Copyright 2007, Welch Allyn Inc. All rights are reserved. No one is
permitted to reproduce or duplicate, in any form, this manual or
any part thereof without permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or
for any illegal or improper use of the product, that may result
from failure to use this product in accordance with the
instructions, cautions, warnings, or statement of intended use
published in this manual.
Welch Allyn
®
is a registered trademark of Welch Allyn, Inc.
Printed in USA
:
USA + 1 315 685 4560
800 535 6663
Canada 800 561 8797China + 86 216 327 9631
European Call Center + 353 46 906 7790France + 331 6009 3366
Germany + 49 747 792 7186Japan + 81 33 219 0071
Latin America + 1 305 669 9003Netherlands + 31 15 750 5000
Thank you for purchasing the Welch Allyn® LS200™ Procedure
Light, designed by a leading worldwide manufacturer of
illuminated diagnostic instrumentation. By following the use and
care guidelines given in this manual, you will be rewarded with
years of dependable, trouble-free service from your new LS200
Procedure Light. Read these instructions thoroughly before use.
Symbol Descriptions
Hazard or Warning
Attention, Caution: Consult user’s manual for
more information.
A warning statement in this manual identifies a condition or
practice, which, if not corrected or discontinued immediately,
could lead to patient injury, illness, or death.
WARNING All users of this procedure light should be
thoroughly trained in the medical procedures appropriate
to the equipment. Furthermore, they should read and
understand the instructions for all other equipment used
in conjunction with the LS200 Procedure Light.
WARNING Lamps are extremely bright. DO NOT stare
directly into lamps when lit.
WARNING DO NOT position luminaire in the upward
direction with the lamps on.
WARNING Risk of explosion if used in the presence of
flammable anesthetics.
WARNING The bezel in front of the lamps may be hot.
Directions for UseChapter 1 Introduction3
Caution
A caution statement in this manual identifies a condition or
practice which, if not corrected or discontinued immediately,
could lead to equipment failure, equipment damage, or data loss.
Caution Remove power cord from electrical outlet and
allow lamps to cool before replacing (replace with Welch
Allyn Lamp #06400 only). Accessible metal parts of this
unit are electrically isolated from the grounding
conductor of the supply cord. If grounding of the
accessible parts is considered necessary, a separate
grounding lead should be employed.
Caution To reduce the risk of electric shock, do not
disassemble unit. Refer servicing to qualified service
personnel.
Caution Operation at closer than 12"
(30.5 cm) for extended periods may result
in erythema. The designed working
distance is 24" (60 cm).
Caution The switch on luminaire is NOT a mains
switch. Transformer energizes when the light is plugged
to mains.
1. Remove stand base, pole and hardware kit from boxes.
2. Place stand base on floor with casters down and lock casters.
3. Insert pole end (A) into hole in base of mobile stand.
4. Rotate pole until it aligns with pin and drops into place.
5. Insert cap-head bolt (F) with lock washer (f) through stand
cross member into pole and tighten with 1/4" Allen wrench.
5
6. Carefully remove LS200 arm/luminaire assembly from box.
Hook transformer onto the factory installed mounting screws
(C) on the pole and slide down until secure.
7. Place arm/luminaire assembly on pole by sliding universal
connector (B) onto top end of the pole and align screw holes.
8. Insert 0.5" (13mm) flathead screw (b) into hole on side of
universal connector and tighten with 5/32" Allen wrench.
9. Push power cord attached to transformer into power cord
clips (D) on pole, ensuring enough slack for full range of arm
motion.
10. Attach power cord (e) to transformer.
11. To lock casters, push down on the tab on the locking casters.
6Chapter 2 Mounting and Operating InstructionsWelch Allyn LS200 Procedure Light
Before moving the LS200 Procedure Light, see “Transport For
Mobile Stand Unit” on page 13.
Caster locks
Align pin in mobile stand with pole end
Directions for UseChapter 2 Mounting and Operating Instructions7
Wall Mount Set-up
Caution DO NOT mount into sheetrock. Mount the
LS200 Procedure Light Arm into a wall stud.
The LS200 Procedure Light arm is designed not to go beyond 90°
and the working distance is 24" (60cm). To ensure proper
placement on the wall, locate the top edge of the wall plate 4"
(10cm) above the working surface.
Note
To mount to a wall stud:
1. Remove wall mount assembly and hardware kit from box.
2. Select mounting location on a stud in the wall that is within 8'
3. Using wall plate as a template, place on selected location on
4. Remove wall plate and drill pilot holes using 0.12" (3mm) drill
Use a carpenter’s level to ensure precise, drift-free
movement of the wall plate.
(2.4m) of electrical outlet.
wall, level plate and mark both holes.
bit.
8Chapter 2 Mounting and Operating InstructionsWelch Allyn LS200 Procedure Light
5. Place wall plate
on wall and drive
top 1.5" (3.8cm)
wood screw
halfway into
wood stud.
6. Align wall plate
using a level and
secure at bottom
with 1.5" (3.8cm)
wood screw.
7. Tighten top wood
screw.
8. Prepare for transformer mounting by following “Mounting The
Transformer” on page 11 instructions through step 4.
9. Carefully remove LS200 arm/luminaire assembly from box.
Complete transformer mounting according to steps 5 and 6 of
instructions on page 11.
10. Place arm/luminaire assembly on
extension arm by sliding universal
connector (B) onto extension arm
(G) and aligning screw holes.
11. Insert 0.5" (13mm) flathead
screw (b) into hole on side of
universal connector and tighten.
Move arm up and down to check
that mounting is secure.
12. Push cord into clips in underside
of extension arm.
Directions for UseChapter 2 Mounting and Operating Instructions9
To mount to a concrete or brick wall:
Caution DO NOT use the center two holes in the wall
plate for concrete walls. Additional holes must be drilled
for masonry surface mounting to assure durable
attachment.
1. Remove wall mount assembly and hardware kit from box.
2. Drill holes in wall plate:
a. Place the wall plate flat on a wooden surface.
b. Mark four holes on wall plate 1" (2.5cm) from each edge in
corners of plate.
c. Use a center punch to start holes.
d. Drill four holes using a 13/64" (5.1mm) drill bit
3. Select mounting location on wall that is within 8' (2.4m) of
electrical outlet.
4. Using wall plate as a template, place on selected location on
wall and mark upper-left hole.
5. Remove wall plate and use a center punch to start hole.
6. Using a hammer drill and 5/32" (4mm) masonry drill bit, drill
hole to a minimum of 1.5" (4cm) deep.
10Chapter 2 Mounting and Operating InstructionsWelch Allyn LS200 Procedure Light
7. Place wall plate on wall and drive 1.25" (3.2cm) masonry
screw halfway into the wall.
8. Align wall plate using a level and secure by tightening screw.
9. Punch the next three holes with wall plate in place.
10. Drill remaining holes.
11. Fasten wall plate with three remaining screws.
12.Prepare for transformer mounting by following “Mounting The
Transformer” on page 11 instructions through step 4.
13.Carefully remove LS200 arm/luminaire assembly from box.
Complete transformer mounting by following steps 5 & 6 of
instructions on page 11.
Directions for UseChapter 2 Mounting and Operating Instructions11
14.Place arm/luminaire assembly on extension arm by sliding
universal connector (B) onto extension arm (G) and aligning
screw holes.
15.Insert 0.5" (13mm) flathead screw (b) into hole on side of
universal connector and tighten. Move arm up and down to
check that mounting is secure.
16.Push cord into clips in underside of extension arm.
Mounting The Transformer
1. Select proper mounting hardware from hardware kit in the
LS200 mounting option box: Choose 1" (2.5cm) wood screws
for mounting into wood stud and 1" (2.5cm) sheet metal
screws and plastic anchors for concrete or brick.
2. Select a location for the transformer that is within 8' (2.4m) of
an electrical outlet and within 1.5' (0.5m) of the wall mount
plate to allow enough slack for full range of arm motion.
12Chapter 2 Mounting and Operating InstructionsWelch Allyn LS200 Procedure Light
3. Using the template indicators on left side of this page (3 1/8"
[8cm] on center), mark the location for drill holes on the wall.
Make sure they are vertical.
3 - 1/8”
8cm
For wooden walls:
No pilot hole is needed. Use 1.5" 3.8cm) wood screws
from transformer hardware kit.
For concrete walls:
Using a hammer drill and 3/16" (2.3mm) masonry drill bit,
drill holes for plastic anchors to a minimum of 1" (2.5cm)
deep. Tap in plastic anchors from transformer hardware kit
until flush.
Directions for UseChapter 2 Mounting and Operating Instructions13
4. Drive screws into the wall, allowing screw heads to protrude
11/32"(2.3mm).
5. Mount transformer by fitting round holes in the back of
transformer over mounting screws on wall, then slide
downward until secure.
6. Attach power cord to transformer.
Operating Instructions
Transport For Mobile Stand Unit
1. To prevent the light from tipping, position arm at a 45° angle
or greater, when rolling the LS200 to a new place.
2. Center arm in range of motion.
3. Unlock casters by lifting up on the tabs and hold unit at top of
pole while moving it.
14Chapter 2 Mounting and Operating InstructionsWelch Allyn LS200 Procedure Light
Arm Adjustment
The LS200 Procedure Light has been designed to allow the user
to adjust the amount of resistance at the joints of the arm. The
main joint features a large knob for easy adjustment of this key
joint. Turn knob to achieve required tension.
Note
The top and lower joints have been preadjusted at Welch Allyn.
Should further adjustment be necessary, the joint resistance can
be increased or decreased by using a screwdriver to adjust the
set screws at point (A) or (B).
There is no need to adjust knob for every use. The set
tension will be maintained over multiple movements.
NEVER completely unscrew large knob.
3
Maintenance
For minor trouble, see “Troubleshooting” on page 16 for possible
causes and corrective action. Only qualified personnel should
make electrical inspections of the LS200 Procedure Light
Repair: To locate qualified service personnel, contact your local,
authorized Welch Allyn distributor, or contact Welch Allyn directly
at 800-535-6663.
Caution Cleaning and disinfection should comply with
local authorities responsible for hygiene and disinfection.
Caution Do not to allow solution to drip into luminaire.
Caution DO NOT spray solution directly into joints.
Caution DO NOT sterilize any part of the unit.
Caution DO NOT immerse any part of the unit in
cleaning solutions.
Turn power switch to the OFF position and unplug the power
cord from the electrical outlet prior to cleaning.
You can wipe the entire unit down using a cloth dampened
slightly with a mild solution of detergent and water or glass
cleaner. Do not to allow the plug prongs to get wet. Wipe the unit
dry with a clean, dry cloth.
DO NOT plug the LS200 Procedure Light into the electrical outlet
until the unit is thoroughly dry.
Directions for UseChapter 3 Maintenance19
Lamp Replacement
WARNING The lamps operate at high temperatures.
DO NOT attempt to replace lamps while they are hot.
Wait until the lamps are cool. Lamps may shatter under
pressure, handle with care. Replace with Welch Allyn
lamp #06400 only.
1. Turn power switch to OFF position and unplug the power cord
from the electrical outlet.
2. Rotate luminaire so that the bezel and handle are facing you
(see A).
3. Turn handle counterclockwise to unscrew (see B). Remove
handle and carefully slide bezel out and allow to rest on wire
tether (see C).
Note
There are (3) color correction filters visible on the bezel
assy once it is removed. These filters also have heat
reduction properties. Do not remove.
4. Locate release lever on lamp holder and push downward (see
D).
5. Grasp expired lamp by its base with two fingers and slide out
of lamp holder (see E).
6. Return release lever to upward position (see F).
7. Remove Welch Allyn Replacement Lamp #06400 from its
package.
8. Grasp replacement lamp between your two forefingers (see
G).
Note
Try not to touch the inside of reflector—dirt or oil can
cause reduced light output. Clean according to
instructions on page 18, if necessary.
Directions for UseChapter 3 Maintenance21
9. Carefully slide the lamp into the lamp holder with two fingers,
making sure the two pins on the base of the lamp slide into
the socket and the reflector slides under the spring (see H).
10. To assure proper placement in socket, gently push lamp into
socket until completely seated and snap into place (see I).
11. Replace bezel. Be certain to align the three smooth
“windows” in bezel with lamps and tabs on edge of bezel
with grooves in white housing (see J).
12.Replace handle and rotate clockwise until secure (see J).
There are two replaceable fuses located in the fuse drawer of the
transformer housing. If either one of these blows, the unit will
not light. Replace only with Welch Allyn part number
236706-3907 for model 44200 or 236706-3114 for models 44202,
44204 and 44206.
1. Turn the power switch to the OFF position and unplug the
power cord from the electrical outlet and transformer.
2. Using a small (1/8") screwdriver, remove the fuse drawer (a)
from the transformer housing by releasing the spring loaded
snap connection (b) on both sides of the component.
3. Pull out fuse drawer.
4. Remove and replace the blown fuse(s). There is no required
orientation of the fuses.
5. Reinsert the drawer by pressing it into the connector until it
snaps into place.
4
Specifications
Electrical Specifications
Wiring Diagram:
A HOSPITAL GRADE POWER CORDD ON-OFF SWITCH
B FUSESE LAMPS
C TRANSFORMER - STEP DOWN,
ISOLATED
Welch Allyn warrants LS200 Procedure Light, when new, to be
free of defects in material and workmanship and to perform in
accordance with manufacturer's specifications for a period of
three years from the date of purchase from Welch Allyn or its
authorized distributors or agents. Welch Allyn will either repair or
replace any components found to be defective or at variance
from manufacturer's specifications within this time at no cost to
the customer. It shall be the purchaser's responsibility to return
LS200 Procedure Light to Welch Allyn or an authorized
distributor, agent, or service representative. This warranty does
not include breakage or failure due to tampering, misuse,
neglect, accidents, modification, or shipping. This warranty is
also void if the instrument is not used in accordance with
manufacturer's recommendations or if repaired by other than
Welch Allyn or an authorized agent. Purchase date determines
warranty requirements. No other express warranty is given.
27
28 WarrantyWelch Allyn LS200 Procedure Light
4341 State Street Road
Skaneateles Falls, NY 13153-0220 U.S.A.
Website: www.welchallyn.com
P.O. Box 220
Telephone: 315-685-4560
FAX: 315-685-3361
Part # 703343 Ver. CPrinted in U.S.A.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.