Yokogawa 2406E Data sheet

Yokogawa 2406E Data sheet

2406E Series of Analog Insulation Testers

 

SERIES

240631

240632

240633

240634

240635 240641

240642

240643

240644

240645

 

Analog models with two and three ratings

AC voltage measurement

 

Automatic discharge

 

 

Sky blue EL backlight

 

Testing Performance Specifications

Model

Suffix

 

Rating

Effective

Central

AC Voltage

Lower limit of

Rated Current

 

Code

 

 

 

Measuring range

Scale Value

Measuring range

measured

 

240631

-E

 

25V/5MΩ

0.001–5MΩ

0.1MΩ

0–300V

0.025MΩ

1mA

240641

-E

 

50V/10MΩ

0.005–10MΩ

0.2MΩ

 

0.05MΩ

1mA

 

 

 

 

125V/20MΩ

0.01–20MΩ

0.5MΩ

 

0.125MΩ

1mA

240632

-E

 

125V/20MΩ

0.01–20MΩ

0.5MΩ

0–300V

0.125MΩ

1mA

240642

-E

 

250V/50MΩ

0.01–50MΩ

1MΩ

 

0.25MΩ

1mA

240633

-E

 

125V/20MΩ

0.01–20MΩ

0.5MΩ

0–600V

0.125MΩ

1mA

240643

-E

 

250V/50MΩ

0.01–50MΩ

1MΩ

 

0.25MΩ

1mA

 

 

 

 

500V/100MΩ

0.05–100MΩ

2MΩ

 

0.5MΩ

1mA

240634

-E

 

250V/50MΩ

0.01–50MΩ

1MΩ

0–600V

0.25MΩ

1mA

240644

-E

 

500V/100MΩ

0.05–100MΩ

2MΩ

 

0.5MΩ

1mA

 

 

 

 

1000V/2000MΩ

1–2000MΩ

50MΩ

 

1MΩ

1mA**

240635

-E

 

250V/500MΩ

0.1–500MΩ

10MΩ

0–600V

0.25MΩ

1mA**

240645

-E

 

500V/1000MΩ

0.5–1000MΩ

20MΩ

 

0.5MΩ

1mA**

 

 

 

 

1000V/2000MΩ

1–2000MΩ

50MΩ

 

1MΩ

1mA**

 

 

 

 

 

* The minimum value at which the rated voltage can be maintained;

EL-backlit

 

Non-backlit

 

 

 

 

 

** 0.55 mA in the case of the first effective measuring range

Standard test conditions:

Ambient temperature/humidity ranges: 23 ±5:/45-75% RH Position of use: Horizontal (5° max. of angle of inclination)

External magnetic fields: None

Battery voltage: Within effective voltage range

(The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)

Tolerances under the above-mentioned conditions:

Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading

Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value

No-load voltage: 130% max. of rated voltage

Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range

Short-circuit current: 12 mA max.

General Specifications

Scale length: Approx. 86 mm (outer scale)

Discharge function: The tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (∞) scale value. Under this condition, the tester is ready to enter voltage measurement mode.

AC voltage measurement: AC voltage measurement is possible wherever the rotary switch is positioned.

Dimensions (main unit): Approx. 120 (W) 110 (H) 60 (D) (mm)

Weight: Approx. 500 g (including batteries)

Batteries: Six AA (R6P) batteries

Accessories: See the list of accessories on the backside of this bulletin for information on accessories.

Standard Accessories

Product

Part Number

Qty

Remarks

Earth and Line probes

98007

1

Earth probe(blake);approx. 1m long

Line probe(vermilion);approx. 1m long

 

 

 

Carrying case

B9075MU

1

w/probe-housing

pocket and neck strap

 

 

 

User's manual

1

Batteries

6

External Dimensions

 

Unit: mm

 

 

 

120

53

7

 

110

 

B9108XA

B9705MU

Accessory-housing case

Hard case