The compact, digital precision meter MultiLab 4010-3 enables you to carry out
pH measurements, ORP measurements, conductivity measurements and dissolved oxygen (D.O.) measurements quickly and reliably.
The MultiLab 4010-3 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The MultiLab 4010-3 supports you in your work with the following functions:
proven calibration procedures
automatic stability control (AR),
automatic sensor recognition
CMC (continuous measurement control)
QSC (sensor quality control).
1 Keypad (antibacterial)
2Display
3 Connectors
Due to its antibacterial properties, the keypad of the
MultiLab 4010-3 is especially suitable for applications in an environment where hygiene is important (see
DATA, page 94).
SECTION 14.2 GENERAL
ba76143e0104/20137
OverviewMultiLab 4010-3
1.2Sensors
1.2.1IDS sensors
IDS sensors
support the automatic sensor recognition
show only the settings relevant to the specific sensor in the setting menu
process signals in the sensor digitally so that precise and interference-free
measurements are enabled even with long cables
facilitate to assign a sensor to a measured parameter with differently colored
couplings
have quick-lock couplings with which to fix the sensors to the meter.
Sensor data from
IDS sensors
IDS sensors transmit the following sensor data to the meter:
SENSOR ID
– Sensor name
– Sensor series number
Calibration data
Measurement settings
The calibration data are updated in the IDS sensor after each calibration procedure. A message is displayed while the data are being updated in the sensor.
In the measured value display, you can display the sensor name
and series number of the selected sensor with the [Info] softkey.
You can then display further sensor data stored in the sensor with
the [More] softkey (see section 4.1.6 S
ENSORINFO, page 17).
8ba76143e0104/2013
MultiLab 4010-3Overview
1.2.2IDS adapter for analog pH sensors
With the aid of an IDS adapter, you can also operate analog sensors
on the MultiLab 4010-3. The combination of the IDS adapter and analog sensor
behaves like an IDS sensor.
The MultiLab 4010-3 has a recess where the IDS adapter (ADA 94/IDS DIN or
ADA 94/IDS BNC), which is available as an accessory, can be permanently
mounted.
The IDS adapter replaces a digital input (channel 2) with a connector for an
analog pH sensor (DIN or BNC plug) and a temperature sensor.
1.2.3Automatic sensor recognition
The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibrating
to operate different IDS sensors at one meter without recalibration
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor series number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to activate the correct cell constant for conductivity sensors automatically
to hide menus automatically that do not concern this sensor
To be able to use the automatic sensor recognition, a meter that supports the
automatic sensor recognition (e.g. MultiLab 4010-3) and a digital IDS sensor
are required.
In digital IDS sensors, sensor data are stored that clearly identify the sensor.
The sensor data are automatically taken over by the meter.
ba76143e0104/20139
SafetyMultiLab 4010-3
2Safety
2.1Safety information
2.1.1Safety information in the operating manual
This operating manual provides important information on the safe operation of
the meter. Read this operating manual thoroughly and make yourself familiar
with the meter before putting it into operation or working with it. The operating
manual must be kept in the vicinity of the meter so you can always find the information you need.
Important safety instructions are highlighted in this operating manual. They are
indicated by the warning symbol (triangle) in the left column. The signal word
(e.g. "Caution") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not
followed.
CAUTION
indicates a possibly dangerous situation that can lead to slight
(reversible) injury if the safety instruction is not followed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
2.1.2Safety signs on the meter
Note all labels, information signs and safety symbols on the meter. A warning
symbol (triangle) without text refers to safety information in this operating
manual.
2.1.3Further documents providing safety information
The following documents provide additional information, which you should
observe for your safety when working with the measuring system:
•Operating manuals of sensors and other accessories
•Safety datasheets of calibration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
2.2Safe operation
2.2.1Authorized use
The authorized use of the meter consists exclusively of the measurement of the
pH, ORP, conductivity and dissolved oxygen in a laboratory environment.
10ba76143e0104/2013
MultiLab 4010-3Safety
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 14 T
ECHNICALDATA, page 94).
Any other use is considered unauthorized.
2.2.2Requirements for safe operation
Note the following points for safe operation:
•The meter may only be operated according to the authorized use specified
above.
•The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
•The meter may only be operated under the environmental conditions
mentioned in this operating manual.
•The meter may not be opened.
2.2.3Unauthorized use
The meter must not be put into operation if:
•it is visibly damaged (e.g. after being transported)
•it was stored under adverse conditions for a lengthy period of time (storing
conditions, see section 14 T
ECHNICALDATA, page 94).
ba76143e0104/201311
CommissioningMultiLab 4010-3
3Commissioning
3.1Scope of delivery
MeterMultiLab 4010-3
USB cable (A plug on mini B plug)
Power pack
Stand with stand base
Short instructions
Detailed operating manual
CD-ROM
3.2Power supply
The MultiLab 4010-3 is supplied with power in the following ways:
Mains operation with the supplied power pack
Operation of the system clock with a buffer battery if there is no mains power
supply (see section 12.1.2 E
XCHANGINGTHEBATTERY, page 86).
3.3Initial commissioning
Perform the following activities:
Connect the power pack
(see section 3.3.1 C
Switch on the meter
(see section 4.2 S
Set the date and time
(see section 4.5.5 E
page 25)
Mount the stand
(see operating manual of the stand)
ONNECTINGTHEPOWERPACK, page 13)
WITCHINGONTHEMETER, page 19)
XAMPLE 2 ONNAVIGATION: SETTINGTHEDATEANDTIME,
12ba76143e0104/2013
MultiLab 4010-3Commissioning
3.3.1Connecting the power pack
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see section 14.2
G
ENERALDATA, page 94).
CAUTION
Use original power packs only (see section 14.2 GENERALDATA,
page 94).
1.Connect the plug of the power pack to the socket for the power pack on
the MultiLab 4010-3.
2.Connect the original power pack to an easily accessible power outlet.
The meter performs a self-test.
ba76143e0104/201313
OperationMultiLab 4010-3
4Operation
4.1General operating principles
4.1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <ENTER>) generally indicates a short keystroke (press
and release) in this operating manual. A long keystroke (hold for approx. 2 sec)
is indicated by the underscore behind the key symbol (e.g. <ENTER
_>).
<F1>
<F4>
<On/Off>Switches the meter on or off
<M>Selects the measured parameter
<CAL>
_>
<CAL
<AR>Freezes the measured value (HOLD function)
<ESC>Switches back to the higher menu level /
<STO>
_>
<STO
<RCL>
<RCL
_>
Softkeys providing situation dependent functions, e.g.:
<F1>/[Info]: View information on a sensor
Calls up the calibration procedure
Displays the calibration data
Switches the AutoRead measurement on or off
Cancels inputs
Saves a measured value manually
Opens the menu for the automatic save function
Displays the manually stored measured values
Displays the automatically stored measured values
<><>
<><>
<ENTER>
<ENTER
<PRT>
<PRT_>
14ba76143e0104/2013
_>
Menu control, navigation
Opens the menu for measurement settings / Confirms
entries
Opens the menu for system settings
Outputs stored data to the interface
Outputs displayed data to the interface at intervals
MultiLab 4010-3Operation
4
3
2
5
7
8
9
6
1
HOLD AR
AutoCal TEC
03.04.2013 08:00Info
YSI
4.1.2Display
Example (pH):
1 Status information (meter)
2 Status information (sensor)
3 Measured value
4 Measured parameter
5 Continuous measurement control (CMC function)
6 Channel display: Plug position of the sensor
7 Sensor symbol (calibration evaluation, calibration interval)
8 Measured temperature (with unit)
9 Softkeys and date + time
4.1.3Status information
AutoCal
e.g. YSI
ConCalCalibration with any buffers
ErrorAn error occurred during calibration
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (<AR> key)
Calibration with automatic buffer recognition, e.g. with the buffer set: YSI buffers
Data are automatically output to the USB-B interface at intervals
Data are output to the USB-A interface (USB flash drive)
Data are output to the USB-A interface (USB printer). If there is a USB-B
connection at the same time (e.g. to a PC), the data are output to the USBB interface only.
Connection to a PC is active (USB-B interface)
ba76143e0104/201315
OperationMultiLab 4010-3
746
351
bac
2
4.1.4Instrument connectors
1 IDS sensors: (pH, ORP, conductivity, D.O.)
a) Channel 1
b) Channel 2
c) Channel 3
2 USB B (device) interface
3 Stirrer (ProOBOD)
4 USB-A (host) interface
5 Power pack
6 Service interface
7 Cover plate
The cover plate closes the mounting slot for the IDS adapter (4010-2/3 pH Adapter DIN or 4010-2/3 pH Adapter BNC) available as an
accessory
CAUTION
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting).
YSI IDS sensors and IDS adapters meet these requirements.
16ba76143e0104/2013
MultiLab 4010-3Operation
1
03.04.2013 08:00
Info
03.04.2013 08:00Info
4.1.5Channel display
The MultiLab 4010-3 manages the connected sensors and displays which sensor is plugged to which connection.
1 Channel display: Display of the plug position for the respective
parameter
The red bar indicates for each connected sensor to which plug position (channel) of the meter it is connected.
4.1.6Sensor info
You can display the current sensor data and sensor settings of a connected
sensor at any time. The sensor data are available in the measured value display with the /[Info] softkey.
1.In the measured value display:
Display the sensor data (sensor name, series number) with [Info].
ba76143e0104/201317
OperationMultiLab 4010-3
03.04.2013 08:00
More
4110
B092500013
03.04.2013 08:00
Man. temperature:25 °C
pH resolution0.001
mV resolution0.1
BufferYSI
Calibration interval7d
Unit for slopemV/pH
QSC:off
Software version1.00
4110
B092500013
03.04.2013 08:00Info
03.04.2013 08:00Info
2.Display further sensor data (settings) with [More].
4.1.7Display of several sensors in the measuring mode
The measured values of the connected sensors can be displayed in the following ways:
Clear display of all connected sensors
Detailed display of one sensor
(e.g. incl. CMC feature with pH sensors)
With the softkey you can very easily switch between the two display types. The
suitable softkey is displayed depending on the operating situation.
18ba76143e0104/2013
MultiLab 4010-3Operation
03.04.2013 08:00
Info
4.2Switching on the meter
1.Switch the meter on with <On/Off>.
The meter performs a self-test.
2.Connect the sensor.
The meter is ready to measure.
If the user administration function is activated, the Login dialog
appears after the meter is switched on (see section 4.4 L
USERNAME, page 19).
OGINWITH
The user administration function is not active in the delivery condition.
4.3Switching off the meter
1.Switch off the meter with <On/Off>.
4.4Login with user name
After activation of the user administration by the administrator, measurements
are only possible after login with a user name. The user name is documented
with the measured values and in records.
All user names entered by the administrator are listed in the User name menu.
The administrator determines for each user whether or not a password is
required for the login to the meter.
If the Password menu item is grayed out, no password is required for the login.
1.Switch on the meter with <On/Off>.
The Login dialog appears.
ba76143e0104/201319
OperationMultiLab 4010-3
User nameAdmin
Password####
Change password
Login
03.04.2013 08:00
2.Using <><>, select the menu item, User name and confirm with
<ENTER>.
The user name is highlighted.
3.Using <><>, select a user name and confirm with <ENTER>.
The login is done immediately if no password is required.
If a sensor is connected the measured value display appears.
4.If a password is required:
Using <><>, select the menu item, Password and confirm with
<ENTER>.
The user specifies the password when he or she first logs in with a
user name.
A valid password consists of 4 digits.
The user can change the password with the next login.
5.Change the digit of the highlighted position with <><>.
Switch to the next position of the password with <><>.
When the password was completely entered, confirm with <ENTER>.
The login takes place. If a sensor is connected the measured value display appears.
Changing the
If the administrator has set up the access with password protection:
password
1.Switch on the meter with <On/Off>.
2.Using <><>, select the menu item, User name and confirm with
3.Using <><>, select a user name and confirm with <ENTER>.
4.Using <><>, select the menu item, Change password and confirm
20ba76143e0104/2013
The Login dialog appears.
<ENTER>.
The user name is highlighted.
with <ENTER>.
MultiLab 4010-3Operation
5.In the Password field, enter the old password with <><> and
<><>and confirm it with <ENTER>.
6.In the New password field, enter the new password with <><> and
<><>and confirm it with <ENTER>.
The password is changed.
The login takes place. If a sensor is connected the measured value display appears.
Forgotten the
password?
Contact the administrator.
4.5Navigation
4.5.1Operating modes
Operating
mode
MeasuringThe measurement data of the connected sensor are shown
CalibrationThe course of a calibration with calibration information, func-
Storage in
memory
Transmittin
g data
SettingThe system menu or a sensor menu with submenus, set-
Explanation
in the measured value display
tions and settings is displayed
The meter stores measuring data automatically or manually
The meter transmits measuring data and calibration records
to a USB interface automatically or manually.
tings and functions is displayed
4.5.2Measured value display
In the measured value display, you can
use <><> to select one of several connected sensors. The selected
sensor is displayed with a colored background.
The following actions / menus refer to the selected sensor
open the menu for calibration and measurement settings with <ENTER>
(short
open the Storage & config menu with the sensor-independent settings by
pressing <ENTER
change the display in the selected measuring screen (e. g. pH <−> mV) by
pressing <M>.
4.5.3Menus and dialogs
The menus for settings and dialogs in procedures contain further subelements.
ba76143e0104/201321
keystroke)
_> (long keystroke, approx. 2 s).
OperationMultiLab 4010-3
General
Interface
Clock function
Service information
Reset
System
03.04.2013 08:00
Language:English
Audio signal:on
Brightness:12
Temperature unit:°C
Stability control:on
General
03.04.2013 08:00
pH
03.04.2013 08:00
Calibration record
Calibration data storage
Buffer:YSI
Single-point calibration:yes
Calibration interval:7 d
Unit for slope:mV/pH
[
i
] 2.00 4.00 7.00 10.00
The selection is made with the <><> keys. The current selection is displayed with a frame.
Submenus
The name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <ENTER>. Example:
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <ENTER>. Subsequently,
the setting can be changed with <><> and <ENTER>. Example:
Functions
Functions are designated by the name of the function. They are immediately
carried out by confirming with <ENTER>. Example: Display the Calibration record function.
22ba76143e0104/2013
MultiLab 4010-3Operation
Calibration record
Calibration data storage
Buffer:YSI
Single-point calibration:yes
Calibration interval:7 d
Unit for slope:mV/pH
[i] 2.00 4.00 7.00 10.00
pH
03.04.2013 08:00
03.04.2013 08:00
Info
System
Data storage
Storage & config
03.04.2013 08:00
Messages
Information is marked by the [
i
] symbol. It cannot be selected. Example:
4.5.4Example 1 on navigation: Setting the language
1.Press the <On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
2.Open the Storage & config menu with <ENTER
The instrument is in the setting mode.
_>.
ba76143e0104/201323
OperationMultiLab 4010-3
General
Interface
Clock function
Service information
Reset
System
03.04.2013 08:00
Language:English
Audio signal:on
Brightness:12
Temperature unit:°C
Stability control:on
General
03.04.2013 08:00
Language:English
Audio signal:on
Brightness:12
Temperature unit:°C
Stability control:on
General
03.04.2013 08:00
3.Select the System submenu with <><>.
The current selection is displayed with a frame.
4.Open the System submenu with <ENTER>.
5.Select the General submenu with <><>.
The current selection is displayed with a frame.
6.Open the General
submenu with <ENTER>.
7.Open the setting mode for the Language with <ENTER>.
24ba76143e0104/2013
8.Select the required language with <><>.
MultiLab 4010-3Operation
Date format:dd.mm.yy
Date:03.04.2013
Time:14:53:40
Clock function
03.04.2013 08:00
9.Confirm the setting with <ENTER>.
The meter switches to the measuring mode.
The selected language is active.
4.5.5Example 2 on navigation: Setting the date and time
The meter has a clock with a date function. The date and time are indicated in
the status line of the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Setting the date,
time and date format
Therefore, check the time at regular intervals.
The date and time are reset (to 01.01.2012 00:00 hours), if the following conditions are met:
the supply voltage failed and
the buffer batteries for the system clock are empty.
The date format can be switched from the display of day, month, year
(dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1.In the measured value display:
Open the Storage & config menu with <ENTER
_>.
The instrument is in the setting mode.
2.Select and confirm the System / Clock function menu with <><>
and <ENTER>.
The setting menu for the date and time opens up.
ba76143e0104/201325
3.Select and confirm the Time menu with <><> and <ENTER>.
The hours are highlighted.
OperationMultiLab 4010-3
4.Change and confirm the setting with <><> and <ENTER>.
The minutes are highlighted.
5.Change and confirm the setting with <><> and <ENTER>.
The seconds are highlighted.
6.Change and confirm the setting with <><> and <ENTER>.
The time is set.
7.If necessary, set the Date and Date format. The setting is made simi-
larly to that of the time.
8.To make further settings, switch to the next higher menu level with
<ESC>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring mode.
26ba76143e0104/2013
MultiLab 4010-3pH value
03.04.2013 08:00Info
5pH value
5.1Measuring
5.1.1Measuring the pH value
The sensor connection and the USB-B (device) interface are galvanically isolated. This facilitates interference-free measurements
also in the following cases:
Measurement in grounded test samples
Measurement with several sensors connected to one
MultiLab 4010-3 in one test sample
1.Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2.If necessary, select the measured parameter with <M>.
Stability control
(AutoRead)
& HOLD function
3.Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without a temperature sensor.
4.If necessary, calibrate or check the IDS pH sensor.
5.Immerse the IDS pH sensor in the test sample.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 9.5.3 A
TROL, page 72) in the System menu.
UTOMATIC STABILITYCON-
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
ba76143e0104/201327
pH valueMultiLab 4010-3
You can terminate the Stability control function and the HOLD function with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manually with <ENTER> at any time. When the Stability control function
is prematurely terminated, the current measurement data are output to the interface (PC, USB memory device or USB printer) without AutoRead info.
Criteria for a stable
measured value
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time intervalStability in the time interval
parameter
pH value15 seconds
Temperature15 seconds
Δ : Better than 0.01 pH
Δ : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
5.1.2Measuring the temperature
For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
When operating a sensor without integrated temperature sensor, e.g.
via an IDS pH adapter, you have to measure and enter the temperature of the
test sample first.
28ba76143e0104/2013
MultiLab 4010-3pH value
The display of the temperature indicates the active temperature measuring
mode:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
Yes0.1 °CAutomatic with temperature
sensor
-1 °CManual
5.2pH calibration
5.2.1Why calibrate?
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
5.2.2When do you have to calibrate?
Routinely within the framework of the company quality assurance
When the calibration interval has expired
5.2.3Carrying out automatic calibration (AutoCal)
Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see section 9.1.1 S
ETTINGSFORPH MEASUREMENTS, page 61).
Use one to five buffer solutions of the selected buffer set in any order.
Below, calibration with YSI buffers (YSI) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the procedure is identical.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH sensor to the meter.
The pH measuring window is displayed.
2.Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
ba76143e0104/201329
pH valueMultiLab 4010-3
03.04.2013 08:00
YSI
03.04.2013 08:00
YSI
10.000
4.Thoroughly rinse the sensor with deionized water.
5.Immerse the sensor in the first buffer solution.
6.For measurements without temperature sensor
(e.g. when using an IDS adapter
):
Enter the temperature of the buffer with <><>.
7.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
8.Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
9.If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
Continuing with two-
point calibration
10.Thoroughly rinse the sensor with deionized water.
11.Immerse the pH sensor in buffer solution 2.
30ba76143e0104/2013
MultiLab 4010-3pH value
03.04.2013 08:00
YSI
12.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
13.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
Continuing with
three- to five-point
calibration
14.Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
15.If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
16.Thoroughly rinse the sensor with deionized water.
17.Immerse the sensor in the next buffer solution.
18.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
19.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
ba76143e0104/201331
pH valueMultiLab 4010-3
03.04.2013 08:00
YSI
4.000
20.Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
21.If necessary, use <M> to finish the calibration.
The calibration record is displayed.
or
Switch to calibration with the next buffer with <ENTER>.
Calibration is automatically completed after the last buffer of a buffer set has been measured. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.4Carrying out a manual calibration (ConCal)
Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is
correctly selected (see section 9.1.1 S
ETTINGSFORPH MEASUREMENTS, page
61).
Use any one to five buffer solutions in ascending or descending order.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH sensor to the meter.
The pH measuring window is displayed.
2.Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
32ba76143e0104/2013
MultiLab 4010-3pH value
03.04.2013 08:00
03.04.2013 08:00
03.04.2013 08:00
4.Thoroughly rinse the sensor with deionized water.
5.Immerse the pH sensor in buffer solution 1.
6.For measurements without temperature sensor
(e.g. when using an IDS adapter
):
Enter the temperature of the buffer with <><>.
7.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
8.Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.
ba76143e0104/201333
pH valueMultiLab 4010-3
03.04.2013 08:00
9.Set the nominal buffer value for the measured temperature with
<><>.
10.Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
11.If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
Continuing with two-
point calibration
12.Thoroughly rinse the sensor with deionized water.
13.Immerse the pH sensor in buffer solution 2.
14.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
15.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
16.Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.
17.Set the nominal buffer value for the measured temperature with
<><>.
18.Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
19.If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
34ba76143e0104/2013
MultiLab 4010-3pH value
03.04.2013 08:00
Continuing with
three- to five-point
calibration
20.Thoroughly rinse the sensor with deionized water.
21.Immerse the sensor in the next buffer solution.
22.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
23.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
24.Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.
25.Set the nominal buffer value for the measured temperature with
<><>.
26.Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
27.If necessary, use <M> to finish the calibration.
The calibration record is displayed.
or
Continue calibrating using the next buffer with <ENTER>.
After the fifth buffer has been measured the calibration is automatically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.5Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
ba76143e0104/201335
pH valueMultiLab 4010-3
CalibrationDetermined val-
ues
1-pointAsy Zero point = Asy
2-pointAsy
Slp.
3-point to 5point
You can display the slope in the units, mV/pH or % (see section
9.1.1 S
5.2.6Calibration data
Asy
Slp.
ETTINGSFORPH MEASUREMENTS, page 61).
Displayed calibration data
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
Zero point = Asy
Slope = Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated by
linear regression.
Displaying the cali-
bration data
The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.
_> key.
36ba76143e0104/2013
MultiLab 4010-3pH value
Menu itemSetting/
function
Calibration /
Calibration data storage /Display
Calibration / Calibration data storage /
Output to USB flash
drive or printer
-Displays the calibration records.
-Outputs the stored calibration data to
Explanation
Further options:
Scroll through the calibration
records with <><>.
Using <PRT>, output the dis-
played calibration record to the
USB-B (PC) interface or the USBA (USB printer) interface.
Using <PRT
bration records to the USB-B (PC)
interface or the USB-A (USB
printer) interface.
Quit the display with <ESC>.
Switch directly to the measured
value display with <M>.
the USB-A interface (USB memory
device/USB printer)
_>, output all cali-
Calibration evalua-
tion
Calibration /
Calibration data storage /
Output to RS232/USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
DisplayCalibration
record
+++
++
+
-Outputs the stored calibration data to
the USB-B interface (PC)
Zero point
[mV]
-15 ... +15-60.5 ... -58.0
-20 ... <-15
or
>+15 ... +20
-25 ... <-20
or
>+20 ... +25
Slope [mV/pH]
>-58.0 ... -57.0
-61.0 ... <-60.5
or
>-57.0 ... -56.0
-
ba76143e0104/201337
-30 ... <-25
or
->+25 ... + 30
-62.0 ... <-61.0
or
>-56.0 ... -50.0
pH valueMultiLab 4010-3
CALIBRATION pH
03.04.2013 07:43:33
4110
Ser. no. B092500013
YSI
Buffer 14.00
Buffer 27.00
Buffer 310.00
Voltage 1184.0 mV
Voltage 23.0 mV
Voltage 3-177.0 mV
Temperature 124.0 °C
Temperature 224.0 °C
Temperature 324.0 °C
Slope-60.2 mV/pH
Asymmetry4.0 mV
Sensor+++
etc...
Calibration record
DisplayCalibration
record
Clean the IDS sensor according
to the sensor operating manual
ErrorError
Error elimination (see section 13
HATTODOIF..., page 89)
W
For pH IDS sensors you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 5.2.8 QSC
(
SENSORQUALITYCONTROL), page 40).
Zero point
[mV]
<-30
or
>+30
Slope [mV/pH]
<-62.0
or
> -50,0
FUNCTION
5.2.7Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is
displayed in the measured value display. Here you can very clearly see
38ba76143e0104/2013
whether or not the current measured value is in the calibrated part of the
measuring range.
MultiLab 4010-3pH value
1
2
3
4
03.04.2013 08:00
Info
The following information is displayed:
1 Currently measured pH value (needle)
2 Marking lines for all nominal buffer values used with the last valid cali-
bration
3 Measuring range for which a valid calibration is available. Measured
values in this range are suitable for documentation.
4 Measuring range for which no valid calibration is available (dark
gray). Measured values in this range are not suitable for documentation. Calibrate the meter with buffers covering this measuring range.
If the current measured value is outside the calibrated range, this area
is displayed in a darker gray.
If a measured value is outside the measuring range pH 0 - 14, overflow arrows are displayed at the left or right edge of the measuring
range.
The limits of the calibrated range are determined by the buffers used for calibration:
Lower limit: Buffer with lowest pH value - 2 pH units
Upper limit: Buffer with highest pH value + 2 pH units
ba76143e0104/201339
pH valueMultiLab 4010-3
03.04.2013 08:00
Info
1
5.2.8QSC function (sensor quality control)
General information
on the QSC function
The QSC function (Quality Sensor Control) is a new sensor evaluation for
digital IDS sensors. It evaluates the condition of an IDS pH sensor individually
and with a very fine grading.
The QSC scale shows the current sensor evaluation with an indicator on the
display.
1 QSC scale
The double arrow on the QSC scale indicates the current sensor evaluation
In the USB output the sensor evaluation is given as a percentage (1-100).
The finely graded sensor evaluation of the QSC function promptly calls your
attention to changes of the sensor.
Thus you can do what is necessary to restore the optimum measuring quality
(e.g. clean, calibrate or replace the sensor).
Sensor evaluation
with / without
QSC function
With QSC functionWithout QSC function (sensor sym-
bol)
Very fine grading of the sensor evaluation (100 grades)
The reference value is individually
determined for each sensor during
Rough grading of the sensor evaluation (4 grades)
A theoretical reference value is used
for all sensors
the QSC initial calibration.
Low tolerances for zero point and
slope when using QSC buffer solutions
Additional QSC calibration required
Greater tolerances for zero point and
slope when using commercial buffer
sets
No additional calibration required
(with special QSC buffer set)
QSC calibrationThe QSC function is enabled by once carrying out an additional three-point cali-
bration with special QSC buffer solutions. It covers the measuring range of the
sensor (pH 2 to pH 11). The QSC initial calibration determines the actual condition of the sensor and stores it as a reference in the sensor.
To meet the high requirements of a QSC initial calibration, the QSC initial calibration should optimally be carried out with the initial commissioning of the
sensor.
40ba76143e0104/2013
MultiLab 4010-3pH value
03.04.2013 08:00
Carry out the normal calibrations for your special measuring range with your
usual standard solutions as previously done.
As soon as the QSC function was enabled for an IDS sensor, it is
not possible to return to the sensor evaluation with the sensor
symbol for this sensor.
Carrying out a QSC
initial calibration
1.Open the menu for measurement settings with <ENTER>.
2.In the QSC menu, select First calibration with <><>.
The calibration display appears. AutoCal QSC-Kit is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC calibration.
3.Calibration with the buffers of the QSC-Kit is done like a normal threepoint calibration.
Follow the user guide.
Carry out the QSC initial calibration very carefully. It determines the
reference value for the sensor. This reference value cannot be
overwritten or reset.
As soon as the QSC function was enabled, it is not possible to
return to the sensor evaluation with the sensor symbol.
4.As soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
initial calibration.
The QSC initial calibration is completed. The sensor is calibrated. If you want
to calibrate with special buffers for your measurements, you can subsequently
carry out a normal calibration with your buffers. The reference values determined with the QSC calibration are also used for the evaluation of normal calibrations. In the measured value display, the QSC scale of the QSC function is
always displayed. A double arrow on the QSC scale indicates the current sen-
ba76143e0104/201341
pH valueMultiLab 4010-3
03.04.2013 08:00
Info
1
sor evaluation.
1 QSC scale
The double arrow on the QSC scale indicates the current sensor evaluation
Carrying out a QSC
control calibration
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).
You can carry out QSC control calibrations at greater intervals than normal calibrations.
1.Open the menu for measurement settings with <ENTER>.
2.In the QSC menu, select Control calibration with <><>.
The calibration display appears. AutoCal QSC-Kit is displayed as the buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other buffers, you will have no valid QSC control calibration.
3.Follow the user guide.
The calibration is carried out like a normal three-point calibration. As soon
as the three-point calibration has been successfully carried out you can
decide whether to accept or discard the calibration as the QSC control
calibration.
42ba76143e0104/2013
MultiLab 4010-3ORP voltage
03.04.2013 08:00
Info
6ORP voltage
6.1Measuring
6.1.1Measuring the ORP
The sensor connection and the USB-B (device) interface are galvanically isolated. This facilitates interference-free measurements
also in the following cases:
Measurement in grounded test samples
Measurement with several sensors connected to one
MultiLab 4010-3 in one test sample
IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.
Stability control
(AutoRead)
& HOLD function
1.Connect the ORP sensor to the meter.
The ORP measuring window is displayed.
2.Adjust the temperature of the solutions and measure the current temperature if the measurement is made without a temperature sensor.
3.Check the meter with the ORP sensor.
4.Immerse the ORP sensor in the test sample.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the
setting for automatic Stability control (see section 9.5.3 A
CONTROL, page 72) in the System menu.
ba76143e0104/201343
UTOMATIC STABILITY
ORP voltageMultiLab 4010-3
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD function with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Criteria for a stable
measured value
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. When the Stability control function
is prematurely terminated, the current measurement data are output to the interface (PC, USB memory device or USB printer) without AutoRead info.
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
parameter
ORP15 seconds
Temperature15 seconds
Time intervalStability in the time
interval
Δ : Better than 0.3 mV
Δ : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
6.1.2Measuring the temperature
For reproducible ORP measurements, it is essential to measure the temperature of the test sample.
When operating a sensor without integrated temperature sensor, you first have
44ba76143e0104/2013
MultiLab 4010-3ORP voltage
to measure and enter the temperature of the sample.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperature
sensor
Resolution of
the temp. dis-
Temp. measurement
play
Yes0.1 °CAutomatic with temperature sensor
-1 °CManual
6.2ORP calibration
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
ba76143e0104/201345
Dissolved oxygen (D.O.)MultiLab 4010-3
03.04.2013 08:00Info
7Dissolved oxygen (D.O.)
7.1Measuring
7.1.1Measuring D.O.
1.Connect the IDS D.O. sensor to the meter.
The D.O. measuring screen is displayed.
2.If necessary, select the measured parameter with <M>.
3.Check or calibrate the meter with the sensor.
4.Immerse the D.O. sensor in the test sample.
Selecting the
displayed
You can switch between the following displays with <M>:
D.O. concentration [mg/l]
measured parameter
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correctionWhen measuring the D.O. concentration of solutions with a salt content of more
than 1 psu, a salinity correction is required. For this, you have to measure and
input the salinity of the measured medium first.
When the salinity correction is switched on, the [Sal] indicator is displayed in
the measuring screen.
You can switch the salinity correction on or off and enter the salinity
in the menu for calibration and measurement settings (see section
Air pressure
correction
(Saturation local
function)
9.3.1 S
The integrated air pressure sensor of the MultiLab 4010-3 measures the
current air pressure.During calibration, the air pressure correction function is
automatically activated. For measurement, the air pressure correction is
applied if the parameter oxygen saturation [%] is displayed and the
ETTINGSFOR D.O. MEASUREMENT, page 66).
Saturation local function is enabled.
46ba76143e0104/2013
MultiLab 4010-3Dissolved oxygen (D.O.)
You can view the current air pressure in the sensor menu when an IDS D.O.
sensor is connected. Press the <ENTER> key in the measured value display.
The current air pressure is displayed as an info message.
The Saturation local function for the parameter oxygen saturation
[%] is switched on or off in the menu for calibration and measurement settings (see section 9.3.3 S
ATURATION LOCAL, page 68).
Stability control
(AutoRead)
& HOLD function
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
Irrespective of the setting for automatic Stability control (see section 9.5.3
A
UTOMATIC STABILITYCONTROL, page 72) in the System menu, you can start a
measurement with Stability control manually at any time.
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD function with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. When the Stability control function
is prematurely terminated, the current measurement data are output to the interface (PC, USB memory device or USB printer) without AutoRead info.
ba76143e0104/201347
Dissolved oxygen (D.O.)MultiLab 4010-3
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured parameterTime intervalStability in the time
interval
D.O. concentration20 seconds
D.O. saturation20 seconds
D.O. partial pressure20 seconds
Temperature15 seconds
Δ : better than 0.03 mg/l
Δ : better than 0.4 %
Δ : Better than 0.8 mbar
Δ : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
7.1.2Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the temperature of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor integrated in the IDS sensor.
7.2Calibration
7.2.1Why calibrate?
D.O. sensors age. This changes the zero point (asymmetry) and slope of the
D.O. sensor. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope.
7.2.2When to calibrate?
When the calibration interval has expired
When your accuracy requirements are especially high
Routinely within the framework of the company quality assurance
7.2.3Calibration procedures
The MultiLab 4010-3 provides 2 calibration procedures:
48ba76143e0104/2013
MultiLab 4010-3Dissolved oxygen (D.O.)
Calibration in water vapor-saturated air.
Calibration via a comparison measurement (e.g. Winkler titration according
to DIN EN 25813 or ISO 5813). At the same time, the relative slope is
adapted to the comparison measurement by a correction factor. When the
correction factor is active, the [Factor] indicator appears in the measuring
window.
Calibration can be carried out either as a 1-point calibration or 2-point calibration procedure.
7.2.41-point calibration
Calibration in water vapor-saturated air (Cal) and calibration via Comparison
meas. according to the Winkler method (Comp) are described jointly here.
1.Connect the D.O. sensor to the meter.
2.Check the selected calibration type and selected number of calibration
points in the sensor setting menu and, if necessary, change them.
3.Put the IDS D.O. sensor into the calibration medium.
When calibrating via comparison measurement
(Comp):
Put the IDS D.O. sensor into the reference solution.
When calibrating in water vapor-saturated air (
Cal):
Put the IDS D.O. sensor into the calibration vessel with water vaporsaturated air.
As the calibration vessel use a BOD bottle that contains a
small amount of clean water (approx. 40 ml).
The sensor must not be immersed in the water.
Leave the sensor in the calibration bottle long enough (5 to
10 minutes) so that the air is water vapor-saturated and the
sensor can adapt to the ambient temperature.
4.Start the calibration with <CAL>.
Cal (calibration in water vapor-saturated air) or Comp (calibration via
Comparison meas. according to the Winkler method) is displayed.
The calibration point (DO Sat) is displayed.
ba76143e0104/201349
Dissolved oxygen (D.O.)MultiLab 4010-3
03.04.2013 08:00
03.04.2013 08:00
5.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
6.Wait for the AutoRead measurement to be completed (audio signal).
When calibrating in water vapor-saturated air (
Cal):
The calibration record is displayed and output to the interface.
When calibrating via comparison measurement
(Comp):
Set the concentration value with <><> and accept it with <ENTER>.
The calibration record is displayed and output to the interface.
7.Switch to the measured value display with <ENTER>.
7.2.52-point calibration
Calibration in water vapor-saturated air (Cal) and calibration via Comparison
meas. according to the Winkler method (Comp) are described jointly here.
1.Connect the D.O. sensor to the meter.
2.Check the selected calibration type and selected number of calibration
points in the sensor setting menu and, if necessary, change them.
50ba76143e0104/2013
MultiLab 4010-3Dissolved oxygen (D.O.)
03.04.2013 08:00
03.04.2013 08:00
3.Place the D.O. sensor in a solution that does not contain any dissolved
oxygen.
A solution not containing any dissolved oxygen can be prepared by
dissolving approx. 8 to 10 g sodium sulfite (Na
) in 500 ml tap-
2SO3
water. Carefully mix the solution. It may take up to 60 minutes until
the solution is free of oxygen.
4.Start the calibration with <CAL>.
Cal (calibration in water vapor-saturated air) or Comp (calibration via
Comparison meas. according to the Winkler method) is displayed.
The calibration point 1 (0%) is displayed.
5.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
6.Wait for the AutoRead measurement to be completed (audio signal).
The display for the second calibration point is displayed.
ba76143e0104/201351
Dissolved oxygen (D.O.)MultiLab 4010-3
03.04.2013 08:00
7.Put the IDS D.O. sensor into the calibration medium.
When calibrating via comparison measurement
(Comp):
Put the IDS D.O. sensor into the reference solution..
When calibrating in water vapor-saturated air (
Cal):
Put the IDS D.O. sensor into the calibration vessel with water vaporsaturated air.
As the calibration vessel use a BOD bottle that contains a
small amount of clean water (approx. 40 ml).
The sensor must not be immersed in the water.
Leave the sensor in the calibration bottle long enough (5
to 10 minutes) so that the air is water vapor-saturated and
the sensor can adapt to the ambient temperature.
8.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
9.Wait for the AutoRead measurement to be completed (audio signal).
When calibrating in water vapor-saturated air (
Cal):
The calibration record is displayed and output to the interface.
When calibrating via comparison measurement
(Comp):
Set the concentration value with <><> and accept it with <ENTER>.
The calibration record is displayed and output to the interface.
10.Switch to the measured value display with <ENTER>.
52ba76143e0104/2013
MultiLab 4010-3Dissolved oxygen (D.O.)
7.2.6Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.
Menu itemSetting/func-
Calibration /
Calibration data
storage /Display
_> key.
Explanation
tion
-Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Using <PRT>, output the dis-
played calibration record to
the USB-B (PC) interface or
the USB-A (USB printer)
interface .
Using <PRT
calibration records to the
USB-B (PC) interface or the
USB-A (USB printer) interface .
_>, output all
Calibration
evaluation
Quit the display with <ESC>.
Switch directly to the mea-
sured value display with <M>.
Calibration / Calibration data storage / Output to
USB flash drive or
printer
Calibration /
Calibration data
storage / Output to
RS232/USB
After calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.
-Outputs the stored calibration
data to the USB-A interface
(USB memory device/USB
printer)
-Outputs the stored calibration
data to the USB-B interface (PC)
ba76143e0104/201353
Dissolved oxygen (D.O.)MultiLab 4010-3
CALIBRATION Ox
03.04.2013 07:43:33
ProOBOD
Ser. no. 12B100016
ProOBOD Cap12B100015
Anzahl der Kalibrierpunkte1
Sensor+++
ProOBOD
calibration
evaluation
Calibration record
DisplayCalibration recordRelative slope
+++S = 0.94 ... 1.06
++S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
+S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
ErrorErrorS < 0.90
or
Error elimination (see section 13 W
HATTODOIF...,
S > 1.10
page 89)
54ba76143e0104/2013
MultiLab 4010-3Conductivity
03.04.2013 08:00Info
8Conductivity
8.1Measuring
8.1.1Measuring the conductivity
The sensor connection and the USB-B (device) interface are galvanically isolated. This facilitates interference-free measurements
also in the following cases:
Measurement in grounded test samples
Measurement with several sensors connected to one
MultiLab 4010-3 in one test sample
1.Connect the conductivity sensor to the meter.
The conductivity measuring window is displayed.
The measuring cell and cell constant for the connected IDS conductivity
sensor are automatically taken over.
Selecting the
displayed
measured parameter
2.If necessary, press <M> to select the measured parameter ϰ.
3.Immerse the conductivity sensor in the test sample.
You can switch between the following displays with <M>:
Conductivity [μS/cm] / [mS/cm]
Resistivity [Ω·cm] / [kΩ·cm] / [MΩ·cm]
Salinity SaL [ ] (
Total dissolved solids TDS [mg/l] / [g/l]
≙ psu)
The factor to calculate the total dissolved solids is set to 1.00 in the factory. You
can adjust this factor to meet your requirements in the range 0.40 ... 1.00. The
factor is set in the menu for the parameter TDS.
Stability control
(AutoRead)
& HOLD function
ba76143e0104/201355
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
ConductivityMultiLab 4010-3
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 9.5.3 A
TROL, page 72) in the System menu.
UTOMATIC STABILITYCON-
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD function with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Criteria for a stable
measured value
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. When the Stability control function
is prematurely terminated, the current measurement data are output to the interface (PC, USB memory device or USB printer) without AutoRead info.
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time intervalStability in the time interval
parameter
Conductivity ϰ10 seconds
Δ ϰ: better than
1.0% of measured value
Temperature15 seconds
Δ : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
56ba76143e0104/2013
MultiLab 4010-3Conductivity
8.1.2Measuring the temperature
For reproducible conductivity measurements, it is essential to measure the
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
8.2Temperature compensation
The calculation of the temperature compensation is based on the preset reference temperature, 20 °C or 25 °C. It appears on the display as Tr20 or Tr25.
You can select one of the following temperature compensation methods:
Nonlinear temperature compensation (nLF) according to ISO 7888
Linear temperature compensation (lin) with adjustable coefficients of
0.000 ... 3.000 %
No temperature compensation (off)
/K
The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 9.4.1 S
TINGSFOR IDS CONDUCTIVITYSENSORS, page 68).
ET-
Application tipsSelect the following temperature compensations given in the table according to
the respective test sample:
Test sampleTemperature compensationDisplay
Natural water
(ground water, surface
nLF
according to ISO 7888
nLF
water, drinking water)
Ultrapure waternLF
nLF
according to ISO 7888
Other aqueous solutions
lin Set linear
temperature coefficient
lin
0.000 ... 10.000 %/K
Salinity (seawater)Automatic nLF according to
Sal, nLF
IOT (International Oceanographic Tables)
ba76143e0104/201357
ConductivityMultiLab 4010-3
03.04.2013 08:00
8.3Calibration
8.3.1Why calibrate?
Aging slightly changes the cell constant, e. g. due to coatings. As a result, an
inexact measured value is displayed. The original characteristics of the cell can
often be restored by cleaning the cell. Calibration determines the current value
of the cell constant and stores this value in the meter.
Thus, you should calibrate at regular intervals.
8.3.2When to calibrate?
After connecting a sensor
Routinely within the framework of the company quality assurance
When the cleaning interval has expired
8.3.3Determining the cell constant (calibration in control standard)
You can determine the actual cell constant of the IDS conductivity sensor by
calibrating with the control standard in the following range:
-1
0.450 ... 0.500 cm
(e.g. 4310, nominal cell constant 0.475cm
-1
)
The cell constant is determined in the control standard, 0.01 mol/l KCl.
In the delivery condition, the calibrated cell constant of the IDS sensor is set to
-1
0.475 cm
(4310 IDS conductivity sensor).
For this calibration procedure, the Type setting must be set to cal. Proceed as
follows to determine the cell constant:
1.Connect the conductivity sensor to the meter.
2.In the measured value display, select the conductivity parameter with
<M>.
3.Start the calibration with <CAL>.
The cell constant that was calibrated last is displayed.
58ba76143e0104/2013
MultiLab 4010-3Conductivity
4.Immerse the conductivity sensor in the control standard solution,
0.01 mol/l KCI.
5.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
6.Wait for the end of the measurement with stability control
or
take over the calibrated value with <ENTER>.
The calibration record is displayed and output to the interface.
7.Switch to the measured value display with <ENTER>.
8.3.4Calibration data
Displaying the cali-
bration record
The calibration record is automatically transmitted to the interface
after calibrating.
The calibration data can be displayed and then output to the interface.
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.
Menu itemSetting/
Calibration /
Calibration data
storage /Display
_> key.
Explanation
function
-Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Using <PRT>, output the dis-
played calibration record to the
USB-B (PC) interface or the
USB-A (USB printer) interface .
Using <PRT
bration records to the USB-B
(PC) interface or the USB-A
(USB printer) interface .
Quit the display with <ESC>.
Switch directly to the measured
value display with <M>.
ba76143e0104/201359
_>, output all cali-
ConductivityMultiLab 4010-3
CALIBRATION Cond
03.04.2013 07:43:33
4310
Ser. no. 09250033
Cell constant 0.476 1/cm
25.0 °C
Sensor +++
Calibration
evaluation
Menu itemSetting/
Explanation
function
Calibration / Calibration data storage / Output to
-Outputs the stored calibration data
to the USB-A interface (USB memory device/USB printer)
USB flash drive or
printer
Calibration /
Calibration data
-Outputs the stored calibration data
to the USB-B interface (PC)
storage / Output to
RS232/USB
After calibration, the meter automatically evaluates the current status of the calibration. The evaluation appears on the display and in the calibration record.
DisplayCalibration recordCell constant [cm-1]
+++Within the range
0.450 ... 0.500 cm
-1
Calibration record
ErrorErrorOutside the range
Error elimination (see section 13
W
HATTODOIF..., page 89)
0.450 ... 0.500 cm
-1
60ba76143e0104/2013
MultiLab 4010-3Settings
9Settings
9.1pH measurement settings
9.1.1Settings for pH measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required measured parameter in the measured value display and press the <ENTER> key.
After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.
Menu itemPossible
setting
Calibration /Calibration
-Displays the calibration record of the last calibration.
record
Calibration / Calibration
-Displays the last calibration records (max. 10)
data storage / Display
Calibration /
-Outputs the stored calibration data to the USB-B interface
Calibration data storage
/ Output to RS232/USB
Calibration / Calibration
-Outputs the stored calibration data to the USB-A interface
data storage / Output to
USB flash drive or
printer
Calibration /BufferYSI
NIST/DIN
ConCal
...
Calibration / Single-point
calibration
yes
no
Explanation
(PC)
(USB memory device/USB printer)
Buffer sets to be used for pH calibration.
More buffers and details: see section 9.1.2 B
FORCALIBRATION, page 62 and section 5.2 PH CALIBRATION,
UFFERSETS
page 29.
Quick calibration with 1 buffer
Calibration /Calibration
interval
1 ... 7 ... 999 dCalibration interval for the IDS pH sensor (in days).
The meter reminds you to calibrate regularly by the flashing sensor symbol in the measuring screen.
Calibration /Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).
QSC /First calibration-Starts the initial calibration with QSC buffers.
This menu item is only available as long as no initial calibration was carried out with the connected IDS sensor.
QSC /Record of first calibration
QSC /Control calibration
-Displays the calibration record of the QSC initial calibra-
tion.
-Starts the control calibration with QSC buffers.
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.
ba76143e0104/201361
SettingsMultiLab 4010-3
Menu itemPossible
setting
Alternative temperatureon
off
Temperature from
channel
▭▭▬
▭▬▭
▬▭▭
Man. temperature-25 ... +25 ...
+130 °C
pH resolution0.001
0.01
0.1
mV resolution0.1
1
Reset-Resets all sensor settings to the delivery condition (see
Explanation
Takes the measured temperature value from an IDS sensor.
This menu item is only available if an IDS adapter and an
IDS sensor with temperature sensor are connected.
Selection of the channel from which the temperature value
should be taken over.
This menu item is only available if two IDS sensors are
additionally connected.
Entry of the manually determined temperature.
This menu item is only available if an IDS adapter is connected.
Resolution of the pH display
Resolution of the mV display
section 9.6.1 R
page 73).
ESETTINGTHEMEASUREMENTSETTINGS,
9.1.2Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
No.Buffer set *pH valuesat
1ConCalAnyAny
2NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
3TEC
Technical buffers
4Merck 1*4.000
1.679
4.006
6.865
9.180
12.454
2.000
4.010
7.000
10.011
7.000
9.000
25 °C
25 °C
20 °C
62ba76143e0104/2013
MultiLab 4010-3Settings
No.Buffer set *pH valuesat
5Merck 2 *1.000
6.000
8.000
13.000
6Merck 3 *4.660
6.880
9.220
7Merck 4 *2.000
4.000
7.000
10.000
8Merck 5 *4.010
7.000
10.000
9DIN 192671.090
4.650
6.790
9.230
10Mettler Toledo USA *1.679
4.003
7.002
10.013
20 °C
20 °C
20 °C
25 °C
25 °C
25 °C
11Mettler Toledo EU *1.995
4.005
7.002
9.208
12Fisher *2.007
4.002
7.004
10.002
13Fluka BS *4.006
6.984
8.957
14Radiometer *1.678
4.005
7.000
9.180
15Baker *4.006
6.991
10.008
16Metrohm *3.996
7.003
8.999
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
ba76143e0104/201363
SettingsMultiLab 4010-3
No.Buffer set *pH valuesat
17Beckman *4.005
7.005
10.013
18Hamilton Duracal *4.005
7.002
10.013
19Precisa *3.996
7.003
8.999
20Reagecon TEC *2.000
4.010
7.000
10.000
21Reagecon 20 *2.000
4.000
7.000
10.000
13.000
22Reagecon 25 *2.000
4.000
7.000
10.000
13.000
25 °C
25 °C
25 °C
25 °C
20 °C
25 °C
23Chemsolute *2.000
20 °C
4.000
7.000
10.000
24USABlueBook
*4.000
25 °C
7.000
10.000
25YSI *4.000
25 °C
7.000
10.000
* Brand names or trade names are trademarks of their respective owners protected
by law.
The buffers are selected in the menu, pH / <ENTER> / Calibration
/ Buffer (see section 9.1.1 S
ETTINGSFORPH MEASUREMENTS, page
61).
64ba76143e0104/2013
MultiLab 4010-3Settings
9.1.3Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC
FUNCTION (SENSORQUALITYCONTROL), page
40).
After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
Setting the calibra-
tion interval
SettingsThe settings are made in the menu for measuring settings of the ORP measure-
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
1.Open the menu for measurement settings with <ENTER>.
2.In the Calibration / Calibration interval menu, set the calibration interval
with <><>.
3.Confirm the setting with <ENTER>.
4.Quit the menu with <M>.
9.2ORP measurement settings
ment. To open the settings, display the required measured parameter in the
measured value display and press the <ENTER> key. After completing the set-
tings, switch to the measured value display with <M>.
Default settings are printed in bold.
Menu itemPossible settingExplanation
mV resolution0.1
Resolution of the mV display
1
Reset-Resets all sensor settings to the
delivery condition (see section
9.6.1 R
SETTINGS, page 73).
ba76143e0104/201365
ESETTINGTHEMEASUREMENT
SettingsMultiLab 4010-3
9.3D.O. measurement settings
9.3.1Settings for D.O. measurement
The settings are available in the menu for measurement and calibration settings. To open the settings, display the required measured parameter in the
measured value display and press the <ENTER> key. After completing the set-
tings, switch to the measured value display with <M>.
Menu itemPossible
setting
Calibration /Calibra-
-Displays the calibration record of the
tion record
Calibration / Cali-
-Displays the last calibration records
bration data storage
/ Display
Calibration / Cali-
-Outputs the stored calibration data to
bration data storage
/ Output to USB
flash drive or printer
Calibration / Cali-
-Outputs the stored calibration data to
bration data storage
/ Output to RS232/
USB
Calibration /
Calibration data
1
2
storage / Number of
cal. points
Explanation
last calibration.
(max. 10)
the USB-A interface (USB memory
device/USB printer)
the USB-B interface (PC)
1-point calibration
or
2-point calibration
Calibration /Calibration interval
Calibration /Comparison meas.
1 ... 180 ...
999 d
on
off
Sal automaticon
off
Calibration interval for the D.O. sensor
(in days).
The meter reminds you to calibrate
regularly by the flashing sensor symbol
in the measuring screen.
Enables to adjust the measured value
with the aid of a comparison measurement, e.g. Winkler titration.
For details, see section 7.2 C
TION, page 48.
ALIBRA-
Automatic salt content correction for
concentration measurements. The
measured salinity value is taken over
by a connected conductivity sensor.
This menu item is only available if an
IDS conductivity sensor is connected.
66ba76143e0104/2013
MultiLab 4010-3Settings
Menu itemPossible
Explanation
setting
Salinity from channel
▭▭▬
▭▬▭
▬▭▭
Selects the channel (sensor), that provides the measured temperature
value.
This menu item is only available if an
IDS adapter and two IDS sensors with
temperature sensor are connected.
Sal correctionon
off
Calibration /
Sensor Cap coeffici
ents
K1
...
K5
KC
Manual salt content correction for concentration measurements.
Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2 E
S
ENSOR CAP COEFFICIENTS, page 67
NTER
Salinity0.0 ... 70.0Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
automatic salt content correction is
switched off and the manual salt content correction is switched on.
Resolution0.1
1
Set a high or low resolution.
The setting of the resolution is separately stored for each measured
parameter.
Saturation localon
off
Saturation local is a procedure that
takes the local air pressure into
account for each saturation measurement.
For details, see section 9.3.3
S
ATURATION LOCAL, page 68.
Reset-Resets all sensor settings to the deliv-
ery condition (see section 9.6.1 R
TINGTHEMEASUREMENTSETTINGS,
ESET-
page 73).
9.3.2Enter Sensor Cap coefficients
The values of the coefficients are provided with the sensor cap.
1.Change the digit of the highlighted position with <><>.
2.Go to the next position with <><>.
3.Confirm with<ENTER> when a coefficient is completely entered.
ba76143e0104/201367
SettingsMultiLab 4010-3
9.3.3Saturation local
Irrespective of the height or air pressure, the calibration value is set to 100%.
Saturation local is ideal for EU compliance.
When Saturation local is enabled, an L will appear next to D.O.% on the
display.
D.O. mg/L readings are unaffected by the selection of Saturation local.
9.4Cond measurement settings
9.4.1Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the measured value display and press the <ENTER> key. After completing the settings,
switch to the measured value display with <M>.
The possible settings are individually displayed for each sensor. Below the setting menu is displayed for two IDS sensors (4310, 4320).
Default settings are printed in bold.
Setting menu of
4310
Menu itemPossible
setting
Calibration /Calibra-
-Displays the calibration record of the
tion record
Calibration / Calibra-
-Displays the last calibration records
tion data storage /
Display
Calibration / Calibra-
-Outputs the stored calibration data to
tion data storage /
Output to USB flash
drive or printer
Calibration /
-Outputs the stored calibration data to
Calibration data storage / Output to
RS232/USB
Calibration /Calibration interval
1 ... 150 ...
999 d
Explanation
last calibration.
(max. 10)
the USB-A interface (USB memory
device/USB printer)
the USB-B interface (PC)
Calibration interval for the IDS conductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor symbol in the measuring screen.
68ba76143e0104/2013
MultiLab 4010-3Settings
Menu itemPossible
Explanation
setting
TypeUsed measuring cell
CalMeasuring cells whose cell constant
is determined by calibration in the KCl
control standard solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is displayed in the status line.
manFreely adjustable cell constant in the
range 0.450 ... 0.500 cm
Man. cell const.0.450 ...
0.475 ...
0.500 cm
Temp. comp. (TC) /
Method
nLF
lin
off
Display and setting options for the cell
constant.
-1
Procedure for temperature compensation (see section 8.2 T
COMPENSATION, page 57).
This setting is only available for the
measured parameters, ϰ and ρ.
-1
.
EMPERATURE
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available when
the linear temperature compensation
is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters, ϰ and ρ.
Multiplier for TDS0.40 ... 1.00Factor for TDS value
Reset-Resets all sensor settings to the deliv-
ery condition (see section 9.6.1
R
ESETTINGTHEMEASUREMENTSET-
TINGS, page 73).
ba76143e0104/201369
SettingsMultiLab 4010-3
Setting menu of
4320
Menu itemPossible
Explanation
setting
Cell constant0.090 ... ...
0.100 ...
-1
Temp. comp. (TC) /
Method
0.110 cm
nLF
lin
off
Display and setting options for the
cell constant
Procedure for temperature compensation (see section 8.2 T
TURECOMPENSATION, page 57).
This setting is only available for the
measured parameters, ϰ and ρ.
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available
when the linear temperature compensation is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters, ϰ and ρ.
Multiplier for TDS0.40 ... 1.00Factor for TDS value
EMPERA-
Reset-Resets all sensor settings to the
delivery condition (see section 9.6.1
R
ESETTINGTHEMEASUREMENTSET-
TINGS, page 73).
70ba76143e0104/2013
MultiLab 4010-3Settings
9.5Sensor-independent settings
9.5.1System
To open the Storage & config menu, press the <ENTER
_> key in the mea-
sured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
Menu itemSettingExplanation
System / General / LanguageDeutsch
Selects the menu language
English
(more)
System / General / Audio signalon
off
Switches on/off the beep on keystroke
System / General / Brightness0 ... 10 ... 22 Changes the display brightness
System / General / Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed
with the selected unit.
System / General / Stability
control
on
off
Switches on or off the automatic stability control during measurement
(see section 9.5.3 A
ITYCONTROL, page 72 )
UTOMATIC STABIL-
System / Interface / Baud rate1200, 2400, 4800,
9600, 19200
System / Interface / Output format
System / Interface / Decimal
separator
ASCII
CSV
Dot (xx.x)
Comma (xx,x)
System / Interface / Output
header
System /Clock functionDate format
Datum
Time
Baud rate of the USB Device interface
Output format for data transmission
For details, see section 11 T
TINGDATA, page 82
RANSMIT-
Decimal separator
Output of a header for Output format:
CSV
Settings of time and date.
For details, see section 4.5.5
XAMPLE 2 ONNAVIGATION: SETTING
E
THEDATEANDTIME, page 25
System /Service informationHardware version and software ver-
sion of the meter are displayed.
System /Reset-Resets the system settings to the
default values.
For details, see section 9.6.2 R
TINGTHESYSTEMSETTINGS, page 75
ESET-
ba76143e0104/201371
SettingsMultiLab 4010-3
9.5.2Data storage
This menu contains all functions to display, edit and erase stored measured
values.
Detailed information on the memory functions of the
MultiLab 4010-3 are given in section 10 D
ATASTORAGE, page 76.
9.5.3Automatic Stability control
The automatic Stability control (AutoRead) function continuously checks the
stability of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
You can activate or switch off the automatic Stability control function (see
section 9.5 S
ENSOR-INDEPENDENTSETTINGS, page 71).
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when you switch over between the measured parameters with <M>.
when the automatic Stability control is switched off.
72ba76143e0104/2013
MultiLab 4010-3Settings
9.6Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1 Resetting the measurement settings
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
pHThe following settings for pH measurements are reset to the default settings
with the Reset function:
SettingDefault settings
Buffer YSI
Calibration interval7 d
Unit for slopemV/pH
Measured parameterpH
pH resolution0.001
mV resolution0.1
Asymmetry0 mV
Slope-59.2 mV
Man. temperature25 °C
Single-point calibrationoff
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
ORPThe following settings for ORP measurements are reset to the default settings
with the Reset function:
SettingDefault settings
mV resolution0.1
Man. temperature25 °C
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
ba76143e0104/201373
SettingsMultiLab 4010-3
Dissolved oxygenThe following settings for D.O. measurements are reset to the default settings
with the Reset function:
SettingDefault settings
Calibration interval180 d
Check interval60 d
Measured parameterD.O. concentration
Relative slope (S
)1.00
Rel
Salinity (value)0.0
Salinity (function)Off
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
ConductivityThe following settings for conductivity measurements are reset to the default
settings with the Reset function:
SettingDefault settings
Calibration interval150 d
Measured parameter
ϰ
Cell constant (c)Depending on the connected mea-
suring cell:
-1
0.475 cm
0.475 cm
0.100 cm
(calibrated)
-1
(set)
-1
Temperature compensationnLF
Reference temperature25 °C
Temperature coefficient (TC) of the
2.000 %/K
linear temperature compensation
Multiplier for TDS1.00
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
74ba76143e0104/2013
MultiLab 4010-3Settings
9.6.2Resetting the system settings
The following system settings can be reset to the default condition:
SettingDefault settings
LanguageEnglish
Audio signalon
Baud rate4800 Baud
Output formatASCII
Decimal separatorDot (xx.x)
Brightness10
Temperature unit°C
Stability controlon
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu, press the <ENTER
measured value display.
_> key in the
ba76143e0104/201375
Data storageMultiLab 4010-3
03.04.2013 11:24:16
pH 7.000 24.8 °C AR +++
ID number:1
continue
Manual data storage 4 von 500
03.04.2013 08:00
10Data storage
You can transmit measured values (datasets) to the data storage:
Manual data storage (see section 10.1 M
Automatic data storage at intervals (see section 10.2 A
RAGEATINTERVALS, page 76)
ANUALDATASTORAGE, page 76)
UTOMATICDATASTO-
10.1Manual data storage
You can transmit a measurement dataset to the data storage as follows. The
dataset is at the same time output to the interface:
1.Press the <STO> key shortly
The menu for manual data storage appears.
.
If the storage is fullWhen all storage locations are occupied, it is not possible to continue storing.
Configuring the
automatic storage
function
2.If necessary, change and confirm the ID number (1 ... 10000) with
<><> and <ENTER>.
The dataset is stored. The meter switches to the measured value display.
Then you can e.g. transmit the stored data to a PC or a USB flash drive (see
section 10.3.1 E
quently erase the storage (see section 10.3.2 E
STORAGE, page 80).
DITINGTHEMEASUREMENTDATASTORAGE, page 79) and subse-
RASINGTHEMEASUREMENTDATA
10.2Automatic data storage at intervals
The storage interval (Interval) determines the time interval between automatic
data storage processes. Each data storage process transmits the current dataset to the interface at the same time.
1.Press the <STO_> key.
The menu for automatic data storage appears.
76ba76143e0104/2013
MultiLab 4010-3Data storage
ID number 1
Interval 30 s
Duration 180 min
continue
0d03h00min
01d17h33min
Automatic data storage
03.04.2013 08:00
1
2
3
1 Adjusted entire storage duration
2 Max. available storage duration
3 Graphical display of the storage usage
SettingsYou can configure the automatic data storage function with the following
settings:
Menu itemPossible settingExplanation
ID number1 ... 10000ID number for the dataset series.
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the storage duration.
Duration1 min ... x minStorage duration.
Specifies after which time the automatic data storage should be terminated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
ba76143e0104/201377
Data storageMultiLab 4010-3
03.04.2013 08:00
0d03h00min
1
03.04.2013 08:00
0d03h00min
2
Stop automatic storage?
yes
no
Warning
03.04.2013 08:00
Starting the
automatic storage
function
To start the automatic storage function, select continue with <><> and con-
firm with <ENTER>. The meter switches to the measured value display.
1 Remaining storage duration
2 Graphical display of the storage duration
The active automatic data storage function can be recognized by the progress
bar in the status line. The progress bar indicates the remaining storage duration.
Terminating the
automatic storage
function prema-
turely
If the automatic data storage function is activated, only the following keys are active: <M>, <STO
_> and <On/Off>. The other keys
and the automatic switch-off function are deactivated.
Proceed as follows to switch off the automatic data storage function before the
adjusted storage duration has expired:
1.Press the <STO_> key.
The following window appears.
78ba76143e0104/2013
2.Using <><>, select yes and confirm with <ENTER>.
The meter switches to the measured value display.
The automatic data storage function is terminated.
MultiLab 4010-3Data storage
10.3Measurement data storage
10.3.1 Editing the measurement data storage
The contents of the manual or automatic measurement data storage can be
shown on the display.
Each of the measurement data storages has a function to erase the entire
contents.
Editing the
data storage
Settings
The storage is edited in the menu, Storage & config / Data storage. To open
the Storage & config menu, press the <ENTER
display.
Open the manual or automatic storage directly with the <RCL> or <RCL
key.
The settings are explained here using the manual data storage as
an example. The same settings and functions are available for the
automatic data storage.
Menu itemSetting/
function
Data storage /
Manual data storage /
Display
-Displays all measurement datas-
_> key in the measured value
_>
Explanation
ets page by page.
Further options:
Scroll through the datasets
with <><>.
Output the displayed dataset
to the interface with <PRT>.
Data storage /
Manual data storage /
Output to USB flash
drive or printer
Data storage /
Manual data storage /
Output to RS232/USB
Data storage /
Manual data storage /
Erase
Quit the display with <ESC>.
-Outputs all stored measured values to the USB-A interface (USB
memory device/USB printer)
-Outputs all stored measurement
data to the USB-B interface (PC)
-Erases the entire manual measurement data storage.
Note:
All calibration data remain stored
when this action is performed.
4110
Ser. no. B092500013
ID number 2
pH 6.012 24.8 °C, AR, Sensor: +++
________________________________________
etc...
Display presentation
of a dataset
Sample printout
Quitting the displayTo quit the display of stored measurement datasets, you have the following
options:
Switch directly to the measured value display with <M>.
Quit the display and move to the next higher menu level with <ESC>.
10.3.2 Erasing the measurement data storage
How to erase the measured value storage is described in section 10.3.1
DITINGTHEMEASUREMENTDATASTORAGE, page 79.
80ba76143e0104/2013
E
10.3.3 Measurement dataset
A complete dataset includes:
Date/time
Meter name, series number
Sensor name, series number
ID number
MultiLab 4010-3Data storage
Measured value of the connected sensor
Measured temperature value of the connected sensor
AutoRead info: AR appears with the measured value if the AutoRead crite-
rion was met while storing (stable measured value). Otherwise, the AR display is missing.
Calibration evaluation:
- 4 levels (+++, ++, +, -, or no evaluation) or
- QSC (percentage)
10.3.4 Storage locations
The MultiLab 4010-3 meter has two measurement data storages. The
measured values recorded either manually or automatic are stored separately
in individual measurement data storages.
StorageMaximum number of datasets
Manual data storage500
Automatic data storage10000
ba76143e0104/201381
Transmitting dataMultiLab 4010-3
11Transmitting data
11.1Outputting current measurement data
1.Output the current measurement data to an interface with <PRT>.
USB-B: e.g.: PC
USB-A: e.g.: printer or USB flash drive
(if no PC is connected)
11.2Transmitting data
The meter has the following interfaces:
USB-B interface (USB Device)
e.g. to connect a PC
USB-A interface (USB Host),
e.g. to connect a USB flash drive/USB printer
Installation of the
USB driver on the
PC
Via the USB-B interface (USB Device) you can transmit data to a PC or printer
and update the meter software.
Via the USB-A interface (USB Host) it is possible to transfer data to an external
USB memory device/USB printer.
11.3Connecting the PC / USB-B interface (USB Device)
Connect the MultiLab 4010-3 to the PC via the USB-B interface.
System requirements of the PC for installation of the USB driver:
PC with at least one free USB connection and CD-ROM drive
Windows 2000, Windows XP, Windows Vista or Windows 7.
1.Insert the supplied installation CD in the CD drive of your PC.
2.Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3.Connect the MultiLab 4010-3 to the PC via the USB-B interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.
82ba76143e0104/2013
MultiLab 4010-3Transmitting data
4.Set the same transmission data at the connected instrument (PC):
Baud rate: to be selected in the range 1200 ... 19200
Handshake: RTS/CTS
Set at the PC only:
– Parity: none
– Data bits: 8
– Stop bits: 2
11.4Connecting a USB memory device/USB printer
(USB-A interface (USB Host))
Connect the USB-A interface (USB Host) of the MultiLab 4010-3 to a USB flash
drive/USB printer.
1.Connect a USB memory device/USB printer to the USB Host interface.
Suitable USB printers:
ModelTypePaper width
Citizen CT-S281Thermal transfer printer58 mm
Star SP700 with USB
Matrix printer76 mm
interface*
*Recommended printer settings for Star SP700:
- CodePage 437
- DIP switch 1...7: =ON, DIP switch 8: OFF
Details: see operating manual of your printer.
ba76143e0104/201383
Transmitting dataMultiLab 4010-3
11.5Options for data transmission to the USB-B (PC)
interface and the USB-A (USB printer) interface
Via the USB-B interface you can transmit data to a PC.
Via the USB-A interface, you can transmit data to a USB printer or USB memory device. How to transmit data to a USB memory device is described in an
extra section (see section 11.6 D
(USB
MEMORYDEVICE), page 85).
The following table shows which data are transmitted to the interface in which
way:
DataControlOperation / description
ATATRANSMISSIONTOTHE USB-A INTERFACE
Current
measured
values of all
connected
sensors
Stored measured values
Calibration
records
Manual With <PRT>.
Simultaneously with every manual data storage process (see
section 10.1 M
Automatic, at
intervals
Manual Displayed dataset with <PRT> after calling up from the storage.
Manual Displayed calibration record with <PRT> (see section 5.2.6 C
With <PRT
Then you can set the transmission interval.
Simultaneously with every automatic data storage process (see
section 10.2 A
All datasets with the Output to RS232/USB function.
(see section 10.3.1 E
page 79).
All datasets with the Output to USB flash drive or printer function
All calibration records with the Output to USB flash drive or
printer function.
ANUALDATASTORAGE, page 76).
_>.
UTOMATICDATASTORAGEATINTERVALS, page 76).
DITINGTHEMEASUREMENTDATASTORAGE,
DITINGTHEMEASUREMENTDATA
ALI-
ALIBRATIONDATA, page 59).
All calibration records with <PRT
For details, see section 11 T
Automatic At the end of a calibration procedure.
The following rule applies: With the exception of the menus, shortly
pressing the <PRT> key generally outputs the display contents to
the interface (displayed measured values, measurement datasets,
calibration records).
84ba76143e0104/2013
RANSMITTINGDATA, page 82.
_>.
MultiLab 4010-3Transmitting data
11.6Data transmission to the USB-A interface
(USB memory device)
Via the USB-A interface, you can transmit data to a USB memory or USB
printer device. How to transmit data to a USB printer device is described in an
extra section (see section 11.6 D
(USB
MEMORYDEVICE), page 85).
The following table shows which data are transmitted to the interface in which
way:
DataControlOperation / description
ATATRANSMISSIONTOTHE USB-A INTERFACE
Stored measured values
ManualAll datasets with the Output to RS232/
USB function.
For details, see section 10.3.1 E
THEMEASUREMENTDATASTORAGE, page
DITING
79
Calibration
data storage
ManualAll stored calibration records of a sensor
with the Calibr. data storage to USB flash drive function.
For details, see menu for calibration and
measurement settings of the sensor.
11.7MultiLab Importer
With the aid of the MultiLab Importer software, you can record and evaluate
measurement data with a PC.
More detailed information can be found in the MultiLab Importer
operating manual.
11.8BOD Analyst Pro
With the aid of the BOD Analyst Pro software you can administrate BOD
measurement on the PC and automatically calculate the measured values.
More detailed information can be found in the BOD Analyst Pro
operating manual.
ba76143e0104/201385
Maintenance, cleaning, disposalMultiLab 4010-3
1
2
12Maintenance, cleaning, disposal
12.1Maintenance
12.1.1 General maintenance activities
The only maintenance activity required is exchanging the battery that buffers
the system clock.
See the relevant operating manuals of the IDS sensors for instructions on maintenance.
12.1.2 Exchanging the battery
The MultiLab 4010-3 has a battery (CR2032 type) to buffer the system clock in
case of a power failure.
To retain the current setting of the date and time during the battery
exchange, supply the meter with power with the power pack while
changing the battery.
To avoid that the system clock is reset in the case of a power failure,
we recommend changing the battery before the end of its service
life (the battery included in the scope of delivery after approx. 8
years).
1.Unscrew the screws (2) of the lid of the battery compartment with a
screwdriver.
86ba76143e0104/2013
MultiLab 4010-3Maintenance, cleaning, disposal
2.Open the battery compartment (1) on the underside of the meter.
3.Remove the battery from the battery compartment.
4.Insert a new battery in the battery compartment.
5.Close the battery compartment (1).
The date (day) flashes on the display.
6.Tighten the screws (2) of the lid of the battery compartment with a
screwdriver.
7.Set the date and time (see 4.5.5 E
THEDATEANDTIME, PAGE 25)
XAMPLE 2 ONNAVIGATION: SETTING
Dispose of used batteries according to the local regulations of your
country.
End users within the European Union are obligated to return used
batteries (even ecologically compatible ones) to a collection point
set up for recycling purposes.
Batteries are marked with the crossed-out waste container
symbol. Therefore, they may not be disposed with the domestic
waste.
12.2Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lintfree cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain solvents. Remove any splashes immediately.
12.3Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
meter against damage during transport.
12.4Disposal
At the end of its operational lifetime, the meter must be returned to the
disposal or return system statutory in your country. If you have any questions,
please contact your supplier.
ba76143e0104/201387
Maintenance, cleaning, disposalMultiLab 4010-3
88ba76143e0104/2013
MultiLab 4010-3What to do if...
13What to do if...
13.1pH
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
CauseRemedy
IDS pH sensor:
– Measured value outside the measur-
ing range
– Air bubble in front of the diaphragm– Remove air bubble
– Air in the diaphragm– Extract air or moisten junction
– Cable broken– Exchange IDS pH sensor
– Gel electrolyte dried out– Exchange IDS pH sensor
CauseRemedy
IDS pH sensor:
– The values determined for zero point
and slope of the IDS pH sensor are
outside the allowed limits.
– Diaphragm contaminated– Clean diaphragm
– Use suitable IDS pH sensor
– Recalibrate
– IDS pH sensor broken– Exchange IDS pH sensor
Buffer solutions:
– The used buffer solutions do not
agree with the set buffer set
– Buffer solutions too old– Use only once.
– Buffer solutions depleted– Change solutions
ba76143e0104/201389
– Set different buffer set
or
– Use different buffer solutions
Note the shelf life
What to do if...MultiLab 4010-3
No stable measured
CauseRemedy
value
IDS pH sensor:
– Diaphragm contaminated– Clean diaphragm
– Membrane contaminated– Clean membrane
Test sample:
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Adjust temperature if necessary
IDS pH sensor + test sample:
– Conductivity too low– Use suitable IDS pH sensor
Obviously incorrect
measured values
– Temperature too high– Use suitable IDS pH sensor
– Organic liquids– Use suitable IDS pH sensor
CauseRemedy
IDS pH sensor:
– IDS pH sensor unsuitable– Use suitable IDS sensor
– Temperature difference between
buffer and test sample too great
– Measurement procedure not suit-
– Adjust temperature of buffer or
sample solutions
– Follow special procedure
able
13.2Dissolved oxygen
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Error message,
CauseRemedy
OFL
Error message,
– Measured value outside the measur-
ing range
CauseRemedy
– Select a suitable measuring
medium
Error
– Sensor contaminated– Clean the sensor
90ba76143e0104/2013
MultiLab 4010-3What to do if...
CauseRemedy
Error message,
OFL
– Measured temperature value outside
the operating conditions
– Keep to the temperature range
for the test sample
(display of OFL/UFL instead of a
temperature value)
– Defective sensor– Calibrate
– Change sensor cap
– Replace sensor
13.3Conductivity
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
CauseRemedy
– Measured value outside the measur-
ing range
– Use suitable IDS conductivity
sensor
Error message,
Error
CauseRemedy
– Sensor contaminated– Clean the sensor and replace it
if necessary
– Calibration solution not suitable– Check the calibration solutions
ba76143e0104/201391
What to do if...MultiLab 4010-3
13.4General information
Sensor symbol
flashes
Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
CauseRemedy
– Calibration interval expired– Recalibrate the measuring sys-
tem
CauseRemedy
– Operating condition undefined or
EMC load unallowed
– Processor reset:
Press the <ENTER> and <On/
Off> key simultaneously
CauseRemedy
– E. g., a question by the service
department
– Switch on the meter.
– Open the menu, <ENTER_> /
Storage & config / System / Service information. The instrument
data are displayed.
or
– Connect the sensor.
Press softkey [Info]/[More] The
sensor data are displayed (see
section 4.1.6 S
ENSORINFO,
page 17)
Data transmission to
USB memory device
does not work
CauseRemedy
– Connected USB memory device was
not recognized
– The USB-B interface is connected to
a PC
– The USB memory device has been
formatted to a file system which is
not supported, e. g. NTFS
– Use other USB memory device
– Disconnect the PC from the
USB-B interface
– Reformat the USB memory
device to FAT 16 or FAT 32 file
system
(Caution
all data on the USB memory
device. Back up all data before
reformatting.)
: Reformatting erases
92ba76143e0104/2013
MultiLab 4010-3What to do if...
Data transmission to
USB printer does not
work
Error message,
Memory error 1
Time gets lost
CauseRemedy
– The USB-B interface is connected to
a PC
– Connected USB memory device
does not work
– Disconnect the PC from the
USB-B interface
– Use a suitable USB printer (see
section 11.4 C
MEMORYDEVICE/USB PRINTER
(USB-A
H
OST)), page 83)
ONNECTINGA USB
INTERFACE (USB
– Check printer settings (see sec-
tion 11.4 C
MEMORYDEVICE/USB PRINTER
(USB-A
H
OST)), page 83)
ONNECTINGA USB
INTERFACE (USB
CauseRemedy
– Instrument storage was not found– Please contact the service.
CauseRemedy
– Buffer battery is empty– Change the buffer battery (see
section 12.1.1 G
NANCEACTIVITIES, page 86)
ENERALMAINTE-
ba76143e0104/201393
Technical dataMultiLab 4010-3
14Technical data
14.1Measuring ranges, resolution, accuracy
Measuring ranges,
accuracy
ParameterMeasuring rangeAccuracy
Air pressure (absolute)*
*Available only if a D.O. sensor is connected
Further data are given in the documentation of your sensor.
300 ... 1100 mbar± 4 mbar
14.2General data
DimensionsApprox. 180 x 80 x 55 mm
(11.22 x 10.04 x 3.15 inches)
WeightApprox. 0.4 kg
Mechanical structureType of protectionIP 43
Electrical safetyProtective classIII
Test certificatesCE, cETLus
Ambient conditionsStorage- 25 °C ... + 65 °C
Operation0 °C ... + 40 °C
Admissible relative
humidity
Power supplyPower pack Kuantech Co. Ltd. KSAC 0900110W1UV-1
Battery
(to buffer the system
clock in the case of a
power failure)
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 270 mA
Output: 9 V = / 1.1 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
3.0 V lithium manganese dioxide battery,
type CR 2032
USB-B (Device), PC
1200, 2400, 4800, 9600, 19200 Baud
Data bits8
Stop bits2
94ba76143e0104/2013
MultiLab 4010-3Technical data
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
Changes or modifications not expressly approved by the
manufacturer could void the user‘s authority to operate the
equipment.
ParityNone
HandshakeRTS/CTS
Cable length Max. 3 m
USB interface (host)TypeUSB 2.0
USB-A (host), USB device
Guidelines
and norms used
EMCEC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safetyEC directive 2006/95/EC
EN 61010-1
UL 61010-1
CAN/CSA-C22.2#61010-1
IP protection classEN 60529
ba76143e0104/201395
Technical dataMultiLab 4010-3
LSN 25/71815
15/07/2004
CERTIFICATE OF ANALYSIS
Sample Ref No.:
Date Received:
The microbiological results demonstrate that the material under test exhibits biocidal activity.
Comment:
03/03/2005
Date Reported:
AUTOTEX AM
Client:
Autotype International Limited
Grove Road
Wantage
Oxon
OX12 7B2
United Kingdom
FILM TEST
Meth. Desc
AUTOTYPESupplier:
Job Ref:
04I0712
x
x
TestResultUnitE st
Salmonella enteritidis%99.6Reduction After 24 Hoursx
x
Klebsiella pneumoniae%99.4Reduction After 24 Hoursx
x
Pseudomonas aeruginosa%99.1Reduction After 24 Hoursx
x
Streptococcus faecalis%99.4Reduction After 24 Hoursx
x
Phoma violacea%99.0Reduction After 48 Hoursx
x
Penicillium purpurogenum%99.3Reduction After 48 Hoursx
x
Bacillus cereus%99.3Reduction After 24 Hoursx
x
Saccharmyces cerevisiae%99.3Reduction After 24 Hoursx
x
1
P.M.Sutton
s
Nutritional Services
Manager
Senior Microbiologist
CChem., MRSC.
BSc. (Hons.)
N.StantonJ.Lloyd
Senior
Microbiologist
Principal
Microbiologist
C.Fuller
MIFST
Company Microbiologist
J.Elliott
Senior
Microbiologist
BSc. (Hons.)
BSc. (Hons.)
R.P.Elliott
Deputy Managing
Director
CChem, MRSC, MIFST
Keypad (antibacterial)
BSc. (Hons.), CBiol., MBiol.,
BSc. (Hons.), CBiol., MBiol
J. Franci
Law Laboratories Ltd Shady Lane, Great Barr, Birmingham B44 9ET England
04I0712/6/1/
96ba76143e0104/2013
MultiLab 4010-3Technical data
LSN 26/38123
24/10/2004
CERTIFICATE OF ANALYSIS
Sample Ref No.:
Date Received:
The microbiological results demonstrate that the material under test exhibits biocidal activity against
the above listed microbial strains.
Comment:
21/02/2005
Date Reported:
AUTOTEX AM AGED 15 YEARS
Client:
Autotype International Limited
Grove Road
Wantage
Oxon
OX12 7B2
United Kingdom
Harmonised JIS Z2801/AATCC 100
Meth. Desc
Job Ref:
05B1760
x
x
x
TestResultUnitE st
Staphylococcus aureus%99.0Reduction After 24 Hoursx
x
Escherichia coli 0157%99.8Reduction After 24 Hoursx
x
Aspergillus niger%99.1Reduction After 48 Hoursx
x
x
3
P.M.Sutton
s
Nutritional Services
Manager
Senior Microbiologist
CChem., MRSC.
BSc. (Hons.)
N.StantonJ.Lloyd
Senior
Microbiologist
Principal
Microbiologist
C.Fuller
MIFST
Company Microbiologist
J.Elliott
Senior
Microbiologist
BSc. (Hons.)
BSc. (Hons.)
R.P.Elliott
Deputy Managing
Director
CChem, MRSC, MIFST
BSc. (Hons.), CBiol., MBiol.,
BSc. (Hons.), CBiol., MBiol
x
x
J. Franci
Law Laboratories Ltd Shady Lane, Great Barr, Birmingham B44 9ET England
05B1760/1/3/
ba76143e0104/201397
Firmware updateMultiLab 4010-3
15Firmware update
15.1Firmware update for the meter MultiLab 4010-3
You can find available firmware update files for your meter on the Internet. With
the "Firmware Update " program and a PC you can update the firmware of the
MultiLab 4010-3 to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB-B interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the MultiLab 4010-3).
1.Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type), the new
data are displayed there.
2.In the windows start menu, open the update folder and start the firmware update program for the meter
3.Using the USB interface cable, connect the MultiLab 4010-3 to a USB
interface (virtual COM port) of the PC.
4.Switch on the MultiLab 4010-3.
5.In the firmware update program, start the update process with OK.
6.Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes up to 15 minutes. A terminatory message is displayed after a successful programming process. The firmware update is completed.
7.Disconnect the MultiLab 4010-3 from the PC.
The MultiLab 4010-3 is ready for operation again.
After switching the meter off and on you can check whether the meter has taken
over the new software version (see Y
VERSIONISINTHEMETEROR IDS SENSOR, PAGE 92).
OUWANTTOKNOWWHICHSOFTWARE
98ba76143e0104/2013
MultiLab 4010-3Firmware update
15.2Firmware-Update for IDS Sensors
With the "Firmware Update" program and a PC you can update the firmware of
an IDS sensor to the newest version.
You can find available firmware update files for your IDS sensor on the Internet.
For updating, connect the IDS sensor to the MultiLab 4010-3, and the
MultiLab 4010-3 to a PC.
For the update via the USB-B interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the MultiLab 4010-3).
1.Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the sensor (or sensor type), the
new data are displayed there.
2.In the windows start menu, open the update folder and start the firmware update program for the IDS sensor
3.Connect the IDS sensor to the MultiLab 4010-3 meter.
4.Using the USB interface cable, connect the MultiLab 4010-3 to a USB
interface (virtual COM port) of the PC.
5.Switch on the MultiLab 4010-3.
6.In the firmware update program, start the update process with OK.
7.Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes up to 5 minutes. A terminatory message is displayed after a successful programming process. The firmware update is completed.
8.Disconnect the MultiLab 4010-3 from the PC.
Meter and sensor are ready for operation again.
After switching the meter off and on you can check whether the sensor has
taken over the new software version (see Y
VERSIONISINTHEMETEROR IDS SENSOR, PAGE 92).
OUWANTTOKNOWWHICHSOFTWARE
ba76143e0104/201399
GlossaryMultiLab 4010-3
16Glossary
pH/ORP
Asymmetrysee zero point
Electromotive force of
a combination
electrode
JunctionThe junction is a porous body in the housing wall of reference elec-
ORP voltage (U)The ORP is caused by oxidizing or reducing substances dissolved in
pH valueThe pH value is a measure of the acidic or basic effect of an aqueous
PotentiometryName of a measuring technique. The signal (depending on the mea-
SlopeThe slope of a linear calibration function.
The electromotive force U of the combination electrode is the measurable electromotive force of an electrode in a solution. It equals the sum
of all the galvanic voltages of the combination electrode. Its dependency on the pH results in the electrode function, which is characterized
by the parameters, slope and zero point.
trodes or electrolyte bridges. It arranges the electrical contact between
two solutions and makes the electrolyte exchange more difficult. The
expression, junction, is also used for ground or junction-less transitions.
water if these substances become effective on an electrode surface (e.
g. a gold or platinum surface).
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
Zero pointThe zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified
temperature is zero. Normally, this is at 25 °C.
Conductivity
Cell constant (C)Characteristic quantity of a conductivity measuring cell, depending on
the geometry.
Conductivity (ϰ)Short form of the expression, specific electrical conductivity.
It corresponds to the reciprocal value of the resistivity.
It is a measured value of the ability of a substance to conduct an electric
current. In water analysis, the electrical conductivity is a dimension for
the ionized substances in a solution.
Reference
temperature
Resistance (ρ)Short name for the electrolytic resistivity. It corresponds to the recipro-
Fixed temperature value to compare temperature-dependent measured
values. For conductivity measurements, the measured value is converted to a conductivity value at a reference temperature of 20 °C or 25
°C.
cal value of the electrical conductivity.
100ba76143e0104/2013
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.