User Guide
General Check
Functional Check
Probe Compensation
Autoset Waveform Display
Getting to Know the Vertical System
Getting to Know the Horizontal System
Getting to Know the Trigger System
Instrument Setups
Setting up the Vertical System
Setting up the Horizontal System
Setting up the Trigger System
Alternate Trigger
Setting up the Sampling System
Setting up the Display System
Save and Recall
Setting up Alternative Functions
Auto Measurement
Cursor Measurement
Using the Run Button
Practical Example Scenarios
Scenario 1: Measuring simple signals
Scenario 2: Observing the delay caused by a sine wave
signal passes through the circuit
Scenario 3: Acquiring single signal
Scenario 4: Reducing random noise of signals
Scenario 5: Using cursors for measurement
Scenario 6: Using the X-Y function
Scenario 7: Video signal triggering
System Prompts and Troubleshooting
Definitions of System Prompts
Troubleshooting
Appendixes
Appendix A: Technical Indicators
Appendix B: Accessories
Oscilloscopes
Appendix C: Maintenance and Cleaning
PageChapter Title
71
74
74
75
77
78
80
82
84
86
86
86
88
88
96
96
97
2
General Safety Rules
To avoid personal injury and damage to this product or
any other connected units, please take time to read the
following safety precautions. To avoid any potential
danger, please use this product strictly in accordance
with use instructions and safety rules.
Maintenance should be carried out only by
qualified personnel.
Avoid fire and personal injury.
Use the correct power cord. Use only a desig
3
nated power cord specified for this product and certified for the country of use.
Use the
plug when the probe or testing cable is connected to
the power source.
Ensure the product is properly grounded.
This product should be properly grounded with the
earth wire of the power cord. To avoid electric shock,
the grounding conductor must be connected to earth
ground. Please ensure that the product is properly
grounded before connecting any input or output terminal.
Connect the oscilloscope probe properly.
Earth wire of the probe is in the same voltage as the
earth. Do not connect the earth wire to high voltage.
Observe all terminal rated values. To avoid fire
and impact caused by excessive electric current,
check all rated values and labels on the product.
Please read detailed information of rated values in the
product manual before connecting the product.
Do not operate this product without Cover.
correct power plug. Do not remove the
When the exterior cover or front panel is open, do not
operate the product.
Use a
ppropriate fuses. Use only the type of fuse
and rated indicator designated for this product.
Avoid exposing circuitry. Upon power connection please do not touch any exposed adaptor or
component.
Do not operate with suspected failure. If you
suspect the product is damaged, have it inspected by
a qualified maintenance technician.
Maintain good ventilation.
Warning: Warning statements identify conditions or
actions that could result in injury or loss of life.
Caution: Caution statements
actions that could result in damage to this product or
other property.
identify conditions or
“Caution” means possible damage to this product or
other properties.
Symbols on the product: The following symbols
may appear on the product:
Do not operate in a humid place.
Do not operate in combustible and explo
sive conditions.
Keep the product surface clean and dry.
Safety Messages and Symbols
Messages on the product: The following mes-
sages may appear on the product:
“Danger” means potential damage that is immediate.
“Warning” means potential damage that is not immediate.
High voltage
Caution! Refer to manual
Protective earth terminal
Earth terminal for chassis
Earth terminal for testing
4
Preface
This manual provides information on the operation of Digital Storage Oscilloscope DSO-1022 M
. Guidance is given in several chapters as follows:
Chapter 1 User Guide: Simple guide to the oscilloscope functions and notes on installation.
Chapter 2 Instrument Setups: Guide to oscilloscope operation.
Chapter 3
Chapter 4 System Prompts and Trouble-shooting
Chapter 5 Servicing and Support
Chapter 6 Appendixes
Appendix A: Technical Indicators
Appendix B: Accessories for DSO-1022 M Oscilloscopes
Appendix C: Maintenance and Cleaning
Practical Example Scenarios: Example scenarios are provided to solve various testing
problems.
5
Digital Storage Oscilloscope DSO-1022 M:
Digital Storage Oscilloscope DSO-1022 M offer user-friendliness, outstanding technical indicators and a host of
advanced features. They are your perfect tools to complete testing tasks swiftly and efficiently.
Introduction
Dear Customer,
Thank you very much for making the excellent decision to purchase a Voltcraft® product.You have acquired an
above-average quality product from a brand family which has distinguished itself in the field of measuring,charging
and network technology by particular competence and permanent innovation.
Whether your are a sophisticated do-it-yourself electronics enthusiast or a professional user, with a Voltcraft®
product you always have the optimal solution at hand, even for solving the most difficult problems. Along with the
remarkable feature that we offer the advanced technology and the robust quality of our Voltcraft® products at a
favourable cost-performance ratio that is almost unbeatable. We are certain that using Voltcraft® will be the
beginning of a long, successful relationship.
We hope you will enjoy using your new Voltcraft® product!
If you have queries about handling the device, that are not answered in this operating instruction,our technical
support is available under the following address and telephone number:
Voltcraft®, 92242 Hirschau, Lindenweg 15, Germany, phone 0180 / 586 582 723 8
6
DSO-1022 M oscilloscopes offer user-friendly front
panel that allows access to all functions easy operation. The scaling of all channels and the positions of
buttons are optimally arranged for direct view operation. As design is based on the mode of traditional
instruments, users can use the new units without
spending considerable time in learning and familiarizing with operation. For faster adjustment to ease
testing, there is an AUTO key. The new units also
feature more appropriate waveform and range scale
positions.
In addition to easy operation, the DSO-1022 M oscilloscopes have all the high performance indicators and
powerful functions that ensure speedy testing and
measurement. With 500MS/s real-time sampling rate ,
these oscilloscopes can display signals much quicker,
while powerful trigger and analytical features enable
easy capture and analysis of waves, while a clear LCD
display and mathematics functions enable the user to
observe and analyse signal problems promptly and
clearly.
The performance features listed below will explain
why the new series can fully satisfy your testing and
measurement requirements:
7
Dual analog channels
mono LCD display system at 320 x 240
resolution
Automatic waveform and status configuration
Storage of
waveforms, setups recurrence
Sophisticated window expansion function to analyse
waveform details and overview precisely
Automatic measurement of 19 waveform parameters
Automatic cursor tracing measurement
Unique wave recording and replay function
Built-in FFT
Multiple waveform mathematics functions (including
add, subtract, multiply and divide)
Edge, video, pulse width and alternate trigger functions
Multilingual menu displays
Help System
waveforms, setups and bit map and
DSO-1022 M Oscilloscope accessories
2 x 1.5m, 1:1/10:1 probe(see passive voltage probe
operating manual),comply with EN61010-031:
2002 standard.
Power supply line conforming to all international
standards
User Manual
DSO-1022 M communications software (USB/RS232C)
USB Lead
8
Digital Storage Oscilloscope DSO-1022 M are small and
compact benchtop oscilloscopes. The user-friendly front
panel enables easy operation for basic testing and measuring tasks.
This chapter provides notes on the following:
General check
Functional check
Probe compensation
Automatic settings for waveform display
Getting to know the vertical system
Getting to know the horizontal system
Getting to know the trigger system
Chapter One User Guide
When starting to use a new oscilloscope, the first step
is always to familiarize yourself with the use of the front
operation panel. This rule of thumb applies to
Storage Oscilloscope DSO-1022 M
describes the operation and functions of the front
panel, so you can learn how to use a Digital Storage
Oscilloscope DSO-1022 M as quickly as possible.
The
Digital Storage Oscilloscope DSO-1022 M
front panel with at-a-glance functions for easy operation. There are buttons and function keys on the front
panel. The functions of buttons are similar to other
oscilloscopes. The row of 5 keys on the right of the
display panel are the menu operation keys (designated
as F1 to F5 from top to down). With these keys you can
set up different options of the current menu. The other
keys are function keys. You can use them to enter
different function menus or access particular functions
. This chapter briefly
Digital
provides a
Indicative definitions in this manual:
Text indications for operation keys given in this manual are
identical to signs on the front panel keys.
Please note that all signs for measurement function keys
appear with frames, e.g. [MEASURE], to represent a front
panel function key marked with the word MEASURE.
Signs for the operation keys on the menu are in shadowed
text, e.g. SAVE WAVEFORM, to indicate the save waveform option of the save menu.
Figure 1-3
interface
Schematic diagram for the display
Figure 1-3
12
General Check
We suggest checking your new
scope DSO-1022 M
1. Check the unit for possible shipping
damages
If the package carton or foam plastic protective lining is
seriously damaged, please do not discard until you
have carried out a check on the entire unit and accessories to ensure satisfactory electrical and mechanical
performance.
2. Check the accessories
A checklist of accessories that come with your
Storage Oscilloscope DSO-1022 M
section Accessories for
DSO-1022 M
missing items against this list.
in the following steps.
Digital Storage Oscilloscope
of this manual. Please check for any
Digital Storage Oscillo-
Digital
is provided in the
13
Functional Check
Carry out a quick functional check in the following
steps to make sure your oscilloscope is operating
normally.
ower on the unit
1. P
Power on the unit. Power supply voltage is 100-240V
AC, 45-440Hz. After connectingto power, let the unit
carry out self-calibration to optimize the oscilloscope
signal path for measurement accuracy. Press the
[FUNCTION] button and then [F1] to start the calibration. Then press [F1] on the next page to display
2 .You have to set the probe attenuation factor of the oscilloscope. This factor changes the vertical range multiple to
ensure the measurement result correctly reflects the amplitude of the measured signal. Set the attenuation factor of the
probe as follows: Press [F4] to display 10X on the menu.
15
Figure 1-6 Deflection factor setting of
the probe on the oscilloscope
Probe ration
Figure 1-6
1 . Connect the probe tip and ground clamp to the corresponding probe compensation signal terminals. Press
[AUTO] and you will see a square wave in the display of
about 3V peak-to-peak at 1kHz in a few seconds. See
Figure 1-7 for details. Repeat these steps to check CH2.
Press [OFF] function button to disable CH1, then press
[CH2] function button to enable CH2. Repeat steps 2 and 3.
Probe Compensation
When connecting the probe to any input channel for the first
time, perform this adjustment to match the probe to the
channel. Skipping the compensation calibration step will
result in measurement error or fault. Please adjust probe
compensation as follows:
Figure 1-7 Probe compensation signal
Figure 1-7
1.
In the probe menu set the attenuation factor to 10X.
Move the switch on the probe to 10X and connect the
probe to CH1. If you are using the probe hook-tip, ensure
a proper and secure connection. Connect the probe tip to
the probe compensator’s signal output connector, then
connect the ground clamp to the earth wire of the probe
compensator. Enable CH1 and press [AUTO].
If you see a “Undercompensation” or “Overcompensa-
3.
tion” waveform
display, adjust the variable capacitor on
the probe with a screwdriver with non-metal handle, until
a “Correct Compensation” waveform illustrated above is
displayed.
Warning: To avoid electric shock when measuring high
voltage with the probe, ensure integrity of the probe’s
Autoset Waveform Display
Digital Storage Oscilloscope DSO-1022 M feature an
autoset function. Your oscilloscope can automatically adjust
the vertical deflection factor, scanning time base and trigger
mode based on the input signal, until the most appropriate
waveform is displayed. The autoset function can only be
operated when the signal to be measured is 50Hz or above
and the duty ratio is larger than 1%.
Using the Autoset Function:
Connect the signal to be measured to the signal input
1.
channel.
2.
Press [AUT
O]. The oscilloscope will automatically set the
vertical deflection factor, scanning time base and trigger
mode. Should you require to make more detailed check,
you can adjust manually after the autoset process until you
get the optimum waveform display.
Getting to Know the Vertical System
As shown in the figure below, there are a series of buttons
and knobs in the vertical control zone. The following steps
will get you familiar with the use of these controls.
Figure 1-9 Vertical control zone on the front panel
1.
Turn the vertical position knob to display the signal in the
centre of the window. The vertical position knob controls
the vertical display position of the signal. When you turn
the vertical position knob, the sign indicating Ground
channel will move up and down with the waveform.
Measurement Tips
If the channel coupling is DC, you can measure the
signal’s DC quickly by checking the difference between
the waveform and signal ground.
In the case of AC coupling, the DC within the signal will
be filtered. With this coupling mode you can display the
AC of the signal with higher sensitivity.
Shortcut key RETURN TO ZERO for resetting the ver
analog channel to zero
tical position of the dual
2.
Change the vertical setups and check changes of status
information.
You can identify changes of any vertical range by reading
the status display column at the lower corner of the waveform window. Turn the vertical scale knob to change the
vertical VOLT/DIV range. You will find that the range in the
current status column has changed accordingly. Press
[CH1], [CH2], [MATH] or [REF] and the screen will show
the corresponding operation menu, sign, waveform and
range status information. Press [OFF] to disable the
selected channel.
Figure 1-10 Horizontal
control zone on the
front panel
Getting to Know the Horizontal System
As shown in the figure on the right hand side, there are
one button and two knobs in the horizontal control
zone. The following steps will get you familiar with
horizontal time base setups.
19
Figure 1-10
Use the horizontal scale knob to change the
1.
horizontal time base setup and check any changes
in status information. Turn the horizontal scale knob
to change the SEC/DIV time base range. Y
find that the time base range in the current status
column has changed accordingly. Range of
horizontal scanning rate is 20ns~50s, in steps of
1-2-5.
ou will
2.
Use the horizontal position knob to adjust the
horizontal position of the waveform window. The
horizontal position knob controls trigger shift of the
signal. When this function is used for trigger shift
and the horizontal position knob is turned, you will
find that the waveform changes with the knob.
3.
Press [MENU] to display the ZOOM menu. In this
menu press [F3] to activate WINDOW EXPANSION.
Then press [F1] to quit WINDOW EXPANSION and
return to the MAIN TIME BASE. You can also set the
HOLDOFF time with this menu.
Shortcut key for resetting the trigger point shift to
horizontal zero position
This shortcut key can quickly return to RETURN TO ZERO and reset the trigger point to the
vertical centre point. You can also turn the horizontal position knob to adjust the horizontal position of
the signal in the waveform window.
Definition:
Trigger point means the actual trigger point
relative to the centre point of the storage device.
By turning the horizontal position knob, you can
move the trigger point horizontally.
Holdof
f means reactivating the time interval of the
trigger circuit. Turn the multi-function control knob
to set the holdoff time.
20
Getting to Know the Trigger System
As shown by Figure 1-11, there are one knob and
three buttons in the trigger menu control zone. The
following steps will get you familiar with trigger
setups.
Figure 1-11 Trigger menu on the front panel
Figure 1-11
21
1.
Use the trigger level knob to change the trigger
level. Y
indicates the trigger level. The sign will move up
and down with the knob. While you move the
trigger level, you will find the trigger level value on
the screen changing accordingly.
Shor
level to zero position
Press 50% to quickly rest the trigger level to zero
(channel vertical reference point). At trigger zero
you get the highest sensitivity. Y
the trigger level knob to reset the trigger point to
zero.
2.
ou will see a trigger sign on the screen that
tcut key for resetting the trigger
ou can also turn
Open the [TRIGGER MENU] (see the figure
below) to change trigger setups.
Press [F1] and select EDGE TRIGGER
Press [F2] and set TRIGGER SOURCE to CH1
Press [F3] and set EDGE TYPE as RISING
Press [F4] and set TRIGGER MODE as AUTO
Press [F5] and set TRIGGER COUPLING as DC
Figure 1-12 Trigger menu
Figure 1-12
3.
Press [FORCE] to generate a compulsory trigger
signal that is mainly used in the normal and single
trigger modes.
22
You should be familiar with basic operation of the
vertical controls, horizontal controls and trigger
system menu of your Digital Storage Oscilloscope
DSO-1022 M by now. After reading the last chapter,
you should be able to use the menus to set up your
oscilloscope. If you are still unfamiliar with these
basic operation and methods, please read Chapter 1.
This chapter will guide you through the following:
Setting up the vertical system ([CH1], [CH2],
[MATH], [REF], [OFF], [VER
[VERTICAL SCALING])
Setting up the horizontal system ([MENU],
[HORIZONTAL POSITION], [HORIZONTAL SCALING])
Setting up the Trigger system ([TRIGGER LEVEL],
[MENU], [50%], [FORCE])
Setting up the sampling method ([ACQUIRE])
Setting up the display mode ([DISPLAY])
Save and exit ([STORAGE])
Setting up the help system ([UTILITY])
Auto measurement ([MEASURE])
Cursor measurement ([CURSOR])
Using the execution buttons ([AUTO], [START/STOP])
TICAL POSITION],
23
Chapter 2 - Instrument Setups
It is recommended that you read this chapter carefully
to understand the various measurement functions and
system operation of your
DSO-1022 M
Setting up the Vertical System
CH1, CH2 and setups
Each channel has its own vertical menu. You should
set up each item for each channel individually. Press
the [CH1] or [CH2] function button and the system will
display the operation menu for CH1 or CH2. For
explanatory notes please see Table 2-1 below:
Digital Storage Oscilloscope
Table 2-1:
Function Menu Setup Explanatory Note
Coupling
Bandwidth
limit
VOLTS/DIV
Explanatory notes for channel menu
Intercept the DC
AC
DC
Ground
On
Off
Coarse
tune
Fine
tune
quantities of the input
signal.
Pass AC and DC
quantities of input
signal.
Disconnect input signal.
Limit bandwidth to
20MHz to reduce
noise display.
Full bandwidth.
Coarse tune in steps of
1-2-5 to set up the
deflection factor of the
vertical system. Fine
tune means further
tuning within the
coarse tune setup
range to improve the
vertical resolution.
Function Menu Setup Explanatory Note
Select either one value
based on the probe
attenuation factor to
keep the vertical
deflection factor
reading correct. There
are four values: 1X,
10X, 100X and 1000X.
Waveform invert
function on.
Normal waveform
display.
Probe
Invert
1X
10X
On
Off
24
1. Setting up channel coupling
Take an example of applying a signal to CH1. The
signal being measured is a sine signal that contains
DC quantities.
Press [F1] to select AC. It is now set up as AC
coupling. DC quantities of the signal being measured
will be intercepted. The waveform display is as
follows:
Figure 2-1 DC quantities of the signal are intercepted
Figure 2-1
25
Press [F1] to select DC. Both DC and AC quantities
of the signal being measured can pass through. The
waveform display is as follows:
Figure 2-2 Both DC and AC quantities of the
signal are displayed
Figure 2-2
Press [F1] to select GROUND. It is now set up as
ground. Both DC and AC quantities contained in the
signal being measured will be intercepted. The
waveform display is as follows:
(Note: In this mode, although waveform is not
displayed, the signal remains connected to the
channel circuit)
Figure 2-3 Both DC and AC quantities of the signal
are intercepted
2. Setting up the channel band
Take an example of applying a signal to CH1. The
signal being measured is a pulse signal that contains
high frequency oscillation.
Press [CH1] to turn CH1 on. Then press [F2] to set
BANDWIDTH LIMIT OFF
bandwidth. The signal being measured can pass
through even if it contains high frequency quantities.
The waveform display is as follows:
Figure 2-4 Waveform display when bandwidth limit is
off
. It is now set up as full
width
Figure 2-3
Figure 2-4
26
Press [F2] to set BANDWIDTH LIMIT ON. All high
frequency quantities higher than 20MHz in the signal
being measured will be limited. The waveform
display is as follows:
Figure 2-5 Waveform display when
bandwidth limit is on
Figure 2-5
27
3. Setting up the probe r
To match the PROBE attenuation factor setup, it is
necessary to set up the probe attenuation factor in
the channel operation menu accordingly. For
example, when the probe attenuation factor is 10:1,
set the probe attenuation factor at 10X in the menu.
Apply this principle to other values to ensure the
voltage reading is correct.
The figure below shows the setup and vertical range
display when the probe is set at 10:1:
ate
Figure 2-6 Setting up the probe attenuation factor in
the channel menu
Figure 2-6
4. Vertical VOLTS/DIV adjustment setup
You can adjust the VOLTS/DIV range of the vertical
deflection factor either in the coarse tune mode or fine
tune mode. In COARSE TUNE mode, the VOLTS/DIV
range is 2mV/div~5V/div
In FINE TUNE mode, you can change the deflection
factor in even smaller steps within the current vertical
range, so as to continuously adjust the vertical
deflection factor within the range of 2mV/div~5V/div
without interruption.
. Tuning is in steps of 1-2-5.
28
Loading...
+ 70 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.