Viasys 3100, 3100A Quick reference card

1. BIAS FLOW - Adjusting Bias Flow will affect mean Paw.
Premature 8-15 LPM Nearterm 10-20 LPM Small child 15-25 LPM Large Child 20-30 LPM Lowering Bias Flow may decrease work of breathing and facilitate weaning.
2. ADJUST - Sets the mean airway pressure
This control directly affects lung volume and oxygenation. Initial setting is slightly higher than conventional ventilation.
3. POWER - Controls piston displacement
4. INSPIRATORY TIME % -
Set to 33% for most applications
For some patients, increasing inspiratory time % to 50 can improve ventilation and increase lung recruitment.
Once set, this control is not typically changed.
5. FREQUENCY - Breath rate is expressed in hertz.
One hertz equals 60 breaths per minute (BPM). Initial frequency settings are dependent on patient size and lung pathophysiology.
Generally, the smaller the patient the higher the frequency. Lower frequencies are employed for larger childen and management strategies designed to minimize the potential for gas trapping
A decrease in frequency = increased tidal volume. An increase in frequency = decreaded tidal volume.
© 2003 VIASYS Healthcare
3100A Quick Reference Card
CONTROLS AT A GLANCE
3
1
4
5
2
WHILE ON HFOV:
Wean FiO2 as tolerated to target FiO2.
Once FiO2 is < target FiO2, begin to wean mean airway pressure by increments of 1-2 cmH2O. Assess for adequate lung inflation with chest x-ray and arterial PO2.
Assess oxygenation with pulse oximetry.
IN AIR LEAK SYNDROME: Mean airway pressures are similar to those used in
conventional ventilation. Higher FiO2's are typically used.
TRANSITION TO CONVENTIONAL VENTILATION, NCPAP, or Nasal Cannula, when each of the following parameters are met:
1) Mean airway pressure is stable and appropriate for the pathology
2) Patient tolerates position changes and procedures well
3) ABG's are acceptable and stable
These guidelines are recommedations only. For any treatment each physician must determine the appropriateness of these guidelines as they apply to specific patients.
Mean airway pressure is regulated by controlling the inflation of the balloon valve in the expiratory limb of the circuit (Figure 1). As inflation pressure inside the balloon increases, the outflow of gas is restricted, providing mean airway pressure.
P/N 775895-101A
WEANING GUIDELINES
VALVE ASSEMBLY LOCATIONS
MEAN AIRWAY PRESSURE
A.
Red - Dump Valve, mid expiratory limb.
B. Green - Mean Airway Pressure, expiratory limb. C. Blue - Limit Valve, inspiratory limb
A
B
C
FIGURE 1
VIASYS Healthcare - Critical Care Division VIASYS Healthcare
22705 Savi Ranch Parkway Rembrandtlaan 1b Yorba Linda, CA 92887 3723 BG Bilthoven 714-283-1830 • 800-520-4368 The Netherlands Fax 714-283-8493 (31) 30 2289711
www.ViasysCriticalCare.com Fax (31) 30 2286244
Loading...