Yamato Scientific External output controller User Manual

HITEC
CR
OUTPUT
Temperature Output
Alarm Output
Time-up Output
Operation Manual
Yamato Scientific America Inc.
1. Specifications
Time-up Output
Alarm Output
Temperature Output
(Analog)
2. Connection
Output at the automatic stop and at the end of program
C contact (Relay contact)
Contact capacity:AC250V 3A
Joint:M4 Screw terminal (Common)
Output at the occurrence of error
C contact (Relay contact)
Contact capacity:AC250V 3A
Joint:M4 Screw terminal (Common)
Voltage output(DC) according to the measured temperature
See table 1 in the next page about output(0℃ at 0V
Resolution:0.1
Load impedance:100KΩ or more
Joint:M4 Screw terminal (Common)
 :DC30V 3A
:DC30V 3A
1. Connect firmly the terminal to use (Outputs which was not specified at the order are unusable)
2. To use the time-up output or the alarm output; Connect to COM (C) and NO in case the contact is “Closed” at output.
Connect to COM (C) and NC in case the contact is “Opened” at output.
3. when you use the temperature output, employ a shielded line as a connecting cable for noise prevention, and cut it as short as possible.