Wavetek Meterman FG2C, Meterman FG3C Operator's Manual

Operator’s Manual
FG2C/FG3C
Function Generator
• Bedienungsanleitung
• Manual de Instrucciones
• Manuel d’Utilisation
TM
Bench Instruament
- 1 -
OPERATORS MANUAL
FG2C/FG3C
Function Generator
English
WARRANTY
The FG2C and FG3C Function Generators are warranted against any defects of material or workmanship within a period of one (1) year following the date of purchase of the Function Generator by the original purchaser or original user. Any Function Generator claimed to be defective during the warranty period should be returned with proof of purchase to a Wavetek authorized Service Center or to the local Wavetek dealer or distributor where the Function Generator was purchased. See maintenance section for details. Any implied warranties arising out of the sale of a Wavetek Function Generator, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited in duration to the above stated one (1) year period. Wavetek shall not be liable for loss of use of the Function Generator or other incidental or consequential damages, expenses, or economical loss or for any claim or claims for such damage, expenses or economical loss. Some states do not allow limitations on how long implied warranties last or the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights that vary from state to state.
CERTIFICATIONS AND PRECAUTIONS
This instrument is EN61010-1:1993 certified for Installation Category II, Pollution Degree II. All inputs are protected against continuous overload conditions up to the limits of each function's stated input protection (see specifications). Never exceed these limits or the ratings marked on the instrument itself. Always inspect your Function Generator, test leads and accessories for signs of damage or abnormality before every use. If an abnormal condition exists (broken or damaged test leads, cracked case, display not reading, etc.), do not use. Never ground yourself when taking measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground and never touch exposed wiring, connections, test probe tips, or any live circuit conductors. Do not operate instrument in an explosive atmosphere (flammable gases, fumes, vapor, dust.) Do not use this or any piece of test equipment without proper training.
DANGER High Voltage
WARNING: To avoid electrical shock, the power cord protective
grounding conductor must be connected to ground.
ATTENTION Refer to Manual
CAUTION: To avoid damaging the instrument, do not use it in a place
where ambient temperature exceeds 40°C.
Protective Conductor Terminal
CAUTION: To avoid damaging the instrument, do not input more than
15VDC to VCF.
Frame or Chassis Terminal
CAUTION: To avoid damaging the instrument, do not input more than
150VAC to Frequency Counter (FG3C only).
CAUTION: To avoid damaging the instrument, do not input more than
10Vpp during EXT modulation operation (FG3C only).
FOR UNITED KINGDOM ONLY NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
Green/Yellow Earth Blue Neutral Brown: Live (Phase)
N
E
L
- 2 -
page Warranty 1 Certifications and Precautions 1 Safety Information 1 Introduction 2
page Control Descriptions 5 Using the Function Generator 5 Maintenance and Repair 8 Specifications 9
CONTENTS
PREPARATION FOR USE – UNPACKING
Your shipping carton should include the Function Generator, 1 meter long BNC to mini- alligator clip test lead, one power cord, one spare fuse, and this manual. If any of the items are damaged or missing, immediately return the complete package to the place of purchase for an exchange.
INTRODUCTION
The FG2C and FG3C Function Generators are stable low distortion instruments that generate sine, triangle, and square waveforms in frequencies up to 3 MHz. The FG2C has amplitude, offset, and duty cycle controls. The outputs are Main(50) and logic(TTL and CMOS). The FG3C has the same capabilities plus internal Sweep(Log/Linear), Modulation (AM/FM), external VCF/MOD and counter inputs and GCV output. The counter can be switched to measure and display the frequency of an external signal up to 150 MHz.
FEATURES COMPARISON TABLE FOR MODELS:
FEATURE / MODEL FG2C FG3C
AM/FM modulation na X SWEEP control na X COUNTER input na X GCV Output na X TTL/CMOS output X X VCF input X X Duty Cycle Control X X
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth
symbol or coloured Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black. The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier. This cable/appliance should be protected by a suitably rated and approved HBC mains fuse : refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if engaged in a live socket. Any rewiring must be carried out in accordance with the information detailed on this label.
- 3 -
230V
115V
INPUTS BEFORE OPENING.
TO AVOID SHOCK, REMOVE
250VIEC1010 CAT
AC
29W 36VA
115
24
T0.16A 250V
T0.315A 250V
FUSE RATING
WARNING
FUSE
Fig. 2 FG2C Rear Panel
1
- +
POWER
ADJ
DUTY
CMOS ADJ
TTL
CMOS
5V 15V
OUTPUT
AMPL
-20dBADJ
OFFSET
MIN MAX
50
1M
100k 10k
1k
100 10
-20dB
ATT
FREQUENCY
FG2C Function Generator
10
VCF
INPUT
TTL/CMOS
OUTPUT
DC 15V
MAX
3.0
0.3
0.6
0.9
1.2
1.8
1.5
2.1
2.4
2.7
7
12a
91
13
8
1211 22
20
21
Fig. 1 FG2C Front Panel
- 4 -
Fig. 3 FG3C Front Panel
Fig. 4 FG3C Rear Panel
Mmk
1
0
1
MAX
- +
6 4 2 7
31 5 9 8
POWER
FG3C Function Generator
S
W
E
E
P
/R
A
TE
MOD/DEPTH
SWEEP ON SWEEP OFF
FREQUENCY
LOG
LIN
SWEEP/TIME
SLOW FAST
AM FM
ADJ
DUTY
CMOS ADJ
CMOS
5V 15V
OUTPUT
AMPL
-20dBADJ
OFFSET
MIN MAX
50
OVER
GATE
0.01Hz0.1 10
MOD
EXT
1M
MOD
ON
100k
GATE
10k 1k
100 10
-20dB
ATT
17
12a
14 1613 15 12
11
22
2a
10
20 21
23
19
FUSE
230V
115V
INPUTS BEFORE OPENING.
TO AVOID SHOCK, REMOVE
INT EXT
TTL/CMOS OUTPUT
DC 15V
MAX
INPUT
VCF/MOD
OUTPUT
GCV
COUNTER
MAX
CAT
INPUT
COUNTER
AC
150Vrms
250VIEC1010 CAT
AC
29W 36VA
115
18 24
T0.16A 250V
T0.315A 250V
FUSE RATING
WARNING
- 5 -
CONTROL DESCRIPTIONS
FRONT PANEL
1. Power Switch Applies line power to the FG2C\FG3C power supply.
2. Gate Time Indicator Gate interval indicator (the gate interval of internal counter is 0.01 second).
2a. Gate Time Selector
Press this button to change gate time when use EXTernal counter mode. The sequence is 0.01s, 0.1s, 1s, or 10s selected by pressing this button.
3. Over Range Indicator In the external counter mode, the indicator is illuminated when the output frequency is greater than the range selected.
4. Counter Display Shows the external frequency via a 6 digit green display, and the internal frequency via a 5 digit green display.
5. Frequency Indicator Indicates the current frequency value multiplier.
6. Gate Time Indicator Flashes at the beginning of each gating interval.
7. Frequency Range Selector Select the required frequency range by pressing the relevant push button on the panel as shown in Table 1 .
Table 1 Push button Frequency Range
1 0.3Hz to 3Hz 10 3Hz to 30Hz 100 30Hz to 300Hz 1k 300Hz to 3kHz 10k 3kHz to 30kHz 100k 30kHz to 300kHz 1M 300kHz to 3MHz
8. Function Selector Press one of the three push buttons to select the desired output waveform.
9. Duty Function
Pull out and rotate the knob to adjust the duty cycle of the waveform.
10. TTL/CMOS Selector When the knob is in, the BNC terminal of (20)
will output a TTL compatible waveform. If the knob is out, rotating the knob
will adjust the CMOS compatible output (5 -15Vpp) at the BNC (20)
terminal .
11. DC Offset Control Pull out the knob to select any DC level of the waveform between ±10V., turn the knob clockwise to select a positive DC offset level of the waveform and counterclockwise for a negative DC offset level of the waveform.
12. Output Amplitude Control with Attenuation Operation Turn clockwise for MAX. output and counterclockwise for MIN output. Pull the knob out for additional 20dB output attenuation.
12a. 20dB Attenuation
Press the button to reduce the output by 20dB .
13 FREQ/SWEEP Selector and Frequency Adjustment
(Sweep On/Off): Turn clockwise the knob clockwise for MAX frequency and counterclockwise for MIN frequency. (Keep the knob pointer within the scale range on the panel). Pull out the knob to start the auto sweep operation; the upper frequency limit is determined by the knob position.
- 6 -
14 SWEEP/TIME
Control and LIN/LOG Selector: (1) Rotate the knob clockwise to adjust sweep time for MAX, or counterclockwise for MIN. (2) Select Linear sweep mode by pushing in the knob or LOG sweep mode by pulling out the knob.
15. MOD ON/off Selector Press the button once, the indicator will light, and the output will be modulated by internal 400Hz sinewave or press the button again, the indicator will be off, and the output will be modulated an external signal via VCF/MOD in connector (21)
.
16. MOD/DEPTH SWEEP RATE and AM/FM Selector: (1) Sweep width can be controlled from 0 to 1000 times. (2) Adjust modulation ratio by turning the knob clockwise for MAX, or counterclockwise for MIN. (3)
Select AM(amplitude modulation) mode by pushing in the knob or FM(frequency modulation) mode by pulling out the knob.
17 MOD EXT
Selector: Press the button once, the indicator will light, and EXTernal MODulation is selected. Press the button again, the indicator will be off, and INT MODulation is selected.
FRONT or REAR PANEL
18. INT/EXT Counter Switch Select internal counter mode (count the frequency of the FG3C output) or select EXT counter mode for an independent counter (input signal from BNC (19)
terminal).
19. EXT. Counter Input Terminal: Accepts external signals for measurement.
20. TTL/CMOS Output Terminal TTL/CMOS compatible signal output
21. VCF/MOD Input Terminal: Connector for the input voltage required to perform the "voltage control frequency" operation or the EXT modulation operation.
22. Main Output Terminal Main signal output.: The Output signal is calibrated for a 50 load.
23. GCV Output This is a DC voltage output and its voltage level will follow the change of Frequency.
24. 115/230 Switch This switch selects the mains power voltage.
The functions of items 2, 2a, 3, 4, 5, 6, 14, 15, 16, 17, 18, 19, and 23 do not apply to model FG2C.
USAGE PROCEDURES
The following section describes the basic setup and controls for operating the functions of this instrument. One of the best ways to observe waveforms is to connect the function generator to an Oscilloscope. Observe the effects that different
controls have on the waveforms displayed on the Oscilloscope when preceding with the following steps:
First-step check:
(1) Ensure that the mains voltage switch (24)
on the rear panel of this instrument of is compatible with the mains supply.
Connect the instrument to the mains supply using the power cord supplied. (2) Press PWR (1)
switch, ensure all the rotary knobs are pushed in, and rotate the AMPL (12) knob to fully counterclockwise.
(3) Rotate the FREQ (13)
control fully counterclockwise.
Triangle, square, and sine waves
(1) Select the Function (8)
, select the Range (7), and rotate the FREQUENCY (13) knob to set the required frequency (read
output in display window (4)
).
(2) Connect the Output 22)
terminal to the oscilloscope for observing the waveform signal.
(3) Rotate the AMPL (12)
knob to control the waveform amplitude.
(4) If attenuation of the output signal is required, pull out the AMPL (12)
knob to obtain 20dB attenuation or press the ATT -20dB
(12a) button for additional 20dB attenuation..
(5) The phase-relation of Output waveform and TTL output is shown in Figure 5 below:
- 7 -
Figure 5 Pulse wave generation
(1) Press the Function (8)
button ( ), select the Range (7),
and rotate the FREQUENCY (13) knob to set the frequency. (2) Connect the OUTPUT terminal (22)
to the oscilloscope for
observing the output signal. (3) Pull out and rotate the Duty (9)
knob to adjust the width of pulse
waveform. (4) Adjust the AMPL (12)
knob to control the pulse amplitude.
(5) Pull out the AMPL (12)
knob to attenuate the output signal by
20dB.
Ramp wave generation
(1) Press the Function (8) button ( ) , select the Range (7) , rotate the FREQUENCY (13) knob to set the frequency. (2) Connect the OUTPUT terminal (22)
to the oscilloscope for observing the output signal.
(3) Pull out and rotate the Duty (9)
knob to adjust the slope of ramp waveform.
(4) Adjust the AMPL (12)
knob to control output amplitude of ramp waveform.
(5) Pull out the AMPL (12)
knob to attenuate the output signal by 20dB.
TTL/CMOS signal output
(1) Select the Range (7)
, and rotate the FREQUENCY (13) knob to set the frequency.
(2) Connect BNC connector of TTL/CMOS (20)
to oscilloscope to observe the output signal. (3) The output is a square waveform fixed at TTL level, suitable for general TTL integrated circuits. (4) If a square waveform of CMOS level is required, pull out the CMOS (10)
knob and adjust to required voltage levels.
Variation of external voltage-controlled frequency(VCF)
This mode of operation allows the user to adjust the frequency of the Function Generator with an external DC control Voltage. (1) Select the Function (8)
, select the Range (7), and rotate the FREQUENCY (13)knob to set the required frequency.
(2) Connect external control voltage (0±10VDC) to the VCF (21)
terminal via a suitable lead, and observe the signal generated from
the Output (22) BNC terminal.
(3) Other adjustments; the AMPL (12)
knob can change the amplitude of signal, or attenuate the signal; the Offset (11) knob is for
DC level changes, rotate the Duty (8) knob to change the output ratio of pulse or ramp waveform …etc..
Auto Sweep
(1) Select the Function (8)
button, select the Range (7), and rotate the FREQUENCY (13) knob to set the required base
frequency.
(2) Connect the OUTPUT terminal (22)
to the oscilloscope for observing the OUTPUT signal.
(3) Rotate the FREQUENCY (13)
knob to determine the upper limit frequency.
(4) Pull out the FREQUENCY (13)
knob to activate auto sweep operation.
(5) Rotate the SWEEP/TIME (14)
knob and SWEEP RATE(16) knob to adjust sweep time and rate.
(6) The LOG sweep mode is selected by pulling out the SWEEP/TIME (14)
knob.
Note: The sweep width can only be adjusted during the sweep cycle and it cannot be stopped.
AM/FM operation
(1) Select the Function (8)
, select the Range (7), rotate the FREQUENCY (13)knob to set required frequency.
(2) Connect OUTPUT terminal (22)
to the oscilloscope for observing the output signal.
(3) For the FM modulation mode, press the MOD (15)
button and pull out the MOD/DEPTH (16) knob.
(4) For the AM modulation mode, press the MOD (15)
button and push in the MOD/DEPTH (16) knob.
(5) Adjust the MOD/DEPTH (16)
knob to achieve required modulation ratio.
TTL PULSE 0V
TRIANGLE 0V
SINE 0V
SQUARE 0V
- 8 -
PRECAUTION ITEMS
(1) The DC OFFSET (11) knob, will provide a DC level voltage of ±10V (no load) or ±5V (50load). However, the signal
amplitude plus the DC level, is limited to ±20V (no load) or ±10V (50load). In case of over voltage, clipping will appear as show n in Figure 6.
(2) The main OUTPUT terminal is calibrated into a 50load. This indicates that the signal source impedance is 50. The OUTPUT
terminal may be connected to any circuit input impedance, but output the voltage and terminal impedance will not be calibrated. To avoid oscillation, the OUTPUT terminal should be connected to a 50load (especially when using high frequency and square wave output), and the connecting cable should be as short as possible.
(3) When the DUTY knob is full counterclockwise, the ratio of positive state to negative state, should not be less than 80:20. When
the DUTY knob is full clockwise, the ratio of positive state to negative state, should not be greater than 20:80. Square waves can be expanded to pulse waves, triangle waves can be expanded to ramp waves, sine waves can be expanded to unsymmetrical sine waves. The Figure 7 shows the effect of the DUTY control adjustment on different waveforms.
Figure 6 Figure 7
FUSE REPLACEMENT (See Figs. 2 and 4) Preliminary: Disconnect test leads from circuit and Function Generator. Push Power Switch off and unplug the power cord from the
rear of the Function Generator.
Warning: Failure to turn off and remove power from the Function Generator before changing the mains power
fuse could result in damage to the instrument and the operator. Fuse replacement: Remove the fuse by inserting a small flat blade screwdriver into the notch at the bottom of the mains power
connector socket. Slide the fuse holder out. Replace the fuse with one of the following: 115 VAC 50/60 Hz: T0.315A/250V fuse (5mm x 20mm), I.R. 100kA – Wavetek p/n 2400-05-0069, Littel Fuse p/n 218.315, or
Bussmann p/n. GDC-315mA. 230 VAC 50/60 Hz: T0.160A/250V fuse (5mm x 20mm), I.R. 100kA – Wavetek p/n 2400-05-0070, Littel Fuse p/n 218.160, or
Bussmann p/n. GDC-160mA.
Warning: Use only an equivalent fuse to the one specified. Use of an incorrect fuse could result in serious injury or even death. Reassembly: Slide the fuse holder back into the fuse cavity and reconnect the power cord. MAINTENANCE & REPAIR
If there appears to be a malfunction during the operation of the Function Generator, the following steps should be performed in order to isolate the cause of the problem: Review the operating instructions for possible mistakes in operating procedure. Inspect and
test the Test Cables for a broken or intermittent connection. Inspect and test the fuse. See Fuse Replacement. Except for the replacement of the fuse, or test cables, repair of the Function Generator should be performed only by a Factory Authorized Service
Center or by other qualified instrument service personnel. The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow the function generator to dry completely before using. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning.
VARY THE DUTY OF SHORTER TIME SECTION BY
ADJUSTING THE FREQUENCY ADJUSTMENT DIAL
VARY THE DUTY OF LONGER TIME SECTION BY
ADJUSTING THE DUTY CONTROLLER
PULSE
(SQUARE WAVE)
RAMP
(TRIANGLE WAVE)
UNSYMMETICRAL
(SINE WAVE)
TTL CMOS
A. Zero DC Offset with Maximum Signal
B. Offset Limit without Clipping
C. Excessive Offset All examples: Output terminated is 50 ohm
POSITIVE
DC OFFSET
NEGATIVE
DC OFFSET
POSITIVE
DC OFFSET
NEGATIVE
DC OFFSET
POSITIVE
DC OFFSET
NEGATIVE
DC OFFSET
- 9 -
SPECIFICATIONS
FG2C FG3C
1.Main
Frequency Range 0.3 Hz to 3 MHz (7 Range) 0.3 Hz to 3 MHz (7 Range) Amplitude >10Vpp (into 50load) >10Vpp (into 50load) Impedance 50±10% 50±10% Attenuator -20dB ±1dB x2 -20dB ±1dB x2 DC Offset <-5V to >+5V (into 50load) <-5V to >+5V (into 50load) Duty Cycle Control 80%:20%:80% to 1MHz Continuously variable 80%:20%:80% to 1MHz Continuously variable LED Display N/A 6 digits; 7.6mm (0.3 in) high Range Accuracy ±5% +1 Hz (at 3.0 position) N/A
2.Sine Wave
Distortion 1 %, 0.3 Hz to 200 kHz 1 %, 0.3 Hz to 200 kHz
THD 35dB below fundamental in all ranges THD 35dB below fundamental in all ranges (Specification applied from MAX. to 1/10 level) (Specification applied from MAX. to 1/10 level)
Flatness < 0.3dB, 0.3 Hz to 300 kHz < 0.3dB, 0.3 Hz to 300 kHz
< 0.5dB, 300 kHz to 3 MHz < 0.5dB, 300 kHz to 3 MHz
3.Triangle Wave
Linear 98 %, 0.3 Hz to 100 kHz 98 %, 0.3 Hz to 100 kHz
95 %, 100 kHz to 3 MHz 95 %, 100 kHz to 3 MHz
4.Square Wave
Symmetry ±2%, 0.3 Hz to 100 kHz ±2%, 0.3 Hz to 100 kHz Rise or Fall Time 100ns at maximum output. 100ns at maximum output.
(into 50load) (into 50load)
5.CMOS Output
Level 4Vpp ±1Vpp to 14.5Vpp 4Vpp ±1Vpp to 14.5Vpp
±0.5Vpp adjustable ±0.5Vpp adjustable
Rise or Fall Time 120ns 120ns
6.TTL Output
Level 3Vpp 3Vpp Fan Out 20 TTL load 20 TTL load Rise or Fall Time 25ns 25ns
7.VCF
Input voltage 0 to 10V ±1V(100:1) 0 to 10V ±1V(100:1)
Input Impedance 10k±10% 10k±10%
8.GCV
Output voltage N/A To set the voltage between
0 to 2V as per different frequency.
in U.S.A. in Canada in Europe Wavetek Meterman Wavetek Meterman Wavetek Meterman 1420 75th Street SW 400 Britannia Rd. E.Unit #1 52 Hurricane Way Everett, WA 98203 Mississauga, ON L4Z 1X9 Norwich, NR6 6JB, U.K. Tel: (877) 596-2680 Tel: (905) 890-7600 Tel: int + 44-1603-404824 Fax: (425) 446-6390 Fax: (905) 890-6866 Fax: int + 44-1603-482409
FG2C FG3C
9.Sweep Operation
Sweep/Manual N/A Switch selector Sweep/Rate N/A 100:1 ratio max. and adjustable Sweep/Time N/A 0.5 Sec to 30 Sec adjustable Sweep/Mode N/A Lin./Log. switch selector
10.Amplitude Modulation
Depth N/A 0 to 100% MODulation Frequency N/A 400 Hz(INT), DC to 1 MHz (EXT) Carrier BW N/A 100 Hz to 3 MHz (-3dB) EXT Sensitivity N/A 10Vpp for 100 % modulation
11.Frequency Modulation
Deviation N/A 0 to ±5% Modulation Frequency N/A 400Hz (INT), DC to 20 kHz (EXT) EXT Sensitivity N/A 10Vpp for 10% modulation
12.Frequency Counter
Int./Ext. N/A Switch selector Range N/A 0.3 Hz to 3 MHz (5 Hz to 150 MHz EXT) Accuracy N/A Time base accuracy ±1count Time base N/A ±20ppm (23°C ±5°C) after 30 minutes warm up Resolution N/A The maximum resolution is 10 nHz for 1Hz
and 1 kHz for 100MHz Input Impedance N/A 1M/150pf Sensitivity N/A 35mVrms (5Hz to 100MHz)
45m Vrms (100MHz to150MHz)
13.General
LED Display N/A 6 digits; 7.6mm (.3 in) high Power Source 115, 230VAC ±15%, 50/60Hz 115, 230VAC ±15%, 50/60Hz Operation Environment Indoor use, Altitude up to 2000m. Ambient Temperature 0°C to 40°C
Relative Humidity 80%(maximum). EN61010-1, Installation category II, Pollution Degree 2 Storage Temperature & Humidity -10 °C to 70°C @ 70% R.H.(maximum). Accessories 1 each RTL-1, Instruction Manual, 2 each RTL-1, Instruction Manual,
Spare Fuse Spare Fuse Dimension 243(w) x 93(H) x 292(D) mm Weight Approx. 2.0 kgs. Approx. 2.2 kgs. Safety: Meets EN61010-1:1993. Installation category II Pollution degree 2
EMC: This product complies with requirements of the following European Community Directives: 89/336/EEC
(Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE Marking). However, electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement
circuit. Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit. Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence of electronic interference.
- 10 -
- 11 -
GEBRAUCHSANLEITUNG
Funktionsgeneratoren FG2C/FG3C
German
GEWÄHRLEISTUNG
Die Funktionsgeneratoren FG2C und FG3C sind ab Kaufdatum gegen Material- und Herstellungsfehler gewährleistet für die Dauer von einem (1) Jahr. Ein Funktionsgenerator der während der Gewährleistungsperiode reparaturbedürftig wird soll mit Kaufnachweis zu einer Wevetek anerkannten Service Stelle oder zum Wavetek Händler, wo das Gerät gekauft wurde, gebracht/geschickt werden. Siehe Abschnitt ‘Wartung und Reparatur’ für Einzelheiten. Irgendwelche implizierte Gewährleistungen welche aus dem Kauf eines Wavetek Funktionsgenerators fortfließen, einschließlich, jedoch nicht beschränkt auf Verkäuflichkeit oder Nutzbarheit zu einem bestimmten Zweck, sind in der Dauer auf ein (1) Jahr beschränkt. Wavetek ist nicht ansprechbar für Gebrauchsverluß oder andere Folgeschäden und verbundene Kosten und Ausgaben. Verschiedene Staten und Länder erlauben keine Beschränkungen auf Folgeschäden oder deren Ausschluß. Deshalb sind diese Beschränkungen vielleicht nicht für Sie zupaßbar. Diese Gewährleistung gibt Ihnen bestimmte Rechte. Vielleicht haben Sie, je nach Staat oder Land, auch noch andere Rechte.
Zertifizierungen und Vorsichtsmaßnahmen
Dieses Gerät ist EN61010-1:1993 zerifiziert für Installationsklasse II, Pollutionsgrad 2. Alle Eingänge sind gegen dauerhafte Überbelastung geschützt, bis zur Grenze welche je Funktion in den Spezifikationen angegeben oder auf dem Gerät markiert ist (Siehe Spezifikationen). Überschreiten Sie niemals diese Grenzen. Prüfen Sie Ihren Funktionsgenerator, Meßkabel und Zubehör vor jeder Messung. Im Problemfall (gebrochene oder beschädigte Meßkabel, gerissenes Gehäuse, keine Anzeige), Gerät nicht verwenden. Vermeiden Sie, beim Vornehmen einer Messung, jeden Kontakt mit Erdpotential. Berühren Sie keine Metallteile, Leiter, usw. Isolieren Sie sich gegenüber Erde (Kleidung, Schuhe, Teppich). Verwenden Sie das Gerät nicht in einer explosiven Umgebung (entflammbare Gase, Dampf, Staub). Verwenden Sie dieses, oder andere Meßgeräte nicht ohne entsprechende Ausbildung.
GEFAHR Hohe Spannung
WARNUNG: Um elektrischen Schlag zu vermeiden, Schutzleiter des
Stromkabels mit Erde verbinden.
ACHTUNG Anleitung lesen
ACHTUNG: Um Schaden am Gerät zu vermeiden, nicht bei über 40°
Messungen vornehmen.
Schutzleiteransschluß
ACHTUNG: Um Schaden am Gerät zu vermeiden, nicht mehr als 15V
DC am VCF Eingang anlegen.
Gehäuseanschluß
ACHTUNG: Um Schaden am Gerät zu vermeiden, nicht mehr als 150V
AC am Frequenzzähler Eingang anlegen (nur FG3C).
ACHTUNG: Um Schaden am Gerät zu vermeiden, nicht mehr als 10Vss
während EXT Modulation anlegen (nur FG3C).
- 12 -
Seite Gewährleistung 11 Zertifikationen und Vorsichtsmaßnahmen 11 Sicherheitsinformation 11 Einleitung 12
Seite Beschreibung der Bedienelemente 15 Anwendung des Funktionsgenerators 16 Wartung und Reparatur 18 Spezifikationen 19
INHALT
GEBRAUCHSVORBEREITUNG – AUSPACKEN
Die Verpackung sollte enthalten: ein Funktionsgenerator, ein Meßkabel von 1 Meter Länge (BNC nach Krokodilklemme), ein Stromkabel, eine Ersatzsicherung und diese Anleitung. Sollte ein Teil fehlen oder beschädigt sein, bringen Sie das Gerät zur Kaufstelle zurück für einen Umtausch.
EINLEITUNG
Die Funktionsgeneratoren FG2C und FG3C sind stabile Instrumente mit wenig Verzerrung, welche Sinus-, Dreieck- und Rechteckwellenformen bis zu 3 MHz erzeugen. Das FG2C hat Kontrollen für Amplitude, DC Komponente und Tastverhältnis. Ausgänge sind Hauptausgang (50) und Logic-Ausgang (TTL oder CMOS). Das FG3C hat die gleichen Eigenschaften, plus interner Wobbel (Log/Linear), Modulation (AM/FM), externer VCF/MOD, Zählereingang und GCV (Spannung proportional zur Frequenz) Ausgang. Der Zähler kann zum Zählen einer externen Frequenz bis zu 150 MHz verwendet werden.
VERGLEICH DER FUNKTIONEN:
FUNKTION / MODELL FG2C FG3C
AM/FM Modulation na X Wobbel Kontrolle na X Zählereingang na X GCV Ausgang na X TTL/CMOS Ausgang X X VCF Eingang X X Tastverhältnis Kontrolle X X
- 13 -
230V
115V
INPUTS BEFORE OPENING.
TO AVOID SHOCK, REMOVE
250VIEC1010 CAT
AC
29W 36VA
115
24
T0.16A 250V
T0.315A 250V
FUSE RATING
WARNING
SICHERUNG
Fig. 2 FG2C Rückseite
1
- +
POWER
ADJ
DUTY
CMOS ADJ
TTL
CMOS
5V 15V
OUTPUT
AMPL
-20dBADJ
OFFSET
MIN MAX
50
1M
100k 10k
1k
100 10
-20dB
ATT
FREQUENCY
FG2C Function Generator
10
VCF
INPUT
TTL/CMOS
OUTPUT
DC 15V
MAX
3.0
0.3
0.6
0.9
1.2
1.8
1.5
2.1
2.4
2.7
7
12a
91
13
8
1211 22
20
21
Fig. 1 FG2C Frontansicht
Loading...
+ 30 hidden pages