Thank you for purchasing this transformer and/or light source for use with
Welch Allyn diagnostic instruments.
The specifications and maintenance
instructions found in this folder should
be followed to ensure many years of
convenient service.
Vaginal/Endoscopic
Diagnostic Instrument Systems
GENERAL
INSTRUCTIONS
Connect light handle
to the No. 73305 transformer. Lamps
in the 73210 and 78000 will illuminate
immediately when connected to the
transformer (optional No. 73308 on-off
switch [formerly 733071] and No. 73400
rheostat controls are available; see
current Welch Allyn catalog).
[73210
or 78000]
No. 56600
No. 56601
No.
58000
No. 53110
78600
No. 76000
No. 73305
REPAIR
For Repairs Send Unit To:
Welch Allyn
Welch Allyn, Inc.
Technical Service Dept.
4341 State Street Road
Skaneateles Falls, N.Y. 13153-0220
Telephone: (315) 685-4577
LAMP REPLACEME
NOTE: If lamp surface is touched or
becomes smudged, wipe clean with a
soft cloth dampened with alcohol. This
will assure longer lamp life and higher
output.
CAUTION: LAMPS BECOME HOT
DURING OPERATION. ALLOW TO
COOL BEFORE REPLACEMENT.
No. 78000 (Vaginal Speculum
Light Handle)
Remove lamp by unscrewing counterclockwise. Replace with Welch Allyn
lamp No. 07800 (formerly No. 00800).
No. 78800 (Series 58600
Vaginal Specula)
This illuminator is used in conjunction
with 58600 and 58601 Disposable
Vaginal Specula, Vaginal Speculum
Light Handle (78000), and Transformer
(73305).
TO PREPARE FOR USE:
1. Place illuminator on light handle.
2. Place illuminator and light handle
into slot on
speculum handle.
Slide illuminator assembly up until it
locks into place.
58600/58601
vaginal
No. 78600 (Series 58600
Vaginal Specula continued)
TO REMOVE ILLUMINATOR:
1.
Place thumb at base of illuminator
assembly and pull down.
2. Pull illuminator off of light handle.
CAUTION: LAMP BECOMES HOT
DURING OPERATION.
TRANSFORMER
No. 73305 (Vaginal and
Endoscopic Instruments)
This is a direct plug-in type transformer
encased in a durable housing and
doubly insulated for electrical safety.
Simply plug it into a 110-130 V.,
Hz., outlet, and the unit will be ready
to power the 78000 Vaginal Speculum
Light Handle or 73210 Endoscopic
Instrument Light Handle.
These units may also be wiped clean
using alcohol or a dry cloth.
The 78000 (purchased after May 1,
1988) may be soaked in the following
solutions:
l
Cidex/Cidex 7
l
Sporicidin
l
70% Isopropyl Alcohol
It will withstand ethylene oxide
o
(130
F., 8
p.s.i.,
4 hr. cycle)
sterilization.
No. 78600 VAGINAL SPECULUM
EXPENDABLE ILLUMINATOR
The 78600 Expendable Illuminator will
withstand over 50 cycles of steam
autoclaving (270
o
F, 27
p.s.i.,
6 min.
cycle), after each time the unit will
show degradation of light transmission,
and will eventually need replacement.
The 78600 Expendable Illuminator may
be soaked in the following solutions:
l
Cidex/Cidex 7
l
Sporicidin
l
70% Isopropyl Alcohol
DO NOT USE IODINE-BASED SOAKS
(e.g. Betadine, Wescodyne) as these
will discolor the illuminator and
adversely affect light output.
No. 73305 TRANSFORMER
The transformer should not be
autoclaved. It may be cleaned by wiping
with a 70% isopropyl alcohol solution
or with a dry cloth. DO NOT SOAK. It
will withstand ethylene oxide
p.s.i.
4 hr. cycle).
8
(130o F.,
No. 73308LINE SWITCH
(replaces 73307)
The line switch should not be
auto-
claved. It may be gas sterilized using
ethylene oxide. DO NOT SOAK IN
GERMICIDAL SOLUTIONS. Unit may
be wiped with alcohol.
MAINTENANCE
TECHNICAL
SPECIFICATIONS
Qualified personnel should perform
periodic electrical inspections. We
recommend that this equipment be
inspected at least every six
months-
more often if used under severe
operating conditions.
LEAKAGE CURRENT TEST
The leakage current measured at
any place on the transformer or instruments with the unit plugged in
should be less than 10 microamps.
This is equivalent to 9.5 MV
in the
following circuit:
10.2 O
-I----+’
.1
5mfd
VOLT METER
No. 73305 TRANSFORMER
Equipped with five-foot cord on secondary voltage with connector for use
with Welch Allyn light sources 73210
and 78000 as noted in this folder.
No. 73210 LIGHT HANDLE FOR
ENDOSCOPIC INSTRUMENTS
For use with Welch Allyn reusable
or disposable anoscopes and
sigmoidoscopes.
Weight: 3.67 oz.
Replacement Lamp: W.A. No. 07800
Size: 3.775
X 1.000
(formerly No. 04000)
in.
No. 78000 LIGHT HANDLE
For use with Welch Allyn disposable
vaginal specula No. 58000, 58001,
58600, 58601.
Weight: