INFORMATION IN THIS DOCUMENT MAY NOT BE
DUPLICATED IN ANY MANNER WITHOUT THE
PRIOR APPROVAL IN WRITING OF WAVETEK.
SAN DIEGO
9045 Balboa Ave., San Diego, CA 92123
P.O. Box 661, San Diego,
’
Tel
7141279-2200
TWX 91w352oo7
Calif.
.
92112
hhuat
Instrument Retease
l3eeion 6181
A-3181
1.1 THE MODEL
Wavetek Model
148A
148A,
20 MHz AM/FM/PM Generator
is a precision source of sine, triangle, square, ramp
and pulse waveforms plus
dc
voltage. The waveforms
may be controlled in symmetry as well as amplitude
and dc offset. A built-in modulation generator can
modulate frequency, phase and amplitude or modulation may be by an external source.
The generator may be run in continuous mode or triggered for a single pulse or gated for a burst of pulses.
Triggering and gating may be by the Model 148 built-in
modulation generator or by
an
external source. The
triggered and gated waveform start/stop point is
- 90’
selectable from
through +
trol plus dc offset control gives haverwave
90’.
Start/stop con-
capability.
External Trigger:
Generator is quiescent until triggered by an external signal, then generates one cycle
at the selected frequency.
External Gate: Same as external trigger, except generator oscillates at the selected frequency for the
duration of the positive state of the external signal
plus the time to complete the last cycle.
Internal Trigger: Same as external trigger, except that
the modulation generator is internally connected to
the trigger input of the main generator.
Internal Gate: Same as external gate, except that the
modulation generator is internally connected to the
trigger input of the main generator.
The main output of waveforms may be
offset. A TTL sync pulse is
available
attenuated
and
at main generator
frequencies, and the modulation generator waveforms
are available at fixed amplitudes.
Frequency of both the main generator and the
modulation generator can be manually controlled at
the front panel or electrically controlled by external
negative square U- and dc. A
TTL sync pulse is provided on a separate output connector. All can be produced with variable symmetry,
amplitude and dc offset.
1.2.1.2
Operational Modes
Continuous: Generator oscillates continuously at the
selected frequency.
3
Modulation Modes
Internal Modulation
Setting a front panel modulation switch in the INT
position routes the selected modulation function from
the modulation amplitude control to the selected mod-
ulating circuits of the main generator.
Amplitude Modulation (AM):
lation functions are used in this
2/
2/ , ‘L
internal
mode. With modulation amplitude ccw, carrier at
function output is not amplitude modulated and ap-
proximately half of normal (AM OFF) amplitude.
Clockwise rotation of modulation amplitude results in
increasing amplitude modulation of the carrier to at
least 100% AM.
,
Frequency Modulation (FM) and Sweep:
m
modulation functions are used to linearly
sweep the main generator frequency. The frequency
dial sets the lower sweep limit and the modulation
amplitude control determines the upper frequency
limit (not to exceed 2.0 x multiplier). A sweep set
mode allows precision setting of upper frequency
limit. For frequency deviation, the dial determines the
center frequency and modulator
varies the main generator frequency above and below
modu-
modulation
M
/\I , 2/
or
and
\
l-l
center by an amount determined by the modulation
amplitude.
-Phase Modulation (PM): As in External Modulation.
Amplitude of modulator
varies phase up to
250’.
2/ , 2/ , \
functions
External Modulation
A BNC feeds an external signal to the modulating circuits when selected by a front panel modulation toggle switch in the EXT position.
peak current and may be attenuated to 60 dB in 20
steps. An additional 20 dB vernier also controls the
waveform amplitudes.
1.2.1.6
Approximately
sine and triangle waveforms only).
1.2.1.7
Adjustable Waveform Start/Stop Point
- 90’
to +
90’
to 2 MHz (operative on
DC Output and DC Offset
dB
Amplitude Modulation (AM):
signals with zero dc component produce suppressed
carrier modulation; i.e., a carrier (at main generator
function output) amplitude of zero. The function output modulated signal has an amplitude sensitivity of
3 volts peak (1.5 Vp into
rier signal level at the function output can be produced at a sensitivity of 3 Vp (I
dc component in. Modulating the dc component mod-
ulates the carrier level. Percent modulation (AM) will
be the ratio of the peak ac to peak dc of the modulating
signal. Input impedance is
Frequency Modulation (FM) and Sweep: Sensitivity is
20% of frequency range/volt peak. Linear behavior
results only when all instantaneous frequencies called for fall within the frequency range (2
to 0.002
called for is the multiplier and dial setting altered by
the instantaneous voltage at the modulation input. Input impedance is 5
Phase Modulation (PM): Sensitivity is
volt peak. Linear behavior results only when all instantaneous transition frequencies called for fall within
the frequency range (2
plier).The
depend heavily on the modulation frequency. and
waveform. Inoperative at frequency multiplier settings
below 100. Input frequencies roll off at 6 dB/octave
above one half of full range frequency and above
150
1.2.1.4,
x
multiplier). The instantaneous frequency
instantaneous frequencies called for will
kHz.
Input impedance is IO
Frequency
5OQ)
kQ.
x
Range
External modulating
per volt peak in. A car-
.5
Vp into 500) per 1 Vp
>2.5 kQ.
x
multiplier
IO0
phase shift/
multiplier to 0.002 x multi-
kQ.
Selectable through function output
by front panel controls to a minimum of
t
14.4 Vdc( t 7.2 Vdc into
offset limited to k 15 Vdc ( t 7.5 Vdc into
set and wave form attenuated proportionately by the
60
dB
output attenuator.
1.2.1.8
AM: Sensitivity of 3 Vp
impedance is
FM: Sensitivity of 20% of frequency range/Vp. Input
impedance is 5
PM
:
pedance
1.2.1.9
Symmetry of all waveform outputs is continuously adjustable from 1
variable duty-cycle pulses, sawtooth ramps and
symmetrical sine waves.
External Modulation Input
>2.5 k0.
kn.
Sensitivity
is 1O
of
kQ.
Symmetry
:1
9 to
When SYMMETRY control is used, indicated frequency is divided by
ima
tely IO.
500)
out/Vp (1.5V
IO0
phase
19:1.
Varying symmetry provides
NOTE
(5OQ).
Controlled
with signal peak plus
5OQ).
DC off-
into 503). Input
shift/VP.
Input im-
non-
approx-
1.2.1.10 Sync Output (TTL)
TTL level pulse which will drive 10
quency and time symmetry are the same as for function output.
TTL
loads. Fre-
0.0002
approximately
1.2.1.5
s
(15V
into
1-2
Hz to 20 MHz in
1
%
Function Output
, 2/ and \ selectableandvariable
p-p into
5On).
5Os1). n
All waveforms and dc can supply 150
IO
overlapping ranges with
vernier control.
(500)
and u up to 15 Vp (7.5 Vp
to 30V
p-p
mA
1.2.1.11 Trlgger and Gate
Input Range: IV p-p to *
Input Impedance: IO
Pulse Width: 25 ns minimum.
Repetition Rate: IO MHz maximum.
Adjustable triggered signal start/stop point: approximately-
90’
to
+90”
IOV.
k0,
33
pF.
to 2 MHz.
1.2.1 .12
Frequency Precision
1.2.2.3
Output
(6OOQ)
Dial Accuracy
+ (I
O/O
of setting + 1 % of full range) on x 100 thru
-
x 1 M ranges.
*+ (2% of setting + 2% of full range) on
-
x 10 and x
10M
ranges.
x
.Ol
Time Symmetry
t
0.5% on x 100 thru x IOOk
2.0 on dial.
t
5% on all other ranges and from 0.02 to 2.0 on dial.
1.2.1
.13
’
Amplitude Precision
Amplitude Change With
Sine variation less than:
t
0.1 dB thru x 1
-+
0.5 dB on x 1 M range;
t
3 dB on x 10M range.
OOk
ranges;
ranges and from 0.2 to
Frequency
’
Step Attenuator Accuracy
-
+
0.3 dB per 20 dB step at 2
1.2.1.14
Waveform Characteristics
kHz.
Sine Distortion
<0.5%
<
All harmonics 30
range.
All harmonics 26 dB below fundamental on x 10
range.
on x 100 Hz to x 10
1
.O%
on x
.01
to x 10 Hz and x 100
kHz.
kHz
ranges.
dB
below fundamental on x I MHz
thru
MHz
2/
, 2/ and ‘L are fixed level
about ground. M and
from 0 to +
1.2.2.4
Voltage control of modulator frequency with sensitivity
of 20% of range per volt. Input impedance is 5
1.2.2.5
5V.
Frequency Modulation (FM IN)
Waveform Characteristics
M
10V
p-p balanced
are fixed level 5 Vp
kQ.
Sine Distortion
<
5%
Time Symmetry
<
O/O
from 1 Hz
5%
from 0.1 Hz to 100
to 10
kHz.
kHz.
1.2.3 General
1.2.3.1
Stability (for amplitude, dc offset and
frequency)
Short Term:t 0.05% for 10 minutes.
Long Term:
1.2.3.2
Specifications apply at
will operate from
1.2.3.3
28.6
cm (11 ?4 in.) wide;
(103/4
in.) deep.
t,O.25O/o
for 24 hours.
Environmental
25°C t 5°C ambient
O°C
to 50°C ambient temperatures.
Dimensions
13.3
Instrument
cm
(5 % in.) high; 27.3 cm
Square Wave Rise/Fall Times
At
FUNCTION
OUT < 25
ns for
15V
p-p into a
5Ost
load.
Triangle Distortion
Odd harmonics within 15% of correct value to 2 MHz.
1.2.2
Modulation Generator
1.2.2.1
Selectable sine 2/ , triangle
ramp M and down ramp M
1.2.2.2
s,2/
ranges.
M, Nl
Waveforms
Frequency Range
,
21
Sweep:
Hzto100
0.1
0.2 Hz
to
2/
,
square ?J , up
.
kHzz
200
kHz
(2 x setting).
in
three
100:1
1.2.3.4
5 kg (11 lb) net; 6.6 kg (14% lb) shipping.
1.2.3.5
90 to
selectable;
Weight
Power
108 to
126V,
105V,,
48 to 400 Hz; less than 40 watts.
Unless otherwise noted, ail specifications
apply from
when FUNCTION OUT is at maximum and
5OQ
terminated, with SYMMETRY control
at OFF. Symmetry and vernier affect frequency calibration. Maximum possible
asymmetry is a function of frequency
setting.
198 to 231V and 216
NOTE
0.1
to 2.0 on frequency dial
to
252V
1-3
SECTION
OPERATION
3.1 CONTROLS AND CONNECTORS
The following item
3.1
.1
Main Generator
1
Frequency Dial
main generator.The setting on this dial
multiplied by the FREQ MULT
basic output frequency of the generator at the
FUNCTION OUT.6 and SYNC OUT 7
The FREQ VERNIER
the modulation generator also affect the main
generator frequency.
Mode
2
Switch
operating mode of the main generator as follows:
a.
EXT GATE Mode
quiescent until a proper gate signal is applied
at the EXTTRIG IN BNC 13 and then outputs
the selected signal for the duration of the gate
signal, plus the time to complete the last cycle
generated.
numbers are keyed
The
-
-
This outer switch selects the
frequency control of the
12
and, in some cases,
-
The main generator is
to figure
12
setting is the
3-1.
BNCs.
b. EXT TRIG
mode, except the main generator output is
one cycle of selected signal only.
c. CONT Mode - The output signal is continuous.
d.
INT TRIG -
cept the trigger signal is applied internally by
the modulation generator.
e. INT GATE
the gate signal is applied internally by the
modulation generator.
TRIGGER LEVEL
is a continuously variable adjustment of the trigger circuitry firing point. When full ccw, a
tive going pulse at approximately quired for triggering (see figure
cw position,
mately +
quired for triggering. In the GATE modes, the‘
generator will run continuously when the control
is cw of 12 o’clock.
Mode
-
Same as for EXT GATE
Same as for EXTTRIG mode, ex-
-
Same as for
Control -
a
positive going pulse at approxi-
7.5V
or more positive voltage is re-
EXT GATE,
This
inner control
7.5V
3-2).
In the full
except
posi-
is
re-
12
13
Figure 3-1
Controls
MOOULATION
and Connectors
GENERATOR
3-1
3.1.2
Modulation
FUNCTION Switch
14
Generator
-
This switch selects the
waveform output of the modulation generator.
Output is at the OUT BNC connector
also available internally to the main generator.
Waveforms are sine
square
(N\).
AMPLITUDE
sweep frequency.
( ‘% ),
ramp up (M) and ramp down
A SWP SET detent holds the OUT BNC at
control 14 level;
( /L ),
triangle
used
to set upper
16
and is
( 2/ ),
.
AM- The instantaneous amplitude of the out-
d
put signal varies with the instantaneous ampli-
tude of the modulation signal.
e.
FM- The instantaneous frequency of the output signal varies with the instantaneous amplitude of the modulation signal.
f
.
PM- The instantaneous phase of the output
signal varies with the instantaneous amplitude of
the modulation signal.
MOD AMPLITUDE Control - This inner con-
trol attenuates the modulation generator signal
that is internally fed to the main generator when
modulating. It has no effect during internal triggering or gating and it has no effect on the
16
OUT
FREQ/PERIOD
15
signal.
MULT Switch - This outer switch
(with ranges given in both frequency and period)
selects the modulation generator frequency/
period range.
-
VARIABLE Control
This inner control sets
the frequency/period within a range.
OUT
(6OOn)
16
Connector - This BNC is the
modulation generator waveform output: sine
( % ),
triangle
level
1OV p-p
balanced about ground and ramp
( 2/ )
and square
(5V p-p into
6OOQ)
waveforms
( ‘L )
fixed
( AA, M )
fixed level 0 to + 5 Vp waveform.
FM IN (5
17
kQ)
Connector - This BNC is
the input
for frequency modulation of the modulation
generator.
Sensitivity is 20% of frequency range
per volt in.
OPERATlON
32
.
convenience,
For
the generator operation has
been grouped in seven basic modes, from which
many variations and combinations are possible.
The following paragraphs give basic switch positions for each mode and requirements and suggestions for operation.
.
g
3.2.1
When setting up the generator, it is advisable to
serve the output on an oscilloscope. Connect
TION
5OQ
-
The dc output can be set from + 15 Vdc
DC
- 15
to
Vdc (+ 7.5 Vdc to - 7.5 Vdc into
5OQ).
Continuous Operation
ob-
FUNC-
OUT to the scope input using a 500 cable and a
load. For continuous waveform output, select a
basic waveform at the desired frequency. Ensure the
modulation switches are set to OFF and the mode is
set to
CONT.
The output amplitude can be as great as
15V p.p;
attenuate as desired.
NOTE
use
For best waveform quality,
TENUATIO N
tion; then use
SWITC H
the
for gross attenua-
AMPLITUDE control for
the AT-
fine attenuation.
The waveform may be skewed to the left or right using
the SYMMETRY control; e.g., making a ramp
from a triangle
(
/G
.)
waveform.
( ,/1 )
/
NOTE
10
The output frequency is divided by
when
the SYMMETRY control is switched from
the OFF position.
The dc level of the waveform may be varied with the
DC OFFSET control, but waveform clipping can occur.
NOTE
The basic modes of operation are:
a.
Continuous- A continuous output signal.
b
.
Triggered
-
One cycle of waveform for each
trigger signal.
C.
Gated
-
A “burst” of waveforms for the dura-
tion of each gate signal.
Waveform clipping will occur (OVERLOAD
LED will light), unless waveform amplitude
is decreased so
that
wavefor m
plus
offset
is less than 7.5 volts at the waveform peak.
3.2. 2
Triggered Operation
In triggered operation, there is one cycle of output for
each trigger signal input. For triggered operation, first
3-3
set up the generator for continuous operation (refer to
paragraph 3.2.1). The main generator may be trig-
gered internally by the modulation generator or trig-
gered by an external source. If an unmodulated
waveform is being output, the modulation generator is
free. to give the desired triggering frequency and
should be selected as the trigger source. Use any
modulation generator waveform as the trigger. Set the
modulation generator frequency for the desired trigger frequency and sync on the modulation generator
output.
NOT E
3.2.4
The TRIGGER LEVEL control
manually triggering or gating the generator. For
manually triggering (single cycles), the generator
mode should be EXT TRIG with no external signal input at the EXT TRIG IN connector. Each time TRIGGER LEVEL is rotated cw through mid-position, one
triggered cycle will be generated. In EXT GATE mode,
the generator runs continuously as long as the TRIGGER LEVEL is
Manual Triggering and Gating
can also be used
CW
of mid-position.
for
1. Trigger frequency must be slower than
the output waveform frequency.
2.
The MOD AMPLITUDE control has no
effect in
Rotate the TRIGGER LEVEL control to obtain a good
trigger. If the waveform start/stop point is to be other
than zero degrees, set the TRIG START/STOP control
as required. (See figure 3-3.) Haverwaves can be set
up using start/stop control and
-90°
s--
If external triggering is to be used, connect a repetitive signal with a positive going transition of greater
than one volt to EXT TRIG IN using a
the TRIGGER LEVEL control for proper triggering.
3.2. 3
In gated operation there is a “burst” of waveforms
lasting the duration of the gate pulse plus the time required to complete the last cycle of waveform started.
Set up as for triggered operation (refer to paragraph 3.2.2)
and select a “trigger” signal whose duty time gives
the desired waveform burst.
AA
Figure
Gated Operation
INT
TRIG
and INT GATE modes.
dc
offset control.
HAVERSINES
_v
3-3..
Waveform Start/Stop Examples
5OQ
cable. Adjust
3.2.5
AM Operation
In amplitude modulation, the instantaneous amplitude
of the output signal varies with the instantaneous
amplitude of the modulation signal.
NOT E
The output waveform will be
LOAD LED will light) if any instantaneous
amplitude greater than
( t
7.5 volts into
Set up the generator as for continuous operation
(refer to paragraph 3.2.1). Switch to internal or external amplitude modulation. If internal, note that the car-
rier (main generator waveform) mean amplitude is
decreased to half. This is to prevent clipping
driving the output amplifier) when the carrier amplitude is modulated (increased and decreased) by
another signal. Set the modulation generator frequency to a lower frequency than the main generator
and sync the scope to the modulation generator output.
Set the modulator amplitude for a desired percentage
of amplitude modulation (0 to greater than 1
and set the main generator ATTENUATION and
AMPLITUDE controls for a desired amplitude of
modulated waveform. If external modulation is
selected, observe that the carrier (main generator
waveform) amplitude drops to zero (null). This is for
suppressed carrier mode of amplitude modulation.
Connect the modulator OUT
IN with a coaxial cable. Set the main generator
ATTENUATION and AMPLITUDE controls
amplitude of suppressed carrier waveform. For best
results when using an external modulation signal at
EXT MOD IN, maintain as near as convenient
amplitude (which does not exceed 5
main generator ATTENUATION and AMPLITUDE controls. An external source may also be used for regular
AM operation, rather than suppressed carrier operation, by supplying a dc component along with the ac
5OQ )
is produced.
clipped(OVER-
t
15 volt s
(6OOQ)
to the EXT MOD
VP);
(over-
00%
range)
for
a desired
a
5 Vp
then use the
3-4
modulating signal (observing 5 Vp limit) to set the car-
rier level.
MAIN DIAL
SETTING
AMPLlTUDE
MOD
MOD IN
EXT
OUTPUT
FREQUENCY
FACTOR*
FM Operation
3.2.6
In FM operation, the instantaneous frequency of the
output signal varies with the instantaneous amplitude
of the modulating signal.
NOTE
The
output
frequency modulation will not
be linear when the instantaneous
modulated frequency exceeds these range
limits.
Upper Limit: 2.0
Lower Limit:
x
FREQ MULT
0.001 x
Upper Limit
Set up the generator as for continuous operation
(refer to paragraph 3.2.1). The frequency setting will
be the center frequency from which the modulated
signal will vary. Switch to internal or external FM
modulation. If internal, set the modulation generator
frequency as desired and sync on the modulation
generator signal. Set the modulation amplitude for the
amount of modulation desired. Because the main
generator is limited in
frequency
range (limited to approximately the dial range for any given multiplier setting),
the
main generator frequency dial and the
modulation amplitude control. must be balanced to
stay within that range. The range is shown in figure 3-4
as the OUTPUT FREQUENCY FACTOR, which, when
multiplied by the FREQ MULT setting, gives the actual
output frequency. Example
swept
over the
full frequency range when the main
1
shows the output being
generator frequency dial is set at midpoint (1 .O), the
MOD AMPLITUDE control set at midrange and the
modulation signal is a balanced waveform
( 2/ , 2/
or \ ); that is, the modulation generator voltage
swings both positive and negative. If the MOD
AMPLITUDE control is rotated toward ccw, the
voltage swing decreases and the angle subtended in
the nomograph decreases. If the MOD AMPLITUDE
control is rotated toward MAX, the angle subtended
would overshoot the OUTPUT FREQUENCY FACTOR
range, indicating that saturation is likely.
Sweep Operation
3.2.7
For sweep operation, set up the generator for con-
tinuous operation (refer to paragraph 3.2.1). The frequency setting will be the lower frequency of the continuously varying (or swept) frequency. Switch
FM/SWP
If
internal set the modulation generator FUNCTION
to internal or external modulation as desired.
2.0-
1.8.
1.6-
1.4-
.8-
.6-
.4-
.2-
.002-
Figure
*Multiply by FREQ MULT for actual output.
3-4.
Frequency Modulation Nomograph
.6
.8
1.6
1.8
switch to SWP SET, monitor the FUNCTION OUT with
a counter or oscilloscope and vary the MOD
AMPLITUDE control for exactly the upper frequency
desired. Note that the main generatoris limited in frequency range (limited to approximately the dial range
for any given multiplier setting). Select either/l/l for
sweep up
rate desired. Keep in mind
orM
for sweep down. Select the sweep
them andb
frequencies are double the indicated frequencies on the
FREQ/PERlOD
control.
Example 2 shows the output being swept from the
bottom of the range to midrange by setting the main
dial fully cw and the VERNIER fully ccw for absolute
bottom of the range. The MOD AMPLITUDE control
)
was left at midrange and the ramp
(/1/1
waveform
used as the modulator. The ramp is a positive going
only waveform. Had a balanced waveform been used,
the angle subtended would have included the dotted
line and resulted in saturation. If an external modulation signal is to be used, the EXT MOD IN values in the
nomograph indicate the signal level required for the
desired results.
3-5
3.1.2
Modulation
FUNCTION Switch
14
Generator
-
This switch selects the
waveform output of the modulation generator.
Output is at the OUT BNC connector
also available internally to the main generator.
Waveforms are sine
square
(N\).
AMPLITUDE
sweep frequency.
( ‘% ),
ramp up (M) and ramp down
A SWP SET detent holds the OUT BNC at
control 14 level;
( /L ),
triangle
used
to set upper
16
and is
( 2/ ),
.
AM- The instantaneous amplitude of the out-
d
put signal varies with the instantaneous ampli-
tude of the modulation signal.
e.
FM- The instantaneous frequency of the output signal varies with the instantaneous amplitude of the modulation signal.
f
.
PM- The instantaneous phase of the output
signal varies with the instantaneous amplitude of
the modulation signal.
MOD AMPLITUDE Control - This inner control attenuates the modulation generator signal
that is internally fed to the main generator when
modulating. It has no effect during internal triggering or gating and it has no effect on the
16
OUT
FREQ/PERIOD
15
signal.
MULT Switch - This outer switch
(with ranges given in both frequency and period)
selects the modulation generator frequency/
period range.
-
VARIABLE Control
This inner control sets
the frequency/period within a range.
OUT
(6OOn)
16
Connector - This BNC is the
modulation generator waveform output: sine
( % ),
triangle
level
1OV p-p
balanced about ground and ramp
( 2/ )
and square
(5V p-p into
6OOQ)
waveforms
( ‘L )
fixed
( AA, M )
fixed level 0 to + 5 Vp waveform.
FM IN (5
17
kQ)
Connector - This BNC is
the input
for frequency modulation of the modulation
generator.
Sensitivity is 20% of frequency range
per volt in.
OPERATlON
32
.
convenience,
For
the generator operation has
been grouped in seven basic modes, from which
many variations and combinations are possible.
The following paragraphs give basic switch positions for each mode and requirements and suggestions for operation.
.
g
3.2.1
When setting up the generator, it is advisable to
serve the output on an oscilloscope. Connect
TION
5OQ
-
The dc output can be set from + 15 Vdc
DC
- 15
to
Vdc (+ 7.5 Vdc to - 7.5 Vdc into
5OQ).
Continuous Operation
ob-
FUNC-
OUT to the scope input using a 500 cable and a
load. For continuous waveform output, select a
basic waveform at the desired frequency. Ensure the
modulation switches are set to OFF and the mode is
set to
CONT.
The output amplitude can be as great as
15V p.p;
attenuate as desired.
NOTE
use
For best waveform quality,
TENUATIO N
tion; then use
SWITC H
the
for gross attenua-
AMPLITUDE control for
the AT-
fine attenuation.
The waveform may be skewed to the left or right using
the SYMMETRY control; e.g., making a ramp
from a triangle
(
/G
.)
waveform.
( ,/1 )
/
NOTE
10
The output frequency is divided by
when
the SYMMETRY control is switched from
the OFF position.
The dc level of the waveform may be varied with the
DC OFFSET control, but waveform clipping can occur.
NOTE
The basic modes of operation are:
a.
Continuous- A continuous output signal.
b
.
Triggered
-
One cycle of waveform for each
trigger signal.
C.
Gated
-
A “burst” of waveforms for the dura-
tion of each gate signal.
Waveform clipping will occur (OVERLOAD
LED will light), unless waveform amplitude
is decreased so
that
wavefor m
plus
offset
is less than 7.5 volts at the waveform peak.
3.2. 2
Triggered Operation
In triggered operation, there is one cycle of output for
each trigger signal input. For triggered operation, first
3-3
set up the generator for continuous operation (refer to
paragraph 3.2.1). The main generator may be trig-
gered internally by the modulation generator or trig-
gered by an external source. If an unmodulated
waveform is being output, the modulation generator is
free. to give the desired triggering frequency and
should be selected as the trigger source. Use any
modulation generator waveform as the trigger. Set the
modulation generator frequency for the desired trigger frequency and sync on the modulation generator
output.
NOT E
3.2.4
The TRIGGER LEVEL control
manually triggering or gating the generator. For
manually triggering (single cycles), the generator
mode should be EXT TRIG with no external signal input at the EXT TRIG IN connector. Each time TRIGGER LEVEL is rotated cw through mid-position, one
triggered cycle will be generated. In EXT GATE mode,
the generator runs continuously as long as the TRIGGER LEVEL is
Manual Triggering and Gating
can also be used
CW
of mid-position.
for
1. Trigger frequency must be slower than
the output waveform frequency.
2.
The MOD AMPLITUDE control has no
effect in
Rotate the TRIGGER LEVEL control to obtain a good
trigger. If the waveform start/stop point is to be other
than zero degrees, set the TRIG START/STOP control
as required. (See figure 3-3.) Haverwaves can be set
up using start/stop control and
-90°
s--
If external triggering is to be used, connect a repetitive signal with a positive going transition of greater
than one volt to EXT TRIG IN using a
the TRIGGER LEVEL control for proper triggering.
3.2. 3
In gated operation there is a “burst” of waveforms
lasting the duration of the gate pulse plus the time re-
quired to complete the last cycle of waveform started.
Set up as for triggered operation (refer to paragraph 3.2.2)
and select a “trigger” signal whose duty time gives
the desired waveform burst.
AA
Figure
Gated Operation
INT
TRIG
and INT GATE modes.
dc
offset control.
HAVERSINES
_v
3-3..
Waveform Start/Stop Examples
5OQ
cable. Adjust
3.2.5
AM Operation
In amplitude modulation, the instantaneous amplitude
of the output signal varies with the instantaneous
amplitude of the modulation signal.
NOT E
The output waveform will be
LOAD LED will light) if any instantaneous
amplitude greater than
( t
7.5 volts into
Set up the generator as for continuous operation
(refer to paragraph 3.2.1). Switch to internal or external amplitude modulation. If internal, note that the car-
rier (main generator waveform) mean amplitude is
decreased to half. This is to prevent clipping
driving the output amplifier) when the carrier amplitude is modulated (increased and decreased) by
another signal. Set the modulation generator frequency to a lower frequency than the main generator
and sync the scope to the modulation generator output.
Set the modulator amplitude for a desired percentage
of amplitude modulation (0 to greater than 1
and set the main generator ATTENUATION and
AMPLITUDE controls for a desired amplitude of
modulated waveform. If external modulation is
selected, observe that the carrier (main generator
waveform) amplitude drops to zero (null). This is for
suppressed carrier mode of amplitude modulation.
Connect the modulator OUT
IN with a coaxial cable. Set the main generator
ATTENUATION and AMPLITUDE controls
amplitude of suppressed carrier waveform. For best
results when using an external modulation signal at
EXT MOD IN, maintain as near as convenient
amplitude (which does not exceed 5
main generator ATTENUATION and AMPLITUDE controls. An external source may also be used for regular
AM operation, rather than suppressed carrier operation, by supplying a dc component along with the ac
5OQ )
is produced.
clipped(OVER-
t
15 volt s
(6OOQ)
to the EXT MOD
VP);
(over-
00%
range)
for
a desired
a
5 Vp
then use the
3-4
modulating signal (observing 5 Vp limit) to set the car-
rier level.
MAIN DIAL
SETTING
AMPLlTUDE
MOD
MOD IN
EXT
OUTPUT
FREQUENCY
FACTOR*
FM Operation
3.2.6
In FM operation, the instantaneous frequency of the
output signal varies with the instantaneous amplitude
of the modulating signal.
NOTE
The
output
frequency modulation will not
be linear when the instantaneous
modulated frequency exceeds these range
limits.
Upper Limit: 2.0
Lower Limit:
x
FREQ MULT
0.001 x
Upper Limit
Set up the generator as for continuous operation
(refer to paragraph 3.2.1). The frequency setting will
be the center frequency from which the modulated
signal will vary. Switch to internal or external FM
modulation. If internal, set the modulation generator
frequency as desired and sync on the modulation
generator signal. Set the modulation amplitude for the
amount of modulation desired. Because the main
generator is limited in
frequency
range (limited to approximately the dial range for any given multiplier setting),
the
main generator frequency dial and the
modulation amplitude control. must be balanced to
stay within that range. The range is shown in figure 3-4
as the OUTPUT FREQUENCY FACTOR, which, when
multiplied by the FREQ MULT setting, gives the actual
output frequency. Example
swept
over the
full frequency range when the main
1
shows the output being
generator frequency dial is set at midpoint (1 .O), the
MOD AMPLITUDE control set at midrange and the
modulation signal is a balanced waveform
( 2/ , 2/
or \ ); that is, the modulation generator voltage
swings both positive and negative. If the MOD
AMPLITUDE control is rotated toward ccw, the
voltage swing decreases and the angle subtended in
the nomograph decreases. If the MOD AMPLITUDE
control is rotated toward MAX, the angle subtended
would overshoot the OUTPUT FREQUENCY FACTOR
range, indicating that saturation is likely.
Sweep Operation
3.2.7
For sweep operation, set up the generator for con-
tinuous operation (refer to paragraph 3.2.1). The frequency setting will be the lower frequency of the continuously varying (or swept) frequency. Switch
FM/SWP
If
internal set the modulation generator FUNCTION
to internal or external modulation as desired.
2.0-
1.8.
1.6-
1.4-
.8-
.6-
.4-
.2-
.002-
Figure
*Multiply by FREQ MULT for actual output.
3-4.
Frequency Modulation Nomograph
.6
.8
1.6
1.8
switch to SWP SET, monitor the FUNCTION OUT with
a counter or oscilloscope and vary the MOD
AMPLITUDE control for exactly the upper frequency
desired. Note that the main generatoris limited in frequency range (limited to approximately the dial range
for any given multiplier setting). Select either/l/l for
sweep up
rate desired. Keep in mind
orM
for sweep down. Select the sweep
them andb
frequencies are double the indicated frequencies on the
FREQ/PERlOD
control.
Example 2 shows the output being swept from the
bottom of the range to midrange by setting the main
dial fully cw and the VERNIER fully ccw for absolute
bottom of the range. The MOD AMPLITUDE control
)
was left at midrange and the ramp
(/1/1
waveform
used as the modulator. The ramp is a positive going
only waveform. Had a balanced waveform been used,
the angle subtended would have included the dotted
line and resulted in saturation. If an external modulation signal is to be used, the EXT MOD IN values in the
nomograph indicate the signal level required for the
desired results.
3-5
3.2.8 PM Operation
In PM operation, the instantaneous phase of the output signal varies with the instantaneous change in
amplitude of the modulating signal. The change in
phase
generator until the correct phase angle change is
made. The modulation circuit differentiates the modulation signal; that is, its output is proportional to the
rate of change of modulation signal amplitude. This
voltage is fed to the ‘main’ generator in exactly the
same manner‘as the FM voltage is. The voltage effects
a change in frequency and, in the case of a step function modulation, for example, exists only long enough
to cause the desired phase shift. Typically, less than
one cycle is required to change the phase. When the
phase angle is increased, the frequency increases to
achieve it. When the phase angle is decreased, the
frequency decreases to achieve it. The frequency required to change the phase also depends upon the
modulation frequency and waveform.
Nominally, the phase of the main generator is shifted
ten degrees for each volt of instantaneous modulation
is
made by changing the frequency of the
NOTE
The output phase will not be linearly modu-
lated
wh
en the instantaneous transition frequencies required to effect the phase
change exceed these range limits.
Upper Limit: 2.0
Lower Limit: 0.001
x
FREQ
MULT
X Upper
Limit
signal. When the main generator is set above a range
midpoint, the modulation signal begins to lose its effectiveness. The effect is that the input signal is rolling
off at 6 dB/octave due to form factor limitations of the
input differentiator. This effect also occurs for
modulation signal frequencies above 150
Set up the generator as for continuous operation
(refer
to
paragraph 3.2.1). Select a range so that the
frequency dial can be set, at midpoint or below (for
linear operation) and switch to internal or external
phase modulation.
NOTE
There is no PM operation for frequency
multipliers of 100 or less.
If internal, set the modulation frequency as desired. (If
other than sine waveform is selected, greater than
150
kHz
modulation frequencies are possible and the
effective roll off must be considered.) Set the modulation amplitude as desired. Full range is 5 Vp and
phase shift is
Because the initial phase reference no
longer exists when the phase shifts, phase
shift measurement will not be possible with
an oscilloscope alone. To measure the
phase shift, an additional circuit
phase modulator will be necessary to
establish a phase angle baseline.
10°
per 1 Vp.
NOTE
kHz.
such as
a
3-6
4.1 FUNCTIONAL BLOCK DIAGRAM ANALYSIS
This
section describes the functions of major circuit
elements and their relationship to one another as
shown in figure 4-1, functional block diagram, and
figure 4-2, basic generator and timing diagram.
Paragraph
circuit
As shown in figure 4-1, the main generator VCG
(Voltage
amplifier receives inputs from the frequency dial, vernier, FM and PM switches which produce a sum cur-
rent. The PM input is provided with a passive differen-
tiator which produces a voltage proportional to the
rate of change of the instantaneous voltage of the
modulating signal.
The VCG summing amplifier is an inverting amplifier
whose output voltage is used to control a complementary current source and current sink. For symmetrical
output waveforms, the currents are equal and directly
proportional to the algebraic sum of the VCG inputs.
The diode gate, controlled by the hysteresis switch,
switches either the current source or sink to the timing
capacitor selected by the frequency multiplier con-
trol. When the current source is switched in, the
charge
the positive-going triangle slope. Likewise, the current
sink produces the negative-going triangle slope.
The triangle amplifier is a unity gain amplifier whose
output is fed to the hysteresis switch and to the
triangle buffer. The hysteresis switch is a bistable
device operating as a window detector with limit
points set to the triangle peaks. When the hysteresis
switch output is +
limit, and the hysteresis switch goes to - 2V. This
switches currents at the diode gate and the negativegoing
reaches the
switch back to positive, repeating the process. As
shown in figure 4-2, this repetitive process results in
the simultaneous generation of a square wave and a
triangle wave at the same frequency.
4.2
provides further descriptions relating
blocks
on the capacitor will rise linearly producing
triangle slope is started. When the triangle
to schematics in section 7.
Control of Generator frequency) summing
2V,
- 1.25V
the triangle rises to the +
limit, the hysteresis switch will
1.25V
selector and by the magnitude of the currents supplied to and removed from it. Since the currents are
linearly proportional to the sum of VCG inputs, so will
be the output frequency. The capacitance multiplier
provides the bottom four frequency ranges.
When the variable symmetry control is rotated, it first
reduces the current sink by a factor of 19, making the
negative-going triangle slope 19 times longer than
normal. This results in an unsymmetrical waveform
output and a frequency division by IO. Continued rotation gradually increases the current sink and reduces
the current source in such a way that the period for
the triangle to complete one cycle remains constant.
This action produces continuously variable symmetry
of the output waveforms over a
while frequency remains constant at one-tenth of dial
and multiplier settings.
The inverted square from the hysteresis switch is fed
to the sync amplifier, where it is buffered and converted to a TTL level output, and to the square
amplifier, where (if square or pulse functions are
selected) a buffered square is sent to the signal
shaper for conditioning.
The triangle buffer provides the
cient drive for the signal shaper and presents a small,
constant load on the triangle amplifier.
, HYSTERESIS
.
C
A
-
1:19
to
$_
1.25 triangle suffi-
I
SWITCH
*
19:1
range
6
The output frequency is determined by the magnitude
of the capacitor selected by the frequency multiplier
Figure
4-2.
Basic Generator Block
and Timing Diagram
4-l
The signal shaper contains switching elements and a
diode array for signal conditioning the buffered
triangle and square inputs into the various waveforms
controlled by the function switch. The selected
waveform is the carrier (+ Y) input to the transconductance multiplier,
an integrated circuit,
four-
quadrant multiplier.
The modulation (+X) input is a positive dc from the
control
amplifier when the
AM switch
is off, providing
a fixed gain reference for the multiplier. Output currents from the multiplier are applied to the summing
node of the preamplifier for conversion to an inverted
voltage signal.
The preamplifier output is then attenuated by the front
panel amplitude control and fed to the output amplifier
summing node along with the dc offset control. The
output amplifier is an inverting amplifier whose output
is fed into step attenuator and then to the function
output connector.
distributed network having
This network provides attenuation in 20 dB
steps to 60
dB.
The attenuator consists of a
5Oa
output impedance.
(1/1
0)
For continuous operation of the basic function
generator loop (bold path in figure 4-1), the trigger
amplifier must maintain a positive level above the
most positive charge on the
integrating
capacitor in
order to reverse bias the start/stop diode. Thus, in
continuous mode the trigger logic senses the continuous control line from the front panel mode selector and holds the inverting trigger amplifier input low.
In triggered and gated modes the trigger amplifier outputs some level below the positive peak charging
level, and the start/stop diode is forward biased to
sink the current source and prevent the timing
capacitor from charging to the positive peak. This
stops waveform generation and holds the triangle output at some dc level called the trigger baseline. The
trigger baseline is the level where a triangle, and thus
sine, waveform starts and stops when triggered or
gated.
The normal trigger baseline is zero volts, analogous to
0°
phase of a sine or triangle waveform. The trigger
start/stop control offsets the trigger amplifier output
and can change the baseline for starting and stopping
a sine or triangle waveform from its negative peak
(- 90°)
to its positive peak (+
90°)
At the extreme
positive peak level setting though, the diode is again
reverse biased and generator operation goes con-
tinuous, independent of generator mode.
While the integrating capacitor is being held from
charging, the start/stop diode must sink the current
source, which has a magnitude variable with VCG inputs. Therefore, a compensation is necessary to the
voltage level output by the trigger amplifier in order to
maintain a constant baseline level as VCG inputs, cur-
rent source magnitude and forward voltage drop by
the start/stop diode are varied. The baseline compensation circuit measures the forward voltage across a
diode placed in the current source and injects an offsetting current into the trigger amplifier to maintain an
equal voltage differential between the baseline level
and trigger amplifier output.
The trigger logic determines that after a waveform
starts, it always stops at a complete cycle and at the
same phase angle at which it started. The trigger logic
receives a trigger stimulus from the signal limiter and
latches the trigger amplifier output positive, allowing
the generator loop to run. When the negative peak of
the last cycle is reached (just one cycle in trigger
mode), the square from the hysteresis switch latches
the trigger amplifier back to its previous level. The integrating capacitor will charge back to the trigger
baseline
where the start/stop diode once again for-
ward biases.
The generator mode switch sets the gated control line
to determine whether the trigger logic is to latch the
generator on for one cycle of for the duration of the
trigger stimulus.
The modulation generator board contains the power
supplies, the modulation generator and various
switching elements to control the source and type of
modulation and triggering signals to the main
generator.
The modulation generator is an integrated circuit
source of sine, triangle and square waveforms, whose
frequency is controllable by front panel multiplier
switch, variable control, and external voltage at the
FM IN input. The triangle and square are applied to a
ramp generator consisting of a balanced modulator
and buffer amplifier to produce ramp waveforms. A
modulation waveform or a SWP SET dc level is sent to
the function buffer via the front panel function selector.
The function buffer output is sent to the modulation
output
(6OOQ)
connector, the generator mode switch
for an internal trigger and gate stimulus, and the
amplitude buffer after being attenuated by the front
panel amplitude control. The amplitude buffer output
goes to the AM, FM and PM switches “internal” positions. The EXT MOD IN connector provides a
connec-
4-2
tion for an external signal to the switches “external”
positions.
The FM and PM switches provide VCG inputs. The AM
switch controls the control amplifier and thus the
transconductance multiplier. When AM is off, the control amplifier produces a positive dc level giving the
multiplier a fixed gain. With internal AM, the dc component from the control amplifier is cut in half, halving
the output amplitude to prevent output clipping when
modulating. The selected
modulation
waveform rides
on the dc. The ac (modulation signal) has a peak value
equal in magnitude to the dc level when the modulation amplitude control is maximum, making the sum of
modulator and carrier signals equal to the maximum
output capability of the output amplifier, and the difference equal to the zero output level, which is
100%
modulation. Then, by varying the modulation signal, a
variable 0 to 100% AM of the carrier (main generator)
signal is produced. With external AM, the dc component is switched to 0 Vdc, resulting in zero amplitude
output, and bipolar signal inputs at the EXT MOD IN
connector will produce suppressed
carrier (4-quadrant)
modulation.
4.2
CIRCUIT ANALYSIS
4.2.1 VCG Amplifier
across series resistors to the supplies equal to the
control voltages. The FET currents will be switched at
the diode gate into a timing capacitor to produce the
triangle waveform.
4.2.2
Symmetry Control
Let the source of Q2 be - 5 Vdc, the wiper of the sym-
-
metry control,
2.5 Vdc, and the source of
The output of U3 will have no current, each
tor will have
1 mA,
and generator frequency will be at
Q1,
R/2
0 Vdc.
resis-
maximum of the range. Open the symmetry switch
and set the potentiometer to its electrical center. The
output of U3 is still at an equipotential point, but now
the total resistance with 5 Vdc across it has changed
from
R
to
10R.
Thus, current will drop to 100 PA and
output frequency will drop to one-tenth of range. If the
potentiometer is rotated, current will flow in U3 output
to maintain the wiper at
-
2.5 Vdc. When the potentio-
meter is ccw, the wiper is at the positive direction and
the upper R/2 will have
source of
which puts
1 mA.
But the lower R/2 is in series with
2.5V
across 19 x the normal resistance.
2.5V
across it with a current
9R,
Now the current sink will have one-nineteenth the
magnitude of the current source. The output waveform
for this condition is shown in figure 4-3. Regardless of
where the symmetry control is rotated, frequency
stays the same (one-tenth of range).
Figure 4-3 is a simplified schematic of the VCG circuitry. The value of a resistor
“R”
is 5 kQ and supplies
are t 15 Vdc. U1 is connected as a summing
amplifier to sum the VCG inputs. A top of range input
produces 1
in
-
5 Vdc at the output of
mA
through the feedback resistor resulting
U1
The negative input of U4 is held at the output level of
U1 by controlling the current through
back. One half the output of U1 is buffered by
Q2
as a feed-
U3
and
applied to the wiper of the variable symmetry control.
The negative input of
U2
is held at 0 Vdc by controlling
the current through Ql as a feedback. As long as the
variable symmetry control is off, the two R/2 resistors
have equal voltage across them and an equal current
through them as through U1 feedback and there is no
current at the output of U3. Since an equal current
exists in the entire resistor string from + to
-
supply,
the result is a positive control voltage relative to the
negative supply at U6 + input and a negative control
voltage
with respect to the positive supply at U5 + input, each of which is proportional to the sum of the inputs to
U1
Similarly, U5 and U6 establish feedback by regulating
current through
FETs,
producing a voltage drop
4.2.3
Range Switching
For frequency ranges associated with multiplier posi-
tions of 100 and
1
K, main board schematic, sheet 1,
the value of the current source and current sink setting
resistors R326, R38, R48 and R330 is 5
kQ,
which provides integrating current sensitivity of 200 PA per volt
of external FM input. With the timing capacitors of 1
and 0.1
pF,
plus the bulk of the top range timing capacitor and the stray capacitance of the multiplier switch,
the generator produces the calibrated output frequency for these ranges. In the top range (multiplier
position of
10M),
the current setting resistors are
paralleled with resistors of one-ninth the value, causing both current sources to run at ten times the usual
current, resulting in 2
When this current is used with the nominal -90
mA
per volt of external FM input.
pF
timing capacitor (fixed value plus strays), the top
range of frequencies result. For the next three ranges
down (multiplier positions
1
M,
100K, 10K),
the nominal
timing capacitor is the fixed top range capacitor plus
strays (i.e.,
910
pF,
0.01
pF).
capacitors of 101
plus the switched values (11
These result in joint timing
pF,
1010 pF and 0.0101
,uF.
pF,
In these
-90
pF)
three ranges the positive and negative current
sources are boosted by 1
%
over the next range down
4-3
VERNIER
+
R/9
OM
I
1
R/2
1
R
v
SYMMETRY
R
R/2
z
Figure 4-3. VCG Simplified Schematic
-
R/9
(multiplier 1 K) by switching 500
with the 5
R38, R48 and R330 producing the output frequencies
for these ranges. The four ranges below multiplier setting 100 all have the same integrating current and timing capacitor as the 100 multiplier range, but for each
of these ranges, 90%, 99%, 99.9% and 99.99% of
the integrating current is subtracted by the
capacitance multiplier circuit.
4.2.4
For the frequency ranges associated with multiplier
positions of
circuit (main board schematic, sheet I), senses the
timing capacitor charging current and subtracts the
appropriate amount so that the effective charging current is a fraction of that delivered by the current
sources. This is accomplished by the connection of
the capacitance multiplier to the timing capacitor with
one input-output terminal through a section of the frequency multiplier switch. The + terminals of U7 and
U8 serve as potentiometric input to these amplifiers.
U7 has a fixed resistive feedback network, giving it a
fixed gain. Capacitor C26 is forced to comply to the
triangle voltage wave being generated, because the
R54
side is driven at the potential of the input/output
terminal and the other side has the same waveform
with some fixed gain from U7. Since the side driven at
the input/output signal is a summing node, it is fed the
necessary
R59, R60 and
selected by the frequency multiplier control, taking on
values which give the correct amplitude to the output
of U8. This output with respect to common is
input/output waveform with a square wave super-
imposed;
be picked up for signal tracing. The input/output waveform is a triangle wave so the differential across
and R63 is a square wave with the correct amplitude
to subtract part of the timing capacitor charging cur-
rent. Since this square wave amplitude is controlled in
decades by the frequency multiplier control via R58,
R59,
R60 and
in decades even though the current sources and timing capacitor remain the same.
kS2
basic current setting resistors
Capacitance
10
current by the feedback resistors
TP1
is the test point where this output can
R61,
Multiplier
through 0.01, a capacitance multiplier
R61.
The feedback resistors are
the instrument frequency is divided
kS2
resistors in parallel
R326,
R58,
the-
R62
a matched duplicate FET,
drain current as Q8 and, therefore, the same
source voltage. In series with Q10 is a duplicate emit-
ter follower,
rent as
base voltage. Since the gate of the dummy FET,
connected to the emitter of the dummy emitter
follower,
voltage. Therefore, within the tolerances of the part
parameters and some unaccounted error for base
current, the active emitter follower output voltage will
be at the value of the input gate. The remaining transistor,
dynamic lead networks at the input of the hysteresis
switch. In this role, it needs no dc integrity, as the output is not directly coupled.
4.2.6
Hysteresis Switch
The hysteresis switch (main board schematic, sheet 2)
consists mainly of
and
Q14/Q15,
of U14 compares an input voltage to common. The input network provides a positive bias to one and a
negative bias to the other; therefore, when the input
terminal (output of the triangle amplifier) is at
the flip-flop changes state. The flip-flop selects which
input comparator of the hysteresis switch will be activated in preparation for the next change of state.
When the timing capacitor is integrating positively,
the positive biased comparator is activated. When the
timing capacitor voltage reaches +
changes state, the negative comparator is activated
and the direction of integration is reversed, so that
when the timing capacitor signal reaches
the flip-flop switches back and the cycle starts over.
In addition to the positive and negative biases at the
comparator inputs, there is a dynamic lead network
on each one. These lead networks are driven by
a separate emitter follower, from the triangle ampli-
fier. They provide the necessary lead to compensate
for the inherent delays of the hysteresis switch,
thereby keeping the higher frequency dial nonlinearity
and sine distortion to a minimum.
4.2.7
Q13. Q13
Q10;
therefore, it has the same
Q13,
the two terminals have the same
Qll,
is a second emitter follower for driving the
U14,
an output flip-flop. Each differential pair
Diode Gate and Timing Capacitor
Q9. Q9
has the identical collector cur-
a double input comparator,
has the identical
gate-to-
emitter-to-
Q9,
is
t 1.25V,
1.25V,
the flip-flop
- 1.25V,
Q11,
4.2.5
Triangle Amplifier
The main board schematic, sheet 2, shows the
amplifier; it uses
Q10,
a bipolar emitter follower, for an open
of
one. It is a fast, very high input impedance circuit
with output impedance low enough to drive the hysteresis switch and the triangle buffer. In series with
Q8,
an FET source follower, to drive
triangle
loop gain
Q8
is
The diode gate (current switch) and the timing
capacitor circuits are shown in the main board
schematic, sheet 2. The current source and sink are
switched to the timing capacitor by the hysteresis
switch via a diode bridge arrangement called the
diode gate. Actually, the hysteresis switch is linked to
the bridge network by two emitter followers,
Q25,
with independent outputs biased to be at the
Q24
and
4-5
same voltage.
lustrated in figure 4-2 shows these points as one terminal at C. When the hysteresis switch output is
positive, CR16 is forward biased, so that the current
sink is sourced by the drive circuit and is ineffective.
CR1
3 and CR1 5 are reverse biased, providing isolation
between the drive circuit and the timing capacitor.
This leaves CR14 forward biased and free to conduct
the current source output to the timing capacitor.
When the timing capacitor voltage rises to the hysteresis switch point
output switches low, forward biasing CR13 which
back biases
be isolated and the sink to discharge the timing capacitor through CR15 This state continues until the
negative switch point is reached and reverts to the
previous state.
4.2.8 Triangle Buffer
The triangle buffer (main board schematic, sheet 2) is
a wide band dc amplifier providing a closed loop gain
of one in potentiometric connection. The input differential stage,
emitters are fed from a current sink
collector load, Q20, is a current source providing
greater open loop gain than a resistive load. Following
this is an emitter follower,
shifter, CR1 2, and another emitter follower, Q22, for
the
output stage. The gain is set to one by the 100%. .
feedback to the input pair feedback side, base of
4.2.9 Signal Shaper
The signal shaper circuit (main board schematic,
sheet 3) is uniquely set’up for each different waveform
by four wafers of the function selector switch. The
t
15 volt power is switched off in the triangle wave
mode and there is virtually no effect on the triangle
wave fed to the circuit. In the positive pulse mode, the
square wave, rather than the triangle wave, is fed to
the circuit and the
a result, the negative swing of the input square wave
is clipped off. The negative pulse is formed, when
selected by the function switch, in a similar manner.
When the square or sine wave is selected, both plus
and minus 15 volt power is applied to the circuit. The
difference in circuit setup for sine and square is the
resistive load at the circuit output and the shape of the
signal fed to the input. For the sine wave mode, the
matched set of -diodes soft clip the input triangle at
three different levels. These signals are resistively
summed to produce a sine wave voltage input to the
multiplier.
square wave is symmetrically hard-clipped by the
The simplified timing diagram il-
(+ 1.25V),
CR14
and CR16 and allows the source to
Q17/Ql8,
-
15 volt power is switched off. As
For the square wave mode, the input
the hysteresis switch
is a monolithic pair. The
Q19.
The active
Q21,
a
zener
diode level
Ql8.
diode network presenting a square wave input voltage
to the multiplier.
4.2.10
Transconductance Multiplier
After the main generator signal passes through the
function selector switch and the signal shaper circuit,
it enters a transconductance multiplier,
board schematic, sheet
by dc from the control amplifier or modulated by ac
from the modulation generator via the AM switch. Cur-
rents in the open collectors of this
a current mirror for optimum gain and fed to the preamplifier summing node for conversion to a voltage
signal at TP7.
4.2.11 Preamplifier
The preamplifier (main board schematic, sheet
the output (power) amplifier, is comprised of a high
frequency ac amplifier combined with a low frequency
dc amplifier. It converts the current from the multiplier
to a voltage signal which is attenuated by the front
panel amplitude control and amplified by the output
amplifier. The
remaining circuitry is the high frequency amplifier.
Again, like the output amplifier, the ac amplifier
symmetrically arranged from the
node to R246 and R249 at the output stage of the
preamplifier. If the input current goes into the node,
the voltage at the summing node will rise by a certain
amount. By capacitive coupling via C92 and C93, the
base voltage of Q40 rises closer to + 15 volts and the
base voltage of
volts. Thus, the emitter base junction of Q40 will be
less forward biased, thereby reducing the emitter current, while the Q41 ‘emitter current increases. The
result is an increase in current in Q42 and a decrease
in current in Q43, causing a decreased voltage output
in
R246/R249.The
summing node tends to cancel the rise in voltage
there, causing the output voltage to stabilize. The
amount of negative voltage at the output required to
pull the summing node back to zero is determined by
the value of R240.
4.2.12 Output
The output amplifier is comprised of a low frequency
dc amplifier and a high frequency ac amplifier. Refer
to the simplified circuit of figure 4-4. The
and Q38 circuit is the dc amplifier and the remaining
circuitry is the ac amplifier. The ac amplifier is symmetrically arranged, top and bottom. The upper
U17
Amplifier
3),
where the amplitude is set
IC
circuit is the dc amplifier and the
R240/R238 summing
Q41
rises further away from - 15
feedback path through R240 to the
U15
(main
are worked into
3),
like
is
U19,
Q37
por-
4-6
tion amplifies the positive swing of the output, while
the lower mirror portion amplifies the negative swing.
Operation is class AB; that is, there is independent
positive half and negative half amplification, with a
small amount of current flow in both sides near zero
swing. The amplifier schematic has been simplified in
figure 4-4 for the following discussion. Assume that
both the input and the output voltages are zero, then
the voltage at point A should also be zero. Because of
the symmetrical configuration of the amplifier, the
current through
output will remain at zero.
the voltage at point A will rise by a certain amount.
This will cause the base voltage of Q47 to
to + 24 volts. Thus, the emitter base junction of Q47
will be less forward biased, thereby reducing its emitter current. The result is that the voltage at point B
Q47
and Q49 will be equal and the
If
the input goes positive,
rise
closer
which is the output voltage, will start to go negative.
Finally, when the output has moved far enough
negative to pull point A back to zero, by pulling current
through the feedback resistor Rfb, the collector current of Q47 and Q49 will again be equal and the output
voltage at point B will stabilize. The amount of nega-
tive voltage at the output required to pull point A back
to zero is controlled by the ratio of
ratio is the closed loop gain of the output amplifier.
The circuit containing
low frequency amplifier used to bias the high frequen-
cy amplifier and to increase the low frequency loop
gain. The high frequency amplifier is isolated from low
frequency signals at the input by capacitance coupl-
ing to the bases of
low frequency amplifier to bias the emitter of
obtain the required dc stability and high loop gain.
U19,
Q37, Q38 is a high gain,
Q47
and Q49. It then employs the
Rfb
to Rin, and this
Q47
to
INPUT 0’
U
19/Q37/Q38
+ 24 Vdc
r
Rin
+
_A
OUTPUT
Figure
- 24
4-4.
Simplified Output Amplifier
I
Vdc
4-7
Other circuit components are shown on the main
board schematic, sheet 3. Emitter followers Q46 and
Q48 increase the driving power to the bases of Q47
and Q49.
the load through R287 and
signalswing. Q51 and Q54 are driven by the collector
of Q47. CR40 through CR43 compensate for the
emitter-base junction voltage drops of
and Q55 to control idling current, reduce crossover
distortion and prevent thermal runaway. The two
resistor-capacitor networks,
R278/C101
high frequency amplifier gain during the transition
time prior to the dc amplifier taking effect. This improves the rise time, since the dc amplifier requires a
few microseconds to respond and stabilize. VR2 and
VR3 are five volt regulators which normally run
saturated to supply the output stage current to the col-
lectors of the output transistors. If the output stage
should demand an abnormal amount of current
through a shorted’transistor or output terminal, the
current through R295 through R298 will generate five
volts of drop. If more current is demanded, the
regulators will simply maintain the five volt drop,
allowing the output collector voltages to collapse,
preventing
amplifier components. The dc offset is fed as a cur-
rent from the front panel control to the output
amplifier summing mode.
4.2.13
The side of the hysteresis switch (main board sche-
matic, sheet 2) not used to drive the current switch
has an inverted square signal which is used to drive
an emitter coupled pair,
output of Q23 is biased to provide a TTL level output,
The sync out signal is connected to the front panel
sync out (TTL) BNC with a coaxial cable.
Next to the sync amplifier, a similar emitter coupled
pair, Q57 and Q58, is connected to the same input
and biased to output a bipolar square wave to the
function switch when square or pulse functions are
selected. In other functions, emitter bias is reversed
so that the square function remains confined to the
hysteresis switch area.
4.2.14
Either an external signal or the modulator function are
selected by the generator mode switch (auxiliary
generator schematic) and summed through R50 and
R51
with the trigger level control. That portion of the
trigger signal more positive than the trigger level is
Q51
and Q54 are a harnessed pair sharing
R291
during the positive
Q51,
Q53, Q54
R268/C1
are emitter bypass circuits to maintain the
excessive power dissipation
Sync Amplifier/Square Amplifier
Q16
and Q23. The collector
Trigger Signal Limiter
00 and
in the
clipped by forward biasing
is clipped by CR2. While
Q3
switches off to a TTL low level. While CR2 is on,
is off and Q3 saturates to a
R58 provide hysteresis to ensure a clean square wave
output.
4.2.15
In continuous mode the continuous control line is low
and
trigger flip-flop (U12) cleared.
sent by emitter follower Q27 to a diode “AND”. A low
is sensed at RI 58, the trigger amplifier inverting input.
The closed loop gain of the trigger amplifier is set by
the ratio of
puts a + 1.5 to + 2 Vdc to reverse bias the start/stop
diode CR27 above the most positive charge on the integrating capacitor.
In trigger mode, both control lines are high, and U13
produces a narrow negative pulse, corresponding to a
high to low transition of the signal limiter output, to
clear
clocked by the negative-going edge of the current
switch
goes high and the trigger amplifier goes to a low level,
forward biasing CR27 which sinks the VCG current
source away from the
charge level on the
voltage drop across CR27 above the trigger amplifier
output. Compensation current enters the trigger
amplifier summing node through
put voltage down exactly the same as the drop
CR27 at a particular magnitude of integrating current.
The 0 Vdc trigger baseline may be modified with
the front panel start/stop control. Whenever a trigger
is received, U13 is cleared and the trigger amplifier
output goes high, allowing the integrating capacitor to
charge. At the positive triangle peak, the hysteresis
switch goes to a negative level and the negative-going
triangle slope is generated. The high-to-low hysteresis
transition clocks
of the square is also fed into the diode “AND” at
CR20 which holds the trigger amplifier output high until the completion of the negative-going slope of the
triangle. When the hysteresis switch returns to
positive, the trigger amplifier returns to its low output,
and the integrating capacitor charges until CR27 forward biases again. The integrating capacitor is again
held at the trigger baseline level.
In gate mode, the gated control line is low and
produces a negative pulse of the same duration as the
signal limiter output. Thus,
Trigger Logic/Trigger Amplifier
U13-8
(main board schematic, sheet 2) holds the
R173
to R158. The trigger amplifier out-
U12.
In the absence of a trigger stimulus,
square
translated by Q26 to TTL levels.
integrating
U12-3
CR1;
the negative portion
CR1
is on, Q1 conducts and
Q1
TTL
high level. R57 and
U12-3
is low, which is
U12
is
U12-3
integrating capacitor. The
capacitor is held at the
R155
to push its out-
across
R4,
high, but the negative portion
U13
U12
is held cleared, the
4-8
signal at CR24 is held low, and the trigger amplifier
output is held high for the trigger duration. The last
triangle cycle started is completed through the action
explained in trigger mode.
the start/stop diode are matched and carry equal cur-
rents, the trigger baseline will be stable with varying
VCG inputs.
4.2.16 Baseline
CR2 (main board schematic, sheet
the current supplied by the VCG current source.
Compensation
1)
is in series with
U9-3
is connected to CR2 anode and, since it is a voltage
follower, it will have the same potential at its pin 6.
U1 O-3 is connected to CR2 cathode and will regulate
the current through
Q7
to make the same potential at
its pin 2; therefore, R64 will have the same voltage
across it as the drop across CR2. The current leaving
Q7
enters the trigger amplifier summing node, and
becomes a voltage offset equal to the drop across
CR2, because R64 and the feedback resistor for the
trigger amplifier have the same value. Since CR2 and
4.2.17
Modulation Generator and Ramp Generator
The function generator used as a modulation source
in the instrument is the lntersil 8038 (U2 on the auxiliary generator board schematic). It is fitted with an
auxiliary current balancing circuit
(U1)
to extend its
useful dynamic range. The ramp output which is not
built into the chip is developed by amplitude
modulating the triangle function with the square func-
tion in a balanced modulator
(
2/
, 2/ ,
‘1_1
,
M,
(U5).
The output signals
M )
are selected in a
function selector switch and fed to the modulation
switches where modulation type is selected. The
ramp signals have a fundamental frequency of two
times that of the other waveforms.
4-9
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.