DPharp EJA Series Transmitter Section ........................CMPL 01C22A01-02E
Model EJA115 Low Flow Transmitter .............................CMPL 01C22K01-01E
REVISION RECORD
iii
IM 01C22K01-01E
1.INTRODUCTION
1. INTRODUCTION
Thank you for purchasing the DPharp electronic
pressure transmitter.
The DPharp Pressure Transmitters are precisely
calibrated at the factory before shipment. To ensure
correct and efficient use of the instrument, please read
this manual thoroughly and fully understand how to
operate the instrument before operating it.
Regarding This Manual
• This manual should be passed on to the end user.
• The contents of this manual are subject to change
without prior notice.
•All rights reserved. No part of this manual may be
reproduced in any form without Yokogawa’s written
permission.
• Yokogawa makes no warranty of any kind with
regard to this manual, including, but not limited to,
implied warranty of merchantability and fitness for a
particular purpose.
• If any question arises or errors are found, or if any
information is missing from this manual, please
inform the nearest Yokogawa sales office.
• The specifications covered by this manual are
limited to those for the standard type under the
specified model number break-down and do not
cover custom-made instruments.
• Please note that changes in the specifications,
construction, or component parts of the instrument
may not immediately be reflected in this manual at
the time of change, provided that postponement of
revisions will not cause difficulty to the user from a
functional or performance standpoint.
• The following safety symbol marks are used in this
manual:
WARNING
Indicates a potentially hazardous situation which,
if not avoided,
injury.
CAUTION
Indicates a potentially hazardous situation which,
if not avoided, may result in minor or moderate
injury. It may also be used to alert against
unsafe practices
IMPORTANT
Indicates that operating the hardware or software
in this manner may damage it or lead to system
failure.
NOTE
Draws attention to information essential for
understanding the operation and features.
Direct current
could
result in death or serious
.
• Yokogawa assumes no responsibilities for this
product except as stated in the warranty.
• If the customer or any third party is harmed by the
use of this product, Yokogawa assumes no responsibility for any such harm owing to any defects in the
product which were not predictable, or for any
indirect damages.
NOTE
For FOUNDATION FieldbusTM, PROFIBUS PA and
HART protocol versions, please refer to IM
01C22T02-01E, IM 01C22T03-00E and IM
01C22T01-01E respectively, in addition to this
manual.
1.1 For Safe Use of Product
For the protection and safety of the operator and the
instrument or the system including the instrument,
please be sure to follow the instructions on safety
described in this manual when handling this instrument. In case the instrument is handled in contradiction
to these instructions, Yokogawa does not guarantee
safety. Please give your attention to the followings.
(a) Installation
• The instrument must be installed by an expert
engineer or a skilled personnel. The procedures
described about INSTALLATION are not permitted
for operators.
1-1
IM 01C22K01-01E
1. INTRODUCTION
• In case of high process temperature, care should be
taken not to burn yourself because the surface of
body and case reaches a high temperature.
• The instrument installed in the process is under
pressure. Never loosen the process connector bolts to
avoid the dangerous spouting of process fluid.
• During draining condensate from the pressuredetector section, take appropriate care to avoid
contact with the skin, eyes or body, or inhalation of
vapors, if the accumulated process fluid may be
toxic or otherwise harmful.
• When removing the instrument from hazardous
processes, avoid contact with the fluid and the
interior of the meter.
• All installation shall comply with local installation
requirement and local electrical code.
(b) Wiring
• The instrument must be installed by an expert
engineer or a skilled personnel. The procedures
described about WIRING are not permitted for
operators.
(f) Modification
• Yokogawa will not be liable for malfunctions or
damage resulting from any modification made to this
instrument by the customer.
1.2 Warranty
• The warranty shall cover the period noted on the
quotation presented to the purchaser at the time of
purchase. Problems occurred during the warranty
period shall basically be repaired free of charge.
• In case of problems, the customer should contact the
Yokogawa representative from which the instrument
was purchased, or the nearest Yokogawa office.
• If a problem arises with this instrument, please
inform us of the nature of the problem and the
circumstances under which it developed, including
the model specification and serial number. Any
diagrams, data and other information you can
include in your communication will also be helpful.
•Responsible party for repair cost for the problems
shall be determined by Yokogawa based on our
investigation.
• Please confirm that voltages between the power
supply and the instrument before connecting the
power cables and that the cables are not powered
before connecting.
(c) Operation
•Wait 10 min. after power is turned off, before
opening the covers.
(d) Maintenance
• Please do not carry out except being written to a
maintenance descriptions. When these procedures
are needed, please contact nearest YOKOGAWA
office.
• Care should be taken to prevent the build up of drift,
dust or other material on the display glass and
name plate. In case of its maintenance, soft and dry
cloth is used.
(e) Explosion Protected Type Instrument
• Users of explosion proof instruments should refer
first to section 2.9 (Installation of an Explosion
Protected Instrument) of this manual.
• The Purchaser shall bear the responsibility for repair
costs, even during the warranty period, if the
malfunction is due to:
- Improper and/or inadequate maintenance by the
purchaser.
- Failure or damage due to improper handling, use or
storage which is out of design conditions.
- Use of the product in question in a location not
conforming to the standards specified by
Yokogawa, or due to improper maintenance of the
installation location.
- Failure or damage due to modification or repair by
any party except Yokogawa or an approved
representative of Yokogawa.
- Malfunction or damage from improper relocation
of the product in question after delivery.
- Reason of force majeure such as fires, earthquakes,
storms/floods, thunder/lightening, or other natural
disasters, or disturbances, riots, warfare, or
radioactive contamination.
•The use of this instrument is restricted to those who
have received appropriate training in the device.
• Take care not to create sparks when accessing the
instrument or peripheral devices in a hazardous
location.
1-2
IM 01C22K01-01E
1. INTRODUCTION
1.3 ATEX Documentation
This procedure is only applicable to the countries in
European Union.
GB
All instruction manuals for ATEX Ex related products
are available in English, German and French. Should
you require Ex related instructions in your local
language, you are to contact your nearest Yokogawa
office or representative.
DK
Alle brugervejledninger for produkter relateret til
ATEX Ex er tilgængelige på engelsk, tysk og fransk.
Skulle De ønske yderligere oplysninger om håndtering
af Ex produkter på eget sprog, kan De rette
henvendelse herom til den nærmeste Yokogawa
afdeling eller forhandler.
I
Tutti i manuali operativi di prodotti ATEX
contrassegnati con Ex sono disponibili in inglese,
tedesco e francese. Se si desidera ricevere i manuali
operativi di prodotti Ex in lingua locale, mettersi in
contatto con l’ufficio Yokogawa più vicino o con un
rappresentante.
E
Todos los manuales de instrucciones para los productos
antiexplosivos de ATEX están disponibles en inglés,
alemán y francés. Si desea solicitar las instrucciones de
estos artículos antiexplosivos en su idioma local,
deberá ponerse en contacto con la oficina o el
representante de Yokogawa más cercano.
NL
SF
Kaikkien ATEX Ex -tyyppisten tuotteiden käyttöhjeet
ovat saatavilla englannin-, saksan- ja ranskankielisinä.
Mikäli tarvitsette Ex -tyyppisten tuotteiden ohjeita
omalla paikallisella kielellännne, ottakaa yhteyttä
lähimpään Yokogawa-toimistoon tai -edustajaan.
P
Todos os manuais de instruções referentes aos produtos
Ex da ATEX estão disponíveis em Inglês, Alemão e
Francês. Se necessitar de instruções na sua língua
relacionadas com produtos Ex, deverá entrar em
contacto com a delegação mais próxima ou com um
representante da Yokogawa.
F
Tous les manuels d’instruction des produits ATEX Ex
sont disponibles en langue anglaise, allemande et
française. Si vous nécessitez des instructions relatives
aux produits Ex dans votre langue, veuillez bien
contacter votre représentant Yokogawa le plus proche.
D
Alle Betriebsanleitungen für ATEX Ex bezogene
Produkte stehen in den Sprachen Englisch, Deutsch
und Französisch zur Verfügung. Sollten Sie die
Betriebsanleitungen für Ex-Produkte in Ihrer
Landessprache benötigen, setzen Sie sich bitte mit
Ihrem örtlichen Yokogawa-Vertreter in Verbindung.
S
Alla instruktionsböcker för ATEX Ex (explosionssäkra)
produkter är tillgängliga på engelska, tyska och
franska. Om Ni behöver instruktioner för dessa
explosionssäkra produkter på annat språk, skall Ni
kontakta närmaste Yokogawakontor eller representant.
Alle handleidingen voor producten die te maken
hebben met ATEX explosiebeveiliging (Ex) zijn
verkrijgbaar in het Engels, Duits en Frans. Neem,
indien u aanwijzingen op het gebied van
explosiebeveiliging nodig hebt in uw eigen taal, contact
op met de dichtstbijzijnde vestiging van Yokogawa of
met een vertegenwoordiger.
GR
ATEX Ex
, .
Ex
Yo ko ga w a .
1-3
IM 01C22K01-01E
1. INTRODUCTION
SK
CZ
LT
PL
SLO
H
LV
EST
BG
RO
M
1-4
IM 01C22K01-01E
2.HANDLING CAUTIONS
2. HANDLING CAUTIONS
This chapter describes important cautions regarding
how to handle the transmitter. Read carefully before
using the transmitter.
The EJA Series pressure transmitters are thoroughly
tested at the factory before shipment. When the
transmitter is delivered, visually check them to make
sure that no damage occurred during shipment.
Also check that all transmitter mounting hardware
shown in Figure 2.1 is included. If the transmitter was
ordered without the mounting bracket, the transmitter
mounting hardware is not included. After checking the
transmitter, repack it in the way it was delivered until
installation.
Mounting bracket
U-bolt nut
(L type)
U-bolt
2.1 Model and Specifications
Check
The model name and specifications are indicated on the
name plate attached to the case. If the reverse operat-
ing mode was ordered (reverse signal), ‘REVERSE’
will be inscribed in field *1; if square root display
mode was ordered, ‘SQRT’ is inscribed in field *2.
: Refer to USER'S MANUAL
F0202.EPS
Figure 2.2 Name Plate Example of TIIS Flameproof Type
2.2 Unpacking
When moving the transmitter to the installation site,
keep it in its original packaging. Then, unpack the
transmitter there to avoid damage on the way.
2.3 Storage
Transmitter
mounting bolt
U-bolt nut
Mounting bracket
(Flat type)
Figure 2.1 Transmitter Mounting Hardware
Transmitter
mounting bolt
U-bolt
F0201.EPS
The following precautions must be observed when
storing the instrument, especially for a long period.
(a) Select a storage area which meets the following
conditions:
• It is not exposed to rain or water.
• It suffers minimum vibration and shock.
• It has an ambient temperature and relative
humidity within the following ranges.
Ambient temperature:
–40 to 85°C without integral indicator
–30 to 80°C with integral indicator
Relative humidity:
5% to 100% R.H. (at 40°C)
Preferred temperature and humidity:
approx. 25°C and 65% R.H.
(b) When storing the transmitter, repack it as nearly as
possible to the way it was packed when delivered
from the factory.
(c) If storing a transmitter that has been used, thor-
oughly clean the chambers inside the cover flanges
and integral flow orifice unit, so that no measured
fluid remains in it. Also make sure before storing
that the pressure-detector and transmitter section are
securely mounted.
2-1
IM 01C22K01-01E
2. HANDLING CAUTIONS
2.4 Selecting the Installation
Location
The transmitter is designed to withstand severe
environmental conditions. However, to ensure stable
and accurate operation for years, observe the following
precautions when selecting an installation location.
(a) Ambient Temperature
Avoid locations subject to wide temperature
variations or a significant temperature gradient. If
the location is exposed to radiant heat from plant
equipments, provide adequate thermal insulation
and/or ventilation.
(b) Ambient Atmosphere
Avoid installing the transmitter in a corrosive
atmosphere. If the transmitter must be installed in a
corrosive atmosphere, there must be adequate
ventilation as well as measures to prevent intrusion
or stagnation of rain water in conduits.
(c) Shock and Vibration
Select an installation site suffering minimum shock
and vibration (although the transmitter is designed
to be relatively resistant to shock and vibration).
(d) Installation of Explosion-protected Transmitters
Explosion-protected transmitters can be installed in
hazardous areas according to the types of gases for
which they are certified. See Subsection 2.9
“Installation of Explosion Protected Type Transmitters.”
2.5 Pressure Connection
WARNING
• Instrument installed in the process is under
pressure. Never loosen the process connector
bolts to avoid the dangerous spouting of
process fluid.
• During draining condensate from the pressuredetector section, take appropriate care to avoid
contact with the skin, eyes or body, or inhalation of vapors, if the accumulated process fluid
may be toxic or otherwise harmful.
The following precautions must be observed in order to
safely operate the transmitter under pressure.
(a) Make sure that the four manifold bolts are tightened
firmly.
(b) Make sure that there are no leaks in the impulse
piping.
(c) Never apply a pressure higher than the specified
maximum working pressure.
2.6 Waterproofing of Cable
Conduit Connections
Apply a non-hardening sealant to the threads to
waterproof the transmitter cable conduit connections.
(See Figure 6.4.2a, 6.4.2b and 6.4.2d.)
2.7 Restrictions on Use of Radio
Transceiver
IMPORTANT
Although the transmitter has been designed to
resist high frequency electrical noise, if a radio
transceiver is used near the transmitter or its
external wiring, the transmitter may be affected
by high frequency noise pickup. To test for such
effects, bring the transceiver in use slowly from a
distance of several meters from the transmitter,
and observe the measurement loop for noise
effects. Thereafter, always use the transceiver
outside the area affected by noise.
2.8 Insulation Resistance and
Dielectric Strength Test
Since the transmitter has undergone insulation resistance and dielectric strength tests at the factory before
shipment, normally these tests are not required.
However, if required, observe the following precautions in the test procedures.
(a) Do not perform such tests more frequently than is
absolutely necessary. Even test voltages that do not
cause visible damage to the insulation may degrade
the insulation and reduce safety margins.
(b) Never apply a voltage exceeding 500 V DC (100 V
DC with an internal lightning protector) for the
insulation resistance test, nor a voltage exceeding
500 V AC (100 V AC with an internal lightning
protector) for the dielectric strength test.
(c) Before conducting these tests, disconnect all signal
lines from the transmitter terminals. Perform the
tests in the following procedure:
• Insulation Resistance Test
1) Short-circuit the + and – SUPPLY terminals in the
terminal box.
2) Turn OFF the insulation tester. Then connect the
insulation tester plus (+) lead wire to the shorted
SUPPLY terminals and the minus (–) leadwire to
the grounding terminal.
2-2
IM 01C22K01-01E
2. HANDLING CAUTIONS
3) Turn ON the insulation tester power and measure
the insulation resistance. The voltage should be
applied short as possible to verify that the insulation resistance is at least 20 MΩ.
4) After completing the test and being very careful not
to touch exposed conductors disconnect the
insulation tester and connect a 100 kΩ resistor
between the grounding terminal and the shortcircuiting SUPPLY terminals. Leave this resistor
connected at least one second to discharge any
static potential. Do not touch the terminals while it
is discharging.
• Dielectric Strength Test
1) Short-circuit the + and – SUPPLY terminals in the
terminal box.
2) Turn OFF the dielectric strength tester. Then
connect the tester between the shorted SUPPLY
terminals and the grounding terminal. Be sure to
connect the grounding lead of the dielectric strength
tester to the ground terminal.
3) Set the current limit on the dielectric strength tester
to 10 mA, then turn ON the power and gradually
increase the test voltage from ‘0’ to the specified
voltage.
4) When the specified voltage is reached, hold it for
one minute.
5) After completing this test, slowly decrease the
voltage to avoid any voltage surges.
2.9 Installation of Explosion
Protected Type
In this section, further requirements and differences and
for explosionproof type instrument are described. For
explosionproof type instrument, the description in this
chapter is prior to other description in this users manual.
For the intrinsically safe equipment and explosionproof
equipment, in case the instrument is not restored to its
original condition after any repair or modification
undertaken by the customer, intrinsically safe construction
or explosionproof construction is damaged and may cause
dangerous condition. Please contact Yokogawa for any
repair or modification required to the instrument.
NOTE
For FOUNDATIONFieldbus and PROFIBUS PA
explosion protected type, please refer to IM
01C22T02-01E and IM 01C22T03-00E respectively.
CAUTION
This instrument is tested and certified as intrinsically safe type or explosionproof type. Please
note that the construction of the instrument,
installation, external wiring, maintenance or
repair is strictly restricted, and non-observance
or negligence of this restriction would result in
dangerous condition.
WARNING
To preserve the safety of explosionproof equipment requires great care during mounting,
wiring, and piping. Safety requirements also
place restrictions on maintenance and repair
activities. Please read the following sections very
carefully.
2.9.1 FM Approval
a. FM Intrinsically Safe Type
Caution for FM intrinsically safe type. (Following
contents refer “DOC. No. IFM012-A12 P.1 and 2.”)
Note 1. Model EJA Series pressure transmitters
with optional code /FS1 are applicable for
use in hazardous locations.
• Entity Installation Requirements
Vmax ≥ Voc or Vt, Imax ≥ Isc or It,
Pmax (IS Apparatus) ≥ Pmax (Barrier)
Ca ≥ Ci + Ccable, La ≥ Li + Lcable
Note 3. Installation
• Barrier must be installed in an enclosure that meets
the requirements of ANSI/ISA S82.01.
• Control equipment connected to barrier must not use
or generate more than 250 V rms or V dc.
• Installation should be in accordance with ANSI/ISA
RP12.6 “Installation of Intrinsically Safe Systems for
Hazardous (Classified) Locations” and the National
Electric Code (ANSI/NFPA 70).
• The configuration of associated apparatus must be
FMRC Approved.
• Dust-tight conduit seal must be used when installed
in a Class II, III, Group E, F and G environments.
• Associated apparatus manufacturer’s installation
drawing must be followed when installing this
apparatus.
• The maximum power delivered from the barrier
must not exceed 0.9 W.
• Note a warning label worded “SUBSTITUTION OF
COMPONENTS MAY IMPAIR INTRINSIC
SAFETY,” and “INSTALL IN ACCORDANCE
WITH DOC. No. IFM012-A12 P.1 and 2.”
Note 4. Maintenance and Repair
• The instrument modification or parts replacement by
other than authorized representative of Yokogawa
Electric Corporation is prohibited and will void
Factory Mutual Intrinsically safe and Nonincendive
Approval.
[Intrinsically Safe]
Hazardous LocationNonhazardous Location
Class I, II, III, Division 1,
Groups A, B, C, D, E, F, G
EJA Series Pressure
Tr ansmittersSafety Barrier
+
–
+
–
Supply
+
–
General
Purpose
Equipment
+
–
[Nonincendive]
Hazardous LocationNonhazardous Location
Class I, II, Division 2,
Groups A, B, C, D, E, F, G
Class III, Division 1.
EJA Series Pressure
Tr ansmitters
+
Supply
–
Not Use
Safety Barrier
General
Purpose
Equipment
F0203-2.EPS
b. FM Explosionproof Type
Caution for FM explosionproof type.
Note 1. Model EJA Series differential, gauge,
and absolute pressure transmitters with
optional code /FF1 are applicable for
use in hazardous locations.
• Explosionproof for Class I, Division 1, Groups B,
C and D.
• Dust-ignitionproof for Class II/III, Division 1,
Groups E, F and G.
• Outdoor hazardous locations, NEMA 4X.
• Temperature Class: T6
• Ambient Temperature: –40 to 60°C
• Supply Voltage: 42 V dc max.
• Output signal: 4 to 20 mA
Note 2. Wiring
• All wiring shall comply with National Electrical
Code ANSI/NEPA70 and Local Electrical Codes.
• When installed in Division 1, “FACTORY
SEALED, CONDUIT SEAL NOT REQUIRED.”
Note 3. Operation
• Keep the “CAUTION” nameplate attached to the
transmitter.
CAUTION: OPEN CIRCUIT BEFORE REMOVING COVER. FACTORY SEALED, CONDUIT
SEAL NOT REQUIRED. INSTALL IN ACCORDANCE WITH THE INSTRUCTION MANUAL
IM 1C22.
• Take care not to generate mechanical sparking
when accessing to the instrument and peripheral
devices in a hazardous location.
Note 4. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void Factory Mutual Explosionproof Approval.
+
–
F0203-1.EPS
2-4
IM 01C22K01-01E
2. HANDLING CAUTIONS
c. FM Intrinsically Safe Type/FM
Explosionproof Type
Model EJA Series pressure transmitters with
optional code /FU1 can be selected the type
of protection (FM Intrinsically Safe or FM
Explosionproof) for use in hazardous locations.
Note 1. For the installation of this transmitter,
once a particular type of protection is
selected, any other type of protection
cannot be used. The installation must
be in accordance with the description
about the type of protection in this
instruction manual.
Note 2. In order to avoid confusion, unnecessary
marking is crossed out on the label other
than the selected type of protection when
the transmitter is installed.
2.9.2 CSA Certification
a. CSA Intrinsically Safe Type
Caution for CSA Intrinsically safe type. (Following
contents refer to “DOC No. ICS003-A12 P.1-1 and
P.1-2.”)
Note 1. Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /CS1 are applicable for use
in hazardous locations
• Intrinsically Safe for Class I, Division 1, Groups A,
B, C & D. Class II, Division 1, Groups E, F & G
and Class III, Division 1 Hazardous Locations.
• Nonincendive for Class I, Division 2, Groups A, B,
C & D, Class II, Division 2, Groups F & G, and
Class III, Hazardous Locations. (not use Safety
Barrier)
• Encl. “Type 4X”
• Temperature Class: T4
• Ambient temperature: –40 to 60°C
• Process Temperature: 120°C max.
Note 2. Entity Parameters
• Intrinsically safe ratings are as follows:
Maximum Input Voltage (Vmax) = 30 V
Maximum Input Current (Imax) = 165 mA
Maximum Input Power (Pmax) = 0.9 W
Maximum Internal Capacitance (Ci) = 22.5 nF
Maximum Internal Inductance (Li) = 730 µH
* Associated apparatus (CSA certified barriers)
Maximum output voltage (Voc) ≤ 30 V
Maximum output current (Isc) ≤ 165 mA
Maximum output power (Pmax) ≤ 0.9 W
Note 3. Installation
• All wiring shall comply with Canadian Electrical
Code Part I and Local Electrical Codes.
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation and Yokogawa
Corporation of America is prohibited and will void
Canadian Standards Intrinsically safe and
nonincendive Certification.
[Intrinsically Safe]
Hazardous LocationNonhazardous Location
Class I, II, III, Division 1,
Groups A, B, C, D, E, F, G
EJA Series Pressure
Transmitters
+
Supply
Hazardous LocationNonhazardous Location
Class I, II, Division 2,
Groups A, B, C, D, E, F, G
Class III
EJA Series Pressure
Transmitters
Supply
–
[Nonincendive]
+
–
Safety Barrier
+
+
–
–
Not Use
Safety Barrier
General
Purpose
Equipment
+
–
General
Purpose
Equipment
+
–
F0204.EPS
b. CSA Explosionproof Type
Caution for CSA explosionproof type.
Note 1. Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /CF1 are applicable for use
in hazardous locations:
• Explosionproof for Class I, Division 1, Groups B,
C and D.
• Dust-ignitionproof for Class II/III, Division 1,
Groups E, F and G.
• Encl “Type 4X”
• Temperature Class: T6, T5, and T4
• Process Temperature: 85°C (T6), 100°C (T5), and
120°C (T4)
• Ambient Temperature: –40 to 80°C
• Supply Voltage: 42 V dc max.
• Output Signal: 4 to 20 mA
2-5
IM 01C22K01-01E
2. HANDLING CAUTIONS
Note 2. Wiring
• All wiring shall comply with Canadian Electrical
Code Part I and Local Electrical Codes.
• In hazardous location, wiring shall be in conduit as
shown in the figure.
CAUTION: SEAL ALL CONDUITS
WITHIN 50 cm OF THE ENCLOSURE.
UN SCELLEMENT DOIT ÊTRE
INSTALLÉ À MOINS DE 50 cm DU
BÎTIER.
• When installed in Division 2, “SEALS NOT
REQUIRED.”
Note 3. Operation
• Keep the “CAUTION” label attached to the
transmitter.
CAUTION: OPEN CIRCUIT BEFORE
REMOVING COVER.
OUVRIR LE CIRCUIT AVANT
D´NLEVER LE COUVERCLE.
• Take care not to generate mechanical sparking
when accessing to the instrument and peripheral
devices in a hazardous location.
Note 4. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation and Yokogawa
Corporation of America is prohibited and will void
Canadian Standards Explosionproof Certification.
Non-Hazardous
Locations
Non-hazardous
Location
Equipment
42 V DC Max.
4 to 20 mA DC
Signal
Non-Hazardous
Locations
Non-hazardous
Location
Equipment
42 V DC Max.
4 to 20 mA DC
Signal
Hazardous Locations Division 1
50 cm Max.
Sealing Fitting
Hazardous Locations Division 2
Sealing Fitting
Conduit
EJA Series
EJA Series
F0205.EPS
c. CSA Intrinsically Safe Type/CSA
Explosionproof Type
Model EJA Series pressure transmitters with
optional code /CU1 can be selected the type of
protection (CSA Intrinsically Safe or CSA
Explosionproof) for use in hazardous locations.
Note 1. For the installation of this transmitter,
once a particular type of protection is
selected, any other type of protection
cannot be used. The installation must be
in accordance with the description about
the type of protection in this instruction
manual.
Note 2. In order to avoid confusion, unnecessary
marking is crossed out on the label other
than the selected type of protection when
the transmitter is installed.
2.9.3 IECEx Certification
Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /SU2 can be selected the type of
protection (IECEx Intrinsically Safe/type n or
flameproof) for use in hazardous locations.
Note 1. For the installation of this transmitter,
once a particular type of protection is
selected, any other type of protection
cannot be used. The installation must be
in accordance with the description about
the type of protection in this instruction
manual.
Note 2. In order to avoid confusion, unnecessary
marking is crossed out on the label other
than the selected type of protection when
the transmitter is installed.
a. IECEx Intrinsically Safe Type / type n
Caution for IECEx Intrinsically safe and type n.
Note 1. Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /SU2 are applicable for use
in hazardous locations
• Type of Protection and Marking Code:
Ex ia IIC T4, Ex nL IIC T4
• Ambient Temperature: –40 to 60°C
• Max. Process Temp.: 120°C
• Enclosure: IP67
Note 2. Entity Parameters
• Intrinsically safe ratings are as follows:
Maximum Input Voltage (Ui) = 30 V
Maximum Input Current (Ii) = 165 mA
Maximum Input Power (Pi) = 0.9 W
Maximum Internal Capacitance (Ci) = 22.5 nF
Maximum Internal Inductance (Li) = 730 µH
2-6
IM 01C22K01-01E
2. HANDLING CAUTIONS
• Type “n” ratings are as follows:
Maximum Input Voltage (Ui) = 30 V
Maximum Internal Capacitance (Ci) = 22.5 nF
Maximum Internal Inductance (Li) = 730 µH
• Installation Requirements
Uo ≤ Ui, Io ≤ Ii, Po ≤ Pi,
Co ≥ Ci + Ccable, Lo ≥ Li + Lcable
Uo, Io, Po, Co, and Lo are parameters of
barrier.
Note 3. Installation
• In any safety barreir used output current must be
limited by a resistor ‘R’ such that Io = Uo/R.
• The safety barrier must be IECEx certified.
• Input voltage of the safety barrier must be less
than 250 Vrms/Vdc.
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation and will void
IECEx Intrinsically safe and type n certification.
• The cable entry devices and blanking elements for
type n shall be of a certified type providing a level
of ingress protection of at least IP54, suitable for
the conditions of use and correctly installed.
• Electrical Connection:
The type of electrical connection is stamped near
the electrical connection port according to the
following marking.
T0202.EPS
Location of the marking
F0213.EPS
Note 4. Operation
• WARNING:
WHEN AMBIENT TEMPERATURE ≥ 55°C,
USE THE HEAT-RESISTING CABLES ≥ 90°C.
Note 5. Special Conditions for Safe Use
• WARNING:
IN THE CASE WHERE THE ENCLOSURE OF
THE PRESSURE TRANSMITTER IS MADE OF
ALUMINUM, IF IT IS MOUNTED IN AN AREA
WHERE THE USE OF ZONE 0 IS REQUIRED,
IT MUST BE INSTALLED SUCH, THAT, EVEN
IN THE EVENT OF RARE INCIDENTS, IGNITION SOURCES DUE TO IMPACT AND
FRICTION SPARKS ARE EXCLUDED.
[Intrinsically Safe]
Hazardous LocationNonhazardous Location
Group I/IIC, Zone 0
EJA Series Pressure
Tr ansmitters
+
Supply
Hazardous LocationNonhazardous Location
Group IIC, Zone 2
EJA Series Pressure
Tr ansmitters
Supply
–
[type n]
+
–
IECEx certified
Safety Barrier
+
–
Not Use
Safety Barrier
+
–
General
Purpose
Equipment
+
–
F0214.EPS
IECEx Certified
Equipment [nL]
+
–
F0215.EPS
b. IECEx Flameproof Type
Caution for IECEx flameproof type.
Note 1. Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /SU2 are applicable for use
in hazardous locations:
• Type of Protection and Marking Code:
Ex d IIC T6...T4
• Enclosure: IP67
• Maximum Process Temperature: 120°C (T4),
100°C (T5), 85°C (T6)
• Ambient Temperature: –40 to 75°C (T4), –40 to
80°C (T5), –40 to 75°C (T6)
• Supply Voltage: 42 V dc max.
• Output Signal: 4 to 20 mA dc
Note 2. Wiring
• In hazardous locations, the cable entry devices shall
be of a certified flameproof type, suitable for the
conditions of use and correctly installed.
• Unused apertures shall be closed with suitable
flameproof certified blanking elements. (The plug
attached is certificated as the flame proof IP67 as a
part of this apparatus.)
• In case of ANSI 1/2 NPT plug, ANSI hexagonal
wrench should be applied to screw in.
Note 3. Operation
• WARNING:
AFTER DE-ENERGIZING, DELAY 10 MINUTES
BEFORE OPENING.
2-7
IM 01C22K01-01E
2. HANDLING CAUTIONS
• WARNING:
WHEN AMBIENT TEMPERATURE ≥ 70°C,
USE THE HEAT-RESISTING CABLES ≥ 90°C.
• Take care not to generate mechanical sparking
when accessing to the instrument and peripheral
devices in a hazardous location.
Note 4. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void IECEx Certification.
2.9.4 CENELEC ATEX (KEMA)
Certification
(1) Technical Data
a. CENELEC ATEX (KEMA) Intrinsically Safe
Type
Caution for CENELEC ATEX (KEMA) Intrinsically safe type.
Note 1. Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /KS2 for potentially explosive atmospheres:
• Type of Protection and Marking code:
EEx ia IIC T4
• Temperature Class: T4
• Enclosure: IP67
• Process Temperature: 120°C max.
• Ambient Temperature: –40 to 60°C
Note 2. Electrical Data
• In type of explosion protection intrinsic safety EEx
ia IIC only for connection to a certified intrinsically
safe circuit with following maximum values:
Ui = 30 V
Ii = 165 mA
Pi = 0.9 W
Effective internal capacitance; Ci = 22.5 nF
Effective internal inductance; Li = 730 µH
Note 3. Installation
• All wiring shall comply with local installation
requirements. (Refer to the installation diagram)
Note 4. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void KEMA Intrinsically safe Certification.
Note 5. Special Conditions for Safe Use
• In the case where the enclosure of the Pressure
Transmitter is made of aluminium, if it is mounted
in an area where the use of category 1 G apparatus
is required, it must be installed such, that, even in
the event of rare incidents, ignition sources due to
impact and friction sparks are excluded.
[Installation Diagram]
Hazardous Location
Transmitter
+
Supply
–
*1: In any safety barriers used the output current must be limited by
a resistor “R” such that Imaxout-Uz/R.
Nonhazardous Location
+
Safety Barrier
–
F0208.EPS
*1
b. CENELEC ATEX (KEMA) Flameproof Type
Caution for CENELEC ATEX (KEMA) flameproof
type.
Note 1. Model EJA Series differential, gauge, and
absolute pressure transmitters with
optional code /KF2 for potentially explosive atmospheres:
• No. KEMA 02ATEX2148
• Applicable Standard: EN50014:1997,
EN50018:2000
• Type of Protection and Marking Code: EEx d IIC
T6···T4
• Temperature Class: T6, T5, and T4
• Enclosure: IP67
• Maximum Process Temperature:
85°C (T6), 100°C (T5), and 120°C (T4)
• Ambient Temperature: T4 and T6; –40 to 75°C,
T5; –40 to 80°C
Note 2. Electrical Data
• Supply voltage: 42 V dc max.
• Output signal: 4 to 20 mA
Note 3. Installation
• All wiring shall comply with local installation
requirement.
• The cable entry devices shall be of a certified
flameproof type, suitable for the conditions of use.
Note 4. Operation
• Keep the “CAUTION” label to the transmitter.
CAUTION: AFTER DE-ENERGIZING,
DELAY 10 MINUTES BEFORE OPENING. WHEN THE AMBIENT
TEMP.70°C, USE HEAT-RESISTING
CABLES90°C.
2-8
IM 01C22K01-01E
2. HANDLING CAUTIONS
• Take care not to generate mechanical sparking
when accessing to the instrument and peripheral
devices in a hazardous location.
Note 5. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void KEMA Flameproof Certification.
c. CENELEC ATEX (KEMA) Intrinsically Safe
Type/CENELEC ATEX (KEMA) Flameproof
Type/CENELEC ATEX Type n
Model EJA Series pressure transmitters with
optional code /KU2 can be selected the type of
protection CENELEC ATEX (KEMA) Intrinsically Safe, Flameproof or CENELEC ATEX
Type n for use in hazardous locations.
Note 1. For the installation of this transmitter,
once a particular type of protection is
selected, any other type of protection
cannot be used. The installation must be
in accordance with the description about
the type of protection in this user’s
manual.
Note 2. In order to avoid confusion, unnecessary
marking is crossed out on the label other
than the selected type of protection when
the transmitter is installed.
• CENELEC ATEX Type of Protection “n”
WARNING
When using a power supply not having a nonincendive circuit, please pay attention not to
ignite in the surrounding flammable atmosphere.
In such a case, we recommend using wiring
metal conduit in order to prevent the ignition.
Note 3. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void Type of Protection “n”.
[Installation Diagram]
Hazardous Location
Tr ansmitter
Ratings of the Power Supply as follows;
(Zone 2 only)
Supply
Maximum Voltage: 30 V
+
–
Nonhazardous Location
+
Power Supply
–
F0209.EPS
• CENELEC ATEX Type of Protection “Dust”
• Applicable Standard: EN50281-1-1:1997
• Type of Protection and Marking Code: II 1D
• Maximum surface temperature: T65°C (Tamb.:
40°C), T85°C (Tamb.: 60°C), and T105°C (Tamb.:
80°C)
Note 1. Installation instructions
The cable entry devices and blanking
elements shall be of a certificated type
providing a level of ingress protection of
at least IP6x, suitable for the conditions
of use and correctly installed.
(2) Electrical Connection
The type of electrical connection is stamped near
the electrical connection port according to the
following marking.
• Applicable Standard: EN60079-15
• Referential Standard: IEC60079-0, IEC60079-11
• Type of Protection and Marking Code:
Ex nL IIC T4
• Temperature Class: T4
• Enclosure: IP67
• Process Temperature: 120°C max.
• Ambient Temperature: –40 to 60°C
Note 1. Electrical Data
Ui = 30 V
Effective internal capacitance; Ci = 22.5 nF
Effective internal inductance; Li = 730 µH
Note 2. Installation
• All wiring shall comply with local installation
requirements. (refer to the installation diagram)
Location of the marking
F0200.EPS
(3) Installation
WARNING
• All wiring shall comply with local installation
requirement and local electrical code.
• There is no need of the conduit seal for both of
Division 1 and Division 2 hazardous locations
because this product is sealed at factory.
• In case of ANSI 1/2 NPT plug, ANSI hexagonal
wrench should be applied to screw in.
2-9
IM 01C22K01-01E
2. HANDLING CAUTIONS
(4) Operation
WARNING
• OPEN CIRCUIT BEFORE REMOVING
COVER. INSTALL IN ACCORDENCE WITH
THIS UWER’S MANUAL
• Take care not to generate mechanical sparking
when access to the instrument and peripheral
devices in hazardous locations.
(5) Maintenance and Repair
WARNING
The instrument modification or parts replacement
by other than authorized Representative of
Yokogawa Electric orporation is prohibited and
will void the certification.
(6) Name Plate
OUTPUT: Output signal.
MWP: Maximum working pressure.
CAL RNG: Specified calibration range.
DISP MODE: Specified display mode.
OUTPUT MODE: Specified output mode.
NO.: Serial number and year of production*1.
TOKYO 180-8750 JAPAN:
The manufacturer name and the address*2.
*1: The third figure from the last shows the last one
figure of the year of production. For example, the
production year of the product engraved in “NO.”
column on the name plate as follows is 2001.
12A819857 132
The year 2001
*2: “180-8750” is a zip code which represents the
following address.
2-9-32 Nakacho, Musashino-shi, Tokyo Japan
Name plate
Tag plate for flameproof type
Tag plate for intrinsically safe type
Tag plate for type n protection
: Refer to USER'S MANUAL
2.9.5 TIIS Certification
a. TIIS Flameproof Type
The model EJA Series pressure transmitter with
optional code /JF3, which has obtained certification
according to technical criteria for explosion-protected
construction of electric machinery and equipment
(Standards Notification No. 556 from the Japanese
Ministry of Labor) conforming to IEC standards, is
designed for hazardous areas where inflammable gases
or vapors may be present. (This allows installation in
Division 1 and 2 areas)
To preserve the safety of flameproof equipment
requires great care during mounting, wiring, and
piping. Safety requirements also place restrictions on
maintenance and repair activities. Users absolutely
must read “Installation and Operating Precautions for
TIIS Flameproof Equipment” at the end of this
manual.
Tag plate for flameproof, intrinsically safe type,
type n protection, and Dust
(For TIIS flameproof type without integral indicator)
When the fill fluid near the sensor part moves
from within, the instrument outputs a failure
signal either high or low of the specific signal. In
that case, generate the alarm to identify that the
failure signal is output since the event may
invalidate the flameproof approval.
If the optional integral indicator is equipped, the
indicator identifies the alarm on its display.
Therefore, no other alarm generation is necessary.
Hazardous LocationNonhazardous Location
4 to 20 mA DC1 to 5 V DC
Transmitter
Figure 2.3 Example of using DCS (Distributed Control
System)
Power
Supply
DCS
Display
F0210.EPS
CAUTION
When selecting cables for TIIS flameproof type
transmitters, determine the cables' maximum
allowable heat resistance depending on the
process and ambient temperature condition on
the transmitter as illustrated in Figure 2.4. Use
cables having a maximum allowable heat
resistance of at least 60°C for the transmitter in
Region A and that of 75°C in Region B.
60
50
45
20
40
20
0
20
020406080100 120
Process Temperature (C)
Ambient
Temperature
(C)
Figure 2.4 Selecting Cables
Region A
Region B
90
F0211.EPS
b. TIIS Intrinsically Safe Type
The model EJA Series pressure transmitter with optional
code /JS3, which has obtained certification according to
technical criteria for explosionprotected construction of
electric machinery and equipment (Standards Notification No.556 from the Japanese Ministry of Labor) conforming to IEC standards, is designed for hazardous areas where explosive or inflammable gases or vapors
may be present. (This allows installation in Division 0, 1
and 2 areas)
To preserve the safety of flameproof equipment requires
great care during mounting,wiring, and piping. Safety
requirements also place restrictions on maintenance and
repair activities. Users absolutely must read “Installation
and Operating Precautions for TIIS Intrinsically Safe
Equipment” at the end of this manual.
CAUTION
For using a safety-barrier with a pressure
transmitter, the safety-barrier must be certified as
a safety-barrier itself.
A safety-barrier must be used under the following condition.
(1) Condition of the current and voltage limits
Maximum output voltage(Uo) ≤ 28V
Maximum output current(Io)≤ 94.3mA
Maximum output power (Po) ≤ 0.66W
(2) Category and Group
Categoryia
GroupII C
(3) Relations between a maximum allowed
inductance and a field wiring inductance,
between a maximum allowed capacitance
and a field wiring capacitance.
Lo ≥ Li + Lw
Co ≥ Ci + Cw
(Li = 730µH, Ci=11nF)
Lo = Maximum external inductance
Li = Maximum internal inductance
Lw = Field wiring inductance
Go = Maximum external capacitance
Ci = Maximum internal capacitance
Cw = Field wiring capacitance
2-11
IM 01C22K01-01E
Hazardous LocationNonhazardous
4 to 20 mA DC
Tr ansmitter
Li=730 µH
Ci=11 nF
Figure 2.5 Diagram for Connecting Safety Barrierre
Lw, Cw
Location
Safety Barrier
Lo, Co
F0212.EPS
2.10 EMC Conformity Standards
EN61326-1 Class A, Table 2 (For use in industrial
locations)
EN61326-2-3
CAUTION
This instrument is a Class A product, and it is
designed for use in the industrial environment.
Please use this instrument in the industrial
environment only.
2. HANDLING CAUTIONS
(2) Installation Category I
“Overvoltage category (Installation category)”
describes a number which defines a transient
overvoltage condition. It implies the regulattion for
impulse withstand voltage. “I” applies to electrical
equipment which is supplied from the circuit when
appropriate transient overvoltage control means
(interfaces) are provided.
NOTE
YOKOGAWA recommends customer to apply
the Metal Conduit Wiring or to use the twisted
pair Shield Cable for signal wiring to conform the
requirement of EMC Regulation, when customer
installs the EJA Series Transmitters to the plant.
2.11 PED (Pressure Equipment
Directive)
EJA series of pressure transmitters are categorized as
pressure accessories of this directive 97/23/EC, which
corresponds to Article 3, Paragraph 3 of PED, denoted
as Sound Engineering Practice (SEP).
2.12 Low Voltage Directive
Applicable standard: EN61010-1
(1) Pollution Degree 2
“Pollution degree” describes the degree to which a
solid, liquid, or gas which deteriorates dielectric
strength or surface resistivity is adhering. “2”
applies to normal indoor atmosphere. Normally,
only non-conductive pollution occurs. Occasionally,
however, temporary conductivity caused by
condensation must be expected.
2-12
IM 01C22K01-01E
3.COMPONENT NAMES
3. COMPONENT NAMES
Vertical impulse piping type
Process connection
Pressure-detector section
Cover flange
CPU
assembly
Integral indicator
(Note 1)
Tr ansmitter section
Setting pin(CN4)
Mounting
Amplifier cover
Range-setting switch
(See Subsection 7.7)
Note 1: See Subsection 10.2, “Model and Suffix Codes,” for details.
Note 2: Insert the pin (CN4) as shown in the figure above to set the burn-out direction. The pin is set to the H side for delivery (unless
option code /C1 is specified in the order).
The setting can be confirmed by calling up parameter D52 using the BRAIN TERMINAL. Refer to Subsection 8.3.3 (10).
Note 3: If optional code /F1 is specified, output signal is –2.5% or lower.
Display mode is ‘square root’. (Display is not lit when ‘proportional’ mode.)
The output signal being zero-adjusted is increasing.
The output signal being zero-adjusted is decreasing.
Select one of these sixteen available engineering units for the display.
T0301.EPS
3-1
IM 01C22K01-01E
4.INSTALLATION
4. INSTALLATION
4.1 Precautions
Before installing the transmitter, read the cautionary
notes in Section 2.4, “Selecting the Installation
Location.” For additional information on the ambient
conditions allowed at the installation location, refer to
Subsection 10.1 “Standard Specifications.”
IMPORTANT
• When welding piping during construction, take
care not to allow welding currents to flow
through the transmitter.
• Do not step on this instrument after installation.
4.2 Mounting
The transmitter can be mounted on a nominal 50
mm (2-inch) pipe using the mounting bracket
supplied, as shown in Figure 4.2.1 and 4.2.2. The
transmitter can be mounted on either a horizontal or
a vertical pipe.
When mounting the bracket on the transmitter,
tighten the (four) bolts that hold the transmitter with
a torque of approximately 39 N·m {4kgf·m}.
The transmitter is shipped with the manifold set up
as per the order specifications.
For correct flow measurement, the flow path must
always be filled with fluid; otherwise, measurement
accuracy cannot be assured.
For the vertical impulse piping type, it is recom-
mended that the manifold be mounted facing up for
liquid flow measurement; facing down for gas flow
measurement, as shown in Figure 4.2.2.
Always turn OFF power, release pressure and
remove a transmitter to non-hazardous area
before disassembling and reassmbling an
indicator.
An integral indicator can be installed in the following
three directions. Follow the instructions in section 9.4
for removing and attaching the integral indicator.
4. INSTALLATION
Figure 4.4Integral Indicator Direction
F0404.EPS
4-3
IM 01C22K01-01E
5. INSTALLING IMPULSE PIPING
5.INSTALLING IMPULSE PIPING
5.1 Process Piping Installation
Precautions
The manifold contains a small-bore orifice. For the
transmitter of a high pressure connection right side, the
orifice is placed facing such a direction as to enable
normal flow measurement when fluid is flowed from
right to left (as viewed from the front). If the orifice is
removed from the manifold, it must be replaced facing
the correct direction. (For disassembly and reassembly
procedures, see Subsection 9.4.4)
Pay careful attention to the following points when
routing the process piping and connection the process
piping to the transmitter.
5.1.1 Connecting Process Piping to the
Transmitter
(1) Confirming the Process Fluid Flow Direc-
tion (Figure 5.1.1)
The mark “” on the manifold indicates the direction
in which the process fluid is flowed (from right to left).
When connecting the process piping to the process
connector, confirm the process fluid flow direction.
Manifold
Flow direction(from right to left)
Process connection
(outflow side)
Orifice name plate
Process connection
(inflow side)
(2) Tightening the Process Connector Mount-
ing Bolts
The transmitter is shipped with the process connector
mounting bolts only loosely tightened. After connecting the process piping, tighten these bolts uniformly to
prevent leaks with a torque of 39 to 49 N·m {4 to 5
kgf·m}.
(3) Removing the Process Connector Port
Dustproof Cap
The process connector port threads are covered with a
plastic cap to exclude dust. This cap must be removed
before connecting the piping. (Be careful not to
damage the threads when removing this cap. Never
insert a screwdriver or other tool between the cap and
port threads to remove the cap.)
5.1.2 Routing the Process Piping
(1) Relationship between Process Fluid and
Manifold Locations (For the vertical impulse piping type)
If condensate (or gas) generated in the process piping
were allowed to accumulate, then it would be necessary to remove it periodically by opening the drain (or
vent) plug. However, this would generate a transient
disturbance in the pressure measurement. Therefore,
the process piping must be routed so that any condensate (or gas) generated in the process piping will not
accumulate in the pressure-sensing assembly of the
transmitter.
Bolt
Process connector
(low pressure side)
Figure 5.1.1 Manifold and Flow Direction Indication
Process connector
(high pressure side)
F0501.EPS
NOTE
• If the process fluid is a gas, then as a rule the
manifold must be located at the downside of
the pressure-sensing assembly. (Figure 5.1.2)
• If the process fluid is a liquid, then as a rule
the manifold must be located at the upside of
the pressure-sensing assembly. (Figure 5.1.3)
5-1
IM 01C22K01-01E
5. INSTALLING IMPULSE PIPING
(2) Pipe Size for Process Piping
Use a 15 mm (1/2-inch) pipe for process piping
connection to the process connector.
(3) Preventing Freezing
If there is any risk that the process fluid in the transmitter pressure-sensing assembly could freeze or
solidify, use a steam jacket or heater to maintain the
temperature of the fluid.
Manifold
Figure 5.1.2 Manifold Location at the Downside (for Gas
Flow Measurement)
F0502.EPS
• The high pressure connecting port on the transmitter is shown on the right (as viewed from the front).
• The transmitter process piping connection is shown
for a vertical impulse piping connection configuration in which the direction of process flow is from
right to left.
• The process piping material used must be compatible with the process pressure, temperature, and
other conditions.
•A variety of process piping-mounted stop valves are
available according to the type of connection
(flanged, screwed, welded), construction (globe,
gate, or ball valve), temperature and pressure.
Select the type of valve most appropriate for the
application.
Gas flow measurement
Union or flange
Manifold
F0503.EPS
Figure 5.1.3 Manifold Location at the Upside (for Liquid
Flow Measurement)
5.2 Process Piping Connection
Examples
Figure 5.2 shows examples of typical process piping
connections. Before connecting the transmitter to the
process, study the transmitter installation location, the
process piping layout, and the characteristics of the
process fluid (corrosiveness, toxicity, flammability,
etc.), in order to make appropriate changes and
additions to the connection configurations.
Manifold
Process piping
Liquid flow measurement
Manifold
Union or flange
Process piping
Figure 5.2 Process Piping Connection Examples
Stop valve
Stop valve
F0504.EPS
Note the following points when referring to these
piping examples.
5-2
IM 01C22K01-01E
6.WIRING
6. WIRING
6.1 Wiring Precautions
IMPORTANT
• Lay wiring as far as possible from electrical
noise sources such as large capacity transformers, motors, and power supplies.
• Remove electrical connection dust cap before
wiring.
• All threaded parts must be treated with waterproofing sealant. (A non-hardening silicone
group sealant is recommended.)
• To prevent noise pickup, do not pass signal
and power cables through the same ducts.
• Explosion-protected instruments must be wired
in accordance with specific requirements (and,
in certain countries, legal regulations) in order
to preserve the effectiveness of their explosionprotected features.
• The terminal box cover is locked by an Allen
head bolt (a shrouding bolt) on CENELEC,
IECEx, and TIIS flameproof type transmitters.
When the shrouding bolt is driven clockwise by
an Allen wrench, it is going in and cover lock is
released, and then the cover can be opened.
See Subsection 9.4 “Disassembly and Reassembly” for details.
Refer to The “Installation and Operating
Precautions for TIIS Flameproof Equipment”
and “Installation and Operating Precautions
for TIIS Intrinsically Safe Equipment” at the
end of this manual for correct wiring.
CAUTION
When selecting cables for TIIS flameproof type
transmitters, determine cables' maximum allowable heat resistance depending on the temperature condition on the transmitter. See Section
2.9.5 TIIS Certification for details.
(d) In environment where oils, solvents, corrosive gases
or liquids may be present, use wires or cables that
are resistant to such substances.
(e) It is recommended that crimp-on solderless terminal
lugs (for 4 mm screws) with insulating sleeves be
used for leadwire ends.
6.3 Connections of External
Wiring to Terminal Box
6.3.1 Power Supply Wiring Connection
Connect the power supply wiring to the SUPPLY +
and – terminals.
Transmitter terminal box
+
Power supply
–
F0601.EPS
Figure 6.3.1 Power Supply Wiring Connection
6.2 Selecting the Wiring Materials
(a) Use stranded leadwires or cables which are the
same as or better than 600 V grade PVC insulated
wire (JIS C3307) or equivalent.
(b) Use shielded wires in areas that are susceptible to
electrical noise.
(c) In areas with higher or lower ambient temperatures,
use appropriate wires or cables.
6.3.2 External Indicator Connection
Connect wiring for external indicators to the CHECK +
and – terminals.
(Note) Use a external indicator whose internal resistance is 10 Ω or
less.
Power supply
+
–
Tr ansmitter terminal box
Figure 6.3.2 External Indicator Connection
6-1
External indicator
IM 01C22K01-01E
F0602.EPS
6. WIRING
6.3.3 BRAIN TERMINAL BT200 Connec-
tion
Connect the BT200 to the SUPPLY + and – terminals
(Use hooks). The communication line requires a
reception resistor of 250 to 600 Ω in series.
Tr ansmitter terminal box
BT200
Figure 6.3.3 BT200 Connection
+
Power supply
–
Ignore the polarity
since the BT200 is
AC-coupled to the
terminal box.
F0603.EPS
6.3.4 Check Meter Connection
Connect the check meter to the CHECK + and –
terminals (use hooks).
•A 4 to 20 mA DC output signal from the CHECK +
and – terminals.
(Note) Use a check meter whose internal resistance is 10 Ω or less.
Power supply
+
–
Check meter
Figure 6.3.4 Check Meter Connection
Tr ansmitter terminal box
F0604.EPS
6.4 Wiring
6.4.1 Loop Configuration
Since the DPharp uses a two-wire transmission system,
signal wiring is also used as power wiring.
DC power is required for the transmitter loop. The
transmitter and distributor are connected as shown
below.
For details of the power supply voltage and load
resistance, see Section 6.6; for communications line
requirements, see Subsection 8.1.2.
(1) General-use Type and Flameproof Type
Hazardous LocationNonhazardous Location
Tr ansmitter terminal box
Figure 6.4.1a Connection between Transmitter and
Distributor
(2) Intrinsically Safe Type
For intrinsically safe type, a safety barrier must be
included in the loop.
Hazardous LocationNonhazardous Location
Tr ansmitter terminal box
Distributor
(Power supply unit)
+
–
Distributor
(Power supply unit)
Receiver
instrument
F0605.EPS
CAUTION
For the intrinsically safe equipment and flameproof equipment, wiring materials and wiring
work for these equipment including peripherals
are strictly restricted. Users absolutely must read
“Installation and Operating Precautions for TIIS
Intrinsically Safe Equipment” and “Installation
and Operating Precautions for TIIS Flameproof
Equipment” at the end of this manual prior to the
work.
+
–
Safety barrier
Figure 6.4.1b Connection between Transmitter and
Distributor
6-2
Receiver
instrument
F0606.EPS
IM 01C22K01-01E
6. WIRING
6.4.2 Wiring Installation
(1) General-use Type and Intrinsically Safe
Type
Make cable wiring using metallic conduit or waterproof glands.
• Apply a non-hardening sealant to the terminal box
connection port and to the threads on the flexible
metal conduit for waterproofing.
Flexible metal conduit
Apply a non-hardening
Wiring metal
conduit
Te e
Drain plug
Figure 6.4.2a Typical Wiring Using Flexible Metal Conduit
(2) Flameproof Type (TIIS)
Wire cables through a flameproof packing adapter, or
using a flameproof metal conduit.
Wiring cable through flameproof packing adapter
for only TIIS flameproof type (see Figure 6.4.2b).
• Use only flameproof packing adapters approved by
Yokogawa.
• Apply a nonhardening sealant to the terminal box
connection port and to the threads on the flameproof
packing adapter for waterproofing.
sealant to the threads for
waterproofing.
F0607.EPS
•Measure the cable outer diameter in two directions to
within 1 mm.
•Calculate the average of the two diameters, and use
packing with an internal diameter nearest to this value
(see Table 6.4.2).
Table 6.4.2Flameproof Packings and Applicable Cable
Optional
Code
G11
G12
Outer Diameters
Wiring Port
Thread
Diameter
G 1/2
Applicable
Cable OD
(mm)
8 to 10
10.1 to 12
Identifying
Mark
16 8-10
16 10-12
Part
Number
G9601AM
T0601.EPS
• Mounting flameproof packing adapter body to conduit
connection (see Figure 6.4.2c)
1) Screw the flameproof packing adapter into the
terminal box until the O-ring touches the wiring
port (at least 6 full turns), and firmly tighten the
lock nut.
2) Insert the cable through the union cover, the union
coupling, the clamp nut, the clamp ring, the gland,
the washer, the rubber packing, and the packing
box, in that order.
3) Insert the end of the cable into the terminal box.
4) Tighten the union cover to grip the cable. When
tightening the union cover, tighten approximately
one turn past the point where the cable will no
longer move up and down.
Proper tightening is important. If it is too tight, a
circuit break in the cable may occur; if not tight
enough, the flameproof effectiveness will be
compromised.
5) Fasten the cable by tightening the clamp nut.
6) Tighten the lock nut on the union cover.
7) Connect the cable wires to each terminal.
Flameproof packing adapter
Flexible metal conduit
Wiring metal
conduit
Te e
Drain plug
Figure 6.4.2b Typical Cable Wiring Using Flameproof
Packing Adapter
Apply a non-hardening
sealant to the threads for
waterproofing.
F0608.EPS
Apply a non-hardnening
sealant to the threads for
waterproofing.
•A seal fitting must be installed near the terminal box
connection port for a sealed construction.
• Apply a non-hardening sealant to the threads of the
terminal box connection port, flexible metal conduit
and seal fitting for waterproofing.
Non-hazardous area
Hazardous area
Flameproof
heavy-gauge
steel conduit
Te e
Drain plug
Figure 6.4.2d Typical Wiring Using Flameproof Metal
Conduit
Gas sealing device
Flameproof flexible
metal conduit
Apply a non-hardening
sealant to the threads of
these fittings for
waterproofing
Seal fitting
After wiring, impregnate the fitting
with a compound to seal tubing.
F0610.EPS
6.5 Grounding
Grounding is always required for the proper operation
of transmitters. Follow the domestic electrical requirements as regulated in each country. For a transmitter
with built-in lightning protector, grounding should
satisfy Class C requirements (ground resistance of 10
or less).
6.6 Power Supply Voltage and
Load Resistance
When configuring the loop, make sure that the external
load resistance is within the range in the figure below.
(Note) In case of an intrinsically safe transmitter, external load
resistance includes safety barrier resistance.
600
External
load
resistance
R (Ω)
250
010.516.424.742
Figure 6.6 Relationship between Power Supply Voltage
and External Load Resistance
E–10.5
R=
0.0236
Power supply voltage E (V DC)
Communication
applicable range
BRAIN and HART
F0612.EPS
Ground terminals are located on the inside and outside
of the terminal box. Either of these terminals may be
used.
WARNING
For TIIS flameproof type and intrinsically safe,
grounding should satisfy Class D requirements
(grounding resistance, 100 or less).
Tr ansmitter terminal box
Ground terminal
(Inside)
Ground terminal
(Outside)
Figure 6.5 Ground Terminals
F0611.EPS
6-4
IM 01C22K01-01E
7.OPERATION
7. OPERATION
7.1 Preparation for Starting
Operation
The Model EJA115 low flow transmitter measures the
flow rates of liquids and gases. This section describes
the operation procedure for the EJA115 as shown in
Figure 7.1 (vertical impulse piping type, high-pressure
connection: right side) when measuring a liquid flow
rate.
(a) Follow the procedures below to introduce process
pressure into the transmitter.
1) Open the stop valve on the downstream side.
2) Gradually open the stop valve on the upstream side
to introduce process fluid into the transmitter
pressure-detector section.
This will cause process fluid to flow into the orifice
built in the manifold, and apply flow-dependent
differential pressure to the high and low pressure
sides of the transmitter.
3) Confirm that there are no pressure leaks in the stop
valves on the upstream and downstream sides,
process piping connection or transmitter, etc.
(b) Venting Gas from the Transmitter Pressure-detector
Section
Since the piping in the example of Figure 7.1 is
constructed to be self-venting, no venting operation
is required. If it is not possible to make the piping
self-venting, refer to Subsection 7.6 for instructions.
(c) Turn ON power and connect the BT200.
Open the terminal box cover, and connect the
BT200 to the SUPPLY + and – terminals.
(d) Using the BT200, confirm that the transmitter is
operating properly. Check parameter values or
change the setpoints as necessary. See Chapter 8 for
BT200 operation.
If the transmitter is equipped with an integral
indicator, its indication can be used to confirm that
the transmitter is operating properly.
Stop valve(downstream side)
Manifold
Stop valve
(upstream side)
F0701.EPS
Figure 7.1Liquid Flow Measurement
Confirming that Transmitter is Operating
Properly
Using the BT200
• If the wiring system is faulty, ‘communication error’
appears on the display.
• If the transmitter is faulty, ‘SELF CHECK ERROR’
appears on the display.
PARAM
C60:SELF CHECK
ERROR
communication error
DATADIAGPRNTESC
Communication error
(Faulty wiring)
Self-diagnostic error
(Faulty transmitter)
F0702.EPS
Using the integral indicator
• If the wiring system is faulty, the display stays blank.
• If the transmitter is faulty, an error code will appear
on the display according to the nature of the error.
7-1
Self-diagnostic error on
the integral indicator
(Faulty transmitter)
F0703.EPS
IM 01C22K01-01E
7. OPERATION
NOTE
If any of the error indications above appears on
the display of the integral indicator or BT200,
refer to Subsection 8.5.2 for corrective action.
Verify and Change Transmitter Parameter
Setting and Values
The following parameters are the minimum settings
required for operation. The transmitter has been
shipped with these parameters. To confirm or change
the values, see Subsection 8.3.3.
• Measuring range ................. See Subsection 8.3.3 (2)
• Output/integral indicator mode
............................................ See Subsection 8.3.3 (4)
• Operation mode .................. See Subsection 8.3.3 (9)
7.2 Zero Point Adjustment
Adjust the zero point after operating preparation is
completed. Make sure to close the stop valves on the
upstream and downstream sides before the adjustment.
IMPORTANT
Do not turn off the power to the transmitter
immediately after a zero adjustment. Powering
off within 30 seconds after a zero adjustment will
return the adjustment back to the previous
settings.
The zero point adjustment can be made in either way:
using the zero-adjustment screw of the transmitter or
the BT200 operation.
For output signal checking, display the parameter A10:OUTPUT (%) in the BT200.
BT200
PARAM
A10:OUTPUT(%)
0.0 %
A11:ENGR OUTPUT
A20:AMP TEMP
DATADIAGPRNTESC
Zero-adjustment Screw
Zero-adjustment
screw
Output signal (%)
display
F0704.EPS
Using the Transmitter Zero-adjustment
Screw
Before adjusting a screw, check that the parameter
J20: EXT ZERO ADJ displays ENABLE. See
Subsection 8.3.3 (13) for the setting procedure.
Use a slotted screwdriver to turn the zero-adjustment
screw. Turn the screw clockwise to increase the output
or counterclockwise to decrease the output. The zero
point adjustment can be made with a resolution of
0.01% of the setting range. Since the degree of zero
adjustments varies with the screw turning speed, turn
the screw slowly for fine adjustment and quickly for
coarse adjustment.
Using the BT200
Zero point can be adjusted by simple key operation of
the BT200.
Select parameter J10: ZERO ADJ, and press the
ENTER key twice. The zero point will be adjusted
automatically to the output signal 0% (4 mA DC).
Confirm that the setting value displayed for the
parameter is ‘0.0%’ before pressing the ENTER key.
See Subsection 8.3.3 (13) for BT200 operating procedures.
SET
J10:ZERO ADJ
–0.0 %
+ 000.0
7-2
A display when parameter
J10 is selected.
Press key
twice for 0% output 4 mA DC.
CLRESC
F0705.EPS
IM 01C22K01-01E
7. OPERATION
7.3 Starting Operation
After completing the zero point adjustment, follow the
procedure below to start operation.
1) Open the stop valve on the upstream side.
2) Gradually open the stop valve on the downstream
side. This places the transmitter in an operational
condition.
3) Confirm the operating status. If the output signal
exhibits wide fluctuations (hunting) due to periodic
variation in the process pressure, use BT200 to
dampen the transmitter output signal. Confirm the
hunting using a receiving instrument or the integral
indicator, and set the optimum damping time
constant. See Subsection 8.3.3 (3), “Damping Time
Constant Setup.”
4) After confirming the operating status, perform the
following.
IMPORTANT
• Remove the BT200 from the terminal box, and
confirm that none of the terminal screws are
loosened.
• Close the terminal box cover and the amplifier
cover. Screw each cover in tightly until it will
not turn further.
• Two covers are required to be locked on the
CENELEC, IECEx, and TIIS Flameproof type
transmitters. An Allen head bolts (shrouding
bolts) are provided under edge of the each
cover for locking. When a shrouding bolts are
driven counterclockwise by an Allen wrench, it
is coming out and locks up a cover. (See page
9-3) After locking, the covers should be
confirmed not to be opened.
• Tighten the zero-adjustment cover mounting
screw to fix the cover in position.
7.4 Shutting Down Operation
Shut down the transmitter operation as follows.
1) Turn off the power.
2) Close the stop valves on the up and downstream
sides.
NOTE
Whenever shutting down the transmitter for a
long period, remove any process fluid from the
transmitter pressure-detector section.
7.5 Transmitter Measurement
Range (Determining Differential Pressure Range)
The following describes the procedure for calculating
the differential pressure range and the calculation
example in low flow measurement.
Conversion factor in pressure unit:
1 Pa=1.0197210
1 mmH2O=9.80665 Pa
1 atm=1.01325102 kPa
7.5.1Determining the Differential Pres-
sure Range
Use the following procedures to determine a differential
pressure range according to the fluid conditions being
measured.
(a) Calculate a water or air equivalent flow from the
flow of the fluid being measured (100% flow).
Equivalent Water Flow Calculation
Qw = 0.03162 . Qf . f(1)
-1
mmH2O
Where, Qw: Water equivalent volumetric flow (m3/h) at
4°C, 1 atm
Qf: Volumetric liquid flow (m3/h) at operating
conditions (t°C, p kPa)
f: Specific liquid density (kg/m3) at operating
conditions (t°C, p kPa)
7-3
IM 01C22K01-01E
7. OPERATION
Equivalent Air Flow Calculation
Qo = 0.5356 . Qn ρn .
273.15 + t
101.325+ pZfZn
.
(2)
Where, Qo: Air equivalent volumetric flow at 0°C,
1 atm (Nm3/h)
Qn: Volumetric gas flow at 0°C, 1 atm (Nm3/h)
n: Specific gas density at 0°C, 1 atm (kg/Nm3)
Zn: Compression factor of gas at 0°C, 1 atm
Zf: Compression factor of gas at operations
conditions (t°C, p kPa)
(b) Obtain a differential pressure from the above
equivalent water or air flow using the nomograph
shown in Figure 7.5.1 or 7.5.2. In this procedure,
multiply Qw or Qo by 1000/60 to convert the flow
unit into liter/min.
(c) Select an orifice bore, taking into considerations
pressure loss, etc.
(d) As necessary, calculate Reynolds number at normal
flow rate and correct the differential pressure
obtained from the procedure (b).
Reynolds Number Calculation
Re = 354(3)
W
D.
Where, Re: Reynolds number at normal flow rate
W: Weight flow at normal flow rate (kg/h)
(Note)
D:Orifice bore (mm)
:Viscosity (mPa·s)
Note: Determination of W
· For liquid, W=Qf·f
· For gas, W=Qn·n
7.5.2 Example of Calculation
Fluid: N2 gas (Nitrogen gas)
Flow range: 0 to 25 Nm
Normal flow rate: 18 Nm
Specific density: 1.251 kg/Nm
1 atm)
Temperature: 30°C
Pressure: 100 kPa
Viscosity: 0.018 mPa·s
From Equation (2), air equivalent volumetric flow Qo
is:
Qo= 0.5356 × 25 1.251 ×
A differential pressure range of 0 to 2400 mmH2O is
obtained from Figure 7.5.2 applying an orifice bore of
6.350 mm (where, Zf/Zn=1 is assumed).
From Equation (3), Reynolds number at normal flow
rate Re is:
Re = 354 ×= 6.97 × 10
18 × 1.251
6.35 × 0.018
Since the correction factor (1.00) is constant at this
Reynolds number, no differential pressure correction is
required. Consequently, the differential pressure range
is determined as 0 to 2400 mmH2O.
3
/h (flow rate at 0°C, 1 atm)
3
/h
3
(specific density at 0°C,
273.15 + 30
101.325 + 100
= 18.38 Nm3/h = 306.3 Nl/min
4
F0714.EPS
Differential Pressure Correction using
Reynolds Number
2
1
.
!P =
Kaf/Ka
!P
0
Where, !P: Corrected differential pressure
!P0: Differential pressure obtained from proce-
dure (b)
Kaf/ka: Correction factor obtained from Figure
7.5.3
For details concerning determination of differential
pressure correction using Reynolds number, pressure loss,
etc., refer to TI 01C20K00-01E.
F0713.EPS
7-4
IM 01C22K01-01E
7. OPERATION
0.05
0.10.5151040
200
100
50
Differential
Pressure
(kPa)
0.5080.8641.5112.5274.0396.350
10
5
1
0.010.050.10.5151040
Equivalent water flow(liter/min at 0C, 1 atm)
Figure 7.5.1 Relationship between Equivalent Water Flow and Differential Pressure
200
100
50
10
5
1
F0707.EPS
200
100
Differential
Pressure
(kPa)
0.51510501005001000
50
0.5080.8641.5112.5274.0396.350
10
5
1
0.51510501005001000
Equivalent air flow(liter/min at 0C, 1 atm)
200
100
50
10
5
1
F0708.EPS
Figure 7.5.2 Relationship between Equivalent Air Flow and Differential Pressure
7-5
IM 01C22K01-01E
1.00
7. OPERATION
Correction
0.95
E
Factor
(Kaf/Ka)
0.90
0.85
0.80
0.75
865862345 8623423
445
10
F
D
C
B
A
2
10
3
Reynolds number(Re)
Figure 7.5.3 Relationship between Reynolds Number and Correction Factor
7.6 Venting or Draining Transmitter Pressure-detector Section
Since this transmitter is designed to be self-draining
and self-venting with vertical impulse piping connections, neither draining nor venting will be required if
the impulse piping is configured appropriately for selfdraining or self-venting operation.
If condensate (or gas) collects in the transmitter
pressure-detector section, the measured pressure may
be in error. If it is not possible to configure the piping
for self-draining (or self-venting) operation, you will
need to loosen the drain (vent) screw on the transmitter
to completely drain (vent) any stagnated liquid (gas).
However, since draining condensate or bleeding off gas
gives the pressure measurement disturbance, this
should not be done when the loop is in operation.
WARNING
Since the accumulated liquid (or gas) may be
toxic or otherwise harmful, take appropriate care
to avoid contact with the body, or inhalation of
vapors.
7.6.2 Venting Gas
1) Gradually open the vent screw to vent gas from the
transmitter pressur-detector section. (See Figure
7.6.2.)
2) When the transmitter is completely vented, close
the vent screw.
3) Tighen the vent screw to a torque of 10 N·m
{1kgf·m}.
Vent screw
Vent screw
When you loosen the vent screw, the gas
escapes in the direction of the arrow.
Figure 7.6.2 Venting the Transmitter
F0711.EPS
7.7 Setting the Range Using the
Range-setting Switch
7.6.1Draining Condensate
1) Gradually open the drain screw or drain plug and
drain the transmitter pressure-detector section. (See
Figure 7.6.1.)
2) When all accumulated liquid is completely removed, close the drain screw or drain plug.
3) Tighten the drain screw to a torque of 10 N·m
{1kgf·m}, and the drain plug to a torque of 34 to 39
N·m {3.5 to 4 kgf·m}.
Drain screw
When you loosen the drain screw or drain
plug, the accumulated loquid will be
expelled in the direction on th earrow.
Figure 7.6.1 Draining the Transmitter
Drain plug
F0710.EPS
With actual pressure being applied to the transmitter,
the range-setting switch (push-button) located on the
optional integral indicator plate and the external zeroadjustment screw allow users to change (re-range) the
low- and high-limit values for the measurement range
(LRV and HRV) without using BT200. However, other
changes in the display settings (scale range and
engineering unit) for the integral indicator requires
BT200.
Follow the procedure below to change the LRV and
HRV settings.
[Example]
Rerange LRV to 0 and HRV to 20 kPa.
1) Connect the transmitter and apparatus as shown in
Figure 9.3.1 and warm up for at least five minutes.
2) Press the range-setting push-button.
The integral indicator then displays “LSET.”
3) Apply a pressure of 0 kPa (atmospheric pressure) to
the transmitter.
(Note 1)
4) Turn the external zero-adjustment screw in the
desired direction. The integral indicator displays the
output signal in %.
(Note 2)
5) Adjust the output signal to 0% (1 V DC) by rotating
the external zero-adjustment screw. Doing so
completes the LRV setting.
7-7
IM 01C22K01-01E
6) Press the range-setting push-button. The integral
indicator then displays “HSET.”
7) Apply a pressure of 20 kPa to the transmitter.
(Note 1)
8) Turn the external zero-adjustment screw in the
desired direction. The integral indicator displays the
output signal in %.
(Note 2)
9) Adjust the output signal to 100% (5 V DC) by
rotating the external zero-adjustment screw. Doing
so completes the HRV setting.
10) Press the range-setting push-button. The transmitter
then switches back to the normal operation mode
with the measurement range of 0 to 20 kPa.
Note 1: Wait until the pressure inside the pressure-detector section
has stabilized before proceeding to the next step.
Note 2: If the pressure applied to the transmitter exceeds the previous
LRV (or HRV), the integral indicator may display error
number “Er.07” (In this case, the output signal percent and
“Er.07” are displayed alternately every two seconds).
Although “Er.07” is displayed, you may proceed to the next
step. However, should any other error number be displayed,
take the appropriate measure in reference to Subsection
8.5.2, “Errors and Countermeasures.”
7. OPERATION
IMPORTANT
• Do not turn off the power to the transmitter
immediately after completion of the change in
the LRV and/or HRV setting(s). Note that
powering off within thirty seconds after setting
will cause a return to the previous settings.
• Changing LRV automatically changes HRV to
the following value.
HRV = previous HRV + (new LRV – previous LRV)
• If the range-setting push-button and external
zero-adjustment screw are not touched during a
range-change operation, the transmitter automatically switches back to the normal operation
mode.
Integral indicator
Note : Use a thin bar which
has a blunt tip, e.g.,
a hexagonal wrench,
to press the rangesetting push-button
Range-setting switch
(Push-button)
Figure 7.7Range-setting Switch
F0712.EPS
7-8
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
8.BRAIN TERMINAL BT200
OPERATION
The DPharp is equipped with BRAIN communications
capabilities, so that range changes, Tag No. setup,
monitoring of self-diagnostic results, and zero point
adjustment can be handled by remote control via
BT200 BRAIN TERMINAL or CENTUM CS console.
This section describes procedures for setting parameters using the BT200. For details concerning the
BT200, see IM 01C00A11-01E, “BT200 User’s
Manual.”
8.1 BT200 Operation Precautions
8.1.1 Connecting the BT200
Connection to the transmitter with the BT200 can be
made by either connecting to the BT200 connection
hooks in the transmitter terminal box or by connecting
to a relaying terminal board.
Relaying
terminals
Control room
Terminal board
8.2
BT200 Operating Procedures
8.2.1 Key Layout and Screen Display
Figure 8.2.1a shows the arrangement of the operating
keys on the BT200 keypad, and Figure 8.2.1b shows
the BT200 screen component.
LCD
(21 character × 8 lines)
Function keys
Movement keys
ENTER key
Power ON/OFF key
Alphanumeric keys
Distributor
F0801.EPS
Figure 8.1.1 Connecting the BT200
8.1.2Conditions of Communication Line
Cable resistance Rc
Power
supply
Load
resistance R
Loop resistance = R + 2Rc
= 250 to 600 Ω
Loop capacitance = 0.22 µF max.
Figure 8.1.2 Conditions of Communication Line
resistance Rc
Cable
cc
DPharp
BT200
F0802.EPS
Figure 8.2.1a BT200 Key Layout
MENU SCREEN
MENU
A:DISPLAY
B:SENSOR TYPE
Screen title
HOMESETADJESC
PARAMETER SCREEN
PARAM
A10:OUTPUT
100.0 %
Parameters
Figure 8.2.1b BT200 Screen Component
A11:ENGR. OUTPUT
1000 mmH20
A20:AMP TEMP
23 deg C
DATADIAGPRNT
8-1
Shift keys
BATTERY
F0803.EPS
Messages
Menu choices
Function
commands
F0804.EPS
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
8.2.2 Operating Key Functions
(1) Alphanumeric Keys and Shift Keys
You can use the alphanumeric keys in conjunction with
the shift keys to enter symbols, as well as alphanumeric keys.
Alphanumeric keys
Shift keys
a. Entering Digits, Symbols, and Spaces
Simply press the alphanumeric keys.
EntryKey-in Sequence
–4
0.3
1 –9
b. Entering Letters (A through Z)
Press an alphanumeric key following a shift key to
enter the letter shown on that side which the shift key
represents. You must press the shift key before
entering each letter.
Left-side letter on the
alphanumeric key
EntryKey-in Sequence
W
IC
J. B
Use the function key [F2]
CAPS
Right-side letter on
the alphanumeric key
to select between
uppercase and lowercase (for letters only). The case
toggles between uppercase and lowercase each time
you press [F2] CAPS.
Entering uppercase
CODECAPSCLRESC
Entry
Boy
( B )( y )( o )
Entering lowercase
CODEcapsCLRESC
Key-in Sequence
to lower case
F0805.EPS
T0801.EPS
F0806 .EPS
T0802.EPS
F0807.EPS
Use the function key [F1]
CODE
to enter symbols. The
following symbols will appear in sequence, one at a
time, at the cursor each time you press [F1] CODE:
/ . – , + * ) ( ’ & % $ # ” !
To enter characters next to these symbols, press [ > ] to
move the cursor.
Entry
symbol command
l/m
Key-in Sequence
( / )
( m )( I )
T0803.EPS
(2) Function Keys
The functions of the function keys depend on the
function commands on display.
MENU
A:DISPLAY
B:SENSOR TYPE
HOMESETADJESC
Function commands
Function keys
F0808.EPS
Function Command List
CommandFunction
ADJDisplays the ADJ menu
CAPS/caps
CODE
CLR
DATA
DEL
DIAG
ESC
HOME
NO
OK
PARM
SET
SLOT
UTIL
*COPY
*FEED
*LIST
*PON/POFF
*PRNT
*GO
*STOP
* Available on BT200-P00 (with printer).
Selects uppercase or lowercase
Selects symbols
Erases input data or deletes all data
Updates parameter data
Deletes one character
Calls the self-check panel
Returns to the most recent display
Displays the menu panel
Quits setup and returns to the previous display
Proceeds to the next panel
Enters the parameter number setup mode
Displays the SET menu
Returns to the slot selection panel
Calls the utility panel
Prints out parameters on display
Paper feed
Lists all parameters in the menu
Automatic printout mode on or off
Changes to the print mode
Starts printing
Cancels printing
T0804.EPS
8-2
IM 01C22K01-01E
8.2.3 Calling Up Menu Addresses Using the Operating Keys
8. BRAIN TERMINAL BT200 OPERATION
STARTUP
SCREEN
INITIAL
DATA
SCREEN
MENU
SCREEN
––WELCOME––
BRAIN TERMINAL
ID: BT200
check connection
push ENTER key
UTILFEED
UTILITY
1.ID
2.SECURITY CODE
3.LANGUAGE SELECT
4.LCD CONTRAST
5.PRINTER ADJUST
The utility screen contains the
following items.
1. BT200 ID settings
2. Security code settings
3. Switching language of messages
(Japanese or English)
esc
4. LCD contrast setting
(UTIL)
PARAM
01:MODEL
EJA115-DM
02:TAG NO.
YOKOGAWA
03:SELF CHECK
GOOD
OK
(ESC)
(SET)
(HOME MENU SCREEN)(SET MENU SCREEN)(ADJ MENU SCREEN)
MENU
A.DISPLAY
B.SENSOR TYPE
HOMESETADJESC
MENU
C.SETTING
D.AUX SET 1
E.AUX SET 2
H:AUTO SET
HOMESETADJESC
5. Adjusting printout tone
(BT200-P00 only)
FUNC
1.MENU
2.UPLOAD TO BT200
3.DOWNLOAD TO INST
4.PRINT ALL DATA
HOMESETADJESC
(ADJ)
MENU
J.ADJUST
K.TEST
M.MEMO
P:RECORD
HOMESETADJESC
PARAMETER
SCREEN
SETUP
SCREEN
PARAM
A60:SELF CHECK
GOOD
PARAM
A21:CAPSULE TEMP
26.5 deg C
DATADIAGPRNTESC
PARAM
A10:OUTPUT(%)
50.0 %
A11:ENGR, OUTPUT
20.0 M
A20:AMP TEMP
24.5 deg C
DATADIAGPRNTESC
SET
C10:TAG NO.
YOKOGAWA
YOKOGAWA
CODECAPSCLRESC
PARAM
C60:SELF CHECK
GOOD
PARAM
C22:HIGH RANGE
100 kPa
DATADIAGPRNTESC
PARAM
C10:TAG NO.
YOKOGAWA
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa
DATADIAGPRNTESC
See “BT200 Instruction Manual” for details
concerning uploading and downloading parameters
and printouts (BT200-P00).
PARAM
J60:SELF CHECK
GOOD
PARAM
J10:ZERO ADJ
0.0 %
DATADIAGPRNTESCDATADIAGPRNTESCDATADIAGPRNTESC
J11:ZERO DEV
22.2 %
J20:EXT. ZERO ADJ
ENABLE
DATADIAGPRNTESC
F0809.EPS
8-3
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
8.3 Setting Parameters Using the BT200
8.3.1Parameter Summary
Instruments to which applicable:
F: Differential pressure transmittersEJA110, EJA120, EJA118W, EJA118N, EJA118Y, and EJA115
P: Pressure transmittersEJA310, EJA430, EJA438W, and EJA438N
L: Liquid level transmittersEJA210 and EJA220
No.Description
Item
MODEL
01
TAG NO.
02
03 SELF CHECK
Model+capsule type
Tag number
Self-diagnostic result
Measured data displayA
Output (in %)A10 OUTPUT (%)
A11 ENGR.
OUTPUT
units)
TEMP
PRESS
A40 INPUT
A60 SELF CHECK
TYPE
Static pressureA30 STATIC
value after zeroing)
Self-diagnostic
messages
Sensor typeB
Model+spanB10 MODEL
Style numberB11 STYLE NO.
Upper range-limitB21 URL
B40
MAX STAT.P.
B60 SELF CHECK
Maximum static
pressure*
Self-diagnostic
messages
6
Setting dataC
Tag numberC10 TAG. NO.
C20 PRESS UNIT
C21 LOW RANGE
Measurement range
units
Measurement range,
lower range value
C22 HIGH
RANGE
Measurement range,
higher range value
Damping time constantC30 AMP
DAMPING
C40 OUTPUT
MODE
C60Same as A60SELF CHECK
Output mode and
integral indicator mode
Self-diagnostic
messages
Auxiliary setting data 1DMenu nameAUX SET 1
Low cutD10 LOW CUT
Low cut modeD11 LOW CUT
MODE
D15 OUT LIMIT(L)–5.0%*
D16 OUT LIMIT(H)110.0%
messages
History of the errorsRECORD
Last errorP10
—
—
Same as A60Self-diagnostic
Display the error
Display the errorOne time beforeERROR REC 2
P12Display the errorTw o time beforeERROR REC 3
P13Display the errorThree time beforeERROR REC 4
P60Same as A60Self-diagnostic
SELF CHECK
messages
—
*1: In case of Model EJA120A, static pressure cannot be measured. The display is always 0 MPa, but this is not a measured value.
*2: When Optional code /F1 is specified, substitute the value with 0.1.
*3: When Optional code /F1 is specified, substitute the value –5 with –2.5.
*4: Not applicable for Model EJA115.
*5: Applicable only for Model EJA118W, EJA118N, EJA118Y, EJA438W, and EJA438N.
*6: See MWP(max. working pressure) on the nameplate. B40 shows an approximate value of maximum pressure for the capsule.
8-5
T0806.EPS
IM 01C22K01-01E
PLF
—
———
—
—
—
—
8. BRAIN TERMINAL BT200 OPERATION
8.3.2Parameter Usage and Selection
Before describing the procedure for setting parameters,
we present the following table showing how the
parameters are used and in what case.
Table 8.3.1 Parameter Usage and Selection
Setup ItemDescription
Tag No. setup
P.8-7
Calibration range setup
P.8-7
Damping time constant setup
P.8-8
Output and integral indicator display
mode setup P.8-9
Output signal low cut mode setup
P.8-9
Change the output limits
P.8-9
Integral indicator scale range and
unit setup
P.8-10
Output status display/setup when
a CPU failure P.8-12
Output status setup when a hardware
error occurs
P.8-12
Range change (while applying
actual inputs)
P.8-13
Zero point adjustment
P.8-13
Span adjustment
P.8-14
Test output (fixed current output)
setup P.8-15
User memo fields
P.8-15
Sets the Tag No. (using 16 alphanumeric characters).
Note: Up to 8 alphanumerics (upper case letters) can be used in the BT100.
Sets the calibration range for 4 to 20 mA DC. Sets three data items: range unit, input value
at 4 mA DC (LRV), and input value at 20 mA DC (HRV).
Note: LRV and HRV can be specified with range value specifications up to 5 digits
(excluding any decimal point) within the range of –32000 to 32000.
Adjusts the output response speed for 4 to 20 mA DC.
Can be set in 9 increments from 0.2 to 64 s.
Sets modes for output signal and integral indicator to “Linear mode”
(proportional to input differential pressure) or to “Square root mode” (proportional to flow).
Used mainly to stabilize output near 0% if output signal is the square root mode.
Two modes are available: forcing output to 0% for input below a specific value, or changing
to proportional output for input below a specific value.
Change the range of normal output.
Sets the following 5 types of integral indicator scale ranges and units:
% scale indicator, user set scale indicator, alternate indication of user set scale
and % scale, input pressure display, alternate indication of input pressure and % scale
When using the user set scale, 4 types of data can be set:
user set scale setting, unit (BT200 only), display value at 4 mA DC (LRV), and display
value at 20 mA DC (HRV).
Note: LRV and HRV can be specified with range value specifications up to 5
digits (excluding any decimal point) within the range of –19999 to 19999.
Sets a unit for temperatures displayed on the BT200.
Sets a unit for static pressure displayed on the BT200.
Reverses the direction for 4 to 20 mA DC output relative to input.
Reverse mode is used for applications in which safety requires that output be
driven toward 20 mA if input is lost.
Displays the status of 4 to 20 mA DC output when a CPU failure. The parameter
of the standard unit is fixed to the high limit value.
Sets the status of the 4 to 20 mA DC output when an abnormal status is detected
with the capsule or the amplifier as the result of self-diagnosis. One of the
following statuses; last held, high limit, and low limit values, can be selected.
Range for 4 to 20 mA DC signal is set with actual input applied. Sets 20 mA DC output
precisely with respect to user’s reference instrument output. Note that DPharp is calibrated
with high accuracy before shipment, so span should be set using the normal range setup.
Adjusts zero point. This can be done either using the external zero-adjustment screw on
the transmitteror using the BT200.
Adjust the characterization curve. All the transmitters are calibrated at factory and this
adjustment is normally not necessary for most cases. Use for specific purposes.
Used for loop checks.
Output can be set freely from –5% to 110% in 1% steps.
Allows user to enter up to 5 items of any desired text in up to 8 uppercase
alphanumeric characters per item.
IMPORTANT
If the transmitter is turned off within 30 seconds
after parameters have been set, the set data will
not be stored and the terminal returns to previous settings.
T0807.EPS
8-6
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
8.3.3Setting Parameters
Set or change the parameters as necessary. After
completing these, do not fail to use the “DIAG” key to
confirm that “GOOD” is displayed for the selfdiagnostic result at _60: SELF CHECK.
(1) Tag No. Setup (C10: TAG NO)
Use the procedure below to change the Tag No. Up to
16 alphanumeric characters can be entered.
• Example: Set a Tag No. to FIC-1a
Press the key to turn on
<When power is off>
––WELCOME––
BRAIN TERMINAL
ID: BT200
check connection
push ENTER key
UTIL FEED
PARAM
01:MODEL
EJA115-DM
02:TAG NO.
YOKOGAWA
03:SELF CHECK
GOOD
MENU
A:DISPLAY
B:SENSOR TYPE
HOMESETADJESC
MENU
C:SETTING
D:AUX SET 1
E:AUX SET 2
H:AUTO SET
HOMESETADJESC
MENU
C10:TAG NO.
YOKOGAWA
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa
DATA DIAG PRNTESC
SET
C10:TAG NO.
YOKOGAWA
YOKOGAWA
CODE CAPSCLRESC
SET
C10:TAG NO.
YOKOGAWA
FIC-1a _
CODE capsCLEESC
the BT200.
Connect DPharp and BT200 using
a communication cable and press
the key.
Displays the name of connected
DPharp model, TAG NO. and
diagnostics information. Press the
(OK) key after confirmation.
OK
Press the (SET) key to
display the SET menu panel.
Select C: SETTING and press the
key.
Select C10: TAG NO. and press
the key.
Set the new TAG NO. (FIC-1a).
FOKOGAWA
FIKOGAWA
FICOGAWA
FIC-GAWA
FIC-1AWA
FIC-1aWA
FIC-1a
Set TAG NO. and press the
key.
SET
C10:TAG NO.
YOKOGAWA
FIC-1a
PRINTER OFF
F2:PRINTER ON
FEED POFFNO
This is the panel for confirming
set data. The set data items flash.
When all items have been confirmed, press the
again. (To go back to the setting
panel, press the (NO) key.
SET
C10:TAG NO.
FIC-1a
The DPharp TAG NO. was
overwritten.
Press the (OK) key to
FEEDNOOK
PARAM
C10:TAG NO.
FIC-1a
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa
DATA DIAG PRNTESC
return to the parameter panel.
Press the (NO) key to
return to the setting panel.
F0811.EPS
(2) Calibration Range Setup
a. Setting Calibration Range Unit
(C20: PRESS UNIT)
The unit is set at the factory before shipment if
specified at the time of order. Follow the procedure
below to change the unit.
• Example: Change the unit from mmH2O to kPa.
SET
C20:PRESS UNIT
mmH20
< mmWG >
< mmHG >
< Torr >
< kPa >
SET
C20:PRESS UNIT
kPa
FEEDNOOK
mmH2O
mmAq
mmWG
mmHg
Torr
kPa
MPa
mbar
bar
2
gf/cm
2
kgf/cm
inH2O
inHg
ftH2O
psi
atm
Pa
hPa
Use the or
key to select “kPa.”
Press the key twice
ESC
to enter the setting.
Press the (OK) key.
F0812.EPS
When you have made an entry mistake, return
the cursor using the key, then reenter.
F0810.EPS
8-7
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
b. Setting Calibration Range Lower
Range Value and Higher Range Value
(C21: LOW RANGE, C22: HIGH
RANGE)
These range values are set as specified in the order
before the instrument is shipped. Follow the procedure
below to change the range.
• The measurement span is determined by the high and
low range limit values. In this instrument, changing
the low range value also automatically changes the
high range value, keeping the span constant.
• Example 1: With present settings of 0 to 30 kPa,
set the lower range value to 0.5 kPa.
SET
C21:LOW RANGE
0 kPa
+ 0.5
DELCLRESC
SET
C21:LOW RANGE
0.5 kPa
FEEDNOOK
Set 0.5.
Press the key twice to
enter the setting.
Press the (OK) key.
• Example 2: With present settings of 0 to 30 kPa,
set the Higher range value to10 kPa.
SET
C22:HIGH RANGE
30 kPa
+ 10
DELCLRESC
SET
C22:HIGH RANGE
10 kPa
FEEDNOOK
PARAM
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa
C22:HIGH RANGE
10 kPa
DATA DIAG PRNTESC
Set 10.
Press the key twice
to enter the setting.
Press the (OK) key.
The low range value is not
changed, so the span changes.
F0814.EPS
(3) Damping Time Constant Setup
(C30: AMP DAMPING)
When the instrument is shipped, the damping time
constant is set at 2.0 seconds. Follow the procedure
below to change the time constant.
SET
C20:PRESS UNIT
kPa
C21:LOW RANGE
0.5 kPa
C22:HIGH RANGE
30.5 kPa
DATA DIAG PRNTESC
The higher range value is changed
while the span remains constant.
Span = Higher range value – Lower range value
F0813.EPS
• Note, however, that changing the higher range value
does not cause the lower range value to change.
Thus, changing the higher range value also changes
the span.
• Calibration range can be specified with range value
specifications up to 5 digits (excluding any decimal
point) for low or high range limits within the range
of –32000 to 32000.
Note1: The damping time constant set here is the damping time
constant for the amplifier assembly. The damping time
constant for the entire transmitter is the sum of the values for
the amplifier assembly and for the capsule assembly. For the
capsule assembly damping time constant (fixed), see the
“General Specifications” found at the end of this manual.
(See Chapter 10.)
Use the or key to
select 4.0 sec.
Press the key twice to
enter the setting.
ESC
Press the (OK) key.
F0815.EPS
8-8
When optional code /F1 is specified, the default value is set
at 0.1 seconds.
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
(4) Output Mode and Integral Indicator Dis-
play Mode Setup (C40: OUTPUT MODE)
The mode setting for the output signal and the integral
indicator coordinate as shown in the table below.
BT200 DisplayOutput Mode
OUT: LIN DSP: LIN
OUT: LIN DSP: SQR
OUT: SQR DSP: SQR
Linear
Linear
Square root
Integral Indicator
Display Mode
Linear
Square root
Square root
T0808.EPS
This mode is set as specified in the order when the
instrument is shipped. Follow the procedure below to
change the mode.
If the instrument is equipped with an integral indicator
and the display mode is “square root”, “
” is dis-
played on the integral indicator.
For details, see Chapter 3.
• Example: Set output mode to Linear and
display mode to Square root.
Use the or key
to select “OUT: LIN, DSP: SQR.”
Press the key twice to
ESC
enter the setting.
Press the (OK) key.
F0816.EPS
(5) Output Signal Low Cut Mode Setup
(D10: LOW CUT, D11: LOW CUT MODE)
Low cut mode can be used to stabilize the output
signal near the zero point. The low cut point can be set
in a range from 0 to 20% of output. (Hysteresis: ±1%)
Either “LINEAR” or “ZERO” can be selected as the
low cut mode.
• Example: Change the low cut setting range
from 10% to 20%, and the low
cut mode from LINEAR to ZERO.
SET
D10:LOW CUT
10.0 %
+ 20.0
CLRESC
SET
D10:LOW CUT
20.0 %
FEEDNOOK
SET
D11:LOW CUT MODE
LINEAR
< LINEAR >
< ZERO >
SET
D11:LOW CUT MODE
ZERO
FEEDNOOK
PARAM
D10:LOW CUT
20.0 %
D11:LOW CUT MODE
ZERO
D20:DISP SELECT
NORMAL %
DATA DIAG PRNTESC
Set “20.”
Press the key twice to
enter the setting.
Press the (OK) key.
Next, the [D11: LOW CUT MODE]
setting panel is displayed.
Use the or key
to select “ZERO.”
Press the key twice to
ESC
enter the setting.
Press the (OK) key.
F0818.EPS
(6) Change Output Limits
(D15:OUT LIMIT(L), D16:OUT LIMIT(H))
The range of normal output is preset at factory from
-5.0 to 110.0% unless otherwise specified, and the
output is limited with these upper and lower values.
This output range can be changed, for example, to
meet the requirements of NAMUR, etc. within the
settable range. Set the lower limit with D15:OUTLIMIT(L) and upper limit with D16:OUT LIMIT(H).
Settable range : -5.0 to 110.0 (%)
Lower limit < Upper limit
Low cut mode “LINEAR”
(%)
50
Output
20
0
Input
Low cut mode “ZERO”
(%)
50
20
50 (%)
050 (%)
LOW CUT at 20%
Input
F0817.EPS
8-9
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
(7) Integral Indicator Scale Setup
The following 5 displays are available for integral
indicators.
D20: DISP SELECT
and Display
NORMAL %
Indicates –5 to 110% range
depending on the Measurement
range (C21, C22).
USER SET
Indicates values depending on the
Engineering range (D22, D23).
Units set using Engineering unit
(D21) are not indicated.
USER & %
Indicates user set and % alternately
in 3 second intervals.
INP PRES
Indicates input pressure.
Indication limits –19999 to 19999.
PRES & %
Indicates input pressure and %
alternately in 3 second intervals.
(Note 1) Scale range can be specified with range limit specifications up
to 5 digits (excluding any decimal point) for low or high range
limits within the range of –19999 to 19999.
The range with decimals is available to the third decimal place.
(Note 2) It indicates the value after zeroing.
Description
and Related parameters
A10:OUTPUT (%)
45.6 %
(Note 1)
A11:ENGR.OUTPUT
20.0 M
A10:OUTPUT (%)
45.6 %
A11:ENGR. OUTPUT
20.0 M
(Note 2)
A40:INPUT
456 kPa
A10:OUTPUT (%)
45.6 %
A40:INPUT
456 kPa
T0809.EPS
See (a.) through (c.) for each setting procedure.
• Example: Change the low cut setting range
from 10% to 20%, and the low
cut mode from LINEAR to ZERO.
SET
D10:LOW CUT
10.0 %
+ 20.0
CLRESC
SET
D10:LOW CUT
20.0 %
FEEDNOOK
SET
D11:LOW CUT MODE
LINEAR
< LINEAR >
< ZERO >
SET
D11:LOW CUT MODE
ZERO
FEEDNOOK
PARAM
D10:LOW CUT
20.0 %
D11:LOW CUT MODE
ZERO
D20:DISP SELECT
NORMAL %
DATA DIAG PRNTESC
Set “20.”
Press the key twice to
enter the setting.
Press the (OK) key.
Next, the [D11: LOW CUT MODE]
setting panel is displayed.
Use the or key
to select “ZERO.”
Press the key twice to
ESC
enter the setting.
Press the (OK) key.
a. Display Selection (D20: DISP SELECT)
Follow the instructions given to the below to change
the range of integral indication scales.
When USER SET is selected, the user set values of
integral indication and A11: ENGR. OUTPUT
parameter are indicated.
• Example: Set the integral indicator scale to
engineering units display.
SET
D20:DISP SELECT
NORMAL %
<NORMAL %>
<USER SET>
<USER & %>
<INP PRES>
SET
D20:DISP SELECT
USER SET
Use the or key
to select “USER SET.”
Press the key twice to
ESC
enter the setting.
Press the (OK) key.
F0818.EPS
8-10
FEEDNOOK
The “%” disappears from the
integral indicator display.
F0820.EPS
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
b. Setting User-set Engineering Unit
(D21: DISP UNIT)
This parameter allows entry of the engineering units to
be displayed on the BT200. When the instrument is
shipped, this is set as specified in the order.
Follow the procedure below to change this setting.
This parameter need not be set for % display.
• Example: Set an engineering unit M.
SET
D21:DISP UNIT
M_
CODE CAPSCLRESC
SET
D21:DISP UNIT
M
FEEDNOOK
Set “M.”
Press the key twice to
enter the setting.
Press the (OK) key.
F0821.EPS
c. Lower and Higher Range Value Setup
in Engineering Unit (D22: DISP LRV,
D23: DISP HRV)
These parameter items are used to set the lower and
higher range values for the engineering unit display.
When the instrument is shipped, these are set as
specified in the order. Follow the procedure below to
change these settings. Note that these parameters need
not be set for % display.
• Example: Set lower range value (LRV) to –50
and higher range value (HRV) to 50.
Setting LRV
SET
D22:DISP LRV
0M
- 50
DELCLRESC
Setting HRV
SET
D23:DISP HRV
100M
+ 50
DELCLRESC
Set “–50.”
Press the key twice to
enter the setting.
Set “50.”
Press the key twice to
enter the setting.
SET
D23:DISP HRV
50M
FEEDNOOK
PARAM
D21:DISP UNT
M
D22:DISP LRV
– 50M
D23:DISP HRV
50M
DATA DIAG PRNTESC
Press the (OK) key.
F0822.EPS
(8) Unit Setup for Displayed Temperature
(D30: TEMP UNIT)
When the instrument is shipped, the temperature units
are set to degC. Follow the procedure below to
change this setting. Note that changing the unit here
changes the unit for A20: AMP TEMP (amplifier
temperature) and A21: CAPSULE TEMP (capsule
temperature).
• Example: Change the unit for the temperature
display.
8-11
SET
D30:TEMP UNIT
deg C
< deg C >
< deg F >
Use the or key to
select “deg F.”
Press the key twice to
ESC
enter the setting.
IM 01C22K01-01E
F0823.EPS
8. BRAIN TERMINAL BT200 OPERATION
(9) Unit Setup for Displayed Static Pressure
(D31: STAT.P.UNIT)
Follow the procedure below to change the static
pressure units.
Changing this parameter changes the unit for the A30:STATIC PRESS (static pressure) display.
• Example: Change the static pressure unit from
2
kgf/cm
SET
D31:STAT.P.UNIT
kgf/cm^2
< MPa >
< mbar >
< bar >
< gf/cm^2 >
mmH2O
mmAq
mmWG
mmHg
Torr
kPa
MPa
mbar
bar
2
gf/cm
2
kgf/cm
inH2O
inHg
ftH
O
2
psi
atm
Pa
hPa
to MPa.
Use the or key
to select “MPa.”
Press the key twice to
ESC
enter the setting.
F0824.EPS
(10) Operation Mode Setup
(D40: REV OUTPUT)
This parameter allows the direction of the 4 to 20 mA
output to be reversed with respect to input. Follow the
procedure below to make this change.
• Example: Change 4 to 20 mA output to 20 to
4 mA output.
SET
D40:REV OUTPUT
NORMAL
< NORMAL >
< REVERSE>
Use the or key
to select REVERSE.
Press the key twice to
ESC
enter the setting.
F0825.EPS
(11) Output Status Display/Setup when a CPU
Failure (D52: BURN OUT)
This parameter displays the status of 4 to 20 mA DC
output if a CPU failure occurs. In case of a failure,
communication is disabled.
Setting of HIGH or LOW is enabled. This is done with
the pin (CN4) on the CPU assembly. See Chapter 3 for
details.
Standard specifications
The parameter is set to HIGH. If a failure, the transmitter outputs the signal of 110% or higher. The parameter
D53: ERROR OUT is set to HIGH from the factory.
Optional code/C1
The parameter is set to LOW. If a failure, output which
1
is –5%*
or lower is generated. The parameter D53:
ERROR OUT is set to LOW from the factory.
*1 : When optional code /F1 is specified, substitute the value with
–5 with –2.5.
• Example: Standard specifications
D52: BURN OUT
HIGH
pin (CN4) position: H
• Example: Optional code/C1
D52: BURN OUT
LOW
pin (CN4) position: L
F0826.EPS
(12) Output Status Setup when a Hardware
Error Occurs (D53: ERROR OUT)
This parameter allows the setting of the output status
when a hardware error occurs. The following three
selections are available.
(a) HOLD; Outputs the last value held before the error
occurred.
(b) HIGH; Outputs an output of 110% when an error
has occurred.
1
(c) LOW; Outputs an output of –5%*
when an error
has occurred.
Note: A hardware error means CAP MODULE FAULT of Er.01 or
AMP MODULE FAULT of Er. 02 which are shown in 8.5.2
“Errors and Countermeasures.”)
*1 : When optional code /F1 is specified, substitute the value with
–5 with –2.5.
• Example: Set the output status to LOW when
a hardware error occurs.
SET
D53:ERROR OUT
HIGH
< HIGH>
< LOW>
< HOLD>
Use the or key
to select “LOW.”
Press the key twice to
ESC
enter the setting.
8-12
F0827.EPS
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
(13) Range Change while Applying Actual
Inputs (H10: AUTO LRV, H11: AUTO HRV)
This feature allows the lower and higher range values
to be set up automatically with the actual input applied.
If the lower and higher range values are set, C21:LOW RANGE and C22: HIGH RANGE are changed
at this same time.
Follow the procedure in the figure below.
The measurement span is determined by the higher and
lower range values. Changing the lower range value
results in the higher range value changing automatically, keeping the span constant.
• Example 1: When changing the lower range
value to 0.5 kPa for the present
setting of 0 to 30 kPa, take the
following action with input pressure
of 0.5 kPa applied.
SET
H10:AUTO LRV
0 kPa
+ 0
SET
H10:AUTO LRV
0.5000 kPa
FEEDNOOK
PARAM
H10:AUTO LRV
0.5000 kPa
H11:AUTO HRV
30.500 kPa
H60:SELF CHEC
GOOD
DATA DIAG PRNTESC
Press the key twice.
The lower range value is changed
to 0.5 kPa.
ESC
Press the (OK) key.
The higher range value is changed
keeping the span constant.
Parameters C21 and C22 are
changed at the same time.
F0828.EPS
Note that changing the higher range value does not
cause the lower range value to change but does change
the span.
• Example 2: When the higher range value is to
be changed to 10 kPa with the
present setting of 0 to 30 kPa, take
the following action with an input
pressure of 10 kPa applied.
SET
H11:AUTO HRV
30 kPa
+ 30
SET
H11:AUTO HRV
10.000 kPa
FEEDNOOK
PARAM
H10:AUTO LRV
0 kPa
H11:AUTO HRV
10.000 kPa
H60:SELF CHECK
GOOD
DATA DIAG PRNTESC
Press the key twice.
The higher range value is changed
to 10 kPa.
ESC
Press the (OK) key.
The lower range value is not
changed, so the span changes.
Parameter C22 is changed at the
same time.
F0829.EPS
(14)Zero Point Adjustment
(J10: ZERO ADJ, J11: ZERO DEV,
J20: EXT ZERO ADJ)
The DPharp supports several adjustment methods.
Select the method best suited for the conditions of your
application.
Note that output signal can be checked by displaying
parameter A10:OUTPUT (%) on the BT200.
Adjustment MethodDescription
Using the BT200Set the present input to 0%.
Adjust for 0% output at input level of
0%.
Adjust output to the reference value
obtained using other means.
If the input level cannot easily be
made 0% (because of tank level,
etc.), adjust output to the reference
value obtained using other means,
such as a sight glass.
Using the external
zero-adjustment
screw
Adjust zero point using the zeroadjustment screw on the transmitter.
This permits zero adjustment without
using the BT200. Accurately adjust
the output current to 4 mA DC or
other target output value using an
ammeter that accuratly reads output
currents.
8-13
T0810.EPS
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
P
A
(a) Follow the procedure below when setting the
present output to 0% (4 mA).
A10:OUTPUT (%)
0.5 %
SET
J10:ZERO ADJ
0.0 %
+ 000.0
CLRESC
SET
J10:ZERO ADJ
0.0 %
FEEDNOOK
A10:OUTPUT (%)
0.0 %
Output is 0.5%.
Press the key twice.
Zero adjustment is completed.
Press the (OK) key.
Output is 0%.
F0830.EPS
(b) Zero Point Adjustment Using the External Zero
Adjustment Screw
• Enabling/inhibiting of zero point adjustment using the
external zero-adjustment screw on the transmitter
(J20: EXT ZERO ADJ)
(15) Span Adjustment
Each DPharp EJA series transmitter is factory characterized according to the specification. Mounting
position effects or zero shifts caused by static pressure
are typically compensated by a zero adjustment.
A span adjustment is a function to correct the slope
error from a zero point in characterizing 100% point
(HRV). This function can be used when span drifts
may be caused or characterization to the specific
pressure standard is required.
Therefore, the zero point adjustment should always be
performed before the upper point adjustment in order
to maintain the pitch between zero and 100% points
within the calibration range.
You can manually perform the trimming procedure by
using J15: SPAN ADJ.
• Span adjustment value
The span adjustment value is calculated as follows.
P
A40
1
Span adjustment value (%) =
: Actual differential pressure/pressure value
1
P
1
100
40: Input (indicated as the value after zeroing)
Follow the procedure below to enable or inhibit zero
Measurement pressure
point adjustment from the zero-adjustment screw on
the transmitter.
This is set to “ENABLE” when the instrument is
A40
shipped.
Applied pressure
F0846.EPS
• Example: Inhibiting zero adjustment by the
P
0
1
external zero-adjustment screw
SET
J20:EXIT ZERO ADJ
ENABLE
< ENABLE >
< INHIBIT>
Use the or key to
select “INHIBIT.”
Press the key twice to
ESC
enter the setting.
F0831.EPS
• Zero point adjustment using external zero-adjustment
screw on the transmitter
Turn the zero-adjustment screw on the outside of the
transmitter case using a slotted screwdriver. Turn the
screw to the right to increase the zero point or to the
left to decrease the zero output; the zero adjusts in
increments of 0.01% of the range setting.
Note that the amount of adjustment to the zero point
changes according to the speed at which the screw is
turned. To make fine adjustments, turn the screw
slowly; to make coarse adjustments, turn the screw
quickly.
Note: When a zero point adjustment has been made, do not turn off
the transmitter less than 30 seconds after adjustment.
8-14
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
• Example: For the range of 0 to 30 kPa.
A40: INPUT = 30.15 kPa
J15: SPAN ADJ = 0.15 %
Suppose that a standard pressure of 30 kPa is applied and
the value of the parameter of A40: INPUT is 30.15 kPa.
Firstly, obtain the slope error for the span as follows;
P
A40
1
Span adjustment value (%) =
P
1
30.0030.15
30.00
100
100 = 0.5 (%)
Add 0.5% to 0.15% of the current value to calculate the
accumulated span adjustment value.
0.15 (0.50) = 0.35
SET
J15:SPAN ADJ
0.15 %
- 0.35
SET
J15:SPAN ADJ
-0.35 %
FEEDNOOK
Set 0.35.
Press key twice.
ESCCLRDEL
Press the (OK) key.
Note: Enter 0.00 to J15: SPAN ADJ to reset the
span adjustment to the initial value at the
shipment.
F0847.EPS
(16)Test Output Setup (K10: OUTPUT X%)
This feature can be used to output a fixed current from
3.2 mA (–5%) to 21.6 mA (110%) for loop checks.
• Example: Output 12 mA (50%) fixed current.
SET
K10:OUTPUT X %
0.0 %
+ 050.0
SET
K10:OUTPUT X %
50.0 % ACTIVE
FEEDNOOK
Note: When optional code /F1 is specified, output range is from
3.6 mA(–2.5%) to 21.6 mA(110%).
Set “50.0%.”
Press the key twice to
output a fixed current at 50%.
ESC
“Active” is displayed while this is
being executed.
Press the (OK) key to cancel
the fixed current output.
F0832.EPS
IMPORTANT
1. Test output is held for approximately 10
minutes, and then released automatically after
the time has elapsed. Even if the BT200 power
supply is turned off or the communication cable
is disconnected during test output, it is held for
approximately 10 minutes.
2. Press the
(OK) key to release test output
immediately.
(17)User Memo Fields (M: MEMO)
This feature provides 5 user memo fields, each holding
up to 8 alphanumeric characters. Up to 5 items such as
inspection date, inspector, and other information can be
saved in these fields.
• Example: Save an inspection date of January
30, 1995.
PARAM
M10:MEMO 1
M20:MEMO 2
M30:MEMO 3
DATA DIAG PRNTESC
SET
M10:MEMO 1
95.1.30_
Set “95.1.30” in the order of year,
month, and day.
Press the key twice to
enter the setting.
ESC
F0833.EPS
8-15
IM 01C22K01-01E
8. BRAIN TERMINAL BT200 OPERATION
8.4 Displaying Data Using the
BT200
8.4.1 Displaying Measured Data
The BT200 can be used to display measured data.
The measured data is updated automatically every 7
seconds. In addition, the display can be updated to the
present data value at any time by pressing the
(DATA) key. For parameters associated with the
display of measured data, see Subsection 8.3.1,
“Parameter Summary.”
The BT200 can be used to display the model and
specifications of the transmitter.
• Example: View transmitter model name.
MENU
A:DISPLAY
B:SENSOR TYPE
Display “A10: OUTPUT (%).”
Data is updated automatically
communi
at 7-second intervals.
Press .
F0834.EPS
8.5 Self-Diagnostics
8.5.1Checking for Problems
(1) Identifying Problems with BT200
The following four areas can be checked.
(a) Whether connections are good.
(b) Whether BT200 was properly operated.
(c) Whether settings were properly entered.
(d) History of the errors.
See examples below.
Press the key.
When the panel shown on the left
appears, press the key.
Since communications will be
unsuccessful if there is a problem
in the connection to the BT200, the
ESC
display at the left will appear.
Recheck the connection.
Press the (OK) key.
The initial data panel shows the
result of current transmitter
diagnostics.
OK
Press the (DIAG) key in the
parameter panel to go to the
diagnostics panel
(C60: SELF CHECK).
An error message is displayed
when an error occurs in the
diagnostics panel.
P10: “ERROR REC 1” displays the last error.
P11: “ERROR REC 2” displays the error one time before
the last error occurred.
P12: “ERROR REC 3” displays the error two times before
the last error occurred.
P13: “ERROR REC 4” displays the error three times before
the last error occurred.
The history of up to four errors can be stored. When the 5th
error has occurred, it is stored in “P10”. The error stored in
“P13” will be deleted, and then, the error in “P12” will be
copied to “P13”. In this sequence, the history of the most
previously occurred error will be removed from memory.
“GOOD” will be displayed if there was no previous error.
Connect the BT200 to the
transmitter, and call item “P.”
Select P10: ERROR REC1 and
press the key to display
the error message.
ESC
(2) Checking with Integral Indicator
NOTE
If an error is detected in the self-diagnostic, an
error number is displayed on the integral indicator. If there is more than one error, the error
number changes at two-second intervals. See
Table 8.5.1 regarding the error numbers.
F0838.EPS
Figure 8.5.1 Identifying Problems Using the Integral
Indicator
For the details of the messages listed below, see Table
8.5.1 Error Message Summary.
CAP MODULE FAULT
AMP MODULE FAULT
OUT OF RANGE
OUT OF SP RANGE
OVER TEMP (CAP)
OVER TEMP (AMP)
OVER OUTPUT
OVER DISPLAY
ILLEGAL LRV
ILLEGAL HRV
ILLEGAL SPAN
ZERO ADJ OVER
Note 1: Press the key twice in the setting panel
(panel 1) to clear all error message (P10 to P13)
information.
Note 2: After two hours from when an error occurs, the error
message of that error will be recorded. Therefore,
if you switch off the transmitter within two hours from
when the error occurs, there is no history of that
error stored in the transmitter, and this function is
meaningless.
F0837.EPS
8-17
IM 01C22K01-01E
8.5.2Errors and Countermeasures
The table below shows a summary of error messages.
Table 8.5.1 Error Message Summary
8. BRAIN TERMINAL BT200 OPERATION
Integral
Indicator
Display
None
----
Er. 01
Er. 02
BT200 DisplayCauseCountermeasure
GOOD
ERROR
CAP MODULE
Capsule problem.
FAULT
AMP MODULE
Amplifier problem.
FAULT
Output Operation
during Error
*1
Outputs the signal
(Hold, High, or Low)
set with parameter
D53.
Outputs the signal
(Hold, High, or Low)
Replace the capsule
when error keeps
appearing even after
*2
restart.
Replace amplifier.
set with parameter
D53.
Er. 03
Er. 04
Er. 05
OUT OF RANGE
OUT OF SP
RANGE
OVER TEMP
(CAP)
Input is outside
measurement range
limit of capsule.
Static pressure
exceeds specified
*3
range.
Capsule temperature
is outside range
(–50 to 130°C).
Outputs high range
limit value or low
range limit value.
Displays present
output.
Displays present
output.
Check input.
Check line pressure
(static pressure).
Use heat insulation or
make lagging to keep
temperature within
range.
Er. 06
OVER TEMP
(AMP)
Amplifier temperature
is outside range
(–50 to 95°C).
Displays present
output.
Use heat insulation or
make lagging to keep
temperature within
range.
Er. 07
Er. 08
OVER OUTPUT
OVER DISPLAY
Output is outside high
or low range limit
value.
Displayed value is
outside high or low
range limit value.
Outputs high or low
range limit value.
Displays high or low
range limit value.
Check input and range
setting, and change
them as needed.
Check input and
display conditions and
modify them as
needed.
Er. 09
ILLEGAL LRV
LRV is outside setting
range.
Holds output
immediately before
Check LRV and
modify as needed.
error occurrence.
Er. 10
ILLEGAL HRV
HRV is outside setting
range.
Holds output
immediately before
Check HRV and
modify as needed.
error occurrence.
Er. 11
ILLEGAL SPAN
SPAN is outside
setting range.
Holds output
immediately before
Check SPAN and
change as needed.
error occurrence.
Er. 12
ZERO ADJ OVER
Zero adjustment is too
large.
Displays present
output.
Readjust zero point.
*1: This error code appears at a capsule problem or when an illegal overpressure is applied to the
pressure sensor.
*2: If the normal pressure is regained, the Er.01 will disappear according to the setting of the parameter
of E50: AUTO RECOVER. When the E50: AUTO RECOVER is set to ON(defalut setting), the Er.01
will disappear automatically. When the E50: AUTO RECOVER is set to OFF, restart the transmitter
to cancel Er.01. If no error code appears then, perform necessary adjustment such as zero-adjust ment to continue the operation. If the error code still exists, replace the capsule assembly.
*3: For Model EJA120A, static pressure cannot be measured. The display is always 0 MPa, but this is
not a measured value.
T0811 .EPS
8-18
IM 01C22K01-01E
9.MAINTENANCE
9. MAINTENANCE
9.1 Overview
WARNING
Since the accumulated process fluid may be
toxic or otherwise harmful, take appropriate care
to avoid contact with the body, or inhalation of
vapors during draining condensate or venting
gas in transmitter pressure-detector section and
even after dismounting the instrument from the
process line for maintenance.
Maintenance of the transmitter is easy due to its
modular construction. This chapter describes the
procedures for calibration, adjustment, and the disassembly and reassembly procedures required for
component replacement.
Since the transmitters are precision instruments,
carefully and thoroughly read the following sections
for proper handling during maintenance.
IMPORTANT
• As a rule, maintenance of this transmitter
should be implemented in a maintenance
service shop where the necessary tools are
provided.
• The CPU assembly contains sensitive parts
that may be damaged by static electricity.
Exercise care so as not to directly touch the
electronic parts or circuit patterns on the board,
for example, by preventing static electrification
by using grounded wrist straps when handling
the assembly. Also take precautions such as
placing a removed CPU assembly into a bag
with an antistatic coating.
9.2 Calibration Instruments Selection
Table 9.2.1 shows the instruments required for calibration. Select instruments that will enable the transmitter
to be calibrated or adjusted to the required accuracy.
The calibration instruments should be handled carefully
so as to maintain the specified accuracy.
9.3 Calibration
Use the procedure below to check instrument operation
and accuracy during periodic maintenance or troubleshooting.
1) Connect the instruments as shown in Figure 9.3.1
and warm up the instruments for at least five
minutes.
IMPORTANT
• To adjust the transmitter for highest accuracy,
make adjustments with the power supply
voltage and load resistance including leadwire
resistances set close to the conditions under
which the transmitter is installed.
• Dismount the manifold assembly(see Subsection 9.4.4) and apply reference pressure on the
high pressure side. (The low pressure side
should be open to atmosphere.)
• Do not perform the calibration procedure until
the transmitter is at room temperature.
2) Apply reference pressures of 0%, 50%, and 100% of
the measurement range to the transmitter. Calculate
the errors (differences between digital voltmeter
readings and reference pressures) as the pressure is
increased from 0% to 100% and is decreased from
100% to 0%, and confirm that the errors are within
the required accuracy.
Note: When the output mode is set to SQRT, apply reference
pressures of 0, 6.25, 25, 56.25, and 100%, instead.
9-1
IM 01C22K01-01E
9. MAINTENANCE
Table 9.2.1 Instruments Required for Calibration
NameRemarksYo kogawa-recommended Instrument
Power
supply
Load
resistor
Voltmeter
Model SDBT or SDBS distributor
Model 2792 standard resistor [250 Ω±0.005%, 3 W]
Load adjustment resistor [100 Ω±1%, 1 W]
Model 2501 A digital multimeter
Accuracy (10V DC range): ±(0.002% of rdg + 1 dgt)
4 to 20 mA DC signal
Model MT220 precision digital manometer
1) For 10 kPa class
for 0 to 10 kPa
Accuracy: ±(0.015% of rdg + 0.015% of F.S.)
±(0.2% of rdg + 0.1% of F.S.)
. . . . .
. . . . . . . . .
for -10 to 0 kPa
2) For 130 kPa class
for 25 to 130 kPa
for 0 to 25 kPa
for -80 to 0 kPa
for 100 to 700 kPa
for 0 to 100 kPa
for -80 to 0 kPa
Select a manometer having
a pressure range close to
that of the transmitter.
Model 7674 pneumatic pressure standard for 200 kPa {2 kgf/cm
Accuracy: ±0.05% of F.S.
Dead weight gauge tester 25 kPa {2500mmH2O}
Accuracy: ±0.03% of setting
that of the transmitter.
Pressure
source
Model 6919 pressure regulator (pressure pump)
Pressure range: 0 to 133 kPa {1000 mmHg}
Prepare the vacuum pump
for negative pressure
ranges.
T0901.EPS
Note: The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and management
procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1% level, there are
difficulties in calibration to this level in the field. For calibration to the 0.1% level, contact Yokogawa representatives from which the
instrument was purchased or the nearest Yokogawa office.
Using pressure
generator
P
Low pressure side
open to atmosphere
Figure 9.3.1 Instrument Connections
Supply pressure
Using pressure
Pressure source
source with
Pressure generator
Reference pressure
High pressure sideHigh pressure side
Load resistance,
250 Ω
Load
adjusting
resistance,
100 Ω
RRc
V
Digital voltmeter
manometer
Low pressure side
open to atmosphere
Power
supply
E
Model MT220
P
precision digital manometer
Reference pressure
Load resistance,
250 Ω
Load
adjusting
resistance,
100 Ω
RRc
V
Digital voltmeter
9-2
Power
supply
E
F0901.EPS
IM 01C22K01-01E
9. MAINTENANCE
9.4 Disassembly and Reassembly
This section describes procedures for disassembly and
reassembly for maintenance and component replacement.
Always turn OFF power and shut off and release
pressures before disassembly. Use proper tools for all
operations. Table 9.4.1 shows the tools required.
Table 9.4.1 Tools for Disassembly and Reassembly
ToolRemarksQuantity
Phillips screwdriver
Slotted screwdriver
Allen wrenches
Wrench
Torque wrench
Adjustable wrench
Socket wrench
Socket driver
Tweezers
CAUTION
Precautions for CENELEC, IECEx, and TIIS
Flameproof Type Transmitters
• Flameproof type transmitters must be, as a
rule, removed to a non-hazardous area for
maintenance and be disassembled and reassembled to the original state. For details, see
“Installation and Operating Precautions for TIIS
Flameproof Equipment” later in this manual.
• On the flameproof type transmitters the two
covers are locked, each by an Allen head bolt
(shrouding bolt). When a shrouding bolt is
driven clockwise by an Allen wrench, it is going
in and cover lock is released, and then the
cover can be opened.
When a cover is closed it should be locked by a
shrouding bolt without fail. Tighten the shrouding bolt to a torque of 0.7 N·m.
1
JIS B4633, No. 2
1
2
JIS B4648
One each, nominal 3 and
5 mm Allen wrenches
1
Width across flats, 17 mm
1
1
1
Width across flats, 16 mm
1
Width across flats, 5.5 mm
1
T0902.EPS
9.4.1 Replacing the Integral Indicator
CAUTION
Cautions for TIIS Flameproof Type Transmitters
Users are prohibited by law from modifying the
construction of a flameproof type transmitter.
This would invalidate the agency approval and
the transmitter’s use in such rated area.
Thus the user is prohibited from using a flameproof type transmitter with its integral indicator
removed, or from adding an integral indicator to
a transmitter. If such modification is absolutely
required, contact Yokogawa.
This subsection describes the procedure for replacing
an integral indicator. (See Figure 9.4.2)
Removing the Integral Indicator
1) Remove the cover.
2) Supporting the integral indicator by hand, loosen its
two mounting screws.
3) Dismount the LCD board assembly from the CPU
assembly.
When doing this, carefully pull the LCD board
assembly straight forward so as not to damage the
connector pins between it and the CPU assembly.
Attaching the Integral Indicator
Integral indicator can be installed in the following three
directions.
F0902-02.EPS
Figure 9.4.1 Installation Direction of Indicator
Shrouding Bolt
Figure 9.4 Shrouding Bolts
Shrouding Bolt
F0902.EPS
9-3
IM 01C22K01-01E
9. MAINTENANCE
1) Align both the LCD board assembly and CPU
assembly connectors and engage them.
2) Insert and tighten the two mounting screws.
3) Replace the cover.
Output terminal cable
Press
forward
LCD board
assembly
Integral
indicator
Boss
Bracket
(for zero-adjustment
screw pin)
Cover
Figure 9.4.2 Removing and Attaching LCD Board Assem-
Mounting
screw
bly and CPU Assembly
Flat cable
CPU assembly
Zero-adjustment
screw pin
F0903.EPS
9.4.2 Replacing the CPU Board Assembly
Mounting the CPU Assembly
1) Connect the flat cable (with black connector)
between the CPU assembly and the capsule.
2) Connect the output terminal cable (with brown
connector).
NOTE
Make certain that the cables are free of pinching
between the case and the CPU assembly edge.
3) Align and engage the zero-adjustment screw pin
with the groove on the bracket on the CPU assembly. Then insert the CPU board assembly straight
onto the post in the amplifier case.
4) Tighten the two bosses. If the transmitter is
equipped with an integral indicator, refer to Subsection 9.4.1 to mount the indicator.
NOTE
Confirm that the zero-adjustment screw pin is
placed properly in the groove on the bracket
prior to tightening the two bosses. If it is not, the
zero-adjustment mechanism will be damaged.
This subsection describes the procedure for replacing
the CPU assembly. (See Figure 9.4.2)
Removing the CPU Assembly
1) Remove the cover. If an integral indicator is
mounted, refer to Subsection 9.4.1 and remove the
indicator.
2) Turn the zero-adjustment screw to the position
(where the screw head slot is horizontal) as shown
in Figure 9.4.2.
3) Disconnect the output terminal cable (cable with
brown connector at the end). When doing this,
lightly press the side of the CPU assembly connector and pull the cable connector to disengage.
4) Use a socket driver (width across flats, 5.5mm) to
loosen the two bosses.
5) Carefully pull the CPU assembly straight forward to
remove it.
6) Disconnect the flat cable (cable with black connector at the end) that connects the CPU assembly and
the capsule.
NOTE
5) Replace the cover.
9.4.3 Replacing the Process Connector
Gaskets
This subsection describes process connector gasket
replacement. (See Figure 9.4.3.)
(a) Loosen the two bolts, and remove the process
connectors.
(b)Replace the process connector gaskets.
(c) Remount the process connectors. Tighten the bolts
securely and uniformly with a torque of 39 to 49
N·m {4 to 5 kgf·m}, and verify that there are no
pressure leaks.
Process connector gasket
Process connector
Bolt
F0904.EPS
Be careful not to apply excessive force to the
CPU assembly when removing it.
Figure 9.4.3 Removing and Mounting the Process
Connector
9-4
IM 01C22K01-01E
9.4.4 Cleaning Manifold Assembly and
Replacing Orifice
9. MAINTENANCE
NOTE
This subsection describes the procedures for cleaning
the manifold assembly and replacing the orifice to
change flow rate. (See Figure 9.4.4.)
䊏 Removing the Manifold Asssembly
1) Remove the process connector as shown in Subsec-
tion 9.4.3.
2) Remove the four bolts that connect the cover flange
with the manifold.
3) Remove the spacer, orifice, and orifice gasket from
inside the manifold.
4) Clean the manifold, spacer, and orifice, or replace
them as necessary.
IMPORTANT
Exercise care as follows when cleaning the
manifold assembly.
• Handle the manifold assembly with care, and
be careful not to damage the inner part of the
manifold, spacer, and orifice. Be especially
careful not to damage or distort the orifice edge
(orifice bore).
• Do not use a chlorinated or acidic solution for
cleaning.
• Rinse thoroughly with clean water after cleaning and dry thoroughly.
Exercise care as follows when reassembling the
manifold assembly. (See Figure 9.4.4.)
• Be careful not to reassemble the orifice in the
wrong direction. Note that the spacer is configured so that it cannot be placed in the reverse
direction.
• When mounting the manifold on the cover
flange, confirm the indication “flow direction”
shown on the manifold surface and the high
and low pressure sides of the pressuredetector section.
Mount the manifold so that the upstream side
of process fluid flow is located at the high
pressure side of the pressure-detection section.
Spacer
Orifice
Process connector
gasket
Orifice gasket
Manifold gasket
Low pressure side
Bolt
Manifold
Process connector
gasket
Cover flange
䊏 Reassembling the Manifold Assembly
1) Reassemble the orifice gasket, orifice, and spacer
into the manifold in that order.
When reassembling, refer to Figure 9.4.4 to ensure
that they are placed in the correct direction.
Replace the orifice gasket with a new gasket.
2) Mount the process connector as shown in Subsection 9.4.3.
3) Mount the manifold on the cover flange with the
four bolts. Tighten the four bolts uniformly to a
torque of 39 to 49 N·m {4 to 5 kgf·m}.
Replace the manifold gaskets with new gaskets.
4) After completing reassembly, a leak test must be
performed to verify that there are no pressure leaks.
High pressure side
F0905.EPS
Figure 9.4.4 Manifold Assembly
9-5
IM 01C22K01-01E
9. MAINTENANCE
9.4.5 Cleaning and Replacing the Capsule
Assembly
This subsection describes the procedures for cleaning
and replacing the capsule assembly. (See Figure 9.4.5.)
CAUTION
Cautions for TIIS Flameproof Type Transmitters
Users are prohibited by law from modifying the
construction of a flameproof type transmitter. If
you wish to replace the capsule assembly with
one of a different measurement range, contact
Yokogawa.
The user is permitted, however, to replace a
capsule assembly with another of the same
measurement range. When doing so, be sure to
observe the following.
• The replacement capsule assembly must have
the same part number as the one being replaced.
• The section connecting the transmitter and
capsule assembly is a critical element in
preservation of flameproof performance, and
must be checked to verify that it is free of
dents, scratches, and other defects.
• After completing maintenance, be sure to
securely tighten the Allen screws that fasten
the transmitter section and pressure-detector
section together.
1) Remove the CPU assembly as shown in Subsection
9.4.2.
2) Remove the two Allen screws that connect the
transmitter section and pressure-detector section.
3) Separate the transmitter section and pressuredetector section.
4) Remove the nuts from the four flange bolts.
5) Hold the capsule assembly by hand and remove the
cover flange.
6) Remove the capsule assembly.
7) Clean the capsule assembly or replace with a new
one.
Reassembling the Capsule Assembly
1) Insert the capsule assembly between the flange
bolts, paying close attention to the relative positions
of the H (high pressure side) and L (low pressure
side) marks on the capsule assembly.
Replace the two capsule gaskets with new gaskets.
2) Install the cover flange on the high pressure side,
and use a torque wrench to tighten the four nuts
uniformly to a torque of 39 N·m {4 kgf·m}.
3) After the pressure-detector section has been reassembled, a leak test must be performed to verify
that there are no pressure leaks.
4) Reattach the transmitter section to the pressuredetector section.
5) Tighten the two Allen screws. (Tighten the screws
to a torque of 5 N·m)
6) Install the CPU assembly according to Subsection
9.4.2.
7) After completing reassembly, adjust the zero point
and recheck the parameters.
Removing the Capsule Assembly
IMPORTANT
Exercise care as follows when cleaning the
capsule assembly.
• Handle the capsule assembly with care, and be
especially careful not to damage or distort the
diaphragms that contact the process fluid.
• Do not use a chlorinated or acidic solution for
cleaning.
• Rinse thoroughly with clean water and dry
thoroughly after cleaning.
Capsule assembly
Capsule
Cover flange
Flange bolt
Figure 9.4.5 Removing and Mounting the Pressure-
detector Section
gasket
Nut
9-6
Allen screw
F0906.EPS
IM 01C22K01-01E
9. MAINTENANCE
9.5 Troubleshooting
If any abnormality appears in the measured values, use
the troubleshooting flow chart below to isolate and
remedy the problem. Since some problems have
complex causes, these flow charts may not identify all.
If you have difficulty isolating or correcting a problem,
contact Yokogawa service personnel.
9.5.1 Basic Troubleshooting
First determine whether the process variable is actually
abnormal or a problem exists in the measurement
system.
If the problem is in the measurement system, isolate
the problem and decide what corrective action to take.
This transmitter is equipped with a self-diagnostic
function which will be useful in troubleshooting; see
Section 8.5 for information on using this function.
: Areas where self-diagnostic offers support
Abnormalities appear in measurement.
YES
Inspect the
process system.
Is process variable
itself abnormal?
NO
Measurement system problem
9.5.2 Troubleshooting Flow Charts
The following sorts of symptoms indicate that transmitter
may not be operating properly.
Example : • There is no output signal.
• Output signal does not change even though
process variable is known to be varying.
• Output value is inconsistent with value
inferred for process variable.
Connect BRAIN TERMINAL and check self-diagnostics.
Does the self-diagnostic
indicate problem location?
NO
Is power supply
polarity correct?
YES
Are power
supply voltage and load
resistance correct?
YES
Are valves opened or
closed correctly?
YES
Refer to error message summary in
Subsection 8.5.2 to take actions.
Refer to Section 6.3 to check/correct
polarity at each terminal from power
supply to the terminal box.
Refer to Section 6.6 for rated voltage
and load resistance.
Fully open the valves on the upstream
and downstream sides.
YES
NO
NO
NO
Isolate problem in
measurement system.
YES
Inspect receiver.
Environmental conditions
Check/correct
environmental conditions.
Does problem exist in
receiving instrument?
NO
Transmitter itself
Check transmitter.
Operating conditions
Check/correct operating
conditions.
Figure 9.5.1 Basic Flow and Self-Diagnostics
F0907.EPS
Is there any pressure leak?
Fix pressure leaks, paying particular
NO
Is there
continuity through the
transmitter loop wiring?
Do the loop numbers
match?
YES
Is orifice clogged with
dust, etc?
NO
Contact Yokogawa service personnel.
attention to connections for process
piping, pressure-detector section, etc.
Find/correct broken conductor or
wiring error.
Refer to Subsection 9.4.4 to clean the
manifold assembly.
YES
NO
YES
F0908.EPS
9-7
IM 01C22K01-01E
9. MAINTENANCE
Output travels beyond 0% or 100%.
Connect BRAIN TERMINAL and check self-diagnostics.
Does the self-
diagnostic indicate problem
location?
NO
Is power supply
polarity correct?
YES
Are valves opened or
closed correctly?
YES
Is there any pressure leak?
NO
Refer to error message summary in
Subsection 8.5.2 to take actions.
Refer to Section 6.3 to check/correct
polarity at each terminal from power
supply to the terminal box.
Fully open the valves on the upstream
and downstream sides.
Fix pressure leaks, paying particular
attention to connections for impulse
piping, pressure-detector section, etc.
YES
NO
NO
YES
Large output error.
Connect BRAIN TERMINAL and check self-diagnostics.
Does the self-
diagnostic indicate problem
YES
location?
Refer to error message summary in
NO
Are valves opened or
Subsection 8.5.2 to take actions.
NO
closed correctly?
Fully open the valves on the upstream
YES
Is the manifold
mounted appropriately for
and downstream sides.
NO
the flow direction?
YES
Refer to Subsection 9.4.4 and mount
the manifold appropriately for the flow
direction of fluid.
Are power supply
NO
voltage and load resistance
correct?
Refer to section 6.6 for rated voltage
YES
and load resistance.
Is the transmitter
installed as appropriate for
the flow direction?
Refer to Subsection 5.1.1 and install
YES
Is zero point
adjusted correctly?
YES
Contact Yokogawa service personnel.
the transmitter as appropriate for the
flow direction of fluid.
Adjust the zero point.
NO
NO
F0909.EPS
Is transmitter
installed where there is
marked variation in
temperature?
Provide lagging and/or cooling, or allow
NO
adequate ventilation.
Were appropriate
instruments used for
calibration?
Refer to Section 9.2 when selecting
YES
instruments for calibration.
Is output adjusted correctly?
YES
Adjust the output.
Is orifice clogged with
dust, etc?
YES
Refer to Subsection 9.4.4 to clean the
manifold assembly.
Contact Yokogawa service personnel.
YES
NO
NO
NO
F0910.EPS
9-8
IM 01C22K01-01E
10. GENERAL SPECIFICATIONS
10.GENERAL SPECIFICATIONS
10.1Standard Specifications
Refer to IM 01C22T02-01E for FOUNDATION
Fieldbus communication type and IM 01C22T0300E for PROFIBUS PA communication type
marked with “䉫”.
䊉 Performance Specifications
See General Specifications sheet, GS 01C22K0100E.
䊉 Functional Specifications
Span & Range Limits
Differential
Pressure Span
L
Capsule
M
Capsule
H
Capsule
Measurement
Range
L
Capsule
M
Capsule
H
Capsule
kPa
1 to 10
2 to 100
20 to 210
Water Equivalent
0.022 to 23.0
inH2O
(/D1)
4 to 40
8 to 400
80 to 830 200 to 2100
Flow l/min
0.016 to 7.2
0.07 to 33.02.0 to 910
mbar
(/D3)
10 to 100
20 to 1000
Zero Adjustment Limits:
Zero can be fully elevated or suppressed, within
the Lower and Upper Range Limits of the capsule.
External Zero Adjustment “䉫”:
External zero is continuously adjustable with
0.01% incremental resolution of span. Span may
be adjusted locally using the digital indicator with
range switch.
Output “䉫”:
Two wire 4 to 20 mA DC output with digital
communications, linear or square root programmable. BRAIN or HART FSK protocol are superimposed on the 4 to 20 mA signal.
Failure Alarm:
Output status at CPU failure and hardware error;
Up-scale: 110%, 21.6 mA DC or more (standard)
Down-scale: –5%, 3.2 mA DC or less
–2.5%, 3.6 mA DC or less (Optional
code /F1)
Note: Applicable for Output signal code D and E
mmH
(/D4)
100 to 1000
200 to 10000
0.05 to
5 kgf/cm
Air Equivalent
Flow Nl/min
0.44 to 198
0.63 to 635
O
2
T1001.EPS
Damping Time Constant (1st order):
The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.
Capsule (Silicone Oil)
Time Constant (approx. sec)0.4 0.3 0.3
LMH
T1002.EPS
Ambient Temperature Limits:
* Safety approval codes may affect limits.
–40 to 85°C (–40 to 185°F)
–30 to 80°C (–22 to 176°F) with LCD Display
Process Temperature Limits:
* Safety approval codes may affect limits.
–40 to 120°C (–40 to 248°F)
Working Pressure Limits (Silicone Oil)
2.7 kPa abs {20 mmHg abs} to maximum working
pressure. See ‘Model and Suffix Codes.’
2
䊉 Installation
Supply & Load Requirements “䉫”:
* Safety approvals can affect electrical requirements.
See Section 6.6, ‘Power Supply Voltage and Load
Resistance.’
Supply Voltage “䉫”:
10.5 to 42 V DC for general use and flameproof type
10.5 to 32 V DC for lightning protector (Optional code
/A)
10.5 to 30 V DC for intrinsically safe, Type n,
nonincendive, or non-sparking type
10.5 to 28 V DC for TIIS intrinsically safe type
EMC Conformity Standards “䉫”: ,
EN61326-1 Class A, Table 2 (For use in industrial
lications)
EN61326-2-3
Communication Requirements “䉫”:
BRAIN
Communication Distance;
Up to 2 km (1.25 miles) when using CEV polyethylene-insulated PVC-sheathed cables.
Communication distance varies depending on type
of cable used.
Load Capacitance;
0.22 µF or less (see note)
Load Inductance;
3.3 mH or less (see note)
Input Impedance of communicating device;
10 kΩ or more at 2.4 kHz.
Note: For general-use and Flameproof type.
For Intrinsically safe type, please refer to
‘Optional Specifications.’
10-1
IM 01C22K01-01E
HART
(
)
Communication Distance;
Up to 1.5 km (1 mile) when using multiple twisted
pair cables. Communication distance varies
depending on type of cable used.
Use the following formula to determine cable
length for specific applications:
6
65 x 10
L=-
R x C
(Cf + 10,000)
C
100{750}
Working
pressure
kPa abs
{mmHg abs}
10{75}
10. GENERAL SPECIFICATIONS
Atmospheric
pressure
Applicable range
Where:
L = length in meters or feet
R = resistance in Ω (including barrier resistance)
C = cable capacitance in pF/m or pF/ft
Cf = maximum shunt capacitance of receiving
devices in pF/m or pF/ft
䊉 Physical Specifications
Wetted Parts Materials:
Diaphragm, Cover flange, Process connector,
Manifold, Orifice, and Drain/Vent Plug;
See ‘Model and Suffix Codes’
Capsule Gasket;
Teflon-coated SUS316L
Process Connector Gasket;
PTFE Teflon
Non-wetted Parts Materials:
Bolting;
SCM435 or SUS630
Housing;
Low copper cast-aluminum alloy with polyurethane
paint (Munsell 0.6GY3.1/2.0)
Degrees of Protection;
IP67, NEMA4X, JIS C0920 immersion proof
Cover O-rings;
Buna-N
Data plate and tag;
SUS304 or SUS316 (option)
Fill Fluid;
Silicone or Fluorinated oil (option)
Weight:
5.6 kg (12.3 lb) without mounting bracket
Connections:
Refer to the ‘Model and Suffix Codes’ to specify
the process and electrical connection type.
2.7{20}
1{7.5}
-40
(-40)
Figure 1. Working Pressure and Process Temperature
0
(32)40(104)80(176)
Process temperature °C (°F)
120
(248)
F1001.EPS
<Settings When Shipped “䉫”>
Tag Number
Output Mode
Display Mode
Operation Mode
Damping Time
Constant
As specified in order
‘Linear’ unless otherwise specified in order
‘Square root’
‘Normal’ unless otherwise specified in order
‘2 sec.’
Note 1: If Tag No. is no more than 16 alphanumeric characters (including - and ·), it will be written into
*1
the tag plate and amplifier memory settings.
Calibration Range
Lower Range Value
Calibration Range
Higher Range Value
Calibration Range
Units
10-2
As specified in order
As specified in order
Selected from mmH2O, mmAq, mmWG,
mmHg, Pa, hPa, kPa, MPa, mbar, bar,
2
, kgf/cm2, inH2O, inHg, ftH2O, or
gf/cm
psi. (Only one unit can be specified)
T1003.EPS
IM 01C22K01-01E
10. GENERAL SPECIFICATIONS
10.2Model and Suffix Codes
䊉 Model EJA115[Style: S3]
Model
EJA115
Output Signal
Measurement span
(capsule)
Wetted parts material
Process flange rating
–
Bolts and nuts material
Installation
Electrical connection
Integral indicator
Mounting bracket
Optional codes
Suffix Codes
. . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . .
-D
. . . . . . . . . . . . . . . . . . . . .
-E
. . . . . . . . . . . . . . . . . . . . .
-F
. . . . . . . . . . . . . . . . . . . . .
-G
. . . . . . . . . . . . . . . . . .
L
. . . . . . . . . . . . . . . . . .
M
. . . . . . . . . . . . . . . . . .
H
. . . . . . . . . . . . . . . . .
S
. . . . . . . . . . . . . . . .
2
. . . . . . . . . . . . . . . .
4
00
. . . . . . . . . . . . . . .
Low Flow transmitter
4 to 20 mA DC with digital communication (BRAIN protocol)
4 to 20 mA DC with digital communication (HART protocol)
Digital communication (FOUNDATION Fieldbus protocol)
Digital communication (PROFIBUS PA protocol)
1 to 10 kPa {100 to 1000 mmH2O}
2 to 100 kPa {200 to 10000 mmH
20 to 210 kPa {2000 to 21000 mmH
(Note 3)
[Body]
[Capsule] [Orifice]
JIS SCS14A JIS SUS316L
Rc1/2 female
1/2 NPT female
Always 00
Description
O}
2
O}
2
(Note 2)
JIS SUS316
(Note 1)
(Note 4)
(Note 5)
[Maximum working pressure]
(L capsule) (M, H capsule)
A
B
-2
-3
-6
-7
-8
-9
0
2
3
4
5
7
8
9
A
C
D
D
E
N
A
B
J
C
D
K
N
. . . . . . . . . . . . . .
. . . . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
JIS SCM435 3.5 MPa {35 kgf/cm
JIS SUS630 3.5 MPa {35 kgf/cm
Vertical impulse piping type, right side high pressure, manifold upside
Vertical impulse piping type, right side high pressure, manifold downside
Vertical impulse piping type, left side high pressure, manifold upside
Vertical impulse piping type, left side high pressure, manifold downside
Horizontal impulse piping type, right side high pressure
Horizontal impulse piping type, left side high pressure
G1/2 female, one electrical connection
1/2 NPT female, two electrical connections without blind plug
Pg 13.5 female, two electrical connections without blind plug
M20 female, two electrical connections without blind plug
G1/2 female, two electrical connections and a blind plug
1/2 NPT female, two electrical connections and a blind plug
Pg 13.5 female, two electrical connections and a blind plug
M20 female, two electrical connections and a blind plug
G1/2 female, two electrical connections and a SUS316 blind plug
1/2 NPT female, two electrical connections and a SUS316 blind plug
M20 female, two electrical connections and a SUS316 blind plug
Digital indicator
Digital indicator with the range setting switch
(None)
JIS SECC 2-inch pipe mounting (flat type)
JIS SUS304 2-inch pipe mounting (flat type)
JIS SUS316 2-inch pipe mounting (flat type)
JIS SECC 2-inch pipe mounting (L type)
JIS SUS304 2-inch pipe mounting (L type)
JIS SUS316 2-inch pipe mounting (L type)
(None)
/䊐
Optional specification
2
} 14 MPa {140 kgf/cm2}
2
} 14 MPa {140 kgf/cm2}
Example: EJA115-DMS400A-92NN/䊐
Note 1: Refer to IM 01C22T01-01E for HART Protocol version.
Note 2: Indicates other wetted parts materials. Diaphragm material is Hastelloy C-276.
Note 3: Indicates material of cover flanges and process connectors. Manifold and vent plugs material
are JIS SUS316.
Note 4: Refer to IM 01C22T02-01E for Fieldbus communication.
Note 5: Refer to IM 01C22T03-00E for PROFIBUS PA communication.
T1004.EPS
10-3
IM 01C22K01-01E
10.3Optional Specifications
For FOUNDATION Fieldbus explosion protected type, see IM 01C22T02-01E.
For PROFIBUS PA explosion protected type, see IM 01C22T03-00E.
10. GENERAL SPECIFICATIONS
Item
Factory Mutual (FM)
CENELEC ATEX
Description
*1
FM Explosionproof Approval
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Hazardous (classified) locations, indoors and outdoors (NEMA 4X)
Division 2, ‘SEALS NOT REQUIRED’,
Temp. Class: T6
Amb. Temp.: –40 to 60C (–40 to 140F)
FM Intrinsically safe Approval
*1
Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1,
Groups E, F & G and Class III, Division 1 Hazardous Locations.
Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division. 2,
Groups E, F & G, and Class III, Division 1 Hazardous Locations.
Enclosure: “NEMA 4X”, Temp. Class: T4, Amb. Temp.: –40 to 60C (–40 to 140F)
Intrinsically Safe Apparatus Parameters
[Groups A, B, C, D, E, F and G]
Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H
[Groups C, D, E, F and G]
Vmax=30 V, Imax=225 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H
Combined FF1 and FS1
CENELEC ATEX (KEMA) Flameproof Approval
*1
*2
Certificate: KEMA 02ATEX2148
II 2G EExd IIC T4, T5, T6
Amb. Temp.: T5; –40 to 80C ( –40 to 176F), T4 and T6; –40 to 75C ( –40 to 167F)
Max. process Temp.: T4; 120C (248F), T5; 100C (212F), T6; 85C (185F)
CENELEC ATEX (KEMA) Intrinsically safe Approval
*2
Certificate: KEMA 02ATEX1030X
II 1G EEx ia IIC T4, Amb. Temp.: –40 to 60C (–40 to 140F)
Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 H
Combined KF2, KS2 and Type n
*2
Type n
II 3G Ex nL IIC T4, Amb. Temp.: –40 to 60°C (–40 to 140°F)
Ui=30 V DC, Ci=22.5 nF, Li=730 µH
Dust
II 1D maximum surface temperature T65°C (149F) {Tamb.: 40°C (104F)},
T85°C (185F) {Tamb.: 60°C (140F)}, T105°C (221F) {Tamb.: 80°C (176F)}
*1:Applicable for Electrical connection code 2 and 7 (1/2 NPT female).
*2:Applicable for Electrical connection code 2, 4, 7 and 9 (1/2 NPT and M20 female).
Code
FF1
FS1
FU1
KF2
KS2
KU2
T1005-1.EPS
10-4
IM 01C22K01-01E
10. GENERAL SPECIFICATIONS
Item
Canadian Standards
Association (CSA)
IECEx
Scheme
TIIS certification
Attached flameproof
packing adapter
Description
CSA Explosionproof Approval
*1
Certificate: 1089598
Explosionproof for Class I, Division 1, Groups B, C and D
Dustignitionproof for Class II/III, Division 1, Groups E, F and G
Division2 ‘SEALS NOT REQUIRED’ , Temp. Class: T4, T5, T6 Encl Type 4x
Max. Process Temp.: T4; 120C (248F), T5; 100C (212F), T6; 85C (185F)
Amb. Temp.: –40 to 80C (–40 to 176F)
Process Sealing Certification
Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
No additional sealing required. Primary seal failure annunciation: at the zero adjustment
screw
CSA Intrinsically safe Approval
*1
Certificate: 1053843
Intrinsically Safe for Class I, Groups A, B, C and D Class II and III, Groups E, F and G
Nonincendive for Class I, Division 2, Groups A, B, C and D Class II, Division 2,
Groups F and G and Class III (not use Safety Barrier)
Encl Type 4x, Temp. Class: T4, Amb. Temp.: –40 to 60C (–40 to 140F)
Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H
Process Sealing Certification
Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
No additional sealing required. Primary seal failure annunciation: at the zero adjustment
screw
*1
Combined CF1 and CS1
IECEx Intrinsically safe, type n and Flameproof Approval
*2
Intrinsically safe and type n
Certificate: IECEx KEM 06.0007X
Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP67
Amb. Temp.: –40 to 60C (–40 to 140F), Max. Process Temp.: 120C (248F)
Electrical Parameters: [Ex ia] Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 H
[Ex nL] Ui=30 V, Ci=22.5 nF, Li=730 H
Flameproof
Certificate: IECEx KEM 06.0005
Ex d IIC T6...T4 Enclosure: IP67
Max.Process Temp.: T4;120C (248F), T5;100C (212F), T6; 85C (185F)
Amb.Temp.: –40 to 75C (–40 to 167F) for T4, –40 to 80C (–40 to 176F) for T5,
–40 to 75C (–40 to 167F) for T6
TIIS Flameproof Approval, Ex do IIC T4X
Certificate: C15296 (Without integral indicator)
C15297 (With integral indicator)
Amb. Temp.: –20 to 60C, Process Temp.: –20 to 120C
TIIS Intrinsically safe Approval, Ex ia IIC T4
Certificate: C14632
Amb. Temp.: –20 to 60C, Process Temp.: –20 to 120C
Electrical connection: G1/2 female
Applicable cable: O. D. 8 to 12 mm
1 pc.
2 pcs.
*1:Applicable for Electrical connection code 2 and 7 (1/2 NPT female).
*2:Applicable for Electrical connection code 2, 4 and 7 (1/2 NPT and M20 female).
Code
CF1
CS1
CU1
SU2
JF3
JS3
G11
G12
T1005-2.EPS
10-5
IM 01C22K01-01E
ItemDescriptionCode
Painting
Color change
Coating change
316 SST exterior parts
Lightning protector
Oil-prohibited use
Oil-prohibited use
with dehydrating treatment
Calibration units
Sealing treatment to
SUS630 nuts
Long vent
Fast response
Failure alarm down-scale *
NAMUR NE43 compliant *
1
1
Stainless steel amplifier
housing
Gold-plate
Stainless steel tag plate
Mill Certificate
Pressure test/Leak test Certificate
* 1:Applicable for Output signal code D and E. The hardware error indicates faulty amplifier or
10. GENERAL SPECIFICATIONS
Amplifier cover only
Amplifier cover and terminal cover, Munsell 7.5 R4/14
Epoxy resin-baked coating
Exterior parts on the amplifier housing (name plates, tag plate, zero-adjustment screw,
stopper screw) will become 316 or 316L SST.
Transmitter power supply voltag: 10.5 to 32 V DC (10.5 to 30 V DC for intrinsically safe
type, 9 to 32 V DC for F
OUNDATION
Fieldbus and PROFIBUS PA communication type.)
Allowable current: Max. 6000 A (140 s), Repeating 1000 A (140 s) 100 times
Degrease cleansing treatment
Degrease cleansing treatment with fluorinated oilfilled capsule.
Operating temperature –20 to 80 °C
Degrease cleansing and dehydrating treatment
Degrease cleansing and dehydrating treatment with fluorinated oilfilled capsule.
Operating temperature –20 to 80 °C
P calibration ( psi unit )
bar calibration ( bar unit )
2
M calibration (kgf/cm
unit )
( See Table for Span and
Range Limits.)
Sealant ( liquid silicone rubber ) is coated on surfaces of SUS630 nuts used for cover
flange mounting.
Total vent plug Length: 112 mm (standard, 32 mm), Material: SUS316
Update time: 0.125 sec or less, see GS for the response time
Output status at CPU failure and hardware error is –5%, 3.2 mA or less.
Failure alarm down-scale: output status at CPU failure and
Output signal limits:
3.8 mA to 20.5 mA
hardware error is –5%, 3.2 mA or less.
Failure alarm up-scale: output status at CPU failure and
hardware error is 110%, 21.6 mA or more.
Amplifier housing material: SCS14A stainless steel
(equivalent to SUS316 cast stainless steel or ASTM CF-8M)
Gold-plated diaphragm
SUS304 tag plate wired onto transmitter
Cover flange, Process connector, Manifold, Orifice, and Spacer
Test Pressure: 3.5 MPa{35 kgf/cm2}
Test Pressure: 14 MPa{140 kgf/cm2}
Nitrogen(N2) Gas
Retention time: 10 minutes
capsule. When combining with Optional code F1, output status for down-scale is –2.5%,
Note: When INSTALLATION CODE ‘2’, ‘3’ or ‘8’ is selected, flow direction arrow mark on above figure
are reversed.
(i. e. Arrow head faces toward left.)
*1:15 mm (0.59 inch) for right side high pressure. (INSTALLATION CODE ‘2’, ‘3’ or ‘8’)
*2:Applicable only for ATEX, IECEx, and TIIS Flameproof type.
10-7
IM 01C22K01-01E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS INTRINSICALLY SAFE EQUIPMENT
INSTALLATION AND OPERATING PRECAUTIONS FOR
TIIS INTRINSICALLY SAFE EQUIPMENT
Apparatus Certified Under Technical Criteria (IEC-compatible Standards)
and from “RECOMMENDED PRACTICES for Explosion-Protected Electrical Installations in General Industries,” published in 1979
1. General
The following describes precautions on electrical apparatus
of intrinsically safe construction (hereinafter referred to as
intrinsically safe apparatus).
Following the Labor Safety and Health Laws of Japan, an
intrinsically safe apparatus must undergo type tests in order
to be certified by the Technical Institute of Industrial Safety,
Inc. These tests are required to satisfy either the technical
criteria for electrical machinery and equipment in compliance
with explosionproof standards involving inflammable gases
or vapors and for machinery and equipment having
explosionproof performance (standards notification no. 556
from the Japanese Ministry of Labor) (hereinafter referred to
as technical criteria), in conformity with IEC Standards, or
the “Recommended Practice for Explosion-Protected
Electrical Installations in General Industries,” published in
1979. Such a certified apparatus can be used in hazardous
locations where inflammable gases or vapors may be present.
Certified apparatus includes a certification label and an
equipment nameplate with the specifications necessary for
explosion requirements as well as precautions on explosion
protection. Please confirm these precautionary items and use
them to meet specification requirements.
For electrical wiring and maintenance servicing, please refer
to “Internal Wiring Rules” in the Electrical Installation
Technical Standards as well as “USER’S GUIDELINES for
Electrical Installations for Explosive Gas Atmospheres in
General Industry,” published in 1994.
To meet intrinsically safe requirements, equipment that can
be termed an “intrinsically safe apparatus” must:
(1) be certified by the Technical Institute of Industrial
Safety, Inc. in accordance with the Labor Safety and
Health Laws of Japan and have the appropriate mark of
certification labeled on its case, and
(2) be used in compliance with the specifications marked on
its certification label, equipment nameplate and
precautionary information furnished.
Note: Intrinsically safe apparatus satisfy their performance under
specific conditions. They are not always absolutely safe under
every operational and environmental condition. In other
words, they are not safe products involved with factors such
as chemical reactions, geographical changes or the like other
than affected by electric energy from the equipment itself.
2. Electrical Apparatus of Intrinsic
Safety Type of Explosion-Protected Construction
The intrinsic safety type of explosion-protected construction
is a method of protection applicable to a circuit or part of a
circuit in which, under prescribed test conditions, no spark or
thermal effect, whether produced normally or accidentally, is
capable of causing a prescribed explosive gas to ignite. In
other words, electrical apparatus of this construction is
intended to suppress electrical energy thereby preventing
ignition of a given explosive gas atmosphere even though
spark or high thermal effect occurs in the electric circuitry.
Intrinsically safe electrical apparatus generally comprise
intrinsically safe apparatus installed in a hazardous location
and a safety barrier (associated apparatus), installed in a nonhazardous location, aimed at preventing electrical energy
from flowing into the electric circuitry of intrinsically safe
apparatus.
However, battery-operated, portable intrinsically safe
apparatus or the like may be used alone.
3. Terminology
(1) Intrinsically safe apparatus: Electrical apparatus in which
all the circuits are intrinsically safe circuits.
(2) Associated apparatus: Electrical apparatus in which there
are both intrinsically safe circuits and non-intrinsically
safe circuits that can affect the safety of intrinsically safe
circuits.
(3) Safety barrier: A specific type of associated apparatus,
which consists mainly of safety barrier elements, and
serves to limit the flow of excessive electrical energy,
which is capable of causing ignition of a given explosive
gas or vapour of a non-intrinsically safe circuit into
concerned intrinsically safe circuits.
(4) Apparatus of category “ia”: Intrinsically safe electrical
apparatus and associated apparatus which are incapable
of causing ignition of a given explosive gas or vapour
with the appropriate safety factors such as:
1
EX-A03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS INTRINSICALLY SAFE EQUIPMENT
— when up to two countable faults are applied and, in
addition,
— when non-countable faults produce an onerous
condition.
(5) Apparatus of category “ib”: Intrinsically safe electrical
apparatus and associated apparatus which are incapable
of causing ignition of a given explosive gas or vapour,
with the appropriate safety factors such as:
— when up to one countable fault is applied and, in
addition,
— when non-countable faults produce an onerous
condition.
(6) Safety rating: A rating to be designated to intrinsically
safe apparatus as well as associated apparatus and is the
maximum rating allowable for maintaining intrinsic
safety of concerned intrinsically safe circuits.
4. Caution on Combining Intrinsically Safe Apparatus and
Safety Barriers
(1) A combination of certified intrinsically safe apparatus
and safety barriers needs to satisfy combination
requirements. If intrinsically safe apparatus specify
safety barriers for combination, safety barriers other than
specified cannot be used (see Note 1 for more details).
(2) Certified intrinsically safe systems specify specific safety
barriers in combination with intrinsically safe apparatus.
So safety barriers other than specified cannot be used
(see Note 2 for more details).
(3) Other than limitations of combining intrinsically safe
apparatus and safety barriers as given in (1) and (2)
above, two or more pieces of apparatus certified under
different standards cannot be combined with each other
(see Note 3 for more details). In addition, bear in mind
that classifications of explosion protection such as “IIA,”
“IIB” and “IIC” and category “ia” and “ib” limit a
combination of intrinsically safe apparatus and safety
barriers.
For more details, see the “Type Certificate Guide for
Explosion-Protected Constructionfor Electrical
Machinery and Equipment,” issued by the Japanese
Ministry of Labour, the Research Institute of Industrial
Safety.
Note 1: Testing Apparatus
Intrinsically safe apparatus and safety barriers are assessed
individually to ensure that their safety requirements are
satisfied. Tested and certified intrinsically safe apparatus and
safety barriers incorporate individual certification numbers.
A combination of intrinsically safe apparatus and safety
barriers involves the following two limitations:
(1) A safety barrier which meets the combination require-
ments by referring to its safety rating and combination
parameters shall be selected.
(2) For pressure transmitters, pH transmitters, temperature
detectors and the like, safety barriers that can be
combined are already specified. Other safety barriers
cannot be used.
Note 2: Testing Intrinsically Safe System
An assembly (as a system) in which intrinsically safe
apparatus and safety barriers are combined is assessed to
ensure that its safety requirements are satisfied. A tested and
certified system incorporates a certification number
(intrinsically safe apparatus and safety barriers have the same
certification number).
Note 3: Impossible Combinations of Apparatus Certified Under
Different Standards
Intrinsically safe apparatus certified under technical criteria
and safety barriers certified under the “Recommended
Practice for Explosion-Protected Electrical Installations in
General Industries” (1979) and vice versa cannot be
combined even if their combination requirements are
satisfied.
5. Installation of Intrinsically Safe
Apparatus and Safety Barriers
(1) Classification of installation location
Intrinsically safe apparatus may be installed, depending upon
applicable gases, in a hazardous area in Zone 0, 1 or 2 (Note
4 below), where the specified gases are present. However,
note that apparatus certified under Technical Criteria, in
category “ib” shall be installed only in Zone 1 or 2. Safety
barriers (associated apparatus) that are combined with these
intrinsically safe apparatus shall be installed only in a nonhazardous area. In cases where safety barriers are installed in
a hazardous area, they shall be enclosed, for example, in a
flameproof enclosure.
Note 4: Hazardous areas are classified in zones based upon the
frequency of the appearance and the duration of an explosive
gas atmosphere as follows:
Zone 0: An area in which an explosive gas atmosphere is
present continuously or is present for long periods.
Zone 1: An area in which an explosive gas atmosphere is
likely to occur in normal operation.
Zone 2: An area in which an explosive gas atmosphere is not
likely to occur in normal operation and if it does
occur it will exist for a short period only.
(2) Ambient temperature limits for intrinsically
safe apparatus
Intrinsically safe apparatus shall be installed in a location
where the ambient temperature ranges from –20° to +40°C
(for those certified under Technical Criteria) or –10° to
+40°C (for those certified under the “Recommended Practice
for Explosion-Protected Electrical Installations in General
Industries” (1979). However, some field-mounted
2
EX-A03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS INTRINSICALLY SAFE EQUIPMENT
intrinsically safe apparatus may be used at an ambient
temperature up to 60°C. So, specifications should be checked
before installing intrinsically safe apparatus.
If the intrinsically safe apparatus are exposed to direct
sunshine or radiant heat from plant facilities, appropriate
thermal protection measures shall be taken.
6. Wiring for Intrinsically Safe
Circuits
In intrinsically safe construction, safety shall be maintained
as an intrinsically safe system involving intrinsically safe
apparatus and safety barriers connected thereto, and electrical
wiring (through intrinsically safe circuits) interconnected
between them. In other words, even when safety
requirements are maintained individually by intrinsically safe
apparatus and safety barriers, they shall not be affected by
electrical or magnetic energy caused by electrical wiring.
To make electrical wiring for intrinsically safe circuits, you
must:
(a) refer to the equipment configuration diagram and make
electrical wiring properly;
(b) prevent intrinsically safe wiring from being contacted
with non-intrinsically safe wiring, and separate the
intrinsically safe circuit from other electrical circuits;
(c) prevent intrinsically safe wiring from being
electrostatically and magnetically affected by nonintrinsically safe wiring;
(d) reduce wiring inductance and capacitance produced
between the intrinsically safe apparatus and safety
barrier where possible, and use a shorter cable between
the intrinsically safe apparatus and safety barrier than
specified if the maximum permissible inductance of the
cable is specified as operating conditions;
(e) conform to conditions of installation such as wiring
method, earthing or the like, if any; and
(f) protect the outer sheath of cables from damage with
appropriate measures.
7. Maintenance and Inspection of
Intrinsically Safe Apparatus
and Safety Barriers
Installations for Explosive Gas Atmospheres in General
Industry” issued in 1994 by the Japanese Ministry of Labour,
the Research Institute of Industrial Safety.
(1) Requirements for maintenance personnel
Maintenance and inspection of intrinsically safe apparatus
and safety barriers shall be conducted by maintenance
personnel skilled in intrinsically safe construction and
installation of electrical devices as well as capable of
applying associated rules.
(2) Maintenance and Inspection
(a) Visual inspection
Visually inspect the external connections of intrinsically
safe apparatus and safety barriers, and cables for damage
or corrosion as well as other mechanical and structural
defects.
(b) Adjustments
Zero, span and sensitivity adjustments shall be made
with applicable adjusting potentiometers and mechanical
adjustment screws.
These maintenance adjustments shall be made in a nonhazardous location.
CAUTION
If intrinsically safe apparatus and safety barriers
require maintenance service and checking, a
gas detector shall be used to ensure that there
is no explosive gas in the location (maintenance servicing shall be conducted in a nonhazardous location).
(3) Repair
Intrinsically safe apparatus and safety barriers shall be
repaired by manufacturers.
(4) Prohibition of modifications and specifica-
tion changes
Do not attempt to make modifications or change specifications which may affect safety.
Maintenance and inspection of intrinsically safe apparatus
and safety barriers shall be limited to within the instructions
described in applicable instruction manuals. If other than this
is required, contact the manufacturers. For more information,
refer to the “USER’S GUIDELINES for Electrical
3
EX-A03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
INSTALLATION AND OPERATING PRECAUTIONS FOR
TIIS FLAMEPROOF EQUIPMENT
Apparatus Certified Under Technical Criteria
(IEC-compatible Standards)
1. General
The following describes precautions on electrical apparatus
of flameproof construction (hereinafter referred to as
flameproof apparatus) in explosion-protected apparatus.
Following the Labour Safety and Health Laws of Japan,
flameproof apparatus is subjected to type tests to meet either
the technical criteria for explosionproof electrical machinery
and equipment (standards notification no. 556 from the
Japanese Ministry of Labour) (hereinafter referred to as
technical criteria), in conformity with the IEC Standards, or
the “Recommended Practice for Explosion-Protected
Electrical Installations in General Industries,” published in
1979. These certified apparatus can be used in hazardous
locations where explosive or inflammable gases or vapours
may be present.
Certified apparatus includes a certification label and an
equipment nameplate with the specifications necessary for
explosion requirements as well as precautions on explosion
protection. Please confirm these precautionary items and use
them to meet specification requirements.
For electrical wiring and maintenance servicing, please refer
to “Internal Wiring Rules” in the Electrical Installation
Technical Standards as well as “USER’S GUIDELINES for
Electrical Installations for Explosive Gas Atmospheres in
General Industry,” published in 1994.
construction is of completely enclosed type and its enclosure
shall endure explosive pressures in cases where explosive
gases or vapours entering the enclosure cause explosion. In
addition, the enclosure construction shall be such that flame
caused by explosion does not ignite gases or vapours outside
the enclosure.
In this manual, the word "flameproof" is applied to the
flameproof equipment combined with the types of protection
"e", "o", "i", and "d" as well as flameproof equipment.
3. Terminology
(1) Enclosure
An outer shell of an electrical apparatus, which encloses live
parts and thus is needed to configure explosion-protected
construction.
(2) Shroud
A component part which is so designed that the fastening of
joint surfaces cannot be loosened unless a special tool is
used.
(3) Enclosure internal volume
This is indicated by:— the total internal volume of the
flameproof enclosure minus the volume of the internal
components essential to equipment functions.
To meet flameproof requirements, equipment that can be
termed “flameproof” must:
(1) Be certified by a Japanese public authority in accordance
with the Labour Safety and Health Laws of Japan and
have a certification label in an appropriate location on its
case, and
(2) Be used in compliance with the specifications marked on
its certification label, equipment nameplate and
precautionary information furnished.
2. Electrical Apparatus of Flameproof Type of Explosion-Protected Construction
Electrical apparatus which is of flameproof construction is
subjected to a type test and certified by the Japanese Ministry
of Labour aiming at preventing explosion caused by electrical
apparatus in a factory or any location where inflammable
gases or vapours may be present. The flameproof
(4) Path length of joint surface
On a joint surface, the length of the shortest path through
which flame flows from the inside to outside of the
flameproof enclosure. This definition cannot be applied to
threaded joints.
(5) Gaps between joint surfaces
The physical distance between two mating surfaces, or
differences in diameters if the mating surfaces are cylindrical.
Note: The permissible sizes of gaps between joint surfaces, the path
length of a joint surface and the number of joint threads are
determined by such factors as the enclosure’s internal
volume, joint and mating surface construction, and the
explosion classification of the specified gases and vapours.
1
EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
4. Installation of Flameproof Apparatus
(1) Installation Area
Flameproof apparatus may be installed, in accordance with
applicable gases, in a hazardous area in Zone 1 or 2, where
the specified gases are present. Those apparatus shall not be
installed in a hazardous area in Zone 0.
Note: Hazardous areas are classified in zones based upon the
frequency of the appearance and the duration of an explosive
gas atmosphere as follows:
Zone 0: An area in which an explosive gas atmosphere is
present continuously or is present for long periods.
Zone 1: An area in which an explosive gas atmosphere is
likely to occur in normal operation.
Zone 2: An area in which an explosive gas atmosphere is not
likely to occur in normal operation and if it does
occur it will exist for a short period only.
(2) Environmental Conditions
The standard environmental condition for the installation of
flameproof apparatus is limited to an ambient temperature
range from –20°C to +40°C (for products certified under
Technical Criteria). However, some field-mounted
instruments may be certified at an ambient temperature up to
+60°C as indicated on the instrument nameplates. If the
flameproof apparatus are exposed to direct sunshine or
radiant heat from plant facilities, appropriate thermal
protection measures shall be taken.
5. External Wiring for Flameproof
Apparatus
Flameproof apparatus require cable wiring or flameproof
metal conduits for their electrical connections. For cable
wiring, cable glands (cable entry devices for flameproof type)
to wiring connections shall be attached. For metal conduits,
attach sealing fittings as close to wiring connections as
possible and completely seal the apparatus. All non-live
metal parts such as the enclosure shall be securely grounded.
For details, see the “USER’S GUIDELINES for Electrical
Installations for Explosive Gas Atmospheres in General
Industry,” published in 1994.
(1) Cable Wiring
• For cable wiring, cable glands (cable entry devices for
flameproof type) specified or supplied with the apparatus
shall be directly attached to the wiring connections to
complete sealing of the apparatus.
• Screws that connect cable glands to the apparatus are
those for G-type parallel pipe threads (JIS B 0202) with
no sealing property. To protect the apparatus from
corrosive gases or moisture, apply nonhardening sealant
such as liquid gaskets to those threads for waterproofing.
• Specific cables shall be used as recommended by the
“USER’S GUIDELINES for Electrical Installations for
Explosive Gas Atmospheres in General Industry,”
published in 1994.
• In necessary, appropriate protective pipes (conduit or
flexible pipes), ducts or trays shall be used for
preventing the cable run (outside the cable glands) from
damage.
• To prevent explosive atmosphere from being propagated
form Zone 1 or 2 hazardous location to any different
location or non-hazardous location through the protective
pipe or duct, apply sealing of the protective pipes in the
vicinity of individual boundaries, or fill the ducts with
sand appropriately.
• When branch connections of cables, or cable connections
with insulated cables inside the conduit pipes are made,
a flameproof or increased-safety connection box shall be
used. In this case, flameproof or increased-safety cable
glands meeting the type of connection box must be used
for cable connections to the box.
(2) Flameproof Metal Conduit Wiring
• For the flameproof metal conduit wiring or insulated
wires shall be used as recommended by the USER’S
GUIDELINES for Electrical Installations for Explosive
Gas Atmospheres in General Industry, published in
1994.
• For conduit pipes, heavy-gauge steel conduits
conforming to JIS C 8305 Standard shall be used.
• Flameproof sealing fittings shall be used in the vicinity
of the wiring connections, and those fittings shall be
filled with sealing compounds to complete sealing of the
apparatus. In addition, to prevent explosive gases,
moisture, or flame caused by explosion form being
propagated through the conduit, always provide sealing
fittings to complete sealing of the conduit in the
following locations:
(a) In the boundaries between the hazardous and non-
hazardous locations.
(b) In the boundaries where there is a different
classification of hazardous location.
• For the connections of the apparatus with a conduit pipe
or its associated accessories, G-type parallel pipe threads
(JIS B 0202) shall be used to provide a minimum of
five-thread engagement to complete tightness. In
addition, since these parallel threads do not have sealing
property, nonhardening sealant such as liquid gaskets
shall thus be applied to those threads for ensuring
waterproofness.
• If metal conduits need flexibility, use flameproof flexible
fittings.
2
EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
6. Maintenance of Flameproof
Apparatus
To maintain the flameproof apparatus, do the following. (For
details, see Chapter 10 “MAINTENANCE OF EXPLOSIONPROTECTED ELECTRICAL INSTALLATION” in the
USER’S GUIDELINES for Electrical Installations for
Explosive Gas Atmospheres in General Industry.)
(1) Maintenance servicing with the power on.
Flameproof apparatus shall not be maintenance-serviced with
its power turned on. However, in cases where maintenance
servicing is to be conducted with the power turned on, with
the equipment cover removed, always use a gas detector to
check that there is no explosive gas in that location. If it
cannot be checked whether an explosive gas is present or not,
maintenance servicing shall be limited to the following two
items:
(a) Visual inspection
Visually inspect the flameproof apparatus, metal
conduits, and cables for damage or corrosion, and other
mechanical and structural defects.
(b) Zero and span adjustments
These adjustments should be made only to the extent
that they can be conducted from the outside without
opening the equipment cover. In doing this, great care
must be taken not to cause mechanical sparks with tools.
(2) Repair
If the flameproof apparatus requires repair, turn off the power
and transport it to a safety (non-hazardous) location. Observe
the following points before attempting to repair the
apparatus.
(a) Make only such electrical and mechanical repairs as will
restore the apparatus to its original condition. For the
flameproof apparatus, the gaps and path lengths of joints
and mating surfaces, and mechanical strength of
enclosures are critical factors in explosion protection.
Exercise great care not to damage the joints or shock the
enclosure.
(b) If any damage occurs in threads, joints or mating
surfaces, inspection windows, connections between the
transmitter and terminal box, shrouds or clamps, or
external wiring connections which are essential in
flameproofness, contact Yokogawa Electric Corporation.
requirements for flameproof apparatus (however, bear in
mind that the apparatus must always be restored to its
original condition). If you attempt to repair the
flameproof apparatus, company-specified components
shall be used.
(d) Before starting to service the apparatus, be sure to check
all parts necessary for retaining the requirements for
flameproof apparatus. For this, check that all screws,
bolts, nuts, and threaded connections have properly been
tightened.
(3) Prohibition of specification changes and
modifications
Do not attempt to change specifications or make
modifications involving addition of or changes in external
wiring connections.
7. Selection of Cable Entry Devices for Flameproof Type
IMPORTANT
The cable glands (cable entry devices for flameproof
type) conforming to IEC Standards are certified in
combination with the flameproof apparatus. So,
Yokogawa-specified cable entry devices for flameproof
type shall be used to meet this demand.
References:
(1) Type Certificate Guide for Explosion-Protected Con-
struction Electrical Machinery and Equipment (relating
to Technical Standards Conforming to International
Standards), issued by the Technical Institution of
Industrial Safety, Japan
(2) USER’S GUIDELINES for Electrical Installations for
Explosive Gas Atmospheres in General Industry (1994),
issued by the Japanese Ministry of Labour, the Research
Institute of Industrial Safet
CAUTION
Do not attempt to re-process threaded connections or
refinish joints or mating surfaces.
(c) Unless otherwise specified, the electrical circuitry and
internal mechanisms may be repaired by component
replacement, as this will not directly affect the
3
EX-B03E
Customer
Maintenance
Parts List
DPharp EJA Series
Transmitter Section
2
11
14
ItemPart No.
13
14
1
2
3
4
5
6
7-1
9
10
11
12
7-2
8
Bellow
F9341RA
F9341RJ
F9341JP
Below
F9341AA
F9341AC
F9341AE
F9341AH
F9341AJ
F9341AR
–
Bellow
F9900RG
F9900RR
F9341KL
Below
F9342AB
F9342AL
F9342AJ
F9342AD
F9342AF
F9342AM
F9342BF
F9342BG
F9900RP
Y9612YU
Below
F9340NW
F9340NX
G9330DP
G9612EB
Bellow
F9341FM
F9341FJ
Below
F9342BL
F9342BM
F9342MK
F9300PB
13
A
12
2
1
DescriptionQty
Cover
2
Cast-aluminum alloy
SCS14A stainless steel
O-ring
2
Case Assembly (Note 1)
1
Name Plate
1
Screw
4
Tag Plate
1
CPU Assembly
1
For F
Cap Screw
2
Screw
2
Plug
1
Cover Assembly
1
LCD Board Assembly
1
Mounting Screw
2
Label
2
4
5
10
3
1
2
A
6
7-1
8
9
Cast-aluminum alloy for G1/2
Cast-aluminum alloy for G1/2 (two electrical connections)
Cast-aluminum alloy for 1/2 NPT (two electrical connections)
Cast-aluminum alloy for M20 (two electrical connections)
Cast-aluminum alloy for Pg13.5 (two electrical connections)
SCS14A stainless steel for 1/2 NPT (two electrical connections)
For cast-aluminum alloy case assembly
For SCS14A stainless steel case assembly
For BRAIN protocol version (Except TIIS Intrinsically safe type)
For HART protocol version (Except TIIS Intrinsically safe type)
For BRAIN protocol version TIIS Intrinsically safe type (Optional code /JS3)
For BRAIN protocol version TIIS Intrinsically safe type with /F1(Optional code /JS3 and /F1)
For BRAIN protocol version (Optional code /F1)
For HART protocol version with write protection switch (Optional code /F1)
For F
OUNDATION
OUNDATION
For Pg13.5
For M20
For G1/2
For 1/2 NPT
Cast-aluminum alloy
SCS14A stainless steel
Without range-setting switch
With range-setting switch
Fieldbus protocol
Fieldbus protocol with PID/LM function (Optional code /LC1)
For integral indicator
5
7-2
Note 1 : Applicable for BRAIN and HART Protocol version (Output signal code D and E).
For F
OUNDATION
Yokogawa Electric Corporation
Fieldbus protocol version (Output signal code F), contact Yokogawa local office.
Changed to Electronic File Format
Revised a book in a new format.
(The location of contents and the associated page numbers may
not coincide with the one in old editions.)
Major Revised Items:
• Add F
OUNDATION
• Add Item 6 to the Wiring Precautions.
• Add FOUNDATOIN Fieldbus protocol.
• Add Output signal code F.
• Add Optional code A1.
• Change the figure of terminal configuration.
• Add Item 7-2.
• Add Optional code K5 and K6.
• Delete EMC Conformity Standards Tables and move the section
to Page 2-14.
• Remove Page 2-15.
• Correction made in BURN OUT figure.
• Add Electrical connection code 7, 8, and 9.
• Add Part No. to Item 3 (For PG13.5 and M20).
• Add Part No. to Item 10 (For 1/2NPT, PG13.5, and M20).
1. Explosion class and option code of JIS flameproof approval.
Explosion class: Ex ds IIC T4(old) to Ex do IIC T4X(new).
Option code: /JF1(old) to /JF3(new)
2. Option code for flameproof packing adapter for JIS flameproof
approval.
Option code: /G1 and /G2(old) to /G11 and /G12(new)
3. Add “Pa” and “hPa” as the unit for calibration range.
4. Part number change for CPU Board Assembly.
Fieldbus protcol version to ‘NOTE’ notice.
2-9
6-1
6-3
8-4
–
10-5
CMPL
2.9
2.10
6.2
6.4.2
8.3.1
–
10.3
CMPL 1C22A1-02E 5th 6th
CMPL 1C22K1-01E 6th
• Add Figure 2.3 Example of using DCS.
• Add Figure 2.4 Selecting Cables.
• Add AS/NZS 2064 1/2 to EMI, EMC Conformity Standards.
• Add selection in the case of JIS flameproof type.
• Change option code for flame packing adapter.
Option code: G1 and G2 G11 and G12
Change Applicable cable O.D. and Identifying mark.
Part number: G9601AH G9601AM
Change the figure of flame proof packing adapter in Figure 6.4.2c.
• Add Pa and hPa to C20 and D31.
• Installation and Operating Precautions for JIS Intrinsically Safe
and Explosionproof Equipment:
EX-A01E EX-A03E, EX-B01E EX-B03E
• Add Optional code F1.
• Change a format.
• Change and add Part No. of Item 7-1, CPU assembly:
Change; F9342BC F9342BB, F9342BK F9342BJ
Add; F9342AF, F9342AM
• Change Part No. of Item 10, Plug:
G9330DK G9330DP
• Change a format.
REVISION RECORD.EPS
IM 01C22K01-01E
EditionDatePageRevised Item
10th
Sep. 2000
2-8
2.9.4b
Change ambient Temperature limit and add Caution for cable
wiring.
2-9
3-1
8-4,8-5
8-8
8-12
8-14
10-1
10-4
2.9.5b
3
8.3.1
8.3.3.(3)
8.3.3.(10)(11)Add Note for /F1.
8.3.3.(14)
10.1
10.3
Add Caution for /JS3
Add Note for /F1
Add parameter E10, E14 and E15. Add Note for /F1.
Add Note for /F1.
Add Note for /F1.
Change the capsule damping time constant.
Add /JS3 and delete /JS1.
Change Amb.Temp. for /KF1
CMPL
CMPL 1C22A1-02E 6th 7th
• Add Parts No. to item 7–1 (For /JS3 and /F1)
• Add Parts No. to item 7–2 (For /LC1)
CMPL 1C22A1-02E 7th 8th
• Add Note for Case Assembly
CMPL 1C22K1-01E 6th 7th
• Change Parts No. of Capsale Assembly for JIS Intrinsically safe
type.
11th
July 2001
2-10
8-4, 8-5
2.10
8.3.1
• Change EMC Conformity number.
• Add footnote (*6) to B40, Maximum static pressure in Parameter
Summary.
CMPL
CMPL 1C22A1-02E 8th 9th (Manual Change)
• Change Part No. of CPU Assembly for BRAIN protocol.
F9342BB F9342AB
CMPL 1C22A1-02E 9th 10th (Manual Change)
• Change Part No. of CPU Assembly for HART protocol.
F9342BH F9342AL
CMPL 1C22A1-02E 10th CMPL 01C22A01-02E 11th
• Delete Part No. of Name Plate.
• Change Part No. of Screw. F9303JU Y9303JU
CMPL 1C22K1-01E 7th CMPL 01C22K01-01E 8th
12th
May 2002
1-2
2-7
10-4
10-5
1.1
2.9.4
10.3
• Add “1.1 For Safety Using.”
• Add descriptions based on ATEX directive.
• Add Optional code K2.
• Add Optional code C2 and C3.
13th
Nov. 2002-
CMPL
14th
15th
Apr. 2003
Apr. 2006
10-4
1-2
1-3
2-6
2-10
2-12
10-4, 10-5
10-5
10-6
CMPL
16thJan. 20081-1
1-4
2-3+
2-12
4-3
9-4
10-1+
10-4, -5
CMPL
• Style change from S2 to S3.
CMPL 01C22K01-01E 8th 9th
• Item 21 F9304V F9340P
• Item 23 D0117 F9340N
10.3 • Delete Opion code K1.
1.1 •
Add (e) Explosion Protected Type Instrument and (f) Modification
1.3 • Add “1.3 ATEX Document”
2.9.3 • Add “IECEx Certification” and delete “SAA Certification”
2.9.5 JIS Certification TIIS Certification
2.12 • Add Low Voltage Directive
10.3 • Add Certificate numbers and Applicable standards
• Add option code /SU2 and delete option code /SU1
• Add option code /PR and /N4
CMPL 01C22A01-02E 11th 12th
JIS Intrinsically safe type TIIS Intrinsically safe type
CMPL 01C22K01-01E 9th 10th
JIS Intrinsically safe type TIIS Intrinsically safe type
•Add direct current symbol.
• Add 11 European languages for ATEX documentation.
2.9.1 • Add applicable standard and certificate number for appovals.
2.10 • Add EMC caution note.
4.4 • Add section of changing the direction of integral indicator.
9.4.1 • Add figure of integral indicator direction.
10.1, 10.2 • Add PROFIBUS PA communication type.
10.3 • Delete applicable standard from the table.
CMPL 01C22A01-02E 12th CMPL 01C22A01-02E 13th
• Delete logo from the tag plate.
REVISION RECORD.EPS
IM 01C22K01-01E
EditionDatePageRevised Item
17th
Oct. 2008
2-9
2-12
7-1
8-4 and 8-5
8-6
8-9 and later
2.9.4
2.10
7.1
8.3.1
8.3.2
8.3.3
Change explosion protection marking for type n from EEx to Ex.
Update EMC coformity standards.
Modify layout.
Add new parameters.
Add items in table 8.3.1.
Add (6) Change Output Limit and (15) Span Adjustment.
Re-number the items.
8-18
9-1
9-6
10-3
10-5, 10-6
10-7
CMPL
8.5.2
9.3
9.4.5
10.2
10.3
10.4
CMPL 1C22A01-02E 13th 14th
Modify descriptions and notes for Er.01.
Add a note for calibration.
Add a note for cleaning.
Add new suffix codes.
Add Sealing statement for CSA.standards. Add /HC.
Correct errors.
Change part No .of item 5 and 8.
REVISION RECORD.EPS
IM 01C22K01-01E
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.