Yamato Scientific DG400 User Manual

Page 1
Instrument Drying Oven
Model DG400
Thank you very much for purchasing this Yamato DG400 instrument drying oven.
Please read the “Operating Instructions” and “Warranty” before operating this unit to assure proper operation. After reading these documents, be sure to store them securely together with the “Warranty” at a handy place for future reference.
Warning:Before o perating the unit, be sure to
read carefully and fully understand important warnings in the operating instructions.
Yamato Scientific Co., Ltd.
Page 2
Table of contents
1.Safety precautions .................................................................. 1
Explanation of pictograms ............................................................. 1
List of symbols ....................................................................... 2
WarningCautions ................................................................... 3
2. Before operating the unit............................................................ 4
Precautions when installing the unit ..................................................... 4
Installation proceduresprecautions .................................................... 7
3. Names and functions of parts........................................................ 9
Main body ........................................................................... 9
Operation panel ..................................................................... 10
Explanation of characters............................................................. 11
4. Operating procedures ............................................................. 12
List of operation modes and functions .................................................. 12
Operation modefunction setting keys and characters ................................... 14
Operating procedures (settings for overheat prevention device) ............................ 15
Operating procedures (fixed temperature operation)...................................... 16
Operating procedures (quick auto stop operation) ........................................ 17
Operating procedures (auto stop operation) ............................................. 19
Operating procedures (auto start operation) ............................................. 21
Useful functions (calibration offset function) ............................................. 23
Useful function (setting lock function)................................................... 24
Useful function (power outage compensation function).................................... 25
5. Cautions on handling .............................................................. 26
6. Maintenance procedures........................................................... 29
Daily inspection/maintenance ......................................................... 29
7. When the unit is not to be used for a long time or when disposing..................... 30
When the unit is not to be used for a long time or when disposing
Notes about disposition .............................................................. 30
8. Troubleshooting................................................................... 31
Safety device and error codes......................................................... 31
When a malfunction is suspected ...................................................... 32
9. After sales service and warranty .................................................... 33
When requesting a repair ............................................................. 33
10. Specifications.................................................................... 34
11. Wiring diagram................................................................... 35
12. List of replacement parts.......................................................... 36
13. List of dangerous materials ....................................................... 37
14. Standard installation manual ...................................................... 38
.......................... 30
Page 3
1. Safety precautions
A
y
r
Explanation of pictograms
variety of pictograms are indicated in this operating instruction and on
products for safe operation. Possible results from improper operation
ignoring them are as follows.
Be sure to fully understand the descriptions below before proceeding to the
text.
Warning Caution
Note 1)Serious injury means a wound, an electrical shock, a bone fracture or intoxication that ma
leave after effects or require hospitalization or outpatient visits for a long time.
Note 2)Minor injury means a wound or an electrical shock that does not require hospitalization o
outpatient visits for a long time.
Indicates a situation which may result in death or serious injury (Note 1.)
Indicates a situation which may result in minor injury (Note 2) and property damages (Note 3.)
About pictograms
Note 3)Property damage means damage to facilities, devices and buildings or other properties.
Meanings of pictograms
This pictogram indicates a matter that encourages the user to adhere to warning (“caution” included). Specific description of warning is indicated near this pictogram.
This pictogram indicates prohibitions Specific prohibition is indicated near this pictogram.
This pictogram indicates matters that the user must perform Specific instruction is indicated near this pictogram.
1
Page 4
Warning
1. Safety precautions
List of symbols
General warnings
Caution
General cautions Electrical shock! Burning!
For water only
Danger!: High
voltage
Poisonous
material
Danger!: High
temperature
Danger!: Moving
part
Caution for no liquid heating!
Danger!: Hazard
of explosion
Caution for water
leak!
Prohibitions
General bans Fire ban
Compulsions
General
compulsions
Connect ground
wire
Do not
disassemble
Install levelly
Do not touch
Pull out the power
plug
Regular
inspection
2
Page 5
1. Safety precautions
WarningCautions
Never operate the unit in an atmosphere containing flammable or explosive gas. Otherwise, an explosion or a fire may result since the unit is not explosion-proof. See section “13. List of dangerous materials” on page 37.
Be sure to connect the ground wire correctly. Otherwise, electrical leak may result and cause an electrical shock or a fire.
When a smoke or an unusual odor is seen or sensed, immediately turn the ELB on the main unit off and pull out the power plug. A fire or an electrical shock may result.
When these are used bundled, they might overheat causing a fire.
Never operate the unit in an atmosphere containing flammable or explosive gas
Be sure to connect the ground wire.
Ban on operation when an abnormality occurs
Never use electrical power cords bundled.
Warning
Take care not to damage electrical power cords.
Avoid tightly bend, pull with a strong force or twist to prevent electrical power cords from damaging. A fire or an electrical shock may result.
Never use an explosive or a flammable material with this unit.
Never use an explosive material, a flammable material or a material containing them. An explosion or an electrical shock may result. See section “13. List of dangerous materials ” on page 37.
Some parts of the unit are hot during and immediately after operation. Take special care for possible burning.
Never try to disassemble or alter the unit. A malfunction, a fire or an electrical shock may result.
Never try to touch a hot part.
Never try to disassemble or alter the unit.
Caution
When a thunder is heard.
When a thunder is heard, turn the main power off immediately. A malfunction, fire or an electrical shock may result.
3
Page 6
2. Before operating the unit
1. Carefully select an installation site.
Take special care not to install the unit at a place described below:
Uneven surfaces or dirty surfaces Where flammable gas or corrosive gas existsWhere the ambient temperature is 35 or more Where temperature changes severely Where humidity is high Where subject to direct sunlight Where vibration is severe
Install this unit at a place with spaces shown below.
Precautions when installing the unit
15c m or mo re
15cm
or
more
2. Install the unit on a level surface.
Install the unit on a level surface. If the whole bottom surface of the unit does not contact the surface evenly, vibrations or noises may result. This might cause unexpected troubles or malfunctions.
15cm
or
more
1m or more
Front
The unit weight is approx. 45 kg. When lifting the unit for transportation and installation, carefully handle it by at least two people.
3. Installation
The unit might fall down or move by an earthquake or an impact resulting in a personal injury. We recommend making safety measures such as to avoid installing the unit at a place other than busy places.
4
Page 7
2. Before operating the unit
Precautions when installing the unit
4. Secure sufficient ventilation for the unit.
Do not operate the unit when its side panels and vent holes are blocked. Internal temperature of the unit will rise degrading the performance and an accident, a malfunction or a fire may result.
5. Do not operate the unit at such a place that may subject to splash.
Do not operate the unit at such a place that may subject to splash. Liquid entering the inside may cause an accident, a malfunction, an electrical shock or a fire.
6. Never operate the unit in an atmosphere containing flammable or explosive gas.
Never operate the unit in an atmosphere containing flammable or explosive gas. Since the
unit is not explosion-proof, an arc is discharged when switching the ELB “ON” and “OFF” and during operation and a fire or an explosion may result.
See the section “13. List of dangerous materials” on page 37 for flammable and explosive gases.
Explosive
gas
Flammable
gas
5
Page 8
2. Before operating the unit
,
Precautions when installing the unit
7. Be sure to connect the pow er plug to the dedicated power distribution panel or a wall outlet.
Use a power distribution panel or a wall outlet that meets the electrical capacity of the unit.
Electrical capacity: DG400 AC100V 10.5A
When the unit will not start even when you turn the Electric Leakage Breaker to “ON”,
check for low main voltage or if the unit is connected to the same power supply line as other devices and connect it to another line if necessary.
Avoid connecting too many devices using a branching outlet or extending a wire with a cord reel or temperature controlling function may degrade due to voltage drop.
Do not connect the unit to any parts or lines other than a correct power supply line such as a gas pipe, a water pipe or a telephone line. Otherwise, an accident or a malfunction may result.
8. Handling of a power cord
Never use electrical power cords bundled. When these are used bundled, they might overheat causing a fire.
Do not convert, forcibly bend, twist or pull the power cord. Otherwise, a fire or an electrical shock may result.
Do not place the power cord under a desk or a chair, or sand between objects to avoid it from being damaged. Otherwise
a fire or an electrical shock may result.
Do not place the power cord close to a stove or other heat generating device. Sheath of the cord may burn and result in a fire or an electrical shock.
If the power cord should be damaged (exposure of core wire or disconnection), immediately turn the ELB off, pull out the power cord (plug) out of the power supply and ask your dealer to replace the cord. If it is left unrepaired, a fire or an electric shock may result.
Connect the power cord to an appropriate wall outlet.
9. Be sure to connect the ground wire.
When the unit has no ground terminal, class D grounding work is necessary and please
consult your dealer or our nearest sales office.
Securely connect to an outlet.
We recommend use of a ground type outlet tap.
Grounded tap
Power plag
When there is no ground terminal.
In this case, class D grounding work is necessary and please consult your dealer or our nearest sales office.
Do not connect the grounding wire to any parts or lines other than a correct grounding terminal such as a gas pipe, a water pipe or a telephone line. Otherwise, an accident or a malfunction may result.
When a bipolar type outlet tap is used
Bipolar receptacle
Insert the ground adaptor included as an option, into a power plug confirming the polarity of the outlet. Connect the grounding wire (green) of the ground adaptor to the ground terminal on the power supply equipment.
Ground wire
6
Page 9
g
A
(1) Select an installation site.
Make sure that all of four legs are securely on a flat surface.Two legs on the unit front are adjustable. Adjust them so that the installed unit will not be
unstable. Turning the adjuster clockwise raises it and turning it anti-clockwise lowers it.
(2) ・Placement of the drain pan.
Be sure to place a drain pan to receive water generated during the drying process.
(3) Install shelf boards.
Install shelf pegs at heights you want on the right and left shelf posts in the internal bath of
the main body.
Completely push shelf boards by sliding to the end. *Take care to put each shelf board on correct pairs of right and left shelf pegs. Make sure that shelf boards will not fall nor rattle.Withstand load of each shelf board is 15kg in even loading. When putting instruments,
arrange them as dispersed as possible.
2. Before operating the unit
Installation proceduresprecautions
Drain pan
Samples
15k
 Shelf board
Put instruments with appropriate spaces between them. Too many instruments may prevent
proper temperature control. To assure proper temperature control, put instruments with a space at least 30% of the shelf board area.
ssure at least 30% of space.
7
Page 10
2. Before operating the unit
Installation proceduresprecautions
(4) Do not put an instrument on the bottom of the internal bath.
Operating the unit with a fixing directly put on the bottom of the internal bath might degrade
its temperature characteristics. This also may cause corrosion, damage or rust of the internal bath. Never put any fixing on the bottom surface.
When putting instruments, take care not to allow them touching the heater, the sensor or
other devices that are installed on the bottom. Put instruments on the shelf board included with the unit.
(5) Take special care for instruments shown below:
Instruments that contain flammable or explosive components or such instruments to which
samples containing those components are attached.
The unit is not explosion proof. Never attempt to dry or process instruments to which
samples that contain flammable or explosive components are attached.
Corrosive instrumentsTake care for handling of corrosive instruments or instruments to which corrosive
components are attached. Although SUS304 stainless steel is used for major components, note that they might corrode with strong acid. Note that packings may corrode with acid, alkali, oil or organic solvents.
(6)
Always operate the unit with the exhaust ports open.
There are two exhaust ports on the top surface of the unit. In regular operation, open both
of two vent holes. Adjust their opening level according to the water amount attached to a specific instrument.
Note that high temperature steam may be blowing out of the exhaust ports. To prevent a burn, never try to look into the exhaust ports or
(7) Always shut the door completely.
Make sure that the clamp on the right side of the door is completely locked before
operating the unit.
(8) About two-tier stacking
Do not stack two DG400 units on each other in two tiers.
touch those parts with bare hands.
8
Page 11
Front panel
Observation
window
3. Names and functions of parts
Operation panel
Main body
Standalone overheat preventive device (hydraulic)
ELB
Rating sticker
Rear panel
Door
Drain pan
Casters with adjusters (2 points)
Exhaust ports
Power cord
9
Page 12
3. Names and functions of parts
R
Operation panel
TIME
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
RUN
STOP
No. Name Operation/action
RUN/STOP key Used for starting/stopping operation. ② ▼▲ keys Used for selecting settings.
TIMER key Key for selecting timer operation settings.
Quick auto stop operation, auto stop operation or auto start operation can be selected.
SUB MENU key
(Long press of the TIMER
Key for setting calibration offset temperature, the key lock function or the power outage compensation function.
key)
RUN lamp Illuminates during fixed value operation and blinks during
timer operation.
HEATER lamp Illuminates while heater power is on. Measured temperature
screen
Displays measured temperature in the bathset characters alarm information.
Set temperature screen Displays a set temperature, timer settings and timer remaining
time.
10
Page 13
3. Names and functions of parts
Explanation of characters
Characters on the controller are explained in this section.
Characters Identifier Name Application
AStP Auto stop setting Used for setting auto stop operation.
AStr Auto start setting Used for setting auto start operation.
Displayed when timer operation has
End Time up
cAL Calibration offset setting
ended. See page 17, 19.
Used for inputting a calibration offset temperature
See section “Using the calibration offset function” on page 23.
Key locks settings to prevent their
Lock Key lock of settings
Pon
* See the section “Operation modefunction setting keys and characters” on page 14 for operation
modes and characters of functions.
Power outage compensation setting
alteration.
See section “Using the lock function” on page 24.
Selects operations after recovery from power outage.
See section “Using the power outage compensation function” on page25.
11
Page 14
4. Operating procedures
List of operation modes and functions
Operation modes of the unit are as shown below:
Name Description Page
Turning the ELB on to enter the operation setting mode.
Fixed temperature
1
operation
Quick auto
2
stop operation
Proceed to temperature setting that uses ▼▲ keys.
Pressing the RUN/STOP key long starts the operation and
pressing it long again stops operation.
Used when you want to “stop fixed temperature operation
being performed automatically in several hours.
Press the TIMER key during fixed value operation to display
“AStP.” Set a duration before stop with the ▼▲ keys.
Pressing the RUN/STOP key starts quick auto stop operation
and activates the timer in the middle of it to automatically
stop it after the set period of time.
P. 1 6
P. 1 7
Used when you want to “set automatic stop for fixed value
operation when making settings for it.”
3 Auto stop operation
4 Auto start operation
* Operation mode cannot be changed while the unit is in operation. First stop operation before
changing the mode.
Press the TIMER key to display “AStP.” Set a duration before stop with the ▼▲ keys.
Pressing the RUN/STOP key starts auto stop operation.
Used when you want to “start operation automatically after
several hours” after power is turned on.
Press the TIMER key to display “AStr.” Set a duration before stop with the ▼▲ keys.
Pressing the RUN/STOP key starts auto start operation.
P. 1 9
P. 2 1
12
Page 15
4. Operating procedures
List of operation modes and functions
Functions of the unit are as shown below:
Name Description Page
Automatic overheat prevention function: This function is linked to the unit set temperature and
has been set to so that it is automatically activated (returned automatically) at a temperature 12 higher
than the set temperature in the bath.
Overheat prevention
1
function
Standalone overheat prevention device:
When the temperature in the bath reaches the set
temperature of the overheat prevention device, controller power is shut off. (Controller display will be eliminated)
Change temperature setting of the hydraulic overheat
preventing device on the right side of the unit to the correct value (set temperature +20), turn the ELB OFF once and then turn it ON again.
P. 1 5
Calibration offset
2
function
3 Setting lock function
Power outage
4
compensation function
Calibration offset function compensates any differences
between the target temperature in the bath and the control temperature of the controller (sensor temperature.)
The function can compensate to either plus or minus side for the whole temperature band of the unit. This compensation can be set with the SUBMENU keys.
This function locks the set operation status. The lock can be set or released with the SUBMENU key.
This function returns the main unit operation to the resume
status after recovery from power outage, or keeps the current stop status. This compensation can be set with the SUBMENU keys.
P. 2 3
P. 2 4
P. 2 5
13
Page 16
4. Operating procedures
Operation modefunction setting keys and characters
Key operations and characters in the diagram below are used for operation mode and function settings.
ELB ON
Fixed temp.
Temperature
Long
press
operaiton
setting
RUN
/
STOP
Operation
start
Quick auto stop
Fixed temp.
operation
TIMER
AStP
Time setting
RUN/
One
STOP
sec.
Timer start
End
RUN/
One
STOP
sec.
Operation
stop
operation
Auto stop
Time setting
One sec.
Timer start
One sec.
Automatic
operation
Timer
TIMER
AStP
RUN/ STOP
End
RUN/ STOP
stop of
Auto start
Time setting
One sec.
Timer start
Auto start of
operation
One sec.
Operation
AStr
RUN/ STOP
RUN/ STOP
stop
Functional
setting
Long press
MENU
Calibaration offset
function
Compensation
setting
Minus
Long press
MENU
SUB
cAL
SUB
Plus
off
Long
press
Setting lock
function
LocK
Lock setting
SUB
MENU
on
Long
press
Power outage compensation
function
Power outage compensation
off
MENU
Pon
setting
SUB
on
14
Page 17
4. Operating procedures
Operating procedures (settings for overheat prevention device)
As a safety measure for preventing overheat, a hydraulic overheat prevention device (manual return) is installed.
Temperature setting range and functions
The temperature setting range for the standalone overheat prevention device is “50℃~120℃.”
When the temperature in the bath keeps rising beyond the controller set temperature and reaches
the set temperature of the overheat prevention device, controller power is shut off. (Controller display will be eliminated) When the overheat prevention device is activated, it will not be released until the ELB is turned on. Change temperature setting to the correct value (set temperature +20), turn the ELB OFF once and then turn it ON again.
How to set temperature
Setting the overheat prevention temperature
Set the temperature scale on the hydraulic overheat
prevention device installed on the upper right side of the unit to the arrow in the diagram shown left.
Turn the ELB to “OFF” and wait for a while without opening
the door.
After a while, turn the ELB ON (set the ELB “ON”.)
Set the temperature scale
to the arrow
Caution
Set temperature as “set temperature +20” as a rough standard and add 5 to the setting if
the device functions improperly.
The temperature setting range for the standalone overheat prevention device is “50℃~
120.” Be sure to set the overheat prevention activation temperature correctly otherwise the
device may not start, the overheat prevention device is activated before temperature in the bath increases completely, or a fire or other unexpected accidents may result.
The temperature is set at 120 on shipping from the factory.
If the temperature for the standalone overheat prevention device is set at around or below the
room temperature, the device may be triggered when the door is opened.
The overheat prevention device has been designed to prevent overheating of devices not to
protect samples. The device does not prevent accidents caused from use of explosive or flammable substances.
15
Page 18
R
How to start fixed temperature operation
MEASURED TEMP.
HEATE
RUN
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
SET TEMP.
RUN
STOP
4. Operating procedures
Operating procedures (fixed temperature operation)
1.Turn the ELB ON. (Turn the ELB to “ON.”)
When the ELB is turned ON, the initial values will be
displayed for about four seconds, then the initial screen
will appear and the current bath temperature and the
previous set temperature are displayed on each of the indicators.
Measured temperature screen: Indicates the current bath temperature
Set temperature screen: Displays the previous set temperature
2. Setting the temperature
Set a temperature using the ▼▲ keys.
RUN
STOP
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
When you want to correct setting errors or change settings
When you want to lower the set temperature during fixed value operation, note
that it takes some time to reach the reset temperature since the unit has no
Caution
cooling capacity.
Immediately after operation has been stopped, the temperature in the bath is
around the set temperature. Operation stop refers only to machine stop and time needed for decreasing the temperature in the bath is not considered.
3. Starting operation
Press the RUN/STOP key longer.
Fixed temperature operation will start and the RUN lamp and the heater lamp come on.
4. Stopping operation
Press the RUN/STOP key longer.
Operation stops, the RUN lamp goes off and the screen switches to the initial setting screen.
When you want to change settings, press the ▼▲ keys on
the current screen to enter the setting mode where you can
change settings. Blinking stops in three seconds after change and setting is completed.
16
Page 19
4. Operating procedures
① ②
Operating procedures (quick auto stop operation)
Used when you want to “stop fixed temperature operation being performed automatically in several hours. Quick auto stop operation is a function to enable auto stop timer setting during operation.
Procedures for quick auto stop operation
MEASURED TEMP.
RUN
STOP
TIMER
SUB MENU
HEATER
RUN
SET TEMP.
1. Setting time period before stop during fixed temperature operation
Make sure that the RUN lamp is illuminated to
indicate the unit is in operation. Press the TIMER key.
Characters AStP are indicated on the measured temperature screen to indicate the auto
stop operation mode and set duration blinks on the set temperature screen.
Set a duration you want using the ▼▲ keys.
About the timer function
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
RUN
STOP
RUN
STOP
The maximum time that can be set for the timer is 999 hours 50 minutes. Up to 99 hours 59 minutes, time can be set in minutes. One hundred hours and over are set only in 10 minutes. Keep the ▼▲ keys pressed to continuously change set
time and you can quickly reach the time you want. Press the ▼▲ keys once at a time for fine adjustment.
2. Starting timer operation
When the time you want is set, press the RUN/STOP key
while the set temperature screen is blinking.
The RUN lamp blinks and timer operation is started. Timer starts counting when the temperature in the bath
reaches the set temperature.
Once timer counting is started, the set temperature screen
changes to the remaining time display.
3. Stopping and ending timer operation
Operation stops automatically when the set temperature
has elapsed.
Characters End
blink on the set temperature
screen to indicate operation has ended.
Press the RUN/STOP key for approx. one second to end
the timer operation mode. The screen switches to the initial setting screen.
17
Page 20
When you want to correct set temperature or set time, or change settings
4. Operating procedures
Operating procedures (quick auto stop operation)
When you want to change settings, press the ▼▲ keys on
the current screen to enter the setting mode where you can
change settings. Blinking stops in three seconds after
change and setting is completed. Note, however, that
temperature changes after timer activation are counted also while temperature is changing.
When you want to change settings before timer activation, press the TIMER key on the current screen to enter the
setting mode where you can change settings. Enter a time
duration from when the set temperature is reached to the time the device shall be stopped.
When you want to change settings after timer activation, press the TIMER key on the current screen to enter the
setting mode where you can change settings. Note,
however, you need to set a time calculated by adding the time already passed to the time to be added.
After change has been made, press the RUN/STOP key to complete the process.
When you want to stop quick auto stop operation in the
middle of it, press the RUN/STOP key long once to stop
device control once, then make settings again in the appropriate mode.
In terms of the remaining time display
dot indicates count down and an illuminating dot indicates a
wait status (while temperature is increasing or decreasing to
the set temperature) during which the timer has stopped counting.
18
a blinking
Page 21
4. Operating procedures
① ②
Operating procedures (auto stop operation)
This mode automatically stops fixed value operation after a certain time from its start set with the timer.
Procedures for auto stop operation
MEASURED TEMP.
RUN
STOP
TIMER
SUB MENU
HEATER
RUN
SET TEMP.
1. Setting a stop timeAfter confirming the temperature you want is set, press
the TIMER key to display characters AStP on the measured temperature screen that indicate auto
stop operation.
The set time is displayed on the set temperature
screen.
Set a time you want using the ▼▲ keys.
Pressing the▼▲ keys makes the set time blink. The time is determined when blinking stops.
About the timer function
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
RUN
STOP
RUN
STOP
The maximum time that can be set for the timer is 999 hours 50 minutes. Up to 99 hours 59 minutes, time can be set in minutes. One hundred hours and over are set only in 10 minutes. Keep the ▼▲ keys pressed to continuously change set
time and you can quickly reach the time you want. Press the ▼▲ keys once at a time for fine adjustment.
2. Starting timer operation
When the time you want is set, press the RUN/STOP key
for about one second while characters AStP that indicate auto stop operation are displayed on the
measured temperature screen and the set time on the set temperature screen.
The RUN lamp blinks and timer operation is started.
Timer starts counting when the temperature in the bath
reaches the set temperature.
Once timer counting is started, the set temperature screen
changes to the remaining time display.
3. Stopping and ending timer operation
Operation stops automatically when the set temperature has elapsed.
Characters End
blink on the set temperature
screen to indicate operation has ended.
Press the RUN/STOP key for approx. one second to end
the timer operation mode. The screen switches to the initial setting screen.
19
Page 22
When you want to correct set temperature or set time, or change settings
4. Operating procedures
Operating procedures (auto stop operation)
When you want to change settings, press the ▼▲ keys on
the current screen to enter the setting mode where you can
change settings. Blinking stops in three seconds after
change and setting is completed. Note, however, that
temperature changes after timer activation are counted also while temperature is changing.
When you want to change settings before timer activation, press the TIMER key on the current screen to enter the
setting mode where you can change settings. Enter a time
duration from when the set temperature is reached to the time the device shall be stopped.
When you want to change settings after timer activation, press the TIMER key on the current screen to enter the
setting mode where you can change settings. Note,
however, you need to set a time calculated by adding the time already passed to the time to be added.
After change has been made, press the RUN/STOP key to complete the process.
Auto stop operation is not available together with auto start operation.
When you want to stop auto stop operation in the middle of
it, press the RUN/STOP key long once to stop device
control once, then make settings again in the appropriate mode.
In terms of the remaining time display
dot indicates count down and an illuminating dot indicates a
wait status (while temperature is increasing or decreasing to
the set temperature) during which the timer has stopped counting.
20
a blinking
Page 23
4. Operating procedures
① ②
Operating procedures (auto start operation)
This mode automatically starts fixed value operation after a certain time from its start set with the timer. However, operation does not stop automatically but needs to be stopped manually.
Procedures for auto start operation
1. Setting an operation start time
After confirming the temperature you want is set,
press the TIMER key to display characters AStr
MEASURED TEMP.
RUN
STOP
TIMER
SUB MENU
HEATER
RUN
SET TEMP.
About the timer function
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
RUN
STOP
RUN
STOP
on the measured temperature screen that
indicate auto start operation.
The set time is displayed blinking on the set
temperature screen.
Set a time you want using the ▼▲ keys. Pressing the▼▲ keys makes the set time blink. The
time is determined when blinking stops.
The maximum time that can be set for the timer is 999 hours 50 minutes. Up to 99 hours 59 minutes, time can be set in minutes. One hundred hours and over are set only in 10 minutes. Keep the ▼▲ keys pressed to continuously change set
time and you can quickly reach the time you want. Press the ▼▲ keys once at a time for fine adjustment.
2. Starting timer operation
When the time you want is set, press the RUN/STOP key
for about one second while characters AStr that indicate auto start operation are displayed on the
measured temperature screen and the set time on the set temperature screen.
Timer starts counting when the RUN/STOP key is
pressed.
Display on the measured temperature screen switches
from set time display to remaining time display.
3. Stopping and ending timer operation
Operation automatically starts at the set time and the
RUN lamp comes on.
To stop operation, press the RUN/STOP key for approx.
one second to end the timer operation mode. The screen switches to the initial setting screen.
21
Page 24
When you want to correct set temperature or set time, or change settings
4. Operating procedures
Operating procedures (auto start operation)
When you want to change the set temperature during timer counting, press the ▼▲ keys during that status to switch the set temperature
screen to the set temperature input mode, which blinks to enable
change of the set temperature with the ▼▲ keys.
When you want to change the set time during timer counting, press
the TIMER key during that status to switch the set temperature
screen to the set time input mode, which blinks to enable change of the set time with the ▼▲ keys.
In either case, the set temperature screen will stop blinking after a
while and switch to the timer count mode and the change made is
determined. Note, however, when you change the set time you need
to set a time calculated by adding the time already passed to the time to be added.
When operation has started after the auto start time, you cannot change the set time.
When you want to stop auto start operation in the middle of it, press
the RUN/STOP key long to stop device control once, then make settings again in the appropriate mode.
In terms of the remaining time display
indicates count down and an illuminating dot indicates a wait status
(while temperature is increasing or decreasing to the set temperature) during which the timer has stopped counting.
a blinking dot
22
Page 25
Useful functions (calibration offset function)
Using the calibration offset function
Calibration offset function compensates any differences between the target temperature in the bath and the control temperature of the controller (sensor temperature.) The function can compensate in parallel to either plus or minus side for the whole temperature band of the unit. The lock can be set or released with the SUBMENU keys.
The temperature is set at “0” on shipping from the factory.
4. Operating procedures
Control temp. after minus side compensation
Current temp.
Start operation at the target set temperature and confirm
Control temp. after plus side compensation
the temperature in the bath with a temperature recorder
SUB MENU
TIMER
MEASURED TEMP.
HEATER
RUN
SET TEMP.
STOP
RUN
after temperature has stabilized.
Confirm the difference between the set temperature and
that in the bath.
Press the TIMER key (SUBMENU key) long to enter the
sub menu mode.
Press the TIMER key (SUBMENU key) several times to
select the characters cAL that indicate the calibration offset function.
Enter the difference between the set temperature and
the temperature in the bath using the ▼▲ keys and press the TIMER key (SUB MENU key) long to exit the sub menu mode. (When you want to set the key lock
function, proceed to character selection process for the key lock function without pressing the TIMER key (SUBMENU key) long.
* You can set either of + or – side for the offset compensation temperature. When compensation is set for the – side, the measured temperature display decreases by the
compensation temperature while the temperature in the bath increases by the same amount.
When compensation is set for the + side, the measured temperature display increases by the
compensation temperature while the temperature in the bath decreases by the same amount.
* Since too large a compensation value may result in larger difference between the actual and
indicated temperatures and may present a danger, consult our nearest sales office before entering a large compensation value.
* The device has, in addition to the calibration offset function, the two-point compensation
function that adjusts offset for the lower temperature range and higher temperature range, for which adjustment temperatures have been input on shipping from the factory.
* Consult the nearest sales office before attempting validation work for the temperature adjusting
device.
23
Page 26
Using the lock function
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
RUN
STOP
RUN
STOP
4. Operating procedures
Useful function (setting lock function)
This function locks the set operation status.
The temperature is set at “off” on shipping from the factory.
Press the TIMER key (SUBMENU key) long to enter
the sub menu mode.
Press the TIMER key (SUBMENU key) several times to select the characters Lock that indicate the
setting lock function.
“Off” is displayed on the set temperature screen. To
lock settings, change to “on” using the key.
Press the TIMER key (SUBMENU key) long to exit the sub menu mode.
RUN
STOP
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
To release lock, press the TIMER key (SUBMENU
key) long again and select the characters Lock that indicate setting lock using the ▼▲ keys.
Lock is released when “off” is selected using the
key.
* When the lock function is “on”, keys other than the
RUN/STOP key and the TIMER key (SUBMENU key) are locked.
24
Page 27
Using the power outage compensation function
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
RUN
STOP
RUN
STOP
4. Operating procedures
Useful function (power outage compensation function)
The power outage compensation function returns the main
unit operation to the resume status after recovery from power outage, or keeps the current stop status.
The function is set at “on” on shipping from the factory.
Press the TIMER key (SUBMENU key) long to enter
the sub menu mode.
Press the TIMER key (SUBMENU key) several times to select the characters Pon that indicate the
setting lock function.
“On” is displayed on the set temperature screen. The
device keeps stop status after recovery from power outage when this setting is set to “off” using the key.
Press the TIMER key (SUBMENU key) long to exit the sub menu mode.
25
Page 28
5. Cautions on handling
1. About handling of flammable or combustible solution
The unit is not explosion proof. Take special care for handling instruments on which
explosive materials, combustible materials or materials containing these are attached.
Flammable or combustible solution will evaporate when left at a room temperature (or at a
lower temperature for some types of solutions) and may be ignited and explode from
switches, lights and other ignitable sources. Be sure to assure sufficient ventilation when
using these materials.
See section “13. List of dangerous materials” on page 37.
2. Ban on use/countermeasures when an error occurs
If smoke is emerges on the unit or an odd odor is felt, immediately turn the ELB on the main
unit off, turn the power supply off and contact your dealer or a Yamato sales office for
inspection. Otherwise, a fire or an electrical shock may result. The user shall never attempt to repair the unit to avoid any possible dangers.
3. Secure sufficient ventilation for the unit.
Do not operate the unit when its side panels and vent holes are blocked. Internal temperature of the unit will rise degrading the performance and an accident, a
malfunction or a fire may result.
4. Do not allow liquid to spill over the unit.
Do not allow liquid to spill over the unit. Pay special attention not to allow liquid to enter into
the vent holes. If liquid is spilt over or into the unit, do not try to operate it any further. Otherwise, an accident, a malfunction, a fire or an electrical shock may result.
5. Do not allow a metal piece to fall into the unit.
Do not allow a clip, a staple, a screw or other metal pieces to fall into the unit. Stop operating the unit if a metal piece has dropped into the unit. Otherwise, an accident, a malfunction, a fire or an electrical shock may result.
6. Do not open the cabinet.
Do not open panels or covers fixed on the unit, or do not operate the unit with any of those open. Otherwise, an accident, a malfunction, or an electrical shock may result.
7. Do not attempt to operate the unit without the drain pan.
Do not attempt to operate the unit without the drain pan. Otherwise, an accident, a malfunction, or an electrical shock may result.
8. Do not attempt to modify the unit.
The user shall never try to modify the unit; other wise, an accident, a malfunction, a fire or an electrical shock may result.
Warning
26
Page 29
5. Cautions on handling
1.Do not step on the unit.
Do not step on the unit. Otherwise, the unit may trip over or be damaged resulting a personal injury or a malfunction.
2.Do not put or drop an object on the unit.
2.Do not put or drop an object on the unit. Since the unit contains high precision devices, vibrations or shock may cause a malfunction.
3. When a thunder is heard.
When a thunder is heard, turn the ELB on the main unit off then turn the main power off immediately. Otherwise, a lightning strike may result and cause a fire.
4. During night and not to be operated for a long period of time.
During the night and when you want to stop the unit for a longer period of time, turn the ELB to “off” and pull out the power cord from the power supply.
5. About recovery from power outage.
When the power is applied again after the unit has stopped due to power outage, the unit
will automatically return to the status immediately before the power outage and resumes operation. Turn the ELB off if you do not want to resume operation by automatic recovery.
6. Always operate the unit at a correct ambient temperature.
The operating temperature range is a room temperature ranging from +5 to 70℃ above room temperature. Never try to operate the unit outside the operating temperature range.
7. When opening or closing the door
When opening or closing the door, do not put your hand or face close to the area the door
moves (space).
The door may touch your hand or face and causing an injury. After operation has been completed, do not leave the unit with its door open in order to,
for example, cool down inside of the bath or dry instruments earlier. Heat from inside the bath may cause deformation of the control panel of a malfunction of the control devices.
8. Do not operate the unit with the door open.
When the unit is operated with the door open, proper temperature control is not possible
and the heater may overheat causing a possible danger. Be sure to operate the unit with the door closed.
Caution
27
Page 30
5. Cautions on handling
Caution
9. About installation of shelf boards and instruments
Correctly place shelf boards and samples according to Installation procedures・precautions
on page 7. If these are not placed correctly, the unit will be unable to perform correctly as well as an accident or a malfunction may result.
10. Do not attempt to do anything other than specified in this operation manual.
Do not attempt to do anything other than specified in this operation manual. Otherwise, an unexpected accident may result.
28
Page 31
6. Maintenance procedures
Daily inspection/maintenance
Be sure to perform daily inspection and maintenance to assure reliable operation of the unit.
Warning
Be sure to pull out the power cord unless necessary before trying to do inspection and
maintenance works.
Start these works after the device has returned to the normal temperature.
Never try to disassemble the unit.
Caution
Wipe off any dirt with a tightly wrung soft cloth. Never try to clean the unit with benzene, thinner
or scouring powder, or rub with a scrubbing brush. Deformation, degradation or discoloration may result.
Every month
Inspect the functions of the ELB.
Test shall be performed with the power cord connected and power is being supplied to the unit.
First turn the ELB to “off.”Then, turn the ELB “on” and press the test button on the
device with a ball-point pen to check whether it is turned off to indicate that it is in the normal state.
Test button
Maintenance of the internal bath
Stop operation and turn the ELB to OFF. Pull out the power cord off the distribution board and the wall outlet. Confirm the temperature in the device and remove shelf boards and clamps.
The internal bath, shelf boards and shelf clamps are made of SUS304 stainless steel and
reinforced glass is used for the observation window. To clean these items, thoroughly wipe with a cloth moistened with cleaning alcohol then wipe gently with a dry cloth.
Never use acid detergent, alkaline detergent, oil or organic solvent, which may cause corrosion or damage to the products.
There are sharp protrusions inside the internal bath, shelf boards and shelf pillars and shall be handled with special care to avoid personal injury. Be sure to wear gloves since handling with bare hands may present danger.
29
Page 32
7. When the unit is not to be used for a long time or when disposing
When the unit is not to be used for a long time or when disposing
Caution
When the unit is not going to be used for a long time
Turn the ELB to off and pull out the power
cord.
Warning
When disposing the unit
Do not leave the unit in the area where
children may have access.
Be sure to remove handles before disposing
the unit to prevent the doors from locking.
In general, dispose the unit as a bulky waste.
Notes about disposition
Always pay attention to the preservation of the global environment.We highly recommend taking the unit apart as far as possible for separation or recycling to
contribute to the preservation of the global environment. Major components and materials for the unit are as follows:
Names of major
components
Major exterior components
Exterior Bonderized steel sheet iron, melamine resin baking finish Internal bath Stainless steel SUS304 Packing Silicon rubber foam Nameplates Polyethylene (PET) resin film
Major electric parts
Switches and relays Resin, cupper Boards Glass fiber and other composite parts Pipe heater SUS304 Power cord Synthesized rubber sheath, cupper, nickel
Major materials
30
Page 33
8. Troubleshooting
Safety device and error codes
The unit has the self diagnostic function with a controller and a separate safety device. Table below shows possible causes and measures when the safety device is triggered.
[Error codes]
When a functional or mechanical abnormality occurs, an error code will be displayed on the
control panel. When an abnormality occurs, confirm the error code and immediately stop operation.
Safety device Symptom Possible causes and measures
Sensor error
Memory error
Measured
temperature error
appears
appears
---- ---- appears
Error in the temperature input circuit
Disconnection or other errors in the temperature
sensor
Measured temperature is outside the
displayable range. Contact our service department.
Memory setting error Contact our service department.
When the upper limit alarm of the temperature
alarm function is triggered. Contact our service department.
31
Page 34
If any of the symptoms below occurs
Symptom Check
8. Troubleshooting
When a malfunction is suspected
Turning the ELB to on will not activate the unit.
Temperature does not rise.
Temperature fluctuates during operation.
Displayed temperature
differs from the measurement.
If power outage occurs
When the power is applied again after the unit has stopped due to power outage, the unit will automatically return to the status immediately before the power outage and resumes operation. Turn the ELB off if you do not want to resume operation by automatic recovery.
If the symptom does not match any of the above, immediately turn the ELB on the main unit
off, pull out the power cord from the power supply and contact your dealer or one of our sales offices.
If the power cord is connected to the power supply securely.
If power outage is occurring.
If the standalone overheat prevention device is working.
If the set temperature is below that in the device.
If the power supply voltage has declined.
If the ambient temperature is low.
If cooling load for inside the bath is large.
If the set temperature is appropriate.
If the power supply voltage has declined.
If ambient temperature fluctuates widely.
If load for inside the bath is large.
If the calibration offset setting is other than “0”. Set it to “0.”
Confirm the settings in ”Using the calibration offset function”
on page23.
32
Page 35
9. After sales service and warranty
When requesting a repair
When requesting a repair
If any trouble occurs, immediately stop operation, turn the ELB off, pull out the power plug and contact your dealer or our sales office.
Information necessary for requesting a repair
Model name of the product
Serial number
Date (y/m/d) of purchase
Description of trouble (as in detail as possible)
Be sure to indicate the warranty card to our service representative.
See the warranty card or the nameplate installed on the unit.
See “3. Names and functions of parts on page 9.
Warranty card (attached separately)
●Warranty card is given by your dealer or one of our sales offices and please fill in your dealer,
date of purchase and other information and store securely.
Warranty period is one full year from the date of purchase. Repair service for free is available
according to the conditions written on the warranty card.
For repairs after the warranty period consult your dealer or one of our sales offices. Paid repair
service is available on your request when the product’s functionality can be maintained by repair.
Minimum holding period of repair parts
The minimum holding period of repair parts for this product is seven years after end of production. Repair parts here refer to parts necessary for maintaining performance of the product.
33
Page 36
10. Specifications
Model DG400
Operating temperature range
Inner material Stainless steel SUS304
Observation window Standard glass 3mm W250 x H300 mm
Configuration
Heater SUS pipe heater 1.0 kW
Control system PID control with a micro computer
Temperature setting / display system
Operation mode
Sensor K thermocouple
Control assembly
Auxiliary functions
Self diagnostic function
Safety
device
Protection device
Room temperature +5℃~70℃
Digital display using up/down keys
Fixed temperature operation, quick auto stop operation
Auto stop operation, auto start operation
Calibration offset function, lock function, power outage
compensation function
Temperature sensor error, memory error, auto overheat
prevention, measured temperature error
ELBI with an over current protector, hydraulic standalone
overheat prevention device
Outer dimensions (W x D x H mm)
Internal dimensions (W x D x H mm)
Number of shelves/ withstand load
Shelf pitch 30 mm
Specifications
Internal volume
Power supply (50/60Hz) AC100V 10.5A
Weight Approx. 45 kg
Included items
*Performance values are for the AC100V power supply with no-load. *Operating environmental temperature range for this device is 5℃~35℃.
Shelf boards x 2, drain pan, operating manual,
504 × 561 × 788
450 × 450 × 450
10 shelves 15 kg/shelf
92
warranty card
34
Page 37
11. Wiring diagram
X2-2
AC100V
ELB
T
1
X1
X2-1
X1
2
H
3
4
Symbol Part name Symbol Part name
X2
OH
1
+
TH
2
SSR
1
2 3
-
4
TB1
5 6
3 4
7 8 9
CONT
24
1
1
CN1CN1
24
PIO
ELB with an over current
Thermostat
ELB
OH
(Standalone overheat prevention
protector
device)
T Terminal block TH Temperature sensor (K)
H Heater CONT Planar circuit board
X1, X2 Relay PIO Display circuit board
SSR Solid state relay
35
Page 38
12. List of replacement parts
Components of DG400
Symbol Part name Code No. Specifications Manufacturer
T Heater LT00008800 100V 1.0 kW Yamato
TH Sensor 1-16-003-0049
OH EGO thermostat LT00008745 55.13225.070 E.G.O
PIO Display circuit board LT00007639 CN40B-Y Yamato
CONT Planar circuit board LT00007640 CN40B-Y Yamato
- Tough card LT00007641 15P 300mm Yamato
X1 VC relay LT00008362 BW22531K Matsushita
X2 Relay 2-05-008-0002 AP3124K Matsushita
SSR SSR 2-16-000-0035 TRS5225 Toho
ELB ELB 2-06-005-0001 BJS153 Matsushita
K thermocouple
LCK-MI-2000Y
Yamato
36
Page 39
13. List of dangerous materials
Never use an explosive substance a flammable substance or a substance containing them for this device.
Nitroglycol, glycerine trinitrate, cellulose nitrate and other explosive nitrate esters
Trinitrobenzen, trinitrotoluene, picric acid and other explosive nitro compounds
Acetyl hydroperoxide, methyl ethyl ketone peroxide, benzoyl peroxide and other
Explosive
substance
Explosive
substance
organic peroxides
Metal “lithium”, metal “potassium”, metal “natrium”, yellow phosphorus, phosphorus
sulfide, red phosphorus, celluloids, calcium carbide (a.k.a, carbide), lime phosphide,
magnesium powder, aluminum powder, metal powder other than magnesium and aluminum powder, sodium dithionous acid (a.k.a., hydrosulphite)
substances
Potassium chlorate, sodium chlorate, ammonium chlorate, and other chlorates
Potassium perchlorate, sodium perchlorate, ammonium perchlorate, and other
perchlorates
Potassium peroxide, sodium peroxide, barium peroxide, and other inorganic
peroxides
Explosive
Potassium nitrate, sodium nitrate, ammonium nitrate, and other nitrates
Sodium chlorite and other chlorites
Oxidizing substances
Calcium hypochlorite and other hypochlorites
Ethyl ether, gasoline, acetaldehyde, propylene chloride, carbon disulfide, and other
substances with ignition point at a degree 30 or more degrees below zero.
n-hexane, ethylene oxide, acetone, benzene, methyl ethyl ketone and other
Flammable substances
Flammable substances
Combustible
substances with ignition point between 30 degrees below zero and less than zero.
Methanol, ethanol, xylene, pentyl acetate, (a.k.a.amyl acetate) and other
substances with ignition point between zero and less than 30 degrees.
Kerosene, light oil, terebinth oil, isopenthyl alcohol(a.k.a. isoamyl alcohol), acetic
acid and other substances with ignition point between 30 degrees and less than 65 degrees.
Hydrogen, acetylene, ethylene, methane, ethane, propane, butane and other
gas
Substance which is a flammable gas at 15, one air pressure.
(Quoted from the separate table 1 in Article 6, the enforcement order of
the Industrial Safety and Health Law)
37
Page 40
14. Standard installation manual
*Install the product according to the following: (Confirm separately for optional items or special specifications)
Model Serial number Date
Specifications
1
2 Installation
Operation-related matters
1
2
Description
1
2 Error codes
3
4
Item Implementation method
Included items
Source voltage
Operation start
Operational descriptions
Maintenance and inspection
Completion of installation Entries
Check for number of staffs against the included item field
Visual check of environmental conditions
Caution: Take care for
environment
Securing a space
Measure the user side voltage
(outlet) with a tester
Measure voltage during
operation (shall meet the specifications)
Caution: Always use a plug
that meets the specification for attaching to the ELB.
Starts operation
Performs fixed value opera-
tion, auto stop operation or auto start operation
Explain operations of each component according to the operational instructions
Explain the customer about error codes and procedures for release according to the opera­tional instructions
Explain operations of each component according to the operational instructions
Fill in the installation date and
the installation mgr. on the nameplate of the main unit
Fill in necessary information to
the warranty card and hand it over to the customer
Explanation of the route for
after-sales service
Installation
mgr.(company name)
TOC No. Reference page of the
operating instruction manual
10.Specifications field P.34
2. Before operating the unit
On the installation site
2. Before operating the unit
Be sure to connect the ground wire.
Power supply is ….
10.Specifications
Specification-power supply
2. Before operating the unit
Installation procedures…
4. Operating procedures
4. Operating procedures
Operating procedures
1. Safety precautions
13. List of dangerous
materials
8. Troubleshooting 9. After sales service and
warranty
6. Maintenance procedures
Daily inspection/ maintenance
9. After sales service and warranty
P.33
Installation mgr.
P. 4
P. 6 P. 6
P.34
P. 7
P. 1 2 25
P.12 25 P. 1
37
P. 3 1
P.29
8
33
Judg ment
Judg ment
38
Page 41
Limited liability
Be sure to use the unit strictly following the handling and operating instructions in this operating instruction. Yamato Scientific Co., Ltd. assumes no responsibility for an accident or a malfunction caused by use of this product in any way not specified in this operating instruction. Never attempt to perform matters prohibited in this operation instruction. Otherwise, an unexpected accident may result.
Notice
Descriptions in this operating instruction are subject to change without notice.
We will replace a manual with a missing page or paging disorder.
Operating instruction Instrument Drying Oven DG400 Second edition Dec.10, 2009
Yamato Scientific Co., Ltd. 103-8432 2-1-6, Nihonbashi, Honcho, Chuo-ku, Tokyo Customer support center
Tool free: 0120-405525
http://www.yamato-net.co.jp
39
Loading...