Periodical Voluntary Inspection of Small-sized Pressure Vessel................................. 58
•
Page 4
Explanation
Illustrated Symbols
Various symbols are used in this safety manual in order to use the unit without
danger of injury and damage of the unit. A list of problems caused by ignoring
the warnings and improper handling is divided as shown below.Be sure that you
understand the warnings and cautions in this manual before operating the unit.
Cautions in Using with Safety
MEANING OF ILLUSTRATED SYMBOLS
WARNING!
CAUTION!
If the warning is ignored, there is the danger of a problem that
may cause a serious accident or even fatality.
If the caution is ignored, there is the danger of a problem that may
cause injury/damage to property or the unit itself.
Meaning of Symbols
This symbol indicates items that urge the warning (including the caution).
A detailed warning message is shown adjacent to the symbol.
This symbol indicates items that are strictly prohibited.
A detailed message is shown adjacent to the symbol with specific actions not to
perform.
This symbol indicates items that should be always performed.
A detailed message with instructions is shown adjacent to the symbol.
1
Page 5
Ta ble of Illustrated Symbols
Warning
Cautions in Using with Safety
Warning,
generally
Caution
Caution,
generally
Caution,
water only
Warning,
high voltage
Caution,
electrical shock
Caution,
deadly poison
Warning,
high temperature
Caution,
Warning,
drive train
Caution,
scald
no road heating
Warning,
explosive
Caution,
not to drench
Prohibit
Prohibit,
generally
Compulsion
Compulsion,
generally
inflammable
Compulsion,
connect to the
Prohibit,
grounding
terminal
to disassemble
Compulsion,
install on a flat
Prohibit,
surface
Prohibit,
to touch
Compulsion,
disconnect the
power plug
Compulsion,
periodical
inspection
2
Page 6
Cautions in Using with Safety
Fundamental Matters of "WARNING!" and "CAUTION!"
WARNING!
Do not use this unit in an area where there is flammable or explosive gas
Never use this unit in an area where there is flammable or explosive gas. This unit is not
explosion-proof. An arc may be generated when the power switch is turned on or off, and
fire/explosion may result. (Refer to page 56 "List of Dangerous Substances".)
Always ground this unit
Always ground this unit on the power equipment side in order to avoid electrical shock due to a power
surge.
If a problem occurs
If smoke or strange odor should come out of this unit for some reason, turn off the circuit breaker right
away, and then disconnect the power plug or power terminal. Immediately contact a service technician
for inspection. If this procedure is not followed, fire or electrical shock may result. Never perform repair
work yourself, since it is dangerous and not recommended.
Do not use the power cord if it is bundled or tangled
Do not use the power cord if it is bundled or tangled. If it is used in this manner, it can overheat and
fire may be caused.
Do not process, bend, wring, or stretch the power cord forcibly
Do not process, bend, wring, or stretch the power cord forcibly. Fire or electrical shock may result.
Substances that can not be used
Never use explosive substances, flammable substances and substances that include explosive or
flammable ingredients in this unit. Explosion or fire may occur. (Refer to page 56 "List of Dangerous
Substances".)
Do not disassemble or modify this unit
Do not disassemble or modify this unit. Fire or electrical shock or failure may be caused.
Do not get close to the vapor outlet / Do not block the outlet
The vapor outlet is provided on the left face of equipment. Do not put your hands or face close to the
outlet. Do not block the outlet. A burn injury or equipment failure may result in.
3
Page 7
Cautions in Using with Safety
Fundamental Matters of "WARNING!" and "CAUTION!"
CAUTION!
When opening the cover...
Make sure that the pressure of equipment has decreased to 0(zero) MPa before
opening the cover. Generally the cover does not open due to the safety lock
mechanism under the high pressure condition. The high-temperature and pressure
vapor blows out if the cover is forced open under high pressure, which may cause a
burn injury. A large amount of vapor blows out from inside of the chamber when
opening the cover just after the sterilizing operation has completed (when the
temperature inside the chamber is high). Do not put your hands and face close to the
cover.
When draining water...
Make sure that the pressure of equipment has decreased to 0(zero) MPa before draining the sterilizing
water. The hot water blows out if the valve is opened under high pressure. The sterilizing water
remains very hot just after the sterilizing operation has completed even the pressure reading is 0(zero)
MPa. Drain the water after it is sufficiently cooled down.
Do not touch the drain bottle during operation
A drain bottle, which contains hot water during and just after operation, is placed inside the door in the
front face of equipment. To avoid a burn injury, remove the bottle after the water is sufficiently cooled
down. Do not open the door during the operation of equipment.
Make sure to drain the water when the water level comes to the seal position
The hot water or vapor may blow out from the drain bottle if the equipment is operated with too much
drain water (water level above the seal position). (Refer to 17 of “Precautions for continuous operation”
in Page 13 for details.)
Securely fix the silicon plug of the drain bottle
Securely fix the silicon plug when installing the drain bottle. The hot water or vapor may blow out from
the drain bottle if the equipment is operated with the plug loosen. (Refer to 6 of “Set the drain bottle” in
Page 10 for details.)
Pour off the water inside the vapor cup after every operation
The cup becomes full with the water after the equipment is operated a few times. Pour off the water
inside the cup after every operation. The hot water or vapor may blow out from the drain bottle if the
cup is full with water. (Refer to 14 of “Attach the vapor cup” in Page 12 for details.)
Do not touch the hot section
Some sections on the equipment such as the circumference of cover or drain bottle are very hot during
or just after the operation of equipment. Do not touch these sections to avoid burn injury.
4
Page 8
Cautions in Using with Safety
Fundamental Matters of "WARNING!" and "CAUTION!"
When taking the sterile samples from the chamber...
Sufficiently remove the vapor inside the chamber before taking the sterile samples from the chamber.
Wear heat-resistant leather gloves to take them from the chamber to protect your hands from
high-temperature samples.
Do not touch the heat releasing outlet
Do not directly touch the heat releasing outlet located around the outer covering. The vapor may blow
out from the safety valve by an accident during sterilizing operation. Do not block the outlet.
During a thunder storm
During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and the main
power. If this procedure is not followed, fire or electrical shock may be caused.
When power failure occurs...
The lock lever on the cover goes into locked state for safety reasons when the power is turned off due
to a power failure. The state is automatically cancelled when the power is turned on and the pressure
inside the equipment decreases.
Do not operate the equipment without supplying sufficient amount of water
Do not operate the equipment without supplying sufficient amount of water. The heater is exposed to
the open-air if the amount of water supplied is insufficient, which causes a deterioration or breakage of
equipment. Make sure before operation that the appropriate amount of water is supplied inside the
chamber. (Refer to 9 of “Pour water into the chamber” in Page 11 for details.)
Do not open the panel on the outer covering
Touching the interior portion of equipment may cause an electric shock, burn injury, fire disaster or
equipment failure.
Do not touch the power plug with a wet hand
An electric shock may result in.
Do not place your hand over the top board
The hand may be stuck in the cover and injured.
5
Page 9
Requirements for Installation
1. Choose a proper place for installation
• Do not install this unit in a place where:
♦ Rough or dirty surface.
♦ Flammable gas or corrosive gas is generated.
♦ Ambient temperature 35°C and above or 5°C and below.
♦ Ambient temperature fluctuates violently.
♦ There is direct sunlight.
♦ There is excessive humidity and dust.
♦ There is a constant vibration.
♦ Not horizontal surface.
♦ The power source is instable.
• Install this unit on a stable place with the space as shown below.
Back side
Before Using This Unit
30cm or more
15cm
or
more
2. Installation on horizontal surface
• Use the equipment on the horizontal and firm place to
keep the water inside the chamber horizontal. If the
equipment tilts and the heater appears from the water
surface, the temperature on the area above the water
rises and a heater failure or operation stop due to
water level detector function may occur.
• The weight of main unit is approximately 65 to 95kg.
Carry and install the equipment carefully by two or
more persons.
15cm
or
more
Front side
15cm or more
6
Page 10
Requirements for Installation
3. Before/after installing
• It may cause injure to a person if this unit falls down or moves by the earthquake and the
impact. etc..To prevent, take measures that the unit cannot fall down, and not install to busy
place.
4. Do not install the equipment near alarm device
• The equipment releases large amount of vapor when the cover
is opened just after the operation is completed. Accordingly,
do not install the equipment on the site over which electrical
equipment especially an alarm device is provided over it.
Before Using This Unit
5. Ventilate the equipment sufficiently
• Do not block the heat releasing outlets on the side face and back face of equipment during
operation. The temperature inside the equipment rises, which may cause the deterioration or
failure of equipment, accident, or fire disaster.
6. Do not use this unit in an area where there is flammable or explosive gas
• Never use this unit in an area where
there is flammable or explosive gas.
This unit is not explosion-proof. An arc
may be generated when the power
switch is turned ON or OFF, and
fire/explosion may result.
• To know about flammable or explosive
gas, refer to page 56 “List of Dangerous
Substances”.
爆発性ガス
可燃性ガス
7
Page 11
Before Using This Unit
Requirements for Installation
7. Choose a correct power distribution board or receptacle
• Choose a correct power distribution board or receptacle that meets the unit’s rated electric
capacity.
• Operating voltage range for respective equipment models are as follows.
SN200/300/500 and SQ500 models: AC100 to 120V
SN210/310/510 and SQ510models: AC200 to 240V
Electric capacity:
NOTE)
There could be the case that the unit does not run even after turning ON the power. Inspect
whether the voltage of the main power is lowered than the specified value, or whether other
device(s) uses the same power line of this unit. If the phenomena might be found, change the
power line of this unit to the other power line.
• Starburst connection with a branching receptacle or extended wiring with a cord reel lowers
electrical power voltage, which may cause the degradation of refrigeration capability.
• Connect the unit to only the power supply. If it is connected to a gas pipe, water pipe or
telephone line, an accident or malfunction may result.
• Do not entangle the power cord. This will cause overheating and possibly a fire.
• Do not bend or twist the power cord, or apply excessive tension to it. This may cause a fire
and electrical shock.
• Do not lay the power cord under a desk or chair, and do not allow it to be pinched in order to
prevent it from being damaged and to avoid a fire or electrical shock.
• Keep the power cord away from any heating equipment such as a room heater. The cord's
insulation may melt and cause a fire or electrical shock.
• If the power cord becomes damaged (wiring exposed, breakage, etc.), immediately turn off the
power at the rear of this unit and shut off the main supply power. Then contact your nearest
dealer for replacement of the power cord. Leaving it may cause a fire or electrical shock.
• Connect the power plug to the receptacle which is supplied appropriate power and voltage.
8
Page 12
Requirements for Installation
9. Always ground this unit
• Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor
or ground terminal to prevent accidents caused by electric leakage.
• Do not connect the earth wire to gas or water pipes. If not, fire disaster may be caused.
• Do not connect the earth wire to the ground for telephone wire or lightning conductor. If not,
fire disaster or electric shock may be caused.
• Please consult your local electrical contractor for power connecting work.
• Do not use a branching receptacle, which may cause the heat generation.
• The D class earth connecting works is required if no ground terminal is provided. In this
case, consult with the selling office where you purchased or our sales office.
• Securely connect it to the switchboard or outlet.
Before Using This Unit
SN200
Earth connecting outlet
G
G
Power plug
An earth connecting outlet is recommended
to be used.
SN210/300/310/500/510, SQ500/510
Rounded terminal
These models do not include the power plug. Correctly
connect the ground to fit with the power supply facility to be
connected.
for M4
Green (to ground)
Black (to power supply)
White (to power supply)
10. Connect the power cord paying attention to the color of each core wire
• When connecting the power cord, do check the breaker
on the electric power equipment be "OFF". Note that
SN210/300/310/500/510 and SQ500/510 do not equip
with the power plug. Select and connect the
appropriate plug or terminal corresponding to the power
Core Wire Color Interior Wiring
Black Power Supply Side
White Power Supply Side
Green Ground Wire Side
capacity that is adjusted to the status of the power
supply equipment side.
9
Page 13
Installation Procedure
1 Determine the installation site
• If there is a bump on the floor, the casters may receive excessive load and get damaged. In
this case, lift and move carefully by two or more persons.
Install the equipment referring to 1 to 5 of "Requirements for Installation" in Page 6.
Before Using This Unit
2 After the unit is placed in the desired position,
lock the stopper button of the casters
• Only the two casters on the front side of the unit
are equipped with a stopper.
3 Connecting the power
• Connect the power referring to 7 to 10 of "Requirements for Installation" in Page 6.
4 Open the cover
• Turn on the electric leakage breaker at the right
side of equipment and turn on the power to
open the cover. The safety lock is released
and the cover can be opened.
• Slide the lock lever on the cover to the right and
grasp the handle of cover to open it.
5 Pour water into the drain bottle
• Pour 1500cc of water into the drain bottle. The water
is used to cool down the high-temperature vapor
generated inside the chamber.
Lock
Turn the lock lever to the left to close the cover.
Turn the lock lever to the right to open the cover.
Release
6 Set the drain bottle
• During operation, high-temperature vapor blows
out from the equipment. Make sure to set the
drain bottle. If the bottle is not set, the safety
device (micro switch) functions to prevent the
equipment from operating.
① Open the door in the front face of equipment.
② Put the exhaust hose silencer into the drain
bottle which contains 1500cc of water in it.
③ Set the bottle to the inside of door. Insert the
exhaust hose into the bottle and fix the silicon
plug onto the opening of bottle.
④ The micro switch is pressed and the equipment
is ready to operate. The temperature display
screen and time display screen are displayed.
Silicon plug
Exhaust hose
Silencer
10
Page 14
Installation Procedure
7 Set the attached drainboard onto the bottom
surface inside the chamber
• The drainboard stabilizes the sterile samples
inside the chamber as well as protects the
heater and sensor. Make sure to set it.
Before Using This Unit
Set the drainboard
horizontally.
8 Close the drain valve
• Close the drain valve at the bottom on the left side face of main
unit. Water leak occurs if not fully closed, which may cause
the burn injury or no-load (water) operation.
• Connect an appropriate hose (inner diameter: 12mm) to the
drain valve and lead it to the draining site.
9 Pour water into the chamber
• Before setting the sterile samples, pour water into the
chamber to the water level gauge (notch) position.
• Insufficient water may cause the no-load (water)
operation. Check the water level every time before
operation. Refill it before the level becomes too low.
Water is required to be poured at dissolution
operation, as well as sterilizing operation.
• When the water level lowers, the equipment detects
an abnormality (“Err20”) and cuts off the heater.
Depending on the conditions of equipment, however,
the detection requires too much time, which may
cause the heater deterioration. Refill water before
the water level becomes too low.
• Refer to the right table for the quantity of water to be
refilled.
Water level
gauge
High
Water QTY for models
SN200/210
SN300/310
SN500/510
SQ500/510
Close
Open
Low
2200~2300 ㏄
3500~3700 ㏄
3500~3700 ㏄
6300~6500 ㏄
10 Use distillated or purified water for sterilizing water
• Fill distillated or purified water inside the chamber. Tap water may be used. Calculus
generate inside the chamber when tap water is used. Frequent cleaning is therefore required.
• Do not use well water. It may cause the corrosion or dirt inside the chamber.
11
Page 15
Installation Procedure
11 Set the sterile samples
• Set the samples to the chamber, putting them into the attached rack or cast (sold separately).
• Put the sample or sterilization bag into the chamber so they should not block or cover the
sensor inside the chamber, exhaust outlet and end connection to pressure gauge. If they are
blocked or covered, the vapor cannot be discharged and the equipment cannot be operated
correctly. Do not spill the samples when taking them out from/putting them into the chamber.
The failure in piping system, bad smell or dirt may result in.
• In case liquid such as medicinal solution or medium is sterilized, the amount of liquid should be
60% or less of the capacity of container. They may be boiled over if too much quantity is
supplied.
• Widely open the opening of sterilization bag when used. If it is closed, the samples are
insufficiently sterilized.
12 Close the door before operation
• Make sure to close the door of equipment before operation. If the door is not fully closed, the
drain bottle falls down and may cause a burn injury. Do not open the door during operation.
Before Using This Unit
Close the cover
13
• Make sure that no foreign objects exist on the
packing of the cover and its contact area before
closing the cover. If any foreign object exists,
the vapor may leak from the inside.
• The lock lever on the cover is held by the hook and does not move when the cover remains
open.
• Fully close the cover and slide the lock lever on the cover to the left side. If it is closed
inappropriately, the vapor blows out from the inside, which may cause a burn injury.
• Do not press the hook and operate the lock lever for purposes other than maintenance of
equipment.
14 Attach the vapor cup
• The equipment is equipped with a vapor cup to
prevent the vapor that generate during air purge
from becoming water droplet and dropping down
onto the floor. Attach the cup onto the vapor
outlet on the left side face of equipment referring
to the right figure. Make sure to pour off the
water inside the vapor cup after every operation.
15 Attach the droplet tray
• Attach the droplet tray to the equipment to
prevent the waterdrops that are made from vapor
that generate during air purge from dropping
down from the packing onto the samples or onto
the top board. Pour off the droplets inside the
tray periodically.
Turn the lock lever to the left to close the cover.
Turn the lock lever to the right to open the cover.
Vapor cup
Droplet tray
12
Page 16
Installation Procedure
16 Precautions for drainage
• Before draining the water, make sure that the pressure, equipment temperature and water
temperature inside the chamber have decreased sufficiently.
17 Precautions for continuous operation
• When operating the equipment continuously after sterilization is completed, leave the
equipment for about 15 minutes with the cover opened to sufficiently lower the temperature
inside the chamber and then close the cover. If the temperature is high, the cover may not
close due to high internal pressure of chamber.
• Before operating the equipment again, check the water level of drain bottle by observing it from
the observation window to make sure that it is below the level indicated by the drain level seal
on the door. In case the water level of drain bottle is above the drain level seal, drain the
water until the water level comes to 1500cc of the water level gauge, indicated on the side face
of bottle. The hot water or vapor may blow out from the drain bottle if the equipment is
operated with too much drain water (above the seal position).
Before Using This Unit
13
Page 17
Before Using This Unit
Installation Procedure
Reference Data
Sterilizing operation using disposal bag for biochemically dangerous object
① Open the opening of sterilization bag so the vapor can be easily entered into it.
Secure the bag with a wire rack so it should not fall down during operation.
② The height of bag should be about two-thirds of chamber. If it is too high, the
vapor cannot be easily entered into it, or it blocks the vapor outlet at the upper
part of chamber, which may cause insufficient sterilization.
③ The preset temperature should be the upper temperature limit of bag to be used.
④ The preset time varies depending on the quality and quantity of sterile samples. Refer to the
following data for the preset time.
• Reference example of SN500 model at room temperature of 25℃
Sterile sample Sterilization temperature Sterilization timeNote
Gauze
Petri dish
121℃
121℃
30 min. Five bolts of dry gauze
40 min. 30 Petri dish with a cover
Sterilization
bag
The data above, however, is used as reference. The actual sterile condition varies depending on the
characteristics and quantity of samples or type of vessels to be used. Confirm the sterile condition
using the biological indicator or chemical indicator.
14
Page 18
Installation Procedure
Time lag
Before Using This Unit
a
℃
Saturated
vapor
temperature
bc
When sterilizing liquid samples, a time lag (b) is made
between the temperature inside the chamber and actual
temperature of liquid by the time when the liquid
temperature reaches the preset sterilization temperature.
For this reason, a longer time than defined by the
Japanese Pharmacopoeia (c) is required to completely
sterilize the samples. Consequently, the actual preset
sterilization time (a) should be set to be extended
The right table shows the time lag between the
temperature inside the chamber and actual temperature of
liquid (water). The table below shows the temperature
rise and cooling time with no load (liquid).
a= Preset sterilization time
b= Time lag
c= sterilization time defined by the
Japanese Pharmacopoeia
Temperature inside the chamber
Water temperature
Time①② ③④
Preset sterilization temperature:121℃/room
temperature: 25℃ (using conical flask)
Time lag
Load
SN200/210
/300/310
SN500/510
SQ500/510
500cc20min. 20min.
1000cc20min. 20min.
2000cc25min. 25min.
3000cc25min. 25min.
4000cc25min. 25min.
5000cc30min. 30min.
Necessary time for sterilization
At room temperature of 25℃(Course: apparatus sterilization/forced cooling: OFF)
1 Manual course key (MANUAL) This key is used to select manual course.
2 Melting & Warm course key
This key is used to select melting & warm course.
(MELT→WARM)
3 Sterilization & Warm course key
This key is used to select sterilization & warm course.
(STERILIZE→WARM)
4 Liquid sterilization course key
This key is used to select liquid sterilization course.
(STERILIZE/LIQUID)
5 Apparatus sterilization course key
This key is used to select apparatus sterilization course.
(STERILIZE/NORMAL)
6 Forced cooling key (COOLING FAN)This key is used to set/cancel the forced cooling.
7 MEMORY (SET MEMORY) This key is used to read/register the memory.
8 Start/stop key (START/STOP) This key is used to start/stop operation.
9 Time setting key (TIME) This key is used to set the time parameter.
10 Temperature setting key (TEMP) This key is used to set the temperature parameter.
11 Sample temperature key (SAMPLE) This key is used to set/cancel the optional sample
temperature sensor (sold separately).
12 Auto start key (AUTO START) This key is used to perform auto start operation. Press it
after selecting the course.
13 Maintenance key (MAINT.) This key is used for maintenance settings.
14 Preheating key (PRE HEAT) This key is used to set the preheating operation.
18
Page 22
Control Panel (Indicators)
16
17
15
Description and Function of Each Part
HEAT
STERILIZE/MELT
PRES.
14
EXHAUST
END
13
12
WARM
11
5
℃
TEMP
WARM
STERILIZE
/NORMAL
TIME
ENTERENTER
COOLING
FAN
STERILIZER
/MELT
SET
MEMORY
WARM
START/
STOP
6 7 8 9 10
18
19
MANUAL
1 2 3
MAINT.
PRE HEAT
MELT
WARM
AUTO
START
SAMPLE
STERILIZE
WARM
4
STERILIZER
/MELT
STERILIZE
/LIQUID
No. Name Function
1 Manual course lamp (MANUAL) This lamp lights up when manual course is selected.
Melting & Warm course lamp
2
This lamp lights up when melting & warm course is selected.
(MELT→WARM)
Sterilization & Warm course lamp
3
This lamp lights up when sterilization & warm course is selected.
(STERILIZE→WARM)
Sterilization/melting & warm
4
This lamp blinks at temperature setting of sterilization/melting & warm.
(STERILIZER/MELT, WARM)
temperature lamp
Liquid sterilization course lamp
5
This lamp lights up when liquid sterilization course is selected.
(STERILIZE/LIQUID)
Apparatus sterilization course lamp
6
This lamp lights up when apparatus sterilization course is selected.
(STERILIZE/NORMAL)
7 Forced cooling lamp (COOLING FAN) This lamp lights up when forced cooling process is selected.
8 MEMORY lamp (SET MEMORY) This lamp lights up at the setting of memory.
Sterilization/melting & warm
9
This lamp blinks at the setting of time for sterilization/melting & warm
(STERILIZER/MELT, WARM)
time lamp
10 Start/stop lamp (START/STOP) This lamp lights up during operation and goes out at operation stop.
11 Time display screen
The display blinks at the setting of time parameters. It displays the
remaining time during operation of equipment.
12 Heat-retention lamp (WARM) This lamp blinks at the setting of heat-retention process.
13 End lamp (END) This lamp blinks when operation is completed.
14 Operation monitoring lamp Refer to Page 20.
15 Pressurization lamp (PRES.) This lamp lights up when the equipment is being pressurized.
16 Temperature display screen
The display blinks at the setting of temperature parameters. It displays
the temperature inside the chamber during operation of equipment.
17 Auto start lamp (AUTO START)
This lamp lights up when auto start key is pressed. It blinks after the
auto start time is set.
Sample temperature lamp
18
(SAMPLE)
This lamp lights up when the sample temperature key (SAMPLE) is
selected when optional sample temperature sensor, which is sold
separately, is attached.
19 Preheating lamp (PRE HEAT)
This lamp lights up when the preheating key (PRE HEAT) is selected. It
blinks during the preheating operation.
19
Page 23
Description and Function of Each Part
Control Panel (Operation monitoring lamp)
STERILIZE/MELT
EXHAUST
HEAT
①
①・②
③・④・⑤
⑥・⑦
The current operating state of equipment can be checked by observing the status of respective lamp.
③ ④ ⑤
②
PRES.
No. Name
⑧
Heat-retention lamp (WARM)
END
Heat lamp (HEAT)
Sterilization/melting lamp
(STERILIZE/MELT)
Exhaust lamp (EXHAUST)
⑥
⑦
WARM
⑧
Lamp state
Operating state
Stop OFF OFF OFFOFFOFFOFFOFFOFF OFF OFF
During preheating Blink OFF OFFOFFOFFOFFOFFOFF OFF OFF
Before pressurization/during
temperature rising
During pressurization/during
temperature rising
When sterilization/melting
starts
After one-thirds of
sterilization/melting process
is completed
After two-thirds of
sterilization/melting process
is completed
When exhaust process
starts
When measured
temperature lowers to the
saturated vapor temperature
without heat-retention
process
When sterilization/melting is
completed without
heat-retention process
During heat-retention ON ON ON ONONONOFFBlink OFF OFF
When heat-retention
process is completed
①②③④⑤⑥⑦⑧
Blink OFF
ON BlinkOFFOFFOFFOFFOFFOFF
ONBlinkOFFOFFOFFOFFOFFONOFF
ON
ON
ONONBlinkOFFOFFOFFOFFONOFF
ON
ONONONBlinkOFFOFFOFFONOFF
ON
ONONONONBlinkOFFOFFOFFOFF
ON
ONONONONONBlinkOFFOFFOFF
ON
OFF OFF OFFOFFOFFOFFOFFOFFBlink
ON
OFF OFF OFFOFFOFFONON OFF Blink
OFFOFFOFFOFFOFFOFFOFFOFF
During
pressurization
ONOFF
End
※ When operation is stopped during pressurization/sterilization by the START/STOP key or a power failure, the lamps ③,
④ and ⑤ light off.
20
Page 24
Description and Function of Each Part
Characters of the Controller
The characters controller shows are as follows:
Character Identifier Name Purpose
yEAr
MntH
_dAy
timE
_Loc
PLoc
_bUZ
_Err
Accm
SmPL
PrEH
Calendar (year)
setting
Calendar (month)
setting
Calendar (day) setting
Time setting
Keylock
Pattern lock
Buzzer sound setting
Error log
Accumulated time
Sample temperature
sensor for sample
Preheating
temperature
Used to set calendar (year).
Used to set calendar (month).
Used to set calendar (day).
Used to set the time.
Locks the key to disable the change of preset
value.
Locks the key to disable the change of memory.
Used to set the buzzer sound.
Used to check the error log.
Used to check the accumulated time.
Used to check the sample temperature sensor
(optional accessory).
Used to set the preheating temperature.
End
bLAc
oUt
※ Refer to “Operation Course/Functional Setting Key and Character” in Page 25 for the operation courses and functional
characters.
Operation end
Reparation at power
failure
Displayed when operation is completed.
Displayed when power is restored during
operation.
21
Page 25
Operation Course/Function List
The equipment can carry out the following operation courses.
Name Description Page
This course sterilizes apparatus made of metal, glass, rubber and
ceramics.
• Press the STERILIZE/NORMAL key to go into the operation setting
mode.
Apparatus
sterilization course
• If other course has been selected, press the STERILIZE/NORMAL key
again.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Press the START/STOP key to start operation.
• Press the START/STOP key again to stop operation.
This course sterilizes liquid such as water, medium, test solution and test
reagent.
Operation Method
27
Liquid sterilization
course
Sterilization &
Warm course
• Press the STERILIZE/LIQUID key to go into the operation setting mode.
• If other course has been selected, press the STERILIZE/LIQUID key
again.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Press the START/STOP key to start operation.
• Press the START/STOP key again to stop operation.
This course sterilizes medium and then keeps them warm.
• Press the STERILIZE→WARM key to go into the operation setting
mode.
• If other course has been selected, press the STERILIZE→WARM key
again.
• Set the sterilization time and sterilization temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Set the heat-retention time and heat-retention temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Press the START/STOP key to start operation.
• Press the START/STOP key again to stop operation.
This course melts medium and then keeps them warm.
29
Melting & Warm
course
• Press the MELT→WARM key to go into the operation setting mode.
• If other course has been selected, press the MELT→WARM key again.
• Set the melting time and sterilization temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Set the heat-retention time and heat-retention temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Press the START/STOP key to start operation.
• Press the START/STOP key again to stop operation.
22
Page 26
Operation Method
Operation Course/Function List
Name Description Page
This course sterilizes and warms other samples other than the above.
• Press the MANUAL key to go into the operation setting mode.If other
course has been selected, press the MANUAL key again.
• Set the sterilization time and sterilization temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
Manual course
The current operation mode can not be changed during operation of equipment. To change the
operation mode, stop the operation of equipment.
• Set the time with the TIME key and then press the ENTER key.
• Set the heat-retention time and heat-retention temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Press the START/STOP key to start operation.
• Press the START/STOP key again to stop operation.
29
The equipment has the following functions.
Name Description Page
Calendar setting
Keylock function
Pattern lock function
Buzzer function
Error log display
Setting of sample
temperature sensor
This function is included in the maintenance mode.
It sets the dominical year, month, date and time.
This function is included in the maintenance mode.
It disables all key operations, except the START/STOP key operation
and cancellation of keylock state.
The “_Loc” is displayed if an unavailable key operation is done. (※)
This function is included in the maintenance mode.
It disables the change related to operation course and memory.
The “PLoc” is displayed if an unavailable key operation is done. (※)
This function is included in the maintenance mode.
It mutes the key operation sound except for the buzzer sounds at
warning and operation end.
This function is included in the maintenance mode.
It displays up to 20 errors occurred in the past, including the error
content and time of occurrence.
This function is included in the maintenance mode.
It enables the sample temperature function. If the setting is turned to
ON when the optional sample sensor is not attached, the “Er.8”, which
indicates disconnection of sample sensor, occurs.
31
Accumulated time
※: When both of keylock and pattern lock functions have been set, "_Loc" is displayed by priority.
This function is included in the maintenance mode.
It displays the accumulated current-carrying time by the hour.
23
Page 27
Operation Course/Function List
The operation functions of this unit are as follows;
Name Description Page
This function turns on the cooling fan during exhaust process to
shorten the cooling time.
The cooling fan is turned on during switching to the exhaust
process in the apparatus sterilization course. In other courses, it
starts to run at the saturated vapor temperature of -2℃ or less.
Forced cooling function
Preheating function
It stops when the equipment goes into the standby state after
operation is completed, or when the temperature inside the
chamber reaches 60℃.
The COOLING FAN key can be set anytime before and during
operation of equipment.
Pressing the COOLING FAN key lights the COOLING FAN lamp
and makes the function available.
This function keeps the temperature of feedwater inside the
chamber with the preset temperature.
The range of preset temperature is from 45℃ to 80℃. The
operation automatically ends after five hours.
Pressing the PRE HEAT key lights the PRE HEAT lamp. The
preset temperature is displayed with blinking. Set the desired
value and then press the ENTER key. This enables the function.
Operation Method
34
Memory function
Auto start function
Sample temperature
function (optional)
Temperature output
terminal (optional)
Time up output terminal
(optional)
Each operation course has three memory banks, where
registration and read of settings are possible. The following
settings can be stored into the memory.
・Sterilization (melting) temperature
・Sterilization (melting) time
・Heat-retention temperature
・Heat-retention time
・ON/OFF of forced cooling function
This function automatically starts the operation of equipment at
the specified time with the selected course.
The time can be set in increments of one minute within the range
from 00:00 to 23:59.
This function counts the sterilization/melting time by the
temperature measured by the sample temperature sensor
(optional).
Pressing the SAMPLE key lights the SAMPLE lamp. The
temperature display screen indicates the temperature measured
with the sample temperature sensor.
This function transmits and output the measured temperature of
controller at 4~20 mA.
This function outputs the relay (“a” contact) at operation end.
Contact spec: AC250V 1A (resistance load)
35
36
37
39
Alarm output terminal
(optional)
This function outputs the relay (“a” contact) at warning of
controller.
Contact spec: AC250V 1A (resistance load)
24
Page 28
Operation Method
Operation Course/Functional Setting Key and Character
The following key operations and characters are used in the setting of operation course and function.
Turn ON the
Breaker
Manual
MANUAL
Set the
sterilization
temp.
ENTER
Set the
sterilization
time
ENTER
Set the
heat-retention
temp.
Melting & Warm
MELT
▼
WARM
Set the
sterilization
temp.
ENTER
Set the
sterilization
time
ENTER
Set the
heat-retention
temp.
Sterilization
& Warm
STERILIZE
▼
WARM
Set the
sterilization
temp.
ENTER
Set the
sterilization
time
ENTER
Set the
heat-retention
temp.
TEMPTEMPTEMP
Liquid
sterilization
STERILIZE
/LIQUID
Set the
steriliza tion
temp.
ENTER
Set the
steriliza tion
time
ENTER
START
/STOP
Operation
starts
Apparatus
sterilization
STERILIZE
/NORMAL
Set the
sterilization
temp.
TEMPTEMPTEMPTEMPTEMP
ENTER
Set the
sterilization
time
TIMETIMETIMETIMETIME
ENTER
START
/STOP
Operation
starts
Maintenance
MAINT.
Calendar (year)
setting
yEAr
Keylock
setting
_Loc
on
TEMPTEMPTEMP
off
ENTER
Calendar
(month) setting
mntH
ENTER
Pattern lock
setting
PLoc
on
TEMPTEMP
off
ENTER
Calendar (day)
setting
ENTER
Buzzer sound
setting
Sample sensor
setting
SmPL
ENTER
Accumulated
time
Accm
ENTER
MAINT.
on
off
ENTER
Set the
heat-retention
time
ENTER
Set the
heat-retention
time
ENTER
Set the
heat-retention
time
_dAy
TEMPTEMP
_bUZ
on
off
TIMETIMETIME
ENTER
START
/STOP
Operation
starts
ENTER
START
/STOP
Operation
starts
ENTER
START
/STOP
Operation
starts
ENTER
Time
setting
timE
TEMPTEMP
ENTER
ENTER
Error log
Err
ENTER
25
Page 29
Operation Method
Operation Course/Functional Setting Key and Character
Turn ON the
Breaker
Forced cooling
function
COOLING
FAN
forced cooling
operation
starts
Preheating
function
PRE HEAT
Set the
preheating
temperature
PrEH
TEMP
ENTER
Operation
starts
Operation
START
/STOP
Operation
starts
Memory
function
Select the
operation course
SET
MEMORY
Select the
memory No.
SET
MEMORY
Registration
Press to select
SET
MEMORY
2 sec.
Auto start
function
Select the
operation course
AUTO
START
Set the
start time
TEMP
ENTER
START
/STOP
Auto start
operation
starts
Sample temperature
function
Select the
operation course
SAMPLE
START
/STOP
Operation
starts
26
Page 30
Operation Method
Apparatus/Liquid Sterilization Course
Follow the procedures below for the setting of apparatus sterilization and liquid sterilization courses.
1. Turn on the earth leakage breaker
• The temperature display screen and time display screen
display the current temperature inside the chamber and
current time (standby state) respectively when the breaker is
turned to on.
MANUAL
MANUAL
MANUAL
STERILIZER
/MELT
PRE HEAT
PRE HEAT
PRE HEAT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
①
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
②
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
displays current temperature inside the chamber.
Time display screen:
displays current time.
2. Select the operation course
• Press either the STERILIZE/NORMAL or STERILIZE/LIQUID
key once.
Press the key twice if other course has been already selected.
3. Set the sterilization temperature
① The sterilization temperature currently set is displayed with
blinking on the temperature display screen. The
STERILIZER/MELT lamp blinks. Press the ∧∨ on the
TEMP key and set the desired sterilization temperature.
② After setting the temperature, press the ENTER key to
determine the value.
Temperature display screen:
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
①
4. Set the sterilization time
① The sterilization time currently set is displayed with blinking
on the time display screen. The STERILIZER/MELT lamp
blinks. Press the ∧∨ on the TIME key and set the desired
sterilization time.
②
② After setting the time, press the ENTER key to determine the
value.
STERILIZE/MELT
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
MANUAL
PRE HEAT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
5. Start operation
• Press the START/STOP key.
The equipment starts operation with the HEAT lamp blinks.
27
Page 31
Operation Method
Apparatus/Liquid Sterilization Course
To change or confirm the preset value during operation…
Preset values can not be changed while the equipment is in operation. Before starting operation, make
sure to confirm them and change them if required.
Press the course key currently being operated to confirm the preset value while the equipment is in
operation. The preset value has been displayed while the key is pressed.
6. End operation
• The time display screen displays “END” when the operation is
completed.
Press the START/STOP key. The equipment goes into the
standby state. The lever on the cover is unlocked and the
cover can be opened.
To abort operation…
Press the START/STOP key. The equipment stops operation.
In case the operation is aborted before the pressurization process, the END lamp lights up and the time
display screen displays “END” without a buzzer sound.
In case the operation is aborted after the pressurization process, the equipment goes into the exhaust
process.
Power-off during operation
If the power is turned off due to a power failure, the operation is aborted. After the recovery, the “bLAc
oUt” is displayed to notify the power failure. Press any key to delete the “bLAc oUt”.
In case the operation is aborted before the pressurization process, the END lamp lights up and the time
display screen displays “END” without a buzzer sound.
In case the operation is aborted after the pressurization process, the equipment goes into the exhaust
process.
28
Page 32
Operation Method
Sterilization & Warm/Melting & Warm/Manual Course
Follow the procedures below for the setting of sterilization & warm, melting & warm and manual course.
1. Turn on the earth leakage breaker
• The temperature display screen and time display screen
display the current temperature inside the chamber and
current time (standby state) respectively when the breaker is
turned to on.
MANUAL
MANUAL
MANUAL
STERILIZER
/MELT
PRE HEAT
PRE HEAT
PRE HEAT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
①
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
②
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
displays current temperature inside the chamber.
Time display screen:
displays current time.
2. Select the operation course
• Press either the STERILIZE→WARM or MELT→WARM or
MANUAL key once.
Press the key twice if other course has been already selected.
3. Set the sterilization temperature
① The sterilization temperature currently set is displayed with
blinking on the temperature display screen. The
STERILIZER/MELT lamp blinks. Press the ∧∨ on the
TEMP key and set the desired sterilization temperature.
② After setting the temperature, press the ENTER key to
determine the value.
Temperature display screen:
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
①
4. Set the sterilization time
① The sterilization time currently set is displayed with blinking
on the time display screen. The STERILIZER/MELT lamp
blinks. Press the ∧∨ on the TIME key and set the desired
sterilization time.
②
② After setting the time, press the ENTER key to determine the
value.
MANUAL
PRE HEAT
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
①
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
②
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
5. Set the heat-retention temperature
① The current heat-retention temperature is displayed with
blinking on the temperature display screen. The
STERILIZER/MELT lamp blinks. Press the ∧∨ on the
TEMP key and set the desired heat-retention temperature.
② After setting the temperature, press the ENTER key to
determine the value. The setting of heat-retention time will
follow.
29
Page 33
Sterilization & Warm/Melting & Warm/Manual Course
MANUAL
PRE HEAT
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
①
6. Set the heat-retention time
① The current heat-retention time is displayed with blinking on
the time display screen. The STERILIZER/MELT lamp
blinks. Press the ∧∨ on the TIME key and set the desired
heat-retention time.
②
Set "00:00" when skipping the heat-retention process in the
manual operation.
② After setting the time, press the ENTER key to determine the
value.
Operation Method
MANUAL
PRE HEAT
STERILIZE/MELT
HEAT
AUTO
MAINT.
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
WARM
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
7. Start operation
• Press the START/STOP key.
The equipment starts operation with the HEAT lamp blinks.
8. End operation
• The time display screen displays “END” when the operation is
completed.
Press the START/STOP key. The equipment goes into the
standby state. The lever on the cover is unlocked and the
cover can be opened.
To change or confirm the preset value during operation…
Preset values can not be changed while the equipment is in operation. Before starting operation, make
sure to confirm them and change them if required.
Press the course key currently being operated to confirm the preset value while the equipment is in
operation. The preset value has been displayed while the key is pressed.
To abort operation…
Press the START/STOP key. The equipment stops operation.
In case the operation is aborted before the pressurization process, the END lamp lights up and the time
display screen displays “END” without a buzzer sound.
In case the operation is aborted after the pressurization process, the equipment goes into the exhaust
process.
Power-off during operation
If the power is turned off due to a power failure, the operation is aborted. After the recovery, the “bLAc
oUt” is displayed to notify the power failure. Press any key to delete the “bLAc oUt”.
In case the operation is aborted before the pressurization process, the END lamp lights up and the time
display screen displays “END” without a buzzer sound.
In case the operation is aborted after the pressurization process, the equipment goes into the exhaust
process.
The operation continues if aborted during the heat-retention process.
30
Page 34
Operation Method
Maintenance Mode
Set the following functions;
Calendar, time, lock, pattern lock, buzzer sound, error log, sample temperature and accumulated
time.
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
FAN
/NORMAL
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
① Press the MAINT. key.
STERILIZER
/MELT
STERILIZER
/MELT
STERILIZER
/MELT
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
WARM
WARM
WARM
WARM
TIME
TIME
TIME
TIME
② The calendar (year) setting screen is displayed.
The time display screen displays the character "yEAr", which
indicates the calendar (year). The temperature display
screen displays the value (year) with blinking.
Press the ∧∨ on the TEMP key and set the desired year.
Press the ENTER key to determine the value.
③ The calendar (month) setting screen is displayed.
The time display screen displays the character "mntH", which
indicates the calendar (month). The temperature display
screen displays the value (month) with blinking.
Press the ∧∨ on the TEMP key and set the desired month.
Press the ENTER key to determine the value.
④ The calendar (date) setting screen is displayed.
The time display screen displays the character "_dAy", which
indicates the calendar (date). The temperature display
screen displays the value (date) with blinking.
Press the ∧∨ on the TEMP key and set the desired date.
Press the ENTER key to determine the value.
⑤ The time setting screen is displayed.
The time display screen displays the character "timE", which
indicates the time. The temperature display screen displays
the value (time) with blinking.
Press the ∧∨ on the TEMP key and set the desired time.
Press the ENTER key to determine the value.
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
⑥ The lock setting screen is displayed.
The time display screen displays the character "_Loc", which
indicates the lock setting. The temperature display screen
displays "on" or "oFF" with blinking.
Press the ∧∨ on the TEMP key to select "on" or "oFF".
Press the ENTER key to determine the setting.
(The setting is "oFF" at factory shipment.)
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
⑦ The pattern lock setting screen is displayed.
The time display screen displays the character "PLoc", which
indicates the pattern lock setting. The temperature display
screen displays "on" or "oFF" with blinking.
Press the ∧∨ on the TEMP key to select "on" or "oFF".
Press the ENTER key to determine the setting.
(The setting is "oFF" at factory shipment.)
31
Page 35
Maintenance Mode
TEMP
ENTERENTER
STERILIZER
/MELT
STERILIZER
/MELT
℃
WARM
WARM
TIME
Operation Method
⑧ The buzzer setting screen is displayed.
The time display screen displays the character "_bUZ", which
indicates the buzzer setting. The temperature display screen
displays "on" or "oFF" with blinking.
Press the ∧∨ on the TEMP key to select "on" or "oFF".
Press the ENTER key to determine the setting.
(The setting is "on" at factory shipment.)
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
⑨ The error log screen is displayed.
The time display screen displays the character "_Err", which
indicates the error log. The temperature display screen
displays nothing.
Press the ∧∨ on the TEMP key to check the error log.
The content of errors and their time of occurrence are
displayed.
Press the ENTER key to go into the setting screen for sample
temperature.
Error log display
For example, Error 1 occurred at 11:59 p.m. on December 31, 2005;
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
⇔
Displayed
alternately every
two seconds
Press the ∧∨ on the TEMP key to select the error No.
℃
STERILIZER
/MELT
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
STERILIZER
/MELT
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
⑩ The sample temperature setting screen is displayed.
The time display screen displays the character "SmPL", which
indicates the sample temperature setting. The temperature
display screen displays "on" or "oFF" with blinking.
Press the ∧∨ on the TEMP key to select "on" or "oFF".
Press the ENTER key to determine the setting.
(Set "on" here when an optional accessory is attached. The
setting is "oFF" at factory shipment.)
32
Page 36
Maintenance Mode
℃
STERILIZER
WARM
/MELT
Ex):
when the accumulated time is 39999
hours
TEMP
ENTERENTER
STERILIZER
/MELT
STERILIZER
/MELT
WARM
WARM
Operation Method
⑪ The accumulated time screen is displayed.
TIME
The time display screen displays the character "Accm", which
indicates the accumulated time. The temperature display
screen displays the accumulated time.
Check the time and press the ENTER key to determine it.
The accumulated time function accumulates the current-carrying
time by the hour.
・When the accumulated time is 10000 hours or more:
→The decimal-point character for one-hour unit lights up.
・When the accumulated time is 20000 hours or more:
→The decimal-point character for ten-hour unit lights up.
・When the accumulated time is 3000 hours or more:
→The decimal-point character for hundred-hour unit lights up.
・This function does not accumulate time more than 4000 hours.(The maximum is 39999 hours.)
The accumulated time can be reset by pressing the ∧ and ∨
on the TEMP key at the same time while the accumulated time is
displayed.
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
MELT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
⑫ The setting returns to Step ①.
Press the MAINT. key to terminate the setting procedures.
33
Page 37
Operation Method
Forced Cooling Function / Preheating Function
Forced cooling function
This function turns on the cooling fan during exhaust process to shorten the cooling time.
The cooling fan is turned on during switching to the exhaust process in the apparatus sterilization
course. In other courses, it starts to run at the saturated vapor temperature (refer to the graph on
Page 15) of -2℃ or less. It stops when the equipment goes into the standby state after operation is
completed, or when the temperature inside the chamber reaches 60℃. The function can be set
anytime before and during operation of equipment.
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
MELT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
MEMORY
FAN
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
① Press the COOLING FAN key.
The COOLING FAN lamp lights up and the function becomes
available.
Preheating function
This function keeps the temperature of feedwater inside the chamber with the preset temperature.
The range of preset temperature is from 45℃ to 80℃. The operation automatically ends after five
hours.
The preheating operation function remains after the power is restored due to a power failure.
STERILIZE/MELT
MANUAL
STERILIZER
/MELT
MAINT.
PRE HEAT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
FAN
/NORMAL
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
℃
TEMP
ENTERENTER
WARM
STERILIZER
/MELT
TIME
WARM
① Press the PRE HEAT key.
② The preheating function screen is displayed.
The time display screen displays the character "PrEH", which
indicates the preheating function. The temperature display
screen displays the value (temperature). The temperature
display screen displays with blinking the preheating
temperature currently set and the STERILIZE/MELT lamp
blinks.
Press the ∧∨ on the TEMP key to set the desired preheating
temperature.
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
FAN
/NORMAL
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
③ After setting the temperature, press the ENTER key or
START/STOP key to determine the value. This starts the
preheating operation.
④ Press the PRE HEAT key to abort the preheating operation.
34
Page 38
Operation Method
Memory Function
Memory function
Each operation course has three memory banks, where registration and read of settings are possible.
The following settings can be stored into the memory.
・Sterilization (melting) temperature
・Sterilization (melting) time
・Heat-retention temperature
・Heat-retention time
・ON/OFF of forced cooling function
1. Register the setting to the memory.
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
FAN
/NORMAL
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
① Make sure that the course where the memory is to be
registered is selected and then press the SET MEMORY key.
The SET MEMORY lamp blinks.
MANUAL
MAINT.
PRE HEAT
MELT
WARM
SAMPLE
AUTO
START
STERILIZE
WARM
STERILIZER
/MELT
STERILIZE
/LIQUID
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
TIME
WARM
START/
STOP
② The memory registration screen is displayed.
The memory No. is displayed on the highest-order digit of
temperature display screen. The temperature/time setting
display for sterilization or melting switches to 1, 2, 3 or normal
mode in this order every time the SET MEMORY key is
pressed.
As for the courses that include the heat-retention setting, the
temperature and time of the setting can be checked by
pressing the ENTER key.
The COOLING FAN lamp lights up when forced cooling
function is set to "on".
STERILIZER
/MELT
STERILIZE
/LIQUID
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
TIME
WARM
START/
STOP
2sec.
③ Select the memory No. that the setting is overwritten and then
press the SET MEMORY key for two seconds. The displayed
preset value changes. This completes the registration of
setting.
MANUAL
MAINT.
PRE HEAT
MELT
WARM
SAMPLE
AUTO
START
STERILIZE
WARM
2. Operation procedures with the registered setting
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
FAN
/NORMAL
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
① Make sure that the course to be operated is selected and then
press the SET MEMORY key. The SET MEMORY lamp blinks.
MANUAL
MAINT.
PRE HEAT
MELT
WARM
SAMPLE
AUTO
START
STERILIZE
WARM
STERILIZER
/MELT
STERILIZE
/LIQUID
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
MEMORY
FAN
STERILIZER
/MELT
SET
TIME
WARM
START/
STOP
STERILIZER
/MELT
STERILIZE
/LIQUID
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
FAN
MEMORY
STERILIZER
/MELT
SET
TIME
WARM
START/
STOP
MANUAL
MAINT.
PRE HEAT
MELT
WARM
SAMPLE
AUTO
START
STERILIZE
WARM
②
② The memory confirmation screen registered is displayed.
Select the memory No. by the SET MEMORY key.
③ Press the START/STOP key. The setting being displayed is
read. The equipment starts operation using the setting.
③
35
Page 39
Operation Method
Auto Start Function
Auto start function
This function automatically starts the operation of equipment at the specified time with the selected
course. The time can be set in increments of one minute within the range from 00:00 to 23:59.
The auto start function remains after the power is restored due to a power failure.
STERILIZE/MELT
MANUAL
MAINT.
PRE HEAT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
MELT
/LIQUID
WARM
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
FAN
/NORMAL
MEMORY
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
TIME
STERILIZER
/MELT
WARM
ENTER
① Make sure that the course to be operated is selected and then
press the AUTO START key. The AUTO START lamp lights up.
② The auto start time screen is displayed.
The time display screen displays the auto start time currently
set with blinking. Press the ∧∨ on the TIME key and set the
desired time.
When starting the operation at 12:00…
TIME
START/
STOP
WARM
ENTER
TIME
START/
STOP
AUTO
START
SAMPLE
STERILIZER
/MELT
STERILIZER
/MELT
WARM
ENTER
③ Set the desired auto start time and then press the ENTER key
or START/STOP key to determine the value.
④ The AUTO START lamp continues blinking until the equipment
starts operation.
Press the START/STOP key while the AUTO START lamp is
blinking to confirm the operation start time.
The procedure above is possible only when the cover on the
main unit is locked.
⑤ Press the AUTO START key to abort the preheating operation.
36
Page 40
Operation Method
Sample Temperature Function
Sample temperature function
This function counts the sterilization/melting time by the temperature measured by the sample
temperature sensor. Set "on" to the setting of sample temperature sensor in the maintenance setting.
STERILIZE/MELT
EXHAUST
PRES.
END
℃
TEMP
ENTERENTER
WARM
COOLING
STERILIZE
/NORMAL
MEMORY
FAN
STERILIZER
/MELT
SET
WARM
TIME
WARM
START/
STOP
MANUAL
MAINT.
PRE HEAT
MELT
WARM
HEAT
AUTO
START
STERILIZER
SAMPLE
/MELT
STERILIZE
STERILIZE
/LIQUID
WARM
① Make sure that the course to be operated is selected and then
press the SAMPLE key.
The SAMPLE lamp lights up and the temperature display
screen displays the temperature measured by the sample
temperature sensor.
② Press the START/STOP key to start the operation. During the
sterilization and melting processes, the display can not be
switched.
37
Page 41
Sample Temperature Function
The preset values at factory shipment are as follows.
The initial setting values of operation
Operation Method
Course
Apparatus
sterilization
Liquid
sterilization
Sterilization
& Warm
Melting
& Warm
Manual
Sterilization
temp.
121℃
121℃
121℃
-- 100℃
121℃
Sterilization
time
20 min.
20 min.
20 min.
20 min.
The initial setting values of function
Function Value
Preheating
Forced cooling OFF
Key lock OFF
Melting
temp.
- - - -
- - - -
- - 50℃
- - 50℃
Melting
time
10 min.
Heat-retention
45℃
Heat-retention
temp.
50℃
time
2 hours
2 hours
2 hours
Pattern lock OFF
Buzzer ON
Error log
Sample temperature OFF
Accumulated time 0 hour
-
38
Page 42
External Output Terminal (optional)
Precautions
• Operate this product according to the procedure described in this Operation Manual. Failure to
follow the operation procedure described herein may result in a problem. The guarantee will
not apply if you operate the product in the wrong manner.
CAUTION!
• Turn off the breaker before connecting the cables.
• Connect a recorder or another appliance of 600 W or less in input impedance to the
temperature output terminal.
• Securely fasten all connections with the screws attached to the terminal block.
Operation Method
Connection procedure
• Connect the cables to the appropriate terminals.
• When using temperature output, use a shielded wire for the cable to be connected to prevent
noise.
ANALOG
+-COMNO
Connection terminal
TIME UP
ALARM
COM NO
39
Page 43
External Output Terminal (optional)
Specification
• The current (mA) corresponding to the measured temperature is output.
• Output temperature range: 0 to 160℃
Temperature Output
(ANALOG)
Time-up Output
(TIME UP)
• Output current: 4 to 20mA
• Load: 600Ω or bellow
• Resolution: ±2℃ (±0.2mA)
• Connection: M4 screw terminal block
• It is output when operation is completed, including an abortion.
• a-contact (relay contact)
• Contact capacity: 250V AC, 1A (resistance load)
• Connection: M4 screw terminal block
Operation Method
• It is output when an abnormality is detected. Refer to "Safety Device and
Error Code" in Page 45. It is not, however, output when an abnormality
Alarm Output
(ALARM)
in the bottle.
• a-contact (relay contact)
• Contact capacity: 250V AC, 1A (resistance load)
• Connection: M4 screw terminal block
Temperature/current output table
Temperature (℃)
0 4
20 6
40 8
60 10
80 12
100 14
120 16
Output current (mA)
140 18
160 20
40
Page 44
Handling Precautions
WARNING!
If a problem occurs
If smoke or strange odor should come out of this unit for some reason, turn off the power key
right away, and then turn off the circuit breaker and the main power. Immediately contact a
service technician for inspection. If this procedure is not followed, fire or electrical shock may
result. Never perform repair work yourself, since it is dangerous and not recommended.
Measure for flammability and handling of flammable solvent
This unit is not designed as the explosion-proof construction. Pay special attention to the
handling of the sample to be handled with this unit on the consumption with the explosive
material, flammable material, and similar ones. The flammable material may be vaporized by
leaving it at the temperature higher than room temperature, and could cause the fire or
explosion. When handling such material, provide ventilation with enough before the operation.
(Refer to page 56 “List of Dangerous Substances”.)
Keep the unit well-ventilated
Keep the heat releasing outlets in the side and back of the unit open during operation. If they
are closed, the inside temperature of the unit may increase, its performance may deteriorate, or
an accident, malfunction or fire may result.
Exercise care not to allow a liquid to get on the unit
Exercise care not to allow a liquid to get on the unit or enter the unit through the heat releasing
outlets in the side or back of the unit. If it enters the unit, stop the operation. Otherwise. an
accident, malfunction, electric shock or fire may result.
Do not drop metallic pieces into the unit
Do not drop metallic pieces, such as clips, staples and screws, into the unit. If such a metallic
piece has dropped into the unit, turn it off. An accident, malfunction, electric shock or fire may
result.
Do not open the panels and covers
Do not operate the unit with the fixed panels and covers open. An accident, malfunction or
electric shock may result.
Do not modify
Do not modify this unit. An accident, malfunction, electric shock or fire may result.
41
Page 45
Handling Precautions
CAUTION!
Do not step on this unit
Do not step on this unit. It will cause injury if this unit fall down or break.
Do not place or drop anything on the unit
Do not place or drop anything on the unit. Since the unit contains precision components, it may
malfunction due to vibration, impact, etc.
During a thunder storm
During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and
the main power. If this procedure is not followed, fire or electrical shock may be caused.
Countermeasure for stop operation during night or long-term stop
Turn off the power of earth leakage breaker and disconnect the power cord from the power
source before stopping the operation of equipment overnight or for a long time.
Do not touch the hot section
The temperature on the cover and top board on the chamber are very hot during operation or just
after operation is completed. Do not touch these sections to avoid a burn injury.
When opening the cover…
Make sure that the pressure gauge reading has decreased to 0(zero) MPa before opening the
cover. Open the cover slow carefully. The high-temperature and pressure vapor blows out if
the cover is opened during high pressure.
When opening/closing the door…
Do not put your hands or face into the traveling range (space) of door when it is opened or
closed. The door may contact, which may cause an injury.
When draining water…
The water in the chamber is very hot just after operation is completed. Be careful not to get a
burn injury. Drain the water after the water is sufficiently cooled down.
Do not drain water during operation. The hot water blows out if the drain valve is opened while
the pressure is increasing.
Do not damage the packing on the cover or flange on the chamber
Damage or dirt on these areas may cause the vapor leakage, which may be the cause of burn
injury. Keep these sections always clean. Do not damage them with the rack when taking out
and putting in the sterile samples. The packing degrades wit time. It must be replaced if vapor
leak occurs frequently. In this case consult with the selling office where you purchased or our
sales office.
Replace the packing early
The packing is a consumable. If it shows the sign of damage or hardening, replace it early.
Please consult with the selling office where you purchased or our sales office for the replacement
of packing.
Do not perform procedures other than described in this document
Do not perform procedures other than described in this document. Otherwise an unexpected
accident may occur.
42
Page 46
r
Maintenance Method
Daily Inspection and Maintenance
For the safety use of this unit, please perform the daily inspection and maintenance without fail.
Using the city water to this unit might attach dirt. Do inspect and maintain this point while performing
daily inspection and maintenance.
WARNING!
• Disconnect the power cable from the power source when doing an inspection or maintenance
unless needed.
• Perform the daily inspection and maintenance after returning the temperature of this unit to the
normal one.
• Do not disassemble this unit.
CAUTION!
• Use a well-drained soft cloth to wipe dirt on this unit. Do not use benzene, thinner or cleanser for
wiping. Do not scrub this unit. Deformation, deterioration or color change may result in.
Monthly maintenance
• Check the earth leakage breaker function.
1. Connect the power cord.
2. Turn the breaker on.
3. Push the red test switch by a ballpoint pen etc.
4. If there is no problem, the earth leakage breaker will be turned off.
• Clean the silencer.
The exhaust hose is equipped with a silencer on its end to reduce
the noise generated daring air purge.
・Remove the silence and wash it with water. It contains a backwater
prevention ball. Make sure not lose the ball when the silencer is
removed.
・After cleaning, put the ball first into the silencer and then fix the
silencer
Silence
Ball
Filter cleaning
• If the filter on the bottom of chamber is clogged with dust or dirt, the
equipment can not drain the water. Clean it appropriately as required.
・The filter is inserted in the drain outlet. Pull it out to sweep it.
・Insert it in place after cleaning.
Cleaning inside the chamber
• Use soft sponge to clean inside the chamber not to damage the surface inside the chamber. Do
not remove the filter on the bottom of chamber at cleaning. If it is removed, the pipe fitting is
clogged with dirt inside the chamber.
• The heater and sensor are provided on the bottom inside the chamber. Make sure not to bend or
damage the filter.
For any questions, contact the dealer who you purchased this unit from, or the nearest sales division in our company.
43
Page 47
Long storage and disposal
When not using this unit for long term / When disposing
CAUTION!
When not using this unit for long term…
• Turn off the power and disconnect the power cord.
WARNING!
When disposing…
• Keep out of reach of children.
• Consult with the specialized disposal services when disposing the equipment.
Environmental protection should be considered
We request you to disassemble this unit as possible and recycle the reusable parts considering to
the environmental protection. The feature components of this unit and materials used are
listed below.
Power cord Synthetic rubber coated wiring materials, copper and nickel
Piping Parts
Hoses Ethylene propylene tube
Pipes Copper, Copper alloy
44
Page 48
In the Event of Failure…
Safety Device and Error Code
This unit has an automatic diagnosis function built in the controller and safety devices independent of the
controller. The table below shows the cause and the solution method when the safety device operates.
Error Code:
If an error in use or equipment a failure occurs, the temperature display screen on the operating panel
displays the corresponding error code and the alarm buzzer sounds. The buzzer stops by pressing any key.
In case an error occurs, check the error code and turn off the earth leakage breaker.
Safety Device Notify Cause/Solution
• Failure in temperature input circuit.
Sensor trouble detection
“Er.01” appears
• Temperature sensor is broken or disconnected.
• The measured temperature is out of the display range.
Make a call for service.
SSR short-circuit
detection
Heater disconnecting
detection
Cover locking error
Cover unlocking error
Incorrect bottle position
Overheat error
Sample sensor (optional)
disconnection
“Er.02” appears
“Er.03” appears
“Er.04” appears
“Er.05” appears
“Er.06” appears
“Er.07” appears
“Er.08” appears
• SSR is in short-circuit
Make a call for service.
• Heater is disconnected.
Make a call for service.
• The cover is unlocked during operation.
Make a call for service.
• The cover is not unlocked at releasing.
Make a call for service.
• Bottle is attached incorrectly.
Open the door and attach the drain bottle correctly.
• The chamber temperature rises to 140℃ or above.
• The temperature of "preset temperature + (plus) 3℃
or above" is continued for one minute during
sterilization process.
• The temperature of "preset temperature + (plus) 10℃
or above" is continued for ten minutes during
heat-retention process.
Make a call for service.
• Disconnection or abnormality of sample sensor.
• The setting is set to "on" when the sensor is not
attached.
Check the setting referring to "Maintenance Mode" in
Page 31.
Make a call for service.
Exhaust valve error
A/D conversion error
Memory error
“Er.09” appears
“Er.14” appears
“Er.15” appears
• Failure in exhaust valve.
Make a call for service.
• Failure in electrical parts.
Make a call for service.
• Failure in preset value of memory.
Make a call for service.
45
Page 49
Safety Device and Error Code
In the Event of Failure…
Internal communication
error
Water level error
Safety valve
“Er.17” appears
“Er.20” appears
Safety valve is
operated.
• Communication error between the control board and
display board.
Make a call for service.
• Lack of water supply
Supply water.
Check the amount of water to be supplied referring to 9 of
"Pour water into the chamber" in Page 11. If the error is
not cancelled, contact our service department.
• Pressure rise inside the chamber or safety valve
failure.
Make a call for service.
46
Page 50
Trouble Shooting
Phenomenon Check point
In the Event of Failure…
The unit does not start to operate
although the leakage breaker is
turned on.
The screen displays the error code
and the alarm buzzer sounds.
Exhaust failure or safety valve is
operated.
Drain failure
Sterilization temperature does not
rise or pressure inside the
chamber does not rise.
Pressure inside the chamber rises
with the solenoid valve opened.
The temperature changes during
operation of equipment.
Too much vapor blows out, or hot
water blows out from the drain
bottle.
Water leaks.
Operation halts in standby state.
Large noise during air purge.
Cover does not open.
In the case if the error other than listed above occurred, turn off the power switch and primary power
source immediately. Contact the shop of your purchase or nearest Yamato Scientific Service Office.
• Check if the power cable is securely connected to the power
supply.
• Check if the power fails.
• Check the power voltage.
• Check the error code.
(refer to "Safety Device and Error Code" on page 45.)
• The hose to the drain bottle is twisted or clogged.
• The exhaust outlet inside the chamber is blocked with the sterile
samples.
• Too much samples are stored.
• The filter is clogged.
• The preset value is lower than the temperature inside the
chamber.
• The power supply voltage is low.
• The ambient temperature is too low.
• The cover is not securely closed.
• The packing or flange is damaged.
• The exhaust outlet inside the chamber is blocked.
• An inadequate preset temperature is set.
• Check if the power supply voltage is low.
• The variation in ambient temperature is too large.
• The drain bottle does not contain water.
• The water in the drain bottle is hot.
• The exhaust hose is removed or broken.
• The silicon plug is not fitted securely.
• The water in the vapor cup is full.
• The water level in the drain bottle is above the drain level.
• The drain valve is not securely closed.
• The drain bottle contains too much water.
• The cover is not securely closed.
Check it referring to the 13 of "Close the cover" in Page 12.
• Check if the silencer is removed,
Check the connection of silencer inside the drain bottle referring
to the 6 of "Set the drain bottle" in Page 10.
• The power is turned off.
• The sterilization process is not completed.
47
Page 51
After Service and Warranty
In Case of Request for Repair
If the failure occurs, stop the operation, turn OFF the power switch, and unplug the power plug.
Please contact the sales agency that this unit was purchased, or the Yamato Scientific's sales
office.
< Check following items before contact >
◆ Model Name of Product
◆ Production Number
◆ Purchase Date
◆ About Trouble (in detail as possible)
Minimum Retention Period of Performance Parts for Repair
The minimum retention period of performance parts for repair of this unit is 7 years after
discontinuance of this unit.
The "performance part for repair" is the part that is required to maintain this unit.
See the production plate attached to this unit.
48
Page 52
Specification
Product name Autoclave
Model SN200/210 SN300/310 SN500/510 SQ500/510
Temperature control
range
Maximum Operational
Pressure
Operational ambient
Performance
temperature
Cover mechanism Manual up and down open/close system (safety lock mechanism attached)
Heater
Exhaust valve For full open and slow exhaust (one each)
Option port
Chamber part
Cooling fan Axial fan motor
Temperature control
system
Setting/display method Digital setting by UP/DOWN key / Digital display
Timer Range: 0 or 1min to 99h59min, Resolution: 1minute
Operation courses
Configurations
Other functions
Safety devices
Pressure vessel standard Small-sized pressure vessel (notification of installation is not required)
Effective dimensions
of chamber
(Diameter×Depth mm)
- Power cord 2130010010 T3-3d-0 3m Yamato Scientific
- Spring (left) LT00014853 Yamato Scientific
- Spring (right) LT00014854 Yamato Scientific
55
Page 59
List of Dangerous Substances
Never use explosive substances, flammable substances and substances that include explosive
or flammable ingredients in this unit.
EXPLOSIVE
Ethylene glycol dinitrate (nitro glycol), Glycerin trinitrate (nitroglycerine), Cellulose
nitrate (nitrocellulose), and other explosive nitrate esters
Reference
EXPLOSIVE:
FLAMMABLE
IGNITING:
OXIDIZING:
Trinitrobenzene, Trinitrotoluene, Trinitrophenol (picric acid), and other explosive
nitro compounds
Acetyl hidroperoxide (peracetic acid), Methyl ethyl ketone peroxide, Benzyl
peroxide, and other organic peroxides
Lithium (metal), Potassium (metal), Sodium (metal), Yellow phosphorus,
Phosphorus sulfide, Red phosphorus, Celluloid compounds, Calcium carbide,
Lime phosphate, Magnesium (powder), Aluminum (powder), Powder of metals
other than magnesium and aluminum, Sodium hydrosulfite
Potassium chlorate, Sodium chlorate, Ammonium chlorate, and other chlorate
Potassium perchlorate, Sodium perchlorate, Ammonium perchlorate, and other
perchlorate
Potassium peroxide, Sodium peroxide, Barium peroxide, and other inorganic
peroxide
Potassium nitrate, Sodium nitrate, Ammonium nitrate, and other nitrate
Sodium chlorite and other chlorites
Calcium hypochlorite and other hypochlorites
Ethyl ether, Gasoline, Acetaldehyde, Propylene chloride, Carbon disulfide, and
other flammable substances having a flash point of lower than -30℃
INFLAMMABLE
LIQUID:
FLAMMABLE
GAS:
Normal hexane, ethylene oxide, acetone, benzene, methyl ethyl ketone, and
other flammable substances having a flash point of -30℃ or higher but lower
than 0℃
Methanol, Ethanol, Xylene, Pentyl acetate (amyl acetate), and other flammable
substances having a flash point of 0℃ or higher but lower than 30℃
Kerosene, Light oil (gas oil), Oil of turpentine, Isopentyl alcohol (isoamyl alcohol),
Acetic acid, and other flammable substances having a flash point of 30℃ or
higher but lower than 65℃
Hydrogen, Acetylene, Ethylene, Methane, Propane, Butane, and other flammable
substances which assume a gaseous state at 15℃ and 1 atm
(Source: Appendix Table 1 of Article 6 of the Industrial Safety and Health Order in Japan)
56
Page 60
Installation Standard Manual
* Install the unit according the procedure described below (check options and special specifications separately).
Model Serial number Date
Person in charge of installation
(company name)
№
Specifications
1 Accessories
Item Method Reference operation manual Judgment
Check the quantities of accessories with the
quantities shown in the Accessory column.
・Visually check the surrounding area.
Note: Pay attention to the ambient
environment.
・Keep space.
・ Measure the customer-specific voltage
(switchboard and outlet) with the tester.
・Measure the voltage at operation (it must be
within the range of standard).
Note: Use the on-spec product when
installed on the plug or breaker.
Specification
Before Using This Unit
"1. Choose a proper place for
installation
Before Using This Unit
"7. Choose a correct power
distribution board or receptacle"
Before Using This Unit
"9. Always ground this unit"
Specification
Person in charge of
installation
Judgment
P.49
P. 6
P. 8
P. 9
P.49
2 Installation
Operation
1 Test operation
Description
Description of
1
operation
2 Error code
Maintenance
3
inspection
Completion of
installation
4
Information to
be entered
Handling Precautions
・ Clean the packing and flange on the
chamber.
・Attach the drain bottle.
Note: Supply water into the bottle.
・Attach the drainboard.
・Close the drain valve.
・Supply water into the chamber.
Note: Supply water to the gauge level on
the drainboard.
・Attach the vapor cup and droplet tray.
・Start operation.
Operate the equipment with the apparatus
sterilization course.
Check: pressure/temperature rise,
Vapor leak is not allowed.
Explain the operation of each unit to the
customer according to this Operation Manual.
Explain error codes and the procedure for
resetting them to the customer according to
this Operation Manual.
Explain the operation of each unit to the
customer according to this Operation Manual.
・Enter the date of installation and the name
of the person in charge of installation on
the face plate on the unit.
・ Enter necessary information on the
guarantee, and pass it to the customer.
・Explain the after-sale service route to the
customer.
"Do not damage the packing on
the cover or flange on the
chamber"
Before Using This Unit
"Installation Procedure"
Operation Method
"Apparatus/Liquid Sterilization
Course"
All
In the Event of Failure…
Maintenance Method
Periodical Inspection
After Service and Warranty
42
P.
P.10
27
P.
P.45
P.43
P.58
P. 48
57
Page 61
Periodical Inspection
Periodical Voluntary Inspection of Small-sized Pressure Vessel
Pursuant to Article 94 of "Ordinance on Safety of Boilers and Pressure Vessels", small-sized pressure
vessels, such as SN 200/210/300/310/500/510 and SQ500/510 models, are obliged to be voluntary
inspected periodically once within a year. The inspection results also must be preserved for three years.
Ordinance on Safety of Boilers and Pressure Vessels
Article 94 (Periodical Voluntary Inspection)
1. With regard to a small-sized boiler or small-sized pressure vessel, the employer shall
implement voluntary inspection, periodically, and at least once for every period not
exceeding one year after the commencement of operation, on the following matters,
provided that this provision shall not apply to a non-use period of small-sized boilers or
small-sized pressure vessels which are not used for a period exceeding one year:
1) For a small-sized boiler, existence of damage or abnormality in the boiler itself,
burning apparatus, automatic control systems and fittings.
2) For a small-sized pressure vessel, existence of damage or wear in the pressure
vessel itself, locking bolts of cover plate, tubes or valves.
2. With regard to the small-sized boilers or small-sized pressure vessels described in the
provision of the preceding paragraph, the employer shall, before the resumption of use,
implement voluntary inspection on the matters as described in each item of the same
paragraph.
3. When having implemented the voluntary inspections set forth by the preceding two
paragraphs, the employer shall record their results and preserve such records for three
years.
Article 95 (Repairs, etc.)
In the event that any abnormality has been found as the result of the voluntary inspection
as set forth by paragraph 1 or 2 of the preceding Article, the employer shall take necessary
measures as repairs, etc
Perform the inspection and preserve the record of inspection results for three years referring to the
inspection procedures described in the next section.
If it is difficult to perform the inspection by yourself, contact the selling office where you purchased or our
sales office.
58
Page 62
Periodical Inspection
Periodical Voluntary Inspection of Small-sized Pressure Vessel
Inspection procedures
- Apparatus necessary for inspection -
① Tools: Cross slot screwdriver, tow monkey wrenches
② Corrected temperature sensor and temperature recorder measurable up to 200℃
③ Corrected pressure gauge with the pressure range of 0.4MPa
④ Pressure device (a compressor capable of pressurizing up to 0.3MPa or high-pressure
Preparation
nonflammable gas cylinder. Automobile inflator may be used.
⑤ Pressure-proof rubber hose (inner diameter: 12 mm), hose band and joint
① Make sure that no damage, rust and deformation exist on the chamber and cover.
② Make sure that no cracks exist on the surface of cover lock.
The cover lock must be replaced if any cracks exist.
③ Make sure that the lock lever of cover does not slide with the cover opened.
④ Make sure that no damage and deformation exist on the packing.
External inspection
The packing must be replaced if any damage or deformation exists.
Back plate
Outer plate removal
① Remove the back plate, right side plate and left side plate in this order.
② For the right and left side plate, unscrew the three bottom screws respectively and remove the
plates by sliding them downward.
Right side plate Left side plate
59
Page 63
Periodical Inspection
Periodical Voluntary Inspection of Small-sized Pressure Vessel
Inspection procedures
① Make sure that the drain valve and exhaust solenoid valve do not rotate by hand. Make sure
that no water droplets are adhered around the fitting area.
If the valves rotate or droplets are adhered, the piping is loose. Tighten it securely.
② Make sure that the nut on the joint does not rotate when turned clockwise with a screw wrench.
Make sure that no water droplets are adhered around it.
Piping
If the valves rotate or droplets are adhered, the piping is loose. Tighten it securely.
③ Make sure that damages do not exist on the piping. Make sure that no water droplets are
adhered around it.
If the conditions above, especially water droplet, are observed on the areas other than the joint, the
piping must be replaced.
Close the cover.
Corrected pressure gauge
Connected to the
drain valve
Pressure
Pressure
device
Pressure-proof hose
① Set up/connect the respective devices referring to the figure above.
② Pressurize the chamber inside. When the corrected pressure gauge reads 0.05MPa, 0.1MPa
and 0.15MPa, compare the readings between the corrected pressure gauge and the pressure
gauge on the main unit.
The tolerance of pressure accuracy compared should be within ±0.005 MPa. If it is beyond this
range, replace the pressure gauge.
① In order to confirm the operating pressure of safety valve, pressurize it to 0.15 MPa first and
then pressurize it gradually.
② The valve should be actuated at the pressure between 0.25 and 0.26 MPa. Do not pressurize
it with 0.27 MPa or more.
If the valve is actuated in the range other than the above, adjust it according to the procedures
below.
Deal drawing
Operating pressure high
Tighten
Operating pressure low
作動圧力低下作動圧力上昇
緩む締る
Loosen
【分解図】
Safety valve
Setscrew
固定ネジ
Spring
スプリング
1) Loosen the setscrew.
2) Adjust the operating pressure with a adjusting screw.
3) After the adjustment is completed, tighten the setscrew.
60
Page 64
Periodical Inspection
Periodical Voluntary Inspection of Small-sized Pressure Vessel
Inspection procedures
Close
Pressure
device
① Close the drain valve as soon as the safety valve operates.
Opening is
穴を塞ぐ
blocked.
Water leak on solenoid valve/exhaust valve
② Pressurize the chamber inside. When the corrected pressure gauge reads 0.05MPa, 0.1MPa
and 0.15MPa, compare the readings between the corrected pressure gauge and the pressure
gauge on the main unit.
The tolerance of pressure accuracy compared should be within ±0.005 MPa. If it is beyond this
range, replace the pressure gauge.
Check the temperatue accuracy by using either of following methods.
① Fix a corrected temperature sensor (a recorder attached) or maximum thermometer near the
chamber temperature sensor inside the chamber. Perform the sterilizing operation at a preset
temperature of 135℃ with no load. Compare the temperature inside the chamber displayed
on the equipment with that indicated by the corrected thermometer. The temperature should
be between 135 and 137℃ to the preset temperature of 135℃. Use the service plug (internal
thread of R1/4) to fix the temperature sensor.
② Perform sterilization and compare the pressure and temperature at sterilization process. The
relation between the temperature inside the chamber and saturated pressure under the
atmosphere of 1013 hPa are as follows.
If parts replacement, repair or correction is difficult to be done by yourselves, contact the selling office
where you purchased or our sales office.
61
Page 65
Periodical Inspection
Periodical Voluntary Inspection of Small-sized Pressure Vessel
- Inspection list -
Item
Chamber
Cover
Cover lock
Packing
Darin valve
Exhaust
solenoid valve
Date
damage, rust,
deformation
damage, rust,
deformation
damage
(crack, rust)
damage,
deformation
Loose in fitting
area, water leak
Loose in fitting
area, water leak
Safety valve
Piping/joint
Pressure
gauge
Temperature
accuracy
Inspected by:
Approved by:
Check (9) the appropriate box when no abnormality is detected. If any abnormality is observed, fill in the
appropriate box with the content of abnormality and its corrective action.
operating
pressure
damage, loose,
water leak
Pressure
accuracy check
Temperature
accuracy check
62
Page 66
Responsibility
Please follow the instructions in this document when using this unit. Yamato Scientific has no
responsibility for the accidents or breakdown of device if it is used with a failure to comply.
Never conduct what this document forbids. Unexpected accidents or breakdown may result in.
Note
◆ The contents of this document may be changed in future without notice.
◆ Any books with missing pages or disorderly binding may be replaced.
Instruction Manual for
Autoclave
Model SN200/210/300/310/500/510, SQ500/510
Third Edition Oct. 27, 2009
Yamato Scientific Co., Ltd.
2-1-6 Nihonbashi Honcho, Chuo-ku,
Tokyo, 103-8432, Japan
http://www.yamato-net.co.jp
63
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.