ist eine geschützte Marke in verschiedenen Ländern.
23 - 42
43 - 62
63 - 82
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d'emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!
2
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
Contents
Contents
1. General information 4
2. Safety 5
3. Specications8
4. Design and function 10
5. Transport, packaging and storage 10
6. Commissioning, operation 11
7. Adjustment of zero point and span 14
8. Spare part kit 15
9. Maintenance and cleaning 18
10. Faults 19
GB
11. Dismounting, return and disposal 20
12. Accessories 21
Appendix 1: Declaration of conformity for model HP-2 22
Declarations of conformity can be found online at www.wika.com.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
3
1. General information
1. General information
■
The pressure transmitter described in the operating instructions has been designed
and manufactured using state-of-the-art technology. All components are subject
GB
to stringent quality and environmental criteria during production. Our management
systems are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are
observed.
■
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused by using the
product contrary to its intended use, non-compliance with these operating instructions, assignment of insuciently qualied skilled personnel or unauthorised modications to the instrument.
■
The general terms and conditions contained in the sales documentation shall apply.
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
1. General information / 2. Safety
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to equipment or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for ecient and
trouble-free operation.
WARNING!
... indicates a potentially dangerous situation that can result in burns, caused
by hot surfaces or liquids, if not avoided.
Abbreviations
2-wireThe two connection lines are used for the voltage supply.
The measurement signal also provides the supply current.
3-wireTwo connection lines are used for the voltage supply.
One connection line is used for the measurement signal.
U
, +5VPositive power supply
B
0V, GNDReference potential
S+Positive output terminal
D+ ,D-Data link USB 2.0
GB
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the appropriate pressure transmitter has been selected in terms of measuring range,
design and specic measuring conditions.
The operator must ensure the compatibility of the uid with the material
used. Where the medium is hydrogen, contact the manufacturer.
Non-observance can result in serious injury and/or damage to the equipment.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
5
2. Safety
WARNING!
■
Open the connections only after the system has been depressurised.
■
Observe the working conditions in accordance with chapter 3 “Specications”.
■
GB
2.1 Intended use
The pressure transmitter is used to convert pressure into an electrical signal.
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The technical specications contained in these operating instructions must be
observed. Improper handling or operation of the instrument outside of its techni-
cal specications requires the instrument to be taken out of service immediately and
inspected by an authorised WIKA service engineer.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
Always operate the pressure transmitter within the overpressure limit.
Further important safety instructions can be found in the individual chapters
of these operating instructions.
2.2 Personnelqualication
WARNING!
Riskofinjuryshouldqualicationbeinsucient!
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried
out by skilled personnel who have the qualications described below.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowl-
edge of country-specic regulations, current standards and directives, are capable of
carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
6
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
2. Safety
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene, ammable or toxic gases
or liquids, and refrigeration plants, compressors, etc., in addition to all standard regulations, the appropriate existing codes or regulations must also be
followed.
WARNING!
No plant conditions are permitted that could lead to the formation of atomic
hydrogen in the connection channel of the pressure transmitter.
WARNING!
Residual media in the dismounted pressure transmitter can result in a risk to
persons, the environment and equipment.
Take sucient precautionary measures.
Do not use this instrument in safety or emergency stop devices. Incorrect use
of the instrument can result in injury.
Should a failure occur, aggressive media with extremely high temperature and
under high pressure or vacuum may be present at the instrument.
2.4 Labelling, safety marking
GB
Product label
Pin assignment
P# Product no.
S# Serial no.
Accuracy
Measuring range
Output signal
Power supply
If the serial number becomes illegible due to mechanical damage or overpainting, traceability will no longer be possible.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
7
2.Safety/3.Specications
Explanation of symbols
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant European directives.
GB
3. Specications
WARNING!
When designing the system, please note that the values given (e.g. burst
pressure, overpressure limit) are dependent upon the material and thread
used.
On request, since the service life depends on the actual pressure prole.
Despite its excellent load cycle stability a permanent load cycle stability is only given
conditionally. This is due to the extreme loadings faced by the highly pressurised
materials. This applies particularly to dynamic applications. Accordingly, the pressureloaded parts of the HP-2 are considered as wear parts that are not covered by the
warranty.
8
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
3.Specications
Output signals
Signal typeSignal
Current (2-wire)4 ... 20 mA
Voltage (3-wire)DC 0 ... 5 V
USBUSB 2.0
Power supply
The power supply depends on the selected output signal
■
4 ... 20 mA:DC 10 ... 30 V
■
DC 0 ... 5 V: DC 10 ... 30 V
■
DC 0 ... 10 V: DC 14 ... 30 V
■
USB 2.0:DC 5 V
LoadinΩ
■
Current output (2-wire): ≤ (power supply - 10 V) / 0.02 A
■
Voltage output (3-wire): > maximum output signal / 1 mA
Ingress protection (per IEC 60529)
■
Angular connector DIN 175301-803 A: IP 65
■
Circular connector M12 x 1 (4-pin):IP 67
■
USB connector type A:Instrument: IP 67, connector: IP 20
■
Cable outlet:IP 67
DC 0 ... 10 V
GB
The stated ingress protection only applies when plugged in using mating connectors
that have the appropriate ingress protection.
Permissible temperature ranges
■
Medium:0 ... +80 °C
■
Ambient: -20 ... +80 °C
■
Storage: -40 ... +85 °C
CE conformity
■
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and immunity
(industrial application)
With USB signal output and strong electromagnetic disturbances in the frequency
range 650 … 750 MHz, the accuracy can be inuenced by up to 1 %.
■
Pressure equipment directive 97/23/EC
RoHS conformity
Yes
For special model numbers, e.g. HP-20000, please note the specications stated on the
delivery note.
For further specications see WIKA data sheet PE 81.53 and the order documentation.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
9
4. Design and function / 5. Transport, packaging and ...
4. Design and function
4.1 Short description
The prevailing pressure is measured at the sensor element through the deformation of
a diaphragm. By supplying power, this deformation of the diaphragm is converted into
GB
an electrical signal. The output signal from the pressure transmitter is amplied and
standardised.
4.2 Scope of delivery
Cross-check scope of delivery with delivery note.
5. Transport, packaging and storage
5.1 Transport
Check the pressure transmitter for any damage that may have been caused during
transportation.
Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
see chapter 3 “Specications”
Store the pressure transmitter in its original packaging in a location that fulls the condi-
tions listed above.
WARNING!
Before storing the instrument (following operation), remove any residual
media. This is of particular importance if the medium is hazardous to health,
e.g. caustic, toxic, carcinogenic, radioactive, etc.
10
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
6. Commissioning, operation
6. Commissioning, operation
CAUTION!
Prior to commissioning, the pressure transmitter must be subjected to a
visual inspection.
Only use the pressure transmitter if it is in perfect condition with respect to
safety.
6.1 Mechanical mounting
■
The sealing faces at the instrument have to be
undamaged and clean.
■
When screwing the instrument in, the force
required to do this must not be applied through
the case, but only through the spanner ats
provided for this purpose and using a suitable
tool.
For the installation point, the valid values for
tightening torque and maximum pressure
should be taken from the high-pressure pipe
supplier's documentation.
GB
Open-ended spanner
(spanner width 27)
Spanner ats
■
When screwing in, do not cross the threads.
■
For information on tapped holes and welding sockets, see Technical information
IN 00.14 at www.wika.com.
6.2 Electrical mounting
■
The instrument must be grounded via the process connection.
■
Use the pressure transmitter with shielded cable, and ground the shield on at least
one end of the lead, if the lines are longer than 30 m (3- or 4-wire) or leave the building.
■
Select a cable diameter that matches the cable gland of the plug. Make sure that
the cable gland of the mounted plug has a tight t and that the seals are present and
undamaged. Tighten the threaded connection and check that the seal is correctly
seated, in order to ensure the ingress protection.
■
For cable outlets, make sure that no moisture enters at the cable end.
■
Observe the specications in accordance with chapter 3 “Specications”.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
Angular connector DIN 175301-803 Amax. 1.5 mm
Circular connector M12 x 1 (4-pin)---
GB
USB connector type A--2 m
Cable outlet0.5 mm
2
2
(AWG 20)6.8 mm1.5 m
6 ... 8 mm-
Connection diagrams
For abbreviations see page 5
Angular connector DIN 175301-803 A
2-wire3-wire
UB11
0V22
S+-3
Cable outlet
2-wire3-wire
UBbrownbrown
0Vgreengreen
S+-white
Circular connector M12 x 1 (4-pin)
2-wire3-wire
4
1
0V33
2
S+-4
UB11
3
USB connector type A
lengths
+5V1
GND4
D+3
D-2
12
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
6. Commissioning, operation
Fitting the DIN 175301-803 angular connector
1.
Loosen the screw .
2.
Loosen the cable gland .
3.
Pull the angled socket + from the instrument.
4.
Via the mounting hole , lever the terminal block
out of the case .
5. Pass the cable with the appropriate cable outer
diameter through the cable gland , ring ,
sealing and the case .
6. Connect the cable ends to the connection terminals on the terminal block in accordance with the
pin assignment (see “Connection diagrams” for the
pin assignment).
7.
Press the terminal block into the case .
8.
Tighten the cable gland around the cable.
Make sure that the cable gland and seal are not
damaged and that they are assembled correctly in
order to ensure ingress protection.
Mounting hole
GB
9.
Place the at, square gasket over the pressure
transmitter's connection pins.
10.
Slide the assembled angled socket + onto
the pressure transmitter's connection pins.
11.
Using the screw , screw the angled socket to the
pressure transmitter, hand-tight.
6.3 Commissioning of USB instruments
■
For installing the driver, administrator rights are required.
■
Connect the USB connector to a USB 2.0 port at your computer.
■
Install the driver via the InstallWizard of the product software.
■
For further operation the P-3x data logger software is available (for details see
instruction manual P-3x data logger).
■
Details on the interface protocol or the DLL (Dynamic Link Library) are available on
the software CD and in the download section at www.wika.com.
For the model HP-2, model P-30 software must be used. All les and
documents are available for download at www.wika.com.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
13
7. Adjustment of zero point and span
7. Adjustment of zero point and span
Only adjust the span-setting potentiometer if calibration equipment is avail-
GB
7.1 Preparation(gureA)
To gain access to the potentiometers, open the instrument as follows:
■
Disconnect the electrical connection from the
instrument.
■
Remove the clamping nut .
■
Carefully pull the instrument connector from the
instrument.
■
Connect the instrument connector to the power
supply and a display unit (e.g. ammeter, voltmeter)
according to the connection diagram.
7.2 Adjustmentofzeropoint(gureB)
■
Go to the start of the measuring range.
■
Using potentiometer “Z”, adjust the minimum output
signal (e.g. 4 mA).
7.3 Settingthespan(gureB)
■
Go to the end of the measuring range.
■
Using potentiometer “S”, adjust the maximum output
signal (e.g. 20 mA).
■
Check the zero point and if there is any deviation,
re-adjust it.
■
Repeat the procedure until the zero point and the
span are set correctly.
able which has at least three times the accuracy of the pressure transmitter.
A
B
7.4 Finishtheadjustment(gureA)
■
Disconnect the instrument connector from the power supply and the display unit.
■
Carefully push the instrument connector onto the instrument, without damaging the
wires or the seals. The seals must be clean and undamaged in order to guarantee the
given ingress protection.
■
Tighten the clamping nut .
After the adjustment, check that the system is functioning correctly.
Recommended recalibration cycle: Half-yearly
14
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
8. Spare part kit
8. Spare part kit
The spare part kit is used for the exchange of process connections and the bae
plate of the models HP-2-D and HP-2-E. It is not compatible with the standard version,
model HP-2-S.
8.1 Contents of the spare part kit
Clamping jawExchangeable process connection
BaeplateforDIPSorEPCO-ring
DIPSEPC
GB
8.2 Tools required
Vice
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
Torque spanner
Tweezers
15
8. Spare part kit
8.3 Mounting
GB
Secure the clamping jaw in the vice as shown, so that
this cannot come loose.
Slide the pressure transmitter between the clamping
jaw, using the groove milled into the process connection.
Loosen the process connection using the torque
spanner.
16
Dry the bae plate incl. the
O-ring using compressed air.
For access, there is a bore
on the process connection.
Bae plate
DIPSEPC
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
8. Spare part kit
Remove the dry bae plate incl. the O-ring using the
tweezers.
GB
Check the sealing cone for damage.
If the sealing cone is damaged, the entire instrument
must be replaced.
Insert the new bae plate with O-ring at an angle, and press down, taking care not to
damage the surface of the bae plate and the O-ring.
DIPSEPC
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
17
8. Spare part kit / 9. Maintenance and cleaning
Tighten the exchangeable
process connection using the
torque spanner.
GB
On the torque spanner, set
the tightening torque to
130 Nm.
9. Maintenance and cleaning
9.1 Maintenance
Repairs must only be carried out by the manufacturer.
When a leak is suspected, as a result of hairline cracks in the process connection, as
described in chapter 8 “Spare part kit”, replace it.
For the order number for the spare part kit see chapter 12 “Accessories”
9.2 Cleaning
CAUTION!
■
Before cleaning, correctly disconnect the pressure transmitter from the
pressure supply, switch it o and disconnect it from the voltage supply.
■
Clean the instrument with a moist cloth.
■
Electrical connections must not come into contact with moisture.
■
Wash or clean the dismounted instrument before returning it, in order to
protect persons and the environment from exposure to residual media.
■
Residual media in the dismounted pressure transmitter can result in a risk
to persons, the environment and equipment. Take sucient precautionary
measures.
18
For information on returning the instrument see chapter 11.2 “Return”.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
10. Faults
10. Faults
In the event of any faults, rst check whether the pressure transmitter is mounted
correctly, mechanically and electrically.
FaultsCausesMeasures
No output signalCable breakReplace connecting cable
No/wrong power supplyCorrect the power supply
No/wrong output signalWiring errorRectify the wiring in accordance
Constant output signal upon
change in pressure
Deviating output signalZero point change through
Signal span too highPressure spikes, cavitationsUse instrument with cavitation and
Signal span too smallWrong power supplyRectify the power supply
Signal span variesEMC interference sources in the
Leakage at the process connectionHairline crack in the pressure portConsult with manufacturer, if
Wrong power supplyReplace instrument
dynamic pressure
environment
Instrument not groundedGround the instrument
Strongly varying pressure of the
medium
with the connection diagram
Re-adjust zero point
pressure-spike protection
Shield the instrument and cables,
remove source of interference
Damping (consulting by the
manufacturer)
required, replace instrument.
GB
If complaint is unjustied, the handling costs will be charged.
CAUTION!
If faults cannot be eliminated by means of the measures listed above, shut
down the pressure transmitter immediately, and ensure that pressure and/or
signal are no longer present, and secure the instrument from being put back
into operation inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
11.2 “Return”.
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
19
11. Dismounting, return and disposal
11. Dismounting, return and disposal
WARNING!
GB
11.1 Dismounting
Only disconnect the pressure gauge once the system has been depressurised!
11.2 Return
Residual media in the dismounted pressure transmitter can result in a risk to
persons, the environment and equipment.
Take sucient precautionary measures.
WARNING!
Risk of burns!
Let the instrument cool down suciently before dismounting it!
During dismounting there is a risk of dangerously hot pressure media
escaping.
WARNING!
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous
substances (acids, bases, solutions, etc.).
When returning the instrument, use the original packaging or a suitable transport
package.
Information on returns can be found under the heading “Service” on our local
website.
11.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specic waste disposal regulations.
20
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
12. Accessories
12. Accessories
Accessories and spare parts
Mating connector
DesignationOrder number
without cablewith 2 m cablewith 5 m cable
Angular connector DIN 175301-803 A
■
with cable gland, metric114275671122579311250186
■
with cable gland, conduit11022485--
Circular connector M12 x 1, 4-pin
■
straight24212621125078011250259
■
angled24212701125079811250232
Sealings for mating connectors
DesignationOrder number
Angular connector DIN 175301-803 A1576240
Spare part kit for models HP-2-D and HP-2-E
Consisting of a replacement connection thread, replacement sealing disc and a mounting aid.
Spare part kit suitable for the following process connections:
Process connectionOrder number
Model HP-2-DModel HP-2-E
M16 x 1.5 female1403989514050403
M20 x 1.5 female1331992314050404
Software
The full software is available to download as freeware from the following path.
www.wika.com / Download / Software / Electronic pressure measurement
GB
11379180.03 09/2013 GB/D/F/E
WIKA operating instructions pressure transmitter, model HP-2
21
Appendix 1: Declaration of conformity for model HP-2
GB
22
WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
Inhalt
Inhalt
1. Allgemeines 24
2. Sicherheit 25
3. Technische Daten 28
4. Aufbau und Funktion 30
5. Transport, Verpackung und Lagerung 30
6. Inbetriebnahme, Betrieb 31
7. Einstellung Nullpunkt und Spanne 34
8. Ersatzteilkit 35
9. Wartung und Reinigung 38
10. Störungen 39
D
11. Demontage, Rücksendung und Entsorgung 40
12. Zubehör 41
Anlage 1: Konformitätserklärung Typ HP-2 42
Konformitätserklärungen nden Sie online unter www.wika.de
11379180.03 09/2013 GB/D/F/E
23WIKA Betriebsanleitung Druckmessumformer, Typ HP-2
1. Allgemeines
1. Allgemeines
■
Der in der Betriebsanleitung beschriebene Druckmessumformer wird nach dem
aktuellen Stand der Technik konstruiert und gefertigt.
Alle Komponenten unterliegen während der Fertigung strengen Qualitäts- und
Umweltkriterien. Unsere Managementsysteme sind nach ISO 9001 und ISO 14001
zertiziert.
D
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraussetzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshinweise und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschriften und allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des
Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig
durchgelesen und verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwen-
dung, Nichtbeachten dieser Betriebsanleitung, Einsatz ungenügend qualizierten
Fachpersonals sowie eigenmächtiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
24WIKA Betriebsanleitung Druckmessumformer, Typ HP-2
11379180.03 09/2013 GB/D/F/E
1. Allgemeines / 2. Sicherheit
Symbolerklärung
WARNUNG!
… weist auf eine möglicherweise gefährliche Situation hin, die zum Tod oder
zu schweren Verletzungen führen kann, wenn sie nicht gemieden wird.
VORSICHT!
… weist auf eine möglicherweise gefährliche Situation hin, die zu geringfügigen oder leichten Verletzungen bzw. Sach- und Umweltschäden führen
kann, wenn sie nicht gemieden wird.
Information
… hebt nützliche Tipps und Empfehlungen sowie Informationen für einen
ezienten und störungsfreien Betrieb hervor.
WARNUNG!
… weist auf eine möglicherweise gefährliche Situation hin, die durch heiße
Oberächen oder Flüssigkeiten zu Verbrennungen führen kann, wenn sie
nicht gemieden wird.
Abkürzungen
2-LeiterDie zwei Anschlussleitungen dienen zur Spannungsversorgung.
Der Speisestrom ist das Messsignal.
3-LeiterZwei Anschlussleitungen dienen zur Spannungsversorgung.
Eine Anschlussleitung dient für das Messsignal.
U
, +5VPositive Hilfsenergie
B
0V, GNDBezugspotential
S
+
,D
D
+
Positiver Messanschluss
Datenleitung USB 2.0
-
2. Sicherheit
D
WARNUNG!
Vor Montage, Inbetriebnahme und Betrieb sicherstellen, dass der richtige
Druckmessumformer hinsichtlich Messbereich, Ausführung und spezischen Messbedingungen ausgewählt wurde.
Der Betreiber muss die Verträglichkeit des Fluides mit dem eingesetzen
Werksto sicherstellen. Bei Medium Wassersto den Hersteller kontaktieren.
Bei Nichtbeachten können schwere Körperverletzungen und/oder
Sachschäden auftreten.
11379180.03 09/2013 GB/D/F/E
25WIKA Betriebsanleitung Druckmessumformer, Typ HP-2
Druckmessumformer immer innerhalb der Überlast-Druckgrenze betreiben.
D
Weitere wichtige Sicherheitshinweise benden sich in den einzelnen
Kapiteln dieser Betriebsanleitung.
2.1 Bestimmungsgemäße Verwendung
Der Druckmessumformer dient zum Umwandeln von Druck in ein elektrisches Signal.
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen Verwendungszweck konzipiert und konstruiert und darf nur dementsprechend verwendet werden.
Die technischen Spezikationen in dieser Betriebsanleitung sind einzuhalten. Eine
unsachgemäße Handhabung oder ein Betreiben des Gerätes außerhalb der techni-
schen Spezikationen macht die sofortige Stilllegung und Überprüfung durch einen
Das Fachpersonal ist aufgrund seiner fachlichen Ausbildung, seiner Kenntnisse der
Mess- und Regelungstechnik und seiner Erfahrungen sowie Kenntnis der landesspezi-
schen Vorschriften, geltenden Normen und Richtlinien in der Lage, die beschriebenen
Arbeiten auszuführen und mögliche Gefahren selbstständig zu erkennen.
Spezielle Einsatzbedingungen verlangen weiteres entsprechendes Wissen, z. B. über
aggressive Medien.
26WIKA Betriebsanleitung Druckmessumformer, Typ HP-2
11379180.03 09/2013 GB/D/F/E
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.