved in File Operation
Mode.
Log data The same information as the log display can be saved by
pressing the SAVE key during log data display.
Backup file If the data save operation is not carried out normally to the
internal memory or external storage medium (CF card or SD card), the data is saved to the backup memory of the XL100. The saved data can be copied to an external storage medium.
5. Introduction of the Main Functions
18
IM XL120P
Triggers
In addition to using the START/STOP key to start or stop the logging, a trigger for starting (or stopping) the save operation of the logging data (measured/ calculated data) can be configured for automatic operation. The trigger for logging can be selected from the list below and configured.
Type Description
None Not set trigger conditions.
External A trigger is activated when a signal is applied to
the external trigger input terminal.
Level High limit (H) A trigger is activated when the measured value is greater
than or equal to the specified value.
Low limit (L) A trigger is activated when the measured value is less than
or equal to the specified value.
Window IN A trigger is activated when the measured value is within the
specified lower and high limits.
Window OUT A trigger is activated when the measured value is outside
the specified lower and high limits.
Alarm A trigger is activated when an alarm occurs on the specified
alarm output channel.
Time A trigger is activated at the specified time.
Timer The time at which the data save operation is stopped can
be specified. Logging is stopped after the specified time elapses.
File Operations
The following file operations are available.
Operation Type Description
Rename Renames files saved on an external storage medium (CF card
or SD card) or internal memory.
Save setting data Saves setting data to an external storage medium (CF card, SD
card, and USB memory), internal memory, or setting memory.
Load setting data Loads the setting data saved on an external storage medium
(CF card, SD card, and USB memory), internal memory, or setting memory and changes the settings.
Copy data Copies the files saved to the an external storage medium (CF
card or SD card), internal memory, or setting memory to an external storage medium (CF card, SD card, or USB memory), internal memory, or setting memory.
Copy backup memory Copies the files saved to the backup memory (memory to which
data is saved when data cannot be saved to an external storage medium or internal memory) to an external storage medium (CF card or SD card) or internal memory.
Format Formats an external storage medium (CF card or SD card),
backup memory, or internal memory.
5. Introduction of the Main Functions
19
IM XL120P
Communication Function (Supported only on the XL121, XL122, and XL124)
FTP Client/Server
The Ethernet interface can be used to automatically transfer measurement data
files to an FTP server connected to the network or access the XL100 from a PC
through FTP to retrieve data on the external storage medium or internal
memory of the XL100.
PC
Ethernet
FTP server
Hub
Hub
Automatically save measured data
Load measured data
FTP client
Ethernet
Hub
Hub
FTP server
Web Server
By configuring the XL100 to be a Web server, the XL100 screen can be shown
on the PC. You can monitor the measured data and switch the display from the
PC.
PC
Ethernet
Hub
Hub
Monitor the measured data
Web server
PC
Monitor the measured data
5. Introduction of the Main Functions
20
IM XL120P
E-mail transmission
An e-mail can be sent automatically from the XL100 when an alarm occurs.
PC
Ethernet
Hub
Hub
Send e-mail
PC
Receive e-mail
Receive e-mail
Serial Communication
The USB interface or serial interface can be used to change the XL100 settings from a PC or retrieve data into the PC through command communication. In addition, Modbus communication is possible on the serial interface. The Modbus master function enables the measured data of a measuring instrument connected as a Modbus slave to be retrieved as communication input data. The data can be assigned to a communication channel and displayed on the XL100 in a similar fashion to measurement and calculation channels.
PC
Change settings and load data
USB, RS-232, RS-485
Command communication
RS-232, RS-485
Modbus communication
Modbus master
Measurement instrument
Modbus slave
For details on the communication functions, see the Communication Function Manual (contained in the CD-ROM).
5. Introduction of the Main Functions
21
IM XL120P
6. Operation Mode and Basic Key Operations
Operation Modes and Switching the Operation Mode
As shown in the figure below, the XL100 has six operation modes: (1) Free
Running Mode in which instantaneous values are measured, (2) Logging Mode
in which continuous measurement is performed while saving data, (3) Logging
& Review Mode in which past measured data can be viewed while the logging
operation is in progress, (4) Review Mode in which saved data is analyzed, (5)
File Operation Mode in which file operations such as saving and loading of the
setup data is performed, and (6) Setting Mode in which various settings such as
the measurement conditions are specified. The keys in the figure below are
used to switch between these modes.
HOME
REVIEW
FILE
SETTING
START / STOP
Review Mode
Setting Mode
Free Running Mode
File Operation Mode
Logging Mode
Logging & Review Mode
Switch the display
Wavef orm
Digital
Bar graph
Alarm summary
Log
Switch the display
Wavef orm
Digital
Bar graph
Alarm summary
Log
Setting menu
Select
Input
Display
Alarm
Calculation
Hardware
Marker measurement
Set
File operation
menu
Select
Rename files
Delete files
Setup file Copy data files
Copy backup memory
Format
Set
REVIEW
Statistical
calculation
System information
System
information
Data Save
Communication
System
Initialize log
Marker
measurement
Statistical calculation
22
IM XL120P
Switching the Display in Free Running Mode or Logging Mode
To switch to a display other than waveform & digital display, press DISPLAY to
show the display switch pop-up menu, select the display using the arrow keys,
and press SELECT.
DISPLAY
SELECT
Display the menu
Select the display
Confirm the display
Switching the menu in Setting Mode
Press the arrow keys to select the desired item, and press SELECT. A
selection list, a setting window, or a setting menu that is one level lower is
displayed. To return to the original setting menu, press ESC.
SELECT
Select the setup item
Confirm the setup item
ESC
6. Operation Mode and Basic Key Operations
23
IM XL120P
Key Operations for Entering Characters
For settings that require characters to be entered, a character entry window
opens as shown below.
To enter a character, press the arrow keys to move the cursor in the character
selection area, select the character, and press SELECT. To set the entered
value, press SET.
Character entry area
Character selection area
Character type selection area
(A: alphabet characters. 0: numeric characters and symbols)
Backspace
Delete
Key Operations for Entering Values
For settings that require a value to be entered, a value entry window opens as
shown below. Press the up and down arrow keys to increment or decrement
the value and the left and right arrow keys to move along the digits. To set the
entered value, press SET. The window will not close when you press SET if a
value outside the range is entered. Enter a value within the range.
6. Operation Mode and Basic Key Operations
24
IM XL120P
7. Signal Wiring
Signal Input Wiring (for Clamp Screws)
Be sure to also read the precautions in section 3.3, “Wiring the Input Signal
Cable” in the User’s Manual when wiring cables.
CAUTION
Do not apply an input exceeding the following values. Otherwise, the XL100 may break down.
• Maximum input voltage
100 mV, 500 mV, and 1 V range and TC input: ±10 VDC
5 V, 10 V, 50 V, and 1-5V/f.s. range: ±60 VDC
• Maximum common mode noise voltage
30 VACrms (50/60 Hz) or ±60 VDC
Wiring Procedure
1. Open the terminal cover of the terminal block unit.
2. Wire the input signal cables to the input terminals.
As shown below, loosen the terminal screws using the screwdriver provided,
insert the signal wires, and fasten the terminal screws.
3. Close the terminal cover of the terminal block unit.
Enlarged view
Accessory screwdriver for terminals
Signal wire
Terminal cover
Terminal block unit
TIP
The terminal block unit can be removed. For the procedure to remove the terminal block unit, see section 3.3, “Wiring the Input Signal Cable” in the User’s Manual.
25
IM XL120P
Wiring Diagram
Use wires of the following specifications.
TC input
DC voltage input
RTD input
+
A
B
b
Lead wire resistance per wire
of 10 or less. Make the
resistance of the three wires equal.
A
b
B
Extension leadwire
DC voltage input
+
+
ch
b
+
ch
b
+
ch
b
DC current input
DC current input
Shunt resistor
Example: For 4 to 20 mA input, use a shunt
resistor of 250 ± 0.1%.
+
+
+
ch
b
Note
• For clamp terminals, use wires of the following specifications.
• Conductive cross-sectional area for single wire: 0.14 mm2 to 2.5 mm2, stranded wire: 0.14 mm2 to 1.5 mm
2
AWG: 26 to 14
• Length of the stripped section of the wire: Approx. 7 mm
Input signal wires whose diameter is 0.3 mm or less may not be securely fastened. Fold over the conducting section of the wire, for example, to make sure that the wire is securely fastened to the clamped terminal.
•RTD input terminals A (+) and B (–) are isolated on each channel. Terminal b is shorted internally across all channels.
Other Wiring
• For a description of the pulse input, logic input, and alarm output wiring, see
section 3.4, “Wiring the Pulse Input, Logic Input, and Signal Cables” in the
User’s Manual.
• For a description of the external trigger input/output wiring, see section 3.5,
“Wiring the External Trigger I/O Signal Cables” in the User’s Manual.
7. Signal Wiring
26
IM XL120P
8. Connecting to the Power Supply and Turning the Power Switch ON/OFF
Connecting the Power Supply
Be sure to also read the precautions in section 3.6, “Connecting the Power
Supply” in the User’s Manual when connecting the power supply.
WARNING
• Use only the power cord and AC adapter supplied by YOKOGAWA Meters & Instruments for the XL100.
• Check that the power source voltage matches the supply voltage rating (100 to 240 VAC), and then connect the power cord.
Connecting the AC Adapter
Follow the steps below to connect the AC adapter.
1. Check that the power switch is OFF.
2. Connect the AC adapter to the AC adapter jack of the XL100.
3. Connect the plug of the power cord supplied with the AC adapter to the
power connector of the AC adapter.
4. Connect the other end of the power cord to the power outlet that meets the
power rating (requirements).
Power supply rating of the AC adapter
Supply voltage rating 100 to 240VAC
Allowable supply voltage range 90 to 264 VAC
Power supply frequency rating 50/60 Hz
2. Connect.
3. Connect.
4. Connect.
1. Check that the power is OFF.
AC adapter
AC adapter jack
Power cord
27
IM XL120P
Turning the Power Switch ON/OFF
The power switch is located on the right side panel of the XL100. To turn the power ON, press the I (ON) side of the power switch. Press the O on the other side to turn the power OFF.
Power switch
Power Switch Operation When the Side Cover Is Attached
(Supported only on the XL121, XL122, and XL124)
You can operate the power switch with the rubber boot and side cover attached by opening the power switch cover. When attaching the accessory side cover, fasten the side cover attachment screw to fix the cover to the XL100.
Side cover
Power switch cover
Side cover attachment screw
Rubber boot
Open from this end.
Display at Power ON
When the power is turned ON, the XL100 shows the startup screen followed by the self-test screen. When the self-test completes normally, the XL100 shows “Self Test OK” followed by the Waveform & Digital display of Free Running Mode. For the corrective action when an error message is displayed, see section 4.1, “Turning ON/OFF the Power Switch” in the User’s Manual.
8. Connecting to the Power Supply and Turning the Power Switch ON/OFF
28
IM XL120P
Language Setting at Startup
When you start up the XL100 for the first time (the first time you turn on the
XL100 after purchase), you must set the language that you are going to use.
Follow the procedure below to set the language.
Once you set the language, the XL100 will start up using the specified language
the next time.
1. When you turn the power switch ON, the following screen appears.
2. Use the arrow keys to select the language, and press SELECT.
3. The language is set, and a self-test starts automatically.
TIP
To change the language once you set it, press SETTING, select HARDWARE > LANGUAGE, and change the setting.
8. Connecting to the Power Supply and Turning the Power Switch ON/OFF
29
IM XL120P
9. Setting the Input Channel
This chapter explains the steps to set the input channel using an example in which a
thermocouple (type E, measurement range: 0.0 to 1000.0°C) is input to CH1. The
settings of other input channels are not changed from the default settings.
For details on setting the input channels, see chapter 5, “Setting the Input
Channels” in the User’s Manual.
Keys
Display
1.
2.
3.
4.
Move the cursor to CH01.
SELECT
SETTING
Change to Setting Mode.
Show the INPUT menu.
SELECT
Show the ANALOG INPUT setting screen.
30
IM XL120P
SELECT
Keys
Display
SELECT
5.
6.
7.
8.
Show the CH01 setting screen.
Show the MODE selection list.
Move the cursor to TC.
9.
SELECT
Select TC.
Move the cursor to MODE.
9. Setting the Input Channel
31
IM XL120P
SELECT
Keys
Display
10.
11.
12.
Move the cursor to RANGE.
Show the RANGE selection list.
Move the cursor to E.
13.
SELECT
Select E.
14.
Move the cursor to SPAN LOWER.
9. Setting the Input Channel
32
IM XL120P
Keys
Display
SELECT
15.
16.
Show the SPAN LOWER entry window.
Enter 0.0.
17.
SET
Set SPAN LOWER settings.
• To perform measurement:
• To configure other settings:
HOME
SETTING
18.
SET
The changed settings are confirmed.
9. Setting the Input Channel
33
IM XL120P
10. Setting the Data Save Operation
This chapter explains the steps to save measured data using an example in which the sampling interval is set to 1 min, the data save destination is set to CF card, and the end trigger is set to timer (seven days later). The settings for saving other measured data are not changed from the default settings. For details on the settings for saving measured data, see section 9.1, “Setting the Save Operation of Measured and Calculated Data” in the User’s Manual.
SELECT
Keys
Display
1.
3.
4.
2.
Move the cursor to DATA SAVE.
SETTING
Change to Setting Mode.
Show the DATA SAVE setting screen.
Show the SAMPLING INTERVAL selection list.
SELECT
34
IM XL120P
5.
SELECT
6.
7.
Move the cursor to 1 min.
Select 1 min.
Move the cursor to SAVE MEDIA.
8.
SELECT
Show the SAVE MEDIA selection list.
9.
Move the cursor to CF CARD.
Keys
Display
10. Setting the Data Save Operation
35
IM XL120P
SELECT
10.
11.
Confirm CF CARD.
Move the cursor to TRIGGER.
12.
SELECT
Show the TRIGGER setting screen.
13.
Move the cursor to END TRIGGER.
Keys
Display
SELECT
14.
Show the END TRIGGER setting screen.
10. Setting the Data Save Operation
36
IM XL120P
15.
16.
SELECT
Show the TRIGGER TYPE selection list.
17.
Move the cursor to TIMER.
Keys
Display
SELECT
18.
Select TIMER.
19.
Move the cursor to TIMER.
SELECT
Show the TIMER SETTINGS window.
10. Setting the Data Save Operation
37
IM XL120P
20.
21.
Enter 0168:00:00.
Keys
Display
SET
Set END TRIGGER settings.
• To perform measurement:
• To configure other settings:
HOME
SETTING
22.
SET
The changed settings are confirmed.
10. Setting the Data Save Operation
38
IM XL120P
11. Confirming the Settings and Performing the Measurement
When you are done with the settings, press
HOME
to switch to Free Running
Mode and check the settings.
The figure below shows the display that appears when the XL100 is configured
as explained in section chapter 9, “Setting the Input Channel” and 10, “Setting
the Data Save Operation.”
Check whether the displayed values are correct. If the measured values are not
correct, switch back to Setting Mode, and check that the input settings are
correct. If the measured values are not correct even though the input settings
are correct, see chapter 14, “Troubleshooting.”
If the CF Card icon is gray, the save destination is not set to the CF card.
Sampling interval
CF card icon
Measured value
Scale upper limit
Scale lower limit
TIP
• To switch to a display other than waveform & digital display, press DISPLAY to show the display switch pop-up menu, select the display using the arrow keys, and press SELECT. (The steps are explained in page 22.)
• Press TIME/DIV to switch the time axis (the time per grid (division)).
• Press RANGE to change the input range or span (scale).
39
IM XL120P
12. Inserting an External Storage Medium and Saving Data
Inserting an External Storage Medium
CF cards (Type II) and SD cards can be used on the XL100 as external storage
media.
For the handling precautions of the external storage media, the estimated
amount of stored data, and other information, see section 4.7, “Inserting and
Removing the External Storage Media” in the User’s Manual.
Inserting or Removing a CF Card
Insert the CF card firmly into the CF card slot on the side panel of the XL100.
To remove the CF card, press the CF card eject button to the left of the CF card
slot, and pull the CF card out.
CF card eject button
CF card slot
CF card
Inserting or Removing a SD Card (Supported only on the XL121,
XL122, and XL124)
Insert the SD card firmly into the SD card slot on the side panel of the XL100.
To remove the SD card, press the SD card, and then pull it out.
SD card slot
SD card
40
IM XL120P
Starting the Data Save Operation
To start the data save (logging) operation, press
START / STOP
. When the data save
operation starts, the START LED illuminates. If a start trigger (see page 18) is
specified, the logging operation starts when the trigger condition is met. The
START LED blinks until the trigger condition is met.
HOME
REVI
TIME
RANGE
START
/ STOP
START/STOP key
START LED
Usage Indication of the External Storage Media
The amount of space used is displayed using a blue bar with respect to the total
space on the storage medium that is specified to be the save destination of the
measured data.
Total space on the save destination medium
Used space
Free space
Stopping the Data Save Operation
Press
START / STOP
. If a stop trigger (see page 18) is specified, the logging operation
stops automatically when the trigger condition is met.
TIP
•A file name “YYMMDDhhmmss.DLO” (YY: year, MM: month, DD: day, hh: hour, mm: minute, ss: second) is automatically assigned to measurement data files (“YYMMDDhhmmss.CSV” if the data type is set to ASCII). You can also assign an arbitrary file name. For the procedure to assign an arbitrary file name, see section 9.1, “Setting the Save Operation of Measured and Calculated Data” in the User’s Manual.
• The measured/calculated data (instantaneous values) can be saved by pressing SAVE in Free Running Mode. For the procedure to save the data manually, see section 9.3, “Manually Saving Measured and Calculated Data” in the User’s Manual.
• The saved data file can be renamed, deleted, or copied to another external storage medium in File Operation Mode. For these operations, see sections 9.7 to 9.9 in the User’s Manual.
• If the data fails to be written to the save destination for some reason, the data is saved to the backup memory. For details on the backup memory, see section 9.1, “Manually Saving Measured and Calculated Data” in the User’s Manual.
12. Inserting an External Storage Medium and Saving Data
41
IM XL120P
13. Analyzing the Saved Data
Loading the Saved Data File
The steps to load measured data saved to a CF card is given below.
TIP
If you press REVIEW in Logging Mode, the XL100 switches to Logging & Review Mode (see page 21).
SELECT
Keys
Display
1.
3.
4.
2.
Change to Review Mode.
Show the LOAD FILE window.
Show the LOAD FILE selection list.
SELECT
REVIEW
Move the cursor to CF CARD.
42
IM XL120P
5.
6.
7.
Move the cursor to the file shown in the FILE area.
Select the file.
Keys
Display
SET
The file is loaded, and the data is displayed as shown in the figure on the right.
Select CF CARD.
SELECT
13. Analyzing the Saved Data
43
IM XL120P
Loading the Measured Data
Two markers (three markers including marker ALM when an alarm is activated) are shown in Review Mode. A measured value at an arbitrary position can be read by moving the markers horizontally.
Active maker
Marker A
Time difference between the markers
Marker B
Measured value at marker A
Measured value at marker B
Difference in the measured values or calculated values
Time at the active marker position
Selecting the Marker
Press
MAR
K
to select marker A and B alternately. The selected marker (active
marker) is displayed with a thick line.
Moving the Marker
Press the left or right arrow key. Hold the key down to move the marker quickly.
Moving the Marker by One Division
Press (fast forward key) to move the active marker to the left or right
by one division.
Displaying Statistical Calculation Values
Press
DISPLA
Y
in Review Mode to display the maximum, minimum, average, peak
(P-P), and rms values from the start to the end of the logging operation. Press
SE
T
to perform statistical calculation between markers A and B. Press
DISPLA
Y
again to return to the marker display.
Statistical calculation values
13. Analyzing the Saved Data
44
IM XL120P
14. Troubleshooting
The table below lists the major symptoms and their corrective actions. For the
procedure to check the items under “Things to Check”, see the referenced
section written in section 12.1, “Troubleshooting” in the User’s Manual. For the
corrective action for error messages, see the referenced section written in
section 12.2, “Messages and Their Corrective Actions” in the User’s Manual.
Symptom Things to Check
Nothing appears even when For AC power operation the power is turned ON. • Check that the power cord is properly connected to the
power outlet.
• Check that the power supply is within the allowed supply voltage range.
For battery operation
• Check that the battery is loaded correctly.
• Check that the battery is charged adequately.
The measurement display • Check that noise is not riding on the input is odd. signal.
• Check that the wires are correctly connected.
• Check that the ambient temperature and humidity are within the allowed specifications.
Keys do not work. Check that the key lock ( ) is not shown at the upper
right corner of the display.
Unable to save/load from the • Turn the power switch OFF and then back internal memory. ON. It may be restored by the power-on self-test.
• There may have been a power problem while the internal memory was being accessed. Format the internal memory in File Operation Mode. Note that the data saved in the memory will be lost in the process.
Unable to save/load from the • Check that the external storage medium is correctly external storage medium. inserted.
• Check that the external storage medium is formatted.
• Check that there is sufficient free space on the external storage medium.
Unable to set or control the • Check that the communication parameters are matched. instrument using • Check that the specifications of the cable is suitable for communication commands. the application.
• Check that the electrical specifications are correct.
Unable to print. • Check that the printer is turned ON.
• Check that the specifications of the connection cable are correct.
• Check that the cable is correctly connected.
• Check that the communication parameters on the XL100 and printer are matched.
• Check that the chart is loaded correctly in the printer.
45
IM XL120P
Index
E
e-mail transmission 20 elapsed time 13 ESC key 7 external storage medium 39 external trigger 18
F
fast forward key 7, 43 FILE key 7, 21 File Operation Mode 21 file operations 18 format 18 Free Running Mode 21 front panel 6 FTP server 19
G
grid 12 GROUP key 7 group name 9
H
hold 11 HOLD key 7 HOME key 7, 21, 38
I
icons 9 input settings 29 input type 15 internal memory 9, 17, 18
K
key lock 11 keys 6
A
AC adapter 11, 26 accessories 4 active channel 12 active marker 43 alarm 16, 18 alarm data 17 alarm function 16 alarm line 12 alarm output status 9 alarm status 9 alarm summary display 14 arrow keys 7
B
backup file 17 backup memory 11 bar graph display 13 battery 8, 11 battery, remaining power 11
C
CF card 10, 17, 39 channel No. 13 character entry 23 CHARGE LED 6 communication function 19 copy backup memory 18 copy data 18
D
data save operation, setting of 33 data save operation, starting of 40 data save operation, stopping of 40 data, saving of 17 date/time 9 DC current 25 DC voltage 15, 25 differential calculation 15 digital display 13 DISPLAY key 7, 22, 43 display switching 22
46
IM XL120P
L
level 18 load setting data 18 log data 17 log display 14 logging & review display 14 Logging & Review Mode 21 logging data 17 Logging Mode 21 logic 15
M
manual print 11 manual sampled data 17 MARK key 7, 43 marker 43 measured data, loading of 43 menu switching 22 Modbus 11, 20 model 4
O
operation mode 9, 21
P
Pen 12 POWER LED 6 power supply, connection of 26 power switch, turning ON/OFF of 27 printer 11 pulse 15
R
RANGE key 7 rear panel 8 rename 18 review display 14 REVIEW key 7, 21, 41 Review Mode 21, 43 RS-232 8, 20 RS-485 8, 20 RTD 15, 25
S
sampling interval 9, 33 SAVE key 7 save setting data 18 saved data file, loading of 41 scale 12 screen image data 17 SD card 10, 39 SELECT key 7 serial communication 20 SET key 7 setting data 17 SETTING key 7, 21, 29, 33 Setting Mode 21 side panel 8 signal input wiring 24 START LED 6, 40 START/STOP key 7, 21, 40 Statistical Calculation 15 statistical calculation 43 status display section 6, 9 storage media, usage indication bar of 12
T
tag 13 terminal block unit 6 terminal cover 24 thermocouple 15, 25, 29 time 18 time axis 12 TIME/DIV key 7 timer 18, 33 triggers 18
U
unit 13 usage indication of the external storage media 40 USB 20 user name 9
V
values, entry of 23
Index
47
IM XL120P
W
waveform 12 Web server 19
Index
48
IM XL120P
Memo