Copyright 2007 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or
any part thereof without permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may
result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use
published in this manual.
Welch Allyn
ParamSet
Masimo
Possession or purchase of a Masimo SpO
®
, Propaq®, Acuity®, Smartcuf®, FlexNet®, and Flexible Monitoring® are registered trademarks of Welch Allyn.
™
is a trademark of Welch Allyn.
®
, SET®, and LNCS® are registered trademarks, and FastSAT™ and APOD™ are trademarks, of Masimo Corporation.
-equipped monitor does not convey any express or implied license to use the
2
device with unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of
one or more of the patents relating to this device.
®
Nellcor
and Oximax® are registered trademarks of Nellcor Puritan Bennett.
Software in this product is copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by United
States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is
entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in
which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise
reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and
ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, call the nearest Welch Allyn representative:
USA 1 800 535 6663
+ 1 315 685 4560
Canada 1 800 561 8797China + 86 216 327 9631
European Call Center + 353 46 906 7790France + 33 1 6009 3366
Germany + 49 747 792 7186Japan + 8133 219 0071
Latin America + 1 305 669 9003Netherlands + 3115 750 5000
Singapore + 656 419 8100South Africa + 2711 777 7555
The Propaq LT Series (802LTAN, 802LTAS, 802LT0N, 802LTRN, 802LT0S, and 802LTRS)
monitors are portable devices intended to be used by clinicians and medically qualified
personnel for single- or multiparameter vital-signs monitoring of ambulatory and
nonambulatory patients, including neonate, pediatric and adult patients. These devices
are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO
likely locations for patients to be monitored by these devices are hospital general
medical-surgical, telemetry, and intermediate care floors, hospital emergency
departments, transport, emergency medical services, and other healthcare applications.
The monitors can be used as standalone devices or as devices networked to an Acuity
Central Station (referred to in this manual as ‘Acuity’) through wireless communication
over a Welch Allyn FlexNet network.
Federal USA law restricts this device to sale, distribution, or use by or on the order of a
licensed healthcare professional.
Even though this manual describes some monitoring techniques, the monitor is intended
for use only by trained and experienced clinicians who know how to measure and
interpret vital signs.
. The most
2
Symbols
Table 1. Directions for use
WARNING Indicates conditions that could lead to illness, injury, or death.
Caution In this manual, indicates conditions that could damage equipment or other property.
Caution On the product, means “Consult the accompanying documentation.”
Power - Turn the monitor power on or off.Down - Move the cursor down to the next display
item.
Action - Act based on what is highlighted.
(See “Using the action button” on page 29.)
Left - Move the
item; decrease the parameter value.
Right - Move the cursor right to the next
display item; increase the parameter value.
Up - Move the cursor up to the next display
item.
Monitor Release - Press and then hold while removing the monitor from the cradle.
cursor left to the next display
Cradle
Silence/Reset - Silence the current alarm tone
for 90 seconds or reset a silenced alarm tone.
Display - Cycle to the next configured display
format, or cancel the current control, setup, or
pop-up menu.
Snapshot - Record a 21-second period of numeric
and waveform data.
Start/Stop NIBP - Start or stop an NIBP
measurement.
Table 3. Status indicators
Monitor
(green) Monitoring normally (no active alarms or alerts).
Connection to Acuity is confirmed and patient identification is confirmed. (Wireless only, Acuity enabled.)
(green flashing) Monitoring normally (no active alarms or alerts). (Standalone only.)
Patient confirmed, and monitor then intentionally disconnected. (Wireless only, Acuity enabled.)
(yellow) At least one alarm is disabled.
Monitor disconnected, connecting or connected; patient not confirmed. (Wireless only, Acuity enabled.)
Upper and lower alarm limits for this parameter are on.
The upper alarm limit for this parameter is on and the lower is off.
The upper alarm limit for this parameter is off and the lower is on.
Upper and lower alarm limits for this parameter are off.
A snapshot exists for this period.
The snapshot for this period has been replaced with a more recent snapshot.
(green) The battery is fully charged.
The battery is partially full.
The battery is partially full and is charging.
(yellow) The battery is low.
The battery is low and is charging.
Page 7
Directions for UseIntroduction3
Table 3. Status indicators (continued)
(red)The battery is near failure; the monitor will shut down soon. If this indicator appears
while the monitor is in the cradle, the battery cannot be charged and must be replaced.
The battery is near failure and is charging. The monitor will shut down if removed from the cradle.
The monitor is communicating wirelessly with the network and with Acuity. (Wireless only, Acuity enabled.)
The monitor is not communicating with the wireless network. (Wireless only, Acuity enabled.)
(Flashing) The monitor is communicating with the network but is not communicating with Acuity.
(Wireless only, Acuity enabled.)
The monitor is communicating by USB cable with a PC.
Cradle
(green) Cradle is powered.
(green) Monitor battery is charging.
NOTE: When the battery is fully charged, this indicator is not lit.
(yellow) Cradle fault or battery fault.
Table 4. Labels
Proceed with caution. If in doubt, refer to the
accompanying documentation.
The monitor or accessory meets all essential
requirements of the European Medical Device
Directive 93/42/EEC for a Class II-b product.
The monitor or accessory meets all essential
requirements of the European Medical Device
Directive 93/42/EEC for a Class I product.
Australian registered importer.Hazard Class 9, IATA/ICAO (International Air
N344
Restrictions for use of wireless device in
Europe. European Communities Class 2 radio
equipment.
FCC ID: PGUWA11A07
IC:4168a-WA11A07
See the accompanying manual.Non-ionizing electromagnetic radiation. This
Recycle the monitor, cradle, and battery
separately from other waste. (Refer to
www.welchallyn.com/weee for collection-
point and additional information.)
High voltage. Do not touch during
defibrillation.
Monitor
Enclosure protection: Drip
-proof. Class IPX1
per EN60529:1991.
The monitor is certified by the Canadian
Standards Association International to comply
with applicable US and Canadian medical
safety standards.
Type CF patient connections, isolated for direct
cardiac application and protected against
defibrillation.
Transport Association/International Civil
Aviation Organization).
This device complies with the 47 CFR Part 15
radiated and conducted emissions
requirements.
This device complies with FCC and Industry Canada requirements for international
radiators (802.11 wireless).
device contains an approved RLAN module of
frequency 5150-5825 MHz.
Non-ionizing electromagnetic radiation. This
device contains an approved RLAN module of
frequency 2402-2480 MHz. (Wireless only,
Acuity enabled; see “EMC” on page 160.)
Recycle the battery separately from other
disposables.
-ion battery.
Lithium
Cradle
Power in (DC).Fuse replacement specification.
T3A/250V
USB cable connector.For indoor use only.
Input power (DC) pin pattern.
Product packaging
Store this way up.Humidity limit.
Temperature limits.Altitude limits.
Rain protection required.Contents are fragile.
Safety
The monitor is safe for patients and clinicians when used in accordance with the
instructions and with the warning and caution statements presented in this manual.
All personnel must read and understand all warning and caution statements
presented in this manual before using the monitor.
•Failure to understand and observe any warning statement in this manual could
•Failure to understand and observe any caution statement in this manual could
General warnings
These statements apply to all aspects of patient monitoring. Statements which apply
specifically to one aspect of monitoring, such as NIBP or SpO
in the corresponding sections of the manual.
Stacking limit.Recycle the packaging material.
lead to patient injury, illness, or death.
lead to equipment damage or loss of patient data.
monitoring, are presented
2
WARNING Many environmental variables, including patient physiology and
clinical application, can affect the accuracy and performance of the monitor. The
clinician must verify all vital-signs information prior to patient intervention.
WARNING Always check the patient mode (adult, pediatric, or neonate) when
monitoring a new patient. The patient mode determines default alarm limits and
internal algorithm settings. Make sure the monitor has settings that are
appropriate before monitoring the patient.
Page 9
Directions for UseIntroduction5
WARNING Make sure Acuity patients, and especially those prone to
arrhythmias, are kept under close surveillance. While monitoring patients with
Acuity, the clinician must review all clinical data before implementing therapy. As
with all computerized arrhythmia analysis systems, Acuity cannot replace skilled
care and proper surveillance by a clinician.
WARNING It is possible for Acuity alarms, alerts, or other events to go
unnoticed if clinical personnel are not present at Acuity or if interruptions occur in
power or system operations. To help reduce this possible occurrence, Acuity
must be installed with redundant power supplies and redundant means of
operator surveillance, such as secondary Acuity Central Stations and hallway
message panels.
WARNING The monitor might not meet its performance specifications if stored
or used outside the specified temperature and humidity ranges.
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 61 on page 55.)
WARNING Do not connect more than one patient to a monitor.
WARNING Do not connect more than one monitor to a patient.
WARNING During defibrillation, keep discharge paddles away from the monitor
ECG lead wires, electrodes, any other monitor sensors, and other conductive
parts in contact with the patient.
WARNING Do not operate this product in the presence of flammable
anaesthetics or other flammable substances in combination with air or
oxygen-enriched environments. Failure to observe this warning can result in an
explosion.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI)
suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in
patient injury and monitor damage.
WARNING Do not operate this monitor near equipment that emits strong
electromagnetic or radio-frequency signals. Electronic equipment of this type can
cause electrical interference with monitor operation, which can distort the ECG
signal and prevent accurate rhythm analysis.
WARNING To comply with Federal Communications Commission (FCC) RF
exposure requirements and to avoid exposure to radio-frequency (RF) radiation,
always use the monitor in accordance with the operating conditions and
instructions provided in this manual.
WARNING Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that “in
some devices, rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate
meter alarms. All pacemaker patients should be kept under close or constant
observation.” See “Pacer pulse rejection” on page 143 for disclosure of the
pacemaker pulse rejection capability of this instrument.
WARNING Use only accessories supplied by Welch Allyn or recommended in
the Welch Allyn Products and Accessories guide (810-0409-XX). Use of any other
accessories can result in inaccurate patient data and damage to the equipment.
Always use accessories according to facility standards and the manufacturer’s
recommendations. Always refer to the manufacturer’s instructions.
WARNING Use of Masimo LNOP sensors/cables will not provide protection in
accordance with IEC defibrillation standards when used with this device.
WARNING Use only ECG cables supplied or specified by Welch Allyn. Use of
any other ECG cables can negate defibrillator protection and can create a risk of
patient injury due to shock.
WARNING Frequently inspect—electrically and visually—all cables, sensors, and
electrode wires. Replace any damaged cables, sensors or wires. Failure to
properly inspect and keep in excellent working order all cables, sensors, and
electrode wires can result in hazards to patients and to equipment failure and
damage.
WARNING Always properly connect the electrosurgery return circuit. Improper
circuit connection can cause current to return through monitor electrodes and
probes, creating a burn hazard for patients.
WARNING Always keep patient motion to a minimum. Motion artifact can cause
inaccurate measurement of patient vital signs.
WARNING Carefully route and secure patient cabling, using the supplied
garment clips. Improperly routed and secured cabling can cause the patient to
become entangled in the cables, creating a strangulation hazard.
WARNING When the patient is wearing the monitor or being transported by
stretcher with the monitor connected, always take care to position the monitor
carrying straps on the patient. Be certain that the straps do not and cannot cross
the neck or throat and cause choking, and the straps do not restrict movement of
the patient’s arms or legs.
WARNING Never use a monitor that is not working properly. If the monitor is not
working properly, patient waveforms might be inaccurate or might not be
displayed.
WARNING If the monitor is damaged, or if you see any indication that the
monitor is not operating properly, disconnect it from the patient. Do not return it
to service until it has been inspected and, if necessary, repaired by qualified
service personnel.
WARNING This wireless medical device was tested and, when used with a
metal-free accessory between the monitor and the patient, complies with FCC
RF Exposure (SAR) guidelines. The use of accessories containing metal may not
ensure compliance with FCC RF exposure guidelines. Specific Absorption Rate
(SAR) is a measurement of radio frequency energy. The FCC permits a maximum
SAR value of 1.6 mW/g. The highest SAR value for this patient monitor, when
worn by a patient in accordance with the directions for use, is 0.560 mW/g.
WARNING High-power radars are allocated as primary users of the bandwidth
between 5.25 GHz and 5.35 GHz and between 5.65 GHz and 5.85 GHz. These
radars can cause interference with this device and can damage this device.
Page 11
Directions for UseIntroduction7
14040/7878
80
12
97
%
HALL, RO BERT E.
3456187
3:00:06P
Adult ,
Rm 23 9
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
14040/7878
80
12
97
%
HALL, RO BERT E.
3456187
3:00:06P
Adult ,
Rm 23 9
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
General cautions
Caution Do not autoclave the monitor.
Caution Autoclave accessories only if the manufacturer's instructions clearly
direct you to do so. Many accessories can be damaged by autoclaving.
Caution Changes or modifications not expressly approved by Welch Allyn could
void the purchaser’s authority to operate the equipment.
Caution This product contains no user-serviceable components. Any
unauthorized changes to the product invalidate Welch Allyn’s warranty and also
invalidate all applicable regulatory certifications and approvals.
Controls, indicators, and connectors
Figure 1. Controls
DisplayCycle to the next configured display format, or cancel
the current control, setup, or pop-up menu.
PowerTurn the monitor power on or off.
HALL, ROBERT E.
II 1mV/cm
SpO2
80
HR/min
NIBP mmHg (102)
@2:47P Manual
Silence/ResetSilence the current alarm tone for 90 seconds or reset a
3456187
3:00:06P
Adult,
Rm 239
ArrowsMove the display cursor up, down, right, or left to
12
97
Resp/min
SpO2
ActionAct based on what is highlighted.
NIBP Start/StopStart or stop an NIBP measurement.
SnapshotRecord 21 seconds of numeric and waveform data.
Monitor releaseRelease the monitor from the cradle.
silenced alarm tone.
highlight an item; change parameter values.
(See “Using the action button” on page 29.)
Figure 2. Indicators: Monitor
GreenConnection to Acuity is confirmed and patient
HALL, ROBERT E.
3456187
3:00:06P
Adult,
II 1mV/cm
SpO2
80
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
Rm 239
12
97
SpO2
Propaq LT
GreenMonitoring normally; no active alarms or alerts.
(flashing)(Standalone.)
Yellow At least one alarm limit is disabled. (Standalone.)
Yellow Equipment alert. (Standalone.)
(flashing)Acuity message windows.
RedPatient alarm.
(flashing)
identification is confirmed.
(Wireless only, Acuity enabled.)
Patient was confirmed and the monitor was then
intentionally disconnected.
(Wireless only, Acuity enabled.)
The following accessories are available for use with the monitor and the cradle:
•Large Color Display Interface and cables
•Propaq LT Monitor PC Utility software (CD)
•Propaq LT Monitor Service Manual (CD)
•Propaq LT Monitor Directions for Use (CD)
•Patient carry strap
•Patient wearable strap
•Transport stretcher strap
•Connector panel plugs
•SpO
•3-lead and 5-lead ECG cables and cable extensions
•ECG electrodes
•NIBP hoses and cuffs
•Battery pack
•AC power adapter
cables and sensors
2
WARNING Use only accessories supplied by Welch Allyn or recommended in
Welch Allyn Products and Accessories.
For ordering information, see Welch Allyn Products and Accessories (810-0409-XX).
USB option
The monitor can be purchased with the optional USB data transfer capability, which
enables communication between the monitor and a PC. (See “About the USB data
transfer option” on page 15.)
WARNING Always use accessories according to your facility’s standards and the
manufacturer’s recommendations.
WARNING Always refer to the manufacturer’s directions for use.
Page 15
Directions for UseIntroduction11
HIPAA considerations
Each medical facility is responsible for creating and enforcing policies and procedures to
guarantee compliance with the regulations defined in 45 CFR 160-164 of the Health
Insurance Portability and Accountability Act (HIPAA) of 1996.
The Propaq LT Monitor, the Large Color Display Interface, the Propaq LT Monitor
Configuration Utility, and the Propaq LT Monitor AutoPrint Utility incorporate security
features that support your implementation of the HIPAA requirements for ensuring that
patient information is kept private and confidential.
Monitor
•Clinicians can lock the monitor display to prevent the display of patient vital signs. For
patient protection when the monitor display is locked, the display is restored instantly
in the event of a button press (unless button-pad lock-out is enabled), an alarm or an
alert.
•Clinicians can lock the monitor buttons to prevent any unauthorized access to the
monitor controls. To protect the patient when the buttons are locked, access to the
buttons is restored instantly in the event of an alarm or an alert.
•No physiological data transmitted by radio from the monitor is in clear text. All data
sent is encoded using Welch Allyn's proprietary PSI Communications Protocol and
can only be decrypted by a Welch Allyn Acuity Central Station residing on the local
Welch Allyn FlexNet network. By default, data transmitted using the 802.11a
standard is also encrypted using AES (Advanced Encryption Standard).
Configuration utility
The Configuration Utility never contains patient data.
AutoPrint utility
•All patient data is stored in a Welch Allyn proprietary data format which is readable
only by machine.
•All patient data is deleted from the PC when it is sent to the printer.
The monitor runs through an operational self-test each time it is powered on. Always
verify that it follows the power-on sequence shown below. When Step 3 completes as
described, the monitor has verified that the visual and audible alarm indicators are
working properly. If Step 4 does not complete as described, remove the monitor from
service and have it examined by a qualified service person.
To turn on the monitor
1.Press and hold until a tone sounds.
2. A tone sounds as the green, yellow, and red lights turn on and off. This occurs twice.
3. The splash screen appears, identifying the product and displaying the message
“Diagnostics in progress”. The green light turns on and off quickly while the low tone
sounds, the yellow light turns on and off quickly while the medium tone sounds, and
the red light turns on and off quickly while the loud tone sounds.
Figure 6. Splash screen
Propaq
LT
Diagnostics in progress
Application Version 1.00.00
Bootcore Version 1.00.00
c
Welch Allyn 2005
4. The splash screen is replaced by one of the power-on screens shown here:
Figure 7. Power-up screens
Portland Westside Hospital
Emergency Department
No data saved.
Select ( ) to Start a New Patient.
Start New Patient
To start monitoring from this point, see “Standalone monitoring” on page 45. To practice
Info
Demo
Portland Westside Hospital
Emergency Department
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete data
Start New Patient
and start a new patient.
Note!
Continue Patient
Info
Demo
using the monitor, see “Using demo mode” on page 38.
Page 18
14Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
English
Deutsch
Français
Español
Italiano
Polski
Nederl
Svenska
Português
Japanese
Selecting a language
To change the language of the monitor interface
1.If the monitor power is on, press to turn it off.
2. Press to turn the monitor power on.
3. Immediately after pressing , and before any screen appears, simultaneously press
and . Keep the buttons pressed until the language selection screen appears.
Figure 8. Language selection screen
English
Deutsch
Français
Español
Italiano
Polski
Nederl
Svenska
Português
Japanese
4. Press , , , and to highlight the language you wish to use.
5. Press.
Power-on continues in the selected language.
Note
If you select French, HR/PR alarm limits cannot be turned off unless you go
through the service menus to change this setting. If you do change this setting,
the change stays in effect through power cycles.
For information about using the service menus, refer to the Propaq LT Monitor Service Manual (810-1811-XX).
The monitor always powers on in this language until the setting is changed again. The
setting can be changed again only if one of the following occurs:
•This procedure is repeated.
•A new configuration is downloaded from a PC. (See “Using the Configuration
Utility to configure the monitor” on page 114).
•A new configuration is downloaded from Acuity.
About the charging/communications cradle
When the charging/communication cradle (the cradle) is attached to AC power (or to
vehicle DC power) and the monitor is seated in the cradle, the cradle does the following:
•Recharges the monitor battery, whether patient monitoring is occurring or not.
•Powers the monitor, conserving the charge on the monitor battery.
•(If the cradle has the USB data transfer option) enables data transfer between a
PC and a monitor. See “About the USB data transfer option” on page 15.
Page 19
Directions for UseOverview of monitor operation15
14040/7878
80
12
97
%
HALL, RO BERT E.
3456187
3:00:06P
Adult ,
Rm 23 9
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
WARNING Never download a configuration file to the monitor while you are
monitoring a patient with the monitor in the cradle.
•Communication with the network (and Acuity) is interrupted if you download
a configuration to the monitor or upload patient data from the monitor.
•Changing the monitor configuration shuts down the monitor and deletes all
patient data from the monitor.
Figure 9. Monitor and cradle
Monitor
HALL, ROBERT E.
3456187
3:00:06P
Adult,
II 1mV/cm
SpO2
HR/min
Rm 239
12
80
NIBP mmHg (102)
@2:47P Manual
Resp/min
97
SpO2
Cradle
The cradle is intended to hold the monitor when the patient is in bed or is other wise
stationary. The cradle can sit on a flat surface near the patient or attach to the bed rail
(using the bed-rail hook).
If the monitor is in the cradle and the AC power adapter is connected to the cradle, the
monitor runs on AC power rather than battery power. This keeps the monitor battery at full
charge so that the monitor can then run on battery power when it is removed from the
cradle to accompany the patient away from the bed.
The cradle can be connected to AC power at all times, whether the monitor is present or
not.
About the USB data transfer option
A cradle configured with the optional USB communication port and connected to a PC
with the Propaq LT Configuration Utility and the Propaq LT AutoPrint Utility can be used
for the following:
•Creating custom monitor configurations on the PC and downloading them to any
number of monitors. (See “Monitor configuration” on page 101.)
•Uploading the configuration file from the monitor to the PC.
•Uploading patient data from the monitor to the PC for printing. (See “Printing
patient data” on page 95.)
Setting up the cradle
1.Place the cradle on a table or shelf, or hang it on a bed rail.
WARNING Place the cradle so that it cannot fall on the patient.
Caution Locate the cradle near the patient but not so close that it interferes
with patient care.
Page 20
16Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
14040/7878
80
12
97
%
HALL, ROBERT E .
345618 7
3:00:0 6P
Adul t,
Rm 2 39
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
14040/7878
80
12
97
%
HALL, ROBERT E .
345618 7
3:00:0 6P
Adul t,
Rm 2 39
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
2. Connect the AC adapter to an AC power outlet and to the cradle. The indicator
(green) on the front of the cradle indicates that the AC power adapter is connected.
Figure 10. Cradle status indicators
HALL, ROBERT E.
II 1mV/cm
SpO2
80
HR/min
NIBP mmHg (102)
@2:47P Manual
3456187
3:00:06P
Adult,
Rm 239
12
97
Resp/min
SpO2
3. Insert the monitor into the cradle. When the monitor is properly seated, it clicks into
place. (green) on the cradle indicates that the monitor battery is charging. (See
“Recharging the battery” on page 137.)
Note
The monitor can be inserted into the cradle when power is on or off. If monitor
power is on, inserting the monitor in the cradle or removing the monitor from the
cradle does not interrupt patient monitoring.
Removing the monitor from the cradle
To remove the monitor, depress the release button on the front of the cradle. With the
button depressed, hold the cradle securely with one hand, grasp the monitor firmly with
the other hand, and lift the monitor out.
Green Monitor battery charging
Yellow Battery or charger failed
(Indicates either that the temperature of
the battery pack is out of range or that
service is required.)
Green AC power connected
Figure 11. Monitor release button
HALL, ROBERT E.
II 1mV/cm
SpO2
80
HR/min
NIBP mmHg (102)
@2:47P Manual
3456187
3:00:06P
Adult,
Rm 239
12
97
Resp/min
SpO2
Monitor release button
Page 21
Directions for UseOverview of monitor operation17
140/78
80
12
97
%
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
Displaying data
Patient vital signs appear on the 3.5-inch (diagonal measurement) monitor display and
optionally on the Large Color Display screen.
Figure 12. Components of the vital-signs display (A)
1
23456
HALL, ROBERT E.
II 1mV/cm
SpO2
80
HR/min
0
Patient name (if available)
1
Waveform source
2
Waveform display scale
3
Patient ID (if available)
4
Connection status icon
5
3456187
Adult
140/78
NIBP mmHg (102)
@2:47P Manual
(blank)Connectivity not enabled.
The monitor is communicating via USB cable with a PC. (Wireless monitor only, Acuity
enabled.)
The monitor is communicating with the network and with Acuity.
(Wireless monitor only, Acuity enabled.)
Rm 239
12
Resp/min
3:00:06P
7
8
97
SpO2
Time of day
6
Patient room number
7
Patient mode
8
Flashing — the monitor is communicating with the network but not with Acuity.
(Wireless monitor only, Acuity enabled.)
Steady — the monitor is not communicating with the network.
(Wireless monitor only, Acuity enabled).
Page 22
18Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
140/78
80
12
97
%
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
Figure 13. Components of the vital-signs display (B)
HALL, ROBERT E.
II 1mV/cm
SpO2
80
HR/min
9
140/78
NIBP mmHg (102)
@2:47P Manual
15141312
Battery status (See Table 3, “Status indicators” on page 2.)
greenpartially to fully charged
yellownearly discharged
reddischarged; the monitor will shut down soon
Adult
3456187
Rm 239
12
Resp/min
3:00:06P
9
10
97
SpO2
11
Color-coded waveforms, ECG Lead identifiers, and vital-signs numerics
10
Green
Cyan
Purple
Yellow
SpO2 pulse amplitude
11
SpO2 numeric data
12
Respiration rate numeric data
13
NIBP numeric data
14
HR (heart rate) is displayed if ECG is active.
15
PR (pulse rate) is displayed if ECG is not active and SpO
The monitor indicates an HR/PR measurement outside the measurable range as follows:
- - -out-of-range low
+ + +out
? ? ?undetermined
ECG and HR/PR
NIBP
Resp
SpO
2
-of-range high
or NIBP is active.
2
Page 23
Directions for UseOverview of monitor operation19
80
97
140/
78
12
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
NIBP mmHg (102) Manual
2:47P
Resp/min
HR/min
SpO2
%
SDM
%
140/78
80
12
97
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
II 1mV/cm
%
12:41
12:40
12:39
12:38
12:37
12:38
12:36
12:35
12:34
12:33
12:32
12:31
125
122
100
75
50
25
100
50
25
100
130/65 (93)
112/87 (87)
192/110 (130)
152/78 (115)
152/78 (115)
22
18
16
17
19
19
19
19
19
19
17
98
98
99
99
98
100
98
98
98
100
152/78 (115)
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
12
SpO2
SEARCH
HALL, ROBERT E.
3456187
12:41:32
Adult
Rm 239
Tabular
Time
140/78
HR/min
NIBP mmHg
Resp/min
Off
140/78
80
12
97
%
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
II 1mV/cm
%
12:41
12:40
12:39
12:38
12:37
12:38
12:36
125
122
100
75
50
25
130/65 (93)
112/87 (87)
192/110 (130)
n/a
n/a
n/a
n/a
22
18
16
17
19
19
98
98
99
99
98
100
n/a
12
SpO2
SEARCH
Tabular
Time
140/78
HR/min
NIBP mmHg
Resp/min
n/a
HALL, ROBERT E.
3456187
12:41:32
Adult
Rm 239
About display formats
The monitor can be configured to display any of these formats:
Blood pressure, heart rate, respiration rate, and
SpO
Bell icons indicating...
Single waveform
•3 seconds of one ECG or SpO
•12 seconds of the Resp signal.
Vital signs displayed in medium-sized numerics.
Tabular trends
98
99
98
99
Current vital signs displayed above the table in small
numerics.
Historical vital signs displayed in a table.
98
98
98
98
measurements displayed in large numerics.
2
Left halfLower alarm limit
Right halfUpper alarm limit
WhiteAlarms enabled
BlackAlarms disabled
signal or ...
2
HALL, ROBERT E.
II 1mV/cm
SpO2
140/78
80
HR/min
NIBP mmHg (102)
@2:47P Manual
HALL, ROBERT E.
II 1mV/cm
Tabular
Time
HR/min
12:41
125
12:40
122
12:39
100
12:38
12:38
12:37
12:36
Adult
Adult
8080140/78
NIBP mmHg
130/65 (93)
n/a
112/87 (87)
75
n/a
n/a
50
25
192/110 (130)
n/a
3456187
Rm 239
12
Resp/min
3456187
Rm 239
Resp/min
3:00:06P
Dual waveform
•6 seconds of an ECG or SpO
signal or 24
2
seconds of the Resp signal or...
•Any two of the following: 3 seconds of an ECG
97
SpO2
or SpO
signal.
Other vital signs displayed numerically below the
signal and 12 seconds of the Resp
2
waveforms.
12:41:32
12
SEARCH
n/a
22
18
16
17
19
19
Tabular trends with single waveform
•3 seconds of an ECG or SpO
•12 seconds of the Resp signal.
SpO2
98
99
98
99
100
n/a
98
Other vital signs displayed numerically below the
waveforms.
Historical vital signs displayed in a table below the
waveform.
signal or...
2
Page 24
20Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
80
97
140/
78
12
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
NIBP mmHg (102) Manual
2:47P
Resp/min
HR/min
SpO2
%
SD M
%
140/78
80
12
97
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
II 1mV/cm
%
12:41
12:40
12:39
12:38
12:37
12:38
12:36
12:35
12:34
12:33
12:32
12:31
125
122
100
75
50
25
100
50
25
100
130/65 (93)
112/87 (87)
192/110 (130)
152/78 (115)
152/78 (115)
22
18
16
17
19
19
19
19
19
19
17
98
98
99
99
98
100
98
98
98
100
152/78 (115)
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
12
SpO2
SEARCH
HALL, ROBERT E.
3456187
12:41:32
Adult
Rm 239
Tabular
Time
140/78
HR/min
NIBP mmHg
Resp/min
Off
The monitor can be configured to cycle quickly through three of the five available display
formats when you press. For example, if the configuration specifies three formats—
Large Numerics, Single Waveform, and Ta b ul a r Tre n d s—you can press repeatedly
to cycle through those formats.
Figure 14. Cycling through the configured display formats
Certain properties of each display type can also be configured. See “Display Format” on
page 127.
To view a tabular display if no tabular trends format is specified
1.Access the Setup menu. (See “To access the setup menus” on page 35.)
2. Highlight Trends and press.
Note
All valid display configurations include at least one waveform format.
Timing out the display and the back light
To conserve battery run life, the monitor display and the back light can be configured to
turn off if no operator activity (that is, a button press) is detected for a specified number of
minutes.
•When the display is off, no patient data is visible.
•When the back light is off, patient data is visible only under direct light.
If the display and the back light are turned off due to a time out, they turn on again
immediately when an alarm or alert occurs or, if the buttons lock-out is not
enabled, when you press any monitor button.
Note
The monitor configuration determines whether the time-out feature is enabled or
disabled, and defines the default time-out period if it is enabled. If the feature is enabled in
the configuration, you can temporarily change the time-out period or disable the time out
through the Setup -> Timings menu.
Page 25
Directions for UseOverview of monitor operation21
Locking out the display, back light, and buttons
If lock-outs are enabled, you can lock out one or more of the following elements:
ButtonsTo prevent unauthorized use
Back Light To extend the battery run life
DisplayTo prevent unauthorized viewing of patient information
The monitor configuration defines which of these, if any, can be locked out. For
information on enabling or disabling lock-outs for these elements, see “Monitor
configuration” on page 101.
To lock out the configured elements
Hold down , , and simultaneously for 5 seconds.
To unlock the configured elements
Hold down , , and simultaneously for 5 seconds.
Note
Alarms and alerts immediately unlock any locked elements.
Large color display interface
Using the Large Color Display Interface (‘interface box’) optional accessory, you can
display patient vital signs from the monitor on a full-size screen. This screen is typically
mounted on a wall within one or two meters of the patient’s bed. It receives video signals
from the cradle via a VGA cable.
Large color display configuration options
•If the interface box is mounted on the large screen (Figure 15), the screen must be
within about 3 meters of the cradle.
•If the interface box is mounted on the cradle (Figure 16), the distance between the
box and the screen is limited by the length of the VGA cable.
•If the interface box, the cradle, and the large screen are mounted together (Figure 17),
the distance from them to the patient’s bed is limited by the length of the SpO
sensor cable, the ECG cable, and the NIBP air hose (whichever is shortest).
Note
Note
The various configurations are not interchangeable. Each requires a unique
combination of screws, USB and power cables, and VGA cable.
If the interface box is attached to the cradle, the bed rail hook cannot be used.
2
Page 26
22Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Figure 15. Interface box mounted on the large screen
Interface box mounted on the large screen
Figure 16. Interface box mounted on the cradle
Interface box mounted
on the cradle
Page 27
Directions for UseOverview of monitor operation23
15
%
60
78
98
120
S
D
M
220
75
110
35
120
50
Manual
120
50
100
90
120
50
(89)
STEWART, AN N
HR/min
NIBP/PSNI-mmHg
Resp/min
SpO2
II
SpO2
Resp
7762940
V
I
Resp
V
II
III
aVR
aVL
aVF
SpO2
15
%
60
78
98
120
S
D
M
220
75
110
35
120
50
Manual
120
50
100
90
120
50
(89)
HR/min
NIBP/PSNI-mmHg
Resp/min
SpO2
STEWART, AN N
7762940
Figure 17. Interface box and cradle mounted on the large screen
Interface box and cradle
mounted on the large screen
Large display viewing options
The large display shows vital signs numerics and either four or nine waveforms.
Figure 18. Large color display viewing options
STEWART, ANN
II
SpO2
Resp
To change between 4-waveform and 9-waveform views, insert a wire into the small hole
in the interface box near the USB connector.
If you change the display format at least 15 seconds before shutting off power to the
interface box, the newly selected display becomes the default display when the interface
box is next powered on.
7762940
HR/min
60
NIBP/PSNI-mmHg
120
78
SpO2
98
Resp/min
15
Manual
(89)
100
90
STEWART, ANN
120
50
220
75
110
35
120
50
120
50
II
III
aVR
aVL
aVF
SpO2
Resp
7762940
HR/min
60
NIBP/PSNI-mmHg
120
78
SpO2
98
Resp/min
15
Manual
(89)
100
90
120
50
220
75
110
35
120
50
120
50
Page 28
24Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Figure 19. Switching between the 4-waveform and 9-waveform views
Waveform view selector
Installing the interface box
The Large Color Display Interface receives DC power from the AC power adapter and
vital-signs data from the cradle. It provides DC power to the cradle and the video signal to
the large display.
Figure 20. Interface box connections
DC power out
(to cradle)
USB in
(from cradle)
Fuse
VGA out to
(large color display)
DC power in
(from AC power adapter)
DC power indicator
When the interface box is attached to the cradle, the default power and USB cables
(references A and B in Table 6, “Cables and AC power adapter”) connect the interface box
and the cradle. (See Figure 17 on page 23.)
If the interface box and the cradle are not attached, longer cables—references C and D in
Table 6, “Cables and AC power adapter”—are used.
The AC power adapter (reference E) is shipped with the cradle.
Note
The AC power adapter used with the Propaq CS monitor or the Propaq Encore
monitor will not power the interface box.
When the interface box is not used, the AC power adapter is attached to the cradle. When
the interface box is used, the AC power adapter is connected to it and not to the cradle.
Page 29
Directions for UseOverview of monitor operation25
Table 6. Cables and AC power adapter
RefPart
Number
A008-0949-00 Power cord, 0.5-meter, box-to-cradle
B008-0947-00 USB cable, 1-foot
C008-0948-00 Power cord, 5-meter, box-to-cradle
D008-0946-00 USB cable, 15-foot
E503-0142-00 AC Power adapter
008-0908-00 AC Power cord, Japan
008-0910-00 AC Power cord, Australia
008-0911-00 AC Power cord, United Kingdom
008-0912-00 AC Power cord, Europe
008-0913-00 AC Power cord, North America
Item Description
Detaching the bed rail hook from the cradle
If you are attaching the interface box to the cradle and if the cradle and the bed rail hook
are attached, you must first detach the bed rail hook from the cradle.
1.Remove the screws from the back of the bed rail hook.
2. Lift the bed rail hook off of the shoulder screws on the back of the cradle.
3. Remove the shoulder screws from the back of the cradle.
Figure 21. Detaching the bed rail hook from the cradle
2
1
3
Page 30
26Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Assembling the large color display
Refer to Figure 22, Figure 23 on page 27, or Figure 24 on page 28.
Figure 22. Mounting the interface box (with face plate) on the large display mounting bracket
60-mm screws
620-0431-00
40-mm screws
620-0432-00
Long cables between box and cradle:
008-0946-00 (USB)
008-0948-00 (Power)
Page 31
Directions for UseOverview of monitor operation27
Figure 23. Mounting the interface box on the cradle
40-mm screws
620-0432-00
18-mm screws
620-0433-00
Short cables between box and cradle:
008-0947-00 (USB)
008-0949-00 (AC Power)
Page 32
28Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Figure 24. Mounting the interface box and the cradle on the large display mounting bracket
60-mm screws
620-0431-00
40-mm screws
620-0432-00
Short cables between box and cradle:
008-0947-00 (USB)
008-0949-00 (AC Power)
Page 33
Directions for UseOverview of monitor operation29
About navigation
You navigate the monitor screens using , , and (arrow buttons), (action button),
and (display button).
Using the arrow buttons
Use , , and to do the following:
•Highlight an item on the display. (See “Using the highlights” on page 29.)
•Select options from a control menu.
•Use and to select options from a pop-up menu.
•Use and to change the values of numeric parameters.
Using the action button
Use to do the following:
•Display the control menu for a blue-highlighted item.
•Return from a control menu to the primary display.
•Access the Setup menu when Setup is highlighted.
•Display tabular and graphical trends when Trends is highlighted.
•Display snapshots when Snapshot is highlighted.
•Turn on the display or the back light if either has been turned off by a time-out.
•Display a pop-up menu.
Using the display button
Use to do the following:
•Cycle through the configured display formats.
•Return from a control menu to the primary display.
•Close a pop-up menu.
Using the highlights
Every screen contains a single element—the current context—highlighted by a blue field.
Some screens also contain elements—parameter values—highlighted by a green field.
About blue highlights
A blue highlight identifies the current context. For example, Figure 25 illustrates a
highlighted row in a trends display and a highlighted setting in the Waveform Size menu.
Page 34
30Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
II 1mV/cm
%
12:41
12:40
12:39
12:38
12:37
12:38
12:36
125
122
100
75
50
25
130/65 (93)
112/87 (87)
192/110 (130)
n/a
n/a
n/a
n/a
22
18
16
17
19
19
98
98
99
99
98
100
n/a
12
SpO2
SEARCH
Tabular
Time
140/78
HR/min
NIBP mmHg
Resp/min
n/a
HALL, ROBERT E.
3456187
12:41:32
Adult
Rm 239
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
140/78
80
12
97
%
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
Figure 25. Examples of highlighted elements
HALL, ROBERT E.
II 1mV/cm
Tabular
Time
12:41
12:40
12:39
12:38
12:38
12:37
12:36
8080140/78
HR/min
125
122
100
75
50
25
3456187
Adult
NIBP mmHg
130/65 (93)
n/a
112/87 (87)
n/a
n/a
192/110 (130)
n/a
Rm 239
12
Resp/min
n/a
22
18
16
17
19
19
12:41:32
SEARCH
SpO2
100
n/a
Drop-down menu
Highlights
98
99
98
99
98
Vital-signs display
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
In a display screen (see “About display formats” on page 19), pressing causes the
monitor to replace the current screen with another screen related to the current context.
For example, if SpO2 is highlighted in the Two waveforms display and you press...
Figure 26. Using the action button ()
HALL, ROBERT E.
II 1mV/cm
SpO2
140/78
80
HR/min
NIBP mmHg (102)
@2:47P Manual
Adult
3456187
Rm 239
12
Resp/min
3:00:06P
97
SpO2
(Two waveforms display)
With SpO2 highlighted, press
...the monitor presents the SpO
About green highlights
Green highlights identify the current values of parameters within a given context. For
example, in the control menu shown in Figure 27, the current settings of the SpO
parameters are highlighted in green.
A control menu includes a topic name for the current context (for example, SpO2); a
column of parameters with one highlighted (for example, SpO2 Monitoring); and a
column of options, with one item in each set of options highlighted (for example, On, On, 100, On, 90, Low).
•The blue highlight indicates the parameter currently enabled for modification.
•The green highlights indicate the current settings for all parameters in the menu.
At the bottom of the screen for all control menus are links to Exit, Trends, Snapshots,
and Setup.
Example: Using a control menu
ExitReturn to the vital-signs display.
TrendsView a tabular history.
SnapshotsView a series of 21-second waveform snapshots of the current patient’s
vital signs.
SetupAccess the setup menu. (See “To access the setup menus” on page 35.)
Using the example (Figure 27), you would do the following to raise the SpO2 lower alarm
limit to 95 (Step 1) and shut off the HR/PR tone (Step 2):
1.W i t h SpO2 Monitoring highlighted, scroll (using ) to highlight Lower Limit, and
Note
press as many times as needed to raise this alarm limit to 95.
If you decrease an upper alarm limit to a value almost as low as the lower limit,
the lower limit decreases so that it is always lower than the upper limit.
If you increase a lower alarm limit to a value almost as high as the upper limit, the
2. Scroll (using ) to HR/PR Tone, and press either or as many times as needed to
upper limit increases so that it is always higher than the lower limit.
highlight Off.
Page 36
32Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Note
When you change a setting (for example, by turning off an alarm limit or by
increasing or decreasing an alarm limit), the change takes effect immediately.
3. Press or to exit the control screen and return to the vital-signs display.
Note
When you exit a control menu, the values displayed at the time you exit are the
values in effect for the monitor. If you change a parameter setting and then decide
before exiting the control menu to keep the previous setting values, you must
return the parameters to the original values before you exit the control menu.
About the HR/PR control menu
ParameterOptionsParameterOptions
Upper Alarm Off OnLower Alarm Off On
Upper LimitsLower Limits
Adult
Pediatric
Neonate
HR/PR Tone Off Low Med HighSelected Source ECG SpO
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt
Ann Jones, MD
503-530-0101 x9999
Patient Mode
Wireless Communications
Start New Patient
Adult
Snapshots
Enabled
Info
none saved
Demo
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt
Ann Jones, MD
503-530-0101 x9999
Patient Mode
Wireless Communications
Start New Patient
Welch Allyn Propaq 802LT0N
Serial# AB72383-1 V1.00.00
Adult
Snapshots
Disabled
Continue Patient
14 of 20 saved
Info
The start-up information screen provides the following:
•Monitor type (LTRN=wireless, LT0N=standalone), serial #, and software version #
(V X.XX.XX)
•Medical facility name and department
•Configuration file name
•Contact person’s name and telephone number
•Current patient mode (Adult, Pediatric, Neonate)
•Number of data snapshots saved (0 - 20)
•Network communications status (Enabled, Disabled)
Demo
To exit the start-up information screen, do one of the following:
•Highlight Continue Patient or Start New Patient to start monitoring.
•Highlight Demo to enter Demo mode.
•Press to turn off the monitor.
Page 41
Directions for UseOverview of monitor operation37
II 1mV/cm
Exit
Trends
Snapshots
Setup
80
HR/min
(100)
NIBP mmHg (148-86)
Resp/min
SpO2
503-530-0101 x9999
Emergency Department
PtldWstsdED10Jun05.mnt
Ann Jones, MD
Portland Westside
12
97
%
HALL, ROBERT E.
3456187
12:41:32
Adult
Rm 239
503-530-0101 x9999
Emergency Department
PtldWstsdED10Jun05.mnt*
Ann Jones, MD
Portland Westside
To view the monitoring information screen
Highlight (in the upper right corner of the vital-signs display) and press.
Figure 31. Monitoring information
HALL, ROBERT E.
II 1mV/cm
80
HR/min
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt
Ann Jones, MD
503-530-0101 x9999
Exit
The monitoring information screen provides the following information:
•Continuous numeric and waveform display of patient vital signs
•Medical facility name and unit
•Configuration file name
•Contact person’s name and telephone number
Note
If a monitor configuration is saved after a configuration file is downloaded, the file
name shown in the information screen is followed by an asterisk (*). This is true
even if the saved configuration is identical to the downloaded configuration.
(100)
NIBP mmHg (148-86)
Trends
3456187
Adult
Snapshots
Rm 239
12
Resp/min
12:41:32
97
SpO2
Setup
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt*
Ann Jones, MD
503-530-0101 x9999
For more information about changing the monitor configuration, see “Changing
the default settings” on page 67.
To exit the monitoring information screen, do one of the following:
•To return to the primary display, press , or highlight Exit and press.
•To see a tabular display of vital signs, highlight Trends and press.
•To view saved snapshots of vital signs, highlight Snapshots and press.
•To view the Setup menu, highlight Setup and press.
•To turn off the monitor, press.
Page 42
38Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Select ( ) to enter Demo Mode
No data saved.
Start New Patient
Info
Demo
(unplug patient cables)
Using demo mode
In Demo mode, the monitor displays simulated patient data for all vital signs. You can use
Demo mode to familiarize yourself with the following:
•Viewing vital-signs displays
•Modifying alarm limits and other settings
•Cycling through display formats
•Responding to alarm conditions
To start Demo mode, all of the following must be true:
•The power-on screen is displayed.
•The monitor is not in NIBP Auto mode.
•Patient data was deleted when the monitor was last shut down; that is, the
monitor contains no stored patient data.
•The SpO
To enter demo mode low
and ECG cables are not connected to the monitor.
2
1.Verify that the SpO
and ECG cables are not connected to the monitor.
2
2. Cycle the monitor off (deleting any saved data) and then on again.
3. When the main screen appears, highlight Demo and press.
Note
If you enter Demo mode with an SpO
or ECG cable connected, the monitor
2
enters Demo mode for only a second or two before shutting down and powering
up in monitor mode.
If you are in Demo mode and you connect an SpO
or ECG cable or select NIBP
2
auto mode, the monitor shuts down and powers up to the start-up screen.
Figure 32. Power-on screen: Demo highlighted
No data saved.
Select ( ) to enter Demo Mode
(unplug patient cables)
Start New Patient
Info
Demo
Demo mode is indicated by the message ‘SIMULATION’ in the upper left corner of
the screen. If the monitor is connected to Acuity, ‘SIMULATION’ also appears on the
Acuity display.
Page 43
Directions for UseOverview of monitor operation39
80
97
12
ID: 01018VDO9PBH
3:00:06P
Adult
Rm 239
NIBP mmHg
Resp/min
HR/min
SpO2
%
SIMULATION
SDM
II 1mV/cm
Exit
Trends
Snapshots
Setup
Low
Med
80
HR/min
NIBP mmHg
Resp/min
SpO2
OffOnOn
Alarm Tone
Suspend Audible Alarms
Setup
(90 sec)
12
97
%
ID: 01018VDO9PBH
3:00:06P
Adult
Rm 239
Alarms
ECG
NIBP
Timings
High
Service
Configuration
Figure 33. Demo mode: initial display
ID: 01018VDO9PBH
SIMULATION
80
HR/min
NIBP mmHg
When you enter Demo mode, the monitor is in ‘Demo mode low’. In this mode, the
simulated vital signs of the patient are steady and do not cause any alarms at the default
alarm limit settings. You can explore the monitor displays and menus, and you can change
the same settings and values in Demo mode that you can change in normal mode.
In Demo mode low, if you adjust the alarm limits to put the simulated patient’s vital signs
out of limits, the monitor simulates an alarm condition. Another way to simulate an alarm
condition is to switch the monitor to Demo mode high, which uses higher numeric values.
To switch to demo mode high
1.Access the Setup menu. (“To access the setup menus” on page 35.)
Adult
97
Rm 239
SpO2
3:00:06P
12
Resp/min
2. Press to highlight Timings.
Figure 34. Setup: timings
ID: 01018VDO9PBH
II 1mV/cm
80
HR/min
Setup
Suspend Audible Alarms
Alarm T one
Exit
Alarms
Service
NIBP mmHg
ECG
Trends
Adult
NIBP
Rm 239
12
Resp/min
Timings
Off
Low
Med
Snapshots
3:00:06P
97
SpO2
Configuration
(90 sec)
High
Setup
%
Tim ings
Default demo mode (Low)
Page 44
40Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Off
II 1mV/cm
Exit
Trends
Snapshots
Setup
80
HR/min
NIBP mmHg
Resp/min
SpO2
Demo Mode
Backlight Timeout (min)
Setup
12
97
%
ID: 01018VDO9PBH
3:00:06P
Adult
Rm 239
Alarms
ECG
NIBP
Timings
Service
Display Timeout (min)2
Low
5
Off2 5
SIMULATION
High
Configuration
80
97
12
ID: 01018VDO9PBH
3:00:06P
Adult
Rm 239
NIBP mmHg
Resp/min
HR/min
SpO2
%
SIMULATION
SDM
The Timings menu specifies the Demo mode—Demo mode low, which simulates
normal vital signs, and Demo mode high, which simulates a higher heart rate (HR),
higher respiration rate (Resp), and lower oxygen saturation (SpO
3. Press to highlight Demo mode, and press or to highlight High.
Figure 35. Setup: timings: demo mode high
).
2
ID: 01018VDO9PBH
SIMULATION
II 1mV/cm
80
HR/min
Setup
Backlight Timeout (min)
Display Timeout (min)
Demo Mode
Exit
4. Press or to return to the main display.
Figure 36. Demo mode high: initial display
ID: 01018VDO9PBH
SIMULATION
Alarms
Service
NIBP mmHg
ECG
Trends
80
HR/min
Adult
NIBP
Snapshots
Adult
Rm 239
12
Resp/min
Timings
101015153030OnOnOff
101015153030OnOnOff
Low
High
Rm 239
97
SpO2
3:00:06P
97
%
SpO2
Configuration
Setup
3:00:06P
Demo Mode: High
The monitor generates an alarm within seconds of reading this new set of simulated
vital signs. The red alarm indicator (Figure 37) illuminates and flashes, the numerics
for all violating vital signs—HR, SpO
moves to the most recent alarming numeric.
With the monitor simulating an alarm, you can practice responding to alarms. (See
“Responding to an alarm” on page 83.)
To exit demo mode
Press . The Demo Mode Power Off screen appears.
Figure 38. Demo mode power off
ID: 01018VDO9PBH
SIMULATION
You have pressed the Power Off button.
Select ( ) to delete patient data
Delete & Shut Down
Adult
and shut down.
Cancel
Rm 239
15:01:24
Setup
•To shut down the monitor, press. (No data can be saved from Demo mode.)
•To resume Demo mode, highlight Cancel and press.
•To access the Setup menu, highlight Setup and press.
Page 46
42Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Select ( ) to delete patient data
You have pressed the Power Off button.
Delete & Shut Down
Cancel
Setup
and shut down.
15:01:24
Adult
Rm 239
There is patient data stored for
Hall, Robert E. ID: 3456187
Save & Shut Down
HALL, ROBERT E.
3456187
Power saving
To maximize battery life, the monitor display shuts off when the following conditions are
all true:
•No button press for a period of n seconds. (n is configurable.) See “Timing out the
display and the back light” on page 20.
•No active alarms or alerts
•No Acuity Message window
•No Patient ID entry window
Turning off the monitor
To turn off the monitor
Press .
The Power-off screen appears:
Figure 39. Power-off screen
Cancel
15:01:24
Setup
HALL, ROBERT E.
You have pressed the Power Off button.
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete patient data
and shut down.
Delete & Shut Down
3456187
Adult
Save & Shut Down
Rm 239
Highlight the desired action and press.
Note
If you press to power down before leaving the power-on screen (Figure 7 on
page 13), the monitor shuts down without presenting the screen shown above.
Communicating with an Acuity Central Station
See “Monitoring in communication with Acuity” on page 71, and see Acuity Directions
for use.
About error detection
The monitor can detect conditions that prevent it from operating properly. If this occurs, it
displays an error message and an error number. Follow the directions displayed on the
screen.
Page 47
Directions for UseOverview of monitor operation43
Transporting the monitor with the patient
An ambulatory patient can wear or carry the monitor using the wearable strap or the
patient carry strap (optional accessories). Medical personnel can use the transport
stretcher carry strap to keep the monitor with a patient during stretcher transport.
WARNING When the patient is wearing or carrying the monitor, carefully route
any patient cabling to reduce the possibility of patient entanglement or
strangulation. Use the supplied garment clips to secure the cable properly.
WARNING When positioning straps on the patient, make sure the straps do not
entangle the patient’s neck or cause choking.
WARNING Make sure the straps do not restrict the movement of the patient’s
limbs or create a hazard for the patient when the patient is walking or moving.
WARNING Never use a strap to carry or pick up both the monitor and the cradle.
The straps are not intended to support, and cannot support, the combined weight
of the monitor and the cradle.
To attach the wearable strap
WARNING Do not put the wearable strap on the patient while the patient is in bed. The intended use of the wearable strap is to keep the monitor—without the
cradle—with the patient when the patient is ambulatory.
Figure 40. Wearable strap
1.Place the wearable strap on the sitting or standing patient and adjust all components
for a comfortable, secure fit (Figure 40 left).
2. Connect the wearable strap securely to the monitor strap mounts (Figure 40 right).
3. Carefully arrange the strap and the monitor on the patient to avoid bruising or other
skin injuries.
Page 48
44Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
To use the patient carry strap
WARNING Do not use the patient carry strap to lift or carry both the monitor and
the cradle. The patient carry strap is not intended to support (and cannot support)
the weight of both the monitor and the cradle. Attempting to carry both the
monitor and the cradle with a patient carry strap could lead to patient injury and to
damage to the monitor and the cradle.
1.Remove the monitor from the cradle.
2. Detach the monitor cables from any accessories.
3. Verify that all cables are disentangled from the bed and any bedside tables.
4. Connect the ends of the carry strap to the strap mounts on the monitor.
Note
Be sure that each end of the carry strap snaps into place.
Figure 41. Patient carry strap
To use the transport stretcher carry strap
1.With the monitor facing away from the stretcher and the patient, attach one end of
the strap to a monitor strap mount.
2. Run the monitor strap under the stretcher restraint straps, near the patient’s waist.
3. Attach the other end of the strap to the other strap mount.
Figure 42. Monitor secured to the patient during stretcher transport
4. Before transporting the patient, verify that all monitor cables are clear.
Page 49
45
3
Standalone monitoring
Overview
The monitor is available in standalone models (802LT0N and 802LT0S) and in wireless
models (802LTAN, 802LTAS, 802LTRN and 802LTRS).
This chapter describes the operation of the standalone monitor and of the wireless
monitor when it is not in communication with the Welch Allyn Flexnet
®
Acuity
Central Station.
About the Model 802LT0N and 802LT0S (standalone) monitors
The model 802LT0N and 802LT0S monitors operate in standalone mode. They measure
and display vital signs, store patient data, and locally indicate alarms and alert conditions.
•The configuration of the monitor is not affected by Acuity.
•The vital signs stored by the monitor are not sent to Acuity.
•Alarms and alerts generated by the monitor do not appear on Acuity.
•Alarms and alerts generated by Acuity do not appear on the monitor.
•Acuity does not update the monitor time and date settings.
®
network and an
About the Model 802LTAN, 802LTAS, 802LTRN and 802LTRS (wireless)
monitors
See “Monitoring in communication with Acuity” on page 71.
Preparing for a new patient
To begin monitoring a new patient
1.If the monitor is on, press to turn it off.
If prompted to save or delete existing data, highlight Delete and press. The monitor
deletes any saved data and temporary settings, and then shuts down.
2. Press to turn on the monitor.
•If the monitor holds stored data from the previously monitored patient, it displays
•If the data from the previously monitored patient was deleted on shut-down, the
“no data saved” start-up screen appears:
Figure 44. Start-up with no saved data
Portland Westside Hospital
Emergency Department
No data saved.
Select ( ) to Start a New Patient.
Start New Patient
Info
Demo
3. Highlight Start New Patient and press. The first configured data display appears.
In place of a patient name, ID: is followed by a string that is generated by the monitor
when you select ‘Start New Patient’. This AutoID string, unique to each new patient,
identifies the patient until you supply appropriate patient identification data.
Note
To bypass patient ID entry, press .
Figure 45. Data display with AutoID
ID: 0101819RO17Q
HR/min
Adult
SpO2
8:82:56
Start-up AutoID
NIBP mmHg
4. In the primary data display, highlight ID: (upper left) and press.
The Patient Information Entry screen appears:
Resp/min
Page 51
Directions for UseStandalone monitoring47
Last
Middle
First
ID#
Rm#
A B C D E F G H I J K L M N O P Q R S T U V
W X Y Z 1 2 3 4 5 6 7 8 9 0 -
Select ( ) to enter highlighted character
Confirm
Cancel
Backspace
Last
Middle
First
ID#
Rm#
A B C D E F G H I J K L M N O P Q R S T U V
W X Y Z 1 2 3 4 5 6 7 8 9 0 -
Select ( ) to enter highlighted character
Confirm
Cancel
HALLERORO
Backspace
3456187
Figure 46. Patient information entry
Last
Middle
First
Select ( ) to enter highlighted character
Confirm
ID#
Backspace
Cancel
Rm#
Entry fields
Character field
5. Enter the ID and name of the new patient.
a. Press or to highlight (blue) a character in the character field (Figure 47).
b. Press to copy it to the green-highlighted location in the entry fields (Figure 46).
c. Repeat from step a until all characters are entered into the field.
d. Press until the insertion point in the entry field changes from green to blue.
e. Press or to move the insertion point to another entry field.
f.Press to return to the character field.
g. Repeat from step a until all fields are complete.
Note
To correct an error in an entry field: Place the cursor (step d and step e) to the
right of the error location, highlight and enter Backspace in the character field to
delete the erroneous character, and then enter the correct character.
Figure 47. Patient information entry (continued)
Last
Middle
HALL
Select ( ) to enter highlighted character
Confirm
First
ID#
3456187
Backspace
Cancel
Rm#
Next letter highlighted
6. Highlight Confirm and press.
\
Note
Name alone is not sufficient to confirm a patient ID; thus, you can confirm the
patient name only after you have entered the patient ID.
7.Verify that the patient mode (adult, pediatric, or neonate) is set correctly for this
patient.
NeonatalTerm birth through 28 days, or up to 44 gestational weeks.
PediatricBetween 29 days and 12 years.
Adult13 years and older.
If the current patient mode setting is not correct:
a. Highlight the current patient mode (Adult, Pediatric, or Neonate) and press.
b. From the Patient Mode selection menu, highlight the appropriate patient mode
Note! You are changing the
patient mode from Adult to Pediatric.
This will change settings to default
Pediatric values and delete previous
If this is correct, highlight “Confirm”
then press Select ( ).
Cancel
Confirm
patient data.
II 1mV/cm
Year
Date Format
HR/min
NIBP mmHg
Resp/min
SpO2
%
14:49:37
Adult
Time Format
Hour
Minute
Month
Day
HALL, ROBERT E.
3456187
Rm 239
12 Hour
24 Hour
14
45
3
17
2005
mm/dd/yy
dd.mm.yy
yy/mm/dd
Set
Cancel
Figure 48. Changing patient mode
Patient Mode
Adult
Pediatric
Neonate
Patient Mode
Adult
Pediatric
Neonate
c. A confirmation screen appears:
Figure 49. Confirming a change of patient mode
HALL, ROBERT E.
Note! You are changing the
patient mode from Adult to Pediatric.
This will change settings to default
Pediatric values and delete previous
HR/min
patient data.
If this is correct, highlight “Confirm”
then press Select ( ).
Cancel
NIBP mmHg
Adult
3456187
Rm 239
SpO2
Confirm
3:00:06P
Resp/min
d. To confirm the change, highlight Confirm and press.
Note
When you change the patient mode and confirm the change:
•All vital-signs data for the patient is lost.
•All monitor settings revert to the defaults for the new patient mode.
8. Verify that the displayed time and date are correct.
•The monitor displays time in the 12-hour (AM/PM) or 24-hour format, and displays
the date in the mm/dd/yy, dd.mm.yy, or yy/mm/dd format.
•The date does not appear on the primary display screens. It appears on the
snapshot list and on the snapshot display.
a. Highlight the time display (in the upper right corner of the screen) and press.
The Time/Date screen appears.
Figure 50. Time/date screen: 24-hour format
Rm 239
Cancel
14:49:37
SpO2
24-hour format
yy/mm/dd
HALL, ROBERT E.
II 1mV/cm
HR/min
Time Format
Hour
Minute
Month
Day
Year
Date Format
NIBP mmHg
Set
12 Hour
14
45
17
2005
mm/dd/yy
Adult
3456187
Resp/min
24 Hour
dd.mm.yy
b. If the displayed time or date is not correct, press or to move the highlight
from one parameter to another, and press or to change the value of the
highlighted parameter. For example, to change the time display format from
24-hour to 12-hour, highlight Time Format and press either or once.
Page 53
Directions for UseStandalone monitoring49
II 1mV/cm
Year
Date Format
HR/min
NIBP mmHg
Resp/min
SpO2
%
14:49:54
Adult
Time Format
Hour
Minute
Month
Day
HALL, ROBERT E.
3456187
Rm 239
12 Hour
24 Hour
2P
45
3
17
2005
mm/dd/yy
dd.mm.yy
yy/mm/dd
Set
Cancel
Select ( ) to delete data
and start a new patient.
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Start New Patient
Continue Patient
Info
Demo
Figure 51. Time/date screen: 12-hour format
HALL, ROBERT E.
II 1mV/cm
HR/min
Time Format
Hour
Minute
Month
Day
Year
Date Format
NIBP mmHg
Set
12 Hour
2P
45
17
2005
mm/dd/yy
Adult
3456187
Resp/min
24 Hour
dd.mm.yy
Rm 239
Cancel
14:49:54
SpO2
12-hour format
yy/mm/dd
c. When the time and date are correct and formatted appropriately, press to
accept the changes and return to the vital-signs display.
Note
If you change the time or date settings and then decide not to accept the
changes, press to cancel the changes and return to the vital-signs display.
Continuing to monitor a patient on power-up
If patient data was saved when the monitor was last turned off (see “To turn off the
monitor” on page 42), you can resume monitoring that patient when the monitor is turned
on again. (When patient data is saved, the monitor settings are also saved.)
To resume monitoring the same patient:
1.Turn on the monitor. The “patient data stored” screen appears:
Figure 52. Start-up with saved data
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete data
Start New Patient
and start a new patient.
Continue Patient
Info
Demo
2. Verify that the displayed name and ID match that of the current patient.
In this manual, lead refers to an ECG electrode or lead wire, and Lead refers to a
waveform source.
You can monitor heart signs (ECG) and respiration rate (Resp) using either a 3-lead or a
5-lead ECG cable. Using a 3-lead cable, you can display one signal waveform for lead I, II,
or III. Using a 5-lead cable, you can display either one or two signal waveforms (I, II, III, V;
and if enabled in the configuration, aV
, aVL, or aVF). You can also display the SpO2 or
R
Resp waveform in place of an ECG waveform.
WARNING Always monitor and set alarms for SpO2 when using impedance
pneumography to monitor respiratory function.
WARNING When monitoring respiration via impedance pneumography, always
select the ECG Lead with the most prominent QRS complex. The monitor rejects
cardiovascular artifact, but this function depends upon accurate ECG R-wave
detection.
WARNING Do not place the monitor near another respiration monitor. Resp
measurement frequencies can cause mutual interference.
WARNING Do not perform impedance pneumography on paced patients.
Pacemaker pulses can sometimes be falsely counted as breaths.
WARNING Always keep patient motion to a minimum. Motion artifact can cause
incorrect readings of breath rate or heart rate.
WARNING If a disconnected lead is too close to other electrical devices, it can
cause a false heart rate, a false respiration rate, or a failure to display a “Lead Fail”
message.
WARNING The monitor displays + + + for HR numerics between 301-350 beats
per minute. For heart rates above 350 beats per minute, it might display
incorrectly low heart readings, due to intermittent picking of R-waves.
WARNING The monitor does not provide internal arrhythmia analysis; therefore,
arrhythmias can cause the monitor to display inaccurate heart rates.
Page 55
Directions for UseStandalone monitoring51
WARNING Motion artifact can cause the monitor to display inaccurate heart
rates. Minimize patient motion whenever possible.
WARNING (1) During a surgical procedure, do not use small ECG electrodes.
(2) Select ECG electrode attachment points remote from the surgical site and
remote from the electrosurgical return electrode. (3) Use electrosurgical return
electrodes with the largest practical contact area. (4) Assure proper application of
the electrosurgical return electrode to the patient.
High-intensity radio-frequency (RF) energy from external sources, such as an
improperly connected electrosurgical unit, can induce heat into electrodes
and cables, which can cause burns on the patient and can lead to
measurement errors.
WARNING Make sure the patient mode is correct. Incorrect patient mode can
result in inaccurate heart rate readings and inappropriate alarm settings.
WARNING Always use the provided garment clips to route ECG cables away
from the patient’s head.
WARNING Use only accessories listed in the Welch Allyn Products and
Accessories guide (810-0409-XX).
Note
WARNING Never use ECG cables with loose or faulty detachable lead wires.
These can cause erratic behavior of the ECG and respiration waveforms due to
intermittent ECG lead wire connections.
WARNING Resp is derived from the same leads as the ECG channel, so the
monitor determines which signals are cardiovascular artifact and which signals are
a result of respiratory effort. If the breath rate is within five per cent of the heart
rate or is a multiple or submultiple of the heart rate, the monitor might ignore
breaths and trigger a respiration alarm.
Caution Never use an ECG cable longer than 10 feet (3 meters) including
extensions. If you use an ECG extension cable with an ECG cable longer than
4 feet, the monitor acts as though no ECG cable is connected.
Caution To protect the monitor from damage during defibrillation or
electrosurgery, for accurate ECG information, and for protection against noise
and other interference, use only ECG electrodes and cables specified or supplied
by Welch Allyn (these cables have the required current-limiting resistors). Follow
recommended application procedures.
The monitor contains type CF fully isolated patient-connected circuitry, but it is
not intended for direct application on a patient’s heart.
Severe artifact and interference (such as defibrillation interference) can cause the
waveform to move off of the display for a few seconds before it is restored.
Impedance pneumography (Resp) is not recommended for use with
high-frequency ventilation.
The monitor counts as breaths any respiratory efforts larger than twice the
background cardiovascular artifact.
1.Inspect the ECG cable. Replace it if it shows any signs of wear, breakage, or fraying.
2. Plug the cable into the monitor.
3. Select electrode sites on the patient, choosing flat areas and avoiding fatty or bony
areas and major muscles.
Figure 54. ECG leads - actual placement
Use only silver/silver chloride electrodes. Other electrodes, such as stainless
steel electrodes, squeeze-bulb electrodes, or electrodes with dissimilar metals,
are subject to large offset potentials due to polarization. Other electrodes can also
have slower recovery time after the application of defibrillator pulses.
V1
RA
RL
LL
LA
V6
Six possible V lead electrode
placement sites for the C lead.
5-lead, adult and pediatric
RA
LL
3-lead, adult and pediatric
RA
LA
3-lead, neonatal
LA
LL
4. Shave or clip the hair from the electrode sites.
5. Thoroughly clean the skin, using soap and water, isopropyl alcohol, or skin preparation
pads, and lightly rub it dry.
Caution To protect the patient from allergic reactions to electrodes, refer to the
electrode manufacturer’s directions for use.
6. If you are using gelled electrodes, verify that the electrode expiration date has not
passed and that the gel is intact and not dried out.
If you are not using gelled electrodes, apply a mound of gel (1/4-inch to 1/2-inch, or
0.6-cm to 1.3-cm) to each electrode contact area.
Page 57
Directions for UseStandalone monitoring53
140/78
80
12
97
%
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
7.Attach lead wires to the electrodes before applying them to the patient.
8. Apply the electrodes to the patient in the proper locations (Figure 54).
Note
At least three appropriate electrode connections are required for ECG/Resp
monitoring.
Figure 55. ECG leads - placement displays, 5-lead and 3-lead
The locations of the circles displayed on the monitor (Figure 55) for each lead are
fixed, and do not indicate the exact placement of the electrodes on the patient.
If the monitor detects that some lead wires are not connected, it displays an “ECG
Fault” equipment alert and a chest diagram indicating the location of the
disconnected lead or leads. If the disconnected lead(s) invalidate the Lead used for
HR determination, then the monitor reassigns, if possible, the Lead used for HR. If
the reassignment succeeds, the monitor then displays another equipment alert with
the message “ECG Lead changed”.
Figure 56. ECG leads - disconnected leads
Disconnected leads
X
X
X
9. When all leads are properly connected, confirm that the monitor displays the ECG
waveform, heart rate, and other patient data.
Lead I
Lead II 6 sec
Lead III
aVR
aVL
aVF
V
Resp
SpO2
Add 2nd Wave
%
STEWART, ANN
7762940
7:05P
Adult
Rm 263
HR/min
Resp/min
SpO2
Resp 4x
60
20
97
130/74
NIBP mmHg (88)
@7:05P Manual
Waveform Source
Lead I
Lead II
Lead III
aVR
aVL
aVF
V
Resp 24 sec
SpO2
Add 2nd Wave
%
HR/min
NIBP mmHg
Resp/min
SpO2
II 1mV/cm
I 4mV/cm
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
80
12
97
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
To change the waveform selection
1.Highlight the current waveform source selection (Lead II, for example) and press .
Figure 58. Waveform source: II
STEWART, ANN
II 1mV/cm
Waveform Source
Lead I
Lead II 6 sec
Lead III
aVR
aVL
aVF
V
Resp
SpO2
80
Add 2nd Wave
HR/min
NIBP mmHg
7762940
Adult
12
Resp/min
Rm 263
15:14:28
97
%
SpO2
2. Highlight your waveform source choice and press or.
Figure 59. Waveform source: Resp
STEWART, ANN
Resp 4x
Waveform Source
Lead I
Lead II
Lead III
aVR
aVL
aVF
V
Resp 24 sec
SpO2
60
Add 2nd Wave
HR/min
NIBP mmHg (88)
@7:05P Manual
130/74
7762940
Adult
20
Resp/min
Rm 263
7:05P
97
%
SpO2
Note
If you scroll to the bottom selection of the Waveform Source menu—either Add
2nd Wave or Remove 2nd wave—the selection takes effect immediately and the
monitor returns to the primary data display.
Note
In the Waveform Source menu, either the source Lead or the waveform period
can be highlighted. If the waveform period is highlighted, a second trace of the
same source will be cascaded to double the period obtained from a single trace.
To change the waveform size
1.Highlight the current waveform scale (1mV/cm, for example) and press.
Figure 60. Waveform size popup menu
STEWART, ANN
II 1mV/cm
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
I 4mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
80
HR/min
NIBP mmHg
Adult
7762940
Rm 263
12
97
Resp/min
12:41:32
%
SpO2
Page 59
Directions for UseStandalone monitoring55
II 1mV/cm
Exit
Trends
Snapshots
Setup
60Hz
50Hz
80
HR/min
140/78
NIBP mmHg (102)
Resp/min
SpO2
OffOnOn
Power source filter
Resp Monitoring
Setup
12
97
%
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
Alarms
ECG
NIBP
Timings
Off
Configuration
ECG Bandwidth
Pacer indicator
Monitor
OffOnOn
Extended
Service
2. Highlight the desired scaling factor and press. (Waveform size does not affect
QRS-detector sensitivity.)
About pacemakers and ECG monitoring
If the patient being monitored has a pacemaker, the monitor detects and can indicate the
occurrence of pacemaker signals. If the Pacer Indicator setting is ON, the monitor displays
and prints vertical dashed lines to indicate detected pacemaker signals. If Pacer Indicator
is OFF, the monitor continues to detect the pacemaker signals but does not display or
print the pacer markers.
WARNING Signals differ between pacemakers. The Association for
Advancement of Medical Instrumentation (AAMI) cautions that “in some devices,
rate meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms.
All pacemaker patients should be kept under close or constant observation.”
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 61 on page 55.)
WARNING Always use fresh ECG electrodes and make sure the ECG cable lead
wires make good connections. The presence of pacer-like noise can cause the
displayed heart rate to be erratic even though the ECG trace might look
undistorted with the pacer indicator off.
To enable or disable the display of pacer indicators
1.Access the Setup menu. (See “To access the setup menus” on page 35.)
2. Highlight ECG.
Figure 61. Turning the pacer indicator off in the ECG setup menu
HALL, ROBERT E.
II 1mV/cm
80
HR/min
Setup
Resp Monitoring
ECG Bandwidth
Power source filter
Pacer indicator
Exit
Alarms
Service
140/78
NIBP mmHg (102)
ECG
Trends
Adult
NIBP
3456187
Rm 239
12
Resp/min
Timings
Off
Monitor
60Hz
50Hz
Off
Snapshots
3:00:06P
97
SpO2
Configuration
Extended
Off
Setup
%
Pacer indicator Off
3. Highlight Pacer indicator and press or to highlight Off or On.
4. To exit the Setup menu, press or, or highlight Exit and press.
If the pacemaker signal is strong enough, the monitor displays it as a waveform
spike. This is true with Pacer indicator ON or OFF.
In accordance with the Pacer Pulse Rejection specification (“Pacer pulse
rejection” on page 143.), pacemaker pulses are not counted as heartbeats
whether Pacer Indicator is On or Off.
Noise on the ECG signal might be detected as pacer signals, causing the pacer
indicator to appear on the display. If you do not need to indicate pacemaker
signals, turn off the pacemaker indicator for a better ECG waveform display.
Improving the waveform display
If the power source filter is off, noise from the power source can cause an unclear or
noisy waveform.
To change the power source filter to reduce noise
1.Access the Setup menu. (See “To access the setup menus” on page 35.)
2. Highlight ECG.
Verify that the setting for Power source filter is correct for the power source in your
facility. If you do not know what this setting should be, consult a qualified service
person.
3. Highlight Power source filter and press or as needed to select 60 Hz, 50 Hz,
or Off.
Figure 62. Turning on the 60-Hz power source filter in the ECG setup menu
HALL, ROBERT E.
II 1mV/cm
80
HR/min
Setup
Alarms
Service
Resp Monitoring
ECG Bandwidth
Power source filter
Pacer indicator
Exit
4. Exit the Setup menu by pressing or.
Monitoring respiration
Resp is based on impedance pneumography, where respirations are sensed from the
ECG electrodes.
All ECG cables listed for the Propaq LT monitor in Products and Accessories
(810-0409-XX) permit respiration monitoring and electrosurgical interference
suppression.
To measure Resp with Lead II selected, the LL lead must be attached to the
patient.
Note
Note
Adult
140/78
NIBP mmHg (102)
ECG
NIBP
Trends
3456187
Rm 239
12
Resp/min
Timings
Off
Monitor
60Hz
50Hz
Off
Snapshots
3:00:06P
97
SpO2
Configuration
Extended
Off
Setup
%
Power source filter 60 Hz
Page 61
Directions for UseStandalone monitoring57
%
HALL, ROBERT E.
3456187
3:08:16
Adult
Rm 239
HR/min
Resp/min
SpO2
Resp 2x2x
80
12
97
NIBP mmHg
%
HALL, ROBERT E.
3456187
3:08:16
Adult
Rm 239
HR/min
Resp/min
SpO2
Resp 8x8x
80
12
97
NIBP mmHg
Resp is part of ECG monitoring. The Resp numeric is displayed (in purple) in the lower
right corner. To view the Resp waveform:
Change the waveform source to Resp. (See “To change the waveform selection” on
page 54.)
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 61 on page 55.)
Figure 63. Resp waveform, default size (2x)
HALL, ROBERT E.
Resp
80
HR/min
For more detail, change the waveform size to 8x. (See “To change the waveform size”
on page 54.)
Note
Figure 64. Resp waveform enlarged for detail (8x)
Waveform size does not affect breath-detector sensitivity.
The monitor can be purchased with the Nellcor or the Masimo SpO2 option. Most of this
chapter pertains to either configuration. Monitors with the Masimo technology, however,
can provide additional SpO
WARNING Accurate measurements of oxygen saturation, when derived from
pulse oximetry, depend to a great extent on patient condition and proper sensor
placement. Patient conditions such as shivering and smoke inhalation can result in
erroneous readings. If you believe a measurement might be inaccurate, verify it
using another clinically accepted measurement method, such as arterial blood gas
measurements using a co-oximeter.
WARNING Use only SpO
Accessories guide (810-0409-XX). Inspect sensors and cables, and discard any
that are damaged. Do not use a sensor with exposed optical components.
WARNING Use of Masimo LNOP sensors/cables will not provide protection in
accordance with IEC defibrillation standards when used with this device.
WARNING If you need to increase the length of the sensor cable, use only one
extension. Use of multiple extensions can adversely affect performance. Do not
attach any cable that is intended for computer use to the SpO
monitor.
WARNING Tissue damage and erroneous measurements can be caused by
incorrect application or use of a sensor. (Examples of bad practices: wrapping the
sensor too tightly, applying supplemental tape, failing to periodically inspect the
sensor site, leaving a sensor on too long in one place.) Refer to the
manufacturer’s directions for specific instructions on application and use, and for
description, warnings, cautions, and specifications.
2
data. (“About the Masimo SpO2 display” on page 62.)
2
accessories listed in the Welch Allyn Products and
2
connector at the
2
WARNING Do not modify the sensor.
WARNING Do not wet the sensor or immerse it in fluid. Do not attempt to
sterilize a sensor.
WARNING Sensors exposed to ambient light while not applied to a patient can
exhibit seminormal saturation readings. Be sure the sensor is securely placed on
the patient and check its application often to ensure accurate readings.
WARNING Inaccurate measurements might be caused by venous pulsations.
WARNING The pulse oximeter can be used during defibrillation, but the
readings might be inaccurate for a short time.
WARNING Do not use the pulse oximeter as an apnea monitor.
WARNING During SpO
pulse rate can result in erroneous pulse rate readings. Always validate the patient
data and patient condition before effecting an intervention or a change in patient
care.
monitoring, a very sudden and substantial change in
2
Page 63
Directions for UseStandalone monitoring59
WARNING Interfering substances: Carboxyhemoglobin can erroneously
increase readings; the level of increase is approximately equal to the amount of
carboxyhemoglobin present. Methemoglobin and other dysfunctional
hemoglobins can also cause erroneous readings. Further assessment beyond
pulse oximetry is recommended. Intravascular dyes, or any substances
containing dyes, that change usual arterial pigmentation can cause erroneous
readings. Darkly pigmented skin can adversely affect SpO
readings.
2
WARNING For a premature infant, high oxygen levels might predispose the
infant to develop retinopathy. Therefore, the upper alarm limit for oxygen
saturation must be carefully selected in accord with accepted clinical standards
and considering the accuracy range of the monitor.
Caution If liquid gets into the SpO2 connector cavity, discontinue SpO2
monitoring until the liquid is removed and the cavity is dry.
1.Inspect the SpO2 cable. Replace it if it shows any signs of wear, breakage, or fraying.
2. Plug the cable into the sensor and the monitor.
Each SpO
sensor is intended for application to a specific site and site size on the
2
patient. To obtain optimal performance, use the right sensor and apply it as instructed
by the sensor manufacturer.
3. Clean the application site. Remove anything, such as nail polish, that could interfere
with the operation of the sensor.
4. Attach the SpO
sensor to the patient according to the manufacturer’s directions for
2
use, observing all warnings and cautions.
5. Confirm that the monitor displays SpO
data within a few seconds of being
2
connected to the patient.
•If ambient light is too bright, shield the sensor site with opaque material. Failure
to do so can result in inaccurate measurements. Light sources that can affect
performance include the following:
-surgical lights (especially those with a xenon light source)
-bilirubin lamps
-fluorescent lights
-infrared heating lamps
-direct sunlight.
•To help reduce unnecessary SpO
simultaneously, place the NIBP cuff and the SpO
•Do not attach the SpO
sensor on the same limb as an arterial catheter or
2
alarms when NIBP and SpO2 are monitored
2
sensor on different limbs.
2
intravascular line.
•The pulse signal can disappear if any of the following conditions exists:
-the sensor is too tight
-ambient light is too bright
-an NIBP cuff is inflated on the same limb as the sensor
-arterial occlusion occurs near the sensor
-the patient is in cardiac arrest or shock
-the patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia
•If poor perfusion affects performance for an adult, consider using the adult nasal
sensor.
•If a sensor is connected and the sensor light does not come on within 3 seconds:
-Verify t h a t SpO
is turned on (Figure 65 on page 60)
2
-Replace the sensor
•If excessive or prolonged patient movement interferes with measurements,
consider the following possible solutions:
-be sure the sensor is secure and properly applied
-use a new sensor with fresh adhesive backing
-select a different type of sensor
-move the sensor to a less active site
6. Periodically verify that the sensor remains properly positioned on the patient.
About SpO2 spot check
Note
The Spot Check feature is available only if it is enabled in the monitor
configuration. Refer to “Using the Configuration Utility to configure the monitor”
on page 114.
When SpO2 is turned on, the monitor generates an alarm condition whenever SpO2
readings are interrupted, such as when the sensor is disconnected from the patient after
the monitor begins taking SpO
Using the SpO
Spot Check feature, however, you can take any number of spot SpO2
2
readings.
2
readings at random intervals, attaching and detaching the sensor repeatedly without
generating alarms.
To prepare to take a spot check reading
1.Highlight SpO2 and press.
2. Highlight SpO2 Monitoring, press to highlight Off, and press to return to the
6. To take another spot check later, repeat from step 2 (above).
To return to continuous SpO
1.Highlight SpO2 @ XX:XX and press.
Spot-check readings are included in trend displays.
2
measurement. For example: SpO2 @ 3:05P.
2
3456187
Adult
Rm 239
98
SpO2 @ 3:05P
text on the display screen now includes the time of the most recent
3:12:13P
Resp/min
monitoring is again turned off.
2
measurements
2
Pulse rate
2. Press to turn automatic SpO
To adjust the SpO
1.Highlight SpO2 and press.
2. Highlight HR/PR Tone.
3. Highlight the desired volume level (Off, Low, Med, or High) and press.
and ECG pulse tone volume
2
About the Masimo SpO2 display
The Masimo SpO2 option provides the following additional features and indicators.
The monitor (models 802LTAS, 802LTRS and 802LT0S) must be configured
specifically to display the Masimo parameters.
Signal IQ (SIQ)(Signal Identification and Quality indicator) A horizontal line
below the SpO
the relative reliability of the oxygen saturation readings.
(Higher spikes indicate greater reliability.)
Perfusion Index (PI) An assessment, expressed as a percentage between 0.02
and 20, of the relative pulse strength at the monitoring
site.
FastSAT™The tracking of rapid changes in arterial oxygen saturation.
SensitivityNormal, APOD, and Maximum.
Note
On. The pop-up menu disappears.
2
waveform, with vertical spikes indicating
2
Page 67
Directions for UseStandalone monitoring63
NormalFor typical monitoring.
APODFor monitoring when the probe is likely to detach from the
patient because of wet skin, motion, or other unusual
circumstances.
MaximumFor monitoring patients with extremely low perfusion.
Averaging TimeThe period (in seconds) over which multiple SpO2
measurements are taken and a result is derived.
Monitoring blood pressure (NIBP)
WARNING Always use a properly fitting cuff, placing it carefully on the patient
according to the instructions presented below. Failure to fit and locate the cuff
correctly can lead to inaccurate pressure readings.
WARNING During monitoring, periodically observe the patient’s limb to make
sure that the circulation is not impaired for a prolonged period. Prolonged
impairment of circulation or improper cuff placement can cause bruising.
WARNING Do not use the monitor to simultaneously measure NIBP on one
patient and monitor ECG on another patient.
WARNING If an NIBP measurement is suspect, repeat the measurement. If you
are still uncertain about the reading, verify it using another method.
WARNING Do not take NIBP measurements during cardiopulmonary bypass.
WARNING When monitoring NIBP, match the monitor patient mode to the NIBP
cuff. For neonates, set the monitor to Neonatal Mode unless the circumference
of the limb is too large for the cuff. In that case, use the Pediatric Mode. Be
aware, however, that the maximum cuff inflation limits are based on the patient
mode, not the cuff; the maximum cuff inflation limits for Pediatric Mode are
greater than for Neonate Mode. (See “NIBP” on page 147 for values.)
Caution Pulse-rate measurements generated through the blood pressure cuff
or through SpO
measurements generated through ECG or through manual palpation.
When the monitor is powered on, the default cuff inflation pressure is based on the
patient mode. (See “Default inflation pressure” on page 147.) After an NIBP
measurement occurs, the monitor adjusts the inflation pressure to optimize subsequent
NIBP measurements.
Note
Always cycle the monitor power before you begin to monitor another patient.
Normal physiological pressure variations affect NIBP measurements from reading
to reading.
If the monitor is in Adult mode and a neonate cuff is connected to the monitor,
the monitor generates an equipment alert.
If the battery charge is low and the monitor is not in the cradle, the battery icon
indicates low battery and NIBP monitoring is disabled.
are subject to artifact and might not be as accurate as heart-rate
1.Select cuff size based on limb circumference. Use only hoses and cuffs listed in the
Welch Allyn Products and Accessories guide.
2. Squeeze all the air from the cuff before placing the cuff on the patient.
3. Place the cuff on the limb, as near heart level as possible.
Figure 70. Cuff placement
Cuff applied evenly and snugly. The
center of the cuff is at heart level,
and the bottom edge is one inch (2.5
cm) above the antecubital fossa.
Possible cuff placements for neonates
•The cuff must fit snugly without being uncomfortably tight.
•The hose must be free of kinks and not pinched.
4. Align the point where the tubing connects to the cuff over the brachial or other
appropriate artery.
Note
If you are simultaneously monitoring blood pressure and SpO
eliminate unnecessary SpO
alarms by placing the cuff and the SpO2 sensor on
2
, you can reduce or
2
different limbs.
5. Screw the hose connector onto the NIBP air connector on the top of the monitor (see
Figure 4 on page 8).
6. Press to start a reading.
Within a few seconds, the pump starts, the cuff begins to inflate, and the manometer
screen appears.
Figure 71. NIBP: initial view
STEWART, ANN
II 1mV/cm
HR/min
NIBP
NIBP Mode:
0 0 50
Exit
NIBP mmHg
Manometer
Manual
100
Trends
Adult
Systolic
150
7762940
Rm 263
Resp/min
Diastolic
200
Snapshots
3:00:06P
SpO2
Mean
250
Setup
%
300
7.When the cuff is fully inflated, the manometer bar dynamically displays the pressure
reading.
Page 69
Directions for UseStandalone monitoring65
NIBP Mode:
II 1mV/cm
Exit
Trends
Snapshots
Setup
HR/min
NIBP mmHg
Resp/min
SpO2
NIBP
Systolic
Diastolic
Mean
%
STEWART, ANN
7762940
3:00:06P
Adult
Rm 263
Manometer
Manual
NIBP in progress.
145
50
100
150
200
250
300
NIBP Mode:
II 1mV/cm
Exit
Trends
Snapshots
Setup
PR/min
NIBP mmHg
Resp/min
SpO2
NIBP
Systolic
Diastolic
Mean
%
STEWART, ANN
7762940
3:00:06P
Adult
Rm 263
Manometer
Manual
50
100
150
200
250
300
(98)
79
132/83
Figure 72. NIBP: reading in progress
STEWART, ANN
NIBP in progress.
II 1mV/cm
HR/min
NIBP
NIBP Mode:
145
Exit
NIBP mmHg
Manometer
Manual
50
Trends
100
Adult
Systolic
150
7762940
Rm 263
Resp/min
Diastolic
200
Snapshots
3:00:06P
SpO2
Mean
250
Setup
%
Manometer pressure indicator bar
300
When the cycle completes, control returns to the previous direct-access display. To
view the final NIBP readings on the manometer scale, select Manometer from the Setup NIBP menu.
Figure 73. NIBP: reading complete
STEWART, ANN
II 1mV/cm
79
PR/min
NIBP
NIBP Mode:
NIBP mmHg
Manometer
Manual
132/83
Systolic
Adult
(98)
7762940
Rm 263
Resp/min
Diastolic
3:00:06P
%
SpO2
Mean
Systolic, diastolic, and MAP
(numeric)
Systolic, diastolic, and MAP
(graphical)
2 2 50
Exit
100
Trends
150
Snapshots
200
250
300
Setup
•If you need to stop the reading and vent the cuff at any time during the
measurement, press.
•If the monitor cannot get a valid NIBP reading, it displays the message
‘NIBP retry in progress’ in the upper left corner of the screen. Then, depending on
settings and patient mode, the monitor attempts one or two more
measurements. If the retries fail to produce a valid reading, the monitor displays
an NIBP error message.
Note
For every inch the cuff is placed above the heart, add 1.9 mmHg (0.253 kPa) to the
displayed NIBP reading. For every inch below the heart, subtract 1.9 mmHg
(0.253 kPa).
8. If motion artifact such as shivering or coughing interferes with NIBP readings, do the
following:
•Position the patient’s limb away from the body so that the applied cuff is not in
contact with the patient’s body or any other object (such as a bed rail). Keep the
cuff as close to heart level as possible.
•Verify that the Smartcuf filter is ON. (See “Improving NIBP accuracy with
Smartcuf” on page 66.)
•Verify that the ECG leads are properly connected to the patient, and monitor ECG
during NIBP. (ECG monitoring is required for Smartcuf.)
The message ???/???/(???) in an NIBP TREND display or printout indicates that
the monitor could not complete an NIBP measurement during that period.
On-demand NIBP readings (manual or turbo) are delayed by 8 seconds if they are
started while the monitor is in a power-saving mode such as display time-out.
Improving NIBP accuracy with Smartcuf
Many factors can adversely affect an NIBP measurement: cardiac arrhythmias, sudden
changes in blood pressure, patient motion such as convulsions or shivering, sudden cuff
movement, vibration, vehicle motion, or a weak pulse. The Smartcuf feature increases
NIBP measurement accuracy in the presence of moderate motion artifact or diminished
pulses.
s
Note
To enable Smartcuf
1.Simultaneously monitor ECG and NIBP.
2. Access the Setup menu. (See “To access the setup menus” on page 35.)
3. Highlight NIBP, highlight Smartcuf, and enable the Smartcuf filter.
Smartcuf can function only when ECG is being monitored.
If Smartcuf is enabled and motion artifact is so severe that it still affects measurement
accuracy, the measurement is marked with the symbol on the display. During certain
types of arrhythmias and other situations where a good ECG signal cannot be obtained,
consider disabling Smartcuf, as follows:
To disable Smartcuf
1.Access the Setup menu.
2. Highlight NIBP, highlight Smartcuf, and disable the Smartcuf filter.
Taking automatic NIBP readings
In the Auto NIBP mode, for intervals shorter than 5 minutes, the monitor immediately
begins taking NIBP readings at the specified interval. For intervals of 5 minutes or longer,
the readings begin when the time of day is a multiple of the interval. (If the interval is 15
minutes, for example, then the readings begin at 00, 15, 30, or 45 minutes after the hour.)
To start automatic NIBP readings
1.Select and apply the appropriate NIBP cuff and hose. (“Monitoring blood pressure
(NIBP)” on page 63.)
2. Highlight NIBP and press.
3. Highlight NIBP Mode and select Auto.
4. Highlight Auto Interval (min) and select an interval.
Note
After you invoke an automatic NIBP, expect a delay before the monitor starts the
first measurement. The delay can be as long as the interval selected.
Page 71
Directions for UseStandalone monitoring67
To stop automatic NIBP readings
1.Highlight NIBP and press.
2. Highlight NIBP Mode and select Manual.
Taking NIBP readings using Turbo mode
In Turbo mode, the monitor starts an NIBP reading and then takes as many more readings
as possible within five minutes.
To use the Turbo mode
1.Select and apply the NIBP cuff and hose. (“Monitoring blood pressure (NIBP)” on
page 63.)
2. Highlight NIBP and press.
3. Highlight NIBP Mode and select Tu r b o .
Note
To end the Turbo mode
If you cycle the monitor power, NIBP returns to auto mode.
Pressor select NIBP Mode Manual. The monitor returns to manual NIBP
measurement mode.
NIBP measurements in power-saving mode
When a manual or turbo NIBP activity awakens the monitor from power-saving mode
(“Power saving” on page 42), cuff inflation pressure is reset to default levels and cuff
inflation is delayed for up to 8 seconds.
NIBP disabled when the battery is low
If the battery is low and the monitor is operating on battery power, NIBP functions are
disabled and the monitor displays the message “NIBP off. Low battery.”
If you attempt to start an NIBP measurement during a low-battery condition, the monitor
displays an equipment alert with the message “Low battery. NIBP disabled.”
Note
Inserting a monitor into a powered cradle during a low-battery condition
immediately enables NIBP monitoring.
Changing the default settings
You can change the default settings (the monitor configuration) using the following
methods:
•Save the current settings as the new default settings.
Select ( ) to use factory settings.
Select ( ) to use file: n/a
Setup
Alarms
ECG
NIBP
Timings
Service
Configuration
ID: 12549M6I9ERF
03:52:57
AdultRmRm
NIBP mmHg
HR/min
SpO2
%
M
Note! You are changing the
default settings to current
If this is correct, highlight
“Confirm” and then press
Cancel
Confirm
settings.
Select ( ).
ID: 12549M6I9ERF
03:53:45
AdultRmRm
NIBP mmHg
HR/min
SpO2
%
M
If this is correct, highlight
“Confirm” and then press
Cancel
Confirm
Select ( ).
Note! You are changing the
default settings to file:
post-op4 15 Dec 06
•Download another configuration from a PC. (See “Using the Configuration Utility
to configure the monitor” on page 114.)
To save the current settings as the new default settings
1.Access the Configuration Setup menu.
80
HR/min
Setup
Alarms
NIBP mmHg
ECG
NIBP
Resp/min
Timings
12
Configuration
97
SpO2
%
Service
Select ( ) to save current settings.
Select ( ) to use factory settings.
Select ( ) to use file: n/a
Exit
Trends
Snapshots
Setup
2. Highlight Select ( ) to save current settings and press .
3. In the confirmation screen, highlight Confirm and press .
ID: 12549M6I9ERF
Note! You are changing the
default settings to current
settings.
If this is correct, highlight
HR/min
“Confirm” and then press
Select ( ).
Cancel
NIBP mmHg
Adult
SpO2
Confirm
S S D D M
03:52:57
%
To restore the last downloaded configuration
1.Access the Configuration Setup menu, highlight Select ( ) to use file:
configuration_file_name, and press .
2. In the confirmation screen, highlight Confirm and press .
ID: 12549M6I9ERF
HR/min
NIBP mmHg
Adult
Note! You are changing the
default settings to file:
post-op4 15 Dec 06
If this is correct, highlight
“Confirm” and then press
Select ( ).
Cancel
S S D D M
SpO2
Confirm
03:53:45
%
To restore the factory configuration
1.Access the Configuration Setup menu, highlight Select ( ) to use factory settings,
and press .
2. In the confirmation screen, highlight Confirm and press .
Page 73
Directions for UseStandalone monitoring69
ID: 12549M6I9ERF
03:53:45
AdultRmRm
NIBP mmHg
HR/min
SpO2
%
M
Note! You are changing the
default settings to factory
If this is correct, highlight
“Confirm” and then press
Cancel
Confirm
settings.
Select ( ).
Select ( ) to delete patient data
You have pressed the Power Off button.
Delete & Shut Down
Cancel
pSetup
and shut down.
15:01:24
Adult
Rm 239
There is patient data stored for
Hall, Robert E. ID: 3456187
Save & Shut Down
HALL, ROBERT E.
3456187
ID: 12549M6I9ERF
Note! You are changing the
default settings to factory
settings.
If this is correct, highlight
HR/min
“Confirm” and then press
Select ( ).
NIBP mmHg
Cancel
S S D D M
Adult
SpO2
Confirm
To discontinue monitoring
1.Press. The Power Off screen appears.
Figure 74. Power-off
HALL, ROBERT E.
You have pressed the Power Off button.
There is patient data stored for
Hall, Robert E. ID: 3456187
Adult
3456187
Rm 239
03:53:45
%
15:01:24
Select ( ) to delete patient data
Delete & Shut Down
and shut down.
Save & Shut Down
Cancel
Setu
•If you intend to continue monitoring the same patient when the monitor is turned
on again, and if you want to save the stored vital-signs data and monitor settings
(to print them at a PC), highlight Save & Shut Down and then press. The
monitor saves the patient data and the monitor settings, and then turns off.
•If you do not intend to continue monitoring the same patient when the monitor is
turned on again, highlight Delete & Shut Down and then press. The monitor
turns off without saving the data and the settings.
2. Disconnect the leads and sensors from the patient.
If you press and then decide that instead of turning off the monitor you want to resume
monitoring the same patient, do one of the following:
•Highlight Cancel and press.
•Wait for 30 seconds.
•Press.
Note
When you power down from Demo mode, you cannot save settings and patient
data.
The model 802LTAN, 802LTAS, 802LTRN and 802LTRS wireless monitors can operate
either in standalone mode or in communication with a Welch Allyn Acuity Central Station.
(See “Standalone monitoring” on page 45.)
While the wireless monitor is connected to the FlexNet network, patient data gathered by
the monitor is continuously stored at Acuity. At the Acuity station you can access this
patient information and perform administrative functions, including the following:
•Admit, transfer, and discharge a patient in the Acuity unit.
•Edit the patient description (name, physician, and so on).
•Review and print patient data such as trends and waveforms.
For more information about Acuity and the FlexNet wireless network, refer to Acuity directions For use.
For Model 802LTAN and 802LTAS monitors, the radio, by default, authenticates against
the WA RADIUS server running on an Acuity Central Station. If the hospital IT department
requires a different authentication/encryption solution, then the monitor, using the factorydefault radio configuration, cannot connect to the network.
WARNING When monitoring a patient who has a pacemaker, position the
monitor to maintain a minimum distance of 6 inches (16 cm) between the monitor
and the pacemaker. If you have any reason to suspect that the monitor is
interfering with the pacemaker, immediately turn the monitor off and
provide appropriate patient care. The Health Industry Manufacturers
Association recommends this minimum distance (6 inches or 16 cm) between a
hand-held wireless radio and a pacemaker, which is consistent with the
independent research by, and recommendations of, Wireless Technology
Research.
WARNING When considering a treatment protocol that involves wireless
communication of patient data, be aware of this limitation:
When the monitor is not connected to the network by wireless connection:
Acuity does not receive patient alarms or alerts from the monitor.
Acuity does not perform arrhythmia and ST analysis on patient data and
therefore does not generate related alarms.
WARNING If you establish a USB link to any active host while the monitor is in
network communication with Acuity, the wireless connection is lost.
Page 76
72Monitoring in communication with AcuityWelch Allyn Propaq LT Vital Signs Monitor
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
Note
When the monitor is used in Canada:
To provide maximum shielding (to prevent radio interference to the licensed
service), the monitor is intended to be operated indoors and away from windows.
Radio equipment (or its transmit antenna) that is installed outdoors is subject to
licensing.
Afin de fournir un écran de blindage maximal (pour empêcher que cet appareil
cause du brouillage au service faisant l’objet d’une licence), l’appareil doit être
utilisé à l’intérieur et devrait être placé loin des fenêtres. Si le matériel (ou son
antenne d’émission) est installé à l’extérieur, il doit faire l’objet d’une licence.
Establishing communication with Acuity
If the monitor is enabled for communication with Acuity, communication is established
automatically whenever the monitor is within range of the network.The monitor always
indicates the status of the connection:
Figure 75. Connection status indicator
The monitor is not enabled for communication with Acuity.(blank)
The monitor is communicating via USB cable with a PC.
The monitor is associated with the network and is
communicating with Acuity.
(Flashing) The monitor is associated with the network but
is not communicating with Acuity.
(Continuous) The monitor is not associated with the
network and is not communicating with Acuity.
HALL, ROBERT E.
Verify that the monitor, within a minute or two, displays to indicate that the connection
to Acuity has been established.
Upon connecting to the network, the monitor uploads the following data to Acuity:
•Patient information and trends stored at the monitor before the network
connection was established
•Monitor settings
At the same time, Acuity downloads some configuration settings to the monitor. (See
“About Acuity-defined configuration” on page 102.)
For information on monitoring vital signs, see the following:
ECG/Resp “Monitoring ECG and Resp” on page 50
SpO
2
NIBP“Monitoring blood pressure (NIBP)” on page 63
“Monitoring SpO2” on page 58
Adult
3456187
Rm 239
3:00:06P
Page 77
Directions for UseMonitoring in communication with Acuity73
Monitoring a patient outside of network range
Monitoring continues locally when the monitored patient goes outside of the range of the
network. The monitor provides local Resp, NIBP, HR/PR and SpO
alerts. The following also happens:
•An alert occurs at Acuity with the message DROPOUT.
•An alert occurs at the monitor with the message Check Acuity/network
connection.
•The monitor emits audible alert tones.
•The appropriate caregiver presses to acknowledge the alert.
WARNING When the monitor moves out of range of the Acuity network,
communication with Acuity is broken immediately, stopping the communication
of patient vital-signs data. If the monitor is out of range of the Acuity network for
several minutes, the radio enters a low-power state. When the monitor is again
within range of the Acuity network, it can take as long as three minutes to restore
communication with Acuity and resume the communication of patient vital-signs
data.
alarms or equipment
2
Note
When the monitored patient is again within network range, the monitor reconnects to the
network and wireless monitoring resumes.
Note
When the monitor is outside of network range, battery operating time is reduced.
When communication with the network is re-established, trends data is uploaded
to Acuity but snapshot data is not.
About Acuity message menus
When the monitor first establishes communication with Acuity and the clinician selects to
start a new patient, the monitor presents a series of screens to determine the location
(unit and room number) and the name and ID number of the monitored patient. If the
patient is already known to Acuity, you can select the identifying information from a series
of lists. If the patient is new to Acuity, you provide the information.
If the clinician is continuing to monitor the same patient, these Acuity message menus
are not presented.
Note
First, the monitor displays a list of units. Highlight the unit for this patient and press.
Some of these menus can contain lists that are too long to fit on the screen. To
view or highlight list items that are not on the screen, press or repeatedly
as needed scroll the list.
Page 78
74Monitoring in communication with AcuityWelch Allyn Propaq LT Vital Signs Monitor
II 1mV/cm
HR/min
NIBP mmHg
Resp/min
SpO2
%
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
SELECT
UNIT
Emergency Department
Pediatric
80
12
97
II 1mV/cm
7762940, STEWART, ANN
3398290, TAYLOR, MIKE
HR/min
NIBP mmHg
Resp/min
SpO2
%
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
SELECT
PATIENT
Enter new patient
4727937, BENET, CLAUDE
9542683, MERKER, HANS
80
12
97
II 1mV/cm
1306B
1306C
HR/min
NIBP mmHg
Resp/min
SpO2
%
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
SELECT
ROOM
Enter new room
1304D
1306A
1306D
80
12
97
Figure 76. Acuity unit list
STEWART, ANN
II 1mV/cm
80
HR/min
SELECT
UNIT
Emergency Department
Pediatric
NIBP mmHg
Adult
7762940
Rm 263
12
Resp/min
12:41:32
%
97
SpO2
The monitor then displays a list of patients.
Figure 77. Acuity patient list
STEWART, ANN
II 1mV/cm
80
HR/min
SELECT
PATIENT
Enter new patient
4727937, BENET, CLAUDE
9542683, MERKER, HANS
7762940, STEWART, ANN
3398290, TAYLOR, MIKE
NIBP mmHg
Adult
7762940
Rm 263
12
Resp/min
12:41:32
97
%
SpO2
•If the patient’s name is already on the list, highlight the name and press.
•If it is not, highlight Enter new patient, press, and use the information entry
screen (Figure 46 on page 47) to enter the patient’s name and/or ID number (if
either is known).
If you select the name of an existing patient, the monitor displays a list of rooms. Select
the patient’s room.
Figure 78. Acuity room list
STEWART, ANN
II 1mV/cm
80
HR/min
SELECT
ROOM
Enter new room
1304D
1306A
1306B
1306C
1306D
NIBP mmHg
Adult
7762940
Rm 263
12
Resp/min
12:41:32
97
%
SpO2
Page 79
Directions for UseMonitoring in communication with Acuity75
aVR 2mV/cm
HR/min
NIBP mmHg
Resp/min
SpO2
%
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
EQUIPMENT ALERT
Communications fault
Check Acuity/network connection
Acknowledge
80
12
97
aVR 2mV/cm
HR/min
NIBP mmHg
Resp/min
SpO2
%
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
CONFIRM PATIENT?
TO 7762940 ED 263
YES
NEW UNIT
IS THIS MONITOR STILL CONNECTED
ANN STEWART
NO
NEW ROOM
80
12
97
If communication is interrupted during monitoring
If communicated monitoring has begun and communication is then interrupted, the
monitor displays an alert message.
Figure 79. Alert message: Communications Fault
STEWART, ANN
aVR 2mV/cm
80
HR/min
NIBP mmHg
Check Acuity/network connection
EQUIPMENT ALERT
Communications fault
Acknowledge
Adult
7762940
Rm 263
12
Resp/min
12:41:32
%
97
SpO2
Press to clear the message.
When the monitor re-establishes communication with Acuity (after an interruption of at
least 30 seconds), it prompts you to confirm that the patient being monitored is the same
patient that was being monitored when communication was interrupted.
Figure 80. Same Patient query
STEWART, ANN
aVR 2mV/cm
Adult
7762940
Rm 263
12:41:32
80
HR/min
CONFIRM PATIENT?
IS THIS MONITOR STILL CONNECTED
TO 7762940 ED 263
ANN STEWART
YES
NO
NEW ROOM
NEW UNIT
Note
NIBP mmHg
If the communication interruption lasts less than 30 seconds, or if you save, cycle
12
Resp/min
97
SpO2
%
power, and reestablish communication within 30 seconds, this confirmation is not
required.
Press to confirm.
To access the Acuity menu after connection has been established
1.Highlight and press . The Wireless mode popup menu appears (Figure 82).
Note
Depending on the level of network traffic, some interactions between the monitor
and Acuity can take a few seconds to complete.
2. Highlight Acuity menu and press.
Page 80
76Monitoring in communication with AcuityWelch Allyn Propaq LT Vital Signs Monitor
II 1mV/cm
TRANSFER
NEW PATIENT
80
HR/min
NIBP mmHg
Resp/min
SpO2
12
97
%
STEWART, ANN
7762940
12:41:32
Adult
Rm 263
ACUITY
MENU
EXIT
END TELE
NEW ROOM
PATIENT INFO
%
80
12
97
STEWART, ANN
7762940
15:20:28
Adult
Rm ED263
HR/min
NIBP mmHg
Resp/min
SpO2
II 1mV/cm
Wireless Mode
Exit
Disconnect
Acuity Menu
Figure 81. Acuity menu
STEWART, ANN
II 1mV/cm
80
HR/min
ACUITY
MENU
EXIT
END TELE
NEW ROOM
TRANSFER
NEW PATIENT
PATIENT INFO
You can use the Acuity menu (Figure 81) to end the telemetry session (END TELE), assign
the patient to a new room (NEW ROOM), transfer the patient to another unit
(TRANSFER), enter the name and ID of a new patient (NEW PATIENT), or update the
information for the current patient (PATIENT INFO).
To close the Acuity menu, highlight EXIT and press .
To disconnect from the network
Note
Disconnecting the monitor from the network while the monitor is outside of
network range greatly extends battery life. However, connection with the network
must first have been established before this is possible.
NIBP mmHg
Adult
7762940
Rm 263
12
Resp/min
12:41:32
97
%
SpO2
1.Highlight and press.
2. Highlight Disconnect, press, and confirm.
Figure 82. Wireless Mode: Disconnect
STEWART, ANN
II 1mV/cm
80
HR/min
changes to and the message ‘DISCONNECTED’ appears, alternating with the
patient ID number.
NIBP mmHg
Adult
7762940
Wireless Mode
Rm ED 263
Exit
Disconnect
Acuity Menu
12
Resp/min
97
15:20:28
SpO2
Wireless Drop-Down menu
Page 81
Directions for UseMonitoring in communication with Acuity77
%
80
12
97
STEWART, ANN
DISCONNECTED
15:20:28
Adult
Rm 263
HR/min
NIBP mmHg
Resp/min
SpO2
II 1mV/cm
Figure 83. Patient disconnected from the network
STEWART, ANN
II 1mV/cm
80
HR/min
Note
To reconnect a patient to the network
1.Highlight and press.
2. Highlight Reconnect and press.
If you disconnect from the network and then turn off the monitor before
reconnecting, and if you then select Continue Patient when you again turn on
the monitor, the monitor powers up in the disconnected state.
The monitor prompts you for information about the connection and the patient.
DISCONNECTED
NIBP mmHg
Adult
12
Resp/min
Rm 263
97
15:20:28
%
SpO2
3. Provide the information requested.
Page 82
78Monitoring in communication with AcuityWelch Allyn Propaq LT Vital Signs Monitor
Page 83
79
5
Alarms and alerts
Overview
An Alarm warns of a patient condition, such as a vital-sign reading that is outside of
acceptable limits. When an alarm occurs, the red light on the monitor flashes and the
numerics of the violating alarm limits on the display turn red. If tones are not suspended,
the alarm tone sounds.
An Alert warns of an equipment condition, such as a low battery or a detached lead.
When an alert occurs, the yellow light on the monitor flashes and a message describing
the error condition appears on the display. If tones are not suspended, the alert tone
sounds.
Alarms have priority over alerts. If an alarm and an alert are detected simultaneously, the
monitor notifies you of the alarm. It then notifies you of the alert only if the alert condition
still exists after the alarm condition is removed.
Silencing an alarm or alert tone
A tone sounds whenever the monitor detects an alarm or alert condition.
To silence the currently sounding tone for 90 seconds
1.Pr e s s.
Note
Silencing the tone does not affect the other alarm or alert indicators.
•The red light (alarm) or the yellow light (alert) flashes.
•After 90 seconds, if the condition is not corrected, the tone starts again.
•If the condition is corrected within 90 seconds of silencing the tone, the monitor
resets the tones for the next alarm or alert.
If a new alarm or alert condition occurs while an earlier alarm or alert is silenced, the
tone sounds again.
Page 84
80Alarms and alertsWelch Allyn Propaq LT Vital Signs Monitor
%
140/78
125
31
88
HALL, ROBERT E.
3456187
3:00:06P
Adult
Rm 239
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
II 1mV/cm
Resp/min
SpO2
I 1mV/cm
HALL, ROBERT E.
3456187
12:41:32
Adult
Rm 239
AudibleAlarmsSilenced
80
140/78
12
97
%
HR/min
NIBP mmHg (102)
@2:47P Manual
EQUIPMENT ALERT
ECG Fault
LL Lead Failed
ACKNOWLEDGE
Figure 84. Sample alarm screen
HALL, ROBERT E.
II 1mV/cm
125
HR/min
140/78
NIBP mmHg (102)
@2:47P Manual
Adult
3456187
Rm 239
31
Resp/min
3:00:06P
88
SpO2
Figure 85. Sample alert screen
HALL, ROBERT E.
Audible Alarms Silenced
I 1mV/cm
140/78
80
NIBP mmHg (102)
HR/min
@2:47P Manual
X
ACKNOWLEDGE
3456187
Adult
Rm 239
12
Resp/min
EQUIPMENT ALERT
ECG Fault
LL Lead Failed
12:41:32
97
SpO2
2. Check the patient and provide appropriate care.
3. Press to silence the audible tone for 90 seconds at the monitor and at Acuity.
•Silencing the audible tone does not remove visual alarm or alert indications.
•Silencing the audible tone causes an alarm or alert suspend at Acuity.
•When the alarm or alert condition is corrected, all alarm or alert indicators cease
and all alarms are immediately rearmed.
4. After caring for the patient, verify that alarm limits are enabled and correctly set.
WARNING If you turn off or modify any alarm limits while responding to an
alarm, restore those alarm limits before you resume monitoring.
Suspending the alarm tone
If this feature is enabled in the monitor configuration (See “Monitor configuration” on
page 101), you can suspend all alarm tones for all parameters—preventing the alarm tone
from sounding if an alarm condition occurs—while monitoring a patient. If an alarm
condition occurs while the alarm tones are suspended, the monitor presents visual alarm
indicators but does not sound the tone.
In the monitor configuration, the alarm tone suspension period can be set to Disabled, to
Always On, or to a period: 90 sec or 2, 3, 4, 5, 10, 15, 30, or 60 minutes.
Note
The factory default suspension period is 4 minutes.
•If it is configured to Disabled, you cannot suspend the alarm tone at the monitor.
Page 85
Directions for UseAlarms and alerts81
II 1mV/cm
Exit
Trends
Snapshots
Setup
Low
Med
HR/min
NIBP mmHg
Resp/min
SpO2
OffOnOn
Alarm Tone
Suspend Audible Alarms
Setup
(90 sec)
%
3:00:06P
Adult
Rm 239
Alarms
ECG
NIBP
Timings
High
Configuration
HALL, ROBERT E.
3456187
Right ( ) to suspend audible alarms.
Service
II 1mV/cm
Exit
Trends
Snapshots
pSetup
Low
Med
80
HR/min
NIBP mmHg
Resp/min
SpO2
OffOnOn
Alarm Tone
Suspend Audible Alarms
Setup
(90 sec)
12
97
%
8765432
3:00:06P
Adult
Rm 239
Alarms
ECG
NIBP
Timings
High
Service
0:01:30 remaining
Audio alarms suspended
ID: 01018VDO9PBH
Configuration
•If it is configured to Always On and you set Suspend Audible Alarms to On,
then the alarm tone remains suspended until:
•you set Suspend Audible Alarms to Off or
•monitor power is turned off and then turned on again or
•the monitor is reconfigured
Note
Suspend Audible Alarms does not affect the behavior of the alarm silence/
reset feature (). Pressing always either silences a sounding alarm tone for 90
seconds or resets the audible alarm if it was already silenced.
To suspend the alarm tone
1.Access the Setup menu. (See “To access the setup menus” on page 35.)
2. Highlight Suspend Audible Alarms).
Figure 86. Suspend Audible Alarms: Off
HALL, ROBERT E.
II 1mV/cm
HR/min
Setup
Suspend Audible Alarms
Alarm T one
Exit
NIBP mmHg
Alarms
ECG
Service
Right ( ) to suspend audible alarms.
Trends
Adult
NIBP
Snapshots
3456187
Rm 239
Resp/min
Timings
Off
Low
Med
3:00:06P
SpO2
Configuration
(90 sec)
High
Setup
%
The configured suspension period—90 seconds in this example—is displayed to the
right of the line.
3. Highlight On.
Note
If audible alarms are suspended, pressing cancels the suspension.
Figure 87. Suspend Audible Alarms: On
ID: 01018VDO9PBH
Audio alarms suspended
II 1mV/cm
80
HR/min
Setup
Alarms
Service
Suspend Audible Alarms
Alarm T one
Exit
NIBP mmHg
ECG
Trends
8765432
Adult
NIBP
Timings
Off
0:01:30 remaining
Low
Snapshots
•The alarm tone is suspended immediately.
If an alarm condition occurs during the suspension period, the alarm tone does
not sound.
Rm 239
12
Resp/min
Med
3:00:06P
97
SpO2
Configuration
(90 sec)
High
Setu
‘Tone suspended’ indicator
%
Configured suspension period
Suspension time remaining
Page 86
82Alarms and alertsWelch Allyn Propaq LT Vital Signs Monitor
II 1mV/cm
10
15
30
60
Exit
Trends
Snapshots
Setup
Auto
Manual
Turbo
HR/min
NIBP mmHg
Resp/min
SpO2
Off
Off
On
On
220
Lower Sys Limit
NIBP Mode
Auto Interval (min)
Upper Sys Alarm
Upper Sys Limit
Lower Sys Alarm
NIBP Manometer
Systolic
Diastolic
Mean
75
%
STEWART, ANN
7762940
3:00:06P
Adult
Rm 263
125
140/78
31
88
(102)
II 1mV/cm
10
15
30
60
Exit
Trends
Snapshots
Setup
Auto
Manual
Turbo
HR/min
NIBP mmHg
Resp/min
SpO2
Off
Off
On
On
210
Lower Sys Limit
NIBP Mode
Auto Interval (min)
Upper Sys Alarm
Upper Sys Limit
Lower Sys Alarm
NIBP Manometer
Systolic
Diastolic
Mean
65
%
STEWART, ANN
7762940
3:00:06P
Adult
Rm 263
(102)
125
140/78
31
88
•A countdown timer appears below the line to indicate the time remaining in the
suspension period.
•‘Audio alarms suspended’ appears in yellow in the upper left corner of the screen.
•When the suspension period elapses, the alarm tone is again enabled.
Changing alarm limits
At the monitor
Typically, each institution determines the appropriate alarm limits for adult, pediatric, and
neonatal patients and then configures the monitor with those alarm limits before putting
the monitor into service. These become the default alarm limits for the monitor—the
alarm limits used each time the monitor is powered on.
The configured alarm limits remain in effect until the monitor configuration is changed.
However, you can temporarily change the current patient’s alarm limits.
To temporarily change alarm limits for the current patient
At Acuity
1.Highlight the vital sign for which you want to set custom limits.
2. Press.
3. Highlight the limit you want to change.
4. Set a new alarm limit.
Figure 88. Changing alarm limits for the current patient
5. Repeat from step 3 for other alarm limits you want to change for the same vital sign.
6. Press.
When the monitor is turned off, you will have a choice of saving the current settings and
stored patient data or deleting the current settings and patient data. Neither choice affects
the default settings for the monitor.
For a wireless monitor, patient alarm limits can also be customized from Acuity. (See the
user manual supplied with any Acuity Central Monitoring Station.)
Page 87
Directions for UseAlarms and alerts83
14040/7878
80
12
97
%
HALL,ROBERTE.
3456187
3:00:06P
Adult,
Rm239
HR/min
NIBPmmHg(102)@2:47PManual
Resp/min
SpO2
SpO22x2x
II1mV/cm
Resp/min
SpO2
II 1mV/cm
HALL, ROBERT E.
3456187
12:41:32
Adu lt
Rm 239
AudibleAlarmsSilenced
80
140/78
12
97
%
HR/min
NIBP mmHg (102)
@2:47P Manual
EQUIPMENT ALERT
ECG Fault
LL Lead Failed
ACKNOWLEDGE
About ParamSet
Using ParamSet, you can quickly widen the alarm limits by a configured percentage
(relative to the patient’s alarming reading) for any vital sign.
For information about ParamSet, see “ParamSet Settings” on page 131.
Responding to an alarm
An alarm condition is indicated on the monitor in the following ways:
•The RED indicator (rightmost of three) above the display screen flashes.
•The numerics for the vital sign in alarm are displayed in RED.
•If alarm tones are not suspended, the alarm-tone sequence sounds repeatedly—3
short tones, a short pause, 2 short tones, and a long pause.
Responding to an alert
An alert condition (“Over view” on page 79) is indicated on the monitor in the following
ways:
•A flashing yellow light above the monitor display.
•A yellow alert message on the monitor display (for example, NO ECG CABLE
DETECTED).
•Repeated sounding of the alert-tone sequence (if tones are not suspended): 3 long
tones and a pause.
Figure 89. Example alert screen
to silence the alert tone for
Press
90 seconds.
HALL, ROBERT E.
HALL, ROBERT E.
Audible Alarms Silenced
II 1mV/cm
II 1mV/cm
140/78
80
NIBP mmHg (102)
HR/min
SpO2
@2:47P Manual
80
HR/min
NIBP mmHg (102)@2:47P Manual
X
ACKNOWLEDGE
3456187
3456187
Adult
Adult,
Rm 239
Resp/min
EQUIPMENT ALERT
ECG Fault
12
Resp/min
LL Lead Failed
Rm 239
12
97
SpO2
Propaq LT
12:41:32
3:00:06P
97
SpO2
The yellow light flashes during an alert
and while the alert is silenced.
Text on a yellow field identifies the
alert condition.
To respond at the monitor to an alert
1.Press to silence the tone for 90 seconds; press to clear the alert.
2. Troubleshoot and correct the problem. (See “Alert messages and status messages”
on page 85.)
Page 88
84Alarms and alertsWelch Allyn Propaq LT Vital Signs Monitor
To respond at Acuity to an alert
For a monitor in communication with Acuity, either the monitor or Acuity can detect
alarms and alerts. The alarms are displayed in both places. See Acuity Directions For use.
About battery charge status
Table 7. Battery status indicators
DisplayStatus/alert
Green
Green
Green
Yellow
Yellow
Red
Red
text
EQUIPMENT ALERT
Low Battery.
Charge battery soon.
EQUIPMENT ALERT
Battery Too Low.
Shutting down.
BatteryMonitorRecommended
Fully chargedOkayNone
Partially fullOkayNone
Partially full; charging OkayNone
LowCan function for up to
30 minutes, but NIBP
is disabled.
Low; chargingNormal function.Do not remove the
Almost completely
discharged
Almost completely
discharged; charging
Shutting down soon.
Can function for up to
5 minutes, but NIBP is
disabled.
Normal function. Do not remove the
action
Prepare to discontinue
monitoring. If possible,
insert the monitor into a
cradle.
monitor from the cradle.
Prepare to discontinue
monitoring. If possible,
insert the monitor into a
cradle.
If this indicator appears
when the monitor is in a
powered cradle, then
the battery is damaged
and must be replaced.
In this case, all stored
patient data will be
deleted when the
monitor is removed
from the cradle.
monitor from the cradle.
Page 89
Directions for UseAlarms and alerts85
Alert messages and status messages
Table 8. Alert messages
Alert typeMessagePossible cause and suggested response
ECGECG Fault.
XX lead failed.
ECG Fault.
XX, XX leads failed.
ECG Fault.
Multiple lead fail.
ECG Fault.
Excessive offset.
ECG Fault.
Cable disconnected.
NIBPNIBP Fault.
Air leak.
Check hose.
NIBP Fault.
Kinked hose.
Check hose.
NIBP Fault.
Overpressure condition.
NIBP Fault.
Weak Pulses.
Can’t find Sys/Dia.
Lead XX (LA, LL, RA, C, or RL) has very poor contact or no
contact with the patient. Check for proper connection and
replace the electrode if needed.
Leads XX and XX (any two leads on a 5-lead cable) have very
poor contact or no contact with the patient. Check for proper
connection; replace electrodes if needed.
At least three leads of a 5-lead cable or at least two leads
of a 3-lead cable have very poor contact or no contact with
the patient. Check for proper connection; replace electrodes
if needed.
At least one channel has excessive offset. At least one
electrode is old, contaminated, or defective. Replace the
electrodes.
The ECG cable is unplugged.
The monitor could not properly inflate the cuff. Check the
hose and cuff for leaks.
The monitor could not properly inflate the cuff. Check for a
hose kink between the monitor and the patient.
The pressure in the cuff exceeded the acceptable limits for
the current patient mode. Check the hose and retry the
measurement.
Not enough pulses to determine the systolic or diastolic
pressures, but a mean pressure is available. Squeeze all air
from the cuff and reapply it.
NIBP Fault.
Artifact.
Can’t find Sys/Dia.
NIBP Fault.
No pulses detected.
NIBP Fault.
Connect ECG to reduce NIBP artifact.
The systolic or diastolic pressures are unreliable due to
artifact, but a mean pressure is available. Usually caused by
patient motion.
The cuff might not be properly applied to the patient, or the
patient might not have detectable pulses due to shock or
arrhythmias.
WARNING The monitor cannot determine whether
this alert has a physiologic cause or a cuff
application cause. Always evaluate the patient for
presence of life
message occurs.
NIBP artifact prevents a valid reading. Connect ECG
electrodes to improve NIBP measurements. (See “Improving
NIBP accuracy with Smartcuf” on page 66.)
-threatening conditions when this
Page 90
86Alarms and alertsWelch Allyn Propaq LT Vital Signs Monitor
Table 8. Alert messages
Alert typeMessagePossible cause and suggested response
Network
Communication
NIBP Fault.
No valid blood pressure found.
NIBP Fault.
Calibrating.
Please wait.
NIBP Fault.
Calibrating.
Minimize motion.
NIBP Fault.
Low battery.
NIBP disabled.
NIBP Fault.
Service required.
NIBP disabled.
NIBP Fault.
Kinked or neonate hose.
NIBP Fault.
Artifact present.
Minimize motion.
Comms Fault.
Check Acuity/network connection.
Comms Fault.
Check USB connection.
Not on Network.
Patient info entry not allowed.
The patient mode setting is incorrect or the wrong hose or
cuff is being used for the current patient mode.
The monitor periodically calibrates (zeroes) the NIBP
channel to make sure it can properly make NIBP
measurements. No NIBP monitoring can be done until the
calibration is completed. Other normal monitor operation
continues during NIBP calibration.
Motion is detected during a periodic NIBP calibration.
Minimize patient motion or motion on the cuff, or
disconnect the cuff. Motion-generated noise on the
pressure transducer can cause the calibration to continue
indefinitely.
The battery is too far discharged to operate the NIBP
channel. Insert the monitor into a powered cradle.
Have the monitor serviced.
A hose is kinked or a neonate hose is detected in the adult
patient mode. Check the hose and the patient mode
selection.
The monitor has detected too much artifact to allow
accurate readings. Take steps to reduce artifact. Position
the patient’s limb away from the body so the applied cuff is
not in contact with the patient’s body or any other object
such as a bed rail.
The monitor detects a network communication problem.
The monitor detects a problem in communication with the
cradle. Detach and reattach the USB cable.
You attempted to select the Name, ID, or Rm field on an
Acuity-enabled monitor that is not connected to the
network.
BatteryLow Battery.
Charge battery soon.
Very Low Battery.
Charge battery now.
Battery Too Low.
Shutting down.
The monitor battery charge is low, and the monitor will shut
down in 30 minutes or less. Insert the monitor into the
cradle. If no cradle is available, find an alternative method
of monitoring the patient before the monitor shuts down.
The monitor battery charge is very low; the monitor will shut
down in 5 minutes or less. Insert the monitor into a cradle or
find another way to monitor the patient before the monitor
shuts down.
The monitor battery charge is too low to support monitor
function. Monitor operation can continue only after the
battery is recharged or replaced or until the monitor is
inserted in a powered cradle.
Page 91
Directions for UseAlarms and alerts87
Table 8. Alert messages
Alert typeMessagePossible cause and suggested response
ChargerCharger Fault.
Service charger.
Charger Disabled.
Battery temperature too high or low.
Battery Fault.
Replace battery.
Service required.
The battery is too cold or too hot to charge. Normalize the
battery temperature before attempting to charge it.
The battery is missing; the battery is discharged too far to
be charged; the charger has timed out; a cell in the battery
pack is overcharged due to cell imbalance; the fuse is
blown. Service required.
SpO
2
Masimo
only
RespResp Fault.
SpO2 Fault.
No sensor detected.
SpO2 Fault.
Defective SpO2 sensor.
SpO2 Fault.
Defective SpO2 sensor/
No sensor detected.
Ambient light.Too much light on the sensor site. Verify that the sensor
Sensor off.The sensor is not properly attached to the patient; reconnect
Interference.An outside signal or outside energy is preventing the
Lead fail.
Resp Fault.
Noisy signal.
Check electrodes.
Resp Fault.
Inappropriate ECG cable.
An SpO2 sensor has been disconnected from the monitor
after being connected for more than a few seconds.
Replace the sensor.
Replace the sensor.
shield is covering the site and, if necessary, reduce the
ambient light, shade the sensor site, or replace the sensor.
it to the monitor and reattach it to the patient. If the sensor
is properly attached, it is defective; replace it.
reading. Remove the interference.
One or more electrodes have very poor or no contact. Check
for proper connection; replace electrodes if needed.
Electrodes have poor contact and might be dried out.
Replace electrodes.
The ECG cable does not contain 1 kΩ current-limiting
resistors, which are required for Resp operation and to
protect the monitor from damage during defibrillation.
Replace the cable with one of the proper type.
GeneralMultiple Faults.Multiple equipment alerts have been triggered
simultaneously.
WARNING If you acknowledge this alert message
before determining which alerts are triggered, you
cannot identify individual alerts.
Table 9. Status messages
MessageNotes
Audio alarms suspended.
Some alarms off.At least one alarm is turned off.
Press Select ( ) for controls.
NIBP off. Low battery.The battery is too low; NIBP monitoring is disabled.
Snapshot in progress.
NIBP in progress.
Page 92
88Alarms and alertsWelch Allyn Propaq LT Vital Signs Monitor
Table 9. Status messages
MessageNotes
NIBP calibrating.The monitor will resume operation when the calibration is complete.
Button pad locked.
Check SpO2 sensor.The sensor cable is faulty, incorrectly inserted, or incompatible. This message
Retry in progress.An NIBP automatic retry is in progress.
Low perfusion index.Masimo only.
SpO2 interference.Masimo only.
SpO2 sensor off.Masimo only.
SpO2 ambient light.Masimo only.
Low Signal IQ.Masimo only.
persists until a valid cable is connected or until monitor power is cycled.
Page 93
89
14040/7878
80
12
97
%
HALL, ROBERT E.
345618 7
3:00:0 6P
Adul t,
Rm 2 39
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
6
Overview
Storing and reviewing patient data
The monitor stores up to 24 hours (at one-minute intervals) of trends information for the
patient being monitored. When data storage is at capacity, the data from each new
reading replaces the data from the oldest.
Table 10. Store and review patient data
All monitorsWireless monitor
Type of dataTRENDS
• Tabular or graphical.
• Include maximum of 24 hours of trend data taken at:
1-minute intervals
NIBP and SpO
Snapshots
Spot Checks
2
SNAPSHOTS
• 21 seconds of numeric and waveform data: 14 seconds
before and 7 seconds after the snapshot request.
•Press
to take snapshots. (Maximum: 5 snapshots
per minute.)
• Maximum of 20 snapshots stored. At capacity, each
new snapshot overwrites the oldest.
Where storedMonitorAcuity via wireless communication
Up to 96 hours (depends on Acuity
system) of full disclosure data
includes:
• numerics and waveforms
• status
Where reviewedMonitor
Printer (with AutoPrint)
Capturing a data snapshot
HALL, ROBERT E.
II 1mV/cm
SpO2
80
HR/min
Press to capture a 21-second period of numeric and waveform patient data. The
monitor captures the 14 seconds preceding and the 7 seconds following the button press.
The monitor can store 20 snapshots. After 20 snapshots have been taken, each new
snapshot replaces the oldest snapshot in memory.
Snapshots and trend data can be reviewed at the monitor.
Reviewing snapshots
To review snapshots
From any main display screen (such as Large numerics, Dual waveform...):
1.Highlight , HR/PR, SpO2, NIBP, Resp, or .
2. Press.
3. Highlight Snapshots and press. The Snapshots display appears.
Figure 90. Snapshots display
STEWART, ANN
Snapshots
2/21/05
80
HR/min
80
II 1mV/cm
Adult
NIBP mmHg
n/a
#5 of 5
1 Sec after
7762940
Rm 263
Resp/min
12
12
15:14:28
97
%
SpO2
97
14:34:58
4. The monitor first displays Select. With Select highlighted, press to view the list of
up to 20 snapshots. To view another snapshot, highlight the snapshot menu (#5 of 5
in the example above), press, and highlight any of the available snapshots.
The blue markers at the bottom of the waveform display represent one-second
intervals. The numeric data above the waveform represents the patient’s vital signs
measured at a specific time relative to that indicated in the lower right corner. For
example: In Figure 90, the data was measured at “1 Sec after” 14:34:58.
5. To view the waveform and the numeric data for a specific time within the 21 seconds
of data, highlight 1Secafter and then use or scroll the display to the desired time.
For example, to see the vital signs taken five seconds after the trigger point of the
snapshot, highlight 1 Sec after and press and hold as needed to increment the
display to 5 Sec after.
Figure 91. Snapshots display: 5 seconds after the button press
STEWART, ANN
Snapshots
2/21/05
80
HR/min
80
II 1mV/cm
Adult
NIBP mmHg
n/a
#5 of 5
5 Sec after
7762940
Rm 263
Resp/min
12
12
15:14:28
97
%
SpO2
97
14:34:58
Page 95
Directions for UseStoring and reviewing patient data91
STEWART, ANN
7762940
15:14:28
Adult
Rm 263
HR/min
NIBP mmHg
Resp/min
SpO2
II1mV/cm
Snapshots
2/21/05
5 Sec after
14:34:58
80
12
97
80
n/a
12
97
%
#5 of 5
Trends
Tabular
Tab/Wave
Graphical
Snapshots
Exit
6. To change the source of the displayed waveform, see “To change the waveform
selection” on page 54.
7.To change the size of the waveform, see “To change the waveform size” on page 54.
8. From this display, you can switch to another type of display or you can exit and return
to the primary display. To do this, highlight Snapshots and press. The Trends menu
appears, from which you can exit or switch to another display—Tabular, Tabular and
Waveform, or Graphical. To switch to the split display, for example, highlight Ta b /Wave and press.
Figure 92. Snapshots display: switching to another display
STEWART, ANN
Trends
Snapshots
Tabular
Tab/Wave
Graphical
II 1mV/cm
Snapshots
Exit
2/21/05
Reviewing trends
To review trends
From any main display screen (such as Large Numerics, Dual Waveform...):
1.Highlight , HR/PR, SpO2, NIBP, Resp, or .
2. Press.
3. Highlight Trends and press. The Trends display appears.
Use and to scroll through the list of trends.
80
HR/min
80
7762940
Adult
NIBP mmHg
n/a
#5 of 5
5 Sec after
Rm 263
12
Resp/min
12
15:14:28
97
%
SpO2
97
14:34:58
•Red indicates an alarm.
• indicates that a snapshot exists for this trend. Highlight and press to
view the snapshot.
• indicates that the snapshot for this trend was overwritten with a more recent
snapshot and is not available for viewing.
•--- indicates a value below the monitor’s measurement range.
•+++ indicates a value above the monitor’s measurement range.
•If the monitor is in communication with an Acuity Central Station, press to print
vital signs at the Acuity system printer.
•If a PC with the Autoprint Utility is connected to a cradle with the USB option, you
can print directly to a printer connected to the PC simply by inserting the monitor
in the cradle.
This chapter tells you how to print with the Autoprint Utility.
The AutoPrint utility uploads from the monitor to the PC, either manually or automatically,
trend data and snapshots stored by the monitor for the current or most recent patient.
The PC then prints the data on the configured laser printer.
Note
Note
Use only a laser printer. The monitor can store a large amount of data. If you use
any other type of printer, it can take a long time to print all of the data.
If you attempt to print patient data during the first minute of monitoring, before
the monitor has recorded a snapshot or generated a trend, the monitor might
display the message, "File transfer failure". If this occurs, cancel the AutoPrint
operation and continue monitoring.
Note
Note
Note
Note
The following message might appear when AutoPrint starts:
“To help protect your computer, Windows Firewall has blocked some features of
this program. Do you want to keep blocking this program?”
To prevent this message from appearing again, select Unblock.
AutoPrint works only when the monitor is powered on. This feature helps you be
certain that only the data for the current patient is being printed or, if no patient is
being monitored, that only the data for the last patient monitored is being printed.
If the monitor is off when it is placed in the cradle, AutoPrint is disabled. To enable
AutoPrint, turn the monitor on and select Continue Patient.
If you turn the monitor on and select Start New Patient, all stored patient data is
deleted from the monitor and cannot be printed.
•Up to 24 hours of HR/PR, RR/BR, and SpO
row, with a 15-minute interval between rows).
•SpO
In the Snapshots printout:
•Up to twenty snapshots (21-second intervals of full-disclosure patient data,
•The utility extracts the available ECG Leads from the patient data and prints a
The utility supports US Letter and European A4 page sizes.
To print the patient data stored in a monitor
1.Install the AutoPrint Utility on a PC.
a. Insert the Propaq LT Monitor PC Utility Program in the CD-ROM drive of the PC.
spot checks and NIBP measurements, interspersed among the rows of
2
trends data.
including waveforms and numerics).
page of strips for each snapshot. Depending on the number of ECG Leads
available, the strips may contain 21 seconds of data for one or three Leads of
ECG data.
trends (one set of measurements per
2
b. Follow the on-screen instructions to install the AutoPrint Utility.
2. Connect the PC to a cradle (Propaq LT Monitor Charging/Communications Cradle),
using the supplied USB cable.
Each time a monitor is inserted into this cradle while the cradle is connected via USB
cable to the PC, the patient data stored in the monitor is printed on the default printer.
Patients monitored continuously for 24 hours typically generate enough data for a 2or 3-page trends table of two panels per page. For example, a 24-hour report with the
following data (192 rows) would cover three pages:
•96 rows of vital-signs data (4 intervals/hr = 4 rows per hour) for 24 hours
•48 NIBP readings (2/hr)
•48 SpO
Figure 98 shows a typical printout.
•Text in italic font indicates one of the trends measurements taken at 15-minute
intervals.
•Text in bold font indicates an NIBP measurement or an SpO
•White text in a black box indicates an alarm condition—a measurement that is
outside of alarm limits.
--- indicates a measurement value below the monitor’s measurement range.
spot-check readings (2/hr)
2
spot check.
2
+++ indicates a measurement value above the monitor’s measurement range.
??? indicates an undetermined measurement value.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.