Yokogawa CA700 Specifications

Bulletin CA700-EN
www.yokogawa.com/ymi
Pressure Calibrator
New Standard for Field Calibration
Pressure Calibrator
CA700
New Standard for Field Calibration
2
New Standard for Field Calibration
3
New Standard for Field Calibration
n High Accuracy and Long Stability
l Achieves the highest accuracy in the portable class
• Basic accuracy: Pressure (measurement) / 0.01% rdg.
Current/voltage (source/measurement) / 0.015% rdg.
n Rangeability
The CA700 is a highly accurate and highly functional pressure
l Achieves the highest resolution and widest range in the portable class
• 0.001 kPa (200.000 kPa range)
n User Support
l Strong support for field calibration and maintenance work
• Calibration procedures of pressure transmitters and pressure switches are
embedded.
• “As Found” , “As Left” data and error rate (%) can be recorded.
n Field Use
l IP54 dustproof and waterproof robust case enables use in harsh
environments.
n Accessories
l Three high-performance hand pump models for different pressure
ranges are available.
calibrator specifically designed for the calibration of differential
pressure and pressure transmitters, which are widely used in plant
processes these days and are improving in terms of accuracy and
functionality. The CA700 has achieved the highest accuracy and
widest range in the portable class, enabling field calibration with
greater accuracy than ever before.
New Standard for Field Calibration
4
New Standard for Field Calibration
5
High Performance and Functionality in a Compact Body
n High Accuracy and Long Stability
The highest accuracy in the portable class achieved with a silicon resonant sensor
The CA700 employs a silicon resonant pressure sensor with which Yokogawa has achieved the highest accuracy in the portable class of a 0.01% of reading for pressure measurement and 0.015% of reading for current and voltage
measurement. This calibrator is ideal for input and output testing of differen­tial pressure and pressure transmitters in that it correctly measures the input and output accuracy and calculates the error rate.
n Rangeability
Has the highest resolution and the widest range in its class
The CA700 has achieved a resolution of 0.001 kPa (200.000 kPa range) which is 10-fold higher than that of general field calibrators. The excellent sensing characteristics of a silicon resonant sensor make it possible to
achieve a very wide range and enable one CA700 calibrator to replace multiple field calibrators.
n User Support
Strong support for field calibration and maintenance work
Calibration procedures for differential pressure and pressure transmitters
are embedded, so routinely used as 50-percent step, up-and-down, 5-point calibration and 25-percent step, up-and-down, as well as 9-point calibration can be performed right out of the box. Calibration operation can be sped up
by following the navigation instructions, resulting in increased efficiency of
field work. Measurement data is automatically recorded to the calibrator so it can be saved to a USB memory stick or transferred to a PC as text data using a USB cable. Input and output values, error rate, date and time, and pass or fail can be displayed on a PC monitor.
Silicon Resonant Sensor
n Field Use
IP54 waterproof and dustproof robust case enables use of this calibrator in harsh environments
Calibration of pressure transmitters is usually carried out in harsh environ­ments where there are many pipes and handrails or in wet areas. The CA700
has a robust case designed according to the IP54 waterproof and
dustproof standards. A shoulder strap can
be attached to the calibrator to make it easy to carry around in the field. In addition, an easy-to-read LCD and tilt stand suitable for outdoor work exposed to direct sunlight extend the range of
use of this calibrator from indoor bench use to outdoor field use.
20 mA SIMULATE (Two-wire Transmitter Simulator)
The CA700 can also be used as a tr ansmitter simulator to carry out a loop test. It can absorb (SIN K) the set current from an ex ternal voltage generating device (e.g., a distributor system or PLC) of instrumentation equipment. 4-20mA current can be sourced with an accur acy of 0.015% of the reading.
Two-wire Transmitter Loop Check
DC mA signals can be measured by supplying power to the transmitter from a 24 V DC power supply. DC mA signal measurement and zero-point check can be performed with an accuracy of 0.015% of the reading. A 250-ohm resistor for HAR T and BR AIN communication is included in this calibrator so there is no need to attach an external resistor when connecting to a handy terminal.
Field Calibration of Differential Pressure and Pressure Transmitters
Calibration of pressure transmitters is required to accurately measure the input and output values and to calculate the error rate. The CA700 ensures reliable calibration with its function to accurately measure the input and output values of pressure and current. Additionally its embedded calibration procedures enable users to perform certain calibration following the prescribed procedure.
Input Command Check and Adjustment of Recorders and Controllers
Instrumentation loop test and operation/command check can be performed by sourcing DC 1-5 V / 4-20 mA ins trument ation signals with an accuracy of 0.015% of the reading. Furthermore, t wo patterns of linear sweep and step sweep can be selected (the sweep time can be specied from 15, 30, 45, and 60 sec).
Check and I/O Adjustment of an Electro-pneumatic Converter
Input and output adjustment of an electro-pneumatic converter is carried out by applying rising and falling currents of 0, 25, 50, 75, and 100% of the span. A reliable test can be car ried out with the CA700 that has a 4-20 mA step function for signal generation and a capability to accurately measure the generated pressure.
Pressure Switch Test
A pressure switch test measures the pressure at the time when the contact opens and closes and the resistance at the time when the dead band contact closes. A test procedure is embedded to enable users to carry out a test following the prescribed procedure.
Distributor
Two-wire transmitter
Two-wire transmitter
Hand pump Differential pressure and
pressure transmitter
Recorder
Electro-pneumatic converter
Hand pumpPressure switch
New Standard for Field Calibration
New Standard for Field Calibration
Supports Various Applications Three High-Performance Hand Pump Models Available
Low Pressure Hand Pump 91051
• Ultra-compact hand pump
• Pressure generation range: -83 to 700 kPa
• Ideal for pressure adjustment in low pressure range
• Smooth pressurization with less internal leaking
• Strainer preventing debris from entering the pump
included
Pneumatic Hand Pump 91056
• High-performance small hand pump with a wide
range of pressure generation
• Pressure generation range: -83 to 4000 kPa
• Pressure generation ranging from low to medium
pressures
• Fine adjustment and pressure reduction possible
with one hand
• Smooth pressurization with less internal leaking
• Strainer preventing debris from entering the pump
included
6
Hydraulic Hand Pump 91061
• High-performance hand pump capable of generat­ing a pressure of up to 70 MPa
• Pressure generation range: 0 to 70 MPa
• Easy grip operation to generate high pressures
• Smooth pressurization with less internal leaking
• Strainer preventing debris from entering the pump
included
7
188
96
264
Unit: mm
96
152
35
Unit: mm
Unit: mm
The external dimensions vary slightly depending on the stroke adjustment of the handle.
144
212
60
258
146 62
Unit: mm
The external dimensions vary slightly depending on the stroke adjustment of the handle.
New Standard for Field Calibration
8
New Standard for Field Calibration
9
n Basic Specifications (Measurement Unit) 23°C±3°C n General Specifications
Pressure Measurement
Model CA700-E-01 CA700–E–02 CA700–E–03
Pressure type Gauge
Measurement range
Measurement display range
Positive pressure: 0 to 200 kPa Negative pressure: -80 to 0 kPa
To 240.000 kPa To 1200.00 kPa To 4200.00 kPa
Positive pressure: 0 to 1000 kPa
Negative pressure: -80 to 0 kPa
Positive pressure: 0 to 3500 kPa
Negative pressure: -80 to 0 kPa
Resolution 0.001 kPa 0.01 kPa 0.01 kPa
Positive pressure 20 to 200 kPa: ±(0.01% of reading + 0.003 kPa)
Measurement accuracy *1 *
(6 months after calibration) (Tested after zero calibration)
Measurement accuracy *1 *
(1 year after calibration) (Tested after zero calibration)
Response time *
2
0 to 20 kPa: ±0.005 kPa
Positive pressure: ±(0.01% of reading + 0.04 kPa) Positive pressure: ±(0.01% of reading + 0.15 kPa)
Negative pressure: ±(0.2% of reading + 0.080 kPa) Negative pressure: ±(0.2% of reading + 0.08 kPa) Negative pressure: ±(0.2% of reading + 0.08 kPa)
Positive pressure 20 to 200 kPa: ±(0.01% of reading + 0.010 kPa)
2
0 to 20 kPa: ±0.012 kPa
Positive pressure: ±(0.01% of reading + 0.08 kPa) Positive pressure: ±(0.01% of reading + 0.30 kPa)
Negative pressure: ±(0.2% of reading + 0.090 kPa) Negative pressure: ±(0.2% of reading + 0.09 kPa) Negative pressure: ±(0.2% of reading + 0.09 kPa)
3
2.5 s or less
Allowable input 2.7 kPa abs to 500 kPa gauge 2.7 kPa abs to 3000 kPa gauge 2.7 kPa abs to 4500 kPa gauge
Internal volume Approx. 6 cm
Temperature coefficient
Influence of positional setup
*4
±(0.001% of reading + 0.001% of full scale)/°C or less
Zero-point drift ±0.3 kPa or less
3
Measurement fluid Gas and liquid (non-corrosive, non-flammable, non-explosive, and non-toxic fluids)
Measurement fluid temperature
*5
-10 to 50°C
Pressure sensor Silicon resonant sensor
Pressure sensor element Diaphragm
Pressure display units
(Pa, hPa, MPa, mbar, bar, atm, mmHg, inHg, gf/cm
2
, kgf/cm2, mmH2O@4°C, mmH2O@20°C, ftH2O@4°C, ftH2O@20°C, inH2O@4°C, inH2O@20°C, Torr, psi )
kPa and other units
Input port Rc 1/4 or 1/4 NPT female thread (selectable)
Measurement unit material
Diaphragm: Hastelloy C276 and input port: SUS316
DC Current Measurement
Range Resolution Measurement range Measurement accuracy (1 year) Remark
20 mA 1 µA 0 to ±20.000 mA 0.015% of reading + 3 µA
100 mA 10 µA 0 to ±100.00 mA 0.015% of reading + 30 µA
Input resistance: 10 Ω or less. The maximum display is 1.2-fold of range.
DC Voltage Measurement
Range Resolution Measurement range Measurement accuracy (1 year) Remark
5 V 0.1 mV 0 to ±5.0000 V 0.015% of reading + 0.5 mV
50 V 1 mV 0 to ±50.000V 0.015% of reading + 5 mV
Input resistance: approx. 1 MΩ. The maximum display is 1.1-fold of range.
Display Dot matrix LCD (320x240 dots)
Backlight LED
Display refresh rate Approx. 3 times/sec
Warm-up time Approx. 5 minutes
Power supply Six alkaline AA batteries
Battery life Approx. 35 hours when measuring current with the 24 V loop power supply OFF and approx 10 hours with the 24 V loop power supply ON
Auto power-off Approx. 60 minutes (the function can be disabled)
Insulation resistance 100 MΩ or more (500 VDC) between the input terminal and case and between the input port and case
Withstand voltage 500 VAC for 1 minute between the input terminal and case and between the input port and case
Protection grade IP54 dustproof and waterproof structure
Dimensions Approx. 264 (W) x 188 (H) x 96 (D) mm, excluding protrusions
Weight Approx. 2kg (including batteries)
Compliance standards
Operating temperature / humidity ranges
Storage temperature / humidity ranges
Safety: EN61010-1, EN61010-2-030, contamination class 2, measurement category 0 (other)
EMC: EN61326-1 Class A, EN55011 Class A Group 1
-10 to 50°C and 20 to 80%RH (no condensation)
-20 to 60°C and 20 to 80%RH (no condensation)
Interfaces Select and switch between USB A mass-storage device, USB mini-B communication device class, and mass storage class
External sensor A dedicated external sensor can be connected via a connector.
Accessories*
* The type of the included conversion connector varies depending on the suffix code (–P1 and –P2). For details, refer to “CA700 Accessories” on the next page.
n Number of Data Records
Device operation status
Measurement / Source
A set of 1.7m long black and red lead wires with alligator clips for generation and measurement, six alkaline AA batteries, R1 1/4” – 1/8” NPT female thread x 1, ferrite core x 2, R 1/4” – 1/4” NPT female thread x 1, accessory case, instruction manual (CD), startup guide, shoulder strap
n CA700 External Dimensions
Number of data records
that can be saved
Save 2000 45
Logging 2000 45
Number of files that can
be saved
Leak test 2000 45
Transmitter calibration (Number of calibration points: 5)
Pressure switch calibration
* The maximum number of files is 250.
* The total data capacity is approx. 3.5 MB
As Found 9 250
As Left 9 250
As Found 1 250
As Left 1 250
24 V Loop Power Supply
Supply voltage Remark
24 V ± 1 V Load current 24 mA when communication resistance OFF
24 V ± 6 V Load current 20 mA when communication resistance ON
n Basic Specifications (Generation Unit) 23°C±3°C
DC Current Source
Range Resolution Source range Accuracy (1 year) Remark (when communication resistance OFF)
20 mA 1 µA 0 to 20.000 mA
20 mA SIMULATE 1 µA 0 to 20.000 mA External power supply: 5 to 28 V. The maximum setting is 1.2-fold of range.
DC Voltage Source
Range Resolution Source range Accuracy (1 year) Remark
5 V 0.1 mV 0 to 5.0000 V 0.015% of setting + 0.5 mV Load resistance: 5 kΩ or more. The maximum setting is 1.1-fold of range.
n Measurement Unit Common Specifications
• CMRR: approx. 120 dB (50/60 Hz)
• NMRR: approx. 60 dB (50/60 Hz)
• Measurement terminal maximum input:
Voltage terminal 50 V DC Current terminal 120 mA
• Current terminal protective input: PTC protection
• Measurement unit voltage to ground: 50 V peak
0.015% of setting + 3 µA
n Generation Unit Common Specifications
• Generation load condition: C ≤ 0.1 µF L ≤ 10 mH
• Generation unit voltage limiter: Approx. 36 V
• Generation unit current limiter: Approx. 36 mA
• Generation unit voltage to ground: 42 V peak
• Sweep function: Step/Linear
Compliance voltage: 24 V. The maximum setting is 1.2-fold of range.
*1: Zero-point calibration condition: Under atmospheric pressure *2: Yokogawa’s pressure standards accuracy is excluded. *3: Conditions of response time measurement: The response time is defined as the time for the readout to settle within ±1% of
the full scale from the time when the positive pressure is released to atmosphere
when it is at its full-scale value (where the input unit is under no load). *4: Full scale of each model CA700-E-01: Positive pressure 200 kPa and negative pressure 80 kPa CA700-E-02: Positive pressure 1000 kPa and negative pressure 80 kPa CA700-E-03: Positive pressure 3500 kPa and negative pressure 80 kPa *5: Liquid temperature 5 to 50°C
n Hand Pump Specifications
Product name Model Pressure generation range MWP (maximum working pressure) Connection port Pressurized media
Hand Pump (Low Pressure) 91051 -83 to 700 kPa 1000 kPa 1/8” NPT female thread Air
Hand Pump (Pneumatic) 91056 -83 to 4000 kPa 5100 kPa 1/8” NPT female thread / 1/4” NPT female thread Air
Hand Pump (Hydraulic) 91061 0 to 70 MPa 70 MPa 1/4” NPT female thread / 1/4” NPT female thread Liquid
n Hand Pump External Dimensions
91051 91056 91061
New Standard for Field Calibration
10
New Standard for Field Calibration
11
n Model and Suffix Code n Model and Suffix Code
CA700
Product name Model Sufx code
CA700 General use type
–E All countries except Japan
–01 Gauge pressure: 200kPa
–02 Gauge pressure: 1000kPa
Pressure Calibrator
*1: Only kPa, Pa, hPa, MPa, mbar, bar, atm are available.
Separately Sold Accessories *
2
Product name Model Specification
Carrying Case 93050 Bag for the calibrator, accessories, and peripheral devices
Grabber Clip 98025 A set of separate red and black clips (for 2 m long wires)
Cleaning Unit 91040 For input connection port –P1 (R 1/4” male thread)
Cleaning Unit 91041 For input connection port –P2 (1/4” NPT male thread)
*2: These accessories are not included in the CA700 calibrator package.
Carrying Case 93050
Grabber Clip 98025
–03 Gauge pressure: 3500kPa
–U1 Metric units*
–U2 Metric units and non-metric units
–P1 Rc 1/4” female thread
–P2 1/4” NPT female thread
1
Cleaning Unit 91040/91041
Pressure Hand Pump Kits *
Product name Model Specification
Hand Pump Kit (Low Pressure) 91050 Hand pump (91051), connector set (91052), case (93052)
Hand Pump Kit (Pneumatic) 91055 Hand pump (91056), connector set (91057), case (93053)
Hand Pump Kit (Hydraulic) 91060 Hand pump (91061), connector set (91062), case (93053)
*6: These accessories are not included in the CA700 calibrator package at the time of purchase.
Hand Pump Kit (Low Pressure) 91050
Hand Pump Kit Accessories *
Product name Model Specification
Hand Pump (Low Pressure) 91051 Hand pump: -83 to 700 kPa (pressure generation range)
Hand Pump (Pneumatic) 91056 Hand pump: -83 to 4000 kPa (pressure generation range)
Hand Pump (Hydraulic) 91061 Hand pump: 0 to 70 MPa (pressure generation range)
Hand Pump Connector (Low Pressure) 91052 Connector set for the 91051 (quick adapter, pressure conversion adapter, flexible hose, sealing tape)
Hand Pump Connector (Pneumatic) 91057 Connector set for the 91056 (quick adapter, sealing cap, flexible hose, sealing tape, and hex wrench)
Hand Pump Connector (Hydraulic) 91062 Connector set for the 91061 (high-pressure hose, quick adapter, conversion adapter, sealing tape, and hex wrench)
Hand Pump Case (Low Pressure) 93052 Case for the 91051 (91051, 91052, and instruction manual)
Hand Pump Case (Pneumatic/Hydraulic) 93053 Case for the 91056/91061 (91056, 91057, 91061, 91062, and instruction manual)
*7: These accessories are included in the hand pump kit (91050, 91055, and 91060) at the time of purchase. They can also be purchased separately.
6
Hand Pump Kit (Pneumatic) 91055
7
Hand Pump Kit (Hydraulic) 91060
CA700 Accessories *
3
Product name Model Specification
Connector *
Connector *
Connector *
4
4
5
Lead Wires for Source / Measurement
91080 R 1/4" male thread to 1/8” NPT female thread conversion connector (for -P1)
91081 R 1/4” male thread to 1/4” NPT female thread conversion connector (for -P1)
91082 1/4” NPT male thread to 1/8” NPT female thread conversion connector (for –P2)
98064 Red and black alligator clip lead wires, 1.7 m long
Accessory Bag B9108XA For lead wires and connector
*3: Included in the CA700 calibrator package at the time of purchase. *4: Included in the package when suffix code –P1 is selected. *5: Included in the package when suffix code –P2 is selected.
Connector 91080 (R 1/4” male thread to 1/8” NPT female thread)
Lead Wires for Source/Measurement 98064
Connector 91081 (R 1/4” male thread to 1/4” NPT female thread)
Accessory Bag B9108XA
Connector 91082 (R 1/4” NPT male thread to 1/8” NPT female thread)
Hand Pump (Low Pressure) 91051
Hand Pump Connector (Low Pressure) 91052
Hand Pump Case (Low Pressure) 93052
Hand Pump (Pneumatic) 91056
Hand Pump Connector (Pneumatic) 91057
Hand Pump Case (Pneumatic/Hydraulic) 93053
Hand Pump (Hydraulic) 91061
Hand Pump Connector (Hydraulic) 91062
YOKOGAWA CORPORATION OF AMERICA (U.S.A.) Phone: +1-770-253-7000 Facsimile: +1-770-254-0928 YOKOGAWA EUROPE B. V. (THE NETHERLANDS) Phone: +31-88-4641000 Facsimile: +31-88-4641111 YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE)
Phone: +7-495-737-7868 Facsimile: +7-495-737-7869
Represented by:
World Wide Web site at
http://www.yokogawa.com/ymi
NOTICE
Before using the product, read the instruction manual
carefully to ensure proper and safe operation.
Tachihi Bld. No.2, 6-1-3 Sakaecho, Tachikawa-shi, Tokyo, 190-8586 Japan Phone: +81-42-534-1413 Facsimile: +81-42-534-1438
YOKOGAWA METERS & INSTRUMENTS CORPORATION International Sales Dept.
Subject to change without notice. [Ed: 01/b] Printed in Japan: Jun. 2013(B)/3,000(KP)
All Rights Reserved. Copyright © 2013, Yokogawa Meters & Instruments Corporation.
n Related Products
NEW
PC-based field device management tool
• Multi-vendor, multi-protocol support BRAIN, FOUNDATION and ISA100.11a
• Automatic device data acquisition upon connection to a device or a segment (Segment Viewer)
• Easy acquisition and diagnosis of device status(Device Viewer)
• Categorization, sorting and filtering (History)
• Multi-parameter set-up (Parameter Manager)
TM
Fieldbus H1, HART®, PROFIBUS
Versatile Device Management Wizard
Highly Accurate All-In-One Calibrator
• Dual display of source and measurement with an accuracy of 0.02%
• Multiple sources and measurements of DCV, DCmA, Ω, TC, RTD, Hz, and PULSE
• 24 V loop power supply and output signal measurement at the same time
• Current sink function to simulate a transmitter
• A variety of sweep functions selectable (Step, Linear, Program, and Sweep)
• Three power supply options available (AA batteries, AC adapter [sold separately], and NiMH batteries [sold separately])
Handy Calibrator
CA150
Handheld Terminal for BRAIN Communication
• Online monitoring and communication.
• Easy setup and alteration procedures in English prompting interactive sequences.
• Diagnostics/security with error messages. ( Self­check function, Security code protection of setpoints, Low battery voltage alarm and Automatic power-off)
• Printer prints out tag numbers and other parameters on the spot.
BRAIN Terminal
BT200
Best-Selling Field Calibrator Model with High Performance and Low Price
• Small, light, and high performance at a low price
• Source and measurement of DCV, DCmA, Ω, TC, RTD, Hz, and PULSE
• DMM-like operation with a rotary switch
• Source and measurement (counting) of dry contact pulses
Handy Calibrator
CA71
Clamp-on Measurement of 4-20 mA Instrumentation Signals
• Process signal measurement with no need to disconnect a loop
• Accuracy 0.2% + 5 dgt and resolution 0.01 mA
• Simultaneous display of percentage (%) of the measured value and span
• LED backlight ideal for measuring in dark places
• Thick signal wire with a diameter of up to 6 mm can be clamped easily.
Clamp-on Process Meter
CL420
Safety DMM Equipped with a Loop Power Supply and 4-20 mA Output
• 24 V loop power supply and measures output signal at the same time
• Resistor (250Ω) for HART and BRAIN communication embedded and selectable.
• Transmitter simulation (current sink) function
• Step, Auto-step, and Linear sweep functions are selectable
• 6,000-count DMM function
• EN61010-1 CATIV600V and CATIII1000V safety design