
ELECTROSURGICAL
VALLEYLAB
GENERATOR
SSE3B
PRODUCT
INFORMATION
INTENDED
The
Model
generator
monopolar
for
cutting
BLEND
degrees
respectively.
procedures
cutting
settings
for
details.
desiccation
independently
instruments,
instruments.
CONTRAINDICATIONS
There
The
patient's
use
use
with
and
#3
or
are
USES
SSE3B
intended
and
and
of
must
no
of
condition
Electrosurgical
bipolar
a
patient
fulguration.
COAG
hemostatis,
The
including
fulguration
be
used
The
bipolar
without
controlled.
it
is
known
external
for
output
return
are
intended
monopolar
transurethral
and
fulguration
also
absolute
or
may
require
Generator
use.
in
connections.
electrode
Its
voltage
and
output
using
output
possible
internal
the
the
surgeon
Although
contraindications
special
the
operating
for
smooth
fulguration
is
resections.
monopolar
is
is
an
isolated
or
cutting.
designed
to
use the
pacemakers,
precautions.
is a general
The
and
is
waveforms,
cutting,
with
powerful
output
referred
The
primarily
bipolar
to
monitoring
room.
monopolar
designed
CUT,
cutting
a
enough
Desiccation
to
output
bipolar
the use
purpose
The
output
for
BLEND.
minimum
is
practical,
the
which
for
output
equipment,
electrosurgical
SSE3B
is
electrosurgical
fl,
with
of
for
Instruction
output
use
of
all
with
is
intended
with
with
electrosurgery.
has
BLEND
increasing
power
monopolar
both
intended
cutting,
common
minimal
but
Manual
bipolar
and
#2,
low
for
is
the
WARNINGS
Never
both
Burns
monopolar
surgeon's
inherently
or
Electrosurgical
一
一
一
pacemaker
the
electrosurgical
mode
increase
the
to
objects.
heart
or
active
the
active
hands.
unsafe
because
which
inhibit
the
and
the
surgeon's
electrode
Electrosurgery
for
units
of
could
current
the
pacemaker
power
use
the
beyond
patient
hands
with
should
danger
cause
can
return
are
is
touched
involves
flammable
be
of
fibrillation.
cause
entirely.
the
normal
electrodes
possible
used
introducing
a
pacemaker
settings
in
to
a
electric
anesthetics
with
caution
without
and
their
most
metal
clinical
instrument
sparking
or
in
electrosurgical
Also,
to
interference
revert
connections.
situations
to
near
flammable
the
to
first
held
tissue
presence
currents
an
asynchronous
checking
in
and
fluids
from
if
of
into
a
the
is
a
the

PRECAUTIONS
The
surgeon
such
standby
a
cannot
higher
Low
return
pass
as
significant
frequency
directly
should
random
generator
detect
than
electrode
normal
component
and
risk
loose
current
should
through
be
aware
failure
accessories
to
the
connections
low
frequency
components
be
the
region
that
patient.
placed
the
or
defective
are
in
the
components
could
so
that
of
the
Model
recommended
The
SSE3B
SSE3B
active
cause
the
heart.
can
malfunction
active
electrode
in
electrosurgical
or
patient
if
generator
Return
the
electrosurgical
fibrillation,
Fault
which
for
electrodes.
malfunction
safety system
may
result
current.
so
the
current
will
reasons
A
is
in
patient
not
POTENTIAL
The
most
unintended
patient
combinations
complication
bipolar
PREVENTATIVE
To
insure
checked
Of
particular
lint
A
steps
Test
thorough
that
to
insure
and
taken
the
Procedure,"
COMPLICATIONS
common
burns
return
of
is
the
electrosurgery
MAINTENANCE
safe
and
for
wear
and
importance
dirt
build-up,
visual
to
correct
generator
outlined
continued
complications
to
the
patient
electrodes,
fault
dependable
inspection
be
efficacy
conditions,
explosion
are
tear.
are:
power
any
inspected
in
very
operation,
printed
cord
of
and
the
and
associated
and
insufficient
of
flammable
unusual.
wear,
the
all
of
at
least twice
service
safety.
surgeon
or
poor
this
circuit
and
generator
the
above
manual,
with
monopolar
caused
patient
surgical
bowel
generator
board
looseness
should
conditions.
a
year
be
by
gas.
should
contact
of
mechanical
be
made
and
performed
electrosurgery
defective
electrode
technique.
Complications
be
finger
and
It
is
that
the
at
active
contact,
À
rare
with
periodically
corrosion,
parts.
appropriate.
recommended
"Acceptance
these
times
are
or
ii

THE
SSE3B
SYSTEM
Congratulations.
generator
available
rivals
effect.
cutting
irrigated
directly
separate,
characteristic
applications.
bipolar
SSE3B
safety
grounded
Electrode
In
order
capabilities
information
facts
understanding
situations
For
the
Sections
and
limitations
should
must
be
ever
in
the
The
and
procedures.
in
isolated
instruments
is
equipped
system
electrosurgery.
Monitoring
to
and
surgeon
1
be
able
used
Your
built.
a
single
best
PURE
watts
spark
variable
CUT
with
bipolar
which
Hand-switch
with
which
instruct
and
must
specifications.
and
and
to
limitations
be
of
his
recognize
the
2.
These
of
the
to
decide
accomplish
new
It
generator.
gap
BLEND
is
exceptionally
For
a
large
is
ideal
as
well
the
combines
System
the
provided.
equipment
potential
most
sections
generator.
what
SSE3B
offers
generators
gives
your
lighted,
output
for
activation
as
conventional
"Balanced
the
The
SSE3B
(REM).
user
Ideally,
important
procedures
them.
is
the
features
The
a
convenience
which
neurosurgery,
may
significant
may
of
an
of
his
The
user
so
he
hazards
contain
After
most
and
SSE3B
and
wide
smooth
digital
has
Output"
also
electrosurgical
he
can
parts
reading
his
has
with
range
and
the
display.
a
non-sparking
be
used with
footswitch
patient
advantages
be
generator,
must
before
the
be
should
use the
of
basic
SSE3B
versatile
performance
COAG
a
minimum
of
starts
output
laparoscopy,
both
equipped
a
given
be
given
they
occur.
these
surgical
these
sections
can
electrosurgical
never
fulguration
of
hemostasis
promptly
is
The
SSE3B
(desiccation)
monopolar
activation.
return
of
isolated
with
generator
great
more
a
generator
instructions
capabilities
handle
and
before
which
cutting
while
even
calibrated
has
or
other
electrode
the
Return
of
deal
than
a
just
complete
in
surgeon
how
in
a
and
The
and
the
of
new
are
it
The
basic
their
Section
electrical
The
technical
increasingly
such
Sections
If
Valleylab
hesitate
Service
Department
problems.
generator
remainder
as
any
problem
is a lack
9
is
a
layman
of
topics
relevant
the
circulating
3
through
questions
representative.
to
call
Department,
are
always
most
short
and
the
which
8
as
arise
or
surgeons
of
primer
Section
Instruction
are
for
the
nurse
well
about using
If
write
Maintenance
our
happy
and
knowledge
on
10
is a glossary
of
less
hospital
and
as
Sections
there
factory.
Service
to
nurses
of
electricity.
electricity
Manual,
concern
engineer.
OR
supervisor
1
and
your
are
questions
The
help
you
have
Sections
new
Department,
in
and
of
terms.
to
will
2.
SSE3B,
he
personnel
with
trying
To
fill
electronics
3
through
the
surgeon
Other
be
call
can't
in
and
your
electrosurgical
to
understand
this
8,
OR
personnel
interested
your
answer,
our
Customer
Engineering
for
but
need,
cover
local
do
the
are
in
not
iii

SECTION
1
ELECTROSURGICAL
Electrosurgery
to
sever
a
low
muscles
could
Although
to
4,000,000
become
confine
operates
extremes.
Medical
is
passed
used
is
never
diathermy"
difference
electric
tissue
frequency,
and
electrocute
a
quite
these
at
diathermy
so
that
high
current.
THEORY
can
or
nerves.
generator
Hz
prominent
radio
750,000
through
the
enough
for
electrosurgery.
between
be
achieve
say
Putting
the
patient.
could
(4
MHz),
frequency
Hz
is
similar
the
current
these
APPLIED
defined
below
at
(750
patient's
density
to
cause
two
TO
THE
as
the
hemostasis,
100,000
it
another
perform
reactive
such
high
currents
KHz)
to
electrosurgery
body.
is
necrosis.
This
techniques
SSE3B
use
of a radio
A
high
Hz
(cycles
way,
electrosurgery
phenomena,
frequencies
to
wires.
which
kept
is
Large
low
The
is
very
is
the
radio
per
a
low
quite
capacitance
and
For
a
compromise
in
that
entry
and
the
British
descriptive.
density
frequency
frequency
second)
frequency
well
it
becomes
these
radio
and
resulting
use
of
at
reasons
between
frequency
exit
the
the
electric
is
used
will
electric
frequencies
and
difficult
electrodes
tissue
term
since
radio
current
because
stimulate
current
inductance,
the
SSE3B
these
current
heating
"surgical
the
major
frequency
up
to
two
are
There
electrosurgery.
Electrosurgical
Electrosurgical
effect.
Electrosurgical
significant
char
The
1.
Electrosurgical
Of
waveform
passed
in
wire
through
are
tissue.
SSE3B
the
the
is
three,
(either
through
tissue.
in a toaster
it.
only
Desiccation
Cutting
Fulguration
cutting.
optimized
Desiccation
desiccation
the
This
three
They
CUT
or
electrical
is
or
surgical
are:
-
-
Electric
Fulguration
to
perform
COAG)
exactly
stove
which
Low
-
Electric
is
technically
can
resistance
the
effects
power
sparking
can
all
three
be
used.
same
becomes
which
coagulation
to
sparking
coagulate
of
these
the
In
of
the
phenomenon
hot
when
can
be
without
tissue
to
large
functions,
simplest
desiccation,
tissue
as
the
achieved
with
tissue
bleeders
because
and
the
current
sparking.
a
cutting
without
current
heat
arises
resistance
passes
with
and
ANY
is

The
and
the
BLEND,
monopolar
thus
surgery
or
very
COAG
in
output
low
the
may
is
primarily
settings
desiccation
be
used
(less
for
designed
than
mode.
desiccation.
30
As
for
watts)
stated
fulguration
must
above,
be
any
and
used
mode,
cutting
to
keep
CUT,
Electrosurgical
In
electrosurgical
so
rapidly
matrix.
conduct
The
through
electrode
exploded
It
is
electrosurgical
tissues,
and
also
much
electrode
considerably
electrosurgical
be
mentioned
current
cells
When
the
that
the
bright
passing
radio
"spark".
intervals
Significant
Cutting
that
heat
is
moved
the
possible,
the
is
pressed
helped
that
it
contacts.
frequency
light
An
"arc"
to
become
role
cutting
they
is
the
explode
dissipated
tissue
and
incision
and
electrode.
way
cooking
against
by
cutting
a
hot
into
in
the
the
is
in
electrosurgery.
the
objective
into
in
to
fresh
is
made.
even
useful,
Moreover,
softens
tissue,
electrosurgical
involves
cautery
tissue)
current
air
gap
a
similar
established
steam
the
dry
tissue
[2]
sparking
wire
can
jumps
is
technically
phenomenon
and
is
leaving
steam
out
is
to
cut
desiccation
a
tough
the
desiccation.
(a
heated
also
across
probably
to
heat
and
therefore
adjacent
contacted,
mechanically
pot
mechanical
to
tissue.
wire
cut
an
known
which
a
cavity
roast.
tissue
air
does
the
cells.
new
softens
cutting
However,
It
with
gap
as an
requires
tissue
in
it
cells
with
If
should
no
by
to
not
cells
the
cell
does
not
When
are
a
sharp
many
a
sharp
may
true
also
electric
exploding
tissue,
electric
longer
play
the
be
a
To
understand
electrosurgical
with
"7"
is
leaves
Voltage
also
voltage
since
a
on
high,
the
air,
small
the
the
the
is
force
is
desiccated
Once
result
the
where
from
the
two
the
the
a
spark
of
two
current
the
spark
energy
spark
types
heating
desiccation
wire
SSE3B.
tissue
tissue,
the
force
that
high
once
tissue.
to
phenomena.
passing
dissipated
is
actually
of
heating
is
the
cutting,
electrode
Because
becomes
the
that
drives
enough,
it is
the
tissue
through
strikes
extremely
phenomenon
difference
let's
the
electrical
drives
electric
a
spark
ionized,
is
The
the
flesh,
in
the
greater
together
concentrated.
alone
between
suppose
at a relatively
power
level
desiccated
resistance
current
sparks
will
jump
makes
established,
first
the
spark
than
are
is
resistance
The
itself.
that
capable
is
practically
the
remainder
that
(heat
very
through
across
to
a
better
of
arising
It
the
one
high
the
the
tissue
the
The
of
exploding
turns
quickly.
cautery
tries
setting
delivered
of
the
a
resistance.
an
air
nearest
conductor
tissue
heating
tissue
of
the
heat
in
the
out
that
impossible.
wire
to
desiccate
of
CUT,
per
As
the
tissue
gap.
moist
than
heating
produced
at
the
heating
originating
tissue.
cells
cutting
[3]
and
say
second)
water
rises,
It
If
the
tissue
the
is
the
point
comes
The
because
by
is
by
in

σος
The
essential
continuous
is
plotted
alternating
the
generator,
ELECTROSURGICAL
Electrode
tissue
Short,
cells
can
scalpei.
TYPICAL
sinewaves.
over
from
is
separated
by
thin
intense
into
steam.
be
minimai
CURRENT
characteristic
time,
positive
750
KHz,
layer
of
sparks
Hemostasis
and
That
a
pure
from
steam.
flash
mimics
と て の
TISSUE
=
0.1
is,
to
AMP
of
if
CUT
negative
CUTTING
the
SPARKS
output
is
a
operating
is
that
of
continuous
they
the
generator
sinewave
frequency
RMS
CUT
the
voltage
waveform
waveforms
at
are
of
EQUAL
PEAK
WAVEFORMS
ーー
A
GUT
TEA-
HH
A
COAG
The
peak
waveforms,
COAG
voitage
but
waveform.
the
τ
==
waveform - a
waveform - bursts
is
ΜΗ
pure
the
power
VOLTAGE
sine
wave.
Crest
HH
af
sina
waves.
Crest
same
is
in
both
about
factor
factar
1/3
1.4
4,3
of
these
in
the
=
~

Electrosurgical
The
SSE3B
sinewaves.
bursts
waveform
had
delivered
most
Now
voltage)
second.
produce
voltages
occur
the
same
of
the
suppose
the
and
is
as
Because
COAG
The
the
(heat
time.
that
the
same
currents
Fulguration
waveform
frequency
31,250
pause
peak
per
the
CUT
the
RMS
consists
times/second.
between
voltage
second)
COAG
waveform
COAG
voltage
during
of
the
each
as
the
is
waveform
and
is
turned
the
less
as
periods
of
sinewave
The
burst.
CUT
had
thus
off
the
short
important
Suppose
waveform,
because
the
could
most
CUT
when
bursts
is
the
same
deliver
of
by
the
of
radio
750
KHz
feature
that
a
but
the
COAG
average voltage
the
having
generator
the
time,
very
is
same
and
of
COAG
average
turned
it
large
is
frequency
the
COAG
the
COAG
waveform
power
off
(RMS
heat
can
on.
per
only
peak
A
good
effect
and
because
the
water
In
this
form
an
waveform,
this
out
general
way
of
EQUAL
A
COAG
second)
COAG
example.
are
equal.
COAG
because
the
waveform
the
in
the
way
the
incision.
it
can
it is
tissue
term
which
POWER
RMS
volts
Peak
volta
ЕТ
ESA.
waveform
to
the
peak
voltage
Note
the
heat
heating
cells
celis
Because
drive
possible
and
includes
AMS
voits
above
aiso
can
are
actually
A
CUT
wavetarm
with
is
that
spark
is
effect
does
dehydrated
of
a
current
to
both
CUT
about
the
more
not
fulgurate
with
low
equal
waveform.
root
to
widely
is
get
the
high
char
desiccation
WAVEFORMS
peak
voltage.
power
three
mean
tissue
intermittent.
high
slowly
through
it
times
dispersed
peak
long
to
a
(energy
Note
square
without
enough
very
after
carbon.
that
higher
to
but are not
voltage
high
and
fulguration.
per
the
in
this
voltages
significant
by
the
The
temperature
flash
torn
of a quality
resistances.
the
water
"Coagulation"
long
into
is
cutting
sparks
steam.
apart
driven
COAG
In
is
of
to
a

Fulguration
First,
always
surprising
Spark
desiccation,
contact
may
For
efficient
only
For
electrode
The
will
out,
Since
jump
the
sparking
produces
and
or.
may
a
given
one
example,
initial
be
its
moist
to
voltage
when
each
between
not
at
fifth
will
tissue
high,
resistance
tissue
the
nearest
is
can
be
to
necrosis
you
spark
the
the
produce
level
producing
the
if
a
begin
typically
high
contrasted
tissue
consider
current
electrode
of
average
ball
in
resistance
rises
is
moist
enough
with
anywhere
that
has
an
is
necrosis,
current
necrosis.
current
electrode
desiccation
0.5
to
until
no
longer
tissue
to
extremely
no
and
is
0.8
make
with
any
practical
the
each
more
the
tissue.
depending
flow,
In
flow
is
pressed
mode,
quite
amperes
the
touching
in
the
a
spark.
desiccation
sparks
cycle
concentrated
general,
of
low
electrical
of
high
As
on
fulguration
desiccation.
against
regardless
and
RMS.
the
fulguration
[1]
in
fulguration
land.
voltage
current
than
a
result
the
current
fulguration
moist
of
the
resulting
As
the
contact
electrode,
mode,
several
This
produces
density.
is
the
generator
is
the
area
desiccation
density.
always
requires
tissue,
waveform.
current
tissue
is
broken.
sparks
as
long
ways.
not
à
new
more
the
dries
will
In
of
as
Electrode
Long
sparks
resuit
coagulation
as
continues.
in
then
deeper
tulguration
hard
and
superficial
TYPICAL
FULGURATION
free
from
tissue.
to
tissue
first,
necrosis
Eschar
black.
is
CURRENT
=
0.1
AMP
RMS
LT

The
on-off
quantity
peak
is
voltage
1.4.
constant
factor
COAG
generator,
ever
is
adjustments
characteristic
called
to
The
crest
over
the
offered
capable
but
or
parts
crest
RMS
factor
entire
of
has
of
factor.
voltage.
of
control
in
a
full
fulgurating
no
spark
replacement.
COAG
The
the
power
Crest
SSE3B
as
gaps
waveforms
factor
crest
COAG
range.
solid
well
and
is
factor
is
This
state
as
does
can
be
defined
of a pure
10.0
is
generator.
a
good
not
described
as
and
is
the
highest
spark
require
with
the
ratio
CUT
sinewave
essentially
The
gap
routine
a
of
crest
SSE3B
type
Blended
As
one
hemostatic
BLEND
the
The
have
current
electrode.
SSE3B
a
sinewave
this
reason
urologists
#1
which
higher
high
crest
solid
state
electrode
surgeons
because
without
the
same area
CUT
might
expect,
effect,
has a crest
crest
cuts
the
find
has a crest
crest
factors,
factor
generators.
where
doing
it
will
having
of
Crest
that
will
factor
too
The
be
finer
cleanly
SSE3B
that
even
factor
that
a
maximum
polypectomy
allow
to
worry
electrode
Factor
the
is,
well
factor
of
CUT
3.6
it
BLEND
them
about
=
BLEND
the
fulgurated,
the
electrode,
of
1.4,
and
crest
the
SSE3B
of
and
is
equivalent
of
hemostasis
through
to
as
more
Peak
RMS
is
walls
1.9
but
the
2.6.
4.4
#3
cut
bleeding.
hemostasis
Voltage
(effective
a
cutting
of
depending
the
in
CUT.
many
surgeons
incision
factor
CUT
is
BLEND
respectively.
to
is
useful
is
an
endoscope
through
average)
the
cleaner
A
bleeds
is
too
#2
"COAG"
for
needed.
the
BLEND
is
waveform
incision
on
pure
slightly
clean
and
BLEND
BLEND
in
cutting
will
pedicle
powers
added.
voltage
with
made
the
fineness
the
cut.
CUT
sinewave
find
that
excessively.
higher.
and
prefer
#3
#3
many
competitive
with
For
example,
prefer
in
are
with
have
has
BLEND
one
equal
moderate
the
would
a
pure
For
Many
BLEND
still
such
a
thin
many
pass
for
of
a
#3
A.
common
that
the
can
be
COAG
knobs.
Valleylab
level
button
COAG
is
The
relationship
next
page.
three
mode
is
tissue
contact
moist
it
will
continue
misunderstanding
degree
varied
This
generators
is
controlled
is
pushed,
unaffected.
Coagulation
modes
is
identical
around
with
the
tissue.
jump
for
sparks
the
of
by
shown
the
Since
hemostasis
dialing
is
by
the
between
and
electrode
metal
the
to
longest
about
in
true
of
have
the
CUT
CUT,
of
as
power
thus
electrode,
peak
more
period
American
(the
various
some
fixed
CUT
waveform
BLEND
liver
plotted
each
becomes
voltage
distant
of
crest
“percentages"
generators
BLEND
knob.
is
and
with
mode
sparks
of
moist
time.
electrosurgical
factor)
made
crest
In
factors
other
converted
COAG
test
electrodes
over
desiccates
too
time.
dry
begin
the
COAG
tissue
The
COAG
of
on
in
words,
to
is
illustrated
equally.
for
to
jump
waveform
and
waveform
the
both
a
The
the
generators
BLEND
Europe,
and
when
BLEND
in
power
good
to
sparking
waveform
the
CUT
but
the
a
and
on
each
of
in
After
electrical
the
nearest
is
highest,
fulgurates
is
and
all
power
BLEND
the
the
the
each
the
will

a
wide area
concentrates
they
explode
One
way
to
electrodes
electrode
the
electrode
output
resistances
designed
available
desiccation
small,
so
bipolar
keep
so
is
that
at
in
power.
with
is
a
the
power
to
produce
the
that
prevented
from
it
which
this
complete.
generators
low
output
sparking
will
way.
It
power
so
that
the
of a generator
the
the
desiccates
desiccation
from
not
sparking
It
designed
density
tissue
cutting
sparking
to
deliver
has
It
is
tissue
starts.
a
very
rapidly
comparable
for
while
cells
effect.
phase
to
tissue.
is
power
low
and
neurosurgery.
CUT
has
are
heated
from
cutting
is
to
design
efficiently
The
SSE3B
output
stops
in
a
low
intensely
is
prolonged
Another
the
to
bipolar
impedance
abruptly
performance
voltage
to
way
those
output
and
until
use
large
and
generator
and
when
to
tissue
to
the
keep
less
the
the
TS
is
MONOPOLAR
The
pin
five
radio
patient's
dispersive
The
detect
If
will
In
addition,
dangerous
PROCEDURES
SSE3B
receptacles
SSE3B
1)
the
also
5)
7)
has
receptacle
frequency
is
the
Patient
2)
Broken
REM
detect:
3)
Insufficient
4)
Patient
ground
Active
6)
Excessive
(Current
than
A
patient's
detected
two
on
jacks
body
electrode
following
and
equipped
patient
System
the
currents actually
electrode
2000
single
WITH
separate
current
plate
plate
SSE3B
THE
the
lower
on
returns
which
with
fault
and
CoHesive
contact
will
current
.flowing
ohms
area
skin
as a fault.
SSE3B
outputs.
the
travels
contacts
a
conditions
cable
plate
not
touching
detect
touching
through
impedance
patient
(less
right.
lower
to
unigue
not
cable.
Return
area
flowing
than
The
The
left.
from
the
a
patient
during
plugged
of
patient's
these
result:
grounded
from
more
between
plate
one
bipolar
monopolar
In
the
generator
large
plate
into
Electrode
the
patient
hazards,
the
than
active
which
cm2
contact
output
monopolar
active
area
surgery:
generator.
skin.
object.
active
120
and
is
outputs
electrode,
via
a
of
monitor
is
being
plate.
provided
electrode
pf
ground.)
barely
area)
is
the
group
are
electrosurgery
patient
the
patient's
system
used,
that
capacitance
contacting
can
frequently
of
the
group
through
plate
which
the
potentially
to
ground.
or
four
of
the
the
or
skin.
will
SSE3B
less
|
the
be
10

When
will
operate,
by
The
To
actually
lights
on
power
on
no
any
illuminate
pushing
CUT
detect
the
until
significant
one
the
the
and
a
working,
will
output
controls
the
of
fault
indicator
COAG
fault
only
jacks.
control
RF
these
and
must
Lights
in
the
illuminate
to
zero
output
seven
the
the
is
generator
be
button.
as
Safety
ACTIVE
CUT
You
can
and
advanced
below
faults
corrected
circuit,
and
when
test
keying
this
will
and
COAG
there
the
to
is
detected,
not
and
Check-Out
or
indicator
is
this
generator.
at
least
setting.
the
a
by
the
operate.
Return
Indicators
to
find
lights
useful
turning
"1",
RETURN
To
make
Fault
out
if
may
radio
either
The
lights
indicating
FAULT
the
indicator
the
generator
be
used.
frequency
the
CUT
will
that
indicator
generator
reset
The
output
or
COAG
not
come
there
is
is
Suppose
light
illuminates,
circuit)
from
Testing
The
inward.
is
is
tone
RETURN
return
simplest
1
the
generator
until
FAULT
detect
The
is
The
generator
conditions
is
the
RETURN
intact.
pressed,
and
1/2"
aluminum
return
applied
return
FAULT
fault
to
the
circuit
faults
the
patient
OFF,
active
generator
the
Return
FAULT
The
indicator
If
the
no
CUT
way
to
10,
should
indicator
area
by
fault
and
should
but
the
you
may
cable
to
Fault
monitor
one
attempts
generator
or
circuitry
system
do
this
then
handle
that
touch
deactivate.
only
have
indicator
noting
alarm
touching
disable
plate
generator
conclude
or
electrode
the
Monitor
system
should
COAG
is
is
is
the
of
is
reset
responds
not
the
can
seems
active
light
to
should
indicator
operational.
to
actually
to
set
active
the
SSE3B.
The
by
to
yet
(REM)
REM
alarm
also
the
electrode
the
generator.
to
be
does
not
that
and
electrode.
System
can
be
and
activate
not
lamps.
either
electrode
RETURN
pressing
actual
produced
can
be
even
be
applied
operate.
the
the
and
tested
this
the
activate,
The
simulate
power
The
it
ground
excessive
tested
tested
to
properly,
problem
current
Electrode
by
verifies
generator
This
only
control
to a grounded
indicator
FAULT
inward.
before
when
the
after
the
If
the
is
cannot
Area
pressing
as
indicated
verifies
way
a
fault
should
indicator
Note,
current
ground
generator
placement
handle.
a
that
while
to
to
flow.
current.
the
the
return
CUT
or
defective
complete
(REM)
the
red
the
the
by
that
most
test
condition.
any
setting
object,
light
should
that
return
is
by
Either
COAG
the
System
indicator
light
indicator
no
the
such
and
stay
the
It
doesn't
electrode
not
keying
of
fault
light
(open
path
bulb
audio
of
the
entire
The
from
the
lit
RETURN
keyed.
the
these
as
Dispersive
Section
(10
square
close
as
as
large
electrode
6.
In
inches)
possible
an
area
summary,
for
to
of
skin
placement
we
each
the
site
as
and
recommend
100
of
surgery
possible.
watts
12
size
using
of
requirements
a
patient
power.
and
should
plate
The
be
plate
in
are
firm
discussed
of
about
should
contact
65
be
in
cm2
as
with

We
recommend
placed
used,
touch
curved
surgery.
should
to
all
firm
relocate
trouble
comes
unless
on
it
is
before
or
be
adhesive
contact
the
with
on,
the
a
the
plate.
critical
it is
convex
Avoid
shaved.
of the
pad
the
the
problem
thin
applied
area
bony
Position
areas.
after
patient
adhesive
obviously
layer
to
prominences
conductive
If
of
pad
of
a
insure
dry
Then
its
cable
good
disposable,
that
to
the
skin,
and
the
pad
press
gel
initial
assembly,
should
has
some
conductive
pregelled,
the
gel
skin.
as
near
fatty
carefully
the
back
with
application.
be
removed
other
pad
Apply
tissue.
of the
the
that
cause.
gel,
is
the
as
possible
and
apply
patient.
is,
from
such
dispersive
the
as
thoroughly
return
Extremely
fingertip
entire
In
case
RETURN
the
skin
Lectrogel
electrode
to
pad
Do
not
electrode
wet
the
hairy
pressure
to
reuse
of
suspected
FAULT
and
replaced
to
site
assure
light
,
be
is
the
to
of
areas
or
a
Limitations
The
SSE3B
devised.
to
the
operating
do
not
occur:
l.
2.
Insufficient
A
patient
than
appears
which
electrode
area.
for
Careless
For
perhaps
("hot")
surgeon
holster
of
RETURN
However,
20
the
should
A
units
Placement
example,
in
at
is
to
the
SSE3B
FAULT
there
room
Patient
electrode
cm2)
same
have
area
special
featuring
if
a
coronary
the
done
be
certain
Return
system
are
staff
Electrode
which
in
good
electrically
at
monitor
dual
REM.
Of
two
active
same
with
an
certain
to
electrical
least
(REM)
section
One
(or
bypass
time
electrode,
that
Fault
is
insure
has
and
it
Safety System
the
faults
Area
only
as a properly
100 cm2
system
patient
Both)
electrodes
procedure,
must
cannot
best
that
of
it
monopolar
it
the
a
small
contact
contact
will
electrode
Two
are
be
guarded
must
touch
cannot
following
area
with
applied
area.
detect
Active
being
both
be
returned
the
patient
safety
detect
unsafe
(small,
the
patient
The
insufficient:
(E7505)
Electrodes
used
electrodes
carefully.
system
and
it
conditions
yet
greater
patient's
electrode
patient
contact
is
required
simultaneously,
are
When
to
an
insulated
accidentally.
ever
is
skin
return
active
one
up
3.
Damaged
For
example,
Sparks
laid
Insulation
can
against
if
jump
the
on
the
active
from
that
patient's
the
Active
cable
metal
skin.
Electrode
to
13
is
frayed
the
or
patient
and
Cable
bare
if
the
metal
damaged
is
exposed,
cable
is

«tt
SENER
a:
A
SEES
에
SWITCHING
(E4045 & E4506)
THE
BIPOLAR
SINGLE
BIPOLAR
3-PIN
FORCEPS
PLUG INTO
OUTPUT
CONNECTOR
BY
A
CONVENTIONAL
(SW2
SERIES)
WITH
THE
INATE
PLATE
THE
IM
BIPOLAR
USED
E0024
BIPOLAR
ADAPTER
NEED
FORCEPS
IN
CONJUNCTION
TO
ELIM-
FOR A PATIENT
ONLY
PROCEDURES
NOTE:
With
instruments
instruments.
output
If
RETURN
@
monopolar
come
pe
one
Bipolar
the
which
attempts
FAULT
on
and
Valleylab
directly
This
does
patient
set
the
instruments
SSE2
is
possible
not
to
use
plate
up
with
generator
can
and
SurgiStat
into
discriminate
bipolar
safety
an
improper
be
disabled.
only
the
because
forceps
system
be
used
models
same
those
between
in
patient
15
output
the
may
with
it
generators
monopolar
interpret
plate.
the
is
permissible
receptacles
monopolar
separate
a
The
have
and
output
bipolar
RETURN
a
bipolar
bipolar
to
plug
as
simple
of a SSE3B,
instrument
FAULT
output.
bipolar
monopolar
isolated
operation.
light
the
as
may

CONNECTING
BIPOLAR
INSTRUMENTS
TO
THE
SSE3B
TN
Applications
general
biterminal
the
conventional
and
thus
Second,
extremely
fine
neural
the
Biopolar
output
forceps.
tissue
electrodes
is
electrode
attached
made
grounded
monopolar
grasped
more
current
patient's
to
to
of
spread
than
ground.
surgery
electrodes.
one
of
the
local
localized.
procedures
used
attached
to
the
wood.
place
radially
electrode.
between
about
is
returning
skin,
for
in
monopolar
the
The
is
necrotized
the
and
“patient"
When
on
the
a
or,
the
bipolar
which
The
most
nature
way
to
the
millimeter
common
In
advantage
it
require
the
the
the
majority
patient's
away
In
two
electrodes
to
at
best,
output
desiccation
bipolar
electrode.
sources
of
the
current
neurosurgery,
of
because
does
with
an
patient's
isolated
"active"
terminal
of
body,
from
ground
true
the
from
it
bipolar
through
is
include
instrument
of
desiccation
the
bipolar
the
monopolar
body
terminal
will
the
the
active
is
desiccated.
the
direct
leaving
is
performed
First,
accidents
flow
desiccation
forceps.
output.
happens
will
appear
current
coagulation
electrode,
operation,
path
some
grounded
the
neurosurgery,
with
has
two
the
patient
is
entirely
means
forceps
appear
no
is
patient's
that
is
commonly
doesn't
If
to
more
returning
of
just
only
If
between
is
be
to
active
the
forceps
basic
plate
the
that
radiate
a
grounded
grounded,
work.
the
as
the
coagulation
the
contact
body
laparoscopy,
or
advantages
is
not
eliminated.
desiccation
done
than
via
tissue
it
tissue
two
via
with
a
minimum
away
generator
only
The
electrode
if
some
will
does
which
electrodes,
point
capacitance
and
other
over
used
very
from
the
it
were
distant
tend
with
spreads
on
the
is
of
a
is
It
is
output.
bipolar
generator.
will
Because
if
cannot
designed
cannot
To
bipolar
available,
available
bipolar
bottom
used.
E0024
bipolar
behave
all
connect
adapter
not
practical
If
this
instrument
Even
like a grounded
of
the
monopolar
do
ali
to
fulgurate
compete
a
plug
it
universally
tool
are not
For
REM
procedure.
is
if
inherent
tools
the
successfully
bipolar
which
is
possible
has.no
needed
system
should
to
attempted,
as
an
the
tasks
or
instrument
has
and
hand
and
equipped
be
used
use
inadeguate
RETURN
output
safety
could
done
cut,
with
four
to
are
switching
only
to
bipolar
the
FAULT
as
of
bipolar
be
eliminated.
with
they
monopolar.
to
the
pins.
use
ordinary
commonly
capability,
the
two
generators
cover
instruments
RETURN
patient
circuit
described
monopolar.
are
clumsy
SSE3B,
If
found
larger
used
both
FAULT
plate
does
above.
procedures,
Unfortunately,
it is
the
special
(large)
on
the
jacks
in
active
with
circuit
connection
not
While
to
use
best
banana plugs
bipolar
small
to
the
and
patient
the
may
trip,
it
would
bipolar
and
to
use the
bipolar
instruments.
jacks
the
left
bipolar
SSE3B
interpret
and
the
bipolar
tools
present
plug
on
and
only
jacks
monopolar
disable
desiccation
be
desirable
can
designs
Valleylab
is
which
If
the
top
right
mode,
during
the
the
tools
be
not
are
the
and
are
the
the
16
一
一
一
一

As
described
that
it
The
bipolar
and
thus
output,
waveforms
earlier,
desiccates
output
the
power
the
BLENDS
will
desiccate,
well
also
can
are
the
but
be
bipolar
has
more
not
there
is
a
very
poor
lower
finely
is
output
for
output
linear
not
differs
cutting
power
controlled.
and
are not
need
for
from
or
fulgurating.
at
When used
BLENDS
the
monopolar
a
given
with
recommended.
with
the
control
the
BIPOLAR
output
setting
bipolar
Since
output.
in
ALL
CONTROL
All
Valleylab
markings
power
available
knob
used
approximately
There
resulting
greatly
impossible
setting.
the
Power
Power
differences
tissue
mode
These
actual
widely
maximum
maximum
a
to a 300
in
skirts
with
are
peak
is
cut
selected
numbers
surgery,
from
given
TAPERS
of O to
that
in
of
the
monopolar
two
output
influenced
to
However,
votage
defined
in
or
moment
power
power
setting
ohm
load.
AND
CALIBRATION
generators
10
mode.
that
predict
occurs
mode.
the
SSE3B
to
the
ways
which
to
energy.
by
power
at a given
as
electrode
fulgurated
is
displayed
are
a
the
to
equals
which
For
example,
and are
output
same
impedance
can
settings
look
the
exactly
available
the
amount
size,
calibration.
moment.
be
at
about
the
prior
correspond
These
at
setting
each
obtained
amount
arbitrary
the
the
Because
size
what
of
this
second.
on
the
of
Consequently,
300
to
the
to
a
setting
identical
SSE3B
on
Valleylab
SSE3B
and
effect
are
the
ohms load
of
CUT
the
configuration
at
the
useful
energy
does
front
They
tissue,
from
power
SSE3B
the
markings
0
to
electrosurgical
will
most
delivered
not
The
panel
are
the
in
were
approximate
of
4
to
those
10
dial
SSE2
and
COAG
be
efficient
guides.
relate
monopolar
with
not
an
and
the
calibration
generator
impedance.
watts
calibrated
was
about
are
on
settings
models.
control
of
achieved
each
well
output
actual
therefore
at
which
percentage
40%
still
the
Model
effect
the
electrode,
by a given
load
second.
to
the
lighted
measurement.
the
that
The
number
would
in
arbitrary
of
of
the
present
SSE3.
also
settings
the
power
is
correspond
produced
resistance
Because
quantity
for
numerals.
power, varies
given
setting.
readout
be
delivered
maximum
power
on
When
and
it
control
and
each
as
the
This
the
the
is
is
of
of
In
at
The
first
of
the
demands
three
ranges
that
for a given
rough
the
of
actual
step
categories
power
amount
in
of
and
powers
deciding
the
the O to
of
knob
what
procedure.
of
low,
in
watts
rotation.
10
power
medium,
dial
are
setting
Table
and
settings
generally
17
1
you
need
divides
high
found
greater
is
various
"power"
on
the
for
to
and
knob
CUT
have
a
procedures
gives
skirts.
modes
rough
typical
than
idea
into
Note
COAG

For
example,
235
watts,
appear
impossible
very
when
while
crude,
to
list
CUT
the
or
COAG
but
exact
COAG
the
are
power
differences
settings.
both
will
be
set
only
to
in
"8",
be
about
technique
the
CUT
90
and
power
watts.
electrodes
will
Table
be
1
make
about
will
it
Output
Voltage
distance
centimeter
can
produce
The
peak
touching
bipolar
Voltage
is
sparks
through
output
tissue)
outputs
the
jump
longer
on
force
depends
air.
sparks.
voltages,
are
plotted
page
which
Because
22.
makes
on
measured
against
sparks
electrode
COAG
open
jump
shape,
has
circuit
control
higher
settings
to
12,000
peak
(with
the
tissue.
volts
voltages
the
for
active
the
can
electrode
monopolar
Although
jump
than
CUT,
about
the
a
it
not
and
18