YOKOGAWA EJX Guide

Communicating Pressure Transmitters
Basic Programming
Yokogawa Corporation of America
2 Dart Road
Newnan, GA 30265
770.254.0400
PG25 Edition 1 xM, December 2003
Error Codes
Integral Display
AL.01 CAP ERROR REPLACE CAPSULE AL.38 SP LOW STATIC PRESSURE EXCEEDS RANGE AL.02 AMP ERR REPLACE AMPLIFIER AL.39 TMP HI TEMP EXCEEDS LIMIT-HIGH AL.10 PRESS CHECK INPUT RANGE AL.40 TMP LO TEMP EXCEEDS LIMIT-LOW AL.11 ST PRESS STATIC PRESSURE>SPECIFIC RANGE AL.50 P LRV ILLEGAL LRV AL.12 CAP TEMP PROCESS TEMP TOO HI OR LOW AL.51 P URV ILLEGAL HRV AL.13 AMP TMP AMBIENT TEMP TOO HI OR LOW AL.52 P SPN VALUE OUTSIDE TRANSMITTER RANGE AL.30 RANGE CHANGE INPUT OR RANGE SETTING AL.53 P ADJ VALUE OUTSIDE TRANSMITTER RANGE AL.31 SP RNG CHANGE INPUT OR RANGE SETTING AL.54 SP RNG ILLEGAL LRV AL.35 P HI CHECK INPUT PRESSURE-TO HIGH AL.55 SP ADJ CHANGE SPAN VALUE AL.36 P LOW CHECK INPUT PRESSURE-TO LOW AL.60 SC CFG CONFIGURATION ERROR-CHECK SETTINGS AL.37 SP HI STATIC PRESSURE EXCEEDS RANGE AL.79 OV DISP DISPLAYED VALUES EXCEED LIMIT
Action or Cause
Integral Display
Action or Cause