Copyright 2007, Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any
form, this manual or any part thereof without permission from Welch Allyn.
Welch Allyn assumes no responsibility for injury or for any illegal or improper use of the product that may
result from failure to use this product in accordance with the instructions, cautions, warnings, or
indications for use published in this manual.
Welch Allyn
®
, Acuity®, Micropaq®, FlexNet® and Flexible Monitoring® are registered trademarks of Welch
Allyn, Inc.
Nellcor
®
is a registered trademark of Nellcor Puritan Bennett. Masimo® and SET® are registered trademarks
of Masimo Corporation.
Software in this product is Copyright Welch Allyn or its vendors. All rights are reserved. The software is
protected by United States of America copyright laws and international treaty provisions applicable
worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated within
this instrument as intended in the operation of the product in which it is embedded. The software may not
be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable
form. This is not a sale of the software or any copy of the software; all right, title and ownership of the
software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, please call Welch Allyn Technical Support:
USA 800 535 6663
315 685 4560
Canada 800 561 8797China 86 21 6327 9631
European Call Center 353 46 9067790France 33 1 55 69 58 49
Germany 49 7477 9271 86Japan 81 3 3219 0071
Latin America 305 669 9003Netherlands 31 157 505 000
Singapore 65 6419 8100South Africa 27 11 777 7555
United Kingdom 44 1332 363812
Australia 61 2 9638 3000
800 074 793
Reorder Part Number 810-1464-XX
Manual Part Number 810-1177-06 Rev. A, 02/2007
Welch Allyn
8500 SW Creekside Place
Beaverton, Oregon 97008-7101
USA
The following table summarizes the new features that are implemented in this software
version 1.5. The version previously released was 1.4.
FeatureReference
Patient alarm tones are now discontinuous, and
life-threatening alarms beep at a faster pace than
other patient alarms.
If you press to silence a patient alarm tone
and a new patient vital sign alarm or equipment
alert occurs, a tone for the new alarm or alert will
break the silence period at the monitor.
Intended Use
The Micropaq™ monitor is intended to be used by clinicians for single or multiparameter
vital signs monitoring of ambulatory and nonambulatory pediatric and adult patients in
health care facilities. It is also intended for patient transport. The monitor is able to
withstand light rain exposure over short periods of time (uniform distribution of
approximately 1 mm of water/ minute for 10 minutes or less).
See “Respond to a Patient Alarm at
Monitor” on page 36.
The Micropaq monitor is intended to operate with an Acuity
wireless communication over Welch Allyn’s FlexNet
devices to the Acuity Central Station through hardwired Ethernet networks and Wireless
Local Area Networks (WLANs). If the Micropaq monitor is moved out of range or loses
communication with the FlexNet network, it continues to monitor the patient, display
patient data, and generate local patient alarms or alert messages.
•The ECG channel is intended primarily for five-lead ECG monitoring, although
three-lead ECG monitoring is supported.
•The Pulse Oximetry channel is intended for continuous noninvasive monitoring of
functional oxygen saturation of arterial hemoglobin (SpO
(measured by an SpO
sensor).
2
™
network. FlexNet connects multiple
®
Central Station through
) and pulse rate
2
2Chapter 1 General InformationWelch Allyn Micropaq Monitor
The most likely locations for patients monitored by this device are step-down units,
telemetry departments, general medical/surgical floors, emergency departments, and inhospital transport.
This guide was written for clinicians. Although this guide may describe some monitoring
techniques, Welch Allyn expects that the operator is a trained clinician who knows how to
take and interpret a patient’s vital signs.
Federal USA law restricts sale of the device identified in this manual to, or on the order of,
a licensed medical practitioner.
Symbols
Warning Warning statements in this manual identify conditions or practices that could result in personal
injury.
Caution Caution statements in this manual identify conditions or practices that could result in damage to the
equipment or other property.
The following symbols appear on the monitor or accessories.
Li++
Direct currentEnclosure Protection Drip proof: Classification
IPX1
IPX1 per EN60529: 1991
Alternating current (battery charger)Fuse
The CE Mark and Notified Body Registration
Number signify the device has met all essential
requirements of European Medical Device
Directive 93/42/EEC
Signifies the device has met all essential
requirements of European Medical Device
Directive 93/42/EEC for a Class 1 product
C
kj
This device has been tested and certified by
the Canadian Standards Association
International to comply with applicable U.S.
US
and Canadian medical safety standards.
Patient connections are Type CF, isolated for
direct cardiac application, and protected
against defibrillation
(battery charger)
Protective earth ground (battery charger)Protected during defibrillation
Lithium Ion batterySeparate batteries from other disposables for
recycling
Li++
Caution: Refer to Directions For Use and
For indoor use only (battery charger)
accompanying documentation
This way upKeep away from rain
Humidity limitStacking limit (by number)
FragileAltitude limit
Temperature limitsIATA/ICAO Hazard Class 9 Package
(International Air Transport Association/
9
International Civil Aviation Organization)
Non-ionizing electromagnetic radiationSingle use only
Directions for UseChapter 1 General Information3
Monitor Front Panel Keys
Select Key and Silence Patient Alarm/
Equipment Alert Key- Selects the choice
highlighted on the menu. During patient alarms,
silences the tone at the monitor and at Acuity (if
connected) for 90 seconds. During equipment
alerts, silences or acknowledges (dismisses) the
alert.
Nurse Call Key - When connected to Acuity,
pressing this key sends Acuity a Nurse Call
message.
Recycle the monitor and battery
separately from other disposables.
www.welchallyn.com/weee
Battery Charger Label and LEDs
Green LED on continuouslyBattery is fully charged.
Scroll Up Key and Reset Alarm Tone Key-
Scrolls up menus on the display. During patient
alarms, resets the tone at the monitor and at
Acuity (if connected).
Scroll Down Key and Main Menu Key- Pressing
this key scrolls down menus on the display, or
causes the Main Menu to appear if no menu is
displayed.
Green LED flashing
Green LED flashing very slowlyBattery detected and waiting to be charged.
Yellow LED on continuouslySomething is wrong with the battery or the
General Warnings and Cautions
Familiarize yourself with all warnings and cautions before using the monitor.
WARNING When considering a treatment protocol that involves wireless
communication of patient data, be sure to recognize some limitations inherent
in wireless communications. When the monitor is not connected to the
network:
•There are no patient alarms or alerts at the Acuity Central Station.
•Acuity does not perform arrhythmia and ST analysis on the patient data
and does not generate related alarms.
•Patient data is not saved.
•Pressing (Nurse Call) on the monitor does not cause any response
at Acuity.
kj
Battery is charging.
charger. (See
Response”
“Battery Status and Possible
on page 44.)
WARNING Do not try to monitor neonatal patients with the monitor. The
monitor is intended for adult or pediatric patients. It is not intended for use with
pediatric patients (or infants) weighing less than 22 lbs (10 kg).
4Chapter 1 General InformationWelch Allyn Micropaq Monitor
WARNING Always check the patient mode at Acuity when monitoring a new
patient. The patient mode determines default alarm limits and internal algorithm
settings.
WARNING The monitor may not meet its performance specifications if stored
or used outside the specified temperature and humidity ranges.
WARNING Do not connect more than one patient to a monitor. Do not connect
more than one monitor to a patient.
WARNING During defibrillation, keep the discharge paddles away from ECG
and other electrodes, as well as other conductive parts in contact with the
patient.
WARNING Do not operate this product in the presence of flammable
anesthetics or other flammable substances in combination with air, oxygenenriched environments, or nitrous oxide; explosion can result.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI)
suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in
patient injury and monitor damage.
WARNING Electronic equipment that emits strong electromagnetic or radio
frequency signals can cause electrical interference with monitor operation. This
interference may distort the ECG signal, thereby preventing accurate rhythm
analysis. Avoid operating this device near equipment of this type.
WARNING Exposure to Radio Frequency (RF) radiation. To comply with Federal
Communications Commission (FCC) RF exposure requirements, this device
shall be used in accordance with the operating conditions and instructions
provided in this manual, including the section “Install the Carrying Pouch” on
page 28.
WARNING Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that
“in some devices, rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate
meter alarms. All pacemaker patients should be kept under close or constant
observation.” See “ECG Specifications” on page 50 for disclosure of the
pacemaker pulse rejection capability of this instrument.
WARNING For patients with a pacemaker, position the monitor to maintain a
minimum 6-inch distance between the monitor and pacemaker. Immediately
turn the monitor off and provide appropriate patient care if you have any reason
to suspect that the monitor is interfering with the pacemaker. The Health
Industry Manufacturers Association recommends this minimum 6-inch distance
between a hand-held wireless radio and a pacemaker, which is consistent with
the independent research by, and recommendations of, Wireless Technology
Research.
Directions for UseChapter 1 General Information5
WARNING Make frequent electrical and visual checks on cables, sensors, and
electrode wires. All cables, sensors, and electrode wires must be inspected and
properly maintained and in proper working order to allow the equipment to
function properly and protect patient safety.
WARNING Avoid electrosurgery burns at monitoring sites by ensuring proper
connection of the electrosurgery return circuit so that the return paths cannot be
made through monitoring electrodes and probes.
WARNING Use of ECG cables not specified by Welch Allyn may negate
defibrillator protection and risk patient injury due to shock.
WARNING To ensure patient safety, the conductive parts of the ECG
electrodes (including associated connectors) and other patient-applied parts
should not contact other conductive parts, including earth ground, at any time.
WARNING Motion artifact can affect the accuracy of patient vital sign
measurements. Minimize patient motion whenever possible.
WARNING For best product performance and measurement accuracy, use only
accessories supplied by Welch Allyn or recommended in the Welch Allyn
Products and Accessories booklet (810-0409-XX). Use accessories according to
your facility’s standards and the manufacturer’s recommendations. Always refer
to the manufacturer’s Directions for Use.
WARNING As with all medical equipment, carefully route the patient cabling to
reduce the possibility of patient entanglement or strangulation. Use the supplied
garment clips to secure the cable properly.
WARNING When positioning the monitor pouch on the patient, make sure the
straps do not entangle the patient’s neck or cause choking. Make sure the straps
do not restrict the movement of the patient’s limbs or create a hazard when
walking or moving.
WARNING If a product has been dropped or severely abused, send it to a
qualified service person to confirm proper operation.
Caution Do not autoclave the monitor. Autoclave accessories only if the
manufacturer’s instructions clearly approve it. Many accessories can be severely
damaged by autoclaving.
It is possible for the monitor to detect a problem that prevents the monitor from operating
properly. If this occurs, the monitor displays an error message and error number. Report
such errors to Welch Allyn. The monitor should be serviced only by a Welch Allyn service
technician while under warranty. Contact Welch Allyn for information about post-warranty
period service.
6Chapter 1 General InformationWelch Allyn Micropaq Monitor
Introducing the Monitor
The monitor is a patient-worn vital signs monitor for use by adult or pediatric ambulatory
patients.
•One or two ECG channels displayed
•Up to 2 ECG leads displayed at the monitor:
I, II, III, V, aV
•Up to 7 ECG leads displayed at Acuity:
I, II, III, V, aV
•One ECG lead displayed at the monitor and
at Acuity: Fixed lead II with 3-lead cable, or
5-lead cable with only RA, LA and LL
electrodes attached.
•Pulse oximetry (SpO
404 only)
•Two-way wireless communication within
Welch Allyn’s FlexNet network
•LCD for display of ECG waveforms, SpO
and heart rate/pulse rate data, and
messages from Acuity
•Standalone operation with patient alarms
when out of range of the network
•Patient alarm limits that can be set at the
monitor or at Acuity
•Configurable formats for single- or dual-waveform ECG display
•Internal antenna
•Nurse call key
•Lightweight (slightly more than 1 pound with battery)
•Rugged and tolerant of brief water exposure
•Rechargeable battery
•Sleep mode to extend battery life
, aVL, or aVF with 5-lead cable
R
, aVL, or aVF with 5-lead cable
R
) monitoring (Model
2
2
Model 402
Model 404
ECG monitoring
ECG monitoring and either one of two pulse oximetry (SpO2) monitoring options:
•SpO
•SpO
with Masimo® SET® technology
2
with NELLCOR® OxiMax™ technology
2
Directions for UseChapter 1 General Information7
Understanding the Monitor and the FlexNet Network
The monitor is intended to operate with an Acuity® Central Station as part of Welch Allyn’s
FlexNet network. FlexNet allows multiple devices to communicate through hardwired
Ethernet networks and Wireless Local Area Networks (WLANs). The Acuity Central
Station provides the primary display and entry of patient data for a patient connected to
the monitor.
To Other Acuity
Central Stations
Acuity Central Station
Access Points
H
R
S
p
O
2
80
97
H
R
S
p
O
2
80
97
H
R
S
p
O
2
80
97
Micropaq monitors
H
R
80
S
p
O
2
97
To Other
Access
Points
FlexNet Network
Each patient-worn monitor supports two-way communication with an Acuity Central
Station through an access point in the FlexNet network. The access point is a digital radio
transceiver that connects to the FlexNet network. During monitoring, the monitor sends
the patient data to Acuity. Acuity and the monitor continuously analyze the data. Acuity
provides appropriate alarm or alert messages at the Central Station and other network
devices such as a hallway message panel and the monitor itself. Acuity also stores the
patient data for viewing or report printing.
If the monitor is moved out of range or loses communication with the FlexNet network
and Acuity, it continues to monitor the patient and display patient data. While not
communicating with Acuity, the monitor continues to generate local patient alarms or
alert messages. Patient data is not stored and Acuity does not perform waveform analysis
or generate arrhythmia messages while the monitor is not communicating with Acuity.
When the monitor is returned to within range of the FlexNet network, it automatically
reconnects to Acuity.
8Chapter 1 General InformationWelch Allyn Micropaq Monitor
Monitor Features
Controls and Connectors
SpO2 Connector
ECG Connector
Visual Alarm Indicator
Liquid Crystal Display
Select Key and
Silence Patient Alarm/
Equipment Alert Key
Scroll Up Key and Reset Patient Alarm Tone Key
Scroll Down Key and Main Menu Key
Nurse Call
Key
Battery pack latch
Battery
Back
Directions for UseChapter 1 General Information9
Visual Alarm Indicator
GreenFlashes slowly during normal operation.
RedFlashes during patient alarm, remains on continuously when alarms
are silenced or suspended.
YellowFlashes during an equipment alert or while not connected to the
network.
Remains on continuously if the operator suspends an alert at Acuity
for 90 seconds or acknowledges (dismisses) a low battery alert from
the monitor or Acuity.
Note
The flashing green LED indicates that the monitor is connected to the network
but not necessarily connected to a patient. If the monitor is actively monitoring a
patient, the green LED indicates no alarms or alerts are detected.
Audible Alarm Indicator
Beeps to indicate a patient alarm, and beeps faster for life-threatening arrhythmia
alarms (see “Patient Alarm and Equipment Alert Specifications” on page 55).
Beeps to indicate when the equipment needs attention. This beep tone is slower than
patient alarm tones (see “Patient Alarm and Equipment Alert Specifications” on
page 55).
Volume can be configured as high, low, or off (configured at Acuity).
Volume can be configured differently for network connection or stand-alone operation
(configured at Acuity).
Nurse Call Key
When connected to Acuity, pressing this key sends a Nurse Call message to Acuity.
Depending on how Acuity is configured, this will cause Acuity to:
Generate a Nurse Call Alert at the Acuity Central Station.
Print a “snapshot” of the patient waveform at the Acuity Central Station.
Scroll Up Key and Reset Patient Alarm Tone Key
Scrolls up menus on the display.
Resets a silenced patient alarm tone.
10Chapter 1 General InformationWelch Allyn Micropaq Monitor
Scroll Down Key and Main Menu Key
Scrolls down menus on the display.
Displays the Main Menu.
Select Key and Silence Patient Alarm/Equipment Alert Key
Selects the choice highlighted on the menu.
During patient alarms, silences the tone at the monitor and Acuity (if connected) for
90 seconds. During equipment alerts, silences or acknowledges (dismisses) the alert
at the monitor and Acuity.
Battery
Insert the battery to turn on power. Remove the battery to turn off power. (While the
battery is removed, the monitor does not perform patient monitoring.)
Recharge the battery while it is removed from the monitor. (See “Recharge a Battery”
on page 43.)
To order a new battery, see “Battery Status and Possible Response” on page 44.
Directions for UseChapter 1 General Information11
Display
Although the Acuity Central Station is the primary location for viewing patient data, the
monitor provides information to support patient care.
Dashed lines indicate the
monitor detects a pacemaker
Waveform scale is
selectable
ECG lead is selectable
HR indicates the heart
rate is from ECG.
PR (pulse rate) is
displayed if SpO
and ECG is not (pulse rate
is derived from SpO
If the monitor detects
a vital sign outside
the measurable range,
- - - (below the range)
+ + + (above the range).
Indicates the monitor
is connected to
This symbol indicates the monitor is not
communicating with Acuity:
Flashing indicates the monitor is associated
with an access point, but not communicating
with Acuity.
Continuous on indicates the monitor is not
communicating with an access point or Acuity.
is active
2
it displays:
Acuity.
).
2
signal (display of pacer
detection can be enabled or
disabled at Acuity)
Indicates one or
more patient alarms
are disabled (off).
Patient name as entered at
Acuity.
If the patient name has not
been entered, the monitor
displays the last four digits of
the monitor serial number, such
as:
ID:6472
Symbol is displayed at the
monitor whenever the Nurse
Call key is pressed.
ECG waveform is displayed
when active.
SpO2 numeric data is a
percentage value.
SpO2 pulse amplitude indicator
(not proportional to pulse
volume)
Low battery icon flashes to
indicate monitor will shut off in
30 minutes or less.
Display Sleep Mode
In order to extend battery life, the display becomes blank after two minutes if no keys are
pressed. The display becomes active again if an alarm or alert occurs, a key is pressed,
the initial Acuity connection occurs, a cable is inserted, or an electrode is attached.
The display will not become blank if a patient alarm is occurring, an Acuity message is
displayed, or the monitor is in Demo mode or Service mode.
12Chapter 1 General InformationWelch Allyn Micropaq Monitor
Main Menu
When you first press , the Main Menu appears:
Press to move through the menu.
Press to select or change the highlighted choice.
EXITExit the Main Menu (the menu disappears).
ACUITY...Access the Acuity Menu with network options. The Acuity Menu
is only accessible while connected to Acuity.
EXITExit all menus and return to the monitoring
screen.
END TELEDiscontinue monitoring a patient.
NEW ROOMReassign a patient to a new room in the same
unit.
TRANSFERTransfer a patient to a new room in a new unit.
NEW PATIENTAssign the monitor to a new patient.
PATIENT INFODisplay patient information such as ID, name,
unit and room.
Whenever the monitor is connected to Acuity and you select
ACUITY... from the Main Menu, the monitor displays the message
ACUITY CONTACTED to confirm that Acuity has been contacted.
The monitor continues to display this message until Acuity
responds, or you press to acknowledge the message and clear
the screen. If the monitor detects an alarm or alert, it clears the
screen to display the appropriate alarm or alert message. The
length of time required for Acuity to respond to your selection at
the monitor can vary widely depending on the amount of network
traffic and other conditions.
ECG LEAD...Access a menu to change the ECG 1 or ECG 2 lead selection (I, II,
III, aV
, aVL, aVF, or V). Available vectors depend on the
R
connected electrodes.
Directions for UseChapter 1 General Information13
ECG SCALE...Change the scale of the ECG waveform. If two waveforms are
displayed, both have the same scale.
1 WAVEFORMThere are four possible ECG waveform display selections:
1 WAVEFORMthe default selection
2 WAVEFORMS
5 SECONDS
FULL SCREEN
Pressing changes to the next selection. This change does not
take effect until after you exit the Main Menu. See “Display” on
page 11 for descriptions.
LIMITS...Enter the Alarm Limits Menu (“Customize Patient Alarm Limits at
the Monitor” on page 37) and change alarm limits.
SYSTEM
INFORMATION
Display information about the network connection and SpO
module.
SERVICE MENU Enter Service Mode for a demonstration mode (Demo, see
“Demonstration Mode” on page 16) or service functions for
technicians. Service Mode is not available if any cables are
plugged in.
Note
To restrict access to the Main Menu, a Menu Lock option can be configured for
the monitor at the Acuity Central Station. When the Menu Lock is enabled, the
operator must press and hold down and for two seconds to gain access
to the Main Menu. The Menu Lock is disabled if the monitor loses
communication with Acuity.
2
14Chapter 1 General InformationWelch Allyn Micropaq Monitor
Waveform Options
There are four ECG waveform options as shown:
1 Waveform
The single ECG 1 (lead II)
waveform is displayed.
5 Seconds
ECG 1 (lead II) cascades from
one line to the other.
The single ECG 1 (lead II) waveform is
allowed to occupy most of the screen.
2 Waveforms
ECG 1 (lead II) and ECG 2
(lead V) are both displayed.
Full Screen
To change the waveform selection during operation:
1.Press to display the Main Menu.
2. Press as needed to highlight the current waveform selection. Then press as
needed to select the desired display.
Directions for UseChapter 1 General Information15
Messages from Acuity
The monitor displays messages sent from Acuity as needed, including patient alarms and
equipment alerts. When Acuity messages are displayed, they temporarily override
information displayed on the lower half of the monitor screen.
sensors (Masimo or Nellcor)Micropaq Directions For Use
2
WARNING For best product performance and measurement accuracy, use only
accessories supplied by Welch Allyn or recommended in the Welch Allyn
Products and Accessories booklet (810-0409-XX). Use accessories according to
your facility’s standards and the manufacturer’s recommendations. Always refer
to the manufacturer’s Directions for Use.
Operating Settings
The following monitor operating settings can be set at the monitor or at the Acuity Central
Station:
•Patient alarm limit settings (ECG and SpO
•ECG lead and scale selection
•ECG display format
Many other monitor operating settings (such as patient mode and alarms volume) can
only be set at the Acuity Central Station. See “Operating Settings” on page 49 for a list of
all settings and where they are set.
).
2
16Chapter 1 General InformationWelch Allyn Micropaq Monitor
Default Settings
When the monitor connects to Acuity for a new patient, the Acuity Central Station
downloads the appropriate default settings stored at Acuity. While the monitor is
connected to Acuity, settings can be changed either at the monitor or at the Acuity Central
Station.
If the monitor is temporarily disconnected from Acuity and the operator changes settings
at the monitor, those settings are transmitted to and stored at Acuity when the monitor
reconnects.
Demonstration Mode
You can practice using the monitor with the Demo mode of operation, including
connection to Acuity.
The Demo mode cannot be activated while you are monitoring a patient or if any cables
have been plugged into the monitor. During the Demo mode, the monitor and Acuity
display the message SIMULATION.
To practice with the monitor in Demo mode:
1.Disconnect all patient cables connected to the monitor.
2. Remove the battery (if installed).
3. Insert the battery and watch for the Power-Up screen.
Power-Up Screen
4. After the Power-Up screen disappears, press to display the Main Menu.
Main Menu
Directions for UseChapter 1 General Information17
5. Press to highlight SERVICE MENU, then press to display the Service Menu.
Service Menu
6. Press to highlight DEMO MENU, then press to display the Demo Menu.
7.Press to highlight DEMO 1 or DEMO 2, then press to start.
SIMULATION indicates the
Demo mode is active.
A simulated waveform and
numerics are displayed.
Demonstration Mode
Demo Mode Display Values and Alarm Limits
DisplayDemo 1Demo 2
ECG WaveformNormal sinus rhythm,
normal ST
ECG Heart Rate
SpO2 Pulse Rate
Saturation%9788Lower 90
SpO
2
a. Demo 2 will cause patient alarms.
80125Lower 50
a
Normal sinus rhythm,
normal ST
Alarm Limits (On)
(not applicable)
Upper 120
Upper 100
8. While in Demo mode you can practice changing settings such as ECG lead selection
and alarm limit adjustment. (These changes only affect the Demo mode and are
erased when you exit the Demo mode.)
18Chapter 1 General InformationWelch Allyn Micropaq Monitor
9. To change to the other Demo selection, press to display the menu, then scroll
down to highlight TOG GLE DE MO MOD E and press .
10. To exit the Demo mode, either insert a patient cable or remove and insert the battery.
The monitor restarts and enters the normal monitoring mode.
19
2
Monitoring
Connect a New Patient
Connect to the Network
1.Insert a battery into the monitor to turn it on. After a few seconds the monitor PowerUp Screen is replaced by an initial monitoring screen.
The monitor is
searching for a
network
connection.
2. After the network connection is established, the monitor may prompt you to select an
Acuity Unit (if your facility has more than one Acuity unit):
If any patient cables are
connected, there will be
some patient data displayed.
The format depends on the
monitor default settings.
3. Press or to highlight the desired Acuity unit, then press .
When you press or to highlight the desired Acuity unit and then press ,
your selection will begin to flash between normal and reverse video to confirm that
the monitor is communicating your selection to Acuity. You cannot scroll to another
selection during this time. The selection continues to flash until Acuity responds back
to the monitor. Then the monitor displays the next appropriate screen (such as a list of
possible patients). The length of time required for Acuity to respond to your selection
at the monitor can vary widely depending on the amount of network traffic and other
conditions.
Be sure to select an Acuity unit. Even though the monitor is
connected to the network (as indicated by the green LED and
network connection symbol), the Acuity Central Station may not
display any indication of this monitor until after you have selected an
Network Connection
Symbol
Acuity unit.
4. The monitor displays a list of possible patients.
If your patient has been pre-admitted to the selected Acuity unit, they will be included
in the list.
Possible
patients to
select.
Select Patient at Central
428-02-2392, Hopkins, Bill J
520-29-0319, Phillips, Mary L
532-94-8372, Smith, Frank R
Example of Patient List
SELECT
PATIENT
▼
5. Scroll through the patient list to look for your patient’s name.
•If your patient is not in the list, highlight Select Patient at Central and press .
The patient name will need to be entered later at the Acuity Central Station.
WARNING If you do not select the patient name at the monitor at this time, do
not adjust any alarm limits until after the patient name and ID are confirmed at
Acuity. When the patient name and ID are confirmed at Acuity, Acuity
downloads the default settings and patient alarm limits for that Acuity unit to the
monitor, thereby overriding any previous settings and alarm limits.
Note
At power-up, the monitor retains the most recent patient mode. The patient
mode can only be changed at Acuity. If the patient is being monitored when the
patient mode is changed, there is a brief interruption in the display and recording
of ECG and SpO
patient data.
2
Directions for UseChapter 2 Monitoring21
•If your patient is in the list, highlight the name and press . Within a few
seconds the monitor displays a list of unassigned rooms.
Possible
rooms to
select.
•If you want to assign the patient to a room, highlight the room and press .
•If you do not want to assign a room at this time, highlight Select Room at
Central and press . The patient room will need to be entered later at the
monitor (see “Reassign a Monitored Patient to a New Room in the Same Unit”
on page 32) or at Acuity (see “Monitor Patient at Acuity” on page 41).
6. If you need to customize alarm limits for your patient, see “Customize Patient Alarm
Limits at the Monitor” on page 37.
Perform ECG Monitoring
WARNING Motion artifact can cause incorrect heart rate readings. Minimize
patient motion whenever possible.
WARNING If a disconnected lead is in too close proximity to other electrical
devices, it may cause false heart rate readings.
WARNING The monitor does not provide internal arrhythmia analysis.
Therefore, arrhythmias may cause the monitor to display inaccurate heart rates.
Select Room at Central
1104A
1104B
1105A
Example of Room List
SELECT
ROOM
▼
WARNING The monitor will show + + + for HR numerics between 301 and
350 beats per minute. Above 350 beats per minute, it may display incorrectly
low heart rates, due to intermittent picking of R-waves.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI)
suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in
patient injury and monitor damage.
WARNING Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that
“in some devices, rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate
meter alarms. All pacemaker patients should be kept under close or constant
observation.” See “ECG Specifications” on page 50 for disclosure of the
pacemaker pulse rejection capability of this instrument.
WARNING For patients with a pacemaker, position the monitor to maintain a
minimum 6-inch distance between the monitor and pacemaker. Immediately
turn the monitor off and provide appropriate patient care if you have any reason
to suspect that the monitor is interfering with the pacemaker. The Health
Industry Manufacturers Association recommends this minimum 6-inch distance
between a hand-held wireless radio and a pacemaker, which is consistent with
the independent research by, and recommendations of, Wireless Technology
Research.
WARNING High-intensity radio frequency (RF) energy from external sources,
such as an improperly connected electrosurgical unit, can induce heat into
electrodes and cables which can cause burns on the patient. Reading errors and
damage to equipment may also result. This hazard can be reduced by (1)
avoiding the use of small ECG electrodes, (2) selecting ECG electrode
attachment points remote from the surgical site and from the electrosurgical
return electrode, (3) using electrosurgical return electrodes with the largest
practical contact area, and (4) assuring proper application of the electrosurgical
return electrode to the patient.
WARNING Verify patient mode at Acuity. Incorrect patient mode may result in
inaccurate heart rates and inappropriate alarm settings.
WARNING To help prevent injury, use the provided garment clips to route the
ECG cables away from the patient’s head.
WARNING Use of ECG cables with loose or faulty detachable lead wires may
cause erratic behavior of the ECG waveform due to intermittent ECG lead wire
connections.
WARNING To ensure patient safety, the conductive parts of the ECG
electrodes (including associated connectors) and other patient-applied parts
should not contact other conductive parts, including earth ground, at any time.
Caution To protect the monitor from damage during defibrillation, for accurate
ECG information, and for protection against noise and other interference, use
only ECG electrodes and cables specified or supplied by Welch Allyn (these
cables have the required current-limiting resistors). Follow recommended
application procedures.
Caution Do not use an ECG cable longer than 10 feet (3 meters). If the nominal
length of the ECG cable, including extensions, exceeds this length, the monitor
is not guaranteed to meet published electromagnetic compatibility (EMC)
performance specifications.
•Even though the monitor contains fully isolated patient-connected circuitry, it has not
been specially designed for direct application on a patient’s heart.
•Use only with accessories provided or recommended in the Welch Allyn Products and Accessories booklet (810-0409-XX).
•Severe artifact and interference (such as defibrillation interference) can cause the
waveform to move off the display for a few seconds before it is restored.
Directions for UseChapter 2 Monitoring23
Perform 5-Lead ECG Monitoring
1.Inspect the ECG cable and replace it if it shows any signs of wear, breakage, or
fraying. Plug the cable into the monitor.
2. Select electrode sites on the patient.
Choose flat areas; avoid fatty or bony areas and major muscles.
3. Shave or clip hair from electrode sites, thoroughly clean skin, and lightly rub dry.
You may use soap and water, isopropyl alcohol or special skin preparation pads. To
avoid allergic reactions to electrodes, refer to the electrode manufacturer’s directions.
4. If you are using pre-gelled electrodes, make sure the electrode date is not expired and
the gel is intact and not dried out. For best results, use only silver/silver chloride
electrodes.
If you are using ungelled electrodes, apply a 1/4- to 1/2-inch mound of gel over the
electrode contact area.
For best product performance and measurement accuracy, do not use stainless steel
needle electrodes, squeeze bulb electrodes, or electrodes with dissimilar metals. Due
to polarization, such electrodes can generate offsets beyond the monitor’s
capabilities. Do not use electrodes from more than one manufacturer on the same
patient.
5. Attach lead wires to the electrodes before applying them to the patient.Apply the
electrodes to the patient in the proper locations.
Flashing circle indicates the
lead is not connected.
Left ArmRight Arm
Right Leg
Left Leg
If the monitor detects that some lead wires are not properly connected, the monitor
displays a chest diagram and indicates which leads are disconnected.
The locations of the circles displayed on the monitor for each lead are fixed, and are
not affected by the exact placement of the electrodes on the patient. For example,
the C lead can be placed on the patient in any one of the V1-V6 locations desired, but
will only be displayed on the monitor in the location shown above.
6. After leads are properly connected, confirm that the monitor displays the ECG
waveform, heart rate, and other patient data.
To change the ECG lead selection, press to display the Main Menu. Then press
Scroll Down to highlight ECG LEAD . . . , then highlight ECG 1 or ECG 2 and press
to change the lead.
3-Lead ECG Application with the 5-Lead ECG Cable
Note
You can perform 3-lead ECG monitoring in a similar manner as 5-lead ECG monitoring. You
may use the 5-lead ECG cable with detachable electrode lead wires, and connect only the
lead wires and electrodes for RA, LA, and LL. Refer to the Welch Allyn Product and Accessories booklet (810-0409-XX) for part numbers.
Follow these steps:
1.P e r f o r m Step 1 through Step 4 on page 23 as described for 5-lead ECG monitoring.
2. Before attaching electrodes to the patient, attach only lead wires for RA, LA, and LL to
3. Apply the electrodes for RA, LA, and LL to the patient in the proper locations.
Be aware that there are some inherent limitations with this application,
especially when compared to 5-lead ECG monitoring. These limitations include
the restriction to only one displayed lead, ECG lead II. Because only one
displayed lead is available (ECG lead II), factors such as a poor electrode
connection at RA, LA, or LL can significantly affect performance. To overcome
these limitations, the 5-lead ECG monitoring is preferred.
The monitor’s 3-lead ECG monitoring is only available for use with Acuity
software versions 6.1 or later.
the 5-lead ECG trunk cable and to the electrodes. Make sure that lead wires for C and
RL are DETACHED from the 5-lead ECG trunk cable.
The monitor displays the chest diagram with two circles blinking confirming that the C
and RL electrodes are not connected.
4. Observe the monitor and visually confirm that within about 30 seconds, the two
circles disappear and the monitor displays the ECG waveform, heart rate, and other
patient data.
Be aware that if you connect the C or RL lead wires to the 5-lead ECG trunk cable and
apply the C or RL electrodes to the patient, the monitor defaults to 5-lead ECG
monitoring and does not enable 3-lead ECG monitoring. To enable 3-lead ECG
monitoring, you must disconnect the ECG cable from the monitor for a few seconds,
and then begin this procedure again.
Directions for UseChapter 2 Monitoring25
Be aware that only ECG lead II is available for display with the monitor’s 3-lead ECG
monitoring. No other ECG lead selections are available.
WARNING Do not try to perform this 3-lead ECG monitoring with any 5-lead
ECG cable that does not have detachable electrode lead wires as described
above. Attempting to perform this procedure with a 5-lead ECG cable which has
lead wires cut off or hanging loose and not connected to the patient would
present a shock hazard to the patient or clinician.
3-Lead ECG Application with the 3-Lead ECG Cable
Note
Refer to the Welch Allyn Product and Accessories booklet (810-0409-XX) for part numbers.
Follow these steps:
1.P e r f o r m Step 1 through Step 4 on page 23 as described for 5-lead ECG monitoring.
2. Attach lead wires to the electrodes before applying them to a patient.
3. Apply the electrodes for RA, LA, and LL to the patient at the proper locations. If the
4. Observe the monitor and visually confirm it displays the ECG waveform, heart rate,
Be aware there are some inherent limitations with this application, especially
when compared to 5-lead ECG monitoring. These limitations include the
restriction to only one displayed lead, ECG II lead. Because only one displayed
lead is available (ECG lead II), factors such as poor electrode connection at RA,
LA, or LL can significantly affect performance. To overcome these limitations,
the 5-lead ECG monitoring is preferred.
The monitor’s 3-lead ECG monitoring is only available for use with Acuity
software versions 6.1 or later.
monitor detects one of the lead wires is not properly connected, it will display a chest
diagram indicating which lead is disconnected.
and other patient data.
Be aware that only ECG lead II is available for display with the monitor’s 3-lead
monitoring. No other ECG lead selections are available. The monitor will not detect
the presence of a 3-lead cable until two or more of its leads are connected to the
patient.
3-Lead ECG Application with the 3-Lead ECG Cable and Cable Extension
This combination functions the same way as the 3-lead ECG application with the 5-lead
cable. For electromagnetic compatibility (EMC) reasons, do not use an ECG cable and
extension cable length of more than approximately 10 feet total.
WARNING Oxygen saturation measurements using pulse oximetry are highly
dependent on proper placement of the sensor and patient conditions. Patient
conditions such as shivering and smoke inhalation may result in erroneous
oxygen saturation readings. If pulse oximetry measurements are suspect, verify
the reading using another clinically accepted measurement method, such as
arterial blood gas measurements on a co-oximeter.
WARNING Use only accessories as listed in the Welch Allyn Products and
Accessories booklet (810-0409-XX). Use only Masimo accessories and sensors
with the Masimo SpO
the Nellcor SpO
WARNING Tissue damage can be caused by incorrect application or use of a
sensor (e.g., wrapping the sensor too tightly, applying supplemental tape, failing
to periodically inspect the sensor site, leaving a sensor on too long in one place).
Refer to the Directions for Use provided with each sensor for specific
instructions on application and use, and for description, warnings, cautions, and
specifications.
WARNING Sensors exposed to ambient light while not applied to a patient can
exhibit semi-normal saturation readings. Be sure the sensor is securely placed
on the patient and check its application often to ensure accurate readings.
2
option. Use only Nellcor accessories and sensors with
2
option.
WARNING Inaccurate measurements may be caused by venous pulsations.
WARNING The pulse oximeter can be used during defibrillation, but the
readings may be inaccurate for a short time.
WARNING The pulse oximeter should NOT be used as an apnea monitor.
WARNING A very sudden and substantial change in pulse rate can result in
erroneous pulse rate readings. Be sure to validate the patient data and patient
condition before intervention or change in patient care.
WARNING Interfering Substances: Carboxyhemoglobin may erroneously
increase readings; the level of increase is approximately equal to the amount of
carboxyhemoglobin present. Methemoglobin may also cause erroneous
readings. Dyes, or any substances containing dyes, that change usual arterial
pigmentation may cause erroneous readings.
Directions for UseChapter 2 Monitoring27
1.Attach the SpO2 sensor to the patient according to the manufacturer’s directions for
use, observing all warnings and cautions.
Each SpO
sensor is designed for application to a specific site on the patient within a
2
certain size range. To obtain optimal performance, use an appropriate sensor and
apply it as described in the sensor’s directions for use.
If excessive ambient light is present, cover the sensor site with opaque material to
block the light. Failure to do so may result in inaccurate measurements. Light sources
that can affect performance include surgical lights (especially those with a xenon light
source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight.
If NIBP will be monitored while using SpO
than the SpO
sensor to help reduce unnecessary SpO2 alarms. For optimal
2
measurements, avoid placing the SpO
, place the NIBP cuff on a different limb
2
sensor on the same limb as an arterial
2
catheter or intravascular line.
Loss of pulse signal can occur if the sensor is too tight, there is excessive ambient
light, an NIBP cuff is inflated on the same limb as the sensor, there is arterial
occlusion proximal to the sensor, the patient is in cardiac arrest or shock, or the
patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
2. Inspect the SpO
cable. Replace it if it shows any signs of wear, breakage, or fraying.
2
Plug the cable into the sensor and the monitor.
3. After the cable is connected, confirm that the monitor displays SpO
data within a
2
few seconds.
4. If excessive patient movement interferes with measurements, consider the following
possible solutions:
•be sure the sensor is secure and properly applied
•use a new sensor with fresh adhesive backing
•select a different type of sensor
•move the sensor to a less active site
The SpO
system is designed to work satisfactorily during normal patient motion.
WARNING As with all medical equipment, carefully route the patient cabling to
reduce the possibility of patient entanglement or strangulation. Use the supplied
garment clips to secure the cable properly.
WARNING When positioning the monitor pouch on the patient, make sure the
straps do not entangle the patient’s neck or cause choking. Make sure the straps
do not restrict the movement of the patient’s limbs or create a hazard when
walking or moving.
Adult Carrying Pouch
The Adult Carrying Pouch is intended for ambulatory adult
patients. It is not intended for use while the patient is in bed.
1.Put the carrying pouch on the patient and insert the monitor.
2. Carefully arrange the pouch and monitor on the patient to
avoid bruising or other skin injuries.
To maximize the monitor’s wireless transmission range,
always make sure that the monitor display is facing out and
away from the patient’s body.
Pediatric Carrying Pouch
The Pediatric Carrying Pouch is intended for ambulatory pediatric
patients (40 to 80 lbs.). It is not intended for use while the patient
is in bed.
1.Insert the monitor into the pouch.
2. Carefully arrange the pouch and the monitor on the patient to
avoid bruising or other skin injuries.
To maximize the monitor’s wireless transmission range,
always make sure that the monitor display is facing out and
away from the patient’s body.
Directions for UseChapter 2 Monitoring29
Monitor a Patient Out of Range of Acuity
While out of range of Acuity, the monitor continues to monitor the patient and provide
local HR/PR and SpO2 alarms or alerts at the monitor as needed.
When the patient wearing the monitor goes out of range of Acuity, do the following:
1.A DROPOUT equipment alert occurs at the Acuity Central Station. Acknowledge the
alert at Acuity.
2. An equipment alert occurs at the monitor with this message:
ACUITY CONNECTION LOST
Depending on how the monitor is configured (as controlled by Acuity), this alert can
also cause the monitor to emit audible alert tones.
If tones are enabled, the authorized person should press on the monitor to
acknowledge (dismiss) the alert and silence this instance of the alert tone.
Note
When the patient returns within range of Acuity, the monitor automatically reconnects to
Acuity. No clinician intervention is required.
The person authorized to press to acknowledge the alert may vary,
depending on the local protocol. Follow the protocol established by your
institution.
If you want to discontinue monitoring the patient, follow these steps.
1.Press to display the Main Menu.
2. Press to highlight ACUITY, then press .
Acuity Menu
3. Press to highlight END TELE, then press .
4. When the monitor displays the message SAFE TO REMOVE BATTERY, remove the
battery.
If the battery is not removed within 30 seconds, the monitor will automatically try to
reconnect to the network.
5. Disconnect the leads and sensors from the patient.
Note
If you do not use END TELE to disconnect from the network as described
above, the Acuity Central Station generates a DROPOUT equipment alert at
Acuity.
If you want to monitor this same patient at a later time, you will need to reselect
the patient name from the monitor or confirm the patient ID at Acuity.
Directions for UseChapter 2 Monitoring31
Reconnect a Recently Monitored Patient
1.Insert a battery into the monitor to turn on the monitor. Confirm that after a few
seconds the monitor Power-Up Screen is replaced by the initial monitoring screen.
2. The monitor will then present a series of menus and messages requesting you to
provide information about the connection and patient. The actual screens presented
depend on how long the patient has been disconnected. Provide the information as
requested. This may include:
•Select an Acuity unit.
•Select a patient from the patient list.
•Select a patient room from the room list.
Note
If you do not select the patient name or room while connecting the patient, you
will need to do that later at the Acuity Central Station. See “Monitor Patient at
Acuity” on page 41 for more information.
•To perform ECG monitoring, see “Perform ECG Monitoring” on page 21.
•To perform SpO
monitoring, see “Perform SpO2 Monitoring” on page 26.
Reassign a Monitored Patient to a New Room in the Same
Unit
If a patient is being monitored and you want to assign them to a new room in the same
unit, follow these steps.
1.Press to display the Main Menu.
2. Press again to highlight ACUITY and press to display the Acuity Menu
screen.
3. Press to highlight NEW ROOM, then press .
.
New Room Selection
Within a few seconds the monitor displays a list of all available rooms, including the
patient’s current room.
•If you decide not to change the patient’s current room assignment, press (the
patient’s current room is the default selection in the list).
•To assign the patient to a new room, highlight the room and press .
•If you want to cancel the patient’s current room assignment, but do not want to
assign a new room at this time, you can highlight Select Room at Central and
press . You can then assign the room later from the Acuity Central Station, or
you can repeat this procedure and assign a new room from the monitor.
Directions for UseChapter 2 Monitoring33
Transfer a Monitored Patient to a New Room in a Different
Unit
If a patient is being monitored and you want to assign them to a new room in a different
unit, follow these steps.
1.Press to display the Main Menu.
2. Press again to highlight ACUITY and press to display the Acuity Menu
screen.
3. Press to highlight TRANSFER, then press .
Transfer a Patient
Within a few seconds the monitor displays a list of units.
4. Press to highlight the new unit, then press .
The patient is not monitored at Acuity during the short time required by Acuity to
process the transfer to the new unit (typically less than one minute). However, the
patient continues to be monitored by the monitor.
(If the selected unit is currently not available, the monitor displays an appropriate
message; press to acknowledge the message and cancel the transfer.)
5. After the patient is assigned to the new unit, the monitor displays a list of unassigned
rooms. (The patient’s previous unit and room assignment is cancelled.)
•To assign the patient to a new room, highlight the room and press .
•If you decide not to assign the patient to a new room at this time, you can
highlight Select Room at Central and press . You can then assign the room
later from the Acuity Central Station, or you can assign a new room from the
monitor later using the procedure on “Reassign a Monitored Patient to a New
If you want to discontinue monitoring a patient and reconnect the monitor to a new
patient, follow these steps.
1.Press to display the Main Menu.
2. Press again to highlight ACUITY and press to display the Acuity Menu
screen.
3. Press to highlight NEW PATIENT, then press .
Select a New Patient
The monitor then presents a series of menus and messages requesting you to
provide information about the connection and patient. The actual screens presented
depend on how the Acuity System is configured.
Provide the information as requested. This may include:
•Select an Acuity unit.
•Select a patient from the patient list. (After you select a new patient, all
monitor operating settings are reset to the Acuity System default power-up
settings.)
•Select a patient room from the room list.
If you do not select the patient name or room while connecting the patient, you will
need to do that later at the Acuity Central Station. See “Monitor Patient at Acuity” on
page 41 for more information.
•To perform ECG monitoring, see “Perform ECG Monitoring” on page 21.
•To perform SpO
monitoring, see “Perform SpO2 Monitoring” on page 26.
2
35
3
Alarms & Alerts
About Alarms and Alerts
Alarms provide a warning about a patient condition (such as a vital sign limit violation).
Alerts provide a warning about an equipment condition that needs attention (such as a
low battery or detached ECG lead).
Alarms and alerts may be detected either by the monitor or by the network. While
connected to the network, alarms or alerts are displayed at the monitor and at the Acuity
Central Station. Alarms have a higher priority than alerts.
Alarm Holdoffs
To help minimize false alarms, the monitor briefly delays or “holds off” triggering alarms
for limit violations for HR/PR or SpO
detects that the patient’s vital sign has returned to acceptable limits, the monitor cancels
the alarm holdoff. The next time a vital sign limit is violated, the monitor starts a new
holdoff period.
. After the alarm holdoff period begins, if the monitor
When a patient alarm occurs, the monitor produces an audible tone (if audible tones are
enabled). Life-threatening arrhythmia alarms beep at a faster pace than other vital sign
alarms (see “Patient Alarm and Equipment Alert Specifications” on page 55). The monitor
also displays a message similar to the following:
Red LED flashes
during alarm.
After the alarm is
silenced (or
suspended at
Acuity), the red
LED is on
continuously.
Flashing numerics
indicate patient
alarm.
Press to silence the alarm tone.
1.Check the patient and provide appropriate care.
2. To silence the alarm tone at the monitor and the Acuity Central Station for 90
seconds, press .
While the alarm tone is silenced, visual alarm indications continue, and the red
alarm indicator on the monitor changes from a flashing display to a continuous
display.
If the alarm condition still exists after 90 seconds, the alarm tone resumes.
Note
If you silence an alarm at the monitor and another patient alarm or an equipment
alert occurs during the silence period, the tone resumes at the monitor. At
Acuity, only life-threatening arrhythmia alarms interrupt the silence period.
If you suspend an alarm at Acuity, only life-threatening arrhythmia alarms interrupt
the silence period at the monitor and Acuity.
To access the Main Menu during silencing, press .
Directions for UseChapter 3 Alarms & Alerts37
3. To reset the alarm tone at the monitor and Acuity before the 90 seconds has elapsed,
press at the monitor, or press Resume at the Acuity Central Station.
4. After caring for the patient, make sure that the appropriate alarm limits are set and
that alarms are on.
Customize Patient Alarm Limits at the Monitor
WARNING If the patient’s name has not yet been assigned to the monitor, do
not adjust any alarm limits until after the patient name and ID are confirmed at
Acuity. When the patient name and ID are confirmed at Acuity, Acuity
downloads the default settings and patient alarm limits for that Acuity unit,
thereby overriding any custom alarm limits that were set at the monitor before
selecting the patient.
1.Press to display the Main Menu.
2. Press to highlight LIMITS, then press to display the Alarm Limits Adjust
Menu:
Alarm limits can be adjusted
for HR/PR and SpO
3. To change an alarm limit, press or to highlight the desired limit, then press
to display the Threshold Adjustment Menu:
Select + or - to change the limit.
•Scroll and select the + or - selections to change the limit as desired.
•To turn the highlighted limit on or off, scroll to ON/OFF and press .
WARNING If you turn off any alarm limits, be sure to restore the appropriate
alarm limits before you resume monitoring. Only life-threatening arrhythmias will be indicated at the monitor and Acuity (if connected) when alarms are turned off.
4. To change other limits, scroll to PREVIOUS MENU, press , then select another
limit to change.
5. When you have completed all changes, scroll to PREVIOUS MENU, then EXIT on the
Alarm Limits Adjust Menu and press to return to the normal monitoring screen.
Note
While the monitor is connected to Acuity, settings can be changed either at the
monitor or at Acuity.
Respond to an Equipment Alert at the Monitor
When the network or the monitor detects an equipment problem, the monitor produces a
an audible alert tone (if audible tones are enabled). Equipment alerts beep at a slower
pace than patient vital sign alarms (see “Patient Alarm and Equipment Alert
Specifications” on page 55).
The monitor also displays a flashing yellow light (LED) and an equipment alert message
similar to the following:
The monitor has detected that
the ECG cable has been
disconnected from the
monitor.
1.In this instance, press to acknowledge (dismiss) the alert tone and clear the
message.
If the message says “PRESS TO SILENCE,” when you press , the tone is
silenced for 90 seconds instead of dismissed.
If the monitor displays a chest diagram with a missing lead flashing, you can press
to silence the tone.
Some alerts do not give you the option to acknowledge the alert or silence the tone.
For these alerts, to remove the message and tone, you must correct the problem.
Directions for UseChapter 3 Alarms & Alerts39
2. If possible, determine what caused the problem and correct it.
Note
After you press to acknowledge or silence some types of alerts, the yellow
LED remains on (either flashing or steady yellow) until you correct the problem.
For low battery alerts and no Acuity connection alerts, specific icons also flash.
For a list of possible messages and suggested responses, see “Alert Messages and
LOW BATTERYThe monitor will shut down within approximately 30 minutes or less due to a low
battery.
• Replace the battery as soon as possible.
VERY LOW BATTERYThe monitor will shut down within approximately 5 minutes or less due to a low
battery.
• Replace the battery as soon as possible.
BATTERY TOO LOW
SHUT DOWN IN PROGRESS
The battery is so low the monitor has to shut down operation.
• Replace the battery immediately.
ACUITY CONNECTION LOSTThe monitor is not connected to the network.
• Press to acknowledge and silence the tone and cancel the message. While
disconnected from the network, the off-network icon and the yellow LED
continue to flash.
NOTE: The monitor will continue to attempt to reconnect until it is successful.
EXCESSIVE ECG OFFSET
REPLACE ELECTRODES
a
Chest icon is displayed with
flashing ECG electrode(s).
The monitor detects poor ECG electrode contact.
• Check and replace ECG electrodes as needed.
The monitor detects that one or more ECG electrodes are disconnected.
a
• Check and replace or reconnect electrodes as needed.
NO ECG CABLE DETECTED• If the ECG cable has been intentionally removed from the monitor, press to
cancel the alert tone.
• If the ECG cable has been unintentionally removed, plug it back into the monitor.
Check the patient and monitor to make sure ECG monitoring resumes properly.
• It is normal for this alert to appear with a 3-lead ECG cable when two or more of
its leads are disconnected from the patient. Reconnect the disconnected lead
wires.
NO SPO2 SENSOR DETECTEDThe SpO
• If disconnection is intentional, press to acknowledge and silence the tone.
sensor has been disconnected for more than 5 seconds.
2
• If disconnection is not intentional, reconnect the sensor or replace the sensor and
reconnect.
a
DEFECTIVE SPO2 SENSOR
or
UNRECOGNIZED SPO2 SENSOR
<key name> KEY STUCK
a
The SpO2 sensor is either defective or not recognized.
• Replace the SpO
a
sensor with a new, compatible SpO2 sensor.
2
During the power-up self test, the monitor detected that a key is stuck ( , , ,
or ). This can happen if you accidentally press a key down before the Main Menu
is displayed during power-up.
• Remove and then reinsert the battery to power up again and see if the key is still
stuck. If it is, contact your biomedical engineering department.
System Error
Thread: <nnn>
The equipment problem is so serious the monitor cannot be used.
• Contact your biomedical engineering department.
Error ID: <nnn>
a. This alert message can be acknowledged from Acuity, but not from the monitor.
41
4
Monitor Patient at Acuity
While the Micropaq is connected to the FlexNet network, patient data gathered by the
monitor is continuously stored at Acuity. You can access this patient information at the
Acuity Central Station and perform administrative functions, including:
•Admit (and discharge) a patient in the Acuity unit.
•Edit the patient description (name, physician, etc.).
•Review and print patient data such as trends and waveforms.
•Suspend patient alarm tones for 90 seconds and resume the alarm tones
For more information about using the Acuity Central Workstation, refer to:
•Acuity Directions For Use
42Chapter 4 Monitor Patient at AcuityWelch Allyn Micropaq Monitor
43
5
Maintenance
This section provides information to help operators of the monitor and the battery charger
perform routine maintenance activities such as changing or recharging batteries,
inspection, and cleaning.
Change the Battery
1.Remove the depleted battery.
2. Insert a fully-charged battery. Use only batteries supplied by Welch Allyn.
WARNING Before installing a battery, carefully inspect the battery case. If
there are any signs of damage, cracks, or leaks, discard the battery properly and
do not use it.
Note
The Acuity unit can be configured to allow you a short time (typically 30 seconds
or more) to change the monitor battery while the monitor is connected to the
network without causing an Acuity equipment alert. If the monitor is connected
to the network and the battery is removed for longer than the allowable battery
changing time, Acuity generates a DROPOUT equipment alert at the Acuity
Central Station.
Recharge a Battery
1.On the monitor battery charger, choose an empty battery well where the LED is off.
2. Insert the depleted battery into the battery well.
3. Confirm that the charger displays a flashing green LED by the battery to indicate the
battery is detected or is charging.
4. When the green LED is on continuously, the battery is fully charged. Remove the
battery.
If the yellow LED is on continuously, the battery may have reached the end of its
useful life. Refer to the table below for suggested responses.
Something is wrong with the battery or the charger. Remove the battery.
• If the LED goes off, it is probably a battery problem. Insert a new
battery into the same battery well. If the new battery charges correctly,
then the battery has a problem; discard the battery. The battery reorder
number is 008-0647-XX. If the same problem occurs with the new
battery, the charger may need repair. Contact biomedical engineering.
• If the LED does not go off when you remove the battery, it is probably a
charger problem. Unplug the charger power cord, wait at least 5
seconds, then plug in the charger power cord again. Insert a new
battery into the same battery well. If the new battery charges correctly,
then the battery has a problem; discard the battery. If the same problem
occurs with the new battery, the charger may need repair. Contact
biomedical engineering.
The charger can accommodate up to eight batteries. The charger charges a maximum of
four batteries at a time. After a battery begins recharging (as indicated by the green LED
that flashes on one second, off one second), it is typically fully recharged within four hours
at room temperature. After a battery is fully charged, the charger continues to maintain
the full charge on the battery until the battery is removed. Leaving a fully-charged battery
in the charger will not harm the battery.
Remove batteries from the battery charger if the battery charger will be disconnected
from ac power for more than a few days. Do not block the cooling vents at the rear of the
battery charger.
The monitor battery charger only charges four batteries at a time. A battery is not fully
charged until the green LED is on continuously. Do not remove a battery until it is fully
charged.
WARNING The monitor battery is Lithium Ion. Do not incinerate, submerge,
crush, disassemble, or autoclave. If a battery has been submerged in liquid,
discard the battery properly; do not try to recharge or reuse the battery. Do not
short the battery terminals. Do not try to connect the battery to any device
except the monitor or the monitor battery charger. Do not expose to high
temperature (above 60° C or 140° F). Use only the specified monitor battery
charger.
Directions for UseChapter 5 Maintenance45
Inspect and Clean the Monitor and Accessories
WARNING Be sure to unplug the monitor battery charger power cord from the
electrical power outlet before inspecting or cleaning the battery charger.
Exposing the battery charger to liquids such as cleaning solutions while
connected to electrical power could result in electrical shock or fire.
WARNING Do not autoclave the monitor, battery, battery charger, or
accessories. Do not immerse the monitor, battery, or battery charger in liquid
when cleaning. Do not immerse accessories in liquid when cleaning unless the
accessory manufacturer’s cleaning instructions explicitly instruct you to do so.
Before cleaning, thoroughly inspect the monitor and all accessories for any signs of
damage, cracks, or improper mechanical function of keys or connectors. While gently
bending and flexing cables, inspect for damage, cracks, cuts, abrasions, extreme wear,
exposed wires or bent connectors. Confirm connectors securely engage. Report damage
or improper function to your service department. At least every 12 months, be sure to
thoroughly inspect the battery charger case and power cord for damage or extreme wear.
To clean the monitor, batteries, or battery charger:
1.Wipe the equipment with a nearly dry cloth moistened with one of the approved
cleaning solutions listed in the table on “Approved Cleaning Solutions” on page 46.
2. Thoroughly wipe off any excess cleaning solution. Do not let the cleaning solution run
into or accumulate in connector openings, latches, or crevices. If liquid gets into a
battery well or connectors, dry the area with warm air, then check the equipment to
confirm that it operates properly.
Caution Use only cleaning solutions which are recommended by Welch Allyn
for this equipment. Use of solutions which are not recommended or which have
a high acid content or are otherwise inappropriate can cause damage to the
equipment, including cracking and deterioration of the plastic case. Do not use
these solutions or similar products: Butyl alcohol, Denatured ethanol, Freon
Mild chlorine bleach solution, Isopropyl alcohol (except for the SpO
Trichloroethane, Trichloroethylene, Acetone, Vesphene II, Enviroquat
Staphene
®
, Misty®, Glutaraldehyde, Formula 409®, or Fantastik®.
• Thoroughly wipe off any excess cleaning
solution. Do not let cleaning solution run into
connector openings or crevices.
Warm water, Liquid soap, Coverage®,
®
Windex
, Ovation®, Hydrogen peroxide
solution, Wex-cide
b
c
®
, T.B.Q.
c
®
ECG cable,
extension cable
cable,
SpO
2
extension cable
Other accessories• Consult manufacturer’s instructions.Consult manufacturer’s instructions.
a. The equipment may be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood and other
body fluids. (Federal OSHA Standard on bloodborne pathogens: 29 CFR 1910.1030, 12/6/91.)
b. If liquid gets into the battery well or connectors, dry the area with warm air, then check the monitoring functions for proper operation.
c. Wex-cide (Wexford Labs, Inc., Kirkwood, MO) and T.B.Q. (Calgon Vestal Lab., Calgon Corp., St. Louis, MO) are disinfectants that meet
OSHA requirements, are EPA approved, and will not harm the outside of the monitor, battery, or battery charger. Wipe away
disinfectants with a water-dampened cloth after the manufacturer’s recommended period.
• Wipe gently with damp cloth moistened with
a mild detergent solution.
• Thoroughly wipe off any cleaning solution.
• Wipe the cable with a 70% isopropyl alcohol
pad and allow it to dry.
Recycling Monitor Components
When the battery, monitor, or battery charger reaches the end of its life,
recycle it locally according to national, state, and local regulations. You can
also return the battery, monitor, or charger to Welch Allyn for recycling.
Li++
Within the European Union
Do not dispose of this product as “unsorted municipal waste.” Prepare it for
reuse or separate collection as specified by Directive 2002/96/EC of the
European Parliament and the Council of the European Union on Waste Electronic
and Electrical Equipment (WEEE).
Mild detergent.
70% isopropyl alcohol pad.
If the monitor or battery (Li++) is contaminated, this directive does not apply. For more
specific information, see www.welchallyn.com/weee, or contact Welch Allyn Customer
Service.
Recycle monitor batteries (Li++) according to the Directive 91/157/EEC
(Batteries and accumulators containing certain dangerous substances) and
Directive 93/86/EEC (Labeling of batteries and accumulators containing
Li++
certain dangerous substances).
Directions for UseChapter 5 Maintenance47
Change the Network Name
This procedure allows you to change the network name assigned to the monitor (as long
as the current network name is one of the pre-set names available in the monitor Network
Name Menu).
Note
To change the network name:
1.P r e s s to access the Main Menu, then repeatedly press until SERVICE MENU
2. Press to display the Service Menu screen.
3. Press and hold and , then press to display the Network Name Menu.
Changing the monitor network name will cause the monitor to re-start and seek
to connect with the FlexNet network corresponding to the new name. Do not
attempt to change the network name unless you are a qualified biomedical
service engineer or technician.
is highlighted.
If the current monitor network name is one of the following pre-set names:
To change the network name, make sure YES is highlighted, then press to display
the following screen:
Press or to highlight the desired network name, then press . The
monitor automatically turns itself off, then turns on and seeks to connect to a FlexNet
network with the new network name.
If the current network name is a custom name, the monitor displays the following
screen:
You cannot change the network name using the Network Name Menu. Press to
return to the Service Menu. Contact Welch Allyn Technical Support for assistance.
49
6
Reference
Operating Settings
The following table lists all of the monitor settings and the default settings.
ParameterSet at
Patient Mode
(Adult [age 13 and older] or
Pediatric [age 12 and younger])
ECG screen mode (Single, Dual, 5
Sec, or Full Screen)
ECG 1 Lead SelectionYesYesYesYesII
ECG 2 Lead SelectionYesYesYesYesV (or III if no V lead)
ECG Size (Scale)YesNoNoYes1 mV/cm
Language NoNoYesYesEnglish
Mains Filter (off, 50, or 60 Hz) NoNoYesYes60
Vital Signs Alarm Volume
(high, low, or off)
With Acuity Connection
Without Acuity ConnectionNoNo
Equipment Alert Volume
(high, low, or off)
With Acuity Connection
Without Acuity ConnectionNoNo
HR/PR Alarm Limits
(Lower, Upper)
SpO
a. Set by clinician at Acuity Central Station.
b. Set by Acuity System Administrator during system installation.
c. SpO2 alarm limit range depends on the software version of the Acuity System to which the monitor is connected. (See “Heart Rate
and Arrhythmia Analysis Option” on page 52 and “Pulse Oximetry (SpO2) Specifications - Nellcor” on page 54.)
c
Set at AcuityPrevious
Monitor
NoYesNoYesAdult
Yes NoYesYesSingle
YesYesYesYesAdult: 50, 120 bpm
YesYesYesYesAdult: 90, 100%
NoNoYesYesU.S.
For Each
Patient
No
No
No
No
a
For Entire
Acuity
Unit
Yes
Yes
Yes
Yes
b
Setting
Retained at
Monitor at
Power-Up
No
Yes
No
Yes
Monitor Default
Setting
off
high
off
low
Ped: 50, 150 bpm
Ped: 90, 100%
50Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Specifications
Monitor Radio Specifications
CharacteristicSpecification
FlexNet Network2.4 GHz Wireless Local Area Network (WLAN) and 10/100 Base-T Ethernet
Frequency
ModulationFrequency Hopping Spread Spectrum (FHSS)
Output Power100 mW
IEEE 802.11 compliantYes
Monitors per Access Point15 (maximum) in most countries. In countries where available frequencies
a. When used within certain countries, authorization for use is restricted as follows:
a
France: The equipment is internally restricted to the 2.448-2.482 GHz frequency range.
Spain: The equipment is internally restricted to the 2.447-2.473 GHz frequency range.
Japan: The equipment is internally restricted to the 2.473-2.495 GHz frequency range.
Italy: Operation requires a user license.
Note: The frequency ranges specified above are subject to geographic-specific regulatory authorities.
network
2.402 to 2.480 GHz
are limited, this number is reduced.
ECG Specifications
The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and
Alarms specified ANSI/AAMI EC13-1992, except for Impulse response at the monitor
(section 3.2.9.8 part (c)), and Standardizing Voltage at the monitor and at Acuity (section
3.2.9.9). The channel also meets the American National Standard, Safe Current Limits for
Electromedical Apparatus (ANSI/AAMI ES1-1993).
Lead Fault IndicatorDisplayed chest icon with flashing indicator for each electrode
ECG Size (sensitivity)0.2, 0.5, 1, 2, 4, and 8 mV/cm
Display Sweep Speed25 mm/sec
Bandwidth
Local display
To Acuity Central Station
Sample Rate364 Hz (182 Hz with turning point decimation to Acuity Central Station)
Input ProtectionElectrosurgery and defibrillator protected when used with ECG cables
Electrosurgery interference suppressionIncluded on all vectors.
Lead Fail Sense Current70 nA dc typical for active leads.
Tall T-wave RejectionMeets AAMI (USA) EC13-1992, section 3.1.2.1.c, up through 1.2 mV
Monitored: II, III, V; Derived: I, aV
Monitored: II
0.5 to 94 Hz independent of patient mode
0.05 to 94 Hz independent of patient mode
specified in the Welch Allyn Products and Accessories booklet (810-0409XX).
140-280 nA dc typical for reference electrode, depending on number of
electrodes attached.
, aVL, aV
R
F
Directions for UseChapter 6 Reference51
CharacteristicSpecification
Common Mode Rejection
FILTER function OFF
FILTER function ON
Input Impedance>2.5 M¾ differential @ 60 Hz
Input Range (ac)10 mV peak to peak (local display)
Input Range (dc)Up to ±500 mV
System Noiseð30 µV peak-to-peak, RTI
QRS DetectorAdult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)
Heart Rate Range
Alarm Limits
Heart Rate Meter Response TimeResponds to change in heart rate within 5 to 9 seconds depending on
HR Display Update Interval at monitor1 second
HR Accuracy±3 beats per minute or 3%, whichever is greater
Heart Rate Response to ineffectively
paced QRS pattern
Heart Rate Averaging MethodHeart rate = 60/ latest average interval in seconds.
Drift Tolerance (AAMI Specification EC131992, 3.2.6.3)
Pacer Pulse DisplayPacer indicator shown on screen if PACER display turned ON; pacer spike
Pacer Pulse RejectionPacer detection range (i.e., will show the dashed vertical marker) for ±3 mV
<1 mV p-p RTI for 10V rms, 50/60 Hz into unbalanced input
<30 µV p-p RTI for 10V rms, 50/60 Hz into unbalanced input
10 mV peak to peak (Acuity Central Station)
Adult Width Range (Duration): 70 to 120 msec
Pediatric Width Range (Duration): 40 to 120 msec
25 to 350 beats per minute (measurement)
25 to 300 beats per minute (display)
25 to 245 beats per minute (lower)
30 to 250 beats per minute (upper)
physiological waveform. (As measured per AAMI standard EC13-1992
clause 4.1.2.1 (f), including 3.1.2.1 parts f. and g. waveforms.) Includes 1
second readout update interval.
Indicates rate of 30 to 46 during AAMI EC13-1992 part 3.1.4.1 part (f) and (g)
tests.
NOTE: AAMI Test 4.1.4 part f and g: Accuracy is affected (i.e., rate
increases) when QRS and pacer spikes are nearly simultaneous as
occasionally is the case during this AAMI test.
For higher heart rates, latest average interval = 7/8 of previous average
interval + 1/8 of latest interval.
For lower heart rates, latest average interval = 3/4 of previous average
interval + 1/4 of latest interval.
Transition rates for choice of formula include hysteresis and are 70 and 80
beats per minute.
80 beats per minute indicated for 80 beats per minute ECG plus drift
waveform
always shown if of sufficient amplitude.
to ±700 mV @ 0.1 ms width, ±2 mV to ±700 mV @ 0.2 to 2 ms pulse width in
electrically quiet environment. Thresholds automatically adjust to reject
repetitive ambient noise. Operates even while pacer indication is disabled.
Will not count as heartbeats approximately 95% of pacemaker pulses
within pacer detection range, with or without AAMI (EC13-1992) tails of 4,
25, 50, 75, or 100 ms decay time constant, whose tail amplitudes are up to
25%, 2mV maximum, whether ventricular only, or A-V sequential pulses
(150 and 250 ms separation), all per AAMI tests 3.1.4.1 and 3.1.4.2
Response to Irregular Rhythm (AAMI Specification EC13-1992, 3.1.2.1. Part e.)
Ventricular Bigeminy (VB)78 to 81 bpm (80 bpm expected)
Slow Alternating VB57 to 65 bpm (60 bpm expected)
Rapid Alternating VB118 to 123 bpm (120 bpm expected)
Bidirectional Systole88 to 93 bpm (90 bpm expected)
52Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Heart Rate and Arrhythmia Analysis Option
Method for Calculating Heart Rate
MonitorDetermined by monitor (displayed at monitor)
Heart rate = 60 / latest average interval in seconds.
For higher heart rates, latest average interval = 7/8 of previous average interval + 1/8 of latest
interval
For lower heart rates, latest average interval = 3/4 (previous average interval) + 1/4 latest interval.
Transition rates for choice of formula include hysteresis and are 70 and 80 beats per minute.
Acuity System with
Arrhythmia Option
Arrhythmia Analysis Option when Connected to Acuity
ST AnalysisST Analysis can be performed for any or all of seven leads, depending on the operator selection.
Determined by Acuity Arrhythmia Option software (displayed at Acuity Central Station)
The beat-to-beat heart rate (HR) value is calculated as follows:
HR = 60000/actual RR (bpm).
Actual RR = time between last detected QRS complex and previously detected QRS complex
(ms).
Average HR is calculated on the basis of the mean RR interval in the last 6 seconds or 8 RR
intervals (whichever is shorter).
The operator can select a measurement offset.
ST segment shifts are recorded in continuous trend data every second. The operator can inspect
trend data to see the duration and elevation or depression for each episode for any time period
recorded. The operator can also inspect a summary of ST segment shift data within tabular trends.
Heart RateHeart rate information is available in the trend data which can be viewed on the display or printed.
The operator can inspect the trend data to see the lowest, highest, and median (averaged) heart
rates. Trend data also includes the total beats per range of time.
Definition of Pause
Arrhythmia Event
A pause is defined as the R-R interval which is greater than or equal to two times the average R-R.
Directions for UseChapter 6 Reference53
Pulse Oximetry (SpO2) Specifications - Masimo
CharacteristicSpecification
Saturation (% SpO2)
Range
Resolution
Alarm Limits
With Acuity 6.0 or higher
With Acuity 5.4X or lower
a
Probe Accuracy (Adults, Pediatrics)
No Motion
During Motion
b
Pulse Rate
Range
Resolution
Alarm Limits
Pulse Rate Accuracy
No Motion
During Motion
2
Display Update Interval at monitor1 second
Alarm Hold-Off Time Period10 seconds; resets if the sensor reports levels within limits before 10
CircuitryMicroprocessor controlled
Electrosurgery interference suppression Yes
Sensor CompatibilityCompatible only with Masimo sensors listed in the Welch Allyn Products
Sensor LEDs
RED Wavelength
INFRARED Wavelength
Sensor Energies (Radiant Power)0.13 mW to 0.79 mW at 50 mA pulsed
alarm limit range depends on the software version of the Acuity System to which the monitor is connected.
a. SpO
2
b. Motion is defined as rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion between 1
to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO
ECG monitor. This variation equals ±1 standard deviation which encompasses 68% of the population.
1% to 100%
1%
50% to 99% (lower); 51% to 100% (upper)
80% to 99% (lower); 81% to 100% (upper)
70% to 100% ±2 counts
0% to 69% unspecified
70% to 100% ±3 counts
0% to 69% unspecified
26 to 239 beats per minute
1 beat per minute
25 to 245 beats per minute (lower)
30 to 250 beats per minute (upper)
±3 beats per minute
±5 beats per minute
seconds elapses
Automatic self-test of oximeter when powered on
Automatic setting of default parameters
Automatic alarm messages
and Accessories booklet (810-0409-XX).
660 nm (nominal)
905 nm (nominal)
against a laboratory co-oximeter and
2
WARNING Interfering Substances: Carboxyhemoglobin may erroneously
increase readings. The level of increase is approximately equal to the amount of
carboxyhemoglobin present. Dyes, or any substances containing dyes, that
change usual arterial pigmentation may cause erroneous readings.
WARNING Although the SpO
saturation (with Acuity 6.0 software or higher), the SpO
alarm limit range can be adjusted down to 50%
2
performance and
2
accuracy is not specified below 70%.
54Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Pulse Oximetry (SpO2) Specifications - Nellcor
CharacteristicSpecification
Saturation (% SpO2)
Range
Resolution
Alarm Limits
Probe Accuracy
Pulse Rate
Range
Resolution
Alarm Limits
Pulse Rate Accuracy±3 beats per minute
Display Update Interval at the monitor1 second
Alarm Hold-Off Time Period10 seconds; resets if the sensor reports levels within limits before 10
CircuitryMicroprocessor controlled
Electrosurgery interference suppressionYes
Sensor CompatibilityCompatible only with Nellcor sensors listed in the Welch Allyn Products and
Sensor LED WavelengthsWithin 500 to 1,000 nm
Sensor Energies (Radiant Power)Does not exceed 15 mW
a. SpO2 alarm limit range depends on the software version of the Acuity System to which the monitor is connected.
b. Refer to the Welch Allyn Products and Accessories guide (810-0409-XX) for accuracy specifications for all Nellcor SpO2 probes
recommended for use.
c. Although some of the listed Nellcor sensors can be used with neonates with other pulse oximetry devices, the monitor is only intended
for use with adult and pediatric patients, not with neonates.
a
With Acuity 6.0 or higher
With Acuity 5.4X or lower
b c
(Adults, Pediatrics)
1% to 100%
1%
50% to 99% (lower); 51% to 100% (upper)
80% to 99% (lower); 81% to 100% (upper)
70% to 100% (0% to 69% unspecified)
OxiMax Max-A, Max-AL±2 counts
OxiCliq N±2.5 counts
D-YS±3 counts
DS-100A±3.5 counts
26 to 239 beats per minute
1 beat per minute
25 to 245 beats per minute (lower)
30 to 250 beats per minute (upper)
seconds elapses
Automatic self-test of oximeter when powered on
Automatic setting of default parameters
Automatic alarm messages
Accessories booklet (810-0409-XX).
WARNING Interfering Substances: Carboxyhemoglobin may erroneously
increase readings. The level of increase is approximately equal to the amount of
carboxyhemoglobin present. Dyes, or any substances containing dyes, that
change usual arterial pigmentation may cause erroneous readings.
WARNING Although the SpO
saturation (with Acuity 6.0 software or higher), the SpO
alarm limit range can be adjusted down to 50%
2
performance and
2
accuracy is not specified below 70%.
Directions for UseChapter 6 Reference55
Patient Alarm and Equipment Alert Specifications
Characteristic Specification
Visual Alarm Indicator at the monitor
Flashing GREEN LED
Flashing RED LED
Continuously ON RED LED
Flashing YELLOW LED
Continuously ON YELLOW LED
Audio Tone LocationsMonitor
Audio Tone Frequency2760 Hz
Life-Threatening Arrhythmia Alarm
Tone Pattern
Patient Alarm Tone Pattern
Equipment Alert Tone Pattern
Audio Tone VolumeThe monitor audio tone volume is configured by the Acuity System to High,
LimitsSetable on all parameters
Alarm ControlAutomatic preset or manual settings
Alarm PriorityHighest priority: Patient alarms
Alarm on TachycardiasMost tachycardias will alarm in less than 8 seconds. These include AAMI
Alarm Holdoff Time Period
Acuity-Configurable Audio Alarm Delay
at the monitor
Patient Alarm Tone Silence from the
monitor or Suspend from Acuity
a
Normal operation
Patient alarm
Patient alarms are silenced
An equipment alert or not connected to the network
Equipment alert suspended for 90 seconds at Acuity or low battery alert
acknowledged (dismissed)
Acuity Central Station (when connected)
1 second on, 1 second off
1second on, 2 seconds off
1second on, 4 seconds off
Low, or Off. The monitor can be configured with separate audio tone volume
settings for when it is connected to an Acuity System and when it is not.
Lowest priority: Equipment alerts
3.1.2.1 part f. waveforms. Certain multifocal tachycardias may initially alarm
as "low rate."
HR = 3 seconds
% SpO
, PR = 10 seconds
2
When a monitor is connected to an Acuity System, the audio alarms at the
monitor can be delayed up to 4 minutes and 15 seconds. The delay time is
selected in Acuity software at the time of Acuity installation. Visual alarm
indications are not delayed.
The monitor LED is continuously ON RED and the audio tone is silenced for 90
seconds (non-adjustable).
If original alarm was silenced from the monitor, new patient alarms or
equipment alerts break the silence at the monitor, but only life-threatening
arrhythmia alarms break the silence at Acuity. If original alarm suspended at
Acuity, only life-threatening arrhythmia alarms break the silence at the
monitor and Acuity.
Equipment Alert Acknowledge from the
monitor
Equipment Alert Suspend from AcuityThe LED is continuously ON YELLOW and the audio tone is silenced for 90
Patient Alarm Tone Reset from the
monitor or Resume from Acuity
Patient Out of Range; Transmitter
Failure
The LED returns to the pre-alert state (except Low Battery remains
continuously ON YELLOW) and the auditory tone is dismissed.
seconds (non-adjustable).
For a patient alarm tone that has been silenced, resets the tone.
An equipment alert is generated whenever the monitor fails to communicate
to an Acuity System after a connection has been successfully established. In
addition, the “No Acuity” icon is displayed on the monitor display.
56Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Characteristic Specification
Transmitter Battery FailureAn equipment alert is generated before the monitor battery becomes
Nurse CallWhen is pressed and the instrument is connected to an Acuity System,
a. To help minimize false alarms, the monitor briefly delays or "holds off" triggering both audible and visual alarms for limit violations
for these vital signs. After the alarm hold-off period begins, if the monitor detects that the patient’s vital sign has returned to
acceptable limits, the monitor cancels the alarm hold-off. The next time a vital sign limit is violated, the monitor starts a new holdoff period.
exhausted.
the monitor sends a Nurse Call message to the Acuity System.
Display Specifications
CharacteristicSpecification
TypeMonochrome passive matrix; LCD module
Resolution320 x 200 pixels
Active Viewing Area2.26 x 1.41 in. (57.5 x 35.9 mm)
Pixel Pitch0.0071 in. (0.18 mm)
Pixel Size0.0065 in. (0.165 mm)
Viewing Angle6 o’clock position
Display Colorblack on white
Operating Temperature0° to 40° C
Shipping and Storage Temperature-20° to 60° C
Operating Altitude-2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude-2,000 to 40,000 ft (-610 to 12,192 m)
Operating Relative Humidity15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shipping and Storage Relative Humidity15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Drop1 meter onto vinyl tile over concrete per EN60601-1
Shock50 g
Vibration, Random0.02g
Degree of Protection Against IngressIPX1 Rating, Drip Proof per EN60529: 1991
Electromagnetic Compatibility (EMC)EN60601-1-2: 1993
Caution The monitor may not meet performance specifications if it is not used
or stored within these environmental specifications.
2
/Hz from 10 to 500 Hz, ramping down to 0.002g2/Hz at 2000 Hz.
Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA DO160D, Category C.
Directions for UseChapter 6 Reference57
Physical Specifications
Protection Classifications, all Configurations
CharacteristicSpecification
Type of Protection against Electric
Shock—Monitor Type: CF
Degree of Protection Against Electric
Shock, for Parts Applied to Patients
Recovery time following defibrillator
discharge
Method of DisinfectionNot suitable for autoclaving (see cleaning instructions on “Inspect and
Flammable AnestheticsNot suitable for use with flammable anesthetics.
Height7.30 in (18.5 cm)
Width3.50 in (8.9 cm)
Depth1.46 in (3.7 cm)
Weight (including battery)
Model 402
Model 404
Battery operation only
Battery must be recharged in separate battery charger.
IEC EN 60601-1, 2nd Edition
See monitor labels. CF defibrillator protected.
IEC EN 60601-1, 2nd Edition
Less than or equal to 10 seconds
Clean the Monitor and Accessories” on page 45).
14.7 oz (0.42 kg)
17.3 oz (0.49 kg)
Battery Specifications
CharacteristicSpecifications
Reorder Number
Lithium Ion Battery
2EA Active A
Battery TypeRechargeable, Lithium Ion
Battery Capacity2 cells, 8.4 V, 1800 mA-hr
Battery Weight4.5 oz (0.13 kg)
Battery ChargerExternal device
Battery Fuse Rating5 A, 125 V (not user-accessible)
Operating Times on Battery
Battery Recharge Time4 hours at 25° C (typical)
Battery Lifetime300 charge/discharge cycles to 70% of original
a. Battery operating times based on these conditions: new fully-charged battery operating at 25° C,
the monitor connected to Acuity, eight patient alarms per hour, minimal motion artifact.
Note
The following factors reduce battery operating time:
•Amount of time not connected to Acuity.
•Frequency and duration of alarms and alerts.
•Amount of operator activity using monitor keys (activates display).
•Age of battery.
•Amount of motion artifact during SpO
008-0647-XX
Note: Internal electronic overload circuitry is used as
the primary method of protection.
WARNING The monitor battery is Lithium Ion. Do not incinerate, submerge,
crush, disassemble, or autoclave. If a battery has been submerged in liquid,
discard the battery properly; do not try to recharge or reuse the battery. Do not
short the battery terminals. Do not try to connect the battery to any device
except the monitor or the battery charger. Do not expose to high temperature
(above 60° C or 140° F). Use only the specified monitor battery charger.
Directions for UseChapter 6 Reference59
Battery Charger Specifications
CharacteristicSpecification
Reorder Number
Universal Battery Charger
Active C
Functional Specifications
CapacityEight charging bays; able to charge four (Lithium Ion) batteries
Protection Classifications
Duty CycleContinuous
Type of Protection Against Electric
Shock
Degree of Protection Against Harmful
Ingress of Water
Method of DisinfectionNot suitable for autoclaving. (See cleaning instructions on “Inspect and
Flammable AnestheticsNot suitable for use with flammable anesthetics.
Environmental Specifications
Operating Temperature0° to 40° C
Shipping and Storage Temperature-20° to 60° C
Operating Altitude-2,000 to 15,000 feet (-610 to 4,572 m)
Shipping and Storage Altitude-2,000 to 40,000 feet (-610 to 12,192 m)
Operating Relative Humidity15% to 95%, noncondensing
Shipping, Storage Relative Humidity15% to 95%, noncondensing
Shock30 g
Vibration0.01g
Electromagnetic Compatibility (EMC)EN60601-1-2: 1993
Physical Specifications
Length15.0 in (38.1 cm)
Width9.0 in (22.9 cm)
Height3.6 in (9.1 cm) including feet
Weight3.5 lb (1.6 kg)
Electrical Specifications
Rated Input100 V-240 V ac, 600 mA, 50/60 ± 3 Hz, Electrical Class I
Rated FusesT1.25 A/250 V, Time-Delay, 5x20 mm
Rated Output per charging bay
Output Over-Current Electronic overload protection
Additional FeaturesDetachable power cord, pilot light
LED Indicators
LED OFFNo battery detected.
Flashing GREEN LED
1 sec ON, 3 sec OFF
1 sec ON, 1 sec OFF
Continuously ON GREEN LEDBattery is charged.
Continuously ON YELLOW LEDBattery or charging bay fault.
a. Per EN 60601-1 unless otherwise stated.
a
008-0651-XX
simultaneously.
Class I, (Protectively Earthed) with Double Insulation
For ordinary, indoor locations only.
Clean the Monitor and Accessories” on page 45.)
2
/Hz from 5 to 500 Hz, 30 minutes per axis
8.4 V ± 100 mV dc @1 A max.
when charge is completed, or fault detected.
Battery detected, waiting to be charged.
Battery is charging.
60Chapter 6 ReferenceWelch Allyn Micropaq Monitor
EMC Compliance
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment. This device complies with IEC EN 60601-1-2:2001.
•All medical electrical equipment must be installed and put into service in accordance
with the EMC information provided in this document and the Micropaq Monitor Directions For Use.
•Portable and mobile RF communications equipment can affect the behavior of
medical electrical equipment.
The monitors and battery charger comply with all applicable and required standards for
electromagnetic interference.
•It does not normally affect nearby equipment and devices.
•It is not normally affected by nearby equipment and devices.
•It is safe to operate the monitor in the presence of high-frequency surgical
equipment.
•However, it is good practice to avoid using the monitor in extremely close
proximity to other equipment.
Directions for UseChapter 6 Reference61
Monitor
Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of the monitor should assure that it is used in such an environment.
The Model 400 Series Monitor uses RF energy only for its
internal function.
a
Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Model 400 Series Monitor is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
No connection to mains
(battery-operated)
No connection to mains
(battery-operated)
network that supplies buildings used for domestic purposes.
IEC 61000-3-3
a. The Model 400 Series Monitor contains a 2.4-GHz frequency-hopping spread-spectrum transmitter for the purpose of wireless
communication. The radio is operated according to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE
Directive (1995/5/EC). The radio is excluded from the EMC requirements of 60601-1-2:2001, but should be considered when
addressing possible interference issues between this and other devices.
62Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of the monitor should assure that it is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic
Environment—Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and voltage
variations on powersupply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
<5% U
t
(>95% dip in Ut)
No connection to mains
(battery-operated). No
other cables requiring
EFT/Burst testing.
No connection to mains
(battery-operated).
No connection to mains
(battery-operated).
Since there is no connection to the
mains, there is no requirement for
mains quality.
for 0.5 cycle
40% U
t
(60% dip in Ut)
for 5 cycles
70% U
t
(30% dip in Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 sec
3 A/m3 A/mPower frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NoteU
is the AC mains voltage prior to application of the test level.
t
Directions for UseChapter 6 Reference63
Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of the monitor should assure that it is used in such an environment.
Immunity TestIEC 60601 Test
Level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance
Level
3 V
rms
3 V/m
Electromagnetic Environment—Guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Model 400 Series
Monitor, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended Separation Distance
d = 1.2
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the
transmitter in watts according to the transmitter
manufacturer and d is the recommended separation distance
in meters.
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey
compliance level in each frequency range
a
, should be less than the
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Note 1At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Model 400 Series Monitor is used exceeds the applicable RF compliance level above, the Model 400 Series
Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the monitor.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
64Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the
Model 400 Series Monitor
The Model 400 Series Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the monitor can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the monitor as
recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output
Power of Transmitter
W
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d = 1.2
PPP
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
Directions for UseChapter 6 Reference65
Battery Charger for the Monitor
Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The Battery Charger for the Model 400 Series Monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the battery charger should assure that it is used in such an environment.
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class B
Class A
Complies
The Battery Charger uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The Battery Charger is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
66Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Battery Charger for the Model 400 Series Monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the monitor should assure that it is used in such an environment.
Immunity TestIEC 60601 Test levelCompliance LevelElectromagnetic Environment—
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
±2 kV common mode
<5% U
t
(>95% dip in Ut)
for 0.5 cycle
40% U
t
(60% dip in Ut)
for 5 cycles
70% U
t
(30% dip in Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 sec
<5% U
t
(>95% dip in Ut)
for 0.5 cycle
40% U
t
(60% dip in Ut)
for 5 cycles
70% U
t
(30% dip in Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Battery Charger requires
continued operation during a power mains
interruption, it is recommended that the
Battery Charger be powered from an
uninterruptible power supply or battery.
3 A/m3 A/mPower frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NoteU
is the AC mains voltage prior to application of the test level.
t
Directions for UseChapter 6 Reference67
Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Battery Charger for the Model 400 Series Monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the battery charger should assure that it is used in such an environment.
Immunity testIEC 60601 Test
Level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance
Electromagnetic Environment—Guidance
Level
Portable and mobile RF communications equipment should
be used no closer to any part of the Battery Charger,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended Separation Distance
3 V
rms
d = 1.2
3 V/md = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the
transmitter in watts according to the transmitter
manufacturer and d is the recommended separation distance
in meters.
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey
compliance level in each frequency range
Interference may occur in the vicinity of equipment marked
with the following symbol:
a
, should be less than the
b
.
Note 1At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Battery Charger is used exceeds the applicable RF compliance level above, the Battery Charger should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the Battery Charger.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
68Chapter 6 ReferenceWelch Allyn Micropaq Monitor
Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the
Battery Charger for the Model 400 Series Monitor
The Battery Charger for the Model 400 Series Monitor is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Battery Charger can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Battery Charger as recommended below, according to the maximum output power of
the communications equipment.
Rated Maximum Output
Power of Transmitter
W
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation Distance According to Frequency of Transmitter
m
Cardiac arrest and SpO2, 27
Cautions, general
Change network name
Change the battery
Change waveform display
Charge a battery
, 5
, 47
, 43
, 14
, 43
Charger
Recharge a battery
Specifications
Cleaning
, 45
Connect a new patient
Connect to the network
Customize alarms
, 43
, 59
, 19, 21
, 19
, 21
D
Default
, 22
70IndexWelch Allyn Micropaq Monitor
Acuity and settings, 16, 49
Operating settings
DEFECTIVE SPO2 SENSOR
, 16, 49
, 40
Defibrillation
During SpO2 monitoring
Warning
Delays for alarms
, 3
, 35
, 26
Demo Mode
Accessing
, 16
Display values, alarm limits
Toggling between modes
Demonstration Mode
Disconnection symbol
, 17
, 16
, 11
Discontinue monitoring a patient
Display
Alert messages
Artifact interference
Blank during sleep mode
Indicator symbols
Pacemaker signals
Sleep mode
Specifications
Waveform display options
DROPOUT alert at Acuity
, 40
, 22
, 11
, 11
, 11
, 11
, 56
, 14
, 30, 43
, 17
, 30
Automatic reconnection
Defined
, 7
Operation out of range
, 7
, 7
Full Screen waveform display
G
General cautions, 5
General Warnings
Green LED
, 3
, 9
H
Hazards
General cautions
General warnings
, 5
, 3
Holdoffs (delays) for alarms
Hyperbaric chamber
Warning, do not use
Hypotension and SpO2
Hypothermia and SpO2
, 4, 21
, 27
, 27
I
, 14
, 35
E
ECG
Alarm holdoff (delay)
Electrode application
Electrode lead off
Lead selection
Scale selection
Size (scale) selection
Electrode application
Electrosurgery warning
END TELE
, 12, 30
Environmental specifications
, 35
, 21
, 23
, 12
, 12
, 12
, 21
, 3
, 56
Equipment alert
Description
Messages
, 38
, 40
Errors
Error message
Error number
Ethernet
, 1
EXCESSIVE ECG OFFSET
, 5
, 5
, 40
F
FCC
Warning, RF
Federal Communications Commission
see FCC
FlexNet
Access points
, 4
, 1, 7
, 7
In-Service
see Demo Mode
Inspection
Intended Use
, 45
, 1
K
Keys
Alarm/Alert Silence
Nurse Call
, 9
Scroll Up/Down
Select
, 10
KEY STUCK
, 40
, 10
, 9
L
Lead off indication, 23
Lead selection
LEDs defined
List of patients
List of rooms
Lithium ion battery
Lockout menu
Low battery icon
LOW BATTERY message
, 12
, 9
, 20
, 21
, 43
, 13, 49
, 11
, 40
M
Magnetic Resonance Imaging
Directions for UseIndex71
see MRI
Main Menu
EXIT
, 12
Menu Lock option
Restricting access
, 13, 49
, 13
Maintenance
Inspection and cleaning
Operator maintenance
, 45
, 43
Masimo
SET technology
Menu Lock option
Messages for alerts
Messages from Acuity
, 6
, 13, 49
, 40
, 15
Micropaq
Acuity messages
Adjust alarm limits
Alarms
, 36
Audible indicators
Automatic reconnection
Battery recharging
Communication with network
Connect to the network
Default operating settings
Default settings
Demo Mode
Equipment alert
Features
, 6
NEW PATIENT
Operating settings
Out of range operation
Reassign to new patient
Specifications
, 15
, 37
, 9
, 7
, 43
, 7
, 19
, 16
, 49
, 16
, 38
, 34
, 15
, 7
, 34
, 50
Monitor
Audible indicators
Features
, 6
Out of range
Recycling
, 46
Specifications
, 9
, 29
, 50
Monitoring
At Acuity
Connect a new patient
, 41
, 19
Reconnect a recently monitored patient
Stop monitoring a patient
Motion interference and SpO2
, 30
, 27
MRI
Warning, do not use monitor
, 4, 21
N
, 31
Automatic reconnection
Connect the monitor
Disconnection symbol
Ethernet
FlexNet
Name change
, 1
, 7
, 47
Operation out of range
NEW PATIENT
NEW ROOM
, 12, 34
, 12, 32
, 7
, 19
, 11
, 7
NIBP
While monitoring SpO2
NO ECG CABLE DETECTED
NO SPO2 SENSOR DETECTED
Nurse Call Key
, 9
, 27
, 40
, 40
O
Operation
Default settings
Settings
Operator maintenance
Out of range
Monitor a patient out of range
OxiMax technology, Nellcor
, 16
, 15
, 43
, 7
, 29
, 6
P
Pacemaker
Signal detection, display
Warning, rate meters
Warning, wireless radio
Pacer
see Pacemaker
Pacer detection
, 49
Patient
Alarms
, 36
Reassign to a new room
Reconnect a recently monitored patient
TRANSFER
, 33
Patient alarms
Download defaults from Acuity
Patient list
, 20
Patient preparation for ECG
Physical specifications
Power specifications
Prepare the patient
Pulse oximetry
see SpO2
, 11
, 3
, 4
, 32
, 21
, 21
, 57
, 57
, 21
, 31
Nellcor
OxiMax technology
Neonatal
Not intended for use
Network
Access points
, 7
, 6
, 3
R
Radio Frequency
see RF
Reassign Micropaq to new patient
, 34
72IndexWelch Allyn Micropaq Monitor
Reassign to a new room, 32
Recharge a battery
, 43
Reconnect a recently monitored patient
Reconnection to network
, 7
Recycling
Batteries
Monitor
Red LED
Replace the battery
Respond to equipment alert
Respond to patient alarms
RF warning, exposure
Room selection
Routine maintenance
, 46
, 46
, 9
, 43
, 38
, 36
, 4
, 21
, 43
S
Scroll Up/Down Keys, 9
Select
Patient
, 20
Room
, 21
Select Patient at Central
Select Room at Central
Select Key
, 10
Select Room at Central
Set patient alarms
, 37
SET technology, Masimo
Settings, operating
Shock and SpO2
, 27
Silence Alarm/Alert Key
Silence alarms
, 36
Simulation
see Demo Mode
Site preparation for ECG
Sleep mode, display
Snapshot print, Nurse Call
Specifications
Alarms
Battery
Battery charger
Display
Environmental
Physical
SpO2
, 50
, 55
, 57
, 59
, 56
, 56
, 57
, 53
SpO2
Alarm holdoff (delay)
Anemia
, 27
Arterial occlusion
Cardiac arrest or shock
Hypotension
Hypothermia
, 27
, 27
Light interference
Monitoring
, 26
Monitoring and defibrillation
Motion interference
, 20
, 21
, 32, 33
, 6
, 49
, 10
, 21
, 11
, 9
, 35
, 27
, 27
, 27
, 26
, 27
, 31
Specifications
Vasoconstriction
Warnings
While monitoring NIBP
Stop monitoring a patient
Symbols, display
System error
, 53
, 27
, 26
, 27
, 30
, 11
, 40
T
Training
see Demo Mode
TRANSFER
, 12, 33
Transfer a patient to new room
U
Unassigned rooms, 21
UNRECOGNIZED SPO2 SENSOR
V
Vasoconstriction and SpO2, 27
Venous pulsations and SpO2
, 26
VERY LOW BATTERY message
Volume of audible indicators
, 9
W
Warnings
General
, 3
SpO2
, 26
Warranty service
Waveform display options
, 5
, 14
Wireless
Warnings about limitations
Wireless Local Area Network
see WLAN
WLAN
, 1, 7
Y
Yellow LED, 9
, 33
, 40
, 40
, 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.