
DFX® Table of Contents
..........................................................................................................................................................................................
...........................................................................................................................................................................................
.........................................................................................................................................................................
........................................................................................................................................................................
......................................................................................................................................................................
..................................................................................................................................................................
3. Tone (Audio Frequency)
......................................................................................................................................................
..........................................................................................................................................................................
........................................................................................................................................................................
.........................................................................................................................................................................
....................................................................................................................................................................
....................................................................................................................................................................
..............................................................................................................................................................................
...................................................................................................................................................................
...................................................................................................................................................................................
..........................................................................................................................................................................................
.............................................................................................................................................................
........................................................................................................................................................................
..............................................................................................................................................................................
..................................................................................................................................................................................
...........................................................................................................................................................................
.............................................................................................................................................................................
.............................................................................................................................................................................
.............................................................................................................................................................................
...........................................................................................................................................................................
...........................................................................................................................................................................
.........................................................................................................................................................................
....................................................................................................................................................................
........................................................................................................................................................................
..........................................................................................................................................................................
......................................................................................................................................................................
..........................................................................................................................................................................
15-16. Learn Accept/Reject
.....................................................................................................................................................
.................................................................................................................................................................
.................................................................................................................................................................
.................................................................................................................................................................
......................................................................................................................................................................
................................................................................................................................................................
.......................................................................................................................................................................................
........................................................................................................................................................................
...................................................................................................................................................................................
................................................................................................................................................................
.........................................................................................................................................................................
........................................................................................................................................................................
..............................................................................................................................................................................
....................................................................................................................................................................................
.............................................................................................................................................................................
......................................................................................................................................................................
.........................................................................................................................................................
30. 2 Frequency (Best Data)
....................................................................................................................................................
31. 2 Frequency (Correlate)
.....................................................................................................................................................
..........................................................................................................................................................
..........................................................................................................................................................
........................................................................................................................................................
................................................................................................................................................................
..........................................................................................................................................................
.........................................................................................................................................................................................
...................................................................................................................................................................................
..........................................................................................................................................................................................
..........................................................................................................................................................................

Chapter 1 DFX™ Assembly
Trigger behind display activates
COIN PROGRAM
SQUEEZE & RELEASE TRIGGER
AFTER BATT. CHECK.
SCROLL OPTIONS
ATER BATT. CHECK USE
TO SCROLL CURRENT SETTINGS
OR MAKE ADJUSTMENTS
GROUND BALANCE ONL
Y
WHILE SEARCHING HOLD THE
TRIGGER AND PRESS
BATTERY CHECK
WHILE SEARCHING, HOLD THE
TRIGGER AND PRESS
REVERSE DISPLA
Y
WHILE SEARCHING. HOLD
THE TRIGGER AND PRESS
PRESS FOR
LIGHT/DARK BACKGROUND.
RELEASE TRIGGER
BACKLIGHT
IN SEARCH MODE, HOLD THE
TRIGGER AND PRESS
RELEASE TRIGGER
PUSH
AIR/GND BALANCE
IN SEARCH MODE PRESS
TO RE-AIR/GND BALANCE
VIEW ANGLE
WHILE SEARCHING HOLD
THE TRIGGER AND PRESS
RELEASE TRIGGER
PUSH
"HOT KEY" SHORTCUTS
for a few seconds then release.
Twist and insert each end of
top of shipping carton into
CLEVIS
LOWER

7.
Remove the protective paper from the two black
elbow cup foam pads. Carefully align pads on the
inside of the elbow cup, one on each side of the
center rod, and press fi rmly into place.
Adjust the elbow cup strap so that it is loose
enough for you to slide your arm in and out with-
out loosening each time you want to set the de-
tector down. The elbow cup strap provides extra
leverage and control. However, some prefer not to
Install battery as described in the next section,
tacts facing toward inside of battery compartment.
It should be noted at this point that the detector
may not work as expected indoors due to the high
degree of metals used in modern construction. It
is best to tune and practice out-of-doors to ensure
buried targets will not produce the normal depth
naturally lost and settled in the ground. Due to the
matrix, and the sophistication of the ground rejec-
tion circuitry, it may take a number of years for
freshly-buried targets to respond at true depths and
discrimination accuracy. The best way to determine
true detection depth is in real search conditions.
Your DFX is designed to automatically turn
itself OFF if the trigger on the grip (or another
control) “is not used” for a period of 10-15
minutes. This protects the battery from damage
in case of an accidental turn on during travel or
pinpointing/depth reading (keeping the detector
operating) is encountered during each 10 minutes
of search. If searching an unusually area, free
of concentrations of metal, the DFX will make
to resume searching. So long as a good battery
remains, the DFX returns to the exact settings prior
Remove all parts from shipping carton and
check the assembly page to make sure all parts are
There are rubber washers between clevis/lower
Use only nonmetallic washers,
fi ber bolt, and thumbnut to secure loop/search
coil to clevis/lower rod.
Unlock "S" rod camlock and insert clevis/low-
er rod into curved "S" rod so that stainless steel
the adjustment holes in the curved "S" rod. Turn
camlock to secure. The second or third adjustment
holes are suitable for average size adults. Individ-
uals 6' or taller should use the fully extended posi-
tion. Individuals well over 6' tall should purchase
Unravel loop cable and wind the cable around
the clevis and rod assembly, fi rst revolution over
the top of the rod. Wind cable all the way to the
top of the curved "S" rod, about fi ve revolutions.
Use the black cable retainers, one near the loop,
and one near the top of the curved "S" rod, to hold
Unlock control box rod camlock and insert
curved "S" rod so that stainless steel spring clip
buttons line up and lock into the rod on top of the
control box. The "S" rod is designed to curve up
toward the display. However, those who prefer to
assemble the "S" rod to curve down toward the
ground. Turn camlock to secure. Plug loop con-
nector into control box, screw lock ring to secure.
8.
9.
Chapter 1 DFX™ Assembly
Grip the instrument by the handle, with your
arm in the elbow cup with strap secure, and sweep
the loop/search coil over the fl oor. If the instrument fi t feels uncomfortable, adjust the elbow cup
by removing and repositioning the bolt/thumbnut
and installing in one of the optional positions. If
necessary, readjust clevis/lower rod length with the
spring clip buttons so that the search coil can be
held near the fl oor without requiring stooping over.

Chapter 2 DFX™ Batteries
The standard battery holder (blue decal) holds
eight “AA” cell batteries. Alkalines are recom-
mended for use with this battery holder.
Non-alkaline batteries can be used in this holder.
When non-alkalines or rechargeable “AA”
cells are used, detecting time (before replacement/
recharge) may be reduced.
"LOW BAT" will automatically appear on the
display when the batteries become too low to prop-
erly operate the detector.
The battery compartment opens by gently pull-
ing down on the front of each of the two latches
catch and hinging open the door.
The non-rechargeable battery holder can use many
different types of batteries, including rechargeable.
This holder is designed for standard size penlight
Battery lengths shorter than this will likely cause
problems with this power supply.
When the instrument is turned on the battery volt-
age will momentarily appear after the opening
display. The detector will then continue to the
ing operation, squeeze and hold the TRIGGER and
press the ARROW DOWN control.

Chapter 2 DFX™ Batteries
A rechargeable battery (green decal) is provided
with your instrument. This battery can be recharged
hundreds of times as long as the battery hasn't been
Full charge can be achieved anytime during the
discharge cycle. When using the QUICK charge
on the following page for charge time. A full charge
will last ten to fi fteen hours of normal use.
Battery life will vary with temperature, the number
of targets found, and the exact settings used. Six
hours is not unusual for extreme high performance
experienced extensive use.
is for USA ONLY. It is not
available on 220-240 volt.)
OVERNIGHT charge options. Always check the
position of this switch prior to charging. Always
follow the charge hours on the chart on the follow-
ing page when the QUICK charge setting is used.
Overcharging with the QUICK charge setting will
battery holder) by applying gentle upward
pressure on the tab of the door so that it unlocks.
exposing the cell positions.
Remove any old cells from the holder. Note the
and (-) for each position marked inside the cell tray.
noting carefully the cor-
rect (+) and (-) positions.
If the cells are installed incorrectly, the detector
may require service by an Authorized
Insert the battery holder into the detector so that
the decal is facing down, with the battery
holder door tab and metal contact points facing
toward the inside of the battery compartment.
Close the battery compartment door and secure
the two latches on the bottom of the case. Hook
the front of each latch fi rst, then press down on the

Chapter 2 DFX™ Batteries
The battery will lose its charge during stor-
age. If stored inserted in your instrument, this loss
will be more noteworthy. It is recommended that
the battery be removed from the instrument dur-
ing periods of storage. It is not advisable to store
rechargeable batteries for long periods of time
without use. If however, storage is necessary, store
without a charge (discharged).
Do not discharge the battery in devices other
than your metal detector. Unnecessary discharging
and/or an absolute discharge will reduce battery
life and may damage the battery. Unlike older re-
chargeable battery designs, the rechargeable battery
provided with your detector can be recharged at any
Regardless of whether or not it already has a
partial charge, memory will not occur.
White's has provided the leading edge of re-
chargeable battery technology with your instru-
ment. Disregard all advice which confl icts with the
above recommendations. Care for batteries pro-
vided by other manufacturers, or with other White's
battery's current condition.
condition by inserting bat-
tery into the instrument and
turning the instrument ON.
If the instrument will not
turn ON, or if voltage tests
eight volts or below, charge
fi ve hours with the QUICK
charge. If the battery voltage tests any other volt-
proper QUICK charge time.
To charge, insert the charger plug into the battery
pack jack, located near the plastic tab and
Plug the charger into a standard wall outlet. (110
Again, the QUICK charge setting uses the above
chart for a specifi c charge time. OVERNIGHT is
designed to charge the battery in as little as fourteen
hours. However, no harm will come to the system
leaving it charging for several days.
It is normal for the battery and charger to get
warm during use. However, if either the battery or
the charger gets too hot to hold or deforms due to
the heat, discontinue use and return for testing.

Chapter 2 DFX™ Batteries
temporarily holds any program
changes or settings not yet saved in a Custom Pro-
or volatile memory is retained so
long as a good battery remains in the detector. To
recover volatile memory immediately squeeze and
release the TRIGGER once the detector is turned
ON. If the battery is removed all volatile memory
memory (programs saved in
Custom Programs) is automatically saved for up to
ten years regardless of whether a battery is in the
When using fresh batteries, the voltage will initially
check somewhere in the 10 to 14 volt area. Unlike
will quickly drop to between 9 and 10 volts and
plateau there for most of its life. Once the recharge-
able battery voltage drops below this plateau, it will
quickly drop below a usable voltage level (eight
volts) and thus require a recharge.
automatically appear on the display when the bat-
tery reaches eight volts.
Like a personal computer, there are times (such as
low battery conditions) when the microprocessor of
a metal detector becomes out of sequence with the
rest of the circuitry. This is often noted by peculiari-
ties in the nondiscrimination or pinpointing (TRIG-
GER squeezed) modes. Symptoms may be blaring
or silent non-discriminate or pinpoint modes, depth
indication inaccuracies or general abnormal opera-
tion. To correct such diffi culties "re-boot" by:
2. Turn ON wait for MAIN MENU to appear.
3. Open battery door and remove battery while
4. Wait one minute, reinstall battery, turn
detector ON, and check for proper function.
Use of maximum backlight may reduce battery life
by up to 50%, depending on battery type.
Rechargeable batteries gradually deteriorate. As
they age they do not provide the life-per-charge
they did when new. This is expected, and not
grounds for replacement under warranty. Addition-
charging with the QUICK option,
use, or a defect due to a problem with a White's

Chapter 3 DFX™ Quick Start
With the TRIGGER in the center position,
press the ON/OFF control and an automatic
The display will momentarily show an
opening screen which lists the
The last automatic display screen to appear
You will be prompted to raise the search
coil (loop) to waist level. Press
asking you to lower the search coil (loop) to
ization will be balanced out.
The last screen will be the live search
ground and listen for a solid repeatable/con-
icons tell what likely coin lies below. V.D.I.
number/chart on top of control box and Sig-
naGraph® provide greater detail. Squeeze
the trigger for pinpointing and depth and it's
1
3
4
2
After you have assembled the DFX™ and inserted the
battery pack, follow these simple steps to start
treasure hunting!
1
2
3
4
5
5
6
7
8
6
7
8

Chapter 3 DFX™ Quick Start
The loop/search coil must be in motion (sweep-
ing from side-to-side) for this instrument to
Practice a smooth sweep of
the loop from side-to-side keeping the loop close
to the ground throughout the swing.
the loop should take approximately two seconds
from right to left, two seconds to return from left
Walk forward slowly. Take small steps no greater
than half normal strides. Make sure each pass of
the loop overlaps the last by at least half the length
of the loop. Do not lift the loop at the end of each
To become comfortable with sweeping the loop
takes some practice. Try to loosen up and fi nd a
comfortable grip on the handle. Premature fatigue
may result from gripping the handle too tightly, im-
properly adjusted rod or elbow support, and limited
body movement. Hold the handle loosely. Adjust
the rod and elbow support for comfort and keep the
elbow strap loose. Use your arm, shoulder and even
your back a little to allow a smooth even sweep of
Now that you're sweeping the loop smoothly over
the ground, you will notice that the detector starts
targets; some trash targets also make the detec-
As the loop is swept over the ground, ignore the
display and concentrate on the sounds the detector
As the loop is passed over metal that is likely trash,
the sound will be inconsistent. Trash targets typi-
cally produce a shorter, sputter-type sound, that
is often broken or double in nature. Place a steel-
pop bottlecap on the ground. Pass the loop over it
different loop sweep speeds. Note that an aluminum
be used as it is a different
type of target. Also note that very old rusty bottle
caps may start reading as quarters due to the elimi-
nation of the iron alloy through deterioration. Once
familiar with the sound typical bottle caps produce,
tion, saving more time for evaluating possible good
As the loop passes over metal that is likely a good
target, a more consistent and smooth sound will be
heard. A good target typically produces a longer,
more solid sound. Place a quarter on the ground
familiar with the sound of a good target.
When the display prompts you to AIR BALANCE
by holding the loop at waist level and press EN-
ground balancing by measuring temperature and
other variables that affect electronic circuits. The
"beeps" and you lower the search coil to the
distance above the ground that you will be search-
ing. Press ENTER to have the DFX
out" or GROUND BALANCE the ground miner-
then automatically "tracks
out" the varying mineralization as you continue to

Chapter 3 DFX™ Quick Start
In the upper left hand-side of the display there is
a V.D.I. number that corresponds to the V.D.I.
control box. It also corresponds to the Discriminate
Edit feature allowing you to reject or accept targets
based on their V.D.I. reference number. There are
and "-" numbers for ferrous (iron) targets. Rejected
V.D.I. numbers may not appear if the VISUAL
DISCRIMINATION feature is ON. Reasonably
consistent V.D.I. reference numbers (± fi ve digits),
in a desirable area of the chart is a vote for digging
2. Possible Target Identities ("Probable or most
To the right of the V.D.I. number, possible target
identities will be represented graphically. These
graphics are called ICONS. A fairly consistent indi-
cation of a desirable target is another vote to dig
the target. One or two possible target icons may
appear. There is signifi cance to which icon appears
fi rst. The fi rst target to appear is always the most
likely, the second is another possibility, slightly less
at the bottom of the display
provides a fi nal vote as to whether or not the target
Sweep the loop over the target several times
doesn't fi ll the screen with information from past
loop sweeps. An operator has limited time to look
information again, sweep the loop over the target
information can be slowed or speeded (FADE
RATE) to operator preference. This is completed
in the PRO OPTIONS under DISPLAY. Automatic
AVERAGING and/or ACCUMULATING of Sig-
information is also available (See PRO
Valuable targets will show up on the positive
the section located to the right of the zero.
Look for consistency. In ideal conditions, coins
zero. Trash produces several bars, sometimes on
In less than ideal conditions, coins may produce
recognizably different pattern than valuable targets.
One of the most visual benefi ts of the Signa-
on iron targets that often fool the other methods
of identifi cation. An iron target will likely show
defi nite bars on both the negative and positive sides
produce such obviously wide patterns. In very bad
ground conditions, a good target may have a few
eralization. However, the pattern will show mostly
positive bars, in a fairly narrow tall group.

Chapter 3 DFX™ Quick Start

Chapter 3 DFX™ Quick Start
Once the decision has been made to dig, move the
loop off to one side of the target area, squeeze and
hold the TRIGGER on the handle, and "X" the loop
over the spot where you believe the target to be.
Note that the TRIGGER also has a locked for-
ward position that accomplishes the same thing as
While the TRIGGER is being held, the loop doesn't
need to be moving to detect the target. The loop
may be moved slowly over the area. The display
will indicate depth in inches and will also show
the strongest reading to aid in pinpointing exactly
where to dig. The shallowest reading on the depth
display, the loudest sound coming from the speaker,
the center of the target. Don't forget to "X" the
target as pinpointing cannot be accurate unless the
target is swept from at least two different directions.
Once pinpointing is complete, release the TRIG-
GER, or return it to the center position.
Pinpointing takes practice. The standard loop pro-
vided with the DFX™ is a high-powered, 12 inch
design. This harmonically tuned loop's strongest
traits are in the detection depth and ground cover-
cal refer to the Advanced Pinpointing Techniques
Advanced Pinpointing Techniques
Targets that are near the surface, because
they give a wider response, are harder to
and the loop swept over the area, you may
slightly above the ground, releasing and re-
squeezing the TRIGGER and again "X" ing the
target will aid pinpointing.
In the Basic Adjustments, DC Sensitivity
In the PRO OPTIONS under AUDIO, V.C.O.
The depth reading has two indication bars.
The top bar shows the current distance from
the target, and the bottom bar shows a memory
of the strongest reading. These two bars will be
even with each other when the loop is directly
over the center of the target.

Chapter 3 DFX™ Quick Start
- Prior to searching and digging you
must have permission to search private property,
from the owner or caretaker.
- Know the laws that apply to the area you
are going to search. Laws vary a great deal with the
City, County, State, and Country, regarding the use
of metal detectors. Be respectful of private prop-
erty, public property, and the laws which govern the
- Care must be taken to dig in a way that is
friendly to the landscape. Tools and methods vary a
great deal with the area, season, and types of target
you are recovering. Check with your dealer for
recommended tools and methods for your area.
- When searching, remove all trash you come
across. This not only makes your future searches of
the area more productive; it promotes the hobby of
Your dealer knows of metal detect-
ing clubs and organizations which promote and
protect the hobby. A club is a great way to not only
learn good detecting habits, but to gain permission
to search areas as a group as well as have organized
Reached from the MAIN MENU, the factory PRE-
general purpose settings, discriminates (re-
jects) most common junk items like nails, foil, pull
tabs, and hot rocks; and responds to most coins and
large jewelry. Use in lawns, parks, and playgrounds
where lots of trash rejection is desired.
less discrimination (less trash
rejection), desirable because of the high degree of
variance found in jewelry alloys. More digging
required. Good program for lawns, parks, and play-
grounds. Use screen in conjunction with sound.
even less discrimination than Coin & Jew-
elry or Jewelry & Beach, all types of metals except
Brass, lead, aluminum, as well as copper, silver,
will also respond . Suitable for all signifi cant targets
NO AUDIO DISCRIMINATION. All
only for metals that could be gold. Dig only V.D.I.
number (possibility gold) targets and avoid iron.
Targets which cause an audio response, without
causing a V.D.I. number to appear on the display,
frequency gold-shooting detectors will respond
better, this mode will offer good results for the oc-
casional nugget hunter by responding to nuggets in
the six-grain and heavier category.

Chapter 4 DFX™ Basic Adjustments
After you have had some fi eld experience, you may
want to make some changes to the basic settings
From the search mode press
pear on the display. Use the ARROW controls to
move the pointer to Basic Adjustments,
through the Basic Adjust-
Using the fi rst adjustment screen (TARGET VOL-
UME) as an example, the screens with a
then use the ARROW up and down controls to
Note: when ENTER is pressed the square
After adjusting press MENU
the ARROW controls to continue viewing/setting
other Basic Adjustments, or squeeze and release the
TRIGGER to begin searching
Adjustment screens with an on/off selection need
only for you to press ENTER to change setting.
Pressing ENTER again changes back to the original
decrease to minimum, hold
How loud a target beeps when detected.
The slight hum or background sound heard continuously during searching.
3. TONE (AUDIO FREQUENCY) -
Selects the frequency or pitch of sound the detector produces.
4. AUDIO DISCRIMINATION -
The ability to reject trash, different sounds for different types of targets.
The ability to operate without the threshold or background hum.
DC non-discriminate mode, working simultaneously with AC discrimination mode.
Degree instrument is responsive to signals in the discriminate (motion) modes.
Degree instrument is responsive to signals in non-discriminate (non-motion)
Used in dark conditions to light the display improving visibility.
Adjusts the display for low or high temperature visibility.

Chapter 4 DFX™ Basic Adjustments
All the MENU items are tied together so that the
ARROW up and down controls scroll through every
adjustment screen. If you continue to press the
you can go beyond the last BASIC
ADJUSTMENT (View Angle) and into the PRO
after VOLUME, you will be scrolling backwards
through the options starting with the end of the Pre-
An important feature of the ARROW controls; If a
BASIC ADJUSTMENT has been made (for exam-
ple Volume) and the trigger has been squeezed and
released to return to a search mode, you can return
to the volume adjustment simply by pressing either
of the ARROW controls. This shortcut returns to
the last adjustment that was made thereby allowing
an operator to switch directly from a search mode
to the adjustment currently being fi ne tuned. This
feature is desirable as you start using BASIC AD-
JUSTMENTS or PRO OPTIONS that are located
further down the menu listings, or any adjustment
that may require some trial and error to fi nd the ap-
If care is taken to use a desired adjustment screen
last (just prior to squeezing and releasing the TRIG-
GER for a search mode), Custom Programs can use
this ARROW RETURN feature to allow quick easy
access to the most used feature. Use that feature
releasing the TRIGGER for searching. Then during
control box for fi eld reference.
must be turned on and air/ground
balanced in any mode, then turned off. "HOT KEY"
Squeeze & release TRIGGER
After battery check, use AR-
ROWS to scroll all the current settings /menus.
press ENTER to re-Air/Ground Balance.
hold the TRIGGER and press ENTER.
While searching, hold the
release TRIGGER to return to searching.
the TRIGGER and press ARROW DOWN. Press
ARROWs for light/dark background. Light or dark
background will not change battery life. It will
make the display easier for some to read, particular-
ly in certain light conditions. It will work in com-
bination with backlight. Reversed display is only
In search mode, hold TRIGGER
While searching, hold the TRIG-

will activate a prior custom pro-
gram stored in that position. After you have
LOAD and press ENTER, to use that program.
saves your current settings in that
custom position with either a generic name or
is the preferred method. Select
NAME and press ENTER. You may now use
the ARROW and ENTER controls to name
your custom program. NAME automatically
Saving Basic and Pro Option
adjustments for future use.
program positions. They will remain permanently
memory regardless if the machine
is turned off or the battery removed. Custom EE-
PROM Programs can be changed at any time by
program. EEPROM is a special type of computer
memory made for this purpose, (Electrical-Eras-
able-Programmable-Read-Only-Memory).
Chapter 4 DFX™ Basic Adjustments
Once all of the changes you desire have been
made to any Preset Program or existing Custom
Program, squeeze and release the TRIGGER as if to
Use the ARROW controls to select one of the
four Custom Programs then press ENTER.
You now must make one of three choices (use the
ARROW controls to make your selection):
Once all of the changes you desire have been
You now must make one of three choices (use the
4
1
2
1
2
3
3

Chapter 4 DFX™ Basic Adjust-
use the ARROW controls to select
the fi rst symbol, number, or letter of the name and
press ENTER. Use the ARROW controls to select
the second symbol, number, or letter of the name,
press ENTER. And so on using up to sixteen digits.
To leave a space, use the ARROWS to select the
point where no symbol or letter appears and press
ENTER. If you make a mistake and press ENTER
when the digit is not as you desire, simply keep
pressing ENTER until that digit is again fl ashing,
then use the ARROWs to select the correct digit and
again press ENTER. It is wise to name the custom
program something that relates to what it is used
for. For example "TRASHY PARKS", "GHOST
TOWN", "NIGHT HUNT", "COMPETITION", etc.
Once the name is fully assembled press MENU.
Once you have SAVED and pressed ENTER, or
NAMED and pressed MENU, there are four direc-
Squeeze and release the TRIGGER to con-
tinue searching using your new custom program.
Press ENTER, select LOAD and press
ENTER to continue searching using your
Press MENU to return to choose or develop a
different program than what you stored.
When the detector is turned back on, regardless
of whether a battery pack was left in the detector or
not, your custom program will be ready for you to
use again and again. Simply select it, press ENTER,
on-screen instructions for Air/Ground Balance and
If you SAVE or NAME a program, then decide
you no longer want to keep it, you can replace it
with a new program using the same procedure
erased when a new program is stored in that posi-
You can NAME a custom program and at a later
date replace the program while maintaining the
gram, then use the SAVE method which maintains
the old name while storing the new program. To
keep the same program with a new name, fi rst
LOAD that custom program, Air/Ground Balance,
then press MENU and go to that custom position
ENTER. You can now develop a new name for the
- When a Custom Program is
that program is also stored. This has advantages
particularly for those who manually set the Ground
Balance for speciality applications. The automatic
Air/Ground Balance sequence will always override
manual settings. To access the last Ground Balance
desired Custom Program then press ENTER. Select
LOAD and press ENTER. Do not Air/Ground Bal-
release the TRIGGER. The last Ground Balance
Ground setting is not available, the instrument will
- The last Basic Adjustment
or Pro Option screen used is remembered by your
Custom Program. From the search mode, either
ARROW control will access the last Basic Adjust-
ment or Pro Option screen used. This allows easy
Factory Preloaded EEPROM PROGRAMS
described on pages 53 and 54.
4

Chapter 4 DFX™ Basic Adjustments
controls and press ENTER. The current volume
level sounds continuously. The number designating
the current level is shown on the right side of the
display. To the left, the graphic knob indicates the
relationship of the current setting to minimum and
Use the ARROW controls to select the volume level
you desire. Volume level will select the loudest
possible sound a shallow target can produce. High
volume levels will slightly reduce battery life.
Press MENU and use the ARROWS to continue
viewing and/or adjusting Basic Adjustments, or
The slight hum or background tone which is nor-
mally heard continuously during searching.
will sound continuously. The number designating
the current level is shown on the right side of the
display. To the left the graphic knob indicates the
relationship of the current setting to minimum and
maximum levels. Note that the maximum threshold
level (42) is well below the minimum VOLUME
level. Thus with the THRESHOLD at maximum,
old levels will slightly reduce battery life.

Chapter 4 DFX™ Basic Adjustments
3. Tone (Audio Frequency)
controls and press ENTER. The current TONE
will sound continuously. The number designat-
ing the current level is shown on the right side of
the display. To the left, the graphic knob indicates
the relationship of the current setting to minimum
dio frequency that you can hear comfortably and
provides the best defi nition for your ears. Press
The ability of the detector to reject trash by
producing different sounds for different types
of targets. Trash is rejected by going silent or
producing a broken "cut-short" sound. Valuables are
detected by a smoother more solid sound.
use the ENTER control to turn AUDIO DISC. ON
or OFF. When ON, specifi c targets will be accepted
or rejected based on the Program currently in use.
Audio Disc. turns ON or OFF the entire audio
discriminate feature. When OFF, all types of metals
under Discrimination, can specifi c targets (V.D.I.
numbers) acceptance or rejection criteria be altered.
detection of all types of

Chapter 4 DFX™ Basic Adjustments
The ability of the detector to be operated without
the threshold or background hum that is normally
heard continuously during operation. The
instrument is silent until a target is detected.
controls and use the ENTER control to turn
To function, SILENT SEARCH requires AUDIO
DISC ON and MIXED MODE OFF. SILENT
AUDIO DISC and MIXED MODE menu selection
to perform the silent search function regardless
of your AUDIO DISC and MIXED MODE
In Pro Options the Discriminate feature can
be used to accept all metal targets while using
ON the all metal pinpointing mode continues to
produce a threshold. This may not be noticed,
target center locating the threshold is not present.
However, releasing, re-squeezing, and holding the
TRIGGER with the loop at waist level a threshold
will be noted. Press ENTER.

A unique hybrid operating mode. It is an all-
metal (DC non-motion, non-discriminate) mode,
working simultaneously with a discriminate (AC
motion discrimination) mode. It is two modes, one
detecting everything and another discriminating,
operating at the same time.
press ENTER control to turn MIXED MODE ON
AUDIO DISC needs to be ON and SILENT
perform properly. See chart on (page 20). MIXED
MODE ON will automatically override AUDIO
DISC selections to perform the MIXED MODE
When Mixed Mode is on, all types of metals will
Discrimination Channel - When the loop is
Discrimination Channel - When the loop is
targets accepted by the discriminate
program will produce a high-pitched
Targets rejected by the discriminate program will
are directly determined by the
discrimination settings. An operator can select
discriminate settings through the selection of an
entire Program or by adjusting the accept and
reject V.D.I. numbers in the Pro Options under
All-Metal Channel -When the loop is not in
All-Metal Channel -When the loop is not in
will produce the same low-pitch
metal channel will by nature detect deeper than
the discrimination channel. Deeper targets will
produce a lower volume sound when the loop is
moved slowly over the area.
gain extra depth by monitoring
the all-metal and discriminate
checking depth and digging
discriminate channel alone.
about the target, Pro Options
TONE I.D. and or V.C.O. can
Chapter 4 DFX™ Basic Adjustments

Chapter 4 DFX™ Basic Adjustments
Used to select the appropriate sensitivity (degree
that the instrument is responsive to signals) while
being used in the discriminate modes (those which
require movement of the loop).
controls, and press ENTER. Use the ARROW
controls to set the level of sensitivity shown by the
number on the right. Press ENTER.
have a direct effect on detector stability. A.C.
to allow stable, predictable performance. Set a
lower level if the detector behaves erratically.
well for most conditions.
detection depth if stability
from 1 - 85 are availabLE.
Used to select the sensitivity (degree that the
detector is responsive to signals) while the
detector is being used in non-discriminate (ALL-
METAL) modes. These are modes that do not
require movement of the loop to respond. D.C.
controls, and press ENTER. Use ARROW
controls to select the desired D.C. SENSITIVITY
level shown by the number on the right. Press
ENTER. D.C. SENSITIVITY levels should be
predictable operation while allowing for reasonable
A.C. and D.C. Sensitivity Adjustments are
traditionally the way to alter detection depth and

Used in dark conditions to light the display,
to select the desired BACKLIGHT level. The
BACKLIGHT level will be visible on the display.
The current level is shown on the right side of
the display. The graphic control knob shows the
relationship of the current setting to minimum
The maximum backlight setting will reduce
battery life by as much as 50% depending on the
type of batteries and how long it is used. Lower
BACKLIGHT settings will have signifi cantly less
When the detector is fi rst turned on, it is
normal for the backlight to be on during the
opening display and BATTERY CHECK. If the
BACKLIGHT is off, it will fade when the MAIN
MENU display appears. If the BACKLIGHT is
ON, it will continue until turned off manually or a
different program is selected. BACKLIGHT can
be saved as part of a custom program, for example
When Backlight is ON and the TRIGGER
is squeezed and released to begin searching,
the display to warn you of the extra battery duty.
you cannot see the display to turn the
BACKLIGHT on, holding the TRIGGER and
pressing MENU will bring up the BACKLIGHT
to select a level you can see the
display. Squeeze and release the TRIGGER to
as needed, for acceptable
display visibility. Backlight
If the instrument is turned ON and the
EMERGENCY BACKLIGHT sequence is
used, the BACKLIGHT will stay ON only
while you stay in that program. Pressing
MENU and selecting another program will
turn BACKLIGHT OFF, if BACKLIGHT
is not also ON in that particular program.
If in the dark at the time the instrument
is turned ON, you may need to squeeze
use the EMERGENCY BACKLIGHT
you desire, press ENTER, press ENTER
for Air Balance, and press ENTER for
Ground Balance. Use the EMERGENCY
BACKLIGHT ON sequence a second time
if the BACKLIGHT fades in that program.
BACKLIGHT is no different than
term volatile memory. However, factory
preset programs use the OFF (0) setting
Chapter 4 DFX™ Basic Adjustments

Adjusts the display for visibility in low or high
trols and press ENTER, use the ARROW controls
to make changes. The current level is shown on
the right side of the display. The graphic control
knob indicates the relationship of the current set-
ting to minimum and maximum levels. Squeeze
VIEWING ANGLE has no impact on battery life.
EMERGENCY VIEWING ANGLE PROCE-
heat prior to use, you may not be able to see the
display to adjust VIEWING ANGLE. Press the
ON/OFF control, hold the TRIGGER and press
trols to fi nd a VIEWING ANGLE level that allows
you to read the display. Squeeze and release the
TRIGGER to begin searching. Like the BACK-
LIGHT, you will lose your custom VIEWING
ANGLE setting if you change Programs. You
may need to use the EMERGENCY VIEWING
ANGLE PROCEDURE to see the display. Select
the program you desire, use the ENTER control
to enter, Air/Ground Balance, then again use the
EMERGENCY VIEWING ANGLE PROCE-
DURE if the display is unreadable. VIEWING
ANGLE is preset at average levels in the factory
preset programs. The display may be unreadable at
either of the extreme settings in a particular envi-
ronment. Custom VIEWING ANGLE settings will
be saved when Custom Programs are stored for
the display typically will be-
come slower at responding.
display and improves vis-
display may become diffi cult
to see. Settings toward MIN
tions result throughout the
day or nights search, you
display visibility. Settings
from 1- 50 are available.
Chapter 4 DFX™ Basic Adjustments

Chapter 5 DFX™ Pro Options
Pinpoint feature, automatically detunes for center location.
Self Adjusting Threshold or Auto-tune, automatically maintains threshold.
Assigns each V.D.I. target number its own special tone or sound.
Pinpoint or non-discriminate feature, increases pitch or tone with target strength.
Motion modes produce the same, or different volume, based on target depth.
Automatically updates Ground Balance during searching.
TRACK appears on right side of display during AUTOTRAC
Positive or negative AUTOTRAC
Prevents tracking the ground during target detection.
(Manual Ground Balance) Coarse viewing, or overriding automatic.
(Manual Ground Balance) Fine viewing, or overriding automatic.
Change V.D.I. (target reference numbers) accepted (detected), or rejected status.
Speeds EDIT by dragging ACCEPT or REJECT with ARROW controls.
Target samples can be used to show or teach ACCEPT discrimination.
Target samples can be used to show or teach REJECT discrimination.
Speeds target responses, so close together targets each respond.
How strongly the instrument rejects or breaks up on iron.
Degree the instrument is responsive to signals in the +95 (hot rock) category
Adjust signal sampling width thus the ideal loop/search coil sweep speed.
How much circuitry (high-pass fi ltering) used to separate ground/trash signals and targets.
Rejected V.D.I. numbers and ICONS do not appear on display.
Graphic display representation of metal targets, ON/OFF.
Response intensity to produce a display indication & 3rd V.D.I. digit@ 86 and higher.
Measurement of ground, or metal target, during pinpointing.
information collects over multiple loop passes.
Emphasizes common or predominate SignaGraph
Clears or fades noncurrent SignaGraph
Selects the intensity of the signal received from the loop.
30. 2 FREQUENCY (BEST DATA) -
Transmits and processes at both 3kHz and 15 kHz frequencies (salt eliminated).
Automatically chooses data from the most reliable frequency (based on both magnitude and phase) for each specifi c target.
31. 2 FREQUENCY (CORRELATE) -
Transmits and processes at both 3kHz and 15 kHz frequencies (salt eliminated).
Compares data at 3 kHz and 15 kHz. Target signals that do not provide reasonably predictable information at both are automati-
cally rejected. Iron typically doesn't compare predictably between frequencies, improved iron rejection can be expected.
32. V.D.I. (NORMALIZED) -
2 Frequency modes (Best Data and Correlate) automatically have V.D.I Normalized ON
as it is required for these modes to operate predictably. Differences at 3 kHz and 15 kHz skew the well known traditional (6.59
kHz) V.D.I. chart/numbers painted on the top of the DFX control box. Normalization recalculates signals for this traditional
V.D.I. number chart. OFF expands/compresses portions of this V.D.I. scale (depending on the 1 frequency used.).
Operates at 3 KHz providing for superior high iron mineral use (no salt present) particularly
for high conducting silver/copper alloys. Normalized ON maintains traditional (6.59 kHz V.D.I. chart/numbers. Normalized OFF
34. 1 FREQUENCY (15 kHz) -
Operates at 15 kHz providing in lower mineralized areas (no salt present) particularly
for lower conducting gold/nickel alloys. Normalize OFF signifi cantly expands lower end of V.D.I. chart/numbers compressing

intricate adjustments available on this model. The
PRO OPTIONS are divided into fi ve major cat-
egories of menus, structured similar to the Basic
Adjustments. Methods of entry, adjustment, exit,
To enter the PRO OPTIONS from a search mode,
press MENU and the MAIN MENU will appear.
Use the ARROW controls to select PRO OPTIONS
PRO OPTIONS menu (PG. 1/2
The ARROW controls are then used to select the
desired PRO OPTION category.
The six major categories have options specifi c
to their titles. For example, all the options under
AUDIO have to do with the way the audio circuits
of the instrument behave. Once a category has been
ROW controls can then be used to scroll through all
the options even beyond that category i.e., beyond
PRO OPTIONS, back to PRESET PROGRAMS,
only reappear if MENU is again pressed, or at spe-
cifi c times during scrolling (to offer short cuts).
From this point forward, we will assume that you
know how to use the MENU control, the Arrow
Controls to make a selection, the ENTER control
to enter or select that option, the Arrow Controls to
make adjustments, and the MENU to exit. Remem-
ber, squeezing and releasing the trigger returns to a
from the search mode to the last menu or adjust-
Chapter 5 DFX™ Pro Options
Be aware that changes you make to a
Program are only in effect as long as
you continue using that Program. If the
detector is turned OFF, the Trigger must
be squeezed and released to recover short
term (volatile) memory upon turning the
detector ON. If you want to keep the
changes you made to a specifi c program
over an extended time period (days or
or through battery changes
or through battery changes
entire Program must be saved in a Custom

Chapter 5 DFX™ Pro Options
Pinpointing feature, automatically detunes for easy
ON - Automatically aids in pinpointing. When the
TRIGGER is squeezed and held for pinpointing,
RATCHET PINPOINTING will shrink the size of
the signal as the loop is passed over the target area
unless the loop is moved too far away from the
OFF - When OFF, the signal will remain original
Manual shrinking is completed by squeezing and
releasing the TRIGGER several times as the loop
is passed over the target. Manual shrinking can
completely eliminate the target. If the target no
longer responds after manual shrinking, move the
loop away from the target, squeeze and release the
TRIGGER, and again sweep the loop over the area.
If the target is a metal, it should again respond.
ting until experienced at
THRESHOLD, OFF (0), or ON (1-10).
Without S.A.T. SPEED (a setting of 0), changes in
the ground (and the passage of time) will tend to
produce changes in the THRESHOLD sound. The
TRIGGER will have to be squeezed and released
repeatedly to maintain the THRESHOLD. This is
particularly noticeable in non-discriminate modes,
although S.A.T. SPEED will also tend to add
metal detectors dating back to the 1970's. It adds
a loop-motion requirement to modes which are
ordinarily non-motion. It is also known to produce
more mineralized than the surrounding ground)
which change with different speed selections.
Only enough S.A.T. SPEED required to maintain
beach hunting and relic hunting require slightly
faster speeds (higher numbers), and prospecting
requires the fastest settings (highest numbers).
greater stability particularly in non-discriminate modes or
when sensitivity levels are maximized. Typically, faster
conditions. Settings from 0 - 10 are available.

Chapter 5 DFX™ Pro Options
Assigns each V.D.I. number its own distinct tone or
pitch. Target ranges can easily be identifi ed by their
pitch of their sound. Reject targets still break up or
Tone I.D. is used in the discriminate or motion modes.
When ON, each V.D.I. target number has its own
audio frequency or pitch (191 different pitches). The
higher the V.D.I. number, the higher the pitch. Where a
target indicates on the V.D.I. scale can be immediately
recognized. The sound that rejected targets produce
will still be canceled or modifi ed (broken up) by the
The 191 different pitches or tones activated with
TONE I.D. cannot be adjusted as to their pitch. Each
V.D.I. number's sound is predetermined by the factory
and is not adjustable, nor will they shift with TONE
elry, and relic searching.
Voltage Controlled Oscillator produces a higher pitched
tone the stronger the target becomes.
Voltage Controlled Oscillator is a feature that will only
work in the non-discriminate modes. When it is ON,
the stronger the response the higher the pitch of the
function when the trigger is squeezed for pinpointing
when activated in a discriminate mode. It will func-
tion full time in the non-discriminate (ALL-METAL)
D.C. SENSITIVITY settings will dramatically change
the performance of V.C.O. D.C. SENSITIVITY set-
tings above a level of 35 may peak the audio pitch too
mended, particularly in trashy areas where targets are
able for pinpointing. Also
can work well in combina-
the non-discriminate chan-
Allows motion modes to produce different volume
levels based on target depth.
Modulation is used in the discriminate or motion
modes. When ON, it allows the depth of the target to
dictate the volume of the response. Thus deep targets
can be easily recognized by their lower-volume sound.
If OFF, the discriminate or motion modes will produce
the same volume of response on all detected targets
regardless of depth. The use of MODULATION al-
lows for the singling out of deep targets in the standard
discriminate mode. It may save time by eliminating
the checking of each target with the depth indication in
out deep targets by their
default (standard setting)
for all factory Preset Pro-
grams. If your hearing isn't
the best, the OFF setting is

allows the instrument to automati-
cally readjust the Ground Balance setting during
occurring changes in the ground mineral of an
detection depth and stability are enhanced. AUTO
is recommended for typical search condi-
tions. If, however, a great deal of decomposed
man-made iron is encountered, AUTOTRAC
operation is affected by TRAC
When ON, "TRACK" appears on right side of dis-
TRAC VIEW offers a way to observe Autotrack-
ing. When ON, "TRACK" will appear momentarily
on the right-hand side of the display just above the
This information is valuable when attempting to
determine an appropriate Trac Speed. It may also
be a valuable aid in determining the relative ground
conditions. Reoccurring "TRACK" would indicate
diffi cult ground. In average ground it is desirable
Chapter 5 DFX™ Pro Options

Chapter 5 DFX™ Pro Options
Trac Speed allows adjustment of the amount of
ground mineral change required to cause Auto-
tracking to occur. At slower speeds (lower num-
bers) it takes a signifi cant change in the ground to
cause tracking to occur. At faster speeds (higher
numbers) it takes very little change in the ground
mineral to cause tracking to occur. The end result,
more tracking occurs at higher numbers than at
lower numbers. Too much tracking can cause errors
in the ground balance setting. Not enough track-
ing can result in the ground balance setting never
catching up with changing ground. TRAC VIEW
is used to see how much tracking is occurring.
Generally, a faster (higher number) TRAC SPEED
is needed for more consistent ground conditions.
for more inconsistent ground conditions. "TRACK"
appearing every three to fi ve sweeps of the loop is
the ideal setting when the ground conditions will
every three to fi ve sweeps
of the loop however, this
than this desired standard.

either positive or negative in relationship to the
correct setting (under, or over kill).
TRAC OFFSET allows the AUTOTRAC
to track the ground slightly more or less than what
would normally be considered perfect, mostly for
experts that prefer a slight offset.
A slightly positive offset can be used to enhance
the responses of small metals (gold nuggets) in
highly-mineralized ground. Positive is indicated by
es the ground in the all-metal mode. Discrimination
A negative offset may be used to eliminate par-
ticularly troublesome "hot rocks" in areas that are
otherwise diffi cult to search. Negative offset is in-
dicated by a loss or slight decrease in the threshold
tors who fully understand
ground rejection. Settings
from -10 - +10 are avail-
TRAC INHIBIT ON prevents the AUTOTRAC
feature from altering Ground Balance during the
detection of targets. This prevents the possibility
of tracking to the corrosion associated with most
metals. TRAC INHIBIT ON is recommended for
most searching conditions. Some metals do not
corrode (GOLD) and since tracking is extremely
important in high mineral conditions, the ability
to turn TRAC INHIBIT OFF is available. OFF is
conditions, OFF for pros-
Chapter 5 DFX™ Pro Options

rent automatic Ground Balance setting (Air/Ground
Balance) and/or manual overriding of the automatic
COARSE & FINE G.E.B. (Ground Exclusion
Balance, or ground rejection) allows manual over-
ride of the automatic balancing features to select a
major adjustments. FINE allows minor adjustments
Before attempting manual Ground Balance adjust-
Balance will begin at the setting selected by the
Automatic Air/Ground Balance performed after
manual adjustment will cancel the manual setting.
To maintain a manual setting, the Automatic Air/
Ground sequence must be avoided by squeezing
Balance instruction appears. Air Balance instruc-
tions may appear during Basic Adjustment or Pro
Options selections following the manual adjust-
When manually adjusting the Ground Balance, it
is necessary to make an adjustment in COARSE
GER. The TRIGGER can be held in, and the loop
lowered and lifted over the ground to check for a
change in sound which would indicate imperfect
Ground Balance. To return to the Ground Balance
display for further adjustment, press either of the
ARROW controls. When perfectly Ground Bal-
als or hot rocks. Settings
from 0 - 255 are available.
Chapter 5 DFX™ Pro Options

Chapter 5 DFX™ Pro Options
FINE G.E.B. is used to make smaller adjustments
to the Ground Balance range on either side of the
COARSE setting. The FINE setting combines
with the COARSE setting to determine the actual
ground rejection level. To determine exactly what
level the automatic Air/Ground sequence selected,
it is necessary to check both COARSE and FINE
Manual Ground Balance will rarely be used for
perfect balancing as the automatic balance is
extremely reliable. It is generally used to off-
In COARSE or FINE, lower numbers indicate a
Ground Balance setting toward the negative side
of the V.D.I. scale (iron). Larger numbers indicate
a Ground Balance setting toward the positive side
of the V.D.I. chart (salt). Bad ground conditions
(negative or low numbers) or
(positive or high numbers). COARSE and
FINE settings of 0 represent a ground rejection
FINE settings of 255 represent a ground rejection
level similar to a V.D.I. number of +10. This V.D.I.
range from -95 to +10 represents the ground rejec-
tion range of this instrument. If ground balance
COARSE setting, signifi cant loss in sensitivity can
be expected to targets which indicate in the V.D.I.
number range of 1 to 10. However, manual ground
balance doesn't exactly work like a discriminator,
where V.D.I. numbers are simply REJECTED. The
actual phase of Ground Balance is out of sequence
with most targets; thus some responses will oc-
cur even if COARSE & FINE G.E.B. are matched
exactly to a particular targets V.D.I. number.
adjusted to the top of the
scale, 255, will automati-
justment can continue un-
to the bottom of the scale,
are available, coarse and

Chapter 5 DFX™ Pro Options
Allows for changing which targets V.D.I. or refer-
ence numbers are ACCEPTed (
the ARROW controls to view the V.D.I. numbers
from negative -95 to positive +95 that appear on
the right side of the display. These are the same
reference numbers that are listed on the top right-
hand side of the control box (V.D.I. SCALE or tar-
get chart). For each V.D.I. number, a √ will appear
to the left indicating whether the current program
will ACCEPT or REJECT targets that display that
Use the ENTER control to change any desired
V.D.I. number to ACCEPT or REJECT status.
If you do not save the entire Program into a Cus-
tom Program, the changes you make in EDIT are
only in use as long as you continue to use that
program. The only exception being the temporary
then switch to COIN & JEWELRY, you lose all
editing you completed in the Coin Program. Each
time a factory Preset Program is ENTERed, all
unsaved settings are overwritten by the factory
recommended settings. Only the Custom Programs
will save your custom settings through Preset Pro-
Accepting a wider range of V.D.I. numbers will
have an impact on detection depth, particularly for
very deep targets that are diffi cult for the display
to identify. Positive +95 is one of the most signifi -
cant V.D.I. numbers. Some Preset Programs are set
The display uses positive +95 for many different
than any other feature. Audio Discriminate
ON/OFF is located in the BASIC
exceptions to the keystroke rule in that
they do not have a graphic control knob.
setups already defi ned and
of these factory programs.
specifi c or unusual targets
you may want to either AC-

types of targets it cannot readily identify. There-
fore, if positive +95 is ACCEPTed, questionable
targets may be dug that indicate this number and
can produce some interesting recoveries.
tip regarding positive +95: If a target indicates
95, and the depth display indicates it as being fairly
being fairly shallow, it is more likely to be trash or
a hot rock. HOT ROCK REJECT in the Pro Options
has 20 levels available for the acceptance or rejec-
tion of +95. An operator can adjust HOT ROCK
REJECTION to suit personal preference and area
The fi rst 30 to 40 negative numbers below zero also
impact detection depth. Most iron indicates further
into the negative numbers. Thus the fi rst 30 to 40
negative numbers can often be ACCEPTed without
digging much iron. ACCEPTing from the posi-
tive numbers down as low as minus 30 to 40 has
the largest impact on the lower end of the positive
numbers. Sensitivity to small jewelry is increased.
However, sensitivity to foil is also increased. On a
beach or in a park that has lots of small foil, AC-
CEPTing down to minus 30 to 40 may not be practi-
cal. While ACCEPTing these minus numbers, a
reduced sensitivity setting may help eliminate some
of these small bits of unwanted foil.
ACCEPTing all V.D.I. numbers will produce the
best overall detection depth however, with the
amount of trash in most areas, searching with no
rejection is seldom practical. TONE I.D. is a more
practical choice. The idea of discrimination is to
ACCEPT V.D.I. numbers most likely to be valu-
ables and to REJECT the V.D.I. numbers most
likely to be trash. Digging ease, and acceptability in
an area will also decide the amount of ACCEPTed
V.D.I. numbers desired. No metal detector can
reject all trash while accepting all valuables. The
deeper you want to detect, the more trash you will
The detector sees metals electronically; thus some
valuable and non-valuable targets may look identi-
cal to the detector. To select ACCEPT and REJECT
V.D.I. numbers is to gamble that the areas you
will be good targets, and the
A factory program which uses the law of averages
usually produces more good targets than trash. The
EDIT feature allows you to fi ne-tune the ACCEPT
desired targets, further improving these odds.
Chapter 5 DFX™ Pro Options

Chapter 5 DFX™ Pro Options
or REJECT with ARROW controls.
Allows major EDIT changes to be completed
quickly. EDIT is for individual V.D.I. number
changes and/or viewing of the current settings.
BLOCK EDIT is for changing multiple V.D.I.
and press ENTER. The ENTER control can be
used to select either ACCEPT or REJECT. The
ARROW controls can then be used to drag that
ACCEPT or REJECT status as far as desired
through the V.D.I. number range. When a V.D.I.
number or range of numbers appear that you want
press ENTER to change the ACCEPT/REJECT
desire. ENTER changes ACCEPT/REJECT; AR-
ROWS drag that ACCEPT/REJECT status; EN-
TER again changes ACCEPT/REJECT; ARROWS
drag that new ACCEPT/REJECT status, etc.
BLOCK EDIT not only saves time if large sections
of the V.D.I. target numbers are to be changed, it
also allows peace of mind in the certainty of how
all the V.D.I. numbers are set. In the same amount
of time it takes to view all the V.D.I. number's cur-
rent status in regular EDIT, you can set them the
way you desire in BLOCK EDIT.
the current settings or the
the arrows are used while
change every V.D.I. number
one appears on the display.

Chapter 5 DFX™ Pro Options
teach the discriminator what metal targets you
desire to ACCEPT or REJECT.
the ARROW controls and press ENTER to turn
LEARN ON. The current selection is indicated by
the √ mark in square on the left of the display.
Squeeze and release the TRIGGER.
the instrument by sweeping them one at a
time in front of the loop until the desired
accept or reject sound is heard.
Press either ARROW control to return to
the LEARN selection display.
Press ENTER control to select LEARN
Squeeze and release the TRIGGER to
return to a searching mode.
target and an exact sample is available,
that specifi c target. If using a stan-
dard program and a specifi c unwanted
trash target is being detected, use the
If you do not return to the LEARN
the detector will continue to learn (AC-
CEPT or REJECT) every metal target it
Selecting LEARN OFF after
you have learned a target is extremely
Iron alloys, such as steel bottle caps
or nails, should never be used for this
LEARN method of programming. Be-
cause of their inconsistent V.D.I. indica-
tions, iron alloys tend to scatter or other-
wise confuse the discriminate program.
If this occurs, it is best to turn the detec-
tor OFF, back ON, and re-ENTER the
desired Program to return to the original
factory preset discriminate settings.
As with EDIT, LEARN changes will only be tem-
porary until saved in a Custom Program.
If LEARN has been used to develop a discriminate
program, the ACCEPT or REJECT V.D.I. numbers
can be reviewed by using the EDIT feature.

Chapter 5 DFX™ Pro Options
close together can each respond.
When a metal is detected, it takes a fraction of a
it can respond to another metal target nearby. The
time it takes to process the fi rst metal target signal
is called RECOVERY SPEED.
There are advantages and disadvantages to fast
well in high trash areas. However, they will have
as double responses on shallow targets. Slower
high trash areas. However, they will have bet-
ter responses on very deep targets. Slower speeds
also have more defi nitive discrimination sounds.
A custom setting needs to be found that suits the
preferences of the individual and the conditions in
the area. As a general rule, the closer together the
metal targets are in an area, the faster the recovery
gets, the slower the speed should be. Don't use the
fast speed if you don't need to.
Ground Filtering adjustments (page 44) will also
dramatically improve or degrade performance in
areas, slower speeds (low-
er numbers) for low trash
areas and/or improved dis-
crimination. Settings from 1

Chapter 5 DFX™ Pro Options
Adjusts how strongly the instrument rejects or
Most starting programs use the minimum setting.
As larger numbers are selected, more
The advantage of higher BOTTLECAP REJECT
iron rejection occurs. Trash becomes easier to iden-
tify by the broken sounds they produce.
The disadvantage of a high Bottlecap Rejection
detector to cancel both responses. Another disad-
vantage is that all targets, iron and non-iron, tend to
TLECAP REJECT. The operator needs to fi ne tune
BOTTLECAP REJECT according to their prefer-
ences and the conditions being searched.
sounds of iron, try increas-

Chapter 5 DFX™ Pro Options
as either negative/magnetic
ductive (hot rock). Typically
“cold rock” will respond in
ever, once the trigger is
squeezed for pinpointing it
ceases to respond. A posi-
tive or conductive “hot rock”
their target like responses
target responses in the +95
Hot rocks (rocks higher or different in mineraliza-
tion than the surrounding ground) respond as +95
on the V.D.I. scale. Unfortunately in some ground
conditions very deep, small, or unusual real metal
targets may also indicate +95. In the past an opera-
tor would either accept +95 with the DISC EDIT
feature and hear hot rocks as a target response
“beep”, or reject +95 and not hear a target like re-
The HOT ROCK REJECT features allows a com-
promise between total rejection or total acceptance
of the + 95 V.D.I. indication. Disc Edit Accept or
Reject of +95 will only control the visual displaying
of +95 with the Visual Discrimination feature ON.
To reject or accept +95 with the audio discrimina-
tion, Audio DISC must be on, and the HOT ROCK
REJECT feature adjusted for your preference-re-
A HOT ROCK REJECT range of 0 to 20 is pro-
vided. At 0 the same characteristics can be expected
Whiteʼs models. At 20 the same characteristics can
be expected as EDIT REJECT +95 provided with
At 10 a compromise between accepting +95 and
rejecting +95 is achieved. A hot rock may or may
not produce a target like response depending upon
many factors such as the actual size of the hot rock,
degree of difference between the hot rockʼs miner-
ground balance setting, and the Sensitivity/Gain
Hot Rock Reject provides for 19 shades of gray be-
tween totally accepting all hot rocks (0), and totally
rejecting all hot rocks (20).

when an operator would de-
sire the best performance
suited to faster loop sweep
speeds are large areas that
Typically high trash areas,
or older areas that deserve
cal search, are ideal for
slower loop sweep speeds.
The quickness the loop or search coil is passed from
to perform properly regarding detection depth and
discrimination. Sweep Speed is often a personal
preference feature. Individuals may like or dis-
like a particular model based on itʼs required ideal
of searching and a slower sweep speed for another
The SWEEP SPEED of the DFX is adjustable in all
A GROUND FILTER setting of 2 has a fi xed (not
An adjustment range from 1 to 20 is available. 1 is
the slowest ideal (best performing) loop or search
coil sweep speed, 20 is the quickest ideal (best per-
mance) loop sweep speed by controlling the width
of the target signal the detectors electronic circuit
dow (lower number settings) a slower loop sweep
dow (higher number settings) a quicker loop sweep
RECOVERY SPEED is highly interactive with
Generally faster RECOVERY SPEEDS are better
RECOVERY SPEEDS are better suited to slower
new dimensions to these features. SWEEP SPEED,
RECOVERY SPEED, and GROUND FILTERING
nation responds to both trash and valued metal
Chapter 5 DFX™ Pro Options

of electronic fi ltering used
to distinguish between the
ground, trash, and valued
fewer the fi lters, the quicker
eral metals targets close to
tection depth in mineralized
ground quickly diminishes
with fewer fi lters. Audio
discrimination sounds are
the electronic fi lters the
fi cult to sort out or respond
separately. Settings from 2
GROUND FILTERING is adjustable with the
DFX. The fi rst two primary fi lters are contained
within the hardware (electrical parts). The remain-
ing fi lters are contained within the computer soft-
ware allowing full user adjustment.
The DFX GROUND FILTERING has an adjust-
ment range from 2 through 6. At a setting of 2
mance in high trash areas of medium to low ground
mineralization. Combined with the RECOVERY
will pull some good targets from previously heavy
performance than 2 in slightly higher ground miner-
end Whiteʼs fi ltering ideal for medium to high min-
fi ltering for extreme ground conditions. Although
in the past, two of those six were dedicated to visual
indications alone. The DFX 6 fi lter selection dedi-
cates all six to ground rejection, discrimination,
mineralized ground conditions the DFX with its
multifrequency salt/conductive mineral cancelation
performance levels never before available.
Chapter 5 DFX™ Pro Options

ed V.D.I. numbers and their
on the display. When OFF, all V.D.I. numbers and
VISUAL DISC. "ON" eliminates both the V.D.I.
the ICON (NAIL, FOIL, PULL TAB,
V.D.I. numbers from appearing
on the display. In other words if Visual Discrimi-
nation is ON, and a specifi c target is rejected (for
example 0-IRON) then 0-IRON will not appear on
the display. If VISUAL DISC. is OFF, all the dis-
play indications will appear regardless of whether
the specifi c V.D.I. number is ACCEPTed or RE-
Visual Discrimination is based on the current dis-
Visual Discrimination has the advantage of elimi-
bers from appearing on the display. The disadvan-
tage is that much of the visual display information
will not appear for you to interpret questionable
target information. Elimi-
Graphic representation of metal targets, "NAIL,
FOIL, PULL TAB, SCREW CAP, COINS".
ICONS are designed for average use inside the
USA. If searching in other countries, or in areas
where such common types of targets are not ex-
pected, ICONS may be distracting. In such cases,
this option allows the ICONS to be turned off.
a quick way of reading the
display information. ON is
OFF will speed the remain-
Chapter 5 DFX™ Pro Options

Controls how strong a target must respond to pro-
duce a display indication, and controls a third digit
V.D.I. Sensitivity controls the intensity (strength)
of the target signal needed to activate the display
indications. A low V.D.I. Sensitivity setting would
require a strong target signal to trigger the display
indications. A higher V.D.I. Sensitivity setting
would require very little target signal to trigger the
To provide greater specifi c target resolution, levels
of 86 and above will provide a third V.D.I. digit
the fi rst two digits can be programmed ACCEPT or
As with all sensitivity levels, if set too high for
the area, the detector will become unstable and the
display indications unreliable.
Low ground mineralization is typically a good
indication that high V.D.I. sensitivity settings are
interference are conditions where a lower V.D.I.
This feature is designed to allow adjustment of the
display stability. However, because the instrument's
display and audio tone are closely associated in
many of the audio modes, the V.D.I. Sensitivity
may increase or decrease overall audio sensitivity.
display indications, but less
above produce a third V.D.I.
digit (.0), providing greater
Chapter 5 DFX™ Pro Options

Measurement of the ground mineralization, or
measuring the phase (V.D.I. reference number) of a
When D.C. PHASE is ON, and the TRIGGER
When a target is located, or when the operator
wants to measure a target or the ground, the TRIG-
GER is squeezed and held as if to pinpoint. When
the loop is held stationary over the target or ground,
the D.C. PHASE is shown on the upper-right side of
the display. The depth reading can be used to help
center the loop over a target for more accurate indi-
cations. The audio tone will also assist in centering
the loop over the target.
Metal targets in the ground will produce a measure-
ment which represents the target indication ± the
ground measurement. To measure the target alone,
move the loop off to one side (being sure to hold
the loop directly on top of the ground mineral),
the loop back over the target. This should allow a
proper D.C. PHASE reading for the target alone.
The relationship between D.C. PHASE, Ground
Balance, V.D.I. number and Discrimination is com-
plex. Results are not always repeatable in varied
ground conditions. Ground and target responses can
distort the detection fi eld of the loop in a number
of ways, by bending, refl ecting, absorbing and thus
relationships of these features.
When considering specifi c numbers it is important
to know D.C. phase is provided by the 15 kHz
channel without normalization. The V.D.I. scale
is distorted/expanded in the lower end of the scale
conditions, certain Basic
also be recognized in other
areas and dealt with by us-
The D.C. Phase of the ground has a relationship
to the proper Ground Balance setting. This is of
COARSE or FINE G.E.B. (Manual Ground Bal-
ance). Manual Ground Balance has a range that
covers V.D.I. numbers from -95 - +10. A Manual
255 equals a D.C. Phase of +10. As mentioned
ance is not the same as EDITing these V.D.I.
V.D.I. -90 will produce operational problems as
the detector will see the ground as a target. With
the same logic, if a target measures -90, and the
ground measures -90, and the detector is Ground
virtually invisible to the detector. These conditions
would be extremely rare however; the point is
there must be a difference between the target and
the ground for the target to be detected.
Chapter 5 DFX™ Pro Options

tiple loop passes, OFF shows SignaGraph
tion only from the last sweep of the loop.
Accumulate allows the SignaGraph
collect information. This collecting of information
continues from one sweep of the loop to the next.
Eventually, if only ACCUMULATE is selected,
will completely fi ll, requiring that
the TRIGGER be squeezed and released to clear it
is suggested to be used in combination with AC-
CUMULATE so that noncurrent information gradu-
required. FADE will gradually reset or clear the
to eliminate old information.
Emphasizes the most common or predominate Sig-
necessarily allow showing of
the information received, and shows this
bination with ACCUMULATE it has more infor-
mation to average, which results in more accurate
indications of trends. FADE is suggested to clear
the display of old information, rather than excessive
as it reduces indications
ticular target, which may
or target positioning in the
If ACCUMULATE is OFF, and AVERAGE
is OFF, then SINGLE SWEEP is active.
information received during the current sweep
of the loop, no more and no less. The next
received only within that sweep. Fade is not
needed in the SINGLE SWEEP mode as each
pass of the loop will clear the display and pro-
vide current information automatically. Sev-
during each sweep of the loop. It is therefore
normal to see one or more SignaGraph
even in the SINGLE SWEEP mode.
Chapter 5 DFX™ Pro Options

Clears or fades noncurrent SignaGraph
FADE allows past information to automatically
clear or fade from the SignaGraph
or with too slow of a FADE rate, the SignaGraph
will fi ll with information that is no longer valid as
you sweep through a search area.
A minimum setting of 1 equals the least degree of
fading. A maximum setting of 14 selects a very
quick fade rate. The SignaGraph
Chapter 5 DFX™ Pro Options

models) is used to promote stability and perfor-
mance. Like a sensitivity control, too much PRE-
AMP GAIN will result in unstable operation and
PREAMP GAIN will have a substantial effect on
For maximum detection depth, fi ne tune the PRE-
AMP GAIN to the highest number setting without
causing OVERLOAD, and/or unstable operation.
More adjustment levels (range) are available than
what may be usable. You may never encounter
used. Electrical interference may make it diffi cult
to receive an Air Balance at high PREAMP GAIN
DISC OFF may not allow as high a PREAMP
GAIN level as other selections. This is normal as
ground, and they may still provide superior per-
formance and depth even with the lower PREAMP
Every time the PREAMP GAIN level is changed,
the detector should be re-ground balanced. Once
in an operating mode, simply press ENTER to re-
als are present. Settings
from 1 - 4 are available.
Chapter 5 DFX™ Pro Options

30. 2 Frequency (Best Data)
30. 2 Frequency (Best Data)
Transmits and processes data continuously at both
Automatically chooses to use only the data from the
most reliable frequency (based on both magnitude
Gold, nickel, and other lower conductive metals re-
more reliably at lower frequencies. 2 Frequency
cel ground and salt and then selects the most ideal
frequency for that particular target alloy on a target
31. 2 Frequency (Correlate)
Transmits and processes data continuously at both
data at 3 kHz and 15 kHz. Target signals that do
not provide reasonably predictable information at
both 3 kHz and 15 kHz are automatically rejected.
Iron typically doesnʼt compare predictably between
frequencies and thus improved iron rejection can be
expected with the use of Correlate.
Differences in the target signal at 3 kHz and 15 kHz
not match within a reasonable range of predictabil-
ity, most likely the target is not a valued metal and
Only one of the four available multifrequency
methods, BEST DATA, CORRELATE, 1 kHz,
and 15 kHz, can be "ON" at any one given
time. For example if BEST DATA is "ON",
the remaining three multifrequency options
are automatically "OFF". To turn BEST DATA
remaining methods. In other words, an opera-
tor never turns "OFF" a multifrequency method,
they simply select the desired method and turn
it "ON", all other methods are automatically
The MULTI FREQUENCY selection in use
appears continuously on the bottom of the LCD
display during searching. The ideal multi fre-
quency or single frequency setting will depend
on many different factors such as the amount,
degree, and type, of ground mineralization, the
types of metal alloys most desired, as well as
personal preferences. No one setting will be ide-
al for all situations. A user should fi rst trust the
factory preset settings for the general types of
ideal settings for that particular area and type
of search. Generally multi frequency settings
are better suited to areas that have both mag-
netic (iron) and conductive (salt) components.
Inland areas may vary in this regard with the
present mostly iron type characteristics possibly
making one of the single frequency modes a
better choice. The same area when the soil is wet
may exhibit conditions similar to a combination
of both iron and salt due to nonferrous metallic
type mineralizationʼs greater electrical activity
when wet. For example soils containing silver,
copper, or nickel oxides/nitrates react differently
in conditions from wet to dry. These soil traits
can occur naturally and are typical of areas with
mineral springs or volcanic origins. However,
be particularly aware of this wet/dry phenomena
in farming areas where the soil mineralization is
likely altered for agricultural purposes.
Chapter 5 DFX™ Pro Options

V.D.I. Normalization - Used in the 1 frequency
operating modes to correct the skewing of the tradi-
tional (6.59 kHz) V.D.I. number display calibration.
2 Frequency modes require Normalization and thus
it is automatically ON in both the 2 Frequency (Best
Data) and 2 Frequency (CORRELATE) operating
modes. It is not an option in 2 Frequency modes.
When operating in the 1 Frequency (3 kHz) mode
Normalization ON recalculates the V.D.I. display
information to represent traditional V.D.I. display
calibration at Whiteʼs traditional frequency 6.59
When OFF in the 1 Frequency (3 kHz) mode the
V.D.I. scale is distorted/expanded in the higher end
of the scale (copper/silver) and compressed in the
lower end of the scale (foil/nickel). This distortion
can be used to provide higher display or discrimina-
tion resolution in this copper/silver coin area of the
advantage of this expanded resolution.
When operating in the 1 Frequency (15 kH
Normalization ON recalculates the V.D.I. display
information to represent traditional V.D.I. display
ration at Whiteʼs traditional frequency 6.59
OFF in the 1 Frequency (15 kHz) mode the
V.D.I. scale is distorted/expanded in the lower end
of the scale (foil/nickel) and compressed in the
higher end of the scale (copper/silver). This distor-
tion can be used to provide higher display or dis-
crimination resolution in this foil/nickel area of the
advantage of this expanded resolution. The advan-
ly. Provides for superior high iron mineral rejection
for better depth on higher conducting metals such as
copper and silver alloys.
mode when iron (without the presence of salt) is
extreme and the primary target is higher conducting
metals such as copper or silver. Some of Whiteʼs
early 1970 models operated at 3 kHz and are still
revered for there performance in high iron black
dium to low mineralization (without the presence of
range targets are the primary targets.
presses the low (foil/nickel) end of the scale. In the 15 kHz mode the opposite occurs.
The high end compresses and the low end expands.
Chapter 5 DFX™ Pro Options

Whiteʼs has pre-loaded each of the four custom
EEPROM Program positions for you. These pro-
grams can be erased simply by developing your
own program, starting with any of the nine available
programs, and using the SAVE or NAME feature
The fi rst fi ve Factory Preset Programs are burnt
into permanent memory EPROM (Electronic,
Program-mable, Read, Only, Memory). Although
you can make changes to these programs while in
use, and you can modify and copy/save them to the
EEPROM positions for future use through battery
changes, the Factory Preset Programs will always
revert to their original factory settings. Only the
programs saved in one of the four available EE-
PROM positions will save through battery changes.
release the trigger) will maintain changes to a fac-
tory program if a good battery remains in the unit.
To assure you retain special settings it is advised to
tions. Once you modify and Save/Name a factory
program to one of the EEPROM positions it is no
longer a Factory Preset Program and thus will save
EEPROM stands for a special type of electronic
memory chip (Electrical, Erasable, Programmable,
Read, Only, Memory). This memory chip allows the
user to store up to four of their own programs for
future use through battery changes and storage.
To use a custom EEPROM Program Simply;
1. Turn the detector ON and wait for the
2. Press ENTER to enter the Preset programs.
3. Use ARROW DOWN to scroll down the
Preset Program pages/screens until you see
thecursor beside the EEPROM program
5. Press ENTER again to LOAD the program.
6. Air/Ground Balance and your ready to search
using the program stored in that position.
To store the current settings you are using do the
NAME. LOAD it when you want to use those set-
tings again. Save or Name if you want to save the
The idea is that once you have determined your
preferences for the particular features and settings,
you can store that entire program in one of the EE-
PROM positions and have it ready for use whenever
you desire, without having to go through all the
options. Save your settings for future turn on and go
Chapter 5 DFX™ Pro Options

EEPROM DP-SILVER (Deep Silver)
A good general purpose program where the
majority of the trash is iron and small/
medium foil and the majority of the ex
pected targets are coins.
Popular performance modifi cations include:
Ground Filtering adjustment 2 for lower
minerals 3 medium minerals.
Remember...if consistently digging a
particular trash target, simply make note
of the typical V.D.I. reference numbers,
go to Disc Edit and change those specifi c
V.D.I. numbers for rejection. Re-save
any changes if future use is desired.
Provides a primarily coin program targeted more
toward trashy public areas. Rejects common
aluminum at the sacrifi cing of some jewelry.
Jewelry 18 K and above in the medium size
category is not likely to respond. However, areas
of common abundant aluminum trash can be
searched with good coin results as well as 14K
jewelry in the medium to small category. The
only way to get all the jewelry is to dig all the
aluminum, lead, and brass, which can be tedious
to the point of frustration in trashy areas.
Popular Performance modifi cations include:
Ground Filtering 2 for lower mineral areas,
4 for more mineralized areas.
Re-save any changes if future use
Re-save any changes if future use
Primarily a showroom or bench program for
testing or demonstrating, likely the fi rst
EEPROM slot an individual would choose to
save their own program in place of.
A high performance general purpose program not
for the meek at heart. Advanced features TONE
I.D., V.C.O., wide open DISC EDIT acceptance
of all nonferrous (valued alloy) targets, and high
Sensitivity/Gain, make this program really sing
for those who can handle the various audio
pitches and depth of information. If you like
silent search, this isnʼt the program for you
Sensitivities and Gains may need to be reduced
in areas of heavy electrical interference.
Popular Performance modifi cations include;
Ground Filtering 2 for lower
mineralization, 4 for higher
Chapter 5 DFX™ Pro Options

Preset Program Settings
TONE (AUDIO FREQ.) 0 - 255
V.D.I. SENSITIVITY 0 - 99
PREAMP GAIN PREAMP GAIN 1 - 4
2 FREQUENCY (BEST DATA) ON/OFF
2 FREQUENCY (CORRELATE) ON/OFF
1 FREQUENCY (15 kHz) ON/OFF

Custom EEPROM Program Settings
TONE (AUDIO FREQ.) 0 - 255
V.D.I. SENSITIVITY 0 - 99
PREAMP GAIN PREAMP GAIN 1 - 4
2 FREQUENCY (BEST DATA) ON/OFF
2 FREQUENCY (CORRELATE) ON/OFF
1 FREQUENCY (15 kHz) ON/OFF

Any mode or control setting allowing total acceptance of all metal types, iron, aluminum, tin,
nickel, gold, brass, lead, copper, silver etc.
Circuitry which produces different audio tones (pitch) for different target's conductivity.
One of the most extreme components of non-conductive, negative ground minerals. Mag-
netic. Also called Magnetite (Fe304) or magnetic iron oxide.
Any intentionally buried or secreted hoard of valuables.
One of the major mineral types which make up the positive ground minerals. Wet
ocean salt/sand will produce a positive response due to its similar conductivity to metal.
The measure of a metal target's ability to allow eddy currents on its surface.
The greatest measure of a metal detectors ability to transmit an electromagnetic fi eld into the
Method of manually or automatically desensitizing a metal detector so that it may locate the
Circuitry which ignores or otherwise indicates, a specifi c target based on its conductiv-
A loss or increase in threshold caused by temperature, time, or battery condition.
Small circulating currents of electricity.
Descriptive of any iron or iron bearing material.
The number of complete alternating current cycles produced by the transmit oscillator per
A state of operation in which specialized circuitry can ignore the masking effect ground
minerals have over metal targets.
A rock which contains a higher concentration of mineralization than the surrounding ground.
Refers to the total volume (average) of ground penetrated by a metal detector.
Series of listings and prompts on a visual display designed to aid the operator in feature selection.
Metallic substances: iron, foil nickel, aluminum, gold, brass, copper, silver, etc.
An electronic component that can be programed to perform certain electronic functions.
Any soil containing conductive or magnetic components.
A condition of operation selected by the operator for specifi c functions.
Any mode that requires loop movement to respond to metals.
Not of iron, any metal that is not iron.
Any mode of operation that doesn't require movement of the loop to respond to metal
The length of time between eddy current generation sustained on a metals surface and the resulting
Finding the exact center of a metal target.
An indication of a target non-acceptance by silence or a broken sound.
The measure or capacity of a metal detector to perceive changes in conductivity within the
An audio or display response alerting the operator that a target has been detected.
The ability of a metal detector to maintain smooth predictable performance.
Refers to any object that causes an audio or display indication.
Glossary

If for any reason you should sell your Spectrum DFX
prior to the date the warranty ex-
pires, the remaining warranty is transferable.
This transfer is authorized by calling 1-800-547-
6911, and getting an Authorization Number.
Simply fi ll out the following information, including the Authorization Number, seal it in a
White's Electronics, 1011 Pleasant Valley Road, Sweet Home,
The remaining warranty period will then be available to the new owner.
The Warranty Statement applies to both the original owner as well as the second owner.
Name: __________________________________________________________
Address (Which appears on the original warranty card):
________________________________________________________________
________________________________________________________________
Instrument Serial Number: __________________________________________
Date Code: ______________________________________________________
Original Purchase Date:_____________________________________________
Name: __________________________________________________________
Address: ________________________________________________________
________________________________________________________________
________________________________________________________________
Comments: ______________________________________________________
________________________________________________________________
________________________________________________________________
________________________________________________________________
________________________________________________________________
Distributor Authorization Code:
_______________________________________________
Owner Info

If within two years (24 months) from the original date of purchase, your White's detector fails due
to defects in either material or workmanship, White's will repair or replace at its option, all necessary parts
without charge for parts or labor.
Simply return the complete detector to the Dealer where you purchased it, or to your nearest
Authorized Service Center. The unit must be accompanied by a detailed explanation of the symptoms of
the failure. You must provide proof of date-of-purchase before the unit is serviced.
This is a transferable manufacturer warranty, which covers the instrument two years from the
original purchase date, regardless of the owner.
Items excluded from the warranty are non-rechargeable batteries, accessories that are not standard
equipment, shipping/handling costs outside the continental USA, Special Delivery costs (Air Freight, Next
Day, 2nd Day, Packaging Services, etc.) and all shipping/handling costs inside the continental USA 90
White's registers your purchase only if the Sales Registration Card is fi lled out and returned to the
factory address soon after original purchase for the purpose of recording this information, and keeping you
up-to-date regarding White's ongoing research & development.
The warranty does not cover damage caused by accident, misuse, neglect, alterations,
modifi cations, unauthorized service, or prolonged exposure to corrosive compounds, including salt.
Duration of any implied warranty (e.g., merchantability and fi tness for a particular purpose)
any incidental or consequential damages. Some states however, do not allow the limitation on the length
of implied warranties, or the exclusion of incidental or consequential damages. Therefore, the above
limitations may not apply to you.
In addition, the stated warranty gives you specifi c legal rights, and you may have other rights
which vary from state-to-state.
The foregoing is the only warranty provided by White's as the manufacturer of your metal detector.
Any "extended warranty" period beyond two years, which may be provided by a Dealer or other third party
on your detector, may be without White's authority, involvement and consent, and might not be honored by
Owner Info

White's Service Centers
White's Authorized Service Centers
nbaker@whiteselectronics.com
Whiteʼs reputation has been built on quality products backed by qual-
ity service. Our Factory Authorized Service Centers are factory trained and
equipped. They offer the same quality service as the factory. Service before
Contact your Dealer. There may be a quick,
Double check the obvious, such as batteries,
and try the detector in another area to be sure there
Be sure to send all necessary parts with your
detector, such as batteries and holders, as these
items can result in symptoms.
Always include a letter of explanation about
your concerns, even if you have talked to the Ser-
vice Center by telephone.
Take care in packaging instruments for ship-
ping. Always insure your package.
1
3
2

White's Electronics (UK) Warranty
WHITE'S ELECTRONICS (UK) Ltd.
The serial number which is unique to your unit is on a white label inside the battery compartment.
Please quote this number on any correspondence regarding your detector.
White's Electronics has always been concerned with the absolute quality of their mineral/metal detectors. Service after the sales is
of extreme importance to us and we always do our utmost to ensure that customers are satisfi ed with our units. If your unit should
require servicing or repair, simply return it to us at the factory in Inverness and we shall carry out the necessary work for you.
Any work carried out by unauthorized persons will automatically nullify the warranty.
If within two years (24 months) from the original date of purchase, your White's detector fails due to de-
fects in either material or workmanship, White's Electronics (UK) Ltd. will repair or replace at its option,
Items excluded from the warranty are non-rechargeable batteries and other accessories.
The warranty is not valid unless the Warranty Registration Card is returned to the factory address within
This warranty does not cover damage to the detector caused by accident, misuse, neglect, alterations,
modifi cations or unauthorized service.
Duration of any implied warranties (e.g., merchantability and fi tness for a particular purpose) shall not be
longer than the stated warranty.
Neither the manufacturer nor the retailer shall be liable for any incidental or consequential damages result-
ing from defects or failures of the instrument to perform.
This warranty does not affect your statutory legal rights.
Whiteʼs Electronics (UK) Ltd
35 Harbour Road ~ Inverness ~ Scotland ~ IV1 1UA
Telephone: (01463) 223456 Fax: (01463) 224048
sales@whelects.demon.co.uk.
Web site: www.whites.co.uk

White's (UK) Warranty Transfer
If for any reason you should sell your White's detector prior to the date the warranty ex-
pires, the remaining warranty may be transferable.
White's Electronics, (UK) Ltd.,
35 Harbour Road, Inverness, Scotland, IV1 1UA.
Whiteʼs will then advise you what, if
any Warranty is available.
The Warranty Statement must be completed with Serial number and information on previ-
Name: __________________________________________________________
Address (as on original warranty card):
________________________________________________________________
________________________________________________________________
Serial Number: (inside battery door)___________________________________
Name: __________________________________________________________
Address: ________________________________________________________
________________________________________________________________
________________________________________________________________
Tel: ____________________ Email:_______________________________
) Printed in U.S.A. (3/2008)