HITEC
CR
Ⅳ
OUTPUT
Temperature Output
Alarm Output
Time-up Output
Operation Manual
Fifth Edition
Yamato Scientific America Inc.

1. Specifications
Time-up Output
Alarm Output
Temperature Output
(Analog)
2. Connection
•
Output at the automatic stop and at the end of program
•
C contact (Relay contact)
•
Contact capacity:AC250V 3A
•
Joint:M4 Screw terminal (Common)
•
Output at the occurrence of error
•
C contact (Relay contact)
•
Contact capacity:AC250V 3A
•
Joint:M4 Screw terminal (Common)
•
Voltage output(DC) according to the measured temperature
•
See table 1 in the next page about output(0℃ at 0V
•
Resolution:0.1
•
Load impedance:100KΩ or more
•
Joint:M4 Screw terminal (Common)
:DC30V 3A
:DC30V 3A
)
℃
1. Connect firmly the terminal to use (Outputs which was not specified at the order are
unusable)
2. To use the time-up output or the alarm output;
Connect to COM (C) and NO in case the contact is “Closed” at output.
Connect to COM (C) and NC in case the contact is “Opened” at output.
3. when you use the temperature output, employ a shielded line as a connecting cable for
noise prevention, and cut it as short as possible.
Lengthwise Arrangement Crosswise Arrangement

3. Caution in use
1. Be sure to turn off the breaker on connection.
2. Use either the time-up and the alarm output under the contact capacity.
3. Be careful not to short-circuit the temperature output terminal.
4. Be sure to connect the temperature output terminal to the recorder etc., with input
impedance of 100kΩ or more.
5. Fasten securely the screw attached to terminal for connection.
Table 1 Output of Each Machine Type
Type (series) Output(mV/
DFS
DHS
DES
DTS
DP
5
DN
DS
SG
IN
IL700
FP 2
SM 10
℃
)