Thank you for purchasing the Current Probe (Model 701930).
This instruction manual contains useful information about the instrument’s functions and
operating procedures as well as precautions that should be observed during use. To
ensure proper use of the instrument, please read this manual thoroughly before
beginning operation.
After reading the manual, keep it in a convenient location for quick reference whenever a
question arises during operation.
The following symbols are used in this manual to alert you to important information.
Affixed to the instrument. Indicates danger to personnel or
instrument and the operator must refer to the User's Manual. The
symbol is used in the User's Manual to indicate the reference.
WARNING
CAUTION
Note
Describes precautions that should be observed to prevent serious
injury or death to the user.
Describes precautions that should be observed to prevent minor or
moderate injury, or damage to the instrument.
Provides important information for the proper operation of the
instrument.
Checking the Contents of the Package
If the wrong instrument or accessories have been delivered, or if some accessories are
missing or defective, contact the dealer from which you purchased them.
Current Probe (Model 701930) Main Body
Accessories
• One user’s manual (this document).
• Carrying case
iiIM 701930-01E
Safety Precautions
This probe meets the Measurement Category II and III, and Pollution Degree 2
requirements as given in the IEC-61010-1 standards.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. If this instrument is used in a manner
not specified in this manual, the protection provided by this instrument may be impaired.
Also, Yokogawa Electric Corporation assumes no liability for the customer’s failure to
comply with these requirements.
The following symbols are used on this instrument.
To avoid injury, death of personnel or damage to the instrument, the operator
must refer to an explanation in the User's Manual or Service Manual.
Both direct and alternating current.
Do not apply around or remove from HAZARDOUS LIVE conductors.
IM 701930-01E
iii
Precautions
In order to ensure safe operation and to obtain maximum performance from the
unit, observe the cautions listed below.
WARNING
• To avoid short circuits and accidents that could result in injury or death, use this
current probe only with power lines carrying 600 V or less (CATII), or 300 V or
less (CATIII).
• Also to avoid short circuits and personal injury when the clamp is open or
measurement is being performed, do not apply the clamp to any bare
conductors.
• When applying the clamp or taking measurements, don’t touch the clamp
anywhere from the edge of the barrier up to the conductor being measured.
• Be careful not to damage the insulation surface while taking measurements.
• This instrument is made for use with the Power Supply (Model 700938 or
701934). It is possible to use a power supply other than the 700938 or 701934,
provided that the connector and pin assignments match, and that voltage and
other electrical specifications are satisfied. Regardless of which power supply is
used, for safety, make sure that it has a
insulation construction.
• Make sure that the waveform measuring equipment connected to this unit’s
output terminal (BNC) is equipped with a power supply having a
earthing with double-insulation construction.
• If the waveform measuring instrument connected to this unit’s output terminal
(BNC) is equipped with any other input terminal, take the following precautions
to ensure that the other instrument does not form a bridge between the probe
and any
1 Isolate the terminal to which the probe is connected from other terminals on
2 If
3 Make sure to ground the measuring instrument.
4 Beware of electric shock when connecting the probe to the item being
5 Before connecting the probe, check that the measuring instrument and power
6 Read and observe all warnings and precautions relating to electrical safety
• To prevent electric shock, do not allow the unit to become wet and do not use
the unit when your hands are wet.
hazardous live parts.
the measuring instrument using
overvoltage category, working voltage, and pollution degree requirements of
the circuit being tested.
basic insulation requirements cannot be met between the terminal to which
this unit is connected and other terminals of the measuring instrument, make
sure that the voltage input to the measurement terminal does not exceed the
safe voltage level(SELV-E)*.
measured. Do not remove the probe from the instrument while the probe is
connected to the item.
supply are properly grounded and that the output connector and power plug
are connected to the BNC connector and power receptacle, respectively.
for the measuring instrument being connected to the probe.
Refer to the following standards regarding the meanings of underlined terms.
IEC61010-1
IEC61010-2-031
IEC61010-2-032
protective earthing with double-
protective
basic insulation conforming to the
* Do not exceed the following levels: 30 Vrms, 42.4 Vpeak, and 60 VDC.
ivIM 701930-01E
CAUTION
• Do not subject the unit to vibrations or shocks during transport or handling. Be
especially careful to avoid dropping the unit.
• Do not store or operate the unit where it will be exposed to direct sunlight, high
temperature, high humidity, or condensation. If exposed to such conditions, the
unit may become damaged, the insulation may deteriorate, and the unit may no
longer satisfy its specifications.
• Before using the unit, inspect it and check the operation to make sure that it was
not damaged due to poor storage or transport conditions. If damage is found,
contact your dealer or YOKOGAWA representative.
• This instrument is not waterproof or dustproof. Do not use the unit in a wet or
dusty environment.
• The sensor head is a precision assembly including a molded component, a
ferrite core, and a Hall effect element. It may be damaged if subjected to
sudden changes in ambient temperature, or mechanical strain or shock, and
therefore great care should be exercised in handling it.
• The matching surfaces of the sensor head are precision ground, and should be
treated with care. If these surfaces are scratched, performance may be
impaired.
• If there is any type of dust or dirt on the matching surfaces of the sensor head,
measurements may be affected. Wipe it away gently with a soft cloth.
• In order to avoid damaging the sensor cables do not bend or pull the sensor
cable or power supply cable.
• Gently wipe dirt from the surface of the unit with a soft cloth moistened with a
small amount of water or mild detergent. Do not try to clean the unit using
cleaners containing organic solvents such as benzene, alcohol, acetone, ether,
ketones, thinners, or gasoline. They may cause discoloration or damage.
• When the power is ON, keep the core section of the sensor closed, except when
clamping it onto the conductor to be measured. The facing surface of the core
section can be scratched while it is open.
IM 701930-01E
Note
Accurate measurement may be impossible in locations subject to strong external magnetic
fields, such as near transformers and high-current conductors, or in locations subject to
strong external electric fields, such as near radio transmission equipment.
v
viIM 701930-01E
Contents
1
Foreword ............................................................................................................................ i
Checking the Contents of the Package ............................................................................. ii
Safety Precautions ............................................................................................................ iii
Corrective Action to be Taken in Case of an Abnormality .............................................. 2-1
IM 701930-01E
vii
Chapter 1701930 Current Probe
Product Overview
This unit can be directly connected to a BNC input connector of a waveform measuring
instrument such as a digital oscilloscope or a DL Series instrument, and by clamping it to
a conductor to be measured, allows the current waveform to be easily captured.
Features
• Highly accurate current detection
• Easy current measurement
• Wideband frequency characteristics; DC to 10 MHz
• Wide clamp permits measurement of high current levels
• Easy protect function during excessive input
1
701930 Current Probe
IM 701930-01E
1-1
Names of Parts
3
1
4 (internal)
Current Direction Indication
2
5
UNLOCK Indication
LOCK Indication
Sensor Cable
2
Terminator
6
8
9
Sensor
Power Supply Cable
7
1-2
IM 701930-01E
Description of Parts
1 Clamp
The part which clamps to the conductor to be measured.
2 Clamp Lever
The lever for opening and closing the clamp. Make sure to open or close the clamp
only by using the clamp lever.
3 Clamp Lock
Locks the clamp in place when it is closed.
4 Sensor Head
This clamps the conductor being measured, and carries out the actual current
measurement. It is a precision assembly including a molded component, a ferrite
core, and a Hall effect element. It may be damaged if subjected to sudden changes in
ambient temperature, or mechanical strain or shock, and therefore great care should
be exercised in handling it.
5 Barrier
A piece which indicates the limit up to which the user may safely touch the clamp, and
reduces the danger of touching the conductor during testing. Do not touch any part of
the clamp beyond the barrier during measurement.
6 Demagnetizing Switch (DEMAG)
If the core has been magnetized by switching the power ON and OFF, or by an
excessive input, this switch demagnetizes it. Always carry out demagnetizing before
measurement. The demagnetizing process takes about three seconds. During
demagnetizing, a demagnetizing waveform is output.
7 Zero Adjustment Dial (ZERO ADJ)
Use the zero adjustment dial to correct for the effect of a voltage offset or temperature
drift on the unit. When performing measurements, always carry out a zero adjustment
after demagnetizing.
8 Output Connector
The current waveform of the measured conductor is output at a constant rate (0.01
VA in conjunction with the waveform measuring instrument with an input impedance
of 1 MΩ.)
Connect to the BNC input connector of the waveform measuring instrument.
1
701930 Current Probe
IM 701930-01E
Note
• The output of this unit is terminated internally. Use a high-impedance input to the
measuring instrument. With an input impedance of 50 Ω, accurate measurement is not
possible.
• If using BNC-banana plug or similar adapters to connect to input terminals other than BNC
connectors, make sure the polarity is correct.
• Turn the clamp lever until it clicks, and check that it is locked securely.
9 Power Plug
Connect the power plug to the receptacle on the Power Supply (Model 70938 or
701934) to supply power to the sensor and terminator.
1-3
Measurement Procedure
Notes on Use
• To avoid short circuits and accidents that could result in injury or death, use this
• Also to avoid short circuits and personal injury when the clamp is open or
• When applying the clamping or taking measurements, don’t touch the clamp
• Be careful not to damage the insulation surface of the conductor being
• This instrument is made for use with the Power Supply (Model 700938 or
• Make sure that the waveform measuring equipment connected to this unit’s
• If the waveform measuring instrument, which is connected to this unit’s output
• To prevent electric shock, do not allow the unit to become wet and do not use
WARNING
current probe only with power lines carrying 600 V or less (CATII), or 300 V or
less (CATIII).
measurement is being performed, do not apply the clamp to any bare
conductors.
anywhere from the edge of the barrier up to the conductor being measured.
measured while taking measurements.
701934). It is possible to use a power supply other than the 700938 and 701934,
provided that the connector and pin assignments match, and that voltage and
other electrical specifications are satisfied. Regardless of which power supply is
used, for safety, make sure that it has a
construction.
output terminal (BNC) is equipped with a power supply having a
earthing with double-insulation construction.
terminal (BNC) is equipped with any other input terminal, take the following
precautions to ensure that the other instrument does not form a bridge between
the probe and any
1 Isolate the terminal to which the probe is connected from other terminals on
the measuring instrument using
overvoltage category, working voltage, and pollution degree requirements of
the circuit being tested.
2 If
basic insulation requirements cannot be met between the terminal to which
this unit is connected and other terminals of the measuring instrument, make
sure that the voltage input to the measurement terminal does not exceed the
safe voltage level(SELV-E)*.
3 Make sure to ground the measuring instrument.
4 Beware of electric shock when connecting the probe to the item being
measured. Do not remove the probe from the instrument while the probe is
connected to the item.
5 Before connecting the probe, check that the measuring instrument and power
supply are properly grounded and that the output connector and power plug
are connected to the BNC connector and power receptacle, respectively.
6 Read and observe all warnings and precautions relating to electrical safety
for the measuring instrument being connected to the probe.
Refer to the following standards regarding the meanings of underlined terms.
IEC61010-1
IEC61010-2-031
IEC61010-2-032
the unit when your hands are wet.
hazardous live parts.
protective earthing with double-insulation
protective
basic insulation conforming to the
1-4
* Do not exceed the following levels: 30 Vrms, 42.4 Vpeak, and 60 VDC.
IM 701930-01E
Preparations for Measurement
CAUTION
• Before turning ON the power, make sure that the voltage of the power supply
being used matches the supply voltage indicated on the rear panel of the
700938 or 701934.
• When using the Current Probe (Model 701930), you can only connect as many
active probes simultaneously to the Power Supply (Model 700938 or 701934) or
the DL Series rear panel probe power supply connector as allowed depending
on the current value of the conductor.
The current consumption of the Current Probe (Model 701930) is dependent
upon the measured current value. Do not let each active probe’s total
current consumption exceed the current regulated by the Power Supply (Model
700938) or the DL Series rear panel probe power supply connector (see figure
below).
1
701930 Current Probe
Current Consumption (mA)
Current to be measured (A)
Current consumption* vs. current to be measured(typical)
* The sum total of a positive and negative current consumption.
1. Have the Power Supply (Model 700938 or 701934), and oscilloscope or recorder for
waveform measurement ready.
2. Turn the power switch OFF and connect the power cord.
3. Connect the power plug of the 701930 to the power receptacle of the 700938 or
701934.
4. Turn the 700938 or 701934 power switch ON, and check that the front panel power
indicator lights.
Note
The output of this unit is terminated internally. Use a high-impedance input to the measuring
instrument. With an input impedance of 50 Ω, accurate measurement is not possible. When
power is supplied to the 701930, a demagnetizing waveform might be generated.
IM 701930-01E
1-5
Demagnetizing and Zero Adjustment
CAUTION
• When disconnecting the output connector, be sure to release the lock before
pulling the connector. Forcibly pulling the connector without releasing the lock,
or pulling on the cable will result in damage to the terminator.
• If using BNC-banana plug or similar adapters to connect to input terminals other
than BNC connectors, make sure the polarity is correct.
• Do not demagnetize when the clamp is connected to a conductor. Doing so
might damage the components of the circuit being measured.
• Do not begin supplying power to the 701930 after the clamp has been
connected to a conductor. Doing so might damage the components of the
circuit being measured.
1. With the waveform measurement instrument input at ground, adjust the trace to the
zero position.
2. Set the input coupling of the waveform measurement instrument to DC.
3. Connect the output connector of the 701930 to the input connector of the waveform
measurement instrument. Turn the clamp lever until it clicks, and check that it is
locked securely.
1-6
4. Without clamping the conductor to be measured, press the clamp lever until the
LOCK indication appears (and the UNLOCK indication disappears), and then check
that the clamp is properly closed.
5. Press the demagnetizing switch (DEMAG) on the terminator.
6. Turn the zero adjustment dial on the terminator to adjust the trace to the zero position.
Note
The output of this unit is terminated internally. Use a high-impedance input to the measuring
instrument. With an input impedance of 50 Ω, accurate measurement is not possible.
IM 701930-01E
Measurement Procedure
CAUTION
• The maximum continuous input range is based on heat that is internally
generated during measurement. Never input current in excess of this level.
Doing so may result in damage to the probe.
• The maximum continuous input range varies according to the frequency of the
current being measured. See page1-9, “Product Specifications.”
• If excess current is input, generated heat activates a built-in safety function that
blocks normal output. If this happens, remove the input immediately (remove
the sensor from the conductor being measured, or reduce the input current to
zero). Wait until the sensor has had sufficient time to cool before resuming
operation.
• If you continuously input a current even if it is below the maximum current value
established by the maximum continuous input range, the self-heating of the
sensor may cause the safety function of the instrument to activate in order to
prevent damage to the instrument.
• At high ambient temperatures, the built-in safety circuit may activate at current
input levels below the rated continuous maximum.
• Continuous input of current exceeding the rated maximum, or repeated
activation of the safety function may result in damage to the unit.
• The probe is rated for maximum input under two conditions in addition to the
input maximums shown in the Product Specifications. These are: (1) 300
Apeak, for non-continuous input, and (2) 500 Apeak at Pulse Width ≤ 30 µs.(1)
indicates an upper waveform response limit of 300 Apeak. Use the sensor at
RMS current input levels that are within the rated continuous maximums. (2)
indicates the upper response limit for a single input pulse. Use the sensor at
levels not exceeding this value.
• Open the clamp only by using clamp lever.
1
701930 Current Probe
IM 701930-01E
1. Check that the system is safe, and that the preparations described in the preceding
section have been carried out.
2. Pull the sensor opening lever so that the sensor head opens.
3. Align the sensor so that the current direction indication corresponds to the direction of
current flow through the conductor to be measured, and clamp so that the conductor
is in the center of the sensor aperture.
1-7
4. Press the clamp lever on the sensor head until the LOCK indicator appears (and the
UNLOCK indication disappears), and check that the opening lever is firmly locked and
the clamp is securely closed.
5. It is now possible to monitor the current waveform. The output voltage rate of the
701930 is 0.01 V/A. Convert the voltage sensitivity of the waveform observation
instrument to current sensitivity. For example, if the voltage sensitivity is 10 mV/DIV,
the current sensitivity is 1 A/DIV.
Note
• The output of this unit is terminated internally. Use a waveform measurement instrument
with an input impedance of at least 1 MΩ.
• Immediately after powering on, this unit may be subject to an appreciable offset drift due
to the effect of self-heating. To counteract this, allow the unit to warm up for about 30
minutes before carrying out measurement.
• When performing continuous measurements, be aware that the offset voltage drifts
depending on factors such as the ambient temperature.
• Under certain circumstances, oscillation may occur if the probe is connected to the Power
Supply (Model 700938 or 701934) while the power supply is ON. In such cases,
oscillation can be stopped and operation restored to normal by opening and closing the
clamp with the clamp lever.
• Depending on the measured current value or its frequency, resonant sounds may be
generated from the clamp. Unless there is dirt or foreign matter on the matching surfaces
of the sensor head, this will have no effect on measurement.
• The reading may be affected by the position within the clamp aperture of the conductor
being measured. The conductor should be in the center of the clamp aperture.
• When carrying out measurement, press the clamp lever until the LOCK indication appears
(and the UNLOCK indication disappears), and check that the clamp is properly closed. If
the clamp is not properly closed, accurate measurement will not be possible.
• At high frequencies, common mode noise may affect measurements taken on the high
voltage side of circuits. If this occurs, reduce the frequency range of the waveform
measuring instrument, or clamp onto the low-voltage side of the circuit, as appropriate.
H
Power sourceLoad
L
• Directly after powering ON, the demagnetizing waveform appears in the output; this is not
a malfunction.
• Accurate measurement may be impossible in locations subject to strong external magnetic
fields such as transformers and high-current conductors, or in locations subject to strong
external electric fields such as radio transmission equipment.
1-8
IM 701930-01E
Specifications
Product Specifications
Accuracy is guaranteed at 23 ± 3 °C (73°F± 5° F) after the power has been on for 30
minutes.
Band width*DC to 10 MHz (–3 dB)
(Typical characteristics shown in Page1-10, Fig.1)
Rise Time*35 ns or less
Maximum Continuous Input Range
150 A(Derating according to frequency shown in Page1-10,
Fig.2)
Maximum Peak Current Value
Non-Continuous 300 Apeak, or 500 Apeak at Pulse Width
≤ 30 µs.
Output Voltage Rate*0.01 V/A
Amplitude Accuracy*To 150 A: ±1%rdg. ±1 mV
150 A to 300 Apeak: ±2% rdg.
(DC, and 45 to 66 Hz)
Noise*Equivalent to 25 mArms or less (for 20 MHz band
measuring instrument)
Insertion Impedance(Typical characteristics shown in Page 1-10, Fig.3)
Temperature Coefficient for Sensitivity*
±2% or less (within a range of 0 to 40°C or 32 to 104°F
when inputting 55 Hz, 150 A)
Maximum Rated Power5.5 VAmax (at the maximum continuous input range)
Rated Supply Voltage±12 V ±1 V
Operating Temperature and Humidity Range
0 to 40°C (32 to 104 °F), 80% RH or less (no condensation)
Storage Temperature and Humidity Range
–10 to 50°C (14 to 122°F), 80% RH or less (no
condensation)
Location for UseIndoor, altitude up to 2000 m (6562 feet)
Effect of External Magnetic Fields
Equivalent to a maximum of 150 mA (in a DC or 60 Hz, 400
Corrective Action to be Taken in Case of an Abnormality
If maintenance service is required, or if the instrument still does not operate properly
even after proper corrective action has been taken, contact your nearest YOKOGAWA
representatives listed on the back cover of this manual.
Symptom Possible CauseCorrective Action
Power is not turned ON.Turn ON the power.
Cannot make DC measurements
(or at low frequencies up to few
hundred Hertz) or the amplitude
is small in the given frequency
region.
Cannot adjust to the zero level
using the zero adjustment dial.
The amplitude is small across
all frequencies.
The measurement instrument,
such as an oscilloscope, is set
to AC coupling.
The sensor is not locked
(it is not hooked securely).
The sensor is magnetized.Demagnetize the sensor
The input impedance of the
measurement instrument,
such as an oscilloscope, is
set to 50Ω.
2
Malfunction? First, Investigate.
Set the instrument to DC
coupling.
Lock the sensor.
before adjusting the zero level.
Set it to 1MΩ.
2-1IM 701930-01E
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.