Viasys Advance Infant Flow System User manual

Operator’s Manual
Infant Flow
Advance™ System
© Copyright 2004, VIASYS Healthcare Critical Care
777077-101 Revision B June 2004
2 Infant Flow Advance™ System
Revision History
Date Revision Pages Changes
December 2003 A All Release
June 2004 B All Release manual in VIASYS Healthcare
template using VIASYS Healthcare Critical Care nomenclature. Add Appendix C, Abdominal Respiratory Sensor Placement. Revise part number list in Appendix B.
777077-101 Revision B June 2004
Operator’s Manual 3
Contact and Ordering Information
United States, Latin America, Asia Pacific:
Sales, Service and Clinical Support: VIASYS Healthcare Critical Care Division 22705 Savi Ranch Parkway Yorba Linda, CA 92887 Phone: (714) 283-8444 (800) 381-3552 Fax: (714) 283-8493
www.VIASYShc.com
United Kingdom:
Sales, Service and Clinical Support: VIASYS Healthcare 3 Welton Rd. Warwick, CV34 5PZ Phone: 01926 490888 Fax: 01926 402262
Europe Sales and Customer Service Technical Service
VIASYS Healthcare VIASYS Healthcare Critical Care Division Leibnizstrasse 7 22705 Savi Ranch Parkway D-97204 Hoechburg Yorba Linda, CA 92887 Germany Phone: (714) 283-8444 Phone +49 (0) 931 4972 – 0 (800) 381-3552 Fax:+49 (0) 931 4972 –423 e-mail: Support.CC.EU@VIASYShc.com website: www.VIASYShc.com
777077-101 Revision B June 2004
4 Infant Flow Advance™ System
CAUTION
Federal law (USA) restricts this device to sale by or on the order of a physician.
CAUTION
Not suitable for use in the presence of flammable anesthetics.
CAUTION
Always read the Operator’s Manual before applying treatment.
CAUTION
The Infant Flow Advance™ has been designed and tested as a complete system using Infant Flow™ accessories. Use only approved accessories (Refer to Appendix B for a list of approved accessories).
CAUTION
Service and/or repair of this instrument is restricted to VIASYS Healthcare authorized or VIASYS Healthcare Trained Personnel only.
777077-101 Revision B June 2004
Operator’s Manual 5
Warranty
The Infant Flow Advance is warranted to be free from defects in material and workmanship and to meet the published specifications for One (1) year from date of shipment.
The liability of VIASYS Healthcare, Critical Care Division, (referred to as the Company) under this warranty is limited to replacing, repairing or issuing credit, at the discretion of the Company, for parts that become defective or fail to meet published specifications during the warranty period; the Company will not be liable under this warranty unless (A) the Company is promptly notified in writing by Buyer upon discovery of defects or failure to meet published specifications; (B) the defective unit or part is returned to the Company, transportation charges prepaid by Buyer; (C) the defective unit or part is received by the Company for adjustment no later than four weeks following the last day of the warranty period; and (D) the Company’s examination of such unit or part shall disclose, to its satisfaction, that such defects or failures have not been caused by misuse, neglect, improper installation, unauthorized repair, alteration or accident.
Any authorization of the Company for repair or alteration by the Buyer must be in writing to prevent voiding the warranty. In no event shall the Company be liable to the Buyer for loss of profits, loss of use, consequential damage or damages of any kind based upon a claim for breach of warranty, other than the purchase price of any defective product covered hereunder.
The Company warranties as herein and above set forth shall not be enlarged, diminished or affected by, and no obligation or liability shall arise or grow out of the rendering of technical advice or service by the Company or its agents in connection with the Buyer's order of the products furnished hereunder.
Limitation of Liabilities
This warranty does not cover normal maintenance such as cleaning, adjustment or lubrication and updating of equipment parts. This warranty shall be void and shall not apply if the equipment is used with accessories or parts not manufactured by the Company or authorized for use in writing by the Company or if the equipment is not maintained in accordance with the prescribed schedule of maintenance.
The warranty stated above shall extend for a period of One (1) year from date of shipment, with the following exceptions:
1. Components for monitoring of physical variables such as temperature, pressure, or flow are warranted for ninety (90) days from date of receipt.
2. Elastomeric components and other parts or components subject to deterioration, over which the Company has no control, are warranted for sixty (60) days from date of receipt.
3. Internal batteries are warranted for ninety (90) days from the date of receipt.
The foregoing is in lieu of any warranty, expressed or implied, including, without limitation, any warranty of merchantability, except as to title, and can be amended only in writing by a duly authorized representative of the Company.
777077-101 Revision B June 2004
6 Infant Flow Advance™ System
777077-101 Revision B June 2004
Operator’s Manual 7
Contents
Revision History....................................................................................2
Contact and Ordering Information ...............................................................................3
Warranty .........................................................................................................................5
Limitation of Liabilities....................................................................................................5
Contents .........................................................................................................................7
Chapter 1: Product Description ...........................................................9
Applications .................................................................................................................10
Intended Use ................................................................................................................10
Chapter 2: Product Specification.......................................................11
ETL Classification........................................................................................................13
EN 60601-1 Classification ...........................................................................................13
Chapter 3: Summary of Warnings and Cautions..............................15
Terms ............................................................................................................................15
Chapter 4: Operating Instructions.....................................................21
Step by Step Instructions............................................................................................22
Chapter 5: Clinical References...........................................................31
Chapter 6: Routine Maintenance........................................................33
Oxygen Analyzer Calibration......................................................................................33
Cleaning........................................................................................................................34
Battery Maintenance....................................................................................................34
Chapter 7: Summary of Symbols.......................................................37
Appendix A: Screen Flow Diagrams.................................................39
Appendix A: Screen Flow Diagrams.................................................39
Appendix B: Approved Accessories .................................................41
Appendix C: Abdominal Respiratory Sensor Placement.................43
Glossary of Terms...............................................................................45
777077-101 Revision B June 2004
8 Infant Flow Advance™ System
777077-101 Revision B June 2004
Operator’s Manual 9
Chapter 1: Product Description
TM
The Infant Flow Advance Analyzer, Flow meter and Pressure Manometer with integral alarm systems designed specifically to be used in conjunction with the Infant Flow generator, and patient interfaces consisting of a selection of masks and prongs. Together these products make up the Infant Flow Advance System.
The Infant Flow Advance Driver is assembled in a sheet metal enclosure from standard and commercial components. These are a Medical Air/Oxygen Blender, a Flow meter calibrated for mixed gas between 0 and 15 L/min, a galvanic Oxygen Fuel Cell with associated electronics, a self activating electronic pressure measuring and limiting system and a power supply.
The Infant Flow Advance Driver is supplied with a pole mount bracket to fit round or square section poles from 10 to 35 mm (0.4 to 1.4 in) diameter.
The Medical Air and Oxygen gas inlets are standard NIST, DISS or other appropriate inlets. High-pressure Medical Air and Oxygen hoses must be ordered separately.
Driver is a combined Air/Oxygen Mixer, Oxygen
TM
nasal CPAP circuits,
The patient outlet is an ISO standard 15 mm female taper fitting and the pressure monitoring inlet is a standard Luer taper fitting. The Oxygen monitor is an integral part of the driver and does not require any additional external fittings.
The Oxygen monitor uses a galvanic fuel cell sensor which is fitted in a pressure controlled path to avoid variations in delivered gas pressure altering the measured value. The gas which is used for the Oxygen analyzer is subsequently vented from the rear of the Infant Flow Driver.
The Infant Flow Advance™ Driver is designed to supply a periodic or patient­triggered additional flow of oxygen-enriched air to the patient, and to provide an apnea monitoring and alarm facility when used in conjunction with the appropriate accessories.
The Infant Flow Advance™ Driver can be used in the following modes:
Nasal CPAP, via the Infant Flow generator, with or without apnea monitoring using a respiration sensor that is attached to the patient’s abdomen.
Pressure Assist (PA), whereby an elevated pressure is intermittently delivered to the patient via the Infant Flow™ generator. This can be applied at a variable rate (R) and for varying lengths of time (Ti) and with or without apnea monitoring using a respiration sensor that is attached to the patient’s abdomen.
Triggered PA (trPA), whereby an elevated pressure is intermittently delivered to the patient via the Infant Flow™ generator. This is triggered by the patient’s own respiratory effort by using a respiration sensor that is attached to the patient’s abdomen and for varying lengths of time (Ti). Apnea monitoring is a feature of this mode and a back-up rate (Rb) is available.
777077-101 Revision B June 2004
10 Infant Flow Advance™ System
Applications
Applications are prescribed by the clinician depending on the needs of the patient.
Intended Use
The Infant Flow Advance System consisting of a driver and generator plus nasal CPAP prongs and masks, is intended for the provision of periodic and patient triggered bi-Level CPAP. The system is for use in hospitals, hospital-type facilities and intra-hospital transport environments.
777077-101 Revision B June 2004
Operator’s Manual 11
Chapter 2: Product Specification
This manual describes the operation and routine maintenance of the M674 models of the Infant Flow Advance™ Driver when fitted with Version 1.00 software and above. Identification of the particular version of software fitted in the equipment may be made at power-up. When the Infant Flow Advance Driver is turned on, the 7 segment displays show first three "8"s, then blanks, then the software version number as part of the start-up procedure.
Gas supply - Nominal 50 PSI (3.5 bar) clean, dry Medical Air and Oxygen.
Range - Minimum 30 psig (2.1 bar), maximum 80 psig (5.6 bar). Maximum
differential pressure 30 psig (2.1 bar).
Power Supply – 95-135 Vac, 0.30A, 50-60 Hz or 200-265 Vac, 0.15A, 50-60 Hz. Internal lead acid battery (2 hours running time when fully charged).
CAUTION
Equipment must be used with approved power supplies. Refer to Appendix B for a list of approved accessories.
Power and Battery status indicators – indicates connection to an external power source and the battery status.
- Top indicator shows green when an external 12V DC power source is
connected and indicates the battery is being charged.
- The full battery indicator shows green when the battery is fully charged.
- The low battery indicator shows red when the battery is low. With this
LED illuminated, approximately 15 minutes of power remain before a complete discharge of the battery occurs.
Dimensions: 11in (27.5 cm) x 8.5 in (22.25 cm) x 5.5 in (13.75 cm) (excluding gas inlets, patient outlets and mounting bracket)
Weight: 16 lbs (7.2 kg)
Air/Oxygen Mixer - Range 21 to 100% Oxygen, accuracy ± 3% of selected
output
Flow meter - Range 0 to 15 L/min, accuracy ± 5% of selected output
Pressure relief valve - 2 systems incorporated
- Patient safety - automatic electronic valve system pre-set to vent to
ambient at 11 cmH2O.
- System and delivery circuit safety - mechanical internal relief valve pre-
set at 205 cmH2O.
Manometer - Range, 0 to 12 cmH
Oxygen Monitor – Range, 0 to 100% Oxygen, accuracy ± 2% of span.
777077-101 Revision B June 2004
O, accuracy ± 1 cmH2O.
2
12 Infant Flow Advance™ System
Alarm System - Four separate alarm systems are provided all of which are automatic. The electronic alarms are set after 2 minutes of operation without operator intervention although the operator can manually set or reset them if required.
1 Supply gases failure: If the differential pressure between the two inlet
gases falls outside of the limit of 20 PSI (1.4 bar) or one gas fails completely, an alarm will sound and the gas at the higher pressure only will be delivered to the patient.
2 High patient pressure: An audible and visual high pressure alarm is
pre-set at 11 cmH relieving solenoid which instantly reduces the pressure in the patient circuit to near zero. The pressure is restored after 3 seconds, but will be reduced to near zero should the cause of the alarm condition still exist. A second high pressure alarm with audible and visual indication is set 3 cmH
O above the measured CPAP pressure.
2
3 Low patient pressure: An audible and visual low pressure alarm is set
at 2 cmH
O below the measured CPAP pressure or at 0 cmH2O if this
2
would otherwise be negative.
4 Failure to deliver correct Oxygen concentration: Audible and visual
alarms are provided at ± 5% of the measured FiO of the alarms with an upper maximum limit of 101% and a lower minimum limit of 20%. There is a low hazard warning at 18% Oxygen or below. Additionally, the Infant Flow Advance has the following characteristics:
O. This alarm automatically activates a pressure
2
at the time of arming
2
Variable augmented flow: 0 to 5 L/min of the mixed gas
Pressure display with time: 0 to 10 cm H
O; accuracy: 10% of full
2
scale
Timing accuracy: 1% of setting
Input connection: 8-way self-locking socket for Transducer Interface
Apnea alarm indicator (red)
Breath indicator (yellow) – indicates inspiration when the Transducer
Interface and abdominal respiratory sensor are properly fitted.
Environment: Keep dry and do not expose to direct sunlight.
- Temperature Operating: 50 to 104 ºF (10 to 40 ºC)
Storage: 32 to 122 ºF (0 to 50 ºC)
- Humidity Operating : <90% non condensing Storage: <90% non condensing
- Atmospheric Pressure Storage: 8.7 to 20.3 PSI (0.6 to 1.4 bar)
777077-101 Revision B June 2004
Operator’s Manual 13
ETL Classification
With respect to applicable requirements of Standard for Safety Medical Electrical Equipment, UL 60601-1 1 General Instructions No. 1, CAN/CSA C22.2 No. 601.1-M90.
st
Ed. 04/25/03 and General Requirements for Safety
EN 60601-1 Classification
Equipment is Class 1 and internally powered, IPX0 Protected, and uses Type B and Type BF applied parts.
Equipment not suitable for use in presence of flammable anesthetics.
NOTE
For use of Model M674A in Australia and New Zealand, Power supply Part Number 777223 must be used. This replaces Power Supply 674-037. With Power Supply Part Number 777223 in use, M674A equipment is Class 2 and internally protected, and uses Type B applied parts.
NOTE
Although the Infant Flow™ Advance meets the requirements of current EMC/RFI legislation, this does not guarantee immunity from all sources of radiated energy. Some mobile telephones and other products containing radio transmitting components may cause malfunction of the Infant Flow™ Advance and should not be used in the vicinity of the device.
777077-101 Revision B June 2004
14 Infant Flow Advance™ System
777077-101 Revision B June 2004
Loading...
+ 32 hidden pages