Specialized high performance model
for Loop Inspection!
Volt mA Calibrator
n Features
l Basic accuracy: 0.015%
(Source&Meas. accuracy of Voltage mA)
l 20mA SIMULATE (SINK) function
l Simultaneously supplies 24V loop power and measure output
signal with high accuracy
l HART/BRAIN comm. resistance (250Ω) embedded
l Sub display displays span% of the source value
l Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
Useful function installed
n Addition of sub display
The sub display additionally displays
span%, source value of voltage or resistance, while the main displays setting value.
n Corresponds to 2WAY
Power supply
Power is supplied by 2 ways :
AA Alkaline batteries or AC Adapter
AC Adapter is sold
separately
2
Source and Measure SimplicitySeries
Specialized high performance model
for TC simulate and RTD simulate!
TC Calibrator
n Features
l Basic accuracy: 0.5°C (Typical of TC type K)
*Including accuracy of internal RJC
l Corresponds to 16 types of TC standard
(JIS/IEC/DIN/ASTM/GOST R)
l Sub display displays value of voltage source and span%
l Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
l Corresponds to other TC types by mV source function
l Measures TC sensor output as a thermometer
RTD Calibrator
n Features
l Basic accuracy: 0.3°C (Typical of PT100)
l Corresponds to 14 types of RTD standard (JIS/IEC/GOST R)
l Sub display displays value of resistance source and span%
l Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
l Corresponds to 2,3,4 wire. Realizes RTD simulation
l Measures output of RTD sensor as a thermometer
3
Supports efficient operation with various types of source pattern
n Step sweep function
Sources by 25% step automatically from 0% to 100% of span
which improves efficiency of operation. It can correspond to
various response time of field devices.
(15/30/45/60 seconds)
n Manual step function
Sources by 25% step manually from 0% to 100% of span.
n Linear sweep function
Sources continuously from 0% to 100% and is capable to
check meter and make linearity tests. Sweep time can be
selected by 15/30/45/60 seconds.
20mA
16mA
12mA
100°C
75°C
50°C
25°C
0°C
20mA
8mA
4mA
KEY
4-20mA Step sweep source example
100%
75%
50%
25%
0%
Selection: 15/30/45/60 sec.
0°C - 100°C Manual step source example
100%
75%
50%
25%
0%
Linear sweep source example
100%
n Span check function
Switches sources 0%ó100% by one touch. With this function, it makes it simple to make adjustment and to inspect
the open and close operation of valves.
4
4mA
20mA
4mA
Selection: 15/30/45/60 sec.
0%
4-20mA Span check example
100%
0%
KEYKEY
Source and Measure SimplicitySeries
Corresponds to various application
20mA SIMULATE
n
CA310 is capable to execute a loop check by simulating a transmit ter, sinking the
current signal from the external source (dis tributor). It achieves the high accuracy
0.015% of set ting to source 4-20m A.
2 wire transmitter
TC SIMULATE
n
CA32 0 corresponds to 16 types of TC for sourcing. It achieves the high basic
accuracy of 0.5°C (t ypical of t ype K ), three times better than the previous model
which enables it to operate a highly reliable test . Additionally, the dif ference of
temperature between object s can be compensated, by using external RJ sensor or
a compensating lead wire.
Distributor
Two-wire Transmitter Loop Check
n
DC mA signals can be measured by supplying power to t he transmitter from a 24 V
DC power supply. DC mA measurement and zero-point check can be performed
with an accuracy of 0.015% of reading. A 25 0-ohm resistor for HART and BR AIN
communication is included in this calibrator so there is no need to attach an
external resistor when connecting to a handy terminal.
2 wire transmitter
TC MEASURING
n
CA32 0 can measure the output of TC like a thermometer. It achieves the basic
accuracy of 0.5°C (t ypical of t ype K ), three times better than the previous model
and is for multiple use for process temperature measuring by corresponding to 16
types of TC.
Controller
RTD SIMULATE
n
CA330 corresponds to 14 types of RT D for sourcing. It achieves the high basic
accuracy of 0.3°C (t ypical of type PT100), twice better than the previous model
which enables it to operate a high reliable test.
Temperature transducer
Process temperature
RTD MEASURING
n
CA330 can measure the output of RT D like a thermometer. It achieves the basic
accuracy of 0.3°C (t ypical of type PT100), twice better than the previous model and
is for multiple use of process temperature measuring by corresponding to 14 types
of RTD.
RTD sensor
5
n Basic Specification (Source function) CA310
DC Current source
RangeResolutionSource rangeAccuracy (1 year)Note
20mA1μA0.000 to 24.000mA0.015% of setting+3μACompliance voltage: 24V
20mA SIMULATE1μA0.000 to 24.000mA0.015% of setting+3μAExternal power supply: 5V to 28V
DC Voltage source
RangeResolutionSource rangeAccuracy (1 year)Note
500mV10μV0.00 to 550.00mV0.015% of setting+50μVMax. current: 10mA
5V0.1mV0.0000 to 5.5000V0.015% of setting+0.5mVMax. current: 10mA
30V1mV0.000 to ±33.000V0.015% of setting+5mVMax. current: 1mA
Accuracy is specified at ambient temperature(Ta) of :23±5°C
Temperature effect: 0.005% or Range/°C is added for other ambient temperature (Ta < 18°C, Ta > 28°C
)
n Basic Specification (Measurement function) CA310
500mV10μV0 to ±550.00mV0.015% of reading+50μVInput resistance: approx. 1MΩ
5V0.1mV0 to ±5.5000V0.015% of reading+0.5mVInput resistance: approx. 1MΩ
30V1mV0 to ±33.000V0.015% of reading+5mVInput resistance: approx. 1MΩ
50V1mV0 to ±55.000V0.015% of reading+5mVInput resistance: approx. 1MΩ
Input resistance: less than 10Ω
24V Loop Power Supply
RangeSupply voltageNote
Loop Power
Accuracy is specified at ambient temperature(Ta) of :23±5°C
Temperature effect: 0.005% or Range/°C is added for other ambient temperature (Ta < 18°C, Ta > 28°C
n Measurement Unit Common Specifications
• CMRR approx. 120dB (50/60Hz)
• NMRR approx. 60dB (50/60Hz)
• Measurement terminal maximum input:
Voltage terminal DC 50V
Current terminal 60mA
• Current terminal protective input: PTC protection
• Maximum allowable applied voltage: Measure terminal
to ground 50V peak
n Generation Unit Common Specifications
• Generation unit voltage limiter: Approx. 36 V
• Generation unit current limiter: Approx. 30 mA
• Sweep function Step (25%)/ Linear
• Step time 15sec/30sec/45sec/60sec
• Generation load condition: C < 0.1μF, L < 10mH
• Output resistance: under 10mΩ
• Output response time: under 300ms
• Maximum allowable applied voltage: Source terminal to
ground 42V peak
24V±1VCommunication resistance OFF: load current 24mA
24V±6VCommunication resistance ON: load current 20mA
)
6
Source and Measure SimplicitySeries
n Basic specification (Source/ Measure) CA320
Thermocouple (TC) Source/Measure (Terminal A: TC plug terminal) t: Temperature of Source/Meas.
TC
K
E
J
T
N
L
U
R
S
B
C
XK
A
D
(W3Re/W25Re)
Extra
TC
Errors of TC are not included
Accuracy is specified at ambient temperature(Ta) of :23±5°C using internal junction compensation.
Temperature effect: 0.05%/°C is added for other ambient temperature (Ta < 18°C, Ta > 28°C)
The display resolution for source / measure is 0.1°C
About formula of accuracy
The accuracy of source or measuring are defined by constant value or formula of linear expression.
Example) Accuracy of type K at measuring point of 1000.0°C is ±(0.5+(1000.0-500)x0.02%)°C=±0.6°C
G
(W/W26Re)
PLATINEL II
Source/Meas. Temperature
-200.0°C ≤ t < 0.0°C0.5+|t|x0.3%0.5+|t|x0.3%
0.0°C ≤ t < +500.0°C0.50.5
+500.0°C ≤ t ≤ +1372.0°C0.5+(t-500)x0.03%0.5+(t-500)x0.02%
-250.0°C ≤ t < -200.0°C1.1+(|t|-200)x2.0%1.1+(|t|-200)x2.0%
-200.0°C ≤ t < 0.0°C0.5+|t|x0.3%0.5+|t|x0.3%
0.0°C ≤ t < +500.0°C0.50.5
+500.0°C ≤ t ≤ +1000.0°C0.5+(t-500)x0.02%0.5+(t-500)x0.02%
-210.0°C ≤ t < 0.0°C0.5+|t|x0.3%0.5+|t|x0.3%
0.0°C ≤ t ≤ +1200.0°C0.5+tx0.02%0.5+tx0.02%
-250.0°C ≤ t < -200.0°C1.1+(|t|-200)x2.5%1.1+(|t|-200)x2.5%
0.0°C ≤ t ≤ +400.0°C0.50.5
-200.0°C ≤ t < 0.0°C0.6+|t|x0.4%0.6+|t|x0.3%
0.0°C ≤ t ≤ +1300.0°C0.60.6
-200.0°C ≤ t < 0.0°C0.5+|t|x0.15%0.5+|t|x0.15%
0.0°C ≤ t ≤ +900.0°C0.50.5
-200.0°C ≤ t < 0.0°C0.5+|t|x0.2%0.5+|t|x0.2%
0.0°C ≤ t ≤ +600.0°C0.50.5
-20.0°C ≤ t < 0.0°C2.02.0
+100.0°C ≤ t ≤ +1767.0°C1.41.4
-20.0°C ≤ t < 0.0°C2.02.0
+100.0°C ≤ t ≤ +1768.0°C1.41.4
+600.0°C ≤ t < +800.0°C1.21.5
+1000.0°C ≤ t ≤ +1820.0°C1.01.1
0.0°C ≤ t < +1000.0°C0.80.8
+1000.0°C ≤ t ≤ +2315.0°C0.8+(t-1000)x0.06%0.8+(t-1000)x0.06%
-200.0°C ≤ t < 0.0°C0.4+|t|x0.2%0.4+|t|x0.2%
+300.0°C ≤ t ≤ +800.0°C0.50.5
0.0°C ≤ t < +1000.0°C1.01.0
+1000.0°C ≤ t ≤ +2500.0°C1.0+(t-1000)x0.06%1.0+(t-1000)x0.06%
0.0°C ≤ t < +300.0°C1.41.8
+1500.0°C ≤ t ≤ +2315.0°C1.82.2
+100.0°C ≤ t < +300.0°C1.41.8
+1500.0°C ≤ t ≤ +2315.0°C1.82.2
0.0°C ≤ t < +100.0°C0.61.8
+1000.0°C ≤ t ≤ +1395.0°C1.02.2
Accuracy (1 year)
Source Accuracy [°C]Meas. Accuracy [°C]
Standard or Regulation
IEC60584-1
JIS C1602
IEC60584-1
IEC60584-1
IEC60584-1 -200.0°C ≤ t < 0.0°C0.5+|t|x0.3%0.5+|t|x0.3%
IEC60584-1
DIN 43710
DIN 43710
IEC60584-1 0.0°C ≤ t < +100.0°C2.01.4
IEC60584-1 0.0°C ≤ t < +100.0°C2.01.4
IEC60584-1 +800.0°C ≤ t < +1000.0°C1.01.2
IEC60584-1
GOST R 8.585-2001 0.0°C ≤ t < +300.0°C0.40.4
IEC60584-1
ASTM E1751/E1751M–09e1 +300.0°C ≤ t < +1500.0°C1.21.2
ASTM E1751/E1751M–09e1 +300.0°C ≤ t < +1500.0°C1.21.2
ASTM E1751/E1751M–09e1 +100.0°C ≤ t < +1000.0°C0.81.8
DC Voltage Source and Measurement
RangeResolution
90mV1μV-11.000 to ±99.999mV0.015% of setting+10μV0.015% of reading+10μVMax. output current: 10mA
Accuracy is specified at ambient temperature (Ta) of 23 ± 5°C
Temperature effect : 0. 005% of Range/°C is added for other ambient temperature (Ta<18°C, Ta > 28°C)
Source Measure
range
SourceMeasure
n Common source specification
• Output resistance: under 40mΩ
• Output response: under 300msec
• Max. load: C<0.1μF, L<10mH
Accuracy (1 year)
Notes
7
n Basic specification (Source/ Measure) CA330
RTD Source/Measure t: Temperature of Source/Meas.
Accuracy (1 year)
RTDCoefficient
3851
3850
PT100
3916
3926
PT2003851
PT5003851
PT10003851
Source/Meas. Temp.
-200.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +800.0°C0.3+tx0.033%0.3+tx0.033%
-200.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +630.0°C0.3+tx0.033%0.3+tx0.033%
-200.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +510.0°C0.3+tx0.033%0.3+tx0.033%
-200.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +630.0°C0.3+tx0.033%0.3+tx0.033%
-200.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +630.0°C0.3+tx0.050%0.3+tx0.050%
-200.0°C ≤ t < 0.0°C0.40.4
0.0°C ≤ t ≤ +630.0°C0.4+tx0.033%0.4+tx0.033%
-200.0°C ≤ t < 0.0°C0.20.2
0.0°C ≤ t ≤ +630.0°C0.2+tx0.033%0.2+tx0.033%
Source
Accuracy [°C]
Meas.
Accuracy [°C]
Cu10427-100.0°C ≤ t ≤ +260.0°C1.51.50.1-3mAMinco Application Aid #18
Ni120627-80.0°C ≤ t ≤ +260.0°C0.20.20.1-3mAMinco Application Aid #18
PT503851
PT50G—
Extra
PT100G—
RTD
Cu50M—
Cu100M—
Accuracy is specified at ambient temperature (Ta) of 23±5°C.
Temperature effect: 0.05°C/°C is added for other ambient temperature (Ta<18°C, Ta > 28°C)
The display resolution for source / measure is 0.1°C
Above accuracy is specified for 4 wire measuring.
Accuracy for 3 wire measuring: 1.0°C to Cu10; 0.6°C to Pt50, Pt50G and Cu50M; 0.3°C to other RTD is each added, on condition the resistance of all cables are the same.
Accuracy for 2 wire measuring: Same with 3 wire measuring excluding resistance of cables.
About formula of accuracy
The accuracy of source or measuring are defined by constant value or formula of linear expression.
Example) Accuracy of PT100(3851) at measuring point of 100.0°C is ±(0.3+100.0x0.033%)°C=±0.333°C
-200.0°C ≤ t < 0.0°C0.40.4
0.0°C ≤ t ≤ +630.0°C0.4+tx0.050%0.4+tx0.050%
-200.0°C ≤ t < 0.0°C0.40.4
0.0°C ≤ t ≤ +800.0°C0.4+tx0.050%0.4+tx0.050%
-200.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +630.0°C0.3+tx0.033%0.3+tx0.033%
-180.0°C ≤ t < 0.0°C0.40.4
0.0°C ≤ t ≤ +200.0°C0.4+tx0.050%0.4+tx0.050%
-180.0°C ≤ t < 0.0°C0.30.3
0.0°C ≤ t ≤ +200.0°C0.3+tx0.033%0.3+tx0.033%
Excitation
current
0.1-3mA
0.1-3mA
0.1-3mA
Standard or Regulation
IEC60751
JIS C 1604
JIS C 1604 1989
(Pt100)
JIS C 1604 1989
(JPt100)
0.1-3mAMinco Application Aid #18
0.05-0.8mAIEC60751
0.05-0.6mAIEC60751
0.05-0.6mAIEC60751
0.1-3mAIEC60751
0.1-3mAGOST R 8.625-2006
0.1-3mAGOST R 8.625-2006
0.1-3mAGOST R 8.625-2006
0.1-3mAGOST R 8.625-2006
Resistance source and measure
RangeResolution
Source and Meas.
Range
SourceMeasurement
500Ω10mΩ0.00 to 550.00Ω0.025% of setting+0.1Ω0.025% of reading+0.1ΩExcitation current 0.1 to 3mA
3000Ω100mΩ0.0 to 3300.0Ω0.025% of setting+0.5Ω0.025% of reading+0.5ΩExcitation current 0.05 to 0.6mA
Accuracy is specified at ambient temperature (Ta) of 23±5°C. Temperature effect: Add the accuracy of ±(0.005% of range) /°C for other ambient temperature (Ta<18°C, Ta > 28°C) ..
Above accuracy is defined for 4-wire measuring.
Accuracy for 3 wire measuring: 0.05Ω to 500Ω range; 0.2Ω to 3000Ω range is added, on condition the resistance of all cables are the same.
Accuracy for 2 wire measuring: Same with 3 wire measuring on condition the resistance of cables are excluded.
n Common measurement specification
• Excitation current: Method of voltage surge current
meausre (typical 0.78mA at 0Ω, 0.6mA at 500Ω,
0.27mA at 3000Ω
• Disconnection detection: Detects when Hi terminal is
open.
n Common source specificaiton
• Response time: Under 5msec (Excluding 3000Ω range,
PT500 and PT1000)
• Max. load: C < 0.1μF, L < 10mH
• Sweep: Step (25%) / linear
• Step time: 15sec / 30sec / 45sec/ 60sec
• Allowable resistance for measuring cables: under 10Ω
8
Accuracy (1 year)
Note
Unit: mm
5.8
Unit: mm
192
(0.8
(3.3
(7.9
Unit: mm
Unit: mm
192
Unit: mm
192
Source and Measure SimplicitySeries
n General Specification
DisplaySegment LCD
BacklightLED (Selection of "Constantly ON", "Constantly OFF" or "Auto off by approx. 2min")
Display refresh rateApprox. 1sec.
Warm-up timeApprox. 5min.
Power supplyFour alkaline AA batteries, Dedicated AC Adapter (Sold separately)
Auto Power OffApprox. 20min. (Disabled by setting)
DimensionsApprox. 90 (W)×192 (H)×42 (D)
WeightApprox. 440g
Standard
Operating temperature / humidity ranges-10 to 55°C 20 to 80%RH (without condensation)
Storage temperature / humidity ranges-20 to 60°C 90% RH or less (without condensation)
Accessories
Safety: EN61010-1 / EN61010-2-030
EMC: EN61326-1 Class A Table 2. EN55011 Class A Group1
CA310: Carrying case (B9108NK)/ Lead cables (a set of black and red lead wires for generation and measurement /98064) /four AA alkaline
batteries/ Instruction manual
CA320: Carrying case (B9108NK)/ Lead cables (a set of black and red lead wires for generation and measurement / 98040) /Binding post (Red
Black 1 piece/ 99045) / four AA alkaline batteries/ Instruction manual
CA330: Carrying case (B9108NK)/Lead cables( a set of 1 black and 3 red lead wires for generation and measurement/98035) /Binding post (Red
Black 1 piece/99045)/ Binding post (Red Red 1 piece / 99046) /four AA alkaline batteries/ Instruction manual
n External Dimensions (CA300 series)
CA310
90
42
n External Dimensions (Accessories)
RJ sensor
90080
3.3
(50)5
35
10
50
1500
CA320CA330
90
TC connector
(2.5) External diameter
42
(4.8)
(2.3)
(7.9)
(3.3)
(16.8)
(20.6)(12.7)
90
42
)
)
)
9
n Model Name and Model Code
Main units (CA300 series)
NameModelSuffix CodeDescription
Volt mA CalibratorCA310Voltage and Current Simulate Model
TC CalibratorCA320Thermocouple Simulate Model
RTD CalibratorCA330RTD Simulate Model
/TEAdd °F setting procedure (for CA320, CA330)
Accessories Sold Separately
*1
NameModelDescription
AC Adapter94013Input: AC 120V, 50/60Hz
AC Adapter94016Input: AC 220V to 240V, 50/60Hz
RJ Sensor
*2
Grabber Clip
Rubber Boots
*3
*4
90080for CA320: RJ (Reference Junction)
98025for CA series: separate type (one set of Red and Black 2.0m)
93060for protection of main unit
Strap97040for hanging main unit on wall with rubber boot
Accessory CaseB9108XAfor accessories
TC Mini Plug Set 1
TC Mini Plug Set 2
*1: These accessories are not included with main unit when purchased
*2: RJ sensor is dedicated for CA320. It is unable to be used for CA71 and CA150
*3: It is impossible to be used with binding post (model no. 99045/99046)
*4: It is impossible to put in the carrying case with rubber boot (93060)
*5: TC mini plugs are dedicated for CA320. Other types of mini plugs are required to be prepared by customer.
*1: These accessories are included with main unit. Included types of accessories are different according to the type of main unit.
*2: Included with CA310 when purchased.
*3: Included with CA330 when purchased.
*4: Included with CA320 when purchased.
*5: Included with CA320/CA330 when purchased.
*6: Included with CA330 when purchased.
*7: It is impossible to put in main unit with rubber boots..
*2
*3
*4
*5
*6
*7
*1
98064 for CA310, Alligator Clip Cable (Red Black 1 set/ 1.7m)
98035 for CA330, Alligator Clip Cable (Red x 3pcs, Black x 1pce 1 set/ 1.7m)
98040 for CA320, Alligator Clip Cable (Red Black 1 set/ 1.7m)
99045 1 short plate attached
99046 1 short plate attached
B9108NK for main unit and lead cables
Lead Cables
98064
Binding Post
99045
Lead Cables
98035
Binding Post
99046
Lead Cables
98040
Carrying Case
B9108NK
11
n Related Products
PC-based field device management tool
• Multi-vendor, multi-protocol support BRAIN, FOUNDATIONTM
Fieldbus H1, HART
• Automatic device data acquisition upon connection to a device or
a segment (Segment Viewer)
• Easy acquisition and diagnosis of device
status (Device Viewer)
• Categorization, sorting and filtering (History)
• Multi-parameter set-up (Parameter Manager)
®
, PROFIBUS and ISA100.11a
Versatile Device Management Wizard
New Standard for Field Calibration
• Achieves the highest accuracy in the portable class !
• Achieves the highest resolution and widest range in portable class
• 0.001kPa (200.000kPa range)
• Calibration procedures of pressure
transmitters and pressure switches are
embedded.
Pressure Calibrator
CA700
Handheld Terminal for BRAIN Communication
• Online monitoring and communication.
• Easy setup and alteration procedures in English
prompting interactive sequences.
• Diagnostics/security with error messages. (Selfcheck function, Security code protection of setpoints,
Low battery voltage alarm and Automatic power-off)
• Printer prints out tag numbers and other
parameters on the spot.
BRAIN Terminal
BT200
Highly Accurate All-In-One Calibrator
• Dual display of source and measurement with an accuracy of 0.02%
• Multiple sources and measurements of DCV, DCmA,
Ω, TC, RTD, Hz, and PULSE
• 24 V loop power supply and output signal
measurement at the same time
• Current sink function to simulate a transmitter
• A variety of sweep functions selectable (Step, Linear,
Program, and Sweep)
• Three power supply options available (AA batteries,
AC adapter [sold separately], and NiMH batteries
[sold separately])
Handy Calibrator
CA150
Safety DMM Equipped with a Loop Power Supply and 4-20 mA
Output
• 24 V loop power supply and measures output signal
at the same time
• Resistor (250Ω) for HART and BRAIN communication
embedded and selectable.
• Transmitter simulation (current sink) function
• Step, Auto-step, and Linear sweep functions are
selectable
• 6,000-count DMM function
• EN61010-1 CATIV600V and CATIII1000V safety design
Process Multimeter
CA450
http://www.yokogawa.com/ymi
YOKOGAWA METERS & INSTRUMENTS CORPORATION
Global Sales Dept. /Phone: +81-422-52-6237 Facsimile: +81-422-52-6462 E-mail: tm@cs.jp.yokogawa.com
YOKOGAWA CORPORATION OF AMERICA Phone: +1-770-253-7000 Facsimile: +1-770-254-0928
YOKOGAWA EUROPE B.V. Phone: +31-88-4641000 Facsimile: +31-88-4641111
YOKOGAWA SHANGHAI TRADING CO., LTD. Phone: +86-21-6239-6363 Facsimile: +86-21-6880-4987
YOKOGAWA ELECTRIC KOREA CO., LTD. Phone: +82-2-2628-3810 Facsimile: +82-2-2628-3899
YOKOGAWA ENGINEERING ASIA PTE. LTD. Phone: +65-6241-9933 Facsimile: +65-6241-2606
YOKOGAWA INDIA LTD. Phone: +91-80-4158-6000 Facsimile: +91-80-2852-8656
YOKOGAWA ELECTRIC CIS LTD. Phone: +7-495-737-7868 Facsimile: +7-495-737-7869
YOKOGAWA AMERICA DO SUL LTDA. Phone: +55-11-5681-2400 Facsimile: +55-11-5681-4434
YOKOGAWA AUSTRALIA PTY. LTD. Phone: +61-2-8870-1100 Facsimile: +61-2-8870-1111
YOKOGAWA MIDDLE EAST & AFRICA B.S.C(c) Phone: +973-17-358100 Facsimile: +973-17-336100
Subject to change without notice.[Ed: 01/b] Printed in Japan: Nov. 2015(B)/3,000(KP)