Yokogawa DLM 2054 Product Overview

DLM 2000DLM 2000
ear Warrant
DLM 2000
Mixed Signal Oscilloscope
SeriesSeries
Series
Lineup includes 200 MHz, 350 MHz, 500 MHz bandwidth models
Lightweight and compact
Large 8.4-inch LCD display
High speed sampling: Up to 2.5 GS/s (1.25 GS/s with 4 ch)
-Year Warrant
y-Year Warranty
Bulletin 7101-00E
Flexible inputs and flexible performanceFlexible inputs and flexible performance
Easy-to-Use & Easy-to-See
Easy-to-Use & Easy-to-See
Easy to use. Portrait body + large screen makes display easy to see.
Easy to use. Portrait body + large screen makes display easy to see.
We elevated the large (8.4-inch) LCD screen up into the line of sight. Also, the portrait format saves space on the desk or test bench. A compact personal oscilloscope designed for easy viewing and ease of use.
Measured values can be accurately read on the 0.1 div sub grid display.
Signal observation on 4 channels or more...
Signal observation on 4 channels or more...
Flexible MSO Input
- Capture a mixed signals of analog and logic signals -
Four channels is not sufficient to view the functioning of digital control circuits. The DLM2000 series converts 4 ch of analog input to 8-bit logic, and functions as a 3 ch analog + 8-bit logic MSO (mixed signal oscilloscope).
Large screen in a compact body
Footprint is approximately 2/3 the size of an 81/2 × 11 sheet of paper (depth of approximately 8 )
Ver tical Postion and Scale
Horizontal Postion and Scale Knob
Dedicated Zoom Keys
Four-Direction Selector Button
Select key moves the cursor up/down/left/right
Jog Shuttle and Rotary Knob
Logic input connector
Trigger Control Keys and Level Knob
3 ch analog + 8-bit logic
The performance of up to 11 inputs by converting to logic
Using logic input, up to 11 input signals can be observed simultaneously as 3 ch of analog and 8-bit logic. It is not only possible to use logic input for observation of data and control signals, or as a trigger source, but also for logic input analysis of I
2
C and SPI serial busses.
Logic probe for the DLM2000
DLM2000 Series Lineup
Model
Item
Analog input channels
Logic input
Maximum sampling rate
Frequency characteristics
Maximum record length
DLM2022
710105
200 MHz 350 MHz 500 MHz 200 MHz 350 MHz 500 MHz
62.5 Mpoints (Single measurement, memory length:/M1S, interleave ON)
DLM2032
710115
DLM2052
710125
2.5 GS/s (interleave ON)
Fast data processing with ScopeCORE
With our proprietary ScopeCORE fast data processing IC, real time display is possible even when simultaneously measuring multichannel signals of 11 inputs.
ScopeCORE fast data processing IC
DLM2024
710110
125 Mpoints (Single measurement, memory length: /M2, interleave ON)
DLM2034
710120
4*2
8bit-
* Or 3 channels when using logic input.
DLM2054
710130
02 03
The DLM2000 ser
ies comes with a varar
iety of easy-to-configure tr
iggers
combining analog and logic inputs such as edge
, enhanced, and B tr
iggers
xample
Sophisticated waveform acquisition engineSophisticated waveform acquisition engine
With long memory and the History function, you'll never miss
With long memory and the History function, you'll never miss an historical waveform. A variety of trigger functions reliably
an historical waveform. A variety of trigger functions reliably capture the waveforms you want.
capture the waveforms you want.
Large capacity (125 Mpoint) memory enables
long-duration measurements
For taking 2 ch measurements in Single mode, you can add the /M2 memory expansion option giving you up to 125 Mpoints of large memory capacity. 10,000 Hz signals can be recorded for up to 5,000 seconds. Even at a sampling rate of 1.25 GS/s, waveforms down to 0.1 seconds can be captured.
Continuous Measurement
2 ch, 4 ch same Standard /M1, /M1S memory option /M2 memory option
Note)The /M1, /M2 memory expansion options are only available on 4ch models. The /M1S option is only available on 2ch models.
1.25 Mpoints
6.25 Mpoints
12.5 Mpoints
With 4 ch (With 2ch for DLM20x2) With 2 ch (With 1ch for DLM20x2)
6.25 Mpoints 25 Mpoints
62.5 Mpoints
Single-Shot Measurement
12.5 Mpoints
62.5 Mpoints 125 Mpoints
You can replay waveforms later on, so you'll never miss
an abnormal waveform
HISTORY
- History Function -
With the DLM2000 series, up to 20,000 previously captured waveforms can be saved in the acquisition memory. With the History function, you can display just one or all of the previously captured waveforms (history waveforms) on screen. You can also perform cursor measurement, computation, and other operations on history waveforms. Using the History function, you can analyze rarely-occurring abnormal signals.
Waveform history
Abnormal signal
Trigger Function capturing combined
Trigger Function capturing combined analog/digital complex waveforms
analog/digital complex waveforms
The DLM2000 series comes with a variety of easy-to-configure triggers
The DLM2000 ser combining analog and logic inputs such as edge
combining analog and logic inputs such as edge, enhanced, and B triggers.
Edge trigger
Enhanced triggers
ies comes with a v
Edge
Edge OR
Edge (qualified)
State
Pulse width
State width
Serial:
TV
iety of easy-to-configure tr
, enhanced, and B tr
(optional) CAN/LIN/UART/I2C/SPI
: (standard) user-defined
: NTSC/PAL/SDTV/
HDTV/user defined
iggers
iggers
.
Trigg er function example
Trigger function example
Example:Trigger on the 7th edge of signal on B. This is
Example :Trigger on an arbitrarily set pattern of up to 128
Example:Trigger on a combination of CAN and LIN bus
A to B(n) trigger:
effective for measurements with shifted timing, such as non-standard video signal vertical/horizontal periods or motor reference position pulses and drive pulses.
Input signal A
Input signal B
Tr igger
Serial pattern trigger (user defined):
bits. This is effective for detecting ID/Data and other portions of proprietary communication formats.
HHHH-LHLH-LLLL (Arbitrary pattern)
Tr i gger
Dual pulse trigger:
triggers. I2C + SPI bus triggers, and other combinations are possible.
Tr igger when either LIN or CAN bus signal conditions become true
Input signal A CAN
B triggers
History search function Replay function
You can search the up to 20,000 previously captured waveforms for history waveforms that meet certain conditions. You can perform cursor measurement and other analyses on the found waveforms.
04 05
Waveforms can be displayed in order, one at a time, by using the rotary knob. With the Replay function, history waveforms can be automatically played back, paused, fast-forwarded, and rewound.
A Delay B
A to B(n)
Dual bus (combination trigger of 2 serial busses)
Input signal B LIN
CAN
LIN
C
apture & Display
Zoom and Search Functions
DLM 2000 Series
Real time filter with optimum noise reduction supports a wide range of frequencies
The DLM2000 series has two types of filters, one processed at the input circuit and one based on MATH functions. These filters are effective for rejecting unwanted signals, allowing observation of only the desired bandwidths.
Real time filters
Each channel has 14 low pass filters available from 8 kHz to 200 MHz. Waveforms of limited bandwidths are stored in internal memory.
Cutoff frequencies : 200 MHz, 100 MHz, 20 MHz, 10 MHz, 5 MHz,
2 MHz, 1 MHz, 500 kHz, 250 kHz, 125 kHz,
62.5 kHz, 32 kHz, 16 kHz, and 8 kHz
Processing with built-in filters
The input waveform can be filtered using an IIR filter, which is a MATH function. Filtered waveforms can be displayed at the same time as the input waveform for comparison. You can select low pass or high pass filters.
Cutoff frequency setting range : 0.01 Hz to 500 MHz
Input signal
Computed waveform
Computed digital filters
Filtering of a PWM waveform using computation
(from 8 kHz to 200 MHz)
Zooms into two different points
— Waveform zoom and search functions —
Zoom two locations simultaneously
Because the DLM2000 series lets you set zoom factors independently, you can display two zoomed waveforms with different time axis scales at the same time. Also, using the Auto Scroll function, you can automatically scroll waveforms captured in long memory and change the zoomed location. With Auto Scroll you can choose forward, backward, fast-forward, scroll speed, and other control options.
Dedicated
Zoom keys
Zoom1 Zoom2
Auto Scroll menu
Large capacity memory gives you a variety of waveform search functions.
Two types of waveform searching
Normally, searching for data takes time and costs money, and long memory is useless without functions for extracting desired data from a large capacity memory. That's why the DLM2000 series does not simply offer long memory, it also provides powerful waveform search functions.
Searching for data in a single screen: the Zoom Search function
This function searches captured waveforms in the long memory and displays waveforms that meet the search criteria in the zoom area. The locations of the found waveforms are marked on screen ( shows the current location).
• Waveform search criteria
Edge, edge (with conditions), state pattern, pulse width, state width, serial bus (only on models with the serial bus analysis option)
:
Search results
marked
Searching for history waveforms: the History Search function
Criteria can be specified for extracting desired waveforms from up to 20,000 previously captured waveforms.
Criterion extraction
Zone created from measured waveforms
Place square
zone and search
Waveform search using edge criterion
06
Searching for waveforms in zones created by moving measured waveforms up/down/left/right.
Search for waveforms that pass through/do not pass through a rectangular zone placed on screen.
U
seful Functions
Fastest and most capable analysis
Displays trends of peak-to-peak or pulse width per cycle
— Measure function and statistics —
Tw enty-eight waveform parameters are included such as: maximum, minimum, peak-to-peak, pulse width, period, frequency, rise/fall time, and duty ratio. Automated measurement can be performed using up to 20 of these waveform parameters. Also, waveform parameters can be measured repeatedly, and the statistical values displayed (mean, maximum, minimum, standard deviation, etc.).
— Trend and histogram displays —
Waveform parameters such as period, pulse width, and amplitude can be measured repeatedly and displayed in graphs. In a single screen you can observe period-by­period fluctuations, compute amplitudes every screen using multiple waveforms, and display amplitudes as trends. You can also display histograms referencing the voltage or time axis using values from repeated automated measurement of waveform parameters.
DLM 2000 Series
Trend display of waveform parameters
Histogram display using the time axis
Measures voltage/time differences automatically
— Cursor Measurement —
Cursors can be placed on the displayed waveform from signal data, and various measurement values at the intersection of the cursor and waveform can be displayed. There are six types of cursor; T, V, T& V, Marker, Degree Cursor.
Simultaneous level and time difference
measurement with the T&V cursor
Keeps waveforms with one push
— Snapshot —
By pressing the SNAPSHOT key to the lower right of the screen, you can freeze a white trace of the currently displayed waveform on the screen. You can press the key repeatedly and conveniently leave traces for comparing multiple waveforms. Also, snapshot data recorded on screen can be saved or loaded as files, and can be recalled for use as reference waveforms when making comparisons.
Using snapshots (white waveforms)
"SNAP SHOT"key
Analyzes frequency spectrums
— FFT analysis —
Up to 2 FFT analyses can be performed simultaneously. FFT can be performed on computed waveforms in addition to the actual waveforms on CH1 to CH4. Analysis can be performaed of the frequency components of waveforms filtered for limited bandwidth, of frequency for changes in period of rotary objects, and other phenomena.
FFT analysis
Displays stored files in thumbnail format
— Thumbnails of saved files —
Thumbnails of waveform data, waveform image data, and Wave­Zone files can be displayed. The image and file names are shown so that you can view screen image contents while copying or
Thumbnails of saved files
deleting files. In addition to normal­sized screens, you can even save wide images that have been zoomed along the time axis.
Zoomed (2x) long image file
Has a GO/NO-GO function
Abnormal waveform detected
— Action on trigger —
GO/NO-GO can be determined using trigger conditions, zone waveforms, measurement parameters, and other criteria. For NO-GO, actions can be carried out at the same time such as sounding a buzzer, saving the current waveform, or sending notification to a designated e-mail address. Wavefor ms in which an abnormality occurred can be saved for confirmation and analysis of the phenomena at a later time.
Action specified for NO-GO
Buzzer Output
to printer
Save waveform
data file
transmission
E-mail
Can check functions with graphical online help
— Graphical online help —
You can view detailed graphical explanations of the oscilloscope's functions by pressing the "?" key in the lower left of the screen. This lets you get help on functions and operations on screen without having to consult the user's manual.
07
S
olutions of the DLM2000
Analysis Applications
DLM 2000 Series
Serial analysis function options
- UART/CAN/LIN/I2C/SPI-
Tr iggers for UART, CAN, LIN, I2C, and SPI bus signals are supported along with decode display analysis (serial bus analysis option only on 4 ch models). Logic input can also be used for serial buses (excluding CAN and LIN).
Simultaneous analyses of different busses:
busses with different speeds can be displayed in individual Zoom screens with different scales.
A wealth of trigger functions:
combinations of serial bus triggers with normal edge triggers.
Simultaneous analyses of I2C and SPI
A wide variety of trigger conditions can be set, such as ID/Data trigger combinations and
Two busses can be analyzed simultaneously. Waveforms and analysis results from
(/F1, /F2, /F3, /F4)
Inputs supported for serial bus analysis
Analog input Logic input
Accessories
PBDH1000 differential probe (model 701924)
1.0 GHz bandwidth 1 M, approximately 1.1 pF Maximum differential input voltage range: ± 25 V
Differential probe (model 701920)
DC to 500 MHz bandwidth 100 k, approximately 2.5 pF Maximum differential input voltage range: ±12V
2
C
I Ye s Ye s
Simultaneous analyses of CAN and LIN
SPI Ye s Ye s
UART
Ye s Ye s
LIN Ye s NA
CAN
Ye s
NA
Power supply analysis option
Dedicated power supply analysis options are available (4 ch models only) for switching loss, joule integral (i2t), SOA (safe operating area) analysis, harmonic analysis of power supply current based on EN61000-3-2, and other operations.
(/G4)
Harmonic analysis of power supply current based on
Switching loss analysis
Voltage and current waveforms can be input to the 62.5 MW (max.) long memory (/M2 models) for computation of switching loss (V(t) X i(t)). A wide variety of switching loss analyses are supported, including turn­on/off loss calculation, loss including continuity loss, and loss over long cycles (50 Hz/60 Hz).
Related Accessories
700924 Differential probe
DC to 100 MHz 1000 Vrms/ ± 1400 V
701928/701929 Current probe
DC to 100 MHz(701928) DC to 50 MHz(701929) 30 Arms
EN61000-3-2
Harmonics determined by the IEC standard that are generated by the target device can be judged for each applicable class (classes A-D). Bar graphs and lists can be displayed for comparing harmonic current limit values with values calculated from actually measured signals.
Harmonic current graph display
701935 Deskew correction signal source
08
C
onnectivity & Software
DLM 2000 Series
Broad Connectivity and Easier Control
Ethernet (optional)
Supports 1000BASE-T, 100BASE-TX, 10BASE-T
GO/NO-GO I/O terminal
Using the GO/NO-GO function, you can input a timing signal for judging a waveform and output the result as a TTL level signal.
RGB video signal output terminal
You can output an image signal and check the waveform on an external monitor.
USB-PC connection terminal
Enables control from a PC.
USB peripheral connection terminal
Supports USB storage, USB keyboards, USB printers.
Sends waveform, screen, and settings data
Remote control
Mail transmission (GO/NO-GO action)
Ethernet
(/C10, /C11 options)
USB
(standard on rear panel)
Probe power terminal (optional)
Power supply output terminal for current probes (701930 and
701931) and differential probes (701920, 701921, 701922, 700924, 700925, and 701926).
GP-IB connection terminal (optional)
Enables control from a PC.
External trigger input
Lets you input a trigger signal separately from the input signal.
Trigger output
Outputs a CMOS 3.3V level trigger signal.
1000BASE-T/100BASE-TX/ 10BASE-T compliant adapters
(hubs and routers)
Sends waveform, screen, and settings data Remote control
Internal storage
(/C8 option)
Software
Xviewer (701992, sold separately)
Xviewer is software for use on a PC. It can be used for display, analysis, and conversion to ASCII of binary waveform data using waveforms captured by the DLM2000 series. By adding the MATH option, you can enter user expressions for performing waveform computations. FFT of up to 2 Mwords can be performed.
USB memory
USB printer
USB keyboard
For details on accessory software, visit
https://y-link.yokogawa.com/YL000.po
Also, you can download free software and trial versions of retail software from this site.
DL series library (freeware)
This is an API that enables you to control a DL or send data from a DL using an external program. The API is offered in the form of a DLL that can be called from a program controlled by the user.
09
Main Specification
Models
Model name
Frequency bandwidth
Input terminal Max. sample rate
DLM2022 (710105) 200MHz
DLM2032 (710115) 350MHz 2 analog channels
DLM2052 (710125) 500MHz (interleave mode off)
DLM2024 (710110) 200MHz
DLM2034 (710120) 350MHz 3 analog channels +
DLM2054 (710130) 500MHz
Basic Specifications
Analog Signal input
Input channels Analog input DLM20x2: CH1, CH2
Input coupling setting AC, DC, DC50 , GND Input impedance Analog input 1 M±1.0%, approximately 20 pF
Voltage axis sensitivity 1 M 2 mV/div to 10 V/div (steps of 1-2-5) setting range 50 2 mV/div to 500 mV/div (steps of 1-2-5) Max. input voltage 1 M 150 Vrms (CAT I)
Max. DC offset 1 MΩ±1V (2 mV/div to 50 mV/div) setting range ±10V (100 mV/div to 500 mV/div)
50 Ω ±1.0% (VSWR 1.4 or less, DC to 500MHz)
50 Must not exceed 5 Vrms or 10 Vpeak
50 Ω±1V (2 mV/div to 50 mV/div)
DC accuracy* Offset voltage accuracy*
1
±(1.5% of 8 div + offset voltage accuracy)
1
2 mV to 50mV/div ±(1% of setting +0.2 mV) 100 mV to 500 mV/div ±(1% of setting + 2 mV) 1 V to 10 V/div ±(1% of setting + 20 mV)
Frequency characteristics (-3 dB attenuation when inputting a sinewave of amplitude ±3div)*
1 M(when using passive probe)
50
Isolation between channels -34 dB@ analog bandwidth (typical value) Residual noise level*
100 mV to 100 V/div DC to 200 MHz DC to 350 MHz DC to 500 MHz 20 mV to 50 mV/div DC to 150 MHz DC to 300 MHz DC to 400 MHz
10 mV to 10 V/div DC to 200 MHz DC to 350 MHz DC to 500 MHz 2 mV to 5 mV/div DC to 150 MHz DC to 300 MHz DC to 400 MHz
3
A/D resolution 8bit (25LSB/div)
Bandwidth limit FULL, 200 MHz, 100MHz, 20 MHz, 10 MHz,
Maximum sample rate Real time sampling mode Interleave OFF 1.25 GS/s
Repetitive sampling mode
Maximum record length 2 ch model Repeat/Single/Single Interleave:
4 ch model Repeat/Single/Single Interleave:
4 ch model Repeat/Single/Single Interleave:
Ch-to-Ch deskew ±100 ns Time axis setting range 1 ns/div to 500 s/div (steps of 1-2-5) Time base accuracy ±0.002% Max. acquisition rate*
Interleave ON 2.5 GS/s
(Standard) 1.25 M/6.25 M/12.5 MPoints 2 ch model Repeat/Single/Single Interleave: (/M1S) 6.25 M/25 M/62.5 MPoints
(Standard) 1.25 M/6.25 M/12.5 MPoints
(/M1) 6.25 M/25 M/62.5 MPoints 4 ch model Repeat/Single/Single Interleave: (/M2) 12.5 M/62.5 M/125 MPoints
4
Dead time in N Single mode Approx. 2.2 µs
Logic Signal Input (4 ch model only)
Number of inputs 8 bit (excl. 4 ch input and logic input) Maximum toggle frequency*
1
Compatible probes 701988, 701989 (8 bit input)
Min. input voltage 701988: 500 mVp-p
Input range Model 701988: ±40 V
Max. nondestructive input voltage ±40 V (DC + ACpeak) or 28 Vrms (when using
Threshold level setting range
Input impedance 701988: Approx. 1 M/approx. 10 pF
Maximum sampling rate 1.25 GS/s Maximum record length Standard Repeat: 1.25 MPoints, Single: 6.25 MPoints
/M1, /M1S option Repeat: 6.25 MPoints, Single: 25 MPoints /M2 option Repeat: 12.5 MPoints, Single: 62.5 MPoints
4 analog channels /
8bit logic
DLM20x4: CH1 to CH4 (CH1 to CH3 when using logic input)
±100V (1 V/div to 10 V/div)
±5V (100 mV/div to 500 mV/div)
DLM202x DLM203x DLM205x
The larger of 0.4 mV rms or 0.05 div rms (typical value)
Max. 12 bit (in High Resolution mode)
5 MHz, 2 MHz, 1 MHz, 500 kHz, 250 kHz, 125 kHz, 62.5 kHz, 32 kHz, 16 kHz, 8 kHz (can be set for each channel)
125 GS/s
Approx. 20,000 waveform/sec/ch (Accumulation mode)
(approx. 450,000 waveforms/sec/ch)
Model 701988: 100 MHz Model 701989: 250 MHz
(701980, 701981 are available)
701989: 300 mVp-p
Model 701989: threshold ±6V
701989) Model 701988: ±40 V (setting resolution of 0.05 V) Model 701989: ±6 V (setting resolution of 0.05 V)
701989: Approx. 100 k/approx. 3 pF
1.25GS/s
2.5GS/s
(interleave mode on)
Triggers
Tr igger modes Auto, Auto Level, Normal, Single, N-Single Tr igger type, trigger source
Tr igger level setting range Tr igger level setting resolution Tr igger level accuracy CH1 to CH4 ±(0.2 div + 10% of trigger level) Window Comparator Center/Width can be set on individual Channels
A triggers Edge CH1 to CH4, Logic, EXT, LINE
AB triggers A Delay B 10 ns to 10 s (Edge, Edge
CH1 to CH4 ±4 div from center of screen CH1 to CH4 0.01 div (TV trigger: 0.1 div)
Edge OR CH1 to CH4 Edge Qualified State CH1 to CH4, Logic Pulse width CH1 to CH4, Logic, EXT State width CH1 to CH4, Logic TV CH1 to CH4 Serial Bus
2
I
C (optional) CH1 to CH4, Logic SPI (optional) UART (optional) CAN (optional) LIN (optional)
User defined CH1 to CH4
A to B(N) 1 to 10
Dual Bus Serial bus only
CH1 to CH4, Logic, EXT
CH1 to CH4, Logic CH1 to CH4, Logic CH1 to CH4 CH1 to CH4
Qualified, State, Serial Bus)
9
(Edge, Edge Qualified,
State, Serial Bus)
from CH1 to CH4
Display
Display 8.4-inch TFT color liquid crystal display
1024 x 768 (XGA)
Functions
Waveform acquisition modes Nor mal, Envelope, Average High Resolution mode Max. 12 bit (the resolution of the A/D converter
1*2
Sampling modes Real time, interpolation, repetitive sampling Accumulation Select OFF, Intensity (waveform frequency by
Roll mode Enabled at 100 ms/div to 500 s/div (depending on
Accumulation time 100 ms to 100 s, Infinite
Zoom function Two zooming windows can be set independently
Zoom factor x2 to 2.5 points/10div (in zoom area) Scroll Auto Scroll Search functions Edge, Edge Qualified, State, Pulse Width, State
History memory Max. data
History search Select Rect, WAVE, Polygon, or Parameter mode Replay function Automatically displays the history waveforms
Cursor Types T, V, T & V, Marker, Degree Snapshot Currently displayed waveform can be retained on
Display Specified or average waveforms
can be improved equivalently by placing a bandwidth limit on the input signal.)
brightness), or Color (waveform frequency by color)
the record length setting)
(Zoom1, Zoom2)
Width
2
I
C (option), SPI (option), UART (option), CAN (option), LIN (option) 2,500 (record length 1.25 kPoints, with standard) 10,000 (record length 1.25 kPoints, with /M1 or /M1S option) 20,000 (record length 1.25 kPoints, with /M2 option)
sequentially
screen
Computation & Analysis Functions
Parameter measurement MAX, MIN, P-P, HIGH, LOW, Rms, Mean, Sdev,
Statistical computation of parameters Min, Max, Ave, Cnt, Sdev Statistics modes Continuous, Cycle, History Trend/Histogram display of wave parameters Up to 2 trend or histgram display of specied wave
Computations (MATH) +, -, x, Filter (Delay, Moving Avg, IIR Lowpass, IIR
Computable no. of traces 2 (Math1, Math2) (1 trace for 2ch model) Max. computable memory length Standard model: 6.25 MPoints, /M1,/M2 memory
Reference function Up to 2 traces (REF1/REF2) of saved waveform
Action ON trigger Modes All Condition, Zone, Param, Rect, Polygon
Analysis XY Displays XY1, XY2 and T-Y simultaneously
Actions Buzzer, Print, Save, Mail, GO-NOGO out
FFT Number of points: 1.25k, 12.5k, 125k, 250k
Histogram Displays a histogram of acquired waveforms User-defined math The following operators can be arbitrarily (/G2 Options) combined in equations:
IntegTY+, IntegTY, +OVER,
-OVER, Pulse Count, Edge Count, V1, V2, T, Freq, Period, Avg Freq, Avg Period, Burst, Rise, Fall, +Width, -Width, Duty, Delay
parameters
Highpass), Integ, Count, user defined math (optional)
expansion option: 25 MPoints, /M2 expansion option: 62.5 MPoints
data can be displayed and analyzed
Window functions: Rectangular, Hanning, Flat-Top FFT Types: PS (LS, RS, PSD, CS, TF, CH are available with /G2 option)
+, -, x, /, SIN, COS, TAN, ASIN, ACOS, ATAN, INTEG, DIFF, ABS, SQRT, LOG, EXP, LN, BIN, DELAY, P2 (power of 2), PH, DA, MEAN, HLBT, PWHH, PWLL, PWHL, PWLH, PWXX, FV, DUTYH, DUTYL,
10
DLM 2000 Series
The maximum record length that can be
Power supply analysis function (/G4 option) The difference in propagation time of voltage and
2
I
C Bus Signal Analysis Functions (/F2 & /F3 Options)
Applicable bus I
2
I
C Trigger modes Every Start, Address & Data, Non-Ack, General
Analyzable signals Analysis results displays Analysis no., time from trigger position (Time
2
C bus Bus transfer rate: 3.4 Mbit/s max.
SM bus Complies with System Management Bus
Auto setup function Auto setting of bit rate, threshold value, time axis
Analyzable no. of data 300,000 bytes max. Search function Searches data that matches specified address
Analysis results save function Analysis list data can be saved to CSV-format files
computed is as well as standard math functions Propagation time difference correction (deskew):
current probe signals can be automatically or manually corrected. Correction range is ±100 ns
(0.01 ns resolution) Automated measurement of power supply analysis parameters: Power supply analysis parameters can be measured automatically and simultaneously with standard measurement items. (Automated measurement of two areas is also possible) Waveform computation of power supply analysis parameters: Wp, Wp+, Wp-, Abs.Wp., P, P+, P-, Abs.P, Z(Impedance) Display of the Area of Voltage-Current Operation: Allows for checking whether it is within the ASO(area of safe operation) Harmonic analysis: Harmonic current emission standard IEC 61000­3-2 edition 2.2(EN61000-3-2 (2000)) Trend display:
Address mode: 7 bit/10 bit
Call, Start Byte, HS Mode Assignable to CH1 to CH4, Logic input, or M1 to M2
(ms)),1st byte address, 2nd byte address, R/W, Data, Presence/absence of ACK, information
scale, voltage axis scale, and display of analysis results
pattern, data pattern, and acknowledge bit condition
SPI Bus Signal Analysis Functions (/F2 & /F3 Options)
Tr igger types 3 wire/4 wire
Byte order MSB/LSB Auto setup function Auto setting of bit rate, threshold value, time axis
Analyzable no. of data 300,000 bytes max. Decode bit length Specify data interval (1 to 32 bits), decode star t
Analysis results displays Analysis no., time from trigger position (Time
Auxiliary analysis functions Data search function Analysis result save function Analysis list data can be saved to CSV-format files
After assertion of CS, compares data after arbitrary byte count and triggers.
scale, voltage axis scale, and display of analysis results
point, and data length
(ms)), Data 1, Data 2
UART Bus Signal Analysis Functions (/F1 & /F3 Options)
Bit rate
Data format Select a data format from the following 8 bit (Non
UART Tr igger modes Every Data, Data, Error (Framing, Parity) Analyzable signals Select CH1 to CH4, logic input, or M1 to M2 Auto setup function Auto setting of bit rate, threshold value, time axis
Analyzable no. of frames 300,000 frames max. Analysis results displays Analysis no., time from trigger position
Auxiliary analysis functions Data search Analysis result save function Analysis list data can be saved to CSV-format files
1200 bps, 2400 bps, 4800 bps, 9600 bps,19200 bps, user defined (an arbitrary bit rate from 1 k to 1 Mbps with resolution of 100 bps)
Parity) / 7 bit Data + Parity / 8 bit + Parity
scale, voltage axis scale, and display of analysis results
(Time(ms)), Data (Bin, Hex) display, ASCII display, and Information.
CAN Bus Signal Analysis Functions (/F4 Option)
Applicable bus CAN version 2.0A/B, Hi-Speed CAN (ISO11898),
Bit rate 1 Mbps/500 kbps/250 kbps/125 kbps/83.3 kbps/
CAN bus Trigger modes SOF, ID/DATA, ID OR, Error(enabled when
Auto setup function Auto setting of bit rate, threshold value, time axis
Analyzable no. of frames 100,000 frames max. Analysis results displays Analysis no., time from trigger position (Time
Low-Speed CAN (ISO11519-2)
33.3 kbps User defined ( an arbitrary bit rate from 10.0 kbps to 1.000 Mbps with resolution of 100 bps)
loading physical values/symbol definitions)
scale, voltage axis scale, and display of analysis results
(ms)), Frame type, ID, DLC, Data, CRC, presence/absence of Ack, information
Auxiliary analysis functions Data search and field jump functions Analysis result save function Analysis list data can be saved to CSV-format files
LIN Bus Signal Analysis Functions (/F4 Option)
Applicable bus LIN Rev. 1.3, 2.0 Bit rate 19.2 kbps, 9.6 kbps, 4.8 kbps, 2.4 kbps, 1.2 kbps
LIN bus Trigger modes Auto setup function Auto setting of bit rate, threshold value, time axis
Analyzable no. of frames 100, 000 frames max. Analysis results displays Analysis no., time from trigger position (Time
Auxiliary analysis functions Analysis result save function
User defined (an arbitrary bit rate from 1000 bps to 200 kbps with resolution of 100 bps) Break Synch, ID/DATA, ID OR, and ERROR trigger
scale, voltage axis scale, and display of analysis results
(ms)), ID, ID-Field, Data, CheckSum, information Data search and field jump functions Analysis list data can be saved to CSV-format files
GP-IB (/C1 & /C11 Options)
Electromechanical specifications Protocol Conforms to IEEE std. 488.2-1987
Conforms to IEEE std. 488-1978 (JIS C 1901-1987)
Auxiliary Input
Rear panel I/O signal External trigger input(DLM20x2: front panel),
Probe interface terminal (front panel) 4 terminals (DLM20x4) Probe power terminal (rear panel) 2 terminals (/P2 option)
external trigger output, GO-NOGO output, video output
4 terminals (/P4 option)
Internal Storage (Standerd model /C8 Option)
Capacity Standard model: 100 MB
/C8 option: 1.8 GB
Built-in Printer (/B5 Option)
Built-in printer 112 mm wide, monochrome, thermal
USB Peripheral Connection Terminal
Connector USB type A connector x 2 (front panel x 1, rear
Electromechanical specifications USB 2.0 compliant Supported transfer standards Low Speed, Full Speed, High Speed Supported devices USB Pr inter Class Ver. 1.0 compliant EPSON/HP
panel x 1)
(PCL) ink jet printers USB Mass Storage Class Ver. 1.1 compliant mass storage devices* Please contact your local Yokogawa sales office for model names of verified devices
USB-PC Connection Terminal
Connector USB type B connector x 1 Electromechanical specifications USB 2.0 compliant Supported transfer standards High Speed, Full Speed
Supported class
USBTMC-USB488 (USB Test and Measurement Class Ver. 1.0)
Ethernet (/C10 & /C11 Options)
Connector RJ-45 connector x 1 Transmission methods Ethernet (1000BASE-T/100BASE-TX/10BASE-T) Supported services Server: FTP, VXI-11
Client: SMTP, SNTP, LPR, DHCP, DNS
General Specifications
Rated supply voltage 100 to 240 VAC Rated supply frequency 50 Hz/60 Hz Maximum power consumption External dimensions 226 (W) x 293 (H) x 193 (D) mm (when printer
Weight Approx.4.2kg With no options Operating temperature range
*1 Measured under standard operating conditions after a 30-minute warm-up followed by calibration.
Standard operating conditions: Ambient temperature: 23˚C ±5˚C
Error in supply voltage and frequency: Within 1% of rating
*2 Value in the case of repetitive phenomenon. The frequency bandwidth of a single-shot phenomenon is the smaller of the two
values, DC to sampling frequency/2.5 or the frequency bandwidth of the repetitive phenomenon.
*3.When the input section is shor ted, the acquisition mode is set to Normal, accumulation is OFF, and the probe attenuation is
set to 1:1.
*4.Acquisition rate does not vary with an increase or decrease in channels.
Ambient humidity: 55 ±10% RH
170 VA
cover is closed, excluding protrusions)
5 ˚C to 40 ˚C
External Dimensions
Unit: mm
11
Model and Suffix Codes Standard Main Unit Accessories
Model Suffix code Description
710105
*1
710110 710115
*1
710120 710125
*1
710130 Power cable
Help language
Option
*1: Logic probes sold separately. Please order the model 701988/701989 accessory logic probes separately. *2: Only one of these may be selected at a time. *3: Specify this option when using current probes or other differential probes such as models 701920 or 701922. *4: Only one of these may be selected at a time. *5: Only one of these may be selected at a time. *6: Only one of these may be selected at a time.
-D
-F
-Q
-R
-H
-HE
-HC
-HK
-HG
-HF
-HI
-HS /LN /B5
/M1
/M2
/M1S
/P2 /P4 /C1 /C10 /C11 /C8 /G2 /G4 /F1 /F2 /F3 /F4
Digital Oscilloscope DLM2022 2ch, 200MHz Mixed Signal Oscilloscope DLM2024 4ch, 200MHz Digital Oscilloscope DLM2032 2ch, 350MHz Mixed Signal Oscilloscope DLM2034 4ch, 350MHz Digital Oscilloscope DLM2052 2ch, 500MHz Mixed Signal Oscilloscope DLM2054 4ch, 500MHz UL/CSA standard VDE standard BS standard AS standard GB standard English Help (Menu and Panel) Chinese Help (Menu and Panel) Korean Help (Menu and Panel) German Help (Menu and Panel) French Help (Menu and Panel) Italian Help (Menu and Panel) Spanish Help (Menu and Panel) No switchable logic input (4 ch model only) Built-in printer "Memory expansion option (4 ch model only)
*2
During continuous measurement: 6.25 Mpoints; Single mode: 25 Mpoints (when interleave mode ON: 62.5 Mpoints)" "Memory expansion option (4 ch model only)
*2
During continuous measurement: 12.5 Mpoints; Single mode:
62.5 Mpoints (when interleave mode ON: 125 Mpoints)" "Memory expansion option (2 ch model only) During continuous measurement: 6.25 Mpoints; Single mode: 25 Mpoints (when interleave mode ON: 62.5 Mpoints)"
*3
Probe power for 2 ch models
*3
Probe power for 4 ch models
*4
GP-IB Interface
*4
Ethernet Interface
*4
GP-IB + Ethernet Interface Internal storage (1.8 GB)
*5
User defined math (4 ch model only)
*5
"Power supply analysis function (includes /G2) (4 ch model only)"
*6
UART trigger and analysis (4 ch model only)
2
*6
I
C + SPI trigger and analysis (4 ch model only)
*6
2
C + SPI trigger and analysis (4 ch model only)
UART + I CAN + LIN trigger and analysis (4 ch model only)
Power cord (with 3-prong to 2-prong adapter) "Passive probe, model 701938 (200 MHz, 1.5 m)
For models 710105, 710110" "Passive probe, model 701939 (500 MHz, 1.3 m)
For models 710115, 710120, 710125, 710130" Protective front cover Soft carrying case for probes Printer roll paper (for /B5 option) User's manuals
Accessory Models
Name Model Specification
Logic probe (PBL100) Logic probe (PBL250) Passive probe Passive probe FET Pprobe Active probe (PBA1000) 100:1 voltage probe 100:1 voltage probe Differential probe Differential probe Differential probe (PBDH1000) Differential probe Differential probe Differential probe Current probe (PBC050) Current probe (PBC100) Current probe Current probe Mini clip converter BNC adapter PCB adapter
Solder-in adapter
Printer roll paper
Xviewer
Probe stand Carrying case
http://www.DLM2000.net/
Mixed Signal Oscilloscope
Part Name Quantity
701988 701989 701938 701939 700939 701912 701944 701945 701921 701922 701924 700924 700925 701920 701929 701928 701930 701931 700971 700972 366945
366946
B9988AE 701992-SP01 701992-GP01 701919 701964
DLM 2000
DLM 2000 Series
Per number of channels
Per number of channels
1 roll 1 set
1 M input resistance, toggle frequency of 100 MHz 100 k input resistance, toggle frequency of 250 MHz 10 M (10:1), 200 MHz, 1.5 m 10 M (10:1), 500 MHz, 1.3 m DC to 900 MHz bandwidth/2.5M/1.8pF DC to 1 GHz bandwidth/100k/0.9pF DC to 400 MHz, 1.2 m, 1000 Vrms DC to 250 MHz, 3 m, 1000 Vrms DC to 100 MHz bandwidth/max. ±700 V DC to 200 MHz bandwidth/max. ±20 V DC to 1 GHz bandwidth/1M/max. ±25 V DC to 100 MHz bandwidth/max. ±1400 V DC to 15 MHz bandwidth/max. ±500 V DC to 500 MHz bandwidth/max. ±12 V DC to 50 MHz bandwidth, 30 Arms DC to 100 MHz bandwidth, 30 Arms DC to 10 MHz bandwidth, 150 Arms DC to 2 MHz bandwidth, 500 Arms For models 701938 and 701939 For models 701938 and 701939 For models 701938 and 701939, 10 per set For models 701938 and 701939, 1 adapter,
red/black cables (3 ea.) Lot size is 10 rolls, 10 meters each For DL/WE series, standard version For DL/WE series, with MATH functions Round base, 1 arm Also for DL1600/DL1700E series
Series
Series
1
1 1
[ DLM is a pending trademark or registered trademark of Yokogawa Electric Corporation.] Any company's names and product names appearing in this document are the registered trademarks or trademarks of their respective companies.
NOTE
"Before operating the product, read the user’s manual thoroughly for
proper and safe operation."
Special Site
Special Site
Product demonstration (Flash) now available
Check here for updated firmware information.
Manual download service! *
* Check here for oscilloscope accessories.
Yokogawa's Approach to Preserving the Global Environment
• Yokogawa's electrical products are developed and produced in facilities that have received ISO14001 approval.
• In order to protect the global environment,Yokogawa's electrical products are designed in accordance with Yokogawa's Environmentally Friendy Product Design Guidelines and Product Design Assessment Criteria.
YOKOGAWA CORPORATION OF AMERICA
2 Dart Road, Newnan, Georgia 30265-1094, U.S.A. Phone: (1)-770-253-7000, Fax: (1)-770-251-6427
YOKOGAWA EUROPE B.V.
Euroweg 2, 3825 HD, Amersfoort, THE NETHERLANDS Phone: (31)-88-4641000, Fax: (31)-88-4641111
YOKOGAWA ENGINEERING ASIA PTE. LTD.
5 Bedok South Road, Singapore 469270 Phone: (65)-62419933, Fax: (65)-62412606
YOKOGAWA MEASURING INSTRUMENTS KOREA CORP.
Phone: (82)-2-551-0660, Fax: (82)-2-551-0665
YOKOGAWA SHANGHAI TRADING CO., LTD.
Phone: (86)-21-5405-0303, Fax: (86)-21-6880-9254
YOKOGAWA ELECTRIC CORPORATION
Communication & Measurement Business Headquarters
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Phone: (81)-422-52-6768, Fax: (81)-422-52-6624
E-mail: tm@cs.jp.yokogawa.com
Subject to change without notice.
All Rights Reserved, Copyright© 2008, Yokogawa Electric Corporation.
Represented by :
[Ed : 02/b] Printed in Japan, 901(KP)
Loading...