Yokogawa CA150 Operating Manual

User's Manual
IM CA150E
CA150 HANDY CAL
1st Edition: Octorber 2006
IM CA150E

Introduction

Thank you for purchasing the CA150 HANDY CAL. This User’s manual contains useful information regarding the instrument's functions and operating procedures, as well as precautions that should be observed during use. To ensure proper use of the instrument, read the manual thoroughly before operating it. After you have finished reading this manual, store it in the carrying case for quick reference whenever a question arises.
Notes
• The information contained in this manual is subject to change without notice. Furthermore, the actual display items may differ slightly from the ones appearing in this manual.
• Every effort has been made to ensure the information contained herein is accurate. However, should any concerns, errors, or emissions come to your attention, or if you have any comments, please contact us.
• Copying or reproduction of any or all of the content of this manual without Yokogawa's permission is strictly prohibited.
• The warranty is included in this manual. Be sure to read the warranty to ensure you understand the terms, and then store it in a safe place. (The warranty cannot be reissued.)
Trademark Acknowledgments
Company names and product names mentioned herein may be trademarks or registered trademarks of their respective companies.
Revision Information
1st Edition: Octorver 2006
Disk No. CA150E
1st Edition: Oct, 2006 (KP)
All Rights Reserved. Copyright © 2006, Yokogawa Meters & Instruments Corporation
1

Checking the Contents of the Package

After opening the package, check the following items before use. If the product is not the one you ordered, any item is missing, or there is a visible defect, contact the dealer from whom you purchased the instrument.
Main Unit
Check that the model name given on the name plate on the back panel of the instrument matches the one on your order.
• Model
Model
CA150
• Serial No.
Should you need to contact the dealer from whom you purchased the instrument, have your unit's serial number handy to give to the person.
Standard Accessories
The following standard accessories are supplied with the instrument. Make sure that all items are present and undamaged.
Source lead cables
(98020)
Measurement lead cables
(RD031)
Carrying case
(93026)
Fuse
(A1566EF)
Terminal adapter
(99022)
Ferrite core
(A1193MN) two
2
AA-size alkaline batteries (six) User’s Manual
(IM CA150E)
Checking the Contents of the Package
Optional Accessories
The following optional accessories are available. Upon receiving these optional accessories, make sure that all the items you ordered have been supplied and are undamaged. If you have any questions regarding optional accessories, or if you wish to place an order, contact the dealer from whom you purchased the instrument.
Name Model Remarks
AC adapter 94010-D UL/CSA standard
94010-F VDE standard 94010-R SAA standard 94010-S BS standard
NiMH battery 94015 RJ sensor B9108WA For reference junction compensation Main body case 93027 With strap and accessories case Accessories case B9108XA
94010-H GB standard
Optional Spare Parts
Name Model Remarks
Source lead cables 98020 Lead Cable for Source Measurement lead cables RD031 Safety Test Lead Carrying case 93026 Terminal adapter 99022 Used for temperature measurement Fuse A1566EF Set of 10 fuses
AC adapter Accessories case RJ sensor
3

Safety Precautions

When operating the instrument, strictly observe the precautions in this manual to ensure its correct and safe operation. If used other than as instructed in this manual, Yokogawa Meters & Instruments Corporation is not liable for any damage that may result.
The following safety symbols are used on the instrument and in the manual:
Danger! Handle with Care.
This symbol indicates that the operator must refer to an explanation in the manual in order to avoid risk of injury or loss of life of personnel or damage to the instrument. This symbol indicates a direct current.
This symbol indicates a power source.
Warning
Indicates a hazard that may result in the loss of life or serious injury of the user unless the described instruction is abided by.
Caution
Indicates a hazard that may result in an injury to the user and/or physical dam­age to the product or other equipment unless the described instruction is abided by.
Note
Indicates information that is essential for handling the instrument or should be noted in order to familiarize yourself with the instrument's operating procedures and/or functions.
Tip
Indicates additional information to complement the present topic.
4
Safety Precautions
Be sure to comply with the following safety precautions. Failure to do so may result in loss of life or injury to personnel from such hazards as electrical shock, or damage to the instrument.
Warning
Prohibition of Use in Gaseous Environments
Do not operate the instrument in the presence of inflammable and explosive gases or vapors. Operating the instrument in such an environment is extremely hazardous.
Protection Feature Defects
Do not operate the instrument if a fuse or other protection feature is defective. Before commencing operation, make sure that protection features are free from defects.
External Connections
When connecting the instrument to the object to be tested or an external con­trol circuit, or if you need to touch any external circuit, turn off the power to the circuit and make sure that no voltage is generated.
Fuses
To prevent a fire, be sure to use fuses with the specified ratings (voltage, cur­rent, and type). Do not short-circuit the fuse holder.
Correct Use of Lead Cables
Be sure to correctly use the measurement lead cables (model: RD031) and source lead cables (model: 98020) without mistaking them.
Removing the Casing
Removing the casing and disassembling or modifying the instrument is strictly prohibited. Some parts inside the instrument are extremely dangerous because they use a high voltage. When the instrument needs an internal inspection or calibration, contact Yokogawa Meters & Instruments Corporation or the dealer from whom you purchased the instrument.
To use the AC adapter (optional) safely, be sure to comply with the following precautions.
Warning
Power Supply
Make sure that the rated power supply voltage of the instrument matches the voltage of the power supply before turning on the power.
5

Contents

Introduction .....................................................................................1
Checking the Contents of the Package........................................ 2
Safety Precautions ..........................................................................4
1. Product Outline ........................................................ 1-1
1.1 Product Outline .................................................................. 1-1
2. Names and Functions of Parts ............................... 2-1
3. Before Starting Source or Measurement ............... 3-1
3.1 Usage Precautions ............................................................. 3-1
3.2 Connecting a Power Supply ............................................. 3-3
3.2.1 Using Alkaline Batteries ............................................. 3-3
3.2.2 Using an AC Adapter ................................................. 3-4
3.2.3 Using an NiMH Battery Pack ..................................... 3-5
3.2.4 Fuse ........................................................................... 3-8
3.3 Turning the Power On and Off .......................................... 3-9
3.3.1 Turning the Power On and Off ................................... 3-9
3.3.2 Auto Power Off .......................................................... 3-9
3.3.3 Turning the Backlight On and Off .............................. 3-9
3.4 Operating Environment ................................................... 3-10
4. Source ................................................................... 4-1
4.1 Connecting the Source Terminals .................................... 4-2
4.2 Source DC Voltage (DCV) Signals .................................... 4-3
4.3 Source DC Current (DCA) Signals .................................... 4-4
4.3.1 Generating DC Current Signals ................................. 4-4
4.3.2 4-20 mA Function ...................................................... 4-5
4.3.3 20 mA SINK Function ................................................ 4-6
4.4 Source Resistance () Signals ......................................... 4-8
4.5 Source Thermocouple (TC) Signals ............................... 4-10
4.5.1 Source Thermocouple (TC) Signals ........................ 4-10
4.5.2 Using an External RJ Sensor .................................. 4-11
4.5.3 Using the Built-in RJ Sensor .................................... 4-12
4.6 Source Resistance Temperature Detector (RTD)
Signals .............................................................................. 4-13
Toc-1
Contents
4.7 Source Frequency and Pulse (PULSE) Signals ............ 4-15
4.7.1 Source a Continuous Pulse Train ............................ 4-16
4.7.2 Source a Pulse Cycle .............................................. 4-17
4.8 Divided Output (n/m) Function ....................................... 4-18
4.9 Sweep Output Functions ................................................. 4-19
4.9.1 Step Sweep Function .............................................. 4-20
4.9.2 Linear Sweep Function ............................................ 4-22
4.9.3 Program Sweep Function ........................................ 4-24
4.10 Temperature Monitor Function ....................................... 4-26
5. Measurement ........................................................... 5-1
5.1 Connecting the Measurement Terminals ......................... 5-2
5.2 Measuring DC Voltage (DCV) ............................................ 5-4
5.3 Measuring DC Current (DCA) ............................................ 5-4
5.3.1 Measuring DC Current ............................................... 5-4
5.3.2 Measuring 24V LOOP ............................................... 5-5
5.4 Measuring Thermocouple (TC) ......................................... 5-6
5.5 Measuring Resistance () ................................................. 5-7
5.6 Measuring Resistance Temperature Detector (RTD) ...... 5-8
5.7 Measuring Frequency (FREQ) and Pulse ........................ 5-9
5.7.1 Measuring Frequency (FREQ) and Pulse ................. 5-9
5.7.2 Measuring Contact Input ......................................... 5-10
6. Memory Function ..................................................... 6-1
6.1 Data Memory Items ............................................................ 6-2
6.1.1 Saving ........................................................................ 6-3
6.1.2 Replacing and Saving ................................................ 6-3
6.1.3 Clearing Memory ....................................................... 6-4
6.1.4 Displaying (Confirming) Saved Data ......................... 6-5
6.2 Setting Memory Items ........................................................ 6-6
6.2.1 Saving ........................................................................ 6-6
6.2.2 Replacing and Saving ................................................ 6-6
6.2.3 Clearing Memory ....................................................... 6-7
6.2.4 Loading ...................................................................... 6-8
7. Setting Mode ............................................................ 7-1
7.1 Source ................................................................................. 7-2
7.2 Measure .............................................................................. 7-3
7.3 Configuration ..................................................................... 7-4
Toc-2
Contents
8. Communication Function ....................................... 8-1
8.1 Cable Connection and Interface Specifications .............. 8-1
8.2 Communication Command List ........................................ 8-2
8.3 Detailed Description of Commands ................................. 8-3
8.4 Error Code List ................................................................. 8-12
8.5 Table of Valid Communication Commands ................... 8-13
8.6 Status Byte Format .......................................................... 8-15
8.7 Output Format of Printer Mode ....................................... 8-16
9. Calibration Mode ..................................................... 9-1
9.1 Calibration of Source Functions ...................................... 9-2
9.2 Calibration of Measurement Functions ........................... 9-5
9.3 Verification after Calibration ............................................. 9-8
9.4 Calibration of Temperature Ranges ................................. 9-8
10. Troubleshooting .................................................... 10-1
11. Specifications ........................................................ 11-1
Appendix 1 Using a Cold Junction
Compensator .................................... App.1-1
Appendix 2 Block Diagram.................................. App.2-1
Appendix 3 Installing Ferrite Core...................... App.3-1
Toc-3

1. Product Outline

1.1 Product Outline

Generation (SOURCE)
Function Range
DC Voltage (DCV) 100 mV, 1 V, 10 V, 30 V DC Current (DCA) 20 mA, 20 mA SINK, 4-20 mA Resistance () 500 , 5 k, 50 k Thermocouple (TC) K, E, J, T, N, L, U, R, S, B Resistance temperature detector (RTD) Pt100, JPt100 Frequency and pulse (PULSE) CPM, 100 Hz, 1000 Hz, 10 kHz,
50 kHz
Measurement (MEASURE)
Function Range
DC Voltage (DCV) 35 V, 5 V, 500 mV DC Current (DCA) 100 mA, 20 mA Resistance () 50 k, 5 k, 500 Thermocouple (TC) K, E, J, T, N, L, U, R, S, B Resistance temperature detector (RTD) Pt100, JPt100 Frequency and pulse (FREQ) 100 Hz, 1000 Hz, 10 kHz, CPM, CPH
24V LOOP (DCA)
1
Product Outline
Other
Divided output (n/m) function Sweep output functions
Step sweep function Linear sweep function Program sweep function
Temperature monitor function
Averaging (measurement)
1-1

2. Names and Functions of Parts

Input terminals
Fuse holder (FUSE)
Measure
keys
Output terminals
2
Names and Functions of Parts
LCD Screen
Output value setting keys
Source keys
Keys
Battery holder
RJ sensor connector
RS232
connector
AC adapter
jack
2-1
Common Keys
POWER Turns on/off the power.
LIGHT Turns on/off the backlight of the LCD screen. (It turns off automatically if
approximately 10 minutes elapse without a key being pressed.)
SAVE Saves measurement values and setting values.
LOAD Displays measurement values and loads setting values.
ENTER Confirms the selected item or displays the temperature monitor.
CHARGE Starts/stops charging of the NiMH battery.
Output Value Setting Keys
Output value setting keys
2-2
Sets the source output value. Each (up) and (down) key corresponds to a digit, and increments or decrements the value by one count. (The ▲ and ▼ marks are used indicate these keys in this manual.) If you attempt to increment or decrement the value 9 or 0, the digit moves up or down one place.
key: Sets the memory number for the memory function. n key: Sets the n (numerator side) for divided output (n/m). m key: Sets the m (denominator side) for divided output (n/m).
SOURCE Keys
MEASURE Keys
FUNCTION Changes the source function. DCV DCA
RANGE Changes the range for each function.
SOURCE ON/OFF Turns on/off the source (setting value output).
n/m Selects/cancels divided output (n/m) mode.
+ ←→ - Toggles the polarity of output.
SWEEP SET Selects step sweep, linear sweep, or program
PULSE SET Selects the mode for generating a pulse and
CLEAR • Restores the setting value to its default.
FUNCTION Changes the source function. DCV DCA
RANGE Changes the range for each function.
MEASURE ON/OFF
24V LOOP Selects/cancels the loop test (24 V output).
HOLD • Holds the display value (measurement value).
TC RTD PULSE (DCV)
(The mark is lit for the selected function.)
sweep.
frequency signals. Refer to "4.7 Generating
Frequency and Pulse (PULSE) Signals."
• Goes back one level in setting mode.
• Clears the memory for the memory function.
TC RTD FREQ (DCV)
(The mark is lit for the selected function.)
OFF turns off the measurement value indication
and ON turns on the measurement value
indication.
• Starts/Stops CPM and CPH measurement.
• Starts communication data output. (When
communication of the setting mode is set to
printer mode.)
2
Names and Functions of Parts
2-3
LCD Screen
2
1
4
3
1 Indicates the function selected with the FUNCTION key of MEASURE. 2 Indicates the measurement value (top row: seven segments).
MEASURE and the unit are also displayed. 3 Indicates the function selected with the FUNCTION key of SOURCE. 4 Indicates the source setting value (bottom row: seven segments).
SOURCE and the unit are also displayed.
2-4
SOURCE: OFF lights when output is off or the protection circuit
has been activated. ON lights when output is on.
Indicates hold.
Indicates that the 4-20 mA range is selected for the source.
Flashes while communication data is being output.
(When communication of the setting mode is set to printer mode.)
Indicates that 24 V DC is being output for a loop test.
Indicates reading of the memory function.
Lights when the memory function or program sweep function is in use.
(Indicates the memory number.)
Lights when the sweep function is in use.
Lights when the divided output (n/m) function or step sweep
function is in use.
Indicates that contact input is selected during pulse measurement.
Lights or flashes during offset or full scale adjustment in
calibration mode.
Indicates the battery level of the alkaline batteries or NiMH battery.
Indicates that the NiMH battery pack is charging.
Indicates the type of thermocouple.
Lights during an RJ compensation calculation.
Lights when the PT100 or JPT100 standard is selected for the
RTD function.
Lights when IPTS-68 (temperature scale standard) is selected for
temperature source and measurement (TC, RTD).
2
Names and Functions of Parts
2-5
Digital Display of Alphanumeric Characters
Since the LCD screen of the instrument has seven segments, alphanumeric characters are displayed as shown below. (Some of the characters are not used.)
0
1
2
3
4
5
6
7
8
9
A B C
c D
E
F G
H h
I i
J K
L M
N
O P
Q
R S T U u V W
X Y
Z
2-6

3. Before Starting Source or Measurement

3.1 Usage Precautions

Safety Precautions
Before using the instrument, be sure to thoroughly read "Safety Precautions" on pages 4 and 5.
Do not remove the casing from the instrument. Some parts inside the instrument are extremely dangerous because they use a high voltage. When the instrument needs an internal inspection or calibration, contact Yokogawa Meters & Instruments Corporation or the dealer from whom you purchased the instrument.
In the case of an abnormality If the instrument begins to emit smoke, give off an unusual odor, or show any other signs of an abnormality, immediately turn off the power switch. If you are using an AC adapter, unplug the power cord from the outlet. Also turn off any object under test that is connected to the input terminals.
AC adapter and power cord Use the designated AC adapter. Do not place anything on the AC adapter or power cord, and prevent heat sources from coming into contact with them. When unplugging the power cord from the outlet, be sure to hold the plug and never pull the actual cord. If the power cord is damaged, contact your dealer.
General Handling Precautions
When carrying the instrument Turn off the power to the object under test. Turn off the power to the instrument and unplug the power cord from the outlet if you are using an AC adapter. Then, disconnect all lead cables from the instrument. When carrying the instrument, use the carrying case.
Keep input terminals away from electrically charged articles as they may damage the internal circuitry.
Do not allow volatile chemicals to come into contact with the casing or operation panel. Also, do not allow the instrument to come into contact with any rubber or vinyl products for prolonged periods. Since the operation panel is made of thermoplastic resin, be careful not to let it come into contact with any heat sources such as a soldering iron.
Before cleaning the case and operation panel, make sure that the power cord is unplugged from the outlet if you are using an AC adapter. Dampen a clean soft cloth with water and gently wipe the surface of the casing and panel. Water getting inside the instrument may result in a failure.
If the AC adapter will not be used for a prolonged period, unplug the power cord from the outlet.
For precautions on handling dry batteries, refer to "3.2.1 Using Alkaline Batteries."
Do not use the instrument with the cover for the battery holder left open.
3
Before Starting Generation or Measurement
3-1
Chapter 3.1 Usage Precautions
Operating Environment
Use the instrument in locations that meet the following conditions:
• Ambient temperature and humidity Ambient temperature: 0 to 40C Ambient humidity: 20 to 80% RH (no condensation)
Do not use the instrument in the following locations:
• In direct sunlight or near heat sources
• Where there is a lot of mechanical vibration
• Near noise sources such as high-voltage equipment or power lines
• Near strong magnetic field sources
• Where an excessive amount of greasy fumes, steam, dust, or corrosive gases are present
• In an unstable place
• Where, for example, fire and explosions caused by inflammable gases and the like are possible
Note
• When you require high source and measurement accuracy, use the instru­ment under the following conditions: Ambient temperature: 23 5C Ambient humidity: 20 to 80% RH (no condensation) When using the instrument in an ambient temperature range of 0 to 18C or 28 to 40C, add the temperature coefficient specified in "11. Specifications" to the accuracy.
• When using the instrument in ambient humidity of 30% or less, use an anti­static mat to prevent static electricity.
• Condensation may occur if the instrument is moved from a location of low temperature and humidity to a location of high temperature and humidity, or if the temperature otherwise changes suddenly. In such a case, leave the instru­ment for at least one hour to ensure it is free from condensation before start­ing operation.
3-2
Chapter 3.2 Connecting a Power Supply

3.2 Connecting a Power Supply

In addition to AA-size alkaline batteries (six), the instrument can use two other types of power supply.
• AA-size (LR6) alkaline batteries (six): 1.5 V
• AC adapter (optional)
• NiMH (nickel hydrogen) battery (optional): Model: 94015

3.2.1 Using Alkaline Batteries

Installing and Replacing Batteries
1 Make sure that the power switch of the instrument is turned off and the
lead cables and AC adapter are not connected.
2 Slide up the lock switch on the left side of the instrument and remove the
alkaline battery holder. 3 Insert the six alkaline batteries into the holder. 4 Insert the holder into the opening on the instrument. 5 Slide down the lock switch to fix the holder in place. (The lock switch
indication changes to " FREE.")
Caution
Insert the batteries with their positive and negative electrodes positioned cor­rectly as indicated on the holder.
3
Before Starting Generation or Measurement
Battery Level Indication
Indicates that the battery voltage is normal.
Indicates that the battery level is above 50%. (Lit)
Replace the batteries when this mark begins flashing.
Lock switch
3-3
Chapter 3.2 Connecting a Power Supply

3.2.2 Using an AC Adapter

Warning
• Use the power cord supplied by Yokogawa Meters & Instruments Corporation for use with the instrument.
• Make sure that the power supply voltage matches the ratedsupply voltage before connecting the power cord.
Ratings
AC adapter power supply ratings (Model: 94010) Rated supply voltage 100 to 240 V AC Allowable supply voltage range 90 to 264 V AC Rated supply frequency 50/60 Hz Allowable supply voltage frequency range 47 to 63 Hz Maximum input current 1.4 A Output voltage rating of AC adapter 12.0 V DC Maximum output current rating of AC adapter 3.0 A
3-4
1 Make sure that the power switch of the instrument is turned off. 2 Connect the AC adapter to the AC adapter jack of the instrument. 3 Connect the plug of the power cord to the power connector of the AC
adapter.
4 Connect the other plug of the power cord to an outlet that meets the
ratings described above.
Chapter 3.2 Connecting a Power Supply

3.2.3 Using an NiMH Battery Pack

Charging type NiMH (nickel hydrogen) battery (optional): Model: 94015 Specifications Voltage: 7.2 V
Capacity: 2100 mAh Number of times can be charged (life cycle): Approx. 300 times (varies depending on the operating environment)
Be sure to observe the following warnings on handling the NiMH battery.
Warning
• The electrolyte solution contained in the NiMH battery pack is alkaline. If it comes into contact with any clothing or skin due to a leakage from or rupture in the battery pack, the clothing or skin may be damaged. In particular, if the solution gets into an eye, it may cause loss of eyesight. In such a case, do not rub the affected eye, but thoroughly wash it immediately with clean water. Then see a doctor quickly for treatment.
• When replacing the NiMH battery pack, always turn off the power switch of the instrument and disconnect the AC adapter power cord from the outlet to avoid possible danger such as a short in the electric circuit or electrical shock.
• Do not use any battery pack other than Yokogawa Meters and Instruments Corporation's NiMH battery pack (model: 94015).
• Do not leave the NiMH battery pack in strong direct sunlight, inside a vehicle under the hot sun, or near a fire, otherwise it may result in a solution leakage or deterioration in the performance and/or life.
• Do not disassemble or modify the NiMH battery pack, otherwise the protective features of the battery pack may be damaged, resulting in heating up or rup­ture.
• Do not short the NiMH battery as this may cause burns due to the battery pack heating up.
• Do not dispose of the battery pack in a fire or apply heat to it, otherwise there is a risk that it will rupture or its electrolyte solution will scatter.
• Do not apply excessive shock to the battery pack, for example, by throwing it. Doing so may cause solution leakage, battery pack heating, or rupture.
• Do not use a defective battery pack, such as one leaking solution, deformed, discolored, or showing any other abnormality.
•Avoid any metal coming into contact with the battery pack when carrying it, as there is a danger of a short.
• Do not immerse the battery pack in water or make it wet. Otherwise, it may heat up or rust, as well as lead to a loss of functions.
3
Before Starting Generation or Measurement
If the battery pack will not be used for a prolonged period, remove it from the instrument and store it in the following environment.
Storage period of 1 year or less: Temperature of -20 to 35C (in locations
with low humidity)
Storage period of 3 months or less: Temperature of -20 to 45C (in locations
with low humidity)
3-5
Chapter 3.2 Connecting a Power Supply
Installing the NiMH Battery
1 Make sure that the power switch of the instrument is turned off and the
lead cables and AC adapter are not connected.
2 If alkaline batteries are in use, slide up the lock switch on the left side of the
instrument and remove the alkaline battery holder before installing the NiMH battery.
3 Insert the holder into the opening on the instrument.
Slide the holder into the opening so that the connector is aligned properly.
4 Slide down the lock switch to fix the holder in place. (The lock switch
indication changes to " FREE.")
Lock switch
3-6
NiMH battery
Battery Level Indication
Indicates that the battery voltage is normal.
Indicates that the battery level is above 50%. (Lit)
Charge the battery when this mark begins flashing.
The battery takes approximately 6 hours to fully charge from the flashing state.
Chapter 3.2 Connecting a Power Supply
Charging the NiMH Battery For safety reasons, the NiMH battery is not sufficiently charged at the time of shipment. Fully charge the NiMH battery prior to use. Use the instrument and AC adapter for charging.
Warning
Be sure to use the CA150 to charge the NiMH battery. Charge the NiMH battery in an environment with a temperature within the range of 10 to 35C. Charging the battery at a temperature that is not in the range above may result in an insufficient charge, solution leakage, or heating up.
Charging Procedure
1 Connect the AC adapter to the instrument when the NiMH battery pack is
installed in accordance with the installation procedure.
2 Press the POWER key to turn on the power. Press the CHARGE key to
start charging. (The CHARGE mark appears.)
3 The CHARGE mark disappears when charging is complete.
(To stop charging, press the CHARGE key again.)
Caution
• Performing generation or measurement is possible during charging, but accu­racy is affected by the heat generated by charging. To ensure accurate mea­surement (generation), use of the charging function at the same time is not recommended.
• For details on the effect of heat generation on accuracy (adding of tempera­ture coefficient), refer to the specifications.
• Performing generation and measurement after charging is complete (after at least 2 hours have elapsed) is recommended.
Usage Guidelines
Usage time differs depending on the source function.
3
Before Starting Generation or Measurement
Source output 5 V DC/10 k or more Approx. 10 hours (Measurement: ON, 24V LOOP: OFF)
Tip
Charging NiMH Battery
Depending on the use condition of the main body of the instrument, charging may aborted.
E.g.) When outputting 24 V DC (loop), press the CHARGE key again if the load cur­rent is high. If charging is aborted again, lower the load current and try charging again.
3-7
Chapter 3.2 Connecting a Power Supply
NiMH Battery Life
The battery can be charged approximately 300 times. (This number varies depending on the operating environment.) The life of the battery is over when the low battery level indication appears only a short time after the battery is fully charged. In such a case, replace the NiMH battery pack with a new one.

3.2.4 Fuse

Caution
Be sure to use the designated fuse.
A fuse for current input protection is inserted in the side (fuse holder) of the instrument. Part number: A1566EF, Rating: 125 mA/250 V FAST
Procedure for Replacing the Fuse
When replacing the fuse, turn the Mpart of the fuse holder with a flat-blade screwdriver and remove the holder. Then, replace the fuse, reinsert the fuse holder, and turn the fuse holder with the screwdriver.
3-8
Fuse holder Fuse
(A1566EF)
Chapter 3.3 Turning the Power On and Off

3.3 Turning the Power On and Off

3.3.1 Turning the Power On and Off

When the power is off, press the POWER key to turn the power on. Press the POWER key again to turn the power off.
Note
Turn the power off before disconnecting the AC adapter from the power supply.
Remove the plug of the AC adapter from the instrument when running the in­strument on batteries.

3.3.2 Auto Power Off

When running the instrument on batteries, the LCD screen flashes (alarm) if approximately 9 minutes 30 seconds elapse without a key being pressed. If no operation is performed within approximately 30 seconds after that, the power turns off automatically. (The auto power off function is set to ON at the time of shipment.) To continue using the instrument after the screen begins flashing, press any key other than the POWER key so that the screen stops flashing and lights.
Tip
If the AC adapter is in use or the CPH range is selected for pulse measurement, the instrument is not turned off automatically regardless of the auto power off setting.
3
Before Starting Generation or Measurement
For details on canceling the auto power off function, refer to "7. Setting Mode."

3.3.3 Turning the Backlight On and Off

The backlight of the LCD screen can be turned on. This makes it easy to see the screen when working in dark places.
Press the LIGHT key to turn the backlight on. Press the LIGHT key again to turn the backlight off.
Note
The backlight turns off automatically if approximately 10 minutes elapse with­out a key being pressed. To continue using the backlight, press any key other than the POWER key.
Using the backlight when the instrument is running on batteries reduces the lifespan of the batteries.
3-9
Chapter 3.4 Opereting Environment

3.4 Opereting Environment

Operating Environment
Ambient Temperature and Humidity
Use the CA150 in the following environment:
• Ambient temperature: 0 to 40C
• Ambient humidity: 20 to 80 % RH (no condesation)
Operating Altitude
2000 m max. above sea level.
Location
Indoors
Measurement Category (CAT.)
The measurement category of the CA150 is I.
Warning
Do not use the CA150 for measurements in locations falling under Measure­ment Categories II, III, and IV.
Measurement Category
Measurement Category Description Remarks
I CAT. I For measurement performed
on circuits not directly connected to MAINS.
II CAT. II For measurement performed Appliances, portable
on circuits directly connected equipments, etc. to the low voltage installation.
III CAT. III For measurement performed Distribution board,
IV CAT. IV For measurement performed Overhead wire,
in the building installation. circuit breaker, etc.
at the source of low-voltage cable systems, etc. installation.
3-10
Chapter 3.4 Opereting Environment
Internal Wiring
Entrance Cable
CAT. IV
Distribution Board
Fixed Equipment, etc.
CAT. I
CAT. III
CAT. II
T
Equipment
Outlet
Pollution Degree
Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface resistivity. The pollution degree of the CA150 in the operating environment is 2. Pollution Degree 2 applies to normal indoor atmospheres. Normally, only non-conductive pollution is emitted. However, a temporary electrical conduction may occur depending on the concentration.
Note
• For accurate source and measurement, operate the CA150 in the 23 ± 5C tem­perature range and 55 ± 10% RH.
• Condensation may occur if the CA150 is moved to another place where the ambient temperature and humidity are higher, or if the temperature changes rapidly. If this happens, let the CA150 adjust to the new environment for at least two hours before using it.
3
Before Starting Generation or Measurement
Strage Location
• We recommend you store the CA150 in an environment with a temperature between 0 and 50C and a relative humidity between 5 to 85% RH.
• When storing the CA150, avoid a location that is:
• exposed to direct sunlight,
• 60C or higher,
• 90% RH or higher,
• close to a heat source,
• exposed to servere vibrations,
• exposed to corrosive or explosive gas,
• exposed to excessive amount of soot, dust, salt, and iron, or
• exposed to water, oil, or chemicals.
3-11

4. Source

The instrument can source DC voltage, DC current (current sink), resistance, thermocouple, resistance temperature detector, and frequency/pulse signals.
Warning
To protect against the risk of electrical shock, do not apply a voltage of 30 V or more to the output terminals.
Also ensure that the circuit-to-ground voltage does not exceed 30V.
Be sure to use the supplied lead cables.
Caution
Do not apply any voltage to the output terminals for ranges other than 20 mA SINK. Otherwise the internal circuitry may be damaged.
The voltage drop component due to the resistance (approximately 0.1 on a round-trip basis) of the lead cables becomes an error on the instrument.
4
Source
4-1

4.1 Connecting the Source Terminals

4.1 Connecting the Source Terminals
Connect the supplied source lead cables (model: 98020) to the output terminals of the instrument. Connect the clips to the input terminals of the target device. Be sure to confirm the polarity to ensure the clips are correctly connected to the input terminals.
Target device terminals
Source lead cables (98020)
Red Black
Source output terminals
Source (setting value) indication
4-2
Output value
setting keys
Function selection
Operation on source side

4.2 Source DC Voltage (DCV) Signals

4.2 Source DC Voltage (DCV) Signals
1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to align the source mark
with DCV. (DC voltage selection) 3 Use the RANGE key to select a range. (100 mV, 1 V, 10 V, or 30 V) 4 Use the output value setting keys to set each digit output value.
Each key corresponds to a digit of the setting value. Each time a key
is pressed the value is incremented or decremented by one count. If you
attempt to increment or decrement the value 9 or 0, the digit moves up or
down one place. A value continues to change when you hold down the
key.
Press the CLEAR key to restore the setting to its initial value (zero). 5 Use the +←→- key to select a polarity if the polarity needs to be changed.
(No sign appears for +.) 6 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 7 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
When using the divided output (n/m) function, refer to "4.8 Divided Output (n/m) Function." When using a sweep output function, refer to "4.9 Sweep Output Functions." When using the temperature monitor function, refer to "4.10 Temperature Monitor Function."
Tip
In the following cases, the protection circuit is activated and the output is turned off.
• If settings are modified with the FUNCTION and RANGE keys.
• If output terminals (or lead cables connected to the terminals) are short-circuited.
• The current becomes excessive (an overload current).
4
Source
4-3

4.3 Source DC Current (DCA) Signals

4.3 Source DC Current (DCA) Signals

4.3.1 Source DC Current Signals

1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to align the source mark
with DCA. (DC current selection) 3 Use the RANGE key to select the 20 mA range. (Max. 22.000 mA) 4 Use the output value setting keys to set each digit output value.
Each key corresponds to a digit of the setting value. Each time a key
is pressed the value is incremented or decremented by one count. If you
attempt to increment or decrement the value 9 or 0, the digit moves up or
down one place. A value continues to change when you hold down the
key.
Press the CLEAR key to restore the setting to its initial value (zero). 5 Use the +←→- key to select a polarity if the polarity needs to be changed.
(No sign appears for +.)
For the - polarity, refer to "4.3.3 20 mA SINK Function." 6 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 7 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
When using the divided output (n/m) function, refer to "4.8 Divided Output (n/m) Function." When using a sweep output function, refer to "4.9 Sweep Output Functions." When using the temperature monitor function, refer to "4.10 Temperature Monitor Function."
Tip
In the following cases, the protection circuit is activated and the output is turned off.
• If settings are modified with the FUNCTION and RANGE keys.
• If output terminals (or lead cables connected to the terminals) are open-circuited.
• The voltage becomes excessive.
4-4
4.3 Source DC Current (DCA) Signals

4.3.2 4-20 mA Function

The source current can be increased or decreased in 4 mA steps.
4 mA steps
1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to align the source mark
with DCA. (DC current selection) 3 Use the RANGE key to select the 4-20 mA range. (4-20 mA lights.) 4 Set the output value.
Use the ▲ ▼ output value setting keys for the two leftmost digits to set the
value in 4 mA steps up and down. (4-8-12-16-20 mA)
Use the ▲ ▼ output value setting keys for the three rightmost digits to
increment or decrement the value by one count. (Use these keys to make
fine adjustments, etc.)
Press the CLEAR key to restore the setting to its initial value (4.000). 5 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 6 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
4
Source
Tip
A step-down operation in which the setting value becomes 3 mA or less is not pos­sible.
• When you use the linear sweep function, the generation start (0%) is set to 4 mA. Refer to "4.9.2 Linear Function."
• The same is the case when you use the divided output (n/m) function. Refer to "4-8 Divided Output (n/m) Function."
4-5
4.3 Source DC Current (DCA) Signals

4.3.3 20 mA SINK Function

The SINK function allows you to use DC current (DCA) with the polarity set to ­(minus). This allows drawing (SINK) the specified value of the current from an external voltage source (distributor, etc.) in the direction of the H terminal. Thus, you can use the instrument in a loop test, for example, as a simulator for two­wire transmitters.
Caution
Use the external power supply of 20 mA SINK within the 5 to 28 V range.
Set the polarity of the applied voltage as shown in the figure below and take care not apply a voltage in the opposite direction.
24V DC
4-20 mA
HLmA HL
MEASURE
SOURCE
CA150
1 Use the FUNCTION key on the SOURCE side to align the source mark
with DCA. (DC current selection)
2 Use the RANGE key to select the 20 mA range. (-22.000 mA to 22.000
mA)
3 Use the output value setting keys to set each digit output value.
Each key corresponds to a digit of the setting value. Each time a key is pressed the value is incremented or decremented by one count. If you attempt to increment or decrement the value 9 or 0, the digit moves up or down one place. A value continues to change when you hold down the key. Press the CLEAR key to restore the setting to its initial value (zero).
4 Use the +←→- key to set the polarity to minus. 5 Connect the terminals as shown in the figure above. 6 Turn on the power of the external voltage source (distributor).
Press the SOURCE ON/OFF key to start generation. (SOURCE OFF changes to ON.)
7 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
Tip
The I/O signals of the distributor can be checked by connecting the terminals as indi­cated by the dashed line in the figure above.
Distributor
AC or DC power supply
Output 1-5 V
4-6
4.3 Source DC Current (DCA) Signals
Input Inductance Component
• The current source function of the instrument may result in an unstable output if the instrument is connected to, for example, a positioner or electro­pneumatic converter having a large input inductance component. Make sure the input inductance component of the equipment to be connected is no greater than 100 H.
• If the equipment's input inductance component is unknown, connect the instrument to the equipment as shown in the figure below, and measure the generated current at the MEASURE side. If the reading does not stabilize or an accuracy error results at that point, the input inductance component is likely to be greater than 100 H.
H
L
Equipment such
as a converter
H L mA H L MEASURE SOURCE
4
Source
DCA
CA150
DCA
• If the equipment's input inductance component is too large, connect a 200- resistor R and a 1-F capacitor C to the instrument's outputs, as shown in the figure below. This setup makes it possible to connect an input having an inductance component of up to 3 H to the instrument. Note, however, that the instrument's response time becomes 1 second (at load resistances no greater than 2 k). Do not use this circuitry for purposes other than current generation (DCA), otherwise errors may be produced.
CA150
HL
SOURCE
DCA
99020
R
C
Equipment such
as a converter
H
L
R: 200  10%, 1/4 W, C: 1 F 10%, 50 V
This combination of a resistor (200 ) and capacitor (1 F) is available as an accessory (Model: 99020).
4-7
4.4 Source Resistance () Signals
4.4 Source Resistance () Signals
Procedure for Generating Resistance Signals
• The instrument generates a resistance signal by receiving the resistance­measuring current I supplied from the device under calibration (resistance meter, RTD thermometer, etc.) and delivering the voltage V = R  I proportional to the preset resistance R between the output terminals, and thus producing the equivalent resistance R = V/I. Consequently, the instrument generates the signal correctly only for such devices that employ this method of measurement.
• The allowable range of the resistance measuring current I that the instrument receives from a resistance measuring device under calibration is rated as
0.01 to 5 mA. (This varies depending on the generated resistance value. For details, refer to the specifications.)
Obtaining Accurate Measurements
• Since the generated resistance value is calibrated without including the voltage drop component of the lead cables, the resistance (approximately 0.1 on a round-trip basis) of the lead cables becomes an error in the case of a load current.
• For accurate generation of resistance signals, use a three-wire connection for output. (Refer to the figure below.)
• If the capacitance between the terminals of a device under calibration is greater than 0.01 F, the instrument may be unable to generate correct resistance values.
Three-wire Connection Output Method
Attach another black source lead cable (98020) to the output terminal L side.
4-8
Red Black
CA150
H L
SOURCE
H
L
L'
Device under calibration (Three-wire measuring equipment)
4.4 Source Resistance () Signals
1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to align the source mark
with . (Resistance selection) 3 Use the RANGE key to select a range. (500 , 5 k, or 50 k) 4 Use the output value setting keys to set each digit output value.
Each key corresponds to a digit of the setting value. Each time a key
is pressed the value is incremented or decremented by one count. If you
attempt to increment or decrement the value 9 or 0, the digit moves up or
down one place. A value continues to change when you hold down the
key.
Press the CLEAR key to restore the setting to its initial value (zero). 5 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 6 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
When using the divided output (n/m) function, refer to "4.8 Divided Output (n/m) Function." When using a sweep output function, refer to "4.9 Sweep Output Functions." When using the temperature monitor function, refer to "4.10 Temperature Monitor Function."
Tip
If settings are modified with the FUNCTION and RANGE keys, the protection circuit is activated and the output is turned off.
4
Source
4-9

4.5 Source Thermocouple (TC) Signals

4.5 Source Thermocouple (TC) Signals

4.5.1 Source Thermocouple (TC) Signals

Generate voltages (mV) corresponding to the following thermocouples. Set the temperature with (C). This enables calibration of a thermometer. Thermocouple (TC) types: K, E, J, T, N, L, U, R, S, B (For the temperature range, refer to the specifications.)
1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to align the source mark
with TC. (Thermocouple selection) 3 Use the RANGE key to select a thermocouple type. 4 Use the output value setting keys to set each digit output value. 5 Each ▼ key corresponds to a digit of the setting value. Each time a key
is pressed the value is incremented or decremented by one count. If you
attempt to increment or decrement the value 9 or 0, the digit moves up or
down one place. A value continues to change when you hold down the
key. 6 Press the SOURCE ON/OFF key to start generation. 7 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
Toggling Display
Press the ENTER key to toggle the display as shown below. Setting value (C) source value (mV) temperature monitor (C)
(The display returns to the setting value (C) if no key is pressed for 10 seconds.)
Temperature Scale Standard Setting
The temperature scale standard (ITS-90/IPTS-68) can be selected in setting mode. Refer to "4.6 Generating Resistance Temperature Detector (RTD) Signals" and "7. Setting Mode."
When using the divided output (n/m) function, refer to "4.8 Divided Output (n/m) Function." When using a sweep output function, refer to "4.9 Sweep Output Functions." When using the temperature monitor function, refer to "4.10 Temperature Monitor Function."
Tip
For details on how to use a cold junction compensator, refer to "Appendix 1."
4-10
4.5 Source Thermocouple (TC) Signals

4.5.2 Using an External RJ Sensor

Use an RJ sensor to measure (detect) the temperature of the device under calibration, and compensate the thermoelectric emf. The instrument outputs (generates) the compensated value. When calibrating a device with a built-in reference junction temperature compensator, connect an optional RJ sensor (model: B9108WA) to the instrument (RJ sensor connector). (RJON appears.)
CA150 unit
RJ sensor (B9108WA)
4
Source
Lead cable
H L
SOURCE
CA150
Calibration of thermometer
External RJ sensor
Device under calibration (thermometer)
H
L
Thermocouple
H L
SOURCE
CA150
Calibration including use of thermocouple
Device under calibration (thermometer)
H
L
External RJ sensor
The generation operating procedure is identical to that described in "4.5.1 Generating Thermocouple (TC) Signals."
For details on using the RJ sensor built into the instrument, refer to "4.5.3 Using the Built-in RJ Sensor."
• Compensation of the output voltage using the temperature measured (detected) with the RJ sensor is executed at a sampling rate of approximately 10-second intervals. (This means that there is a delay of up to 10 seconds before the first compensation starts after the external RJ sensor is connected to the connector of the instrument.)
• To perform accurate measurement, leave enough time for the temperature to stabilize after connecting the RJ sensor to the instrument.
4-11
4.5 Source Thermocouple (TC) Signals

4.5.3 Using the Built-in RJ Sensor

Although accurate temperature output (reference junction compensation) requires an external RJ sensor to be used, you can use the RJ sensor built into the instrument. The default setting (at shipment) is OFF. Set the built-in RJ sensor to ON (enable) in setting mode.
ENTER
ENTER
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. SEt appears on the top row and SrC (SOURCE) appears on the bottom row.
2 Select source.
Three types of setting modes are available. Use the ▲ ▼ key to select a mode. Source: SrC, Measure: MEAS, Common: ConF
3 Press the ENTER key to confirm the selection. 4 Select the RJ sensor setting.
Three types of source settings are available. Use the ▲ ▼ key to select a setting. Interval: IntVAL, RJ sensor: rJC, Calibration: CAL
5 Press the ENTER key to confirm the selection. 6 Use the key to select ON. 7 Simultaneously press the CLEAR and ENTER keys to switch back to
normal mode. (Press the CLEAR key to go back one level.)
8 Use the FUNCTION key on the SOURCE side to select TC. 9 Confirm that RJON appears.
The generation operating procedure is identical to that described in "4.5.1 Generating Thermocouple (TC) Signals." For details on setting mode, refer to "7. Setting Mode."
Tip
The built-in RJ sensor measures (detects) the internal temperature of the measure­ment terminals. Accuracy is affected by temperature rises due to, for example, charg­ing. For accurate temperature output, use an external RJ sensor or cold junction com­pensator.
4-12

4.6 Source Resistance Temperature Detector (RTD) Signals

4.6 Source Resistance Temperature Detector (RTD) Signals
Resistance Temperature Detector: Select from PT100 and JPT100. Temperature scale standard: Select from ITS-90 and IPTS-68. (The default setting: ITS-90) The temperature scale standard can be set in setting mode.
1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to align the source mark
with RTD. (Resistance temperature detector selection) 3 Use the RANGE key to select PT100 or JPT100. 4 Use the output value setting keys to set each digit output value.
Each key corresponds to a digit of the setting value. Each time a key
is pressed the value is incremented or decremented by one count. If you
attempt to increment or decrement the value 9 or 0, the digit moves up or
down one place. A value continues to change when you hold down the
key.
Press the CLEAR key to restore the setting to its initial value (zero). 5 Use the +←→- key to select a polarity if the polarity needs to be changed.
(No sign appears for +.) 6 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 7 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
Toggling Display
Press the ENTER key to toggle the display as shown below. Setting value (C) resistance value corresponding to temperature () temperature monitor (C)
(The display returns to the setting value (C) if no key is pressed for 10 seconds.)
When using the divided output (n/m) function, refer to "4.8 Divided Output (n/m) Function." When using a sweep output function, refer to "4.9 Sweep Output Functions." When using the temperature monitor function, refer to "4.10 Temperature Monitor Function."
Tip
For details on the procedures for resistance generation and three-wire connection output method, refer to "4.4 Generating Resistance () Signals."
4
Source
4-13
4.6 Source Resistance Temperature Detector (RTD) Signals
Setting the Temperature Scale Standard
You can select from ITS-90 and IPTS-68 in setting mode. (The default setting: ITS-90) ITS-90: 1990 International Temperature Scale IPTS-68: 1968 International Practical Temperature Scale
ENTER
ENTER
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. SEt appears on the top row and SrC (SOURCE) appears on the bottom row.
2 Select common.
Three types of setting modes are available. Use the ▲ ▼ key to select a mode.
Source: SrC, Measure: MEAS, Common: ConF 3 Press the ENTER key to confirm the selection. 4 Select the temperature scale standard setting.
Four types of common settings are available. Use the ▲ ▼ key to select a
setting.
Auto power off: PWr.oFF, Communication: CoM, Date: dAtE,
Temperature scale standard: t.Std 5 Press the ENTER key to confirm the selection. 6 Use the key to select ITS-90 or IPTS-68. 7 Simultaneously press the CLEAR and ENTER keys to switch back to
normal mode.
(Press the CLEAR key to go back one level.) 8 Use the FUNCTION key on the SOURCE side to select RTD. 9 IPTS68 appears if IPTS-68 is set.
4-14
For details on setting mode, refer to "7. Setting Mode."

4.7 Source Frequency and Pulse (PULSE) Signals

4.7 Source Frequency and Pulse (PULSE) Signals
Frequency
(continuous)
Pulse cycle
Amplitude voltage
(setting value)
SOURCE
ON
Frequency
(setting value)
Amplitude voltage
(setting value)
SOURCE
ON
1
Frequency
(setting value)
2
3
n
End
n = pulse number
(setting value)
PULSE SET Key
When the generation of frequency and pulse signals is selected with FUNCTION of SOURCE, each press of the PULSE SET key toggles the mode as shown below.
Frequency
Amplitude
setting mode
setting mode
4
Source
Pulse number
setting mode
Amplitude Voltage Setting Values
Default value: 0.1000 V Upper value: 11.0000 V Lower value: 0.0000 V Resolution: 0.0001 V
4-15
4.7 Source Frequency and Pulse (PULSE) Signals

4.7.1 Source a Continuous Pulse Train

1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to select frequency and
pulse (PULSE). ( PULSE appears.) 3 Use the RANGE key to select a frequency setting mode.
(Default setting: 100 Hz)
(100 Hz, 1000 Hz, 10 kHz, 50 kHz, 1000 CPM)
Use the ▲ ▼ output value setting keys to set the frequency.
Each time a key is pressed the value is incremented or decremented by
one count.
If you attempt to increment or decrement the value 9 or 0, the digit moves
up or down one place. A value continues to change when you hold down
the key.
Press the CLEAR key to restore the frequency setting of each range to its
default setting. 4 Press the PULSE SET key once to switch to amplitude setting mode.
Use the ▲ ▼ output value setting keys to set the voltage.
Press the CLEAR key to restore the setting to 0.10000 V. 5 Press the PULSE SET key again to switch to pulse number setting mode.
(Default setting: Cont, continuous pulse)
For generating a continuous frequency, no setting is necessary.
For generating a pulse cycle, set the pulse number. 6 Press the PULSE SET key again to return back to frequency setting mode. 7 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 8 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
Tip
When frequency or pulse (PULSE) is selected using FUNCTION, the sweep functions (step, linear, and program) and the divided output (n/m) function cannot be used.
4-16
4.7 Source Frequency and Pulse (PULSE) Signals

4.7.2 Source a Pulse Cycle

1 Connect the terminals. 2 Use the FUNCTION key on the SOURCE side to select frequency and
pulse (PULSE). ( PULSE appears.)
3 Use the RANGE key to select a frequency setting mode.
(Default setting: 100 Hz) (100 Hz, 1000 Hz, 10 kHz, 50 kHz, 1000 CPM) Use the ▲ ▼ output value setting keys to set the frequency. Each time a key is pressed the value is incremented or decremented by one count. If you attempt to increment or decrement the value 9 or 0, the digit moves up or down one place. A value continues to change when you hold down the key. Press the CLEAR key to restore the frequency setting of each range to its default setting.
4 Press the PULSE SET key once to switch to amplitude setting mode.
Use the ▲ ▼ output value setting keys to set the voltage. Press the CLEAR key to restore the setting to 0.10000 V.
5 Press the PULSE SET key again to switch to pulse number setting mode.
(Default setting: Cont, continuous pulse) Use the ▲ ▼ output value setting keys to set the pulse number.
Press the CLEAR key to restore the pulse setting to Cont. 6 Press the PULSE SET key again to return back to frequency setting mode. 7 Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.) 8 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
Tip
When frequency or pulse (PULSE) is selected using FUNCTION, the sweep functions (step, linear, and program) and the divided output (n/m) function cannot be used.
4
Source
4-17

4.8 Divided Output (n/m) Function

4.8 Divided Output (n/m) Function
The divided output (n/m) function outputs a value n/m times the setpoint of a voltage, current, resistance, thermocouple or resistance temperature detector (RTD) signal. Output value = setting value n/m.
Setting value
n/m key
1 Set the source value to output (FUNCTION, RANGE, Setting Value). 2 Press the n/m key to switch to n/m mode. (n/m appears.) 3 Use the ▼ n key of the output setting keys to set the value for the
numerator, and the ▲ ▼ m key of the output setting keys to set value for the denominator. The initial value is 1/1. n is settable from 0 to 19 and m is settable from 1 to
19. (nm)
Press the CLEAR key to restore the setting to the initial value 1/1. 4 Press the SOURCE ON/OFF key to start generation. 5 Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.) 6 Press the n/m key again to cancel the n/m function. (The n/m indication
disappears.)
Tip
• When frequency or pulse (PULSE) is selected using FUNCTION, the divided output (n/m) function cannot be used.
• When the divided output (n/m) function is enabled, a setting mode cannot be se­lected. Disable the divided output (n/m) function by pressing the n/m key and switch to the normal mode.
• The n and m setting values can even be changed during output (SOURCE ON).
• Divide the range between the source setting value and the default value (4 mA):
E.g.: When the source setting value is 3 mA and the default value is 4 mA, the output value is (3 mA - 4 mA) 1/2 + 4 mA = 3.5 mA.
• In the case of a B type TC source, a 600C offset is added to the source value.
4-18

4.9 Sweep Output Functions

4.9 Sweep Output Functions
Three types of sweep output (generation) are available.
Each press of the (SWEEP SET) key toggles the mode as shown below.
Step Sweep Linear Sweep Program Sweep Cancel
Tip
When frequency or pulse (PULSE) is selected using FUNCTION, the sweep functions (step, linear, and program) cannot be used.
4
Source
4-19
4.9 Sweep Output Functions

4.9.1 Step Sweep Function

Set divided output (n/m) to be generated automatically as shown in the figure below.
Setting value
OFF
SOURCE ON
key operation
Source value indication
n
Press
(ON)
33
22
1
m
Actual output
Interval time setting value
2
1
0
1 Set the source value to output (FUNCTION, RANGE, Setting Value). 2 Press the SWEEP SET key to select Step Sweep. (SWEEP and n/m
appear.)
3 Use the m key to set the step number. The default value is 1. m is
settable from 1 to 19.
4 Use the n key to set the generation start step.
The default value is 1. n is settable from 0 to 19. (A value higher than the m value cannot be set.)
5 Press the SOURCE ON/OFF key to start the steps as shown in the figure
above. 6 Press the SOURCE ON/OFF key again to stop the output. 7 Press the SWEEP SET key three times to cancel the sweep output
function.
Linear Sweep Program Sweep Cancel
If you press the SOURCE ON/OFF key again after output is stopped in Step 6 above, output resumes from the stopped step. If you change the n and m values while output is stopped and then press the ON/ OFF key again, output restarts from the first step.
Tip
The default setting for the interval time is 5 seconds. For details on changing the setting, refer to the next page.
4-20
4.9 Sweep Output Functions
Sweep Interval Time
You can change the sweep interval time. The default setting (at shipment) is 5 seconds (FAST). Set 5 seconds (FAST) or 10 seconds (SLOW) in setting mode. When a sweep function is enabled, a setting mode cannot be selected. Disable the sweep function by pressing the SWEEP SET key and switch to the normal mode.
1 In the normal mode, simultaneously press the CLEAR and ENTER keys to
switch to setting mode. SEt appears on the top row and SrC (SOURCE) appears on the bottom row.
2 Select source.
Three types of setting modes are available. Use the ▲ ▼ key to select a mode.
Source: SrC, Measure: MEAS, Common: ConF 3 Press the ENTER key to confirm the selection. 4 Select the interval time setting.
Three types of source settings are available. Use the ▲ ▼ key to select a
setting.
Interval: IntVAL, RJ sensor: rJC, Calibration: CAL 5 Press the ENTER key to confirm the selection. 6 Use the key to select SLOW or FAST. 7 Simultaneously press the CLEAR and ENTER keys to switch back to
normal mode.
(Press the CLEAR key to go back one level.)
ENTER ENTER
4
Source
Tip
The interval setting (FAST, SLOW) of setting mode is common to the following three items. (It cannot be set individually for each item.)
Step sweep, linear sweep, and program sweep
For details on setting mode, refer to "7. Setting Mode."
4-21
4.9 Sweep Output Functions

4.9.2 Linear Sweep Function

Output can be varied in a continuous manner as shown in the figure below.
Setting value
Source value indication
OFF
SOURCE ON/OFF
key operation
OFF
SOURCE ON/OFF
key operation
SOURCE ON indication
Flashing
Press (OFF)
Actual output
Press
(ON)
Approx. 2 sec Approx. 3 sec
0
Press
(ON)
Press
(ON)
SOURCE OFF indication
Flashing
Press
(OFF)
When the SOURCE ON key is pressed while output
(
is being varied)
OFF
(
1 Set the source value to output (FUNCTION, RANGE, Setting Value). 2 Press the SWEEP SET key twice to select Linear Sweep.
Step Sweep Linear Sweep
(SWEEP appears.) 3 Use the output setting key to set the setting value (upper value). 4 Press the SOURCE ON/OFF key to start output automatically as shown in
the figure above. 5 Press the SOURCE ON/OFF key again to stop the output. 6 Press the SWEEP SET key twice to cancel the sweep output function.
Program Sweep Cancel
Tip
• The default setting for the linear sweep time is 16 seconds. For details on changing the setting, refer to the next page.
• When output reaches the lower limit, the state is retained for 3 seconds, and then output is turned off and the first sweep operation ends.
•To change the direction for varying output, press the SOURCE ON/OFF key during the sweep operation. The direction for varying output changes and the sweep op­eration continues. (For example, pressing the SOURCE ON/OFF key during rising output results in the output switching to a falling direction.)
4-22
4.9 Sweep Output Functions
Linear Sweep Time
You can change the linear sweep time. The default setting (at shipment) is 16 seconds (FAST). Set 16 seconds (FAST) or 32 seconds (SLOW) in setting mode. When a sweep function is enabled, a setting mode cannot be selected. Disable the sweep function by pressing the SWEEP SET key and switch to the normal mode.
1 In the normal mode, simultaneously press the CLEAR and ENTER keys to
switch to setting mode. SEt appears on the top row and SrC (SOURCE) appears on the bottom row.
2 Select source.
Three types of setting modes are available. Use the ▲ ▼ key to select a mode.
Source: SrC, Measure: MEAS, Common: ConF 3 Press the ENTER key to confirm the selection. 4 Select the sweep time (interval) setting.
Three types of source settings are available. Use the ▲ ▼ key to select a
setting.
Interval: IntVAL, RJ sensor: rJC, Calibration: CAL 5 Press the ENTER key to confirm the selection. 6 Use the key to select SLOW or FAST. 7 Simultaneously press the CLEAR and ENTER keys to switch back to
normal mode.
(Press the CLEAR key to go back one level.)
ENTER ENTER
4
Source
Tip
The interval setting (FAST, SLOW) of setting mode is common to the following three items. (It cannot be set individually.)
Step sweep, linear sweep, and program sweep
For details on setting mode, refer to "7. Setting Mode."
4-23
4.9 Sweep Output Functions

4.9.3 Program Sweep Function

Multiple setting values saved with the SAVE key can be output in order as shown in the figure below.
Memory Number Function Range
0
1
2
3
4
5
99
SOURCE
ON
5V
0V
Memory No.
When the start number was set to 0
01234321012•
Unused
DCV
DCV
DCV
DCV
DCV
10 V
10 V
10 V
10 V
10 V
Unused
Source Value
(Setting Value)
5.0000V
1.0000V
4.0000V
2.0000V
3.0000V
Unused
Tip
• This is only valid for the saved data memory information of SOURCE. (The saved information of measurement conditions is ignored.)
• If there is an unused number during the step operation or a number saved with different FUNCTION and RANGE conditions, the step operation returns to the start­ing number and repeats the step operation.
• The frequency or pulse source values saved in the data memory cannot be used for program sweep.
4-24
You can change the sweep interval time. The default setting (at shipment) is 5 seconds (FAST). Set 5 seconds (FAST) or 10 seconds (SLOW) in setting mode. The interval setting (FAST, SLOW) is also common to step sweep and linear sweep. When a sweep function is enabled, a setting mode cannot be selected. Disable the sweep function by pressing the SWEEP SET key and switch to the normal mode.
4.9 Sweep Output Functions
Saving Source Values
1 Set the source value to output (FUNCTION, RANGE, Setting Value). 2 Press the SAVE key. (MEM No. appears.)
3 Use the key to select the memory number. 4 Press the ENTER key to confirm the selection. (MEM No. disappears.)
Source value
Program Sweep
1 Press the SWEEP SET key three times to select Program Sweep.
Step Sweep Linear Sweep Program Sweep (SWEEP and MEM NO. appear.)
2 Use the key to set the memory number (Start).
(The source value that corresponds to a memory number is displayed)
3 Press the SOURCE ON/OFF key to start output automatically.
(Output moves back and forth from the start to the last memory number.) 4 Press the SOURCE ON/OFF key again to stop the output. 5 Press the SWEEP SET to cancel the sweep output function.
Cancel
4
Source
If you press the SOURCE ON/OFF key again after output is stopped in Step 4 above, output resumes from the stopped step. If you change the starting number while output is stopped and the press the ON/ OFF key again, output restarts from the first step.
4-25

4.10 Temperature Monitor Function

4.10 Temperature Monitor Function
The ambient temperature (temperature measured with the built-in RJ sensor) can be displayed when you are using a source function. A reading higher than the room temperature may be displayed because of a temperature rise within the instrument.
When the Voltage (DCV), Current (DCA), Resistance (), or Frequency/Pulse (PULSE) source is selected
Press the ENTER key to display the ambient temperature (C). (C flashes.)
When the Thermocouple (TC) or Resistance Temperature Detector (RTD) source is selected
1 The source value [setting value] (C) is displayed. 2 Press the ENTER key once.
Thermocouple (TC): The thermoelectric power (mV) corresponding to the temperature is displayed. Resistance Temperature Detector (RTD): The resistance value () corresponding to the temperature is displayed. (The offset by the RJ sensor is not included.)
3 Press the ENTER key once again.
The ambient temperature (C) is displayed.
4 Press the ENTER key once again.
The source value [setting value] (C) is displayed.
Tip
• If an external RJ sensor is connected, the temperature measured with the external RJ sensor is displayed.
• The source value is redisplayed automatically after approximately 10 seconds in the case of thermocouple and resistance temperature detector, and approximately 2 seconds in the case of other sources.
4-26

5. Measurement

Warning
•Turn off the power supply to the object to be tested before connecting it to the instrument. It is extremely dangerous to connect and disconnect measure­ment lead cables while power is being supplied to the object.
• It is extremely dangerous to incorrectly connect the voltage input terminal H and the current input terminal mA. Make sure that the measurement function (FUNCTION) selection and terminal connections are correct. An incorrect con­nection may not only cause damage to the circuit or device under test and the instrument, but also result in injury to the operator.
• The maximum allowable applied voltage for the grounding of all input/output terminals is 42 Vpeak. Be sure not to exceed this voltage because doing so may not only cause damage to the instrument, but also result in injury to the operator.
Caution
About the Fuse
There is a built-in current input protection fuse for the current input terminal. The fuse blows when an excessive current flows. If the fuse does blow, be sure to replace it with the designated fuse (model: A1566EF). For details on the pro­cedure for replacing the fuse, refer to "3.2.4 Fuse."
5
Measurement
About Display
HOLD key: Enables you to hold the display value (measurement value). MEASURE ON/OFF key: Switches display on and off. (Shows/hides) Display update interval: Approximately 1 second Overrange: "------" is displayed if the input value is overrange.
5-1

5.1 Connecting the Measurement Terminals

5.1 Connecting the Measurement Terminals
This shows the lead cable connections for when DCV, &ohm&, TC, RTD, or
Measurement lead cables (RD031)
Measurement input
terminals
Measurement
indication
Function
selection
Red
Black
FREQ is selected with FUNCTION. For the connections when DCA or 24V LOOP (DCA) is selected, refer to the next page.
Operation on
measurement side
5-2
DC Current (DCA)
Measurement lead cables (RD031)
5.1 Connecting the Measurement Terminals
Black
Red
Measurement input terminals
24V LOOP (DCA)
Measurement lead cables (RD031)
Measurement input terminals
3W (, RTD)
Terminal adapter (99022)
5
Measurement
Red
Black
Measurement input terminals
5-3

5.2 Measuring DC Voltage (DCV)

5.2 Measuring DC Voltage (DCV)
Red
Measurement lead cables (RD031)
Measurement input terminals
Black
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with DCV. (DC voltage selection)
3 Use the RANGE key to select a range. (35 V, 5 V, or 500 mV)
Set the range in accordance with the object to be measured.
4 The measurement value and unit (mV, V) appear.

5.3 Measuring DC Current (DCA)

5.3.1 Measuring DC Current

5-4
Black
Measurement lead cables (RD031)
Measurement input terminals
Red
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with DCA. (DC current selection)
3 Use the RANGE key to select a range. (100 mA, 20 mA)
Set the range in accordance with the object to be measured.
4 The measurement value and unit (mA) appear.
5.3 Measuring DC Current (DCA)

5.3.2 Measuring 24V LOOP

This function is for measuring the current flowing when a 24 V DC constant voltage is applied. With 24V LOOP measurement, you can perform a transmitter loop test.
Measurement lead cables (RD031)
Measurement input terminals
Red
Black
Two-wire transmitter
4-20 mA
24 V DC
HLmA
24V LOOP
MEASURE
Converter
Distributor
CA150
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with DCA. (DC current selection)
3 Press the 24V LOOP key.
A constant voltage (24 V DC) is output between the H and mA
measurement terminals. 4 The measurement value and unit (mA) appear. 5 Press the 24V LOOP key again to cancel measurement.
Tip
A constant voltage (24 V DC) is not output if the measurement function is set to other than DCA.
5
Measurement
5-5

5.4 Measuring Thermocouple (TC)

5.4 Measuring Thermocouple (TC)
Thermocouple
+ leg - leg
Terminal adapter (99022)
Measurement input terminals
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with TC. (Thermocouple selection)
3 Use the RANGE key to select a thermocouple type.
(K, E, J, T, N, L, U, R, S, B)
4 The measurement value and unit (C) appear.
Tip
• The measurement value (display value) when the built-in RJ sensor or an external RJ sensor is used (set) is the value after temperature compensation has been per­formed.
• When there is an overrange or measurement is not possible, "------" appears.
• If the operating environment changes suddenly, wait for RJ compensation to stabi­lize before beginning measurement.
• It may take a while for the temperature of the terminal parts to stabilize after a termi­nal adapter is attached or a thermocouple is connected.
5-6
About Temperature Scale Standards
You can select from ITS-90 and IPTS-68 in setting mode. (IPTS-68 appears if IPTS-68 is set.) For details, refer to "4.6 Generating Resistance Temperature Detector (RTD) Signals" and "7. Setting Mode."
5.5 Measuring Resistance ()
Two-wire method Three-wire method
Red
Measurement lead cables (RD031)
Black
5.5 Measuring Resistance ()
Terminal adapter (99022)
Measurement input terminals
Measurement input terminals
Measurement input terminals
Terminal adapter (99022)
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with V. (Resistance selection)
3 Use the RANGE key to select a range. (50 k, 5 k, 500 )
Set the range in accordance with the object to be measured.
4 The measurement value and unit () appear.
Tip
For accurate measurement in the 500 range, use the three-wire method.
5
Measurement
5-7

5.6 Measuring Resistance Temperature Detector (RTD)

5.6 Measuring Resistance Temperature Detector
(RTD)
Two-wire method Three-wire method
Measurement lead cables (RD031)
Red
Black
Terminal adapter (99022)
Measurement input terminals
Measurement input terminals
Measurement input terminals
RTD
Terminal adapter (99022)
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with RTD. (Resistance temperature detector selection)
3 Use the RANGE key to select a range. (PT100, JPT100)
You can select the temperature scale standard (ITS-90, IPTS-68).
4 The measurement value and unit (C) appear.
Tip
• For accurate measurement, use the three-wire method.
• About Temperature Scale Standards
You can select from ITS-90 and IPTS-68 in setting mode.
(IPTS-68 appears if IPTS-68 is set.)
For details, refer to "4.6 Generating Resistance Temperature Detector (RTD) Sig­nals" and "7. Setting Mode."
5-8

5.7 Measuring Frequency (FREQ) and Pulse

5.7 Measuring Frequency (FREQ) and Pulse

5.7.1 Measuring Frequency (FREQ) and Pulse

Red
Measurement lead cables (RD031)
Measurement input terminals
1 Connect the terminals as shown in the figure above. 2 Use the FUNCTION key on the MEASURE side to align the
measurement mark with FREQ. (Frequency and pulse selection)
3 Use the RANGE key to select a range. Set the range in accordance with
the object to be measured. <For frequency measurement> Set the range to 100 Hz, 1000 Hz, or 10 kHz. The measurement value and unit (Hz or kHz) appear. <For pulse measurement> Set the range to 100000 CPM or 100000 CPH. After the measurement value is reset to 0, the unit (CPM or CPH) and HOLD are displayed. The instrument is in a standby state for measurement.
4 Press the HOLD key to start pulse count (measurement).
In the CPM range, pulses are counted (measured) for 1 minute. In the CPH range, pulses are counted (measured) for 1 hour. (HOLD lights up.)
Black
Tip
• In the case of the CPM and CPH ranges, pressing the HOLD key while HOLD is lit after the count ends will restart the count from 0.
• In the case of the CPM and CPH ranges, pressing the HOLD key partway through before the selected time (CPM: 1 minute, CPH: 1 hour) is reached will end the count at that point in time. (The count value until the end is displayed.)
• In the case of the CPM and CPH ranges, "------" is displayed and measurement ends if the count number exceeds the measurement range (limit).
• Auto power off does not work while the 100000 CPH range is selected.
5
Measurement
5-9
5.7 Measuring Frequency (FREQ) and Pulse

5.7.2 Measuring Contact Input

The instrument can measure transistor contact on/off signals. Set contact input to ON in setting mode. The default setting (at shipment) is OFF.
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. SEt appears on the top row and SrC (SOURCE) appears on the bottom row.
2 Select MEAS.
Three types of setting modes are available. Use the ▲ ▼ key to select a mode.
Source: SrC, Measure: MEAS, Common: ConF 3 Press the ENTER key to confirm the selection. 4 Select the contact input setting.
Three types of measure settings are available. Use the ▲ ▼ key to select
a setting.
Averaging: AVE, Contact input: Cont, Measurement calibration: CAL 5 Press the ENTER key to confirm the selection. 6 Use the key to select ON. 7 Simultaneously press the CLEAR and ENTER keys to switch back to
normal mode.
(Press the CLEAR key to go back one level.) 8 Use the FUNCTION key on the MEASURE side to align the
measurement mark with FREQ. (Frequency and pulse selection). The
mark appears. 9 The measurement value and unit (Hz, kHz, CPM, CPH) appear.
5-10
ENTER
ENTER
For details on setting mode, refer to "7. Setting Mode."

6. Memory Function

Two types of information are saved to memory: data memory items and setting memory items. (Data memory items and setting memory items are saved to separate memory areas.)
Data Memory Items
Information such as records of source values and measurement values and program sweep output data can be stored.
Press the SAVE key to save data and the LOAD key to display
(confirm) data.
Number of storable items: 100 (No. 0 to No. 99)
Stored items: Save date and time
Source FUNCTION, RANGE, and source values
When PULSE is selected with FUNCTION: Frequency, amplitude, and pulse number
Measurement MEASURE ON/OFF, FUNCTION, RANGE,
and measurement values
Setting mode Contact input and temperature scale
standard
Setting Memory Items
Conditions that have been previously saved in setting mode can be reproduced. Press the SAVE + ENTER keys to save data and the LOAD + ENTER key to load data.
Number of storable items: 21 (No. 0 to No. 20) No. 0 becomes the initial setting (default value) at power on.
Stored items: Save date and time
Source FUNCTION, RANGE, and source values
When PULSE is selected with FUNCTION: Frequency, amplitude, and pulse number
Measurement MEASURE ON/OFF, FUNCTION, RANGE,
and measurement values
Setting mode Source: Interval and built-in RJ sensor
Measurement: Averaging and contact input Common: Auto power off, communication, and temperature scale standard
Tip
The following items cannot be saved:
SOURCE ON/OFF state, 24V LOOP ON/OFF state, divided output (n/m) state (n and m setting values), sweep output state (setting), backlight on/off state, NiMH battery charge state
6
Memory Function
6-1

6.1 Data Memory Items

6.1 Data Memory Items
Memory number setting
Memory operations
6-2
6.1 Data Memory Items

6.1.1 Saving

1 Press the SAVE key when a source value
or measurement value is displayed. MEM No. (memory number) appears. The next number after the largest used (saved) number appears.
2 Use the key to change the
memory number.
3 Press the ENTER key. The data is saved and save mode is canceled.
(To not save the data, press the SAVE key to cancel save mode.)
Tip
To confirm saved data, press the LOAD key. After pressing the LOAD key, use the
key to select the memory number to confirm. The save date and time appear and
then the data appears.

6.1.2 Replacing and Saving

Follow the procedure below to replace saved data (a memory number).
1 Press the SAVE key. (Save mode)
MEM No. (memory number) appears.
2 Use the key to select a memory
number.
3 Press the ENTER key. If data exists for
the memory number, a replace confirmation indication appears.
4 Press the ENTER key again to replace the data.
The instrument returns to normal (source/measurement) mode. (To not replace the data, press the SAVE key to cancel the operation.)
Replace confirmation indication
6
Memory Function
6-3
6.1 Data Memory Items

6.1.3 Clearing Memory

Follow the procedures below to clear (delete) saved data.
Clearing Selected Memory Number
1 Press the SAVE key. (Save mode)
MEM No. (memory number) appears.
2 Use the key to select a memory
number. 3 Press the CLEAR key. 4 The clear confirmation indication appears. 5 Press the ENTER key to clear the data.
(To not clear the data, press the SAVE key to cancel the operation.)
Clearing All Memory Numbers
1 Press the SAVE key. (Save mode)
MEM No. (memory number) appears.
2 Use the key to select a memory
number. 3 Press the CLEAR key. 4 Press the CLEAR key again. 5 The clear all confirmation indication appears. 6 Press the ENTER key to clear all the memory numbers.
(To not clear the data, press the SAVE key to cancel the operation.)
Tip
To return to normal (source/measurement) mode from save mode, press the SAVE key again while MEM No. (memory number) is displayed.
Clear confirmation indication
Clear confirmation indication
6-4

6.1.4 Displaying (Confirming) Saved Data

Follow the procedures below to display (confirm) saved data. 1 Press the LOAD key. (Confirmation
mode) MEM No. (memory number) and LOAD appear.
2 Use the key to select a memory
number. 3 The save date and time appear. 4 The saved data appears.
6.1 Data Memory Items
Follow the procedure below to clear saved data. 1 Press the LOAD key. (Confirmation
mode)
MEM No. (memory number) and LOAD
appear.
2 Use the key to select a memory
number. 3 The save date and time appear. 4 The saved data appears. 5 Press the CLEAR key. 6 The clear confirmation indication appears. 7 Press the ENTER key to clear the data.
The indication changes to "-----."
(To not clear the data, press the LOAD key to cancel the operation.)
For details, refer to "6.1.3 Clearing."
Tip
To return to normal (source/measurement) mode from load mode, press the LOAD key again while MEM No. (memory number) and LOAD are displayed.
6
Memory Function
6-5

6.2 Setting Memory Items

6.2 Setting Memory Items
In addition to normal mode (source/measurement) FUNCTION, RANGE, and source values (measurement values), the conditions of setting mode can also be saved. This enables you to load saved conditions and reproduce them (reflect the state) in source and measurement modes. The data saved to MEM No. 0 of setting memory becomes the initial setting (default value) at power on.
You can save data in MEM No. 0 to No. 20

6.2.1 Saving

1 Simultaneously press the SAVE and
ENTER keys while a source or measurement value is displayed.
2 SAVE/SETUP appears and then the
MEM No. (memory number) appears.
3 Use the key to change the
memory number.
4 Press the ENTER key. The data is saved
and setting memory mode is canceled.
Saving to Memory Number 0
While MEM No. 1 is displayed in Step 3 above, press the (down) part of the
key for at least 1 second. The number changes to MEM No. 0 and data can
be saved as initial settings at power on.
Tip
To confirm saved data, simultaneously press the LOAD and ENTER keys and then
use the ▲ ▼ key to select the memory number to confirm. The saved data appears.

6.2.2 Replacing and Saving

Follow the procedure below to replace saved data (a memory number).
1 Simultaneously press the SAVE and
ENTER keys. (Setting memory mode) MEM No. (memory number) appears.
2 Use the key to select a memory
number.
3 Press the ENTER key. If data exists for
the memory number, a replace confirmation indication appears.
4 Press the ENTER key again to replace the data.
The instrument returns to normal (source/measurement) mode. (To not replace the data, press the SAVE key to cancel the operation.)
6-6
Replace confirmation indication
6.2 Setting Memory Items

6.2.3 Clearing Memory

Follow the procedure below to clear (delete) saved data.
Clearing Selected Memory Number
1 Simultaneously press the SAVE and
ENTER keys. (Setting memory mode)
MEM No. (memory number) appears.
2 Use the key to select a memory
number. 3 Press the CLEAR key. 4 The clear confirmation indication appears. 5 Press the ENTER key to clear the data.
(To not replace the data, press the SAVE key to cancel the operation.)
Clearing All Memory Numbers
1 Simultaneously press the SAVE and
ENTER keys. (Setting memory mode)
MEM No. (memory number) appears.
2 Use the key to select a memory
number. 3 Press the CLEAR key. 4 Press the CLEAR key again. 5 The clear all confirmation indication appears. 6 Press the ENTER key to clear MEM No. 1 to No. 20.
(No. 0 is not cleared.)
(To not clear the data, press the SAVE or CLEAR key to cancel the
operation.)
Tip
• If the data of No. 0 is cleared, the data is replaced with the setting conditions (initial settings) at the time of shipment.
•To return to normal (source/measurement) mode from setting memory mode, press the SAVE key again while MEM No. (memory number) is displayed.
Clear confirmation indication
Clear confirmation indication
6
Memory Function
6-7
6.2 Setting Memory Items

6.2.4 Loading

Loading
1 Simultaneously press the LOAD and
ENTER keys. (Load mode) MEM No. (memory number) and LOAD appear.
2 LOAD/SETUP appears.
3 Use the key to select a memory
number. If no data exists (empty) for the memory
number, "-----" appears. 4 The saved data (settings) appears. 5 Press the ENTER key. 6 The setting conditions are reproduced
(state is reflected) and the instrument
returns to normal (source/measurement)
mode.
Clearing Saved Data
1 Simultaneously press the LOAD and ENTER keys. (Load mode)
MEM No. (memory number) and LOAD appear. 2 SETUP appears.
3 Use the key to select a memory number. 4 The saved data (settings) appears. 5 Press the CLEAR key. 6 The clear confirmation indication appears. 7 Press the ENTER key to clear the data.
The indication changes to "-----."
(To not clear the data, press the LOAD key to cancel the operation.)
For details, refer to "6.1.3 Clearing."
Tip
To return to normal (source/measurement) mode from load mode, press the LOAD key again while MEM No. (memory number) and LOAD are displayed.
6-8

7. Setting Mode

Level 1
Level 2
Level 3
ENTER
CLEARCLEAR
ENTER
CLEAR/ENTER
CLEAR + ENTER
Setting modeNormal mode
* You can press the CLEAR + ENTER keys in any level (display) in setting mode to return to normal mode.
In the normal mode, simultaneously press the CLEAR and ENTER keys to switch to setting mode. The following table shows the items that can be set in setting mode. In the normal mode, simultaneously press the CLEAR and ENTER keys and switch to setting mode. (When the divided output (n/m) or a sweep function is enabled, a setting mode cannot be selected.)
Level 1
Source
Measure
Conf
Level 2: Detailed Items
Interval Time
Built-in RJ Sensor
Calibration (Source)
Averaging
Contact Input
Calibration (Measure)
Auto Power Off
Communication
Date/Time
Temperature Scale
Level 3
SLOW/FAST
ON/OFF
ON/OFF
ON/OFF
AUTO/MAN
Normal/Print
In the case of Print: Transfer interval
Year/Month/Day
Hour/Minute/Second
ITS-90/IPTS-68
Basic Operating Procedure
1 Simultaneously press the CLEAR and ENTER keys. (Setting mode) 2 Use the ▼ keys to select an item. (Level 1: Source, Measure, or
Configuration) 3 Press the ENTER key. 4 Use the keys to select a detailed item. (Level 2) 5 Press the ENTER key. 6 Use the keys to select a setting for the item. (Level 3)
7
Setting Mode
7-1

7.1 Source

7.1 Source
(1) Interval Time
This sets the interval time for step sweep and program sweep, as well as the sweep time for linear sweep. (These items cannot be set individually.)
Selection
FAST
SLOW
Interval Time
5 seconds
10 seconds
Default value: FAST
Linear Sweep Time
16 seconds
32 seconds
For details, refer to "4.9.1 Step Sweep Function," "4.9.2 Linear Sweep Function," and "4.9.3 Program Sweep Function."
(2) Built-in RJ Sensor
This turns on or off the RJ sensor built into the instrument.
For details, refer to "4.5.3 Using the Built-in RJ Sensor."
(3) Calibration (Source)
This is a calibration mode for adjusting (calibrating) source.
For details, refer to "9. Calibration Mode."
7-2

7.2 Measure

ENTER
ENTER
(1) Averaging
This enables (ON) or disables (OFF) the moving averaging process for measurement data. (default value: ON) If the reading (measurement value) fluctuates as a result of, for example, noise being included in the input signal, use the averaging function.
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. 2 Use the keys to select Measure. 3 Press the ENTER key to confirm the selection. 4 Use the keys to select the averaging setting. 5 Press the ENTER key to confirm the selection. 6 Use the keys to select ON or OFF.
7.2 Measure
7
Setting Mode
To return to normal mode from setting mode, simultaneously press the CLEAR and ENTER keys. (Press the CLEAR key to go back one level.)
(2) Contact Input
The instrument can measure transistor contact on/off signals during frequency measurement (FREQ). The initial value is OFF. When ON is set and frequency (FREQ) is set with FUNCTION, the mark appears.
For details, refer to "5.7.2 Measuring Contact Input."
(3) Calibration (Measure)
This is a calibration mode for adjusting (calibrating) measurement.
For details, refer to "9. Calibration Mode."
7-3
7.3 Common

7.3 Configuration

(1) Auto Power Off
When auto power off is set, the whole LCD screen flashes if approximately 9 minutes 30 seconds elapse without a key being pressed. If no operation is performed within approximately 30 seconds after that, the power turns off automatically. (This function is for when the instrument is running on batteries.) (default value: AUTO = the power turns off automatically)
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. 2 Use the keys to select configuration. 3 Press the ENTER key to confirm the selection. 4 Use the keys to select the auto power off setting. 5 Press the ENTER key to confirm the selection. 6 Use the keys to select AUTO or MANUAL.
ENTER
ENTER
To return to normal mode from setting mode, simultaneously press the CLEAR and ENTER keys. (Press the CLEAR key to go back one level.)
Tip
If the AC adapter is in use or the CPH range is selected for pulse measurement, the instrument is not turned off automatically regardless of the auto power off setting.
7-4
7.3 Common
(2) Communication
You can select normal mode and printer mode for communication. (default value: Normal mode) Normal mode: Enables normal sending and receiving. Printer mode: Enables a source value and measurement value to be
output to a printer at a specified interval* (0 to 3600 seconds).
*:If the transfer interval of printer mode is set to 0 seconds, 1 data item is output
each time the HOLD key is pressed. If a transfer interval other than 0 seconds is set, pressing the HOLD key starts communication and outputs data in accordance with the transfer interval, and
pressing the HOLD key again stops communication.
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. 2 Use the keys to select Configuration. 3 Press the ENTER key to confirm the selection. 4 Use the keys to select the communication setting. 5 Press the ENTER key to confirm the selection. 6 Use the keys to select Normal or Printer.
If you select Printer, next set the transfer interval.
ENTER
7
Setting Mode
ENTER
ENTER
To return to normal mode from setting mode, simultaneously press the CLEAR and ENTER keys. (Press the CLEAR key to go back one level.)
7-5
7.3 Common
(3) Date/Time
You can set the date and time. (These settings are set to Japan time at the time of shipment.) Top row: Year (2 digits)/Month/Day Bottom row: Hour/Minute/Second
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
2 Use the keys to select Configuration. 3 Press the ENTER key to confirm the selection. 4 Use the keys to select the date/time setting. 5 Press the ENTER key to confirm the selection. 6 Use the keys to set year/month/day on the top row. (Flashes) 7 Press the ENTER key to confirm the setting. 8 Use the ▼ keys to set hour/minute/second on the bottom row. (Flashes)
To return to normal mode from setting mode, simultaneously press the CLEAR and ENTER keys. (Press the CLEAR key to go back one level.)
mode.
ENTER
ENTERENTER
7-6
(4) Temperature Scale Standard
You can select from ITS-90 and IPTS-68 for the temperature scale standard. (default setting: ITS-90)
ITS-90: 1990 International Temperature Scale IPTS-68: 1968 International Practical Temperature Scale
For details, refer to "4.6 Generating Resistance Temperature Detector (RTD) Signals"

8. Communication Function

You can configure the instrument and confirm setting values and measurement values from a personal computer.
Note
You can use a communication cable to connect the instrument to an RS232 compliant serial port of a personal computer, etc.
In printer mode, you can output source values and measurement values at pre­set intervals.

8.1 Cable Connection and Interface Specifications

Cable Connection
Recommended communication cable RS232 (D-SUB 9-pin) cross cable
Connect the communication cable to the RS232 connector on the side of the instrument.
RS232 Settings
Baud rate: 9600 bps Parity: None Stop bit: 2 bits Data length: 8 bits Flow control: None (Xon/Xoff control for printing only) Terminator: Fixed to CrLf
8
Communication Function
8-1

8.2 Communication Command List

8.2 Communication Command List
Command Description When When
AS Sets and queries the source/SINK of the current (DCA)
BL Queries whether the back light is on or off
BT Starts charging the batteries when the AC adapter is connected
CD Sets the source value during calibration
CL Sets and queries the calibration item
CMF Sets and queries the calibration measurement function
CP Sets the calibration point
CS Sets the measurement value during calibration
CSF Sets and queries the calibration source function
CW Writes the calibration data
DT Sets and queries the date and time
DW Decreases the m(th) digit of the source value by 1 digit 䊊䊊
ESC C/RC Initializes the setting information (setting conditions)
ESC S Outputs the status byte 䊊䊊
H Sets and queries the output header of the OD and OM commands 䊊䊊
HD Holds and queries the measurement value indication
IM Sets and queries the mask of status byte 䊊䊊
MF Sets and queries the measurement function
MO Turns on/turns off and queries the measurement value indication
MR Sets and queries the measurement range 䊊䊊
ND Sets and queries the n and m values for divided output (n/m)
NM Sets and queries divided output (n/m)
OB Queries the battery charge state
OD Outputs the measurement values 䊊䊊
OE Outputs the error information 䊊䊊
OM Requests sending of memory data
OR Queries whether an external RJ sensor is connected
OS Outputs the setting information (conditions)
PU Sets and queries display of PULSE (source)
SD Sets and queries the source values
SF Sets and queries the source function
SO Starts/stops and queries source 䊊䊊
SR Sets and queries the source range 䊊䊊
SY Switches and queries normal mode and calibration mode 䊊䊊
TE Sets and queries TC and RTD (source) display
TT Sets and queries the international temperature standard
UP Increases the m (th) digit of the source value by 1 digit 䊊䊊
VO Sets and queries start/stop of 24 V DC (LOOP) power supply
8-2
Normal Setting

8.3 Detailed Description of Commands

8.3 Detailed Description of Commands
Setting and Control
Command: Send command syntax Answer: Return data syntax of command (setting, control) with no
response. When an error occurs, the same data as that of the error message ERRm (m = error number) displayed on the LCD is returned.
Query
Command: Send command syntax Return: Return data syntax of command (query) with response.
Operating Condition by Mode
Normal: The command can be used during normal measurement/source
operation.
Setting: The command can be used in setting mode (Chapter 7) or
calibration mode (Chapter 9). For details on whether each command can be used in other modes, refer to "8.5 Table of Valid Communication Commands."
8
Communication Function
AS Normal SettingSets and queries the source/SINK of the current (DCA)
Command = ASm<Delimiter> Answer = ASm <Delimiter>
Command = AS?<Delimiter> Return = ASm <Delimiter>
Parameter m=0: Source (Generation)/1:SINK (Draw in)
When the current source setting value is 0 mA and the measurement
function is other than DCA, an error is returned.
BL Normal SettingQueries whether the back light is on or off
Command = BLm<Delimiter> Answer = BLm <Delimiter>
Command = BL?<Delimiter> Return = BLm <Delimiter>
Parameter m=0: Off/1: On
default value of m = 0 (Off)
BT Normal SettingStarts charging the batteries when the AC adapter is connected
Command = BT<Delimiter>
An error is generated in the following cases (LCD indication)
• AC power supply is not connected (Err 14)
• Charging has already started (Err 12)
Use the OB command to confirm the charge state.
8-3
8.3 Detailed Description of Commands
CD Normal SettingSets the source value during calibration
During generation in calibration mode, sets the current output source
values as the calibration values for the selected function, range, and
scale (+FS/0).
CL Normal SettingSets and queries the calibration item
Command = CL?<Delimiter> Return = CLm <Delimiter>
Parameter m=3: (Calibration) Source/4: (Calibration) Measure
CMF Normal SettingSets and queries the calibration measurement function
Command = CMF?<Delimiter> Return = CMFm <Delimiter>
Parameter
m=0: DCV/1: DCA/2:
default Value of m=0 (DCV)
CP Normal SettingSets the calibration point
Parameter
m=0: + FS Calibration/1: + Zero Calibration/2: -FS Calibration
CS Normal SettingSets the measurement value during calibration
During measurement in calibration mode, sets the current input
measurement values as the calibration values for the selected function,
range, and scale (+FS/0/-FS ).
*Command = CD<Delimiter> Answer = OD <Delimiter>
*Command = CLm<Delimiter> Answer = CLm <Delimiter>
*Command = CMFm<Delimiter> Answer = CMFm <Delimiter>
*Command = CPm<Delimiter> Answer = CPm <Delimiter>
*Command = CS<Delimiter> Answer = CS <Delimiter>
8-4
CSF Normal SettingSets and queries the calibration source function
䊊*Command = CSFm<Delimiter> Answer = CSFm <Delimiter>
Command = CSF?<Delimiter> Return = CSFm <Delimiter>
Parameter m=0: DCV/1: DCA/2:
default Value of m=0 (DCV)
CW Normal SettingWrites the calibration data
䊊*Command = CW<Delimiter>
Answer = CW, OK<Delimiter> (Normal End)
Writes to EEPROM after calibrating each function and range.
If the power is turned off without executing this command, the values that
were calibrated just before are discarded.
*: Can only be used in calibration mode of setting mode.
8.3 Detailed Description of Commands
DT Normal SettingSets and queries the date and time
Answer = DTyyyymmddhhmmss<Delimiter>
Command = DT? (CrLf)
Return = yyyyy/mm/dd, hh:mm:ss<Delimiter>
Parameter (default Value)
yyyy: Year (2006) 4 byte, mm: Month (04) 2 byte, dd: Day (01) 2 byte
hh: Hour (00) 2 byte, mm: Minute (00) 2 byte, ss: Second (00) 2 byte
Setting (confirmation) of the 2 leftmost digits of the year is only possible
for the communication function.
(Using the operation keys to change the values of the 2 leftmost digits on
the LCD will not move up or down to the 2 rightmost digits.)
DW Normal SettingDecreases the m (th) digit of the source value by 1 digit
Parameter m = 1 to 5 (1: Least Significant Digit to 5: Most Significant
Digit)
ESC C
or
RC
("ESC" = ASCII 0 &batu&1B)
or
Command = RC <Delimiter>
The following settings (common setting item) are not initialized.
• Auto power off setting
• Communication setting
• International temperature standard selection (ITS-90/IPTS-68)
• Time setting
Command = DTyyyymmddhhmmss<Delimiter>
䊊䊊Command = DWm<Delimiter> Answer = DW, OK<Delimiter>
8
Communication Function
Normal SettingInitializes the setting information (setting conditions)
Command = ESC C<Delimiter
ESC S Normal SettingOutputs the status byte
("ESC" = ASCII 0 x 1B)
䊊䊊Command = ESC S<Delimiter>
The status byte is output. (Decimal number)
(Refer to "8.6 Status Byte Format.")
H Normal SettingSets and queries the output header of the OD and OM commands
䊊䊊Command = Hm<Delimiter> Answer = Hm <Delimiter>
Command = H?<Delimiter> Return = Hm <Delimiter>
Parameter m=0: No/1: Yes
default Value of m=0 (No)
HD Normal SettingHolds and queries the measurement value indication
Command = HDm<Delimiter> Answer = HDm <Delimiter>
Command = HD?<Delimiter> Return = HDm <Delimiter>
Parameter m=0: Indication Update/1: Indication Hold
8-5
8.3 Detailed Description of Commands
IM Normal SettingSets and queries the mask of status byte
Command = IM?<Delimiter> Return = IMm <Delimiter>
Performs detection or sets mask for each bit of status byte.
If IM0 is set, all information bits are masked.
If IM63 is set, the current operating conditions are reflected for all
information bits.
(No mask)
Parameter m = 0 to 63 (Decimal number)
1: Detects bit 0 (Measurement Ends)
2: Detects bit 1 (Output Change Ends)
4: Detects bit 2 (Syntax Error)
8: Detects bit 3 (Overrange)
16: Detects bit 4 (24 V Loop Output Error)
32: Detects bit 5 (Error During Output)
(Bits 6 and 7 of the status byte are fixed bits.)
default Value of m = 63 (No Mask)
MF Normal SettingSets and queries the measurement function
Command = MF?<Delimiter> Return = MFm <Delimiter>
Parameter
m=0: DCV/1: DCA/2: &ohm&/3: TC/4: RTD/5: Freq
MO Normal SettingTurns on/turns off and queries the measurement value indication
Command = MO?<Delimiter> Return = MOm <Delimiter>
Parameter m=0: Turn off/1: turn on
䊊䊊Command = IMm<Delimiter> Answer = IMm <Delimiter>
Command = MFm<Delimiter> Answer = MFm <Delimiter>
Command = MOm<Delimiter> Answer = MOm <Delimiter>
8-6
MR Normal SettingSets and queries the measurement range
䊊䊊Command = MRm<Delimiter> Answer = MRm <Delimiter>
Command = MR?<Delimiter> Return = MRm <Delimiter>
Parameter
[DCV] m = 0: 500 mV/1: 5 V/2: 35 V
[DCA] m = 0: 20 mA DC/1: 100 mA DC
[] m = 0: 500 /1: 5 k/2: 50 k
[TC] m = 0: K/1: E/2: J/3: T/4: R/5: B/6: S/7: N/8: L/9: U
[RTD] m = 0: PT100/1: JPT100
[Freq] m = 0: 100 Hz/1: 1000 Hz/2: 10 kHz/3: CPM/4: CPH
ND Normal SettingSets and queries the n and m values for divided output (n/m)
Command = NMnm<Delimiter> Answer = NMnm <Delimiter>
Command = NM?<Delimiter> Return = NMnm <Delimiter>
Parameter
n = n value (2 digits from 00 to 19, n m)
m = m value (2 digits from 01 to 19, n m)
default Value of n = 01 and m = 01
8.3 Detailed Description of Commands
NM Normal SettingSets and queries divided output (n/m)
Command = NM?<Delimiter> Return = NMm <Delimiter>
Parameter m = 0: Off/1: On
default Value of m = 0 (Off)
OB Normal SettingQueries the battery charge state
Parameter
m = 0: Off
1: On (Charging)
A query of the charge state can executed approximately 2 seconds after
the CHARGE key has been pressed (or the BT command has been sent).
OD Normal SettingOutputs the measurement values
Parameter
<Header Section (4 bytes)>
a = V: Voltage/A: Current/O: Resistance/T: Temperature/F: Frequency
b = DC: Direct Current/ AC: Alternating Current/R3: Three-wire Method
Resistance Measurement
c = N: Normal/O: Overrange/E: No Data/B: Burnout
<Data Section (10 Bytes)>
d = Measurement Value (7 digits)
e = E+0/E+3/E-3
(When overrange, no data, and burnout: de = 99999.E+3)
Command = NMm<Delimiter> Answer = NMm <Delimiter>
Command = OB<Delimiter> Return = m <Delimiter>
䊊䊊Command = OD<Delimiter> Return = abcde <Delimiter>
8
Communication Function
OE Normal SettingOutputs the error information
䊊䊊Command = OE<Delimiter> Return = ERRm <Delimiter>
Outputs the last error generated
After the output is returned, the save error number is overwritten with
"ERR00<Delimiter>."
If no error occurs, "ERR00<Delimiter>" is returned.
Parameter
m = Error Code Number
8-7
8.3 Detailed Description of Commands
OM Normal SettingRequests sending of memory data
Query of memory data
Parameter
m = Memory Data Number (0 to 99)
n = Date, Time, Measurement Values, Source Values, [PULSE Source
Amplitude] <Delimiter>
= yyyy/mm/dd, hh:mm:ss, abcde, fghij [, fghij] <Delimiter>
Date
yyyy/mm/dd = yyyy (Year) mm (Month) dd (Day)
hh:mm:ss = Time
Measurement Value Header Section
a = V: Voltage/A: Current/O: Resistance/T: Temperature/F: Frequency
b = DC: Direct Current/ AC: Alternating Current/Blank: Nothing Applicable
c = N: Normal/O: Overrange/E: No Data
(When no data: abc = E)
Measurement Data Section
d = Measurement Value (7 digits)
e = E+0/E+3/E-3
(When overrange and no data: de = 99999.E+3)
Source Value Header Section
f = V: Voltage/A: Current/O: Resistance/T: Temperature/F: Frequency
g = DC: Direct Current/ AC: Alternating Current/Blank: Nothing Applicable
h = N: Normal/E: No Data
(When no data: abc = E)
Source Value Data Section
i = Source Value (8 Digits)
j = E+0/E+3/E-3
(When no data: de = 99999.E+3)
Command = OMm<Delimiter> Return = n <Delimiter>
8-8
OR Normal SettingQueries whether an external RJ sensor is connected
Command = OR<Delimiter> Return = m <Delimiter>
Parameter
m = 0: disconnected/1: connected
8.3 Detailed Description of Commands
OS Normal SettingOutputs the setting information (conditions)
Parameter
a (Measure) = ON/OFF
b (Measure Function) = DCV/DCA/OHM/TC/RTD/FREQ
c (Measure Range) = (DCV) 35 V/5 V/500 mV
(DCA) 100 mA/20 mA
(OHM) 50 kOHM/5 kOHM/500 OHM
(TC) K/E/J/T/R/B/S/N/L/U
(RTD) PT100/JPT100
(FREQ) 100 Hz/1000 Hz/10 kHz/CPM/CPH
d (Source) = ON/OFF
e (Source Function) = DCV/DCA/OHM/TC/RTD/PULSE
f (Source Range) = (DCV) 100 mV/1 V/10 V/30 V
(DCA) 20 mA/4-20 mA
(OHM) 500 OHM/5 kOHM/50 kOHM
(TC) K/E/J/T/R/B/S/N/L/U
(RTD) PT100/JPT100
(FREQ) 100 Hz/1000 Hz/10 kHz/50 kHz/CPM
g (Source Value)
h (Output for 24 V LOOP Measurement) = ON/OFF
i (Backlight) = ON/OFF
j (Charging) = OFF/ON
Function b<Delimiter>
Command = OS<Delimiter> Return = Measure a<Delimiter>
Range c<Delimiter>
Source d<Delimiter>
Function e<Delimiter>
Range f<Delimiter>
Data g<Delimiter>
24 V Output h<Delimiter>
Light i<Delimiter>
Charge j<Delimiter>
8
Communication Function
PU Normal SettingSets and queries display of PULSE (source)
Command = PUm<Delimiter> Answer = PUm <Delimiter>
Command = PU?<Delimiter> Return = PUm <Delimiter>
Parameter m = 0: Frequency/1: Amplitude/2: Pulse Number
default Value m = 0 (Frequency)
This can only be set and queried when the source function is PULSE.
8-9
8.3 Detailed Description of Commands
SD Normal SettingSets and queries the source values
Command = SD?<Delimiter> Return = SDm <Delimiter>
Parameter
m = Source Values
100m V m = 0 to 110.000 mV
1 V m = 0 to 1.10000 V
10 V m = 0 to 11.0000 V
30 V m = 0 to 30.00 V
20 mA, 4-20 mA m = 0 to +22.000 mA
20 mA SINK m = 0 to -22.000 mA
500 m = 0 to 550.00
5k m = 0 to 5.5000 k
50 k m = 0 to 55.000 k
PT100 m = -200.0 to 850.0 C
PT100 (ITPS-68) m = -200.0 to 650.0 C
JPT100 m = -200.0 to 500.0 C
Km = -200.0 to 1372.0 C
Em = -200.0 to 1000.0 C
Jm = -200.0 to 1200.0 C
Tm = -200.0 to 400.0 C
Nm = -200.0 to 1300.0 C
Lm = -200.0 to 900.0 C
Um = -200.0 to 400.0 C
Rm = 0 to 1768 C
Sm = 0 to 1768 C
Bm = 600 to 1820 C
100 Hz (*1) m = 1.00 to 110.00 Hz
1000 Hz (*1) m = 90.0 to 1100.0 Hz
10 kHz (*1) m = 0.9 kHz to 11.0 kHz
50 kHz (*1) m = 9 kHz to 50 kHz
1000 CPM m = 1.0 to 1100.0 CPM
PULSE DCV (*2) m = 0 to 11.0000 V
PULSE Cycle (*3) m = 0 (cont), 1 to 60000 cycle
Command = SDm<Delimiter> Answer = SDm <Delimiter>
8-10
*1: When the PULSE source display setting is frequency, only (PU0) can
be set
*2: When the PULSE source display setting is amplitude, only (PU1) can
be set
*3: When the PULSE source display setting is pulse number, only (PU2)
can be set
SF Normal SettingSets and queries the source function
Command = SFm<Delimiter> Answer = SFm <Delimiter>
Command = SF?<Delimiter> Return = SFm <Delimiter>
Parameter
m = 0: DCV/1: DCA/2: /3: TC/4: RTD/5: Pulse
Set the PULSE source DCV and cycle of each parameter with the PU
command.
8.3 Detailed Description of Commands
SO Normal SettingStarts/stops and queries source
Command = SO?<Delimiter> Return = SOm <Delimiter>
Parameter
m = 0: Stop/1: Start
SR Normal SettingSets and queries the source range
Command = SR?<Delimiter> Return = SRm <Delimiter>
Parameter
[DCV] m = 0: 100 mV/1: 1 V/2: 10 V/3: 30V
[DCA]
(When normal) m = 0: 20 mA/1: 4-20 mA
(When Calibrating) m = 0: 20 mA/2: 20 mA SINK
[]
(When normal) m = 0: 500 /1: 5 k/2: 50 k
(When Calibrating) m = 0: 500 (1 mA)/1: 5 k (0.1 mA)/
2: 50 k (0.05 mA)/
3: 500 (5 mA)/4: 5 k (0.5 mA)
[TC] m = 0: K/1: E/2: J/3: T/4: R/5: B/6: S/ 7: N/8: L/9: U
[RTD] m = 0: PT100/1: JPT100
[PULSE] m = 0: 100 Hz/1: 1000 Hz/2: 10 kHz/3: 50 kHz/4: 1000 CPM
SY Normal SettingSwitches and queries normal mode and setting mode
Command = SY?<Delimiter> Return = SYm <Delimiter>
Parameter
m = 0: Normal Mode/1: Setting Mode
default Value
m = 0 (Normal Mode)
䊊䊊Command = SOm<Delimiter> Answer = SOm <Delimiter>
䊊䊊Command = SRm<Delimiter> Answer = SRm <Delimiter>
8
Communication Function
䊊䊊Command = SYm<Delimiter> Answer = SYm <Delimiter>
TE Normal SettingSets and queries TC and RTD (source) display
Command = TEm<Delimiter> Answer = TEm <Delimiter>
Command = TE?<Delimiter> Return = TEm <Delimiter>
Parameter
m = 0: Temperature Value/1: mV Value (Resistance Value)/2: Room
Temperature
default Value
m = 0 (Temperature Value)
TT Normal SettingSets and queries the international temperature standard
Command = TTm<Delimiter> Answer = TTm <Delimiter>
Command = TT?<Delimiter> Return = TTm <Delimiter>
Parameter m = 0: IPTS-68/1: ITS-90
default Value m = 1 (ITS-90)
8-11
8.3 Detailed Description of Commands
UP Normal SettingIncreases the m (th) digit of the source value by 1 digit
Parameter
m = 1 to 5 (1: Least Significant Digit to 5: Most Significant Digit)
VO Normal SettingSets and queries start/stop of 24 V DC (LOOP) power supply
Command = VO?<Delimiter> Return = VOm <Delimiter>
This controls the ON/OFF of output of 24 V LOOP measurement power
supply (output from measurement terminals).
This can only be set when the measurement function is DCA.
Parameter m = 0: Stop Supply/1: Start Supply
Initial Value m = 0 (Stop Supply)

8.4 Error Code List

Indication Description
Err 00 No error (No error code is displayed on the LCD)
Err 11 Received a command that is not used by the instrument
Err 12 The parameter specified for the command is incorrect
Err 13 Received a command that cannot be executed due to the instrument state
Err 14 Attempted to charge the battery when no AC power supply is connected
Err 15 Attempted to charge the battery when no NiMH battery is connected
Err 16 Detected error during calibration
Err 20 24 V DC (LOOP) measurement power supply error
Err 23 The current or voltage of the source output has become excessive
Err 60 The setting value storage information of the EEPROM is inappropriate
Err 61 The measurement calibration value storage information of the EEPROM is
inappropriate
Err 62 The source calibration value storage information of the EEPROM is
inappropriate
Err 79 ROM check error at power on
Err 80 RAM check error at power on
䊊䊊Command = UPm<Delimiter> Answer = UP, OK <Delimiter>
Command = VOm<Delimiter> Answer = VOm <Delimiter>
8-12

8.5 Table of Valid Communication Commands

8.5 Table of Valid Communication Commands
X: Invalid (restricted) Blank: Valid
Sweep
Mode
Measure
n/m
Mode
*3
Program
Mode
*3
Step
Sweep
Mode
*3
Normal
Calibration
Mode
AS
BL
BT
CD
CL
CMF
CP
CS 
CSF
CW
DT 
DW
ESC C
ESC S
H
HD?
HDm
IM
MF
MO?
MOm
MR?
MRm
ND?
NDm
NM?
NMm
*3
Mode
*3
Memory
Mode
Setting
Mode
Linear Sweep
*3: When the temperature (TC, RTD) display function is selected for both
source and measurement, the settings selected for the thermocouple and resistance temperature detector on the source side take priority.
ment
Off
8
Communication Function
8-13
8.5 Table of Valid Communication Commands
X: Invalid (restricted) Blank: Valid
Normal
Calibration
Mode
OB
OD
OE
OM
OR
OS
PU?
PUm 
SD?
SDm
SF 
SO?
SOm
SR?
SRm
SY?
SYm
TE?
TEm
TT
UP
VO
Mode
Memory
Mode

Setting
Mode
Linear
Sweep
Mode
Step
Sweep
Mode



*3: When the temperature (TC, RTD) display function is selected for both
source and measurement, the settings selected for the thermocouple and resistance temperature detector on the source side take priority.
n/m
Mode
Program
Sweep
Mode
Measure
ment
Off
8-14

8.6 Status Byte Format

8.6 Status Byte Format
Description of the ESC S Command
bit 7
(Fixed)
bit 6
0
(Fixed)
bit 5
0
Output
Error
bit 4
24 V Power
Supply
Error
bit 3
Overrange
bit 2
Syntax
Error
bit 1
Output
Change
Complete
bit 7: Fixed to 0 bit 6: Fixed to 1 bit 5: Becomes 1 if an error occurs during output.
(MAIN CPU [OVERLOAD] terminal input reflected)
The information is retained until the status byte is retrieved. bit 4: Becomes 1 if an error occurs with 24 V loop measurement power
supply.
(MAIN CPU [/24VOVF] terminal input reflected)
The information is retained until the status byte is retrieved. bit 3: Becomes 1 if a measurement value overrange occurs.
The information is retained until the status byte is retrieved. bit 2: Becomes 1 when a prohibited operation or command was
processed, a command could not be interpreted, or a parameter is
out of the setting range.
The information is retained until the status byte is retrieved. bit 1: Becomes 1 when output is stable after an output value is changed
in the output ON state.
The information is retained until the status byte is retrieved. bit 0: Becomes 1 when the measurement data is confirmed during
measurement.
The information is retained until the status byte is retrieved.
bit 0
Measurem
ent End
8
Communication Function
8-15

8.7 Output Format of Printer Mode

8.7 Output Format of Printer Mode
Source setting values and measurement values are output in printer mode (when a printer is connected). The output format is 29 characters per line.
Example of Output
When Source OFF and Measure OFF
Source :OFF Measure :OFF
When Source OFF and Measure ON
Source :OFF Measure :Range 35V Data 0.000V Average OFF
When Source ON and Measure OFF
Source :Function DCV Range 100mV Data 0.000mV Measure :OFF
When Source ON and Measure ON
Source :Function DCV Range 100mV Data 0.000mV Measure :Range 35V Data 0.000V Average OFF
8-16

9. Calibration Mode

To maintain high accuracy, it is recommended to calibrate the instrument once a year. This section describes the calibration procedure using the standard devices recommended in "Standard Device Selection."
(1) Standard Device Selection Source Function
Function
DCV
DCA
Measurement Function
Function
DCV
DCA
Standard Device Range Accuracy
• Digital multimeter
• Digital multimeter
• DC voltage/current
standard
Standard Device Range Accuracy
• DC voltage/current
standard
Reference resistor
100 mV
1 V
10 V
30 V
20 mA
20 mA SINK
500
5 k
50 k
500 mV
5 V
35 V
20 mA
100 mA
500
5 k
50 k
0.002% + 1 V
0.002% + 5 V
0.002% + 0.05 mV
0.002% + 1 V
0.0035% + 0.5 A
0.0035% + 0.5 A
0.002% + 0.01
0.005% + 0.15
0.01% + 5
0.002% + 5 V
0.002% + 50 V
0.0025% + 0.5 mV
0.0025% + 0.4 A
0.004% + 3 A
0.0055% + 7.5 m
0.0055% + 0.75
0.055% + 1
9
Calibration Mode
(2) Required Environment and Conditions for Calibration
Ambient temperature: 23 1C
Relative humidity: 45% to 75%
Warm-up time: The standard devices must be warmed up for the
specified times and the instrument must be warmed up
for at least 2 hours (for calibration).
9-1

9.1 Calibration of Source Functions (Adjustment)

9.1 Calibration of Source Functions (Adjustment)
(1) Calibration Points and Calibration Ranges
Using the ▲ ▼ output setting value keys, adjust the output values so that the readings on the standard device (CA150 source values) match the calibration points shown below.
Function
DCV
DCA
Range
100 mV
1 V
10 V
30 V
20 mA
20 mA SINK
L 500
H 500
L 5 k
H 5 k
50 k
Calibration Point
ZERO
0 mV
0 V
0 V
0 V
0 mA
0 mA
0 mV*
0 mV*
0 mV
0 mV
0 m V
Full Scale
100.000 mV
1.00000 V
10.0000 V
30.00 V
20.000 mA
-20.000 mA
500.00 mV
2500.00 mV
500.00 mV
2500.00 mV
2500.00 mV
Condition
External voltage of 28 V
Excitation current of 1 mA
Excitation current of 5 mA
Excitation current of 0.1 mA
Excitation current of 0.5 mA
Excitation current of 0.05 mA
Connection
Diagram
<1>
<2>
<3>
<4>
Caution
About Resistance (500 ) Internal Offset Calibration *: When performing zero point calibration, make sure the voltage between the
H and L terminals is within approximately20V (0.02 mV). If this value is exceeded, the instrument needs to be repaired (internal calibration).
About Resistance Excitation Current
When calibrating the 500 and 5 kranges, two types of calibration are required because of differences of current (excitation current) inflowing from an external device.
L 500 , 1 mA Calibration with the resistance measurement range of a L 5 k, 0.1 mA digital multimeter is possible. During calibration, make
sure the resistance measurement current is the current value shown on the left.
H 500 , 5 mA Apply the current shown on the left from an external device H 5 k, 0.5 mA as shown in connection diagram <4> and then measure
the voltage drop and perform calibration.
9-2
9.1 Calibration of Source Functions (Adjustment)
Tip
•You can select to calibrate just the functions and ranges that require calibration. (Be sure to perform both zero point and full scale (FS) calibration of the same range together.)
• Calibration of thermocouple (TC) and resistance temperature detector (RTD) takes place at the same time as calibration of the100 mV and 500 ranges.
(2) Connection Diagrams
<1> DCV
CA150
<3> 20 mA SINK
CA150
H L
SOURCE
H L
SOURCE
Digital
multimeter
H (V)
L
Digital
multimeter
I (A)
L
H
L
DC voltage/current
standard
<2> DCA
CA150
<4> &ohm&, RTD
CA150
H L
SOURCE
H L
SOURCE
Digital
multimeter
I (A)
Excitation
current
L
Digital
multimeter
DC voltage/current
standard
(3) Calibration Procedure
Connect the instrument in accordance with the function and range to be calibrated. Refer to "(2) Connection Diagrams."
1 Simultaneously press the CLEAR and ENTER keys to switch to setting
mode. 2 Use the key to select source (SEt SrC). 3 Press the ENTER key to confirm the selection. 4 Use the key to select the calibration (SrC CAL) setting. 5 Press the ENTER key to confirm the selection. 6 Use the FUNCTION and RANGE keys to set the range to calibrate.
(The output full scale value of the selected range appears on the bottom
row.) 7 Press the ENTER key to confirm the setting. 8 Start zero-point calibration:
The CAL and ZERO segments light, the output setting value of the
instrument appears on the bottom row, and the zero point calibration value
of the selected range appears on the top row.
9
Calibration Mode
H (V)
L
H
L
9-3
Loading...