2Indicating and adjustment module for IPT-1* sensors
30
30
31
31
32
33
31549-EN-081211
1 About this document
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important
instructions for maintenance and fault rectification. Please read this
information before putting the instrument into operation and keep this
manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained qualified
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can
result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors3
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must be
carried out only by trained specialist personnel authorised by the plant
operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
The pluggable indicating and adjustment module is used for remote
measured value indication and parameter adjustment for IPT-1*
pressure transmitters.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to system
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of
standard regulations and guidelines. The user must take note of the
safety instructions in this operating instructions manual, the countryspecific installation standards as well as all prevailing safety
regulations and accident prevention rules.
The instrument must only be operated in a technically flawless and
reliable condition. The operator is responsible for trouble-free
operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the required occupational safety measures with the
current valid rules and regulations and also take note of new
regulations.
2.5 Safety approval markings and safety tips
The safety approval markings and safety tips on the device must be
observed.
4Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
2 For your safety
2.7 Compatibility according to NAMUR NE 53
With respect to compatibility, NAMUR recommendation NE 53 is met.
The parameter adjustment of the basic sensor functions is indepen-
dent of the software version. The range of available functions depends
on the respective software version of the individual components.
2.8 Safety instructions for Ex areas
Please note the Ex-specific safety information for installation and
operation in Ex areas. These safety instructions are part of the
operating instructions manual and come with the Ex-approved
instruments.
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors5
1
2
3 Product description
3 Product description
3.1 Configuration
Scope of delivery
Equipment
The scope of delivery encompasses:
l Indicating and adjustment module
l Documentation
- this operating instructions manual
The indicating and adjustment module is equipped with a display with
full-dot matrix as well as four keys for adjustment. An integrated
backlight can be switched via the adjustment menu. The following
hardware versions are necessary:
l Indicating and adjustment module …- 01 or higher
l Sensor electronics 4 … 20 mA …- 02 or higher
l Sensor electronics P rofibus PA or Foundation Fieldbus …- 03 or
higher
l
Fig. 1: Indicating and adjustment module
1Display
2Keys
6Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
1
2
Fig. 2: Rear of the indicating/adjustment module
1Integrated seal ring
2Gold plated contact paths
3.2 Principle of operation
3 Product description
Application area
The indicating and adjustment module is used for measured value
indication, adjustment, and diagnostics for the following WIKA
sensors:
l IPT-10 vers. 2.0 (ceramic sensor)
l IPT-1* vers. 3.0 (metallic sensor)
l IPT-11 vers. 4.0 (ceramic sensor)
The indicating and adjustment module is mounted into the respective
sensor housing or into the external housing. After mounting, the sensor
and the external housing are also splash water proof without housing
cover.
The operation of two indicating and adjustment modules in parallel in
the sensor and in the external housing is not supported.
Power supply
Power supply directly via the respective sensor or the external
housing. An additional connection is not necessary.
3.3 Operation
The adjustment is carried out via the integrated keys. The entered
parameters are generally saved in the respective sensor. A copy
function enables loading of the parameters into the indicating and
adjustment module.
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors7
®
3 Product description
3.4 Packaging, transport and storage
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE foam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l Not in the open
l Dry and dust free
l Not exposed to corrosive media
l Protected against solar radiation
l Avoiding mechanical shock and vibration
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 … 85 %
8Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
4 Mounting
4.1 Mounting steps
4 Mounting
Mount/Dismount indicating and adjustment
module
The indicating and adjustment module can be inserted or removed at
any time. It is not necessary to interrupt the voltage supply.
For mounting, proceed as follows:
1 Unscrew the housing cover
2 Place the indicating/adjustment module in the requested position
on the electronics
Information:
Four different positions are possible, each displaced by 90°.
Fig. 3: Mounting the indicating and adjustment module
3 Press the indicating/adjustment module lightly onto the electronics
and turn it to the right until it snaps in
4 Screw housing cover with inspection window tightly back on
Note:
If you intend to retrofit the instrument with an indicating and adjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors9
4 Mounting
Dismounting is carried out in reverse order.
10Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
5 Set up
1.1
2
3
1
5.1 Adjustment system
Fig. 4: Indicating and adjustment elements
1LC display
2Indication of the menu item number
3Adjustment keys
5 Set up
Key functions
l [OK] key:
- Move to the menu overview
- Confirm selected menu
- Edit parameter
- Save value
l [->] key to select:
- menu change
- list entry
- Select editing position
l [+] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- jump to the next higher menu
Adjustment system
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors11
The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not confirmed with [OK] will not be sa ved.
5 Set up
5.2 General functions
Introduction
Measured value indica-
tion
Menu overview
IPT-1* pressure transmitters have various functions. Hence they can
be adapted to the respective application.
Some of these functions are sensor-specific. These are described in
the operating instructions manual of the corresponding sensor. Other
functions, however, have a general character, i.e. they are available in
different sensors.
The general functions are described in this paragraph. The functions of
the indicating/adjustment module are determined by the sensor and
correspond to the respective software version of the sensor.
The following presentations are available in the measured value
display:
l Level or pressure as digital value, sensor-TAG
l Level or pressure as digital value and bar graph, sensor TAG
l Level or pressure as digital value, temperature value
With [->] you select the different presentations of the measured value.
You can reach the menu overview from any presentation with [OK].
With [ESC] you return from the menu overview to the measured value
display.
In the menu overview you select the appropriate menu with [->] and
open it with [OK]. Then the individual menu items are available.
▶
Basic adjustment
Display
Diagnostics
Service
Info
Menu section, basic adjustment
Damping
To damp process-dependent measured value fluctuations, you have to
set an integration time of 0 … 999 s in this menu item.
Depending on the sensor type, the factory setting is 0 s or 1 s.
Damping
0 s
Linearisation curve
In this menu item you select the linearization curve:
l linear
l Cylindrical tank
l Spherical tank
l User programmable
User programmable means: Switching on a linearization curve
programmed via PC and PACTware
12Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
5 Set up
The linearization curve creates a correlation between height and
volume. It takes into account the vessel geometry for the displayed
measured value and current output.
Factory setting is linear.
Linearisation curve
linear
Edit sensor TAG
Lighting
In the menu item "Sensor-TAG" you edit a 12-digit measurement loop
name. An unambiguous designation can hence be assigned to the
sensor, e.g. the measurement loop name or the tank or product
designation. In digital systems and in the documentation of larger
plants, a singular designation should be entered for exact identification
of individual measuring sites.
The available digits comprise:
l Letters from A … Z
l Numbers from 0 … 9
l Special characters +, -, /, -
Factory setting is "Sensor".
Sensor-TAG
Sensor
Menu section, display
An integrated background lighting can be switched via the adjustment
menu. The following version is necessary:
l Indicating and adjustment module …- 01 or higher
l Sensor electronics 4 … 20 mA …- 01 or higher
l Sensor electronics pressure transmitter 4 … 20 mA …- 02 or
higher
l Sensor electronics P rofibus PA or Foundation Fieldbus …- 03 or
higher
The version is stated on the type label of the indicating and adjustment
module or the sensor electronics. The function depends also on the
height of the supply voltage, see operating instructions manual of the
respective sensor.
In the default setting, the lightning is switched off.
Pointer
Min. and max. measured values are saved in the sensor. The values
are displayed in the menu item "Peak values".
l Min. and max. pressure
1)
Pressure: -50 … +150 % of the nominal measuring range.
1)
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors13
5 Set up
Sensor status
Curve presentation
l Min. and max. temperature
Pointer
2)
In this menu item, the device status is displayed. If the sensor detects
a fault, "OK" will be displayed. If a fault is detected, a flashing failure
message is outputted sensor-specifically, e.g. "E013". The fault is also
displayed in clear text, e.g. "No measurement value".
Information:
The fault message as well as the clear text indication are also carried
out in the measured value display.
Meas. reliability
Sensor status
Available with the trend curve:
l "X-Zoom": Resolution in minutes, hours or days
l "Stop/Start": Interrupt a recording or start a new recording
l "Unzoom": Reset the resolution to minutes
Trend curve
Menu section, service
Simulation of
measured values
In this menu item you simulate a user-defined level or pressure value
via the current output. This allows you to test the signal path, e.g.
through connected indicating instruments or the input card of the
control system.
The following simulation variables are available:
l Percent
l Current
l Pressure
With Profibus PA sensors, the selection of the simulated value is made
via the "Channel" in the menu "Basic adjustments".
How to start the simulation:
2)
Temperature: -50 … +150 °C.
14Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
5 Set up
1 Push [OK]
2 Select the requested simulation variable with [->] and confirm with
[OK]
3 Set the requested value with [+] and [->].
4 Push [OK]
The simulation is now running, with 4 … 20 mA/HART a current is
outputted and with Profibus PA or Foundation Fieldbus a digital value.
How to interrupt the simulation:
à Push [ESC]
Information:
The simulation is terminated automatically 10 minutes after the last key
has been pushed.
Simulation
Start simulation?
Reset
Unit of measurement
With the reset function, modified values are reset. Three subfunctions
are available:
l Basic adjustment
- Reset the values modified with the indicating and adjustment
module to the basic setting
l Factory setting
- As basic adjustment, but also reset of special parameters to
the delivery status
l Peak values measured value and temperature
- Reset the min./max. values of pressure and temperature to the
current values
Reset
Select reset?
For pressure transmitters more comprehensive units are available.
They are described in the operating instructions manual of the
respective sensor in the menu "Basic adjustments".
Unit of measurement
bar
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors15
5 Set up
Language
Copy sensor data
The sensor is already set to the ordered national language. In this
menu item you can change the language. The following languages are
available, e.g. in software version 3.50:
l Deutsch
l English
l Français
l Espanõl
l Pycckuu
l Japanese
l Italiano
l Netherlands
l Japanese
l Chinese
Language
Deutsch
With this function
l Load parameter adjustment data from the sensor into the
indicating and adjustment module
l Write p arameter adjustment data from the indicating and adjust-
ment module into the sensor
The data are permanently saved in an EEPROM memory in the
indicating and adjustment module and remain there even in case of
power failure. From there, they can be written in one or several
sensors or kept as backup for a probable sensor exchange.
Kind and volume of the copied data is stated in the operating
instructions manual of the respective sensor.
Information:
Before writing the data into the sensor, it is checked if the data fit the
sensor. If data do not fit, a fault signal is triggered or the function is
blocked. When writing data into the sensor, you will see from which
instrument type the data originate and which TAG-no. this sensor had.
The following items are checked:
l Software version
l WHG approval
l SIL activated
l Measuring principle
l Signal output
l Pressure measuring range
16Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
5 Set up
PIN
Info
In this menu item, the PIN is activated/deactivated permanently.
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modifications. If the PIN is activated
permanently, it can be deactivated temporarily (i.e. for approx. 60 min.)
in any menu item. The instrument is delivered with the PIN set to 0000.
PIN
Disable permanently?
Only the following functions are permitted with activated PIN:
l Select menu items and show data
l Read data from the sensor into the indicating/adjustment module.
Menu section, info
In this menu item the most important sensor information can be
displayed:
l Sensor type
l Serial number: 8-digit number, e.g. 12345678
Sensor type
Serial number
12345678
l Date of manufacture: Date of the factory calibration, e.g. 10.
January 2008
l Software version: Edition of the sensor software, e.g. 3.50
Date of manufacture
10. January 2008
Software version
3.50
l Date of last change using PC: Date of the last change of sensor
parameters via PC, e.g. 10. January 2008
Last change using PC
10. January 2008
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors17
5 Set up
l Sensor details, e.g. approval, process fitting, seal, measuring cell,
measuring range, electronics, housing, cable entry, plug, cable
length etc.
Sensor characteristics
Display now?
5.3 Special functions - 4 … 20 mA/HART
Introduction
Display
The 4 … 20 mA/HART special functions are briefly described in this
paragraph. The respective range of functions of the indicating and
adjustment module is determined by the sensor and the sensor
software version.
In the menu item "Display" you can define how the measured value
should be presented on the display.
The following indication values are available:
l Height
l Pressure (only with pressure transmitters)
l Current
l Scaled
l Percent
l Lin. percent
l Temperature
The selection "scaled" opens the menu items "Display unit" and
"Scaling". In "Display unit" there are the following options:
l Height
l Mass
l Flow
l Volume
l Without unit
Depending on selection, the different units are in turn available.
In the menu item "Scaling", the requested numerical value with
decimal point is entered for 0 % and 100 % of the measured value.
There is the following relation between the indication value in the menu
"Display" and the adjustment unit in the menu "Basic adjustment":
l With pressure, displayed value "Pressure" or " Height" means:
presentation of the measured value in the selected adjustment
unit, e.g. bar or m.
Displayed value
31549-EN-081211
Scaled
18Indicating and adjustment module for IPT-1* sensors
Current output
5 Set up
Display unit
Volume
l
Scaling
0 % = 0.0 l
100 % = 100.0 l
Menu section, service
In the menu item "Current output" you determine the behaviour of the
current output during operation and in case of failure. The following
options are available:
Current output
Characteristics curve4 … 20 mA
Failure mode
Min. current
Max. current
3)
4)
5)
20 … 4 mA
Hold value
20.5 mA
22 mA
< 3.6 mA
3.8 mA
4 mA
20 mA
20.5 mA
The values in bold font represent the data of the factory setting.
In HART multidrop mode, the current is constantly 4 mA. This value
does not change even in case of failure.
Current output
Functional safety (SIL)
Output mode: 4-20 mA
Failure mode: 22 mA
Min. current 3 .8 mA
The functional safety is already activated Ex factory for instruments
▼
▼
▼
with SIL qualification. For instruments Ex factory without SIL
qualification, the functional safety must be activated by the user f or
applications according to SIL via the indicating and adjustment
module. The SIL factory setting cannot be deactivated by the user.
The activation of SIL has the following impact:
3)
Value of the current output in case of failure, e.g. if no valid measured value
is delivered.
4)
This value is not underrun during operation.
5)
This value is not exceeded during operation.
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors19
5 Set up
l In the menu item "Failure mode" under "Current output", the
parameters "Hold value" and "20.5 mA" are blocked
l In the menu item "HART mode", the function "Multidrop" is blocked
Note:
For such applications, it is absolutely necessary to take note of "Safety
Manual".
HART mode
Introduction
Sensor address
HART offers standard and multidrop mode.
The mode standard with the fixed address 0 means output of themeasured value as 4 … 20 mA signal.
In Multidrop mode, up to 15 sensors can be operated on one two-wire
cable. An address between 1 and 126 must be assigned to each
6)
sensor.
In this menu item you determine the HART mode and enter the
address for multidrop.
HART mode
Standard
Address 0
The default setting is standard with address 0.
5.4 Special functions - Profibus PA
The Profibus PA special functions are briefly described in this
paragraph. The respective range of functions of the indicating and
adjustment module is determined by the sensor and the sensor
software revision.
Menu section, basic adjustment
Level and pressure sensors operate as slaves on the Profibus PA. To
be identified as a bus participant, each sensor must have a unique
address. Each instrument is delivered with address 126. With this
address, it can at first be connected to an existing bus. However, the
address must be changed. This can be done in this menu item.
Sensor address
126
6)
The 4 … 20 mA signal of the HART sensor is switched off. The sensor
consumes a constant current of 4 mA. The measuring signal is transmitted
exclusively as digital HART signal.
20Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
5 Set up
Channel
Display
The channel is the input selector switch for function block (FB) of the
sensor. Within the function block, additional scalings (Out-Scale) are
carried out. In this menu item, the value fir the function block is
selected:
l SV1 (Secondary Value 1):
- Pressure or height
l SV2 (Secondary Value 2):
- Percent
l PV (Primary Value):
- Linearised percentage value
Channel
PV lin. value
A pressure transmitter delivers the following measured values:
l SV1 (Secondary Value 1): P ressure or height value before
adjustment
l SV2 (Secondary Value 2): P ercentage value after the adjustment
l PV (Primary Value): Linearised percentage value
l PA-Out (value after passing the function block): PA output
l Temperature
In the menu item "Display" you can define which measured value
should be presented on the display.
Displayed value
PA-Out
Additional PA value
Profibus transmits two values cyclically. The first value is determined in
the menu item "Channel". The additional cyclical value is selected in
the menu item "Additional PA value".
With pressure transmitters the following values are available:
l SV1 (Secondary Value 1): P ressure or height value before
adjustment
l SV2 (Secondary Value 2): P ercentage value after the adjustment
l PV (Primary Value): Linearised percentage value
Additional PA value
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors21
5 Set up
Determine Out-Scale
Here, you determine the unit and scaling for PA-Out. These settings
also apply to the values displayed on the indicating and adjustment
module if in the menu item "Displayed value" PA-Out was selected.
In the menu item "PV-Out-Scale", the requested numerical value with
decimal point is entered for 0 % and 100 % of the measured value.
Out-Scale-Unit
PV-Out-Scale
7)
7)
The following display values are available in "Out-Scale-Unit": Pressure,
height, mass, flow, volume, others (without unit, %, mA).
22Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
Basic adjustment
P
▶
Basic adjustment
Display
Diagnostics
Service
Info
5 Set up
5.5 Menu schematic
Information:
Depending on the version and application, the highlighted menu
windows are not always available.
1
Unit
Unit of measurement
Temperature unit°C
▼
Damping
Display
Basic adjustment
▶
Display
Diagnostics
Service
Info
Displayed value
Pressure
Lighting
Switched off
bar
2
Position correction
Offset
=
0.2 mbar
0000 mbar
Linearisation curve
linear
▼
Displayed value
Scaled
1.2
1.6
▼
2.1
1.1
▼
1.5
1 s
2.1
▼
2.4
Min. adjustment
000.0 %
=
0.0 mbar
0.0 mbar
Sensor-TAG
Sensor
Display unit
Volume
▼
l
▼
1.3
1.7
2.2
Max. adjustment
100.00 %
=
100.00 mbar
0.0 mbar
Scaling
0 % = 0.0
l
100 % = 100.0
l
1.4
2.3
▼
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors23
5 Set up
Diagnostics
Basic adjustment
Display
▶
Diagnostics
Service
Info
3
Pointer
p-min.: -5.8 mbar
p-max.: 167.5 mbar
T-min.: -12.5 °C
T-max.: +85.5 °C
Service
Basic adjustment
Display
Diagnostics
▶
Service
Info
Current output
Output mode: 4-20 mA
Fail.mode: < 3.6 mA
Min. current: 3.8 mA
max. current: 20.5 mA
SIL
Deactivated!
Application
▼
Level
3.1
4.1
▼
Sensor status
OK
4
Simulation
Start simulation
3.2
4.2
▼
Trend curve
Start trend curve?
Reset
Select reset
3.3.1
4.3
▼
Language
Deutsch
4.4
▼
▼
▼
▼
4.5
▼
4.9
HART mode
Standard
Address 0
4.6
Copy sensor data
Copy sensor data?
4.7
PIN
4.8
Enable?
Info
Basic adjustment
Display
Diagnostics
Service
▶
Info
Sensor type
Serial number
12345678
5.1
5
Date of manufacture
e.g. 16. May 2008Software version
e.g. 3.50
5.2
Last change using PC
e.g. 16. May 2008
5.3
Sensor characteristics
Display now?
5.4
24Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
5 Set up
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors25
P
5 Set up
Basic adjustment
▶
Basic adjustment
Display
Diagnostics
Service
Info
5.6 Menu schematic
Information:
Depending on the version and application, the highlighted menu
windows are not always available.
1
Sensor address
126
Max. adjustment
100.00 %
=
100.00 mbar
0.0 mbar
Display
Basic adjustment
▶
Display
Diagnostics
Service
Info
Displayed value
PA-Out
Diagnostics
Basic adjustment
Display
▶
Diagnostics
Service
Info
1.1
1.4
2.1
2
3
Unit
Unit of measurement
bar
▼
Temperature unit°C
▼
Damping
0 s
Lighting
Switched off
▼
1.1
1.5
2.4
Position correction
Offset
=
0.2 mbar
0000 mbar
Linearisation curve
linear
▼
1.2
1.6
Min. adjustment
000.0 %
=
0.0 mbar
0.0 mbar
Sensor-TAG
Sensor
1.3
1.7
Pointer
p-min.: -5.8 mbar
p-max.: 167.5 mbar
T-min.: -12.5 °C
T-max.: +85.5 °C
3.1
Sensor status
OK
3.2
Trend curve
Start trend curve?
3.3.1
26Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
Service
Basic adjustment
Display
Diagnostics
▶
Service
Info
5 Set up
4
Additional PA value
Secondary Value 1
Reset
Select reset
Application
Level
▼
Info
Basic adjustment
Display
Diagnostics
Service
▶
Info
Sensor type
Serial number
12345678
5
Out-Scale-Unit
Volume
l
Language
Deutsch
Date of manufacture
e.g. 16. May 2008Software version
e.g. 3.50
4.2
4.6
5.2
4.1
4.3
▼
4.9
5.1
PA-Out-Scale
100.00 lin %
= 0.0 l
0.00 lin %
= 100.0 l
Copy sensor data
Copy sensor data?
Last change using PC
e.g. 16. May 2008
4.3
4.7
5.3
Simulation
Start simulation
PIN
Enable?
Sensor characteristics
Display now?
4.2
▼
4.8
5.4
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors27
P
5 Set up
Basic adjustment
▶
Basic adjustment
Display
Diagnostics
Service
Info
5.7 Menu schematic
Information:
Depending on the version and application, the highlighted menu
windows are not always available.
1
Unit
Unit of measurement
bar
Temperature unit
°C
▼
Damping
Display
Basic adjustment
▶
Display
Diagnostics
Service
Info
Displayed value
AI-Out
Diagnostics
Basic adjustment
Display
▶
Diagnostics
Service
Info
2
3
Position correction
Offset
=
0.2 mbar
0000 mbar
Linearisation curve
linear
▼
Lighting
Switched off
▼
1.2
1.6
2.4
1.1
▼
1.5
0 s
2.1
Min. adjustment
000.0 %
=
0.0 mbar
0.0 mbar
Sensor-TAG
Sensor
1.3
1.7
Max. adjustment
100.00 %
=
100.00 mbar
0.0 mbar
1.4
Pointer
p-min.: -5.8 mbar
p-max.: 167.5 mbar
T-min.: -12.5 °C
T-max.: +85.5 °C
3.1
Sensor status
OK
3.2
Trend curve
Start trend curve?
3.3.1
28Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
Service
Basic adjustment
Display
Diagnostics
▶
Service
Info
5 Set up
4
Simulation
Start simulation
PIN
Info
Basic adjustment
Display
Diagnostics
Service
▶
Info
Device-ID
Sensor-TAG
Sensor characteristics
Display now?
Enable?
4.2
▼
4.5
5.1
5.4
Reset
Select reset
Application
Level
5
Sensor type
Serial number
12345678
4.3
▼
4.6
5.2
Language
Deutsch
Date of manufacture
e.g. 16. May 2008Software version
e.g. 3.50
4.3
5.2
Copy sensor data
Copy sensor data?
Last change using PC
e.g. 16. May 2008
4.4
5.3
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors29
6 Maintenance and fault rectification
6 Maintenance and fault rectification
6.1 Maintenance
When used in the correct way, no special maintenance is required in
normal operation.
6.2 Instrument repair
You can download a return form (24 KB) in the Internet from our
homepage
If a repair is necessary, please proceed as follows:
l Print and fill out one form per instrument
l If necessary, state a contamination
l Clean the instrument and pack it damage-proof
l Attach the completed form and probably a safety data sheet to the
instrument
l Please contact the agency serving you for the address of the
return shipment
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
www.wika.com under the item "Service".
30Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
7 Dismounting
7 Dismounting
7.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
7.2 Disposal
The indicating and adjustment module consists of materials which can
recycled by specialised recycling companies. We have purposely
designed the components to be easily separable.
WEEE directive 2002/96/EG
This indicating and adjustment module is not subject to the WEEE
directive 2002/96/EG and the respective national laws (in Germany, e.
g. ElektroG). Pass the indicating and adjustment module directly on to
a specialised recycling company and do not use the municipal
collecting points. They may only be used for privately used products
according to the WEEE directive.
Correct disposal avoids negative effects to persons and environment
and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no possibility to dispose of the old instrument
professionally, please contact us concerning return and disposal.
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors31
8 Supplement
8 Supplement
8.1 Technical data
General data
Weightapprox. 150 g (0.33 lbs)
Ambient conditions
Ambient temperature-15 … +70 °C (+5 … +158 °F)
Storage and transport temperature-40 … +80 °C (-40 … +176 °F)
Indicating and adjustment module
Voltage supply and data transmissionthrough the sensor
IndicationLC display in dot matrix
Adjustment elements4 keys
Protection
- unassembledIP 20
- mounted into the sensor without coverIP 40
Materials
- HousingABS
- Inspection windowPolyester foil
32Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
8.2 Dimensions
27,6mm
(1
3
/
32
")
ø66,3mm
(2
39
/
64
")
45,1mm
(1
25
/
32
")
9,7mm
(
3
/
8
")
Indicating and adjustment module
8 Supplement
Fig. 5: Indicating and adjustment module
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors33
8 Supplement
34Indicating and adjustment module for IPT-1* sensors
31549-EN-081211
8 Supplement
31549-EN-081211
Indicating and adjustment module for IPT-1* sensors35