Yamato Scientific WG221S User Manual

Page 1
Demineralizers
Still, water, models
WG221
WG221S
First Edition
Thank you very much for buying Yamato Scientific AUTO STILL series WG221/221S model.
Warning:
signs in this instruction manual prior to use.
Read and apprehend the important warning
Yamato Scientific CO.,LTD
Page 2
Contents
For Safety Use................................................................................................................................................1
Explanation of Graphic Indication...............................................................................................1
Table of Symbol Mark..................................................................................................................2
Caution for Safety........................................................................................................................3
Table of Dangerous Element.......................................................................................................5
Safety Precaution ..........................................................................................................................................6
Installation and Preparation for Use............................................................................................6
Caution for Use............................................................................................................................................16
Identification of Parts..................................................................................................................................17
Body...........................................................................................................................................17
Piping Diagram..........................................................................................................................18
Operation System......................................................................................................................19
Operation Panel.........................................................................................................................22
Emergent Troubleshooting ........................................................................................................................27
Run Menu .....................................................................................................................................................28
Preparation and Check before Use...........................................................................................28
Run Menu ..................................................................................................................................29
Repeated Collection of Pure Water ..........................................................................................30
Collection of The Fixed Volume of Ion Exchanged Water........................................................31
Collection of The Fixed Volume of Distilled Water(WG221S)..................................................32
Indication of Water Quality........................................................................................................33
Display of Total Flow/Resetting.................................................................................................35
Safety Devices and Error Codes................................................................................................................36
Displays and Details of Error Code...........................................................................................36
Trouble Displays other than Error Codes/Details .....................................................................37
Troubleshooting.........................................................................................................................38
Maintenance.................................................................................................................................................39
Check and maintenance ...........................................................................................................39
Exchange of Pre-Treatment Cartridge......................................................................................39
Exchange of Ion Exchange Resin Cartridge.............................................................................39
Cleaning Still..............................................................................................................................40
Exchange Heaters.....................................................................................................................44
Clean Water Supply Hose Filter................................................................................................45
Exchange Hoses .......................................................................................................................45
When you do not use the unit for a long time...........................................................................46
After-sales Service and Warranty..............................................................................................................47
Specifications ..............................................................................................................................................48
Wiring Diagram ............................................................................................................................................49
WG221/221S.............................................................................................................................49
Page 3
For Safety Use
Explanation of Graphic Indication
This instruction manual and our products apply various graphic indications for safety use. Ignoring these indicators or improper handling of the unit can cause such situations as indicated below. Please read and apprehend the following contents.
Warning
Caution
(1) Serious injuries mean any injuries , electric shocks, bone fractures and poisoning causing after effects or requiring hospitalization or long outpatient t r eat ment (2) Injuries mean any injuries and electric shocks not requiring hospitalization or long outpatient treatment. (3) Damages on property mean any damages on facilities, machiner y, buildings, etc.
Warning indicates possibility of serious injuries (1) or death.
Caution indicates possibility of injuries (2) on person or damages on property ( 3).
Meaning of graphic indications
It shows conducts to be surely done.
It shows warnings or cautions.
It shows prohibited conducts.
‑1‑
Page 4
Table of Symbol Mark
Warning
For Safety Use
General Warning
Caution
General Caution
Water only
Prohibition
Warning of High Potential voltage
Caution of electric shock
Caution of deadly
poison
Warning of High Potential voltage
Caution of scald
Warning of drive
section
Caution of
water boiling
low-
Warning of
explosion
Caution of water
leak
General Prohibition Prohibition of fire
Compulsion
General
Compulsion
Ground the unit
Prohibition of disassembling
Install on level
area
Prohibition of
touch
Pull electric power
plug
Fixed time check
or maintenance
‑2‑
Page 5
Caution for Safety
Warning
Do not use this unit in flammable or explosive gas environments.
This unit is not explosive proof. Never use this unit in flammable or explosive environments to prevent fire or explosion. See Page5 for dangerous element → Table of Dangerous Element
Never fail to ground the unit.
Connect with the grounded outlet. If you do not have grounded outlet, use the adapter for grounding that is non-standard accessories, and then be sure to ground the lead.
Do not use this unit if malfunction occurs.
If smoke or any strange odor should disburse from the unit, switch off power to prevent fire or electric shock.
Do not bundle the power cable during use.
Overheat or fire can occur if the power code is bundled in use of the unit.
For Safety Use
Do not damage the power code.
Forcibly bending, pulling or wrenching damages the power code, and can cause a fire or electric shock.
Do not disassemble or remodel this unit.
Disassembling this unit can cause a fire, electric shock or other crisis. Never disassemble or remodel this unit.
Do not touch hot area.
Pay attention not to get scalded. Some areas of the boiler are hot during or after operat ion. Make sure if the boiler is cool enough before you start maintenance of the heater or other devices.
Fasten the tap when you do not use the unit
When you do not use the unit (in the nighttime or holidays) , certainly fasten the tap to prevent the accident of water leak .
‑3‑
Page 6
Caution for Safety
Caution
If it begins to thunder.
If it begins to thunder, switch the power off. Neglecting this can result in fire or electric shock.
Be careful for using the cleanser (Organosol).
The main ingredients of the cleanser (organosol) is sulfamic acid and the solution is about PH1. Wear protective gloves, mask and glasses to handle the cleanser. Wash immediately by pure water the cleanser splashing on body.
For Safety Use
‑4‑
Page 7
Table of Dangerous Element
Nitroglycol, Nitroglycerin, Nitrocellulose, and other explosive nitric esters.
Trinitrobenzene, Trinitrotoluene, Picric acid, and other explosive nitro compounds. Peracetic acid, Methyl ethyl ketone peroxide, Benzoyl peroxide, and other organic peroxides. Sodium azide, and other metallic azides Metallic lithium, Metallic potassium, Metallic sodium, Yellow phosphorus, Phosphorus sulfide, Red phosphorus, Celluloid, Calcium carbide, Lime phosphate, Magnesium powder, Aluminum powder, and other combustible metal powders and sodium dithionite (hydrosulfite). Potassium chlorate, Sodium chlorate, Ammonium chlorate, and other chlorates. Potassium perchlorate, Sodium perchlorate, Ammonia perchlorate, and other perchlorates. Potassium peroxide, Sodium peroxide, Barium peroxide, and other inorganic peroxides.
Potassium nitrate, Sodium nitrate, Ammonia nitrate, and other nitrates.
Explosive
Explosive
Substance
Combustible
Substance
Oxidant
For Safety Use
Flammable
Ignitable
Substance
Combustible
Gas
Sodium chlorite and other chlorites.
Calcium hypochlorite and other hypochlorites. Ethyl ether, Gasoline, Acetaldehyde, Propylene Oxide, Carbon
disulfide, and other flammable substances with a flash point below minus 30ºC. Normal hexane, Ethylene oxide, Acetone, Benzene, Methyl ethyl ketone, and other flamm able substances with a flash point between minus 30ºC and 0 Methanol, Ethanol, Xylene, Pentyl acetate (amyl acetate), and other flammable substance with a flash point between 0ºC and 30ºC. Kerosene, Light oil, Turpentine oil, Isoamyl alcohol, Acetic acid, and other flammable substances with a flash point between 30ºC and 65ºC Hydrogen, Acetylene, Ethylene, Methane, Ethane, Propane, Butane and other flammable gas at 15℃ degree and under 1 atmosphere.
o
C.
‑5‑
Page 8
Installation and Preparation for Use
Warning
Do not use this unit in flammable or explosive gas environments.
This unit is not explosion proof. Never use this unit in flammable or explosive gas
environments. Electric arc occurs when you switch the unit “ON” and “OFF”, and can cause a fire or explosion.
See Page 5 (Table of dangerous element) about flammable or explosive gas.
Safety Precaution
Select carefully the space to install the unit.
Do not install this unit in any of the following environments.
In inflammable or corrosive gas environments
In temperature beyond 35
In severe change in temperature
In very humid locations[
Under direct sun
In frequently vibrating areas
Provide ample space surrounding the unit as indicated below. Install the unit in the place with t he sink facilities if you can.
‑6‑
Page 9
Installation and Preparation for Use
Warning
Do not remodel this unit.
Never disjoint this unit to avoid electric shock.
There are some high-pressure areas inside the unit. Ask your seller or Yamato Scientific’s
Technical Service Department for internal adjustment or repair.
You must follow this instruction manual for your daily check or maintenance. Never remodel
the unit by yourself to prevent any troubles.
Safety Precaution
Install this unit on level areas.
●
If you do not, unexpected trouble can occur.
‑7‑
Page 10
Safety Precaution
Installation and Preparation for Use
Warning
Ground the unit.
If you do not ground the unit, the earth leakage circuit breaker would not work in case of electric leakage. Be sure to connect the ground lead to prevent electric shock.
Ground the ground lead to the ground wire or terminal of the power source. Contact an electrician if you have no equipment for grounding.
Never connect with gas or water pipe.
Be
●
sure to ground the lead wire when you use the adapter of non-standard accessories for
grounding.
Use exclusively the outlet.
Use an outlet with adequate capacity(more than 15 A).
If you connect to those without enough capacity, you would not take enough distilled water or
control properly the unit due to decrease of voltage from the power source. Connect the power code to the sufficient power source.
Necessary capacity: AC100V single-phase 15A
Precaution for connecting the power cord.
Check certainly that the circuit breaker of power source is “off “ and connect the power cord
for this unit. Power plug of this unit is three core cable. So the plug has grounding line. If you do not have adaptable outlet (See Page2), Use the adapter for grounding that is non­standard accessories. If you use it, be sure to ground the lead wire.
Precaution for proper use of the power cord
●
Do not bundle the power cord during operation, or overheat or fire can occur.
●
Do not process, or bend, wrench or pull forcibly the cord. Such conduct can cause a fire or
electric shock.
●
Do not damage the cord, by placing a desk or chair on it or placing it between objects. Such
conducts can result in fire or electric shock.
●
Do not place the cord near any stoves or heaters, or the cable coating can melt causing a fire
or electric shock.
If the power cord is damaged (core exposure or disconnection), turn immediately the circuit
breaker and the power supply off, and request the seller to change the power cord. Neglecting this procedure can cause a fire or electric shock.
●
Connect the cord to an exclusive outlet.
‑8‑
Page 11
Installation and Preparation for Use
Warning
Connect securely the water supply hose.
If you do not connect the joint securely, the water-supply hose or connector would come off,
causing water leak or spouting.
Use the connector and water-supply hose out of accessories for this unit. Install the unit on
the level and stable place near a tap and drainage.
Connect securely the water-supply hose in the following procedure.
Connect to a tap.
(1) Slide the joint sleeve of the socket ③ to the direction the arrow shows. Then you can
separate the connector and the hose. (2) Loosen the plug ② from the ring ①. (3) Fasten 4 screws ⑤ in the same way while pushing the ring ① lightly and evenly so that the
packing ⑥ touches evenly the tap. (4) Turn right and fasten securely the plug ②.Then the packing ⑥ seals the tap and connector. (5) Slide the sleeve to the direction the allow shows and insert the socket ③ securely into the
plug ②. Make sure the sleeve goes back to the original position when you loose your hold.
Then you can finish the connection.
Safety Precautions
‑9‑
Page 12
Installation and Preparation for Use
Warning
Connection to the unit
(1) Remove the rubber cap from the plug ⑧. (2) Slide the sleeve to the direction the arrow
shows, and insert securely the socket
into the plug ⑧.
Make sure the sleeve returns to the original
position when you loose your hold. Then,
you can finish the connection.
The socket has a built-in valve which would
not open for water run until the plug is
connected.
Connect the water-supply hose to a tap with drainage system.
Connect the hose to a tap with drainage system. If you connect the water-supply hose to a tap without drainage system, damage on the hose could result in overflow of water.
Safety Precautions
‑10‑
Page 13
Installation and Preparation for Use
Caution
If drainage facility is away from the tap, use the Water Supply Unit, a non­standard accessory.
The “Water Supply Unit” includes all items from the sleeve to the joint. Refer to the instruction manual for the “Water Supply Unit” about connection to the faucet and for more information.
The “Water Supply Unit” has the structure to keep tight connection to the pipe compared to a normal hose set even when the water pressure changes.
Safety Precautions
Joint
Joint
Keep the following pressure of water service.
Keep water pressure within 0.5 to 5×100kPa (0.5 to 5kgf/cm2) even for use at night.
Keep the same water pressure even when you use the non-standard accessory “Water Supply Unit”,
‑11‑
Page 14
Installation and Preparation for Use
Warning
Connect securely the drainage hose.
Connect securely the drainage hose. If the hose comes off due to incomplete connection, water would leak inside the unit and might cause machine trouble.
(1) Take the drainage hose and the hose band out of accessories for the unit.
(2) Make sure the circuit breaker is off.
(3) Take out the rubber stopper from the drain .
(4) Pass the hose band over the hose. Then, insert the hose into the drain and fasten the
hose band.
Safety Precautions
Be careful to arrange the drainage hose.
Do not bend the hose nor make any convex part.
Arrange the hose lower than the drain of the unit. And also avoid the connection which make water accumulate in the drainage hose or drain exit and become resistance for drain.
Lead the end of the drain hose to drainage capable of about 2
Drainage with larger capacity is required when boiler water is drained.
/min. of cooling water drained.
‑12‑
Page 15
Installation and Preparation for Use
Warning
Check the temperature of drained cooling water.
The drained water sometimes exceeds 60℃ especially in summer. Do not touch the water carelessly, or you can get scalded. Drain the water away from your work areas.
Drain hot water away from chloroethylene pipe, if used, in the drainage facility, or the drained water can deteriorate the pipe. You must connect the drain trap, a non-standard accessory for the unit, to the chloroethylene pipe in case the pipe temperature does not get lower than 60℃. Use the drain trap of the non-standard accessory even under 60℃ when you do not use the above-said pipe and joint.
Safety Precautions
If the drain pipe of the drainage facility does not get lower than 60
Connect the drain trap, a non-standard accessory for the unit.
The trap keeps the drained water for a while to cool the water naturally. In addition, the trap mixes tap water and cooled drained water. Then mixed water is drained to drainage facility after mixed water is cooled down.
If you want to know about the details of the drain trap, ask your seller or Yamato Service Department.
AUTOSTILL
AUTOSTILL
AUTOSTILLAUTOSTILL
℃℃℃℃
.
Draintrap
Draintrap
DraintrapDraintrap (non‑standardaccessories)
(non‑standardaccessories)
(non‑standardaccessories)(non‑standardaccessories)
Carrier
Carrier
CarrierCarrier (non‑standardaccessories)
(non‑standardaccessories)
(non‑standardaccessories)(non‑standardaccessories)
‑13‑
Page 16
Installation and Preparation for Use
Warning
Connect securely the Ion-exchange resin cartridge.
Connect securely the ion-exchange resin cartridge(CPC-N) as follows.
If you do not connect securely, the connected hose would come off, causing water leak.
(1) Make sure the earth leakage circuit breaker or the unit is switched off, and the tap is being
fastened. (2) Use the ion-exchange resin cartridge out of accessories for the unit. (3) Put the ion-exchange resin cartridge on the receive table.(See Fig.01) (4) Fix the ion-exchange resin cartridge with the band of the receive table.(See Fig.02)
Safety Precautions
(5) Remove the rubber cap from the entrance and exit of the ion-exchange resin cartridge. (6) Insert the coupler which marked “IN” into the entrance(left side) of the ion-exchange resin
cartridge as far as making sounds like “katchi”. (7) Insert the coupler which marked “OUT” into the exit(right side) of the ion-exchange resin
cartridge as far as making sounds like “katchi”.
Sometimes the coupler is not easy to be inserted. If you insert the coupler hard, you may break the insertion. Be careful not to insert the coupler curved.
(8) When you want to remove the coupler from the ion-exchange resin cartridge, it makes
easy to remove the coupler that y ou pull the coupler forward wit h pushing the black part of
the coupler deeply .
‑14‑
Page 17
Installation and Preparation for Use
Warning
Connect securely the pre-treatment cartridge.
Connect securely the inside hose as follows.
If you do not connect securely, the connected hose would come off, causing water leak.
(1) Make sure the earth leakage circuit breaker or the unit is switched off, and the tap is being
fastened. (2) Use the pre-treatment cartridge ① out of accessories for the unit. (3) Take off the caps on the IN and OUT of the cartridge ①. (4) Open the unit front door. Then you can find one connecting hose with the IN coupler
and other with the OUT. Connect the respective hoses to I N and OUT of the cartridge ①. (5) To connect each hose, slide the blue part ③ of the coupler back to the hose side, and fit
the end of the coupler to the respective end of the cartridge. Then release the blue part
.(See Fig.03)
(6) After you finish the connection, set the pre-treatment cartridge at the left side before the
distilled water tank, as the picture shows. Be careful not to bend the hose. (See Fig. 04)



Safety Precautions
Fig.03
Fig.03
Fig.03Fig.03
Fig.04
Fig.04
Fig.04Fig.04
‑15‑
Page 18
Caution for Use
Warning
No use at occurrence of abnormality/troubleshooting
If smoke or any strange odor should disburse from this unit for any reasons, turn off immediately unit source breakers, extract the plug, and request the seller or Yamato Scientific to check the unit. Neglecting this procedure can result in fire or electric shock. Never try repairing the unit yourself. It is too dangerous.
If it begins thundering.
If it begins thundering, turn the circuit breaker off immediately.
Neglecting this procedure can result in circuit disorder, fire or electric shock due to
thunderbolts.
When you dispose of the unit.
Dispose of the unit as large-sized discarded article.
Do not leave the unit in child playground etc.
Be careful to use the cleanser(Organosol).
Keep the cleanser in an airtight container, and avoid preservation in hot and humid areas.
The main ingredient of the cleanser “Organosol 10” is sulfamic acid (PH1 acid solution).
Wear protective gloves, mask and glasses to handle the cleanser.
Wash by pure water the cleanser splashing on body.
Neutralize cleanser after use with sodium hydroxide.
Check by a PH indicator if cleanser is neutralized irrigation channel or fields.
Do not use the vacant cleanser container as a drink bottle.
Do not drain the cleanser directly into an agricultural irrigation channel or fields. The cleanser can kill rice plants and other farm products.
When you do not use the unit in the nighttime or holidays
Turn the circuit breaker off.
Be sure to
The change of water pressure may cause some accident like water leakage.
Fasten the tap.
‑16‑
Page 19
Body
This drawing identifies parts.
Identification of Parts
valve
outlet
outlet
‑17‑
Page 20
Piping Diagram
Identification of Parts
Decompressor Pressure switch Solenoid valve for cooling water Solenoid valve for raw water
Pre-treatment cartridge
Ion-exchange resin cartridge Float receptacle Float switch to detect boiler overheat Float switch to control boiler water level
Float switch to control the heater Solenoid valve to drain boiler water
Valve to drain boiler water Boiler
Heater Condenser
Electrode for ion-exchanged water Electrode for distilled water examiner Solenoid valve for boiler water supply Solenoid valve to drain the initial distilled water Solenoid valve to take ion-exchanged water Solenoid valve to take distilled water Ion-exchanged water flow sensor
Distilled water flow sensor
(WG221S only)
Ion-exchanged water outlet Distilled water outlet
Distilled water tank Float switch to check the water level Air filter Pump to take distilled water Drain for distilled water tank
‑18‑
Page 21
Identification of Parts
Operation System
This drawing identifies WG221S operation system by the process. (There are some different parts used for WG221.)
(1) Drain boiler water
Turn the circuit breaker on and push the power switch ,then after 15sec.
solenoid valve to drain
boiler water opens for about 40sec. and at the same time for about 30sec
(2)Supply and distillation of boiler water
.
solenoid valve for cooling water opens
After finished boiler water drainage, water opens at the same time ,and then
When
sufficient,
Following water supply, when
closing
float switch (to control the boiler water level) in the float receptacle finds water level
heater becomes on to start distillation.
solenoid valve for raw water and
solenoid valve for raw water and
water is supplied into boiler.
float switch controls water supply into the boiler by opening and
solenoid valve for boiler water
solenoid valve for boiler
.
‑19‑
Page 22
Operation System
●
(3)Flow of cooling water
Identification of Parts
In distillation of boiler water ,cooling water flows and drains from the water solenoid valve to
exchanged water , distillation stops and water flow and dr ain stops automatically.
(4)Collection of distilled water
condenser. When distilled water tank is full of water or taking ion-
decompressor, cooling
For about 10mim. from the beginning of distillation of boiler water , distilled water condensed in
condenser is stored into distilled water tank through electrode for distilled water, after initial water is drained by water full and stops distillation when the highest
Distilled water stored is taken out by
take distilled water
,
distilled water flow sensor (WG221S only)
solenoid valve (to drain the initial distilled water)
pump(to take distilled water) through
float switch inside the tank is activated.
and
opened. The unit judges
solenoid valve to
distilled water outlet
.
‑20‑
Page 23
Operation System
(5)Collection of ion-exchanged water
Ion-exchanged water is taken out by through
pre-treatment cartridge exchanged water and
,
Ion-exchanged water outlet
,
solenoid valve to take ion-exchanged water
Identification of Parts
decompressor,
Ion-exchange resin cartridge(CPC-N)
,
.
solenoid valve for raw water
Ion-exchanged water flow sensor
,
,
electrode for ion-
‑21‑
Page 24
Operation Panel
Identification of Parts
20
Power switch
Operation indication lamp
Select switch
Lamp to identify the measured electrode
Unit of water quality measurement
Indicator (1)
Trouble indication lamp
Distilled water level indicator
Switch to change methods to take pure water
Display lamp to identify the method to collect water
Switch to take ion-exchanged water
Lamp to indicate collection of ion-exchanged water
⑫ ⑬
Switch to take distilled water
Lamp to indicate collection of distilled water
Unit of amount to collect
Indicator (2)
Function key
Enter key
Up key
Down key
‑22‑
Page 25
Operation Panel
Power Switch…
This is the switch for power supply. The STAND-BY lamp is lit when the power cable is connected and the earth leakage circuit breaker is on. W hen you push the switch in this situation, power becomes “ON”. If you push the switch again, then power becomes “OFF”.
Operation Display Lamp…
Stand-By: (1) The yellow lamp is lit when the power source is “ON”. But the power switch is “OFF”. (2) The yellow lamp blinks in case of suspension of water supply or decrease of raw
DISTILLATION: The green light is lit during distillation. DRAINAGE: The green light is lit during automatic drainage of boiler water.
Select Switch…
This switch is to select either electrode for water quality check. Each one push automatically and alternatively select either sensor for ion-exchanged water or the other for distilled water. Then the indicator (1) shows either electrical conductivity or resistivity of the selected electrode.
Lamps to identify the electrode to measure…
Either lamp is lit to show the electrode selected by the SELECT switch.
ION EXCHANGED WATER: Water quality at the exit of the ion-exchange resin cartridge DISTILLED WATER:
×10-4S/m
4 
×10
Ω・m
Quality of distilled water at the exit of the condenser
Unit to display water quality…
Either unit of the value the indicator (1) shows is identified.
(The resistivity is shown by the integral number within 1 to 18.
Identification of Parts
water pressure. In addition, the indicator (1) shows
: represents the electrical conductivity (=µS/cm) : represents the resistivity (=MΩ・cm)
.
‑23‑
Page 26
Operation Panel
Indicator (1) …
The displayed (selected) water quality is the converted value at 25℃. Another water quality not indicated (selected) is always being checked. If either water deteriorates its quality to the specific value, the indicator repeatedly indicates the sensor No. The indicator also gives the respective error codes in case of stop of raw-water supply, decrease of water pressure or any other trouble in the device.
<The standard values to determine det er ioration of water quality>
Identification of Parts
Ion-exchanged water: Distilled water:
<Example of indication of water deter i or at i on>
Incase you select the ion-exchanged water by SELECT key: When quality of ion-exchanged water is 0.10×10
Electric conductivity of the selected electrode C3 (ion-exchanged water)
<Stop of raw-water supply or decrease of water pressure>
comes up and the STAND-BY, an operation display lamp, blinks.
<In trouble>
The display identifies the trouble. Refer to the Trouble Shooting Guide (P.27) for more information.
-4
S/m while distilled water is 10×10-4S/m or more. (deterioration of water)
or more
or more
Electric conductivity of the electrode C1 (deteriorated distilled water)
‑24‑
Page 27
Identification of Parts
Operation panel
Trouble Indication lamp…
The lamp blinks when detects abnormality in operation or trouble of the unit. The indicator (1) identifies such a situation.
Distilled Water Level indicator…
The indication lamps show 5 levels of the distilled water in the tank. When the water storage is less than 2 distilled water for preventing the empty pump operation. When the lowest lamp start blinking, you can take distilled water again.
<No water> <Full water>
, the lowest yellow lamp is lit. In this case, you can not take
Operational range of the pump to take water
Switch to change methods to collect pure water…
This is the switch to select taking water repeatedly or by the fixed volume. To select taking the fixed amount, push once. If you push again, you can take water repeatedly.
Lamp to identify the water-collection method…
This lamp is lit only while you select t he repeat ed collection. On the other hand, the lamp goes out when you select collection of the fixed volume.
Switch to collect ion-exchanged water…
1) In case of repeated collection Switch ON (one push) to collect ion-exchanged water. Switch OFF (one more push) to stop collection.
2) In case of collecting the fixed volume Switch ON (one push) to start collection of ion-exchanged water. The unit stops automatically its operation when the collection reaches the set amount. If you push the switch during collection, the unit would stop operation even before the collection reaches the set amount.
Lamp to indicate collection of ion-exchanged water…
The green lamp is lit during collection of ion-exchanged water.
Switch to take distilled water…
1) In case of repeated collection Switch ON (one push) to collect distilled wat er. Switch OFF (one more push) to stop collection.
2) In case of collecting the fixed volume Switch ON (one push) to start collection of distilled water. The unit stops its operation automatically when the collection reaches the set amount. If you push the switch during collection, the operation would stop even before the collection reaches the set amount. WG221 can change to collection mode of the fixed volume by using switch (to change methods to take pure water). But in collection of the fixed volume mode, switch to take distilled water does not work.
‑25‑
Page 28
Operation panel
Lamp to indicate collection of distilled water…
The green lamp is lit during collection of distilled water.
Unit of volume to collect…
When the indicator (2) shows the volume to take, the unit of ( case the indicator shows the total flow, the unit of ×10
Indicator (2)…
The indicator shows the following.
1) In case of repeated collection The indicator shows
2) In case of collecting the fixed amount
It shows the set volume. The initially set value is 0.1 decreasing as water is being collected. At the completion of collection, the set value comes up again.
3) Identification of function
The indicator also identifies the function selected. a) b) : Mode to change the unit for water quality check
4) Display of total flow
The total water flow shows the aggr egate volume of distilled water collected.
Function key…
Push the select functions. The indicator (2) identifies the selected function.
Identification of Parts
) is displayed. In other
is displayed.
.
. The display set value is
: Indication of the total flow
Enter key…
A push on the key completes your selection of function. In other word, you need to push
this key to escape from the function-selecting mode.
Up key…
Push the key to set the following
1) At collection of the set volume
Set the volume to collect. Each one push increases the set value in the indicator (2). The set value goes up successively when y ou keep pushing the key.
2) At the selection of a function
You can select the following functions by pushing this key. a) ON or OFF to display the total flow of ion-exchanged water. b) Change of units to indicate water quality.
Down key…
Push this key to set the following
1) At collection of the set volume
Set the volume to collect. Each one push decreases the set value in the indicator (2). The set value goes down successively when you keep pushing the key.
2) At the selection of a function
You can select the following functions by pushing this key. a) ON or OFF to display the total flow of ion-exchanged water. b) Change of the unit to indicate water quality.
‑26‑
Page 29
Emergent Troubleshooting
Identification of “TROUBLE” / Identification of causes
This unit has self-diagnosis function. If any operational abnormalities or machine troubles should occur, the alarm lamp in operation panel blinks. f this occurs, immediately disconnect the breaker and turn the tap off. See page 36 “Display and Details of Error Code” ,if you need to know about the details of identification of “TROUBLE”.
The TROUBLE lamp also blinks.
Water leakage
The TROUBLE lamp also blinks.
Abnormal boiler water level
The TROUBLE lamp also blinks.
Heater overheating
The TROUBLE lamp also blinks.
Defect of the float switch to
control the boiler water level.
The TROUBLE lamp also blinks.
Defect of the float switch for the
distilled water tank.
The TROUBLE lamp also blinks.
Defect in the electric circuit
‑27‑
Page 30
Preparation and Check before Use
Warning
1. Check water supply.
z
Check if the water-supply hose is securely connect ed.
z
Then, open the tap.
z
Make sure if no water leaks from the joint of the hose.
2. Check drain.
z
Check if the drain hose is securely connected.
z
Make sure if the hose is not bent or if there are no conv ex part s.
z
If the hose is bent, not only the unit does not run correctly, but also it may cause the accident of
water leak. Sometimes check if the drain is done correctly.
3. Check the power source.
z
Make sure that the pow er cable is connect ed t o an appropriat e out let .
4. Just after run
z
After turning on electricity initially or drain from distilled water tank, It takes much time to start
collecting of distilled water because of the air in pump or pipe.
z
Just after exchanging pre-treatment cartridge or Ion-exchange resin cartridge , It also takes much
time to start collecting of distilled water. And when you change the some of cartridge, drain about 5 liters to remove t he initial impurities.
Run Menu
‑28‑
Page 31
Run Menu
Run menu as follows when you are ready for operation.
1. Power-on
1
Switch ON the earth leakage circuit breaker.
2 Push the POWER switch. Operation panel display in about 15 seconds after you
push the power switch until the dist illat ion st ar ts.
Run Menu
The indicator (1) blinks
The STAND-BY, an operational indication lamp, goes
out. (However, the indication (1) does not show
you push the power key some time after switching on the earth leakage circuit breaker.)
2. Drain boiler water after you push the power switch.
Display of operation panel…
●
Operation indication lamp lit with the green light during drainage of boiler water . (about 30 sec.)
(However, drainage of boiler does not work only if you push the power key after switching on the earth leakage circuit breaker.)
.
if
3. 15 minutes after you push the power switch.
Display of operation panel…
The indicator (1) shows electric conductivity of ion-
exchanged water. (Ex.
The ION EXCHANGED WATER lamp is lit to identify the electrode being measured.
Among distilled water level indicators, the yellow lamp is lit to show no water storage.
Water is supplied to the specific level. At t he completion of initial water supply, distillation starts. Then, the green DISTILLATION lamp is lit among operation indication lamps.
)
‑29‑
Page 32
Repeated collection of ion-exchanged water
Push once the ION EXCHANGED WATER switch while the CON. lamp, identifying the water collection method, is lit.
Among the operation panel indicators;
The lamp indicating collection of ion-exchanged water
blinks If you push again the ION EXCHANGED WATER switch; Among the operation panel indicators;
The lamp indicating collection of ion-exchanged water
goes out.
Run Menu
Repeated Collection of Pure Water
Repeated collection of distilled water
Push once the DISTILLED WATER switch while the STAND BY lamp, identifying the method of water collection, is lit but the yellow “no water” lamp, among indicators to show the volume of distilled water, is not lit.
Among the operation panel indicators;
The lamp, indicating collection of distilled water, is lit. If you push again the DISTILLED WATER switch; Among the operation panel indicators;
The lamp indicating collection of distilled water goes out.
(If, among indicators to show the volume of distilled water stored, the yellow “no water” lamp gets lit during collect ion of water, the unit automatically stops taking water and the lamp showing collection of distilled water goes out.)
‑30‑
Page 33
Run Menu
Collection of The Fixed Volume of Ion Exchanged Water
1. Push the switch to change the method to collect water. Then the CON. lamp goes out.
Among the operation panel indicators;
The CON. lamp, a collection method indicator, goes
out.
The indicator (2) shows the initial value of
a unit of collection, blinks.
. ( ),
2. Push either
▲▲▲▲
or
key to set the volume to take.
▼▼▼▼
Among the operation panel indicators;
The indicator (2) changes the value according to your
key touch. (You can set up to 99.9
3. Push the ION EXCHANGED WATER switch. Ion exchanged water comes out from the outlet
Among the operation panel displays;
The lamp indicating collection of ion exchange water is
lit.
The indicator (2) starts decreasing the displayed value.
.)
4. When the indicator (2) decreases the displayed value to and the remaining water flow of 0.1
is completed;
Among the operation panel indicators;
The lamp indicating the collection of ion exchanged water goes out.
The indicator (2) returns to the set value.
(The actual amount of collected water may increase or decrease between maximum±10%)
‑31‑
Page 34
Run Menu
Collection of The Fixed Volume of Distilled Water(WG221S)
1. Push the switch to change the collection method and put out the CON. lamp.
Among the operation panel indicators;
The CON. lamp, the collection method indicator, goes out.
The indicator (2) shows the initial value of
At the same time, (
), a unit of water collection, is lit.
.
2. Push
▲▲▲▲
or
key to set the volume to take.
▼▼▼▼
Among the operation panel indicators;
The indicator (2) changes the value according to each your key operation.
(You can set up to 99.9
3. Push the DISTILLED WATER switch.
Distilled water comes out from the outlet.
Among the operation panel indicators;
The lamp indicating collection of distilled water is lit.
The indicator (2) starts decreasing the displayed value.
(If the yellow “no water” lamp, one of indicators to show t he level of distilled water stored, is lit during water collection, the unit automatically stops collection and the lamp indicating collection of distilled water goes out. In addition, the indicator (2) returns to the set value.)
.)
4. When the indicator (2) decreases the displayed value down to and the remaining water flow of 0.1 is completed;
Among the operation panel indicators;
The lamp indicating collection of distilled water goes out.
The indicator (2) returns to the set value. (The actual amount of collected water may increase or decrease between maximum±10%)
‑32‑
Page 35
Run Menu
Indication of Water Quality
Selection of either electrode for ion-exchanged or distilled water
Each one push on SELECT switch selects alternatively the electrode of either water quality examiner to be measured.
Among the operation panel indicators;
Either ION EXCHANGED WATER and DISTILLED WATER is lit alternatively by each push to identify one electrode measured.
(The electric conductivity is shown by t he unit of “
Change of the unit to indicate water quality.
1. Each one push on the FUNCTION key changes displays of indicator (2).
Operation panel display
The indicator (2) changes its displays as follows.
” when the power switch is “ON”.)
Repeated collection Mode to display the total water flow
Mode to change the unit to indicate water quality
Select the display of .
2. Push either
▲▲▲▲
or
key once.
▼▼▼▼
Among the operation panel indicators;
The unit for water quality indication changes. Then the
unit of resistivity “
The indicator (1) also indicates the resistivity. The
integral number shows from 1 to 18 within the range of
1.0 to 18.2 (
” is shown.
).
3. Push the ENTER key, Then, you can fix the unit and get out from the function mode.
Among the operation panel indicators;
The indicator (2) shows
The unit to indicate water quality returns to the electric
conductivity when you switch the POWER key OFF and ON again.
.
‑33‑
Page 36
Run Menu
Indication of Water Quality
Measurement of Electric conductivity
The water quality examiner in the operation panel shows the electric conductivity at the exit of ion­exchange resin cartridge and of distilled water condenser. Use the indication as the standard for exchanging ion-exchange resin cartridge. When you read the indication, be sure that the part of electrode is in the full of water; that is in the flowing of Ion-exchanged water. Besides that the part of electrode is not in the water, the indication is not correct for the influence of bubbles in the following case.
1. At the beginni ng of r unni ng or dur ing the unit stopped
2. Just after exchanging pre-treatment cartridge or ion-exchange resin cartridge
3. Just after the beginning of distillation of water
Electric conductivity
Electric conductivity shows how much electricity can pass. As water contains more electrolyte or impurity, more electricity can pass through and the value of conductivity gets larger. On the contrary, the value gets smaller as water contains less electrolyte.
●That means water is more pure as the conductivity is smaller. However, the value of conductivity shows containment of electrolyte only but does not indicate any containment of non-electrolyte such as organic, colloid, dissolved gas and microorganism. Therefore, conductivity is just one of indicators to shows water purit y.
Resistivity is to indicate the same property as the electric conductivity shows. Resistivity is reciprocal to electric conductivity. The larger resistivity indicat es higher purit y.
To calculate the resistivity from the elect r ic conduct ivity;
R
= Resistivity ρ=Electric conductivity
R[Ω・m]
R[Ω・m]R[Ω・m]
Pure water theoretically has the following value.
R
=18.3×10 (The resistivity is displayed by the integral number within 18 to 1×10 number indication.)
ρ=0.055×10
4
Ω・
-4
=
ρ
    SSSSm]
m (18.3MΩ・cm) 25
S/m (0.055・/cm) 25
[/
m]
m]m]
or
R[×10
R[×10 Ω・m]
R[×10R[×10
R[Ω・m]
1111
4444
Ω・m]
Ω・m]Ω・m]
=
    ×10
ρ
4
Ω・
1111
−4
−4
−4−4
×10 SSSSm]
[/
×10×10
m. Note there is no decimal
m]
m]m]
Quality of ion-exchanged/distilled water
Ion-exchanged water and distilled water have the following properties respectively. Select either water according to your requirement.
It is ideal to use distilled water just after you collect. If you do not use the water soon, drain the distilled water tank. When you keep distilled water in the tank for a while, drain the tank and store new distilled water in the tank to use.
(1) Ion-exchanged water
Since almost all electrolyte are removed from ion-exchanged water, you can get water with the lowest electric conductivity. But it is impossible to remove non-electrolytic impurity. Purity is a little low with a new resin or at the re-start of flow after operation stops.
(2) Distilled water
The mode can remove averagely electrolyte and non-electrolyte except ammonia and other substances with low boiling point. However, water absorbs carbonic acid gas from air and creates carbonic gas during its production process (in condensation and storage). Accordingly, water quality is worse than ion-exchanged water and becomes weak acid (pH5 to 6) with its electric conductivity about 1 t o 2× 10
-4
S/m 25℃ (1 to 2.5・/cm).
‑34‑
Page 37
Run Menu
Display of Total Flow/Resetting
Check the total flow to know the time to exchange ion-exchange resin or pre-treatment cartridge.
1. Each one push on the FUNCTION key changes display of indicator (2).
Among operation panel indicators, the indicator (2) changes indications as follows.
Repeated collection Mode to display the total flow
Mode to change the unit for indicating water quality
Select and display .
2. Push either
▲▲▲▲
or
key once.
▼▼▼▼
The indicator (2) in the operation panel shows the total flow.
3. Push the ENTER key to cancel the function.
The indicator (2) in the operation panel shows .
Resetting the total flow.
Select the total flow in the indicator (2). Then, push the same time.
and key at the
The indicator (2) shows .
‑35‑
Page 38
Safety Devices and Error Codes
Displays and Details of Error Code
Safety device Indicators Indication Cause Effect on operation Measure
Detecting leakage
Heater overheat protection
Detecting trouble of float switch
Detecting abnormal water level
Detecting float switch trouble
Indicator(1)
is lit
is lit
is lit
is lit
is lit
Water leakage
Heater overheat
Trouble of the float switch for distilled water tank
Abnormal water level in the float receptacle
Trouble of the float switch for float receptacle
Heaters, solenoid valves and any other controllers become off.
Heaters, solenoid valves and any other controllers become off.
Heaters, solenoid valves and any other controllers become off.
Heaters, solenoid valves and any other controllers become off.
Heaters, solenoid valves and any other controllers become off.
Turn the breaker off and check piping parts. (See P38 for troubleshooting)
Contact the seller or Yamato Scientific’s Technical Service Department.
Contact the seller or Yamato Scientific’s Technical Service Department.
Check if the boiler water drain valve is open. If this abnormality is indicated with the valve closed, contact the seller or Yamato Scientific’s Technical Service Department.
Contact the seller or Yamato Scientific’s Technical Service Department.
Detecting trouble of the electric circuit
”TROUBLE” blinks to alarm abnormal situation while the indicator (1) shows a certain error code.
is lit
Trouble in the electric circuit
Heaters, solenoid valves and any other controllers become off.
Contact the seller or Yamato Scientific’s Technical Service Department.
‑36‑
Page 39
Safety Devices and Error Codes
Trouble Displays other than Error Codes/Details
Safety devise Indicator indication Cause Effect on operation Troubleshooting
Detecting abnormal water quality
Indicator (1)
Ion-exchanged water deteriorates lower than the standard. See page24 about the standard.
Distilled water deteriorates lower than the standard. (When the value reaches 10×10 or more.)
Water quality is too low to measure.
Disconnection of the measured electrode without water.
-4
S/m
The indicator repeatedly alarms
. However, operation continues as usual.
The indicator repeatedly alarms
. However, operation continues as usual.
The usual operation continues. However, the sign is continuously shown when you select the electrode.
The usual operation continues. However the sign is continuously shown when you select the electrode.
It is not machine trouble. Change the old ion-exchange resin into new one.
Contact the seller or Yamato Scientific’s Technical Service Department.
Indications of and are shown at the same time. Follow the respective instructions.
This situation sometimes occurs in operation of this device. However, if this indication would not disappear for a long time when water flows, request service.
Detecting sensor trouble
Lamp to identify the electrode measured.
STAND BY
lamp is lit
Blinking
Interruption/pressure decrease of original water flow. When pressure of water service gets lower than 50kPa (0.5kgf/cm
Disconnection of the thermistor for temperature compensation of the measured electrode.
2
)
Distillation and collection of ion­exchanged water stops temporarily.
The temperature without compensation is shown when you select the electrode.
Make sure if the water pressure is normal and the tap is open. Distillation starts again when the original water pressure is re-gained.
It is better to request earlier service.
‑37‑
Page 40
Safety Device and Error Codes
Troubleshooting
Troubleshooting when is lit to indicate water leak.
(1) Turn the earth leakage circuit breaker in the right side of the unit OFF. (2) After the defect is repaired, to re-start the operation, wipe the water at the bottom areas of the device
and dry it, and also remove the leak detecting electrode and dry it . (3)Set the electrode in the original position. (4) Close the door . (5) Turn the breaker ON, and push the POWER switch. Normal operation starts as the defect is repaired.
‑38‑
Page 41
Check and maintenance
Maintenance
Time for periodical check and maintenance
condition.)
Item to check and maintain Period Note
Exchange of pre-treatment cartridge
Exchange of ion-exchange resin cartridge Cleaning the still 3 months
Cleaning the water supply hose filter 6 months Hose exchange 2 years Check the joint once a month. Distilled water tank 3 months
Each 6 or 12 months
When the indicator (1) shows
(Check periodically the unit to use in stable
Durability: About 5000
in Tokyo (with quality of 200×10 You need earlier exchange of the cartridge when you use raw water with worse quality. Durability:
About 650 200×10
You need earlier cleaning when you use raw water with worse quality.
in case of using water service
of raw water with quality of
-4
S/m
-4
S/m)
Exchange of Pre-Treatment Cartr idge
Refer to Page 15 “Connect securely the pre-treatment cartridge” in the “Safety Precaution”. Dispose of the old pre-treatment cartridge as nonflammable. If you use the unit without changing the pre-treatment cartridge, the span of life of the ion-exchange resin cartridge is made short.
Exchange of Ion Exchange Resin Cartridge
Because the keeping of the cartridge for long time makes the water quality and durability worse, prepare the spare cartridge for the time to exchange. The standard of keeping in about four months. Refer to Page 14 “Connect securely ion-exchange resin cartridge” in the “Safety Precaution”. Dispose of the exchanged cartridge as nonflammable. If you use the unit without changing the cartridge, stone attaches increasingly to the inside of the boiler and heaters, which can decrease collection of distilled water and damage heat er s.
‑39‑
Page 42
Maintenance
Cleaning Still
Removal the still
1.
Turn “OFF” the circuit breaker.
2.
Close the faucet.
3.
Make sure if the boiler cools down enough (about more than 30 min. later). Open the front door of this
unit and the valve to drain boiler water.
4.
Remove both hose connected to the ① boiler and ② condenser respectively. Twist the hose-band by
using tools, move the part of notches and you can remove the exit of distilled water and outlet/drain of boiler. When you remove them, be careful not to damage the glasses with giving excessive power. Remove the nut of entrance and exit of cooling water by rotating counterclockwise. Be sure not to lose the inside packing.
5. Take off the four connecting screws by cross screwdriver, and take off the left side board.
6. By cross screwdriver , loosen the four connecting screws in right side of terminal rack that is in the upper
right of framework made by removing the left side board. Pull the heater terminal.
‑40‑
Page 43
Maintenance
Cleaning Still
Removal the still
7.(1) Pull the heater terminal from the grommet. But do not bend or pull leads from heaters beyond the necessity. (2) Remove the two screws of band for boiler fixed by cross screwdriver, and take off the boiler
condenser.
(3) Loosen the three rolled screw of metal fittings connection, and remove the boiler condenser.
Cleaning the boiler
1. Preparing scale detergent (1) Prepare hot water with temperature between 50 and 60℃ about 2 liters.
(2) Mix well “Organosol 10” about 200g, an accessory for this unit, and hot water prepared (1).
2. Stop the hose joint in the lower part of the boiler (outlet/drain of boiler) by rubber stopper etc.
3. Set the boiler at the stable place not to pour scale detergent from the boiler.
4. Pour the scale detergent from joint into the condenser with the heater connected.
5. Almost all scales are removed in about 4 to 5 hours. Drain the scale detergent from the boiler. If a lot of scales in the boiler are not removed, add new scale detergent to repeat cleaning. (1) If you finished removing the scales, take off the heaters from the boiler and wash them thoroughly by
the tap water. Wash them in the big beaker with full of water not to be wet of leads and drain.
Do not wash them by the tap water directly.
(2) If solid scales is not removed by the scale detergent, do the following treatment.
Boiler : Wipe off by brush etc.
Heater: Wipe off by soft material like chip or plastic etc. The scales in the heater must be wiped equally and the solid scales must not be left partly. Extremely the part only become a large heat resistance, therefore the heater may be broken.
‑41‑
Page 44
Maintenance
Cleaning Still
Cleaning the condenser
1. Pour the scale detergent into the cooling pipe of the condenser . (See page41 about the
2. If the cleanser comes out from the hose joint, stop the hose joint by silicone stopper etc. Almost all fur is removed in about 4 or 5 hours.
3. After the scale detergent drained, wash the inside of condenser by tap water thoroughly.
About the treatment of the cleanser ( See page16 “Caution for Use”)
1. Wash the boiler and the heater a little early. The more the amount of the scales adhesion is, the more
difficult you remove it. And it may cause that the amount of taking distilled water decreases and the heater is broken.
2. After washing, drain the cleanser from them and neutralize cleanser by the neutralizer. Check by a PH
indicator if cleanser is neutralized irrigation channel or fields. (the main ingredient of the scales cleanser : sulfamic acid , PH of the solution : about PH1 acid)
3. Keep the cleanser in an airtight container, avoid preservation in hot and humid areas, and keep in cool
and dark areas.
4. Wear protective gloves, mask and glasses to handle the cleanser.
5. Wash by pure water the cleanser splashing on body.
6. Do not use the vacant cleanser container as a drink bottle.
7. Do not drain the cleanser directly into an agricultural irrigation channel or fields. The cleanser can kill
rice plants and other farm products.
preparing scale detergent )
Installation of the boiler
1. Set the joint (to the condenser) level by the band for boiler fixed. After you check that there is packing in the cap nut, insert it into the boiler with the characters “YK-W-3” upward.
‑42‑
Page 45
Maintenance
Cleaning Still
Installation of the boiler
2. Connect the four heater terminals with the terminal rack.
3. Install the left side board on the unit.
4. Insert the tube into supply and drain for boiler and fix it by hose band.
Installation of the Condenser
1. Insert the packing into the joint with boiler and condenser, and fix boiler and condenser in the same direction by metal fittings connection.
2. Connect the each hose to entrance of cooling water, exit of cooling water , and exit of distilled water.
3. Close the valve to drain boiler water.
‑43‑
Page 46
Maintenance
Exchange Heaters
●
If the heater snapped or was broken, exchange the heater as follows.(Refer to P.40〜43 for exchanging)
1. Turn “OFF” the circuit breaker .
2. Close the faucet.
3. About more than 30 min. later after turned “OFF” the circuit breaker, open the front door of this unit and the valve to drain boiler water.
4. Open the left side board, loosen the four connecting screws in right side of terminal rack, and pull the heater terminal.
5. Pull the heater terminal from the grommet.
6. Remove the cap nut of the heater, and pull t he heat er.
7. Remove the packing and the cap nut from the broken heater.
8. Install the packing and the cap nut to the new heater. At this time, do not touch it with empty hands not t o apply the finger marks.
9. Insert it into the boiler with the heater mark “YK-W-3” upward.
10. Insert the heater terminal into the grommet, and fix it in the terminal rack with checking the position of installing heater lead.
11. Install the left side board on the unit.
12. Close the valve to drain boiler water.
13. Open the faucet after closing the front door.
14. Turn “ON” the circuit breaker.
15. Check that the STAND-BY lamp lit.
16. Push the calibrate switch in the upper part of the circuit breaker. (The calibrate switch is a switch to memorize in the inside controller the standard temperature which the sensor inside the each heaters shows at normal running. It senses malfunction when the temperature become 20℃ high more than the standard temperature for som e reasons at the normal running. )
17. At the calibrate running for 5 min. long, the indicator (1) of the operation panels blinks CAL.
18. After 5 min. calibrate running, the running become normal automatically.
‑44‑
Page 47
Clean Water Supply Hose Filter
(1) Close the faucet and remove the water supply
hose after turning off the unit breaker. (2) Remove ② plug from ① ring with turning. (3) Clean ⑦ filter in the plug by tap water. (4) Push the filter by flat end of a pencil etc. from
the hole at the side of sleeve in case of terrible
stoppage. (5) Brush the filter . (6) Assemble the filter in the reverse procedure.
Maintenance
Exchange Hoses
Use the specified hose for exchange.
‑45‑
Page 48
When you do not use the unit for a long time
If you do not use the unit for a long time, turn off the unit breaker for safety and close the faucet. Water in the boiler and the distilled water tank deteriorates its quality by the growth of unwanted bacteria and algae. Drain as follow.
Draining boiler water
(1) Before draining the boiler water, turn off the unit
breaker , check if faucet is closed , and wait for
30 minutes or longer. (2) Open the valve to drain the boiler water. (3) Make sure if all water is drained from the boiler
and float receptacle. (4) Close the boiler water draining valve. If the
drain valve remains open in the next use of the
unit, water would not flow into the boiler or start
distillation.
Maintenance
Draining distilled water
(1)Make sure if the unit breaker is turned
off. Open the front door.
(2) Pull out the hose to drain distilled water
set at the upper left on the back of front door.
(3) Remove the silicone stopper to drain the
distilled water from the tank. Check in advance the remaining water level in the tank, and prepare receptacle for drainage.
(4) Be careful not to set the drain hose
higher than the bottom level of the distilled water tank, or more tank water remains in the tank.
(5) Install the silicone stopper at the point
tightly,and Insert stopper securely in the original folder after completing drainage.
‑46‑
Page 49
After-sales Service and Warranty
.
When you request repair Warranty (attachment)
If any troubles should occur, note the error code, stop the operation immediately. Turn the power switch off, pull the power cord out and contact the seller or Yamato Scientific’s Technical Service Department.
(Necessary information)
Model number
Serial number
Date of purchase
See the warranty or nameplate on the unit
Detailed information on difficulties
The service man requests you to show the
warranty on his visit.
When you encounter the following
Fault indication Check points
STAND BY switch is not lit when the breaker is turned on.
No water is supplied.
Water supply can not be stopped.
No water is supplied into the boiler.
Heater can not be turned on. No cooling water flows.
Distilled water is not drained. Boiler water is not drained.
Distilled water is not stored. Distillation would not stop.
No water is collected.
The seller or Yamato Scientific’s gives you the warranty.
Keep your warranty card for future references. Check the
name of distributor, date of purchase and any other contents of warranty.
The term of warranty is one year limited commencing the date of purchase. Repair is made without charge according to the contents of warranty.
As for repair after expiration of the warranty period, consult the seller or our Technical Service Department. As long as the function of the unit can be maintained by repair, we will repair the unit upon your request.
Minimum period to keep repair parts in stock
Minimum period to keep repair parts is 7 years limited after we stop the production of this model of units. The repair parts mean any necessary parts to maintain the performance of the unit.
Is the power supply cable is securely connected with the outlet?
●
●
Is there any power failure?
●
Defect of solenoid valve to supply raw water.
●
Low pressure or interruption of water service.
●
Defect of pressure switch.
●
Stoppage in the pre-treatment cartridge.
●
Defect of float switch.
●
Defect of solenoid valve to supply raw water.
●
Defect of solenoid valve to supply boiler water
●
Defect of float switch.
●
Defect of solenoid valve to drain boiler water.
●
Defect of float switch.
●
Heater disconnection.
●
Defect of solenoid valve for cooling water.
●
Defect of solenoid valve to drain initial distilled water.
Defect of solenoid valve to drain boiler water.
●
Defect of solenoid valve to drain initial distilled water.
●
Defect in piping.
Defect of float switch.
●
Defect of solenoid valve to collect ion-exchanged or distilled water
●
Defect of pump to take distilled water.
●
Defect in piping.
‑47‑
Page 50
Specifications
Type WG221 WG221S
Collect
ion method ion exchange ⇒ Distillation
Pure water to collect Distilled / Ion exchanged water
Volume of distilled water to produce
Water flow to collect 1
/min. or more (Distilled / Ion exchanged water)
About 1.8 /h
Boiler Super hard gl ass
Condenser Super hard glassStill
heater Ceramic heater 1.9kw
Distilled water tank Made of pol
Raw water filter
Activated carbon and hollow fiber membrane (0.1
Pre-treatment cartridge (ACF0827)
yethylene 20
Ion-exchange resin cartridge Cartridge type of one-touch connect (CPC-N)
Water quality examiner
Digital display (selection of either electric conductivity or resisitivity)
0.05
300×10
18〜0.1×10
-4
S/m・25℃ (in case of electric conductivity)
4
Ω・m・25℃
(The resistivity is shown by the integral number
from 1 to 18.)
Water supply pump Magnet pump
Indication of tank water levels 5 levels by LED display
Water volume to take 0.1
Range of raw water pressure 0.5
Power requirement
External dimensions
(W×D×H)
(50/
60 Hz
*1
About 570×455(590:including outlet)×730 (mm
Weight About 43
Water abnormality alarm
Automatic boiler drainage ●Automatic adjustment of cooling water level
Prevention of low-water boiling ●Overheat prevention
Supplemental functions
Water leak detector ●No water supply detector
Earth leakage circuit breaker ●Initial distilled water drainage
Abnormality detection of the float switch for the distilled water tank
System to set the volume of distilled water to take(WG221S only
System to set the volume of ion-exchanged water to take
〜5 ×
100 kPa(0.5〜5 kgf/cm
AC100V About 15A
99.9
(kg)
Water supply hose , Drain hose , Joint hose 1 (for each) Instruction manual this book Warranty 1
Accessories
Hose clamp 1 Scale detergent (1kg ) 1 Pre-treatment cartridge 1 Ion-exchange resin cartridge 1 Type Part No.
Water outlet unit -------------------------- (OWH10) 253686
Drain trap ----------------------------------- (OWI10) 253211
Pressure pump (OWJ10) --------------- (OWJ10) 253220
Non-standard accessories
WQ500/501 joint unit -------------------- (OWQ10) 253695
Carrier --------------------------------------- (AS200) 253175
WL100 joint unit --------------------------- (OWL20) 253668
Joint unit of membrane filter ------------ (OWM30) 253667
Membrane filter ---------------------------- (OWN10) 253670
*1…The exterior dimensions exclude projected areas.
μm
Mix bed Resin 3
2
)
‑48‑
Page 51
WG221/221S
1 3
1 30
1 3 4 5
Wiring Diagram
1 5 6 10
Symbol Name of Parts Symbol Name of Parts Symbol Part of Name
Cal_SW Calibrat e SW SV1 Solenoid valve to supply boiler water T1 Terminal
E1 Electrode for ion-exchanged water SV2 Solenoid valve to supply raw water T2 Terminal E2 Electrode for distilled water SV3 E2 2 Electrode for distilled water SV4 Solenoid valve to drain boiler water Tf2 Transformer
ELB Earth leak age circuit breaker SV5
FL1 Flow sensor (ion-exchanged water) SV6 Solenoid valve for cooli ng water X1 Relay FL2 Flow sensor(distilled water)WG221S
only FS1 Float switch to control heater OH Overheat detector FS2 Float switch to control boiler water level P1 Pump to tak e water FS3 Float switch for boiler overheat PIO-5 Operation board FS4 Float switch to check water level PLANAR P l anar board FS5 Float switch to check water level RELAY1 Relay board FS6 Float switch to check water level RELAY2 Relay board FS7 Float switch to check water level SSR SSR board FS8 Float switch to check water level
H1 Heater
MOS RELAY
MOS relay board
SV7 Solenoid valve to take distilled water PS Pressure switch
Solenoid valve to drain initial distilled water
Solenoid valve to take ion-exchanged water
Tf1 Transformer
WL
Water leak detector
‑49‑