Yokogawa 2406E Data sheet

2406E Series of Analog Insulation Testers
240631 240632 240633 240634 240635 240641 240642 240643 240644 240645
Analog models with two and three ratings
AC voltage measurement
Automatic discharge
Sky blue EL backlight
Testing Performance Specifications
Model
240631 240641
240632 240642 240633 240643
240634 240644
240635 240645
Suffix
Code
-E
-E
-E
-E
-E
-E
-E
-E
-E
-E
Rating
25V/5M
50V/10M 125V/20M 125V/20M 250V/50M 125V/20M 250V/50M
500V/100M
250V/50M
500V/100M
1000V/2000M
250V/500M
500V/1000M
1000V/2000M
Non-backlitEL-backlit
Effective
Measuring range
0.001–5M
0.005–10M
0.01–20M
0.01–20M
0.01–50M
0.01–20M
0.01–50M
0.05–100M
0.01–50M
0.05–100M 1–2000M
0.1–500M
0.5–1000M 1–2000M
* The minimum value at which the rated voltage can be maintained; ** 0.55 mA in the case of the first effective measuring range
Central
Scale Value
0.1M
0.2M
0.5M
0.5M 1M
0.5M 1M 2M 1M 2M
50M 10M 20M 50M
Lower limit of
AC Voltage
Measuring range
0–300V
0–300V
0–600V
0–600V
0–600V
measured
0.025M
0.125M
0.125M
0.125M
0.05M
0.25M
0.25M
0.5M
0.25M
0.5M 1M
0.25M
0.5M 1M
Rated Current
1mA 1mA 1mA 1mA 1mA 1mA 1mA 1mA 1mA 1mA 1mA** 1mA** 1mA** 1mA**
Standard test conditions:
Ambient temperature/humidity ranges: 23 ±5:/45-75% RH
Position of use: Horizontal (5° max. of angle of inclination)
External magnetic fields: None
Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerances under the above-mentioned conditions:
Resistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.
SERIES
General Specifications
Scale length: Approx. 86 mm (outer scale)
Discharge function: The tester automatically begins discharge when you turn off the
MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite () scale value. Under this condition, the tester is ready to enter voltage measurement mode.
AC voltage measurement: AC voltage measurement is possible wherever the rotary switch is positioned.
Dimensions (main unit): Approx. 120 (W) ҂ 110 (H) ҂ 60 (D) (mm)
Weight: Approx. 500 g (including batteries)
Batteries: Six AA (R6P) batteries
Accessories: See the list of accessories on the backside of this bulletin for information
on accessories.
Standard Accessories
Product Part Number Qty
Earth and Line probes
Carrying case
Batteries
98007
B9075MU
Earth probe(blake);approx. 1m long
1
Line probe(vermilion);approx. 1m long
w/probe-housing
1
pocket and neck strap
1
6