Welch Allyn Diatek 600 User manual

Diatek® Model 600 Thermometer
Thermo Menu
Operation Manual
Technical Manual
Diatek is a trademark of Diatek Instruments, Inc. All rights reserved.
Model 600
Technical Manual

TABLE OF CONTENTS

WARRANTY.............................................................................................................................3
MODEL 600 OPERATIONAL CHARACTERISTICS................................................................4
Normal Mode.........................................................................................................................4
Monitor Mode.........................................................................................................................4
Pulse Timer............................................................................................................................4
Backlight................................................................................................................................4
Error Indicators......................................................................................................................5
Broken Probe .................................................................................................................................5
Low Battery ....................................................................................................................................5
Instrument Malfunction...................................................................................................................5
Self-Tests...............................................................................................................................5
Battery Installation Display Test.............................................................................................5
Thermometer Startup Tests...................................................................................................5
Display and Horn Test....................................................................................................................5
Internal Calibration Test .................................................................................................................5
Internal Microprocessor Test..........................................................................................................6
Battery Life.............................................................................................................................6
Shelf Life........................................................................................................................................6
Temperature Taking Without Backlight...........................................................................................6
Temperature Taking With Backlight ...............................................................................................6
C/F Mode Switch....................................................................................................................6
Normal/Monitor Mode Switch.................................................................................................6
MODEL 600 THEORY OF OPERATION..................................................................................7
Technical Overview................................................................................................................7
Temperature Measurement and Display................................................................................7
Probe Temperature Measurement.........................................................................................8
Monitor Mode Operation........................................................................................................8
Normal Mode Operation ........................................................................................................8
SELF TESTS.........................................................................................................................9
Internal Calibration Test .................................................................................................................9
Internal Microprocessor Memory Test.............................................................................................9
Broken Probe Test .........................................................................................................................9
DISPLAY................................................................................................................................9
Display Voltage Reference.............................................................................................................9
Display............................................................................................................................................9
Horn ...............................................................................................................................................9
Backlight.........................................................................................................................................9
Power Consumption ............................................................................................................10
Low Battery Detection..........................................................................................................10
Pulse Timer Operation.........................................................................................................10
SPECIFICATIONS..................................................................................................................18
Model 600 1
Diatek, A Welch Allyn Company
OPERATING INSTRUCTIONS...............................................................................................19
Set-up Procedure.................................................................................................................19
Preparation for temperature measurement:.........................................................................19
Normal Mode Operation ......................................................................................................19
Oral Temperature Measurement ..................................................................................................19
Rectal Temperature Measurement...............................................................................................20
Monitor Mode Operation......................................................................................................20
Backlight Operation .............................................................................................................20
Pulse Timer..........................................................................................................................20
°F/°C Conversion.................................................................................................................20
Error Indicators....................................................................................................................21
Broken Probe ...............................................................................................................................21
Low Batteries................................................................................................................................21
Probe Position Error.....................................................................................................................21
Malfunction...................................................................................................................................21
Self Tests.............................................................................................................................21
Initial Power Up Display................................................................................................................21
Calibration Check.........................................................................................................................21
Microprocessor Self Check...........................................................................................................22
Cleaning and Sterilization....................................................................................................22
Battery Replacement...........................................................................................................22
Calibration Key Procedure...................................................................................................22
Test Procedure.............................................................................................................................23
M9600 Procedure................................................................................................................24
MODEL 600 TROUBLESHOOTING.......................................................................................24
PARTS LIST (units with serial numbers equal to or greater than 193309) ......................26
PARTS LIST (units with serial numbers less than 193309)...............................................27
RECOMMENDED SPARE PARTS FOR MODEL 600 ...........................................................28
MODEL 600 PREVENTATIVE MAINTENANCE ....................................................................29
TABLE OF FIGURES
Figure 1 - Block Diagram........................................................................................................11
Figure 2 - Schematic...............................................................................................................12
Figure 3 - Schematic (earlier serial numbers).........................................................................13
Figure 4 - Printed Circuit Board Component Layout...............................................................14
Figure 5 - Final Assembly.......................................................................................................15
Figure 6 - Operating Controls and Components.....................................................................17
2 Model 600
Technical Manual

W ARRANTY

1-YEAR LIMITED WARRANTY ON NEW INSTRUMENTS
Diatek Instruments, Inc. (Diatek) is warranted to be free from original defects in material and workmanship under normal use and service for a period of one year from the date of first shipment from Diatek. This warranty shall be fulfilled by Diatek or its authorized representative repairing or replacing at Diatek's discretion, any such defect, free of charge for parts and labor.
Diatek should be notified via telephone of any defective product and the item should be immediately returned, securely packaged and postag e prepaid to Diatek. Loss or dam age in shipment shall be at purchaser's risk.
Diatek will not be responsible for loss associated with t he use of any Diatek product that (1) has had the serial number defaced, (2) has been repaired by anyone other than an authorized Diatek Service Representative, (3) has been altered, or (4) has been used in a manner other than in accordance with instructions.
The information in this manual has been car efully reviewed and is believed to be accurate; however, no responsibility is assumed for inaccuracies. Furthermore, this information does not convey to the purchaser of Welch Allyn or Diatek devices any license under the patent rights to the manufacturer.
: Instrumentation purchased new from
Model 600 3
Diatek, A Welch Allyn Company

MODEL 600 OPERATIONAL CHARACTERISTICS

The Diatek Model 600 Digital Thermometer is a portable instrument for measuring patient temperatures over the 84°F to 108°F (28.9°C to 42.2°C) range. The Model 600 features an LCD readout with backlight, interchangeable probes, disposable prove covers, disposable batteries with a capacity for up to 45,000 temperatures, automatic power shutdown, both normal and monitor modes of operation and self-calibration/test capability.

Normal Mode

Removal of the probe automatically turns on the Model 600 and places it in the predict mode when selected by the normal/monitor slide switch. All segments of the display except the pulse timer are lighted simultaneously for 6 seconds as a display test and the horn is tested for 0.1 seconds. The display will read 84.0 (28.9) with the down arrow on until the probe exceeds that temperature and will ten display a rising temperature for about 30 seconds.
When the readout has stabilized, the horn will sound for 1 second, the “F” or “C” symbol will light and the display will remain fixed indicating the end of the temperature taking cycle. The displayed temperature will be the actual probe temperature plus a computed correction factor. The probe temperature is sensed about 4 seconds from turn on to correct for ambient probe temperature variations. If the temperature measurement cycle is not completed within 5 minutes from turn on, the thermometer will turn off at that time. A temperature in excess of 108°F (42.2°C) will cause an error condition to be indicated by the horn beeping twice every 10 seconds and the up arrow blinking. Five minutes of this error condition causes the thermometer to turn off. Returning the probe to its holder and then removing it will restart the thermometer.
Note: The normal mode display is peak reading and will not read down.

Monitor Mode

Removal of the probe automatically turns on the Model 600 and places it in the monitor mode when selected by the normal/monitor slide switch. All segments of the display except the pulse timer are lighted simultaneously for 6 seconds as a display test and the horn is tested for 0.1 seconds. The display will then indicate the actual probe temperature, updating every 1.5 seconds as long as it remains within the 84°F to 108°F (28.9°C to 42.2°C) range, and will show the appropriate up or down arrow when outside this range. Five minutes of operation outside the operating range will cause the thermometer to turn off.

Pulse Timer

Pressing the pulse timer switch at any time causes the 30 second clock type display to start. One additional segment is turned on each second until 30 seconds has been reached at which time the timer display goes blank. The horn sounds for 0.1 second at 0,15, and 30 seconds. Pressing the timer switch while the timer is running causes the timer to restart. The switch will not respond if pressed sooner than 1.5 seconds since the last timer activation.

Backlight

The backlight is activated as long as the backlight switch is pressed and the thermometer function is active. In the predict mode, activating the backlight any time prior to the end of the temperature taking cycle will cause the backlight to turn on for 5 seconds when the final temperature has been reached and the horn sounds. No backlight is available when only the pulse timer is in use.
4 Model 600
Technical Manual

Error Indicators

Probe Position Error—During normal mode operation the visual probe position error indicator will be displayed as long as tissue contact is broken and the probe temperature is falling. This display indicates that a condition exists which may not provide accurate temperatures. The error indicator does not affect the temperature taking cycle in any way.
No temperatures may be taken when any of the following error conditions exist:

Broken Probe

A probe error will be indicated if the probe circuit is open (below 10° greater than 60,000 ohms) or shorted (above 150° F or less than 2,000 ohms). The horn sounds a double beep repeating every 10 seconds with the probe error display flashing. The unit will shut off automatically after 5 minutes of error indication. The error will be detected any time during predict or monitor mode operation.

Low Battery

A low battery condition exists when the battery voltage drops to approximately 3.0 volts. A low battery indication may occur any time during monitor mode operations but only at the start of a predict mode temperature. If a low battery condition exists prior to starting the pulse timer, the low battery error will be displayed when the timer is started and any normal mode temperature in progress will abort at that time. The low battery error displays a flashing symbol with a double horn beep sounding every 10 seconds. The unit will shut off automatically after five minutes of error indication.

Instrument Malfunction

A malfunction error will be indicated if either the internal calibration self-test or the internal microprocessor self-test fails. The malfunction indicator flashes with a double horn beep sounding every 10 seconds. The thermometer will automatically shut off after five minutes of error indication.

Self-Tests

Battery Installation Display Test

Immediately following the installation of batteries into the Model 600, a one time only display test is provided. All segments of the entire display (including the pulse timer and error indicators) are sequentially lighted for 0.4 seconds. The entire test takes 25 seconds and must occur prior to starting any temperature or timer functions.

Thermometer Startup Tests

Display and Horn Test

Each time the probe is removed from its holder all display segments (except for the pulse timer) are lighted for 6 seconds as a visual display test. During the display test the horn sounds for 0.1 second.

Internal Calibration Test

During each thermometer startup display test a precision resistor is used to check the thermometer calibration at 100.5°F (38.1°C). A failure of the thermometer to measure the calibration resistor within ±0.2°F will cause the malfunction error to flash, the horn will sound with a double beep every 10
Model 600 5
Diatek, A Welch Allyn Company
seconds and no temperature measurement will be allowed. The thermometer will automatically shut off after 5 minutes of the error display.

Internal Microprocessor Test

During each thermometer display test a check sum of all program memory is calculated and compared against a reference stored in the program memory. An error in the check sum will cause the malfunction error to flash, the horn will sound with a double beep every 10 seconds, and no temperature measurement will be allowed. The thermometer will shut off after 5 minutes of the error display.

Battery Life

Shelf Life

The shelf life of 70°F. The storage life degrades rapidly with increased temperature.
carbon zinc chloride
or
alkaline
batteries should be at least three years when stored at

Temperature Taking Without Backlight

The Model 600 should operate continuously for about 2,000 hours when the backlight is not used. If an operating time of 2 minutes per temperature is assumed then the potential number of temperatures that can be taken should be about 2,000 hours @ 30 temperatures/hour = 60,000 temperatures.

Temperature Taking With Backlight

The backlight uses about 20 times more power than the rest of the thermometer. If 20 seconds of backlight usage is assumed for each temperature taken then the number of temperatures that can be taken should be about 17,300 temperatures.

C/F Mode Switch

The display mode may be changed between Fahrenheit and Celsius at any time prior to, during, or after taking a temperature by activating the C/F slide switch. The numbers will change as the switch is activated, however, the C/F indicator is only visible at the end of a normal mode temperature.

Normal/Monitor Mode Switch

The selection of normal or monitor mode operation should be made prior to turning on the thermometer by setting the normal monitor switch to the desired position. Once the thermometer has been turned on, a change from normal to monitor mode may be made at any time, however, a change from monitor to normal will cause the horn to beep twice and the thermometer to shut off. This error indication and shut off prevents a normal mode temperature from being started with the probe already in place.
6 Model 600
Technical Manual

MODEL 600 THEORY OF OPERATION

Note: Refer to Model 600 Thermometer Block Diagram and Model 600 Schematic Diagram.

Technical Overview

The heart of the Model 600 is comprised of two custom integrated circuits which provide the microprocessor and analog circuit functions. All control and display functions are governed by the microprocessor (U2) and all analog interfacing to the microprocessor, probe, horn and backlight is provided by U1. Probe resistance measurements are made by ratioing pulse widths generated by sequentially switching in two calibration resistors and the probe thermistor. These pulse widths are measured by the microprocessor which calculates the probe resistance. The actual probe temperature is then calculated from the probe resistance.
During monitor mode operation, the actual probe temperature appears directly on the liquid crystal display which is driven by the integral LCD driver on the microprocessor.
During normal mode operation, the shape of the rising temperature curve is monitored and a continuously computed correction factor is added to the actual probe temperature. The normal mode temperature cycle is terminated when the predicted temperature remains stable.
All switch inputs are monitored by the microprocessor which in turn activates the required functions. Inputs to the microprocessor are A/D output pulse widths, switches, low battery detection, power on
reset, microprocessor clock and interval timer clock. Outputs from the microprocessor control the display, analog IC and display power, backlight, horn and
calibration resistor selection. To extend battery life, the power to both the analog lC (U1) and the LCD reference are turned off and
the microprocessor is put to sleep when possible. Power to the thermometer is always connected and is not turned on and off by the probe or pulse
switch functions.

Temperature Measurement and Display

When batteries are installed in the thermometer, C4 and RN1-B provide a power on reset pulse to the microprocessor (U2-56). C3 and RN1-A set the microprocessor clock frequency at approximately 200 kHz. Following the high going reset pulse, the microprocessor initiates the power on display test which causes display segments (DS1 ) to be sequentially lighted for 0.4 seconds and then extinguished. After the completion of the display test which lasts approximately 25 seconds, the microprocessor then goes into a sleep mode awaking 8 times per second to test the probe switch S1. If the probe is in the thermometer (S1 open) or the probe is disconnected, then the microprocessor returns to sleep for another 1/8 of a second. The pulse timer switch (S5) is connected to an interrupt input on the microprocessor (U2-63) and will wake up the microprocessor and start the 30 second pulse timer.
In the sleep mode and even when only the pulse timer is active, the 200 kHz clock (U2-59) will be seen to turn on for about 2 milliseconds every 1/8 of a second. If when tested, the probe switch circuit is found to be active, the microprocessor turns on display power by setting U2-10 low and initiates a 6 second display test which turns on all display segments except the 30 second pulse timer simultaneously. During the display test the following events take place.
1. The A/D power control line (U2-3) is set high by the microprocessor to turn on the power to U1.
Model 600 7
Diatek, A Welch Allyn Company
2. The low battery detector output (Ul-9) is tested by the microprocessor. A low logic level on Ul-9 indicates a low battery condition (less than 3.0 volts).
3. The horn drive signal (U2-11) goes high activating the horn test for 0.1 second.
4. The backlight switch, normal/monitor switch and F/C switches are tested by the microprocessor.
5. A calibration cycle is performed.
6. The calibration test is performed.
7. The microprocessor memory test is performed.
8. The probe resistance is tested for an open or shorted condition.
Following the events which take place during the display test, a continuous monitor mode temperature measurement or a predict mode temperature taking cycle is started.

Probe Temperature Measurement

Q3-Q6, Q8, R3-R5, R-16, C2 and U1 comprise a unique resistance to pulse width conversion circuit which allows any one of 4 resistance’s to be measured by the microprocessor. The circuit allows the microprocessor to select precision resistors with equivalent temperatures at opposite ends of the temperature measurement range and measure their relative pulse widths.
The microprocessor sets U2- 16 to 19 high as required to select the appropriate resistance to be measured. R3 provides a pulse width calibration at 93.2 °F, R5 provides calibration at 106.9 °F and R3 is used for a calibration test at 100.5 °F. A ratiometeric calculation is used to compute the probe resistance from its pulse width relative to those provided by R3 and R5. The accuracy of each calibration resistor is .05 °F. A resistance to pulse width conversion is initiated by the microprocessor setting U2-64 high for approximately 0.5 seconds with the appropriate FET switch selected sometime within that period. U2-64 is then set low and the time from this transition until the A/D output (U1-2) goes high is measured by the microprocessor. The crystal X1 provides the time base for the pulse width measurements. The probe temperature is computed using the probe resistance value obtained from the ratiometeric pulse width calculations.

Monitor Mode Operation

In the monitor mode a new probe temperature measurement is completed every 1.5 seconds and a new calibration cycle is completed every 3 seconds. The computed probe temperature is displayed directly on the liquid crystal display and will follow both upward and downward movement.

Normal Mode Operation

In the normal mode. a new probe temperature measurement is completed every 1.5 seconds. At least one calibration cycle is performed at the start of a predict cycle and continue every 3 seconds until the probe temperature rises above 84.0 °F. The calibration cycle is stopped at this point to reduce quantizing errors associated with the measurement of the slowly changing probe temperature.
The normal mode displayed temperature is the sum of the actual probe temperature and a computed correction factor which is based on the shape of the changing temperature curve at that point in time. The values of the correction factor can vary from 0 °F to 2.3 °F with a typical value when the horn sounds of about 1 °F. During the display test prior to starting a normal mode temperature cycle, the ambient temperature of the probe is measured and is used in the correction factor computation to adjust for differing probe starting conditions. The normal mode display is peak reading and therefore the highest temperature is displayed even though the present predicted probe temperature may drop.
8 Model 600
Loading...
+ 22 hidden pages