Xylem 980 User Manual

SenTix® 950 SenTix® 980
pH combination electrodes with refillable liquid reference system
ba75843e01 08/2009
Operating manual
SenTix
Note
The latest version of the present operating manual can be found on the Internet under www.WTW.com.
Copyright © Weilheim 2009, WTW GmbH
Reprinting - even as excerpts - is only allowed with the explicit written authorization of WTW GmbH, Weilheim. Printed in Germany.
ba75843e01 08/2009
SenTix General information

General information

Automatic sensor
recognition
The sensor electronics with the stored sensor data is in the connecting head of the electrode. The data include, among other things, the sensor type and series number. With each calibration, the calibration data is written in the sensor and the calibration history is recorded. The data is recalled by the meter when the sensor is connected and is used for measurement and for measured value documentation. Storing the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with several meters.
The digital transmission technique guarantees the failure-free communi­cation with the meter even with long connection cables. If the sensor firmware is enhanced by WTW, it can be updated via the meter.
ba75843e01 08/2009
Technical data SenTix

Technical data

General data

WTW model Reference electrolyte Junction NTC Special features
SenTix SenTix
WTW model pH measuring range Allowed temperature
SenTix SenTix
®
950 KCl 3 mol/L, Ag+ free Ceramic Yes Plastic shaft
®
980 KCl 3 mol/L, Ag+ free Platinum Yes

Measurement and application characteristics

range
®
950 0.000 ... 14.000 0 ... 80 °C (32 ... 176 °F)
®
980 0.000 ... 14.000 0 ... 100 °C
(32 ... 212 °F)
WTW model Membrane resistance
Typical application
at 25 °C
SenTix SenTix
®
950 < 1 GOhm Field
®
980 < 600 MOhm Laboratory

Shaft dimensions, shaft material, electrical connection

WTW model Shaft Electrical connection
SenTix SenTix
Length [mm]
®
950 120 12 Polyamide Fixed cable Digital plug 1.5 m
®
980 120 12 Glass Fixed cable Digital plug 1.5 m
Ø [mm] Material Electrode
connection
Meter con­nection

Connection cable

Diameter 4.3 mm Smallest allowed bend radius Fixed installation: 20 mm
Flexible use: 60 mm
Plug type Socket, 4 pins
Cable length

Accuracy of the IDS measuring technique

Measured parameter Accuracy (± 1 digit)
pH ± 0.004 U [mV] ± 0.2 T [°C] ± 0.1
ba75843e01 08/2009
SenTix Commissioning, measuring, calibration
Refilling opening
Refilling opening

Commissioning, measuring, calibration

Commissioning The electrode is filled with reference electrolyte solution in the factory. Pre-
pare the electrode for measuring as follows: z Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop­per or a slider.
The refilling opening must always be open during calibration and measurement!
Calibration
and measurement:
General rules
z Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics and can easily be removed with deionized water.
Note
Please keep the watering cap. It is required for the electrode to be stored. Always keep the watering cap clean.
z SenTix
®
950: Remove any gas bubbles behind the pH membrane by
shaking.
z Connect the electrode to the meter. z Calibrate the electrode according to the operating manual of the meter
and observe the following rules while doing so:
z Make sure the refilling opening for the reference electrolyte solution is
open.
z Avoid the carryover of any solution (sample or buffer solution) from one
measurement to the next by taking the following measures: – Shortly rinse the calibration and sample beakers with the solution the
beakers are to be filled with next.
ba75843e01 08/2009
– Between measurements, rinse the electrode with the solution that fol-
lows. Alternatively, you can also rinse the electrode with deionized wa­ter and then carefully dab it dry.
z Immerse the electrode in the solution in a vertical or slightly tilted position.
Make sure the immersion depth is correct. The junction must be complete­ly submersed in the solution. The junction is in the area of the bottom end of the shaft (see figure). At the same time, the level of the reference elec-
Storage SenTix
Junction
Referenceelectrolyte
(Example)
Glass membrane
Shaft
trolyte in the electrode must be at least 2 cm above the level of the solu­tion.
z Provide approximately the same stirring conditions for measuring as for
calibrating.
Note
Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane.
Subsequent
calibrations
During short
measuring breaks
Overnight or
longer
The frequency of subsequent calibrations depends on the application. The meters provide an option where you can enter a calibration interval. After the calibration interval has expired, the meter will automatically remind you of the due calibration.

Storage

With the refilling opening open, immerse the electrode in reference electrolyte (KCl 3 mol/L, Ag rinse the electrode with the test sample or deionized water.
Note
Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane.
Put the clean electrode in the watering cap that is filled with reference electrolyte (KCl 3 mol/L, Ag
Note
pH electrodes must not be stored dry or in deionized water. The electrode could be permanently damaged by this. If the liquid in the watering cap has dried up, condition the electrode in reference electrolyte (KCl 3 mol/L,
+
free) for at least 24 hours.
Ag+
+
free). Prior to the next measurement, shortly
+
free) and close the refilling opening.
ba75843e01 08/2009
SenTix Aging
Note
During longer storing periods, salt sediments may develop on the watering cap. They do not affect the measuring characteristics and can easily be re­moved with deionized water when the electrode is put into operation again.

Aging

Every pH electrode undergoes a natural aging process. With aging, the re­sponding behavior becomes slower and the electrode slope and asymmetry change. Moreover, extreme operating conditions can considerably shorten the lifetime of the electrode. These are:
z Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins
z High temperatures z High changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and mechanical damage.
Refilling the
reference
electrolyte

Maintenance and cleaning

During operation, a small amount of reference electrolyte leaks through the junction from the electrode into the test sample. If the level of reference elec­trolyte becomes too low with time, refill it through the refilling opening. Refill­ing is very easy with the enclosed dropping bottle. Proceed as follows:
z Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
z Open the refilling opening of the electrode z Press the tip of the dropping bottle into the refilling opening while turning
it slightly
z Pump the reference electrolyte in the shaft using the dropping bottle z Pull the dropping bottle out of the refilling opening while turning it slightly
as necessary.
ba75843e01 08/2009
Maintenance and cleaning SenTix
Cleaning Remove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:
Contamination Cleaning procedure
Fat and oil Rinse with water containing household wash-
ing-up liquid Lime and hydroxide deposits Rinse with citric acid (10 % by weight) Proteins Immerse in pepsin cleaning solution PEP/pH
for approx. 1 hour.
Note:
Make sure the level of the reference electrolyte is above that of the cleaning solu­tion.
Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy the glass membrane.
After cleaning Rinse the electrode with deionized water and condition it in reference
electrolyte solution for at least 1 hour. Then recalibrate the electrode.
ba75843e01 08/2009
SenTix Wear parts and accessories

Wear parts and accessories

Description Model Order no.
Reference electrolyte solution 250 mL (KCl 3 mol/L, Ag
Pepsin cleaning solution 3 x 250 ml PEP/pH 109 648
Note
Detailed information on our wide range of buffer solutions and more acces­sories is given in the price list of the WTW catalog "Laboratory and field in­strumentation".
+
free)
KCl-250 109 705

Disposal

When the electrode is at the end of its operational lifetime, please dispose of it as electronic waste and not with unsorted municipal waste. It is illegal to dispose of it with the household refuse.
ba75843e01 08/2009
Disposal SenTix
ba75843e01 08/2009
Loading...