Vanguard EZCT-2000B User Manual

TM
EZCT-2000B
DIGITAL CURRENT-TRANSFORMER TESTER
USER’S MANUAL
1520 S. Hellman Ave. Ontario, California 91761, USA
TEL: (909) 923-9390 FAX: (909) 923-9391
March 2014
Revision 1
EZCT-2000B USER’S MANUAL REV 1
SAFETY SUMMARY
FOLLOW EXACT OPERATING PROCEDURES
Any deviation from the procedures described in this User’s Manual may create one or more safety hazards, may damage the EZCT-2000B, or cause errors in the test results. Vanguard Instruments Company, Inc. assumes no liability for unsafe or improper use of the EZCT-2000B.
All safety precautions provided in this manual must be observed during all phases of testing including test preparation, test lead connection, actual testing, and test lead disconnection.
SAFETY WARNINGS AND CAUTIONS
The EZCT-2000B can produce a voltage up to 2,000 Vac that can cause severe injury and/or equipment damage. Due to this reason, the EZCT-2000B shall be used only by trained operators.
The EZCT-2000B’s X output terminals are rated to 2,000 Vac working voltage. Any voltage above 2,000 Vac will damage the input circuitry. Please see section 3.2 for further information.
All devices under test shall be off-line and fully isolated. Never attempt to test any current transformer still connected to a circuit. All current transformer terminals shall be isolated before conducting any test with the EZCT-2000B.
Always ground the EZCT-2000B to a substation ground before connecting the test cables to a transformer.
DO NOT MODIFY TEST EQUIPMENT
To avoid the risk of introducing additional or unknown hazards, do not install substitute parts or perform any unauthorized modification to any EZCT-2000B test unit. To ensure that all designed safety features are maintained, it is highly recommended that repairs be performed only by Vanguard Instruments Company factory personnel or by an authorized repair service provider. Unauthorized modifications can cause safety hazards and will void the manufacturer’s warranty.
WARNING
Do not remove test leads during a test. Failure to heed this warning can result in electrical shock to personnel and damage to the equipment.
i
REV 1 EZCT-2000B USER’S MANUAL
TABLE OF CONTENTS
CONVENTIONS USED IN THIS DOCUMENT .................................................................................................................... 1
1.0 INTRODUCTION ................................................................................................................................................ 2
1.1 General Description and Features ............................................................................................................... 2
1.2 Furnished Test Accessories .......................................................................................................................... 4
1.3 Technical Specifications ............................................................................................................................... 5
1.4 EZCT-2000B Controls and Indicators............................................................................................................ 6
2.0 PRE-TEST SETUP ................................................................................................................................................ 8
2.1 Operating Voltages ...................................................................................................................................... 8
2.2 LCD Screen Contrast Control ....................................................................................................................... 8
2.3 Printer Paper Control ................................................................................................................................... 8
2.4 Printer Paper ................................................................................................................................................ 8
3.0 OPERATING PROCEDURES .............................................................................................................................. 10
3.1 EZCT-2000B Cable Connections ................................................................................................................. 10
3.2 EZCT-2000B X Input Voltage Warning ....................................................................................................... 15
3.3 Performing Tests ........................................................................................................................................ 16
3.3.1. Entering Test Record Header Information ........................................................................................ 16
3.3.2. Performing Resistance, Excitation, and Ratio Tests .......................................................................... 19
3.3.3. Performing a CT Burden Test ............................................................................................................ 30
3.3.4. Performing an Insulation Resistance Test ......................................................................................... 35
3.3.5. Performing a Current Source Test .................................................................................................... 40
3.4 Working With Test Records ....................................................................................................................... 42
3.4.1. Restoring and Printing a Test Record From Flash EEPROM .............................................................. 42
3.4.2. Restoring and Printing a Test Record From a USB Flash Drive ......................................................... 46
3.4.3. Printing a Restored Test Record ....................................................................................................... 48
3.4.4. Printing a Directory of Test Records Stored in the EZCT-2000B’s Memory ...................................... 50
3.4.5. Printing a Directory of Test Records Stored in a USB Flash Drive ..................................................... 53
3.4.6. Copying Test Records to a USB Flash Drive ....................................................................................... 55
3.4.7. Erasing Test Records From the Flash EEPROM ................................................................................. 58
3.4.8. Erasing Test Records From a USB Flash Drive ................................................................................... 61
3.5 Working With Test Plans ............................................................................................................................ 64
3.5.1. Extracting the Test Plan From a Test Record .................................................................................... 64
3.5.2. Printing a Directory of Test Plans Stored in the EZCT-2000B’s Memory .......................................... 66
3.5.3. Printing a Directory of Test Plans Stored in a USB Flash Drive ......................................................... 68
3.5.4. Printing a Test Plan ........................................................................................................................... 70
3.5.5. Erasing Test Plans From the Flash EEPROM ...................................................................................... 72
3.5.6. Erasing Test Plans From a USB Flash Drive ....................................................................................... 75
3.5.7. Loading a Test Plan from the EZCT-2000B’s Flash EEPROM ............................................................. 78
3.5.8. Loading a Test Plan from a USB Flash Drive ...................................................................................... 80
3.5.9. Running a Test Using a Loaded Test Plan ......................................................................................... 82
3.5.10. Unloading a Test Plan from the Working Memory ........................................................................... 87
4.0 CHANGING SETUP PARAMETERS .................................................................................................................... 88
4.1 Setting the Knee Point Marker .................................................................................................................. 88
4.2 Enabling and Disabling the Buried CT in Transformer Delta Option .......................................................... 90
4.2.1. Enabling the Buried CT in Transformer Delta Option ....................................................................... 90
4.2.2. Disabling the Buried CT in Transformer Delta Option ...................................................................... 91
4.3 Setting the Clock ........................................................................................................................................ 95
5.0 DIAGNOSTICS, VERIFICATION, AND TROUBLESHOOTING .............................................................................. 97
5.1 Performing a Diagnostics Test .............................................................................................
5.2 Verifying the EZCT-2000B’s V
5.3 Verifying the EZCT-2000B’s I
Sense Circuit Using an External Meter ...................................................... 99
x
Sense Circuit Using an External Meter ..................................................... 100
x
...................... 97
ii
EZCT-2000B USER’S MANUAL REV 1
5.4 Quickly Verifying the EZCT-2000B’s Turns Ratio Circuit .......................................................................... 101
5.5 Troubleshooting Guide ............................................................................................................................ 102
6.0 UPGRADING FIRMWARE ............................................................................................................................... 103
7.0 Appendix A - Calculating Turns Ratio on a Shunt Reactor ............................................................................ 107
LIST OF TABLES
Table 1. EZCT-2000B Technical Specifications ............................................................................................................... 5
Table 2. Functional Descriptions of EZCT-2000B Controls and Indicators ..................................................................... 7
Table 3. Descriptions of Tabulated Test Results Elements .......................................................................................... 28
Table 4. Description of Test Plan Elements ................................................................................................................. 71
LIST OF FIGURES
Figure 1. EZCT-2000B Controls and Indicators .............................................................................................................. 6
Figure 2. Typical EZCT-2000B Excitation and Ratio Test Cable Connection ................................................................. 10
Figure 3. EZCT-2000B Insulation Resistance Test Cable Connection ........................................................................... 11
Figure 4. EZCT-2000B CT Load Burden Test Cable Connection .................................................................................... 12
Figure 5. EZCT-2000B Typical Current Source Connection .......................................................................................... 13
Figure 6. Bushing CT Connection on Delta Transformer .............................................................................................. 14
Figure 7. Bushing CT Connection on Y Transformer .................................................................................................... 14
Figure 8. Sample CT Name Plate .................................................................................................................................. 15
Figure 9. Typical EZCT-2000B Tabulated Report Printout ............................................................................................ 27
Figure 10. Typical EZCT-2000B Graphic Report with Multiple Plot Curves .................................................................. 29
Figure 11. EZCT-2000B Typical Burden Test Results Report Printout .......................................................................... 34
Figure 12. EZCT-2000B Typical Insulation Resistance Test Results Report Printout .................................................... 39
Figure 13. Typical Internal Test Record Directory Printout ......................................................................................... 52
Figure 14. Typical USB Flash Drive Record Directory Printout ..................................................................................... 54
Figure 15. Typical EZCT-2000B Flash EEPROM Test Plan Directory Printout ............................................................... 67
Figure 16. Typical USB Flash Drive Test Plan Directory Printout ................................................................................. 69
Figure 17. Typical Test Plan Printout ........................................................................................................................... 71
Figure 18. Graphic Report Showing Knee Point Marker .............................................................................................. 89
Figure 19. Buried CT in a Delta Transformer Illustration 1 .......................................................................................... 93
Figure 20. Buried CT in a Delta Transformer Illustration 2 .......................................................................................... 93
Figure 21. EZCT-2000B V Figure 22. EZCT-2000B I
Figure 23. EZCT-2000B Turns Ratio Verification Test Connections ........................................................................... 101
Figure 24 .................................................................................................................................................................... 107
Figure 25 .................................................................................................................................................................... 108
Figure 26 .................................................................................................................................................................... 109
Figure 27 .................................................................................................................................................................... 110
Figure 28 .................................................................................................................................................................... 111
Figure 29 .................................................................................................................................................................... 112
Figure 30 .................................................................................................................................................................... 113
Figure 31 .................................................................................................................................................................... 113
Verification Test Connections ............................................................................................ 99
x
Verification Test Connections ............................................................................................ 100
x
iii
REV 1 EZCT-2000B USER’S MANUAL
CONVENTIONS USED IN THIS DOCUMENT
This document uses the following conventions:
A key or switch on the EZCT-2000B is indicated as
[KEY].
Menu options are referenced as (MENU OPTION).
Screen and menu names are referenced as “SCREEN/MENU NAME”.
The terms “test record” and “test shot” are used interchangeably.
The terms “USB Flash drive” and “USB thumb drive” are used interchangeably.
EZCT-2000B LCD screen output is shown as:
1. OPTION 1
2. OPTION 2
3. OPTION 3
4. OPTION 4
5. OPTION 5
Warning messages are indicated as:
Warning message
WARNING
Important notes are indicated as:
Note details
NOTE
1
EZCT-2000B USER’S MANUAL REV 1
1.0 INTRODUCTION
1.1 General Description and Features
The EZCT-2000B is Vanguard’s third-generation microprocessor-based current transformer test set. Designed specifically for CT testing, the EZCT-2000B has the following outstanding features that can greatly increase productivity and save time during the commissioning stage:
Performs CT excitation, current-ratio, polarity, and phase angle tests
Measures insulation resistance and winding resistance of the CT secondary windings
Measures the CT’s secondary burden
The EZCT-2000B’s test leads can be connected to all the CT output terminals, and the complete CT test can be performed automatically without any operator intervention.
Saturation Test
The CT saturation test is performed in compliance with the ANSI/IEEE C57.13.1 and IEC 60044 standards. Test voltage ranges from 50, 300, 500,1200 and 2000 Vac can be selected for the saturation test. The test voltage is raised and lowered automatically by the EZCT-2000B. The saturation test voltage and current data is collected and stored in the EZCT-2000B’s internal memory. Knee point voltages (ANSI 10/50, IEC 10/50, IEEE-30, IEEE-45) are calculated and printed on the test report.
All of the EZCT-2000B’s test leads can be connected to the CT output terminals (X1, X2, X3, X4 and X5), and there is no lead switching required during testing. This convenient arrangement allows for testing any of the 10 possible combinations of X1 to X5. Up to 10 saturation tests can be stored in one record. Once the test is completed, the test report and CT excitation curves can be printed on the built-in thermal printer.
CT Ratio and Polarity Test
The CT current-ratio is determined using the ANSI/IEEE C57.12.90 measurement method. A test voltage is applied on any two terminals (X1 to X5) of the CT, and the induced voltage is measured through the H1 and H2 terminals of the CT. The CT current-ratio is displayed and also stored in memory. The current-ratio is measured from 0.8 to 5,000. The CT winding polarity is displayed as a “+” sign (in-phase) or a “-” sign (out-of-phase) and is annotated with the phase angle in degrees. The CT current-ratio error and phase displacement is also calculated based on the CT burden (or rated power) and rated current.
CT Winding Resistance Test
The EZCT-2000B can measure the DC resistance of the CT winding under test. The DC winding resistance reading range is from 100 micro-ohms to 10 Ohms. Equivalent winding resistance values at different temperatures are also provided in the report.
CT Winding Insulation Resistance Test
The EZCT-2000B can also measure the insulation resistance of the CT’s secondary winding using a test voltage up to 1000 Vdc. The DC winding resistance reading range is from 2 to 500 Mega­ohms. The insulation resistance test results are displayed and printed on the report.
2
REV 1 EZCT-2000B USER’S MANUAL
CT Burden Test
The EZCT-2000B can measure the CT’s secondary burden by injecting a 1A or 5A test current into the load. The CT burden measurements (Voltage, current, Cos ϕ, and burden impedance) are displayed on the screen and printed on the test report. This important test verifies the actual CT burden before putting the CT in service, thus avoiding any potential configuration conflicts.
Current Source
The EZCT-2000B’s programmable current source (0-20A, 0-15Vac) can be used to verify CT loads. The on-time timer and output current are displayed on the LCD screen.
Test Record Header Information
Test record header information, including the company, substation name, circuit ID, manufacturer, mode, CT serial number, and the operator’s name, can be stored with each record. In addition to the test record header, a 20-character test description for each test in the record (10 tests per record) can also be entered.
User Interface and Display
The EZCT-2000B features a back-lit LCD screen (128 x 64 dot graphic) that is clearly viewable in both bright sunlight and low-light levels. An alpha-numeric keypad is used to enter test information and to control the unit’s functions.
Built-in Thermal Printer
A 4.5-inch wide built-in thermal printer can print the CT test results and saturation curves.
Internal Test Record Storage Capacity
The EZCT-2000B can store up to 140 test records in Flash EEPROM. Each test record may contain up to 10 saturation curves, current-ratio readings, polarity and DC resistance readings, insulation test results, and load burden test results. Test records can be recalled and printed by the built-in thermal printer. Test records can also be transferred to a PC using the RS-232C port, USB port, or via the USB Flash drive interface port.
Internal Test Plan Storage Capacity
The EZCT-2000B can store up to 128 CT test plans in Flash EEPROM. A test plan is comprised of the saturation test voltage, current range selection, CT nameplate ratios, and CT winding terminal combinations (X1 to X5) for each test and also includes the insulation and CT load burden test definitions. Up to 10 test definitions can be stored per test plan. The ability to store test plans makes CT testing an extremely simple process. To perform a test, the EZCT-2000B is connected to the CT terminals and a test plan is selected to run.
Creating test plans for the EZCT-2000B is also a simple process. A test plan can be created using the EZCT-2000B’s keypad or can be created on a PC (with provided software) and then downloaded to the EZCT-2000B via the RS-232C or USB ports. For added convenience, test plans can also be copied from a USB Flash drive to the EZCT-2000B via the USB Flash drive interface.
3
EZCT-2000B USER’S MANUAL REV 1
Computer Interface
The EZCT-2000B can be used as a stand-alone unit or can be computer-controlled. It can be connected to a PC via the standard RS-232C or USB port. In computer-controlled mode, test records can be downloaded from the EZCT-2000B’s memory, or CT tests can be run from the PC. A Windows® XP/Vista-based Current Transformer Analysis software application is provided with the EZCT-2000B and can be used to transfer test records from the EZCT-2000B to a PC. Test plans can also be created with the provided software. Additionally, tabulated test records can be exported to Microsoft® Excel.
1.2 Furnished Test Accessories
The EZCT-2000B comes furnished with the following:
1 Power Cord
5 20-foot X Cable Sets
1 35-foot H Cable Set
Current Source Cables
Insulation Test Cables
Ground Cables
Transportation Case
4
REV 1 EZCT-2000B USER’S MANUAL
1.3 Technical Specifications
Table 1. EZCT-2000B Technical Specifications
TYPE Portable current-transformer test set
PHYSICAL SPECIFICATIONS 19”W X 13”H x 15.8” D (48.3 cm x 33.0 cm x 40.1 cm); Weight: 73 lbs (33.1
kg)
INPUT POWER 100 – 120 Vac or 200 – 240 Vac (factory pre-set), 50/60 Hz
MEASUREMENT METHOD ANSI/IEEE C57.12.90, IEC 60044-1, and ANSI/IEEE C57.13.1 standards
TEST OUTPUT VOLTAGES 0 – 50 Vac @ 10A max; 0 – 300 Vac @ 10A max; 0 – 500 Vac @ 5A max;
0 – 1200 Vac @ 1.2A max; 0 – 2000 Vac @ 1A max
CURRENT SOURCE 1 – 20A @ 0 – 15 Vac
CURRENT SOURCE DISPLAY Test current and current on-time
VOLTAGE READING RANGE CURRENT READING RANGE CT CURRENT RATIO RANGE 0.8 – 999: 0.1%; 1000 – 1999: 0.3%; 2,000 – 5,000: 1%
PHASE ANGLE
MEASUREMENT
RESISTANCE READING
RANGE
INSULATION RESISTANCE
READING RANGE
DISPLAY Back-lit LCD Screen (128 x 64 dot graphic); viewable in bright sunlight and
PRINTER Built-in 4.5-inch wide thermal printer
COMPUTER INTERFACES One RS-232C (115,000 Baud) port, One USB port
EXTERNAL DATA STORAGE One USB Flash drive interface port
PC SOFTWARE Windows® XP/Vista-based CT Analysis software is included with the
INTERNAL TEST RECORD
STORAGE
INTERNAL TEST PLAN
STORAGE
SAFETY Designed to meet UL 6110A-1 and CAN/CSA C22.2 No. 1010.1-92
ENVIRONMENT
CABLES Five 20-foot X cable sets, One 35-foot H cable set, current source cables,
WARRANTY One year on parts and labor
0 – 2,200 Vac; Accuracy: ±1.0% of reading, ±1 V 0 – 10A; Accuracy: ±1.0% of reading, ±0.02A
0 – 360 degrees; Accuracy: ± 1.0 degree)
100 micro-ohms – 10 ohms; Accuracy: 2% of reading, ± 1 count
2 Mega-ohms – 500 Mega-ohms; Accuracy: 3% of reading, 500 Vdc or 1000 Vdc test voltage
low-light levels
purchase price
Stores 140 test records. Each test record may contain up to 10 saturation curves, current-ratio readings, polarity and DC resistance readings, insulation resistance test results, and CT load burden test results.
Stores 128 test plans. A test plan consists of the saturation test voltage, current selection, CT nameplate ratios, and CT winding terminal combinations (X1 to X5) for each test.
standards Operating: -10°C to 50°C (15°F to +122°F); Storage: -30°C to 70°C (-22°F
to +158°F)
insulation test cables, power cord, ground cable. A transportation case is included with the purchase price.
The above specifications are valid at nominal operating voltage and at a temperature of 25°C (77°F). Specifications may change without prior notice.
NOTE
5
EZCT-2000B USER’S MANUAL REV 1
1.4 EZCT-2000B Controls and Indicators
The EZCT-2000B’s controls and indicators are shown in Figure 1 below. A leader line with an index number points to each control and indicator, which is cross-referenced to a functional description in Table 2. The table describes the function of each item on the control panel. The purpose of the controls and indicators may seem obvious, but users should become familiar with them before using the EZCT-2000B. Accidental misuse of the controls will usually cause no serious harm. Users should also be familiar with the safety summary found on the front page of this User’s Manual.
Figure 1. EZCT-2000B Controls and Indicators
6
REV 1 EZCT-2000B USER’S MANUAL
Table 2. Functional Descriptions of EZCT-2000B Controls and Indicators
Item
Number
1
2-6
7
8
9
10
11
12
13
14
15
16
17
18
19
Panel Markings Functional Description
INSULATION
RESISTANCE
METER
X1, X2, X3, X4, X5
H Current transformer primary input test cable connector.
GROUND Grounding stud.
None AC receptacle.
100-120Vac, 12A
50-60Hz
None
None Rugged, alpha-numeric membrane keypad
HIGH VOLTAGE
PRESENT
EMERGENCY
TURN OFF
“PUSH”
USB MEM USB Flash drive interface.
USB PC USB PC interface.
RS-232C RS-232C PC interface. Baud rate is set for 115KB, 8 data bits, 2 stop bits.
None Built-in 4.5-inch wide thermal printer.
AC CURRENT
SOURCE
0 – 20A
Insulation resistance meter connectors.
Current transformer excitation voltage connectors. Each set of connectors contains a test voltage connector and sensing connector. The EZCT-2000B’s X output terminals are rated to 2000 Vac working voltage. Any voltage above 2000 Vac will damage the input circuitry.
Power switch with built-in circuit breaker.
Back-lit graphic LCD screen (128 x 64); viewable in bright sunlight and low light levels.
LED warning indicator that is illuminated when high voltage is present.
Emergency turn off switch.
AC current source connectors.
7
EZCT-2000B USER’S MANUAL REV 1
2.0 PRE-TEST SETUP
2.1 Operating Voltages
The EZCT-2000B’s operating voltage is preset at the factory for 100-120 Vac, 50/60 Hz or 200­240 Vac, 50/60 Hz.
2.2 LCD Screen Contrast Control
To increase the LCD screen contrast, press and hold the [PAPER Contrast] key for two seconds. Release the button when the desired contrast level has been reached.
To decrease the LCD screen contrast, press and hold the [PAPER Contrast] key for two seconds. Release the button when the desired contrast level has been reached.
2.3 Printer Paper Control
To advance the thermal printer paper, press and release the [PAPER Contrast] key.
To retract the thermal printer paper, press and release the [PAPER Contrast] key.
2.4 Printer Paper
The EZCT-2000B’s built-in thermal printer uses 4.5-inch wide thermal paper for printing test results. To maintain the highest print quality and to avoid paper jams, the use of thermal paper supplied by Vanguard Instruments Company is highly recommended. Additional paper can be ordered from the following sources:
Vanguard Instruments Co, Inc.
1520 S. Hellman Avenue Ontario, CA 91761 Tel: 909-923-9390 Fax: 909-923-9391 Part Number: VIC TP-4 paper
BG Instrument Co.
13607 E. Trent Avenue Spokane, WA 99216 Tel: 509-893-9881 Fax: 509-893-9803 Part Number: VIC TP-4 paper
8
REV 1 EZCT-2000B USER’S MANUAL
2.5 Replacing the Thermal Printer Paper
The roll of thermal paper is housed inside a dispenser underneath the printer cover. To replace the paper, follow the steps below:
Unscrew the two large printer cover screws and remove the printer cover.
Remove the leftover thermal paper roll from the paper holder.
Unroll the new thermal paper roll.
Feed the thermal paper into the slot between the paper pocket and the rubber roller.
The printer will automatically pull the paper under the thermal head.
Place the paper roll into the paper holder.
Lift the thermal head and align the thermal paper if necessary.
Re-install the printer cover.
Thermal paper has a chemical coating on one side of the paper. This side should be facing the thermal print head. Incorrect paper loading may result in blank output on
NOTE
the thermal paper.
The thermal paper will show a red stripe to indicate that the roll is about to run out of paper.
2.6 Computer Interface Ports
The EZCT-2000B features one USB and one RS-232C PC interface port. A Windows-based “Current Transformer Analysis” software application is supplied with the EZCT-2000B. For further information, please see the software user’s manual.
9
EZCT-2000B USER’S MANUAL REV 1
3.0 OPERATING PROCEDURES
3.1 EZCT-2000B Cable Connections
Always connect the EZCT-2000B to the substation ground before connecting any test cables. The EZCT-2000B is supplied with five 20-foot X test cables and one 35-foot H cable. The X cable connections are required to run the current transformer excitation test. The H and X cable connections are required to run the transformer turns-ratio test. A typical excitation and ratio test connection is shown in Figure 2. The insulation resistance test connection is shown in Figure 3. The burden test connection is shown in Figure 4. The current source test connection is shown in Figure 5. Transformer bushing CT connections for Delta and Y transformers are shown in Figure 6 and Figure 7, respectively.
Figure 2. Typical EZCT-2000B Excitation and Ratio Test Cable Connection
10
REV 1 EZCT-2000B USER’S MANUAL
Figure 3. EZCT-2000B Insulation Resistance Test Cable Connection
11
EZCT-2000B USER’S MANUAL REV 1
Figure 4. EZCT-2000B CT Load Burden Test Cable Connection
12
REV 1 EZCT-2000B USER’S MANUAL
Figure 5. EZCT-2000B Typical Current Source Connection
13
EZCT-2000B USER’S MANUAL REV 1
Figure 6. Bushing CT Connection on Delta Transformer
Figure 7. Bushing CT Connection on Y Transformer
14
REV 1 EZCT-2000B USER’S MANUAL
3.2 EZCT-2000B X Input Voltage Warning
The EZCT-2000B X output terminals are rated to 2,000 Vac working voltage. Any voltage present at these terminals above 2,000 Vac may damage the X sense circuitry, cause false readings, or both. An example of a typical situation where this may occur is shown in Figure 8 below.
Figure 8. Sample CT Name Plate
For the above example CT, the turns ratio between X1-X4 is 3150 to 1. The turns ratio between X1-X2 is 400 to 1. The turns ratio between X1-X4 and X1-X2 is 7.88 (3150/400). If a test voltage of 300 Vac is applied to the X1-X2 terminals, a voltage of 2,364 Vac (300 Vac x 7.88) will be induced at the X1-X4 terminals. If all the test leads are connected to the EZCT-2000B and the excitation test is performed on the X1-X2 terminals, the voltage induced at the X1-X4 terminals will exceed 2,000 Vac as the voltage across the X1-X2 terminals increases above 250 Vac. A “Flash-Over” condition may occur and damage the EZCT-2000B. In this case, the user should only connect the X1-X2 leads and run its excitation test, then connect the X4 lead before running the X1-X4 excitation test.
15
EZCT-2000B USER’S MANUAL REV 1
3.3 Performing Tests
3.3.1. Entering Test Record Header Information
You can enter the test record header information before performing tests. The record header includes identifying information such as the company, station, circuit, model number, etc. Once the header information has been entered, it will apply to all subsequent test records. To enter the header information:
a. When the unit is turned on and the firmware has been loaded, you will be presented
with the “START-UP” menu as shown below:
1. RUN TEST 06/17/09
2. SETUP 10:14:25
3. TEST PLANS
4. DIAGNOSTICS
5. CURRENT SOURCE
Press the [2] key (SETUP).
b. The following screen will be displayed:
1. RECORD ID
2. PRINT RECORD
3. RECORD DIRECTORY
4. SAVE/RESTORE RECORD
5. ERASE RECORD
6. NEXT PAGE
Press the [1] key (RECORD ID)
c. The following screen will be displayed:
COMPANY:
vanguard
↑↓ TO POSITION
"enter" to accept
Type the company name using the alpha-numeric keypad.
When pressing a key, the corresponding number on the key will be displayed first. Pressing the key again will display the first letter on the key. Pressing the key again will display the second letter on the key. For example, to type the letter “A”, you must press
[2] key twice. To erase the character at the cursor position, press the [CLEAR] key.
the Press the
Contrast]
[PAPER Contrast] key to move to the next character. Press the [PAPER
key to move to the previous character. Press the [ENTER] key when you
are done typing the company name.
16
REV 1 EZCT-2000B USER’S MANUAL
d. The following screen will be displayed:
STATION:
LAB
↑↓ TO POSITION
"enter" to accept
Type the station name using the alpha-numeric keypad and then press the [ENTER] key.
e. The following screen will be displayed:
CIRCUIT:
CIRCUIT 1
↑↓ TO POSITION
"enter" to accept
Type the circuit information using the alpha-numeric keypad and then press the
[ENTER] key.
f. The following screen will be displayed:
MANUFACTURER:
ABB
↑↓ TO POSITION
"enter" to accept
Type the manufacturer name using the alpha-numeric keypad and then press the
[ENTER] key.
g. The following screen will be displayed:
MODEL:
EZCT2000B
↑↓ TO POSITION
"enter" to accept
Type the model information using the alpha-numeric keypad and then press the
[ENTER] key.
17
h. The following screen will be displayed:
SERIAL NUMBER:
26002
↑↓ TO POSITION
"enter" to accept
Type the serial number using the alpha-numeric keypad and then press the [ENTER] key.
i. The following screen will be displayed:
COMMENTS:
no comment
↑↓ TO POSITION
"enter" to accept
EZCT-2000B USER’S MANUAL REV 1
Enter any relevant comments using the alpha-numeric keypad and then press the
[ENTER] key.
j. The following screen will be displayed:
OPERATOR:
vn
↑↓ TO POSITION
"enter" to accept
Type the operator’s name using the alpha-numeric keypad and then press the [ENTER] key. All header information will be saved, and you will be returned to the “START-UP” menu.
18
REV 1 EZCT-2000B USER’S MANUAL
3.3.2. Performing Resistance, Excitation, and Ratio Tests
The following procedure describes the general steps for performing excitation, resistance, and ratio tests.
a. When the EZCT-2000B is turned on, it will first go through a start-up cycle and load the
firmware. Then the “START-UP” menu will be displayed as shown below:
1. RUN TEST 06/17/09
2. SETUP 11:18:25
3. TEST PLANS
4. DIAGNOSTICS
5. CURRENT SOURCE
Press the [1] key (RUN TEST) to start a test.
b. The following screen will be displayed:
1. Res, excit & ratio
2. excitation & ratio
3. excitation only
4. ratio only
5. resistance & excit
6. burden test only
7. insulation tst only
Select the test type by pressing the corresponding key ([1] - [5])
c. The following screen will be displayed:
SELECT TAP:
1. X1-X2
2. X1-X3
3. X1-X4
4. X1-X5
5. NEXT PAGE
Select the tap connection by pressing the corresponding key (
[1] - [4]). If the tap
connection is not listed, press the [5] key to view the next page of options.
d. If the selected test included an excitation test, the following screen will be displayed:
SELECT VOLTAGE RANGE:
1. 50V
2. 300V
3. 500V
4. 1200V
5. 2000V
Select a test voltage range by pressing the corresponding key (
19
[1] - [5]).
EZCT-2000B USER’S MANUAL REV 1
e. If the selected test included an excitation test, the following screen will be displayed:
SET TEST CURRENT:
1. 0.2A
2. 0.5A
3. 1A
4. 2A
5. 5A
6. 10A
Select the maximum test current for the excitation test by pressing the corresponding key (
[1] - [6]
).
f. If the selected test included a ratio test, the following screen will be displayed:
XFMR NAME PLATE RAT.
1. YES
2. NO
1. YES
Press the
[1]
key if you would like to enter the CT nameplate values. The
following screen will be displayed:
ENTER PLATE RATIO:
0 :
Type the first number using the keypad.
You can press the
NOTE
[CLEAR]
key to restart a field entry if necessary.
20
REV 1 EZCT-2000B USER’S MANUAL
Press the [ENTER] key. The following screen will be displayed:
ENTER PLATE RATIO:
1000 : 0.0
Type the second number using the keypad and then press the Continue to step g.
2. NO
Press the
[2] key if you do not want to enter the CT nameplate values. Continue
to step g.
g. The following screen will be displayed:
ENTER TEST 1 NOTE:
_
↑↓ to position
"ENTER" TO ACCEPT
ENTER PLATE RATIO:
1000 : 5.0
[ENTER] key.
Use the alpha-numeric keys on the keypad to enter a test note. The test note field is 20 characters long. One test note can be saved for each test.
When pressing a key, the corresponding number on the key will be displayed first. Pressing the key again will display the first letter on the key. Pressing the key again will display the second letter on the key. For example, to type the letter “A”, you must press
[2] key twice. To erase the character at the cursor position, press the [CLEAR] key.
the Press the
[PAPER Contrast]
[PAPER Contrast] button to move to the next character. Press the
key to move to the previous character. Press the [ENTER] key
when you are done typing the note.
h. The following screen will be displayed:
calc err vs burden?
1. yes
2. no
Press the
[1] key (YES). Selecting this option will print the current ratio error table and
current phase error table as part of the tabulated test results. Please see Figure 9, items 19 and 20.
21
i. The following screen will be displayed:
ENTER BURDEN VA:
(500.0 max)
0.0
EZCT-2000B USER’S MANUAL REV 1
Type the burden value using the keypad and then press the
j. The following screen will be displayed:
ENTER cos ϕ:
(0.00 - 1.00)
0.0
Type the Cos ϕ value using the keypad and then press the
k. The following screen will be displayed:
rated secondary cur:
1. 5A 2)1A
[ENTER] key.
[ENTER] key.
Select the rated secondary current by pressing the
[1] key (5A) or [2] key (1A).
l. The following screen will be displayed showing a summary of the test parameters:
TEST 1 parameters: 300V 1.0A x1-x2
"START" to begin
Press the [START] key to start the test.
22
REV 1 EZCT-2000B USER’S MANUAL
m. If the selected test included a resistance test, the following screen will be displayed
momentarily:
CABLES ENERGIZED! 300V 1.0A x1-x2 DC RESISTANCE TEST
The following screen will then be displayed:
CABLES ENERGIZED! 300V 1.0A x1-x2
RESISTANCE = 215.1
Ω
The “HIGH VOLTAGE PRESENT” light will be illuminated to indicate that high voltage is present.
NOTE
n. If the selected test included a ratio test, the following screen will be displayed
momentarily:
CABLES ENERGIZED! 300V 1.0A x1-x2 I= 0.0004 AMPS V= 34.2 VAC
The following screen will then be displayed:
RATIO TEST
V
= 99.5 VH = 0.497
x
I=0.0402 RAT=+199.88
23
When the testing is finished, the test results graph will be displayed:
Press any key on the keypad to continue.
o. The following screen will be displayed:
print test results?
1. YES
2. NO
EZCT-2000B USER’S MANUAL REV 1
Press the [1] key (YES) if you would like to print the test results. The test results will be printed on the thermal printer. A typical EZCT-2000B tabulated test report printout is shown in Figure 9. A typical graphic report is shown in Figure 10.
Press the [2] key (NO) if you do not want to print the test results.
p. The following screen will be displayed:
KEEP THIS TEST?
1. YES
2. NO
Press the
[1] key (YES) to keep the test results.
q. The following screen will be displayed:
TEST 1 SAVED
Press any key to continue.
24
REV 1 EZCT-2000B USER’S MANUAL
r. The following screen will be displayed:
RUN ANOTHER TEST?
1. YES
2. NO
Press the
[2] key (NO).
s. The following screen will be displayed:
RUN INSUL RES TEST?
1. YES
2. NO
Press the
[2] key (NO).
If you would like to run the insulation resistance test as well, press the [1] key and then follow the instructions in section 3.3.4, steps c to f.
t. The following screen will be displayed:
RUN BURDEN TEST? (REQUIRES LEAD CHNG)
1. YES
2. NO
Press the
[2] key (NO).
If you would like to run the burden test as well, press the [1] key and then follow the instructions in section 3.3.3, steps c to e.
u. The following screen will be displayed:
SAVE THIS RECORD?
1. YES
2. NO
Press the [1] key (YES) to save the record.
25
The following screen will be displayed momentarily:
SAVING RECORD...
PLEASE WAIT...
v. The following confirmation screen will then be displayed:
RECORD NUMBER 1 has been saved!
The test record number is automatically assigned to each test record stored in the EZCT-2000B’s Flash EEPROM.
NOTE
EZCT-2000B USER’S MANUAL REV 1
Press any key to return to the “START-UP” menu.
26
Loading...
+ 88 hidden pages