WIKA LS-10 Specifications

Electronic pressure measurement
Submersible pressure transmitter for level measurement Model LS-10, standard version
WIKA data sheet PE 81.55
Applications
Level measurement in rivers and lakes
Level measurement in vessel and storage systems
Control of sewage lift and pumping stations
Monitoring of sewage, settling and stormwater retention basins
Special features
Robust
Reliable
Economical
Cable supports up to 220 lbs. of strain
Rated IP68 for permanent submersion
Description
For simple measuring tasks
The model LS-10 submersible pressure transmitter has been optimised for simple measuring requirements in level
measurement. It oers excellent quality, is cost-eective and
reliable.
It has been designed to the current demands of the industry and has a 4 ... 20 mA output as standard, an accuracy of
0.5% and PUR cable. With IP 68 ingress protection, it is suitable for permanent level measurement up to 328 ft (100 m) water column.
Submersible pressure transmitter model LS-10
Reliable and long-lasting
The submersible pressure transmitter features a hermetically
sealed and exceptionally robust stainless steel case. The
proven, fully-welded construction ensures a long service life and permanent sealing.
Data sheets showing similar products: High-performance submersible pressure transmitter; model LH-20; see data sheet PE 81.56 Intrinsically safe submersible pressure transmitter; model IL-10; see data sheet PE 81.23
-
Optional WIKA LevelGuard Anti-clog attachment for submersible level transmitters. For use in lift stations,
wet wells and other dicult level applications. For more
information request bulletin LG-1.
Page 1 of 5WIKA data sheet PE 81.55 ∙ 01/2016
Measuring ranges
Relative pressure
bar Measuring range 0 ... 0.25 0 ... 0.4 0 ... 0.6 0 ... 1 0 ... 1.6
Overpressure limit 2 2 3 5 8
Burst pressure 2.4 2.4 4 6 10
Measuring range 0 ... 2.5 0 ... 4 0 .. 6 0 ... 10
Overpressure limit 8 10 10 10
Burst pressure 10 10 10 10
inWC Measuring range 0 ... 100 0 ... 150 0 ... 250
Overpressure limit 750 750 1,100
Burst pressure 950 950 1,600
psi Measuring range 0 ... 5 0 ... 10 0 ... 15 0 ... 25 0 ... 50
Overpressure limit 30 45 70 120 150
Burst pressure 35 60 90 180 150
Measuring range 0 ... 100 0 ... 160
Overpressure limit 150 160
Burst pressure 150 160
O Measuring range 0 ... 2.5 0 ... 4 0 ... 6 0 ... 10 0 ... 16
mH
2
Overpressure limit 20 20 30 50 80
Burst pressure 24 24 40 60 100
Measuring range 0 ... 25 0 ... 40 0 ... 60 0 ... 100
Overpressure limit 80 100 100 100
Burst pressure
100 100 100 100
The given measuring ranges are also available in mbar, kPa and MPa.
Output signal
Analogue signal
4 ... 20 mA
Load in Ω
≤ (power supply - 10 V) / 0.02 A - (cable length in m x 0.14 Ω)
Reference conditions
Temperature
59 ... 77 °F (15 ... 25 °C)
Atmospheric pressure
12.5 ... 15.4 psi (860 ... 1,060 mbar)
Humidity
45 ... 75 % relative
Voltage supply
Power supply
DC 24 V
Power supply
DC 10 ... 30 V
WIKA data sheet PE 81.55 ∙ 01/2016Page 2 of 5
Accuracy data
Electrical connection
Accuracy at reference conditions
≤ ±0.5 % of span
Including non-linearity, hysteresis, zero oset and end value
deviation (corresponds to measured error per IEC 61298-2).
Non-linearity (per IEC 61298-2)
≤ ±0.2 % of span
Non-repeatability
≤ ±0.1 % of span
Temperature error at 32 ... 176 °F (0 ... 50 °C)
Mean temperature coecient of zero point
Measuring ranges ≤ 3.6 psi: ≤ ±0.4 % of span/10 K
Measuring ranges > 3.6 psi: ≤ ±0.2 % of span/10 K
Mean temperature coecient of span ≤ ±0.2 % of span/10 K
Long-term stability at reference conditions
≤ ±0.2 % of span/year
Short-circuit resistance
S
vs. U
-
+
Reverse polarity protection
U+ vs. U
-
Insulation voltage
DC 500 V
Cable lengths
Available cable lengths
Meter (m) 1.5 3 5 10 15
20 25 30 40 50 60 80 100
Feet (ft) 5 10 20 30 40
50
Other cable lengths on request
Connection diagram
Cable outlet
U
+
U
-
Shield grey
brown
green
Operating conditions
Ingress protection (per IEC 60529)
IP 68
Permissible temperature ranges
Medium: 14 ... 122 °F (-10 ... +50 °C)
Ambient: 14 ... 122 °F (-10 ... +50 °C)
Storage: -22 ... 176 °F (-30 ... +80 °C)
Immersion depth
up to 328 ft. (100 m)
Maximum tensile strength of the cable
without strain relief: up to 350 N
with strain relief: up to 1,000 N
Weight
Level probe: approx. 0.4 lbs. (180 g)
Cable: approx. 0.054 lbs/ft (80 g/m)
Additional weight (accessories): approx. 1.11 lbs. (500 g)
Materials
Wetted parts
Case from stainless steel
Sensor out of stainless steel
Protection cap from PA
Cable from PUR
Approvals, directives and certicates
Approval
CSA
GOST-R
for further approvals, see local website
CE conformity
EMC directive 2004/108/EC, EN 61326 emission (group 1,
class B) and immunity (industrial application)
Page 3 of 5WIKA data sheet PE 81.55 ∙ 01/2016
Dimensions in mm
Cable outlet Cable from PUR
Accessories
Description Order number
Cable strain relief clamp
The cable strain relief clamp enables easy and secure mechanical fastening of the submers­ible pressure transmitter’s cable at the measuring point. It acts as a guide for the cable, in order to avoid mechanical damage and to reduce the tensile stress.
Additional weight
The additional weight increases the dead weight of the submersible pressure transmitter.
It simplies the lowering into monitoring wells, narrow shafts and deep wells. It eectively reduces negative environmental inuences on the measuring result from the measured medium (e.g. turbulent ow).
CrNi-Stahl 316L, approx. 500 g, length (L) 130 mm
Junction box
The junction box, with IP 67 ingress protection and watertight ventilation element, provides
a moisture-free electrical termination for the submersible pressure transmitter. It should be mounted in dry environment or directly in the switch cabinet.
Filter element
The lter element prevents dirt and moisture from entering the venting tube. The watertight diaphragm also oers a reliable protection for the submersible pressure transmitter.
Ordering information
Model / Measuring range / Cable length / Accessories
14052336
14052341
14052339
14052344
WIKA data sheet PE 81.55 ∙ 01/2016Page 4 of 5
Accessories
7.67" [194.82mm]
4.45" [113.03mm]
0.86" [21.84mm]
Dimensions in inches (mm)
0.28”
(7 mm)
3.70”
(94 mm)
0.43”
(11 mm)
5.12”
(130 mm)
6.89”
(175 mm)
Vent tube lter
Part# 7193131
The optional Teon
®
vent
tube lter protects the
vent opening and protects against the entry of dirt and moisture.
Cable junction box
Part# 2459686
The cable junction box is rated NEMA 4 / IP 67 and
is suitable for mounting outside tanks or shafts or
inside dry control boxes.
Can be wall or DIN rail mounted.
Cable clamp
Part# 2074257 The cable clamp secures the cable without bending or kinking that can damage the cable vent tube or outer jacket.
Desiccant drying cartridge part # 9836700 The desiccant drying cartridge helps prevent moisture buildup inside the vent tube.
7.67” (195mm)
(22mm)
4.45” (113mm)
NPT adapter Part# 1631322 The 316 SS G1/2 adapter replaces the removable protective cap
and converts the threads to 1/2”NPT male external, 1/4” female
internal threads. Includes O-ring.
O-ring
1.23” (31mm)
1.81” (46mm)
1/2” NPT male
1/4” NPT female
internal
Conduit adapter Part# 50476114 316 SS 1/2” NPT male cable conduit adapter. Must be factory installed.
1/2 NPT
27mm
between
ats
Additional weight
Part# 1524399 The additional weight replaces the protective
LevelGuard Anti-clog attachment Part # 50077091
cap and helps to stabilize the transmitter in turbulent conditions. Weight: approximately 1.1 lb,
5.12”
(130 mm)
© 2016 WIKA Alexander Wiegand SE & Co. KG, all rights reserved. The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
316 SS.
1.85” (47mm)
1.17”
29.7mm
The stainless steel LevelGuard attachment must be factory installed and calibrated.
WIKA data sheet PE 81.55 ∙ 01/2016Page 5 of 5
WIKA Instrument, LP
1000 Wiegand Blvd. Lawrenceville, GA 30043 Tel: 888-WIKA-USA • 770-513-8200 Fax: 770-338-5118 info@wika.com www.wika.com
01/2016 US