Yokogawa CA300 Specifications

Process Calibrator
CA300 series
Process Calibrator
Source and Measure Simplicity
series
Bulletin CA300-EN
www.yokogawa.com/ymi
Specialized high performance model for Loop Inspection!
Volt mA Calibrator
n Features
l Basic accuracy: 0.015%
(Source&Meas. accuracy of Voltage mA)
l 20mA SIMULATE (SINK) function l Simultaneously supplies 24V loop power and measure output
signal with high accuracy
l HART/BRAIN comm. resistance (250Ω) embedded l Sub display displays span% of the source value l Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
Useful function installed
n Addition of sub display
The sub display additionally displays span%, source value of voltage or resis­tance, while the main displays setting value.
n Corresponds to 2WAY
Power supply
Power is supplied by 2 ways : AA Alkaline batteries or AC Adapter
AC Adapter is sold separately
2
Source and Measure Simplicity Series
Specialized high performance model for TC simulate and RTD simulate!
TC Calibrator
n Features
l Basic accuracy: 0.5°C (Typical of TC type K)
*Including accuracy of internal RJC
l Corresponds to 16 types of TC standard
(JIS/IEC/DIN/ASTM/GOST R)
l Sub display displays value of voltage source and span% l Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
l Corresponds to other TC types by mV source function l Measures TC sensor output as a thermometer
RTD Calibrator
n Features
l Basic accuracy: 0.3°C (Typical of PT100) l Corresponds to 14 types of RTD standard (JIS/IEC/GOST R) l Sub display displays value of resistance source and span% l Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
l Corresponds to 2,3,4 wire. Realizes RTD simulation l Measures output of RTD sensor as a thermometer
3
Supports efficient operation with various types of source pattern
n Step sweep function
Sources by 25% step automatically from 0% to 100% of span which improves efficiency of operation. It can correspond to various response time of field devices. (15/30/45/60 seconds)
n Manual step function
Sources by 25% step manually from 0% to 100% of span.
n Linear sweep function
Sources continuously from 0% to 100% and is capable to check meter and make linearity tests. Sweep time can be selected by 15/30/45/60 seconds.
20mA
16mA
12mA
100°C
75°C
50°C
25°C
0°C
20mA
8mA
4mA
KEY
4-20mA Step sweep source example
100%
75%
50%
25%
0%
Selection: 15/30/45/60 sec.
0°C - 100°C Manual step source example
100%
75%
50%
25%
0%
Linear sweep source example
100%
n Span check function
Switches sources 0%ó100% by one touch. With this func­tion, it makes it simple to make adjustment and to inspect the open and close operation of valves.
4
4mA
20mA
4mA
Selection: 15/30/45/60 sec.
0%
4-20mA Span check example
100%
0%
KEY KEY
Source and Measure Simplicity Series
Corresponds to various application
20mA SIMULATE
n
CA310 is capable to execute a loop check by simulating a transmit ter, sinking the current signal from the external source (dis tributor). It achieves the high accuracy
0.015% of set ting to source 4-20m A.
2 wire transmitter
TC SIMULATE
n
CA32 0 corresponds to 16 types of TC for sourcing. It achieves the high basic accuracy of 0.5°C (t ypical of t ype K ), three times better than the previous model which enables it to operate a highly reliable test . Additionally, the dif ference of temperature between object s can be compensated, by using external RJ sensor or a compensating lead wire.
Distributor
Two-wire Transmitter Loop Check
n
DC mA signals can be measured by supplying power to t he transmitter from a 24 V DC power supply. DC mA measurement and zero-point check can be performed with an accuracy of 0.015% of reading. A 25 0-ohm resistor for HART and BR AIN communication is included in this calibrator so there is no need to attach an external resistor when connecting to a handy terminal.
2 wire transmitter
TC MEASURING
n
CA32 0 can measure the output of TC like a thermometer. It achieves the basic accuracy of 0.5°C (t ypical of t ype K ), three times better than the previous model and is for multiple use for process temperature measuring by corresponding to 16 types of TC.
Controller
RTD SIMULATE
n
CA330 corresponds to 14 types of RT D for sourcing. It achieves the high basic accuracy of 0.3°C (t ypical of type PT100), twice better than the previous model which enables it to operate a high reliable test.
Temperature transducer
Process temperature
RTD MEASURING
n
CA330 can measure the output of RT D like a thermometer. It achieves the basic accuracy of 0.3°C (t ypical of type PT100), twice better than the previous model and is for multiple use of process temperature measuring by corresponding to 14 types of RTD.
RTD sensor
5
n Basic Specification (Source function) CA310
DC Current source
Range Resolution Source range Accuracy (1 year) Note
20mA 1μA 0.000 to 24.000mA 0.015% of setting+3μA Compliance voltage: 24V
20mA SIMULATE 1μA 0.000 to 24.000mA 0.015% of setting+3μA External power supply: 5V to 28V
DC Voltage source
Range Resolution Source range Accuracy (1 year) Note
500mV 10μV 0.00 to 550.00mV 0.015% of setting+50μV Max. current: 10mA
5V 0.1mV 0.0000 to 5.5000V 0.015% of setting+0.5mV Max. current: 10mA
30V 1mV 0.000 to ±33.000V 0.015% of setting+5mV Max. current: 1mA
Accuracy is specified at ambient temperature(Ta) of :23±5°C Temperature effect: 0.005% or Range/°C is added for other ambient temperature (Ta < 18°C, Ta > 28°C
n Basic Specification (Measurement function) CA310
DC Current measurement
Range Resolution Measurement range Accuracy (1 year) Note
20mA 1μA 0 to ±24.000mA 0.015% reading+3μA
50mA 1μA 0 to ±60.000mA 0.015% reading+3μA
DC Voltage measurement
Range Resolution Measurement range Accuracy (1 year) Note
500mV 10μV 0 to ±550.00mV 0.015% of reading+50μV Input resistance: approx. 1MΩ
5V 0.1mV 0 to ±5.5000V 0.015% of reading+0.5mV Input resistance: approx. 1MΩ
30V 1mV 0 to ±33.000V 0.015% of reading+5mV Input resistance: approx. 1MΩ
50V 1mV 0 to ±55.000V 0.015% of reading+5mV Input resistance: approx. 1MΩ
Input resistance: less than 10Ω
24V Loop Power Supply
Range Supply voltage Note
Loop Power
Accuracy is specified at ambient temperature(Ta) of :23±5°C Temperature effect: 0.005% or Range/°C is added for other ambient temperature (Ta < 18°C, Ta > 28°C
n Measurement Unit Common Specifications
• CMRR approx. 120dB (50/60Hz)
• NMRR approx. 60dB (50/60Hz)
• Measurement terminal maximum input: Voltage terminal DC 50V Current terminal 60mA
• Current terminal protective input: PTC protection
• Maximum allowable applied voltage: Measure terminal to ground 50V peak
n Generation Unit Common Specifications
• Generation unit voltage limiter: Approx. 36 V
• Generation unit current limiter: Approx. 30 mA
• Sweep function Step (25%)/ Linear
• Step time 15sec/30sec/45sec/60sec
• Generation load condition: C < 0.1μF, L < 10mH
• Output resistance: under 10mΩ
• Output response time: under 300ms
• Maximum allowable applied voltage: Source terminal to ground 42V peak
24V±1V Communication resistance OFF: load current 24mA
24V±6V Communication resistance ON: load current 20mA
6
Source and Measure Simplicity Series
n Basic specification (Source/ Measure) CA320
Thermocouple (TC) Source/Measure (Terminal A: TC plug terminal) t: Temperature of Source/Meas.
TC
K
E
J
T
N
L
U
R
S
B
C
XK
A
D
(W3Re/W25Re)
Extra
TC
Errors of TC are not included Accuracy is specified at ambient temperature(Ta) of :23±5°C using internal junction compensation. Temperature effect: 0.05%/°C is added for other ambient temperature (Ta < 18°C, Ta > 28°C) The display resolution for source / measure is 0.1°C
About formula of accuracy
The accuracy of source or measuring are defined by constant value or formula of linear expression. Example) Accuracy of type K at measuring point of 1000.0°C is ±(0.5+(1000.0-500)x0.02%)°C=±0.6°C
G
(W/W26Re)
PLATINEL II
Source/Meas. Temperature
-200.0°C ≤ t < 0.0°C 0.5+|t|x0.3% 0.5+|t|x0.3%
0.0°C ≤ t < +500.0°C 0.5 0.5
+500.0°C ≤ t ≤ +1372.0°C 0.5+(t-500)x0.03% 0.5+(t-500)x0.02%
-250.0°C ≤ t < -200.0°C 1.1+(|t|-200)x2.0% 1.1+(|t|-200)x2.0%
-200.0°C ≤ t < 0.0°C 0.5+|t|x0.3% 0.5+|t|x0.3%
0.0°C ≤ t < +500.0°C 0.5 0.5
+500.0°C ≤ t ≤ +1000.0°C 0.5+(t-500)x0.02% 0.5+(t-500)x0.02%
-210.0°C ≤ t < 0.0°C 0.5+|t|x0.3% 0.5+|t|x0.3%
0.0°C ≤ t ≤ +1200.0°C 0.5+tx0.02% 0.5+tx0.02%
-250.0°C ≤ t < -200.0°C 1.1+(|t|-200)x2.5% 1.1+(|t|-200)x2.5%
0.0°C ≤ t ≤ +400.0°C 0.5 0.5
-200.0°C ≤ t < 0.0°C 0.6+|t|x0.4% 0.6+|t|x0.3%
0.0°C ≤ t ≤ +1300.0°C 0.6 0.6
-200.0°C ≤ t < 0.0°C 0.5+|t|x0.15% 0.5+|t|x0.15%
0.0°C ≤ t ≤ +900.0°C 0.5 0.5
-200.0°C ≤ t < 0.0°C 0.5+|t|x0.2% 0.5+|t|x0.2%
0.0°C ≤ t ≤ +600.0°C 0.5 0.5
-20.0°C ≤ t < 0.0°C 2.0 2.0
+100.0°C ≤ t ≤ +1767.0°C 1.4 1.4
-20.0°C ≤ t < 0.0°C 2.0 2.0
+100.0°C ≤ t ≤ +1768.0°C 1.4 1.4
+600.0°C ≤ t < +800.0°C 1.2 1.5
+1000.0°C ≤ t ≤ +1820.0°C 1.0 1.1
0.0°C ≤ t < +1000.0°C 0.8 0.8
+1000.0°C ≤ t ≤ +2315.0°C 0.8+(t-1000)x0.06% 0.8+(t-1000)x0.06%
-200.0°C ≤ t < 0.0°C 0.4+|t|x0.2% 0.4+|t|x0.2%
+300.0°C ≤ t ≤ +800.0°C 0.5 0.5
0.0°C ≤ t < +1000.0°C 1.0 1.0
+1000.0°C ≤ t ≤ +2500.0°C 1.0+(t-1000)x0.06% 1.0+(t-1000)x0.06%
0.0°C ≤ t < +300.0°C 1.4 1.8
+1500.0°C ≤ t ≤ +2315.0°C 1.8 2.2
+100.0°C ≤ t < +300.0°C 1.4 1.8
+1500.0°C ≤ t ≤ +2315.0°C 1.8 2.2
0.0°C ≤ t < +100.0°C 0.6 1.8
+1000.0°C ≤ t ≤ +1395.0°C 1.0 2.2
Accuracy (1 year)
Source Accuracy [°C] Meas. Accuracy [°C]
Standard or Regulation
IEC60584-1 JIS C1602
IEC60584-1
IEC60584-1
IEC60584-1 -200.0°C ≤ t < 0.0°C 0.5+|t|x0.3% 0.5+|t|x0.3%
IEC60584-1
DIN 43710
DIN 43710
IEC60584-1 0.0°C ≤ t < +100.0°C 2.0 1.4
IEC60584-1 0.0°C ≤ t < +100.0°C 2.0 1.4
IEC60584-1 +800.0°C ≤ t < +1000.0°C 1.0 1.2
IEC60584-1
GOST R 8.585-2001 0.0°C ≤ t < +300.0°C 0.4 0.4
IEC60584-1
ASTM E1751/E1751M–09e1 +300.0°C ≤ t < +1500.0°C 1.2 1.2
ASTM E1751/E1751M–09e1 +300.0°C ≤ t < +1500.0°C 1.2 1.2
ASTM E1751/E1751M–09e1 +100.0°C ≤ t < +1000.0°C 0.8 1.8
DC Voltage Source and Measurement
Range Resolution
90mV 1μV -11.000 to ±99.999mV 0.015% of setting+10μV 0.015% of reading+10μV Max. output current: 10mA
Accuracy is specified at ambient temperature (Ta) of 23 ± 5°C Temperature effect : 0. 005% of Range/°C is added for other ambient temperature (Ta<18°C, Ta > 28°C)
Source Measure
range
Source Measure
n Common source specification
• Output resistance: under 40mΩ
• Output response: under 300msec
• Max. load: C<0.1μF, L<10mH
Accuracy (1 year)
Notes
7
n Basic specification (Source/ Measure) CA330
RTD Source/Measure t: Temperature of Source/Meas.
Accuracy (1 year)
RTD Coefficient
3851
3850
PT100
3916
3926
PT200 3851
PT500 3851
PT1000 3851
Source/Meas. Temp.
-200.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +800.0°C 0.3+tx0.033% 0.3+tx0.033%
-200.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +630.0°C 0.3+tx0.033% 0.3+tx0.033%
-200.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +510.0°C 0.3+tx0.033% 0.3+tx0.033%
-200.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +630.0°C 0.3+tx0.033% 0.3+tx0.033%
-200.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +630.0°C 0.3+tx0.050% 0.3+tx0.050%
-200.0°C ≤ t < 0.0°C 0.4 0.4
0.0°C ≤ t ≤ +630.0°C 0.4+tx0.033% 0.4+tx0.033%
-200.0°C ≤ t < 0.0°C 0.2 0.2
0.0°C ≤ t ≤ +630.0°C 0.2+tx0.033% 0.2+tx0.033%
Source
Accuracy [°C]
Meas.
Accuracy [°C]
Cu10 427 -100.0°C ≤ t ≤ +260.0°C 1.5 1.5 0.1-3mA Minco Application Aid #18
Ni120 627 -80.0°C ≤ t ≤ +260.0°C 0.2 0.2 0.1-3mA Minco Application Aid #18
PT50 3851
PT50G
Extra
PT100G
RTD
Cu50M
Cu100M
Accuracy is specified at ambient temperature (Ta) of 23±5°C. Temperature effect: 0.05°C/°C is added for other ambient temperature (Ta<18°C, Ta > 28°C) The display resolution for source / measure is 0.1°C Above accuracy is specified for 4 wire measuring. Accuracy for 3 wire measuring: 1.0°C to Cu10; 0.6°C to Pt50, Pt50G and Cu50M; 0.3°C to other RTD is each added, on condition the resistance of all cables are the same. Accuracy for 2 wire measuring: Same with 3 wire measuring excluding resistance of cables.
About formula of accuracy
The accuracy of source or measuring are defined by constant value or formula of linear expression. Example) Accuracy of PT100(3851) at measuring point of 100.0°C is ±(0.3+100.0x0.033%)°C=±0.333°C
-200.0°C ≤ t < 0.0°C 0.4 0.4
0.0°C ≤ t ≤ +630.0°C 0.4+tx0.050% 0.4+tx0.050%
-200.0°C ≤ t < 0.0°C 0.4 0.4
0.0°C ≤ t ≤ +800.0°C 0.4+tx0.050% 0.4+tx0.050%
-200.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +630.0°C 0.3+tx0.033% 0.3+tx0.033%
-180.0°C ≤ t < 0.0°C 0.4 0.4
0.0°C ≤ t ≤ +200.0°C 0.4+tx0.050% 0.4+tx0.050%
-180.0°C ≤ t < 0.0°C 0.3 0.3
0.0°C ≤ t ≤ +200.0°C 0.3+tx0.033% 0.3+tx0.033%
Excitation
current
0.1-3mA
0.1-3mA
0.1-3mA
Standard or Regulation
IEC60751 JIS C 1604
JIS C 1604 1989 (Pt100)
JIS C 1604 1989 (JPt100)
0.1-3mA Minco Application Aid #18
0.05-0.8mA IEC60751
0.05-0.6mA IEC60751
0.05-0.6mA IEC60751
0.1-3mA IEC60751
0.1-3mA GOST R 8.625-2006
0.1-3mA GOST R 8.625-2006
0.1-3mA GOST R 8.625-2006
0.1-3mA GOST R 8.625-2006
Resistance source and measure
Range Resolution
Source and Meas.
Range
Source Measurement
500Ω 10mΩ 0.00 to 550.00Ω 0.025% of setting+0.1Ω 0.025% of reading+0.1Ω Excitation current 0.1 to 3mA
3000Ω 100mΩ 0.0 to 3300.0Ω 0.025% of setting+0.5Ω 0.025% of reading+0.5Ω Excitation current 0.05 to 0.6mA
Accuracy is specified at ambient temperature (Ta) of 23±5°C. Temperature effect: Add the accuracy of ±(0.005% of range) /°C for other ambient temperature (Ta<18°C, Ta > 28°C) .. Above accuracy is defined for 4-wire measuring. Accuracy for 3 wire measuring: 0.05Ω to 500Ω range; 0.2Ω to 3000Ω range is added, on condition the resistance of all cables are the same. Accuracy for 2 wire measuring: Same with 3 wire measuring on condition the resistance of cables are excluded.
n Common measurement specification
• Excitation current: Method of voltage surge current meausre (typical 0.78mA at 0Ω, 0.6mA at 500Ω,
0.27mA at 3000Ω
• Disconnection detection: Detects when Hi terminal is open.
n Common source specificaiton
• Response time: Under 5msec (Excluding 3000Ω range, PT500 and PT1000)
• Max. load: C < 0.1μF, L < 10mH
• Sweep: Step (25%) / linear
• Step time: 15sec / 30sec / 45sec/ 60sec
• Allowable resistance for measuring cables: under 10Ω
8
Accuracy (1 year)
Note
Unit: mm
5.8
Unit: mm
192
(0.8
(3.3
(7.9
Unit: mm
Unit: mm
192
Unit: mm
192
Source and Measure Simplicity Series
n General Specification
Display Segment LCD
Backlight LED (Selection of "Constantly ON", "Constantly OFF" or "Auto off by approx. 2min")
Display refresh rate Approx. 1sec.
Warm-up time Approx. 5min.
Power supply Four alkaline AA batteries, Dedicated AC Adapter (Sold separately)
Battery lilfe CA310: 50 hours (5V source, load over 10kOhm), 25 hours (20mA source, load under 5V ) / CA320 : 55 hours / CA330 : 55 hours
Auto Power Off Approx. 20min. (Disabled by setting)
Dimensions Approx. 90 (W)×192 (H)×42 (D)
Weight Approx. 440g
Standard
Operating temperature / humidity ranges -10 to 55°C 20 to 80%RH (without condensation)
Storage temperature / humidity ranges -20 to 60°C 90% RH or less (without condensation)
Accessories
Safety: EN61010-1 / EN61010-2-030
EMC: EN61326-1 Class A Table 2. EN55011 Class A Group1
CA310: Carrying case (B9108NK)/ Lead cables (a set of black and red lead wires for generation and measurement /98064) /four AA alkaline batteries/ Instruction manual
CA320: Carrying case (B9108NK)/ Lead cables (a set of black and red lead wires for generation and measurement / 98040) /Binding post (Red Black 1 piece/ 99045) / four AA alkaline batteries/ Instruction manual
CA330: Carrying case (B9108NK)/Lead cables( a set of 1 black and 3 red lead wires for generation and measurement/98035) /Binding post (Red Black 1 piece/99045)/ Binding post (Red Red 1 piece / 99046) /four AA alkaline batteries/ Instruction manual
n External Dimensions (CA300 series)
CA310
90
42
n External Dimensions (Accessories)
RJ sensor 90080
3.3 (50)5
35
10
50
1500
CA320 CA330
90
TC connector
(2.5) External diameter
42
(4.8)
(2.3)
(7.9)
(3.3)
(16.8)
(20.6)(12.7)
90
42
)
)
)
9
n Model Name and Model Code
Main units (CA300 series)
Name Model Suffix Code Description
Volt mA Calibrator CA310 Voltage and Current Simulate Model
TC Calibrator CA320 Thermocouple Simulate Model
RTD Calibrator CA330 RTD Simulate Model
/TE Add °F setting procedure (for CA320, CA330)
Accessories Sold Separately
*1
Name Model Description
AC Adapter 94013 Input: AC 120V, 50/60Hz
AC Adapter 94016 Input: AC 220V to 240V, 50/60Hz
RJ Sensor
*2
Grabber Clip
Rubber Boots
*3
*4
90080 for CA320: RJ (Reference Junction)
98025 for CA series: separate type (one set of Red and Black 2.0m)
93060 for protection of main unit
Strap 97040 for hanging main unit on wall with rubber boot
Accessory Case B9108XA for accessories
TC Mini Plug Set 1
TC Mini Plug Set 2
*1: These accessories are not included with main unit when purchased *2: RJ sensor is dedicated for CA320. It is unable to be used for CA71 and CA150 *3: It is impossible to be used with binding post (model no. 99045/99046) *4: It is impossible to put in the carrying case with rubber boot (93060) *5: TC mini plugs are dedicated for CA320. Other types of mini plugs are required to be prepared by customer.
*5
*5
90040
90045 K (Yellow) /E (Violet) /J (Black) /T (Blue)
AC Adapter 94013
K (Yellow) /E (Violet) /J (Black) /T (Blue) /R,S (Green) /B,U (White) /G (Red/Green) /D (Red/White) /C (Red) /N (Orange)
AC Adapter 94016
RJ Sensor 90080
Grabber Clip 98025
Accessory Case B9108XA
10
Rubber Boots 93060
TC Mini Plug Set 1 90040
Strap 97040
TC Mini Plug Set 2 90045
Source and Measure Simplicity Series
n Model Name and Model Code
Accessories (included with main unit)
Name Model Description
Lead Cables
Lead Cables
Lead Cables
Binding Post (Red Black)
Binding Post (Red Red)
Carrying Case
*1: These accessories are included with main unit. Included types of accessories are different according to the type of main unit. *2: Included with CA310 when purchased. *3: Included with CA330 when purchased. *4: Included with CA320 when purchased. *5: Included with CA320/CA330 when purchased. *6: Included with CA330 when purchased. *7: It is impossible to put in main unit with rubber boots..
*2
*3
*4
*5
*6
*7
*1
98064 for CA310, Alligator Clip Cable (Red Black 1 set/ 1.7m)
98035 for CA330, Alligator Clip Cable (Red x 3pcs, Black x 1pce 1 set/ 1.7m)
98040 for CA320, Alligator Clip Cable (Red Black 1 set/ 1.7m)
99045 1 short plate attached
99046 1 short plate attached
B9108NK for main unit and lead cables
Lead Cables 98064
Binding Post 99045
Lead Cables 98035
Binding Post 99046
Lead Cables 98040
Carrying Case B9108NK
11
n Related Products
PC-based field device management tool
• Multi-vendor, multi-protocol support BRAIN, FOUNDATIONTM Fieldbus H1, HART
• Automatic device data acquisition upon connection to a device or a segment (Segment Viewer)
• Easy acquisition and diagnosis of device status (Device Viewer)
• Categorization, sorting and filtering (History)
• Multi-parameter set-up (Parameter Manager)
®
, PROFIBUS and ISA100.11a
Versatile Device Management Wizard
New Standard for Field Calibration
• Achieves the highest accuracy in the portable class !
• Basic accuracy: Pressure (measurement) 0.01% rdg
Current/voltage (source/meas.) 0.015% rdg
• Achieves the highest resolution and widest range in portable class
• 0.001kPa (200.000kPa range)
• Calibration procedures of pressure transmitters and pressure switches are embedded.
Pressure Calibrator
CA700
Handheld Terminal for BRAIN Communication
• Online monitoring and communication.
• Easy setup and alteration procedures in English prompting interactive sequences.
• Diagnostics/security with error messages. (Self­check function, Security code protection of setpoints, Low battery voltage alarm and Automatic power-off)
• Printer prints out tag numbers and other parameters on the spot.
BRAIN Terminal
BT200
Highly Accurate All-In-One Calibrator
• Dual display of source and measurement with an accuracy of 0.02%
• Multiple sources and measurements of DCV, DCmA, Ω, TC, RTD, Hz, and PULSE
• 24 V loop power supply and output signal measurement at the same time
• Current sink function to simulate a transmitter
• A variety of sweep functions selectable (Step, Linear, Program, and Sweep)
• Three power supply options available (AA batteries, AC adapter [sold separately], and NiMH batteries [sold separately])
Handy Calibrator
CA150
Safety DMM Equipped with a Loop Power Supply and 4-20 mA Output
• 24 V loop power supply and measures output signal at the same time
• Resistor (250Ω) for HART and BRAIN communication embedded and selectable.
• Transmitter simulation (current sink) function
• Step, Auto-step, and Linear sweep functions are selectable
• 6,000-count DMM function
• EN61010-1 CATIV600V and CATIII1000V safety design
Process Multimeter
CA450
http://www.yokogawa.com/ymi
YOKOGAWA METERS & INSTRUMENTS CORPORATION
Global Sales Dept. /Phone: +81-422-52-6237 Facsimile: +81-422-52-6462 E-mail: tm@cs.jp.yokogawa.com
YOKOGAWA CORPORATION OF AMERICA Phone: +1-770-253-7000 Facsimile: +1-770-254-0928 YOKOGAWA EUROPE B.V. Phone: +31-88-4641000 Facsimile: +31-88-4641111 YOKOGAWA SHANGHAI TRADING CO., LTD. Phone: +86-21-6239-6363 Facsimile: +86-21-6880-4987 YOKOGAWA ELECTRIC KOREA CO., LTD. Phone: +82-2-2628-3810 Facsimile: +82-2-2628-3899 YOKOGAWA ENGINEERING ASIA PTE. LTD. Phone: +65-6241-9933 Facsimile: +65-6241-2606 YOKOGAWA INDIA LTD. Phone: +91-80-4158-6000 Facsimile: +91-80-2852-8656 YOKOGAWA ELECTRIC CIS LTD. Phone: +7-495-737-7868 Facsimile: +7-495-737-7869 YOKOGAWA AMERICA DO SUL LTDA. Phone: +55-11-5681-2400 Facsimile: +55-11-5681-4434 YOKOGAWA AUSTRALIA PTY. LTD. Phone: +61-2-8870-1100 Facsimile: +61-2-8870-1111 YOKOGAWA MIDDLE EAST & AFRICA B.S.C(c) Phone: +973-17-358100 Facsimile: +973-17-336100
Subject to change without notice. [Ed: 01/b] Printed in Japan: Nov. 2015(B)/3,000(KP)
All Rights Reserved. Copyright © 2015, Yokogawa Meters & Instruments Corporation.
Clamp-on Measurement of 4-20 mA Instrumentation Signals
• Process signal measurement with no need to disconnect a loop
• Accuracy 0.2% + 5 dgt and resolution 0.01 mA
• Simultaneous display of percentage (%) of the measured value and span