Theory of Operation, Calibration
and PM procedures
June 2002 X3 All Added Extended Functions and
UVT Test Screens. Added Error
Codes.
August 2002 A All Final review & engineering
release. Modified OVP,
Calibration and screens.
December 2004 B 39 Replaced Figure 3.13 Umbrella
Check Valve with Figure 3.13
Umbrella Check Valve (P/N
21950)
L1534 Revision B December 2004
Page 3
Service Manual 3
Warranty
THE Vela ventilator systems are warranted to be free from defects in material and workmanship and to
meet the published specifications for two (2) years or 8,000 hours, whichever occurs first. The turbine
only is warranted to be free from defects in material or workmanship for five (5) years or 40,000 hours
whichever occurs first.
The liability of VIASYS Healthcare, Critical Care Division, (referred to as the Company) under this
warranty is limited to replacing, repairing or issuing credit, at the discretion of the Company, for parts
that become defective or fail to meet published specifications during the warranty period; the Company
will not be liable under this warranty unless (A) the Company is promptly notified in writing by Buyer
upon discovery of defects or failure to meet published specifications; (B) the defective unit or part is
returned to the Company, transportation charges prepaid by Buyer; (C) the defective unit or part is
received by the Company for adjustment no later than four weeks following the last day of the warranty
period; and (D) the Company’s examination of such unit or part shall disclose, to its satisfaction, that
such defects or failures have not been caused by misuse, neglect, improper installation, unauthorized
repair, alteration or accident.
Any authorization of the Company for repair or alteration by the Buyer must be in writing to prevent
voiding the warranty. In no event shall the Company be liable to the Buyer for loss of profits, loss of
use, consequential damage or damages of any kind based upon a claim for breach of warranty, other
than the purchase price of any defective product covered hereunder.
The Company warranties as herein and above set forth shall not be enlarged, diminished or affected
by, and no obligation or liability shall arise or grow out of the rendering of technical advice or service by
the Company or its agents in connection with the Buyer's order of the products furnished hereunder.
Limitation of Liabilities
This warranty does not cover normal maintenance such as cleaning, adjustment or lubrication and
updating of equipment parts. This warranty shall be void and shall not apply if the equipment is used
with accessories or parts not manufactured by the Company or authorized for use in writing by the
Company or if the equipment is not maintained in accordance with the prescribed schedule of
maintenance.
The warranty stated above shall extend for a period of FIVE (5) years from date of shipment or 40,000
hours of use, whichever occurs first, with the following exceptions:
1. Components for monitoring of physical variables such as temperature, pressure, or flow are
warranted for ninety (90) days from date of receipt.
2. Elastomeric components and other parts or components subject to deterioration, over which the
Company has no control, are warranted for sixty (60) days from date of receipt.
3. Internal batteries are warranted for ninety (90) days from the date of receipt.
The foregoing is in lieu of any warranty, expressed or implied, including, without limitation, any warranty
of merchantability, except as to title, and can be amended only in writing by a duly authorized
representative of the Company.
L1534 Revision B December 2004
Page 4
4 Vela Ventilators
Contents
Revision History ..............................................................................................................2
Limitation of Liabilities..................................................................................................................................3
Chapter 2 Theory of Operation.....................................................................................17
General Device Description ........................................................................................................17
Pneumatic System Overview ......................................................................................................17
Flow Delivery System ................................................................................................................................17
Safety System ...........................................................................................................................................18
Inspiratory Hold Valve................................................................................................................................18
User interface module (UIM) .....................................................................................................................18
Power System ...........................................................................................................................................20
Main Controller System .............................................................................................................................20
The Watchdog Timer and Hardware Fault.................................................................................................21
Disassembly and Reassembly Procedures............................................................................... 24
Power Cable..............................................................................................................................................24
Top Cover Part number 15893 .................................................................................................................. 25
Battery Tray Part Number 10633...............................................................................................................26
Switch Test ................................................................................................................................................56
Alarm Test .................................................................................................................................................57
Version Information ...................................................................................................................................64
Date & Time...............................................................................................................................................65
Low Min Volume Off ..................................................................................................................................67
Language Selection...................................................................................................................................70
Test Set-up................................................................................................................................................75
Power Up Verification/Service Verification Tests.......................................................................................75
Replacing the software PCMCIA cards .....................................................................................................89
Replacing the external A/C fuses. .............................................................................................................90
L1534 Revision B December 2004
Page 7
Service Manual 7
The Annual 5000 hour P.M. Procedure...................................................................................... 91
Replacing the ambient air filter. .................................................................................................................91
Replacing the fan filter...............................................................................................................................91
Replacing the filter, retaining ring & check valve on the low pres O2 fitting ..............................................92
Replacing the filter, retaining ring and check valve on the hi pressure fitting ............................................92
Replacing the muffler and muffler filter assemblies ...................................................................................93
Replacing the left panel w/power PCB. .....................................................................................................95
AC Power Supply ....................................................................................................................................105
DC Power Supply ....................................................................................................................................105
Data Input / Output.................................................................................................................... 106
Analog Inputs ..........................................................................................................................................106
Analog Outputs........................................................................................................................................ 106
Digital Communication.............................................................................................................................107
Video Output ...........................................................................................................................................107
This work is protected under Title 17 of the U.S. Code and is the sole property of the Company. No part
of this document may be copied or otherwise reproduced, or stored in any electronic information
retrieval system, except as specifically permitted under U.S. Copyright law, without the prior written
consent of the Company. For more information, contact:
World Headquarters European Office
1100 Bird Center Drive Rembrandtlaan 1b
Palm Springs, CA 92262-8099 3723 BG Bilthoven
U.S.A. P.O. Box 299, 3720 AG Bilthoven
Phone: (760) 778-7200 The Netherlands
(800) 328-4139 Phone: (31) 30 2289 711
Fax: (760) 778-7274 Fax: (31) 30 2286 244
www.ViasysCriticalCare.com
Trademark Notices
Vela is a registered trademark of VIASYS Healthcare, Critical Care Division in the U.S. and some
other countries. All other brand names and product names mentioned in this manual are trademarks,
registered trademarks, or trade names of their respective holders.
EMC Notice
This equipment generates, uses, and can radiate radio frequency energy. If not installed and used in
accordance with the instructions in this manual, electromagnetic interference may result. The
equipment has been tested and found to comply with the limits set forth in EN60601-1-2 for Medical
Products. These limits provide reasonable protection against electromagnetic interference when
operated in the intended use environments described in this manual.
The ventilator has been tested to conform to the following specifications:
This ventilator is also designed and manufactured to comply with the safety requirements of IEC 601-1,
IEC 601-2-12, CAN/CSA-C22.2 No. 601.1-M90, and UL 2601-1.
L1534 Revision B December 2004
Page 10
10 Vela Ventilators
MRI Notice
This equipment contains electromagnetic components whose operation can be affected by intense
electromagnetic fields.
Do not operate the ventilator in a MRI environment or in the vicinity of high-frequency surgical
diathermy equipment, defibrillators, or short-wave therapy equipment. Electromagnetic interference
could disrupt the operation of the ventilator.
Intended Use Notice
The Vela Ventilators are designed to provide ventilator support for the critical care management of
infant, pediatric or adult patients with compromised lung function. They are intended to provide
continuous respiratory support in an institutional health care environment. They should only be
operated by properly trained clinical personnel, under the direction of a physician.
Regulatory Notice
Federal law restricts the sale of this device except by or on order of a physician.
IEC Classification
Type of Equipment: Medical Equipment, Class 1 type B
Adult/Pediatric/Infant Lung Ventilator
Declaration of Conformity Notice
This medical equipment complies with the Medical Device Directive,
93/42/EEC, and the following Technical Standards, to which
Conformity is declared:
EN60601-1
EN60601-1-2
ISO 9001, EN 46001
EU Notified Body:
BSI (Reg. No. 0086)
Tradenames:
Vela
If you have a question regarding the Declaration of Conformity for this product, please contact VIASYS
Healthcare, Critical Care Division at the number given in Appendix A.
L1534 Revision B December 2004
Page 11
Service Manual Front Matter 11
Safety Information
Please review the following safety information prior to operating the ventilator. Attempting to
operate the ventilator without fully understanding its features and functions may result in unsafe
operating conditions.
Warnings and Cautions which are general to the use of the ventilator under all circumstances are
included in this section. Some Warnings and Cautions are also inserted within the manual where they
are most meaningful.
Notes are also located throughout the manual to provide additional information related to specific
features.
If you have a question regarding the installation, set up, operation, or maintenance of the ventilator,
contact VASYS Healthcare Customer Care as shown in Appendix A, Contact & Ordering Information.
Terms
WARNINGS identify conditions or practices that could result in serious adverse reactions or
potential safety hazards.
CAUTIONS identify conditions or practices that could result in damage to the ventilator or other
NOTES identify supplemental information to help you better understand how the ventilator
Warnings
Warnings and Cautions appear throughout this manual where they are relevant. The Warnings and
Cautions listed here apply generally any time you operate the ventilator.
• The Vela Ventilator is intended for use by a trained practitioner under the direction of a
• When the ventilator is connected to a patient, a trained health care professional should be in
• Alarm loudness must be set above ambient sound in order to be heard.
• Always have an alternate means of ventilation available whenever the ventilator is in use.
• The operator should not touch the electrical connectors of the ventilator or accessories, and
• Due to possible explosion hazard, the ventilator should not be used in the presence of
equipment.
works.
qualified physician.
attendance at all times to react to an alarm or other indications of a problem.
the patient simultaneously.
flammable anesthetics.
• An audible alarm indicates an anomalous condition and should never go unheeded.
• Anti-static or electrically conductive hoses or tubing should not be used within the patient
circuit.
• If a mechanical or electrical problem is recognized while running the Operational Verification
Tests, or while operating the ventilator, the ventilator must be removed from use and referred
to qualified personnel for servicing. Using an inoperative ventilator may result in patient injury.
L1534 Revision B December 2004
Page 12
12 Vela Ventilators
• When a low gas supply alarm occurs, the oxygen concentration delivered to the patient will
differ from that set on the O2 control setting.
• A source gas failure will change the FIO2 and may result in patient injury.
• The functioning of this equipment may be adversely affected by the operation of other
equipment nearby, such as high frequency surgical (diathermy) equipment, defibrillators,
short-wave therapy equipment, “walkie-talkies,” or cellular phones.
• Water in the air supply can cause malfunction of this equipment.
• Do not block or restrict the Oxygen bleed port located on the instrument back panel.
Equipment malfunction may result.
• Electric shock hazard - Do not remove any of the ventilator covers or panels. Refer all
servicing to an authorized VIASYS Healthcare service technician.
• A protective ground connection by way of the grounding conductor in the power cord is
essential for safe operation. Upon loss of protective ground, all conductive parts including
knobs and controls that may appear to be insulated, can render an electric shock. To avoid
electrical shock, plug the power cord into a properly wired receptacle, use only the power cord
supplied with the ventilator, and make sure the power cord is in good condition.
Cautions
The following cautions apply any time you work with the ventilator.
• When replacing fuses, ensure that new fuses are of the same type and value as those being
replaced. Incorrect fuses can cause damage to the ventilator.
• A battery that is fully drained (i.e. void of any charge) may cause damage to the ventilator and
should be replaced.
• All accessory equipment that is connected to the ventilator must comply with
CSA/IEC601/UL2601.
• To avoid damage to the equipment, clean the air filter regularly.
The following cautions apply when cleaning the ventilator or when sterilizing ventilator
accessories.
• Do not sterilize the ventilator. The internal components are not compatible with sterilization
techniques.
• Do not gas sterilize or steam autoclave tubing adapters or connectors in place. The tubing will,
over time, take the shape of the adapter, causing poor connection and possible leaks.
• DO NOT submerge the ventilator or pour cleaning liquids over or into the ventilator.
L1534 Revision B December 2004
Page 13
Service Manual Front Matter 13
Equipment Symbols
The following symbols may be referenced on the ventilator or in accompanying documentation
Symbol Source/Compliance Meaning
Symbol #03-02 IEC 60878 Indicates ATTENTION, consult ACCOMPANYING DOCUMENTS
Symbol #5016 IEC 60417
Symbol #5034 IEC 60417
Symbol #01-36 IEC 60878
Symbol #5035 IEC 60417
Symbol #01-37 IEC 60878
Symbol #5019 IEC 60417
Symbol #01-20 IEC 60878
Symbol #5021 IEC 60417
Symbol # 01-24 IEC 60878
Symbol # 5333 IEC 60417
Symbol #02-03 IEC 60878
Symbol #5032 IEC 60417
Symbol #01-14 IEC 30878
Symbol# 5049 IEC 60417
Symbol #5007 IEC 60417
Symbol #01-01 IEC 60878
Symbol #5008 IEC 60417
Symbol #01-02 IEC 60878
This symbol indicates a FUSE.
This symbol indicates INPUT.
This symbol indicates OUTPUT
This symbol indicates protective EARTH (ground).
This symbol indicates the EQUIPOTENTIAL connection used to
connect various parts of the equipment or of a system to the same
potential, not necessarily being the earth (ground) potential (e.g., for
local bonding).
This symbol indicates TYPE BH equipment, which indicates equipment
that provides a particular degree of protection against electric shock,
particularly with regards to allowable leakage current and reliability of
the protective earth connection.
This symbol indicates the equipment is suitable for alternating current.
This Symbol indicates the ON condition for a part of the equipment.
When pressed the ventilator will operate from the MAINS voltage (if
connected) or internal or external batteries if the battery charge is
within operating specifications.
Indicates ON (Power)
Indicates OFF (Power)
Horizontal return with line feed. Indicates ACCEPT entered values for
a specific field.
This symbol indicates CANCEL. Do not accept entered values. The
ventilator continues to operate at previous settings.
Pressing the button with this symbol will FREEZE the current display.
This symbol indicates a CONTROL LOCK.
ACCEPT
CANCEL
Symbol #0651 ISO 7000
Graphical Symbol in
general use internationally
for “DO NOT”
Symbol #5467 IEC 60417
Symbol #5569 IEC 60417
L1534 Revision B December 2004
Page 14
14 Vela Ventilators
Symbol Source/Compliance Meaning
VIASYS Healthcare symbol
This symbol represents a NEBULIZER.
Symbol #5319 IEC 60417
Symbol #5307 IEC 60417
VIASYS Healthcare symbol
This symbol indicates ALARM SILENCE
This symbol indicates ALARM RESET
Increase OXYGEN
VIASYS Healthcare symbol
Symbol #5031 IEC 60417
Indicates VARIABLE ORIFICE FLOW SENSOR
This symbol indicates DIRECT CURRENT (DC)
Symbol #5546 IEC 60417
This symbol indicates the INTERNAL BATTERY STATUS display
VIASYS Healthcare symbol
This symbol indicates INSPIRATORY HOLD
VIASYS Healthcare symbol
This symbol indicates EXPIRATORY HOLD
VIASYS Healthcare symbol
This symbol indicates MANUAL BREATH
L1534 Revision B December 2004
Page 15
Chapter 1 Introduction
General Instructions
When disassembling or assembling the Vela, refer to the pneumatic schematic, tubing diagram, and the
wiring diagram shown in Appendix B and the appropriate schematics and assembly drawings for each
assembly. The illustrations shown in this manual are for reference only, current revisions of these
diagrams and schematics are available to qualified personnel from VIASYS Healthcare, Critical Care
Division, Technical Support.
Always take standard ESD precautions when working on Vela ventilator systems.
Ensure the ventilator is disconnected from the AC power supply before performing and repairs or
maintenance. When you remove any of the ventilator covers or panels, immediately disconnect the
internal battery “quick release” connector (see figure 3.1) before working on the ventilator.
Recommended Tools & Equipment
Note
Before using any test equipment [electronic or pneumatic] for calibration procedures, the accuracy of the
instruments must be verified by a testing laboratory. The laboratory master test instruments must be traceable to
the NIST (National Institute of Standards Technology) or equivalent.
When variances exist between the indicated and actual values, the calibration curves [provided for each
instrument by the testing laboratory] must be used to establish the actual correct values. This certification
procedure should be performed at least once every six months. More frequent certification may be required
based on usage.
Long & short Philips screwdrivers
Flat bladed screwdriver
¼” Nut Dirver
5/16” Nut Driver
7/8” Nut Driver
11/32” Nut Drivel
Digital Volt Meter
Tack puller or Needle nosed pliers
Diagonal cutters
1” and ¾” open ended wrenches
Pressure Manometer (cmH2O and psig)
Adult Test Lung P/N 33754
Adult Patient Circuit P/N 10684
Variable Orifice Flow Sensor assembly P/N 15972
Valve Body P/N 20005
Tapered nipple P/N 00680
Hex nut P/N 00822
Regulator P/N 6754
1/8” ID Tubing tee P/N 00358 D (10pk)
1/8” ID silicone tubing P/N 04029 X (50ft)
L1534 Revision B December 2004
Page 16
16 Vela Ventilators
Recommended Maintenance Schedules
Schedules
Every 500 hours , the fan and ambient air filters should be cleaned and replaced if necessary.
Every 5000 hours, VIASYS recommends that the following Preventive Maintenance procedure be
performed (see chapter 4 for instructions). This procedure includes:
• Replacement of the fan filter, the ambient air filter, the cone filters in the high & low pressure
gas inlets and the turbine mufflers/filter.
• Perform verification procedures described in Chapter 4
• Calibration of the transducers & solenoids if necessary.
Every 10,000 hours or every two years, whichever occurs sooner, the internal oxygen sensor should
be replaced.
Maintenance on the Vela should only be carried out by a trained and authorized service technician.
VIASYS Healthcare will make available to qualified technicians, service manuals and such items as
circuit diagrams, component parts lists, calibration instructions and other information to assist in repair
of those parts of the ventilator designated by the manufacturer as repairable items.
The drawings, diagrams and schematics included in this manual are for reference only and may be
updated separately from this manual after publication. For current revisions of all documentation,
contact VIASYS Healthcare Tech Support at the numbers provided in Appendix A.
L1534 Revision B December 2004
Page 17
Chapter 2 Theory of Operation
General Device Description
The Vela Ventilator uses a revolutionary turbine gas delivery system with sophisticated microprocessor
control. Its Graphical User Interface provides support for pediatric to adult patients. The Vela can
deliver clinically advanced modes of ventilation like Pressure Support and can be powered with an
internal battery or AC power for an more extensive patient range.
Pneumatic System Overview
The Vela ventilator pneumatic system is electromechanical and is comprised of four major subsystems,
each containing several components. These systems are the flow delivery system, the exhalation
system, the safety system and the inspiratory hold valve. Individual subsystems are discussed in detail
below.
Flow Delivery System
This electromechanical system controls all inspiratory flow to the patient. The system delivers flow to
satisfy criteria for many breath types, including volume controlled, pressure controlled, and pressure
supported. The system comprises a turbine, differential pressure transducer, 2 auto-zero valves, and an
optical encoder speed transducer. When a breath is initiated, the controller controls the speed of the
turbine to achieve the required flow rate.
The speed and differential pressure transducer signals function as control inputs to ensure that the
proper flow rate is delivered even when backpressure varies. Periodically, the auto zero valves activate
to reference both sides of the differential pressure transducer to ambient pressure. The offset is
recorded by the controller, and is used as a correction for future pressure measurements. This
compensates for long term and temperature drift. Materials exposed to patient gases include
compatible plastics, aluminum, and plated steel.
Exhalation System
The exhalation system controls the flow of gas from the patient’s lungs during the exhalation phase of a
breath. This electromechanical subsystem is made up of an exhalation valve, a flow transducer, a
differential pressure transducer, an airway pressure transducer, and three auto zero solenoid valves.
During exhalation, the outflow of gases is regulated by the exhalation valve to achieve the set PEEP.
The exhalation valve is comprised of an electromagnetic linear actuator operating against a mechanical
poppet/seat. The gas flow travels through the flow transducer. The flow transducer is a variable orifice
type and creates a differential pressure proportional to flow. This differential pressure is transmitted to
the differential pressure transducer, which converts the pressure signal to an electrical signal. The
controller uses this signal for flow triggering and to monitor exhaled tidal volume. The airway pressure
transducer reads pressure in the exhalation leg of the patient circuit. This signal is used as a feedback
signal for controlling PEEP, pressure control, pressure support, and various pressure monitors.
Periodically, the auto zero valves activate to reference the differential and airway pressure transducers
to ambient pressure. The offset is recorded by the controller, and is used as an offset for future
pressure measurements. This compensates for long term and temperature drift. Materials exposed to
patient gases include compatible plastics, aluminum, and stainless steel.
L1534 Revision B December 2004
Page 18
18 Vela Ventilators
Safety System
The mechanical safety system ensures that the patient can breath spontaneously from room air and
that the patient pressure is limited to a maximum preset value in the event of a ventilator malfunction.
This mechanical system consists of a pressure relief valve and a sub ambient relief valve. In the event
of a ventilator malfunction that results in high pressure, the pressure is limited by a relief valve. The
relief valve consists of a user-adjustable, spring-loaded poppet acting against a seat.
In the event the ventilator fails to deliver a breath, the patient may inspire spontaneously by drawing
room air through the sub ambient relief valve.
Materials exposed to patient gas are aluminum, compatible rubber, and compatible plastics.
Inspiratory Hold Valve
The inspiratory hold valve is an electromechanical solenoid valve. If activated, the inspiratory hold
valve blocks flow between the flow delivery system and the patient. This valve is activate during
inspiratory hold and maximum inspiratory pressure maneuvers. Materials exposed to patient gases are
aluminum and compatible rubber and plastic.
Oxygen Blending System
The optional oxygen blending system is made up of an O2 Inlet Transducer, five solenoid valves, five
flow orifices, an inlet filter, and an accumulator. When a breath is initiated, the turbine draws mixed gas
from the accumulator. Filtered air is drawn into the accumulator through the filter. Oxygen is supplied
to the accumulator through the solenoids and orifices. The controller opens and closes the valves as
required to supply the correct amount of oxygen to satisfy the O2 setting and the flow demand. The
signal from the O2 inlet pressure transducer is used to compensated delivered O2 for O2 inlet pressure
variations. Surfaces exposed to patient gas are constructed from compatible plastics, plated steel, and
aluminum.
There is also an optional oxygen inlet port, which allows for low-flow titration of oxygen into the gas
output of this device.
Electronic Overview
The Vela ventilator electronic system is comprised of several subsystems, each containing numerous
components. These subsystems are the GUI System, the Power System, the Main Controller System,
and the Exhalation and Flow Delivery systems. Individual subsystems are discussed in detail.
User interface module (UIM)
The UIM consists of a 10.4-inch, 800x600 active matrix LCD with an analog resistive touch screen
overlay, a back light inverter, a set of membrane key panels, an optical encoder, and a Control PCB.
Software and the touch screen provide a set of context sensitive soft keys. The membrane panel
provides a set of hard (permanent) keys for dedicated functions. Selecting the function with a soft key
and adjusting the setting using the optical encoder changes a parameter. The parameter is accepted or
canceled by pressing the appropriate membrane key.
L1534 Revision B December 2004
Page 19
Service Manual Theory of Operation 19
(3 & 4)
LCD & TOUCHSCREEN
(1 & 2)
MEMBRANE PANEL WITH
(5)
EMBEDDED LED'S
BACK LIGHT AC VOLTAGE
DIGITAL
TOUCH
SCREEN
BACK LIGHT INVERTER
(6)
CONTROL PCB
(7)
DC-DC
CONVERTERS
24VDC
5V
5V, PRN PORT
3.3V, DIG LOGIC
HIGH SPEED
SERIAL
DIGITAL
UNIVERSAL SERIAL BUS (FUTURE)
PRINTER
RS232
RS232
RS232 (MIB)
(CRT)
Figure 1 -- User Interface Design Module Block Diagram
The UIM performs all ventilator control functions, gas calculations, monitoring and user interface
functions. The UIM uses a Graphical User Interface (GUI) via the active matrix SVGA LCD and resistive
touch screen to provide system and patient information to the user and to allow the user to modify
ventilator settings. The Monitor MCU handles all user interface requirements, including updating the
active matrix liquid crystal display (LCD), monitoring the membrane keypad, analog resistive touch
screen, and optical encoder for activity. The Monitor MCU also performs all the input/output functions of
the UIM, including RS-232, printer, video output, and IEEE 1073 Medical Information Bus (MIB).
Communication between the Control and Monitor MCU’s is accomplished via an 8 bit dual port SRAM.
Liquid Crystal Display
The liquid crystal display (LCD) provides graphical and digital feedback to the clinician. The panel is a
10.4” SVGA, 800x600 pixel, active matrix LCD. The LCD is used to implement the graphical user
interface (GUI). It provides all of the adjustable controls and alarms, as well as displays waveforms,
loops, digital monitors and alarm status in real time.
L1534 Revision B December 2004
Page 20
20 Vela Ventilators
Touch Screen
The touch screen is a 10.4” analog resistive overlay on a piece of glass, which is placed over an LCD
screen. The touch screen and the LCD together provide a set of software configurable soft keys. The
software enables the keys to be context sensitive. The touch screen has a resolution of 1024x1024.
Physically, the touch screen consists of two opposing transparent resistive layers separated by
insulating spacers. Touching the screen brings the two opposing layers into electrical contact. The Y
coordinate is determined by applying a voltage from top to bottom on the top resistive layer. This
creates a voltage gradient across this layer. The point of contact forms a voltage divider, which is read
by the analog-to-digital converter. The X coordinate is determined by applying a voltage from left to
right on the bottom resistive layer. Again this creates a voltage gradient and the point of contact forms a
divider, which is read with an analog-to-digital converter.
Membrane Panel
The membrane panel provides a set of permanent dedicated keys, which enable control of ventilator
functions. The membrane panel also provides visual display using embedded light emitting diodes
(LEDs). The membrane panel consists of membrane switches, which are read by the monitor CPU. The
switches form a matrix of rows and columns. A key closure causes an interrupt to the monitor CPU,
which responds by scanning the key matrix to determine which key has been pressed.
Light Emitting Diodes (Leds)
Some of the membrane keys require LED’s to indicate when the key is active. The LED’s are
embedded into the membrane panels.
Optical Encoder
The optical encoder allows settings to be modified. The setting is selected by pressing a soft key on the
LCD and then modified by turning the optical encoder (data dial) to change the value. When the
encoder is rotated two pulse streams are generated, phase A and B. When the encoder is turned
clockwise, phase A leads B by 90 degrees. When the direction is counter clockwise, phase B leads A
by 90 degrees. The electronics uses the phase information to drive an up-down counter, which is read
by the monitor CPU. The optical encoder is not interrupt-driven and therefore must be polled by the
monitor CPU.
Back Light Inverter
The back light inverter converts 5 VDC into the high frequency AC voltage necessary to power the LCD
back light, which is used to illuminate the LCD.
Power System
The Power System conditions and controls electrical energy from the AC line input and the internal
battery. When energy is available from the AC line, the ventilator operates from this source, and also
recharges the internal battery. When AC line power is not available, the power system draws energy
from the internal battery. The power system uses energy efficient DC-to-DC converter technology to
convert energy from the AC line or battery to appropriate voltages and currents to supply power to
ventilator components and systems.
Main Controller System
The Main Controller System is comprised of three Pressure Transducers, an Analog-to-Digital
Converter, two Digital-to-Analog Converters, the Input-Output Processor, Solenoid Valves, and the
Watchdog and Hardware Fault Monitors.
L1534 Revision B December 2004
Page 21
Service Manual Theory of Operation 21
One of the pressure transducers measures the patient circuit pressure. This pressure is an input to the
controller. A differential pressure transducer measures the pressure across the turbine. This pressure is
also an input to the controller. A second differential pressure transducer is used to measure the flow at
the outlet of the exhalation valve. This pressure is also an input to the controller.
Analog to digital converters are used to change the analog pressure signals into measured binary
numeric values for use by the microprocessor in the controller.
Digital to Analog converters are used to change the binary numeric commands generated by the
microprocessor in the controller into analog signals which drive the turbine and exhalation valve.
The Input-Output Processor is a small microcontroller which, under software control, performs several
repetitive tasks such as generating the refresh signals for the display system, cycling the A-to-D
converters through a pattern of measurements from the multiple signal sources, and scanning the
control panel for pressed buttons. Such repetitive tasks are thereby off loaded from the Main Processor.
Solenoid Valves and Valve Drivers (including the Auto Zero valves) are employed on the Circuit
Pressure transducer and on the Turbine Differential Pressure Transducer. These valves allow the
controller software to compensate for long term drift and temperature induced zero shift in the pressure
transducers by periodically rechecking the zero pressure readings. Similar solenoid valves are
employed in the Oxygen Blending System. The valve drivers for the Auto Zero and Blender valves are
similar.
The Main Processor is a 386-type CPU which controls all ventilator functions. All user settings for
alarms, controls, ventilation mode, waveform, and monitoried data are stored here and are combined
with measured pressure, flow, and speed data to cause the ventilator to function. The algorithms,
formulae, and control functions which define ventilator behavior are contained in the software program
executed by the CPU.
The Watchdog Timer and Hardware Fault
Monitors shut down the ventilator if a malfunciton is detected. The Watchdog Timer consists of two
timers and a PAL containing a state machine. The Main Controller CPU must communicate with a state
machine at intervals within a time window set by the two timers. The CPU must obtain a key from the
PAL and send the correct address and data response back to the state machine at each interval. If the
response is incorrect, or comes at an invalid time, the Watchdog shuts down the CPU and forces the
ventilator hardware to a safe state. The Hardware Fault Monitors check the status of the power supplies
to the ventilator electornics. If any is out of the safe operationg range, the ventilator will shut down and
cannot be made to operate until the fault is corrected.
Exhalation System
The electrical portion of the exhalation system is comprised of the Exhalation Valve Driver Circuitry.
The driver converts the low voltage signal output by a D-to-A converter into a controlled constant
current which energizes the linear solenoid positioner in the exhalation valve.
Flow Delivery System
The electrical portion of the flow delivery system is comprised of a 3 Phase Brushless Motor Driver, and
an Optical Speed Transducer.
The 3 Phase Brushless DC Motor Driver converts the low voltage signal output by a D-to-A converter
into three controlled currents which energize the three motor phases and cause the motor to create a
torque, resulting in motor rotation. The torque generated is a function of current, and therefore of the
control voltage from the D-to-A converter. The speed of rotation is monitored by the optical Speed
Transducer. The transducer outputs a train of pulses with a frequency proportional to the rotational
speed of the motor. This pulse train is a control feedback input to the controller.
L1534 Revision B December 2004
Page 22
22 Vela Ventilators
Oxygen Blending System
The electrical portion of the optional oxygen blending system is made up of a safety solenoid, a
pressure regulator set to 40 PSI, an O2 Inlet Transducer, five Solenoid Valves, one Nebulizer solenoid,
and the driver circuitry for the solenoid valves.
The Oxygen Inlet Pressure Transducer measures the incoming gas pressure so that O2 delivery can be
compensated for inlet pressure fluctuations.
The Solenoid Valves are energized and deenergized under software control by the Main controller to
supply the correct amount of oxygen to satisfy the current O2 setting and current gas flow demand.
The driver circuitry translates the binary logic signals presented by the controller to larger voltage and
currents suitable for energizing the Solenoid Valves.
L1534 Revision B December 2004
Page 23
Chapter 3 Disassembly & Assembly
General Instructions and Warnings
When performing the procedures in this chapter, refer to the Vela wiring and tubing diagrams.
Reference copies of these are located in Appendix B of this manual. Ensure that you follow these safety
warnings and precautions:
Figure 3.1 Battery Disconnect Molex Connector
WARNING
Always disconnect the main power cable before removing the instrument cover and disconnect the
battery once the top cover and battery tray have been removed to prevent injury and/or damage to the
VELA Ventilator System (see figure 3.1).
CAUTION
The Vela contains ESD susceptible components. Ensure you are properly grounded before performing
any service or maintenance procedures and store ESD susceptible electrical components in an antistatic bag to prevent damage to the component.
Note
When the batteries are disconnected, the system will automatically re-set the battery status memory and will
initiate an 18-hour re-charge cycle upon re-connect. During this period, the red DC status light will remain on.
If the DC status light remains lit after the initial charging period has expired, contact VIASYS tech support as
described in Appendix A for a replacement battery.
Note
Do not cut the main harness tie wraps since the entire harness is regarded as one component.
Note
The terms left and right refer to a view from the front of the unit looking towards the rear.
L1534 Revision B December 2004
Page 24
24 Vela Ventilators
Required Tools
Long & short Philips screwdrivers
Flat bladed screwdriver
¼” Nut Dirver
5/16” Nut Driver
7/8” Nut Driver
11/32” Nut Driver
Digital Volt Meter
Tack puller or Needle nosed pliers
Diagonal cutters
1” and ¾” open ended wrenches
Disassembly and Reassembly Procedures
To perform a complete disassembly of the unit, follow all of the steps in each removal section, in the
order presented in this chapter. To reassemble the unit, follow all of the steps in each installation
section, starting with the last component and finishing with the power cable installation instructions.
Power Cable
Figure 3.2 Power Cable Connector with Guard
Removal
1. Remove the (2)Phillips pan-head screws in the top portion of the power cable guard at the rear of
the unit (see figure 3.2)
2. Remove the guard and unplug the power cable.
Installation
1. Plug the power cable into the rear of the unit.
2. Install the power cable guard with the (2) Phillips pan-head screws in the top portion.
L1534 Revision B December 2004
Page 25
Service Manual Disassembly & Reassembly Procedures 25
Top Cover Part number 15893
Removal
1. Remove the power cable.
2. Remove the (3) Phillips pan-head screws in the upper rear of the front panel (if present).
3. Remove the (4) Phillips pan-head screws in the back panel.
4. Remove the (3) screws from the bottom of each side of the ventilator (if present).
Note
The configuration of the screws securing the cover will depend on the model of Vela that you are working with.
5. When all screws have been removed, slide the top cover towards the rear of the unit and lift off.
CAUTION
The SVGA output ribbon cable lies directly beneath the top cover assembly running from the front of the Vela to
the rear. It can be damaged if care is not taken when removing the cover. Make sure that the cover clears this
cable before sliding it back.
Installation
1. Install the top cover by sliding it toward the front of the unit, over the side rails, and snap into place.
2. Install the (4) Phillips pan-head screws in the back panel.
3. Do not re-install the (3) Phillips pan-head screws in the upper rear of the front panel if present.
They have been removed from later versions of the Vela.
4. Install the power cable.
Cover mounting screws
on rear of instrument
Figure 3.3 Cover Assembly from the rear
L1534 Revision B December 2004
Page 26
26 Vela Ventilators
Battery Tray Part Number 10633
Removal
1. Remove the power cable and top cover.
2. Remove the (4) Phillips pan-head screws in the battery tray.
3. Lift the battery tray out of the unit.
4. Disconnect the batteries from the white Molex DC power connector (see figure 3.4
Figure 3.4 Large white Molex battery disconnect
To release connector, push in locking tabs
To attach push until locking tabs engage.
Tab
Note
When the batteries are disconnected, the system will automatically re-set the battery status memory and will
initiate an 18-hour re-charge cycle upon re-connect. During this period, the red DC status light will remain on.
If the DC status light remains lit after the initial charging period has expired, contact VIASYS tech support as
described in Appendix A for a replacement battery.
Installation
1. Connect the white Molex DC power connector.
2. Ensure the Molex connector is pushed down into the unit behind the muffler tubes and install the
battery tray. The tray is correctly installed when the wires run towards the rear of the unit.
3. Secure the battery tray to the left & right panels with (4) Phillips pan-head screws.
4. Install the top cover and re-connect the power cable and the power cable guard.
Battery Tray Parts List
Item Part number Description QtyUM
1 21363 Tray, battery 1 Ea
2 52000-00190 Grommet 11/16 id 0 Ea
3 21543 Battery 12v, 2.7 Ahr, ni-mh 4 Ea
4 15719 Cable assy, battery tray 1 Ea
5 20742 Bracket, capacitor 2 Ea
6 21068 Foam pad, capacitor bracket 2 Ea
7 64084 Capacitor elect 0.1f 63vdc 1 Ea
8 04383 Washer, lock int #6 4 Ea
9 07212 Nut, 6-32 x .093 depth 8 Ea
10 05038 Tie, strap 4 Ea
11 08231 Mounting bracket, cable tie 2 Ea
12 07803 Cable tie 2 Ea
14 09349x Tape, adh-2s,2" x .045" 0 Ft
L1534 Revision B December 2004
Page 27
Service Manual Disassembly & Reassembly Procedures 27
Figure 3.5 Battery Tray and large capacitor assembly
L1534 Revision B December 2004
Page 28
28 Vela Ventilators
PCMCIA Cards
The PCMCIA card is located inside the unit. It contains the Vela software.
Card release
button
Figure 3.6 PCMCIA Card in place
Removal
1. Remove the power cable, top cover, and battery tray.
2. Push in the button under the card to release. (see figure 3.6)
3. Gently slide the card out of its socket.
Installation
1. Slide the card into the socket in the direction of the arrow printed on the card.
2. Click card into place.
3. Install the battery tray, top cover, and power cable.
Left Panel
The left panel is structural only and has no components attached to it.
Removal
1. Remove the power cable, top cover, and battery tray.
2. Remove the (2) Phillips countersink screws at the rear of the panel.
3. Remove the (1) Phillips countersink screw in the upper front of the panel.
4. Lift out the left panel.
Installation
1. Slide the front of the left panel behind the lip on the front panel and align the back of the panel with
the rear panel edge.
2. Install the (2) Phillips countersink screws in the rear of the right panel.
3. Install the (1) Phillips countersink screw in the upper front of the right panel.
4. Install the battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 29
Service Manual Disassembly & Reassembly Procedures 29
Right Panel Containing the Power PCB P/N 15894
Note
The internal battery fuse is located on the power PCB. Call Viasys if this fuse needs replacing.
Removal
1. Remove the power cable, top cover, and battery tray.
2. Remove the (2) Phillips countersink screws from the right side of the rear panel.
3. Remove the (1) Phillips countersink screw in the upper front of the right panel.
4. Remove the (1) Phillips pan-head screw in the lower center of the right panel.
5. Gently lift out the panel and the power PCB. Lay the panel flat and make the following
disconnections:
Ref on Pwr
PCB
J1 26-pin ribbon cable to J2 on main PCB P/N 15494
J2 3-pin to fan assembly
J5 14-pin to J5 on main PCB (in main harness)
J6 10-pin to P2 on turbine motor driver PCB
J7 small 2-pin to main power switch
J300 large 2-pin to battery DC power (in main harness)
J400 4-pin quick disconnect to SVGA PCB
P1-L/N AC line in to power (black-power, white-neutral) P/N 15890
to ground lug (ground)
Connector Desc Connection made
Cable Assy
P/Number
Note
Do NOT remove the connector from J4 of the power PCB.
Installation
Make the following connections to the Power PCB.
Ref on Pwr
PCB
J1 26-pin ribbon cable from J2 on main PCB P/N 15494
J2 3-pin from fan assembly
J5 14-pin from J5 on main PCB (in main harness)
J6 10-pin from P2 on turbine motor driver PCB
J7 small 2-pin from main power switch
J300 large 2-pin from battery DC power (in main harness)
J400 4-pin quick disconnect from SVGA PCB
P1-L/N AC line in to power (black-power, white-neutral) P/N 15890
from ground lug (ground)
Connector Desc Connection made
WARNING
Ensure that the power cables are correctly connected to prevent injury and/or damage to the unit.
Cable Assy
P/Number
L1534 Revision B December 2004
Page 30
30 Vela Ventilators
CAUTION
Ensure that J5 is pinned properly since an incorrect connection could damage the power PCB.
6. Slide the right panel into the front panel, ensuring that it rests on the bottom rail.
7. Install the (1) Phillips pan-head screw in the lower center of the panel.
8. Install the (1) Phillips countersink screw in the upper front of the panel.
9. Install the (2) Phillips countersink screws in the side of the rear panel.
10. Install the battery tray, top cover, and power cable.
Note
The Vela right panel assembly with the power board attached is supplied as a single complete assembly part
number 15894. Do not attempt to disassemble or to replace individual components.
L1534 Revision B December 2004
Page 31
Service Manual Disassembly & Reassembly Procedures 31
Figure 3.7 Right Panel and Power PCB
L1534 Revision B December 2004
Page 32
32 Vela Ventilators
Front Panel Part number 15869
Removal
1. Remove the power cable, top cover, battery tray, right panel, and left panel.
2. Disconnect the following from the main PCB:
JP2 40-pin ribbon cable to JP1 on the SVGA out PCB
J301 12-pin to J301 on the blender PCB P/N 15497
Note
Disconnect the video cable from JP2 on the main PCB rather than from JP1 on the SVGA out PCB.
3. Disconnect the green oxygen nebulizer tube.
4. Remove the (2) Phillips pan-head screws behind the front panel at the lower left and right.
5. Disconnect the 1/8” silicone tube #4 from P2 and #5 from P1 on the exhalation differential
transducer, and #1 from the turbine.
6. Disconnect the 1/16” silicone tubes from the common (front) ports of the turbine outlet auto-zero
solenoid (to the main tubing harness), the turbine inlet auto-zero solenoid (to the turbine), and the
airway pressure auto-zero solenoid (Nafian tube to the exhalation valve receptacle).
7. Disconnect the following from the main PCB:
J4 8-pin to J1 on the FIO2 monitoring PCB
J900 2-pin (alarm loudness) to alarm potentiometer
J701 5-pin to turbine optic encoder and thermistor
J401 4-pin to exhalation valve
U719 fiberoptic to rear panel connector
J5 14-pin to J5 on power PCB
J302 2-pin to inspiratory hold solenoid on manifold base
J2 26-pin to J1 on power PCB P/N 15494
J905 3-pin to J2 on remote alarm (nurse call) PCB
AL901 (hard-wired) 2-prong to alarm cable connector
Installation
1. Connect the green oxygen nebulizer tube.
2. Connect the following to the main PCB:
J401 4-pin from exhalation valve
J701 5-pin from turbine optic encoder and thermistor
U719 fiberoptic from rear panel connector
J5 14-pin from J5 on power PCB
J2 26-pin from J1 on power PCB P/N 15494
J905 3-pin from J2 on remote alarm (nurse call) PCB
3. Slide the front panel into place and secure the (2) Phillips panhead screws behind the front panel
at the lower left and right.
4. Connect the following to the main PCB:
J302 2-pin from inspiratory hold solenoid on manifold base
J4 8-pin from J1 on the FIO2 monitoring PCB
J301 12-pin from J301 on the blender PCB P/N 15497
AL901 (hard-wired) 2-prong from alarm cable connector
J900 2-pin alarm loudness cable from alarm potentiometer
L1534 Revision B December 2004
Page 33
Service Manual Disassembly & Reassembly Procedures 33
5. Connect the 1/16” silicone tubes to the common (front) ports of the turbine outlet auto-zero
solenoid (from the main tubing harness), the turbine inlet auto-zero solenoid (from the turbine), and
the airway pressure auto-zero solenoid (Nafian tube from the exhalation valve receptacle).
6. Connect the 1/8” silicone tube #4 to P2 and #5 to P1 on the exhalation differential transducer, and
#1 to the turbine.
7. Connect the following to the main PCB:
JP2 40-pin ribbon cable from JP1 on the SVGA out PCB
Note
Ensure that the fiber optic cable runs under the turbine muffler, that the tubing runs over the nebulizer line and is not
pinched in between the front and base panels, and that the cable from J701 is securely connected to the turbine.
30 15961 CABLE ASSY, FLOW SENSOR - MAIN BOARD 1 EA
31 15962 CABLE ASSY, VIDEO BOARD TO POWER 1 EA
32 15963 CABLE ASSY, VIDEO BOARD TO MAIN BOARD 1 EA
33 21560 KNOB, SPINNER-1.50 1 EA
34 53002-46204 SCR 6-32X1/4 SS PPH PNH 4 EA*
35 10632 TUBING ASSY, 1/8 X 1/16 3 EA
* Commonly found hardware items are not supplied by VIASYS Healthcare.
L1534 Revision B December 2004
Page 34
34 Vela Ventilators
Figure 3.8 Front panel with main PCB
Main PCB (on Front Panel) P/N 52130
Removal
1. Remove the power cable, top cover, battery tray, right panel, left panel, and front panel.
2. Disconnect the following from the main PCB:
J801 20-pin to J801 on the flow sensor PCB (in harness)
J1210 5-pin (gray top, green) to the flow sensor PCB (in P/N 15961 harness)
J3 10-pin to the flow sensor PCB (in P/N 15961 harness)
J1211 5-pin (blue wire-top) to the SVGA PCB
J1 5-pin (orange wire) to SVGA PCB (position wire on 2
J1 5-pin (gray-left) to SVGA PCB
J601 5-pin to SVGA PCB
3. Loosen (7) Phillips pan-head screws on the main PCB.
4. Lift the main PCB out of the front panel.
Installation
1. Position the main PCB onto the standoffs in the front panel.
2. Tighten (7) Phillips pan-head screws on the main PCB.
nd
pin from right)
L1534 Revision B December 2004
Page 35
Service Manual Disassembly & Reassembly Procedures 35
3. Connect the following to the main PCB:
J801 20-pin from J801 on the flow sensor PCB (in harness)
J1210 5-pin (gray top, green) from the flow sensor PCB (in P/N 15961 harness)
J3 10-pin from the flow sensor PCB (in P/N 15961 harness)
J1211 5-pin (blue wire-top) from the SVGA PCB
J1 5-pin (orange wire) from SVGA PCB (position wire on 2
nd
pin from right)
J1 5-pin (gray-left) from SVGA PCB
J601 5-pin from SVGA PCB
Note
The J1 position on the main PCB is located above the J601 connector and accepts two 5-pin connectors; the
connector with one wire above the connector with three wires.
4. Install the front panel, left panel, right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 36
36 Vela Ventilators
Figure 3.9 Front Panel wiring diagram
L1534 Revision B December 2004
Page 37
Service Manual Disassembly & Reassembly Procedures 37
Flow Sensor PCB assembly P/N 15991
Figure 3.10 Flow Sensor PCB assembly
Removal
1. Remove the power cable, top cover, battery tray, right panel, left panel, front panel, and main PCB.
2. Disconnect the (2) 1/16” ID tubes to the main tubing harness.
3. Disconnect the following from the flow sensor PCB:
J1 6-pin (grey) to main PCB P/N 15961
4. Remove the (2) 1/4” nuts on the flow sensor PCB.
5. Lift off the flow sensor PCB.
Installation
1. Position the flow sensor PCB onto the two standoffs.
2. Install the (2) 1/4” nuts on the flow sensor PCB.
3. Connect the following to the flow sensor PCB:
J1 6-pin (grey) from the main PCB P/N 15961
4. Connect the (2) 1/16” ID tubes from the main tubing harness, #2 and #3 (with a black line on it).
5. Install the main PCB, front panel, left panel, right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 38
38 Vela Ventilators
Exhalation Valve Assembly P/N 15871
Dryer tube
Figure 3.11 Exhalation Valve Assembly
Removal
1. Remove the power cable, top cover, battery tray, right panel, left panel, and front panel.
2. Disconnect the following from the FIO2 monitoring PCB:
J2 2-pin (clear conn.) to oxygen sensor
J3 2-pin (blue conn.) to oxygen inlet solenoid on blender assembly
J4 2-pin (red conn.) to nebulizer solenoid assembly on upper rear of blender
3. Remove (2 Phillips pan-head screws on the valve cover.
4. Lift out the exhalation valve assembly.
Installation
1. Ensure the dryer tube is securely attached to the nipple on the exhalation valve assembly.
2. Seat the valve onto the standoffs, with the PCB towards the inside of the unit.
3. Install the valve cover using (2) Phillips panhead screws.
4. Connect the following to the FIO2 monitoring PCB:
J2 2-pin (clear conn.) from oxygen sensor
J3 2-pin (blue conn.) from oxygen inlet solenoid on blender assembly
J4 2-pin (red conn.) from nebulizer solenoid assembly on upper rear of blender
5. Install the front panel, left panel, right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 39
Service Manual Disassembly & Reassembly Procedures 39
Inspiratory Hold Solenoid and Check Valve Assembly P/N 11382
Removal
1. Remove the power cable, top cover, battery tray, and
left panel.
2. If the front panel is installed, disconnect the cable from
J302 on the main PCB.
3. Remove (2) Phillips pan-head screws and lift out the
solenoid.
4. Lift out the spring and check valve assembly.
Figure 3.12 Inspiratory Hold Solenoid
Installation
Figure 3.13 Umbrella check valve (P/N 21950)
1. Place the spring (P/N 21971) into the center of the check valve (P/N 21950).
2. Slip the solenoid over the spring using the guide and seat into place.
3. Install (2) Phillips pan-head screws.
4. If the front panel is installed, connect the cable from J302 on the main PCB.
5. Install the left panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 40
40 Vela Ventilators
Manifold Base assembly
Replace three o-rings
Figure 3.14 Manifold base assembly (early model Vela units only),
Nut securing O2
sensorl
bottom panel & O2 hose removed
O2 sensor placed
horizontally for easy
access from rear of
ventilator
Figure 3.15 Manifold Base assembly showing horizontal O2 cell for easy removal.
L1534 Revision B December 2004
Page 41
Service Manual Disassembly & Reassembly Procedures 41
Removal
1. Remove the power cable, top cover, and
battery tray.
2. Carefully turn the unit onto its side.
3. Remove (2) Phillips countersink screws that
hold the safety check valve guard, and remove
the guard.
4. Remove (4) rubber feet.
5. If the left panel is installed, remove (3) Phillips
pan-head screws on the left side of the bottom
panel, and if the front panel is installed,
remove (2) Phillips pan-head screws at the
front of the bottom panel.
Figure 3.15 Removal of safety check valve guard
6. Remove (7) Phillips pan-head screws on the outside of the bottom panel.
7. Gently remove the bottom panel.
8. Remove the three visible o-rings from the underside of the unit, part number 30018, 30023 and
04358see figure 3.X.
9. Pull off the oxygen diffuser tube from the nipple and remove it from the channel base, see figure
3.15
Figure 3.16 Removing the O2 diffuser tube
Installation
1. Carefully turn the unit upside down.
2. Attach the curved portion of the oxygen diffuser tube to the nipple and push into place. Place the
remaining portion of the tube inside the channel base in the underside of the unit.
3. Insert the three o-rings- into the underside of the unit, noting that each o-ring has a different shape.
4. Ensure that all three o-rings are in place and position the bottom panel over the three screw holes
at the front.
L1534 Revision B December 2004
Page 42
42 Vela Ventilators
5. Install (7) Phillips panhead screws on the bottom panel.
6. If the left panel is installed, install (3) Phillips pan-head screws on the left side of the bottom panel,
and if the front panel is installed, install (2) Phillips pan-head screws at the front of the bottom
panel.
7. Install (4) rubber feet.
8. Install the safety check valve guard with (2) Phillips countersink screws.
9. With the unit upright, install the battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 43
Service Manual Disassembly & Reassembly Procedures 43
Oxygen Sensor part number 15972
For early model Vela units do the following:
Removal
1. Remove the power cable, top cover, battery tray, left panel, and bottom panel.
2. Turn the unit on its side.
3. Remove the 1” mounting nut from the manifold base (see figure 3.14) and pull the sensor from the
manifold.
Note
To remove the O2 sensor only when complete disassembly is not required, simply grasp the sensor in
place and unscrew from the retaining nut leaving the nut in place in the base.
Installation
1. Ensure the o-ring is installed into the oxygen senor.
2. With the unit on its side, place the sensor into the manifold and install the 1” mounting nut.
3. Install the bottom panel, left panel, battery tray, top cover, and power cable.
For later model Vela units (usually 1400 series or later) do the following:
Removal
On the bottom rear of the Vela, unscrew the EMI shield containing the foam ambient air filter and lift out
the shield and filter together.
The O2 sensor is visible on the left . Using needle nose pliers, disconnect the O2 filter connector. Pull
firmly straight back on the O2 sensor to remove.
Installation
Seat the new O2 filter and re-connect.
Replace the EMI shield containing the foam ambient air filter and re-attach the screws securing it in
place.
L1534 Revision B December 2004
Page 44
44 Vela Ventilators
Rear Panel part number 15892
Removal
1. Remove the power cable, top cover, battery tray, right panel, and left panel.
2. With the unit upright, disconnect the following from the turbine motor driver PCB:
P2 10-pin to main PCB and power PCB via main harness
3. Cut the two tie wraps that hold the main harness and the tie wrap that holds the blender cable to
the turbine.
4. Turn the unit on its side.
5. Remove (3) Phillips pan-head screws on the rear panel.
6. Slide the fiber optic cable from under the turbine muffler assembly.
7. Remove the rear panel from the manifold base.
Installation
1. Carefully place the unit on its side and run the fiber optic cable under the muffler tubes.
2. Direct the blender nipple on the rear panel through the hole in the manifold base.
3. Install (3) Phillips pan-head screws on the rear panel, ensuring that the blender nipple is centered.
4. With the unit upright, install the blender cable tie wrap to the turbine, and install the two tie wraps
that hold the main harness onto the tie mounts, noting that the tie wraps are positioned where the
wires all run parallel.
5. Connect the following to the turbine motor driver PCB:
P2 10-pin from main PCB and power PCB via main harness
6. Install the left panel, right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 45
Service Manual Disassembly & Reassembly Procedures 45
L1534 Revision B December 2004
Page 46
46 Vela Ventilators
Figure 3.17 Rear Panel views
L1534 Revision B December 2004
Page 47
Service Manual Disassembly & Reassembly Procedures 47
Turbine and Muffler Assembly
Removal
1. Remove the power cable,
top cover, battery tray, right
panel, left panel, and
blender.
2. If the front panel is installed,
disconnect the cable from
J701 on the main PCB,
disconnect the green
oxygen nebulizer tube, and
disconnect the blender
cable.
3. If installed, use a 3/4"
wrench to remove the high
and low pressure oxygen
fittings from the rear of the
unit, noting that the high
pressure fitting is located
above the low pressure
fitting (see figure 3.18).
Figure 3.18 Turbine & Muffler assembly
4. Remove (5) Phillips pan-head screws.
5. Disconnect the gray silicone intake and output
elbows by gently pulling them out from the muffler
tubes.
6. Lift the assembly off the manifold base.
7. Gently pull up the elbows to remove.
Installation
1. Insert the elbows into the manifold base.
2. Set the turbine and muffler assembly onto the
manifold.
3. Install (5) Phillips pan-head screws.
Figure 3.19 High & low pressure
O2 fittings on back panel
4. If removed during this procedure, use a 3/4" wrench to install the high and low pressure oxygen
fittings to the rear of the unit, noting that the high pressure fitting is located above the low pressure
fitting.
5. Connect the gray silicone intake and output elbows by inserting them into the muffler tubes. Push
once in the center of each elbow to properly seat.
6. If the front panel is installed, connect the cable to J701 on the main PCB, connect the green
oxygen nebulizer tube, and connect the blender cable.
7. Install the blender, left panel, right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 48
48 Vela Ventilators
Over Pressure Relief Valve assembly
Figure 3.20 Over Pressure Relief Valve assembly with o-ring
Removal
1. Remove the power cable, top cover, battery tray, right panel, left panel, and front panel.
2. Use a flat screwdriver to pry the valve from its location at the front left side of the manifold base.
3. Remove the o-ring (P/N 08963).
4. Using a flat screwdriver in the middle slot, unscrew the adjuster in the middle of the valve and
remove the two o-rings (P/N 30005 in the relief valve poppet and P/N 30017 on the brass adjuster).
Installation
1. Install the two o-rings (P/N 30005 in the relief valve poppet and P/N 30017 on the brass adjuster).
2. Slide the housing over the poppet and screw clockwise into the housing, tightening with a flat
screwdriver.
3. Slide the valve into place in the manifold base, ensuring that the adjustment screw is in the notch.
CAUTION
Do not adjust the screw in the relief valve, since it must be factory calibrated.
4. Push until the valve is seated into place.
5. Install the front panel, left panel, right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 49
Service Manual Disassembly & Reassembly Procedures 49
Turbine Motor Driver PCB part number 71597
Removal
1. Remove the power cable, top cover, battery tray, right panel.
2. Disconnect the following from the turbine motor driver PCB:
P1 8-pin to the turbine
P2 10-pin to J6 on the power PCB
3. Remove the (2) keps screws.
4. Lift off the PCB.
Note
Do not remove the turbine from the turbine muffler assembly. They are supplied complete.
Installation
1. Position the turbine motor driver PCB on the right side of the turbine assembly, on top of the filter
holder, with the board toward the turbine.
2. Install (2) keps screws.
3. Connect the following to the turbine motor driver PCB:
P1 8-pin from the turbine
P2 10-pin from J6 on the power PCB
4. Install the right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 50
50 Vela Ventilators
Muffler/Filter Assemblies part number 10364
Removal
1. Remove the power cable, top cover,
battery tray, and left panel.
2. Remove (6) Phillips pan-head screws on
the filter cover.
Figure 3.21
3. Remove the cover and the two o-rings (P/N
30020).
O-rings P/N 30020
Figure 3.22
4. Use needle-nosed pliers to pull out both the left
and right filter assemblies. The right assembly
has the filter sock
Figure 3.24
Figure 3.23
Installation
1. Insert the muffler (output) into the left tube (see
picture) and the sock filter (input) into the right tube.
The right and left muffler/filter assemblies are sold
together as part number 10365.
2. Install two o-rings (P/N 30020).
3. Install the cover using (6) Phillips pan-head
screws.
4. Install the left panel, battery tray, top cover, and
power cable.
L1534 Revision B December 2004
Page 51
Service Manual Disassembly & Reassembly Procedures 51
Blender Assembly Part Number 15895
Removal
1. Remove the power cable, top cover, battery tray, and right panel.
2. If the rear panel is installed, remove the high and low pressure oxygen fittings from the rear panel
using a 3/4" wrench, noting that the high pressure fitting is located above the low pressure fitting.
3. If the rear panel is installed, disconnect the 40-pin ribbon cable from the SVGA out PCB.
4. If the front panel is installed, disconnect the green oxygen nebulizer tube.
5. Disconnect the following from the blender PCB:
J301, 12-pin connector to J301 on the main
PCB
6. Remove the bottom strip ambient air inlet filter
by pinching and pulling it out.
Figure 3.25 Ambient air filter
7. Remove (3) Phillips pan-head screws on the right side of the rear panel.
8. Use needle-nosed pliers to disconnect the oxygen diffuser tube as you remove the blender
assembly.
9. If the exhalation valve assembly is installed, disconnect the following from the FIO2 PCB:
J3 2-pin (blue conn.) to oxygen inlet solenoid on blender assembly
J4 2-pin (red conn.) to nebulizer solenoid assembly on upper rear of blender
Installation
1. If the exhalation valve assembly is installed, connect the following to the FIO2 PCB:
J3 2-pin (blue conn.) from oxygen inlet solenoid on blender assembly
J4 2-pin (red conn.) from nebulizer solenoid assembly on upper rear of blender
2. Position the blender assembly inside the unit.
3. Use needle-nosed pliers to feed the oxygen diffuser tube through the hole in the rear panel and
connect it to the blender nipple.
4. Install (3) Phillips pan-head screws on the right side of the rear panel.
5. Install the bottom strip ambient air inlet filter.
6. Connect the following to the blender PCB:
301 12-pin from J301 on the main PCB
7. If the front panel is installed, connect the green oxygen nebulizer tube.
8. If the rear panel is installed, connect the 40-pin ribbon cable to the SVGA out PCB.
9. If the rear panel is installed, install the high and low pressure oxygen fittings to the rear panel using
a 3/4" wrench, noting that the high pressure fitting is located above the low pressure fitting.
10. Install the right panel, battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 52
52 Vela Ventilators
Figure 3.26 Blender Assembly
L1534 Revision B December 2004
Page 53
Service Manual Disassembly & Reassembly Procedures 53
Fan Assembly part number 15500
Fan
EMI Shield Filter holderFilter Filter Grid
Figure 3.27 Fan Assembly blow out
Removal
1. Remove the power cable, top cover, and battery tray.
2. Use a 5/16” nut driver to remove the inside nuts and a short (#2) Phillips screwdriver to remove the
(4) Phillips pan-head outside screws on the fan assembly.
3. Remove the fan and ESD shield from the inside of the unit, and the filter grid from the rear of the
unit.
4. Disconnect the fan cable from J2 on the power PCB.
Installation
1. Place the fan filter inside the external fan cover. Align with 4 attachment holes, and hold it against
the rear panel.
2. Hold the ESD shield to the inside of the rear panel with the copper contact strips touching the rear
panel.
3. Place the fan onto the ESD shield with the label to the inside of the unit and the wire at the bottom.
4. Use a 5/16” nut driver to install the nuts from the inside and a short (#2) Phillips screwdriver to
install the (4) Phillips pan-head machine screws onto the fan assembly from the outside.
5. Connect the fan cable to J2 on the power PCB.
6. Install the battery tray, top cover, and power cable.
L1534 Revision B December 2004
Page 54
54 Vela Ventilators
L1534 Revision B December 2004
Page 55
Chapter 4 Operational Verification and
Calibration
Operational Verification Testing
Prior to using the Vela ventilator on a new patient the following checks should be carried out to ensure
optimum performance. Verification testing should always be performed “off patient”.
WARNING
Disconnect patient from the ventilator before performing verification testing.
Note
All personnel performing preventive maintenance and product repair must betrained and certified by VIASYS
Healthcare to service the product
Note
If any portion of the following performance check fails, and you are unable to correct the problem, contact your
VIASYS Healthcare Certified Service Technician.
First perform the User Verification Tests (UVTs). To access the UVTs, do the following’
1. After disconnecting the patient, turn the ventilator OFF (i.e., STANDBY).
2. Press and hold the Accept button.
3. While holding the Accept button, turn the ventilator ON. Continue to hold the button until the
ventilator completes the Power On Self Tests (POST).
4. Release the Accept button when the UVT menu appears in the screen. The Audible Alarm
sounds. Press Alarm Reset button to clear.
Figure 4.1 UVT Startup Screen
5. Press the Patient Removed touch screen icon, the UVT test selection screen displays (see
figure 4.2).
L1534 Revision B December 2004
Page 56
56 Vela Ventilators
6. Press the appropriate touch screen icon to begin a test.
Lamp Test
Run this test to check the front lamps to make sure they are functioning properly.
1. Press the LampTest touch screen icon to start the test. The ventilator illuminates all front panel
LEDs.
2. Press the LampTest touch screen icon again to turn the LEDs off and exit the test.
Switch Test
Run this test to check the front panel membrane switches to make sure they are working properly.
1. Press the SWITCH TEST icon.
2. Press each membrane switch control in turn. Watch for the name of the control to appear in the
message bar at the bottom left of the touch screen as follows:
Figure 4 2 The UVT Screen with the Main screen in Service mode
L1534 Revision B December 2004
Page 57
Service Manual Chapter 4 Operational Testing and Calibration 57
3. Press the SWITCH TEST icon again to exit the test.
Alarm Test
Run this test to check the audible alarm.
1. Press the Alarm Test touch screen icon to start the test. The audible alarm sounds.
2. Press the Alarm Test touch screen icon again to silence the audible alarm and exit the test.
Filter Test
Run this test to check the differential pressure across the turbine inlet filter. A high differential pressure
may indicate a dirty or occluded filter.
1. Press the Filter Test touch screen icon. The turbine accelerates flow to 140 lpm and checks the
where x.x is the average turbine differential pressure and P indicates the test passed.
If the test fails, the following message appears:
differential pressure to make sure it is within range. At the end of the test, the ventilator displays
a pass or fail message. If the test passes, the following message appears:
x.x Passed
x.x Failed
where x.x is the average turbine differential pressure and F indicates the test failed:
L1534 Revision B December 2004
Page 58
58 Vela Ventilators
2. If the test fails, check the rotation of the turbine inlet filter to make sure that the seam is
horizontally oriented and run the test again. If the test fails again, contact a Bird Certified Service
Technician.
Leak Test
Note
This test should be performed with all circuit accessories installed (e.g., humidifier, water traps, and so on.)
Make sure al connections are secure and all openings occluded before beginning the test.
Run this test to make sure the patient breathing circuit is not leaking.
1. Attach a one-liter test lung (P/N 33754) at the patient breathing circuit wye.
2. Press the Leak Test touch screen icon to run the test. The test begins by increasing the
pressure in the patient breathing circuit to 60 cmH2O. The ventilator then displays the following
messages in sequence:
Leak test requested
Leak test in progress
3. The ventilator then waits 10 seconds and measures the circuit pressure again. If the difference
between the starting and ending measurements is less than or equal to 6 cmH2O, the test
passes and the ventilator displays the following message:
Exit
xx.x Passed
where xx.x is the ending measurement.
4. Otherwise, if the difference is more than 6 cmH2O, the test fails and the ventilator displays the
following message:
xx.x Failed
5. If the test fails, check all connections to make sure there are no leaks and repeat the test.
6. It the test fails again, refer to Chapter 5, for troubleshooting procedures
To exit the UVTs press the EXIT touch screen icon.
L1534 Revision B December 2004
Page 59
Service Manual Chapter 4 Operational Testing and Calibration 59
Extended Functions
The Extended Functions menu may be accessed via the UVT Test Select screen as described in the
previous section. Press the EXTENDED FUNCTIONS icon from the UVT screen.
Figure 4.3 Accessing the Extended Functions Screen from the UVT menu
You can also access the extended functions screen via the Screens menu during normal operation.
Press the screen display area on the top center of the touch screen to activate Screen Select menu.
Figure 4.4 Accessing the Extended Functions screen from the Screen Select menu
Press the EXTENDED FUNCTIONS touch screen icon to display the Extended Functions screen.
L1534 Revision B December 2004
Page 60
60 Vela Ventilators
Figure 4.1 Extended Functions screen menu
Events
To access the Events list, press the Events screen icon
Figure 4.2 Press the Events touch screen icon
A list of chronological events displays with the latest event at the top as shown in figure 4.11
Figure 4.3 The Events screen
L1534 Revision B December 2004
Page 61
Service Manual Chapter 4 Operational Testing and Calibration 61
Transducer Data
Press the Transducer Data screen icon.
Figure 4.4 Press the Transducer Data screen icon
The Transducer Selection screen displays.
Figure 4.5 Transducer data selection screen
From here you can set the analog data display associated with each transducer. For example, if you
press the SET ANALOG 0 screen icon as shown in figure 4.X, the following options display in a drop
down menu.
In this example, the transducer associated with Analog 0 is set to display Turbine Speed.
Set each analog value to one of the transducers in the drop down menu. Press EXIT to return to the
Extended Functions screen.
For instructions on how to view this data as a displayed waveform, see Chapter 5, Maintenance &
Troubleshooting.
Transducer Tests
To access the transducer test screen press the Transducer Tests screen icon.
Figure 4.7 Press transducer tests icon
The transducer tests screen appears with icons for the Turbine Differential, Exhalation Differential and
Circuit Pressure transducers.
Figure 4.8 Transducer test screen
Press the Turbine Differential (TURB DIFF) screen icon.
L1534 Revision B December 2004
Page 63
Service Manual Chapter 4 Operational Testing and Calibration 63
Figure 4 9 Press TURB DIFF icon
The following messages appear in sequence in the message area at the bottom left of the touch
screen.
Turb Diff Pressure Xdcr Test – Requested
Turb Diff Pressure Xdcr Test – In Progess
Turb Diff: XXX Passed (where XXX is the transducer reading within range)
Or
Turb Diff: XXX Failed (where XXX is the transducer reading OUT of range)
Press the Exhalation Differential Pressure (EXHL DIFF) screen icon.
The following messages appear in sequence in the message area at the bottom left of the touch
screen.
Exhl Diff Pressure Xdcr Test – Requested
Exhl Diff Pressure Xdcr Test – In Progess
Exhl Diff: XXX Passed (where XXX is the transducer reading within range)
Or
Exhl Diff: XXX Failed (where XXX is the transducer reading OUT of range)
Press the Circuit Pressure (CIRC PRESS) screen icon.
The following messages appear in sequence in the message area at the bottom left of the touch
screen.
Circ Pressure Xdcr Test – Requested
Circ Pressure Xdcr Test – In Progess
Circ Press: XXX Passed (where XXX is the transducer reading within range)
Or
Circ Press: XXX Failed (where XXX is the transducer reading OUT of range)
L1534 Revision B December 2004
Page 64
64 Vela Ventilators
Version Information
To access the software version information screen, press the VERSION INFO screen icon.
Figure 4.10 Press the Version Info icon
The Versions screen, showing the versions of all software installed on the machine, displays.
Note
The version information shown in figure 4.X is an example only. Your machine will have different versions of
software depending on the date of sale and the updates loaded.
Figure 4 11 Sample Version information screen
L1534 Revision B December 2004
Page 65
Service Manual Chapter 4 Operational Testing and Calibration 65
Date & Time
The date and time can be set and the ventilator hours of use (Hour Meter) can be viewed by pressing
the Date/Time screen icon.
Figure 4.12 Press Date/Time icon
Figure 4.13 Date/Time and Hour Meter screen
To adjust the date and time, select by pressing the touch screen directly over the control to be adjusted
(see figure 4.X).
Figure 4.14 Select the month control
L1534 Revision B December 2004
Page 66
66 Vela Ventilators
Adjust by turning the Data Dial clockwise to increase, or counter clockwise to decrease, the value.
Figure 4.15 Data Dial
FiO2 Calibration
See the Calibration section of this Chapter.
Ventilator Setup
To access the ventilator setup options, press the Vent Setup screen icon.
Figure 4.16 Press Vent Setup icon
L1534 Revision B December 2004
Page 67
Service Manual Chapter 4 Operational Testing and Calibration 67
Figure 4.17 Ventilator Setup Screen
Low Min Volume Off
There are two ranges possible for the Low Minute Volume Alarm. They are OFF to 99.9 liters and 0.1
to 99.9 liters. The Low Min Vol Off Disable/Enable screen icon, allows you to toggle between the two
ranges.
Figure 4.18 Low Min Vol Off Enabled
L1534 Revision B December 2004
Page 68
68 Vela Ventilators
Locks Disabled
Toggles the control Locks on and off.
Figure 4.19 Locks Enable/Disable
FiO2 Monitor Disabled
Disables or Enables the FiO2 Monitor Function.
Figure 4.20 Monitor enable/disable
L1534 Revision B December 2004
Page 69
Service Manual Chapter 4 Operational Testing and Calibration 69
Altitude Adjustment
Allows for adjustment of altitude in feet or meters above sea level.
Figure 4.21 Altitude feet/meters
Figure 4.22 Altutude in Meters
To adjust the value, select the control,
The Altitude screen icon toggles between Feet and Meters
and the selection is reflected in the control label (see figure
4.22)
then adjust using the data dial.
L1534 Revision B December 2004
Page 70
70 Vela Ventilators
Language Selection
A choice of 5 languages is available (see figure 4.23). Select a language by pressing the appropriate
screen icon. Press EXIT to leave the Ventilator Setup screen and return to the Extended Functions
screen
Figure 4.23 Exit Ventilator Setup
Press EXIT again to exit the Special Functions menu
Figure 4.24 Exit Special Functions
The ventilator will begin operation at previously selected settings.
Set up the patient and ventilator operational settings per the instructions your Operator’s Manual.
L1534 Revision B December 2004
Page 71
Service Manual Chapter 4 Operational Testing and Calibration 71
Verification tests
1. Before attaching the Vela to a new patient, perform the following Operational verification checks.
2. Turn on the Vela, choose New Patient and Accept. This will return all settings to the defaults.
3. To check monitor performance, allow the ventilator to operate for two minutes. View the
monitored parameters. The values should appear as follows:
Parameter Value
Minute volume 6 L+ 1.2 L
Tidal Volume 500 ml + 100 ml
I:E Ratio 1:6.1 + 10%
Breath Rate 12 bpm + 2 bpm
PIP Should equal manometer display + 5 cmH2O
MAP N/A
PEEP 5 cmH2O + 2 cmH2O
Inspiratory Time 0.68 seconds + .05 seconds
3. Check the alarms as follows:
a. Power Fail Check
i. Remove the power cord from the wall. The ventilator should do the following:
1. Switch to battery power.
2. Sound the audible alarm.
3. Turn the AC Power Source indicator OFF.
4. Display the BATTERY ON message in the alarm window.
5. LED for internal battery will light.
ii. Press the Alarm Reset button to clear the alarm.
iii. Plug the AC power cord back into the wall socket.
b. High Pressure Limit Check
i. Lower the High Pressure Alarm setting to 5 cmH2O below the Peak Inspiratory
Pressure (PIP). When the ventilator cycles to inspiration and the high pressure
limit is violated, the high pressure alarm should occur. When this happens the
ventilator should:
6. Immediately cycle into the expiratory phase.
7. Sound the audible alarm.
L1534 Revision B December 2004
Page 72
72 Vela Ventilators
8. Display the HIGH PRES message in the alarm window.
ii. Return the High Pressure Alarm setting to 5 cmH2O above PIP, and press the
Alarm Reset button to clear the alarm.
This completes the performance check. The following Ventilator Performance Checklist may be used to
document each Performance Check for your records.
L1534 Revision B December 2004
Page 73
Service Manual Chapter 4 Operational Testing and Calibration 73
Vela Ventilator Performance Checklist
This checklist is for use during the Vela Operational Verification Procedure.
Serial Number_______________ Hours______________ Date_____________
Verification Steps:
Verification Step Check & Initial
1. Inspect the ventilator and components for appearance and cleanliness.
Confirm the Exhalation valve, diaphragm, and air intake filter and test lungs
are correctly installed. Wipe the ventilator clean if needed using a cloth
moistened with an approved cleaning solution.
2. Enter the User Verification Test (UVT). Press the Patient Removed touch
screen icon.
3. Test the alarm volume. Adjust as required.
4. Confirm the proper functioning of the front panel lamps and LEDs.
5. Confirm the proper functioning of the membrane switches.
6. Complete the internal filter performance check.
7. Check the patient breathing circuit on the ventilator and conduct a leak test.
Make sure all needed components are firmly attached in the circuit.
8. Exit the UVT and conduct performance test.
Set the parameters to the following recommended settings:
Mode
Tidal Volume
Peak Flow
A/C
500ml
60 lpm
Inspiratory Pause
Sigh
Square Waveform
OFF
OFF
OFF
Breath Rate
PEEP/CPAP
Sensitivity
Pressure Support
Pressure Control
% O2
12 bpm
5 cmH2O
OFF
OFF
OFF
21
Bias Flow
Altitude
Inspiratory Time
Termination Sensitivity
Variable Time Termination
Flow Termination
10 lpm
Current
0.3 sec
25%
3.0 sec
OFF
L1534 Revision B December 2004
Page 74
74 Vela Ventilators
Set the variable alarm controls as follows:
Alarm Control Setting
High Pressure PIP plus 5 cmH2O
Low Pressure PIP minus 10 cmH2O
Low Minute Volume 4 Liters
High Breath Rate 16 bpm
Apnea Interval 20 secs
Verification Step Check & Initial
9. After at least two minutes of operation compare the displayed readings to the
following:
Service Manual Chapter 4 Operational Testing and Calibration 75
Calibration
Note
All personnel performing preventive maintenance and product repair must betrained and certified by VIASYS
Healthcare to service the product
Test Set-up
To calibration port
To pressure manometer
(0 to 60cmH2O)
To pressure manometer
O-60 psig
Figure 4.34 Callibration setup #1
High pressure gas
source (wall or bottled)
with 0-50 psig regulator
attached
To calibration port
Figure4.35 Calibration setup #2
Power Up Verification/Service Verification Tests
1. Remove the top cover and battery tray as described in Chapter 3.
2. Locate DIP Switch 1 on the Main PCBA on the rear of the front panel. Place it in the ON position,
as shown in figure 3.
L1534 Revision B December 2004
Page 76
76 Vela Ventilators
PT 501
P1P1P2P2
11
TURBINE EEPROM
S/N: XXXXXXXXXXXX
BIRD C 1995
PT 500
TBIRDTM 1995
PAL P/N 15532
3. Connect the Vela to a properly grounded AC source. Turn on power to the unit.
Figure 4.36 Main PCB
MeasurementMod
Turbine
Differential
Transducer
P1 P2 P1 P2
Exhalation
Pressure
Transducer
Exhalation
Flow
Transducer
Overrang
e
Underrang
e
12345678
ON
MANOMETER
GAS PRESSURE
Low
Range
20
cmH2O
250
cmH2O
185 mm
Hg
GAS FLOW RATE
Low
Range
5 lpm
2(%)/N2
O
5 lpm
Air
5 lpm
2(%)/N2O
O
e
PRESSURE
High
Range
20
psi
100
psi
5170
mmHg
High
Range
5 lpm
2(%)/N2
O
180 lpm
Air
20 lpm
2(%)/N2O
O
P-
VOLUMERATE/TIME
3 liters
2(%)/N2
O
3 liters
Air
Minute
2(%)/N2O
O
Lo
wPressur
eRang
e
4 psi
maxdifferenti
al
789
123
0
F
1
P
+
15
psima
x
Rate
BPM
inspirator y
T
expiratory
T
I:E
Ratio
8
7
6
5
4
3
2
1
ON
Zer
o
Peak
654
/Continuou
s
Rese
t
Ente
r
F
2
Hig
hPre ssur
eRang
P
P-
e
+
100 psi
maxdifferenti
al
100
psima
x
Lo
Hig
wFlo
hFlo
wRang
wRang
e
e
Clean,
Clean, dry
drygas
gasonl
only5
y180
lpmmax
lpmmax
.
.
Attach to the
tran sducer needin g
P/N 33903
calibration
Figure 4.37 Location of transducers
4. The ventilator performs its Power On Self Test. When the self-test is complete, the screen displays
“flat” waveforms and the ventilator makes no attempt to deliver a breath. This indicates that you are
in a “Service” mode
5. The audible alarm should sound, press the alarm reset button to silence alarm.
6. Press the MAIN icon at the top of the touch
screen
7. The SVT TEST SELECT screen displays.
8. Press CALIBRATION to access the
transducer calibration screen (see figure 4.7)
Figure 4.38 SVT Test Select Screen
L1534 Revision B December 2004
Page 77
Service Manual Chapter 4 Operational Testing and Calibration 77
Figure 4.39 Calibration selection screen
Exhalation Pressure Transducer
1. Press the EXH PRESS XDCR touch screen icon from the CALIBRATION SELECT Screen. The
EXHL PRESS CALIB screen displays showing both pressures as “Pending” and with the
instruction “Apply 0 cmH2O and press button”.
2. Disconnect tubing from the exhalation pressure transducer before continuing with the procedure
(see figure 4.6 for location of the transducers ).With no tubing attached to the exhalation pressure
transducer, press the PRESSURE APPLIED touch screen icon for zero calibration.
3. When the calibration is complete, the screen legend changes to read “Apply 60 cmH2O, Press
Button” and the zero pressure value now says “OK” with an A/D value added.
8. When the calibration is completed, the screen displays OK for both calibrations. SAVE
CALIBRATION and CANCEL CALIBRATION. Press SAVE CALIBRATION to save and return to
the transducer calibration selection screen.
CAUTION
Make sure that your pneumatic connection is to the Exhalation Pressure transducer before applying 60 cmH2O
pressure. If the 60.0 cmH2O pressure is inadvertently applied to the Exhalation Differential (Flow) transducer it
will permanently damage the transducer.
Turbine Pressure Transducer
1. From the CALIBRATION SELECT screen, press TURB PRESS XDCR.
2. The first Turbine Pressure calibration screen displays showing both calibration pressures as
pending and with the legend “Apply 0 cmH2O and press button”.
Figure 4.43 Turbine Pressure Zero calibration screen #1
2. Disconnect tubing from the transducer before continuing with the procedure. See figure 4.6 for
location of the transducers
3. With no tubing attached to the turbine pressure transducer, press the PRESSURE APPLIED touch
screen icon for zero calibration.
4. The screen legend changes to read “Apply 60 cmH2O, Press Button” and the status of the zero
calibration line now reads “OK” with a recorded A/D value.
L1534 Revision B December 2004
Page 79
Service Manual Chapter 4 Operational Testing and Calibration 79
9. Reconnect tubing to the port after calibration.
CAUTION
Make sure that your pneumatic connection is to the Turbine Differential transducer before applying 60 cmH2O
pressure. If the 60.0 cmH2O pressure is inadvertently applied to the Exhalation Differential (Flow) Transducer it
will be permanently damaged and require replacement..
L1534 Revision B December 2004
Page 80
80 Vela Ventilators
Exhalation Flow Transducer Calibration
1. From the CALIBRATION SELECT screen, press EXL FLOW XDCR. The first Exhalation
Differential Pressure (Flow) calibration screen displays.
2. Disconnect tubing from the transducer before continuing with the procedure. See figure 4.6 for
location of the transducers
3. With no tubing attached to the transducer, press the PRESSURE APPLIED touch screen icon for
zero calibration.
4. When the calibration is complete, the status of the zero pressure line reads “OK” and an A/D value
is recorded. The screen legend changes to read “Apply 3 cmH2O, Press Button.”
5. Attach the calibration setup shown in figure 1 to the Exhalation Differential (Flow) Transducer port
P2 (see figure 4.6 for location).
6. Turn the calibration syringe handle clockwise to generate positive pressure.
7. When the pressure manometer reads 3 cmH2O +/- 0.6 cm H2O, press the PRESSURE APPLIED
touch screen icon.
8. When the calibration is complete, the status of the zero pressure line reads “OK” and an A/D value
is recorded. The screen displays SAVE CALIBRATION and CANCEL CALIBRATION. Press SAVE
CALIBRATION to save and return to the transducer calibration selection screen.
L1534 Revision B December 2004
Page 81
Service Manual Chapter 4 Operational Testing and Calibration 81
9. Remove calibration setup from exhalation differential pressure (flow) transducer port and reconnect
tubing before continuing.
Oxygen Pressure Transducer Calibration
1. From the CALIBRATION SELECT screen, press O2 PRESS. The Oxygen Pressure Transducer
Calibration screen #1 appears.
Figure 4.49 O2 Pressure calibration screen #1
2. With no tubing attached to the Vela, press the PRESSURE APPLIED touch screen icon for zero
calibration.
4. When the calibration is complete, the status of the zero pressure line reads “OK” and an A/D value
is recorded (see figure 4.19). The screen legend changes to read “Apply 50 cmH2O, Press Button.”
L1534 Revision B December 2004
Page 82
82 Vela Ventilators
Figure 4.50 O2 Pressure calibration screen #2
5. Attach the calibration setup tube to the
nebulizer port on the front of the Vela (see
figure 4.20).
6. Press the Nebulizer membrane button on the
front panel to open the solenoid.
7. Apply gas from the high pressure gas source
at 50 psig pressure.
8. When the pressure manometer reads 50
psig, press the PRESSURE APPLIED touch
screen icon.
9. When the calibration is complete, the status
of the 50 psig pressure line reads “OK” and
an A/D value is recorded (see figure
4.21).The screen displays SAVE CALIBRATION and
CANCEL CALIBRATION. Press SAVE CALIBRATION to
save and return to the transducer calibration
selection screen.
Figure 4.51 Nebulizer port on the Vela
Figure 4.52 O2 Pressure calibration screen #3
L1534 Revision B December 2004
Page 83
Service Manual Chapter 4 Operational Testing and Calibration 83
psig
p
Regulator calibration:
Connect to high
High pressure gas
source (wall or bottled)
with 0-50 psig regulator
attached
pressure inlet on rear
Measure pressure at
Nebulizer
ort on Vela
Figure 4.53 Calibration setup for Regulator Calibration
To Pressure
Manometer 0-60
1. Enter the SVT mode as detailed in this procedure then press the MAIN icon. From the
CALIBRATION SELECT screen, press the SOLENOID touch screen icon to enter the solenoid
mode.
2. Connect gas source to the high-pressure port on the rear panel (see figure 18). Using a 0-50 psig
regulator, apply 50 psig gas pressure. In turn, press the O2-0, O2-1, and O2-2 solenoid touch
screen icons.
3. Using a calibrated manometer, measure the pressure at the Nebulizer port on the front of the
ventilator for each solenoid (see figure 17). It must read 40 psig +/- 1. Adjust the internal adjustment regulator to achieve the correct pressure at the Nebulizer port (see figure 4.22)
High Pressure
O2 Port
Figure 4.54 Vela Rear Panel
3. Perform the Operational Verification Procedure described in this chapter once the calibration is
complete. .
L1534 Revision B December 2004
Page 84
84 Vela Ventilators
Note
All personnel performing preventive maintenance and product repair must betrained and certified by VIASYS
Healthcare to service the product
Service Manual Chapter 4 Operational Testing and Calibration 85
Figure 4.57 Location of O2 Blender solenoids
L1534 Revision B December 2004
Page 86
86 Vela Ventilators
Oxygen Sensor Calibration
From the Special Functions menu, press the FIO2 MON CALIBRATION touch screen icon
Figure 4.58 Special Functions menu,
The following screen will appear.
Figure4.59 FiO2 Sensor Calibration Ambient Air
Attach 100% O2 at 50psig to the high-pressure inlet on the rear of the Vela.
Press the START AMB CALIBRATION touch screen icon. The machine will take 3-5 minutes to
perform the calibration and the following screen will appear while the calibration is in progress.
Figure 4.60 Ambient Calibration in progress.
L1534 Revision B December 2004
Page 87
Service Manual Chapter 4 Operational Testing and Calibration 87
When calibration is complete, the following screen will appear showing the new ambient value.
Figure 4.61 Ambient calibration complete showing new value
Press the START 100% CALIBRATION touch screen icon. The calibration will take 3-5 minutes during
which time the following screen will appear.
Figure 4.62 100 %FIO2 calibration in progress
When the calibration is complete the following screen will appear.
Figure 4.63 FIO2 calibration complete, new values showing.
If the calibration values are acceptable, press EXIT to exit the calibration sequence.
L1534 Revision B December 2004
Page 88
88 Vela Ventilators
L1534 Revision B December 2004
Page 89
Chapter 5 Maintenance & Troubleshooting
Routine Maintenance Procedures
Replacing the software PCMCIA cards
Occasional software upgrades may require replacement of the internal PCMCIA card that drives the
Vela. This is easily access by removing the screws holding the ventilator cover and sliding the cover off
towards the rear as shown in figure 5.1.
Figure 5.1 Slide cover towards rear
CAUTION
The SVGA Output ribbon cable lies directly beneath the top cover assembly running from the front of the Vela to
the rear. It can be damaged if care is not taken when removing the cover. Make sure that the cover clears this
cable before sliding it back.
The card is located towards the front of the Vela beneath the right panel. To replace the card, refer to
figure 5.2 and do the following:
Card eject button
Figure 5.2 PCMCIA cards in place
L1534 Revision B December 2004
Page 90
90 Vela Ventilators
Note
If a new version of software is installed, it can take up to two minutes for the software to download to the
ventilator. During this time the ventilator is inoperative and the audible alarm will sound. The ventilator will
resume normal operation once the download is complete.
1. Press the protruding button beneath the card to eject.
2. To install, make sure the arrow on the card is pointing up then push the card into the
applicable slot until it clicks into place.
WARNING
It is essential when working inside the ventilator to protect against static discharge, which can damage the
ventilator and cause it to become inoperative. ALWAYS take standard ESD precautions before removing the
ventilator cover.
Replacing the external A/C fuses.
Remove the power cord protective cover. Unplug the
A/C power cord.
Figure 5.3 Remove the power cord guard
The AC power module is a universal module for A/C
voltages from 100-240 volts. Using a flat bladed
screwdriver, pry open the cover. Using the same
screwdriver, loosen and pull out the red fuse holder as
shown in figure 5.4.
Remove the fuses from both sides of the fuse holder
and replace with fuses P/N 71612(available from
VIASYS Healthcare customer support.
Figure 5.4 Pry out the fuse holder
WARNING
It is important that the fuses are replaced with the same type and value as those removed. Failure to do so can
result in ventilator malfunction.
L1534 Revision B December 2004
Page 91
Service Manual Chapter 5 Maintenance & Troubleshooting 91
To insert the fuse holder containing the new fuses, make sure
the 4 metal tabs are facing up as shown in figure 5.5 and
carefully push the fuse holder in to the A/C power module until it
seats
Figure 5.5 Insert fuse holder with metal tabs showing
The Annual 5000 hour P.M. Procedure
Replacing the ambient air filter.
The Ambient Air Filter is located at the rear of the ventilator below the rear panel. It is a high-density
foam insert. To remove it, pinch and pull away from the ventilator as shown in figure 5.6. The filter will
lift out. Replace with a clean filter. To prevent the entry of dust into the ventilator, press the filter in
around the edges to ensure a snug fit.
Figure 5.6 Remove high density foam ambient air fIlter
Replacing the fan filter
Pop out the fan filter retaining grid, using screwdriver.
Pinch & pull out the foam filter (see figure 5.7)
To install a new filter, carefully place the new filter over
the fan housing and place the retaining grid back onto
the fan housing. Check to make sure there are no
creases or folds in the foam filter as this will reduce its
efficiency and could allow contaminants into the
ventilator body. Push the retaining grid until all four
“tabs” click into place.
Figure 5.7 Pinch and remove foam fan filter
L1534 Revision B December 2004
Page 92
92 Vela Ventilators
Replacing the filter, retaining ring & check
valve on the low pres O2 fitting
Using a ¾ inch open-ended wrench, unscrew the lowpressure oxygen fitting from the back panel as shown in
figure 5.8.
Figure 5.8 Unscrew the low pressure O2 fitting
Remove the white cone filter and dislodge & remove the white Delrin
retaining ring (use needle nose pliers if needed).
Using needle nose pliers, the black duck billed check valve should come
out easily.
Figure 5.9 White Delrin retaining washer
and duck billed check valve
To replace, place the black duck billed check valve and insert it into the fitting. Tap lightly with the end
of a screwdriver to make sure it is seated then place the Delrin retaining ring over it with the “lip” facing
into the Vela as shown in figure X.X. Place the new cone filter onto the Delrin ring and push gently until
you are sure it has seated correctly. Replace the fitting and tighten down with the ¾ inch wrench.
Replacing the filter, retaining ring and check
valve on the hi pressure fitting
Using a ¾ inch open-ended wrench, unscrew the twopiece high-pressure oxygen fitting. Remove the white
cone filter, the white Delrin washer & the black duck
billed check valve as described above.
Install the black check valve and make sure it is seated. Install the
white Delrin washer with the lip down as shown in figure 5.10.
Place the cone filter on the delrin washer and push gently to seat.
Replace and tighten down the fitting with a ¾ inch open-ended
wrench.
Figure 5.10 Installing the Delrin Washer
Figure 5.9 Unscrew the high pressure O2 fitting
L1534 Revision B December 2004
Page 93
Service Manual Chapter 5 Maintenance & Troubleshooting 93
Replacing the muffler and muffler filter assemblies
Note
Always take precautions against static discharge when working with the
ventilator. Retain all hardware for re-assembly.
From the back panel, remove the power cord guard by
unscrewing the two retaining screws as shown in figure
5.11. Unplug the power cord.
Unscrew the screws holding on the Vela cover. Slide cover towards the rear of the machine and off.
Figure 5.11 Removing the power cord guard
Figure 5.12 Slide cover off towards the rear
Remove the 4 screws securing the battery tray & lift it, and the attached large capacitor assembly out.
Lean the battery tray and capacitor assembly against the left side of the Vela while you disconnect the
large white Molex DC power connector. Set the complete assembly aside.
Note
When the batteries are disconnected, the system will automatically re-set the battery status memory and will
initiate an 18-hour re-charge cycle upon re-connect. During this period, the red DC status light will remain on.
If the DC status light remains lit after the initial charging period has expired, contact VIASYS tech support as
described in Appendix A for a replacement battery.
Remove the right panel. There are two screws in the rear of the ventilator, one in the front top and one
on the lower left side in the center.
Pull the panel out from the rear first, then ease to the back & tilt out. Lay the assembly flat. The
Mufflers are now accessible.
L1534 Revision B December 2004
Page 94
94 Vela Ventilators
Replacing the muffler/filter assemblies.
Silicone Elbow
To remove & replace the mufflers and filter, first ease out the two
silicone elbows from the muffler assembly front panel.
Loosen the 6 screws on the front plate,
& remove the front plate. Remove and
replace the two black o-rings on the
front panel.
Using needle nose pliers, firmly grasp the metal
grid inside the R side muffler/filter assy. Pull
firmly towards you and the muffler/filter should
slide out. Repeat with the left side muffler.
You will see that only the Right side muffler has a filter attached,
the L side has wire muffler only.
L1534 Revision B December 2004
Page 95
Service Manual Chapter 5 Maintenance & Troubleshooting 95
To install the new filter/muffler
assemblies, push the left side
muffler (with no filter) firmly in and
make sure it is pushed all the way
into the housing. Insert the right
side muffler with the filter, ensure
that the seam on the filter lays
horizontally and press home.
Replace the front cover with the
new O-rings and reattach with six
screws.
Install the two silicone elbows into the muffler
cover and push in the center to make sure
they are seated.
Replacing the left panel w/power PCB.
1. Lean the right panel up to the Vela, slide in the front then push in the rear. Reattach the screws
on the rear (2), front (1) and at the lower edge of the right side (1).
2. To install the battery tray, re-attach the large white Molex connector and make sure it lays
between the muffler elbows as you replace the battery tray onto the Vela. This will ensure that
there is enough space for the large capacitor attached to the Battery tray. Re-attach the 4
screws holding the battery tray in place. Reconnect the ribbon cable.
3. Slide the cover in place from the rear and push the top front edge home under the lip. Attach the
screws.
4. Reattach the power cord and re-attach the power cord guard with the two screws.
L1534 Revision B December 2004
Page 96
96 Vela Ventilators
Troubleshooting
If The Ventilator Does not Turn ON
If you turn the power switch ON and the ON indicator does not illuminate, perform the troubleshooting
procedures given in Table 5.1.
Table 5.1
Troubleshooting Power-Up Problems
PROBLEM POSSIBLE CAUSE ACTION
Ventilator plugged into an AC
source but does not power up.
Ventilator attached to alternate
external DC power source but
does not power up.
No power at AC outlet, or the
AC/DC select switch is in the ALT
PWR SOURCE position, or the AC
Line Voltage switch is set to the
wrong voltage.
If the external source is a battery,
the battery may not be charged, or
the AC/DC select switch may be in
the AC position.
Try connecting to a known good AC
power source. Make sure the AC/DC
switch is in the AC position. Make
sure, the voltage setting of the
ventilator matches the voltage of your
power source. Check the fuse
assembly if the ventilator still does not
power up, Contact your Bird Products
Certified Service Technician. Check
the DC voltage output supply from
transformer at J9. Check DC voltage
at J3. Black is ground. Red is 5 volts
DC and Yellow is 14.7 to 20 DC. If
voltage is not present, replace Power
Supply PCB.
Plug the ventilator into a known good
AC source, or to a known good
battery and see if it powers up. If
using a 12VDC power source, set the
AC/DC select switch to the ALT PWR
SOURCE position. Check the fuse
assembly. If the ventilator still does
not power up, contact your Bird
Products Certified Service
Technician. Check DC voltage at J3.
Black is ground. Red is 5 volts DC
and Yellow is 14.7 to 20 DC. If
voltage is not present, replace Power
Supply PCB.
L1534 Revision B December 2004
Page 97
Service Manual Chapter 5 Maintenance & Troubleshooting 97
Reviewing the Transducer Waveforms
To review the transducer waveforms set in Chapter 4, in normal operational mode press the touch
screen immediately over the waveform heading display on the MAIN screen.
Figure 5.13 Select waveform displayed
A scrolling drop down menu displays all the possible waveform choices. To scroll through the choices
use the Data Dial.
Figure 5.14 Drop down menu
From the choices displayed, select the transducer you wish to review (in this case, transducer 0) and
press the Accept membrane button.
Figure 5.15 Selecting Xducr
L1534 Revision B December 2004
Page 98
98 Vela Ventilators
The waveform immediately changes to display the signal associated with the selected transducer (in
this case it is Turbine Speed).
Figure 5.16 Turbine Speed waveform displayed
L1534 Revision B December 2004
Page 99
Appendix A Contact & Ordering Information
How to Call for Support
To get help on performing any of the preventive maintenance routines, or to request service on your
ventilator, contact VIASYS Healthcare Customer Care:
Technical Support
Hours: 7:00 AM to 3:30 PM (PST) Monday through Friday
Phone: (760) 778-7200
Fax: (760) 778-7377
VIASYS Healthcare Customer Care Helpline
Hours: 24 hours, seven days a week
Phone: (800) 934-2473 (From within the US)
Fax: (760) 778-7377
VIASYS Healthcare Critical Care
1100 Bird Center Drive
Palm Springs, CA 92262-8099
U.S.A.
Phone: (760) 778-7200
(800) 328-4139
Fax: (760) 778-7274
L1534 Revision B December 2004
Page 100
100 Vela Ventilators
Ordering Parts
To obtain Vela ventilator parts contact customer service at:
Customer Service:
Hours: 7:00 Am to 3:30 PM (PST)
Monday through Friday
Phone: (800) 328-4139
(760) 778-7200
Fax: (760) 778-7274
L1534 Revision B December 2004
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.