Xylem 430 User Manual

Turb 430 IR
Turb 430 T
Operating manual
ba75507e02 12/2005
Handheld Turbidimeter
Turb 430 IR/T
going to press
Firmware
Operating manual
The use of advanced technology and the high quality standard of our instruments are the result of continuous development. This may result in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
Part of the process of consequently improving our products is the continuos further development of instrument firmware. All current data for the Turb 430 IR/T can be found on the Internet under http://
www.WTW.com:
! Firmware
! Operating manual
You can easily transfer new firmware to your instrument with the aid of the AK 540/B cable and a PC. More detailed information can be found in the appendix of the detailed operating manual on the CD-ROM provided.
Copyright
© 2005, WTW GmbH
Reprinting - even as excerpts - is only allowed with the explicit written authorization of WTW GmbH, Weilheim. Printed in Germany.
2
ba75507e02 12/2005
Turb 430 IR/T Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 General features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5 LabStation (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Connecting the LabStation . . . . . . . . . . . . . . . . . . . . . . . 15
3.4 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.1 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.2 Inserting a cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.3 General operating principles . . . . . . . . . . . . . . . . . . . . . 21
4.3.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3.2 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3.3 Navigation example 1: setting the language . . . 24
4.3.4 Navigation example 2: setting the date and time 25
4.3.5 Menu overview . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.4 System settings (System menu) . . . . . . . . . . . . . . . . . . 28
4.4.1 Measured value memory . . . . . . . . . . . . . . . . . . 29
4.4.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.4.3 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.4.4 Date/time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.5 Turbidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.5.1 General information . . . . . . . . . . . . . . . . . . . . . . 33
4.5.2 Aligning and marking a cell . . . . . . . . . . . . . . . . 33
4.5.3 Measuring turbidity . . . . . . . . . . . . . . . . . . . . . . 34
4.5.4 Settings for turbidity measurements . . . . . . . . . 36
4.5.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.6 Saving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.6.1 Storing measurement datsets . . . . . . . . . . . . . . 40
4.6.2 Filtering measurement datsets . . . . . . . . . . . . . 41
4.6.3 Displaying measurement datsets . . . . . . . . . . . 42
4.6.4 Download the measurement datsets to the RS232
interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.6.5 Erasing stored measurement datasets . . . . . . . 43
4.7 Transmitting data (RS 232 interface) . . . . . . . . . . . . . . . 43
4.7.1 Connecting a PC/external printer . . . . . . . . . . . 44
ba75507e03 07/2006
3
Contents Turb 430 IR/T
4.7.2 Configuring the RS232 interface . . . . . . . . . . . . 44
4.7.3 Selecting the output format of datasets . . . . . . . 45
4.7.4 Transmitting data . . . . . . . . . . . . . . . . . . . . . . . . 47
4.8 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.8.1 Resetting the system settings . . . . . . . . . . . . . . 48
4.8.2 Resetting turbidimeter settings . . . . . . . . . . . . . 48
4.9 Meter information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.10 Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 51
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.1 Inserting/exchanging the batteries . . . . . . . . . . . 51
5.1.2 Retrofitting the accumulator pack . . . . . . . . . . . 52
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.2.1 Cleaning the cell shaft . . . . . . . . . . . . . . . . . . . . 53
5.2.2 Cleaning the cells . . . . . . . . . . . . . . . . . . . . . . . 53
5.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.1 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.2 Turbidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.1.1 Turb 430 IR/T . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.1.2 LabStation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.2 Turbidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
7.2.1 Turb 430 IR . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
7.2.2 Turb 430 T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
8 Accessories, options . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
8.1 WTW accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
8.1.1 Connection cable: . . . . . . . . . . . . . . . . . . . . . . . 61
9 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
10 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Appendix: Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . 69
Appendix: Turbidity values under 1 FNU/NTU . . . . . . . . . .70
4
ba75507e03 07/2006
Turb 430 IR/T Overview

1Overview

1.1 General features

The compact Turb 430 IR/T handheld precision turbidimeter enables you to carry out turbidity measurements quickly and reliably.
The Turb 430 IR/T handheld meter provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
4
3
2
1
1 Keypad
2 Display
3 Cell shaft
4 Socket field
Note
If you need further information or application notes, you can obtain the following material from WTW:
! Application reports
! Primers
ba75507e03 07/2006
! Safety datasheets.
Information on available literature is given in the WTW catalog or on the Internet under www.WTW.com.
5
Overview Turb 430 IR/T

1.2 Keypad

.
Key functions
PRT
Switch to the measured value display
<M>
Start calibration
<CAL/ZERO>
Open menus / confirm entries / start measurement
.
<START/ENTER>
Call up the Configuration menu (all settings are made here)
<MENU>
Switch the measuring instrument on/off
<ON/OFF>
Output display contents to RS232 interface (e.g. print)
<PRT>
Open the Store menu: <STO> Quick storing: 2 x <STO>
Highlight menu items or selection Set values <>, <>
Switch to the next higher menu level / cancel input
<ESC>
6
ba75507e03 07/2006
Turb 430 IR/T Overview
Note
Keys with an additional number printed on are assigned doubly. This enables to directly enter numbers in special menus. Thus, you can, for example, conveniently enter the date and time via the number keys.

1.3 Display

The graphic display shows all information of the current measurement in the measured value display. The illumination enables to read the display even in the darkness.
Example
Turbidity
157.0
01.02.05 15:12

1.4 Socket field

FNU NTU
Measured value (with unit)
Status line with date and time
Identifying the
connectors
ba75507e03 07/2006
12 3
1 Power pack
2 Contacts for operation on the LabStation
3 RS232 serial interface
7
Overview Turb 430 IR/T

1.5 LabStation (optional)

With the LabStation, which is available as an accessory, you can conveniently use the Turb 430 IR/T in the laboratory. Laboratory operation with the LabStation enables the following additional functions:
! Line power operation is possible to save the batteries or
accumulator pack
! The accumulator pack in the Turb 430 IR/T is automatically charged
as soon as the meter is placed in the LabStation.
LabStation
LabStation
8
ba75507e03 07/2006
Turb 430 IR/T Safety

2 Safety

This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual carefully before working with the meter. The operating manual must always be available within the vicinity of the meter.
Target group The meter was developed for work in the field and in the laboratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take when handling chemicals.
The personnel responsible for the commissioning, operation and maintenance must have the necessary qualifications for this work. If the personnel do not have the required skills they have to be instructed. Furthermore, it must be ensured that the personnel read and completely understand the present operating manual.
Safety instructions Safety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g. "Caution") indicates the level of danger:
Caution
indicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to the instrument or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
ba75507e03 07/2006
9
Safety Turb 430 IR/T

2.1 Authorized use

This meter is authorized exclusively for carrying out turbidity measurements in the field and laboratory.
The technical specifications as given in chapter 7 T
ECHNICAL DATA must
be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized. Any other use is considered to be unauthorized.

2.2 General safety instructions

This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter 7 T It left the factory in a safe and secure technical condition.
Opening the photometer or adjustment, maintenance and repair work must only be performed by specialist personnel authorized by the manufacturer.
The only exceptions to this are the activities described in chapter 5M
AINTENANCE, CLEANING, DISPOSAL. Non-compliance results in the
loss of warranty claims.
Follow the points listed below when operating the photometer:
ECHNICAL DATA).
Function and
operational safety
! Follow the local safety and accident prevention regulations
! Observe the enclosed instructions of reagents and accessories
! Observe the regulations when dealing with dangerous substances
! Follow the operating instructions at the workplace
! Use only original spare parts.
The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the meter can only be guaranteed under the environmental conditions that are specified in chapter 7 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the instrument reaches room temperature before putting the instrument back into operation.
10
ba75507e03 07/2006
Turb 430 IR/T Safety
Safe operation It is the responsibility of the operator to continuously observe the
overall technical condition (externally recognizable deficits and damage as well as alterations to the operational behavior) of the meter.
If safe operation is no longer possible, the instrument must be taken out of service and secured against inadvertent operation! Safe operation is no longer possible if the meter:
! has been damaged in transport
! has been stored under adverse conditions for a lengthy period of
time
! is visibly damaged
! no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Caution
Danger of eye damage by visible and invisible LED radiation. In the cell shaft of the Turb 430 IR there are light emitting diodes (LED) of the 1M class. Do not look at the radiation using optical instruments. With normal, authorized use there is no hazard.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and guidelines are observed when using dangerous substances:
! EEC directives for protective labor legislation
! National protective labor legislation
! Safety regulations
! Safety datasheets of the chemical manufacturers.
ba75507e03 07/2006
11
Safety Turb 430 IR/T
12
ba75507e03 07/2006
Turb 430 IR/T Commissioning

3 Commissioning

3.1 Scope of delivery

! Handheld turbidimeter Turb 430 IR or Turb 430 T
! 4 batteries, 1.5 V type AA (in the battery compartment)
! Optional: Accumulator pack and power pack with Euro plug
and exchange plugs for USA, UK, and Australia
! Optional: LabStation
! 5 empty cells 28 mm
with label to mark the cell
! AMCO
! Microfiber cloth to clean the meter
! Compact operating manual and short operating manual
! CD-ROM with detailed operating manual
®
-Clear turbidity standard
Charging time of the
accumulator pack
Note
The optional parts of the scope of delivery are available as accessories (see section 8.1).

3.2 Power supply

You can operate the meter either with batteries, accumulator pack or a power pack. The power pack supplies the meter with low voltage (9 V DC). At the same time, the accumulator pack is charged. The accumulator pack is charged even while the meter is switched off.
The LoBat display indicator appears when the batteries or accumulator pack are nearly discharged.
approx. 36 hours.
Caution
The line voltage at the operating site must lie within the input voltage range of the original power pack (
DATA).
SEE chapter 7 TECHNICAL
ba75507e03 07/2006
Caution
Use original power packs only (see chapter 7 TECHNICAL DATA).
13
Commissioning Turb 430 IR/T
Note
The accumulator pack should not be completely discharged. If you do not operate the instrument for a longer period of time you should charge the accumulator pack every six months.
Automatic switchoff
function
The meter has an automatic switch-off function in order to save the batteries or accumulator pack (see section 4.4).
Display illumination During operation with the batteries or accumulator pack the meter
automatically switches off the display illumination if no key is pressed for 30 seconds. The illumination is switched on with the next keystroke again. The display illumination can also be switched off completely (see section 4.4.2).
Note
Power pack and accumulator pack are available as an accessory (see section 8.1).
Connecting the power
pack
(optional)
2
4
3
14
1
1 If necessary, replace the Euro plug (1) on the power pack (2)
by the country-specific plug suitable for your country.
2 Connect the plug (3) to the socket (4) of the turbidimeter.
3 Connect the power pack to an easily accessible power socket.
ba75507e03 07/2006
Turb 430 IR/T Commissioning

3.3 Connecting the LabStation

Note
The LabStation is available as an accessory (see section 8.1).
In order to use the functions of the LabStation for operation in the laboratory, connect the LabStation and place the Turb 430 IR/T in the LabStation.
Connecting the
LabStation (optional)
5
4
2
3
1
1 If necessary, replace the Euro plug (1) on the power pack (2)
by the country-specific plug suitable for your country.
2 Connect the plug (3) to the socket (4) of the LabStation.
3 Connect a PC or printer to the socket (5) of the LabStation as
necessary.
4 Connect the power pack to an easily accessible power socket.
ba75507e03 07/2006
5 Place the Turb 430 IR/T in the LabStation.
15
Commissioning Turb 430 IR/T
LabStation
16
ba75507e03 07/2006
Turb 430 IR/T Commissioning

3.4 Initial commissioning

Perform the following activities:
! For
– accumulator operation: insert the accumulator pack (see
section 5.1.2)
– line power operation and charging the accumulator pack: connect
the power pack (see section 3.2)
– operation with LabStation: connect the LabStation and place the
Turb 430 IR/T in the LabStation (see section 3.3)
! Switch on the meter (see section 4.1)
! Set the language as necessary (see section 4.3.3)
! Set the date and time as necessary (see section 4.3.4)
Note
When you set the language, date and time according to the mentioned sections of this operating manual you will quickly become familiar with the simple operation of the Turb 430 IR/T.
ba75507e03 07/2006
17
Commissioning Turb 430 IR/T
18
ba75507e03 07/2006
Turb 430 IR/T Operation

4 Operation

4.1 Switching on the meter

Switching on Press the <ON/OFF> key.
The Start menu appears for 30 seconds.
The status line indicates the meter designation and the version number of the software.
Start
Turbidity
i Turb 430 IR V 0.24
Switching off Press the <ON/OFF> key.
Automatic switchoff
function
Note
in das Menü Start gelangen Sie bei eingeschaltetem Gerät durch ggf. mehrfaches Drücken der Taste <ESC>.
After a few seconds the meter automatically switches to the measuring mode.
Turbidity
i Insert sample i Press and hold <START> i Align sample
The meter has an automatic switch-off function in order to save the batteries or accumulator pack (see section 4.4). The automatic switchoff switches off the meter if no key is pressed for an adjustable period.
ba75507e03 07/2006
The automatic switchoff is not active
! if the power is supplied by the power pack (optional),
! if the power is supplied by the LabStation (optional),
! if the Timer function is running.
19
Operation Turb 430 IR/T
Display illumination
with battery-powered
operation
During operation with batteries or accumulator pack the meter automatically switches off the display illumination if no key is pressed for 30 seconds. The illumination is switched on again with the next keystroke.

4.2 Inserting a cell

To be able to insert cells in the Turb 430 IR/T, the cell shaft has to be prepared to take in a cell.
1 Push the dust cover (1) upward.
The cell shaft for 28 mm cells is open.
1
Inserting a 28 mm cell
2 Insert the cell so that it is positioned on the bottom of the cell
shaft. The cell is ready to be measured.
3 Align the cell (see section 4.5.2).
20
ba75507e03 07/2006
Turb 430 IR/T Operation

4.3 General operating principles

This section contains basic information on the operation of the Turb 430 IR/T.
Operating elements,
display
Operating modes,
navigation
An overview of the operating elements and the display is given in section 1.2 and section 1.3.
An overview of the operating modes of the Turb 430 IR/T and the navigation through menus and functions can be found in section 4.3.1 and section 4.3.2.

4.3.1 Operating modes

The instrument has the following operating modes:
! Measurement
The display indicates measurement data in the measured value display
! Calibration
The display indicates a calibration procedure with calibration information
! Data transmission
The meter transmits measuring datasets or calibration records to the serial interface
! Configuration
The display indicates a menu with further menus, settings and functions

4.3.2 Navigation

Measured value display In the measured value display, open the menu with <MENU>.
Menus and dialogs The menus for settings and dialogs in courses contain further
submenus. The selection is made with the <><> keys. The current selection is displayed in reverse video.
! Menus
The name of the menu is displayed at the upper edge of the frame. Menus are opened by confirming with <START/ENTER>. Example:
ba75507e03 07/2006
21
Operation Turb 430 IR/T
Configuration
Turbidity
System
Info
! Settings
Settings are indicated by a colon. The current setting is displayed on the right-hand side. With <START/ENTER>, the selection of the possible settings is opened. Subsequently, the setting can be changed with <><> and<START/ENTER>. Example:
System
Language: Deutsch
Beep:
Illumination Contrast:
Temperature unit
Switchoff time
: On
:
:
Off
48 %
°C
30 min
! Functions
Functions are designated by the name of the function. They are immediately carried out by confirming with <START/ENTER>. Example: display the Calibration record function (in the Turbidity menu).
Turbidity
Calibration record
Calibration interval:
Reset
:
i 2.00 4.01 7.00 10.01
090 d
! Messages
Information or operating instructions are designated by the i symbol. They cannot be selected. Example:
22
ba75507e03 07/2006
Turb 430 IR/T Operation
Turbidity \ Calibration
i Insert standard
1000 FNU/NTU
i Press and hold <START> i Align sample
The i indicates info texts, e.g. messages, notes or instructions
Note
The principles of navigation are explained in the two following sections by reference of examples:
! Setting the language (section 4.3.3)
! Setting the date and time (section 4.3.4).
ba75507e03 07/2006
23
Operation Turb 430 IR/T

4.3.3 Navigation example 1: setting the language

Note
The following example describes in the language of the country how to set the language. On delivery, English is set as the language in the Turb 430 IR/T. During initial commissioning, the language is set in the menu, Configuration / System / Language.
1 In the measured value display:
Open the Configuration menu with <MENU>. The instrument is in the configuration mode.
2 Select the System menu with <><>.
The current selection is displayed in reverse video.
3 Open the System menu with <START/ENTER>.
4 Select the Language menu with <><>.
The current selection is displayed in reverse video.
System
Language: Deutsch
Store
Display
Reset
Interface
Continue ...
5 Open the setting of the Language with <START/ENTER>.
System
Language:
Store
Display
Reset
Interface
Continue ...
Deutsch
6 Select the required language with <><▼>.
24
7 Confirm the setting with <START/ENTER>.
The setting is active. The menu is displayed in the selected language.
ba75507e03 07/2006
Turb 430 IR/T Operation
8 To make further settings, switch to the next higher menu level
with <ESC>.

4.3.4 Navigation example 2: setting the date and time

The meter has a clock with a date function. The date and time are indicated in the status line of the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well.
Numerals are generally entered via the number keys.
The correct setting of the date and time and date format is important for the following functions and displays:
! Current date and time
Setting the date, time
and date format
! Calibration date
! Identification of stored measured values.
Therefore, check the time at regular intervals.
Note
After a fall of the supply voltage (empty batteries or accumulator pack), the date and time are reset to 01.01.2003, 00:00 hours.
The data format can be switched from the display of day, month, year (dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1 In the measured value display:
Open the Configuration menu with <MENU>. The instrument is in the configuration mode.
2 Select and confirm the System / Continue ... / Date/time menu
with <><> and <START/ENTER>.
Date/time
Time:
:
Date
Date format
: dd.mm.yy
14:53:40
30.10.03
ba75507e03 07/2006
25
Operation Turb 430 IR/T
3 Select and confirm the Time menu with <><▼> and
<START/ENTER>.
A display for the entry of numerals with the number keys opens up.
Time
14:53:40
4 Enter the time using the number keys.
The digit to be changed is displayed underlined.
Note
In the case of wrong entries, you can cancel the procedure with <ESC>. After canceling with <ESC>, it is possible to enter all digits once again. The new digits are only taken over by confirming with <START/ ENTER>.
5 Confirm the setting with <START/ENTER>.
The time is set.
6 Set the current Date as necessary. The setting is made
similarly to that of the time.
7 Change the date format as necessary.
8 To make further settings, switch to the next higher menu level
with <ESC>. or Switch to the measured value display with <M> (short
pressure). The instrument is in the measurement mode.
26

4.3.5 Menu overview

Turbidity Calibration record
Calibration interval
Reset
Timer
ba75507e03 07/2006
Turb 430 IR/T Operation
System Language Deutsch
English Français Español
Measured value
Display
memory
RS232 download
Data filter Filter
ID Date
Delete
i 4 of 1000 occupied
i Filter: No filter
Display Illumination Auto off
On Off
Contrast 0 ... 100 %
Brightness 0 ... 100 %
Reset
Interface Baud rate 1200, 2400, 4800,
9600, 19200
Info
ba75507e03 07/2006
Continue ... /
Date/time
Continue ... /
Switchoff time
Continue ... /
Beep
Output format ASCII
CSV
Time hh:mm:ss
Date
Date format dd.mm.yy
mm.dd.yy mm/dd/yy
10, 20, 30, 40, 50 min, 1, 2, 3, 4, 5, 10, 15, 20, 24 h
On Off
27
Operation Turb 430 IR/T

4.4 System settings (System menu)

The following instrument features and general functions can be found in the Configuration / System menu:
! Language selection (Language)
! Memory and database functions (Store)
! Display settings (Display)
! Restore basic settings (Reset)
! Configuration of the interface for PC/printer (Interface)
! Setting the date/time (Date/time)
! Setting the switch-off time (Switchoff time)
! Setting the keyboard sound (Beep)
Settings/functions The settings are given in the menu, Configuration / System.
Move to the Configuration menu with the <MENU> key.
Menu item Setting Description
Language Deutsch
English
Select the language (see section 4.3.3)
Français Español
Store Display
RS232 download
Memory and database functions (see section 4.6.2)
Data filter Delete
Display Illumination
Contrast Brightness
Switch on/off the display illumination (see section 4.4.2)
Reset - Resets all system settings
to default (see section 4.8.1)
Interface Baud rate
Output format
Baud rate of the data interface (see section 4.4.3)
28
Continue ... /
Date/time
Time Date Date format
Settings of time and date (see section 4.3.4)
ba75507e03 07/2006
Turb 430 IR/T Operation
Menu item Setting Description
Continue ... /
Switchoff time
10, 20, 30, 40, 50 min, 1, 2, 3, 4, 5, 10, 15, 20, 24 h
The automatic switchoff function switches the meter off if no entry is made for a specified period of time (Switchoff time). This saves the batteries or accumulator pack.
Continue ... /
Beep
On Off
Switch on/off the beep on keystroke

4.4.1 Measured value memory

In the Measured value memory menu, you find functions to display and edit the stored measurement datasets:
! Display the measurement datsets on the screen (Display)
! Download the measurement datsets to the RS232 interface (RS232
download)
! Set up filter rules for the stored measurement datsets (Data filter)
! Erase all stored measurement datsets (Delete)
! Information on the number of occupied memory locations
The settings are given in the menu, Configuration / System / Measured value memory.
Move to the Configuration menu with the <MENU> key.
ba75507e03 07/2006
29
Operation Turb 430 IR/T
Settings/functions
Menu item Setting/
Description
function
Display - Displays in pages all
measurement datasets that correspond to the filter settings.
Further options: ! Scroll through the
datasets with <><>.
! Output the displayed
dataset to the interface with <PRT>.
! Quit the display with
<ESC>.
RS232 download - Downloads to the interface all
measurement datasets that correspond to the filter settings. The download is ordered according to the date and time.
The process can take several minutes. To terminate the process prematurely, press <ESC>.
Data filter see
section 4.6.2.
Allows to set filter criteria in order to display and download datasets to the interface.
Delete - Erases the entire contents of
the measuring data memory, independent of the filter settings.
Note: All calibration data remains
stored when performing this action.
All details on the subjects of memory and stored data is given in section 4.6.2.
30
ba75507e03 07/2006
Turb 430 IR/T Operation

4.4.2 Display

In the Configuration / System / Display menu, you set the display features:
! Switching on/off the display illumination (Illumination)
! Display contrast (Contrast)
The settings are given in the menu, Configuration / System / Display. Move to the Configuration menu with the <MENU> key.
Settings
Menu item Setting Description
Illumination Auto off The display illumination is
automatically switched off if no key has been pressed for 30 seconds.
On Off
Switches the display illumination on or off permanently
Contrast 0 ... 100 % Changes the display
contrast
Brightness 0 ... 100 % Changes the display
brightness

4.4.3 Interface

In the Interface menu, you set the features of the interface:
! Transmission speed (Baud rate)
Settings
ba75507e03 07/2006
! Output format (Output format)
The settings are given in the menu, Configuration / System / Interface. Move to the Configuration menu with the <MENU> key.
Menu item Setting Description
Baud rate 1200, 2400,
4800, 9600,
Baud rate of the data interface
19200
Output format ASCII
CSV
Output format for data transmission
For details, see section 4.7
31
Operation Turb 430 IR/T

4.4.4 Date/time

In the Configuration / System / Continue ... / Date/time menu, you set the system clock:
! Current time (Time)
! Current date (Date)
! Format of the date display (Date format)
The settings are given in the menu, Configuration / System / Continue ... / Date/time.
Move to the Configuration menu with the <MENU> key.
Settings
Menu item Setting Description
Time hh:mm:ss Enter the time with the
number keys
Date Enter the date with the
number keys
Date format dd.mm.yy
Settings of time and date.
mm.dd.yy mm/dd/yy
32
ba75507e03 07/2006
Turb 430 IR/T Operation

4.5 Turbidity

4.5.1 General information

Venting the sample Air bubbles in the sample affect the measuring result to a massive
extent because they have a large scattering effect on the incident light. Larger air bubbles cause sudden changes in the measured values whereas smaller air bubbles are recorded by the instrument as turbidity. Therefore, avoid or remove air bubbles:
Avoiding or removing
air bubbles
Aligning the cell
! During sampling, ensure all movement is kept to a minimum
! If necessary, vent the sample (ultrasonic baths, heating or adding a
surface-active substance to reduce the surface tension)
Note
For turbidity values under 1 FNU/NTU, follow the instructions given in the appendix (see A
PPENDIX: TURBIDITY VALUES UNDER 1 FNU/NTU).

4.5.2 Aligning and marking a cell

Even perfectly clean quality cells exhibit tiny directional differences in their light transmittance. Therefore, if you want to achieve accurate and reproducible measurement results, it is necessary to always index the sample cells and cells for calibration standards in the same way (see section 2130 of "Standard Methods for the Examination of Water and Wastewater", 19th edition). To do so, the optimum alignment of the cell is determined.
1 Clean the cell (see section 5.2.2).
ba75507e03 07/2006
2 Insert the cell (see section 4.2).
3 Align the cell:
! Press and hold the <START/ENTER> key.
! Slowly and in small steps rotate the cell by one complete
rotation (by 360 °). After each step wait for a short time until the displayed measured value is stable.
! Turn the cell back to the position with the lowest measured
value.
Note
To keep the drift as low as possible, the time for aligning the cell while pressing and holding the <START/ENTER> key is limited to 30 seconds. After this time, the meter starts measuring automatically.
33
Operation Turb 430 IR/T
4 Release the <START/ENTER> key.
Measurement starts. The measured value is displayed.
Marking a cell To be able to quickly bring a cell into the optimum position, it is helpful
to mark the optimum position of the cell once it is determined. This shortens each measurement or calibration procedure with this cell considerably.
The marking can, e. g., be done on a label on the cap of the cell.
5 Mark the optimum position of the cell.
The cell is prepared for the shortened measuring and calibration procedures.
Measuring

4.5.3 Measuring turbidity

Caution
Never pour any liquids directly into the cell shaft. Always use a cell for measurement. The meter only measures precisely if the cell is closed with the black light protection cap (WTW cells).
Note
The outside of the cell always has to be clean, dry, and free of fingerprints and scratches. Clean the cells before starting to measure (see section 5.2.2). Only hold the cells by the top or by the black light protection cap.
1 Rinse out a clean cell with the sample to be measured:
Pour approximately 10 ml sample into the cell. Close the cell and rotate it several times before throwing the sample away.
2 Repeat the rinsing procedure twice more.
3 Fill the cell with the sample to be measured (approx. 15 ml).
Close the cell with the black light protection cap.
4 Clean the cell (see section 5.2.2).
34
5 Insert the cell (see section 4.2).
ba75507e03 07/2006
Turb 430 IR/T Operation
6 Align the cell:
! Marked cell
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and for a short time hold the <START/ENTER>
key until the measured value is displayed.
! Unmarked cell (see page 33)
Press and hold the <START/ENTER> key.
Slowly and in small steps rotate the cell by one complete
rotation (by 360 °). After each step wait for a short time until the displayed measured value is stable.
– Turn the cell back to the position with the lowest
measured value.
Note
To keep the drift as low as possible, the time for aligning the cell while pressing and holding the <START/ENTER> key is limited to 30 seconds. After this time, the meter automatically starts measuring or calibrating.
Display with measuring
range overflow
7 Release the <START/ENTER> key.
Measurement starts. The measured value is displayed.
Turbidity
157.0
8 Repeat the steps 2 to 8 for further samples.
If the measured value is outside the measuring range of the Turb 430 IR/T, it is indicated on the display:
Turbidity
> 1100
FNU NTU
01.02.05 15:12
FNU NTU
ba75507e03 07/2006
01.02.05 15:12
35
Operation Turb 430 IR/T

4.5.4 Settings for turbidity measurements

Overview For turbidity measurements, the following settings are possible in the
menu, Configuration / Turbidity:
! Calibration record (display, print)
! Entering the Calibration interval
! Reset
Settings/functions The settings are given in the menu, Configuration / Turbidity.
Move to the Configuration menu with the <MENU> key.
Menu item Possible
setting
Calibration record - Displays the calibration
Calibration interval 1 ... 999 d Calibration interval for
Reset Reset all settings for the

4.5.5 Calibration

When to calibrate? ! After the calibration interval has expired
! With a temperature change
Description
record of the last calibration.
turbidity measurement (in days).
If the calibration interval has expired, the meter reminds you to calibrate before each measurement.
Turbidity measuring mode (see section 4.8.2)
Calibration procedures
and calibration
standards
36
For the menu-guided three-point calibration you need the following three calibration standards in the mentioned order:
Standard no. FNU/NTU
1 1000
2 10,0
30,02
ba75507e03 07/2006
Turb 430 IR/T Operation
Calibration record At the end of each calibration procedure a calibration info (i symbol)
and the calibration record is displayed.
Displaying and
downloading calibration
data to interface
Sample printout of a
record
Preparing the
calibration
You can view the data of the last calibration on the display. Subsequently, you can download the displayed calibration data to the interface, e. g. to a printer or PC, with the <PRT> key.
The calibration record of the last calibration can be found under the menu item, Configuration / Turbidity / Calibration record.
31.10.03 16:13 Turb 430 IR Ser. no. 12345678
Calibration Turbidity Calibration date 31.10.03 16:13:33 Calibration interval 90 d
Perform the following preparatory activities when you want to calibrate:
1 Keep the cells with the required calibration standards ready
and mark them as necessary (see page 33).
2 Clean the cell (see section 5.2.2).
3 Insert the cell (see section 4.2).
Carrying out calibration
Note
For turbidity values under 1 FNU/NTU, follow the instructions given in the appendix (see A
PPENDIX: TURBIDITY VALUES UNDER 1 FNU/NTU).
1 Press the <CAL/ZERO> key.
The menu-guided calibration begins. Follow the instructions on the display.
Turbidity \ Calibration
i Insert standard
1000 FNU/NTU
i Press and hold <START> i Align sample
2 Insert the cell with the displayed calibration standard (here e.g.
1000 FNU/NTU) in the cell shaft (see section 4.2).
ba75507e03 07/2006
37
Operation Turb 430 IR/T
3 Align the cell:
! Marked cell:
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and hold the <START/ENTER> key until the
measured value is displayed.
! Unmarked cell (see page 33)
Press and hold the <START/ENTER> key.
Slowly and in small steps turn the cell by one complete
rotation (by 360 °).
After each step wait for a short time until the displayed
measured value is stable.
– Turn the cell back to the position with the lowest
measured value.
Turbidity \ Calibration
i Turb. = 1000 FNU/NTU i Start calibration
by releasing <START>
4 Release the <START/ENTER> key.
Measurement of the calibration standard begins.
Note
Before measuring the third calibration standard of 0.02 FNU/NTU you can exit the calibration with <ESC> at any time. The new calibration data is discarded. The old calibration data is further used.
5 Repeat the steps 4 - 6 with the calibration standards, 10.0 FNU/
NTU and 0.02 FNU/NTU. After measuring the 0.02 FNU/NTU calibration standard, the
calibration result is displayed. Calibration is completed.
38
6 Confirm the calibration result with <START/ENTER>.
The calibration record is displayed.
7 Confirm the calibration record with <START/ENTER>.
The display shows instructions for the first measurement.
ba75507e03 07/2006
Turb 430 IR/T Operation
Turbidity
i Insert sample i Press and hold <START> i Align sample
Note
If i Calibration error! was displayed as the calibration result, a note appears on the display to recalibrate before measuring. Should a valid calibration not be possible the meter also offers to continue measuring with the last valid calibration data.
ba75507e03 07/2006
39
Operation Turb 430 IR/T

4.6 Saving

The meter has 1000 memory locations for measurement datsets.
You can transmit measured values (datasets) to the data memory with the <STO> key. Each storage process transmits the current dataset to the interface at the same time.
The number of memory locations that are still free is displayed in the Store menu. The number of memory locations that are occupied is displayed in the System Measured value memory menu.
Measurement dataset A complete dataset consists of:
! Date/time
! ID number (ID)
! Measured value

4.6.1 Storing measurement datsets

Proceed as follows to transmit to the data memory and simultaneously output to the interface a measurement dataset:
1 Press the <STO> key.
The Store display appears.
Store (996 free)
i 02.02.2005 11:24:16
16.80 FNU/NTU
ID:
Store (ID: 1)
1
2 Using <><>, <START/ENTER> and the number keys,
change and confirm the ID number (ID) as necessary (0 ...
999).
3Using <START/ENTER> or <STO>, confirm Store.
The dataset is stored. The instrument switches to the measured value display.
40
Note
A measurement dataset is stored quickly by twice pressing <STO>. It is stored with the ID last set.
ba75507e03 07/2006
Turb 430 IR/T Operation
If the memory is full You can erase the entire memory (see section 4.6.5), or overwrite the
oldest dataset with the next storing procedure. A security prompt appears before a dataset is overwritten.

4.6.2 Filtering measurement datsets

The functions to display and download stored measurement datsets (see section 4.4.1) refer to all stored measurement datsets that correspond to the specified filter criteria.
The settings are given in the menu, Configuration / System / Measured value memory / Data filter. Move to the Configuration menu with the <MENU> key.
Data filter
Menu item Setting/function Description
Filter Filter criteria:
No filter Data filter switched off
ID Selection according to ID
number
Date Selection according to
period
ID + Date Selection according to
period and ID number
ID Entry of filter criteria
These menu items are made visible by selecting
Date
the filter criteria in the Filter menu.
ba75507e03 07/2006
41
Operation Turb 430 IR/T

4.6.3 Displaying measurement datsets

You can read out stored datasets to the display. Only those datasets are displayed that correspond to the selected filter criteria (see section 4.6.2).
Start reading out the data to the display in the menu, Configuration / System / Measured value memory / Display.
Representation of a
dataset
02.02.2005 11:24:16 ID: 1
16.80 FNU/NTU
i Scroll with
▲ ▼
Further datasets that correspond to the filter criteria are displayed with the <><> keys.
Quitting the display To quit the display of stored measurement datasets, you have the
following options:
! Switch directly to the measured value display with <M> (short
pressure).
! Leave the display and switch to the superordinate menu with <ESC>
or <START/ENTER>.

4.6.4 Download the measurement datsets to the RS232 interface

42
You can download stored datasets to the RS232 interface. Only those datasets are downloaded that correspond to the selected filter criteria (see section 4.6.2).
The datasets are downloaded in the specified output format (see section 4.7.3).
The data download to the interface is started in the menu, Configuration / System / Measured value memory / RS232 download.
ba75507e03 07/2006
Turb 430 IR/T Operation

4.6.5 Erasing stored measurement datasets

You can erase the stored measurement datsets altogether if you do no longer need them.
Erasing all measurement datsets is done in the menu, Configuration / System / Measured value memory / Delete.
Note
Erasing individual datasets is not possible. If all memory locations are occupied, however, it is possible to overwrite the oldest dataset at a time. A security prompt appears before a dataset is overwritten.

4.7 Transmitting data (RS 232 interface)

Via the RS 232 interface, you can transmit data to a PC or an external printer.
The data can be transmitted to a PC, for example, with the aid of a so­called terminal program.
Generally, a terminal program serves to establish a connection to a device on a data interface and to communicate with the device via a console on the display. A terminal program usually offers the possibility to save the contents of the console in a text file or print it. If the terminal program is connected to the meter, it can receive data from the meter and display it on the console.
Terminal programs are available for different operating systems by different manufacturers. Windows (version 95 to XP) contains the "HyperTerminal" terminal program. It is in the program menu under Accessories.
For more detailed information please refer to the user information of the terminal program.
The settings required for the use of the "HyperTerminal" program are given in section 4.7.1.
Note
When using the "HyperTerminal" program you can load the transmission data automatically using the *.ht file provided on the CD.
ba75507e03 07/2006
43
Operation Turb 430 IR/T

4.7.1 Connecting a PC/external printer

Connect the interface to the devices via the AK540/B (PC) or AK540/S (external printer) cable.
Caution
The RS232 interface is not galvanically isolated. When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Baud rate can be selected from: 1200, 2400, 4800, 9600,
19200 The baud rate must agree with the baud rate set on the PC/printer.
Handshake RTS/CTS
Socket assignment
PC only:
Parity none
Data bits 8
Stop bits 1s
3
4
6
1
5
1 RTS 2 RxD
2
3 TxD 4 DTR 5 SG
RS 232 REC
6 CTS

4.7.2 Configuring the RS232 interface

For error-free data transmission, the RS232 interface should be set to the same transmission speed (Baud rate) on the Turb 430 IR/T and PC/printer.
44
You can set the following values for the baud rate on the Turb 430 IR/ T: 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu, Configuration / System / Interface / Baud rate.
ba75507e03 07/2006
Turb 430 IR/T Operation

4.7.3 Selecting the output format of datasets

For downloading data to the interface you can select the output format.
It is selected in the menu, Configuration / System / Interface / Output format.
The ASCII output format delivers formatted datasets. The CSV output format delivers datasets separated by ";".
Output format, ASCII
Turb 430 IR Ser. no. 12345678
31.10.04 09:56:20 ID: 1
16.01 FNU/NTU
________________________________________
Turb 430 IR Ser. no. 12345678
31.10.04 15:48:08 ID: 1
26.01 FNU/NTU
________________________________________
etc...
ba75507e03 07/2006
45
Operation Turb 430 IR/T
Output format, CSV
123456
15.01.05;15:06:49;1;26.5;FNU/NTU;VALID;
15.01.05;16:06:49;1;16.5;FNU/NTU;VALID;
Data Description
1 Date Date of storing
2 Time Time of storing
3 ID adjusted ID
4 Measured value ! Measured value or
! Upper/lower measuring range limit
(only with measured value status, OFL/ UFL)
5 Unit of 4 ! Unit of the measured value
6 Measured value
status of 4
! VALID: Measured value valid
! INVALID: Measured value invalid
! UFL: Measured value below the lower
measuring range limit
! OFL: Measured value above the upper
measuring range limit
46
ba75507e03 07/2006
Turb 430 IR/T Operation

4.7.4 Transmitting data

The following table shows which data are transmitted to the interface in which way:
Data Operation / description
Current measured value
! Press <PRT>.
! Simultaneously with every manual storage
process.
Stored measured values
! Display stored dataset and press <PRT>.
! All datasets according to the filter criteria via
the RS232 download function (see section 4.6.2).
Note
With the <PRT> key you output data that is being shown on the display to the interface (displayed measured values, stored measurement datsets, calibration record).
ba75507e03 07/2006
47
Operation Turb 430 IR/T

4.8 Reset

You can reset (initialize) all system and measurement settings.

4.8.1 Resetting the system settings

With the System / Reset function, all resettable settings are reset.
! Settings for Turbidity (see section 4.8.2)
! System settings
System setting Default settings
Baud rate 4800 Baud
Output format ASCII
Illumination Auto off
Contrast 50 %
Brightness 50 %
Switchoff time 30 min
Beep On

4.8.2 Resetting turbidimeter settings

With the Turbidity / Reset function, all turbidimeter settings are reset.
Setting Default settings
Calibration interval 90 d
48
ba75507e03 07/2006
Turb 430 IR/T Operation

4.9 Meter information

The following meter information is listed in the Configuration / Info menu:
! Model designation
! Software version
! Series number of the meter
Info
i Model: Turb 430 IR i Software: V 0.35 i Ser. no.
: 04280003

4.10 Software update

With a software update you obtain the current instrument software (see appendix).
The current software version can be found on the Internet under www.WTW.com.
The proceeding for updating the software is given in the appendix (see A
PPENDIX: FIRMWARE UPDATE).
ba75507e03 07/2006
49
Operation Turb 430 IR/T
50
ba75507e03 07/2006
Turb 430 IR/T Maintenance, cleaning, disposal

5 Maintenance, cleaning, disposal

5.1 Maintenance

The meter is almost maintenance-free. The only maintenance task is replacing the batteries or accumulator pack.

5.1.1 Inserting/exchanging the batteries

Caution
Make sure that the poles of the batteries are the right way round. The ± signs on the batteries must correspond to the ± signs in the battery compartment.
1
2
1 Open the battery compartment:
Unscrew the two screws (1) on the underside of the meter,
Remove the lid of the battery compartment (2).
ba75507e03 07/2006
2 If necessary, take four old batteries out of the battery
compartment.
3 Insert four batteries (3) in the battery compartment.
4 Close the battery compartment and fix it with the screws.
51
Maintenance, cleaning, disposal Turb 430 IR/T

5.1.2 Retrofitting the accumulator pack

Caution
Use original WTW accumulator packs only.
Together with the power pack the accumulator pack is available as an accessory (see section 8.1).
1
2
3
1 Open the battery compartment:
Unscrew the two screws (1) on the underside of the meter,
Remove the lid of the battery compartment (2).
2 If necessary, take four old batteries out of the battery
compartment.
3 Connect the cable of the accumulator pack with the socket (3)
on the bottom of the battery compartment and insert the accumulator pack in the battery compartment.
4 Close the battery compartment and fix it with the screws.
52
ba75507e03 07/2006
Turb 430 IR/T Maintenance, cleaning, disposal

5.2 Cleaning

Occasionally wipe the outside of the meter with a damp, lint-free cloth. Disinfect the housing with isopropanol as required.
Caution
The housing components are made out of synthetic materials (polyurethane, ABS and PMMA). Thus, avoid contact with acetone and similar detergents that contain solvents. Remove any splashes immediately.

5.2.1 Cleaning the cell shaft

If liquid is in the cell shaft (e.g. due to a spilled cell), clean the cell shaft as follows:
1 Switch the Turb 430 IR/T off and pull out the power plug.
2 Rinse the cell shaft with distilled water.

5.2.2 Cleaning the cells

Cells have to be clean, dry, and free of fingerprints and scratches. Therefore, clean them regularly:
1 Clean the cells inside and out with hydrochloric acid or
laboratory soap.
2 Rinse out several times with distilled water.
3 Let them dry in the air.
4 Only hold the cells by the top or by the light protection cap so
that the optical path is not impaired.
5 Before measuring, clean the cell with the enclosed cleaning
cloth.
Note
Scratches in the glass change the optical characteristics of the cell and falsify the measured value. For this reason, never use scratched cells!
ba75507e03 07/2006
53
Maintenance, cleaning, disposal Turb 430 IR/T

5.3 Disposal

Packing This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the instrument against damage during transport.
Batteries/accumulator
pack
NiMH
Meter Dispose of the meter without the batteries and accumulator pack as
Remove the batteries or accumulator pack from the meter (see section 5.1). Dispose of the batteries or accumulator pack at a suitable facility according to local legal requirements. It is illegal to dispose of the accumulator pack with household refuse.
electronic waste at an appropriate collection point.
54
ba75507e03 07/2006
Turb 430 IR/T What to do if...

6 What to do if...

6.1 General errors

Display, LoBat
Instrument does not
react to keystroke
Error message,
Error
0, 8, 16, 16384
Cause Remedy
– The batteries or accumulator
pack are largely depleted
Cause Remedy
Software error
Operating condition undefined
or EMC load unallowed
Cause Remedy
Instrument error Repeat measurement
Insert new batteries
Charge the accumulator pack
(see section 3.2)
– Processor reset:
Press the <START/ENTER> and <PRT> key simultaneously.
– Meter defective,
send meter to WTW for repair and quote the error number
Error message,
Measured values
obviously incorrect
Measured value display
< 0.01 FNU
ba75507e03 07/2006

6.2 Turbidity

Cause Remedy
Cell not correctly inserted – Lock cell into place
Cell contaminated Clean the cell
Calibration too old Carry out calibration
Cause Remedy
Calibration defective Carry out calibration
Measured value outside the
measuring range
– not possible
55
What to do if... Turb 430 IR/T
56
ba75507e03 07/2006
Turb 430 IR/T Technical data

7 Technical data

7.1 General data

7.1.1 Turb 430 IR/T
Dimensions approx. 236 x 86 x 117 mm
Weight approx. 0.6 kg (without batteries)
Mechanical structure Type of protection IP 67
Electrical safety Protective class III
Test certificates cETLus, CE, FCC
Ambient
conditions
Allowable relative
humidity
Power
supply
Storage - 25 °C ... + 65 °C
Operation 0 °C ... + 50 °C
Climatic class 2
Yearly mean: 30 days /year: other days:
Batteries 4 x 1.5 V, type AA
Operating time with battery operation
Accumulator pack (optional)
Power pack Charging device (optional)
75 % 95 % 85 %
Turb 430 IR: approx. 3000 measurements Turb 430 T: approx. 2000 measurements
5 x 1.2 V nickel metal hydride (NiMH), type AA
FRIWO FW7555M/09, 15.1432.500-00 Friwo Part. No. 1883259 Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA Output: 9 V = / 1.5 A Connection max. overvoltage category II
Primary plugs contained in the scope of delivery: Euro, US, UK and Australian.
interface
ba75507e03 07/2006
Serial
Connection of the cable AK 540/B or AK 540/S
Baud rate adjustable:
1200, 2400, 4800, 9600, 19200 Baud
Type RS232
Data bits 8
Stop bits 2
Parity None
Handshake RTS/CTS
Cable length max. 15 m
57
Technical data Turb 430 IR/T
Guidelines
and norms used
EMC EC guideline 89/336/EEC
EN 61326-1/A3:2003 FCC Class A
Instrument safety EC guideline 73/23/EEC
EN 61010-1 :2001 UL STD 61010-1 CAN/CSA-C22.2 No. 61010-1
Climatic class VDI/VDE 3540
IP protection EN 60529:1991
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
58

7.1.2 LabStation

Dimensions approx. 236 x 82 x 170 mm
Weight approx. 0.6 kg
ba75507e03 07/2006
Turb 430 IR/T Technical data

7.2 Turbidity

7.2.1 Turb 430 IR
Measuring principle Nephelometric measurement according to DIN EN ISO 7027
Light source Infrared LED
Measuring range 0.01 ... 1100 FNU/NTU
Resolution in the range
0,01 ... 9,99
in the range 10,0 ... 99,9
in the range 100 ... 1100
max. 0.01 FNU/NTU
max. 0.1 FNU/NTU
max. 1 FNU/NTU
Accuracy in the range
0 ... 1100 FNU/NTU
± 2 % of the measured value or ± 0.01 FNU/NTU
Reproducibility 0,5% of the measured value
Response time 4 seconds
Calibration Automatic 3-point calibration
7.2.2 Turb 430 T
Measuring principle Nephelometric measurement according to US EPA 180.1
Light source White light tungsten lamp
Measuring range 0.01 ... 1100 NTU
Resolution in the range
0,01 ... 9,99
max. 0.01 NTU
in the range 10,0 ... 99,9
max. 0.1 NTU
in the range 100 ... 1100
max. 1 NTU
Accuracy in the range
Reproducibility 1% of the measured value
Response time 7 seconds
Calibration Automatic 3-point calibration
ba75507e03 07/2006
0 ... 500 NTU
in the range 500 ... 1100 NTU
± 2 % of the measured value or ± 0.01 NTU
± 3 % of the measured value
59
Technical data Turb 430 IR/T
60
ba75507e03 07/2006
Turb 430 IR/T Accessories, options

8 Accessories, options

8.1 WTW accessories

Description Model Order no.
LabStation pHotoFlex LS 251 301
Accumulator with Turb 430 IR/T
pHotoFlex BB 251 300
power pack
3 empty cuvettes, 28 x 60 mm LKS28-Set 251 302
Calibration standard kit for
Kal.Kit Turb 430 IR 600 560
Turb 430 IR
Calibration standard kit for
Kal.Kit Turb 430 T 600 561
Turb 430 T
Thermoprinter
Needle printer
* a connection cable is required to connect the printer (see
section 8.1.1)
*
*
P3001 250 045
LQ 300+ 250 046

8.1.1 Connection cable:

PC You can connect a PC (serial COM interface) to the Turb 430 IR/T as
follows:
ba75507e03 07/2006
Description Model Order no.
! Connection PC - Turb 430 IR/T
– Cable AK 540/B 902 842
+ USB adapter
Ada USB 902 881
(for USB connection on PC)
! Connection PC - LabStation
– Zero modem cable included in the scope of
delivery of the LabStation
+ USB adapter
Ada USB 902 881
(for USB connection on PC)
61
Accessories, options Turb 430 IR/T
Thermoprinter
You can connect the P3001 to the Turb 430 IR/T in the following ways:
Description Model Order
no.
! Connection P3001 - Turb 430 IR/T
Cable AK 540/S 902 843
! Connection P3001 - LabStation
– Cable AK 3000 250 745
in conjunction with an adapter
Specialist shops
(socket - socket) [GenderChanger]
or:
– Cable, 2 x 9-pin (socket - plug) Specialist shops
Needle printer You can connect an LQ300 needle printer to the Turb 430 IR/T in one
of the following ways:
Description Model Order
no.
! Connection LQ300 - Turb 430 IR/T
– Cable AK 540/B 902 842
with adapter
Specialist shops
9-pin (plug) - 25-pin (plug)
! Connection LQ300 - LabStation
– Cable AK/LQ300 250 746
in conjunction with an adapter
Specialist shops
(socket - socket) [GenderChanger]
or:
– Zero modem cable,
Specialist shops
9-pin (socket) - 25-pin (plug)
62
ba75507e03 07/2006
Turb 430 IR/T Lists

9Lists

This chapter provides additional information and orientation aids.
Abbreviations The list of abbreviations explains the indicators and the abbreviations
that appear on the display and in the manual.
Specialist terms The glossary briefly explains the meaning of the specialist terms.
However, terms that should already be familiar to the target group are not described here.
ba75507e03 07/2006
63
Lists Turb 430 IR/T
Abbreviations
Cal Calibration
dDay
h Hour
j Year
LoBat Batteries almost empty (Low battery)
mMonth
s Second
S Slope (internat. k)
SELV Safety Extra Low Voltage
Slp. Slope determined with calibration
64
ba75507e03 07/2006
Turb 430 IR/T Lists
Glossary
Adjusting To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains within the tolerance.
Calibration Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
Cell Vessel that takes a liquid sample for measurement.
LED Light Emitting Diode
LEDs are used as the light source in the Turb 430 IR/T.
Measured parameter The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or DO concentration.
Measured value The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring system The measuring system comprises all the devices used for measuring,
e. g. meter and sensor. In addition, there is the cable and possibly an amplifier, terminal strip and armature.
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.
Reset Restoring the original condition of all settings of a measuring system.
Resolution Smallest difference between two measured values that can be
displayed by a meter.
Standard solution The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Test sample Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed.
ba75507e03 07/2006
65
Lists Turb 430 IR/T
66
ba75507e03 07/2006
Turb 430 IR/T Index

10 Index

A
Accumulator pack
Charging time
Air ..........................................................33
Aligning and marking a cell .......................33
Authorized use .........................................10
Automatic switchoff function ................14, 19
.....................................13
C
Calibration ...............................................36
Calibration order ......................................36
Calibration points and measuring ranges ...36
Calibration standards ...............................36
Cleaning .................................................53
Connecting a PC ......................................44
Connecting a printer .................................44
Connecting sensors ...................................7
D
Data filter ................................................41
Dataset ...................................................40
Date and time ....................................25, 32
Default settings
System settings
Turbidimeter .......................................48
Display ................................................7, 31
Display illumination ..............................7, 20
..................................48
F
Filter .......................................................41
Firmware update ......................................69
I
Initial commissioning ................................17
Initialization .............................................48
Inserting a cell .........................................20
Interface ..................................................31
J
Jack field ...................................................7
M
Measured value display ........................... 21
Measured value memory ......................... 40
Measurement dataset .............................. 40
Measuring range overflow ........................ 35
Measuring turbidity .................................. 34
Menus (navigation) .................................. 21
Messages ............................................... 22
N
Navigation .............................................. 21
O
Obligations of the purchaser ..................... 11
Operating modes .................................... 21
Operational safety ................................... 10
P
Power pack ............................................. 13
Print ....................................................... 47
R
Reset ..................................................... 48
RS232 socket assignment ....................... 44
S
Safe operation ........................................ 11
Safety ...................................................... 9
Saving .................................................... 40
Scope of delivery .................................... 13
Storage .................................................. 29
Switching on ........................................... 19
System settings ...................................... 28
T
Target group ............................................. 9
Transmitting data .................................... 43
Transmitting measured values .................. 43
Turbidity ................................................. 33
K
Keys .........................................................6
ba75507e03 07/2006
67
Index Turb 430 IR/T
68
ba75507e03 07/2006
Turb 430 IR/T

Appendix: Firmware update

General information With the "Firmware Update Turb430" program you can update the
firmware of the Turb 430 IR/T to the latest version with the aid of a Personal Computer.
A free serial interface (COM port) on your PC and an interface cable is required for this (see chapter 8 A
Program installation Install the firmware update program on your PC with the
"Turb430_Vx_yy_English.exe" installation program.
Program start Start the "Firmware Update Turb430" program from the WTW directory
in the Windows start menu. The program automatically selects the first free serial interface (COM port). The selected interface is displayed on the left side of the status line on the screen bottom.
Via the language menu you can change the adjusted language.
CCESSORIES, OPTIONS).
Firmware update Proceed as follows:
1 With the aid of an interface cable, connect the Turb 430 IR/T to
the serial interface (COM port) of the PC named in the status line.
2 Make sure the Turb 430 IR/T is switched on.
3 To start the updating process click the OK button.
4 Then follow the instructions of the program.
A corresponding message and a progress bar (in %) appear during the programming procedure.
The programming procedure takes approx. four minutes. A final message appears after the successful programming
procedure. The firmware update is completed with this.
5 Disconnect the meter from the PC.
The instrument is ready for operation.
After switching the meter off and on again you can check on the start display whether the meter has taken over the new software version.
ba75507e03 07/2006
69
Turb 430 IR/T
Appendix: Turbidity values under 1 FNU/NTU
With turbidity values (under 1 FNU/NTU), the measured value is strongly influenced by the cell and its alignment.
In order to increase measuring accuracy with turbidity values under 1 FNU/NTU, calibration in the 0.02 FNU/NTU standard and later measurement should be carried out in the same cell. For calibration in the standards 10.0 and 1000 FNU/NTU follow the instructions on the display.
Proceed as follows to measure turbidity values under 1 FU/NTU:
Calibration procedure:
Measuring
1 Press the <CAL/ZERO> key.
The menu-guided calibration begins.
2 Carry out calibration of the standards 1000 FNU/NTU and
10.0 FNU/NTU in a clean and unscratched cell according to the menu guidance.
3 Fill the cleaned cell with the 0.02 FNU/NTU standard and
calibrate.
4 Mark the alignment of the cell.
Calibrate
! after the calibration interval has expired
! with a temperature change
5 Fill the marked and cleaned cell with test sample, align it
with the marking and measure.
6 If necessary, fill the marked and cleaned cell once more
with test sample and carry out further measurements.
70
ba75507e03 07/2006
Wissenschaftlich-Technische Werkstätten GmbH
Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim
Germany
Tel: +49 (0) 881 183-0
+49 (0) 881 183-100 Fax: +49 (0) 881 183-420 E-Mail: Info@WTW.com Internet: http://www.WTW.com
Loading...