Withings BP-801 User Manual

Read this manual before use. Keep it for further reference.
Disclaimer: Information in this guide may change without notice.
The manufacturer assumes no responsibilities for errors that may appear in this guide.
WIRELESS BLOOD PRESSURE MONITOR
Requirements
In order to use your Wireless Blood Pressure Monitor, you need an iOS (7.0 or higher) or Android (4.0 and higher) device.
Package content
Withings Wireless Blood Pressure Monitor Four AAA alkaline (LR3) cells (already inserted) USB adapter
Important safety information
Consult your doctor during pregnancy, or if you suer from arrhyth­mia and arteriosclerosis. Please read this section carefully before using the Withings Wireless blood pressure monitor.
Intended use
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches (22cm- 42cm). This device is not intended to be a diagnostic device.
Contact your physician if hypertensive values are indicated.
General safety and precautions
– Do not forcibly bend the arm cu –Do not inate the arm cu when it is not wrapped around your arm –Do not apply strong shocks and vibrations to the blood pressure
monitor or drop it
–Do not take measurement after bathing, drinking alcohol, smoking,
exercising or eating.
–Do not immerse the arm cu in water
INTRODUCTION
General usage
Always consult your doctor. Self-diagnosis of measurement results and self treatment are dangerous People with severe blood ow problems, or blood disorders, should consult a doctor before using the blood pressure monitor. Cu ination can cause internal bleeding. Operational factors such as common arrhythmias, ventricular premature beats, atrial brillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-eclampsia or renal disease can aect the performance of the automated sphygmomanometer and/or its blood pressure reading. This device is a precision measuring equipment liable to be understood by lay user but it still should be handled with care. A long exposure of the device to lint, dust or sunlight might reduce its life time or damage it. Damaged cu or sensor might lead to incorrect measurements.
– Do not leave the blood pressure monitor unattended with
infants or persons who cannot express their consent.
– Do not use the blood pressure monitor for any purpose other than
measuring blood pressure. – Do not disassemble the blood pressure monitor. – Do not operate the blood pressure monitor in a moving vehicle
(car, airplane). – Do not use a cellular phone near the deivce. It may result in
an operational failure. – Improper continuous pressure of cu or too frequent measurements
may interfere blood ow and result harmful injury to the users.
– Do not apply the cu over the users’ arm bearing a wound or
medical treatment and so on, as this can cause further inucence on the therapy.
AAA alkaline cells usage
– If AAA alkaline cells uid should get on your skin or clothing,
immediately rinse with plenty of clean water.
– Use only four AAA alkaline cells with this blood pressure monitor.
Do not use any other types of AAA alkaline cells. – Do not insert AAA alkaline cells with their polarities incorrectly aligned. – Replace old AAA alkaline cells with new ones immediately.
Replace all four AAA alkaline cells at the same time. – Do not use new and used AAA alkaline cells together
Replacing the AAA alkaline cells
If the low battery symbol appears in the Withings application, replace all four AAA alkaline cells at the same time.
1. Remove the AAA alkaline cells cover at the lower end of
the aluminum tube.
2. Install or replace four AAA alkaline cells so that the + (positive) and – (negative) polarities match the polarities indicated on the AAA alkaline cells compartment.
3. Put the AAA alkaline cells cover back in place.
If the device will not be used for a long period of time, it is advised to remove the alkaline cells.
CAUTION
Cleaning
– Do not use an alcoholic-base or solvent agent to clean the device – Clean the device with a soft and dry cloth – The dirt on the cu can be cleaned by a moisten cloth and soap – Do not ush the device and cu with much water – Do not dismantle the device or disconnect the cu or try to repair
by yourself. If any problem happens, refer to the distributor
– Do not operate the device under severe environment of extreme
temperature or humidity, or direct sunshine – Do not shake the unit violently – Do not submerge the device or any of the components in water – Do not let the device under strong shocks, such as dropping
the unit on the oor
Storage
– If you are not using the device for an extended period, remove
the alkaline cells from the aluminium tube for storage. – Store the device and the components in a clean and safe location
Maintenance
If you can’t x the problems using the troubleshooting instructions , request service from your dealer. Manufacturer will make available on requested circuit diagrams, component part lists, descriptions, calibration instructions, or other information that will assist manufacturer’s sta or authorized representative for repair. It is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy.
AFTER USE
Error message Causes Countermeasures
Measurement could not be performed. Please try again. If the problem occurs again, please contact customer service.
Hardware auto- ‐check error. 1- Measurement could not be performed.
Please try again. If the problem occurs again, please contact customer service.
Connection error. Quit the application and recon­nect your blood pressure monitor.
Communication error between the Blood Pressure Monitor and the collateral device.
1- Push the blood pressure monitor button for 3 seconds to turn it o. Quit the application and push the blood pressure monitor button to turn it on. 2- If the problem occurs again, please contact customer service.
Please wait until the cu is totally deated before starting a new measurement. Stay still during measurement.
- Cu is not deated before measurement.
- Noise interference.
- Excessive user movements.
1- Please wait until the cu is totally deated before starting a new measurement. Stay still during measurement. 2- If the problem occurs again, please contact customer service.
ERROR MESSAGES