Withings WPM05 Users Manual

Withings BPM Connect
INSTRUCTIONS MANUAL
MANUEL D'UTILISATION
EN | FR | DE
|
BETRIEBSANLEITUNG
THANK YOU FOR CHOOSING
WITHINGS BPM CONNECT
VIELEN DANK, DASS SIE SICH FÜR WITHINGS BPM CONNECT ENTSCHIEDEN HABEN
This guide is available in additional languages at: www.withings.com/guides
FR Ce guide est disponible dans plusieurs langues sur www.withings.com/guides
DE Diese Anleitung finden Sie in weiteren Sprachen unter: www.withings.com/guides
IT Questa guida è disponibile in altre lingue all'indirizzo: www.withings.com/guides
ES Esta guía está disponible en otros idiomas en: www.withings.com/guides
Overview
Vue d'ensemble
A
B
C
|
Übersicht
A
Button
FR Bouton ON/OFF DE Knopf ON/OFF
B
Cuff
FR Brassard DE Manschette
C
USB plug
FR Prise USB DE USB-Stecke
3
Quick start from your smartphone
Démarrage rapide depuis votre smartphone | Schnellstart über Ihr Smartphone
1
Visit go.withings.com from your smartphone or tablet, and follow the instructions to install the Health Mate app.
FR Visitez go.withings.com depuis votre smartphone
ou votre tablette et suivez les instructions pour installer l'application Health Mate.
DE Besuchen Sie mit dem Browser Ihres Smartphones
oder Tablets go.withings.com und folgen Sie den Anweisungen zum Installieren der Health Mate App.
go.withings.com
2
In the app, follow the instructions to install Withings BPM Connect.
FR Dans l'application, suivez les instructions pour
installer Withings BPM Connect.
DE Folgen Sie in der App den Anweisungen zur
Installation von Withings BPM Connect.
4 – 5
How to take a blood pressure measurement
Comment prendre une mesure de tension artérielle | Eine Blutdruckmessung vornehmen
1
Sit down in a comfortable position.
FR Asseyez-vous confortablement.
DE Setzen Sie sich in einer komfortablen Position.
2
Unroll cuff. Place your arm inside it.
FR Déroulez votre brassard. Placez votre bras à
l'intérieur.
DE Entfalten Sie Ihre Manschette. Bitte legen Sie
Ihren Arm drinnen.
6 – 7
How to take a blood pressure measurement
Comment prendre une mesure de tension artérielle | Eine Blutdruckmessung vornehmen
3
Tighten the cuff around your arm. The tube should be positioned against your inner arm.
FR Serrez le brassard autour de votre bras. Le tube
doit être placé dans le creux du bras.
DE Ziehen Sie die Manschette um den Arm. Der
Schlauch sollte zur Innenseite Ihres Armes positioniert werden.
4
Place your arm on a table and level with your heart.
FR Placez votre bras sur une table, de sorte que le
brassard soit au même niveau que le cœur.
DE Platzieren Sie Ihren Arm auf einem Tisch auf
gleicher Ebene mit Ihrem Herz.
8 – 9
How to take a blood pressure measurement
Comment prendre une mesure de tension artérielle | Eine Blutdruckmessung vornehmen
5
Turn on your blood pressure monitor to start the measurement. You can select a simple or triple measurement with a long press on the button. Press the button again to start the measurement.
FR Allumez votre tensiomètre pour commencer la
mesure. Vous pouvez sélectionner une mesure simple ou triple avec un appui long sur le bouton. Appuyez à nouveau sur le bouton pour commencer à prendre la mesure.
DE Schalten Sie Ihr Blutdruckmessgerät ein, um die
Messung zu beginnen. Sie können eine einfache oder dreifache Messung durch langes Drücken der Taste auswählen. Drücken Sie die Taste erneut, um die Messung zu beginnen.
10
Box Contents
Contenu de la boîte | Lieferumfang
11
Blood Pressure Monitor
FR Tensiomètre
DE Blutdruckmessgerät
Charging cable
FR Câble de recharge
DE Ladekabel
Instructions manual
FR Manuel d'utilisation
DE Betriebsanleitung
Important information
12
Requirements
In order to use your BPM Connect, you need an iOS (10.0 or higher) or Android (6.0 and higher) device to install it. Thereafter, the product can be used in stand-alone, thanks to the Wi-Fi and Bluetooth® connections.
Important safety information
Consult your doctor during pregnancy, or if you suer from arrhythmia and arteriosclerosis. Please read this section carefully before using the Wi-Fi Smart Blood Pressure monitor.
Intended use
The BPM Connect is a digital monitor intended for use in measuring blood pressure, heart rate. The device is intended to be used in a human adult population with an arm circumference between 9 inches to 17 inches (22 cm to 42 cm). The BPM Connect is a medical device. Contact your physician if hypertensive values are indicated.
General safety and precautions
- Do not forcibly bend the arm cu.
- Do not inflate the arm cu when it is not wrapped around your arm.
- Do not apply strong shocks and vibrations to the blood pressure monitor or drop it.
- Do not take measurement after bathing, drinking alcohol, smoking, exercising or eating.
- Do not immerse the arm cu in water.
- Do not use with a pacemaker, a defibrillator or other electric implant.
- Use on adults only.
- Do not use on children or pets.
Android is a trademark of Google Inc. iOS is a trademark of Apple Inc.
Caution
13
General usage
Always consult your doctor. Self-diagnosis of measurement results and self treatment are dangerous. People with severe blood flow problems, or blood disorders, should consult a doctor before using the blood pressure monitor. Cu inflation can cause internal bleeding. Operational factors such as common arrhythmias, ventricular premature beats, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-eclampsia or renal disease can aect the performance of the automated sphygmomanometer and/ or its blood pressure reading. This device is a precision measuring equipment liable to be understood by lay user but it still should be handled with care. A long exposure of the device to lint, dust or sunlight might reduce its life time or damage it. Damaged cu or sensor might lead to incorrect measurements. No usage for USB port. The user is an intended operator. Parts in contact with the skin: cu. Measurements can be aected by extreme temperatures, humidity & altitude.
- Do not leave the blood pressure monitor unattended with infants or persons who cannot express their consent.
- Do not use the blood pressure monitor for any purpose other than measuring blood pressure.
- Do not disassemble the blood pressure monitor.
- Do not operate the blood pressure monitor in a moving vehicle (car, airplane).
- Do not use a cellular phone near the device.
- Do not use the device with the USB cable plugged.
- Improper continuous pressure of cu or too frequent measurements may interfere blood flow and result harmful injury to the users. Check if the use of the device does not result in prolonged impairment of your blood circulation.
- Do not apply the cu over the users’ arm bearing a wound or medical treatment and so on, as this can cause further injury.
- Cu should be used on the arm on the side of a mastectomy.
- Use of the device can temporarily cause loss of function of simultaneously used equipment on the same limb.
After use
14
Cleaning
- Do not use an alcoholic-base or solvent agent to clean the device
- Clean the device with a soft and dry cloth
- The dirt on the cu can be cleaned by a moisten cloth and soap
- Do not flush the device and cu with much water
- Do not dismantle the device or disconnect the cu or try to repair by yourself. If any problem happens, refer to the distributor
- Do not operate the device under severe environment of extreme temperature or humidity, or direct sunshine
- Do not shake the unit violently
- Do not submerge the device or any of the components in water
- Do not let the device under strong shocks, such as dropping the unit on the floor
Storage
- Store the device and the components in a clean and safe location
- If storage conditions are dierent from the usage conditions indicated in this document, please wait 30 minutes before taking a measurement.
Maintenance
If you can’t fix the problems using the troubleshooting instructions , request service from your dealer. Manufacturer will make available on requested circuit diagrams, component part lists, descriptions, calibration instructions, or other information that will assist manufacturer’s sta or authorized representative for repair. It is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Do not use the device while doing maintenance steps.
Error messages
Error message Causes Countermeasures
Measurement could not be performed. Please try again. If the problem occurs again, please contact customer service.
Connection error. Quit the application and reconnect your blood pressure monitor.
Please wait until the cuff is totally deflated before starting a new measurement. Stay still during measurement.
Hardware auto- check error. Measurement could not be performed. Please
Communication error between the Blood Pressure Monitor and the collateral device.
- Cuff is not deflated before measurement.
- Noise interference.
- Excessive user movements.
try again. If the problem occurs again, please contact customer service.
1. Push the blood pressure monitor button
for 3 seconds to turn it off. Quit the appli­cation and push the blood pressure monitor button to turn it on.
2. If the problem occurs again, please contact
customer service.
1. Please wait until the cuff is totally deflated
before starting a new measurement. Stay still during measurement.
2. If the problem occurs again, please contact
customer service.
15
Error messages
Error message Causes Countermeasures
Check that the Blood Pressure Monitor is correctly positioned on your arm and that mea­surement is performed in good conditions.
Low battery level. Battery level is too low. Load the device with the provided charging
- User movement, or not relaxed, or talking …
- Cuff is not correctly fastened
- Cuff broken
- Pump or valve failure
- Pressure overflow
- Noise interference
- User movement, or not relaxed, or talking …
- User special characteristics
- Battery level is too low
1. Check that the Blood Pressure Monitor
is correctly positioned on your arm and that measurement is performed in good conditions.
2. Low battery level. Load your BPM Connect
with a micro USB cable.
3. If user has special characteristics, please
contact your physician.
4. If the problem occurs again, please contact
customer service.
cable.
16
Troubleshooting
Problem Remedy
The inflation action cannot be done or the air pressure cannot rise
The low battery icon is displayed on the device Reload the device with the cable provided
The blood pressure cannot be taken and the application shows an error message
Under normal measuring circumstance, the reading at home is different from that of the clinics or each measurement has various reading
1. Check the cuff position and fasten the cuff correctly ad remeasure the pressure again
2. Check the blood pressure monitor connection to the iOS or Android device
1. Re-fasten the cuff
2. Relax yourself and sit down
3. Keep the cuff and heart at the same level during the measurement period
4. Keep silent and still during measurement
5. If the patient has sever heart beat problem, then the blood pressure may not be read correctly
1. The variation is due to the different environments
2. The blood pressure is changing according to the physiological or psychological status of
the human body
3. Show your recorded values to your physician
17
Specifications Technical data
Product description: Digital automatic Blood Pressure Monitor
Model: WPM05
Blood pressure measurement method: Cuff oscillometric method
Cuff inflation: Automatic inflation with air pump at 15 mmHg/s
Pressure sensor: Gauge sensor
Measurement range (pressure): 0 to 285 mmHg, DIA 40 to 130 mmHg, SYS 60 to 230 mmHg
Measurement range (pulse): 40 to 180 beats/min
Accuracy (pressure): Within +- 3 mmHg or 2% of reading
Accuracy (pulse): Within +-5% of reading
Sensor: Semiconductor pressure sensor
Operating conditions: 5 to 40°C, 15 to 90% RH, atmospheric 86Kpa~106kpa, altitude: 2000m
Storage and transport conditions: -25 to 70°C, 10 to 95% RH, atmospheric 86Kpa~106kpa, altitude: 2000m