VOLTCRAFT DSO-1104F, DSO-1204E User guide

DSO1000E/F Series
Digital Storage Oscilloscope
User Manual
Version 1.5

Copyright Decl aration

Digital Storage Oscilloscope I

General Safety Summary

Read the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To evade potential hazards, use this product only as specified.
Only qualified personnel should perform maintenance. Avoid fire or personal injury. Use suitable power cord. Use only the power cord specified for this product and
certified for the country of use. Connect and disconnect properly. Connect a probe with the oscillosc ope before it is connected to measured circuits; disconnect the probe from the oscilloscope after it is disconnected from measured circuits. Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. Connect the probe in a r igh t way. The probe ground lead is at ground potential. Do not connect the ground lead to an elevated voltage. Check all terminal ratings. To avoid fire or shock hazard, check all ratings and markings on the product. Refer to the product manual for detailed information about ratings before making connections to the product. Do not operate without covers. Do not operate this product with covers or panels removed. Avoid exposed circuitry. Do not touch exposed connections and components when power is present. Do not operate with suspected failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Assure good ventilation. Do not operate in wet/damp environments. Do not operate in an explosive atmosphere. Keep product surfaces clean and dry.
Digital Storage Oscilloscope II

Safety Terms and Symbols

Terms on the product. The following terms may appear on the product:
Danger It represents that harms may be caused to you at once if you perfor m the
operation.
Warning It represents that latent harms may be caused to you if you perform the
operation.
Notice It represents the damage possibly caused to the product or other
properties if you perform the operation.
Characters on the product. The following characters may appear on the product:
Notice
Please read
the manual
Protective
ground terminal
Measuring
ground terminal
Chassis
ground terminal

Product Scrapping

Device Recycling
We need extract and utilize natural resources to produce this device. If you do not reclaim the device in a proper way, some substances it contains may become harmful or poisonous to environments or human bodies. To avoid them being released outside and to minimize the waste of natural resources, we suggest you reasonably call back this device to ensure proper recovery and recycling of most materials within it.
Digital Storage Oscilloscope III

Brief Introduction to the Series of digital storage oscilloscope

DSO1000E/F Series osc illoscopes cover t he bandwidths from 80MHz t o 250MHz, and provide the real-time up to 1GSa/s. In addition, they have 7 inch color TFT LCD as well as WINDOWS-style interfaces and menus for easy operation. What’s more, the plenty menu information and the easy-to-operate buttons allow you to gain information as much as possible in measurement; the multifunctional knobs and the powerful shortcut keys help you save a lot of time in operation; the Auto Scale function lets you detect sine and square waves automatically; By using the three methods the oscilloscope provides (context-sensitive, hyperlinks, and an index), you may master all operations on the device in quite a short time so as to greatly improve your efficiency in production and development.
Model Channels Bandwidth Sample Rate AFG
DSO-1084E DSO-1104E DSO-1204E DSO-1254E DSO-1084F DSO-1104F DSO-1204F DSO-1254F
4 80MHz 1GS/s ­4 100MHz 1GS/s ­4 200MHz 1GS/s ­4 250MHz 1GS/s ­4 80MHz 1GS/s Yes 4 100MHz 1GS/s Yes 4 200MHz 1GS/s Yes 4 250MHz 1GS/s Yes
Digital Storage Oscilloscope IV

Contents

Copyright Declaration ................................................................................................ I
General Safety Summary ....................................................................................... II
Safety Terms and Symbols .................................................................................... III
Product Scrapping ................................................................................................. III
Brief Introduction to the Series of digital storage oscilloscope ............................... IV
Contents ................................................................................................................... V
Chapter 1 Introduction ........................................................................................ 1
1.1 General Inspection ................................................................................... 2
1.2 Prepare Instrument for Use ....................................................................... 3
1.3 Brief Introduction of Front Panel ............................................................... 4
1.4 Accidence of the user interface ................................................................. 5
1.5 Functional Check ...................................................................................... 6
1.5.1 Connect the oscilloscope ...................................................................... 6
1.5.2 Observe the waveform ......................................................................... 6
1.6 Probe Introduction .................................................................................... 7
1.6.1 Safety ................................................................................................... 7
1.6.2 Probe Check Wizard ............................................................................. 7
1.6.3 Manual Probe Compensation ............................................................... 7
1.6.4 Probe Attenuation Setting ..................................................................... 8
1.7 Remote Control via USB ........................................................................... 9
1.7.1 Install Software ..................................................................................... 9
1.7.2 Connect DSO scope to PC ................................................................. 10
1.7.3 User’s Interface of PC ........................................................................ 11
Chapter 2 Function Introduction ...................................................................... 13
2.1 Menu and Control Keys .......................................................................... 14
2.2 Connector ............................................................................................... 15
Digital Storage Oscilloscope V
2.3 Universal Knobs and Softkeys ................................................................ 15
2.4 Oscilloscope Setup ................................................................................. 16
2.5 Horizontal Controls ................................................................................. 17
2.6 Vertical System ....................................................................................... 18
2.6.1 Vertical Controls ................................................................................. 18
2.6.2 Math Operation ................................................................................... 20
2.7 Trigger System ....................................................................................... 24
2.7.1 Edge Trigger ....................................................................................... 26
2.7.2 Pulse Trigger ...................................................................................... 26
2.7.3 Video Trigger ...................................................................................... 28
2.7.4 Slope Trigger ...................................................................................... 29
2.7.5 Overtime Trigger ................................................................................. 30
2.7.6 Window Trigger .................................................................................. 31
2.7.7 Pattern Trigger .................................................................................... 32
2.7.8 Interval Trigger ................................................................................... 33
2.7.9 Under Amp Trigger ............................................................................. 34
2.7.10 UART Trigger ..................................................................................... 35
2.7.11 LIN Trigger ......................................................................................... 39
2.7.12 CAN Trigger ....................................................................................... 41
2.7.13 SPI Trigger ......................................................................................... 44
2.7.14 IIC Trigger .......................................................................................... 47
2.8 Save/Recall ............................................................................................ 51
2.8.1 Internal Save and Recall ..................................................................... 51
2.8.2 External save and recall ..................................................................... 52
2.9 Measure System ..................................................................................... 53
2.9.1 Scale measurement ............................................................................ 53
2.9.2 Cursor measurement .......................................................................... 54
2.9.3 Automatic Measurement ..................................................................... 55
2.10 Acquire ................................................................................................... 59
2.10.1 XY Mode ............................................................................................ 62
2.10.2 Roll Mode ........................................................................................... 62
2.11 Display ................................................................................................... 62
Digital Storage Oscilloscope VI
2.12 Utility System .......................................................................................... 63
2.12.1 Update Firmware ................................................................................ 64
2.12.2 Self Calibration ................................................................................... 64
2.12.3 Pass/Fail ............................................................................................ 65
2.13 Fast Action Buttons ................................................................................. 66
2.13.1 AUTO SCALE ..................................................................................... 66
2.13.2 Default Setup...................................................................................... 68
Chapter 3 Waveform Generator ........................................................................ 71
3.1 Set Wave Type and Parameters ............................................................. 72
3.2 Edit Arbitrary Waveform .......................................................................... 73
3.3 Output Arbitrary Waveform...................................................................... 75
Chapter 4 Troubleshooting ............................................................................... 76
4.1 Problem Settlement ................................................................................ 76
Chapter 5 General Care and Cleaning ............................................................. 78
5.1 General Care .......................................................................................... 78
5.2 Cleaning ................................................................................................. 78
Appendix A: T echnical Specifications ................................................................... 79
Appendix B: Accessories ....................................................................................... 86
Appendix C Harmful and Poisonous Substances or Elements ............................ 87
Appendix D To replace the fuse ............................................................................. 88
Appendix E Open Source Information ................................................................ 90
Digital Storage Oscilloscope VII

Chapter 1 Introduction

General Inspection
Prepare Instrumen t for Use
Brief Introduction of Front Panel
Accidence of the user inter face
Functional Check
Probe Introduction
Remote Control via USB
Digital Storage Oscilloscope 1

1.1 General Inspection

Please check the instrument as following steps after receiving an oscilloscope:
Check the shipping container for damage:
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
Check the accessories:
Accessories supplied with the instrument are listed in "Accesso r i es " in t h is manual. If the contents are incomplete or damaged, please notify the franchiser.
Check the instrument:
In case there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, please notify the franchiser.
Digital Storage Oscilloscope 2

1.2 Prepare Instrument for Use

Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation.
Connect the Power Cord
Connect the power cord as desired.
This osc illoscop e c an accept 100-240 V, 45-440 Hz AC power supply. Please use the power cord supplied with the accessories to connect the oscillo scope to the pow er source as shown in the figure below.
Turn the instrument on by pressing the power switch in the lower left corner of front panel. If the instrument does not turn on, verify that the power cord is firmly connected. Also make sure that the instrument is connected to an energized power source. Power Switch:
To turn off the instrument, please press power switch.
Digital Storage Oscilloscope 3

1.3 Brief Introduction of Front Panel

The content below simply describes and introduces the front panel and the back part of this series of digital oscilloscope so that you can get famili ar with this seri es of d igi tal oscilloscope well within the shortest time.
1. Universal Knob
2. Function Button
3. Waveform Generator
4. Horizontal System
5. Trigger System
6. Fast Action Button(Run/Stop, Single, Auto Scale)
7. Vertical System
8. External Trigger Input Channel
9. CH1~CH4 Input Channel for signal
10. Waveform Generator Output
11. Function Softkeys
12. Menu Butt on
13. Probe Compensation terminal
14. USB Port
15. Power Button
Digital Storage Oscilloscope 4

1.4 Accidence of the user interface

This section will make you understand the front operation panel of this series of digital oscilloscope at first before use.
Figure 1-2 Interface display
1. Voltcraft logo
2. Trigger Status Auto: The oscill oscope works in auto mode and is acquiring waveforms in the
absence of triggers.
Ready: All pre-triggered data have been acquired and the oscilloscope is ready to
accept a trigger.
Roll: The oscilloscope is acquiring and displaying waveform data continuously in
roll mode.
Stop: The oscilloscope has stopped acquiring waveform data.
3. Readout shows main time base setting.
4. Sample rate.
5. Main Time Base Window
Digital Storage Oscilloscope 5
CH1: to connect with the probe
PROBE COMP
6. Tigger time
7. Trigger level, Readout tells trigger level.
8. Operating Menu shows different information for respective function keys.
9. If this icon lights up/active, it means the USB disk has been connected.
10. If this icon lights up/active, it means the wave generator works
11. The inf orm ation of coupling, B andwidth and volt/ di v of CH1~CH4.
12. Channel Marker
13. Window displays waveform.

1.5 Functional Check

Follow the steps below to perform a quick functional check to your oscilloscope.

1.5.1 Connect the oscilloscope

Set the switch on the probe to 10X and connect the probe to Channel 1 on the oscilloscope. First, align the slot in the probe connector with the protuberance on the CH1 BNC and push to c onnect; then, turn to right to lock the probe in place; aft er t hat, connect the probe tip and reference lead to the PROBE COMP connectors. The r e i s a mark on the panel: Probe COMP ~2V@1KHz.

1.5.2 Observe the waveform

Press the [Auto Scale] button and y ou should see within a few seconds a square wave of about 2V peak-to-peak at 1KHz in the display. Press the [CH1] MENU button twice to remove Channel 1. Push the [CH2] MENU button and repeat the st eps to observe CH2, CH3 and CH4.
Digital Storage Oscilloscope 6

1.6 Probe Introduction

1.6.1 Safety

When using the probe, keep your fingers behind the guard on the probe body to avoid electric shock. Do not touch metallic portions of the probe head while it is connect ed to a voltage source. Connect the probe to the oscilloscope and connect the ground terminal to ground before you start any measurements.

1.6.2 Probe Check Wizard

Every time you connect a probe to an input channel, you s hould use the probe check wizard to verify that this probe is operating correctly. Use the vertical menu (for example, push the [CH1] MENU button) to set the Probe attenuation factor.

1.6.3 Manual Probe Compensation

Upon the first connection of a probe and an input channel, you should manually perform this adjustment to match the probe to the input channel. Uncompensated or
Digital Storage Oscilloscope 7
Compensated correctly
Undercompensated
miscompensated probes may lead to errors or faults i n measurement. To adjust the probe compensation, follow the steps below.
1. Set the Probe option attenuation in the channel menu to 10X. Set the switch on the
probe to 10X and connect the probe to Channel 1 on the oscilloscope. If you use the probe hook-tip, ensure it is firmly inserted onto the probe. Attach the probe tip to the PROBE COMP ~2V@1KHz connector and the reference lead to the PROBE COM P Ground connector. Display the channel and then press the Auto Scale button.
2. Check the shape of the displayed waveform.
Overcompensated
3. If necessary, use a nonmetallic screwdriver to adjust the variable capacity of your
probe until the shape of the waveform turns to be t he same as the above figure. Repeat this step as necessary. See the figure below for the way of adjustment.

1.6.4 Probe Attenuati on Setting

Probes are of various attenuation factors which affect the vertical scale of the signal. The Probe Check function is used to verify if t he P robe attenuation option matches the attenuation of the probe. You can push a vertical menu button (such as the CH1 MENU button) and select the Probe option that matches the attenuation factor of your probe. Make sure that the Attenuation switch on the probe matches the Probe opti on in the oscilloscope. Switch settings are 1X and 10X. When the Attenuation switch is set to 1X, the probe limits the bandwidth of the oscilloscope to 6MHz. To use the f ull bandwidth of the oscilloscope, be sure to set the
Digital Storage Oscilloscope 8
switch to 10X.

1.7 Remote Control via USB

1.7.1 Install Software

Caution: You must install the software before using the oscilloscope on PC.
1. While in Windows, insert the installation CD into the CD-ROM drive.
2. The installation should start up automatically. Otherwise in Windows Explorer, switch
to the CD-ROM drive and run Setup.exe. According to the installation information, please install it step by step.
Download and install IO libraries suite.
Please download latest IO Libraries Suite from the website:
http://www.keysight.com/main/software.jspx?ckey=2175637&lc=chi&cc=CN&nid =-11143.0.00&id=2175637
Also you can install the software named IOLibSuite in the CD. Double click IO Libraries Suite program and install it. According to the installation information, please install it step by step. The installation maybe takes several minutes.
Digital Storage Oscilloscope 9
After installation, you can see it in bottom right corner.

1.7.2 Connect DSO scope to PC

1. Power on. Connect DSO scope to PC via USB cable. The driver will install by itself.
When the driver is installed successfully, there will have a prompt as follows:
2. After installing successfully, users can see driver information in device manager.
3. Click Keysight IO libraries suite icon in bottom right corner, and choice “Connection
Expert”. Now users can see the device information.
Digital Storage Oscilloscope 10
4. If installing unsuccessfully, right click to update the driver manually, and choice the
correct route: C:\Program Files\IVI Foundation\VISA\IVI USB Staging\b841aa80\Windows
5. Double click “DSO-xxxx” shortcut on desktop to open the software of DSO device.
The device is connected with PC successfully. You can see “Connected” tip in status column.

1.7.3 User’s Interface of PC

Click the DSO software icon on the desk after you finished the software setting and equipment connecting. Then a user interface will be showed as follows:
In addition to displaying waveforms, the display area is filled with many details about
Digital Storage Oscilloscope 11
the waveform and the oscilloscope control settings.
1. The Main Menu
2. The Toolbar
3. Control Panel
4. The Horizontal Trigger Position
Drag Horizontal Position icon on screen to adjust the trigger position against screen.
Double click the icon to reset the trigger point back to the screen center.
5. The Trigger Level
Drag the trigger level icon on screen to adjust the trigger level position against
screen.
6. It shows the system time.
7. Connect Status
8. The markers show the CH1/CH2/CH3/CH4 reference points of the displayed
waveforms. If there is no marker, the channel is not displayed.
Digital Storage Oscilloscope 12

Chapter 2 Function Introduction

This chapter provides some general information that you need to learn before using an oscilloscope. It contains:
Menu and Control Keys
Connector
Universal Knobs and Softkeys
Oscilloscope Setup
Horizontal Controls
Vertical System
Trigger System
Save/Recall
Measure System
Acquire
Display
Utility System
Fast Action Buttons
Digital Storage Oscilloscope 13

2.1 Menu and Control Keys

As shown in the figure below:
Figure 2-1 Control keys
All the keys are described as follows:
[CH1], [CH2] , [CH3], [CH4]: display setup menus of channel 1 and channel 2.  [Math]: display “Arithmetical operation” waveform menu.  [Horizontal]: Set Horizontal system.  [Trig Menu]: display Trigger control menu.  [Force Trig]: It is used for finishing acquisition of the current waveform no matter
whether the oscilloscope detects trigger, and it is mainly applied to “Normal” and “Single” in the trigger mode.
[Default]: recall the default factory setup.  [Help]: enter the on-line help system.  [Utility]: display “UTILITY FUNCTION” menu.  [Cursors]: display the “CURSOR” menu. The [V0] knob can be used for
regulating the position of the cursor when the “CURSOR” menu is displayed and the cursor is triggered.
[Meas]: show the “Measure” menu.  [Wave Gen]: show the waveform generator menu.  [Save Recall]: show the “Save/Recall” menu of setups and waveform.  [Display]: show the “Display” menu.  [Auto Scale]: automatically set the control state of the oscilloscope so as to
display suitable waveform.
Digital Storage Oscilloscope 14
V0: menu options ( Press this knob to reset data (trigger holdoff trigger and slope trigger), select menu options and so on. Easy to operate.
Press the waveform generator function.
[Run/Stop]: continuously acquire waveform or stop acquisition  [Single]: Acquire a single trigger, finish acquisition and then stop.

2.2 Connector

Figure 2-2 Connector
CH1, CH2, CH3, CH4: for an input connector of a measured signal.  EX T TRIG: be used as an input connector of an external trigger sourc e. Use
[Trig Menu] button to select “External” trigger source (only for Edge trigger), and the trigger signal source can be used for triggering in the third channel while acquiring data.
Gen Out: Waveform signal output.  Probe compensation: T he probe compensation signal is output and grounded
so that the probe is matched with the channels of the oscilloscope.

2.3 Universal Knobs and Softkeys

Universal Knob. Under different menu options, it supports selecting
MEASURE), moving cursors and levels (Slope Trigger).
, overtime of the overtime
Wave Gen button on the front panel to open the arbitrary
Digital Storage Oscilloscope 15
of the screen and give a full screen display of waveforms. Push it again to show the menu options.
F1
functional. They are in charge of selecting corresponding menu options on the screen in different menu modes.
This functional softkey is mainly used to turn pages and confirm a
selection, such as
Hide/Show softkey. Push it to hide the menu options on the right side
-F5: These five softkeys are all multi-
‘next page’, ‘previous page’.

2.4 Oscilloscope Setup

While operating the oscilloscope, you may often use four features: Auto Scale, saving a setup, recalling a setup and default setup. Hereinafter they are introduced one by one. Auto Scale: This function can be used t o adjust the horizontal and vertical scales of the oscilloscope automatically and set the trigger coupling, type, positi on, slope, level and mode, etc., to acquire a stable waveform display. Saving a Setup: By default, the oscilloscope will save the setup each time before being closed, and automatic ally recall the setup once being turned on. (Note: If you modify the setup, please wait for more than 10 seconds before turning off the oscilloscope to ensure the proper storage of new settings.) You can save 10 settings permanently in the oscilloscope and reset them as necessary. Recalli a Setup: The oscilloscope can recall any of your saved setups or the default factory setup. Default Setup: The oscilloscope is preset for normal operations when it is shipped from the factory. This is the default setup. You may recall this setup at any time for your requirements.
Digital Storage Oscilloscope 16

2.5 Horizontal Controls

Use the horizontal controls to change the horizontal scale and position of waveforms. The horizontal position readout shows the time represented by the center of the screen, using the trigger time as zero. When you change the horizontal scale, the waveform will expand or contract to the screen center. The readout near the upper right of the screen shows the current horizontal position in second. The oscill oscope also has an arrow icon at the top of the graticule to indicate the horizontal pos iti on.
1. Horizontal Position Knob: Used to control t he trigger position against the screen center. Push this button to reset the trigger point back to the screen center. AN: Used to set the horizontal position as zero.
Universal Knob
2.SEC/DIV Knob: Used to change the horizontal time scale so as to magnify or compress the waveform horizontally. If the waveform acquisition is stopped (by using the [Run/Stop] or [Single] butt on), the SEC/DIV control will expand or compress the waveform.
Note: Press SEC/DIV Knob to enter Dual-window Mode.
Digital Storage Oscilloscope 17
Location of expanded window data in memory
Minor Window
(Expanded Window)
Single-window Mode
Dual-window Mode (Full Screen)
Major Window
Press SEC/DIV Knob again to exit Dual-window Mode.

2.6 Vertical System

2.6.1 Vertical Controls

Vertical controls can be used to display and remove waveforms, adjust vertical scale and position, set input parameters and perform math calculati ons. Each channel has a separate vertical menu to set. See below for menu description.
Digital Storage Oscilloscope 18
resolution to small steps between the Coarse
1X
Vertical Position
Volts/div
1. Vertical Position Knob: Move the channel waveform up and down on the screen. In dual-window mode, move the waveforms in both windows at the same time in a same direction. Push this knob to return waveforms to the vertical center position on the screen. Two channels correspond to two knobs. Also you can press this knob to switch between coarse and fine.
2. VOLTS/DIV Knob Control the oscilloscope to magnify or attenuate the source signal of the channel waveform. The vertical size of the display on the screen will change (increase or decrease) to the ground level.
3. Menu (CH1, CH2, CH3, CH4): Display vert ical menu options; turn on or off the display of channel waveforms.
Options Settings Comments
DC passes both DC and AC components of the input
Coupling
BW 20MHz
Div
DC AC GND
OFF ON
Coarse Fine
signal. AC blocks the DC component of the input signal and attenuates signals below 10Hz. Ground disconnects the input signal.
Limits the bandwidth to reduce display noise; filters the signal to eliminate noise and other unnecessary HF components.
Selects the resolution of the VOLTS/DIV knob. Coarse defines a 1-2-5 sequence. Fine changes the
settings.
Probe
10X 100X 1000X
Selects a value according to the probe attenuation factor so as to ensure correct vertical readouts. Reduce bandwidth to 6MHz when using a 1X probe.
Digital Storage Oscilloscope 19
The invert function turns the displayed waveform 180
trigger is also inverted.
Invert
Coupling
If the channel adopts a DC coupling mode, you can quickly measure the DC
component of the signal by observing the difference between the waveform and the signal ground.
If the channel adopts an AC coupling mode, the DC component in the signal is
filtered. By this mode, the AC component of the signal is displayed at a higher sensitivity.
If the channel adopts a GND coupling mode, cut off the input signal. Inside the
channel, the channel input is connected with a zero volt reference electric level.
Fine Resolution
In the fine resolution setting, the vertical scale readout di splays the actual VOLTS/DIV setting. The vertical scale changes only after you adjust the VOLT S/DIV control and set to coarse.
Remove Waveform Display
To remove a waveform from the screen, first push the menu button to display the vertical menu, then push again to remove the waveform. A channel waveform which is unnecessary to be displayed can be used as a trigger source or for math operations.
OFF ON
degrees, with respect to the ground level. When the oscilloscope is triggered on the inverted signal, the

2.6.2 Math Operation

The series of scope supports many math operations between analog channels waveforms, including addition (+), subtraction (-), multiplication (*), division (/) and FFT. You can use cursors to measure it. The contents of this chapter:
Units for Math Waveforms
Math Operators
To Adjust the Math Waveform Scale and Offset Note: If the analog channel or the math function display is truncated (waveforms do not display on the screen completely), the resulting math will also be truncated.
Digital Storage Oscilloscope 20
Operation Unit
Addition (+)or subtraction (-) V multiplication (*) division (/) None FFT dB, Vrms,
Addition or Subtraction
Math operators perform arithmetic operations - add or subtract operation - on any two analog input channels. When you select addition or subtraction, the Source A and Source B values are added or subtracted point by point, and the result is displayed.
1. Press the [Math] button on the front panel to enter the MATH function menu.
2. Press the Source 1 and Source 2 softkey respectively, and then turn the Universal Knob to select the source to do math operation. Analog channels (CH1~CH4) can be used as Source 1 or source 2.
3. Press the Operation softkey and then turn the universal to select + or -to make addition or subtraction operation. The resulting math waveform is displayed on the screen and labeled with “M”.
V2
Scale: Press the Scale s oftkey, and then turn the Universal Knob to select the vertical scale.
Digital Storage Oscilloscope 21
Multiplication and Division
Math operators perform arithmetic operations multiplication or divisi on operation on any two analog input channels. When you select multiplication or division, the Source 1 and Source 2 values are multiplied or divided point by point and the result is displayed.
1. Press the [Math] button on the front panel to enter the MATH function menu.
2. Press the Source 1and Source 2 softkey respectively, and then turn the Universal Knob to select the source to do math operation. Analog channels (CH1 ~CH4) can be used as Source 1 or source 2.
3. Press the Operation softkey and then turn the universal to select * or / to make multiplication or division operation. The resulting math waveform is displayed on the screen and labeled with “M”. Scale: Press the Scale s oftkey, and then turn the Universal Knob to select the vertical scale.
FFT Operation
FFT is used to compute the fast Fourier transform using analog input channels or reference waveforms. FFT takes the digitized time record of the specified source and transforms it to the frequency domain. When the FFT function is selected, the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency. The readout for the horizontal axis changes from time to frequency (Hertz) and the vertical readout changes from volts to dB. FFT operation can facilitate the following works:
Measure harmonic components and distortion in the system Measure the characteristics of the noise in DC power Analyze vibration
To display a FFT waveform:
1. Press the [Math] butt on on the front panel to open the MATH function menu.
2. Press the Operation softkey and then turn the Universal Knob to select FFT. The resulting math waveform is displayed on the screen and labeled with “M”.
3. Press the Source soft k ey, and then turn the Universal Knob to select the source to do FFT operation. Analog channels (CH1~CH4) can be used as the source.
Digital Storage Oscilloscope 22
Loading...
+ 78 hidden pages