Welch Allyn Micropaq 402, Micropaq 404 User manual

Micropaq
®
Monitor
Directions for Use
Model 402 and Model 404 Software version 1.5X
ii Welch Allyn Micropaq Monitor
Copyright 2007, Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or any part thereof without permission from Welch Allyn.
Welch Allyn assumes no responsibility for injury or for any illegal or improper use of the product that may result from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.
Welch Allyn
®
, Acuity®, Micropaq®, FlexNet® and Flexible Monitoring® are registered trademarks of Welch
Allyn, Inc.
Nellcor
®
is a registered trademark of Nellcor Puritan Bennett. Masimo® and SET® are registered trademarks
of Masimo Corporation.
Software in this product is Copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated within this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, please call Welch Allyn Technical Support:
USA 800 535 6663
315 685 4560
Canada 800 561 8797 China 86 21 6327 9631
European Call Center 353 46 9067790 France 33 1 55 69 58 49
Germany 49 7477 9271 86 Japan 81 3 3219 0071
Latin America 305 669 9003 Netherlands 31 157 505 000
Singapore 65 6419 8100 South Africa 27 11 777 7555
United Kingdom 44 1332 363812
Australia 61 2 9638 3000
800 074 793
Reorder Part Number 810-1464-XX
Manual Part Number 810-1177-06 Rev. A, 02/2007
Welch Allyn 8500 SW Creekside Place Beaverton, Oregon 97008-7101 USA
www.welchallyn.com
Printed in USA
Contents
1 - General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
iii
Summary of New Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
General Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Introducing the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Model 402 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Model 404 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Understanding the Monitor and the FlexNet Network . . . . . . . . . . . . . . . . . . . . . 7
Monitor Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Controls and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Operating Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Demonstration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2 - Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Connect a New Patient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Connect to the Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Perform ECG Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Perform SpO2 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Install the Carrying Pouch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Monitor a Patient Out of Range of Acuity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Stop Monitoring a Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Reconnect a Recently Monitored Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Reassign a Monitored Patient to a New Room in the Same Unit . . . . . . . . . . . . 32
Transfer a Monitored Patient to a New Room in a Different Unit . . . . . . . . . . . . 33
Reassign the Monitor to a New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3 - Alarms & Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
About Alarms and Alerts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Alarm Holdoffs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Respond to a Patient Alarm at Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Customize Patient Alarm Limits at the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . 37
Respond to an Equipment Alert at the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . 38
Alert Messages and Display Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
iv Contents Welch Allyn Micropaq Monitor
4 - Monitor Patient at Acuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5 - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Change the Battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Recharge a Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Inspect and Clean the Monitor and Accessories . . . . . . . . . . . . . . . . . . . . . . . . 45
Recycling Monitor Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Within the European Union . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Change the Network Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6 - Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Operating Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Monitor Radio Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
ECG Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Heart Rate and Arrhythmia Analysis Option . . . . . . . . . . . . . . . . . . . . . . . . . 52
Pulse Oximetry (SpO
Pulse Oximetry (SpO
Patient Alarm and Equipment Alert Specifications . . . . . . . . . . . . . . . . . . . . 55
Display Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Environmental Specifications (with Battery Installed) . . . . . . . . . . . . . . . . . 56
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Battery Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Battery Charger Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
EMC Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
) Specifications - Masimo . . . . . . . . . . . . . . . . . . . . . 53
2
) Specifications - Nellcor . . . . . . . . . . . . . . . . . . . . . . 54
2
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
1
1

General Information

Summary of New Features

The following table summarizes the new features that are implemented in this software version 1.5. The version previously released was 1.4.
Feature Reference
Patient alarm tones are now discontinuous, and life-threatening alarms beep at a faster pace than other patient alarms.
If you press to silence a patient alarm tone and a new patient vital sign alarm or equipment alert occurs, a tone for the new alarm or alert will break the silence period at the monitor.

Intended Use

The Micropaq™ monitor is intended to be used by clinicians for single or multiparameter vital signs monitoring of ambulatory and nonambulatory pediatric and adult patients in health care facilities. It is also intended for patient transport. The monitor is able to withstand light rain exposure over short periods of time (uniform distribution of approximately 1 mm of water/ minute for 10 minutes or less).
See “Respond to a Patient Alarm at
Monitor” on page 36.
The Micropaq monitor is intended to operate with an Acuity wireless communication over Welch Allyn’s FlexNet devices to the Acuity Central Station through hardwired Ethernet networks and Wireless Local Area Networks (WLANs). If the Micropaq monitor is moved out of range or loses communication with the FlexNet network, it continues to monitor the patient, display patient data, and generate local patient alarms or alert messages.
The ECG channel is intended primarily for five-lead ECG monitoring, although three-lead ECG monitoring is supported.
The Pulse Oximetry channel is intended for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO (measured by an SpO
sensor).
2
network. FlexNet connects multiple
®
Central Station through
) and pulse rate
2
2 Chapter 1 General Information Welch Allyn Micropaq Monitor
The most likely locations for patients monitored by this device are step-down units, telemetry departments, general medical/surgical floors, emergency departments, and in­hospital transport.
This guide was written for clinicians. Although this guide may describe some monitoring techniques, Welch Allyn expects that the operator is a trained clinician who knows how to take and interpret a patient’s vital signs.
Federal USA law restricts sale of the device identified in this manual to, or on the order of, a licensed medical practitioner.
Symbols
Warning Warning statements in this manual identify conditions or practices that could result in personal injury.
Caution Caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property.
The following symbols appear on the monitor or accessories.
Li++
Direct current Enclosure Protection Drip proof: Classification
IPX1
IPX1 per EN60529: 1991
Alternating current (battery charger) Fuse
The CE Mark and Notified Body Registration Number signify the device has met all essential requirements of European Medical Device Directive 93/42/EEC
Signifies the device has met all essential requirements of European Medical Device Directive 93/42/EEC for a Class 1 product
C
kj
This device has been tested and certified by the Canadian Standards Association International to comply with applicable U.S.
US
and Canadian medical safety standards. Patient connections are Type CF, isolated for direct cardiac application, and protected
against defibrillation (battery charger) Protective earth ground (battery charger) Protected during defibrillation
Lithium Ion battery Separate batteries from other disposables for
recycling
Li++
Caution: Refer to Directions For Use and
For indoor use only (battery charger) accompanying documentation
This way up Keep away from rain
Humidity limit Stacking limit (by number)
Fragile Altitude limit
Temperature limits IATA/ICAO Hazard Class 9 Package
(International Air Transport Association/
9
International Civil Aviation Organization) Non-ionizing electromagnetic radiation Single use only
Directions for Use Chapter 1 General Information 3
Monitor Front Panel Keys
Select Key and Silence Patient Alarm/ Equipment Alert Key- Selects the choice highlighted on the menu. During patient alarms, silences the tone at the monitor and at Acuity (if connected) for 90 seconds. During equipment alerts, silences or acknowledges (dismisses) the alert.
Nurse Call Key - When connected to Acuity, pressing this key sends Acuity a Nurse Call message.
Recycle the monitor and battery separately from other disposables. www.welchallyn.com/weee
Battery Charger Label and LEDs
Green LED on continuously Battery is fully charged.
Scroll Up Key and Reset Alarm Tone Key-
Scrolls up menus on the display. During patient
alarms, resets the tone at the monitor and at
Acuity (if connected).
Scroll Down Key and Main Menu Key- Pressing
this key scrolls down menus on the display, or
causes the Main Menu to appear if no menu is
displayed.
Green LED flashing
Green LED flashing very slowly Battery detected and waiting to be charged.
Yellow LED on continuously Something is wrong with the battery or the
General Warnings and Cautions
Familiarize yourself with all warnings and cautions before using the monitor.
WARNING When considering a treatment protocol that involves wireless communication of patient data, be sure to recognize some limitations inherent in wireless communications. When the monitor is not connected to the network:
There are no patient alarms or alerts at the Acuity Central Station.
Acuity does not perform arrhythmia and ST analysis on the patient data
and does not generate related alarms.
Patient data is not saved.
Pressing (Nurse Call) on the monitor does not cause any response
at Acuity.
kj
Battery is charging.
charger. (See
Response”
“Battery Status and Possible
on page 44.)
WARNING Do not try to monitor neonatal patients with the monitor. The monitor is intended for adult or pediatric patients. It is not intended for use with pediatric patients (or infants) weighing less than 22 lbs (10 kg).
4 Chapter 1 General Information Welch Allyn Micropaq Monitor
WARNING Always check the patient mode at Acuity when monitoring a new patient. The patient mode determines default alarm limits and internal algorithm settings.
WARNING The monitor may not meet its performance specifications if stored or used outside the specified temperature and humidity ranges.
WARNING Do not connect more than one patient to a monitor. Do not connect more than one monitor to a patient.
WARNING During defibrillation, keep the discharge paddles away from ECG and other electrodes, as well as other conductive parts in contact with the patient.
WARNING Do not operate this product in the presence of flammable anesthetics or other flammable substances in combination with air, oxygen­enriched environments, or nitrous oxide; explosion can result.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI) suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in patient injury and monitor damage.
WARNING Electronic equipment that emits strong electromagnetic or radio frequency signals can cause electrical interference with monitor operation. This interference may distort the ECG signal, thereby preventing accurate rhythm analysis. Avoid operating this device near equipment of this type.
WARNING Exposure to Radio Frequency (RF) radiation. To comply with Federal Communications Commission (FCC) RF exposure requirements, this device shall be used in accordance with the operating conditions and instructions provided in this manual, including the section “Install the Carrying Pouch” on page 28.
WARNING Pacemaker signals can differ from one pacemaker to the next. The Association for Advancement of Medical Instrumentation (AAMI) cautions that “in some devices, rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms. All pacemaker patients should be kept under close or constant observation.” See “ECG Specifications” on page 50 for disclosure of the pacemaker pulse rejection capability of this instrument.
WARNING For patients with a pacemaker, position the monitor to maintain a minimum 6-inch distance between the monitor and pacemaker. Immediately turn the monitor off and provide appropriate patient care if you have any reason to suspect that the monitor is interfering with the pacemaker. The Health Industry Manufacturers Association recommends this minimum 6-inch distance between a hand-held wireless radio and a pacemaker, which is consistent with the independent research by, and recommendations of, Wireless Technology Research.
Directions for Use Chapter 1 General Information 5
WARNING Make frequent electrical and visual checks on cables, sensors, and electrode wires. All cables, sensors, and electrode wires must be inspected and properly maintained and in proper working order to allow the equipment to function properly and protect patient safety.
WARNING Avoid electrosurgery burns at monitoring sites by ensuring proper connection of the electrosurgery return circuit so that the return paths cannot be made through monitoring electrodes and probes.
WARNING Use of ECG cables not specified by Welch Allyn may negate defibrillator protection and risk patient injury due to shock.
WARNING To ensure patient safety, the conductive parts of the ECG electrodes (including associated connectors) and other patient-applied parts should not contact other conductive parts, including earth ground, at any time.
WARNING Motion artifact can affect the accuracy of patient vital sign measurements. Minimize patient motion whenever possible.
WARNING For best product performance and measurement accuracy, use only accessories supplied by Welch Allyn or recommended in the Welch Allyn Products and Accessories booklet (810-0409-XX). Use accessories according to your facility’s standards and the manufacturer’s recommendations. Always refer to the manufacturer’s Directions for Use.
WARNING As with all medical equipment, carefully route the patient cabling to reduce the possibility of patient entanglement or strangulation. Use the supplied garment clips to secure the cable properly.
WARNING When positioning the monitor pouch on the patient, make sure the straps do not entangle the patient’s neck or cause choking. Make sure the straps do not restrict the movement of the patient’s limbs or create a hazard when walking or moving.
WARNING If a product has been dropped or severely abused, send it to a qualified service person to confirm proper operation.
Caution Do not autoclave the monitor. Autoclave accessories only if the manufacturer’s instructions clearly approve it. Many accessories can be severely damaged by autoclaving.
It is possible for the monitor to detect a problem that prevents the monitor from operating properly. If this occurs, the monitor displays an error message and error number. Report such errors to Welch Allyn. The monitor should be serviced only by a Welch Allyn service technician while under warranty. Contact Welch Allyn for information about post-warranty period service.
6 Chapter 1 General Information Welch Allyn Micropaq Monitor

Introducing the Monitor

The monitor is a patient-worn vital signs monitor for use by adult or pediatric ambulatory patients.
One or two ECG channels displayed
Up to 2 ECG leads displayed at the monitor: I, II, III, V, aV
Up to 7 ECG leads displayed at Acuity: I, II, III, V, aV
One ECG lead displayed at the monitor and at Acuity: Fixed lead II with 3-lead cable, or 5-lead cable with only RA, LA and LL electrodes attached.
Pulse oximetry (SpO 404 only)
Two-way wireless communication within Welch Allyn’s FlexNet network
LCD for display of ECG waveforms, SpO and heart rate/pulse rate data, and messages from Acuity
Standalone operation with patient alarms when out of range of the network
Patient alarm limits that can be set at the monitor or at Acuity
Configurable formats for single- or dual-waveform ECG display
Internal antenna
Nurse call key
Lightweight (slightly more than 1 pound with battery)
Rugged and tolerant of brief water exposure
Rechargeable battery
Sleep mode to extend battery life
, aVL, or aVF with 5-lead cable
R
, aVL, or aVF with 5-lead cable
R
) monitoring (Model
2
2
Model 402
Model 404
ECG monitoring
ECG monitoring and either one of two pulse oximetry (SpO2) monitoring options:
•SpO
•SpO
with Masimo® SET® technology
2
with NELLCOR® OxiMax™ technology
2
Directions for Use Chapter 1 General Information 7

Understanding the Monitor and the FlexNet Network

The monitor is intended to operate with an Acuity® Central Station as part of Welch Allyn’s FlexNet network. FlexNet allows multiple devices to communicate through hardwired Ethernet networks and Wireless Local Area Networks (WLANs). The Acuity Central Station provides the primary display and entry of patient data for a patient connected to the monitor.
To Other Acuity
Central Stations
Acuity Central Station
Access Points
H R
S p
O 2
80
97
H R
S p O
2
80
97
H R
S p
O 2
80
97
Micropaq monitors
H R
80
S p
O 2
97
To Other Access Points
FlexNet Network
Each patient-worn monitor supports two-way communication with an Acuity Central Station through an access point in the FlexNet network. The access point is a digital radio transceiver that connects to the FlexNet network. During monitoring, the monitor sends the patient data to Acuity. Acuity and the monitor continuously analyze the data. Acuity provides appropriate alarm or alert messages at the Central Station and other network devices such as a hallway message panel and the monitor itself. Acuity also stores the patient data for viewing or report printing.
If the monitor is moved out of range or loses communication with the FlexNet network and Acuity, it continues to monitor the patient and display patient data. While not communicating with Acuity, the monitor continues to generate local patient alarms or alert messages. Patient data is not stored and Acuity does not perform waveform analysis or generate arrhythmia messages while the monitor is not communicating with Acuity.
When the monitor is returned to within range of the FlexNet network, it automatically reconnects to Acuity.
8 Chapter 1 General Information Welch Allyn Micropaq Monitor

Monitor Features

Controls and Connectors
SpO2 Connector
ECG Connector
Visual Alarm Indicator
Liquid Crystal Display
Select Key and Silence Patient Alarm/ Equipment Alert Key
Scroll Up Key and Reset Patient Alarm Tone Key
Scroll Down Key and Main Menu Key
Nurse Call
Key
Battery pack latch
Battery
Back
Directions for Use Chapter 1 General Information 9
Visual Alarm Indicator
Green Flashes slowly during normal operation.
Red Flashes during patient alarm, remains on continuously when alarms
are silenced or suspended.
Yellow Flashes during an equipment alert or while not connected to the
network.
Remains on continuously if the operator suspends an alert at Acuity for 90 seconds or acknowledges (dismisses) a low battery alert from the monitor or Acuity.
Note
The flashing green LED indicates that the monitor is connected to the network but not necessarily connected to a patient. If the monitor is actively monitoring a patient, the green LED indicates no alarms or alerts are detected.
Audible Alarm Indicator
Beeps to indicate a patient alarm, and beeps faster for life-threatening arrhythmia alarms (see “Patient Alarm and Equipment Alert Specifications” on page 55).
Beeps to indicate when the equipment needs attention. This beep tone is slower than patient alarm tones (see “Patient Alarm and Equipment Alert Specifications” on page 55).
Volume can be configured as high, low, or off (configured at Acuity).
Volume can be configured differently for network connection or stand-alone operation (configured at Acuity).
Nurse Call Key
When connected to Acuity, pressing this key sends a Nurse Call message to Acuity. Depending on how Acuity is configured, this will cause Acuity to:
Generate a Nurse Call Alert at the Acuity Central Station.
Print a “snapshot” of the patient waveform at the Acuity Central Station.
Scroll Up Key and Reset Patient Alarm Tone Key
Scrolls up menus on the display.
Resets a silenced patient alarm tone.
10 Chapter 1 General Information Welch Allyn Micropaq Monitor
Scroll Down Key and Main Menu Key
Scrolls down menus on the display.
Displays the Main Menu.
Select Key and Silence Patient Alarm/Equipment Alert Key
Selects the choice highlighted on the menu.
During patient alarms, silences the tone at the monitor and Acuity (if connected) for 90 seconds. During equipment alerts, silences or acknowledges (dismisses) the alert at the monitor and Acuity.
Battery
Insert the battery to turn on power. Remove the battery to turn off power. (While the battery is removed, the monitor does not perform patient monitoring.)
Recharge the battery while it is removed from the monitor. (See “Recharge a Battery” on page 43.)
To order a new battery, see “Battery Status and Possible Response” on page 44.
Directions for Use Chapter 1 General Information 11
Display
Although the Acuity Central Station is the primary location for viewing patient data, the monitor provides information to support patient care.
Dashed lines indicate the monitor detects a pacemaker
Waveform scale is
selectable
ECG lead is selectable
HR indicates the heart
rate is from ECG.
PR (pulse rate) is
displayed if SpO
and ECG is not (pulse rate
is derived from SpO
If the monitor detects
a vital sign outside
the measurable range,
- - - (below the range)
+ + + (above the range).
Indicates the monitor
is connected to
This symbol indicates the monitor is not communicating with Acuity: Flashing indicates the monitor is associated with an access point, but not communicating with Acuity. Continuous on indicates the monitor is not communicating with an access point or Acuity.
is active
2
it displays:
Acuity.
).
2
signal (display of pacer detection can be enabled or disabled at Acuity)
Indicates one or more patient alarms are disabled (off).
Patient name as entered at Acuity. If the patient name has not been entered, the monitor displays the last four digits of the monitor serial number, such as:
ID:6472
Symbol is displayed at the monitor whenever the Nurse Call key is pressed.
ECG waveform is displayed when active.
SpO2 numeric data is a percentage value.
SpO2 pulse amplitude indicator (not proportional to pulse volume)
Low battery icon flashes to indicate monitor will shut off in 30 minutes or less.
Display Sleep Mode
In order to extend battery life, the display becomes blank after two minutes if no keys are pressed. The display becomes active again if an alarm or alert occurs, a key is pressed, the initial Acuity connection occurs, a cable is inserted, or an electrode is attached.
The display will not become blank if a patient alarm is occurring, an Acuity message is displayed, or the monitor is in Demo mode or Service mode.
12 Chapter 1 General Information Welch Allyn Micropaq Monitor
Main Menu
When you first press , the Main Menu appears:
Press to move through the menu.
Press to select or change the highlighted choice.
EXIT Exit the Main Menu (the menu disappears).
ACUITY... Access the Acuity Menu with network options. The Acuity Menu
is only accessible while connected to Acuity.
EXIT Exit all menus and return to the monitoring
screen.
END TELE Discontinue monitoring a patient.
NEW ROOM Reassign a patient to a new room in the same
unit.
TRANSFER Transfer a patient to a new room in a new unit.
NEW PATIENT Assign the monitor to a new patient.
PATIENT INFO Display patient information such as ID, name,
unit and room.
Whenever the monitor is connected to Acuity and you select ACUITY... from the Main Menu, the monitor displays the message ACUITY CONTACTED to confirm that Acuity has been contacted. The monitor continues to display this message until Acuity responds, or you press to acknowledge the message and clear the screen. If the monitor detects an alarm or alert, it clears the screen to display the appropriate alarm or alert message. The length of time required for Acuity to respond to your selection at the monitor can vary widely depending on the amount of network traffic and other conditions.
ECG LEAD... Access a menu to change the ECG 1 or ECG 2 lead selection (I, II,
III, aV
, aVL, aVF, or V). Available vectors depend on the
R
connected electrodes.
Directions for Use Chapter 1 General Information 13
ECG SCALE... Change the scale of the ECG waveform. If two waveforms are
displayed, both have the same scale.
1 WAVEFORM There are four possible ECG waveform display selections:
1 WAVEFORM the default selection
2 WAVEFORMS
5 SECONDS
FULL SCREEN
Pressing changes to the next selection. This change does not take effect until after you exit the Main Menu. See “Display” on page 11 for descriptions.
LIMITS... Enter the Alarm Limits Menu (“Customize Patient Alarm Limits at
the Monitor” on page 37) and change alarm limits.
SYSTEM INFORMATION
Display information about the network connection and SpO module.
SERVICE MENU Enter Service Mode for a demonstration mode (Demo, see
“Demonstration Mode” on page 16) or service functions for
technicians. Service Mode is not available if any cables are plugged in.
Note
To restrict access to the Main Menu, a Menu Lock option can be configured for the monitor at the Acuity Central Station. When the Menu Lock is enabled, the operator must press and hold down and for two seconds to gain access to the Main Menu. The Menu Lock is disabled if the monitor loses communication with Acuity.
2
14 Chapter 1 General Information Welch Allyn Micropaq Monitor
Waveform Options
There are four ECG waveform options as shown:
1 Waveform
The single ECG 1 (lead II)
waveform is displayed.
5 Seconds
ECG 1 (lead II) cascades from
one line to the other.
The single ECG 1 (lead II) waveform is
allowed to occupy most of the screen.
2 Waveforms
ECG 1 (lead II) and ECG 2
(lead V) are both displayed.
Full Screen
To change the waveform selection during operation:
1. Press to display the Main Menu.
2. Press as needed to highlight the current waveform selection. Then press as needed to select the desired display.
Directions for Use Chapter 1 General Information 15
Messages from Acuity
The monitor displays messages sent from Acuity as needed, including patient alarms and equipment alerts. When Acuity messages are displayed, they temporarily override information displayed on the lower half of the monitor screen.

Accessories

5-lead ECG cable Battery
ECG electrodes Battery charger (8-battery)
3-lead ECG cable (optional) ECG extension cable (optional)
SpO
cable (Masimo or Nellcor) Carrying pouch
2
SpO
sensors (Masimo or Nellcor) Micropaq Directions For Use
2
WARNING For best product performance and measurement accuracy, use only accessories supplied by Welch Allyn or recommended in the Welch Allyn Products and Accessories booklet (810-0409-XX). Use accessories according to your facility’s standards and the manufacturer’s recommendations. Always refer to the manufacturer’s Directions for Use.

Operating Settings

The following monitor operating settings can be set at the monitor or at the Acuity Central Station:
Patient alarm limit settings (ECG and SpO
ECG lead and scale selection
ECG display format
Many other monitor operating settings (such as patient mode and alarms volume) can only be set at the Acuity Central Station. See “Operating Settings” on page 49 for a list of all settings and where they are set.
).
2
16 Chapter 1 General Information Welch Allyn Micropaq Monitor
Default Settings
When the monitor connects to Acuity for a new patient, the Acuity Central Station downloads the appropriate default settings stored at Acuity. While the monitor is connected to Acuity, settings can be changed either at the monitor or at the Acuity Central Station.
If the monitor is temporarily disconnected from Acuity and the operator changes settings at the monitor, those settings are transmitted to and stored at Acuity when the monitor reconnects.

Demonstration Mode

You can practice using the monitor with the Demo mode of operation, including connection to Acuity.
The Demo mode cannot be activated while you are monitoring a patient or if any cables have been plugged into the monitor. During the Demo mode, the monitor and Acuity display the message SIMULATION.
To practice with the monitor in Demo mode:
1. Disconnect all patient cables connected to the monitor.
2. Remove the battery (if installed).
3. Insert the battery and watch for the Power-Up screen.
Power-Up Screen
4. After the Power-Up screen disappears, press to display the Main Menu.
Main Menu
Directions for Use Chapter 1 General Information 17
5. Press to highlight SERVICE MENU, then press to display the Service Menu.
Service Menu
6. Press to highlight DEMO MENU, then press to display the Demo Menu.
7. Press to highlight DEMO 1 or DEMO 2, then press to start.
SIMULATION indicates the Demo mode is active.
A simulated waveform and numerics are displayed.
Demonstration Mode
Demo Mode Display Values and Alarm Limits
Display Demo 1 Demo 2
ECG Waveform Normal sinus rhythm,
normal ST
ECG Heart Rate SpO2 Pulse Rate
Saturation% 97 88 Lower 90
SpO
2
a. Demo 2 will cause patient alarms.
80 125 Lower 50
a
Normal sinus rhythm, normal ST
Alarm Limits (On)
(not applicable)
Upper 120
Upper 100
8. While in Demo mode you can practice changing settings such as ECG lead selection and alarm limit adjustment. (These changes only affect the Demo mode and are erased when you exit the Demo mode.)
18 Chapter 1 General Information Welch Allyn Micropaq Monitor
9. To change to the other Demo selection, press to display the menu, then scroll down to highlight TOG GLE DE MO MOD E and press .
10. To exit the Demo mode, either insert a patient cable or remove and insert the battery. The monitor restarts and enters the normal monitoring mode.
19
2

Monitoring

Connect a New Patient

Connect to the Network
1. Insert a battery into the monitor to turn it on. After a few seconds the monitor Power­Up Screen is replaced by an initial monitoring screen.
The monitor is
searching for a
network
connection.
2. After the network connection is established, the monitor may prompt you to select an Acuity Unit (if your facility has more than one Acuity unit):
If any patient cables are connected, there will be some patient data displayed. The format depends on the monitor default settings.
Example of Initial Monitor Screen
Possible Acuity Unit
selections.
Press to view more.
Example of Acuity Unit Selection
20 Chapter 2 Monitoring Welch Allyn Micropaq Monitor
3. Press or to highlight the desired Acuity unit, then press .
When you press or to highlight the desired Acuity unit and then press , your selection will begin to flash between normal and reverse video to confirm that the monitor is communicating your selection to Acuity. You cannot scroll to another selection during this time. The selection continues to flash until Acuity responds back to the monitor. Then the monitor displays the next appropriate screen (such as a list of possible patients). The length of time required for Acuity to respond to your selection at the monitor can vary widely depending on the amount of network traffic and other conditions.
Be sure to select an Acuity unit. Even though the monitor is connected to the network (as indicated by the green LED and network connection symbol), the Acuity Central Station may not display any indication of this monitor until after you have selected an
Network Connection
Symbol
Acuity unit.
4. The monitor displays a list of possible patients.
If your patient has been pre-admitted to the selected Acuity unit, they will be included in the list.
Possible
patients to
select.
Select Patient at Central
428-02-2392, Hopkins, Bill J 520-29-0319, Phillips, Mary L 532-94-8372, Smith, Frank R
Example of Patient List
SELECT
PATIENT
5. Scroll through the patient list to look for your patient’s name.
If your patient is not in the list, highlight Select Patient at Central and press .
The patient name will need to be entered later at the Acuity Central Station.
WARNING If you do not select the patient name at the monitor at this time, do not adjust any alarm limits until after the patient name and ID are confirmed at Acuity. When the patient name and ID are confirmed at Acuity, Acuity downloads the default settings and patient alarm limits for that Acuity unit to the monitor, thereby overriding any previous settings and alarm limits.
Note
At power-up, the monitor retains the most recent patient mode. The patient mode can only be changed at Acuity. If the patient is being monitored when the patient mode is changed, there is a brief interruption in the display and recording of ECG and SpO
patient data.
2
Directions for Use Chapter 2 Monitoring 21
If your patient is in the list, highlight the name and press . Within a few
seconds the monitor displays a list of unassigned rooms.
Possible
rooms to
select.
If you want to assign the patient to a room, highlight the room and press .
If you do not want to assign a room at this time, highlight Select Room at
Central and press . The patient room will need to be entered later at the monitor (see “Reassign a Monitored Patient to a New Room in the Same Unit” on page 32) or at Acuity (see “Monitor Patient at Acuity” on page 41).
6. If you need to customize alarm limits for your patient, see “Customize Patient Alarm
Limits at the Monitor” on page 37.
Perform ECG Monitoring
WARNING Motion artifact can cause incorrect heart rate readings. Minimize patient motion whenever possible.
WARNING If a disconnected lead is in too close proximity to other electrical devices, it may cause false heart rate readings.
WARNING The monitor does not provide internal arrhythmia analysis. Therefore, arrhythmias may cause the monitor to display inaccurate heart rates.
Select Room at Central
1104A 1104B 1105A
Example of Room List
SELECT
ROOM
WARNING The monitor will show + + + for HR numerics between 301 and 350 beats per minute. Above 350 beats per minute, it may display incorrectly low heart rates, due to intermittent picking of R-waves.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI) suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in patient injury and monitor damage.
WARNING Pacemaker signals can differ from one pacemaker to the next. The Association for Advancement of Medical Instrumentation (AAMI) cautions that “in some devices, rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms. All pacemaker patients should be kept under close or constant observation.” See “ECG Specifications” on page 50 for disclosure of the pacemaker pulse rejection capability of this instrument.
22 Chapter 2 Monitoring Welch Allyn Micropaq Monitor
WARNING For patients with a pacemaker, position the monitor to maintain a minimum 6-inch distance between the monitor and pacemaker. Immediately turn the monitor off and provide appropriate patient care if you have any reason to suspect that the monitor is interfering with the pacemaker. The Health Industry Manufacturers Association recommends this minimum 6-inch distance between a hand-held wireless radio and a pacemaker, which is consistent with the independent research by, and recommendations of, Wireless Technology Research.
WARNING High-intensity radio frequency (RF) energy from external sources, such as an improperly connected electrosurgical unit, can induce heat into electrodes and cables which can cause burns on the patient. Reading errors and damage to equipment may also result. This hazard can be reduced by (1) avoiding the use of small ECG electrodes, (2) selecting ECG electrode attachment points remote from the surgical site and from the electrosurgical return electrode, (3) using electrosurgical return electrodes with the largest practical contact area, and (4) assuring proper application of the electrosurgical return electrode to the patient.
WARNING Verify patient mode at Acuity. Incorrect patient mode may result in inaccurate heart rates and inappropriate alarm settings.
WARNING To help prevent injury, use the provided garment clips to route the ECG cables away from the patient’s head.
WARNING Use of ECG cables with loose or faulty detachable lead wires may cause erratic behavior of the ECG waveform due to intermittent ECG lead wire connections.
WARNING To ensure patient safety, the conductive parts of the ECG electrodes (including associated connectors) and other patient-applied parts should not contact other conductive parts, including earth ground, at any time.
Caution To protect the monitor from damage during defibrillation, for accurate ECG information, and for protection against noise and other interference, use only ECG electrodes and cables specified or supplied by Welch Allyn (these cables have the required current-limiting resistors). Follow recommended application procedures.
Caution Do not use an ECG cable longer than 10 feet (3 meters). If the nominal length of the ECG cable, including extensions, exceeds this length, the monitor is not guaranteed to meet published electromagnetic compatibility (EMC) performance specifications.
Even though the monitor contains fully isolated patient-connected circuitry, it has not been specially designed for direct application on a patient’s heart.
Use only with accessories provided or recommended in the Welch Allyn Products and Accessories booklet (810-0409-XX).
Severe artifact and interference (such as defibrillation interference) can cause the waveform to move off the display for a few seconds before it is restored.
Directions for Use Chapter 2 Monitoring 23
Perform 5-Lead ECG Monitoring
1. Inspect the ECG cable and replace it if it shows any signs of wear, breakage, or fraying. Plug the cable into the monitor.
2. Select electrode sites on the patient.
Choose flat areas; avoid fatty or bony areas and major muscles.
3. Shave or clip hair from electrode sites, thoroughly clean skin, and lightly rub dry.
You may use soap and water, isopropyl alcohol or special skin preparation pads. To avoid allergic reactions to electrodes, refer to the electrode manufacturer’s directions.
4. If you are using pre-gelled electrodes, make sure the electrode date is not expired and the gel is intact and not dried out. For best results, use only silver/silver chloride electrodes.
If you are using ungelled electrodes, apply a 1/4- to 1/2-inch mound of gel over the electrode contact area.
For best product performance and measurement accuracy, do not use stainless steel needle electrodes, squeeze bulb electrodes, or electrodes with dissimilar metals. Due to polarization, such electrodes can generate offsets beyond the monitor’s capabilities. Do not use electrodes from more than one manufacturer on the same patient.
5. Attach lead wires to the electrodes before applying them to the patient. Apply the electrodes to the patient in the proper locations.
Flashing circle indicates the lead is not connected.
Left ArmRight Arm
Right Leg
Left Leg
If the monitor detects that some lead wires are not properly connected, the monitor displays a chest diagram and indicates which leads are disconnected.
The locations of the circles displayed on the monitor for each lead are fixed, and are not affected by the exact placement of the electrodes on the patient. For example, the C lead can be placed on the patient in any one of the V1-V6 locations desired, but will only be displayed on the monitor in the location shown above.
24 Chapter 2 Monitoring Welch Allyn Micropaq Monitor
6. After leads are properly connected, confirm that the monitor displays the ECG waveform, heart rate, and other patient data.
To change the ECG lead selection, press to display the Main Menu. Then press Scroll Down to highlight ECG LEAD . . . , then highlight ECG 1 or ECG 2 and press
to change the lead.
3-Lead ECG Application with the 5-Lead ECG Cable
Note
You can perform 3-lead ECG monitoring in a similar manner as 5-lead ECG monitoring. You may use the 5-lead ECG cable with detachable electrode lead wires, and connect only the lead wires and electrodes for RA, LA, and LL. Refer to the Welch Allyn Product and Accessories booklet (810-0409-XX) for part numbers.
Follow these steps:
1. P e r f o r m Step 1 through Step 4 on page 23 as described for 5-lead ECG monitoring.
2. Before attaching electrodes to the patient, attach only lead wires for RA, LA, and LL to
3. Apply the electrodes for RA, LA, and LL to the patient in the proper locations.
Be aware that there are some inherent limitations with this application, especially when compared to 5-lead ECG monitoring. These limitations include the restriction to only one displayed lead, ECG lead II. Because only one displayed lead is available (ECG lead II), factors such as a poor electrode connection at RA, LA, or LL can significantly affect performance. To overcome these limitations, the 5-lead ECG monitoring is preferred.
The monitor’s 3-lead ECG monitoring is only available for use with Acuity software versions 6.1 or later.
the 5-lead ECG trunk cable and to the electrodes. Make sure that lead wires for C and RL are DETACHED from the 5-lead ECG trunk cable.
The monitor displays the chest diagram with two circles blinking confirming that the C and RL electrodes are not connected.
4. Observe the monitor and visually confirm that within about 30 seconds, the two circles disappear and the monitor displays the ECG waveform, heart rate, and other patient data.
Be aware that if you connect the C or RL lead wires to the 5-lead ECG trunk cable and apply the C or RL electrodes to the patient, the monitor defaults to 5-lead ECG monitoring and does not enable 3-lead ECG monitoring. To enable 3-lead ECG monitoring, you must disconnect the ECG cable from the monitor for a few seconds, and then begin this procedure again.
Directions for Use Chapter 2 Monitoring 25
Be aware that only ECG lead II is available for display with the monitor’s 3-lead ECG monitoring. No other ECG lead selections are available.
WARNING Do not try to perform this 3-lead ECG monitoring with any 5-lead ECG cable that does not have detachable electrode lead wires as described above. Attempting to perform this procedure with a 5-lead ECG cable which has lead wires cut off or hanging loose and not connected to the patient would present a shock hazard to the patient or clinician.
3-Lead ECG Application with the 3-Lead ECG Cable
Note
Refer to the Welch Allyn Product and Accessories booklet (810-0409-XX) for part numbers.
Follow these steps:
1. P e r f o r m Step 1 through Step 4 on page 23 as described for 5-lead ECG monitoring.
2. Attach lead wires to the electrodes before applying them to a patient.
3. Apply the electrodes for RA, LA, and LL to the patient at the proper locations. If the
4. Observe the monitor and visually confirm it displays the ECG waveform, heart rate,
Be aware there are some inherent limitations with this application, especially when compared to 5-lead ECG monitoring. These limitations include the restriction to only one displayed lead, ECG II lead. Because only one displayed lead is available (ECG lead II), factors such as poor electrode connection at RA, LA, or LL can significantly affect performance. To overcome these limitations, the 5-lead ECG monitoring is preferred.
The monitor’s 3-lead ECG monitoring is only available for use with Acuity software versions 6.1 or later.
monitor detects one of the lead wires is not properly connected, it will display a chest diagram indicating which lead is disconnected.
and other patient data.
Be aware that only ECG lead II is available for display with the monitor’s 3-lead monitoring. No other ECG lead selections are available. The monitor will not detect the presence of a 3-lead cable until two or more of its leads are connected to the patient.
3-Lead ECG Application with the 3-Lead ECG Cable and Cable Extension
This combination functions the same way as the 3-lead ECG application with the 5-lead cable. For electromagnetic compatibility (EMC) reasons, do not use an ECG cable and extension cable length of more than approximately 10 feet total.
26 Chapter 2 Monitoring Welch Allyn Micropaq Monitor
Perform SpO2 Monitoring
WARNING Oxygen saturation measurements using pulse oximetry are highly dependent on proper placement of the sensor and patient conditions. Patient conditions such as shivering and smoke inhalation may result in erroneous oxygen saturation readings. If pulse oximetry measurements are suspect, verify the reading using another clinically accepted measurement method, such as arterial blood gas measurements on a co-oximeter.
WARNING Use only accessories as listed in the Welch Allyn Products and Accessories booklet (810-0409-XX). Use only Masimo accessories and sensors
with the Masimo SpO the Nellcor SpO
WARNING Tissue damage can be caused by incorrect application or use of a sensor (e.g., wrapping the sensor too tightly, applying supplemental tape, failing to periodically inspect the sensor site, leaving a sensor on too long in one place). Refer to the Directions for Use provided with each sensor for specific instructions on application and use, and for description, warnings, cautions, and specifications.
WARNING Sensors exposed to ambient light while not applied to a patient can exhibit semi-normal saturation readings. Be sure the sensor is securely placed on the patient and check its application often to ensure accurate readings.
2
option. Use only Nellcor accessories and sensors with
2
option.
WARNING Inaccurate measurements may be caused by venous pulsations.
WARNING The pulse oximeter can be used during defibrillation, but the
readings may be inaccurate for a short time.
WARNING The pulse oximeter should NOT be used as an apnea monitor.
WARNING A very sudden and substantial change in pulse rate can result in
erroneous pulse rate readings. Be sure to validate the patient data and patient condition before intervention or change in patient care.
WARNING Interfering Substances: Carboxyhemoglobin may erroneously increase readings; the level of increase is approximately equal to the amount of carboxyhemoglobin present. Methemoglobin may also cause erroneous readings. Dyes, or any substances containing dyes, that change usual arterial pigmentation may cause erroneous readings.
Directions for Use Chapter 2 Monitoring 27
1. Attach the SpO2 sensor to the patient according to the manufacturer’s directions for use, observing all warnings and cautions.
Each SpO
sensor is designed for application to a specific site on the patient within a
2
certain size range. To obtain optimal performance, use an appropriate sensor and apply it as described in the sensor’s directions for use.
If excessive ambient light is present, cover the sensor site with opaque material to block the light. Failure to do so may result in inaccurate measurements. Light sources that can affect performance include surgical lights (especially those with a xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight.
If NIBP will be monitored while using SpO than the SpO
sensor to help reduce unnecessary SpO2 alarms. For optimal
2
measurements, avoid placing the SpO
, place the NIBP cuff on a different limb
2
sensor on the same limb as an arterial
2
catheter or intravascular line.
Loss of pulse signal can occur if the sensor is too tight, there is excessive ambient light, an NIBP cuff is inflated on the same limb as the sensor, there is arterial occlusion proximal to the sensor, the patient is in cardiac arrest or shock, or the patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
2. Inspect the SpO
cable. Replace it if it shows any signs of wear, breakage, or fraying.
2
Plug the cable into the sensor and the monitor.
3. After the cable is connected, confirm that the monitor displays SpO
data within a
2
few seconds.
4. If excessive patient movement interferes with measurements, consider the following possible solutions:
be sure the sensor is secure and properly applied
use a new sensor with fresh adhesive backing
select a different type of sensor
move the sensor to a less active site
The SpO
system is designed to work satisfactorily during normal patient motion.
2
28 Chapter 2 Monitoring Welch Allyn Micropaq Monitor
Install the Carrying Pouch
WARNING As with all medical equipment, carefully route the patient cabling to reduce the possibility of patient entanglement or strangulation. Use the supplied garment clips to secure the cable properly.
WARNING When positioning the monitor pouch on the patient, make sure the straps do not entangle the patient’s neck or cause choking. Make sure the straps do not restrict the movement of the patient’s limbs or create a hazard when walking or moving.
Adult Carrying Pouch
The Adult Carrying Pouch is intended for ambulatory adult patients. It is not intended for use while the patient is in bed.
1. Put the carrying pouch on the patient and insert the monitor.
2. Carefully arrange the pouch and monitor on the patient to avoid bruising or other skin injuries.
To maximize the monitor’s wireless transmission range, always make sure that the monitor display is facing out and away from the patient’s body.
Pediatric Carrying Pouch
The Pediatric Carrying Pouch is intended for ambulatory pediatric patients (40 to 80 lbs.). It is not intended for use while the patient is in bed.
1. Insert the monitor into the pouch.
2. Carefully arrange the pouch and the monitor on the patient to avoid bruising or other skin injuries.
To maximize the monitor’s wireless transmission range, always make sure that the monitor display is facing out and away from the patient’s body.
Directions for Use Chapter 2 Monitoring 29

Monitor a Patient Out of Range of Acuity

While out of range of Acuity, the monitor continues to monitor the patient and provide local HR/PR and SpO2 alarms or alerts at the monitor as needed.
When the patient wearing the monitor goes out of range of Acuity, do the following:
1. A DROPOUT equipment alert occurs at the Acuity Central Station. Acknowledge the alert at Acuity.
2. An equipment alert occurs at the monitor with this message:
ACUITY CONNECTION LOST
Depending on how the monitor is configured (as controlled by Acuity), this alert can also cause the monitor to emit audible alert tones.
If tones are enabled, the authorized person should press on the monitor to acknowledge (dismiss) the alert and silence this instance of the alert tone.
Note
When the patient returns within range of Acuity, the monitor automatically reconnects to Acuity. No clinician intervention is required.
The person authorized to press to acknowledge the alert may vary, depending on the local protocol. Follow the protocol established by your institution.
30 Chapter 2 Monitoring Welch Allyn Micropaq Monitor

Stop Monitoring a Patient

If you want to discontinue monitoring the patient, follow these steps.
1. Press to display the Main Menu.
2. Press to highlight ACUITY, then press .
Acuity Menu
3. Press to highlight END TELE, then press .
4. When the monitor displays the message SAFE TO REMOVE BATTERY, remove the battery.
If the battery is not removed within 30 seconds, the monitor will automatically try to reconnect to the network.
5. Disconnect the leads and sensors from the patient.
Note
If you do not use END TELE to disconnect from the network as described above, the Acuity Central Station generates a DROPOUT equipment alert at Acuity.
If you want to monitor this same patient at a later time, you will need to reselect the patient name from the monitor or confirm the patient ID at Acuity.
Directions for Use Chapter 2 Monitoring 31

Reconnect a Recently Monitored Patient

1. Insert a battery into the monitor to turn on the monitor. Confirm that after a few seconds the monitor Power-Up Screen is replaced by the initial monitoring screen.
2. The monitor will then present a series of menus and messages requesting you to provide information about the connection and patient. The actual screens presented depend on how long the patient has been disconnected. Provide the information as requested. This may include:
Select an Acuity unit.
Select a patient from the patient list.
Select a patient room from the room list.
Note
If you do not select the patient name or room while connecting the patient, you will need to do that later at the Acuity Central Station. See “Monitor Patient at
Acuity” on page 41 for more information.
To perform ECG monitoring, see “Perform ECG Monitoring” on page 21.
To perform SpO
monitoring, see “Perform SpO2 Monitoring” on page 26.
2
32 Chapter 2 Monitoring Welch Allyn Micropaq Monitor

Reassign a Monitored Patient to a New Room in the Same Unit

If a patient is being monitored and you want to assign them to a new room in the same unit, follow these steps.
1. Press to display the Main Menu.
2. Press again to highlight ACUITY and press to display the Acuity Menu screen.
3. Press to highlight NEW ROOM, then press .
.
New Room Selection
Within a few seconds the monitor displays a list of all available rooms, including the patient’s current room.
If you decide not to change the patient’s current room assignment, press (the
patient’s current room is the default selection in the list).
To assign the patient to a new room, highlight the room and press .
If you want to cancel the patient’s current room assignment, but do not want to
assign a new room at this time, you can highlight Select Room at Central and press . You can then assign the room later from the Acuity Central Station, or you can repeat this procedure and assign a new room from the monitor.
Directions for Use Chapter 2 Monitoring 33

Transfer a Monitored Patient to a New Room in a Different Unit

If a patient is being monitored and you want to assign them to a new room in a different unit, follow these steps.
1. Press to display the Main Menu.
2. Press again to highlight ACUITY and press to display the Acuity Menu screen.
3. Press to highlight TRANSFER, then press .
Transfer a Patient
Within a few seconds the monitor displays a list of units.
4. Press to highlight the new unit, then press .
The patient is not monitored at Acuity during the short time required by Acuity to process the transfer to the new unit (typically less than one minute). However, the patient continues to be monitored by the monitor.
(If the selected unit is currently not available, the monitor displays an appropriate message; press to acknowledge the message and cancel the transfer.)
5. After the patient is assigned to the new unit, the monitor displays a list of unassigned rooms. (The patient’s previous unit and room assignment is cancelled.)
To assign the patient to a new room, highlight the room and press .
If you decide not to assign the patient to a new room at this time, you can
highlight Select Room at Central and press . You can then assign the room later from the Acuity Central Station, or you can assign a new room from the monitor later using the procedure on “Reassign a Monitored Patient to a New
Room in the Same Unit” on page 32.
34 Chapter 2 Monitoring Welch Allyn Micropaq Monitor

Reassign the Monitor to a New Patient

If you want to discontinue monitoring a patient and reconnect the monitor to a new patient, follow these steps.
1. Press to display the Main Menu.
2. Press again to highlight ACUITY and press to display the Acuity Menu screen.
3. Press to highlight NEW PATIENT, then press .
Select a New Patient
The monitor then presents a series of menus and messages requesting you to provide information about the connection and patient. The actual screens presented depend on how the Acuity System is configured.
Provide the information as requested. This may include:
Select an Acuity unit.
Select a patient from the patient list. (After you select a new patient, all monitor operating settings are reset to the Acuity System default power-up settings.)
Select a patient room from the room list.
If you do not select the patient name or room while connecting the patient, you will need to do that later at the Acuity Central Station. See “Monitor Patient at Acuity” on page 41 for more information.
To perform ECG monitoring, see “Perform ECG Monitoring” on page 21.
To perform SpO
monitoring, see “Perform SpO2 Monitoring” on page 26.
2
35
3

Alarms & Alerts

About Alarms and Alerts

Alarms provide a warning about a patient condition (such as a vital sign limit violation).
Alerts provide a warning about an equipment condition that needs attention (such as a
low battery or detached ECG lead).
Alarms and alerts may be detected either by the monitor or by the network. While connected to the network, alarms or alerts are displayed at the monitor and at the Acuity Central Station. Alarms have a higher priority than alerts.

Alarm Holdoffs

To help minimize false alarms, the monitor briefly delays or “holds off” triggering alarms for limit violations for HR/PR or SpO detects that the patient’s vital sign has returned to acceptable limits, the monitor cancels the alarm holdoff. The next time a vital sign limit is violated, the monitor starts a new holdoff period.
. After the alarm holdoff period begins, if the monitor
2
Vital Sign Alarm Holdoff Period
HR 3 seconds
% SpO2 or PR 10 seconds
36 Chapter 3 Alarms & Alerts Welch Allyn Micropaq Monitor

Respond to a Patient Alarm at Monitor

When a patient alarm occurs, the monitor produces an audible tone (if audible tones are enabled). Life-threatening arrhythmia alarms beep at a faster pace than other vital sign alarms (see “Patient Alarm and Equipment Alert Specifications” on page 55). The monitor also displays a message similar to the following:
Red LED flashes
during alarm.
After the alarm is
silenced (or
suspended at
Acuity), the red
LED is on
continuously.
Flashing numerics
indicate patient
alarm.
Press to silence the alarm tone.
1. Check the patient and provide appropriate care.
2. To silence the alarm tone at the monitor and the Acuity Central Station for 90 seconds, press .
While the alarm tone is silenced, visual alarm indications continue, and the red alarm indicator on the monitor changes from a flashing display to a continuous display.
If the alarm condition still exists after 90 seconds, the alarm tone resumes.
Note
If you silence an alarm at the monitor and another patient alarm or an equipment alert occurs during the silence period, the tone resumes at the monitor. At Acuity, only life-threatening arrhythmia alarms interrupt the silence period.
If you suspend an alarm at Acuity, only life-threatening arrhythmia alarms interrupt the silence period at the monitor and Acuity.
To access the Main Menu during silencing, press .
Directions for Use Chapter 3 Alarms & Alerts 37
3. To reset the alarm tone at the monitor and Acuity before the 90 seconds has elapsed, press at the monitor, or press Resume at the Acuity Central Station.
4. After caring for the patient, make sure that the appropriate alarm limits are set and that alarms are on.

Customize Patient Alarm Limits at the Monitor

WARNING If the patient’s name has not yet been assigned to the monitor, do not adjust any alarm limits until after the patient name and ID are confirmed at Acuity. When the patient name and ID are confirmed at Acuity, Acuity downloads the default settings and patient alarm limits for that Acuity unit, thereby overriding any custom alarm limits that were set at the monitor before selecting the patient.
1. Press to display the Main Menu.
2. Press to highlight LIMITS, then press to display the Alarm Limits Adjust Menu:
Alarm limits can be adjusted
for HR/PR and SpO
3. To change an alarm limit, press or to highlight the desired limit, then press
to display the Threshold Adjustment Menu:
Select + or - to change the limit.
Scroll and select the + or - selections to change the limit as desired.
To turn the highlighted limit on or off, scroll to ON/OFF and press .
.
2
38 Chapter 3 Alarms & Alerts Welch Allyn Micropaq Monitor
WARNING If you turn off any alarm limits, be sure to restore the appropriate alarm limits before you resume monitoring. Only life-threatening arrhythmias will be indicated at the monitor and Acuity (if connected) when alarms are turned off.
4. To change other limits, scroll to PREVIOUS MENU, press , then select another limit to change.
5. When you have completed all changes, scroll to PREVIOUS MENU, then EXIT on the Alarm Limits Adjust Menu and press to return to the normal monitoring screen.
Note
While the monitor is connected to Acuity, settings can be changed either at the monitor or at Acuity.

Respond to an Equipment Alert at the Monitor

When the network or the monitor detects an equipment problem, the monitor produces a an audible alert tone (if audible tones are enabled). Equipment alerts beep at a slower pace than patient vital sign alarms (see “Patient Alarm and Equipment Alert
Specifications” on page 55).
The monitor also displays a flashing yellow light (LED) and an equipment alert message similar to the following:
The monitor has detected that the ECG cable has been disconnected from the monitor.
1. In this instance, press to acknowledge (dismiss) the alert tone and clear the message.
If the message says “PRESS TO SILENCE,” when you press , the tone is silenced for 90 seconds instead of dismissed.
If the monitor displays a chest diagram with a missing lead flashing, you can press
to silence the tone.
Some alerts do not give you the option to acknowledge the alert or silence the tone. For these alerts, to remove the message and tone, you must correct the problem.
Directions for Use Chapter 3 Alarms & Alerts 39
2. If possible, determine what caused the problem and correct it.
Note
After you press to acknowledge or silence some types of alerts, the yellow LED remains on (either flashing or steady yellow) until you correct the problem.
For low battery alerts and no Acuity connection alerts, specific icons also flash.
For a list of possible messages and suggested responses, see “Alert Messages and
Display Information” on page 40.
40 Chapter 3 Alarms & Alerts Welch Allyn Micropaq Monitor

Alert Messages and Display Information

Message and Display
Possible Cause(s) and Suggested Response
Information
LOW BATTERY The monitor will shut down within approximately 30 minutes or less due to a low
battery.
• Replace the battery as soon as possible.
VERY LOW BATTERY The monitor will shut down within approximately 5 minutes or less due to a low
battery.
• Replace the battery as soon as possible.
BATTERY TOO LOW SHUT DOWN IN PROGRESS
The battery is so low the monitor has to shut down operation.
• Replace the battery immediately.
ACUITY CONNECTION LOST The monitor is not connected to the network.
• Press to acknowledge and silence the tone and cancel the message. While disconnected from the network, the off-network icon and the yellow LED continue to flash. NOTE: The monitor will continue to attempt to reconnect until it is successful.
EXCESSIVE ECG OFFSET REPLACE ELECTRODES
a
Chest icon is displayed with flashing ECG electrode(s).
The monitor detects poor ECG electrode contact.
• Check and replace ECG electrodes as needed.
The monitor detects that one or more ECG electrodes are disconnected.
a
• Check and replace or reconnect electrodes as needed.
NO ECG CABLE DETECTED • If the ECG cable has been intentionally removed from the monitor, press to
cancel the alert tone.
• If the ECG cable has been unintentionally removed, plug it back into the monitor. Check the patient and monitor to make sure ECG monitoring resumes properly.
• It is normal for this alert to appear with a 3-lead ECG cable when two or more of its leads are disconnected from the patient. Reconnect the disconnected lead wires.
NO SPO2 SENSOR DETECTED The SpO
• If disconnection is intentional, press to acknowledge and silence the tone.
sensor has been disconnected for more than 5 seconds.
2
• If disconnection is not intentional, reconnect the sensor or replace the sensor and reconnect.
a
DEFECTIVE SPO2 SENSOR
or
UNRECOGNIZED SPO2 SENSOR
<key name> KEY STUCK
a
The SpO2 sensor is either defective or not recognized.
• Replace the SpO
a
sensor with a new, compatible SpO2 sensor.
2
During the power-up self test, the monitor detected that a key is stuck ( , , , or ). This can happen if you accidentally press a key down before the Main Menu is displayed during power-up.
• Remove and then reinsert the battery to power up again and see if the key is still stuck. If it is, contact your biomedical engineering department.
System Error Thread: <nnn>
The equipment problem is so serious the monitor cannot be used.
• Contact your biomedical engineering department.
Error ID: <nnn>
a. This alert message can be acknowledged from Acuity, but not from the monitor.
41
4

Monitor Patient at Acuity

While the Micropaq is connected to the FlexNet network, patient data gathered by the monitor is continuously stored at Acuity. You can access this patient information at the Acuity Central Station and perform administrative functions, including:
Admit (and discharge) a patient in the Acuity unit.
Edit the patient description (name, physician, etc.).
Review and print patient data such as trends and waveforms.
Suspend patient alarm tones for 90 seconds and resume the alarm tones
For more information about using the Acuity Central Workstation, refer to:
Acuity Directions For Use
42 Chapter 4 Monitor Patient at Acuity Welch Allyn Micropaq Monitor
43
5

Maintenance

This section provides information to help operators of the monitor and the battery charger perform routine maintenance activities such as changing or recharging batteries, inspection, and cleaning.

Change the Battery

1. Remove the depleted battery.
2. Insert a fully-charged battery. Use only batteries supplied by Welch Allyn.
WARNING Before installing a battery, carefully inspect the battery case. If there are any signs of damage, cracks, or leaks, discard the battery properly and do not use it.
Note
The Acuity unit can be configured to allow you a short time (typically 30 seconds or more) to change the monitor battery while the monitor is connected to the network without causing an Acuity equipment alert. If the monitor is connected to the network and the battery is removed for longer than the allowable battery changing time, Acuity generates a DROPOUT equipment alert at the Acuity Central Station.

Recharge a Battery

1. On the monitor battery charger, choose an empty battery well where the LED is off.
2. Insert the depleted battery into the battery well.
3. Confirm that the charger displays a flashing green LED by the battery to indicate the battery is detected or is charging.
4. When the green LED is on continuously, the battery is fully charged. Remove the battery.
If the yellow LED is on continuously, the battery may have reached the end of its useful life. Refer to the table below for suggested responses.
44 Chapter 5 Maintenance Welch Allyn Micropaq Monitor
Charger LED Battery Charger
Label - LEDs
Green LED on continuously
Green LED flashing
Green LED flashing very slowly
LED off No battery is detected.
Yellow LED on continuously
Battery Status and Possible Response
Battery is fully charged.
Battery is charging.
Battery is detected and waiting to be charged.
Something is wrong with the battery or the charger. Remove the battery.
• If the LED goes off, it is probably a battery problem. Insert a new battery into the same battery well. If the new battery charges correctly, then the battery has a problem; discard the battery. The battery reorder number is 008-0647-XX. If the same problem occurs with the new battery, the charger may need repair. Contact biomedical engineering.
• If the LED does not go off when you remove the battery, it is probably a charger problem. Unplug the charger power cord, wait at least 5 seconds, then plug in the charger power cord again. Insert a new battery into the same battery well. If the new battery charges correctly, then the battery has a problem; discard the battery. If the same problem occurs with the new battery, the charger may need repair. Contact biomedical engineering.
The charger can accommodate up to eight batteries. The charger charges a maximum of four batteries at a time. After a battery begins recharging (as indicated by the green LED that flashes on one second, off one second), it is typically fully recharged within four hours at room temperature. After a battery is fully charged, the charger continues to maintain the full charge on the battery until the battery is removed. Leaving a fully-charged battery in the charger will not harm the battery.
Remove batteries from the battery charger if the battery charger will be disconnected from ac power for more than a few days. Do not block the cooling vents at the rear of the battery charger.
The monitor battery charger only charges four batteries at a time. A battery is not fully charged until the green LED is on continuously. Do not remove a battery until it is fully charged.
WARNING The monitor battery is Lithium Ion. Do not incinerate, submerge, crush, disassemble, or autoclave. If a battery has been submerged in liquid, discard the battery properly; do not try to recharge or reuse the battery. Do not short the battery terminals. Do not try to connect the battery to any device except the monitor or the monitor battery charger. Do not expose to high temperature (above 60° C or 140° F). Use only the specified monitor battery charger.
Directions for Use Chapter 5 Maintenance 45

Inspect and Clean the Monitor and Accessories

WARNING Be sure to unplug the monitor battery charger power cord from the electrical power outlet before inspecting or cleaning the battery charger. Exposing the battery charger to liquids such as cleaning solutions while connected to electrical power could result in electrical shock or fire.
WARNING Do not autoclave the monitor, battery, battery charger, or accessories. Do not immerse the monitor, battery, or battery charger in liquid when cleaning. Do not immerse accessories in liquid when cleaning unless the accessory manufacturer’s cleaning instructions explicitly instruct you to do so.
Before cleaning, thoroughly inspect the monitor and all accessories for any signs of damage, cracks, or improper mechanical function of keys or connectors. While gently bending and flexing cables, inspect for damage, cracks, cuts, abrasions, extreme wear, exposed wires or bent connectors. Confirm connectors securely engage. Report damage or improper function to your service department. At least every 12 months, be sure to thoroughly inspect the battery charger case and power cord for damage or extreme wear.
To clean the monitor, batteries, or battery charger:
1. Wipe the equipment with a nearly dry cloth moistened with one of the approved cleaning solutions listed in the table on “Approved Cleaning Solutions” on page 46.
2. Thoroughly wipe off any excess cleaning solution. Do not let the cleaning solution run into or accumulate in connector openings, latches, or crevices. If liquid gets into a battery well or connectors, dry the area with warm air, then check the equipment to confirm that it operates properly.
Caution Use only cleaning solutions which are recommended by Welch Allyn for this equipment. Use of solutions which are not recommended or which have a high acid content or are otherwise inappropriate can cause damage to the equipment, including cracking and deterioration of the plastic case. Do not use these solutions or similar products: Butyl alcohol, Denatured ethanol, Freon Mild chlorine bleach solution, Isopropyl alcohol (except for the SpO Trichloroethane, Trichloroethylene, Acetone, Vesphene II, Enviroquat Staphene
®
, Misty®, Glutaraldehyde, Formula 409®, or Fantastik®.
cable),
2
®
,
,
46 Chapter 5 Maintenance Welch Allyn Micropaq Monitor
Equipment Cleaning Instructions Approved Cleaning Solutions
a
Monitor
a
Battery Battery Charger
• Wipe with a nearly-dry cloth moistened with
a
cleaning solution.
• Thoroughly wipe off any excess cleaning solution. Do not let cleaning solution run into connector openings or crevices.
Warm water, Liquid soap, Coverage®,
®
Windex
, Ovation®, Hydrogen peroxide
solution, Wex-cide
b
c
®
, T.B.Q.
c
®
ECG cable, extension cable
cable,
SpO
2
extension cable
Other accessories • Consult manufacturer’s instructions. Consult manufacturer’s instructions.
a. The equipment may be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood and other
body fluids. (Federal OSHA Standard on bloodborne pathogens: 29 CFR 1910.1030, 12/6/91.) b. If liquid gets into the battery well or connectors, dry the area with warm air, then check the monitoring functions for proper operation. c. Wex-cide (Wexford Labs, Inc., Kirkwood, MO) and T.B.Q. (Calgon Vestal Lab., Calgon Corp., St. Louis, MO) are disinfectants that meet
OSHA requirements, are EPA approved, and will not harm the outside of the monitor, battery, or battery charger. Wipe away
disinfectants with a water-dampened cloth after the manufacturer’s recommended period.
• Wipe gently with damp cloth moistened with a mild detergent solution.
• Thoroughly wipe off any cleaning solution.
• Wipe the cable with a 70% isopropyl alcohol pad and allow it to dry.

Recycling Monitor Components

When the battery, monitor, or battery charger reaches the end of its life, recycle it locally according to national, state, and local regulations. You can also return the battery, monitor, or charger to Welch Allyn for recycling.
Li++
Within the European Union
Do not dispose of this product as “unsorted municipal waste.” Prepare it for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE).
Mild detergent.
70% isopropyl alcohol pad.
If the monitor or battery (Li++) is contaminated, this directive does not apply. For more specific information, see www.welchallyn.com/weee, or contact Welch Allyn Customer Service.
Recycle monitor batteries (Li++) according to the Directive 91/157/EEC (Batteries and accumulators containing certain dangerous substances) and Directive 93/86/EEC (Labeling of batteries and accumulators containing
Li++
certain dangerous substances).
Directions for Use Chapter 5 Maintenance 47

Change the Network Name

This procedure allows you to change the network name assigned to the monitor (as long as the current network name is one of the pre-set names available in the monitor Network Name Menu).
Note
To change the network name:
1. P r e s s to access the Main Menu, then repeatedly press until SERVICE MENU
2. Press to display the Service Menu screen.
3. Press and hold and , then press to display the Network Name Menu.
Changing the monitor network name will cause the monitor to re-start and seek to connect with the FlexNet network corresponding to the new name. Do not attempt to change the network name unless you are a qualified biomedical service engineer or technician.
is highlighted.
If the current monitor network name is one of the following pre-set names:
com.protocol demo.protocol
com1.protocol com2.protocol
com3.protocol com4.protocol
com5.protocol com6.protocol
com7.protocol com8.protocol
then the monitor displays the following screen
48 Chapter 5 Maintenance Welch Allyn Micropaq Monitor
To change the network name, make sure YES is highlighted, then press to display the following screen:
Press or to highlight the desired network name, then press . The monitor automatically turns itself off, then turns on and seeks to connect to a FlexNet network with the new network name.
If the current network name is a custom name, the monitor displays the following screen:
You cannot change the network name using the Network Name Menu. Press to return to the Service Menu. Contact Welch Allyn Technical Support for assistance.
49
6

Reference

Operating Settings

The following table lists all of the monitor settings and the default settings.
Parameter Set at
Patient Mode (Adult [age 13 and older] or Pediatric [age 12 and younger])
ECG screen mode (Single, Dual, 5 Sec, or Full Screen)
ECG 1 Lead Selection Yes Yes Yes Yes II ECG 2 Lead Selection Yes Yes Yes Yes V (or III if no V lead) ECG Size (Scale) Yes No No Yes 1 mV/cm Language No No Yes Yes English Mains Filter (off, 50, or 60 Hz) No No Yes Yes 60 Vital Signs Alarm Volume (high, low, or off)
With Acuity Connection Without Acuity ConnectionNoNo
Equipment Alert Volume (high, low, or off)
With Acuity Connection
Without Acuity ConnectionNoNo HR/PR Alarm Limits (Lower, Upper) SpO
Alarm Limits
2
(Lower, Upper) Regulatory settings (U.S., Europe, Japan) Pacer Detection Enable No Yes No Yes On Menu Lockout No No Yes No Off Display Backlight Timeout No No Yes Yes 120 seconds
a. Set by clinician at Acuity Central Station. b. Set by Acuity System Administrator during system installation. c. SpO2 alarm limit range depends on the software version of the Acuity System to which the monitor is connected. (See “Heart Rate
and Arrhythmia Analysis Option” on page 52 and “Pulse Oximetry (SpO2) Specifications - Nellcor” on page 54.)
c
Set at Acuity Previous
Monitor
No Yes No Yes Adult
Yes No Yes Yes Single
Yes Yes Yes Yes Adult: 50, 120 bpm
Yes Yes Yes Yes Adult: 90, 100%
No No Yes Yes U.S.
For Each Patient
No No
No No
a
For Entire Acuity Unit
Yes Yes
Yes Yes
b
Setting Retained at Monitor at Power-Up
No Yes
No Yes
Monitor Default Setting
off high
off low
Ped: 50, 150 bpm
Ped: 90, 100%
50 Chapter 6 Reference Welch Allyn Micropaq Monitor

Specifications

Monitor Radio Specifications
Characteristic Specification
FlexNet Network 2.4 GHz Wireless Local Area Network (WLAN) and 10/100 Base-T Ethernet
Frequency Modulation Frequency Hopping Spread Spectrum (FHSS) Output Power 100 mW IEEE 802.11 compliant Yes Monitors per Access Point 15 (maximum) in most countries. In countries where available frequencies
a. When used within certain countries, authorization for use is restricted as follows:
a
France: The equipment is internally restricted to the 2.448-2.482 GHz frequency range. Spain: The equipment is internally restricted to the 2.447-2.473 GHz frequency range. Japan: The equipment is internally restricted to the 2.473-2.495 GHz frequency range. Italy: Operation requires a user license.
Note: The frequency ranges specified above are subject to geographic-specific regulatory authorities.
network
2.402 to 2.480 GHz
are limited, this number is reduced.
ECG Specifications
The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and Alarms specified ANSI/AAMI EC13-1992, except for Impulse response at the monitor (section 3.2.9.8 part (c)), and Standardizing Voltage at the monitor and at Acuity (section
3.2.9.9). The channel also meets the American National Standard, Safe Current Limits for Electromedical Apparatus (ANSI/AAMI ES1-1993).
Characteristic Specification
Connector Hypertronics D01 latching connector Selectable Leads
5-Lead Cable
3-Lead Application (using 3-lead ECG
cable, or 5-lead ECG cable with
detachable lead wires; only RA, LA,
LL electrodes connected)
Lead Fault Indicator Displayed chest icon with flashing indicator for each electrode ECG Size (sensitivity) 0.2, 0.5, 1, 2, 4, and 8 mV/cm Display Sweep Speed 25 mm/sec Bandwidth
Local display
To Acuity Central Station
Sample Rate 364 Hz (182 Hz with turning point decimation to Acuity Central Station) Input Protection Electrosurgery and defibrillator protected when used with ECG cables
Electrosurgery interference suppression Included on all vectors. Lead Fail Sense Current 70 nA dc typical for active leads.
Tall T-wave Rejection Meets AAMI (USA) EC13-1992, section 3.1.2.1.c, up through 1.2 mV
Monitored: II, III, V; Derived: I, aV Monitored: II
0.5 to 94 Hz independent of patient mode
0.05 to 94 Hz independent of patient mode
specified in the Welch Allyn Products and Accessories booklet (810-0409­XX).
140-280 nA dc typical for reference electrode, depending on number of electrodes attached.
, aVL, aV
R
F
Directions for Use Chapter 6 Reference 51
Characteristic Specification
Common Mode Rejection
FILTER function OFF
FILTER function ON
Input Impedance >2.5 M¾ differential @ 60 Hz Input Range (ac) 10 mV peak to peak (local display)
Input Range (dc) Up to ±500 mV System Noise ð30 µV peak-to-peak, RTI QRS Detector Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)
Heart Rate Range
Alarm Limits
Heart Rate Meter Response Time Responds to change in heart rate within 5 to 9 seconds depending on
HR Display Update Interval at monitor 1 second HR Accuracy ±3 beats per minute or 3%, whichever is greater Heart Rate Response to ineffectively
paced QRS pattern
Heart Rate Averaging Method Heart rate = 60/ latest average interval in seconds.
Drift Tolerance (AAMI Specification EC13­1992, 3.2.6.3) Pacer Pulse Display Pacer indicator shown on screen if PACER display turned ON; pacer spike
Pacer Pulse Rejection Pacer detection range (i.e., will show the dashed vertical marker) for ±3 mV
<1 mV p-p RTI for 10V rms, 50/60 Hz into unbalanced input <30 µV p-p RTI for 10V rms, 50/60 Hz into unbalanced input
10 mV peak to peak (Acuity Central Station)
Adult Width Range (Duration): 70 to 120 msec Pediatric Width Range (Duration): 40 to 120 msec
25 to 350 beats per minute (measurement) 25 to 300 beats per minute (display) 25 to 245 beats per minute (lower) 30 to 250 beats per minute (upper)
physiological waveform. (As measured per AAMI standard EC13-1992 clause 4.1.2.1 (f), including 3.1.2.1 parts f. and g. waveforms.) Includes 1 second readout update interval.
Indicates rate of 30 to 46 during AAMI EC13-1992 part 3.1.4.1 part (f) and (g) tests. NOTE: AAMI Test 4.1.4 part f and g: Accuracy is affected (i.e., rate increases) when QRS and pacer spikes are nearly simultaneous as occasionally is the case during this AAMI test.
For higher heart rates, latest average interval = 7/8 of previous average interval + 1/8 of latest interval. For lower heart rates, latest average interval = 3/4 of previous average interval + 1/4 of latest interval. Transition rates for choice of formula include hysteresis and are 70 and 80 beats per minute.
80 beats per minute indicated for 80 beats per minute ECG plus drift waveform
always shown if of sufficient amplitude.
to ±700 mV @ 0.1 ms width, ±2 mV to ±700 mV @ 0.2 to 2 ms pulse width in electrically quiet environment. Thresholds automatically adjust to reject repetitive ambient noise. Operates even while pacer indication is disabled.
Will not count as heartbeats approximately 95% of pacemaker pulses within pacer detection range, with or without AAMI (EC13-1992) tails of 4, 25, 50, 75, or 100 ms decay time constant, whose tail amplitudes are up to 25%, 2mV maximum, whether ventricular only, or A-V sequential pulses (150 and 250 ms separation), all per AAMI tests 3.1.4.1 and 3.1.4.2
Response to Irregular Rhythm (AAMI Specification EC13-1992, 3.1.2.1. Part e.)
Ventricular Bigeminy (VB) 78 to 81 bpm (80 bpm expected) Slow Alternating VB 57 to 65 bpm (60 bpm expected) Rapid Alternating VB 118 to 123 bpm (120 bpm expected) Bidirectional Systole 88 to 93 bpm (90 bpm expected)
52 Chapter 6 Reference Welch Allyn Micropaq Monitor
Heart Rate and Arrhythmia Analysis Option
Method for Calculating Heart Rate
Monitor Determined by monitor (displayed at monitor)
Heart rate = 60 / latest average interval in seconds. For higher heart rates, latest average interval = 7/8 of previous average interval + 1/8 of latest interval For lower heart rates, latest average interval = 3/4 (previous average interval) + 1/4 latest interval. Transition rates for choice of formula include hysteresis and are 70 and 80 beats per minute.
Acuity System with Arrhythmia Option
Arrhythmia Analysis Option when Connected to Acuity
ST Analysis ST Analysis can be performed for any or all of seven leads, depending on the operator selection.
Determined by Acuity Arrhythmia Option software (displayed at Acuity Central Station)
The beat-to-beat heart rate (HR) value is calculated as follows:
HR = 60000/actual RR (bpm). Actual RR = time between last detected QRS complex and previously detected QRS complex (ms). Average HR is calculated on the basis of the mean RR interval in the last 6 seconds or 8 RR intervals (whichever is shorter).
The operator can select a measurement offset.
ST segment shifts are recorded in continuous trend data every second. The operator can inspect trend data to see the duration and elevation or depression for each episode for any time period recorded. The operator can also inspect a summary of ST segment shift data within tabular trends.
Heart Rate Heart rate information is available in the trend data which can be viewed on the display or printed.
The operator can inspect the trend data to see the lowest, highest, and median (averaged) heart rates. Trend data also includes the total beats per range of time.
Definition of Pause Arrhythmia Event
A pause is defined as the R-R interval which is greater than or equal to two times the average R-R.
Directions for Use Chapter 6 Reference 53
Pulse Oximetry (SpO2) Specifications - Masimo
Characteristic Specification
Saturation (% SpO2)
Range
Resolution
Alarm Limits
With Acuity 6.0 or higher With Acuity 5.4X or lower
a
Probe Accuracy (Adults, Pediatrics)
No Motion
During Motion
b
Pulse Rate
Range
Resolution
Alarm Limits
Pulse Rate Accuracy
No Motion
During Motion
2
Display Update Interval at monitor 1 second Alarm Hold-Off Time Period 10 seconds; resets if the sensor reports levels within limits before 10
Circuitry Microprocessor controlled
Electrosurgery interference suppression Yes Sensor Compatibility Compatible only with Masimo sensors listed in the Welch Allyn Products
Sensor LEDs
RED Wavelength
INFRARED Wavelength
Sensor Energies (Radiant Power) 0.13 mW to 0.79 mW at 50 mA pulsed
alarm limit range depends on the software version of the Acuity System to which the monitor is connected.
a. SpO
2
b. Motion is defined as rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion between 1
to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO ECG monitor. This variation equals ±1 standard deviation which encompasses 68% of the population.
1% to 100% 1%
50% to 99% (lower); 51% to 100% (upper) 80% to 99% (lower); 81% to 100% (upper)
70% to 100% ±2 counts 0% to 69% unspecified 70% to 100% ±3 counts 0% to 69% unspecified
26 to 239 beats per minute 1 beat per minute 25 to 245 beats per minute (lower) 30 to 250 beats per minute (upper)
±3 beats per minute ±5 beats per minute
seconds elapses
Automatic self-test of oximeter when powered on Automatic setting of default parameters Automatic alarm messages
and Accessories booklet (810-0409-XX).
660 nm (nominal) 905 nm (nominal)
against a laboratory co-oximeter and
2
WARNING Interfering Substances: Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substances containing dyes, that change usual arterial pigmentation may cause erroneous readings.
WARNING Although the SpO saturation (with Acuity 6.0 software or higher), the SpO
alarm limit range can be adjusted down to 50%
2
performance and
2
accuracy is not specified below 70%.
54 Chapter 6 Reference Welch Allyn Micropaq Monitor
Pulse Oximetry (SpO2) Specifications - Nellcor
Characteristic Specification
Saturation (% SpO2)
Range
Resolution
Alarm Limits
Probe Accuracy
Pulse Rate
Range
Resolution
Alarm Limits
Pulse Rate Accuracy ±3 beats per minute Display Update Interval at the monitor 1 second Alarm Hold-Off Time Period 10 seconds; resets if the sensor reports levels within limits before 10
Circuitry Microprocessor controlled
Electrosurgery interference suppression Yes Sensor Compatibility Compatible only with Nellcor sensors listed in the Welch Allyn Products and
Sensor LED Wavelengths Within 500 to 1,000 nm Sensor Energies (Radiant Power) Does not exceed 15 mW
a. SpO2 alarm limit range depends on the software version of the Acuity System to which the monitor is connected. b. Refer to the Welch Allyn Products and Accessories guide (810-0409-XX) for accuracy specifications for all Nellcor SpO2 probes
recommended for use.
c. Although some of the listed Nellcor sensors can be used with neonates with other pulse oximetry devices, the monitor is only intended
for use with adult and pediatric patients, not with neonates.
a
With Acuity 6.0 or higher With Acuity 5.4X or lower
b c
(Adults, Pediatrics)
1% to 100% 1%
50% to 99% (lower); 51% to 100% (upper) 80% to 99% (lower); 81% to 100% (upper) 70% to 100% (0% to 69% unspecified) OxiMax Max-A, Max-AL ±2 counts OxiCliq N ±2.5 counts D-YS ±3 counts DS-100A ±3.5 counts
26 to 239 beats per minute 1 beat per minute 25 to 245 beats per minute (lower) 30 to 250 beats per minute (upper)
seconds elapses
Automatic self-test of oximeter when powered on Automatic setting of default parameters Automatic alarm messages
Accessories booklet (810-0409-XX).
WARNING Interfering Substances: Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substances containing dyes, that change usual arterial pigmentation may cause erroneous readings.
WARNING Although the SpO saturation (with Acuity 6.0 software or higher), the SpO
alarm limit range can be adjusted down to 50%
2
performance and
2
accuracy is not specified below 70%.
Directions for Use Chapter 6 Reference 55
Patient Alarm and Equipment Alert Specifications
Characteristic Specification
Visual Alarm Indicator at the monitor
Flashing GREEN LED
Flashing RED LED
Continuously ON RED LED
Flashing YELLOW LED
Continuously ON YELLOW LED
Audio Tone Locations Monitor
Audio Tone Frequency 2760 Hz Life-Threatening Arrhythmia Alarm
Tone Pattern Patient Alarm Tone Pattern Equipment Alert Tone Pattern Audio Tone Volume The monitor audio tone volume is configured by the Acuity System to High,
Limits Setable on all parameters Alarm Control Automatic preset or manual settings Alarm Priority Highest priority: Patient alarms
Alarm on Tachycardias Most tachycardias will alarm in less than 8 seconds. These include AAMI
Alarm Holdoff Time Period
Acuity-Configurable Audio Alarm Delay at the monitor
Patient Alarm Tone Silence from the monitor or Suspend from Acuity
a
Normal operation Patient alarm Patient alarms are silenced An equipment alert or not connected to the network Equipment alert suspended for 90 seconds at Acuity or low battery alert acknowledged (dismissed)
Acuity Central Station (when connected)
1 second on, 1 second off 1second on, 2 seconds off 1second on, 4 seconds off
Low, or Off. The monitor can be configured with separate audio tone volume settings for when it is connected to an Acuity System and when it is not.
Lowest priority: Equipment alerts
3.1.2.1 part f. waveforms. Certain multifocal tachycardias may initially alarm as "low rate." HR = 3 seconds % SpO
, PR = 10 seconds
2
When a monitor is connected to an Acuity System, the audio alarms at the monitor can be delayed up to 4 minutes and 15 seconds. The delay time is selected in Acuity software at the time of Acuity installation. Visual alarm indications are not delayed. The monitor LED is continuously ON RED and the audio tone is silenced for 90 seconds (non-adjustable).
If original alarm was silenced from the monitor, new patient alarms or equipment alerts break the silence at the monitor, but only life-threatening arrhythmia alarms break the silence at Acuity. If original alarm suspended at Acuity, only life-threatening arrhythmia alarms break the silence at the
monitor and Acuity. Equipment Alert Acknowledge from the monitor Equipment Alert Suspend from Acuity The LED is continuously ON YELLOW and the audio tone is silenced for 90
Patient Alarm Tone Reset from the monitor or Resume from Acuity Patient Out of Range; Transmitter Failure
The LED returns to the pre-alert state (except Low Battery remains
continuously ON YELLOW) and the auditory tone is dismissed.
seconds (non-adjustable).
For a patient alarm tone that has been silenced, resets the tone.
An equipment alert is generated whenever the monitor fails to communicate
to an Acuity System after a connection has been successfully established. In
addition, the “No Acuity” icon is displayed on the monitor display.
56 Chapter 6 Reference Welch Allyn Micropaq Monitor
Characteristic Specification
Transmitter Battery Failure An equipment alert is generated before the monitor battery becomes
Nurse Call When is pressed and the instrument is connected to an Acuity System,
a. To help minimize false alarms, the monitor briefly delays or "holds off" triggering both audible and visual alarms for limit violations
for these vital signs. After the alarm hold-off period begins, if the monitor detects that the patient’s vital sign has returned to acceptable limits, the monitor cancels the alarm hold-off. The next time a vital sign limit is violated, the monitor starts a new hold­off period.
exhausted.
the monitor sends a Nurse Call message to the Acuity System.
Display Specifications
Characteristic Specification
Type Monochrome passive matrix; LCD module Resolution 320 x 200 pixels Active Viewing Area 2.26 x 1.41 in. (57.5 x 35.9 mm) Pixel Pitch 0.0071 in. (0.18 mm) Pixel Size 0.0065 in. (0.165 mm) Viewing Angle 6 o’clock position Display Color black on white
Environmental Specifications (with Battery Installed)
Characteristic Specification
Operating Temperature 0° to 40° C Shipping and Storage Temperature -20° to 60° C Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m) Shipping and Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m) Operating Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural Shipping and Storage Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural Drop 1 meter onto vinyl tile over concrete per EN60601-1 Shock 50 g Vibration, Random 0.02g
Degree of Protection Against Ingress IPX1 Rating, Drip Proof per EN60529: 1991 Electromagnetic Compatibility (EMC) EN60601-1-2: 1993
Caution The monitor may not meet performance specifications if it is not used or stored within these environmental specifications.
2
/Hz from 10 to 500 Hz, ramping down to 0.002g2/Hz at 2000 Hz. Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA DO­160D, Category C.
Directions for Use Chapter 6 Reference 57
Physical Specifications
Protection Classifications, all Configurations Characteristic Specification
Type of Protection against Electric Shock—Monitor Type: CF
Degree of Protection Against Electric Shock, for Parts Applied to Patients Recovery time following defibrillator discharge Method of Disinfection Not suitable for autoclaving (see cleaning instructions on “Inspect and
Flammable Anesthetics Not suitable for use with flammable anesthetics. Height 7.30 in (18.5 cm) Width 3.50 in (8.9 cm) Depth 1.46 in (3.7 cm) Weight (including battery)
Model 402 Model 404
Battery operation only Battery must be recharged in separate battery charger. IEC EN 60601-1, 2nd Edition See monitor labels. CF defibrillator protected. IEC EN 60601-1, 2nd Edition Less than or equal to 10 seconds
Clean the Monitor and Accessories” on page 45).
14.7 oz (0.42 kg)
17.3 oz (0.49 kg)
Battery Specifications
Characteristic Specifications
Reorder Number Lithium Ion Battery 2EA Active A
Battery Type Rechargeable, Lithium Ion Battery Capacity 2 cells, 8.4 V, 1800 mA-hr Battery Weight 4.5 oz (0.13 kg) Battery Charger External device Battery Fuse Rating 5 A, 125 V (not user-accessible)
Operating Times on Battery
Battery Recharge Time 4 hours at 25° C (typical) Battery Lifetime 300 charge/discharge cycles to 70% of original
a. Battery operating times based on these conditions: new fully-charged battery operating at 25° C,
the monitor connected to Acuity, eight patient alarms per hour, minimal motion artifact.
Note
The following factors reduce battery operating time:
Amount of time not connected to Acuity.
Frequency and duration of alarms and alerts.
Amount of operator activity using monitor keys (activates display).
Age of battery.
Amount of motion artifact during SpO
008-0647-XX
Note: Internal electronic overload circuitry is used as the primary method of protection.
a
Monitor w/ ECG only: 25 hrs Monitor w/ ECG and SpO
capacity (typical)
: 10 hrs
2
monitoring.
2
58 Chapter 6 Reference Welch Allyn Micropaq Monitor
WARNING The monitor battery is Lithium Ion. Do not incinerate, submerge, crush, disassemble, or autoclave. If a battery has been submerged in liquid, discard the battery properly; do not try to recharge or reuse the battery. Do not short the battery terminals. Do not try to connect the battery to any device except the monitor or the battery charger. Do not expose to high temperature (above 60° C or 140° F). Use only the specified monitor battery charger.
Directions for Use Chapter 6 Reference 59
Battery Charger Specifications
Characteristic Specification
Reorder Number Universal Battery Charger Active C
Functional Specifications
Capacity Eight charging bays; able to charge four (Lithium Ion) batteries
Protection Classifications
Duty Cycle Continuous Type of Protection Against Electric
Shock Degree of Protection Against Harmful
Ingress of Water Method of Disinfection Not suitable for autoclaving. (See cleaning instructions on “Inspect and
Flammable Anesthetics Not suitable for use with flammable anesthetics.
Environmental Specifications
Operating Temperature 0° to 40° C Shipping and Storage Temperature -20° to 60° C Operating Altitude -2,000 to 15,000 feet (-610 to 4,572 m) Shipping and Storage Altitude -2,000 to 40,000 feet (-610 to 12,192 m) Operating Relative Humidity 15% to 95%, noncondensing Shipping, Storage Relative Humidity 15% to 95%, noncondensing Shock 30 g Vibration 0.01g Electromagnetic Compatibility (EMC) EN60601-1-2: 1993
Physical Specifications
Length 15.0 in (38.1 cm) Width 9.0 in (22.9 cm) Height 3.6 in (9.1 cm) including feet Weight 3.5 lb (1.6 kg)
Electrical Specifications
Rated Input 100 V-240 V ac, 600 mA, 50/60 ± 3 Hz, Electrical Class I Rated Fuses T1.25 A/250 V, Time-Delay, 5x20 mm Rated Output per charging bay
(Continuous) Charge Time 4 hours typical for fully discharged battery. Automatic charge termination
Output Over-Current Electronic overload protection Additional Features Detachable power cord, pilot light
LED Indicators
LED OFF No battery detected. Flashing GREEN LED
1 sec ON, 3 sec OFF
1 sec ON, 1 sec OFF Continuously ON GREEN LED Battery is charged. Continuously ON YELLOW LED Battery or charging bay fault.
a. Per EN 60601-1 unless otherwise stated.
a
008-0651-XX
simultaneously.
Class I, (Protectively Earthed) with Double Insulation
For ordinary, indoor locations only.
Clean the Monitor and Accessories” on page 45.)
2
/Hz from 5 to 500 Hz, 30 minutes per axis
8.4 V ± 100 mV dc @1 A max.
when charge is completed, or fault detected.
Battery detected, waiting to be charged. Battery is charging.
60 Chapter 6 Reference Welch Allyn Micropaq Monitor
EMC Compliance
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical electrical equipment. This device complies with IEC EN 60601-1-2:2001.
All medical electrical equipment must be installed and put into service in accordance with the EMC information provided in this document and the Micropaq Monitor Directions For Use.
Portable and mobile RF communications equipment can affect the behavior of medical electrical equipment.
The monitors and battery charger comply with all applicable and required standards for electromagnetic interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is safe to operate the monitor in the presence of high-frequency surgical
equipment.
However, it is good practice to avoid using the monitor in extremely close
proximity to other equipment.
Directions for Use Chapter 6 Reference 61
Monitor
Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment—Guidance
RF emissions CISPR 11
Group 1
The Model 400 Series Monitor uses RF energy only for its internal function.
a
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Model 400 Series Monitor is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
No connection to mains (battery-operated)
No connection to mains (battery-operated)
network that supplies buildings used for domestic purposes.
IEC 61000-3-3
a. The Model 400 Series Monitor contains a 2.4-GHz frequency-hopping spread-spectrum transmitter for the purpose of wireless
communication. The radio is operated according to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/EC). The radio is excluded from the EMC requirements of 60601-1-2:2001, but should be considered when addressing possible interference issues between this and other devices.
62 Chapter 6 Reference Welch Allyn Micropaq Monitor
Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment—Guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/ burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions, and voltage variations on power­supply input lines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
<5% U
t
(>95% dip in Ut)
No connection to mains (battery-operated). No other cables requiring EFT/Burst testing.
No connection to mains (battery-operated).
No connection to mains (battery-operated).
Since there is no connection to the mains, there is no requirement for mains quality.
for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Note U
is the AC mains voltage prior to application of the test level.
t
Directions for Use Chapter 6 Reference 63
Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
Compliance Level
3 V
rms
3 V/m
Electromagnetic Environment—Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Model 400 Series Monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance
d = 1.2
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distance in meters. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey compliance level in each frequency range
a
, should be less than the
b
. Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Model 400 Series Monitor is used exceeds the applicable RF compliance level above, the Model 400 Series Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the monitor.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
64 Chapter 6 Reference Welch Allyn Micropaq Monitor
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Model 400 Series Monitor
The Model 400 Series Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the monitor as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output Power of Transmitter W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.21.22.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation Distance According to Frequency of Transmitter m
150 kHz to 80 MHz d = 1.2
P P P
80 MHz to 800 MHz d = 1.2
800 MHz to 2.5 GHz d = 2.3
Directions for Use Chapter 6 Reference 65
Battery Charger for the Monitor
Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The Battery Charger for the Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the battery charger should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment—Guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Group 1
Class B
Class A
Complies
The Battery Charger uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Battery Charger is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
66 Chapter 6 Reference Welch Allyn Micropaq Monitor
Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Battery Charger for the Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Immunity Test IEC 60601 Test level Compliance Level Electromagnetic Environment—
Guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions, and voltage variations on power-supply input lines IEC 61000-4-11
Power frequency (50/ 60 Hz) magnetic field IEC 61000-4-8
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
±2 kV common mode
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Battery Charger requires continued operation during a power mains interruption, it is recommended that the Battery Charger be powered from an uninterruptible power supply or battery.
3 A/m 3 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location in a typical commercial or hospital environment.
Note U
is the AC mains voltage prior to application of the test level.
t
Directions for Use Chapter 6 Reference 67
Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Battery Charger for the Model 400 Series Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the battery charger should assure that it is used in such an environment.
Immunity test IEC 60601 Test
Level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
Compliance
Electromagnetic Environment—Guidance
Level
Portable and mobile RF communications equipment should be used no closer to any part of the Battery Charger, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance
3 V
rms
d = 1.2
3 V/m d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distance in meters. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol:
a
, should be less than the
b
.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Battery Charger is used exceeds the applicable RF compliance level above, the Battery Charger should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Battery Charger.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
68 Chapter 6 Reference Welch Allyn Micropaq Monitor
Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the Battery Charger for the Model 400 Series Monitor
The Battery Charger for the Model 400 Series Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Battery Charger can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Battery Charger as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output Power of Transmitter W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.21.22.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation Distance According to Frequency of Transmitter m
150 kHz to 80 MHz d = 1.2
P P P
80 MHz to 800 MHz d = 1.2
800 MHz to 2.5 GHz d = 2.3

Index

69
1-Waveform display, 14 2-Waveform display 5-Second waveform display
, 14
, 14
A
Accessories, 15
Inspection and cleaning
Use only recommended Access points Acuity
, 7
, 7
Acuity Central Station
Alarm holdoff at Acuity
Default settings for Micropaq
Download default settings
DROPOUT equipment alert
Messages displayed on Micropaq
Micropaq operating settings
Monitoring at Acuity
Silence Micropaq alarms ACUITY CONNECTION LOST Acuity Menu
END TELE
NEW PATIENT
NEW ROOM
TRANSFER
, 12
, 12
, 12
, 12
Adjust alarm limits at Micropaq Alarm/Alert Silence Key Alarm holdoff with Acuity Alarms
Adjust alarm limits at Micropaq
Alarm holdoffs (delays)
Alarms off symbol
Download defaults from Acuity
Priority
, 55
Respond to patient alarms
Silence Alarms Key
Specifications Alerts
, 35
Messages
Micropaq
, 55
, 40
, 38
Silence Alerts Key Anemia and SpO2
, 27
Anesthetics
, 45
, 5
, 1, 7
, 55
, 21 , 30
, 41
, 36 , 40
, 10
, 55
, 35
, 11
, 36
, 10
, 10
, 49
, 16
, 37
, 37
, 21
, 15
Warning
Arterial occlusion and SpO2
, 3
, 27
Artifact interference and waveform display Audible indicator Autoclave, caution about use
, 9
, 5, 45
B
Battery
Changing Low battery icon Recharge Recycling Specifications
, 43
, 11
, 43
, 46
, 57
Battery charger
Specifications Use
, 43
, 59
Battery life
Sleep mode BATTERY TOO LOW Beeps
, 9
Bell symbol Blank display during sleep
, 11
, 40
, 11
, 11
C
Cardiac arrest and SpO2, 27 Cautions, general Change network name Change the battery Change waveform display Charge a battery
, 5
, 47
, 43
, 14
, 43
Charger
Recharge a battery
Specifications Cleaning
, 45
Connect a new patient Connect to the network Customize alarms
, 43
, 59
, 19, 21
, 19
, 21
D
Default
, 22
70 Index Welch Allyn Micropaq Monitor
Acuity and settings, 16, 49 Operating settings
DEFECTIVE SPO2 SENSOR
, 16, 49
, 40
Defibrillation
During SpO2 monitoring Warning
Delays for alarms
, 3
, 35
, 26
Demo Mode
Accessing
, 16
Display values, alarm limits
Toggling between modes Demonstration Mode Disconnection symbol
, 17
, 16
, 11
Discontinue monitoring a patient Display
Alert messages
Artifact interference
Blank during sleep mode
Indicator symbols
Pacemaker signals
Sleep mode
Specifications
Waveform display options DROPOUT alert at Acuity
, 40
, 22
, 11
, 11
, 11
, 11
, 56
, 14
, 30, 43
, 17
, 30
Automatic reconnection Defined
, 7
Operation out of range
, 7
, 7
Full Screen waveform display
G
General cautions, 5 General Warnings Green LED
, 3
, 9
H
Hazards
General cautions General warnings
, 5
, 3
Holdoffs (delays) for alarms Hyperbaric chamber
Warning, do not use Hypotension and SpO2 Hypothermia and SpO2
, 4, 21
, 27
, 27
I
, 14
, 35
E
ECG
Alarm holdoff (delay) Electrode application Electrode lead off Lead selection Scale selection
Size (scale) selection Electrode application Electrosurgery warning END TELE
, 12, 30
Environmental specifications
, 35 , 21
, 23
, 12
, 12
, 12
, 21
, 3
, 56
Equipment alert
Description
Messages
, 38
, 40
Errors
Error message
Error number Ethernet
, 1
EXCESSIVE ECG OFFSET
, 5
, 5
, 40
F
FCC
Warning, RF Federal Communications Commission
see FCC FlexNet
Access points
, 4
, 1, 7
, 7
In-Service
see Demo Mode Inspection Intended Use
, 45
, 1
K
Keys
Alarm/Alert Silence
Nurse Call
, 9
Scroll Up/Down
Select
, 10
KEY STUCK
, 40
, 10
, 9
L
Lead off indication, 23 Lead selection LEDs defined List of patients List of rooms Lithium ion battery Lockout menu Low battery icon LOW BATTERY message
, 12
, 9
, 20
, 21
, 43
, 13, 49
, 11
, 40
M
Magnetic Resonance Imaging
Directions for Use Index 71
see MRI
Main Menu
EXIT
, 12
Menu Lock option Restricting access
, 13, 49
, 13
Maintenance
Inspection and cleaning Operator maintenance
, 45
, 43
Masimo
SET technology Menu Lock option Messages for alerts Messages from Acuity
, 6
, 13, 49
, 40
, 15
Micropaq
Acuity messages
Adjust alarm limits
Alarms
, 36
Audible indicators
Automatic reconnection
Battery recharging
Communication with network
Connect to the network
Default operating settings
Default settings
Demo Mode
Equipment alert
Features
, 6
NEW PATIENT
Operating settings
Out of range operation
Reassign to new patient
Specifications
, 15
, 37
, 9
, 7
, 43
, 7
, 19
, 16
, 49
, 16
, 38
, 34
, 15
, 7
, 34
, 50
Monitor
Audible indicators
Features
, 6
Out of range
Recycling
, 46
Specifications
, 9
, 29
, 50
Monitoring
At Acuity
Connect a new patient
, 41
, 19
Reconnect a recently monitored patient
Stop monitoring a patient Motion interference and SpO2
, 30
, 27
MRI
Warning, do not use monitor
, 4, 21
N
, 31
Automatic reconnection Connect the monitor Disconnection symbol Ethernet FlexNet Name change
, 1
, 7
, 47
Operation out of range NEW PATIENT NEW ROOM
, 12, 34
, 12, 32
, 7
, 19
, 11
, 7
NIBP
While monitoring SpO2 NO ECG CABLE DETECTED NO SPO2 SENSOR DETECTED Nurse Call Key
, 9
, 27 , 40
, 40
O
Operation
Default settings
Settings Operator maintenance Out of range
Monitor a patient out of range OxiMax technology, Nellcor
, 16
, 15
, 43
, 7
, 29
, 6
P
Pacemaker
Signal detection, display
Warning, rate meters
Warning, wireless radio Pacer
see Pacemaker Pacer detection
, 49
Patient
Alarms
, 36
Reassign to a new room
Reconnect a recently monitored patient
TRANSFER
, 33
Patient alarms
Download defaults from Acuity Patient list
, 20
Patient preparation for ECG Physical specifications Power specifications Prepare the patient Pulse oximetry
see SpO2
, 11
, 3
, 4
, 32
, 21
, 21
, 57
, 57
, 21
, 31
Nellcor
OxiMax technology
Neonatal
Not intended for use
Network
Access points
, 7
, 6
, 3
R
Radio Frequency
see RF Reassign Micropaq to new patient
, 34
72 Index Welch Allyn Micropaq Monitor
Reassign to a new room, 32 Recharge a battery
, 43
Reconnect a recently monitored patient Reconnection to network
, 7
Recycling
Batteries
Monitor Red LED Replace the battery Respond to equipment alert Respond to patient alarms RF warning, exposure Room selection Routine maintenance
, 46
, 46
, 9
, 43
, 38
, 36
, 4
, 21
, 43
S
Scroll Up/Down Keys, 9 Select
Patient
, 20
Room
, 21
Select Patient at Central
Select Room at Central Select Key
, 10
Select Room at Central Set patient alarms
, 37
SET technology, Masimo Settings, operating Shock and SpO2
, 27
Silence Alarm/Alert Key Silence alarms
, 36
Simulation
see Demo Mode Site preparation for ECG Sleep mode, display Snapshot print, Nurse Call Specifications
Alarms
Battery
Battery charger
Display
Environmental
Physical
SpO2
, 50 , 55 , 57
, 59
, 56
, 56
, 57
, 53
SpO2
Alarm holdoff (delay) Anemia
, 27
Arterial occlusion Cardiac arrest or shock Hypotension Hypothermia
, 27
, 27
Light interference Monitoring
, 26
Monitoring and defibrillation Motion interference
, 20
, 21
, 32, 33
, 6
, 49
, 10
, 21
, 11
, 9
, 35
, 27
, 27
, 27
, 26
, 27
, 31
Specifications Vasoconstriction Warnings
While monitoring NIBP Stop monitoring a patient Symbols, display System error
, 53
, 27
, 26
, 27 , 30
, 11
, 40
T
Training
see Demo Mode TRANSFER
, 12, 33
Transfer a patient to new room
U
Unassigned rooms, 21 UNRECOGNIZED SPO2 SENSOR
V
Vasoconstriction and SpO2, 27 Venous pulsations and SpO2
, 26
VERY LOW BATTERY message Volume of audible indicators
, 9
W
Warnings
General
, 3
SpO2
, 26
Warranty service Waveform display options
, 5
, 14
Wireless
Warnings about limitations Wireless Local Area Network
see WLAN WLAN
, 1, 7
Y
Yellow LED, 9
, 33
, 40
, 40
, 3
Loading...