WIKA A-RB-1 Operating Instructions Manual

Operating Instructions
Digital Indicator Model A-RB-1
2
Contents
Page Section Topic
4 1. General 5 2. Layout of front panel 5 2.1 LED display 'A' 5 2.2 MODE key 'H' 5 2.2.1 Real mode 'h' 5 2.2.2 HOLD mode 'h' 6 2.2.3 MIN mode (minimum memory) 'h' 6 2.2.4 MAX mode (maximum memory) 'h'
6 2.3 Set key descending value ( ) 'B' 6 2.4 Set key ascending value ( ) 'C'
6 2.5 Programming key PROG 'F'
(see also appendix A)
7 2.5.1 Programming the indication
Low end of scale High end of scale Decimal point Input signal Output signal Damping Reference for damping
Baud rate setting 10 2.5.2 Verification of programmed settings 10 2.5.3 Compensation of zero offset 11 2.6 Model A-RB-1-D with 2 alarms
(see also appendix B) 11 2.6.1 Setting of alarm 1 14 2.6.2 Setting of alarm 2 14 2.6.3 Verification of set points of alarm 1 14 2.6.4 Verification of set points of alarm 2 14 2.7 Reset key 'G' 14 2.7.1 Erase data memory 15 2.7.2 Exit programming mode 15 2.7.3 Exit verification mode
3
Contents
Page Section Topic
15 3. Layout of back panel terminals 16 3.1 Layout of 15-pin plug 'J' 16 3.1.1 Opening the case 17 3.1.2 Scheme of line power settings 17 3.2 Layout of 9-pin plug 'K' 18 3.3 Wiring examples 19 3.4 Layout of 9-pin Sub-D plug 'L' 19 4. Option serial interface RS-232 20 4.1 Transfer of data and parameters 22 4.2 Programming of parameters 23 5. Error messages 24 5.1 Error messages E1/-E1 24 5.2 Error messages E2/-E2 24 5.3 Error messages E3/-E3 24 5.3.1 Error message E3 25 5.3.2 Error message -E3 26 6. Preparing for installation 26 7. Environment 26 8. Setting of physical unit 26 9. Maintenance 27 10. Specifications
28 App. A Schematic description of settings and data
transmission
32 App. B Schematic description of alarm settings
and verifications
34 App. C Dimensions
36 App. D Layout of front and back panel
4
Safety Instructions
Preface
Many thanks for buying our digital indicator Model A-RB-1.
This operating manual includes instructions for the operation of the digital indicator and information on its functionality. Read these operat­ing instructions thoroughly prior to starting up the digital indicator. In order to avoid any damage or injuries that might be caused by any non­observance of the appropriate regulations, please ensure that the person operating this digital indicator receives these operating instructions.
The appropriate national safety regulations (e.g. VDE 0100) must be observed when mounting, starting up and operating these displays. Serious injuries and/or damage can occur should the appropriate regulations not be observed. Only appropriately qualified persons should work on these instruments.
1. General
The digital indicators A-RB-1 are precision instruments for the measure­ment of current and voltage signals of pressure or other transmitters. The instruments are normally DIN-Size panel mounting (96 x 48 x 190 mm) per IEC 61 554.
Indication is made via a 3 1/2 -digit LED display, covering a range from
-1999 to +1999 digits. The actual span to be indicated can be easily programmed anywhere within this range. The same applies to decimal point, signal input, analogue output damping and baud rate of the data interface. All programming can be made while the instrument is operative.
This versatility is achieved by means of a powerful microprocessor, which also controls all other functions.
An inbuilt isolated transformer provides power supply of DC 24 V max. 30 mA to energise transmitters connected.
Analogue output of 0 … 10 V, 0 … 20 or 4 … 20 mA, adjustable damping, as well as HOLD memory and MIN and MAX memory are serial standard.
5
Optionally available are 2 alarm contacts and also a serial RS-232 interface.
2. Layout of front panel
The position of the operation and connection elements is shown in appendix C.
2.1 LED display 'A'
By means of the MODE key (see section 2.2), the display can be set to optionally read the REAL value measured, or either one of the values stored in the HOLD, MIN or MAX memory.
2.2 MODE key 'H'
The MODE key is found at the right-hand side next to the display, indicated by REAL, HOLD, MIN and MAX. Hitting changes the modes in consecutive order. A red LED indicates the active mode.
2.2.1 REAL mode 'h'
Indicates the current value measured.
2.2.2 HOLD mode 'h'
Hold the value indicated at the very moment the key is pressed. Measurement continues in the background, meaning that the memories of Minimum and Maximum as well as the alarm contacts continue to operate. HOLD discontinues upon further hitting of the MODE key or hitting of the RESET key (see 2.7).
6
2.2.3 MIN mode (minimum memory) 'h'
The lowest value indicated since last hitting of the RESET key is memorised and will be displayed in this mode. Measurement continues in the background, meaning that the memories of MIN and MAX as well as the alarm contacts continue to operate. MIN discontinues upon further hitting of the MODE key. Hitting of the RESET key erases the memory (see 2.7).
2.2.4 MAX mode (maximum memory) 'h'
The highest value indicated since last hitting of the RESET key is memorised and will be displayed in this mode. Measurement continues in the background, meaning, that the memories of MIN and MAX as well as the alarm contacts continue to operate. MAX discontinues upon further hitting of the MODE key. Hitting of the RESET key erases the memory (see 2.7).
2.3 Set key descending value ( ) 'B'
Selects the next lower value or individual parameter during program­ming.
2.4 Set key ascending value ( ) 'C'
Selects the next higher value or individual parameter during program­ming.
2.5 Programming key PROG 'F'
Hitting of this key actuates the programming mode, at which all operative parameters can be set and verified.
2.5.1 Programming the indication (see also appendix A)
All programming is made in consecutive order by initially holding the PROG key pressed for approx. 5 seconds, until the message SCL (for scaling) appears instead of PRO for programming.
7
Press PROG once more: -A-appears, standing for "Low end of scale". Press PROG once more: The "MIN" LED flashes and the corresponding value is displayed. Change this value as desired by hitting the (
) and
(
) keys. Press PROG once more: The new value will be memorised.
This is indicated by 3 dashes "---" appearing for a few seconds.
The "MIN" LED extinguishes and -E-appears, standing for "High end of scale". Press PROG once more: The "MAX" LED flashes and the corresponding value is displayed. Change this value as desired by hitting the (
) and ( ) keys. Press PROG once more: The new value
will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds.
The "MAX" LED extinguishes and dP appears, standing for "Decimal point". Change the decimal point as desired by hitting the (
) and (  )
keys. Press PROG once more: The new position of the decimal point will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds after which InX appears, standing for "Signal input X", where "X" stands for figures 1 to 3 as explained below. Change this value as desired by hitting the (
) and ( ) keys.
X = 1: In1 = Input signal voltage 0 ... 10 V X = 2: In2 = Input signal current 0 ... 20 mA X = 3: In3 = Input signal current 4 ... 20 mA
Press PROG once more: The new value will be memorised.
8
This is indicated by 3 dashes "---" appearing for a few seconds after which OuX appears, standing for "Output signal X", where "X" stands for figures 1 to 3 as explained below. Change this value as desired by hitting the (
) and ( ) keys.
X = 1: Ou1 = Output signal voltage 0 ... 10 V X = 2: Ou2 = Output signal current 0 ... 20 mA X = 3: Ou3 = Output signal current 4 ... 20 mA
Press PROG once more: The new value will be memorised .This is indicated by 3 dashes "---" appearing for a few seconds.
Subsequently "-d- (damping)" appears automatically. After hitting the key once more the current setting of the damping is displayed. The damping can be set within a range of 0.1 s ... 50.0 s. The damping can be changed in 0.1 s steps using the (
) and ( ) keys.
The damping is adapted from the behaviour of a capacitor. After the set time (=t) has passed, approx. 63% of the changed value is applied. After 5 x t has passed, approx. 97% of the changed value is applied.
Press PROG once more: The new damping value will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds.
Then dXX (reference of damping) is displayed. XX stands for the selected setting. The damping can be activated for the display, the analogue output signal, the alarm contacts and the MIN / MAX memory in any combination as specified in the following table:
9
Setting Display Analogue Alarm MIN / MAX d XX output contacts memory d011000 d020100 d031100 d040010 d051010 d060110 d071110 d080001 d091001 d100101 d111101 d120011 d131011 d140111 d151111
(0: Damping deactivated, 1: Damping activated)
The value for dXX can be changed using the (
) and ( ) keys. Press
PROG once more: The new value will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds.
With instruments not featuring the serial data interface, programming is now complete and standard operating mode will be automatically selected.
With instruments featuring the serial data interface, -b- appears. After hitting the key once more the current setting of the Baud rate of the serial interface (RS 232) is indicated. Change this value as desired by hitting the (
) and ( ) keys.
Display Baud rate
01.2 1200 Baud
02.4 2400 Baud
04.8 4800 Baud
09.6 9600 Baud
19.2 19200 Baud
38.4 38400 Baud
10
Press PROG once more: The new value will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds. Programming is now complete and standard operating mode will be automatically selected.
Programming can be terminated at any time by hitting the RESET key. In this instance, only such changes are accepted that have been acknowl­edged by appearance of "---". Otherwise, previously set values remain in effect.
2.5.2 Verification of programmed settings
(see also appendix A)
Short hitting of the PROG key initiates "Pro" to appear at the display, followed by all current settings in consecutive order, where:
-A- Low end of scale
-E- High end of scale dP Decimal point InX Input signal OuX Output signal
-d- Damping dxx Reference of damping
-b- At digital interface: baud rate setting
Indication can be terminated at any time by hitting the RESET key. (see also 2.7)
2.5.3 Compensation of zero offset
Despite careful calibration, the instrument may indicate a zero offset in operation. This may be caused by a static head acting on the transmitter or other process conditions. Preferably this should be compensated by shifting the zero signal of the transmitter. If this cannot be accomplished, true indication can be achieved by means of shifting low end and high end indication correspondingly as per examples below.
11
Example 1: Scaling: 0 … 400 bar Zero offset: 4 bar Corrective: -4 … 396 bar
Example 2: Scaling: 0 … 400 bar Zero offset: -7 bar Corrective: 7 … 407 bar
However, it is more favourable to compensate the offset by adjustment the connected transmitter.
2.6 Model A-RB-1-D with 2 alarms
(see also appendix B)
The keys SET 1 (alarm 1) and SET 2 (alarm 2) actuate the programming mode to enter and verify the settings of the alarms. The max. loading capacity of the alarms is AC 250 V / 8 A.
2.6.1 Setting of alarm 1
(see also appendix B)
Programming of alarm 1 is made by initially holding the SET 1 key pressed until SP1 (Set point 1) disappears and the message SE (Set point makE) appears.
Press SET 1 once more: The red LED in the upper left hand corner of the SET 1 key flashes and the corresponding value is displayed. Change this value as desired by hitting the (
) and ( ) keys.
Press SET 1 once more: The new value will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds. The LED extinguishes and SA appears, standing for "Set point breAk".
Press SET 1 once more: The red LED in the upper right corner of the SET 1 key flashes and the corresponding value is displayed. Change this value as desired by hitting the (
) and ( ) keys.
12
Press SET 1 once more: The new value will be memorised. This is indicated by 3 dashes "---" appearing for a few seconds and the LED extinguishes.
Contact function can be selected to HIGH ALARM (meaning make on rising value), or LOW ALARM (meaning make on falling value). This is easily achieved by programming the make (SE) value either above or below the corresponding break (SA) value.
Setting (SE) above (SA) means HIGH ALARM. (SE) will energise the alarm circuit, which will remain energised until the display figure decreases to reach the value of (SA).
Setting (SE) below (SA) means LOW ALARM. (SE) will energise the alarm circuit, which will remain energised until the display figure increases to reach the value of (SA).
The difference between (SE) and (SA) represents the hysteresis across make and break points of the contact. (This must not be confused with any hysteresis across approach of the set points with rising and falling values. This sort of mechanical delay is not apparent with a digital instrument).
Both values can be programmed without limitations, as the case demands. Setting both, (SE) and (SA) at the same values, will automati­cally create HIGH ALARM function.
The LED's in the upper corners of the SET keys are intended to indicate the alarm configuration together with the relay status. The left LED, when lit, indicates energised alarm circuit at HIGH ALARM programmed. The right LED, when lit, indicates energised alarm circuit at LOW ALARM programmed.
SET 1 key corresponds to alarm 1, SET 2 key corresponds to alarm 2.
Programming of alarm contacts can be terminated at any time by hitting the RESET key. In this instance, only such changes are accepted that have been acknowledged by appearance of "---". Otherwise, previously set values remain in effect.
13
hysteresis
Relay on
Relay off
"-E-"
makE point
400
"-A-"
breAk point
500
value
"-A-"
breAk point
800
"-E-"
makE point
1000
hysteresis
Relay on
Relay off
value
Examples:
a) HIGH ALARM (make at 1000, break at 800)
The circuit relay is energised once the display value rises to 1000. It remains energised until the display value falls to 800.
The left hand LED at the SET 1 key comes on at 1000 and is turned off at 800.
b) LOW ALARM (make at 400, break at 500)
The circuit relay is energised once the display value falls to 400. It remains energised until the display value rises to 500.
The right hand LED at the SET 1 key comes on at 400 and is turned off at 500.
14
2.6.2 Setting of alarm 2
(see also appendix B)
Programming of the alarm 2 contact is initiated by hitting key SET 2, otherwise fully identically to the programming of no. 1.
2.6.3 Verification of set points of alarm 1
(see also appendix B)
Short hitting of the SET 1 key initiates "SP1" to appear at the display, followed by the current settings of (SE) and (SA) of alarm contact no. 1
Indication can be terminated at any time by hitting the RESET key. (see also 2.7)
2.6.4 Verification of set points of alarm 2
(see also appendix B)
Verification of the alarm 2 contact is initiated by hitting key SET 2, otherwise fully identical to the verification of no. 1.
2.7 RESET key 'G'
The RESET key enables to - erase memories - exit programming mode ­exit verification mode.
2.7.1 Erase data memory
Hitting the RESET key while the instrument is operative erases the data memories as explained below:
- When pressed in REAL mode (indication of value measured), the MIN and MAX memories will be erased.
- When pressed in HOLD mode, this will reset the instrument into REAL mode.
- When pressed in MIN mode, only the MIN memory will be erased.
- When pressed in MAX mode, only the MAX memory will be erased.
15
2.7.2 Exit programming mode
(see also 2.5.1, 2.6.1 and 2.6.2)
Programming can be terminated at any time by hitting the RESET key. In this instance, only such changes are accepted, that have been acknowl­edged by appearance of "---". Otherwise, previously set values remain in effect.
2.7.3 Exit verification mode
(see also 2.5.2, 2.6.3 and 2.6.4)
The verification mode proceeds automatically and can be terminated at any time by hitting the RESET key.
3. Layout of back panel terminals
(see illustration of appendix D)
The back panel features the 15-pin plug 'J', containing the terminals for power supply and both alarm contacts, and the 9-pin plug 'K', contain­ing the terminals of input signal, transmitter supply and analogue output.
Both racks are of the plug-terminal type for ease of wiring.
Every 2nd pin of the 15-pin plug remains blank to enable safe wiring of the line voltage.
Instruments incorporating the serial interface will additionally feature a 9­pin Sub-D plug 'L'.
16
3.1 Layout of 15-pin plug 'J'
Pin Designation
1 3 Alarm 1 contact max. AC 250 V / 8 A 5
7 9 Alarm 2 contact max. AC 250 V / 8 A 11
13 15 AC 230 V ± 10 % or AC 115 V ± 10 %, 50 / 60 Hz
Internal jumpers provide adaption to line voltage AC 230 V ± 10 %, 50 / 60 Hz or AC 115 V ± 10 %, 50 / 60 Hz. To adjust, open the enclosure and arrange jumpers next to the fuse as indicated in the drawing. A replacement of the fuse is not required.
3.1.1 Opening the case
Remember to disconnect the power supply prior to opening the enclosure!
First strip the terminal blocks from the digital indicator. Then remove the retaining screw in the centre of the case back and loosen the electronic rack from the locking device by exerting a constant pressure on the 9­pin female plug towards the front. Now you can pull the rack out of the case.
For assembly please carry out these steps in reverse order.
17
3.1.2 Scheme of line power settings
Power setting 230 VAC ± 10 %: Power setting 115 VAC ± 10 %:
3.2 Layout of 9-pin plug 'K'
Pin Designation
1+U
in
Voltage signal input
2-U
in
3+IinCurrent signal input 4-I
in
5 -out Common minus of current and voltage output
signal (pins 8 and 9)
6 +24 V Transmitter supply 7 -GND
8+U
out
Analogue output signal voltage
9+I
out
Analogue output signal current
18
+ I-in 3
- I-in 4
GND 7 + 24 V 6
+ signal
- power supply + power supply
+ U-in 1
- U-in 2
GND 7 + 24 V 6
+ signal
- power supply + power supply
+ U-in 1
- U-in 2
GND 7 + 24 V 6
+ signal
- signal
- power supply + power supply
+ I-in 3
- I-in 4
GND 7 + 24 V 6
- terminal
+ terminal
3.3 Wiring examples:
a) 2-wire, 4 ... 20 mA transmitter signal
Digital indicator Transmitter
b) 3-wire, 0 ... 20 mA transmitter signal
Digital indicator Transmitter
c) 3-wire, 0 ... 10 V transmitter signal
Digital indicator Transmitter
d) 4-wire, 0 ... 10 transmitter signal
Digital indicator Transmitter
19
3.4 Layout of 9-pin Sub-D plug 'L'
Layout of the serial interface is identical to that of commonly found Personal Computers. This simplifies data input into these widely used machines.
Pin Designation
2 TX DATA 3 RX DATA 5 GROUND
4. Option serial interface RS-232
The instrument can be optionally equipped with a serial RS-232 data interface. The interface transmits data measured as well as enabling programming of the instrument. Data transmission and programming follows the same rule in principle. Data reading requires a 5-byte command. The instrument responds by acknowledging the command together with the respective data. Programming requires a 5-byte command followed by a word of 1 to 6 bytes length. All commands transmitted and received are followed by
<CR>, Dec 13, () (carriage Return).
20
4.1 Transfer of data and parameters
Com- Response Meaning mand No. of byte
Example
RREAL Value active value measrued incl.
5 + 6 byte + sign and decimal point
RREAL+12.34
RMODE 1 = REAL mode status message, indicating
2 = HOLD mode current mode of instrument 3 = MIN mode 4 = MAX mode
5 + 1 byte +
RMODE1
RMINM Min value contents of MIN memory incl.
5 + 6 byte + sign and decimal point
RMINM+01.23
RMAXM Max value contents of MAX memory incl.
5 + 6 byte + sign and decimal point
RMAXM+14.56
RST1E SE of alarm 1 starting point alarm 1, incl.
5 + 6 byte + sign and decimal point
RST1E+10.00
RST1A SA of alarm 1 end point alarm 1, incl.
5 + 6 byte + sign and decimal point
RST1A+09.50
RST2E SE of alarm 2 starting point alarm 2, incl.
5 + 6 byte + sign and decimal point
RST2E+08.00
RST2A SA of alarm 2 end point alarm 2, incl.
5 + 6 byte + sign and decimal point
RST2A+08.50
21
Com- Response Meaning mand No. of byte
Example
RSCLA Low scale low end of scale value incl.
5 + 6 byte + sign and decimal point
RSCLA+00.00
RSCLE High scale high end of scale value incl.
5 + 6 byte + sign and decimal point
RSCLE+16.00
RINPX 1 = 0 ... 10 V input signal selected
2 = 0 ... 20 mA 3 = 4 ... 20 mA
5 + 1 byte +
RINPX3
ROUTX 1 = 0 ... 10 V output signal selected
2 = 0 ... 20 mA 3 = 4 ... 20 mA
5 + 1 byte +
ROUTX3
RDAEM Damping set value of damping in seconds
5 + 4 byte +
RDAEM00.1
RDBEZ Damping reference set reference of damping
5 + 2 byte +
RDBEZ01
All commands transmitter to the instrument must terminate with <CR> Dec. 13 (
). All date received
from the instrument will terminate with <CR> Dec. 13 (
↵ ).
22
4.2 Programming of parameters
Com- Response Meaning
mand No. of byte
Example
PMODE 1 = REAL mode signal to set
2 = HOLD mode mode of instrument 3 = MIN mode 4 = MAX mode
5 + 1 byte +
PMODE1
PMINM 0 = Reset erase MIN memory
5 + 1 byte +
PMINM0
PMAXM 0 = Reset erase MAX memory
5 + 1 byte +
PMAXM0
PST1E SE of alarm 1 set switch-on value alarm 1, incl.
5 + 6 byte + sign and decimal point
PST1E+10.00
PST1A SA of alarm 1 set switch-off value alarm 1, incl.
5 + 6 byte + sign and decimal point
PST1A+09.50
PST2E SE of alarm 2 set switch-on value alarm 2, incl.
5 + 6 byte + sign and decimal point
PST2E+08.00
PST2A SA of alarm 2 set switch-off value alarm 2, incl.
5 + 6 byte + sign and decimal point
PST2A+08.50
PSCLA Low scale set low end of scale value incl.
5 + 6 byte + sign and decimal point
PSCLA+00.00
23
Com- Response Meaning
mand No. of byte
Example
PSCLE High scale set high end of scale value incl.
5 + 6 byte + sign and decimal point
PSCLE+16.00
PINPX 1 = 0 ... 10 V set input signal at instrument
2 = 0 ... 20 mA 3 = 4 ... 20 mA
5 + 1 byte +
PINPX3
POUTX 1 = 0 ... 10 V set output signal (analogue output)
2 = 0 ... 20 mA at instrument 3 = 4 ... 20 mA
5 + 1 byte +
POUTX3
PDAEM Damping set value in seconds for damping
5 + 4 byte +
PDAEM00.1
PDBEZ Damping reference set reference of damping
5 + 2 byte +
PDBEZ01
All commands transmitted to the instrument must terminate with <CR> Dec. 13 (
). The decimal
point must be entered with both, low end and high end scale values, where the last value entered will determine the actual decimal point, irrespective of this being the low or high end value. Note correct decimal point when entering the alarm contact settings.
24
5. Error messages
Altogether 6 different error messages may be displayed:
5.1 Error messages E1/-E1
The error messages E1/-E1 will appear, if the actual input signal exceeds the programmed maximum signal value (10 V or 20 mA) by more than 9% or if it falls below the minimum signal value (0 V, 0 mA or 4 mA) by more than 9 % of the maximum signal value, because the inbuilt A/D converter is overloaded. E1 indicates a too high signal, -E1 indicates a too low signal. Below table indicates the actual values at which this error message will appear:
Message Set input signal Actual signal
E1 0 ... 10 V > 10.9 V
0 ... 20 mA > 21.8 mA 4 ... 20 mA > 21.8 mA
-E1 0 ... 10 V < -0.9 V 0 ... 20 mA < -1.8 mA 4 ... 20 mA < 2.2 mA
5.2 Error messages E2/-E2
E2 appears if the input value exceeds the corresponding indication of +1999 digits. -E2 appears if the input value exceeds the corresponding indication of -1999 digits.
5.3 Error messages E3/-E3
5.3.1 Error message E3
E3 will appear, if the actual input signal exceeds the maximum signal value (10 V or 20 mA) by more than 6 %. This is still within the capacity of the inbuilt A/D converter (see also 5.1). If the value to display is still within the capacity of the indicator, the value and the respective error message will flash intermittently. If the value exceeds ±1999, E3 will be displayed permanently.
25
Below table indicates the actual values at which this error message will appear:
Message Set input signal Actual signal
E3 flashes 0 ... 10 V > 10.6 V and < 10.9 V intermittently 0 ... 20 mA > 21.2 mA and < 21.8 mA with measured 4 ... 20 mA > 21.2 mA and < 21.8 mA value
E3 0 ... 10 V > 10.6 V and < 10.9 V and
display > +1999 or < -1999
0 ... 20 mA > 21.2 mA and < 21.8 mA and
display > +1999 or < -1999
4 ... 20 mA > -21.2 mA and < 21.8 mA
display > +1999 or < -1999
5.3.2 Error message -E3
-E3 will appear, if the actual input signal is below the minimum signal
value (0 V, 0 mA or 4 mA) by more than 6 % of the maximum signal value. This is still within the capacity of the inbuilt A/D converter (see also 5.1). If the value to display is still within the capacity of the indicator, the value and the respective error message will flash intermittently. If the value exceeds ±1999, -E3 will be displayed permanently.
Below table indicates the actual values at which this error message will appear:
Message Set input signal Actual signal
-E3 flashes 0 ... 10 V < -0.6 V and > -0.9 V
intermittently 0 ... 20 mA < -1.2 mA and > -1.8 mA with measured 4 ... 20 mA < 2.8 mA and > 2.2 mA value
-E3 0 ... 10 V < -0.6 V and > -0.9 V and
display > +1999 or < -1999
0 ... 20 mA < -1.2 mA and > -1.8 mA and
display > +1999 or < -1999
4 ... 20 mA < 2.8 mA and > 2.2 mA and
display > +1999 or < -1999
26
6. Preparing for installation
The instrument is designed to fit panels of 40 mm maximum thickness. Panel cut out per IEC 61 554, 92 +0.8 mm wide and 45 + 0.6 mm high. Panel clamps are supplied with the instrument.
7. Environment
The front panel of the instrument is protected against moisture and dust (IP65) by means of a sealed foil. Full ingress protection may be achieved by fitting an appropriate gasket between panel and instrument.
The ambient operating temperature should be maintained within the range 0 to 50 °C. When the permissible ambient temperature is ex­ceeded make sure that there is sufficient ventilation.
8. Setting of physical unit
The instrument is supplied with a variety of labels to suit most commonly used units of pressure and temperature. A number of blank labels may be used to indicate customised units.
The pocket 'I' above the MODE key is intended to accept the unit label. To insert the label, pry the pocket open with a tipped instrument. Take care not to loosen or damage the protective foil.
9. Maintenance
No subject to wear and tear parts or components requiring any regular maintenance are contained in the instrument. In case of obvious malfunction, it is recommended that you return the instrument to an authorised WIKA service for repair.
The front foil may be cleaned using a moist cloth and some non abrasive household detergent.
27
Display
- Design 7-Segment-LED, red, 3 1/2 -digit
- Size of digits 14.56 mm
- Indication range -1999 ... +1999
Accuracy ± 0.05 % of span ± 2 digit Measuring rate 10 measurements/s, damping selectable in 100 ms steps up to max. 50.0 s Error messages E1: A/D converter overflow
E2: Display overflow (measured value exceeds maximum possible display value) E3: Input signal is below the minimum signal value or exceeds the maximum signal value
Scaling adjustment Menu driven, initial value and final value free adjustable between -1999 and +1999;
Adjustable decimal point
Signal input Selectable as:
0 ... 20 mA, 4 ... 20 mA 0 ... 10 V
Analogue output Selectable as:
0 ... 20 mA, 4 ... 20 mA
0 ... 10 V Response time 100 ms (10 ... 90 %) {Serial interface} RS-232 Transmitter supply DC 24 V ± 5 %, max. 30 mA, galvanically isolated, short-circuit proof (for
approx. 8 minutes) {Alarm contacts}
- Number 2, independently settable
- Function MAX/MIN-alarm adjustable by setting of the switch-on and switch-off value
- Switching point Adjustable over the complete indication range
- Hysteresis Adjustable over the complete indication range
- Accuracy True value by means of digital control
- Contacts 1 potential-free relay change over contact for each alarm contact
- Load AC 250 V 8 A with resistive load;
AC 250 V 1 A with cos ϕ = 0.1
HOLD memory Displayed value is fix, measurement and control of MIN and MAX values as
well as alarm contacts goes on in the background. MIN/MAX memory Two separately working memories for MIN and MAX values;
Individual or common reset enabled by pressing the RESET key;
Unlimited data storage by digital memory Power supply AC 230 V, 50/60 Hz, ± 10 % or
AC 115 V, 50/60 Hz, ± 10 %, changeable by means of internal jumper Electrical connection Detachable screw terminals
-Max. cable diameter 2.5 mm
2
Permissible ambient 0 °C ... 50 °C temperature CE Conformity Conformity in accordance with 89/336/EWG
Interference emission per EN 60 000-6-4
Interference compatibility per EN 61 000-6-2
For cable lengths of > 30 m, shielded cables are to be used Case According to IEC 61 554
- Material PC, ABS-Blend, black
- Ingress protection Front: IP65; Back: IP00 (according to IEC 60 529 / EN 60 529)
- Mass Approx. 530 g
- Mounting Removable screw elements for a wall thickness up to 40 mm
10. Specifications
Specification Model A-RB-1
{ } Items in curved brackets are optional extras for additional price.
28
Appendix A Schematic description of settings and data transmission
Initial status: REAL mode
Press PROG and hold down
Pro appears at display and will change to SCL indication programming mode.
Press PROG
- A- appears at display
Press PROG
current low scale value is displayed and MIN-LED flashs
Adjust value using (
) and ( ) keys
low scale value is changing.
Press PROG
- - - appears, indicating acceptance of new lower scale value
-E- appears at display
Press PROG
current high scale value is displayed and MAX-LED flashs
Adjust value using (
) and ( ) keys
high scale value changes
Press PROG
- - - appears, indicating acceptance of new upper scale value
dP. appears and current position of decimal point is
displayed
29
Adjust value using ( ) and ( ) keys
decimal point changes
Press PROG
- - - appears, indicating acceptance of new decimal position
InX appears and current index of input signal is displayed
Adjust value using ( ) and ( ) keys
X index figure changes
Press PROG
- - - appears, indicating acceptance of new input signal
OuX appears at display
Adjust value using ( ) and ( ) keys
X index figure changes
Press PROG
- - - appears, indicating acceptance of new output signal
- d - appears at display
Press PROG
set value for damping in seconds appears
Adjust value using ( ) and ( ) keys
set time for damping is changed
30
Press PROG
- - - appears, indicating acceptance of the time of damping
dXX appears automatically with the current reference of
damping
Adjust value using ( ) and ( ) keys
reference of damping changes
Press PROG
- - - acceptance of the reference of damping
with - b - appears at display RS232
Press PROG
set baud rate appears (values in kBaud).
Adjust value using ( ) and ( ) keys
stored value for the baud rate is changed.
Press PROG
- - - appears, indicating acceptance of new baud rate
Hitting RESET immediately terminates the programming mode at any time. In this instance, the instrument resumes REAL mode.
Only such changes that have been acknowledged by appearance of "---" will become effective. Otherwise, previously set values remain valid.
31
Short hitting of the PROG key, while the instrument is in REAL mode, initiates “PRO“ to appear at the display, followed by all current settings in consecutive order.
Below flow diagram explains appearance of key words and values, presuming settings of:
Examples:
Scale expansion: 0 ... 6.00 bar Input signal: 4 ... 20 mA Signal output: 0 ... 10 V Damping
: 1.0 s
Reference of damping: Display Baud rate 9600 baud
Initial status: REAL mode
Press PROG shortly
Pro SCL -A- 0.00 -E- 6.00
d.P In3 Ou1 -d- 01.0 d01
-b- 09.6
Hitting RESET immediately terminates verification mode at any time. In this instance, the instrument resumes REAL mode.
32
Appendix B Schematic description of alarm settings and verification
Initial status: REAL mode
Press SET 1 and hold down
SP1 appears at display and will change to SE indicating programming mode
Press SET 1
current Set point make value no. 1 appears and the
- LED of alarm 1 flashs
Adjust value using (
) and ( ) keys
make value changes
Press SET 1
- - - appears, indicating acceptance of new Set point make value
SA of alarm 1 appears at display
Press SET 1
current Set point break value no. 1 appears and the
- LED of alarm 1 flashs
Adjust value using (
) and ( ) keys
break value changes
Press SET 1
- - - appears, indicating acceptance of new Set point break value
33
Hitting RESET immediately terminates the programming mode at any time. In this instance, the instrument resumes REAL mode.
Only such changes that have been acknowledged by appearance of "---" will become effective. Otherwise, previously set values remain valid.
Set points of alarm 2 are entered in the same manner.
Short hitting of the SET 1 (SET 2, respectively) key, while the instrument is in REAL mode, initiates verification of the current settings of alarm 1 (2, respectively).
Example:
Set point make (SE): 4.00 Set point break (SA): 3.80
Initial status: REAL mode
Press SET 1 shortly
SP1 SE 4.00 SA 3.80
Hitting RESET immediately terminates verification mode at any time. In this instance, the instrument resumes REAL mode.
Set points or alarm 2 are verified in the same manner.
34
89.5
47.5
95.7
173.5
Appendix C Dimensions
Dimensions in mm
35
8.9
42.3
45
+0.6
92
+0.8
Panel cutout in mm
Dimensions in mm
36
AB CD EFG HI
d e h
Appendix D Layout of front and back panel
37
A LED-display B Decrease value key C Increase value key D Check / set contacts no. 1
d - LED = MAX value; - LED = MIN value
E Check / set contacts no. 2
e - LED = MAX value; - LED = MIN value
F Select programming mode. Continue with programming G RESET memories, CANCEL programming H Select display mode (toggle between LED’s) h REAL = display true value
HOLD = hold value displayed MIN = contents of minimum memory displayed
MAX = contents of maximum memory displayed I Pocket window holding unit label J Terminal block power supply and contacts output K Terminal block signal input, signal output and transmitter supply L Sub-D serial port (optional)
38
For your notes
39
For your notes
40
WIKA Alexander Wiegand GmbH & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg / Germany Phone (+49) 93 72/132-9986 Fax (+49) 93 72/132-217 E-Mail testequip@wika.de www.wika.de
1287974 04/2005 GB
Loading...