Note: Do not cut transducer cables to alter length. This will void the
factory warranty. Cables are available in several lengths. Assess the
installation location prior to ordering to determine the optimum length.
If the wrong length is ordered, contact the factory.
FST1, 2, 3
Temp Sensors
FST4, 5
DANGER
HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH
• Follow safe electrical work practices. See NFPA 70E in the USA, or applicable local codes.
• This equipment must only be installed and serviced by qualified electrical personnel.
• Read, understand and follow the instructions before installing this product.
• Turn off all power supplying equipment before working on or inside the equipment.
• Use a properly rated voltage sensing device to confirm power is off.
DO NOT DEPEND ON THIS PRODUCT FOR VOLTAGE INDICATION
Failure to follow these instructions will result in death or serious injury.
A qualied person is one who has skills and knowledge related to the construction and
operation of this electrical equipment and the installation, and has received safety
training to recognize and avoid the hazards involved. NEC2009 Article 100
No responsibility is assumed by Veris Industries for any consequences arising out of the
use of this material.
DIMENSIONS
(163 mm)
(153 mm)
6.4”
6.0”
FSR Monitor
4.1”
(105 mm)
4.3”
(110 mm)
0.75”
(19 mm)
2.1”
(53 mm)
2x 0.5”
NOTICE
• This product is not intended for life or safety applications.
• Do not install this product in hazardous or classied locations
other than those listed in Specications.
• Read and understand the instructions before installing
this product.
• Turn off all power supplying equipment before working on it.
• The installer is responsible for conformance to all applicable codes.
No responsibility is assumed by Veris Industries for any consequences arising out of the
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INSTALLATION GUIDE
FST(1, 2, 3) Insert Temperature SensorFST(1, 2, 3) Transducer
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
QUICK INSTALL
This manual contains detailed operating instructions for all aspects of the FSR Series.
The following condensed instructions are provided to assist the operator in getting
the instrument started up and running as quickly as possible. This pertains to basic
operation only. If specic instrument features are to be used or if the installer is
unfamiliar with this type of instrument, refer to the appropriate sec tion in the
manual for complete details.
Note: The following steps require informati on supplied by the meter itself, so it is necessary to
supply power to the unit, at least temp orarily, to obtain setup informatio n.
1 - Transducer Location
1. In general, selec t a mounting location on the piping system with a minimum of
10 pipe diameters (10x the pipe inside diameter) of straight pipe upstream and 5
straight diameters downstream. See Table 2.1 for additional congurations.
2. Select a mounting method for the transducers based on pipe size and liquid
characteristics. See Table 2.2. Transducer congurations are illustrated in Figure
Q.1 below. The V-mount conguration is usually the rst choice, with W-mount
and Z-mount used if needed to boost signal strength.
Note: All FST1xxxx, FST2xxx x, and FST3xxxx transducers use V-Mount congu ration.
3. Enter the following data into the meter via the integral keypad or the software
utility (if not entered by the factory)
INSTALLATION GUIDE
2 - Electrical Connections
Transducer/Power Connections
1. Route the transducer cables from the transducer mounting location back to the
enclosure. Connect the transducer wires to the terminal block in the enclosure.
Note: Do not cut transducer cables to alter length. This will void the
factory warranty. Cables are available in several lengths. Assess the
installation location prior to ordering to determine the optimum
length. If the wrong length is ordered, contact the factory.
2. Verify that power supply is correct for the meters power option.
AC units require 95 to 265 VAC, 47 to 63 Hz @ 17 VA maximum. DC units
require 10 to 28 VDC @ 5 Watts maximum.
3. Connect power to the ow meter.
Downstream+
DownstreamUpstreamUpstream+
1. Transducer mounting method
2. Pipe O.D. (Outside Diameter)
3. Pipe wall thickness
4. Pipe material
5. Pipe sound speed*
6. Pipe relative roughness*
7. Pipe liner thickness
8. Pipe liner material
9. Fluid type
10. Fluid sound speed*
11. Fluid viscosity*
12. Fluid specic gravity*
* Nominal values for these parameters are included within the operating system. Modify if the
exact system values a re known.
TOP VIEW
OF PIPE
TOP VIEW
OF PIPE
TOP VIEW
OF PIPE
Figure Q.2 - Transducer Connections
3 - Pipe Preparation and Transducer Mounting
FST4xxxx, FST5xxxx Transducers
1. Place the ow meter in signal strength measuring mode. This value is available on
the display (Service Menu) or in the data display of the software utility.
2. The pipe surface, where the transducers are to be mounted, must be clean and dry.
Remove scale, rust or loose paint to ensure satisfactory acoustic conduction. Wire
brushing the rough surfaces of pipes to smooth bare metal may also be useful.
Plastic pipes do not require preparation other than cleaning.
3. Apply a single ½” (12 mm) bead of acoustic couplant grease to the upstream
transducer and secure it to the pipe with a mounting strap.
4. Apply acoustic couplant grease to the downstream transducer and press it onto
the pipe using hand pressure at the lineal distance calculated in Step 1.
5. Space the transducers according to the recommended values found on the product
conguration sheet or from the software utility. Secure the transducers with the
mounting straps at these locations.
W-MountV-MountZ-Mount
Figure Q.1 - Transducer Mounting Congurations
4. Record the value calculated and displayed as Transducer Spacing (FST4, FST5 only).
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
FST1xxxx, FST2xxxx, and FST3xxxx Transducers
1. Place the ow meter in signal strength measuring mode. This value is available on
the display (Service Menu) or in the data display of the software utility.
2. The pipe surface, where the transducers are to be mounted, must be clean and dry.
Remove scale, rust or loose paint to ensure satisfactory acoustic conduction. Wire
brushing the rough surfaces of pipes to smooth bare metal may also be useful.
Plastic pipes do not require preparation other than cleaning.
3. Apply a single ½” (12 mm) bead of acoustic couplant grease to the top half of the
transducer and secure it to the pipe with bottom half or U-bolts.
4. Tighten the nuts so that the acoustic coupling grease begins to ow out from the
edges of the transducer and from the gap between the transducer and the pipe.
Do not over tighten.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INTRODUCTION
General
The Veris ultrasonic ow meter is designed to measure the uid velocity of liquid
within a closed conduit. The transducers are a non-contacting clamp-on or clamparound type that does not foul and is easy to install.
The Veris family of transit time ow meters utilize two transducers that function
as both ultrasonic transmitters and receivers. The transducers are clamped on
the outside of a closed pipe at a specic distance from each other. The transducers
can be mounted in V-Mount where the sound transverses the pipe two times,
W-Mount where the sound transverses the pipe four times, or in Z-Mount where the
transducers are mounted on opposite sides of the pipe and the sound crosses the pipe
once. The selection of mounting method is based on pipe and liquid characteristics
that both have an eect on how much signal is generated. The ow meter operates
by alternately transmitting and receiving a frequency modulated burst of sound
energy between the two transducers and measuring the time interval that it takes
for sound to travel between the two transducers. The dierence in the time interval
measured is directly related to the velocity of the liquid in the pipe.
INSTALLATION GUIDE
Because the transducers are non-contacting and have no moving parts, the ow
meter is not aected by system pressure, fouling, or wear. FST4 and FST5 transducers
are rated to a pipe surface temperature of -40 to +250 °F (-40 to +121 °C). FST1, FST2,
and FST3 small pipe transducers are rated from -40 to +185 °F (-40 to +85 °C).
FrequencyTransducersTransmission
Modes
2 MHzAll ½” thru 1½”
2” Tubing
1 MHz
500 kHzlarger than 24”W, V, and Z24” and Greater
2” ANSI and Copper
all 2” to 24”W, V, and Z2” to 24”
Selected by
Firmware
Selected by
Firmware
User Safety
The FSR Series employs modular construction and provides electrical safety for the
operator. The display face contains voltages no greater than 28 VDC. The display face
swings open to allow access to user connections.
Pipe Size and
Type
Specic to
Transducer
Specic to
Transducer
TOP VIEW
OF PIPE
W-MountV-MountZ-Mount
Figure 1.1 - Ultrasound Transmission
TOP VIEW
OF PIPE
TOP VIEW
OF PIPE
Application Versatility
The FSRxxxx ow meter can be successfully applied on a wide range of metering
applications. The simple-to-program monitor allows the standard product to be used
on pipe sizes ranging from ½ inch to 100 inches (12 mm to 2540 mm) pipe*. A variety
of liquid applications can be accommodated:
ultrapure liquids
potable water
chemicals
sewage
reclaimed water
cooling water
river water
DANGER
HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH
• Disconnect electrical power before opening the instrument enclosure.
• Wiring mus t conform to applicable codes.
Failure to follow these instructions will result in death or serious injury.
Data Integrity
Non-volatile ash memory retains all user-entered conguration values in memory
for several years at 77°F (25°C), even if power is lost or turned o. Password
protection is provided as part of the Security menu (SEC MENU) and prevents
inadvertent conguration changes or totalizer resets.
Product Identification
The serial number and complete model number of the monitor are located on the top
outside surface of the housing. If technical assistance is required, please provide the
Customer Service Department with this information.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
PART 1 - MONITOR INSTALLATION
After unpacking, save the shipping carton and packing materials in case the
instrument is stored or re-shipped. Inspect the equipment and carton for damage. If
there is evidence of shipping damage, notify the carrier immediately.
Mount the enclosure in an area that is convenient for servicing, calibration, and
observation of the LCD readout.
1. Locate the monitor within the length of transducer cables supplied. If this is
not possible, it is recommended that the cable be exchanged for one that is of
proper length. If additional cable is added, utilize RG59 75 Ω coaxial cable and
BNC connections. Transducer cables that are up to 990 feet (300 meters) can be
accommodated.
2. Mount the monitor in a location:
• Where little vibration exists.
• That is protected from corrosive uids.
• That is within the monitor’s ambient temperature limits -40 to +185°F
(-40 to +85°C).
• That is out of direct sunlight. Direct sunlight may increase monitor
temperature to above the maximum limit.
3. Mounting - Refer to Figure 1.2 for enclosure and mounting dimension details.
Ensure that enough room is available to allow for door swing, maintenance and
conduit entrances. Secure the enclosure to a at surface with two appropriate
fasteners.
INSTALLATION GUIDE
Transducer Connections
To access terminal strips for wiring, loosen the two screws in the enclosure door and
open.
Guide the transducer terminations through the monitor conduit hole located in the
bottom-left of the enclosure. Secure the transducer cable with the supplied conduit
nut (if exible conduit was ordered with the transducer).
The terminals within the unit are of a screw-down barrier terminal type. Connec t
the appropriate wires at the corresponding screw terminals in the monitor. Observe
upstream and downstream (+/–) orientation. See Figure 1.3.
372
VE
D
ACL
C US
E167432
PRODUCT SERVICE
TUV
RoHS
R
AC IN : 100-240VAC,50/60Hz
DC OUT :
PWC-15E
R
+15V / 0.3A
$
0.15A
R2807
-Vo
1500mA250V
C US
W
www.astrodyne.com
strodyne
ACN
+Vo
O
N
1 2 3 4
+
+
-
-
-
-
+
+
Downstream
Downstream
Upstream
Upstream
Modbus B
Modbus A
TFX Rx
TFX Tx
Reset Total
Modbus Gnd
Modbus
Signal Gnd.
Control 1
Control 2
Frequency Out
4-20 mA Out
95 - 264 VAC
AC Neutral
Figure 1.2 - FSR Dimensions
6.4”
(163 mm)
6.0”
(153 mm)
1.4”
4.1”
(105 mm)
4.3”
(110 mm)
0.75”
(19 mm)
2x 0.5”
(13 mm)
2.1”
(53 mm)
To Transducers
Note: Wire colors may va ry!
(+) connectio n with be either red or blue;
(–) connection will be e ither black or clear.
Figure 1.3 - Transducer Connections.
Note: The transducer cable carries low level, high freque ncy signals. Do not add length to the cable
supplied with the transducers. If additional cable is required, contact the manufacturer to arrange
an exchange for a transducer with the a ppropriate length of cable. Cables to 990 feet (300 meters)
are available. If adding cabl e, ensure that it is RG59 75 Ω compatible and uses BNC terminations.
Connect power to the screw terminal block in the monitor. See Figure 1.4 and Figure
1.5. Utilize the conduit hole on the right side of the enclosure for this purpose. Use
wiring practices that conform to local and national codes (e.g., The National Electrical
Code® Handbook in the U.S.).
CAUTION
• HAZARD OF IMPROPER OR UNSAFE OPERATION
• This instrument requires clean electric al line power. Do not operate this unit on
circuit s with noisy compone nts (e.g., fluores cent lights, relays, compressors, or
variable frequency drives).
• Do not use with high current step-down transformers from high voltage sources.
• Do not run signal wires with line power in the s ame wiring tray or conduit.
Any other wiring method may be unsafe or cause improper operation of the
instrument.
4. Conduit Holes - Conduit holes should be used where cables enter the enclosure.
Holes not used for cable entry should be sealed with plugs.
Note: Use NEMA 4 [IP-65] rated ttings/plugs to maintain the watertight integrity o f the
enclosure. Generally, the right conduit hole (viewed from fro nt) is used for power, the left conduit
hole for transducer connectio ns, and the center hole is utilized fo r I/O wiring.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INSTALLATION GUIDE
AC Power Connections
Connect 90 to 265 VAC, AC Neutral and Chassis Ground to the terminals referenced in
Figure 1.4. Do not operate without an earth (chassis) ground connection.
ACN
1500mA250V
372
W
C US
VE
D
R
AC IN : 100-240VAC,50/60Hz
DC OUT :
C US
ACL
95 - 264 VAC
95 - 264 VAC
AC Neutral
AC Neutral
www.astrodyne.com
PWC-15E
E167432
Signal Gnd.
Control 1
Control 2
Frequency Out
4-20 mA Out
Reset Total
Modbus Gnd
Modbus B
Modbus A
O
1 2 3 4
N
Figure 1.4 - AC Power Connections
strodyne
+15V / 0.3A
$
R
TUV
PRODUCT SERVICE
+Vo
-Vo
0.15A
R2807
RoHS
Modbus
TFX Rx
TFX Tx
Downstream
Upstream
-
-
+
+
DC Power Connections
The device can be operated from a 10 to 28 VDC source, as long as the source is
capable of supplying a minimum of 5 Watts of power.
Connect the DC power to 10 to 28 VDC In, Power Gnd., and Chassis Gnd., as in Figure
1.5.
10 - 28 VDC
10 - 28 VDC
Power Gnd.
Power Gnd.
Signal Gnd.
Control 1
Control 2
Frequency Out
4-20 mA Out
Reset Total
O
1 2 3 4
N
Modbus Gnd
Modbus B
Modbus A
Modbus
TFX Rx
TFX Tx
Downstream
Upstream
-
-
+
+
Power
Ground
10 -28 VDC
Note: In electrically noisy ap plications, ground the meter to the pipe where the transducers are
mounted to provide additio nal noise suppression. This approach is only eective with conduc tive
metal pipes. Remove the ear th (chassis) ground derived from the line voltage p ower supply at the
meter and connect a new ea rth ground between the meter and the pipe b eing measured.
Note: The terminal blocks accomodate wire ga uges up to 14 AWG.
Note: AC powered versions are protected by a eld replaceabl e fuse. This fuse is equivalent to
Littelfuse/Wick mann P.N. 3720500041 or 37405000410.
Figure 1.5 - DC Power Connections
Note: DC powered versions are protected by an au tomatically reset ting fuse. This fuse does not
require replacement.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
PART 2 - TRANSDUCER INSTALLATION
INSTALLATION GUIDE
General
The FST transducers contain piezoelectric crystals for transmitting and receiving
ultrasonic signals through walls of liquid piping systems. FST transducers are
relatively simple and straightforward to install, but spacing and alignment of the
transducers is critical to the system’s accuracy and performance. Take care to ensure
that these instructions are carefully executed. FST1, FST2, and FST3 small pipe
transducers have integrated transmitter and receiver elements that eliminate the
requirement for spacing measurement and alignment.
Mounting of the FST4 and FST5 clamp-on ultrasonic transit time transducers is a
three-step process:
1. Select the optimum location on a piping system.
2. Enter the pipe and liquid parameters into either the sof tware utility or key the
parameters into the transmitter using the keypad. The software utility or the
monitor’s rmware calculates proper transducer spacing based on these entries.
3. Pipe preparation and transducer mounting.
BTU meters require two RTDs to measure heat usage. The ow meter utilizes 1,000 Ω,
three-wire, platinum RTDs in two mounting styles. Surface mount RTDs are available
for use on well insulated pipes. Mounting the RTD in an uninsulated area causes
inconsistent temperature readings. Insertion (wetted) RTDs should be sued in these
areas instead.
Step 1 - Mounting Location
The rst step in the installation process is the selection of an optimum location for
the ow measurement to be made. This requires a basic knowledge of the piping
system and its plumbing.
Piping Configuration and Transducer PositioningUpstream
Pipe
Diameters
***
255
Flow
*
Flow
*
**
145
**
105
Flow
*
**
105
Flow
*
Flow
*
Flow
*
**
105
**
245
**
Downstream
Pipe
Diameters
An optimum location is dened as:
• A piping system that is completely full of liquid when measurements are
being taken. The pipe may become completely empty during a process
cycle, which results in the error code 0010 (Low Signal Strength) being
displayed on the ow meter while the pipe is empty. This error code clears
automatically once the pipe rells with liquid. It is not recommended to
mount the transducers in an area where the pipe may become par tially
lled. Partially lled pipes cause erroneous and unpredictable operation
of the meter.
• A piping system that contains lengths of straight pipe such as
those described in Table 2.1. The optimum straight pipe diameter
recommendations apply to pipes in both horizontal and vertical
orientation. The straight runs in Table 2.1 apply to liquid velocities that
are nominally 7 FPS (2.2 MPS). As liquid velocity increases above this
nominal rate, the requirement for straight pipe increases proportionally.
• Mount the transducers in an area where they will not be inadvertently
bumped or disturbed during normal operation.
• Avoid installations on downward owing pipes unless adequate
downstream head pressure is present to overcome partial lling of or
cavitation in the pipe.
• Note: Do not cut transducer cables to alter length. This will
void the factory warranty. Cables are available in several
lengths. Assess the installation location prior to ordering to
determine the optimum length. If the wrong length is ordered,
contact the factory.
Table 2.1 - Piping Conguration and Transducer Positioning
The ow meter system provides repeatable measurements on piping systems that do
not meet these requirements, but accuracy of these readings may be inuenced to
various degrees.
Step 2 - Transducer Spacing
Transit time ow meters can be used with two dierent transducer types. Meters
that utilize the FST4 and FST5 transducer sets consist of two separate sensors
that function as both ultrasonic transmitters and receivers. FST1, FST2, and FST3
transducers integrate both the transmitter and receiver into one assembly that xes
the separation of the piezoelectric crystals. These transducers are clamped on the
outside of a closed pipe at a specic distance from each other.
The FST4 and FST5 transducers can be mounted in:
W-Mount where the sound traverses the pipe four times. This mounting
method produces the best relative travel time values but the weakest signal
strength.
V-Mount where the sound traverses the pipe twice. V-Mount is a compromise
between travel time and signal strength.
Z-Mount where the transducers are mounted on opposite sides of the pipe
and the sound crosses the pipe once. Z-Mount will yield the best signal
strength but the smallest relative travel time.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
For further details, reference Figure 2.1. The appropriate mounting conguration is
based on pipe and liquid characteristics. Selection of the proper transducer mounting
method is not entirely predictable and many times is an iterative process. Table 2.2
contains recommended mounting congurations for common applications. These
recommended congurations may need to be modied for specic applications
if such things as aeration, suspended solids, out of round piping or poor piping
conditions are present. Use of meter diagnostics in determining the optimum
transducer mounting is covered later in this section.
Table 2.3 - Transducer Mounting Mod es — FST1, FST2, FST3
Step 3 - Entering Pipe and Liquid Data
The system calculates proper transducer spacing by utilizing piping and liquid
information entered by the user. Enter this information via the keypad or via the
optional software utility.
The best accuracy is achieved when transducer spacing is exactly what the meter
calculates, so use the calculated spacing if signal strength is satisfactory. If the pipe is
not round, the wall thickness is not correct, or the actual liquid being measured has a
dierent sound speed than the liquid programmed into the transmitter, the spacing
can vary from the calculated value. If that is the case, place the transducers sat the
highest signal level obser ved by moving the transducers slowly around the mount
area.
Note: Transducer spacing is calculated on “ideal” pipe. Ideal p ipe is almost never found so the
transducer spacing distances may need to be altered. An eec tive way to maximize signal strength
is to congure the display to show signal strength, x o ne transducer on the pipe and then starting
at the calculated spacing, move the rema ining transducer small distances forward and back to
nd the maximum signal s trength point.
Important! Enter all of the data on this list, save the data, and reset the
meter before mounting transducers.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INSTALLATION GUIDE
The following information is required before programming the instrument:
Pipe liner thickness (if present)Pipe liner material (if present)
Fluid typeFluid sound speed
Fluid viscosity
1
Fluid specic gravity
1
1
Note: Much of the data relating to material sound spee d, viscosity, and specic gravity is
pre-programmed into the o w meter. This data only needs to be modied if it is kn own that a
particular ap plications data varies from the reference values. Refer to Part 4 of this man ual for
instructions on entering conguration data into the ow meter via the monitor’s keypad. Refer to
Part 5 for data entry via the sof tware.
1
Nominal values for these parameters are included within the operating system. The nominal
values may be used as they appear o r may be modied if exact system values are kno wn.
After entering the data listed above, the meter calculates proper transducer spacing
for the particular data set. This distance is in inches if it is congured in English units,
or millimeters if congured in metric units.
Before mounting the transducers onto the pipe surface, clean an area slightly larger
than the at surface of each transducer to eliminate all rust, scale and moisture.
For pipes with rough surfaces, such as ductile iron pipe, wire brush the surface to a
shiny nish. Paint and other coatings need not be removed unless aked or bubbled.
Plastic pipes typically do not require surface preparation other than soap and water
cleaning.
Properly orient the transducers and spaced them on the pipe to provide optimum
reliability and per formance. On horizontal pipes, when Z-Mount is required, mount
the transducers 180 radial degrees from one another and at least 45 degrees from
the top-dead-center and bottom-dead-center of the pipe. See Figure 2.2 Also see
Z-Mount Transducer Installation. On vertical pipes the orientation is not critical.
Measure the spacing between the transducers using the two spacing mark s on the
sides of the transducers. These marks are approximately 0.75” (19 mm) back from the
nose of the FST4/FST5 transducers. See Figure 2.3.
Mount FST1, FST2, and FST3 transducers with the cable exiting within ±45 degrees
of the side of a horizontal pipe. See Figure 2.2. On vertical pipes the orientation does
not apply.
Figure 2.3 - Transducer Alignment Marks
V-Mount and W-Mount Installation
Application of Couplant
For FST4 and FST5 transducers, place a single bead of couplant, approximately ½
inch (12 mm) thick, on the at face of the transducer. See Figure 2.4. Generally, a
silicone-based grease is used as an acoustic couplant, but any grease-like substance
that is rated not to “ow” at the temperature that the pipe may operate at will be
acceptable. For pipe surface temperature over 130°F (55°C) Sonotemp® (P.N. D0022011-010) is recommended.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INSTALLATION GUIDE
Transducer Positioning
1. Place the upstream transducer in position and secure with a mounting strap. Place
straps in the arched groove on the end of the transducer. A screw is provided to
help hold the transducer onto the strap. Tighten the transducer strap securely.
2. Place the downstream transducer on the pipe at the calculated transducer spacing.
See Figure 2.5. Apply rm hand pressure. If signal strength is greater than 5,
secure the transducer at this location. If the signal strength is not 5 or greater
then using rm hand pressure, slowly move the transducer both towards and
away from the upstream transducer while observing signal strength. Clamp
thetransducer where the highest signal strength is observed. Signal levels much
less than 5 may not yield acceptable data.
Note: Signal strength readings upd ate only every few seconds, so it is advisable to move
the transducer 1/8”, wait, see if signal is increasing or decreasing and then repeat unti l
the highest level is achieved.
3. If after adjustment of the transducers the signal strength does not rise to above 5,
then select an alternate transducer mounting method. If the mounting method
was W-Mount, then re-congure the monitor for V-Mount, move the downstream
transducer to the new spacing distance and repeat Step 4.
Small Pipe Transducer Installation
The small pipe transducers are designed for specic pipe outside diameters. Do not
attempt to mount a transducer onto a pipe that is either too large or too small for the
transducer.
FST1, FST2, and FST3 installation consists of the following steps:
1. Apply a thin coating of acoustic coupling grease to both halves of the transducer
housing where the housing will contact the pipe. See Figure 2.6.
2. On horizontal pipes, mount the transducer in an orientation such that the cable
exits at ±45 degrees from the side of the pipe. Do not mount with the cable
exiting on either the top or bottom of the pipe. On vertical pipes the orientation
does not matter. See Figure 2.2.
Transducer
Spacing
Figure 2.5 - Transducer Positioning
3. Tighten the wing nuts or “U” bolts so that the acoustic coupling grease begins to
ow out from the edges of the transducer or from the gap between the transducer
halves. Do not over tighten.
4. If signal strength is less than 5, remount the transducer at another location on the
piping system.
5. Conguration Procedure:
a. Establish communications with the transit tme meter. See Part 5 Software Utility.
b. From the tool bar, select calibration.
USP - Device Addr 127
ConfigurationCalibrationStrategy
Device Addr 127
1350 Gal/Min
Flow:
Pos:
Neg:
Margin:
Delta T:
0 OB
0 OB
0 OB
15.6%
100%
-2.50 ns
09:53:39
Totalizer Net:
Sig. Strength:
Last Update:
Errors
2000
1600
1200
HelpWindowCommunicationsViewEditFile
!
Print PreviePrint
Scale:60 MinTime:
200
c. On the pop-up screen, click Next twice to get to page 3 of 3. Click Edit in this
screen.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INSTALLATION GUIDE
Calibration (Page 3 of 3) - Linearization
28.2
Gal/M
Delta Time
1) Please establish a
reference flow rate.
1FPS / 0.3MPS Minimum.
2) Enter the reference flow
rate below. (Do not enter 0)
3) Wait for flow to stabilize.
4) Press the Set button.
Flow:
Set
Edit
Export...
d. In the Calibration Points Editor screen, use the Add and Remove buttons as
needed. Click OK when complete.
e. The display returns to Calibration (Page 3 of 3). Click nish
f. Power cycle the unit to activate the new settings.
Mounting Transducers in Z-Mount Configuration
Installation on larger pipes requires careful measurements of the linear and radial
placement of the FST4 and FST5 transducers. Failure to properly orient and place the
transducers on the pipe may lead to weak signal strength and/or inaccurate readings.
1. Place the transducers on opposite sides of the pipe. This distance around the pipe
is calculated by multiplying the pipe diameter by 1.57. The transducer spacing
along the pipe is the same as found in the Transducer Positioning section.
2. For FST4 and FST5 transducers, place a single bead of couplant, approximately ½
inch (12 mm) thick, on the at face of the transducer. See Figure 2.4. Generally,
a silicone-based grease is used as an acoustic couplant, but any good quality
grease-like substance that is rated to not “ow” at the temperature that the pipe
may operate at is acceptable.
3. Place the upstream transducer in position and secure with a stainless steel strap
or other fastening device. Place straps in the arched groove on the end of the
transducer. A screw is provided to help hold the transducer onto the strap. Tighten
transducer strap securely. Larger pipes may require more than one strap to reach
the circumference of the pipe.
4. Place the downstream transducer on the pipe at the calculated transducer spacing.
See Figure 2.7. Using rm hand pressure, slowly move the transducer both towards
and away from the upstream transducer while observing signal strength. Clamp
the transducer at the position where the highest signal strength is observed.
Signal strength of between 5 and 98 is acceptable. The fac tory default signal
strength setting is 5, however there are many application specic conditions that
may prevent the signal strength from attaining this level.
A minimum signal strength of 5 is acceptable as long as this signal level is
maintained under all ow conditions. On certain pipes, a slight twist to the
transducer may cause signal strength to rise to acceptable levels.
5. Certain pipe and liquid characteristics may cause signal strength to rise to greater
than 98. At this level, the signals may saturate the input ampliers and cause
erratic readings. To lower the signal strength, change the transducer mounting
method to the next longest transmission path. For example, if there is excessive
signal strength and the transducers are mounted in a Z-Mount, try changing to
V-Mount or W-Mount. Finally you can also move one transducer slightly o line
with the other transducer to lower signal strength.
6. Secure the transducer with a stainless steel strap or other fastener.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
1234
O
N
PART 3 - INPUTS/OUTPUTS
FSRxxxx SERIES
INSTALLATION GUIDE
General
The FSR1 is available in two congurations: the ow model and the energy model.
The ow model is equipped with a 4-20 mA output, two open collec tor outputs, a
rate frequency output, and RS-485 communications using the Modbus RTU command
set. The energy version has inputs for two 1,000 Ω RTD sensors in place of the rate
frequency and alarm outputs. This version allows the measurement of pipe input and
output temperatures for calculating energy usage calculations.
4-20 mA Output
The 4-20 mA output interfaces with most recording and logging systems by
transmitting an analog current signal that is propor tional to system ow rate. The
4-20 mA output is internally powered (current sourcing) and can span negative to
positive ow/energy rates.
For AC powered units, the 4-20 mA output is driven from a +15 VDC source located
within the meter, isolated from earth ground connections. The AC powered model
accommodates loop loads up to 400 Ω. DC powered meters utilize the DC power
supply voltage to drive the current loop. The current loop is not isolated from DC
ground or power. Figure 3.1 shows graphically the allowable loads for various input
voltages. The combination of input voltage and loop load must stay within the
shaded area of Figure 3.1.
Supply Voltage - 7 VDC
1100
1000
900
800
700
600
500
400
Loop Load (Ohms)
300
200
100
10121416182022242628
0.02
Supply Voltage (VDC)
= Maximum Loop Resistance
Operate in the
Shaded Regions
90-265 VAC
Loop
Resistance
Figure 3.2 - 4-20 mA Output
AC Neutral
Signal Gnd.
Control 1
Control 2
Frequency Out
4-20 mA Out
Reset Total
Signal Ground
Meter Power
7 VDC
Drop
The 4-20 mA output signal is available between the 4-20 mA Out and Signal Gnd.
terminals as shown in Figure 3.2.
Control Outputs
Two independent open collector transistor outputs are included with the ow meter.
Each output can be congured for one of the following four functions:
Rate Alarm
Signal Strength Alarm
Totalizing/Totalizing Pulse
Errors
None
Both control outputs are rated for a maximum of 100 mA and 10 to 28 VDC. A pull-up
resistor can be added externally or an internal 10 kΩ pull-up resistor can be selected
using DIP switches on the power supply board.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
Square wave
output
Simulated turbine
output
TM
FSRxxxx SERIES
INSTALLATION GUIDE
Set the on/o values for the Rate Alarm and Signal Strength Alarm using either the
keypad or the software utility.
Typical control connections are illustrated in Figure 3.4. Please note that only
the Control 1 output is shown. Control 2 is identical except the pull-up resistor is
governed by SW2.
VCC
10K
O
90-265 VAC
AC Neutral
Signal Gnd.
Control 1
Control 2
Frequency Out
4-20 mA Out
Reset Total
N
1 2 3 4
SW1/SW2
Figure 3.4 - Typical Control Connections
10 - 28
VDC
100 mA Maximum
90-265 VAC
AC Neutral
Signal Gnd.
Control 1
Control 2
Frequency Out
4-20 mA Out
Reset Total
O
N
VCC
10K
1 2 3 4
SW1/SW2
Alarm Output
The ow rate output permits output changeover at two separate ow rates allowing
operation with an adjustable switch deadband. Figure 3.5 illustrates how the setting
of the two set points inuences rate alarm operation.
A single-point ow rate alarm places the ON setting slightly higher than the OFF
setting, establishing a switch deadband. If a deadband is not established, switch
chatter (rapid switching) may result if the ow rate is very close to the switch point.
Minimum
Flow
Maximum
Flow
Totalizer Output for Energy Meter
Energy units can be ordered with a totalizer pulse output option. This option is
installed in the position where the Ethernet option would normally be; therefore,
the totalizer pulse output option and the Ethernet communications output cannot be
installed simultaneously.
Optional totalizing pulse specications:
Signal1 pulse for each increment of the totalizer’s least signicant digit
TypeOpto-isolated, open collector transistor
Pulse Width30 msec, max. pulse rate 16 Hz
Voltage28 VDC max.
Current100 mA max. (current sink)
Pull-up Resistor2.8 kΩ to 10 kΩ
Wiring and conguring this option is similar to the totalizing pulse output for the
ow only version. This option must use an external current limiting resistor.
Totalizing
Pulse Output
Option
Internal
RxD
TB1
Total Pulse
100 mA
Maximum
V
CC
2.8K to 10K
Isolated Output
Total Pulse
Pull-up
Resistor
Set OFF
Set ON
Output ON
Output OFF
Deadband
Figure 3.5 - Single Point Alarm Operation
Batch/Totalizer Output
Totalizer mode congures the output to send a 33 mSec pulse each time the display
totalizer increments divided by the TOT MULT. The TOT MULT value must be a whole,
positive, numerical value.
Fir example, If the totalizer exponent (TOTL E) is set to E2 (×100) and the totalizer
multiplier (TOT MULT) is set to 1, then the control output pulses each time the display
totalizer increments or once per 100 measurement units totalized.
Signal Strength Alarm
The SIG STR alarm provides an indication that the signal level reported by the
transducers has fallen to a point where ow measurements may not be possible.
It can also be used to indicated that the pipe has emptied. Like the rate alarm
described previously, the signal strength alarm requires that two points be entered,
establishing an alarm deadband. The ON value must be lower than the OFF value. If a
deadband is not established and the signal strength decreases to approximately the
value of the switch point, the output may “chatter.”
Error Alarm Outputs
When a control output is set to ERROR mode, the output activates when an error
causes the meter to stop measuring reliably. See the Appendix of this manual for a list
of potential error codes.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
FSRxxxx SERIES
INSTALLATION GUIDE
Frequency Output [FSRxxx1x only]
The frequency output is an open-collector transistor circuit that outputs a pulse
waveform that varies proportionally with ow rate. This type of frequency output is
also know as a “Rate Pulse” output. The frequency output is proportional to the max
ow rate entered into the meter. The maximum output frequency is 1000 Hz.
In addition to the control outputs, the frequency output can be used to provide total
information by use of a K-factor that relates the number of pulses from the frequenc y
output to the number of accumulated pulses that equates to a specic volume.
This relationship is described by the following equation: K-factor = 60,000 / full scale units. The 60,000 relates to measurement units in volume/min. Measurement units in
seconds, hours or days would require a dierent numerator.
If the frequency output is to be used as a totalizing output, then the meter and the
receiving instrument must have identical K-factor values programmed into them
to ensure that accurate readings are recorded by the receiving instrument. Unlike
standard mechanical ow meters such as turbines, gear or nutating disk meters, the
K-factor can be changed by modifying the MAX RATE ow rate value.
Note: For a full treatment of K-factors p lease see the Appendix of this manual.
There are two frequency output types available:
1.) Turbine meter simulation - This option is utilized when a receiving
instrument is capable of interfacing directly with a turbine ow meter’s
magnetic pickup. The output is a relatively low voltage AC signal whose
amplitude swings above and below the signal ground reference. The
minimum AC amplitude is approximately 500 mV peak-to-peak. To activate
the turbine output circuit, turn SW4 OFF .
To interconnect meters, utilize three-wire shielded cable such as Belden® 9939 or
equal. In noisy environments, connect the shield on one end to earth ground. Use a
USB to RS-485 converter to communicate with a PC running Windows 98, Windows
ME, Windows 2000, Windows NT, Windows XP, Windows Vista, or Windows 7.
For computers with RS-232C serial ports, use an RS-232C to RS-485 converter to
interconnect the RS-485 network to a communication port on a PC. If monitoring
more than 126 meters, use an additional converter and communication port.
4-20 mA Out
Reset Total
Modbus Gnd
Modbus B
Modbus A
RS232 to RS485
Model 485USBTB-2W
USB to RS485
TD(A)-
TD(B)+
GND
GND
4-20 mA Out
Reset Total
A (-)
B (+)
A (-)
B (+)
GND
Modbus Gnd
Modbus B
Modbus A
+12V
RS-485
To 12 VDC
Supply
Model 485SD9TB
RS-232
RS-485 Converter
Figure 3.9 - RS-485 Network Connectio ns
Heat Flow [BTU meters only]
The BTU meter allows the integration of two 1000 Ω, 3-wire, platinum RTDs with the
ow meter, providing a means of measuring energy consumed in liquid heating and
cooling systems. The RTDs are attached at the factory to a simple plug-in connector
eliminating the possibility of mis-wiring. Simply install the RTDs on or in the pipe as
recommended, and then plug the RTDs into the meter. The surface mount versions
are available in standard lengths of 20 feet (6 meters), 50 feet (15 meters) and 100
feet (30 meters) of attached shielded cable.
BACK OF
CONNECTOR
500 mV
0
p-p
Figure 3.7 - Frequency Output Waveform (Simulated Turbine)
2.) Square-wave frequency - This option is utilized when a receiving
instrument requires that the pulse voltage level be either of a higher
potential and/or referenced to DC ground. The output is a square-wave with
a peak voltage equaling the instrument supply voltage when the SW3 is ON.
If desired, an external pull-up resistor and power source can be utilized by
leaving SW3 OFF. Set SW4 to ON for a square-wave output.
+V
0
Figure 3.8 - Frequency Output Waveform (Square Wave)
RS-485
The RS-485 feature allows up to 126 metering systems to be placed on a single threewire cable bus. Each meter is assigned a unique numeric address that allows all of the
meters on the cable network to be independently accessed. A Modbus RTU command
protocol is used to interrogate the meters. An explanation of the command structure
is detailed in the APPENDIX of this manual. Flow rate, total, signal strength and
temperature (if so equipped) can be monitored over the digital communications bus.
Baud rates up to 9600 and cable lengths to 5,000 feet (1,500 meters) are supported
without repeaters or “end of line” resistors.
1000 Ω
RETURN LINE
RTD #2
1000 Ω
SUPPLY LINE
RTD #1
Figure 3.10 - RTD Schematic
Installation of Surface Mount RTDs
Only use surface mount RTDs on well insulated pipe. Installing the RTD in an
uninsulated area causes inconsistent temperature readings.
Select areas on the supply and return pipes to mount the RTDs. Remove or peel back
the insulation all the way around the pipe in the installation area. Clean an area
slightly larger than the RTD down to bare metal on the pipe.
Place a small amount of heat sink compound on the pipe in the RTD installation
location. See Figure 3.11. Press the RTD rmly into the compound. Fasten the RTD to
the pipe with the included stretch tape.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
TM
Area to Bare Metal Surface
FSRxxxx SERIES
INSTALLATION GUIDE
Route the RTD cables back to the ow meter and secure the cable so that it will not be
pulled on or abraded inadvertently. Replace the insulation on the pipe, ensuring that
the RTDs are not exposed to air currents.
Heat Tape
MINCO
Heat Sink
Clean RTD Mounting
Compound
Figure 3.11 - Surface Mount RTD Installation
Installation of Insertion RTDs
Insertion RTDs are typically installed through ¼ inch (6 mm) compression ttings
and isolation ball valves. Inser t the RTD suciently into the ow stream such that a
minimum of ¼ inch (6 mm) of the probe tip extends into the pipe diameter. Only use
insertion (wetted) RTDs on pipes that are not insulated.
Mount RTDs within ±45 degrees of the side of a horizontal pipe. On vertical pipes the
orientation is not critical. Route the RTD cables back to the ow meter and secure the
cable so that it will not be pulled on or abraded inadvertently.
If the cables are not long enough to reach the meter, route the cables to an elec trical
junction box and add additional cable from that point. Use three-wire shielded cable,
such as Belden® 9939 or equal, for this purpose.
Note: Adding cable adds to the resistance the meter reads and may h ave an eect on absolute
accuracy. If cable is added, ensure that the same length is ad ded to both RTDs to minimize errors
due to changes in cable resistance.
Wiring to Meter
After the RTDs have been mounted to the pipe, route the cable back to the meter
through the middle hole in the enclosure. Connect to the meter inserting the RTD
connector into the mating connector on the circuit board.
ACN
AC IN : 100-240VAC,50/60Hz
DC OUT :
C US
ACL
E167432
95 - 264 VAC
AC Neutral
Signal Gnd.
4-20 mA Out
Reset Total
Modbus Gnd
Modbus B
Modbus A
www.astrodyne.com
PWC-15E
1500mA250V
372
W
C US
VE
D
R
Figure 3.13 - Connecting RTDs
strodyne
+15V / 0.3A
R
+Vo
-Vo
0.15A
R2807
$
TUV
RoHS
PRODUCT SERVICE
RTD 1
RTD 2
Exc.
Exc.
Sig.
Sig.
Gnd.
Gnd.
Shield
Shield
0 to 50°C
TEMP. SET
0 to 100°C
-40 to 200°C
Modbus
TFX Rx
TFX Tx
Downstream
Upstream
-
-
+
+
RTD’s
SUPPLY LINE
MINCO
MINCO
RETURN LINE
RTD #1
RTD #2
Replacement RTDs
Complete RTD kits, including the energy meters plug-in connector and calibration
values for the replacements, are available from the manufacturer.
It is also possible to use other manufacturer’s RTDs. The RTDs must be 1000 Ω
platinum RTDs suitable for a three-wire connection. A connection adapter is available
to facilitate connection to the meter. See Figure 3.14.
WHITE
PIN #8
PIN #6
PIN #4
PIN #2
PIN #5
PIN #3
PIN #1
RED
BLACK
GREEN
BROWN
BLUE
DRAIN
RTD2
RTD1
WHITE
BLACK
RED
DRAIN
Figure 3.12 - Insertion Style RTD Installation
GREEN
BLUE
BROWN
Figure 3.14 - RTD Adapter Connections
Note: It will be necessary to calibrate third par ty RTDs to the meter for proper operation. See the
Appendix of this manual for the calibration procedure.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
PIN#5
PIN#3
PIN#1
PIN#8
PIN#6
PIN#4
PIN#2
TM
FSRxxxx SERIES
PART 4 - STARTUP AND CONFIGURATION
Before Starting the Instrument
Note: Flow meter systems require a full pipe of l iquid before a successful start-up can be
completed. Do not at tempt to make adjustmen ts or change congurations until a full pipe is
veried.
Note: If Dow 732 RTV was used to couple the transducers to the pipe, the adhesive must be
fully cured before readin gs are attempted. Dow 732 requires 24 hours to cure satisfactoril y. If
Sonotemp® acoustic coupling grease was utilized as a coupla nt, curing is not required.
INSTALLATION GUIDE
Instrument Startup
1. Verify that all wiring is properly connec ted and routed, as described in Part 1 of
this manual.
2. Verify that the transducers are properly mounted, as described in Part 2 of this
manual.
3. Apply power. The display briey shows a software version number and then all of
the segments illuminate in succession.
Important!: In order to complete the installation of the ow meter, the
pipe must be full of liquid.
To verify proper installation and ow measurement operation:
1. Go to the SER MENU and conrm that signal strength (SIG STR) is between 5 and
98. If the signal strength is lower than 5, verify that proper transducer mounting
methods and liquid/pipe characteristics have been entered. To increase signal
strength, if a W-Mount transducer installation was selected, re-congure for a
V-Mount installation (standard from factory); if V-Mount was selected, recongure for Z-Mount.
Note: Mounting conguration change s apply only to FST4, FST5 transducer sets.
2. Verify that the actual measured liquid sound speed is within 2% of the value
entered as FLUID SS in the BSC MENU. The measured liquid sound speed (SSPD
FPS and SSPD MPS) is displayed in the SER MENU. The pipe must be full of liquid in
order to make this measurement.
Keypad Programming
Congure units with keypads using the keypad interface or by using the Windows®
compatible software utility. Units without a keypad can only be congured using the
software utility. See Part 5 of this manual for software details. Of the two methods
of conguration, the sof tware utility provides more advanced features and oers
the ability to store and transfer meter congurations between meters. All entries
are saved in non-volatile ash memor y and are retained indenitely in the event of
power loss.
The four-key tactile feedback keypad interface allows the user to view and change
conguration parameters used by the operating system.
Mode
Indicators
Figure 4.1 - Keypad Interface
Keypad
1. Press the MENU key from RUN mode to enter PROGRAM mode. Press the MENU key
in PROGRAM mode to exit from conguration parameter selection and menus. If
changes to any conguration parameters are made, the user is prompted with a
SAVE? when returning to RUN mode. Choose YES to save the new parameters in
program memory.
2. The arrow ▼ keys are used to scroll through menus and conguration
parameters. The arrow keys are also used to adjust parameter numerical values.
3. The ENTER key functions are:
• Pressed from the RUN mode to view the current software version
operating in the instrument.
• Used to access the conguration parameters in the various menus.
• Used to initiate changes in conguration parameters.
• Used to accept conguration parameter changes.
Menu Structure
The ow meter rmware uses a hierarchical menu struc ture. A map of the user
interface is included in the Appendix of this manual. The map provides a visual path
to the conguration parameters that users can access. Employ this tool each time
conguration parameters are accessed or revised.
The seven menus used in the software are as follows:
MenuDescription
BSCBasic. Contains all of the conguration parameters necessary to initially program
CH1Channel 1. Congures the 4-20 mA output.
CH2Channel 2. Congures the type and operating parameters for channel 2 output
SENSensor. Used to select the sensor type (i.e. FST1, FST2, etc.).
SECSecurity. Used for resetting totalizers, returning ltering to factory settings, and
SERService. Contains system settings used for advanced conguration and zeroing
DSPDisplay. Used to congure meter display functions.
the meter to measure ow.
options. Channel 2 parameters are specic to the model used.
Alta Labs, Ene rcept, Enspector, Hawk eye, Trustat, Aerospon d, Veris, and the Veris ‘V ’ logo are tradem arks or registe red trademarks o f Veris Industries, L .L.C. in the USA and/or ot her countries.
Loading...
+ 41 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.