VEGA SWING83 User Manual

Page 1
Level and Pressure
Operating Instruction
VEGASWING 81, 83 fail safe and signal conditioning instrument VEGATOR 526 Ex
VEGATOR 526 Ex
on
Page 2
Contents
Safety information ........................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 3
1.2 Approvals ............................................................................. 5
1.3 Recurring test acc. to WHG ................................................. 6
1.4 Technical data and dimensions ........................................... 7
2 Mounting
2.1 VEGASWING ...................................................................... 13
3 Electrical connection
3.1 VEGASWING ...................................................................... 18
3.2 Connection plan VEGATOR 526 Ex ................................... 19
4 Set-up
4.1 VEGASWING 8… W5␣ Ex.C................................................. 22
4.2 Signal conditioning instrument VEGATOR 526 Ex ............ 23
Contents
5 Diagnosis
5.1 Failure removal ................................................................... 26
The described module must only be installed and operated as described in this operating instruction. Please note that other action can cause damage for which VEGA does not take responsibility.
2 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 3
Product description

1 Product description

1.1 Function and configuration

VEGASWING 81␣ W5␣ Ex.C and 83␣ W5␣ Ex.C fail safe vibrating level switches detect levels of liquids with a viscosity of 0,2 mPa␣ s to
10.000␣ mPa␣ s and a density ␣ 0,5␣ g/cm3. The modular construction enables the use in vessels, tanks and pipelines. Typical applications are overfill protections according to AK5.
Due to the simple and rugged measuring system, VEGASWING can be used virtually unaffected by the chemical and physical features of the liquid. It operates also under arduous conditions such as turbulences, air bubbles, foam generation, build-up and changing medium.
With the mechanical connections thread and flange, VEGASWING 81␣ W5␣ Ex.C or 83␣ W5␣ Ex.C can be used in manifold applications.
For the use in very aggressive products, Hastelloy C4 and ECTFE (halar), Säkaphen and enamel-coatings are used for VEGASWING 81 and 83.
VEGASWING 83 F… can be coated with enamel or up to 1200 mm and Säkaphen up to 4␣ m tube length.
Function
The tuning fork is piezoelectrically energized and vibrates at a resonance frequency of approx. 400␣ Hz. A second piezoelectrical element measures the resonance frequency of the tuning fork. When the tuning fork is covered with the medium, the resonance frequency changes. This frequency change is continuously detected by the integral oscillator and converted into a switching command by the connected signal conditioning instrument. With a mode switch you can determine the switching mode (overfill protection or protection against dry running of pumps).
Note that VEGASWING 8…␣ W␣ Ex.C only meets the fail safe requirements in mode A (overfill protection).
Fail safe / Self-monitoring
A special feature of this instrument configuration VEGATOR 526␣ Ex with VEGASWING 81␣ W5␣ Ex.C or 83␣ W5␣ Ex.C is the integral self-monitoring. It is based on a red­undant microcomputer system in the signal conditioning instrument and includes the complete measuring chain from the tuning fork via the oscillator, the two-wire line as well as the signal conditioning instrument (see also “1.3 Recurring test acc. to WHG“).
For medium temperatures of more than 100°C to 130°C VEGASWING 81␣ W5␣ Ex.C and 83␣ W5␣ Ex.C must be equipped with a temperature adapter.
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 3
The self-monitoring includes two routines:
- the permanent monitoring
- the cyclical self-test.
During the permanent monitoring the signal conditioning instrument induces the oscillator of the sensor to carry out a test cycle in 960␣ msec-cycle. Hence defined test signals are transmitted via the two-wire line to the signal conditioning instrument. Probable errors in the sensor electronics as well as line break, shortcircuit and mechanical damages on the fork are signalled via these test signals.
Page 4
Product description
During the cyclical self-test safety relevant components and procedures (e.g. level relay, CPU) are checked in a very narrow time pattern in the signal conditioning instrument. The fault monitoring of the level relay is made without interruption of the circuit of connected systems.
The integral fault monitoring detects:
- interruption of the connection line to the piezoelements
- strong abrasion on the tuning fork
- break or damage of the tuning fork
- vibration failure
- all safety relevant components (sensor and signal conditioning instrument) are permanently monitored (drift, component failure, time failure etc.)
- meets requirement class 5 (AK5) acc. to DIN␣ V␣ 19␣ 250
- two-wire line on line break and shortcircuit.
Overfill protection with VEGASWING 83 W5 Ex.C
If a so called failure is determined or in case of voltage failure, the electronics takes a defined switching condition, i.e. the failure is signalled via a defined current to the connected signal conditioning instrument VEGATOR 526␣ Ex and the level and fail safe relays de-energize.
Configuration
A measuring system consists of:
- a VEGASWING with oscillator SWING␣ E86␣ Ex
- a signal conditioning instrument VEGATOR 526 Ex
- external instruments switched by VEGATOR (signallers, horns etc.).
Ex-Zone 1
VEGASWING
Two-wire line
4 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
(intrinsically safe circuit)
Not Ex-area
VEGATOR 526 Ex
VEGATOR 526 W Ex
on
Relay for max. signal
Relay for fault
!
signal
Page 5
Product description

1.2 Approvals

VEGASWING 81 W5␣ Ex.C and 83␣ W5␣ Ex.C fail safe are certified acc. to CENELEC Ex­protective regulations. An approval for the use in PTB Zone 0 in the scope of ElexV (Germany) is available.
VEGASWING 81␣ W5␣ Ex.C and 83␣ W5␣ Ex.C fail safe are approved as part of an overfill protection acc. to Water Resources Law (WHG).
Overfill protection acc. to Water Resources Law (WHG)
Level detection with fault monitoring
Instrument Oscillator Level switch General type
VEGASWING SWING E86 Ex 526 Ex Z-65.11-31 81 W5 Ex.C, 83 W5 Ex.C
Level measuring instruments for the use in hazardous areas PTB Zone 0 (Germany)
Level detection with fault monitoring
Instrument Oscillator Conformity Level switch Conformity
certificate VEGATOR certificate PTB-no. VEGATOR BVS-no.
For these applications, the appropriate legal documents must be noted (general type approval and conformity certificates). These are supplied with the appropriate instrument.
DIN The overfill protection meets requirement class 5 (AK5) acc. to DIN␣ V␣ 19␣ 250.
VEGATOR approval
VEGASWING SWING Ex-97.D.2046 526 Ex 93.C.2015 X 81 W5 Ex, E86 Ex 83 W5 Ex
Level measuring instruments acc. to CENELEC for the use in hazardous areas
Level detection with fault monitoring
Instrument Oscillator Conformity Level switch Conformity
VEGASWING SWING Ex-97.D.2046 526 Ex 93.C.2015 X 81 W5 Ex, E86 Ex 83 W5 Ex
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 5
certificate VEGATOR certificate PTB-no. BVS-no.
Page 6
Product description

1.3 Recurring test acc. to WHG

Acc. to the general type approval Z-65.11-31 the sensor must neither be removed nor controlled for the annual recurring test acc. to WHG.
Due to the certified fail-safe condition of the overfill protection (AK5 acc. to DIN␣ V 19␣ 250) the recurring test for the instrument and the transmitter (signal conditioning instrument) is not required. Hence the test of the overfill protection is reduced to the function test of the connected system parts. This can be made, e.g. by interruption of the circuit to the signal or control system.
Note the instruction under “3.4 Connection plan VEGATOR 526 Ex“.
6 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 7
Product description

1.4 Technical data and dimensions

VEGASWING 81 W5 Ex.C and 83 W5 Ex.C
Housing
Housing material PBT (Polyester) Protection IP 66 Cable entry 1 x Pg 13,5 Terminals max. 1 x 1,5 mm
Mechanical connection
Thread G 1 A or 1“ NPT
- material 1.4571 (V4A) or Hastelloy C4 (2.4610) Flanges DIN and ANSI from DN 50
- material 1.4571, 1.4571 with Hastelloy C4 plated DN 50 PN 40 steel enamelled
Tuning fork
Material 1.4581, Hastelloy C4, Hastelloy C4 enamelled
1.4581 with Säkaphen or ECTFE plated
Extension tube (VEGASWING 83 W5 Ex)
Material 1.4571, Hastelloy C4, Hastelloy C4 enamelled
1.4571 with Säkaphen or ECTFE plated
Length
- 1.4571, Hastelloy C4 200 mm … 4000 mm
- Hastelloy C4 enamelled 200 mm … 1200 mm
- 1.4571, Halar 200 mm … 1200 mm
- 1.4571, Säkaphen 200 mm … 4000 mm
2
Weight
Plastic housing approx. 1,5 kg Tube extension approx. 0,11 kg/m (VEGASWING 83 W5 Ex.C)
Ambient conditions
Ambient temperature on housing -40°C … +65°C Storage and transport temperature -40°C … +70°C Medium temperature -40°C … +100°C Medium temperature with temperature adapter of 1.4571 (option) -40°C … +130°C
Operating pressure
Operating pressure max. 25 bar
(VEGASWING 83 W5 Ex.C: with locking ARV2EX0.C… to 4 bar, otherwise unpressurized)
Test pressure max. 40 bar
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 7
Page 8
Product description
Medium
Viscosity
- dynamic 0,2 … 10.000 mm2/s Density 0,5 or 1,0 g/cm3 changeable
Function
Mode A/B-mode via signal conditioning instrument
Integration time approx. 0,5 s Meas. frequency approx. 400 Hz
CE-conformity
VEGASWING 81 W5 Ex.C and 83 W5 Ex.C vibrating level switches meet the protective regulations of EMVG (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following standards: EMVG Emission EN 50 081 - 1: 1992
Susceptibility EN 50 082 - 2: 1992
NSR EN 61 010: 1993
A: Max. detection, overfill protecion (fail␣ safe) B: Min. detection, dry run protection (not fail safe)
Approvals VEGASWING 8… W5 Ex.C
Water Resources Law (WHG)
Approval as overfill protection acc. to WHG Ambient temperature on the housing -40°C … +65°C Medium temperature -40°C … +100°C
- with temperature adapter to 130°C Operating pressure max. 25 bar
(general type approval Z-65.11-31)
Oscillator
Two-wire output (SWING E86 Ex)
Supply voltage 12 … 18,9 V DC (via VEGA-signal conditioning
Output two-wire output Required signal conditioning instrument VEGATOR 526 Ex Current consumption 4 … 25 mA Protection class II Overvoltage category II Modes definition via signal conditioning instrument Classification EEx ia IIC T4
8 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
instrument VEGATOR 526 Ex)
Page 9
Product description
Signal conditioning instrument VEGATOR 526 Ex
General
Series module unit for carrier BGT 596 Ex Dimensions W = 50,8 mm (10 TE), H = 128,4 mm, D = 162 mm Weight approx. 440 g
Ambient conditions
Ambient temperature -20°C … +60°C Storage and transport temperature -40°C … +70°C
Power supply
Operating voltage UN = 24 V AC (20 … 28 V AC); 50/60 Hz
Power consumption approx. 11 VA or 8 W Electrical connection multipoint connector DIN 41 612,
Electrical protective measures
Protection class II (in housing type 589-10 Ex) Overvoltage category II Protection
- mounted in housing type 589… IP 30
Protection (mounted into carrier BGT 596 Ex with Ex-module)
- front side (completely equipped) IP 30
- upper and lower side IP 20
- wiring side IP 00
UN = 24 V DC (16 … 60 V DC
series F (d, z) 32-pole
Inputs
Number 1 Data transmission analogue, 4 … 20 mA Current limitation 24 mA (permanently shortcircuit proof) Sensor supply voltage approx. 15 … 18 V DC Connection line 2-wire (screened) Resistance per conductor max. 35
Function
Integration time
- switch on delay 0,5 s
- switch off delay 1,0 s
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 9
Page 10
Product description
Relay output (fail safe)
Number 1 switching relay (closing contact, floating) Mode A/B-switch
A - max. detection or overfill protection
B - min. detection or dry run protection Contact 1 spdt each Contact material AgNi and Au plated Turn-on voltage min. 12 V DC
max. 250 V AC, 48 V DC Switching current min. 10 mA DC
max. 2 A AC, 1 A DC Breaking capacity max. 125 VA AC, 60 W DC
Fail safe relay (not fail safe)
Number 1 fail safe relay (spdt, floating) Contact material AgCd and au plated Turn-on voltage min. 10 mV
max. 250 V AC, 48 V DC Switching current min. 10 µA
max. 2 A AC, 1 A DC Breaking capacity max. 125 VA AC, 54 W DC
Indicating elements
LED in front plate
- green on operating voltage on
- yellow switch point control
- red fault signal Set-up key (green) ready for set-up
Approvals
Max. values UO = 18,9 V; IK = 95 mA; P = 647 mW Characteristics linear
EEx ia IIC EEx ia IIB Max. permissible outer inductance (mH) 0,5 1 1,5 2 0,5 1 1,5 2 Max. permissible outer capacitance (nF) 52 40 27 15 200 190 177 160
EEx ib IIC EEx ib IIB Max. permissible outer inductance 4 mH 15 mH Max. permissible outer capacitance 240 nF 1,1 µF
The intrinsically safe circuits are reliably galvanically isolated from all not-intrinsically safe circuits up to a peak value of 375␣ V of the nominal voltage.
10 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 11
Product description
162
100
12
4
128,4
50,8
10 TE
Electrical connection
Mounted in
- carrier BGT 596 Ex 32-pole multipoint connector, series F (d, z)
- in housing type 589… screw terminal, max. für 1,5 mm
with holes
2
CE-conformity
VEGATOR 526 Ex signal conditioning instrument meets the protective regulations of EMVG (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following standards: EMVG Emission EN 50 081 - 1: 1993
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
Potential stages and galvanic isolation
Sensor Ex-range
Meas. data input
P1
Microcomputer control and conditioning
P3 P4 P5
Relay contacts
Voltage supply
Power supply unit
P2
Relay contacts
P1 … P5 Potential stages
Fail safe relay
Level relay
Signal conditioning instrument VEGATOR 526 Ex
Circuit board 100 x 160 x 1,5 European size
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 11
Multiple plug
Page 12
VEGASWING 81 W5 Ex.C and 83 W5 Ex.C WHG
Product description
VEGASWING 81 … VEGASWING 83 …
VEGASWING 81 G1 / N1
Thread G 1 A or 1"␣ NPT
153
128
29
Mounting boss acc. to DIN (accessory)
SW 41
100
26
VEGASWING 81 F
ø60
VEGASWING 81 F
157
118
enamelled
170
~130
75
Flange DN 50
25
25
26
80x110
VEGASWING 83 G1 / N1
Thread G 1 A or 1"␣ NPT
25
Switch point
L (min. 200, max. 3000 mm)
Switch point
20
65
~30
Temperature adapter of 1.4571 (option)
51
ø41
ø29
100
26
SW 41
SW 60
G11/2 A or 11/2"␣ NPT
VEGASWING 83 F
Flange DN 50
39
L (min. 200, max. 3000 mm)
39
Switch point
25
VEGASWING 83 F
enamelled
Locking unpressurized
SW 60 SW 60
L (min. 200, max. 1200 mm)
49
20
G11/2 A or 11/2"␣ NPT
26
Switch
25
point
Locking up to 4 bar
12 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 13
Mounting

2 Mounting

2.1 VEGASWING

Generally VEGASWING can be mounted in any individual position. The instrument must just be mounted such, that the tuning fork is at the height of the requested switch point. However note that VEGASWING is only fail safe as overfill protection (mode A). Note the following for mounting:
Transport
Do not hold VEGASWING on the tuning fork. Especially on flange and tube versions the tuning fork can be damaged by the instrument weight.
Switch point
Laterally on the tuning fork there are markings for the switch point in vertical installation in relation to the medium water. Note when mounting VEGASWING that the marking is at the height of the requested level. Note that the switch point of the instrument is shifted when the medium has a density deviating from water (water = 1,0 g/cm3). Adjust the appropriate density range on VEGATOR 526␣ Ex. See also “4.2 Signal conditioning instrument VEGATOR 526␣ Ex“ (see fig. 2.1).
Vertical mounting
from top, from bottom
Switch point with low density
Switch point
approx. 25 mm
Switch point with higher density
Switch point
75 mm
Horizontal mounting
Switch point
recommended installation position for adhesive products:
Marking on top
Switch point
Fig. 2.2
Adhesive mediums
For horizontal mounting in adhesive and viscous mediums, the surface of the tuning fork should be vertical to reduce build-up on the fork (see fig. 2.2). The position of the tuning fork is marked on the hexagon of VEGASWING. Hence you can check the position of the tuning fork during mounting. When the hexagon is placed to the seal, the thread can be turned by half a turn. This is sufficient to reach the recommended installation position.
For adhesive and viscous mediums the tuning fork should protrude into the vessel to avoid build-up. Sockets for flanges and mounting bosses should therefore not exceed a certain length. With flange versions, the socket should have a length of max. 40␣ mm, with threads max. 30␣ mm. The slide switch for the density adjustment must be set to position 2.
Fig. 2.1
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 13
Page 14
Mounting
Pressure
In case of overpressure or gauge pressure in the vessel, the mounting boss must be sealed on the thread. Cover the thread with Teflon tape, hemp etc. or use a seal ring.
Vibrations
Extreme vibrations and shocks e.g. by stirrers and turbulences in the vessel can energize the extension tube of VEGASWING 83 to resonance frequencies. This causes an increased wear on the upper weld joint.
Therefore provide a suitable support or fixing directly above the tuning fork to fasten the extension tube (see fig. 2.3).
This is mainly valid for applications in Ex-area zone 0 or acc. to VbF. Note that the tube is not bent by this measure.
Humidity
Turn the cable entry of horizontally mounted instruments to the bottom to avoid humidity ingress. The plastic housing is therefore rotational by approx. 330°. For vertically mounted instruments loop the connection line to the instrument housing to the bottom so that rain and condensation water can drain off. This is mainly valid for mounting outside, in humid areas (e.g. by cleaning processes) or on cooled or heated vessels (see fig. 2.4).
Fig. 2.4
Lockings
For VEGASWING 83…␣ G and 83…␣ N a locking is available for height adjustment. The switch point can be adjusted after mounting and modified afterwards.
Adjust the switch point of the tuning fork to the requested height. By tightening the pressure screw you fix the tube to this position.
Fig. 2.3
Stirrers
Due to stirrers etc. the level switch can be subjected to strong lateral forces. For this reason the extension tube of VEGASWING 83␣ W5␣ Ex.C should not be too long, better check if it is possible to mount a level switch VEGASWING 81␣ W5␣ Ex.C laterally in horizon­tal position.
14 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
For pressure applications you have to secure the pressure screw with the fixing screw.
VEGASWING 83…␣ G and 83…␣ N can be used in vessels with up to 4␣ bar pressure. Therefore use the locking (G␣ 11/2␣ A, 1/2“␣ NPT) up to 4␣ bar (ARV2EX0.C…). Note the operating instruction of the locking.
Page 15
Mounting
Cable entries
Use cable with round cross section area of conductor and tighten the cable entry. The cable entry is suitable for cable diameters from 5␣ mm to 12␣ mm.
Lateral load
Note that the vibrating element is not subjected to lateral forces. Mount the instrument on a position in the vessel where no interfering influences such as e.g. by stirrers, filling openings etc. occur. This is mainly due for instrument types with extension tube (see fig. 2.5). The surfaces of the tuning fork should be in parallel to the product movement so that there is only little resistance by the tuning fork of VEGASWING in case of product movements.
VEGASWING with enamel
Treat the instruments with enamel-coating very carefully and avoid strong shocks or impacts. Unpack VEGASWING just before mounting. Insert VEGASWING carefully into the provided vessel opening and avoid any contact with sharp-edged vessel parts.
Flow
(e.g. in tubes) When mounting in tubes or in vessels with certain flow direction, VEGASWING should be mounted such that the surfaces of the tuning fork must be along the flow direction.
Mounting boss
VEGASWING has a defined beginning of thread. This means that each VEGASWING is always in the same position after screwing in. Hence remove the supplied seal from the thread of VEGASWING. This thread is not necessary when using a mounting boss. Screw VEGASWING into the mounting boss.
You can determine the later position of VEGASWING already before welding (see fig.
2.2). Mark the appropriate position of the mounting boss. Before welding, you have to unscrew VEGASWING and remove the rubber ring from the mounting boss.
Optionally a mounting boss is available which is already provided with the marking. Weld this mounting boss with the marking to the top (see fig. 2.6).
Fig. 2.5
Chemical resistance
For applications in Ex-areas or acc. to WHG and VbF, VEGASWING must
only be used in liquids to which the materials of the tuning fork system are sufficiently chemically resistant.
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 15
Marking (option)
Fig. 2.6
Page 16
Mounting
8 0
Installation examples
Mounting not suitable for adhesive products
Fig. 2.7
Signal conditioning instrument VEGATOR 526 Ex
VEGATOR can be either mounted with a module into a carrier BGT 596␣ Ex or into a single housing type 589-10␣ EX.
Module-Ex (Ex 32)
for VEGATOR 526 Ex Multipoint connector DIN 41 612, series F, 32­pole (d, z) with coded pins, Ex-separating chamber and mounting material for mounting in carrier BGT 596 Ex.
The multipoint connection is available as follows:
- Wire-Wrap standard connection 1,0 x 1,0 mm
- Plug connection 2,8 x 0,8 mm
- Termi-Point standard connection 1,6␣ x␣ 0,8␣ mm
- Soldering connection
- Screw terminals 2 x 0,5 mm
2
For further information concerning mounting, see the operating instruction of the carrier.
Coding
The multiple plug of the instrument is provided with a coding hole to avoid interchanging of the various instruments.
Due to a fixed coded pin on the Ex-module, it is ensured that only Ex-instruments can be inserted.
2 4 6
8 10 12 14 16 18 20
(top view)
22 24 26 2 3 32
Fig. 2.8
Coded pin here
Coding of the multipoint connector
Mounting carrier
Mount the appropriate module (Ex-version) to your carrier BGT 596␣ Ex. Wire the connection of the multipoint connector according to the wiring plans on the following pages.
Transparent cover
The front plate of VEGATOR 526 Ex can be provided with a lockable transparent cover to avoid unauthorized adjustment. The cover is attached to the instrument.
16 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 17
Mounting
Ex-version
Ex-separating chamber
To ensure sufficient “Air and creeping distances“, an Ex-separating chamber must be mounted to the connections of VEGATOR. Loop the lines through the Ex-separating chamber and connect the lines. Fasten the Ex-separating chamber with the lower holding screw. Note the operating instruction of the carrier BGT 596␣ Ex.
Module with multipoint connector acc. to DIN 41␣ 612 for carrier (back view)
d b z
2 4 6
8 10 12 14 16 18 20 22 24 26
28 30 32
Mounting in carriers
When you mount your VEGATOR with Ex­approval in a carrier, you have to use a VEGA­Ex-module. Keep a distance of at least 10 mm (2 TE) to modules of other manufacturers. When you want to mount VEGATOR to the complete left in the carrier, you have to mount a blind cover of at least 20 mm (4 TE) in front of the module of the instrument.
Blind cover
Fig. 2.10
Protection with Ex-applications
In Ex-applications a protection of IP 20 must be kept in mounted condition. Hence cover the gaps or free modules from the front by appropriate blind covers.
Ex-separating chamber
Fig. 2.9
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 17
Page 18

3 Electrical connection

Electrical connection

3.1 VEGASWING

Danger
Before starting connection work, switch off the power supply.
Connect mains voltage acc. to the following connection diagrams.
Note
Generally use screened cable. The screening of the cable must be earthed at the sensor side (VEGASWING) via terminal 3.
Overvoltage arresters are recommended in case of danger of overvoltages. For connection note the operating instruction of the overvoltage arresters.
Generally connect VEGASWING to vessel ground (PA) or in case of plastic vessels to the next earth potential. Laterally on the hexagon of the mounting boss or flange there is a thread (screw M4␣ x␣ 5). This connection is used to drain off electrostatic charges. Normally in case of metal vessels, the sensor has already connection to vessel earth via the mechanical connection.
For Ex-applications the mounting regulations for hazardous areas must be noted.
Two-wire output (SWING E86 Ex)
For connection to a signal conditioning instrument VEGATOR 526␣ Ex
Voltage supply via the connected signal conditioning instrument VEGATOR (for further information see technical data)
When VEGASWING is used in Ex-area or as part of an overfill protection acc. to WHG, note the regulations of the conformity certificates, type approvals and the general construction approval, see “1.2 Approvals“.
18 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 19
Electrical connection

3.2 Connection plan VEGATOR 526 Ex

The following figures show the currentless condition.
Coordination of the terminal socket or the multipoint connector
d b z
2 L1 (+)
N (-)
2
8
8
18
20
30 32
10
Fail safe relay
Level relay
+
-
Recurring test acc. to WHG
Supply voltage
Meas. data input (intrinsically safe signal circuit)
Terminal socket with multipoint connector (part of housing type 589-10 Ex) Type 589… (WHG etc.)
E Ex i
z 2
4 6
8 10 12 14 16 18
20 22 24
28 30 32
z
d
2 4 6
8 10 12 14
16 18
Coded
20
pin
22 24
28 30 32
d
Separating wall
Key (opener) to interrupt the circuit to the signalling or control facility
18
20
Not Ex-area
Ex-area
Sensor 1
1
Overvoltage arrester
Overvoltage arrester
2 3
Oscillator of VEGASWING 8… W5 Ex
Connection with overvoltage arresters
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 19
Page 20
Connection examples for overvoltage arresters
Vessel without cathodic corrosion protection
Electrical connection
VEGASWING
1
E86 Ex
2
Overvoltage arrester type B62 - 27 Ex 0 or B62 - 27 Ex
Outer isolation
Line length
␣ 50 m
Screen
min. 4 mm
Overvoltage arrester
A1
Typ
B62-27
A2
2
Cu
Ex-area
Metal housing isolated mounted or plastic housing
Outer isolation
E1 E2
Screen
Earth / PA
Not Ex-area (control room)
optionally overvoltage arrester
E1
Typ
B62-27G
A1 A2E2
d30 d32
VEGATOR 526 Ex
connection terminal
External earth terminal
to potential
equalisation line
Note :
- Use suitable cable, optionally with metal cover or screen. Cable with metal cover or screen (as shown) do not require a potential equalization along the complete intrinsically safe circuit. Just connect the screen or metal cover to the overvoltage arrester (test voltage ␣ 500␣ V).
- The capacity of the screen must correspond at least to that of a 1,5 mm2 Cu-line.
20 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 21
Electrical connection
Vessel with cathodic corrosion protection
VEGASWING
E86 Ex
External earth terminal
Overvoltage arrester type B62 - 27 Ex 0 or B62 - 27 Ex
Outer isolation
1 2
Line length
␣ 50 m
Screen
Cathodic tank protection 2 … 24 V (object voltage)
min. 4 mm
Overvoltage arrester
A1
Typ
B62-27
A2
2
Cu
Ex-area
Metal housing isolated mounted or plastic housing
Outer isolation
E1 E2
Screen
Earth / PA connection terminal
Not Ex-area (control room)
optionally overvoltage arrester
A1
E1
Typ
B62-27G
A2E2
d30 d32
VEGATOR 526 Ex
Note:
- Use suitable cable, optionally with metal cover or screen. Screen or metal cover must only be connected to the overvoltage arrester. Screen or metal cover must not be earthed.
- The cable must have an outer isolation.
- The capacity of the screen must at least correspond to that of a 1,5 mm2 Cu-line.
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 21
Page 22
Set-up

4 Set-up

4.1 VEGASWING 8… W5 Ex.C

Functional table
The following table gives a survey on the switching conditions dependent on the adjusted mode and the level.
Level Level Relay Fail safe Fail safe
relay control relay control (closing lamp (spdt) lamp contact) VEGATOR
526 Ex
Mode A
(fail safe)
d20d18
Relay Relay
energized energized
d20d18
Relay Relay de-energize
Mode B
(not
d energized
d20d18
fail safe) Relay Relay
energized energized
d20d18
Relay Relay de-energized
energized
Failure of the voltage supply Mode A / B individual
d20d18
Relay Relay de-energized de-energized
Failure Mode Acc. to Acc. to
A / B individual mode mode
“see “see Relay above“ above“
de-energized
z8z10 d8
z10z8 d8
z8z10 d8
z8z10 d8
z8z10 d8
z8z10 d8
22 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 23
Set-up
VEGASWING 8… W5 Ex.C
1
2
3
4
1 Type plate (oscillator) 2 Terminals 3 Oscillator 4 Housing
No pre-adjustments must be made on VEGASWING 8…␣ W5␣ Ex.C.

4.2 Signal conditioning instrument VEGATOR 526 Ex

Adjustment elements
1
2
VEGATOR 526 W Ex
3
on
6
A B
3 2 1
4 5
1
7
8
9
1 Screw 2 LED-relay control lamps 3 Set-up key 4 LED-fault signal 5 LED-supply voltage 6 Multiple plug 7 Hook switch:
Mode A (overfill protection) fail safe Mode B (protection against dry running of pumps) not the fail safe
8 Slide switch:
Position 3 – Density of the product 0,5 … 0,85 g/cm Position 2 – Density of the product >0,85 g/cm Position 1 – not allowed
9 Front plate
3
3
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 23
Page 24
Set-up
Pre-adjustments
Before connecting VEGATOR 526␣ Ex to voltage supply, the mode of the meas. system and the sensitivity of the sensor must be adjusted. The adjustment elements are on the board of the signal conditioning instrument.
Mode
Mode A or B can be adjusted by means of the hook switch (6).
Mode A (fail safe)
means with covered vibrating element
- Relay de-energizes
- LED-relay output lights means with uncovered vibrating element
- Relay energizes
- LED-relay output extinguishes
Mode B (not fail safe)
means with covered vibrating element
- Relay energizes
- LED-relay output extinguishes means with uncovered vibrating element
- Relay de-energizes
- LED-relay output lights
Density of the product
With the slide switch (7) you can modify the sensitivity of the meas. system. Note that the switch is never in position 1.
Position 3 - Density of the product
0,5␣ …␣ 0,85␣ g/cm
Position 2 - Density of the product
>␣ 0,85␣ g/cm
Position 1 - not allowed
The information on the position of the switch point (see “2 Mounting“) relates to water in conjunction with the slide switch on position 2.
3
3
In case of products with deviating density, this switch point shifts dependent on the density and the installation to the direction of the housing or tuning fork end (see following table).
A density change by 0,1␣ g/cm3 effects e.g. a shifting of the switch point by approx. 2,5␣ mm.
To avoid faulty switching by probable adhesions, the slide switch (7) must be set to position 2 in case of adhesive, high viscosity products.
By means of the slide switch (7) you can also detect products below 1␣ g/cm3 (0,5␣ …␣ 0,85␣ g/ cm3). In this case choose position 3. The switch point is then again at the height of the marking (25␣ mm above the tip of the fork).-
Note
- Modifications in the switch position are only effective after the next voltage change-over.
- Modifications can be carried out any time after the set-up.
Control lamps
LED in the front plate show on/off, switch condition and failure.
Green
- mains control lamp (5)
- mains voltage on, instrument in operation
Red
- signal lamp (4)
- failure on the sensor circuit by failure of the sensor of line defect.
- when the fail safe relay is de-energized, the red signal lamp lights.
24 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
Page 25
Set-up
Yellow
- relay control lamp (2)
- the yellow control lamp reacts acc. to the adjusted mode (A/B).
- generally the relay control lamp shows the safe (currentless) condition of the relays.
- a dark control lamp means that the relay is in activated (current) condition.
Set-up
After connecting VEGASWING 8…␣ W5␣ Ex and inserting VEGATOR 526␣ Ex in the carrier or the single housing, the meas. system is ready for operation.
• Screw the signal conditioning instrument.
With the lowered set-up key (3) you can set­up the level detection.
Yellow LED lights (= level relay de-energized)
VEGATOR 526 W Ex
Green LED in set-up key flashes (= set-up not yet carried out)
on
Green LED lights (= supply voltage on)
Ready condition before set-up
Pre-requirements to carry out the set-up
- VEGASWING is mounted in its final position and the tuning fork is not covered with the medium
- the failure-LED is dark
• Push the lowered set-up key (3) for approx. 4␣ …␣ 6␣ seconds.
With the set-up an adaption of VEGATOR to VEGASWING is made. The actual resonance frequency is detected by VEGATOR as current and saved as reference value.
Condition up to approx. 10 secs. after set-up
Yellow LED lights
VEGATOR 526 Ex
Green LED in set-up key lights
on
Green LED lights
Condition approx. 10 secs. after set-up
VEGATOR 526 Ex
on
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 25
Green LED lights
Page 26

5 Diagnosis

VEGATOR 526 Ex
on
Diagnosis

5.1 Failure removal

Via the condition of the LED in the front plate of VEGATOR two failures can be detected:
- failure during set-up
- failure detection within the self-monitoring.
In both cases the level and fail safe relays de­energize.
In case of set-up failures, a new start must be made (switch off voltage supply and switch on again) and repeat the set-up.
When failures are detected, first of all switch off the voltage supply, then remove the failure (e.g. line break) and switch on the voltage supply again. It is not necessary to repeat the set-up.
Failure Reason / Removal
Before pushing the set-up key (3) Check the following failure reasons:
- line break
- interpolation of the sensor lines
lights
- short-circuit
- undefined position of the hook switch (6) or the slide switch (7)
- electronics defect on VEGASWING or on VEGATOR
off
lights
lights
In case of an electronics defect, call our service department.
After set-up - set-up key was not pushed long enough
Repeat the set-up (push set-up key (3) for 4␣ …␣ 6␣ s)
lights
VEGATOR 526 Ex
flashes
on
26 VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe
off
lights
Page 27
Diagnosis
Failure Reason / Removal
After pushing the set-up key (3) - set-up was carried out with covered tuning fork.
lights
VEGATOR 526 Ex
tuning fork.
- set-up key was pushed too long (>␣ 6␣ s)
Carry out a new start and repeat the set-up with uncovered
lights for approx. 2 s
on
off
lights
Carry out a new start and repeat the set-up. Note the correct period for pushing the set-up key (3).
During operation Check the following reasons:
- line break
- short-circuit
- electronics defect
VEGATOR 526 Ex
on
lights
aus
leuchtet
leuchtet
In case of an electronics defect, call our service department.
VEGASWING 81 W5 Ex.C, 83 W5 Ex.C fail safe 27
Page 28
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail info@vega-g.de
ISO 9001
Technical data subject to alterations 2.21 489 / June ’97
Loading...