YOKOGAWA AXW080, AXW100, AXW040, AXW125 AXW150, AXW250 User Manual

...
User’s Manual
ADMAG TI Series AXG and AXW Magnetic Flowmeter ATEX Explosion Protection Type
IM 01E21A03-02EN
IM 01E21A03-02EN
1st Edition
ADMAG TI Series AXG and AXW Magnetic Flowmeter ATEX Explosion Protection Type
IM 01E21A03-02EN 1st Edition

Contents

1. Introduction ...................................................................................................1
2. Explosion Protection Type ..........................................................................3
2.1 Technical Data ...................................................................................................... 3
2.2 Cable Entry ........................................................................................................... 6
2.3 Installation ............................................................................................................ 6
2.4 Operation .............................................................................................................. 7
2.5 Maintenance and Repair ..................................................................................... 7
2.6 Name Plate ............................................................................................................ 8
2.7 Specic Condition of Use ................................................................................... 9
Revision Information
1st Edition: Oct. 2018 (KP) All right Reserved Copyright © 2018. Yokogawa Electric Corporation
<1. Introduction>

1. Introduction

This manual provides the basic guidelines for explosion protection type of ADMAG TI (Total
Insight) Series AXG and AXW magnetic owmeters.
For the items which are not covered in this manual, read the applicable user’s manuals and general
specications as listed in IM 01E21A11-01EN
(Read Me First (Optional Code EC)). These documents can be downloaded from the website of YOKOGAWA. To ensure correct use of the product, read these manuals thoroughly and fully understand how to operate the product before maintaining it. For
method of checking the model and specications, read the applicable general specications in
IM 01E21A11-01EN (Read Me First (Optional Code EC)).
Website address: http://www.yokogawa.com/d/doc/
These manuals can be downloaded from the website of YOKOGAWA or purchased from the YOKOGAWA representatives.
NOTE
A CAUTION sign denotes a hazard. It calls attention to procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or the entire product.
An IMPORTANT sign denotes that attention is required to avoid damage to the product or system failure.
A NOTE sign denotes information necessary for essential understanding of operation and features.
CAUTION
IMPORTANT
NOTE
1
1
When describing the model name like “AXG###” in this manual, “###” means any of the following. For AXG###:
002, 005, 010, 015, 025, 032, 040, 050, 065, 080, 100, 125, 150, 200, 250, 300, 350, 400
For AXW###:
025, 032, 040, 050, 065, 080, 100, 125, 150, 200, 250, 300, 350, 400
Precautions Related to the Protection,
Safety, and Alteration of the Product
The following safety symbol marks are used in this manual and product.
WARNING
A WARNING sign denotes a hazard. It calls attention to procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in injury or death of personnel.
The following symbols are used in the product and the manual to indicate the accompanying safety precautions:
Protective grounding terminal
Functional grounding terminal (This terminal should not be used as a protective grounding terminal.)
Alternating current
Direct current
Caution This symbol indicates that the operator must read an explanation in the user’s manual in order to avoid the risk of injury or death of personnel or damage to the product.
For the protection and safe use of the product and the system in which this product is incorporated, be sure to follow the instructions and precautions on safety that is stated in user's manual as listed in IM 01E21A11-01EN (Read Me First (Optional Code EC)) whenever you handle the product. Take special note that if you handle the product in a manner that violated these instructions, the protection functionality of the product may be damaged or impaired. In such cases, YOKOGAWA does not guarantee the quality, performance, function, and safety of product.
• Do not modify this product.
• The product should be disposed of in accordance with local and national legislation/ regulations.
<1. Introduction>
2
2
Regarding This User’s Manual
• This manual should be provided to the end user.
• The contents of this manual are subject to change without prior notice.
• All rights reserved. No part of this manual may be reproduced in any form without YOKOGAWA’s written permission.
• YOKOGAWA makes no warranty of any kind with regard to this manual, including, but not limited to, implied warranty of merchantability
and tness for a particular purpose.
• If any question arises or errors are found, or if any information is missing from this manual,
inform the nearest YOKOGAWA sales ofce.
• The specications covered by this manual are
limited to those for the standard type under the
specied model number break-down and do not
cover custom-made products.
• Note that changes in the specications,
construction, or component parts of the
product may not immediately be reected in
this manual at the time of change, provided that postponement of revisions will not cause
difculty to the user from a functional or
performance standpoint.
• This manual is intended for the following personnel; Engineers responsible for installation and wiring of the product.
• To ensure correct use, read this manual and the applicable user’s manuals as listed in IM 01E21A11-01EN (Read Me First (Optional Code EC)) thoroughly before starting operation.
Read the general specications as listed in
IM 01E21A11-01EN (Read Me First (Optional
Code EC)) for its specication.
Trademark
• All the brands or names of Yokogawa Electric’s products used in this manual are either trademarks or registered trademarks of Yokogawa Electric Corporation.
• All other company and product names mentioned in this manual are trade names, trademarks or registered trademarks of their respective companies.
• In this manual, trademarks or registered trademarks are not marked with ™ or ®.
For Safe Use of Product
For the protection and safe use of the product and the system in which this product is incorporated, be sure to follow the instructions and precautions on safety that is stated in user's manual as listed in IM 01E21A11-01EN (Read Me First (Optional Code EC)) whenever you handle the product. Take special note that if you handle the product in a manner that violated these instructions, the protection functionality of the product may be damaged or impaired. In such cases, YOKOGAWA shall not be liable for any indirect or consequential loss incurred by either using or not being able to use the product.
<2. Explosion Protection Type>

2. Explosion Protection Type

3
3
WARNING
AXG### and AXW### magnetic owmeter
(Integral Flowmeter and Remote Sensor), and AXG4A and AXW4A Remote Transmitter are
products which have been certied as explosion
protection type products if model code for
explosion protection is specied. Strict limitations
are applied to the structures, installation locations, external wiring work, maintenance and
repairs, etc. of these products. Sufcient care
must be taken, as any violation of the limitations may cause dangerous situations. Be sure to read this manual before handling the explosion protection type products.
WARNING
The terminal box cover and display cover are locked by hexagon socket head cap screw. In the case of opening the cover, use the hexagonal wrench (nominal size 3). Read the Maintenance Manual, IM 01E22A01-02EN or IM 01E24A01­02EN, for the procedure. Before opening the cover, be sure to check that the power of
owmeter has been turned off. Once the cover is
closed, be sure to re-lock the product. Be sure to lock the cover with the screw by using the hexagonal wrench after tightening the cover.
IMPORTANT
For multiple approval types: (Transmitter Wiring Terminal: Clamp type) For the installation of this transmitter, once a particular type of protection is selected, any other type of protection cannot be used. Apply a permanent mark in the check box of the selected
approval type on the certication label on the
transmitter to distinguish it from unused approval types.

2.1 Technical Data

Applicable Standard:
EN 60079-0: 2012+A11:2013, EN 60079-1: 2014, EN 60079-7: 2015, EN 60079-11: 2012, EN 60079-31: 2014
Certicate:
FM 17ATEX0031X
Integral Flowmeter
• Type of Gas Atmosphere Protection Group: II Category: 2G Type of Protection:
Ex db eb ia IIC T6...T3 Gb (Transmitter Wiring Terminal: M4 screw type) Ex db eb ia IIC T6...T3 Gb Terminal Compartment: Ex db or Ex eb (Transmitter Wiring Terminal: Clamp type)
• Type of Dust Atmosphere Protection Group: II Category: 2D Type of Protection:
Ex tb IIIC T75°C...T130°C Db
• Enclosure:
IP66/IP67 in accordance with EN 60529
• Maximum Surface Temperature: See Table 2.1 and Table 2.2.
• Ambient Temperature: See Table 2.1 and Table 2.2.
• Process Temperature: See Table 2.1 and Table 2.2.
• Power Supply: 100 to 240 Va.c. (50/60 Hz) / 100 to 120 Vd.c. 24 Va.c. (50/60 Hz) / 24 Vd.c.
• Um: 250 V
• Current I/O: 4 to 20 mA, 32 Vd.c. max.
• Digital I/O: 30 Vd.c. max., 200 mA max.
<2. Explosion Protection Type>
4
4
Remote Sensor
• Type of Gas Atmosphere Protection Group: II Category: 2G Type of Protection:
Ex db eb ia IIC T6…T3 Gb
• Type of Dust Atmosphere Protection Group: II Category: 2D Type of Protection:
Ex tb IIIC T75°C...T150°C Db
• Enclosure:
IP66/IP67 in accordance with EN 60529
• Maximum Surface Temperature: See Table 2.1 and Table 2.2.
• Ambient Temperature: See Table 2.1 and Table 2.2.
• Process Temperature: See Table 2.1 and Table 2.2.
• Um: 250 V
Remote Transmitter
• Type of Gas Atmosphere Protection Group: II Category: 2G Type of Protection:
Ex db IIC T6 Gb (Transmitter Wiring Terminal: M4 screw type) Ex db IIC T6 Gb or Ex db eb IIC T6 Gb (Transmitter Wiring Terminal: Clamp type)
• Type of Dust Atmosphere Protection Group: II Category: 2D Type of Protection:
Ex tb IIIC T75°C Db
• Enclosure:
IP66/IP67 in accordance with EN 60529
• Maximum Surface Temperature: T75°C
• Ambient Temperature: –40°C to +60°C
• Power Supply: 100 to 240 Va.c. (50/60 Hz) / 100 to 120 Vd.c. 24 Va.c. (50/60 Hz) / 24 Vd.c.
• Current I/O: 4 to 20 mA, 32 Vd.c. max.
• Digital I/O: 30 Vd.c. max., 200 mA max.
Table 2.1 Temperature Table
Model Size
AXG002, AXG005, AXG010, AXG015
AXG025, AXG032, AXG040, AXG050, AXG065, AXG080, AXG100, AXG125
AXG150, AXG200, AXG250, AXG300, AXG350, AXG400
AXG015, AXG025, AXG032, AXG040, AXG050, AXG065, AXG080, AXG100, AXG125
AXG002, AXG005, AXG010, AXG015, AXG025, AXG040, AXG050, AXG080, AXG100, AXG150, AXG200
AXW025, AXW032, AXW040, AXW050, AXW065, AXW080, AXW100, AXW125
AXW150, AXW200, AXW250, AXW300, AXW350, AXW400
2.5 to 15 mm
(0.1 to 0.5 in.)
25 to 125 mm
(1 to 5 in.)
150 to 400 mm
(6 to 16 in.)
15 to 125 mm
(0.5 to 5 in.)
2.5 to 200 mm (0.1 to 8 in.)
25 to 125 mm
(1 to 5 in.)
150 to 400 mm
(6 to 16 in.)
Process
Connection
Wafer, Flange PFA Lining Table A Table C
Wafer, Flange PFA Lining Table B Table D
Wafer, Flange PFA Lining Table A Table C
Clamp, Union,
Butt Weld Joint
Wafer Ceramics Tube Table G Table H
Flange PTFE Lining Table I Table I
Flange PTFE Lining Table J Table J
Lining
PFA Lining Table E Table F
Temperature Table
Integral
Flowmeter
Remote
Sensor
<2. Explosion Protection Type>
Table 2.2 Ambient Temperature and Process Temperature
5
5
Table No.
A -40°C to +60°C
B
C -40°C to +60°C
D
E
F
G -10°C to +60°C
H -10°C to +60°C
I
J -10°C to +60°C
Ambient
Temperature
-40°C to +45°C T6 T75°C -40°C to +50°C
-40°C to +60°C
-40°C to +45°C T6 T75°C -40°C to +50°C
-40°C to +60°C
-10°C to +45°C T6 T75°C -10°C to +50°C
-10°C to +60°C
-10°C to +45°C T6 T75°C -10°C to +50°C
-10°C to +60°C
-10°C to +45°C T6 T75°C -10°C to +50°C
-10°C to +60°C
Temperature
Class
T6 T75°C -40°C to +75°C
T5 T90°C -40°C to +90°C
T4 T120°C -40°C to +120°C
T3 T130°C -40°C to +130°C
T5 T90°C -40°C to +75°C
T4 T120°C -40°C to +120°C
T3 T130°C -40°C to +130°C
T6 T75°C -40°C to +75°C
T5 T90°C -40°C to +90°C
T4 T120°C -40°C to +120°C
T3 T150°C -40°C to +150°C
T5 T90°C -40°C to +75°C
T4 T120°C -40°C to +120°C
T3 T150°C -40°C to +150°C
T5 T90°C -10°C to +75°C
T4 T120°C -10°C to +120°C
T3 T130°C -10°C to +130°C
T5 T90°C -10°C to +75°C
T4 T120°C -10°C to +120°C
T3 T150°C -10°C to +150°C
T6 T75°C -10°C to +75°C
T5 T90°C -10°C to +90°C
T4 T120°C -10°C to +120°C
T3 T130°C -10°C to +130°C
T6 T75°C -10°C to +75°C
T5 T90°C -10°C to +90°C
T4 T120°C -10°C to +120°C
T3 T150°C -10°C to +150°C
T5 T90°C -10°C to +75°C
T4 T120°C -10°C to +120°C
T3 T130°C -10°C to +120°C
T6 T75°C -10°C to +75°C
T5 T90°C -10°C to +90°C
T4 T120°C -10°C to +100°C
T3 T150°C -10°C to +100°C
Maximum Surface
Temperature
Temperature
Process
<2. Explosion Protection Type>
F0201.ai
Marking
F0202.ai
Marking
F0203.ai
Marking
6
6

2.2 Cable Entry

The type of cable entry is stamped near the cable entry port according to the following codes.
Marking Screw Size
M ISO M20 x 1.5 Female N ASME 1/2 NPT Female
Integral Flowmeter Remote Sensor
Remote Transmitter

2.3 Installation

Read the Installation Manual, IM 01E22A01-01EN (for AXG### and AXG4A) or IM 01E24A01-01EN (for AXW### and AXW4A), for basic installation procedure.
• All wiring shall comply with EN 60079-14, and local electric codes and requirements.
• Unused apertures shall be closed with suitable
certied blanking elements.
(The plug attached is certied.)
• If the magnetic owmeter is mounted in an area
in the presence of combustible dust, it shall be installed in such a way that the risk from electrostatic discharges and propagating brush
discharges caused by rapid ow of dust is avoided.
• The sensor is not surrounded by pipe insulation material.
• Cable glands, adapters and/or blanking elements with a suitable IP rating shall be of Ex db IIC/Ex tb
IIIC or Ex eb IIC/Ex tb IIIC certied by ATEX and shall be installed so as to maintain the specic
degree of protection (IP code) of the product.
• Take care the following warning marking “POTENTIAL ELECTROSTATIC CHARGING HAZARD”.
• In order to prevent the grounding conductor from loosening, the conductor must be secured to the terminal, tightening the screw with appropriate torque. Care must be taken not to twist the conductor.
• For multiple types of protection for terminal compartment, permanently mark the protection type installed. Once the type is marked, it cannot be changed.
• If the product is installed as the protection type Ex e,
terminate all the cable nish with crimp terminal of
a rod shape which of conductor length is 5 to 6 mm and cross section is 0.8 to 2.5 mm2, and connect them reliably.
• For the installation of multi protection type, tick the box of the selected type of protection type on the label in order to avoid confusion. e.g. In the case of selecting “db”, not “eb”
Ex db eb ia IIC T6...T3 Gb Ex tb IIIC T75°C...T130°C Db Terminal Compartment: Ex db Ex eb
<2. Explosion Protection Type>
F0204.ai
F0205.ai
F0206.ai
7
7
The grounding terminals are located on the inside and outside of the terminal area. Connect the cable to the grounding terminal in accordance with wiring procedure (a) or (b).
(a) Internal grounding terminal (b) External grounding terminal
Integral Flowmeter
(a)
(b)
Remote Sensor

2.4 Operation

When installation of explosion protection type product, read installation manual, IM 01E22A01­01EN (for AXG### and AXG4A) or IM 01E24A01­01EN (for AXW### and AXW4A).
Integral Flowmeter, Remote Sensor and Remote Transmitter
• Take care not to generate mechanical spark when access to the product and peripheral devices in hazardous locations.
• Take care the following warning marking “POTENTIAL ELECTROSTATIC CHARGING HAZARD”.
Take care the following warning marking when opening the cover.
Integral Flowmeter and Remote Transmitter
• Take care the following warning marking
“AFTER DE-ENERGIZING, DELAY 20 MIN­UTES BEFORE OPENING”.
(a)
Remote Transmitter
(a)
(b)
Remote Sensor
• Take care the following warning marking
“DE-ENERGIZING BEFORE OPENING”.

2.5 Maintenance and Repair

Only personnel authorized by Yokogawa Electric Corporation can repair the product. For maintenance of explosion protection type product, read maintenance manual, IM 01E22A01­02EN (for AXG### and AXG4A) or IM 01E24A01­02EN (for AXW### and AXW4A).
(b)
<2. Explosion Protection Type>
F0207.ai
F0209.ai

2.6 Name Plate

Example for name plates of Integral Flowmeter, Remote Sensor or Remote Transmitter.
Integral Flowmeter
*1
Remote Sensor
*1
F0208.ai
Remote Transmitter
8
8
*1
- MODEL: Specied model code
- SUFFIX: Sufx codes of the model code
- STYLE: Specied style code
- SIZE: Nominal size of apparatus
- METER FACTOR: Sensor constant number of apparatus
- SUPPLY: Power supply voltage of apparatus
- OUTPUT: Output signal of apparatus
- FLUID TEMP.: Fluid temperature of apparatus
- FLUID PRESS: Fluid pressure of apparatus
- AMB. TEMP.: Ambient temperature
- NO.: Manufacturing serial number
- No.: FM 17ATEX0031X:
ATEX type examination certicate number
-
Protection type and temperature class for gas Ex db eb ia IIC T6...T3 Gb Ex db IIC T6 Gb Ex db eb IIC T6 Gb:
- Protection type and maximum surface
temperature for dust
Ex tb IIIC T75°C...T130°C Db
Ex tb IIIC T75°C...T150°C Db
Ex tb IIIC T75°C Db:
- Um: Maximum r.m.s. a.c. or d.c. voltage
- ENCLOSURE: Enclosure protection code
- WARNING: Warning to apparatus
- : Name of manufacturer
*1: The product-producing country
<2. Explosion Protection Type>
2.7 Specic Condition of Use
Integral Flowmeter, Remote Sensor and Remote Transmitter
• Electrostatic charges on the non-metallic parts (excluding glass parts) or coated parts of the product shall be avoided.
• The ameproof joints differ from the standard
values in EN 60079-1. Only personnel authorized by the manufacturer of the product can repair the
ameproof joints.
Integral Flowmeter and Remote Sensor
• The applicable temperature class, ambient temperature range and process temperature range of the product are shown in Table 2.1 and Table 2.2.
• The property class of the fasteners used to fasten the transmitter enclosure onto the neck part is at least A2-70.
• In the case the electrodes and/or grounding rings are made of titanium, the sensor should be kept away from impacts and frictions in hazardous locations.
• From the safety point of view, the intrinsically safe circuit of the AXG and AXW series shall be assumed to be connected to earth.
• The product shall be installed as overvoltage category II.
9
9
Integral Flowmeter and Remote Transmitter
• Multiple types of protection for terminal compartment, permanently mark the protection type installed. Once the type is marked, it cannot be changed. For the installation of multi protection type, tick the box of the selected type of protection on the label in order to avoid confusion.

Revision Information

Title : ADMAG TI Series
AXG and AXW Magnetic Flowmeter ATEX Explosion Protection Type
Manual No. : IM 01E21A03-02EN
Edition Date Page Revised Item
1st Oct. 2018 New Publication
10
Loading...