Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d'emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!
Prima di iniziare ad utilizzare lo strumento, leggere il manuale d'uso!
Conservare per future consultazioni!
2
WIKA operating instructions air2guide model A2G-60
40202798.02 08/2012 GB/D/F/E/I
Contents
Contents
1.General information4
2.Safety5
3.Specications6
Design and function
4.
5.Transport, packaging and storage9
6.Commissioning, operation9
7.Maintenance and cleaning12
8.Disposal12
8
GB
40202798.02 08/2012 GB/D/F/E/I
WIKA operating instructions air2guide model A2G-60
3
1. General information
1. General information
■
The duct/immersion temperature sensor described in the operating
instructions has been designed and manufactured using state-of-theart technology. All components are subject to stringent quality and
GB
environmental criteria during production. Our management systems
are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on
handling the instrument. Working safely requires that all safety
instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and
general safety regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept
in the immediate vicinity of the instrument and readily accessible to
skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the
operating instructions prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused
by using the product contrary to its intended use, non-compliance
with these operating instructions, assignment of insuciently
qualied skilled personnel or unauthorised modications to the
instrument.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modications.
■
Further information:
- Internet address:www.wika.de / www.wika.com
www.air2guide.com
- Relevant data sheet: TE 62.90
4
WIKA operating instructions air2guide model A2G-60
40202798.02 08/2012 GB/D/F/E/I
1. General information / 2. Safety
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation which can
result in serious injury or death if not avoided.
Information
... points out useful tips, recommendations and information
for ecient and trouble-free operation.
2. Safety
WARNING!
Before installation, commissioning and operation, ensure
that the appropriate duct/immersion temperature sensor
has been selected in terms of measuring range, design
and specic measuring conditions.
Non-observance can result in serious injury and/or damage
to the equipment.
Further important safety instructions can be found in the
individual chapters of these operating instructions.
2.1 Intended use
■
For measuring the temperature of gaseous media in heating,
ventilation and air-conditioning systems
■
Designed for connection to control and display systems
■
In combination with a stem also suitable for measuring liquid
media (e.g. in pipeline systems)
The instrument has been designed and built solely for the intended use
described here, and may only be used accordingly.
GB
40202798.02 08/2012 GB/D/F/E/I
WIKA operating instructions air2guide model A2G-60
5
2. Safety / 3. Specications
The manufacturer shall not be liable for claims of any type based on
operation contrary to the intended use.
2.2 Personnel qualication
GB
Skilled personnel
Skilled personnel are understood to be personnel who, based on their
technical training, knowledge of measurement and control technology
and on their experience and knowledge of country-specic regulations,
current standards and directives, are capable of carrying out the work
described and independently recognising potential hazards.
Explanation of symbols
3. Specications
Electrical connection
Connection terminal, 3-pin (3-wire)
Cable gland: One-way, M16 for cable with max. D = 8 mm
Power supply
DC 15 ... 24 V / AC 24 V ±10 %
Power consumption
Typically 0.35 W / 0.82 VA
Output
TRV 0 ... 10 V, min. load 5 kΩ (3-wire)
TRA 4 … 20 mA (2-wire)
6
WARNING!
Risk of injury if qualication is insucient!
Improper handling can result in considerable injury and
damage to equipment.
■
The activities described in these operating instructions
may only be carried out by skilled personnel who have
the qualications described below.
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant
European directives.
WIKA operating instructions air2guide model A2G-60
40202798.02 08/2012 GB/D/F/E/I
3. Specications
Dimensions in mm
Insertion length L
Standard 192 mm
Mounting ange
■
Material PA6.6, colour black
■
Max. +130 °C
GB
Ingress protection
IP 65 per EN 60529 / lEC 529
For further specications see WIKA data sheet TE 62.90 and the order
documentation.
40202798.02 08/2012 GB/D/F/E/I
WIKA operating instructions air2guide model A2G-60
7
4. Design and function
4. Design and function
4.1 Description
Design
CE conformity:
GB
2004/108/EG Electromagnetic compatibility (EMC)
2002/95/EC RoHS (restriction of the use of certain hazardous
substances in electrical and electronic equipment)
Accuracy @ 21°C
Typically ±1 % of measuring range
Scale ranges
Standard temperature setting: 0 … +50 °C
to be set via jumper: -50 … +50 °C, -15 … +35 °C, -10 … +120 °C,
0 … +160 °C, 0 … +250 °C
Permissible temperature
■
Head temperature
-35 … +90 °C passive (Pt1000 and Ni1000)
-35 … +70 °C active (TRV and TRA)
■
Sensor sleeve
Standard: -50 … +160 °C
High temperature version: -50 ... + 260 °C
4.2 Scope of delivery
Cross-check the scope of delivery with the delivery note.
8
WIKA operating instructions air2guide model A2G-60
Check the duct/immersion temperature sensor for any damage that
may have been caused by transport. Obvious damage must be reported
immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage
Storage temperature: -20 ... +85 °C
Protect the instruments from moisture and dust.
6. Commissioning, operation
Installation and mechanical connection
■
Protect measuring instruments from contamination, high temperature
changes and vibrations
■
In order to avoid any additional heating, the instruments must not
be exposed to direct solar irradiation while in operation.
■
The duct/immersion temperature sensor can be fastened directly
to the ventilation duct using either a mounting ange or screws.
GB
40202798.02 08/2012 GB/D/F/E/I
WIKA operating instructions air2guide model A2G-60
The instruments are designed to operate with safety extra-low voltage
(SELV). When wiring up the instruments, the technical specications for
GB
those instruments should be followed. Particularly for passive sensors,
e.g. Pt1000, etc., in a 2-wire conguration, the sensor lead resistance
should be taken into account. If necessary, this must be corrected in the
connected electronics. Through self-heating, the measurement current
aects the measuring accuracy and should, therefore, not be above
1 mA.
In the case of a sensor with transmitter, as a rule, the transmitter should
be operated in the middle of the measuring range, since deviations can
occur at the range limits.
The ambient temperature of the transmitter electronics should be kept
constant. The transmitters must be operated with a constant power
supply.
The measuring range is altered by changing the positions of the jumper
links (see connection diagram). The initial value of the new measuring
range is then available after approx. 2 s.
SensorSensorSensor
4-wire
red
red
40202798.02 08/2012 GB/D/F/E/I
white
2-wire3-wire
10
white
brown
WIKA operating instructions air2guide model A2G-60
brown
green
white
white
6. Commissioning, operation
Scale range adjustment for a TRV sensor
Scale range adjustment
-50 ... +50 °C
-10 ... +120 °C
0 ... +50 °C
0 ... +160 °C
0 ... +250 °C
-15 ... +35 °C
Offset adjustment
Terminals
Screw terminals, max. 1,5 mm²
Cable entry
Single, M20 for cable of max. D = 8 mm
Double, M20 for 2 cables of max. D = 7 mm
GB
40202798.02 08/2012 GB/D/F/E/I
WIKA operating instructions air2guide model A2G-60
11
7. Maintenance and cleaning / 8. Disposal
7. Maintenance and cleaning
The duct/immersion temperature sensor is maintenance-free and oers
long service life provided it is handled and operated properly.
GB
Clean the instruments with a moist cloth (soap water).
Repairs must only be carried out by the manufacturer or appropriately
qualied skilled personnel.
8. Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an
environmentally compatible way and in accordance with the country-
specic waste disposal regulations.
12
WIKA operating instructions air2guide model A2G-60
40202798.02 08/2012 GB/D/F/E/I
Inhalt
Inhalt
1.Allgemeines14
2.Sicherheit15
3.Technische Daten16
Aufbau und Funktion
4.
5.Transport, Verpackung und Lagerung19
6.Inbetriebnahme, Betrieb19
7.Wartung und Reinigung22
8.Entsorgung22
18
D
40202798.02 08/2012 GB/D/F/E/I
WIKA Betriebsanleitung air2guide Typ A2G-60
13
1. Allgemeines
1. Allgemeines
■
Der in der Betriebsanleitung beschriebene Kanal-/Tauchtempe-
raturfühler wird nach dem aktuellen Stand der Technik konstruiert
und gefertigt. Alle Komponenten unterliegen während der Fertigung
strengen Qualitäts- und Umweltkriterien. Unsere Managementsy-
steme sind nach ISO 9001 und ISO 14001 zertiziert.
D
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit
dem Gerät. Voraussetzung für sicheres Arbeiten ist die Einhaltung
aller angegebenen Sicherheitshinweise und Handlungsanwei-
sungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfall-
verhütungsvorschriften und allgemeinen Sicherheitsbestimmungen
einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittel-
barer Nähe des Gerätes für das Fachpersonal jederzeit zugänglich
aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller
Arbeiten sorgfältig durchgelesen und verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestim-
mungswidrige Verwendung, Nichtbeachten dieser Betriebsanleitung,
Einsatz ungenügend qualizierten Fachpersonals sowie eigenmäch-
tiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufs-
… weist auf eine möglicherweise gefährliche Situation hin,
die zum Tod oder zu schweren Verletzungen führen kann,
wenn sie nicht gemieden wird.
Information
… hebt nützliche Tipps und Empfehlungen sowie Infor-
mationen für einen ezienten und störungsfreien Betrieb
hervor.
2. Sicherheit
WARNUNG!
Vor Montage, Inbetriebnahme und Betrieb sicherstellen,
dass der richtige Kanal-/Tauchtemperaturfühler hinsichtlich
Messbereich, Ausführung und spezischen Messbedingungen ausgewählt wurde.
Bei Nichtbeachten können schwere Körperverletzungen
und/oder Sachschäden auftreten.
Weitere wichtige Sicherheitshinweise benden sich in den
einzelnen Kapiteln dieser Betriebsanleitung.
2.1 Bestimmungsgemäße Verwendung
■
Zur Temperaturmessung in gasförmigen Medien von Heizungs-,
Lüftungs- und Klimaanlagen
■
Ausgelegt zur Aufschaltung an Regler- und Anzeigesystemen
■
In Verbindung mit einer Tauchhülse auch zur Messung in
üssigen Medien (z. B. Rohrleitungssysteme) geeignet
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen Verwendungszweck konzipiert und konstruiert und darf nur
dementsprechend verwendet werden.
40202798.02 08/2012 GB/D/F/E/I
WIKA Betriebsanleitung air2guide Typ A2G-60
D
15
2. Sicherheit / 3. Technische Daten
Ansprüche jeglicher Art aufgrund von nicht bestimmungsgemäßer
Verwendung sind ausgeschlossen.
2.2 Personalqualikation
WARNUNG!
Verletzungsgefahr bei unzureichender Qualikation!
D
Fachpersonal
Das Fachpersonal ist aufgrund seiner fachlichen Ausbildung, seiner
Kenntnisse der Mess- und Regelungstechnik und seiner Erfahrungen
sowie Kenntnis der landesspezischen Vorschriften, geltenden Normen
und Richtlinien in der Lage, die beschriebenen Arbeiten auszuführen
und mögliche Gefahren selbstständig zu erkennen.
Symbolerklärung
3. Technische Daten
Elektrischer Anschluss
Anschlussklemme, 3-polig (3-Leiter)
Kabelverschraubung: Einfach, M16 für Kabel mit max. D = 8 mm
Hilfsenergie
DC 15 ... 24 V / AC 24 V ±10 %
Leistungsaufnahme
Typ. 0,35 W / 0,82 VA
Ausgang
TRV 0 … 10 V, min. Belastung 5 kΩ (3-Leiter)
TRA 4 … 20 mA (2-Leiter)
16
Unsachgemäßer Umgang kann zu erheblichen Personenund Sachschäden führen.
■
Die in dieser Betriebsanleitung beschriebenen
Tätigkeiten nur durch Fachpersonal nachfolgend
beschriebener Qualikation durchführen lassen.
CE, Communauté Européenne
Geräte mit dieser Kennzeichnung stimmen überein mit den
zutreenden europäischen Richtlinien.
WIKA Betriebsanleitung air2guide Typ A2G-60
40202798.02 08/2012 GB/D/F/E/I