Vanguard DigiTMR User Manual

TM
DIGITMR
DIGITAL CIRCUIT BREAKER ANALYZER
USER’S MANUAL
1520 S. Hellman Ave. Ontario, California 91761, USA
TEL: (909) 923-9390 FAX: (909) 923-9391
May 2009
Revision 3
DIGITMR USER’S MANUAL REV 3
SAFETY SUMMARY
FOLLOW EXACT OPERATING PROCEDURES
Any deviation from the procedures described in this User’s Manual may create one or more safety hazards, may damage the DigiTMR, damage the test transformer, or cause errors in the test results. Vanguard Instruments Company, Inc. assumes no liability for unsafe or improper use of the DigiTMR.
All safety precautions provided in this manual must be observed during all phases of testing including test preparation, test lead connection, actual testing, and test lead disconnection.
SAFETY WARNING AND CAUTIONS
The DigiTMR shall be used only by trained operators. When using the DigiTMR in Time-versus­Travel mode, all circuit breakers under test shall be off-line and fully isolated.
DO NOT MODIFY TEST EQUIPMENT
To avoid the risk of introducing additional or unknown hazards, do not install substitute parts or perform any unauthorized modification to any DigiTMR test unit. To ensure that all designed safety features are maintained, it is highly recommended that repairs be performed only by Vanguard Instruments Company factory personnel or by an authorized repair service provider. Unauthorized modifications can cause safety hazards and will void the manufacturer’s warranty.
WARNING
Do not remove test leads during a test. Failure to heed this warning can result in electrical shock to personnel and damage to the equipment.
i
REV 3 DIGITMR USER’S MANUAL
TABLE OF CONTENTS
CONVENTIONS USED IN THIS DOCUMENT ..................................................................................... 1
1.0 INTRODUCTION .................................................................................................................. .. 2
1.1 General Description and Features ................................................................................... 2
1.2 Operational Modes .......................................................................................................... 6
1.2.1. Quick-Shot Mode ...................................................................................................... 6
1.2.2. Time-versus-Travel Circuit Breaker Analyzer Mode ................................................. 6
1.3 Supplied Cables ................................................................................................................ 6
1.4 Technical Specifications ................................................................................................... 7
1.5 DigiTMR Controls and Indicators ..................................................................................... 8
2.0 PRE-TEST SETUP ................................................................................................................. 10
2.1 Operating Voltages ........................................................................................................ 10
2.2 Main Power Fuse ........................................................................................................... 10
2.3 LCD Screen Contrast Control .......................................................................................... 10
2.4 Printer Paper Control ..................................................................................................... 11
2.5 Printer Paper .................................................................................................................. 11
3.0 OPERATING PROCEDURES ................................................................................................. 12
3.1 DigiTMR Cable Connections ........................................................................................... 12
3.1.1. Contact Cable Connections .................................................................................... 12
3.1.2. Initiate Cable Connections ..................................................................................... 13
3.1.3. Analog and Digital Voltage Monitor Connections .................................................. 15
3.1.5. AC Current Sensor Connection ............................................................................... 17
3.1.6. Transducer Connection .......................................................................................... 18
3.2 General Procedures ....................................................................................................... 19
3.3 Timing Breakers in Quick-Shot Mode ............................................................................ 19
3.4 Timing Breakers in Time-Travel Mode ........................................................................... 20
3.5 Performing Timing Tests ................................................................................................ 21
3.5.1. Timing an OPEN Operation With No Insertion Resistors ....................................... 21
3.5.2. Timing an OPEN Operation WITH Insertion Resistors ............................................ 24
3.5.3. Timing a CLOSE-OPEN Operation Using Contact Channel #1 ................................. 26
3.5.4. Timing an OPEN-CLOSE-OPEN Operation ............................................................... 28
3.6 Performing Diagnostic Tests .......................................................................................... 30
3.6.1. Performing a Slow-Close Test ................................................................................. 30
3.6.2. Performing a Transducer Self-Test ......................................................................... 32
3.6.3. Checking Cable Hookups ........................................................................................ 33
3.7 Printing Test Results ...................................................................................................... 34
3.7.1. Printing the Tabulated Test Results ....................................................................... 34
3.7.2. Printing the Graphic Test Results ........................................................................... 37
3.8 Working with Timing Shots ............................................................................................ 43
3.8.1. Saving Timing Shots in the DigiTMR’s EEPROM ..................................................... 43
3.8.2. Printing a Timing Shot Directory of Shots Stored in the EEPROM ......................... 45
3.8.3. Recalling a Timing Shot From the DigiTMR’s EEPROM .......................................... 47
ii
DIGITMR USER’S MANUAL REV 3
3.8.4. Erasing a Timing Shot From the DigiTMR’s EEPROM ............................................. 49
3.8.5. Erasing All Timing Shots from the DigiTMR’s EEPROM .......................................... 51
3.8.6. Entering Shot Description Information .................................................................. 53
3.9 Working with Breaker Test Plans ................................................................................... 56
3.9.1. Recalling a Breaker Test Plan ................................................................................. 56
3.9.2. Printing a Breaker Test Plan ................................................................................... 58
3.9.3. Printing the Breaker Test Plan Directory ................................................................ 60
3.10 Changing Setup Parameters ........................................................................................... 61
3.10.1. Setting Up Analysis Points ...................................................................................... 61
3.10.2. Setting the Measurement Units ............................................................................. 64
3.10.3. Setting the DigiTMR’s Internal Clock ...................................................................... 66
3.10.4. Setting the Preferred Printing Method .................................................................. 67
4.0 GLOSSARY OF TERMS ......................................................................................................... 68
5.0 DIGITMR TROUBLE-SHOOTING GUIDE ............................................................................... 69
LIST OF TABLES
Table 1. DigiTMR Technical Specifications ...................................................................................... 7
Table 2. Functional Descriptions of DigiTMR Controls and Indicators ........................................... 9
Table 3. Voltage Selection Jumper Settings .................................................................................. 10
Table 4. Descriptions of Tabulated Test Results Elements ........................................................... 36
iii
REV 3 DIGITMR USER’S MANUAL
LIST OF FIGURES
Figure 1. DigiTMR Controls and Indicators ..................................................................................... 8
Figure 2. Voltage Selection Jumper Settings and Initiate Relay ................................................... 10
Figure 3. Contact Cable Connections ............................................................................................ 12
Figure 4. DC Trip and DC Close Initiate Circuit Cable Connections ............................................... 13
Figure 5. DC Trip and AC Close Initiate Circuit Cable Connections ............................................... 14
Figure 6. Analog and Digital Voltage Monitoring Cable Connections .......................................... 15
Figure 7. External Trigger Cable Connections ............................................................................... 16
Figure 8. Typical CT Connection .................................................................................................... 17
Figure 9. Typical Transducer Connection ...................................................................................... 18
Figure 10. DigiTMR General Operating Procedures ...................................................................... 19
Figure 11. Typical "Slow-Close" Test Results Printout .................................................................. 31
Figure 12. Typical Tabulated Test Results Printout of an OPEN Operation .................................. 35
Figure 13. Typical Graphic Test Results Printout .......................................................................... 39
Figure 14. Expansion Graph from 0 to 200 ms ............................................................................. 40
Figure 15. Graphic Interpretation of an OPEN Timing Shot ......................................................... 41
Figure 16. Graphic Interpretation of a CLOSE Timing Shot ........................................................... 41
Figure 17. Graphic Interpretations of OPEN-CLOSE and CLOSE-OPEN Timing Shots ................... 42
Figure 18. Typical Shot Directory Printout .................................................................................... 46
Figure 19. Typical Test Plan Printout ............................................................................................ 59
Figure 20. Typical Test Plan Directory Printout ............................................................................ 60
Figure 21. Transducers Compatible with the DigiTMR ................................................................. 76
Figure 22. Linear Transducer on an OCB ...................................................................................... 76
Figure 23. Rotary Transducer on Alstom HE 1014 Circuit Breaker ............................................... 77
Figure 24. Rotary Transducer on ABB HMB Mech ........................................................................ 77
Figure 25. Westinghouse SFA Transducer Installation ................................................................. 78
Figure 26. ABB 72PM Circuit Breaker with Linear Transducer and Special Timing Rod ............... 78
Figure 27. Rotary Transducer on Megna Blast Breaker ................................................................ 79
Figure 28. Rotary Transducer on Federal Pacific Breaker ............................................................. 79
Figure 29. Linear Transducer on Mitsubishi 500-SFMT ................................................................ 80
Figure 30. Linear Transducer on Areva DT1-72.5F ....................................................................... 80
APPENDIX
Appendix A. ITE Circuit Breaker Model 14.4K Timing Chart ......................................................... 71
Appendix B. DigiTMR Test Plan for ITE Circuit Breaker Model 14.4K ........................................... 72
Appendix C. DigiTMR Timing Chart for ITE Circuit Breaker Model 14.4K ..................................... 73
Appendix D. DigiTMR Timing Chart for ITE Circuit Breaker Model 14.4K ..................................... 74
Appendix E. Timing Chart and Plan for Siemens Type SPS2, 121 Circuit Breaker ........................ 75
Appendix F. DigiTMR Transducer Illustrations .............................................................................. 76
iv
DIGITMR USER’S MANUAL REV 3
CONVENTIONS USED IN THIS DOCUMENT
This document uses the following conventions:
• A key or switch on the DigiTMR is indicated as [KEY].
Menu options are referenced as (MENU OPTION).
Screen and menu names are referenced as “SCREEN/MENU NAME”.
The terms “test record” and “test shot” are used interchangeably.
DigiTMR LCD screen output is shown as:
1. OPTION 1
2. OPTION 2
3. OPTION 3
4. OPTION 4
Warning messages are indicated as:
Warning message
WARNING
Important notes are indicated as:
Note details
NOTE
1
REV 3 DIGITMR USER’S MANUAL
1.0 INTRODUCTION
1.1 General Description and Features
The DigiTMR is a microprocessor-based, time-versus-travel, circuit breaker analyzer. In Quick­Shot Mode (optional), the DigiTMR has the capability of online testing and can capture the breaker’s trip or close time, the trip/close-coil current “fingerprint,” and the battery supply voltage while the breaker is still in service. In Time-Travel mode, the DigiTMR can fully analyze a circuit-breaker’s performance by testing the contact time, stroke, velocity, over-travel, and contact wipe. Contact-motion analysis can be performed for all breaker contact operations (Open, Close, Open – Close, Close – Open, and Open – Close – Open). The DigiTMR’s timing window is selectable between 1-second, 10-second, or 20-second periods. The 10-second and 20-second timing windows are ideal for timing long duration events such as circuit-switcher contact testing.
The DigiTMR can also perform a Slow-Close test. A Slow-Close test determines the contact touch distance and penetration by using the travel transducer and dry contact channels.
The DigiTMR can be computer-controlled via its RS-232C interface. Computer control is fully covered in the Vanguard Circuit Breaker Analyzer (VCBA) software manual. This manual focuses on operating the DigiTMR in stand-alone mode.
The DigiTMR features:
Three contact-input channels
One analog voltage monitoring input channel
One digital voltage monitoring input channel
One external trigger input channel
One digital travel transducer input channel
One Trip/Close current monitoring channel
One CT input channel (optional)
Contact Timing Channels
A breaker’s interruption of contacts is timed with the dry-contact input channels. Each channel can detect the main contact’s and the insertion resistor contact’s times in milliseconds and cycles.
In Time Travel Mode, each DigiTMR contact channel measures contact close time, contact open time, insertion resistor contact open and close times,
contact bounce time, resistor on time, and contact spread time.
The DigiTMR can detect insertion resistances ranging from 10 to 7,000 Ohms. Resistance over 7,000 Ohms is detected as an open contact. In Quick-Shot mode,
NOTE
the DigiTMR can provide the contact Close time and contact Cleared time using the circuit breaker CT.
2
DIGITMR USER’S MANUAL REV 3
Analog Voltage Monitoring Channel
One analog voltage monitoring channel, designated as “V1”, is dedicated to monitoring the breaker’s DC power supply, or coil voltages from 0 to 255 Vdc or peak Vac. Test results can be printed and include nominal voltage (when the supply voltage has no load), and minimum voltage (when the supply voltage is under load or when a coil is energized).
The “V1” connection can also monitor the DC power supply’s voltage fluctuations during a breaker operation. This feature is effective in detecting breaker power supply problems that normally cannot be detected using a voltmeter.
Digital Voltage Monitoring Channel
One digital voltage monitoring channel, designated as “V2”, monitors the status voltage as “ON” or “OFF”. Voltages from 30 to 255 Vdc are considered “ON”, whereas voltages less than 30 Vdc are considered “OFF”. This input channel is dedicated to monitoring the breaker’s auxiliary switch action during a breaker operation. Input voltage range is from 1 to 300 Vac/dc.
External Trigger Input Channel
The external trigger input can be used to trigger timing functions when the DigiTMR senses a voltage. Voltage levels can range from 30 to 300 Vdc or peak Vac. Different trigger voltages can be set at the factory per specific requests from users.
Travel Transducer Input Channel
The DigiTMR features one digital transducer input channel. It can interface with linear and rotary transducers as well as with special transducers developed by Vanguard Instruments Company, Incorporated. Transducers offered by Vanguard include 10-inch, 25-inch, and 30-inch linear transducers, a 15-inch universal transducer, a rotary transducer, an ABB AHMA-8 type transducer, an ABB AHMB type transducer, and a Westinghouse SF/SFA type transducer.
Please see Appendix F for transducer illustrations.
In Time Travel Mode, the digital transducer measures the breaker stroke, over-travel, bounce back, contact wipe, and contact velocity. Breaker contact velocity calculations are based upon a contact’s travel distance over a period of time, which is defined by the manufacturer’s specifications. The DigiTMR not only calculates simple distance-over-time velocities, but also supports other formulas. An example is a Siemens SPS2 breaker velocity calculation formula shown below:
V = 12.2 x a/b
The formula and analysis points can be stored in a test plan for later use in the field. Please see section 3.9 for further information about test plans.
3
REV 3 DIGITMR USER’S MANUAL
If a breaker test plan is used during testing, test results are compared with the specifications in the test plan, and a “Pass” or “Fail” indicator is printed on the test
NOTE
report based on the results of this comparison.
Breaker Initiate Capability
A built-in, solid-state, initiate device is used to operate a breaker from the DigiTMR. Operational modes include Open, Close, Open-Close, Close-Open, and
Open-Close-Open. The multiple operations of Open-Close, Close-Open, and Open-Close-Open can be initiated with a programmable delay time or by sensing the breaker’s contact state.
The solid-state switching implemented in the initiate circuitry allows the DigiTMR to switch a breaker’s AC or DC control circuit (30 A max, 250 Vac/dc). Each trip and close circuit is protected by a 5-Ampere slow-blow fuse. An interlocking “ARM” switch also protects the initiate circuit.
The DigiTMR not only plots the trip or close current waveforms, but also prints the steady-state current readings of the trip and close coils during an operation. The DigiTMR features one, non­contact, Hall-effect sensor with a range of 0-20 Amperes dc to 5Khz.
If the DigiTMR initiates the OPEN test or the CLOSE test in Quick-Shot Mode,
NOTES
the test report will show the OPEN or CLOSE coil current readings and current waveform.
If the user operates the breaker in Quick-Shot mode using the external trigger,
no coil current reading or waveform will be recorded.
Trip/Close Current Monitoring
A built-in, hall-effect, current sensor records the trip and close coil current levels and duration. The coil current waveform is graphically plotted on the DigiTMR’s built-in thermal printer. The current reading is also printed on the tabulated test results report.
Built-in Thermal Printer
The DigiTMR features a built-in 4.5-inch wide thermal printer that can be used to print the breaker’s contact analysis results in both tabular and graphics formats. Please see section 2.5 for thermal paper ordering information.
4
DIGITMR USER’S MANUAL REV 3
Computer Interface
The DigiTMR can be computer-controlled via its RS-232C interface. A Windows® XP/Vista-based Circuit Breaker Analysis software application (VCBA) is provided with each unit. Using this software, circuit breakers can be timed from the PC. Test records can be retrieved from the DigiTMR and then stored on the PC for future analysis and
report generation. Circuit breaker test plans can also be created on the PC and transferred to the DigiTMR. Additionally, test records can be exported in Microsoft® Excel format for further analysis.
User Interface
The DigiTMR features a back-lit LCD screen (20 characters by 4 lines) that is viewable in both bright sunlight and low-light levels. A rugged, 16-key, membrane keypad is used to control the unit.
Internal Test Record Storage
The DigiTMR can store up to 100 test records in Flash EEPROM. Test records can be retrieved and printed on the built-in thermal printer, or they can be transferred to a PC via the unit’s RS­232C interface.
Internal Breaker Test Plan Storage
The DigiTMR can store up to 99 circuit-breaker test plans. Test plans are comprised of all circuit­breaker performance specifications (stroke, velocity, and contact time). A test plan can be used to immediately test a circuit-breaker. A pass/fail report is then generated by comparing actual performance with the specifications in the stored test plan. Test plans can also be generated on a PC and transferred to the DigiTMR via the unit’s RS-232C interface.
Diagnostic Capabilities
The DigiTMR is designed with self diagnostic capabilities to check the integrity of the electronics. Self-testing of the contact cable connections and transducers allows the user to examine the DigiTMR’s components whenever verification is required.
5
REV 3 DIGITMR USER’S MANUAL
1.2 Operational Modes
1.2.1. Quick-Shot Mode
In Quick-Shot (online) mode, the trip/close time is derived from the time of trip, or close-coil initiation, to the breaker’s bushing current-break-or-make as detected by an AC clamp-on current sensing probe. The DIGITMR can time a circuit breaker’s OPEN operation by sensing the time when the OPEN coil is energized until the bushing CT current returns to zero. The breaker’s CLOSE time is detected when the close coil is energized until the bushing CT current is detected. The Quick-Shot mode provides a quick way to diagnose a breaker’s performance and can reduce maintenance costs.
1.2.2. Time-versus-Travel Circuit Breaker Analyzer Mode
In Time-versus-Travel mode, the DigiTMR applies a test voltage of 35 Vdc to each of the contact channels, thus allowing an analog to digital converter (A/D) to determine a close, open, or an insertion resistor as the state of the contact. The DigiTMR records into memory 10,000 readings from the A/D’s and the transducer position counters. The contact time, circuit breaker stroke and velocity are then derived from the stored data.
The DigiTMR uses an up-down counter to sense the breaker’s transducer reading. The DigiTMR’s travel transducer employs optical encoders to send quadrature signals to the DigiTMR’s counters. With the use of digital transducers and counters, the need to setup or calibrate the transducers is eliminated.
1.3 Supplied Cables
The DigiTMR comes furnished with the following cables:
1 Ground Cable
1 Power Cord
1 Contact Lead
1 Contact Extension Cable
1 Transducer Cable
3 Voltage and Trigger Leads
3 Voltage and Trigger Extension Cables
1 AC Current Transducer and Cable (Quick-Shot option only)
1 Initiate Cable
1 RS-232C cable
6
DIGITMR USER’S MANUAL REV 3
1.4 Technical Specifications
Table 1. DigiTMR Technical Specifications
TYPE Circuit-breaker analyzer
PHYSICAL SPECIFICATIONS 18”W x 7”H x 15”D (45.7 cm x 17.8 cm x 38.1 cm); Weight: less than 19 lbs
(8.6 kg)
INPUT POWER 3 Amps, 100 – 120 Vac or 200 – 240 Vac (factory pre-set), 50/60 Hz
DRY-CONTACT INPUTS 3 dry-input channels; each channel detects main and insertion-resist or
contacts
TIMING WINDOWS 1-second, 10-seconds, or 20-seconds
TIMING RESOLUTIONS ±100 micro-seconds @ 1-second duration
±1.00 milli-seconds @ 10-second duration ±2.00 milli-seconds @ 20-second duration
TIMING ACCURACY 0.05% of reading ±0.1 ms @ 1-second duration
DRY-CONTACT CHANNEL
PROTECTION
CONTACT DETECTION RANGE Closed: less than 20 ohms; Open: greater than 5,000 ohms
RESISTOR DETECTION RANGE 50 – 5,000 ohms
TRIGGER INPUT VOLTAGE Open/Close: 30 – 300V, DC or peak AC VOLTAGE SENSING INPUT
RANGE
BREAKER OPERATIONS Initiate Open, Close, Open – Close, Close – Open, Open – Close – Open
BREAKER INITIATE CAPACITY 30A, 250 Vac/dc max
INITIATE CURRENT READING
RANGE
TRAVEL TRANSDUCER INPUT 1 digital travel transducer channel; Linear range: 0.0 – 60.0 in (±0.005 in.);
CONTACT TRAVEL POINT
DIFFERENCE
DISPLAY Back-lit LCD Screen (20 characters by 4 lines); viewable in bright sunlight
PRINTER Built-in 4.5-inch wide thermal printer can print both graphic contact travel
INTERNAL TEST RECORD
STORAGE
COMPUTER INTERFACE RS-232C port (19,200 baud)
PC SOFTWARE Windows® XP/Vista-based Circuit Breaker Analysis software is included
SAFETY Designed to meet UL 61010A-1 and CAN/CSA C22.2 No. 1010.1-92
ENVIRONMENT Operating: -10°C to 50° C (15°F to +122° F); Storage: -30° C to 70° C (-
OPTIONS Transportation case (available for the DigiTMR and the travel transd ucers)
WARRANTY One year on parts and labor
Fuses protect all isolated power supplies; all contact inputs are grounded until test; input channels are protected against static discharge
V1: analog input; 0 – 255V, DC or peak AC; Sensitivity: ±1V V2: voltage presence/absence detector input; 30 – 300V, DC or peak AC
One, non-contact, Hall-effect sensor, 0 – 20 amp range, dc to 5Khz
Rotary range: 0 – 360 degrees (±0.006 degrees) Measures “slow-close” contact-point distances; results can be printed
and low-light levels
waveforms and tabulated test results
Stores up to 100 test records and 99 breaker test plans
with purchase price
standards
22°F to +158° F)
The above specifications are valid at nominal operating voltage and at a temperature of 25°C (77°F). Specifications may change without prior notice.
NOTE
7
REV 3 DIGITMR USER’S MANUAL
1.5 DigiTMR Controls and Indicators
The DigiTMR’s controls and indicators are shown in Figure 1 below. A leader line with an index number points to each control and indicator, which is cross-referenced to a functional description in Table 2. The table describes the function of each item on the control panel. The purpose of the controls and indicators may seem obvious, but users should become familiar with them before using the DigiTMR. Accidental misuse of the controls will usually cause no serious harm. Users should also be familiar with the safety summary found on the front page of this User’s Manual.
8
Figure 1. DigiTMR Controls and Indicators
DIGITMR USER’S MANUAL REV 3
Table 2. Functional Descriptions of DigiTMR Controls and Indicators
Item
Number
1 2
3
4 5 6
7
8 9
10
11
12
13
14
Panel Markings Functional Description
RS-232C 9-pin serial computer interface
CONTACT INPUT 16-pin female connector for the 3 contact channels.
300V MAX V1 and V2
300V MAX
TRIGGER
CT AC current sensor input channel.
TRAVEL
TRANSDUCER
100-120Vac, 2A,
50-60Hz
Fuse: 250Vac, 3A
Fast-Blow GROUND Safety ground terminal. 5/16-18 th read stud with wing nut.
INITIATE
CLOSE, OPEN
250V, 5A
SLOW-BLOW
“PUSH” TO
ARM
None
None
None
3-pin connectors. Each input is a voltage detector fo r the ti ming vol tages of the trip or close coil switching during the active period. Voltage detection levels range from 30 to 300 Vdc or peak.
3-pin connector triggers timing functions. Voltage levels rangi ng from 30 to 300 Vac, dc or peak.
16-pin male connector. Travel en coder input used for contact motion data.
3-wire power plug. 3 Ampere , AC fuses. The ON/OFF switch is a 2-pole rocker.
4-pin connector used for the switching circuit for op era ting the circuit breaker under test.
Close circuit fuse: 5 Ampere, 250V, Slow-Blow . Open circuit fuse: 5 Ampere, 250V, Slow-Blow.
Spring-loaded, pushbutton switch . Press and hold to compl e te the Tri p or Close circuits for breaker tests.
Thermal printer. Microprocessor controlled printer that uses specially treated,
4.5-inch wide thermal paper. Membrane keypad, 10 alpha-numeric keys and 6 function keys (START,
STOP, CLEAR, ENTER, and CONTRAST/PAPER positioning UP and DOWN)
Back-lit LCD Screen (20 characters by 4 lines); viewable in bright sunlight and low-light levels
9
REV 3 DIGITMR USER’S MANUAL
2.0 PRE-TEST SETUP
2.1 Operating Voltages
The DigiTMR’s operating voltage is preset at the factory and is selectable between 100-120 Vac, 50/60 Hz or 200-240 Vac, 50/60 Hz. To set the proper operating voltage, the voltage selection jumper must be configured properly as shown in Table 3. The initiate relay with the correct operating voltage is also required. The jumper connector is shown in Figure 2 and is marked as “JP3/VOLT SEL” on the circuit board.
Table 3. Voltage Selection Jumper Settings
Voltage Selection PINS Relay Part Number
100 – 120 Vac 200 – 240 Vac
1 & 2, 3 &4 HE2aN-Q-AC120V
2 & 3 HE2aN-Q-AC-240V
Figure 2. Voltage Selection Jumper Settings and Initiate Relay
2.2 Main Power Fuse
The DigiTMR uses an AC input module containing the AC receptacle, power switch, and protective fuses. The fuses are 3-Ampere, 20mm, 250Vac, fast-blow types. The OPEN and CLOSE initiate uses 5-Ampere, 3AG, 250Vac, slow-blow type fuses.
2.3 LCD Screen Contrast Control
To increase the LCD screen contrast, press and hold the
[PAPER Contrast] key for two
seconds.
To decrease the LCD screen contrast, press and hold the [PAPER Contrast] key for two seconds.
10
DIGITMR USER’S MANUAL REV 3
2.4 Printer Paper Control
To advance the DigiTMR’s printer paper, press and release the [PAPER Contrast] key.
To retract the DigiTMR’s printer paper, press and release the [PAPER Contrast] key.
2.5 Printer Paper
The DigiTMR’s built-in thermal printer uses 4.5-inch wide thermal paper for printing test results. To maintain the highest print quality and to avoid paper jams, the use of thermal paper supplied by Vanguard Instruments Company is highly recommended. Additional paper can be ordered from the following sources:
Vanguard Instruments Co, Inc.
1520 S. Hellman Avenue Ontario, CA 91761 Tel: 909-923-9390 Fax: 909-923-9391 Part Number: VIC TP-4 paper
BG Instrument Co.
13607 E. Trent Avenue Spokane, WA 99216 Tel: 509-893-9881 Fax: 509-893-9803 Part Number: VIC TP-4 paper
2.6 Replacing the Thermal Printer Paper
The roll of thermal paper is housed inside a dispenser underneath the printer cover. To replace the paper, follow the steps below:
Unscrew the two large printer cover screws and remove the printer cover.
Remove the leftover thermal paper roll from the paper holder.
Unroll the new thermal paper roll.
Feed the thermal paper into the slot between the paper pocket and the rubber roller.
The printer will automatically pull the paper under the thermal head.
Place the paper roll into the paper holder.
Lift the thermal head and align the thermal paper if necessary.
Re-install the printer cover.
Thermal paper has a chemical coating on one side of the paper. This side should be facing the thermal print head. Incorrect paper loading may result in blank output on
NOTE
the thermal paper.
The thermal paper will show a red stripe to indicate that the roll is about to run out of paper.
11
REV 3 DIGITMR USER’S MANUAL
3.0 OPERATING PROCEDURES
3.1 DigiTMR Cable Connections
3.1.1. Contact Cable Connections
A typical contact cable connection to a circuit breaker is shown in Figure 3. Red clips are connected to phase A, B, and C of the breaker’s bushings. The black clips are connected to the ground or common side of the bushings. If the common side of the bushing is not grounded, jumpers should be installed between the bushings.
It is advisable to ground one side of the contacts for most testing purposes. If a breaker is floating or ungrounded, ensure that the contact input channels are
NOTE
protected against static discharge.
12
Figure 3. Contact Cable Connections
DIGITMR USER’S MANUAL REV 3
3.1.2. Initiate Cable Connections
The DigiTMR will trip or close breakers through a solid-state device which will operate on any AC or DC control voltage ranging from 10 to 300 Volts. Both the trip and close circuits are protected by 5-Ampere fuses. A typical DC trip and DC close control circuit test connection is shown in Figure 4. A typical DC trip and AC close control circuit is shown in Figure 5.
Figure 4. DC Trip and DC Close Initiate Circuit Cable Connections
13
REV 3 DIGITMR USER’S MANUAL
Figure 5. DC Trip and AC Close Initiate Circuit Cable Connections
14
DIGITMR USER’S MANUAL REV 3
3.1.3. Analog and Digital Vo ltage Monitor Connections
The analog voltage input “V1” can monitor a breaker’s DC control voltage during an operation. The analog input records the nominal DC voltage at no load and the minimum DC voltage while the Trip or Close coil is energized.
Nominal and minimum voltage readings can be printed on a tabulated report. Analog waveforms can also be plotted graphically. Using the reports, the user can see the breaker’s DC control voltage “dip” under load conditions and can easily detect problems such as a poor connection or an excessive voltage drop.
The digital voltage input “V2” monitors the voltage status as “ON” or “OFF” states and can graph the results.
A typical analog voltage monitoring connection (V1) and a typical digital monitoring connection (V2) are shown in Figure 6.
Maximum voltage recorded is 255 Vdc.
NOTE
Figure 6. Analog and Digital Voltage Monitoring Cable Connections
15
REV 3 DIGITMR USER’S MANUAL
3.1.4. External Trigger Input Connections
The External Trigger Mode can be used to start recording when the DigiTMR senses a voltage. A typical application for the External Trigger Mode is to time a circuit breaker in a close operation and to start timing only when the close coil is energized, thus bypassing the 52X relay delay time.
Since the 52X relay carries the close coil current, the initiate cable must be connected to the close terminal as shown in Figure 4. The DigiTMR will energize the 52X relay to start the close operation and will then start timing when it senses the voltage across the closing coil. Please see Figure 7 for a typical test connection. Another application for the External Trigger Mode is to start timing the breaker when the user trips or closes the breaker remotely.
Minimum trigger voltage is set at 30 Vac/dc. Maximum continuous voltage is set
NOTES
at 300 Vac/dc.
Different trigger voltages can be set at the factory based on specific user
requests.
The DigiTMR will start looking for the external trigger voltage when the
message “AWAITING TRIGGER…” is shown on the LCD. If the external trigger voltage is not sensed by the DigiTMR within ten seconds after the initiate sequence has begun, the DigiTMR will return to the main menu.
16
Figure 7. External Trigger Cable Connections
DIGITMR USER’S MANUAL REV 3
3.1.5. AC Current Sensor Connection
The DigiTMR can sense a breaker’s bushing current by using a clamp-on AC current sensor connected to the breaker’s bushing CT. The AC current sensor is used in Quick-Shot Mode only. The AC current sensor requires a minimum of 100 mA to operate. The maximum current is 25 Amperes. A typical AC sensor connection is shown in Figure 8.
Figure 8. Typical CT Connection
17
REV 3 DIGITMR USER’S MANUAL
3.1.6. Transducer Connection
A typical transducer connection is shown in Figure 9. See Appendix F for additional transducer illustrations.
18
Figure 9. Typical Transducer Connection
DIGITMR USER’S MANUAL REV 3
3.2 General Procedures
The main steps for timing a circuit breaker are listed in Figure 10 below.
Connect the DigiTMR
Ground the DigiTMR (see note)
Connect the DigiTMR cables to the circuit breaker
Install the transducer and cable
Apply Power
Turn on the power for the DigiTMR. Once the unit
goes through its start-up cycle, the “Start-Up” menu will be displayed.
Perform self test (hook-up test, cable test) if
necessary
Time Breaker
Select DigiTMR test
Operate Breaker
Get Test Results
View or print tabulated results
Print graphic results
Save timing shot
Figure 10. DigiTMR General Operating Procedures
To reduce the possibility of damaging the DigiTMR due to static discharge in high voltage substations, it is recommended that one side of the breaker’s bushings be
NOTE
grounded for the contact channels.
3.3 Timing Breakers in Quick-Shot Mode
The DigiTMR can provide OPEN or CLOSE times only in Quick-Shot Mode. The DigiTMR can initiate the OPEN or CLOSE tests, and then the test report will show the OPEN or CLOSE coil current reading and waveform. If the DigiTMR is operated in Quick-Shot mode using the External Trigger, no coil current reading or waveform will be recorded.
19
REV 3 DIGITMR USER’S MANUAL
3.4 Timing Breakers in Time-Travel Mode
The DigiTMR can initiate the breaker operation and perform a timing test on the following operations:
OPEN
CLOSE
OPEN-CLOSE
CLOSE-OPEN
OPEN-CLOSE-OPEN
The DigiTMR can start the OPEN-CLOSE operations without a delay or by using a
NOTES
programmable delay between the OPEN and CLOSE commands.
The CLOSE-OPEN operation can be started by the DigiTMR using several
options:
o Contact #1 Close
The DigiTMR can initiate a CLOSE command and then an OPEN command after detecting that the breaker’s contact was closed. The DigiTMR can detect when the breaker’s contact is closed via contact channel #1. The selection of this option is recommended for the CLOSE-OPEN operation since it truly represents that the circuit breaker closed and then opened.
o Set DELAY
The DigiTMR can initiate a CLOSE command and then an OPEN command after a programmable delay (in milliseconds).
o No DELAY
The DigiTMR can initiate a CLOSE command and then an OPEN command without a delay. Since the circuit breaker is in the open state, the circuit breaker’s 52B contact allows the close coil to be energized, which can start the close command. When the circuit breaker is making the transition from an open state to a close state, the 52A contact will close, allowing the breaker to initiate the open command.
The OPEN-CLOSE-OPEN operations can be initiated by the DigiTMR using a
programmable delay between each of the operations. The delays can be set between the OPEN to CLOSE and between the CLOSE to OPEN operations.
20
DIGITMR USER’S MANUAL REV 3
3.5 Performing Timing Tests
3.5.1. Timing an OPEN Operation With No Insertion Resistors
To time an OPEN operation with no insertion resistors:
a. Start from the “START-UP” menu:
1. TIME BRKR 04/17/09
2. GET RSLT 10:49:23
3. SET-UP
4. DIAGNOSTICS
Press the [1] key (TIME BRKR) to start a test.
If a previous test shot is in the DigiTMR’s memory, the following screen will be displayed:
PREV SHOT NOT SAVED
1. SAVE SHOT
2. CONTINUE (NO SAVE)
1. SAVE SHOT
If you would like to save the previous shot in memory, press the [1] key (SAVE SHOT). The following screen will be displayed:
SAVE SHOT # 0001 TEST: OPEN DATE: 04/12/09 “ENTER” TO PROCEED
Press the [ENTER] key. The following screen will be displayed:
SAVING IN PROGRESS
PLEASE WAIT
Once the record is saved, the following screen will be displayed:
SHOT SAVED
Press the
[ENTER] key and you will be returned to the “START-UP” menu. Re-
start at step a.
2. CONTINUE (NO SAVE)
21
Loading...
+ 60 hidden pages