Xylem 3017M Quick Start Guide

3017M Quick Start Guide
DPD CHLORINE ANALYZER
Quick Start Guide
330028
Table of Contents
1 What’s Included?
2 Inspection
3 Materials Required
4 Mounting Instructions
6 Reagent Preparation
7 Sample Inlet Device - Overview, Warnings & Mounting Location
9 Sample Inlet Device - Step-by-Step Installation Procedure
11 Sample Pump Tube Installation
12 Reagent Tube Installation
13 Startup & Operation
What’s Included?
Item Part # Qty.
3017M Chlorine Analyzer 332266 1
Sample Pump Tube* 332405 1
Reagent Pump Tubes* 332268 2
Reagent Bottle Cap Assembly* 332270 2
1/2” ID Drain Tube, 6 ft. 327112 1
1/8” OD Sample Inlet Tube, 3 ft. 147901 1
3017M Operators Manual 332100 1
Accessories that may be included:
Sample Inlet Device 327114 1
Kit - Total Chlorine Reagent or Kit - Free Chlorine Reagent
Reagent Mixing Instructions 330030 1
330006 330007
*Installed on the 3017M
1
1
Inspection
Remove the analyzer and Sample Inlet Device from the shipment boxes. Inspect the shipment for any damage or missing parts. Contact YSI Customer Support immediately to report any damage or discrepancies with the shipment. Any questions should be directed to YSI Customer Support at (800) 765-4974 or (937) 767-7241.
2
5”
1/2” ID Urethane tubing
As close as possible 1m
(3.28 ft) max.
Figure 1: YSI 3017M DPD Chlorine Analyzer and Sample Inlet Device shown in the recommended installation configuration.
open to atmospheric drain (item #327112)
NOTE: Do not apply power to the unit until all of the following steps have been completed.
The 3017M was shipped from the factory as shown in Figure 2. It will be necessary to attach the mounting tabs, mount the analyzer, mix the reagents, position the sample pump tube, tension the reagent pump tubes and connect the analyzer to the sample.
Materials Required
• #1 Phillips screwdriver
• 1/16” (or smaller) flat-head screwdriver
• 3/8” drive ratchet
• 7/16” socket
• 7/16” wrench
• Scissors
• Rag or paper towels
Figure 2: The 3017M Chlorine Analyzer
as it is shipped from the factory
Display & Control
Sample Pump
Reagent Pump
Measuring Cell
I/O Terminal Block
1. Power
2. Communication
3
Reagents
Figure 3: 3017M Chlorine
Analyzer components
Mounting Instructions
1. Attach the enclosure’s 4 mounting tabs to the back of the 3017M
using the flat head screws in the hardware kit.
2. Mount the 3017M in the desired location with user supplied
mounting hardware.
3. Install the Sample Inlet Device in its designated location (if
applicable). Below and to the left of the analyzer is an ideal position. See Figure 1.
4. Power and RS 485/4 to 20 mA connections are made through the
cable glands that are supplied with the analyzer. The cable glands can be found on the left-hand side of the analyzer.
5. Wire the main power and any other signal or alarm connections.
6. The terminal block connectors are opened by pushing the 1/16”
screwdriver into the small, square opening adjacent to the opening for the wire. See Figure 4.
External Power Switch (recommended)
4
1 2 3
4
5 6 7 8
9 10 11 12 13 14 15 16
Figure 4: 3017M terminal block and external power switch
(Wiring reference table on next page)
Loading...
+ 11 hidden pages