VOLTCRAFT DSO-1022 M User guide

Version 03/08
Item No.: 12 24 94
DIGITAL STORAGE OSCILLOSCOPE DSO-1022 M
DSO-1022 M
Table of Contents
PageChapter Title
General Safety Rules Preface
Chapter 1
Chapter 2
User Guide General Check Functional Check Probe Compensation Autoset Waveform Display Getting to Know the Vertical System Getting to Know the Horizontal System Getting to Know the Trigger System Instrument Setups Setting up the Vertical System Setting up the Horizontal System Setting up the Trigger System Alternate Trigger Setting up the Sampling System Setting up the Display System Save and Recall Setting up Alternative Functions Auto Measurement Cursor Measurement
3 5 9 13 13 16 17 18 19 21 23 23 35 39 46 51 55 56 59 62 70
1
Chapter 3
Chapter 4
Chapter 5
Index
Table of Contents
Using the Run Button Practical Example Scenarios Scenario 1: Measuring simple signals Scenario 2: Observing the delay caused by a sine wave signal passes through the circuit Scenario 3: Acquiring single signal Scenario 4: Reducing random noise of signals Scenario 5: Using cursors for measurement Scenario 6: Using the X-Y function Scenario 7: Video signal triggering System Prompts and Troubleshooting Definitions of System Prompts Troubleshooting Appendixes Appendix A: Technical Indicators Appendix B: Accessories Oscilloscopes Appendix C: Maintenance and Cleaning
PageChapter Title
71 74 74
75 77 78 80 82 84 86 86 86 88 88 96
96 97
2
General Safety Rules
To avoid personal injury and damage to this product or any other connected units, please take time to read the following safety precautions. To avoid any potential danger, please use this product strictly in accordance with use instructions and safety rules.
Maintenance should be carried out only by qualified personnel.
Avoid fire and personal injury.
Use the correct power cord. Use only a desig
3
nated power cord specified for this product and certi­fied for the country of use.
Use the
plug when the probe or testing cable is connected to the power source.
Ensure the product is properly grounded. This product should be properly grounded with the earth wire of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Please ensure that the product is properly grounded before connecting any input or output termi­nal.
Connect the oscilloscope probe properly. Earth wire of the probe is in the same voltage as the earth. Do not connect the earth wire to high voltage.
Observe all terminal rated values. To avoid fire and impact caused by excessive electric current, check all rated values and labels on the product. Please read detailed information of rated values in the product manual before connecting the product.
Do not operate this product without Cover.
correct power plug. Do not remove the
When the exterior cover or front panel is open, do not operate the product.
Use a
ppropriate fuses. Use only the type of fuse
and rated indicator designated for this product.
Avoid exposing circuitry. Upon power connec­tion please do not touch any exposed adaptor or component.
Do not operate with suspected failure. If you suspect the product is damaged, have it inspected by a qualified maintenance technician.
Maintain good ventilation.
Warning: Warning statements identify conditions or
actions that could result in injury or loss of life.
Caution: Caution statements actions that could result in damage to this product or other property.
identify conditions or
“Caution” means possible damage to this product or other properties.
Symbols on the product: The following symbols may appear on the product:
Do not operate in a humid place.
Do not operate in combustible and explo sive conditions.
Keep the product surface clean and dry.
Safety Messages and Symbols Messages on the product: The following mes-
sages may appear on the product: “Danger” means potential damage that is immediate. “Warning” means potential damage that is not immedi­ate.
High voltage
Caution! Refer to manual
Protective earth terminal
Earth terminal for chassis
Earth terminal for testing
4
Preface
This manual provides information on the operation of Digital Storage Oscilloscope DSO-1022 M . Guidance is given in several chapters as follows: Chapter 1 User Guide: Simple guide to the oscilloscope functions and notes on installation.
Chapter 2 Instrument Setups: Guide to oscilloscope operation.
Chapter 3
Chapter 4 System Prompts and Trouble-shooting
Chapter 5 Servicing and Support
Chapter 6 Appendixes
Appendix A: Technical Indicators
Appendix B: Accessories for DSO-1022 M Oscilloscopes
Appendix C: Maintenance and Cleaning
Practical Example Scenarios: Example scenarios are provided to solve various testing
problems.
5
Digital Storage Oscilloscope DSO-1022 M:
Digital Storage Oscilloscope DSO-1022 M offer user-friendliness, outstanding technical indicators and a host of advanced features. They are your perfect tools to complete testing tasks swiftly and efficiently.
Introduction
Dear Customer,
Thank you very much for making the excellent decision to purchase a Voltcraft® product.You have acquired an above-average quality product from a brand family which has distinguished itself in the field of measuring,charging and network technology by particular competence and permanent innovation.
Whether your are a sophisticated do-it-yourself electronics enthusiast or a professional user, with a Voltcraft® product you always have the optimal solution at hand, even for solving the most difficult problems. Along with the remarkable feature that we offer the advanced technology and the robust quality of our Voltcraft® products at a favourable cost-performance ratio that is almost unbeatable. We are certain that using Voltcraft® will be the beginning of a long, successful relationship.
We hope you will enjoy using your new Voltcraft® product!
If you have queries about handling the device, that are not answered in this operating instruction,our technical support is available under the following address and telephone number: Voltcraft®, 92242 Hirschau, Lindenweg 15, Germany, phone 0180 / 586 582 723 8
6
DSO-1022 M oscilloscopes offer user-friendly front panel that allows access to all functions easy opera­tion. The scaling of all channels and the positions of buttons are optimally arranged for direct view opera­tion. As design is based on the mode of traditional instruments, users can use the new units without spending considerable time in learning and familiariz­ing with operation. For faster adjustment to ease testing, there is an AUTO key. The new units also feature more appropriate waveform and range scale positions. In addition to easy operation, the DSO-1022 M oscillo­scopes have all the high performance indicators and powerful functions that ensure speedy testing and measurement. With 500MS/s real-time sampling rate , these oscilloscopes can display signals much quicker, while powerful trigger and analytical features enable easy capture and analysis of waves, while a clear LCD display and mathematics functions enable the user to observe and analyse signal problems promptly and clearly. The performance features listed below will explain why the new series can fully satisfy your testing and measurement requirements:
7
Dual analog channels mono LCD display system at 320 x 240 resolution Automatic waveform and status configuration Storage of waveforms, setups recurrence Sophisticated window expansion function to analyse waveform details and overview precisely Automatic measurement of 19 waveform parameters Automatic cursor tracing measurement Unique wave recording and replay function Built-in FFT Multiple waveform mathematics functions (including add, subtract, multiply and divide) Edge, video, pulse width and alternate trigger func­tions Multilingual menu displays Help System
waveforms, setups and bit map and
DSO-1022 M Oscilloscope accessories
2 x 1.5m, 1:1/10:1 probesee passive voltage probe operating manual),comply with EN61010-031 2002 standard. Power supply line conforming to all international standards User Manual DSO-1022 M communications software (USB/RS­232C) USB Lead
8
Digital Storage Oscilloscope DSO-1022 M are small and compact benchtop oscilloscopes. The user-friendly front panel enables easy operation for basic testing and measur­ing tasks.
This chapter provides notes on the following:
General check Functional check Probe compensation Automatic settings for waveform display Getting to know the vertical system Getting to know the horizontal system Getting to know the trigger system
Chapter One User Guide
When starting to use a new oscilloscope, the first step is always to familiarize yourself with the use of the front
operation panel. This rule of thumb applies to Storage Oscilloscope DSO-1022 M
describes the operation and functions of the front panel, so you can learn how to use a Digital Storage Oscilloscope DSO-1022 M as quickly as possible.
The
Digital Storage Oscilloscope DSO-1022 M front panel with at-a-glance functions for easy opera­tion. There are buttons and function keys on the front panel. The functions of buttons are similar to other oscilloscopes. The row of 5 keys on the right of the display panel are the menu operation keys (designated as F1 to F5 from top to down). With these keys you can set up different options of the current menu. The other keys are function keys. You can use them to enter different function menus or access particular functions
. This chapter briefly
Digital
provides a
Indicative definitions in this manual: Text indications for operation keys given in this manual are identical to signs on the front panel keys. Please note that all signs for measurement function keys appear with frames, e.g. [MEASURE], to represent a front panel function key marked with the word MEASURE.
Signs for the operation keys on the menu are in shadowed text, e.g. SAVE WAVEFORM, to indicate the save wave­form option of the save menu.
Figure 1-3 interface
Schematic diagram for the display
Figure 1-3
12
General Check
We suggest checking your new scope DSO-1022 M
1. Check the unit for possible shipping damages
If the package carton or foam plastic protective lining is seriously damaged, please do not discard until you have carried out a check on the entire unit and acces­sories to ensure satisfactory electrical and mechanical performance.
2. Check the accessories
A checklist of accessories that come with your Storage Oscilloscope DSO-1022 M section Accessories for DSO-1022 M
missing items against this list.
in the following steps.
Digital Storage Oscilloscope
of this manual. Please check for any
Digital Storage Oscillo-
Digital
is provided in the
13
Functional Check
Carry out a quick functional check in the following steps to make sure your oscilloscope is operating normally.
ower on the unit
1. P
Power on the unit. Power supply voltage is 100-240V AC, 45-440Hz. After connectingto power, let the unit carry out self-calibration to optimize the oscilloscope signal path for measurement accuracy. Press the [FUNCTION] button and then [F1] to start the calibra­tion. Then press [F1] on the next page to display
2 .You have to set the probe attenuation factor of the oscillo­scope. This factor changes the vertical range multiple to ensure the measurement result correctly reflects the ampli­tude of the measured signal. Set the attenuation factor of the probe as follows: Press [F4] to display 10X on the menu.
15
Figure 1-6 Deflection factor setting of the probe on the oscilloscope
Probe ration
Figure 1-6
1 . Connect the probe tip and ground clamp to the corre­sponding probe compensation signal terminals. Press [AUTO] and you will see a square wave in the display of about 3V peak-to-peak at 1kHz in a few seconds. See Figure 1-7 for details. Repeat these steps to check CH2. Press [OFF] function button to disable CH1, then press [CH2] function button to enable CH2. Repeat steps 2 and 3.
Probe Compensation
When connecting the probe to any input channel for the first time, perform this adjustment to match the probe to the channel. Skipping the compensation calibration step will result in measurement error or fault. Please adjust probe compensation as follows:
Figure 1-7 Probe compensation signal
Figure 1-7
1.
In the probe menu set the attenuation factor to 10X. Move the switch on the probe to 10X and connect the probe to CH1. If you are using the probe hook-tip, ensure a proper and secure connection. Connect the probe tip to the probe compensator’s signal output connector, then connect the ground clamp to the earth wire of the probe compensator. Enable CH1 and press [AUTO].
Observe the shape of the displayed waveform.
2. Overcompensation Correct Compensation Undercompensation
16
Figure 1-8 Probe compensation calibration
Overcompensation Correct Compensation Undercompensation
Figure 1-8
If you see a “Undercompensation” or “Overcompensa-
3. tion” waveform
display, adjust the variable capacitor on the probe with a screwdriver with non-metal handle, until a “Correct Compensation” waveform illustrated above is displayed.
Warning: To avoid electric shock when measuring high voltage with the probe, ensure integrity of the probe’s
Autoset Waveform Display
Digital Storage Oscilloscope DSO-1022 M feature an autoset function. Your oscilloscope can automatically adjust the vertical deflection factor, scanning time base and trigger mode based on the input signal, until the most appropriate waveform is displayed. The autoset function can only be operated when the signal to be measured is 50Hz or above and the duty ratio is larger than 1%.
Using the Autoset Function:
Connect the signal to be measured to the signal input
1. channel.
2.
Press [AUT
O]. The oscilloscope will automatically set the vertical deflection factor, scanning time base and trigger mode. Should you require to make more detailed check, you can adjust manually after the autoset process until you get the optimum waveform display.
Getting to Know the Vertical System
As shown in the figure below, there are a series of buttons and knobs in the vertical control zone. The following steps will get you familiar with the use of these controls.
Figure 1-9 Vertical control zone on the front panel
1.
Turn the vertical position knob to display the signal in the centre of the window. The vertical position knob controls the vertical display position of the signal. When you turn the vertical position knob, the sign indicating Ground channel will move up and down with the waveform.
Measurement Tips
If the channel coupling is DC, you can measure the signal’s DC quickly by checking the difference between the waveform and signal ground. In the case of AC coupling, the DC within the signal will be filtered. With this coupling mode you can display the AC of the signal with higher sensitivity.
Shortcut key RETURN TO ZERO for reset­ting the ver analog channel to zero
tical position of the dual
2.
Change the vertical setups and check changes of status information. You can identify changes of any vertical range by reading the status display column at the lower corner of the wave­form window. Turn the vertical scale knob to change the vertical VOLT/DIV range. You will find that the range in the current status column has changed accordingly. Press [CH1], [CH2], [MATH] or [REF] and the screen will show the corresponding operation menu, sign, waveform and range status information. Press [OFF] to disable the selected channel.
Figure 1-10 Horizontal control zone on the front panel
Getting to Know the Horizontal System
As shown in the figure on the right hand side, there are one button and two knobs in the horizontal control zone. The following steps will get you familiar with horizontal time base setups.
19
Figure 1-10
Use the horizontal scale knob to change the
1. horizontal time base setup and check any changes in status information. Turn the horizontal scale knob to change the SEC/DIV time base range. Y find that the time base range in the current status column has changed accordingly. Range of horizontal scanning rate is 20ns~50s, in steps of 1-2-5.
ou will
2.
Use the horizontal position knob to adjust the horizontal position of the waveform window. The horizontal position knob controls trigger shift of the signal. When this function is used for trigger shift and the horizontal position knob is turned, you will find that the waveform changes with the knob.
3.
Press [MENU] to display the ZOOM menu. In this menu press [F3] to activate WINDOW EXPANSION. Then press [F1] to quit WINDOW EXPANSION and return to the MAIN TIME BASE. You can also set the HOLDOFF time with this menu.
Shortcut key for resetting the trigger point shift to horizontal zero position This shortcut key can quickly return to RETURN TO ZERO and reset the trigger point to the vertical centre point. You can also turn the horizon­tal position knob to adjust the horizontal position of the signal in the waveform window.
Definition: Trigger point means the actual trigger point relative to the centre point of the storage device. By turning the horizontal position knob, you can move the trigger point horizontally. Holdof
f means reactivating the time interval of the trigger circuit. Turn the multi-function control knob to set the holdoff time.
20
Getting to Know the Trigger System
As shown by Figure 1-11, there are one knob and three buttons in the trigger menu control zone. The following steps will get you familiar with trigger setups.
Figure 1-11 Trigger menu on the front panel
Figure 1-11
21
1.
Use the trigger level knob to change the trigger level. Y indicates the trigger level. The sign will move up and down with the knob. While you move the trigger level, you will find the trigger level value on the screen changing accordingly.
Shor level to zero position
Press 50% to quickly rest the trigger level to zero (channel vertical reference point). At trigger zero you get the highest sensitivity. Y the trigger level knob to reset the trigger point to zero.
2.
ou will see a trigger sign on the screen that
tcut key for resetting the trigger
ou can also turn
Open the [TRIGGER MENU] (see the figure below) to change trigger setups. Press [F1] and select EDGE TRIGGER Press [F2] and set TRIGGER SOURCE to CH1 Press [F3] and set EDGE TYPE as RISING Press [F4] and set TRIGGER MODE as AUTO Press [F5] and set TRIGGER COUPLING as DC
Figure 1-12 Trigger menu
Figure 1-12
3.
Press [FORCE] to generate a compulsory trigger signal that is mainly used in the normal and single trigger modes.
22
You should be familiar with basic operation of the vertical controls, horizontal controls and trigger system menu of your Digital Storage Oscilloscope DSO-1022 M by now. After reading the last chapter, you should be able to use the menus to set up your oscilloscope. If you are still unfamiliar with these basic operation and methods, please read Chapter 1.
This chapter will guide you through the following:
Setting up the vertical system ([CH1], [CH2], [MATH], [REF], [OFF], [VER [VERTICAL SCALING]) Setting up the horizontal system ([MENU], [HORIZONTAL POSITION], [HORIZONTAL SCAL­ING]) Setting up the Trigger system ([TRIGGER LEVEL], [MENU], [50%], [FORCE]) Setting up the sampling method ([ACQUIRE]) Setting up the display mode ([DISPLAY]) Save and exit ([STORAGE]) Setting up the help system ([UTILITY]) Auto measurement ([MEASURE]) Cursor measurement ([CURSOR]) Using the execution buttons ([AUTO], [START/STOP])
TICAL POSITION],
23
Chapter 2 - Instrument Setups
It is recommended that you read this chapter carefully to understand the various measurement functions and
system operation of your DSO-1022 M
Setting up the Vertical System
CH1, CH2 and setups Each channel has its own vertical menu. You should set up each item for each channel individually. Press the [CH1] or [CH2] function button and the system will display the operation menu for CH1 or CH2. For explanatory notes please see Table 2-1 below:
Digital Storage Oscilloscope
Table 2-1
Function Menu Setup Explanatory Note
Coupling
Bandwidth limit
VOLTS/DIV
Explanatory notes for channel menu
Intercept the DC
AC DC Ground
On Off
Coarse tune Fine tune
quantities of the input signal. Pass AC and DC quantities of input signal. Disconnect input signal.
Limit bandwidth to 20MHz to reduce noise display. Full bandwidth.
Coarse tune in steps of 1-2-5 to set up the deflection factor of the vertical system. Fine tune means further tuning within the coarse tune setup range to improve the vertical resolution.
Function Menu Setup Explanatory Note
Select either one value based on the probe attenuation factor to keep the vertical deflection factor reading correct. There are four values: 1X, 10X, 100X and 1000X.
Waveform invert function on. Normal waveform display.
Probe
Invert
1X 10X
On Off
24
1. Setting up channel coupling
Take an example of applying a signal to CH1. The signal being measured is a sine signal that contains DC quantities.
Press [F1] to select AC. It is now set up as AC coupling. DC quantities of the signal being measured will be intercepted. The waveform display is as follows:
Figure 2-1 DC quantities of the signal are intercepted
Figure 2-1
25
Press [F1] to select DC. Both DC and AC quantities of the signal being measured can pass through. The waveform display is as follows:
Figure 2-2 Both DC and AC quantities of the signal are displayed
Figure 2-2
Press [F1] to select GROUND. It is now set up as ground. Both DC and AC quantities contained in the signal being measured will be intercepted. The waveform display is as follows: (Note: In this mode, although waveform is not displayed, the signal remains connected to the channel circuit)
Figure 2-3 Both DC and AC quantities of the signal are intercepted
2. Setting up the channel band
Take an example of applying a signal to CH1. The signal being measured is a pulse signal that contains high frequency oscillation. Press [CH1] to turn CH1 on. Then press [F2] to set BANDWIDTH LIMIT OFF bandwidth. The signal being measured can pass through even if it contains high frequency quantities. The waveform display is as follows:
Figure 2-4 Waveform display when bandwidth limit is off
. It is now set up as full
width
Figure 2-3
Figure 2-4
26
Press [F2] to set BANDWIDTH LIMIT ON. All high frequency quantities higher than 20MHz in the signal being measured will be limited. The waveform display is as follows:
Figure 2-5 Waveform display when bandwidth limit is on
Figure 2-5
27
3. Setting up the probe r
To match the PROBE attenuation factor setup, it is necessary to set up the probe attenuation factor in the channel operation menu accordingly. For example, when the probe attenuation factor is 10:1, set the probe attenuation factor at 10X in the menu. Apply this principle to other values to ensure the voltage reading is correct.
The figure below shows the setup and vertical range display when the probe is set at 10:1:
ate
Figure 2-6 Setting up the probe attenuation factor in the channel menu
Figure 2-6
4. Vertical VOLTS/DIV adjustment setup
You can adjust the VOLTS/DIV range of the vertical deflection factor either in the coarse tune mode or fine tune mode. In COARSE TUNE mode, the VOLTS/DIV range is 2mV/div~5V/div In FINE TUNE mode, you can change the deflection factor in even smaller steps within the current vertical range, so as to continuously adjust the vertical deflection factor within the range of 2mV/div~5V/div without interruption.
. Tuning is in steps of 1-2-5.
28
Figure 2-7 Coarse tuning and fine tuning the vertical deflection factor
Figure 2-7
29
5. Waveform inversion setup
Waveform inversion: The displayed signal is inverted 180 degrees with respect to the ground level. Figure 2-8 shows the uninverted waveform. Figure 2-9 shows the inverted waveform.
Figure 2-8 Inversion setup for vertical channel inversion (uninverted)
Figure 2-8
Figure 2-9 Inversion setup for vertical channel inversion (inverted)
Figure 2-9
I. Opera
Math functions are displays of +, -, ×, and FFT mathematical results of CH1 and CH2. The menu is as follows:
Figure 2-10 Math functions
ting Math Functions
Figure 2-10
30
Table 2-2 Explanatory notes for the math menu
Function Menu Setup Explanatory Note
Type
Signal source 1
Operator
Signal source 2
Math
CH1 CH2
+
­×
CH1 CH2
To carry out +, -, ×, functions
Set signal source 1 as CH1 waveform Set signal source 1 as CH2 waveform
Signal source 1 + Signal source 2 Signal source 1 ­Signal source 2 Signal source 1 × Signal source 2 Signal source 1 ÷ Signal source 2
Set signal source 2 as CH1 waveform Set signal source 2 as CH2 waveform
31
FFT spectrum analysis By using the FFT (Fast Fourier Transform) algorithm, you can convert time domain signals (YT) into frequency domain signals. With FFT, you can conveniently observe the following types of signals: Measure the harmonic wave composition and distortion of the system Demonstrate the noise characteristics of the DC power Analyse oscillation
Table 2-3 Explanatory notes for the FFT menu
Function Menu Setup Explanatory Note
Type
Signal source
Window
FFT
CH1 CH2
Hanning Hamming Blackman Rectangle
To carry out FFT algorithm functions
Set CH1 as math waveform Set CH2 as math waveform
Set Hanning window function Set Hamming window function Set Blackman window function Set Rectangle window function
How to operate FFT functions
Signals with DC quantities or DC offset will cause error or offset of FFT waveform quantities. To reduce DC quantities, select AC coupling. To reduce random noise and frequency aliasing resulted by repeated or single pulse event, set the acquiring mode of your oscilloscope to average acquisition.
Select the FFT Window
Assuming the YT waveform is constantly repeating itself, the oscilloscope will carry out FFT conversion of time record of a limited length. When this cycle is a whole number, the YT waveform will have the same amplitude at the start and finish. There is no waveform interruption. However, if the YT waveform cycle is not a whole number, there will be different amplitudes at the start and finish, resulting in transient interruption of high frequency at the connection point. In frequency domain, this is known as leakage. To avoid leakage, multiply the original waveform by one window function to set the value at 0 for start and finish compulsively. For application of the window function, please see the table below:
32
Table 2-4
33
FFT Window Feature
The best frequency resolution, the
Rectangle
Hanning
Hamming
Blackman
worst amplitude resolution. Basi­cally similar to a status without adding window
Frequency resolution is better than the rectangle window, but amplitude resolution is poorer
Frequency resolution is marginally better than Hanning window.
The best amplitude resolution and the poorest frequency resolution.
.
Most Suitable Measurement Item
emporary or fast pulse. Signal level is
T generally the same before and after. Equal sine wave of very similar frequency. There is broad-band random noise with slow moving wave spectrum.
Sine, cyclical and narrow-band random noise.
emporary or fast pulse. Signal level
T varies greatly before and after.
Mainly for single-frequency signals to search for higher-order harmonic wave.
Definition FFT resolution means the quotient of the sam-
pling and math points. When math point value is fixed, the sampling rate should be as low as possible relative to the FFT resolution. Nyquist frequency: To rebuild the original waveform, at least 2f sampling rate should be used for waveform with a maximum frequency of f. This is known as Nyquist stability criterion, where f is the Nyquist frequency and 2f is the Nyquist sampling rate.
II. Reference Waveform
Display of the saved reference waveforms can be set on or off in the [REF] menu. The waveforms are saved in the non volatile memory of the oscilloscope and identified with the following names: RefA, RefB. To display (recall) or hide the reference waveforms, take the following steps:
Press the [REF] menu button on the front panel.
1. Press RefA (RefA
2.
signal source and then select the position of the
reference option). Select the
signal source by turning the multi-function control knob on the upper part of the front panel. You can choose from 1 to 10. After selecting a numeral for saved wave­form, e.g. form which was originally stored in that position. After displaying the waveform, press the CANCEL button [F5] to go back to the previous menu.
3.
Press RefB (RefA reference option). Select the second signal source for the math function by repeating step 2. In actual application, when using your Digital Storage Oscilloscope DSO-1022 M to measure and observe such waveforms, you can compare the current waveform with the reference waveform for analysis. Press [REF] to display the reference waveform menu. Setup is as follows:
1, press the recall button to display the wave-
34
Table 2-5 Selecting the storage position
Function Menu Setup Explanatory Note
1~10 stand for
Signal source selection
Disk
Close
Recall
1~10
DSO
--
--
positions of 10 groups of waveforms respec­tively
Select an internal storage position
Close the recalled waveform
Recall the selected waveform
T
o select an internal storage position, choose between 1 and 10. To save a waveform, see the [ST
Setting up the Horizontal System
Horizontal Control Knob
You can use the horizontal control knob to change the horizontal graticule (time base) and trigger the horizontal position of the memory (triggering position). The vertical centre point above the horizontal orienta­tion of the screen is the time reference point of the waveform. Changing the horizontal graticule will cause the waveform to increase or decrease in size relative to the screen centre. When the horizontal position changes, the position with respect to the waveform triggering point is also changed.
Horizontal position: Adjust the horizontal positions of channel waveforms (including math waveforms).
Resolution of this control button changes with thetime base.
ORAGE] menu.
35
Cancel
--
Go back to the previous menu
Horizontal scaling: Adjust the main time base, i.e sec/div. When time base extension is on, you can use
the horizontal scaling knob to change the delay scanning time base and change the window width. For details see notes on time base extension.
Horizontal control knob menu: Display the horizontal menu (see the table below).
Table 2-6
Function Menu Setup Explanatory Note
1. Enable main time base
2. If window extension is
Main time base
--
--
enabled, press main time base to disable window extension
Figure 2-11 Horizontal system interface
Figure 2-1
1
Window extension
--
f
Holdof
--
Enable time base
Adjust holdof
f time
36
Icon definitions
represents the memory position of the current waveform. represents the memory position of the triggering point. represents the position of the triggering point in the current waveform window. horizontal time base (main time base), i.e sec/div. horizontal distance between the triggering position and the window centre point.
37
Definitions Y-T Mode: In this mode the Y axis indicates
voltage and the X axis indicates time. X-Y Mode: In this mode the X axis indicates CH1 voltage and the Y axis indicates CH2 voltage. Slow Scanning Mode: When horizontal time base control is set at 50ms/div or slower, the unit will operate in the slow scan sampling mode. When observing low frequency signals in slow scanning mode, it is advised to set the channel coupling as DC. Sec/Div: A horizontal scaling (time base) unit. If waveform sampling is stopped (by pressing the [RUN/STOP] button), time base control can expand or compress the waveform.
Window Extension
Window extension can be used to zoom in or zoom out a band of waveform to check image details. The window extension setting must not be slower than that of the main time base.
Figure 2-12 Display with the window extended
In divided into two zones as shown above. The upper part displays the original waveform. You can move this zone left and right by turning the horizontal POSITION knob, or increase and decrease the selected zone in size by turning the horizontal SCALE knob.
time base extension mode, the display is
the
Figure 2-12
The lower of the selected original waveform zone. Please note that the resolution of extended time base relative to the main time base is now higher (as shown in the above figure). Since the waveform shown in the entire lower part corresponds to the selected zone in the upper part, you can increase the extended time base by turning the horizontal SCALE knob to decrease the size of the selected zone. In other words, you can increase the multiple of waveform extension.
X-Y Mode
This mode is suitable for CH1 and CH2 only. After selecting the X-Y will display CH1 voltage, while the vertical axis will display CH2 voltage.
part is the horizontally extended waveform
display mode, the horizontal axis
38
Figure 2-13 Waveform display in X-Y
Figure 2-13
39
mode
Caution: In the normal X-Y mode, the oscilloscope can use the random sampling rate to acquire wave­forms. To adjust sampling rate and channel vertical range in the X-Y mode, the omitted sampling rate is 100MS/s. Generally,through adjusting time base range, lower the sampling rate appropriately will result in lissajous figures of better display quality. The following functions have no effect in the X-Y display mode: Auto measurement mode Cursor measurement mode Reference or math waveform Vector display type Horizontal position knob Trigger control
Setting up the Trigger System
Triggering decides when the oscilloscope collects data and display waveforms. Once the trigger is correctly set up, it can convert unstable display into significant waveforms. When beginning to collect data, the oscilloscope first collects sufficient data to draw a waveform on the left of the triggering point. While waiting for the triggering condition to occur, it will
continuously collect data. When trigger is detected, the oscilloscope will continuously collect sufficient data to draw a waveform on the right of the triggering point. The trigger control zone on the operation panel of your oscilloscope comprises a trigger level adjust­ment knob, a trigger menu button [MENU], [50%] for setting up the trigger level at the vertical centre point of the signal, and a compulsory trigger button [FORCE]
Pulse Trigger: When the pulse width of the trigger signal reaches a preset trigger condition, trigger occurs.
Video T
standard video signals. Alternate Trigger: Applicable to triggering signals without frequency coherence.
Below are notes for various trigger menus.
rigger: Carry out field or line trigger to
Trigger level: Trigger level sets the signal voltage with respect to the triggering point. [50%]: Setting the trigger level at the vertical centre point of the trigger signal amplitude. [FORCE]: To generate a compulsory trigger signal. Mainly used in the trigger mode and “Normal” and “Single” modes. [MENU]: Button for the trigger setup menu
ger Control
Trig
Trigger modes: edge, pulse, video and alternate Edge Trigger: When the edge of the trigger signal reaches a given level, trigger occurs.
Edge Trigger
Edge trigger means triggering at the trigger thresh­old. When selecting “edge trigger”, you are triggering at the rising and falling edges of the input signal.
40
Table 2-8
Function Menu Setup Explanatory Note
Type Edge
Set CH1 as the signal source trigger signal Set CH2 as the signal source trigger signal Set the external trigger input channel as the signal source trigger signal Set the external trigger source divided by 5 to extend the external trigger level range
Signal source selection
CH1 CH2 EXT
EXT/5
41
Inclination
rigger mode
T
T
rigger coupling
AC Line Alternate
Rising Falling
Auto Normal Single
DC AC H/F Reject L/F Reject
Set up as AC Line trigger CH1, CH2 trigger their own signals alternately
Set to trigger on the signal’ Set to trigger on the signal’
Set to sample waveform only if no trigger condition is detected Set to sample waveform only if trigger condition is satisfied Set to sample waveform once when detecting one trigger and then stop
Intercept DC quantities of the input signal Allow AC and DC quantities of the input signal to pass Reject high frequency quantities above 80kHz of the signal Reject low frequency quantities below 80kHz of the signal
s rising edge s falling edge
Pulse Trigger
Pulse trigger means determining the triggering time based on the pulse width. You can acquire abnor­mal pulse by setting the pulse width condition.
Table 2-9 (page 1)
Function Menu Setup Explanatory Note
Type
Trigger source
Pulse width condition
Setting
Next ½
Pulse
CH1 CH2 EXT EXT/5
AC Line Alternate
Larger Smaller Equal
20ns - 10s
--
Set CH1 as the signal source trigger signal Set CH2 as the signal source trigger signal Set the external trigger input channel as the signal source trigger signal Set the external trigger source divided by 5 to extend the external trigger level range Set up as AC Line trigger CH1, CH2 trigger their own signals alternately
Trigger when pulse width is larger than default value
rigger when pulse width is smaller than default value
T Trigger when pulse width equals to default value
Set the pulse width at 20ns~10s and adjust by turning the control knob on the upper front panel
Go to next page
42
Table 2-10 (page 2)
Function Menu Setup Explanatory Note
43
Trigger polarity
rigger mode
T
Previous 2/2
Positive pulse width Negative pulse width
Auto
Normal
Single
--
Set positive pulse width as the trigger signal Set negative pulse width as the trigger signal
The system automatically samples waveform data when there is no trigger signal input. The scan baseline is shown on the display. When the trigger signal is generated, it automatically turns to trigger scan. The system stops acquiring data when there is no trigger signal. When the trigger signal is generated, trigger scan occurs. One trigger will occur when there is an input trigger signal. Then trigger will stop.
Go to previous page
Video Trigger
By selecting video trigger, you can carry out field or line trigger with NTSC or PAL standard video signals. Default trigger coupling is DC. Trigger menus are as follows:
Table 2-11 Video trigger setup
Function Menu Setup Explanatory Note
Type Pulse
Set CH1 as the trigger signal Set CH2 as the trigger signal Set the external trigger input channel as the trigger signal Attenuate the external trigger source 5 times as the trigger signal Set ALL Line as the trigger signal Set CH1 and CH2 as alternate trigger signals
Suitable for video signals of low black level Suitable for video signals of high black level
Set the TV line to synchronize with trigger Set synchronized trigger on the specified line and adjust by turning the control knob on the upper front panel Set the video odd field to synchronized trigger Set the video even field to synchronized trigger
Trigger source
Standard
Synchronization
CH1 CH2 EXT EXT/5 AC Line Alternate
PAL NTSC
All lines Specified lines Odd field Even field
44
When PAL is selected for STANDARD format and SYNCHRONIZATION mode is either All Line or Specified Line, you will see a screen display as shown in Figure 2-14. When SYNCHRONIZATION mode is either Odd Field or Even Field, you will see a screen display as shown in Figure 2-15. Figure 2-14 Video trigger: line synchronization
Figure 2-14
45
Figure 2-15 V
ideo trigger: field synchronization
Figure 2-15
Alternate Trigger
When alternate trigger is selected, the trigger signal will be present in two vertical channels. This triggering mode is suitable for observing two signals of unre­lated signal frequencies. The figure below shows the alternate trigger waveform. Trigger menus are listed in Table 2-12.
Figure 2-16 Observing two signals of different frequencies in the alternate trigger mode
able 2-12 Alternate trigger setup
T
Function Menu Setup Explanatory Note
Type
T
rigger source
Edge
Alternate
Set trigger mode to edge
Set CH1 and CH2 to alternate trigger
Figure 2-16
Inclination
Trigger mode
T
rigger coupling
Rising
Auto
AC
Set trigger inclination as rising edge
Set trigger mode to automatic
Set trigger coupling mode to AC
46
Setup for Trigger coupling mode
Enter the trigger setup menu to set up the trigger coupling mode and achieve the most stable synchro­nization. The trigger coupling menus are as follows:
Function Menu Setup Explanatory Note
Type
Edge
Set CH1 and CH2 to
Trigger source
Inclination
Alternate
Rising
alternate trigger
Set trigger inclination as rising edge
Adjusting the Holdoff Time
You can adjust the holdoff time to observe compli­cated waveforms (e.g. pulse string series). Holdoff time means the waiting time for the trigger to be ready for use again. During this time the oscilloscope will not trigger until the holdoff is complete. For example, if you wish to trigger one group of pulse series at the first pulse, set the holdoff time to the pulse string width as shown in Figure 2-18. For holdoff menus please see the table below:
Trigger mode
Coupling
47
Auto
DC AC H/F Reject L/F Reject
Set trigger mode to automatic
Intercept DC quantities Allow all quantities to pass Intercept high frequency quantities of the signal, only allow low frequency quantities to pass Intercept low frequency quantities of the signal, only allow high frequency quantities to pass
Table 2-15
Function Menu Setup Explanatory Note
1. Enable main time base
2. If window exten-
Main time base
--
sion is enabled, press main time base to disable window extension
Figure 2-17 Use the holdoff function to synchronize complicated signals
Window exten sion
--
Holdoff
-
--
Enable time base extension
Adjust holdof
f time
Figure 2-17
48
Operation
Follow the normal signal synchronization procedure
1. and select the edge, trigger source and inclination in the trigger [MENU]. Adjust the trigger level to make the waveform display as stable as possible. Press the key in the horizontal [MENU] to display
2. the horizontal menu. Adjust the multi-function control knob in the upper
3. front panel. The holdoff time will change accordingly until the waveform display is stable.
Definitions
1.Trigger source: Trigger can be obtained from
various sources: input channel (CH1, CH2), external trigger (EXT, EXT/5), grid. Input Channel: The most common trigger source is input channel (choose either one). The selected trigger source can operate normally whether the input is displayed or not. External Trigger: This type of trigger source can trigger in a third channel while acquiring data in two other channels. For example, you can use an external clock or the signal from a circuit to be measured
49
as the trigger source. Both EXT and EXT/5 trigger sources use external trigger signals from the EXT the signals directly. You can use EXT within the trigger level range of –1.6V and +1.6V. EXT/5 divide the trigger by 5. As a result, trigger range is extended to –8V to +8V, enabling the oscilloscope to trigger at a large signal. AC Line: It means the AC Line power source. This trigger mode is suitable for observing signals related to the AC Line
- e.g. the correlation between lighting equipment and power source equipment - to achieve stable synchro­nization.
2.Trigger mode: Determine the action of your oscilloscope when there is no trigger. This oscillo­scope offers three trigger modes for selection: auto, normal and single.
Auto Trigger: The system will sample waveform data when there is no trigger signal input. The scan baseline is shown on the display. When the trigger signal is generated, it automatically turns to trigger scan for signal synchronization.
TRIG jack. EXT can use
Note: When time base of the scan waveform is set to 50ms/div or slower, the Auto mode allows no trigger signal. Normal Trigger: In this mode, your oscillo­scope samples waveforms only when triggering conditions are met. The system stops acquiring data and waits when there is no trigger signal. When the trigger signal is generated, trigger scan occurs. Single Trigger: In this mode, you only have to press the [RUN] button once and the oscilloscope will wait for trigger. When the oscilloscope detects one trigger, it will sample and display the acquired waveform, then stop.
3.Trigger coupling: Trigger coupling determines which quantities of the signal are transmitted to the trigger circuit. Coupling modes are DC, AC, low frequency reject and high frequency reject. DC: Allowing all quantities to pass. AC: Intercepting DC quantities and attenuating signals under 10Hz. Low Frequency Reject: Intercepting DC quantities and attenuating low frequency quantities under 80kHz.
High Frequency Reject: Attenuating high frequency quantities over 80kHz.
4.Pretrig
before/after triggering. The trigger position is typically set at the horizontal center of the screen. In this case, you are able to view five divisions of pretrigger and delayed trigger information. Use the horizontal position button to adjust the horizontal shift of the waveform to see more pretrigger information. By observing pretrigger data, you can see the waveform before trigger occurs. For example, you can detect the glitch that occurs when the circuitry starts. Observation and analysis of trigger data can help you identify the cause of glitch.
ger/Delayed Trigger: Data sampled
50
Setting up the Sampling System
As shown below, [ACQUIRE] button in the control zone is the function key for the sampling system.
Figure 2-18 Function key for the sampling system.
Figure 2-18
51
Press the [ACQUIRE] button to pop out the sampling setup menu. You can use this menu to adjust the sampling mode. Table 2-16 Sampling menu
Function Menu Setup Explanatory Note
Turn on the ordinary sampling mode Turn on the peak detect mode Set to average sampling and display the average number of times
Set the average number of times in multiples of 2, i.e. 2, 4, 8, 16, 32, 64, 128, 256. To change the average number of times, use the multi-function control knob on the left of figure 2-18.
Acquisition mode
A
verage number of times
--
--
--
Sample Peak detect Average
2~256
By changing acquisition setup, you can observe the consequent changes in waveform display. If the signal contains considerable noise, you will see the following waveform displays when average sampling is not selected and when 32-time average sampling is selected, see figure 2-19 and 220:
Figure 2-19 Waveform without average sampling
Figure 2-19
Figure 2-20 Waveform when 32-time average sampling is selected
Figure 2-20
52
Note:
1.To avoid mixed envelop, select Peak Detect.
2.To reduce random noise of the displayed signal, select average sampling and increase the average number of times in multiples of 2, i.e selecting from 2 to 256.
53
Definitions: Real time sampling: In this mode, the system makes full acquisition to fill the memory. Maximum sampling
rate is 250MS/s. At a setting of 50ns or faster, the oscilloscope will automatically carry out interpolation, i.e. inserting a sampling point between other sampling points. Peak detect mode: In this mode, the oscilloscope identifies the maximum and minimum values of the input signals at each sampling interval and use these values to display the waveform. In effect, the oscilloscope can acquire and display narrow pulse which would otherwise be omitted in the sampling mode. Noise seems to be more significant in this mode. Average mode: The oscilloscope acquires several waveforms and take the average value to display the final waveform. You can use this mode to reduce random noise.
54
Setting up the Display System
As shown below, the [DISPLAY] button in the control zone is the function key for the display system.
Figure 2-21 Function key for the sampling system (display).
Figure 2-21
Press the [DISPLAY] button to pop out the setup menu shown below the display mode.
. You can use this menu to adjust
Table 2-17 Display menu
Function Menu Setup Explanatory Note
Sampling points are
Display type
Format
Persist
Vector dots
YT XY
Off
Infinite
linked for display Sampling points are directly displayed
Operating mode of the oscilloscope X-Y is the display mode; CH1 is X input, CH2 is Y input
The waveform on the screen is refreshed at higher speed The original waveform on the screen remains on display. New data will be added continu ously until this function is disabled
-
55
Contrast
+, -
Setting the waveform contrast
(STORAGE).
[STORAGE]
[STORAGE]
[STORAGE]
Length
French
German
Dutch
Digital Storage Oscilloscope DSO-1022 M
Digital Storage
Oscilloscope DSO-1022 M
Digital Storage Oscilloscope DSO-1022 M
Digital Storage Oscilloscope DSO-1022 M
Digital Storage Oscilloscope DSO­1022 M
250MS/s
≤14ns
DSO-1022 M
Digital Storage Oscilloscope DSO-1022 M
DSO-1022 M
USB Lead
RS-232 Cable
97
VOLTCRAFT IM INTERNET http://www.voltcraft.de
Imprint
These operating instructions are published by Voltcraft®, Lindenweg 15, D-92242 Hirschau/ Germany, Phone +49 180 586 582 723 8. No reproduction (including translation) is permitted in whole or part e.g. photocopy, microfilming or storage in electronic data processing equipment, without the express written consent of the publisher. The operating instructions reflect the current technical specifications at time of print. We reserve the right to change the technical or physical specifications.
© Copyright 2008 by Voltcraft
®
.
*02_03/08_01-SP
Loading...