LE HOT. WAIT UNTIL LAMP IS COOL.
WHI
Lamp Replacement
1. Exposing the Lamp (See Figure #1)
Turn power switch to the off (“0”) position and unplug power cord from outlet
.
Rotate Light Assembly so that the lamp points up (towards ceiling).
Completely loosen screw in Lamp Housing and lift Lamp Housing to expose lamp.
2. Freeing the Lamp and Lamp Socket (See Figure #2)
Unclip the Lamp Retaining Wire from the Lamp Housing by pushing on wire “ears” towards center and swing out as shown to free the Lamp and Lamp Socket.
3. Removing the Lamp Socket (See Figures #3 and 4)
Hold the Lamp reflector by its outer rim. Grasp the white sides of the lamp Socket as shown (Figures 3 and 4) pulling on the Lamp Socket while rocking it
slightly to remove it from the Lamp. Do not pull on Lamp Socket Wires.
4. Install the new Lamp
Verify that the lamp number located on the metal plate inside the Light Assembly (Figure 2) matches the replacement lamp number.
Replace Lamp and reattach Lamp Retaining Wire to Lamp Housing.
Align Lamp pins to Lamp Socket holes and reconnect Lamp Socket.
Reinstall Lamp Housing onto Light Assembly by first engaging Lamp Housing
tab (opposite screw) into Light Assembly. Fully tighten screw.
Reconnect Power cord and verify Lamp operation.