Wavetek 148A User Manual

INSTRUCTION MANUAL
MODEL 148A
MHz
20
AM/FM/PM GENERATOR
@ -
1981. - Wavetek
THIS DOCUMENT CONTAINS INFORMATION PRO-
PRIETARY TO WAVETEK AND IS SOLELY FOR IN-
STRUMENT OPERATION AND MAINTENANCE. THE
INFORMATION IN THIS DOCUMENT MAY NOT BE DUPLICATED IN ANY MANNER WITHOUT THE PRIOR APPROVAL IN WRITING OF WAVETEK.
SAN DIEGO
9045 Balboa Ave., San Diego, CA 92123 P.O. Box 661, San Diego,
Tel
7141279-2200
TWX 91w352oo7
Calif.
.
92112
hhuat
Instrument Retease
l3eeion 6181
A-3181
1.1 THE MODEL
Wavetek Model
148A
148A,
20 MHz AM/FM/PM Generator is a precision source of sine, triangle, square, ramp and pulse waveforms plus
dc
voltage. The waveforms may be controlled in symmetry as well as amplitude and dc offset. A built-in modulation generator can modulate frequency, phase and amplitude or modula­tion may be by an external source.
The generator may be run in continuous mode or trig­gered for a single pulse or gated for a burst of pulses. Triggering and gating may be by the Model 148 built-in modulation generator or by
an
external source. The
triggered and gated waveform start/stop point is
- 90’
selectable from
through +
trol plus dc offset control gives haverwave
90’.
Start/stop con-
capability.
External Trigger:
Generator is quiescent until trig­gered by an external signal, then generates one cycle at the selected frequency.
External Gate: Same as external trigger, except gen­erator oscillates at the selected frequency for the
duration of the positive state of the external signal plus the time to complete the last cycle.
Internal Trigger: Same as external trigger, except that
the modulation generator is internally connected to the trigger input of the main generator.
Internal Gate: Same as external gate, except that the
modulation generator is internally connected to the trigger input of the main generator.
The main output of waveforms may be offset. A TTL sync pulse is
available
attenuated
and
at main generator frequencies, and the modulation generator waveforms are available at fixed amplitudes.
Frequency of both the main generator and the modulation generator can be manually controlled at the front panel or electrically controlled by external
voltages.
1.2
SPECIFICATIONS
1.2.1
1.2.1.1
Main Generator
Waveforms
Selectable sine % , triangle 2/ , square % positive square
J-L ,
negative square U- and dc. A TTL sync pulse is provided on a separate output con­nector. All can be produced with variable symmetry, amplitude and dc offset.
1.2.1.2
Operational Modes
Continuous: Generator oscillates continuously at the selected frequency.
3
Modulation Modes
Internal Modulation
Setting a front panel modulation switch in the INT position routes the selected modulation function from the modulation amplitude control to the selected mod-
ulating circuits of the main generator.
Amplitude Modulation (AM):
lation functions are used in this
2/
2/ , ‘L
internal
mode. With modulation amplitude ccw, carrier at
function output is not amplitude modulated and ap-
proximately half of normal (AM OFF) amplitude.
Clockwise rotation of modulation amplitude results in
increasing amplitude modulation of the carrier to at least 100% AM.
,
Frequency Modulation (FM) and Sweep:
m
modulation functions are used to linearly
sweep the main generator frequency. The frequency
dial sets the lower sweep limit and the modulation
amplitude control determines the upper frequency limit (not to exceed 2.0 x multiplier). A sweep set
mode allows precision setting of upper frequency
limit. For frequency deviation, the dial determines the center frequency and modulator varies the main generator frequency above and below
modu-
modulation
M
/\I , 2/
or
and
\
l-l
center by an amount determined by the modulation amplitude.
-Phase Modulation (PM): As in External Modulation. Amplitude of modulator varies phase up to
250’.
2/ , 2/ , \
functions
External Modulation
A BNC feeds an external signal to the modulating cir­cuits when selected by a front panel modulation tog­gle switch in the EXT position.
peak current and may be attenuated to 60 dB in 20 steps. An additional 20 dB vernier also controls the waveform amplitudes.
1.2.1.6
Approximately sine and triangle waveforms only).
1.2.1.7
Adjustable Waveform Start/Stop Point
- 90’
to +
90’
to 2 MHz (operative on
DC Output and DC Offset
dB
Amplitude Modulation (AM): signals with zero dc component produce suppressed carrier modulation; i.e., a carrier (at main generator function output) amplitude of zero. The function out­put modulated signal has an amplitude sensitivity of 3 volts peak (1.5 Vp into
rier signal level at the function output can be pro­duced at a sensitivity of 3 Vp (I dc component in. Modulating the dc component mod-
ulates the carrier level. Percent modulation (AM) will be the ratio of the peak ac to peak dc of the modulating signal. Input impedance is
Frequency Modulation (FM) and Sweep: Sensitivity is 20% of frequency range/volt peak. Linear behavior
results only when all instantaneous frequencies call­ed for fall within the frequency range (2 to 0.002 called for is the multiplier and dial setting altered by the instantaneous voltage at the modulation input. In­put impedance is 5
Phase Modulation (PM): Sensitivity is volt peak. Linear behavior results only when all instan­taneous transition frequencies called for fall within
the frequency range (2
plier).The
depend heavily on the modulation frequency. and
waveform. Inoperative at frequency multiplier settings
below 100. Input frequencies roll off at 6 dB/octave
above one half of full range frequency and above
150
1.2.1.4,
x
multiplier). The instantaneous frequency
instantaneous frequencies called for will
kHz.
Input impedance is IO
Frequency
5OQ)
kQ.
x
Range
External modulating
per volt peak in. A car-
.5
Vp into 500) per 1 Vp
>2.5 kQ.
x
multiplier
IO0
phase shift/
multiplier to 0.002 x multi-
kQ.
Selectable through function output by front panel controls to a minimum of
t
14.4 Vdc( t 7.2 Vdc into offset limited to k 15 Vdc ( t 7.5 Vdc into set and wave form attenuated proportionately by the 60
dB
output attenuator.
1.2.1.8
AM: Sensitivity of 3 Vp
impedance is FM: Sensitivity of 20% of frequency range/Vp. Input impedance is 5 PM
:
pedance
1.2.1.9
Symmetry of all waveform outputs is continuously ad­justable from 1 variable duty-cycle pulses, sawtooth ramps and symmetrical sine waves.
External Modulation Input
>2.5 k0.
kn.
Sensitivity
is 1O
of
kQ.
Symmetry
:1
9 to
When SYMMETRY control is used, in­dicated frequency is divided by ima
tely IO.
500)
out/Vp (1.5V
IO0
phase
19:1.
Varying symmetry provides
NOTE
(5OQ).
Controlled
with signal peak plus
5OQ).
DC off-
into 503). Input
shift/VP.
Input im-
non-
approx-
1.2.1.10 Sync Output (TTL)
TTL level pulse which will drive 10 quency and time symmetry are the same as for func­tion output.
TTL
loads. Fre-
0.0002
approximately
1.2.1.5
s
(15V
into
1-2
Hz to 20 MHz in
1
%
Function Output
, 2/ and \ selectableandvariable
p-p into
5On).
5Os1). n
All waveforms and dc can supply 150
IO
overlapping ranges with
vernier control.
(500)
and u up to 15 Vp (7.5 Vp
to 30V
p-p
mA
1.2.1.11 Trlgger and Gate
Input Range: IV p-p to * Input Impedance: IO Pulse Width: 25 ns minimum. Repetition Rate: IO MHz maximum. Adjustable triggered signal start/stop point: approxi­mately-
90’
to
+90”
IOV.
k0,
33
pF.
to 2 MHz.
1.2.1 .12
Frequency Precision
1.2.2.3
Output
(6OOQ)
Dial Accuracy
+ (I
O/O
of setting + 1 % of full range) on x 100 thru
-
x 1 M ranges.
*+ (2% of setting + 2% of full range) on
-
x 10 and x
10M
ranges.
x
.Ol
Time Symmetry
t
0.5% on x 100 thru x IOOk
2.0 on dial.
t
5% on all other ranges and from 0.02 to 2.0 on dial.
1.2.1
.13
Amplitude Precision
Amplitude Change With
Sine variation less than:
t
0.1 dB thru x 1
-+
0.5 dB on x 1 M range;
t
3 dB on x 10M range.
OOk
ranges;
ranges and from 0.2 to
Frequency
Step Attenuator Accuracy
-
+
0.3 dB per 20 dB step at 2
1.2.1.14
Waveform Characteristics
kHz.
Sine Distortion
<0.5%
< All harmonics 30 range. All harmonics 26 dB below fundamental on x 10
range.
on x 100 Hz to x 10
1
.O%
on x
.01
to x 10 Hz and x 100
kHz.
kHz
ranges.
dB
below fundamental on x I MHz
thru
MHz
2/
, 2/ and ‘L are fixed level about ground. M and from 0 to +
1.2.2.4 Voltage control of modulator frequency with sensitivity
of 20% of range per volt. Input impedance is 5
1.2.2.5
5V.
Frequency Modulation (FM IN)
Waveform Characteristics
M
10V
p-p balanced
are fixed level 5 Vp
kQ.
Sine Distortion
<
5%
Time Symmetry <
O/O
from 1 Hz
5%
from 0.1 Hz to 100
to 10
kHz.
kHz.
1.2.3 General
1.2.3.1
Stability (for amplitude, dc offset and
frequency)
Short Term:t 0.05% for 10 minutes.
Long Term:
1.2.3.2
Specifications apply at
will operate from
1.2.3.3
28.6
cm (11 ?4 in.) wide;
(103/4
in.) deep.
t,O.25O/o
for 24 hours.
Environmental
25°C t 5°C ambient
O°C
to 50°C ambient temperatures.
Dimensions
13.3
Instrument
cm
(5 % in.) high; 27.3 cm
Square Wave Rise/Fall Times
At
FUNCTION
OUT < 25
ns for
15V
p-p into a
5Ost
load.
Triangle Distortion
Odd harmonics within 15% of correct value to 2 MHz.
1.2.2
Modulation Generator
1.2.2.1
Selectable sine 2/ , triangle ramp M and down ramp M
1.2.2.2
s,2/
ranges.
M, Nl
Waveforms
Frequency Range
,
21
Sweep:
Hzto100
0.1
0.2 Hz
to
2/
,
square ?J , up
.
kHzz
200
kHz
(2 x setting).
in
three
100:1
1.2.3.4 5 kg (11 lb) net; 6.6 kg (14% lb) shipping.
1.2.3.5
90 to selectable;
Weight
Power
108 to
126V,
105V,,
48 to 400 Hz; less than 40 watts.
Unless otherwise noted, ail specifications apply from when FUNCTION OUT is at maximum and
5OQ
terminated, with SYMMETRY control at OFF. Symmetry and vernier affect fre­quency calibration. Maximum possible asymmetry is a function of frequency setting.
198 to 231V and 216
NOTE
0.1
to 2.0 on frequency dial
to
252V
1-3
SECTION
OPERATION
3.1 CONTROLS AND CONNECTORS
The following item
3.1
.1
Main Generator
1
Frequency Dial
main generator.The setting on this dial multiplied by the FREQ MULT basic output frequency of the generator at the
FUNCTION OUT.6 and SYNC OUT 7 The FREQ VERNIER the modulation generator also affect the main generator frequency.
Mode
2
Switch
operating mode of the main generator as follows: a.
EXT GATE Mode
quiescent until a proper gate signal is applied at the EXTTRIG IN BNC 13 and then outputs
the selected signal for the duration of the gate
signal, plus the time to complete the last cycle
generated.
numbers are keyed
The
-
-
This outer switch selects the
frequency control of the
12
and, in some cases,
-
The main generator is
to figure
12
setting is the
3-1.
BNCs.
b. EXT TRIG
mode, except the main generator output is one cycle of selected signal only.
c. CONT Mode - The output signal is continuous.
d.
INT TRIG -
cept the trigger signal is applied internally by
the modulation generator.
e. INT GATE
the gate signal is applied internally by the modulation generator.
TRIGGER LEVEL
is a continuously variable adjustment of the trig­ger circuitry firing point. When full ccw, a
tive going pulse at approximately ­quired for triggering (see figure
cw position,
mately + quired for triggering. In the GATE modes, the‘ generator will run continuously when the control
is cw of 12 o’clock.
Mode
-
Same as for EXT GATE
Same as for EXTTRIG mode, ex-
-
Same as for
Control -
a
positive going pulse at approxi-
7.5V
or more positive voltage is re-
EXT GATE,
This
inner control
7.5V
3-2).
In the full
except
posi-
is
re-
12 13
Figure 3-1
Controls
MOOULATION
and Connectors
GENERATOR
3-1
3.1.2
Modulation
FUNCTION Switch
14
Generator
-
This switch selects the waveform output of the modulation generator. Output is at the OUT BNC connector also available internally to the main generator. Waveforms are sine square
(N\).
AMPLITUDE
sweep frequency.
( ‘% ),
ramp up (M) and ramp down
A SWP SET detent holds the OUT BNC at
control 14 level;
( /L ),
triangle
used
to set upper
16
and is
( 2/ ),
.
AM- The instantaneous amplitude of the out-
d
put signal varies with the instantaneous ampli-
tude of the modulation signal.
e.
FM- The instantaneous frequency of the out­put signal varies with the instantaneous ampli­tude of the modulation signal.
f
.
PM- The instantaneous phase of the output signal varies with the instantaneous amplitude of the modulation signal.
MOD AMPLITUDE Control - This inner con-
trol attenuates the modulation generator signal that is internally fed to the main generator when modulating. It has no effect during internal trig­gering or gating and it has no effect on the
16
OUT
FREQ/PERIOD
15
signal.
MULT Switch - This outer switch (with ranges given in both frequency and period) selects the modulation generator frequency/
period range.
-
VARIABLE Control
This inner control sets
the frequency/period within a range.
OUT
(6OOn)
16
Connector - This BNC is the
modulation generator waveform output: sine
( % ),
triangle
level
1OV p-p
balanced about ground and ramp
( 2/ )
and square
(5V p-p into
6OOQ)
waveforms
( ‘L )
fixed
( AA, M )
fixed level 0 to + 5 Vp waveform. FM IN (5
17
kQ)
Connector - This BNC is
the input for frequency modulation of the modulation generator.
Sensitivity is 20% of frequency range
per volt in.
OPERATlON
32
.
convenience,
For
the generator operation has been grouped in seven basic modes, from which many variations and combinations are possible. The following paragraphs give basic switch posi­tions for each mode and requirements and sug­gestions for operation.
.
g
3.2.1 When setting up the generator, it is advisable to
serve the output on an oscilloscope. Connect TION
5OQ
-
The dc output can be set from + 15 Vdc
DC
- 15
to
Vdc (+ 7.5 Vdc to - 7.5 Vdc into
5OQ).
Continuous Operation
ob-
FUNC-
OUT to the scope input using a 500 cable and a
load. For continuous waveform output, select a basic waveform at the desired frequency. Ensure the modulation switches are set to OFF and the mode is set to
CONT.
The output amplitude can be as great as
15V p.p;
attenuate as desired.
NOTE
use
For best waveform quality, TENUATIO N tion; then use
SWITC H
the
for gross attenua-
AMPLITUDE control for
the AT-
fine attenuation.
The waveform may be skewed to the left or right using the SYMMETRY control; e.g., making a ramp from a triangle
(
/G
.)
waveform.
( ,/1 )
/
NOTE
10
The output frequency is divided b y
when
the SYMMETRY control is switched from
the OFF position.
The dc level of the waveform may be varied with the
DC OFFSET control, but waveform clipping can occur.
NOTE
The basic modes of operation are:
a.
Continuous- A continuous output signal.
b
.
Triggered
-
One cycle of waveform for each
trigger signal.
C.
Gated
-
A “burst” of waveforms for the dura-
tion of each gate signal.
Waveform clipping will occur (OVERLOAD
LED will light), unless waveform amplitude
is decreased so
that
wavefor m
plus
offset
is less than 7.5 volts at the waveform peak.
3.2. 2
Triggered Operation
In triggered operation, there is one cycle of output for
each trigger signal input. For triggered operation, first
3-3
set up the generator for continuous operation (refer to paragraph 3.2.1). The main generator may be trig-
gered internally by the modulation generator or trig-
gered by an external source. If an unmodulated waveform is being output, the modulation generator is
free. to give the desired triggering frequency and
should be selected as the trigger source. Use any
modulation generator waveform as the trigger. Set the modulation generator frequency for the desired trig­ger frequency and sync on the modulation generator output.
NOT E
3.2.4
The TRIGGER LEVEL control manually triggering or gating the generator. For
manually triggering (single cycles), the generator
mode should be EXT TRIG with no external signal in­put at the EXT TRIG IN connector. Each time TRIG­GER LEVEL is rotated cw through mid-position, one triggered cycle will be generated. In EXT GATE mode, the generator runs continuously as long as the TRIG­GER LEVEL is
Manual Triggering and Gating
can also be used
CW
of mid-position.
for
1. Trigger frequency must be slower than the output waveform frequency.
2.
The MOD AMPLITUDE control has no effect in
Rotate the TRIGGER LEVEL control to obtain a good trigger. If the waveform start/stop point is to be other than zero degrees, set the TRIG START/STOP control as required. (See figure 3-3.) Haverwaves can be set
up using start/stop control and
-90°
s--
If external triggering is to be used, connect a repeti­tive signal with a positive going transition of greater than one volt to EXT TRIG IN using a the TRIGGER LEVEL control for proper triggering.
3.2. 3
In gated operation there is a “burst” of waveforms lasting the duration of the gate pulse plus the time re­quired to complete the last cycle of waveform started. Set up as for triggered operation (refer to paragraph 3.2.2) and select a “trigger” signal whose duty time gives the desired waveform burst.
AA
Figure
Gated Operation
INT
TRIG
and INT GATE modes.
dc
offset control.
HAVERSINES
_v
3-3..
Waveform Start/Stop Examples
5OQ
cable. Adjust
3.2.5
AM Operation
In amplitude modulation, the instantaneous amplitude of the output signal varies with the instantaneous amplitude of the modulation signal.
NOT E
The output waveform will be LOAD LED will light) if any instantaneous amplitude greater tha n ( t
7.5 volts into
Set up the generator as for continuous operation (refer to paragraph 3.2.1). Switch to internal or exter­nal amplitude modulation. If internal, note that the car-
rier (main generator waveform) mean amplitude is decreased to half. This is to prevent clipping driving the output amplifier) when the carrier ampli­tude is modulated (increased and decreased) by another signal. Set the modulation generator fre­quency to a lower frequency than the main generator and sync the scope to the modulation generator output. Set the modulator amplitude for a desired percentage of amplitude modulation (0 to greater than 1 and set the main generator ATTENUATION and AMPLITUDE controls for a desired amplitude of modulated waveform. If external modulation is selected, observe that the carrier (main generator waveform) amplitude drops to zero (null). This is for suppressed carrier mode of amplitude modulation. Connect the modulator OUT
IN with a coaxial cable. Set the main generator ATTENUATION and AMPLITUDE controls amplitude of suppressed carrier waveform. For best
results when using an external modulation signal at
EXT MOD IN, maintain as near as convenient amplitude (which does not exceed 5 main generator ATTENUATION and AMPLITUDE con­trols. An external source may also be used for regular AM operation, rather than suppressed carrier opera­tion, by supplying a dc component along with the ac
5OQ )
is produced.
clipped(OVER-
t
15 volt s
(6OOQ)
to the EXT MOD
VP);
(over-
00%
range)
for
a desired
a
5 Vp
then use the
3-4
Loading...
+ 14 hidden pages